Smiths Medical CADD-Solis: troubleshooting and error codes
The ServiceForge knowledge base holds 35 documented symptoms for the Smiths Medical CADD-Solis, with 44 ranked causes, 58 diagnostic checks with test points and expected values, and 10 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
LCD backlight not visible
Display content cannot be seen; LED backlight is off or too dim
Ranked causes
- 1. Microprocessor has not enabled LCD (unknown)
- 2. Brightness setting too low (common)
- 3. LED backlight failed (occasional)
Verify LCD is enabled by turning on pump display
- Test point
- Front panel controls
- Expected
- Display activates and shows content
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data loss from memory during power-off patient safety
Delivery and record keeping parameters may be lost if backup battery is discharged
Ranked causes
- 1. Backup battery discharged or defective (rare)
- 2. Insufficient charge time for completely discharged backup battery (occasional)
Verify backup battery charge status
- Test point
- Battery backed RAM circuit
- Expected
- Minimum 10 months backup capacity
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD display not visible
LCD panel cannot be read even though power is on
Ranked causes
- 1. LED backlight disabled to conserve power (common)
- 2. LCD drive circuitry disabled by microprocessor (occasional)
Increase brightness setting
- Test point
- Display settings menu
- Expected
- Display becomes visible
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD display not visible or dim
LCD display is dark or difficult to read; LED backlight dim or off
Ranked causes
- 1. LED backlight disabled by microprocessor to conserve battery (common)
- 2. LCD circuit power supply fault (rare)
Check LCD brightness setting
- Test point
- Pump display menu
- Expected
- Brightness level at or above 50%
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery alarm activating unexpectedly patient safety
Pump emits beeps and displays low battery message despite adequate battery level; pump may shut down prematurely
Ranked causes
- 1. High internal resistance battery or battery type mismatch (common)
- 2. Battery voltage noise causing false low battery detection (occasional)
Verify battery type installed
- Test point
- Battery compartment label or inserted batteries
- Expected
- Four AA alkaline batteries
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump shuts down due to low operating voltage patient safety
Pump turns off; no operation possible
Ranked causes
- 1. Battery voltage dropped below minimum operating threshold (common)
- 2. Use of carbon-zinc batteries or other high internal resistance batteries (occasional)
Verify battery type installed
- Test point
- Pump battery compartment
- Expected
- Four AA alkaline batteries (not carbon-zinc or other types)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump shuts down due to low voltage patient safety
Pump powers off unexpectedly; operation stops
Ranked causes
- 1. Operating voltage below minimum threshold (common)
- 2. Carbon-zinc or high-resistance batteries installed (occasional)
Verify battery type installed
- Test point
- Battery compartment
- Expected
- Four AA alkaline batteries
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump shuts down unexpectedly patient safety
Pump powers off due to insufficient operating voltage
Ranked causes
- 1. Operating voltage dropped below minimum threshold (common)
- 2. High internal resistance in battery (carbon-zinc or depleted battery) (occasional)
Check battery type in use
- Test point
- Battery compartment
- Expected
- Four AA alkaline batteries or rechargeable pack
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm inoperative after power loss patient safety
Audible alarm does not sound when pump loses all power during RUN mode
Ranked causes
- 1. Capacitor backup discharged or failed; pump was not powered for minimum backup charge time (occasional)
Verify alarm circuitry capacitor charge status
- Test point
- Pump power and alarm function
- Expected
- Audible alarm sounds for 2 minutes after power loss if pump powered for 2+ minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm volume not functioning
Pump should sound alarm at three selectable volumes: low, medium, and high
Ranked causes
- 1. Audible alarm circuit malfunction (unknown)
Verify low volume alarm
- Test point
- Audible alarm output
- Expected
- Audible alarm sound at low volume
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backup memory loss during extended off-time patient safety
Delivery and record keeping parameters lost when pump powered off
Ranked causes
- 1. Battery backup exhausted after more than 10 months with pump off (rare)
Verify pump retains parameters after power down
- Test point
- Pump memory after 10+ months off
- Expected
- Delivery and record keeping parameters intact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backup RAM losing settings after power loss patient safety
Pump delivery and record keeping parameters lost when pump powered off
Ranked causes
- 1. Backup battery fully discharged or failed (rare)
Verify pump parameters retained after power cycle
- Test point
- Pump display and memory
- Expected
- Delivery parameters and records preserved after power off and on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backup memory failure patient safety
Delivery and record keeping parameters lost after extended power loss
Ranked causes
- 1. Internal lithium-manganese dioxide backup battery discharged or failed (rare)
Power cycle pump and verify stored parameters
- Test point
- Pump display menu
- Expected
- Delivery rate and parameters retained after power off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery compartment damaged or contains debris caution
Battery compartment shows cracks, breaks, corrosion, or contains foreign material; battery contacts damaged
Ranked causes
- 1. Physical damage, corrosion, or contamination of battery compartment (unknown)
Visually inspect battery compartment
- Test point
- Battery compartment interior, contacts, housing
- Expected
- No damage, cracks, debris, or corrosion; battery contacts intact and clean
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery depleted message and continuous beeping patient safety
Display shows battery depleted warning; pump emits continuous beep tone; pump may not operate
Ranked causes
- 1. Battery voltage too low for safe operation (common)
Measure battery voltage
- Test point
- Battery pack terminal
- Expected
- Below 4.5 V nominal; depleted
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery door does not fit snugly or is cracked caution
Battery door loose, does not close properly, cracked or broken; seal improperly positioned
Ranked causes
- 1. Battery door damage, misalignment, or seal degradation (unknown)
Close and latch battery door
- Test point
- Battery door
- Expected
- Fits snugly in place, not cracked or broken, seal in proper position
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fallout alarm not activating
Opening battery door should trigger continuous two-tone alarm for minimum 2 minutes
Ranked causes
- 1. Battery fallout alarm malfunction (unknown)
Verify pump has power
- Test point
- LED indicators
- Expected
- LED flashing periodically
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery fallout alarm not functioning patient safety
Pump fails to sound continuous two-tone alarm when battery power is lost and AC adapter disconnected
Ranked causes
- 1. Battery fallout alarm circuit malfunction (unknown)
Program continuous rate to 1.0 mL/hr
- Test point
- pump display
- Expected
- 1.0 mL/hr set
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette latch does not operate smoothly or cassette not firmly attached patient safety
Latch mechanism rough, stiff, or cassette loose; cassette does not seat firmly when latched
Ranked causes
- 1. Latch mechanism damage or wear; cassette seating issue (unknown)
Attach medication cassette to pump
- Test point
- Cassette latch handle
- Expected
- Latch handle in upright position; smooth operation; status bar shows 'Reservoir Cassette Latched'
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette lock does not turn or engage patient safety
Lock does not turn smoothly clockwise; cassette cannot be secured in locked position
Ranked causes
- 1. Lock mechanism jamming, damage, or wear (unknown)
Insert key into cassette lock and turn clockwise
- Test point
- Cassette lock
- Expected
- Key turns smoothly clockwise into locked position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Depleted battery condition
Battery depleted message appears on display; pump beeps continuously
Ranked causes
- 1. Battery fully depleted (common)
Verify pump shuts down due to low voltage
- Test point
- Power supply circuit
- Expected
- Pump shuts off automatically to prevent data loss
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display brightness affecting battery life
Raising LCD brightness setting decreases pump battery life; low brightness conserves power
Ranked causes
- 1. LED backlight power consumption from brightness setting (common)
Monitor battery discharge rate at current brightness setting
- Test point
- Battery voltage during operation
- Expected
- Slower discharge at lower brightness; faster at higher brightness
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Downstream occlusion pressure not within range
Pump should detect high pressure condition and trigger alarm when tubing is clamped during PCA dose delivery
Ranked causes
- 1. Occlusion pressure sensor calibration drift (unknown)
Prepare pump and cassette for occlusion test
- Test point
- Medication cassette and tubing
- Expected
- Cassette attached, locked, tubing primed with distilled water free of air bubbles, distal clamp closed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Downstream occlusion pressure out of range patient safety
Pump does not generate expected pressure or high pressure alarm does not sound when delivering PCA dose against clamped distal tubing
Ranked causes
- 1. Occlusion pressure sensor offset or gain calibration error (unknown)
Turn pump on and attach medication cassette with distilled water
- Test point
- cassette connector
- Expected
- Cassette latched and locked
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad keys feel flat or lack tactile response
Keys do not have distinctive tactile feel when pressed; feel flat instead of responsive
Ranked causes
- 1. Keypad mechanical wear or damage (unknown)
Press each key on keypad
- Test point
- All keypad keys
- Expected
- Distinctive tactile feel, keys not flat
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LCD backlight flashing during motor operation
Backlight blinks or flashes when pump motor activates
Ranked causes
- 1. Normal motor operation flashing behavior (common)
Measure backlight flash duration during motor operation
- Test point
- LCD backlight
- Expected
- Flash for 12 mS (milliseconds) per motor cycle
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED indicator not functioning
Amber LED should blink every 3 seconds in STOP mode; Green LED should blink every 3 seconds in RUN mode
Ranked causes
- 1. LED indicator circuit malfunction (unknown)
Verify Amber LED in STOP mode
- Test point
- LED indicator
- Expected
- Blink approximately every 3 seconds
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery alarm activated
Pump displays low battery message and emits 3 beeps every 5 minutes; pump continues to operate but with reduced battery life warning
Ranked causes
- 1. Battery capacity between 25% and low threshold (common)
Verify battery status display
- Test point
- Pump LCD display
- Expected
- Battery low message displayed with 3 beeps every 5 minutes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery condition
Battery low message appears on display; 3 beeps every 5 minutes; LCD backlight flashes 12 mS during motor operation
Ranked causes
- 1. Battery state between 25% and low threshold (common)
Verify battery state measurement
- Test point
- Power supply circuit
- Expected
- Between 25% and low threshold before depletion
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery message and 3 beeps every 5 minutes
Battery low warning appears on display; pump emits beep pattern every 5 minutes; pump continues to operate
Ranked causes
- 1. Transition to low battery condition; supply voltage below threshold (common)
Measure battery voltage
- Test point
- Battery pack terminal or pump power input
- Expected
- Low battery threshold (typically 4.5–5.0 V nominal range); below full charge (6.0 V) but above depleted
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery warning at 25–100% battery state
No alarm, pump operates normally
Ranked causes
- 1. Battery level is adequate (common)
Pump does not power on or system fault appears on startup
No audible beeps on power up, system fault screen displays before home screen, or LCD shows error
Ranked causes
- 1. Electrical or mechanical fault in pump (unknown)
Insert 4 AA batteries and press and hold ON/OFF key
- Test point
- LCD display and audio output
- Expected
- Six audible beeps occur; home screen displays without prior fault screen
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System fault alarm displayed high voltage
Unrecoverable hardware or software fault detected. Amber indicator LED continuously illuminated with audible two-tone alarm.
Ranked causes
- 1. Hardware or software fault detected (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System fault appears on power-up
LCD displays system fault screen before home screen appears during power-up sequence
Ranked causes
- 1. Electrical or mechanical fault (unknown)
Insert 4 AA batteries and press ON/OFF key; observe LCD during power-up
- Test point
- LCD display
- Expected
- No system fault screen; six audible beeps occur; pump displays home screen
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Visible damage to pump housing, LCD, sensors, or connectors caution
Cracks, breaks, or damage visible on LCD, occlusion sensor seals, valves, expulsor, cassette hinge, latch handle, lock, sensors, keypad, lights, AC jack, remote dose jack, USB connector, air detector, or housings
Ranked causes
- 1. Physical damage from mishandling, drop, or wear (unknown)
Visually inspect pump housing, LCD, and all seals
- Test point
- Entire pump exterior
- Expected
- No cracks, breaks, or visible damage; all seals intact
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Battery depleted | Battery completely depleted; pump will shut down soon | Battery state reaches fully depleted threshold | critical |
Battery depleted | Battery voltage is critically low; pump cannot continue safe operation | Battery voltage drops below depleted threshold after low battery condition | critical |
Depleted battery | Battery voltage has dropped to critically low level | Battery state falls below operational threshold after low battery condition | critical |
Depleted battery | Battery charge below minimum operating threshold | Battery voltage drops below depleted level during pump operation | critical |
Depleted battery warning | Battery power is too low to operate the pump | Battery voltage drops below operational threshold | critical |
Low battery | Battery capacity has fallen to low threshold | Battery state between 25% and low threshold | warning |
Low battery | Battery power is low but pump remains operable | Battery state transitions below 25% capacity threshold | warning |
Low battery | Battery power is depleted to 25% of full charge or below | Battery voltage drops to low battery threshold during pump operation | warning |
Pump shutdown due to low voltage | Pump automatically powers off because operating voltage is too low | Operating voltage drops below minimum threshold required for safe pump operation | critical |
Shutdown due to low voltage | Pump has shut down because operating voltage is insufficient | Supply voltage falls below minimum operating threshold during motor operation | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI