Smiths Medical CLINITEK 500: troubleshooting and error codes

The ServiceForge knowledge base holds 72 documented symptoms for the Smiths Medical CLINITEK 500, with 175 ranked causes, 112 diagnostic checks with test points and expected values, and 85 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Error 25 displayed

Readhead could not find calibration chip during initialization

Ranked causes

  1. 1. Misalignment or incorrect installation of fixed table or holddown (common)
  2. 2. Flex cable not fully seated in connector J7 (occasional)
  3. 3. Lamp malfunction (occasional)
  4. 4. Preamp malfunction (occasional)
  5. 5. Stepper motor drive circuit failure (occasional)
  6. 6. Mother PCB defective (rare)
  7. 7. PREAMP PCB mechanical interference (occasional)
First check

Perform 3 power on/off cycles ignoring errors before 3rd cycle

Test point
Instrument power
Expected
Error 25 no longer displayed after 3rd cycle

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 06-2 - Reagent strip lost at second readhead

Strip detected at first readhead but missing at second readhead; strip may have fallen or misaligned during transport.

Ranked causes

  1. 1. Strip fell from platform or misaligned in transport (common)
  2. 2. Bent or broken pins on Moving Table (occasional)
  3. 3. Loose or misseated Fixed Table (occasional)
  4. 4. Loose or misseated Moving Table (occasional)
  5. 5. Drive housing failure (rare)
  6. 6. Worn table guides (rare)
First check

Locate any dropped strip in waste area

Test point
Waste bin and platform
Expected
Strip visible or instrument clear

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 08-1, 08-2 - Reagent strip misalignment during processing

Strip position drifted during transport; detected correction factor of 0 or 8 indicates significant misalignment relative to calibration chip.

Ranked causes

  1. 1. Strip not positioned correctly during loading (common)
  2. 2. Bent or broken Moving Table pins (occasional)
  3. 3. Instrument not level (occasional)
  4. 4. Loose or misseated tables (occasional)
  5. 5. Worn or misaligned table guides (occasional)
  6. 6. Drive housing misalignment or failure (rare)
First check

Verify strip positioning during loading

Test point
Platform rear wall
Expected
Strip end touching rear wall, not touching bottom

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 24 displayed

Push bar not finding home in proper time during instrument startup or operation

Ranked causes

  1. 1. Misalignment or incorrect installation of push bar or fixed table (common)
  2. 2. Optical sensor malfunction (occasional)
  3. 3. Stepper motor drive circuit failure (occasional)
  4. 4. Push bar shaft dirty or dry (occasional)
  5. 5. Mother PCB defective (rare)
  6. 6. PREAMP PCB mechanical interference (occasional)
First check

Visually inspect push bar and fixed table alignment

Test point
Push bar and fixed table mechanical assembly
Expected
No visible misalignment or improper installation

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 26 displayed

Fixed table is missing or not installed properly

Ranked causes

  1. 1. Fixed table missing or not fully engaged (common)
  2. 2. Drive housing or fixed table guide wear or damage (occasional)
  3. 3. Connector not fully seated (occasional)
  4. 4. Sensor malfunction or fixed table positioning issue (occasional)
  5. 5. Mother PCB defective (rare)
  6. 6. Processor #1 on main PCB defective (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 50 displayed

Printer error message indicates external printer is not responding or interface connection is faulty

Ranked causes

  1. 1. Printer power off or not ready (common)
  2. 2. Interface cable loose or disconnected (common)
  3. 3. Printer out of paper (common)
  4. 4. Defective printer mother PCB (rare)
  5. 5. Defective printer cable (rare)
  6. 6. Defective main PCB (rare)
First check

Verify printer is powered and ready

Test point
External printer
Expected
Printer turned on and on-line with paper loaded

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 23 - Moving Table misaligned

Moving table position error detected; table may not be installed, loose, or mechanically misaligned.

Ranked causes

  1. 1. Moving table not installed or loose (common)
  2. 2. Fixed table misaligned or loose (occasional)
  3. 3. Worn or damaged Drive Housing or table guides (occasional)
  4. 4. Stepper motor drive circuit failure (rare)
  5. 5. Optical sensor failure (rare)
First check

Verify Moving Table is installed

Test point
Moving Table position
Expected
Table present and visually aligned

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 51 displayed

Control results memory is almost full with nearly 200 control result sets stored without computer transfer

Ranked causes

  1. 1. Computer not connected or powered off (common)
  2. 2. External port not enabled (common)
  3. 3. Incorrect computer interface setup parameters (occasional)
  4. 4. Defective mother PCB (rare)
  5. 5. Defective main PCB (rare)
First check

Transfer control results to computer

Test point
RS 232 serial communication port
Expected
Records transfer successfully

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 52 displayed

Sample results memory is almost full with nearly 500 patient result sets stored without computer transfer

Ranked causes

  1. 1. Computer not connected or powered off (common)
  2. 2. External port not enabled (common)
  3. 3. Incorrect computer interface setup parameters (occasional)
  4. 4. Defective mother PCB (rare)
  5. 5. Defective main PCB (rare)
First check

Transfer sample results to computer

Test point
RS 232 serial communication port
Expected
Records transfer successfully

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 53 displayed

Control results memory is completely full with 200 control result sets stored without computer transfer

Ranked causes

  1. 1. Computer not connected or powered off (common)
  2. 2. External port not enabled (common)
  3. 3. Incorrect computer interface setup parameters (occasional)
  4. 4. Defective mother PCB (rare)
  5. 5. Defective main PCB (rare)
First check

Transfer control results to computer

Test point
RS 232 serial communication port
Expected
Records transfer successfully

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 54 displayed

Sample results memory is completely full with 500 patient result sets stored without computer transfer

Ranked causes

  1. 1. Computer not connected or powered off (common)
  2. 2. External port not enabled (common)
  3. 3. Incorrect computer interface setup parameters (occasional)
  4. 4. Defective mother PCB (rare)
  5. 5. Defective main PCB (rare)
First check

Transfer sample results to computer

Test point
RS 232 serial communication port
Expected
Records transfer successfully

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Test results not being printed by internal printer

Internal printer fails to produce output; no paper or no data transmission

Ranked causes

  1. 1. Internal printer set to OFF (common)
  2. 2. No paper installed in printer (common)
  3. 3. Paper installed backwards or incorrect paper used (occasional)
  4. 4. Loose electrical connection to printer (occasional)
  5. 5. Defective printer (occasional)

11 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 01, 02, 03, 04, 05 - Instrument optical error

Optical error detected during measurement; instrument may not read strips correctly.

Ranked causes

  1. 1. Problem with pre-amp or A/D PCB (common)
  2. 2. Lamp failure (occasional)
  3. 3. Defective pre-amp module (occasional)
  4. 4. Defective A/D PCB (occasional)
First check

Power cycle the instrument

Test point
Power switch
Expected
Instrument restarts without error

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 09-1, 09-2 - Reagent strip skewed during processing

Strip rotated or tilted relative to normal orientation during transport through measurement stations.

Ranked causes

  1. 1. Jammed or stacked strips in outfeed area (common)
  2. 2. Bent or broken Moving Table pins (occasional)
  3. 3. Instrument not level (occasional)
  4. 4. Worn or misaligned table guides (occasional)
First check

Inspect waste bin for jammed strips

Test point
Outfeed area and waste bin
Expected
No strips blocking outlet

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Push bar does not move back to left after strip movement caution

Push bar fails to return to left position after advancing a strip; remains extended

Ranked causes

  1. 1. Last strip placed in loadlist run or waiting for ID entry (common)
  2. 2. Very dark urine prevents strip detection until first readhead (occasional)
  3. 3. Strip detector malfunction (occasional)
  4. 4. Defective push bar mechanism (rare)

10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bar code reader power or communication failure caution

Bar code port does not provide correct voltages or bar code reader communications test fails

Ranked causes

  1. 1. Auto-reset fuses damaged (common)
  2. 2. Bar code reader defective (occasional)
  3. 3. Processor PCB bar code control circuit defective (rare)
First check

Measure bar code port +5V

Test point
Connector J5 pin 6 to ground (pin 1 or 3)
Expected
5 VDC
Tools
DVM, bar code test fixture

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display is blank

No image or text visible on display screen; may indicate power or system initialization issue

Ranked causes

  1. 1. No power (common)
  2. 2. Improperly inserted program card (common)
  3. 3. Defective program card or defective system electronics (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 07-1 - Reagent strip not fully wetted or upside-down

Strip not properly saturated with specimen or placed upside-down on platform; incomplete wetting prevents valid test results.

Ranked causes

  1. 1. Upside-down strip placement by operator (common)
  2. 2. Strip not fully immersed in specimen (common)
  3. 3. Physical damage to platform components (occasional)
First check

Visually inspect platform and guides for physical damage

Test point
Push Bar, Fixed Table, Holddown
Expected
No visible cracks or deformation

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 10-1, 10-2 - Instrument optical error

Optical measurement system error; may indicate lamp failure, dirty calibration bars, or sensor/PCB failure.

Ranked causes

  1. 1. Dirty or contaminated Fixed Table and calibration bars (common)
  2. 2. Pre-amp subassembly failure (occasional)
  3. 3. Defective A/D PCB (occasional)
First check

Power cycle and clean calibration bars

Test point
Fixed Table calibration bars
Expected
No visible dirt or contamination

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 20-1 - Automatic strip type detection failed

Instrument unable to identify strip type when Auto Detect mode selected in Setup Routine; applies to Kanji units only.

Ranked causes

  1. 1. Unrecognized or invalid strip type (common)
  2. 2. Defective strip (occasional)
  3. 3. Pre-amp subassembly failure (rare)
First check

Verify strip type is recognized

Test point
Strip package and Setup Routine
Expected
Strip type listed in recognized types

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 21 - Internal memory error

Instrument memory error detected; likely program card or processor failure.

Ranked causes

  1. 1. Defective program card (common)
  2. 2. Processor module #1 failure (occasional)
  3. 3. Main PCB failure (occasional)
First check

Power cycle the instrument

Test point
Power switch
Expected
Error clears and normal operation resumes

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 27 displayed

Holddown is improperly installed, missing, or dirty

Ranked causes

  1. 1. Holddown missing, dirty, or incorrectly installed (common)
  2. 2. Fixed table or holddown wear and discoloration (occasional)
  3. 3. Readhead performance or alignment issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 28 displayed

Reagent strip detected at loading station but not found at first readhead

Ranked causes

  1. 1. Moving table missing or not installed (common)
  2. 2. Strip sensor malfunction or cable issue (occasional)
  3. 3. Readhead performance issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 29 displayed

Calibration bar error detected

Ranked causes

  1. 1. Calibration bars damaged, misaligned, or platform not fully installed (common)
  2. 2. Fixed table calibration chip damage (occasional)
  3. 3. Stepper motor drive or connector issue (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 40 displayed

Instrument displays Error 40 indicating communications error between processor 1 and processor 2

Ranked causes

  1. 1. Improper software version installed (occasional)
  2. 2. Defective processor 1 (occasional)
  3. 3. Defective main PCB (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 56 displayed

System error indicating general internal communications problem; error repeats after power cycle

Ranked causes

  1. 1. Defective processor 1 (occasional)
  2. 2. Defective main PCB (occasional)
  3. 3. Defective program card (occasional)
First check

Power cycle instrument

Test point
Power switch and display
Expected
System restarts normally without error

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Push bar moves right when no strip placed caution

Push bar advances without strip on platform; false trigger of movement mechanism

Ranked causes

  1. 1. Strip detector accidentally triggered by hand, sleeve or foreign object (common)
  2. 2. Ambient light changed significantly (occasional)
  3. 3. Strip detector problem (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

A-D converter reading out of range

A-D counts for one or more channels outside specified limits; lamp on or lamp off readings incorrect

Ranked causes

  1. 1. Readhead or light source defective (occasional)
  2. 2. A-D converter PCB defective or miscalibrated (occasional)
First check

Run Read A-D Counts test with lamp on

Test point
IR, RED, GREEN, BLUE channels at Readhead 1 and Readhead 2
Expected
All channels 800-2046 counts, 800-2046

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display fails test or is not powered caution

Display does not pass test or is blank; display interface PCB not receiving proper voltages

Ranked causes

  1. 1. Display assembly defective (common)
  2. 2. Main Processor PCB defective (occasional)
First check

Verify display receives power

Test point
Display interface PCB connector P2 pin 20 (NEG), pins 5, 16, 17 (POS)
Expected
+5 VDC
Tools
DVM

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 55 displayed

Instrument unable to store setup changes to program card; default values loaded instead on power-on

Ranked causes

  1. 1. Defective program card (occasional)
  2. 2. Defective main PCB (occasional)
First check

Print setup configurations to determine which to use

Test point
Display and printer output
Expected
Both configurations printed and readable

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer communication failure

Test message sent to printer; no output or corrupted output; instrument or external printer not responding

Ranked causes

  1. 1. Printer not connected or powered off (common)
  2. 2. Printer interface defective (occasional)
First check

Verify external printer connected and powered on

Test point
Auxiliary port and printer power switch
Expected
Printer on and receiving commands

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printout does not contain all reports

Missing reports not included in printed output; flagged reports not printing

Ranked causes

  1. 1. Missing reports flagged for confirmatory report with Edit flag ON (common)
  2. 2. Printer malfunction or test failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Push bar does not move to the right after strip placed caution

Push bar mechanism fails to advance after reagent strip placed on platform; no rightward motion

Ranked causes

  1. 1. Other strips being moved along platform (common)
  2. 2. Strip detector problem (occasional)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

RS232 port communication failure caution

RS232 port (RJ45 or DB-25) does not communicate; loop-back test fails

Ranked causes

  1. 1. Mother PCB RS232 circuit defective (common)
  2. 2. Processor PCB RS232 control circuit defective (occasional)
First check

RS232 port loop-back test

Test point
RS232 port (RJ45 or DB-25)
Expected
Loop-back test passes
Tools
Loop-back connector, test software

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensor interrupt circuit failure caution

Sensor does not change status when interrupted; sensor test fails after replacement

Ranked causes

  1. 1. Sensor defective (common)
  2. 2. Mother PCB or Main Processor PCB sensor circuit defective (occasional)
First check

Sensor status test

Test point
Sensor interrupt (all sensors)
Expected
Status changes when sensor interrupted

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Serial port communication failure

Test message sent to port; received message does not match or first and last characters do not correspond

Ranked causes

  1. 1. Loop-back connector not installed or defective (common)
  2. 2. Serial port interface defective (occasional)
First check

Install loop-back connector on serial port to be tested

Test point
Computer port, Bar Code Reader port, or Auxiliary port
Expected
Loop-back connector seated and secured

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Stepper motor drive output incorrect caution

Stepper motor drive circuits (IC 5, 6, or 7) do not produce proper square-wave output; waveforms missing or distorted

Ranked causes

  1. 1. Stepper motor driver chip defective (common)
  2. 2. Processor PCB not providing input signal (occasional)
First check

Measure stepper motor driver input signal

Test point
Stepper motor driver IC pin 11 (see table 7-4-2 for duration)
Expected
Square wave pulse signal
Tools
Oscilloscope

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Strip detector not detecting presence of strip in load zone

Strip Verify Test does not respond with beep when strip placed in load zone

Ranked causes

  1. 1. Detector setup values out of range (not 10 to 254) (occasional)
  2. 2. Strip detector sensor defective or misaligned (occasional)
First check

Print detector setup values

Test point
Detector output on instrument printer
Expected
Four values between 10 and 254

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Strip not centered under readheads

Strip centering test fails; error code #34 set; step count shows 00

Ranked causes

  1. 1. Strip centering values outside limits (not 4 to 40 steps) (unknown)
  2. 2. Strip centering values differ by more than 10 steps between readheads (unknown)
First check

Verify strip centering test values within acceptable range

Test point
Readhead 1 and Readhead 2 centering steps
Expected
4 to 40 steps

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Strip walk position out of specification

Strip Walk Test result is B or C for new assembly, or C for used assembly; strip position at readhead not centered at ideal value of 4

Ranked causes

  1. 1. Strip position 0 or 8 detected at either readhead (rare)
  2. 2. Strip position 2 or 6 occurred six or more times; average < 3 or > 5.5; or position 1 or 7 detected (occasional)
First check

Run Strip Walk Test with 10 MULTISTIX-10SG reagent strips

Test point
Strip position at Readhead 1 and Readhead 2
Expected
New assembly: Strip Walk Test = A; Used assembly: Strip Walk Test = A or B

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch screen does not respond correctly

Screen does not respond to touch inputs as expected during operation.

Ranked causes

  1. 1. Screen needs to be recalibrated (common)
  2. 2. Defective screen (occasional)
First check

Verify cables are properly seated in Bezel Assembly

Test point
Bezel Assembly connectors
Expected
All cables fully seated

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch screen not responding to input

Touch screen fails to respond to key presses or touch input; display may be functional

Ranked causes

  1. 1. Touch screen requires recalibration (common)
  2. 2. Defective touch screen (rare)
First check

Cycle power

Test point
Power switch and touch screen display
Expected
Touch screen responds to input after restart

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

A/D PCB reference voltage incorrect caution

A/D PCB does not provide correct reference voltages to pre-amp PCB and lamps; voltages out of specification at test points

Ranked causes

  1. 1. A/D PCB defective (common)
First check

Measure lamp unloaded voltage

Test point
Connector P4 and P5 pins 1 and 2
Expected
+5.8 to +6.24 VDC, +5.8 to +6.24 V
Tools
DVM

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 5-volt power supply output high voltage

Adjust potentiometer VR1 to set the 5V output within specification using a voltmeter at TB2 terminals 1 and 4

Ranked causes

  1. 1. 5V supply output out of range (common)
First check

Measure 5V power supply output voltage

Test point
TB2 terminal 1 (positive) and terminal 4 (DC Ground)
Expected
+5 volts DC VDC, 4.75 to 5.25 VDC
Tools
Voltmeter, Display Test Stand Fixture (p/n 71647014)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Horizontal plate adjustment

Align horizontal plate position and Moving Table fingers to ensure proper strip centering and movement

Ranked causes

  1. 1. Horizontal plate or Moving Table misaligned (common)
First check

Horizontal plate screw tension

Test point
Two screws in horizontal plate
Expected
Loose for adjustment

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Printer mechanism cleaning

Clean printer rails and platen to maintain print quality and prevent operation issues

Ranked causes

  1. 1. Dirt buildup on printer rails or platen (occasional)
First check

Print head rail cleanliness

Test point
Rails that print-head rides upon, especially ends
Expected
Clean, no visible dirt buildup
Tools
Cotton tip applicators, Isopropyl Alcohol

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pusher bar slide lubrication

Clean and lubricate pusher bar shaft and slide arm to ensure smooth motion and prevent wear

Ranked causes

  1. 1. Pusher bar shaft and slide arm require lubrication (occasional)
First check

Push Bar Shaft removal

Test point
Phillips head screw securing Push Bar Shaft to Motor Casting
Expected
Screw loose and shaft removable
Tools
#1 Phillips screwdriver

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Readhead alignment

Align readhead carrier position and belt tension using alignment fixture to ensure proper optical function and calibration

Ranked causes

  1. 1. Readhead misalignment or belt tension incorrect (common)
First check

Readhead carrier position

Test point
Readhead carrier front position
Expected
Flush against alignment fixture

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Sensor positioning

Position optical sensors for optimal performance: table sensor fully forward, crank arm sensor fully down, moving table sensor fully left with flag centered

Ranked causes

  1. 1. Optical sensors require optimal positioning (occasional)
First check

Table in place sensor position

Test point
Table in place sensor
Expected
Fully forward

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display does not update to run mode

Display screen remains in standby or menu mode; does not transition to active test result display mode

Ranked causes

  1. 1. Unknown display or mode transition fault (unknown)
First check

Perform memory reset and defaults reset

Test point
Instrument test card and setup menu
Expected
System transitions to run mode normally

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display pixel test failure

Display does not turn on all pixels or does not turn off all pixels during test sequence

Ranked causes

  1. 1. Display panel defective or not responding to test command (occasional)
First check

Run Test Display from Instrument Tests Level 2

Test point
Screen pixel state during test sequence
Expected
All pixels on for 5 seconds, then all pixels off for 5 seconds

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 30 displayed

Readhead alignment error detected

Ranked causes

  1. 1. Readhead misalignment or fixed table not fully installed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 31 displayed

Instrument displays Error 31 indicating strip detector set up error or mechanical failure during strip detection initialization

Ranked causes

  1. 1. Fixed table not installed completely to stops (common)
First check

Verify strip detector operation

Test point
Strip detector assembly
Expected
Strip detector responds and cycles pusher bar

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 34 displayed

Instrument displays Error 34 indicating strip centering error or instrument factor error during strip positioning

Ranked causes

  1. 1. Fixed table not installed completely to stops (common)
First check

Perform strip centering test

Test point
Strip centering mechanism
Expected
Strip centers correctly under readheads

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exercise all motions test error detection

Diagnostic test runs motors and actions through normal run cycles; operator selects number of cycles; error detected and displayed on screen during operation

Ranked causes

  1. 1. Motor or actuator malfunction during cycling (occasional)
First check

Select number of cycles and run exercise all motions test

Test point
Test menu screen
Expected
All cycles complete without error; cycles per hour rate displayed as approximately 500 cycles/Hr

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fixed table cannot be installed

Unable to install fixed table on instrument; moving table position issue prevents installation

Ranked causes

  1. 1. Moving table not in lowest position (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hardware homing test failure

Instrument Test Card self-check; moving table, blotter and readhead homing by Processor #1 fails; FAIL message with two-digit error code displayed

Ranked causes

  1. 1. Moving table, blotter or readhead mechanical jam or sensor failure (occasional)
First check

Observe hardware homing test execution

Test point
Self-check screen
Expected
PASS displayed; table and readhead reach home position without error

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leukocyte Correction Factor generation range exceeds limit patient safety

LEU Correction Factor test range value exceeds acceptable threshold; instrument normalization fails

Ranked causes

  1. 1. Reagent strip alignment or readhead position variation (common)
First check

Run 10 Multistix 10 SG strips dipped in CT500 Calibration Solution

Test point
Leukocyte test pads
Expected
Range value <= 34

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No motor movement on power-up caution

Readhead, moving table, or blotter motor does not move when unit is powered on with test card installed

Ranked causes

  1. 1. SIMM module (Processor 1) defective (common)
First check

Verify motor movement on power-up

Test point
Readhead, moving table, blotter motors
Expected
Motor movement occurs

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply or system initialization issue

Instrument fails to respond to power commands or system initialization

Ranked causes

  1. 1. Power supply malfunction or connection failure (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Power supply output voltage incorrect
  • Pre-amp PCB output voltage incorrect
  • Printer power missing or incorrect
  • Processor software version mismatch on startup
  • Pusher bar not at left position
  • ROM checksum test failure
  • Strip detector does not detect strip or cycle pusher bar
  • Strip detector power missing or incorrect
  • Test Readheads calibration readings out of acceptable range
  • Test Readheads readhead #2 calibration edge correction out of range
  • Touch screen position sensing error
  • XRAM memory data corruption detected

12 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
21ROM Checksum ErrorCalculated checksum for the ROM does not equal 0error
23Table Movement ErrorTable sensor fails to respond correctly at 2.5 or 3.0 seconds into cycleerror
24Push Bar Movement ErrorPush bar sensor fails to respond at 2.5 or 3.8 seconds in cycleerror
25Readhead Scan Movement ErrorReadhead not positioned on calibration chip during RUN initializationerror
26Table Not In PlaceTable sensor is not activatederror
27Holddowns Not In PlaceReflectance between pads 10 and 11 not below 40%error
28False Strip DetectStrip detected but not sensed at 1st readhead or areas between pads 2-3 and 9-10 do not both read >65%error
29Cal Chip NOT found errorCalibration chip not found under readhead 1 within 750 steps during initializationerror
30Read-head Alignment errorReadhead 2 calibration chip edge varies more than ±3 motor steps from readhead 1error
31Instrument mechanical error - strip detector set up error during initializationError detected by processor 2 during strip detector initialization or fixed table positioningerror
34Instrument mechanical error - strip centering error or instrument factor errorError detected during strip positioning and centering processerror
#34Strip centering test failedStrip centering values outside range 4-40 steps or differ by more than 10 steps between readheadserror
40Internal Communications ErrorCommunication failure between the two processorscritical
40Improper software installed or processor communication failureCommunications error between processor 1 and processor 2 during operationerror
50Printer error - external printer not responding or interface disconnectedInstrument unable to communicate with or receive status from external printererror
50Printer ErrorExternal printer has an error or is out of paperwarning
51Control results memory almost full - approximately 200 result sets stored without transferInstrument memory reaches threshold with stored control results not transferred to external computerwarning
51Control Results Almost FullControls database has space for only 10 more resultswarning
52Sample results memory almost full - approximately 500 patient result sets stored without transferInstrument memory reaches threshold with stored patient results not transferred to external computerwarning
52Sample Results Almost FullSample database has space for only 10 more resultswarning
53Control results memory completely full - 200 result sets stored without transferInstrument control results memory reaches maximum capacity with data not transferred to computererror
53Control Results FullControls database is fullerror
54Sample results memory completely full - 500 patient result sets stored without transferInstrument patient results memory reaches maximum capacity with data not transferred to computererror
54Sample Results FullSample database is fullerror
55Setup Information ProblemSetup information files corrupted during data integrity check at power ONerror
55System unable to store setup changes to program card; default values loaded insteadInstrument fails to write setup configuration to program card during setup modification or power-offerror
56General internal communications error - generic watchdog timeoutSystem detects communications failure between internal processors or moduleserror
56System ErrorSerious system error or all communication problems between Processor 1 and Processor 2critical
61Low Dark Value Readhead 1Dark value < 1 count for any channelerror
62High Dark Value Readhead 1Dark value >= 400 counts on any channelerror
63A/D Converter Overrange Readhead 1Read value at full scale (>2040 counts) on any channelerror
64Low Lamp Level Readhead 1All channels read < 200 counts above dark value at calibrationerror
65Low Channel Output Readhead 1One or more channels read < 200 counts above dark at calibrationerror
68Misaligned Strip Readhead 1Strip end position reflectance crosses 30% threshold at position 0 or 8error
69Skewed Strip Readhead 1Reflectance >65% at one area but <65% in other between pads 2-3 and 9-10error
70Reflectance >100% Readhead 1Reflectance for any pad measures more than 100%error
81Low Dark Value Readhead 2Dark value < 1 count for any channelerror
82High Dark Value Readhead 2Dark value >= 400 counts on any channelerror
83A/D Converter Overrange Readhead 2Read value at full scale (>2040 counts) on any channelerror
84Low Lamp Level Readhead 2All channels read < 200 counts above dark value at calibrationerror
85Low Channel Output Readhead 2One or more channels read < 200 counts above dark at calibrationerror
86Missing Strip Readhead 2Both areas between pads 2-3 and 9-10 do not read >50%error
88Misaligned Strip Readhead 2Strip end position reflectance crosses 30% threshold at position 0 or 8error
89Skewed Strip Readhead 2Reflectance >65% at one area but <65% in other between pads 2-3 and 9-10error
90Reflectance >100% Readhead 2Reflectance for any pad measures more than 100%error
93Result Data Integrity Readhead 2Checksum sent by Processor 2 does not match checksum calculated by Processor 1critical
DT001Display blank - no video outputPower applied but display shows no outputcritical
DT002Display malfunction - garbled or incorrect outputDisplay shows corrupted, partial, or unreadable characterserror
DT003Display character string or touch screen operation corrected by power cycleAfter power cycle and touch screen recalibration display/touch responds correctlyinfo
DT004System memory and defaults reset corrected problemAfter memory reset and defaults reset system responds normallyinfo
DT005Display and processor issue resolved after troubleshootingAfter troubleshooting display, main PCB and program card system returns to normalinfo
DT006Display or processor issue persists after troubleshootingDisplay does not update to run mode despite troubleshooting stepserror
DT007Touch screen response corrected by power cycle and recalibrationAfter power cycle and touch screen recalibration touch screen responds to inputinfo
DT008Touch screen not responding after power cycleTouch screen still unresponsive after power cycle restarterror
DT009Touch screen malfunction - adjacent areas around keys not respondingTouch screen shows localized dead spots or areas not respondingerror
DT010Pusher bar not at left starting positionPusher bar position indicator shows bar is not fully retracted to left stoperror
DT011Strip detector does not detect strip or pusher bar does not cycleStrip detector fails to sense test strip presence or initiate pusher bar movementerror
DT012System locked up - unresponsive to inputSystem becomes unresponsive to user input; processing haltedcritical
DT013Display does not update to run mode after data transfer attemptDisplay remains in standby or menu mode; does not show active test resultserror
DT014Touch screen hardware defective and requires replacementTouch screen does not respond to input despite power cycle and recalibration attemptserror
DT015Instrument setup parameters incorrect or reset to defaults neededPrint out shows incorrect setup configuration or customer settings conflict with defaultswarning
Error 01Instrument optical errorOptical measurement system failure during strip readingerror
Error 02Instrument optical errorOptical measurement system failure during strip readingerror
Error 03Instrument optical errorOptical measurement system failure during strip readingerror
Error 04Instrument optical errorOptical measurement system failure during strip readingerror
Error 05Instrument optical errorOptical measurement system failure during strip readingerror
Error 06-2Reagent strip lost between readheadsStrip detected at readhead 1 but not at readhead 2error
Error 07-1Reagent strip not fully wetted or upside-downInsufficient wetting or inverted strip orientation detected during processingwarning
Error 08-1Reagent strip misaligned during processing (readhead 1)Strip position correction factor of 0 or 8 detected at readhead 1error
Error 08-2Reagent strip misaligned during processing (readhead 2)Strip position correction factor of 0 or 8 detected at readhead 2error
Error 09-1Reagent strip skewed during processing (readhead 1)Strip rotation or tilt detected at readhead 1 during transporterror
Error 09-2Reagent strip skewed during processing (readhead 2)Strip rotation or tilt detected at readhead 2 during transporterror
Error 10-1Instrument optical errorOptical measurement system failure during strip readingerror
Error 10-2Instrument optical errorOptical measurement system failure during strip readingerror
Error 20-1Automatic strip type detection failed (Kanji units only)Strip type could not be auto-detected when Auto Detect mode is enabledwarning
Error 21Internal memory errorMemory integrity check failed or program card communication lostcritical
Error 23Moving Table misalignedMoving table position sensor detects misalignmenterror
Error 24Push bar mechanical error - not finding home in proper timeDuring instrument startup or operation when push bar fails to reach home positionerror
Error 25Readhead could not find calibration chip during initializationDuring startup calibration sequence when readhead fails to locate cal chiperror
Error 26Fixed table is missing or not installed properlyDuring initialization when fixed table is not detected or not fully engagederror
Error 27Holddown is improperly installed, missing, or dirtyDuring specimen processing when holddown is not properly positionedwarning
Error 28Reagent strip detected at loading station but not found at first readheadDuring specimen processing when strip is placed but not transported to first readheaderror
Error 29Calibration bar error detectedDuring startup calibration when calibration bars are not detected or damagederror
Error 30Readhead alignment error detectedDuring startup when readhead position is incorrect or misalignederror
General Error 30Readhead #2 calibration edge correction out of rangeRH2 Calibration Edge Correction value outside +/- 3 rangeerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI