Skytron 3501B: troubleshooting and error codes
The ServiceForge knowledge base holds 52 documented symptoms for the Skytron 3501B, with 207 ranked causes, 78 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Back section function moves improperly patient safety
Back section articulation slow, jerky, or unresponsive in one or both directions
Ranked causes
- 1. Incorrect speed adjustment (common)
- 2. Spool valve not centered (occasional)
- 3. Defective check valves (occasional)
- 4. Low on oil (occasional)
- 5. Pinched hose (occasional)
- 6. Defective mini-valve (occasional)
- 7. Pressure relief valve not set properly (occasional)
- 8. Defective solenoid (occasional)
- 9. Defective relay box or pendant control (occasional)
- 10. Kidney bridge raised (occasional)
- 11. Loose retaining hardware (occasional)
- 12. Microswitch out of adjustment (occasional)
- 13. Loose table top (occasional)
Leg function moves improperly patient safety
Leg section articulation slow, jerky, or unresponsive in one or both directions
Ranked causes
- 1. Incorrect speed adjustment (common)
- 2. Spool valve not centered (occasional)
- 3. Defective check valves (occasional)
- 4. Low on oil (occasional)
- 5. Pinched hose (occasional)
- 6. Defective mini-valve (occasional)
- 7. Pressure relief valve not set properly (occasional)
- 8. Defective solenoid (occasional)
- 9. Defective relay box or pendant control (occasional)
- 10. Loose retaining hardware (occasional)
- 11. Microswitch out of adjustment (occasional)
- 12. Loose table top (occasional)
Slide function moves improperly patient safety
Slide motion slow, jerky, or unresponsive; table does not slide smoothly toward head or foot
Ranked causes
- 1. Incorrect speed adjustment (common)
- 2. Spool valve not centered (occasional)
- 3. Defective check valve (occasional)
- 4. Low on oil (occasional)
- 5. Pinched hose (occasional)
- 6. Defective mini-valve (occasional)
- 7. Pressure relief valve not set properly (occasional)
- 8. Defective solenoid (occasional)
- 9. Defective relay box or pendant control (occasional)
- 10. Loose retaining hardware (occasional)
- 11. Microswitch out of adjustment (occasional)
- 12. Loose table top (occasional)
Back section or trendelenburg flex function moves improperly patient safety
Flex system slow, jerky, or unresponsive; back section and trendelenburg not coordinating
Ranked causes
- 1. Incorrect speed adjustment for flex, trendelenburg, or back section (common)
- 2. Spool valve not centered (occasional)
- 3. Defective check valves (occasional)
- 4. Low on oil (occasional)
- 5. Pinched hose (occasional)
- 6. Defective mini-valve (occasional)
- 7. Pressure relief valve not set properly (occasional)
- 8. Defective solenoid (occasional)
- 9. Defective relay box or pendant control (occasional)
- 10. Kidney bridge raised (occasional)
- 11. Worn ball joint or retainer nut (occasional)
Kidney lift moves improperly caution
Kidney lift cylinder does not move smoothly or at incorrect speed during operation
Ranked causes
- 1. Incorrect speed adjustment (common)
- 2. Spool valve not centered (common)
- 3. Defective check valve (occasional)
- 4. Low on oil (common)
- 5. Pinched hose (occasional)
- 6. Defective mini-valve (occasional)
- 7. Pressure relief valve not set properly (occasional)
- 8. Defective solenoid (occasional)
- 9. Defective relay box or pendant control (occasional)
- 10. Bypass valve out of adjustment (occasional)
Trendelenburg function moves improperly patient safety
Trendelenburg position change slow, jerky, or unresponsive in one or both directions
Ranked causes
- 1. Incorrect speed adjustment (common)
- 2. Spool valve not centered (occasional)
- 3. Defective check valves (occasional)
- 4. Low on oil (occasional)
- 5. Pinched hose (occasional)
- 6. Defective mini-valve (occasional)
- 7. Pressure relief valve not set properly (occasional)
- 8. Defective solenoid or wiring (occasional)
- 9. Defective relay box or pendant control (occasional)
- 10. Worn ball joint or retainer nut (occasional)
Brakes will not set properly patient safety
Brake cylinders fail to lock and hold the surgical table in position
Ranked causes
- 1. Emergency brake release valve open or defective (common)
- 2. Spool valve not centered (occasional)
- 3. Defective check valve (occasional)
- 4. Low on oil (common)
- 5. Pressure relief valve not set properly (occasional)
- 6. Pinched hose (occasional)
- 7. Defective mini-valve (occasional)
- 8. Defective relay box or pendant control (occasional)
- 9. Loose retaining hardware (occasional)
Lateral tilt function moves improperly patient safety
Lateral tilt slow, jerky, or unresponsive; table does not tilt smoothly left or right
Ranked causes
- 1. Incorrect speed adjustment (common)
- 2. Spool valve not centered (occasional)
- 3. Defective check valves (occasional)
- 4. Low on oil (occasional)
- 5. Pinched hose (occasional)
- 6. Defective mini-valve (occasional)
- 7. Pressure relief valve not set properly (occasional)
- 8. Defective solenoid (occasional)
- 9. Defective relay box or pendant control (occasional)
Table will not elevate properly patient safety
Elevation function insufficient or non-responsive; table cannot raise or raises slowly
Ranked causes
- 1. Pressure relief valve not set properly (common)
- 2. Low on hydraulic oil (common)
- 3. Spool valve not centered (occasional)
- 4. Defective pump (occasional)
- 5. Defective mini-valve (occasional)
- 6. Defective solenoid or wiring (occasional)
- 7. Defective relay box or pendant control (occasional)
- 8. Leaking cylinder hose (occasional)
- 9. Uneven weight distribution (rare)
Brakes will not retract properly caution
Brake cylinders do not fully retract when brake unlock is activated
Ranked causes
- 1. Incorrect speed adjustment (occasional)
- 2. Defective check valve (occasional)
- 3. Spool valve not centered (occasional)
- 4. Defective mini-valve (occasional)
- 5. Pinched hose (occasional)
- 6. Defective solenoid or wiring (occasional)
- 7. Defective relay box or pendant control (occasional)
- 8. Defective brake cylinder (occasional)
Table will not descend properly patient safety
Descent function insufficient or non-responsive; table cannot lower or lowers slowly
Ranked causes
- 1. Incorrect speed adjustment (common)
- 2. Defective check valve (common)
- 3. Spool valve not centered (occasional)
- 4. Galled slider assembly (occasional)
- 5. Defective solenoid or wiring (occasional)
- 6. Defective relay box or pendant control (occasional)
- 7. Uneven weight distribution (rare)
Kidney lift chatters or loses position caution
Kidney lift cylinders chatter during operation or fail to maintain position
Ranked causes
- 1. Defective or dirty check valve (common)
- 2. Oil leakage in circuit (common)
- 3. Air inside cylinder (occasional)
- 4. Pinched hose (occasional)
- 5. Low on oil (common)
- 6. Lift rods binding (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back section function chatters or loses position patient safety
Back section unstable; function moves erratically or loses position when locked
Ranked causes
- 1. Defective or dirty check valves (common)
- 2. Oil leakage in circuit (common)
- 3. Air inside cylinder (occasional)
- 4. Pinched hose (occasional)
- 5. Low on oil (occasional)
Back section or trendelenburg flex function chatters or loses position patient safety
Flex function unstable; back section and trendelenburg move erratically or lose position
Ranked causes
- 1. Defective or dirty check valves (common)
- 2. Oil leakage in circuit (common)
- 3. Air inside cylinder (occasional)
- 4. Pinched hose (occasional)
- 5. Low on oil (occasional)
Lateral tilt function chatters or loses position patient safety
Lateral tilt unstable; function moves erratically or loses position when locked
Ranked causes
- 1. Defective or dirty check valves (common)
- 2. Oil leakage in circuit (common)
- 3. Air inside cylinder (occasional)
- 4. Pinched hose (occasional)
- 5. Low on oil (occasional)
Leg function chatters or loses position patient safety
Leg section unstable; function moves erratically or loses position when locked
Ranked causes
- 1. Defective or dirty check valves (common)
- 2. Oil leakage in circuit (common)
- 3. Air inside cylinder (occasional)
- 4. Pinched hose (occasional)
- 5. Low on oil (occasional)
Slide function chatters or loses position patient safety
Slide motion unstable; function moves erratically or loses position when locked
Ranked causes
- 1. Defective or dirty check valve (common)
- 2. Oil leakage in circuit (common)
- 3. Air inside cylinder (occasional)
- 4. Pinched hose (occasional)
- 5. Low on oil (occasional)
Trendelenburg function chatters or loses position patient safety
Trendelenburg position unstable; function moves erratically or loses position when locked
Ranked causes
- 1. Defective or dirty check valve (common)
- 2. Oil leakage in circuit (common)
- 3. Air inside cylinder (occasional)
- 4. Pinched hose (occasional)
- 5. Low on oil (occasional)
Brakes will not stay locked patient safety
Brakes release or lose pressure after being set, allowing table to drift
Ranked causes
- 1. Emergency brake release valve open or defective (common)
- 2. Defective or dirty check valve (common)
- 3. Oil leakage in circuit (common)
- 4. Leaking O-ring inside cylinder (occasional)
Table does not respond to control inputs
One or all control functions fail to operate; table may not move in any direction
Ranked causes
- 1. Motor/Pump Assembly defective (unknown)
- 2. Main Switch Circuit and Wiring fault (unknown)
- 3. Pendant control defective (unknown)
- 4. Relay Box defective (unknown)
Switch operating mode from AC120V to BATTERY
- Test point
- Main power mode selector
- Expected
- Function operates in alternate mode
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Table loses elevation position patient safety
Table descends or sinks after being elevated and locked; cannot maintain set height
Ranked causes
- 1. Defective check valve (common)
- 2. Leaking mini-valve (common)
- 3. Loose fittings, joints, or hoses (occasional)
- 4. Leaking O-ring inside cylinder (occasional)
Casters dirty or not rotating freely caution
Casters rough or sticky when rotated; debris and old grease present; table positioning difficult
Ranked causes
- 1. Debris and old grease accumulation in casters (common)
- 2. Casters not lubricated or worn bearing surfaces (occasional)
- 3. Casters worn or defective (occasional)
Inspect each caster for debris and old grease
- Test point
- Caster assembly
- Expected
- Clean casters with no debris or old grease
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main wire harness continuity broken
Wiring between major components (battery, relay box, pump, solenoids) has open circuit or bad connection
Ranked causes
- 1. Battery circuit breaker (15 amp) open or in line fault between CN4 pin 1 and battery (+) (common)
- 2. Open or broken wire between CN4 and HN4 connectors (pump power) (occasional)
- 3. Open or broken wire between CN4 and HN3 (capacitor/rectifier unit) connectors (occasional)
Continuity from CN4 pin 1 to battery (+) connector
- Test point
- CN4 pin 1(RED) and battery positive terminal
- Expected
- continuity (0 ohms)
- Tools
- Ohmmeter
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor/pump assembly not producing hydraulic pressure high voltage
Hydraulic pump not delivering oil flow when power applied, affecting all table functions
Ranked causes
- 1. Motor or wiring defective; motor resistance out of specification (common)
- 2. No input power to motor due to relay box or wiring fault (common)
- 3. Thermal protector tripped after continuous operation (occasional)
Motor input voltage (AC120V mode, all connectors connected)
- Test point
- HN4 pins 1(+) RED and 2(-) BLUE
- Expected
- approximately 24-28 volts when function button activated V, 24-28 V
- Tools
- DC voltmeter
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Return-to-level function not working patient safety
LEVEL button on pendant control does not automatically level major table functions (lateral tilt, Trendelenburg, back section, leg section)
Ranked causes
- 1. Micro-switch circuit disconnected or faulty at CN10 connector (common)
- 2. Defective micro-switch (NS-1 through NS-7) (occasional)
- 3. Riser cord or wiring continuity broken between CN10 and terminal strips TM1/TM2 (occasional)
Verify table functions work without CN10 connected
- Test point
- CN10 connector at relay box
- Expected
- All functions operate properly using Pendant Control
11 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary switches not functioning
Switch controls for table movements (up, down, tilt, trend, back, kidney, leg, brake) not responding to activation
Ranked causes
- 1. Wires loose or bad connection at CN2 connector or switch circuit board (common)
- 2. Individual switch defective (occasional)
Continuity at CN2 connector pins
- Test point
- CN2 pins (pin 1A common)
- Expected
- 0 ohms
- Tools
- Ohmmeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No AC120V operation available high voltage
Table functions fail when operating in AC120V mode
Ranked causes
- 1. Power cord not connected or faulty (common)
- 2. Transformer output voltage incorrect (occasional)
Confirm 120VAC input at main switch output
- Test point
- Connector HN1 pins 1 and 2
- Expected
- 120 VAC
- Tools
- AC voltmeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No BATTERY mode operation available
Table functions fail when operating in BATTERY mode
Ranked causes
- 1. Battery voltage below minimum (common)
- 2. Battery voltage drops below 11VDC under load (occasional)
Test individual battery voltage at rest
- Test point
- Battery 1 and Battery 2 terminals
- Expected
- 12 VDC ± 1 V each, ±1 V
- Tools
- DC voltmeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rectifier or charging system fault high voltage
DC voltage outputs from Capacitor/Rectifier Unit incorrect
Ranked causes
- 1. Capacitor/Rectifier Unit defective (unknown)
- 2. Transformer defective (unknown)
Test DC voltage output pin 1 (purple) to ground
- Test point
- Connector HN3 pins 1 (purple) and 2 (black, ground)
- Expected
- 30 VDC
- Tools
- DC voltmeter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Side rails loose or gravity stops restricted patient safety
Retaining hardware loose on side rail sections or gravity stops do not move freely
Ranked causes
- 1. Retaining hardware loose or corroded (common)
- 2. Gravity stops stuck or damaged (occasional)
Inspect side rail sections for loose retaining hardware
- Test point
- Side rail fasteners
- Expected
- All hardware tight and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Single control function inoperative
One directional or function button does not respond
Ranked causes
- 1. Solenoid for that function defective (common)
- 2. Auxiliary switch or relay for function faulty (occasional)
Test function with auxiliary switches
- Test point
- Auxiliary switch for the specific function
- Expected
- Function operates when auxiliary switch is used
Solenoid coil defective
Solenoid not energizing or holding, preventing hydraulic valve spool movement in one or more directions
Ranked causes
- 1. Solenoid coil resistance out of specification or open circuit (common)
- 2. Thermal fuse inside solenoid blown from continuous operation (occasional)
Solenoid coil resistance
- Test point
- Between both pins at solenoid connector
- Expected
- approximately 16 ohms at room temperature, approximately 16 ohms
- Tools
- Ohmmeter R x 1 scale
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Top slide function not moving toward foot end patient safety
Slide mechanism remains stationary or will not move toward foot end when control is activated
Ranked causes
- 1. Back section below horizontal (CS-1 inhibit switch active) (common)
- 2. CS-1 Back-Down Inhibit switch malfunction (occasional)
Check back section position
- Test point
- Back section visual position or angle indicator
- Expected
- Back section at or above horizontal
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Top slide function not moving toward head end patient safety
Slide mechanism remains stationary or will not move toward head end when control is activated
Ranked causes
- 1. Leg section lowered more than 45° below horizontal (CS-2 inhibit active) (common)
- 2. CS-2 Leg-Down 45° Inhibit switch malfunction (occasional)
Check leg section position
- Test point
- Leg section visual position or angle indicator
- Expected
- Leg section within 45° below horizontal
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Trendelenburg ball joint has visible play patient safety
Movement observed when table top is slid to extremities and load applied; no visible play is permitted
Ranked causes
- 1. Ball joint collar or locking allen bolt loose (common)
- 2. Ball socket wear or metal particulate damage (occasional)
Slide table top to each extremity and apply load
- Test point
- Trendelenburg ball joint
- Expected
- No visible play or movement in joint
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back section cannot be lowered below horizontal when slide is toward foot patient safety
Back section control non-responsive or locked when table top is slid toward foot end
Ranked causes
- 1. CS-1 Back-Down Inhibit switch functioning correctly (protective inhibit) (common)
Check slide position
- Test point
- Top slide position (center/head/foot)
- Expected
- Slide at center or toward head end
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back section cannot be raised more than 45° when kidney lift is up patient safety
Back section movement is restricted; stops at 45° above horizontal when kidney lift is in raised position
Ranked causes
- 1. CS-3 Back-Up Inhibit switch functioning correctly (protective inhibit active) (common)
Verify kidney lift position
- Test point
- Kidney lift position indicator or visual inspection
- Expected
- Kidney lift fully down
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CS-5 kidney lift inhibit switch verification patient safety
Verify CS-5 kidney lift inhibit switch operation at different back section angles to ensure kidney lift is properly blocked when back section exceeds 45 degrees.
Ranked causes
- 1. Switch contact or circuit malfunction (unknown)
Measure switch continuity with back section less than 45 degrees
- Test point
- CN10 pins 12 & 24
- Expected
- Infinity
- Tools
- DMM
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Micro-switch adjustment for level return
Table does not properly return to level, or improper readings obtained during test. Requires switch bracket adjustment.
Ranked causes
- 1. Micro-switch bracket misalignment (common)
Level table top using pendant control
- Test point
- TRENDELENBURG and LATERAL-TILT function buttons
- Expected
- Table top level
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charging Indicator LEDs not illuminating
Charging indicator does not show battery charge level
Ranked causes
- 1. No signal from relay box to charging indicator (unknown)
Test charging indicator signal from relay box
- Test point
- Relay box CN3 pin 11 (common) to pin 1 (lowest charge)
- Expected
- approximately 1.8 VDC
- Tools
- DC voltmeter
Head section flexing throughout range of travel patient safety
Head section exhibits flexing during articulation; not rigid throughout travel range
Ranked causes
- 1. Bearing block misalignment or loose fasteners (common)
Place head section in level position and remove top
- Test point
- Head section
- Expected
- Head section level, top removed
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Kidney lift cannot be raised when back section is above 45° patient safety
Kidney lift remains stationary; audible alarm sounds when attempting to raise kidney lift while back section is elevated above horizontal
Ranked causes
- 1. CS-3 Back-Up Inhibit switch functioning correctly (common)
Test back section position
- Test point
- Back section angle indicator or visual position
- Expected
- Back section at or below 45° above horizontal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Kidney lift cylinder piston extension incorrect caution
Kidney lift piston-to-cylinder end distance not at specification when fully down; prevents proper operation
Ranked causes
- 1. Kidney lift cylinder removed and not reset to proper extension (common)
Measure kidney lift piston extension with kidney lift all the way down
- Test point
- Kidney lift cylinder, head end
- Expected
- 82 mm
- Tools
- Ruler or measuring tape
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Kidney lift cylinders misaligned at end of stroke caution
One kidney lift cylinder reaches full down or up position before the other
Ranked causes
- 1. Bypass valve requires adjustment (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lateral tilt thrust bearing loose or improper torque caution
Lateral tilt M8 x 20 allen bolt not properly secured; only for serial numbers before 3501B-81-357
Ranked causes
- 1. M8 x 20 lateral tilt thrust bearing allen bolt loose or improper torque (common)
Remove M8 x 20 lateral tilt thrust bearing allen bolt
- Test point
- Lateral tilt thrust bearing bolt
- Expected
- Bolt removed without damage
- Tools
- 6mm allen wrench
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leg section cannot be lowered more than 45° when slide is toward head patient safety
Leg section control non-responsive or locked at 45° when table top is slid toward head end
Ranked causes
- 1. CS-2 Leg-Down 45° Inhibit switch functioning correctly (protective inhibit) (common)
Check slide position
- Test point
- Top slide position (center/head/foot)
- Expected
- Slide at center or toward foot end
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leg section twists or flexes at full extension patient safety
Leg section shows twisting or flexing as it approaches stalled position; both sides do not stop moving at same time
Ranked causes
- 1. One cylinder not reaching fully extended position simultaneously (common)
Position leg section all the way up
- Test point
- Leg section
- Expected
- Both sides stop moving at same time with no twisting or flexing
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output voltage to pump motor high voltage
Pump/motor relay not supplying 24V power when function buttons activated
Ranked causes
- 1. Broken continuity in HN4 to CN4 wiring (red or black wires) (common)
Output voltage to pump (AC120V mode, all connectors connected)
- Test point
- HN4 pins 1(+) RED and 2(-) BLUE
- Expected
- approximately 24-28 volts when function button activated V, 24-28 V
- Tools
- DC voltmeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output voltage to solenoids high voltage
Solenoid energizing circuit not supplying 24V to hydraulic valve actuators when function buttons activated
Ranked causes
- 1. Relay box defective (common)
Solenoid output voltage at CN1A
- Test point
- CN1A connector pins (activate each function button)
- Expected
- 24 volts V
- Tools
- DC voltmeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Relay box input power absent or low high voltage
Power supply not delivering correct voltage to relay box circuits
Ranked causes
- 1. Problem in input wiring, connectors or power supply components (common)
Relay box input power voltage (BATTERY mode)
- Test point
- CN4 pins 1(+) BLUE and 2(-) BLUE
- Expected
- approximately 24-28 volts V, 24-28 V
- Tools
- DC voltmeter
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slide cylinder piston extension incorrect caution
Slide piston-to-cylinder end distance not at specification; prevents proper slide range of motion
Ranked causes
- 1. Slide cylinder removed and not reset to proper extension (common)
Measure slide piston extension with table top centered
- Test point
- Slide cylinder, head end
- Expected
- 82 mm
- Tools
- Ruler or measuring tape
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slide roller binding or uneven contact caution
Rollers not contacting slide bars evenly; table top moves with up or down movement or binds during slide
Ranked causes
- 1. Slide roller adjustment eccentric cam out of position (common)
Remove seat section top, hose cover, and lateral tilt frame covers
- Test point
- Slide roller adjustment area
- Expected
- Access to top rollers obtained
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI