Smiths Medical CLINITEK 500: troubleshooting and error codes
The ServiceForge knowledge base holds 72 documented symptoms for the Smiths Medical CLINITEK 500, with 175 ranked causes, 112 diagnostic checks with test points and expected values, and 85 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Error 25 displayed
Readhead could not find calibration chip during initialization
Ranked causes
- 1. Misalignment or incorrect installation of fixed table or holddown (common)
- 2. Flex cable not fully seated in connector J7 (occasional)
- 3. Lamp malfunction (occasional)
- 4. Preamp malfunction (occasional)
- 5. Stepper motor drive circuit failure (occasional)
- 6. Mother PCB defective (rare)
- 7. PREAMP PCB mechanical interference (occasional)
Perform 3 power on/off cycles ignoring errors before 3rd cycle
- Test point
- Instrument power
- Expected
- Error 25 no longer displayed after 3rd cycle
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 06-2 - Reagent strip lost at second readhead
Strip detected at first readhead but missing at second readhead; strip may have fallen or misaligned during transport.
Ranked causes
- 1. Strip fell from platform or misaligned in transport (common)
- 2. Bent or broken pins on Moving Table (occasional)
- 3. Loose or misseated Fixed Table (occasional)
- 4. Loose or misseated Moving Table (occasional)
- 5. Drive housing failure (rare)
- 6. Worn table guides (rare)
Locate any dropped strip in waste area
- Test point
- Waste bin and platform
- Expected
- Strip visible or instrument clear
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 08-1, 08-2 - Reagent strip misalignment during processing
Strip position drifted during transport; detected correction factor of 0 or 8 indicates significant misalignment relative to calibration chip.
Ranked causes
- 1. Strip not positioned correctly during loading (common)
- 2. Bent or broken Moving Table pins (occasional)
- 3. Instrument not level (occasional)
- 4. Loose or misseated tables (occasional)
- 5. Worn or misaligned table guides (occasional)
- 6. Drive housing misalignment or failure (rare)
Verify strip positioning during loading
- Test point
- Platform rear wall
- Expected
- Strip end touching rear wall, not touching bottom
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 24 displayed
Push bar not finding home in proper time during instrument startup or operation
Ranked causes
- 1. Misalignment or incorrect installation of push bar or fixed table (common)
- 2. Optical sensor malfunction (occasional)
- 3. Stepper motor drive circuit failure (occasional)
- 4. Push bar shaft dirty or dry (occasional)
- 5. Mother PCB defective (rare)
- 6. PREAMP PCB mechanical interference (occasional)
Visually inspect push bar and fixed table alignment
- Test point
- Push bar and fixed table mechanical assembly
- Expected
- No visible misalignment or improper installation
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 26 displayed
Fixed table is missing or not installed properly
Ranked causes
- 1. Fixed table missing or not fully engaged (common)
- 2. Drive housing or fixed table guide wear or damage (occasional)
- 3. Connector not fully seated (occasional)
- 4. Sensor malfunction or fixed table positioning issue (occasional)
- 5. Mother PCB defective (rare)
- 6. Processor #1 on main PCB defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 50 displayed
Printer error message indicates external printer is not responding or interface connection is faulty
Ranked causes
- 1. Printer power off or not ready (common)
- 2. Interface cable loose or disconnected (common)
- 3. Printer out of paper (common)
- 4. Defective printer mother PCB (rare)
- 5. Defective printer cable (rare)
- 6. Defective main PCB (rare)
Verify printer is powered and ready
- Test point
- External printer
- Expected
- Printer turned on and on-line with paper loaded
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 23 - Moving Table misaligned
Moving table position error detected; table may not be installed, loose, or mechanically misaligned.
Ranked causes
- 1. Moving table not installed or loose (common)
- 2. Fixed table misaligned or loose (occasional)
- 3. Worn or damaged Drive Housing or table guides (occasional)
- 4. Stepper motor drive circuit failure (rare)
- 5. Optical sensor failure (rare)
Verify Moving Table is installed
- Test point
- Moving Table position
- Expected
- Table present and visually aligned
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 51 displayed
Control results memory is almost full with nearly 200 control result sets stored without computer transfer
Ranked causes
- 1. Computer not connected or powered off (common)
- 2. External port not enabled (common)
- 3. Incorrect computer interface setup parameters (occasional)
- 4. Defective mother PCB (rare)
- 5. Defective main PCB (rare)
Transfer control results to computer
- Test point
- RS 232 serial communication port
- Expected
- Records transfer successfully
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 52 displayed
Sample results memory is almost full with nearly 500 patient result sets stored without computer transfer
Ranked causes
- 1. Computer not connected or powered off (common)
- 2. External port not enabled (common)
- 3. Incorrect computer interface setup parameters (occasional)
- 4. Defective mother PCB (rare)
- 5. Defective main PCB (rare)
Transfer sample results to computer
- Test point
- RS 232 serial communication port
- Expected
- Records transfer successfully
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 53 displayed
Control results memory is completely full with 200 control result sets stored without computer transfer
Ranked causes
- 1. Computer not connected or powered off (common)
- 2. External port not enabled (common)
- 3. Incorrect computer interface setup parameters (occasional)
- 4. Defective mother PCB (rare)
- 5. Defective main PCB (rare)
Transfer control results to computer
- Test point
- RS 232 serial communication port
- Expected
- Records transfer successfully
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 54 displayed
Sample results memory is completely full with 500 patient result sets stored without computer transfer
Ranked causes
- 1. Computer not connected or powered off (common)
- 2. External port not enabled (common)
- 3. Incorrect computer interface setup parameters (occasional)
- 4. Defective mother PCB (rare)
- 5. Defective main PCB (rare)
Transfer sample results to computer
- Test point
- RS 232 serial communication port
- Expected
- Records transfer successfully
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test results not being printed by internal printer
Internal printer fails to produce output; no paper or no data transmission
Ranked causes
- 1. Internal printer set to OFF (common)
- 2. No paper installed in printer (common)
- 3. Paper installed backwards or incorrect paper used (occasional)
- 4. Loose electrical connection to printer (occasional)
- 5. Defective printer (occasional)
11 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 01, 02, 03, 04, 05 - Instrument optical error
Optical error detected during measurement; instrument may not read strips correctly.
Ranked causes
- 1. Problem with pre-amp or A/D PCB (common)
- 2. Lamp failure (occasional)
- 3. Defective pre-amp module (occasional)
- 4. Defective A/D PCB (occasional)
Power cycle the instrument
- Test point
- Power switch
- Expected
- Instrument restarts without error
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 09-1, 09-2 - Reagent strip skewed during processing
Strip rotated or tilted relative to normal orientation during transport through measurement stations.
Ranked causes
- 1. Jammed or stacked strips in outfeed area (common)
- 2. Bent or broken Moving Table pins (occasional)
- 3. Instrument not level (occasional)
- 4. Worn or misaligned table guides (occasional)
Inspect waste bin for jammed strips
- Test point
- Outfeed area and waste bin
- Expected
- No strips blocking outlet
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Push bar does not move back to left after strip movement caution
Push bar fails to return to left position after advancing a strip; remains extended
Ranked causes
- 1. Last strip placed in loadlist run or waiting for ID entry (common)
- 2. Very dark urine prevents strip detection until first readhead (occasional)
- 3. Strip detector malfunction (occasional)
- 4. Defective push bar mechanism (rare)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bar code reader power or communication failure caution
Bar code port does not provide correct voltages or bar code reader communications test fails
Ranked causes
- 1. Auto-reset fuses damaged (common)
- 2. Bar code reader defective (occasional)
- 3. Processor PCB bar code control circuit defective (rare)
Measure bar code port +5V
- Test point
- Connector J5 pin 6 to ground (pin 1 or 3)
- Expected
- 5 VDC
- Tools
- DVM, bar code test fixture
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is blank
No image or text visible on display screen; may indicate power or system initialization issue
Ranked causes
- 1. No power (common)
- 2. Improperly inserted program card (common)
- 3. Defective program card or defective system electronics (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 07-1 - Reagent strip not fully wetted or upside-down
Strip not properly saturated with specimen or placed upside-down on platform; incomplete wetting prevents valid test results.
Ranked causes
- 1. Upside-down strip placement by operator (common)
- 2. Strip not fully immersed in specimen (common)
- 3. Physical damage to platform components (occasional)
Visually inspect platform and guides for physical damage
- Test point
- Push Bar, Fixed Table, Holddown
- Expected
- No visible cracks or deformation
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 10-1, 10-2 - Instrument optical error
Optical measurement system error; may indicate lamp failure, dirty calibration bars, or sensor/PCB failure.
Ranked causes
- 1. Dirty or contaminated Fixed Table and calibration bars (common)
- 2. Pre-amp subassembly failure (occasional)
- 3. Defective A/D PCB (occasional)
Power cycle and clean calibration bars
- Test point
- Fixed Table calibration bars
- Expected
- No visible dirt or contamination
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 20-1 - Automatic strip type detection failed
Instrument unable to identify strip type when Auto Detect mode selected in Setup Routine; applies to Kanji units only.
Ranked causes
- 1. Unrecognized or invalid strip type (common)
- 2. Defective strip (occasional)
- 3. Pre-amp subassembly failure (rare)
Verify strip type is recognized
- Test point
- Strip package and Setup Routine
- Expected
- Strip type listed in recognized types
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 21 - Internal memory error
Instrument memory error detected; likely program card or processor failure.
Ranked causes
- 1. Defective program card (common)
- 2. Processor module #1 failure (occasional)
- 3. Main PCB failure (occasional)
Power cycle the instrument
- Test point
- Power switch
- Expected
- Error clears and normal operation resumes
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 27 displayed
Holddown is improperly installed, missing, or dirty
Ranked causes
- 1. Holddown missing, dirty, or incorrectly installed (common)
- 2. Fixed table or holddown wear and discoloration (occasional)
- 3. Readhead performance or alignment issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 28 displayed
Reagent strip detected at loading station but not found at first readhead
Ranked causes
- 1. Moving table missing or not installed (common)
- 2. Strip sensor malfunction or cable issue (occasional)
- 3. Readhead performance issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 29 displayed
Calibration bar error detected
Ranked causes
- 1. Calibration bars damaged, misaligned, or platform not fully installed (common)
- 2. Fixed table calibration chip damage (occasional)
- 3. Stepper motor drive or connector issue (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 40 displayed
Instrument displays Error 40 indicating communications error between processor 1 and processor 2
Ranked causes
- 1. Improper software version installed (occasional)
- 2. Defective processor 1 (occasional)
- 3. Defective main PCB (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 56 displayed
System error indicating general internal communications problem; error repeats after power cycle
Ranked causes
- 1. Defective processor 1 (occasional)
- 2. Defective main PCB (occasional)
- 3. Defective program card (occasional)
Power cycle instrument
- Test point
- Power switch and display
- Expected
- System restarts normally without error
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Push bar moves right when no strip placed caution
Push bar advances without strip on platform; false trigger of movement mechanism
Ranked causes
- 1. Strip detector accidentally triggered by hand, sleeve or foreign object (common)
- 2. Ambient light changed significantly (occasional)
- 3. Strip detector problem (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
A-D converter reading out of range
A-D counts for one or more channels outside specified limits; lamp on or lamp off readings incorrect
Ranked causes
- 1. Readhead or light source defective (occasional)
- 2. A-D converter PCB defective or miscalibrated (occasional)
Run Read A-D Counts test with lamp on
- Test point
- IR, RED, GREEN, BLUE channels at Readhead 1 and Readhead 2
- Expected
- All channels 800-2046 counts, 800-2046
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display fails test or is not powered caution
Display does not pass test or is blank; display interface PCB not receiving proper voltages
Ranked causes
- 1. Display assembly defective (common)
- 2. Main Processor PCB defective (occasional)
Verify display receives power
- Test point
- Display interface PCB connector P2 pin 20 (NEG), pins 5, 16, 17 (POS)
- Expected
- +5 VDC
- Tools
- DVM
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 55 displayed
Instrument unable to store setup changes to program card; default values loaded instead on power-on
Ranked causes
- 1. Defective program card (occasional)
- 2. Defective main PCB (occasional)
Print setup configurations to determine which to use
- Test point
- Display and printer output
- Expected
- Both configurations printed and readable
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer communication failure
Test message sent to printer; no output or corrupted output; instrument or external printer not responding
Ranked causes
- 1. Printer not connected or powered off (common)
- 2. Printer interface defective (occasional)
Verify external printer connected and powered on
- Test point
- Auxiliary port and printer power switch
- Expected
- Printer on and receiving commands
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printout does not contain all reports
Missing reports not included in printed output; flagged reports not printing
Ranked causes
- 1. Missing reports flagged for confirmatory report with Edit flag ON (common)
- 2. Printer malfunction or test failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Push bar does not move to the right after strip placed caution
Push bar mechanism fails to advance after reagent strip placed on platform; no rightward motion
Ranked causes
- 1. Other strips being moved along platform (common)
- 2. Strip detector problem (occasional)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RS232 port communication failure caution
RS232 port (RJ45 or DB-25) does not communicate; loop-back test fails
Ranked causes
- 1. Mother PCB RS232 circuit defective (common)
- 2. Processor PCB RS232 control circuit defective (occasional)
RS232 port loop-back test
- Test point
- RS232 port (RJ45 or DB-25)
- Expected
- Loop-back test passes
- Tools
- Loop-back connector, test software
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensor interrupt circuit failure caution
Sensor does not change status when interrupted; sensor test fails after replacement
Ranked causes
- 1. Sensor defective (common)
- 2. Mother PCB or Main Processor PCB sensor circuit defective (occasional)
Sensor status test
- Test point
- Sensor interrupt (all sensors)
- Expected
- Status changes when sensor interrupted
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serial port communication failure
Test message sent to port; received message does not match or first and last characters do not correspond
Ranked causes
- 1. Loop-back connector not installed or defective (common)
- 2. Serial port interface defective (occasional)
Install loop-back connector on serial port to be tested
- Test point
- Computer port, Bar Code Reader port, or Auxiliary port
- Expected
- Loop-back connector seated and secured
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Stepper motor drive output incorrect caution
Stepper motor drive circuits (IC 5, 6, or 7) do not produce proper square-wave output; waveforms missing or distorted
Ranked causes
- 1. Stepper motor driver chip defective (common)
- 2. Processor PCB not providing input signal (occasional)
Measure stepper motor driver input signal
- Test point
- Stepper motor driver IC pin 11 (see table 7-4-2 for duration)
- Expected
- Square wave pulse signal
- Tools
- Oscilloscope
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Strip detector not detecting presence of strip in load zone
Strip Verify Test does not respond with beep when strip placed in load zone
Ranked causes
- 1. Detector setup values out of range (not 10 to 254) (occasional)
- 2. Strip detector sensor defective or misaligned (occasional)
Print detector setup values
- Test point
- Detector output on instrument printer
- Expected
- Four values between 10 and 254
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Strip not centered under readheads
Strip centering test fails; error code #34 set; step count shows 00
Ranked causes
- 1. Strip centering values outside limits (not 4 to 40 steps) (unknown)
- 2. Strip centering values differ by more than 10 steps between readheads (unknown)
Verify strip centering test values within acceptable range
- Test point
- Readhead 1 and Readhead 2 centering steps
- Expected
- 4 to 40 steps
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Strip walk position out of specification
Strip Walk Test result is B or C for new assembly, or C for used assembly; strip position at readhead not centered at ideal value of 4
Ranked causes
- 1. Strip position 0 or 8 detected at either readhead (rare)
- 2. Strip position 2 or 6 occurred six or more times; average < 3 or > 5.5; or position 1 or 7 detected (occasional)
Run Strip Walk Test with 10 MULTISTIX-10SG reagent strips
- Test point
- Strip position at Readhead 1 and Readhead 2
- Expected
- New assembly: Strip Walk Test = A; Used assembly: Strip Walk Test = A or B
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen does not respond correctly
Screen does not respond to touch inputs as expected during operation.
Ranked causes
- 1. Screen needs to be recalibrated (common)
- 2. Defective screen (occasional)
Verify cables are properly seated in Bezel Assembly
- Test point
- Bezel Assembly connectors
- Expected
- All cables fully seated
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen not responding to input
Touch screen fails to respond to key presses or touch input; display may be functional
Ranked causes
- 1. Touch screen requires recalibration (common)
- 2. Defective touch screen (rare)
Cycle power
- Test point
- Power switch and touch screen display
- Expected
- Touch screen responds to input after restart
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
A/D PCB reference voltage incorrect caution
A/D PCB does not provide correct reference voltages to pre-amp PCB and lamps; voltages out of specification at test points
Ranked causes
- 1. A/D PCB defective (common)
Measure lamp unloaded voltage
- Test point
- Connector P4 and P5 pins 1 and 2
- Expected
- +5.8 to +6.24 VDC, +5.8 to +6.24 V
- Tools
- DVM
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: 5-volt power supply output high voltage
Adjust potentiometer VR1 to set the 5V output within specification using a voltmeter at TB2 terminals 1 and 4
Ranked causes
- 1. 5V supply output out of range (common)
Measure 5V power supply output voltage
- Test point
- TB2 terminal 1 (positive) and terminal 4 (DC Ground)
- Expected
- +5 volts DC VDC, 4.75 to 5.25 VDC
- Tools
- Voltmeter, Display Test Stand Fixture (p/n 71647014)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Horizontal plate adjustment
Align horizontal plate position and Moving Table fingers to ensure proper strip centering and movement
Ranked causes
- 1. Horizontal plate or Moving Table misaligned (common)
Horizontal plate screw tension
- Test point
- Two screws in horizontal plate
- Expected
- Loose for adjustment
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Printer mechanism cleaning
Clean printer rails and platen to maintain print quality and prevent operation issues
Ranked causes
- 1. Dirt buildup on printer rails or platen (occasional)
Print head rail cleanliness
- Test point
- Rails that print-head rides upon, especially ends
- Expected
- Clean, no visible dirt buildup
- Tools
- Cotton tip applicators, Isopropyl Alcohol
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pusher bar slide lubrication
Clean and lubricate pusher bar shaft and slide arm to ensure smooth motion and prevent wear
Ranked causes
- 1. Pusher bar shaft and slide arm require lubrication (occasional)
Push Bar Shaft removal
- Test point
- Phillips head screw securing Push Bar Shaft to Motor Casting
- Expected
- Screw loose and shaft removable
- Tools
- #1 Phillips screwdriver
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Readhead alignment
Align readhead carrier position and belt tension using alignment fixture to ensure proper optical function and calibration
Ranked causes
- 1. Readhead misalignment or belt tension incorrect (common)
Readhead carrier position
- Test point
- Readhead carrier front position
- Expected
- Flush against alignment fixture
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sensor positioning
Position optical sensors for optimal performance: table sensor fully forward, crank arm sensor fully down, moving table sensor fully left with flag centered
Ranked causes
- 1. Optical sensors require optimal positioning (occasional)
Table in place sensor position
- Test point
- Table in place sensor
- Expected
- Fully forward
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display does not update to run mode
Display screen remains in standby or menu mode; does not transition to active test result display mode
Ranked causes
- 1. Unknown display or mode transition fault (unknown)
Perform memory reset and defaults reset
- Test point
- Instrument test card and setup menu
- Expected
- System transitions to run mode normally
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display pixel test failure
Display does not turn on all pixels or does not turn off all pixels during test sequence
Ranked causes
- 1. Display panel defective or not responding to test command (occasional)
Run Test Display from Instrument Tests Level 2
- Test point
- Screen pixel state during test sequence
- Expected
- All pixels on for 5 seconds, then all pixels off for 5 seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 30 displayed
Readhead alignment error detected
Ranked causes
- 1. Readhead misalignment or fixed table not fully installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 31 displayed
Instrument displays Error 31 indicating strip detector set up error or mechanical failure during strip detection initialization
Ranked causes
- 1. Fixed table not installed completely to stops (common)
Verify strip detector operation
- Test point
- Strip detector assembly
- Expected
- Strip detector responds and cycles pusher bar
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 34 displayed
Instrument displays Error 34 indicating strip centering error or instrument factor error during strip positioning
Ranked causes
- 1. Fixed table not installed completely to stops (common)
Perform strip centering test
- Test point
- Strip centering mechanism
- Expected
- Strip centers correctly under readheads
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exercise all motions test error detection
Diagnostic test runs motors and actions through normal run cycles; operator selects number of cycles; error detected and displayed on screen during operation
Ranked causes
- 1. Motor or actuator malfunction during cycling (occasional)
Select number of cycles and run exercise all motions test
- Test point
- Test menu screen
- Expected
- All cycles complete without error; cycles per hour rate displayed as approximately 500 cycles/Hr
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fixed table cannot be installed
Unable to install fixed table on instrument; moving table position issue prevents installation
Ranked causes
- 1. Moving table not in lowest position (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hardware homing test failure
Instrument Test Card self-check; moving table, blotter and readhead homing by Processor #1 fails; FAIL message with two-digit error code displayed
Ranked causes
- 1. Moving table, blotter or readhead mechanical jam or sensor failure (occasional)
Observe hardware homing test execution
- Test point
- Self-check screen
- Expected
- PASS displayed; table and readhead reach home position without error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leukocyte Correction Factor generation range exceeds limit patient safety
LEU Correction Factor test range value exceeds acceptable threshold; instrument normalization fails
Ranked causes
- 1. Reagent strip alignment or readhead position variation (common)
Run 10 Multistix 10 SG strips dipped in CT500 Calibration Solution
- Test point
- Leukocyte test pads
- Expected
- Range value <= 34
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No motor movement on power-up caution
Readhead, moving table, or blotter motor does not move when unit is powered on with test card installed
Ranked causes
- 1. SIMM module (Processor 1) defective (common)
Verify motor movement on power-up
- Test point
- Readhead, moving table, blotter motors
- Expected
- Motor movement occurs
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply or system initialization issue
Instrument fails to respond to power commands or system initialization
Ranked causes
- 1. Power supply malfunction or connection failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Power supply output voltage incorrect
- Pre-amp PCB output voltage incorrect
- Printer power missing or incorrect
- Processor software version mismatch on startup
- Pusher bar not at left position
- ROM checksum test failure
- Strip detector does not detect strip or cycle pusher bar
- Strip detector power missing or incorrect
- Test Readheads calibration readings out of acceptable range
- Test Readheads readhead #2 calibration edge correction out of range
- Touch screen position sensing error
- XRAM memory data corruption detected
12 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
21 | ROM Checksum Error | Calculated checksum for the ROM does not equal 0 | error |
23 | Table Movement Error | Table sensor fails to respond correctly at 2.5 or 3.0 seconds into cycle | error |
24 | Push Bar Movement Error | Push bar sensor fails to respond at 2.5 or 3.8 seconds in cycle | error |
25 | Readhead Scan Movement Error | Readhead not positioned on calibration chip during RUN initialization | error |
26 | Table Not In Place | Table sensor is not activated | error |
27 | Holddowns Not In Place | Reflectance between pads 10 and 11 not below 40% | error |
28 | False Strip Detect | Strip detected but not sensed at 1st readhead or areas between pads 2-3 and 9-10 do not both read >65% | error |
29 | Cal Chip NOT found error | Calibration chip not found under readhead 1 within 750 steps during initialization | error |
30 | Read-head Alignment error | Readhead 2 calibration chip edge varies more than ±3 motor steps from readhead 1 | error |
31 | Instrument mechanical error - strip detector set up error during initialization | Error detected by processor 2 during strip detector initialization or fixed table positioning | error |
34 | Instrument mechanical error - strip centering error or instrument factor error | Error detected during strip positioning and centering process | error |
#34 | Strip centering test failed | Strip centering values outside range 4-40 steps or differ by more than 10 steps between readheads | error |
40 | Internal Communications Error | Communication failure between the two processors | critical |
40 | Improper software installed or processor communication failure | Communications error between processor 1 and processor 2 during operation | error |
50 | Printer error - external printer not responding or interface disconnected | Instrument unable to communicate with or receive status from external printer | error |
50 | Printer Error | External printer has an error or is out of paper | warning |
51 | Control results memory almost full - approximately 200 result sets stored without transfer | Instrument memory reaches threshold with stored control results not transferred to external computer | warning |
51 | Control Results Almost Full | Controls database has space for only 10 more results | warning |
52 | Sample results memory almost full - approximately 500 patient result sets stored without transfer | Instrument memory reaches threshold with stored patient results not transferred to external computer | warning |
52 | Sample Results Almost Full | Sample database has space for only 10 more results | warning |
53 | Control results memory completely full - 200 result sets stored without transfer | Instrument control results memory reaches maximum capacity with data not transferred to computer | error |
53 | Control Results Full | Controls database is full | error |
54 | Sample results memory completely full - 500 patient result sets stored without transfer | Instrument patient results memory reaches maximum capacity with data not transferred to computer | error |
54 | Sample Results Full | Sample database is full | error |
55 | Setup Information Problem | Setup information files corrupted during data integrity check at power ON | error |
55 | System unable to store setup changes to program card; default values loaded instead | Instrument fails to write setup configuration to program card during setup modification or power-off | error |
56 | General internal communications error - generic watchdog timeout | System detects communications failure between internal processors or modules | error |
56 | System Error | Serious system error or all communication problems between Processor 1 and Processor 2 | critical |
61 | Low Dark Value Readhead 1 | Dark value < 1 count for any channel | error |
62 | High Dark Value Readhead 1 | Dark value >= 400 counts on any channel | error |
63 | A/D Converter Overrange Readhead 1 | Read value at full scale (>2040 counts) on any channel | error |
64 | Low Lamp Level Readhead 1 | All channels read < 200 counts above dark value at calibration | error |
65 | Low Channel Output Readhead 1 | One or more channels read < 200 counts above dark at calibration | error |
68 | Misaligned Strip Readhead 1 | Strip end position reflectance crosses 30% threshold at position 0 or 8 | error |
69 | Skewed Strip Readhead 1 | Reflectance >65% at one area but <65% in other between pads 2-3 and 9-10 | error |
70 | Reflectance >100% Readhead 1 | Reflectance for any pad measures more than 100% | error |
81 | Low Dark Value Readhead 2 | Dark value < 1 count for any channel | error |
82 | High Dark Value Readhead 2 | Dark value >= 400 counts on any channel | error |
83 | A/D Converter Overrange Readhead 2 | Read value at full scale (>2040 counts) on any channel | error |
84 | Low Lamp Level Readhead 2 | All channels read < 200 counts above dark value at calibration | error |
85 | Low Channel Output Readhead 2 | One or more channels read < 200 counts above dark at calibration | error |
86 | Missing Strip Readhead 2 | Both areas between pads 2-3 and 9-10 do not read >50% | error |
88 | Misaligned Strip Readhead 2 | Strip end position reflectance crosses 30% threshold at position 0 or 8 | error |
89 | Skewed Strip Readhead 2 | Reflectance >65% at one area but <65% in other between pads 2-3 and 9-10 | error |
90 | Reflectance >100% Readhead 2 | Reflectance for any pad measures more than 100% | error |
93 | Result Data Integrity Readhead 2 | Checksum sent by Processor 2 does not match checksum calculated by Processor 1 | critical |
DT001 | Display blank - no video output | Power applied but display shows no output | critical |
DT002 | Display malfunction - garbled or incorrect output | Display shows corrupted, partial, or unreadable characters | error |
DT003 | Display character string or touch screen operation corrected by power cycle | After power cycle and touch screen recalibration display/touch responds correctly | info |
DT004 | System memory and defaults reset corrected problem | After memory reset and defaults reset system responds normally | info |
DT005 | Display and processor issue resolved after troubleshooting | After troubleshooting display, main PCB and program card system returns to normal | info |
DT006 | Display or processor issue persists after troubleshooting | Display does not update to run mode despite troubleshooting steps | error |
DT007 | Touch screen response corrected by power cycle and recalibration | After power cycle and touch screen recalibration touch screen responds to input | info |
DT008 | Touch screen not responding after power cycle | Touch screen still unresponsive after power cycle restart | error |
DT009 | Touch screen malfunction - adjacent areas around keys not responding | Touch screen shows localized dead spots or areas not responding | error |
DT010 | Pusher bar not at left starting position | Pusher bar position indicator shows bar is not fully retracted to left stop | error |
DT011 | Strip detector does not detect strip or pusher bar does not cycle | Strip detector fails to sense test strip presence or initiate pusher bar movement | error |
DT012 | System locked up - unresponsive to input | System becomes unresponsive to user input; processing halted | critical |
DT013 | Display does not update to run mode after data transfer attempt | Display remains in standby or menu mode; does not show active test results | error |
DT014 | Touch screen hardware defective and requires replacement | Touch screen does not respond to input despite power cycle and recalibration attempts | error |
DT015 | Instrument setup parameters incorrect or reset to defaults needed | Print out shows incorrect setup configuration or customer settings conflict with defaults | warning |
Error 01 | Instrument optical error | Optical measurement system failure during strip reading | error |
Error 02 | Instrument optical error | Optical measurement system failure during strip reading | error |
Error 03 | Instrument optical error | Optical measurement system failure during strip reading | error |
Error 04 | Instrument optical error | Optical measurement system failure during strip reading | error |
Error 05 | Instrument optical error | Optical measurement system failure during strip reading | error |
Error 06-2 | Reagent strip lost between readheads | Strip detected at readhead 1 but not at readhead 2 | error |
Error 07-1 | Reagent strip not fully wetted or upside-down | Insufficient wetting or inverted strip orientation detected during processing | warning |
Error 08-1 | Reagent strip misaligned during processing (readhead 1) | Strip position correction factor of 0 or 8 detected at readhead 1 | error |
Error 08-2 | Reagent strip misaligned during processing (readhead 2) | Strip position correction factor of 0 or 8 detected at readhead 2 | error |
Error 09-1 | Reagent strip skewed during processing (readhead 1) | Strip rotation or tilt detected at readhead 1 during transport | error |
Error 09-2 | Reagent strip skewed during processing (readhead 2) | Strip rotation or tilt detected at readhead 2 during transport | error |
Error 10-1 | Instrument optical error | Optical measurement system failure during strip reading | error |
Error 10-2 | Instrument optical error | Optical measurement system failure during strip reading | error |
Error 20-1 | Automatic strip type detection failed (Kanji units only) | Strip type could not be auto-detected when Auto Detect mode is enabled | warning |
Error 21 | Internal memory error | Memory integrity check failed or program card communication lost | critical |
Error 23 | Moving Table misaligned | Moving table position sensor detects misalignment | error |
Error 24 | Push bar mechanical error - not finding home in proper time | During instrument startup or operation when push bar fails to reach home position | error |
Error 25 | Readhead could not find calibration chip during initialization | During startup calibration sequence when readhead fails to locate cal chip | error |
Error 26 | Fixed table is missing or not installed properly | During initialization when fixed table is not detected or not fully engaged | error |
Error 27 | Holddown is improperly installed, missing, or dirty | During specimen processing when holddown is not properly positioned | warning |
Error 28 | Reagent strip detected at loading station but not found at first readhead | During specimen processing when strip is placed but not transported to first readhead | error |
Error 29 | Calibration bar error detected | During startup calibration when calibration bars are not detected or damaged | error |
Error 30 | Readhead alignment error detected | During startup when readhead position is incorrect or misaligned | error |
General Error 30 | Readhead #2 calibration edge correction out of range | RH2 Calibration Edge Correction value outside +/- 3 range | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.