Smiths Medical 1818R: troubleshooting and error codes
The ServiceForge knowledge base holds 13 documented symptoms for the Smiths Medical 1818R, with 23 ranked causes, 13 diagnostic checks with test points and expected values, and 2 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No pulse shown on bar graph patient safety
Pulse waveform not displayed on oximeter bar graph
Ranked causes
- 1. Patient cable or sensor disconnected from oximeter (common)
- 2. Sensor incorrectly positioned on patient (common)
- 3. Poor patient perfusion (occasional)
- 4. Defective sensor or patient cable (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No printout on optional printer
Printer connected but not producing output
Ranked causes
- 1. Printer power not connected or switch off (common)
- 2. Printer interface assembly not securely connected (common)
- 3. Printer interface malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No trends in memory
Trend data not stored or accessible in oximeter memory
Ranked causes
- 1. Trend data not collected (common)
- 2. Sensor disconnected during data collection (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oximeter does not turn on
Unit does not power up when turned on
Ranked causes
- 1. Weak batteries (common)
- 2. Batteries not installed or incorrectly installed (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oximeter turns off unexpectedly
Unit powers down without operator command
Ranked causes
- 1. Automatic shutoff feature activated (common)
- 2. Weak or dead batteries (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulse rate erratic, intermittent, or incorrect patient safety
Pulse rate display unstable, skips beats, or shows inaccurate readings
Ranked causes
- 1. Sensor incorrectly positioned (common)
- 2. Patient motion (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 value erratic, intermittent, or incorrect patient safety
Oxygen saturation reading unstable or shows inaccurate values
Ranked causes
- 1. Poor patient perfusion (common)
- 2. Patient motion (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clock frequency out of specification
Measured oscillator frequency at U11-71 does not match specified 3.072 MHz
Ranked causes
- 1. Oscillator frequency drift or crystal failure (occasional)
Measure clock frequency at U11-71
- Test point
- U11-71
- Expected
- 3.072 MHz, 50% duty cycle
- Tools
- Oscilloscope or frequency counter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Current draw exceeds specification
Total battery current through J9-1 exceeds nominal 110 mA ±15 mA under standard operating conditions
Ranked causes
- 1. Excessive power consumption from circuit or sensor (common)
Measure total battery current at J9-1
- Test point
- J9-1
- Expected
- 110 mA, ±15 mA
- Tools
- DC power supply, ammeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low battery LED does not illuminate
LED indicator fails to turn on when battery voltage drops to expected threshold
Ranked causes
- 1. Battery voltage not dropping to LED threshold (common)
Measure voltage at J9-1 during battery discharge
- Test point
- J9-1
- Expected
- 3.00 VDC ±0.2 VDC, ±0.2 VDC
- Tools
- DC power supply, voltmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oximeter does not shut off at low battery
Unit continues operating below shutdown threshold voltage instead of powering down
Ranked causes
- 1. Battery voltage below shutdown threshold (common)
Monitor oximeter shutdown during simulated battery discharge
- Test point
- J9-1
- Expected
- Oximeter off at 2.7 VDC ±0.2 VDC, ±0.2 VDC
- Tools
- DC power supply, voltmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oximeter does not turn off after timeout
Unit remains powered after finger removed from sensor or sensor disconnected, beyond expected 2-minute timeout
Ranked causes
- 1. Auto-shutoff timer malfunction (rare)
Measure auto-shutoff time with finger removal
- Test point
- Sensor input
- Expected
- Shutdown in 2 minutes ±5 seconds
- Tools
- Timer, sensor
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage level out of specification at test point
DC voltage measured at specified test point does not match nominal value or range
Ranked causes
- 1. Power supply regulation failure (common)
Measure T19 output voltage
- Test point
- T19
- Expected
- 5 VDC, ±0.1 VDC
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
E00 | ROM error | Read-only memory failure detected | critical |
E01 | RAM error | Random-access memory failure detected | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI