Siemens Healthineers Orbic: troubleshooting and error codes

The ServiceForge knowledge base holds 72 documented symptoms for the Siemens Healthineers Orbic, with 84 ranked causes, 111 diagnostic checks with test points and expected values, and 6 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Insufficient accuracy of 3D navigation patient safety

3D navigation accuracy falls below required tolerance

Ranked causes

  1. 1. Interface at calibration phantom not clean or damaged (common)
  2. 2. Navigation marker ring on I.I. not mounted correctly (common)
  3. 3. Grid holder at I.I. is dirty (occasional)
First check

Inspect calibration phantom interface

Test point
phantom interface
Expected
clean and undamaged

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Acquisition image chain software warning during framegrabbing radiation

Software warning during image framegrabbing, usually self-corrected but may indicate component issues

Ranked causes

  1. 1. Exposure footswitch problem (occasional)
  2. 2. D1 power supply board failure in image container (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

C-arm carbon fiber structure damage caution

Visual damage to carbon fiber structure detected during inspection, ranging from superficial paint damage to cracks or flaked surfaces

Ranked causes

  1. 1. Superficial damage in painted area or running surface edges (common)
  2. 2. Large-scale destruction of carbon fiber structure (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

C-arm carbon structure damage caution

Visual inspection reveals superficial or extensive damage to carbon fiber structure, including paint damage, projecting carbon fibers, cracks, or flaked-off areas on running surfaces

Ranked causes

  1. 1. Superficial damage to painted area or running surface edges (occasional)
  2. 2. Extensive damage indicating general or large-scale destruction of carbon structure (rare)
First check

Visually inspect C-arm for damage to carbon structure

Test point
Entire C-arm carbon structure
Expected
No visible damage, cracks, flaking, or projecting fibers

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration unsuccessful radiation

Calibration process fails to complete

Ranked causes

  1. 1. Dose unstable (common)
  2. 2. Phantom positioned incorrectly (common)
First check

Verify dose stability

Test point
dose monitoring
Expected
stable dose reading

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Disk full error from image chain radiation

Disk storage capacity exceeded, causing general stop failure in image chain

Ranked causes

  1. 1. Insufficient free disk space on drive C (common)
  2. 2. D1 power supply board failure in image container (occasional)
First check

Verify free space on drive C

Test point
drive C:
Expected
more than 600MB free space

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

I.I. laser light localizer function check

Laser diodes or laser light fan beam positioning do not function or are misaligned

Ranked causes

  1. 1. Laser diode malfunction or failure (unknown)
  2. 2. Laser light fan beam misalignment (unknown)
First check

Test laser diode function

Test point
Laser diodes on localizer
Expected
Laser diodes illuminate when activated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

I.I. laser light localizer mechanical failure radiation

Mechanical damage detected on integrated or external I.I. laser light localizer, or I.I. ring loose or damaged

Ranked causes

  1. 1. Mechanical damage to localizer assembly (occasional)
  2. 2. I.I. ring loose or damaged (occasional)
First check

Test laser diode function

Test point
Integrated I.I. laser light localizer
Expected
Laser diodes activate and project light

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lifting column movement abnormality caution

Lifting column produces noises, exhibits bearing play, or fails to stop at end positions

Ranked causes

  1. 1. Bearing wear or mechanical wear (occasional)
  2. 2. End position limit switch failure (rare)
First check

Electrically move lifting column full range and listen for noise

Test point
Basic unit lifting column motor and bearings
Expected
Smooth movement with no excessive noise or play

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor trolley lifting column movement abnormality caution

Monitor trolley lifting column produces noises, exhibits bearing play, or fails to stop at end positions

Ranked causes

  1. 1. Bearing wear or mechanical wear (occasional)
  2. 2. End position limit switch failure (rare)
First check

Move motorized monitor trolley column full range and note movement

Test point
Monitor trolley lifting column motor and bearings
Expected
Smooth movement with no excessive noise or bearing play

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Radiation release switch not functioning radiation

Hand or foot switch for radiation release does not activate fluoroscopy, or cables damaged

Ranked causes

  1. 1. Electrical switch failure or contact corrosion (occasional)
  2. 2. Cable breakage or damage (occasional)
First check

Test hand switch for radiation release

Test point
Hand-operated radiation release switch
Expected
Switch activates fluoroscopy when pressed

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3D reconstruction calibration required

Routine 3D reconstruction calibration and limiting resolution check as part of maintenance

Ranked causes

  1. 1. Routine maintenance requirement (common)
First check

Calibrate 3D reconstruction

Test point
3D reconstruction system
Expected
Calibration complete per IQ quick test procedure

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3D reconstruction motor-controlled orbital movement does not stop on emergency stop patient safety

When emergency stop switch is pressed during 3D scan, motor-controlled orbital movement continues instead of stopping

Ranked causes

  1. 1. Emergency stop circuit malfunction (unknown)
First check

Press emergency stop switch during 3D scan

Test point
Emergency stop switch and orbital motor control
Expected
Motor-controlled orbital movement stops immediately

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

3D scan does not execute properly after emergency stop reset caution

After disengaging emergency stop switch, 3D scan fails to execute or orbital movement does not stop at end position

Ranked causes

  1. 1. Motor control or limit switch malfunction (unknown)
First check

Start 3D scan after emergency stop disengagement

Test point
Motor control and C-arm end-position detection
Expected
3D scan executes properly and orbital movement stops at C-arm end position

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

3D scan release blocked during image quality test

Error messages displayed when testing 3D function; 3D scan cannot be released

Ranked causes

  1. 1. C-arm angulation perpendicular rotated during transport (occasional)
First check

Verify gear mark alignment at 0 degree C-arm angulation

Test point
Cable module gears
Expected
Two marks on both gears exactly across from one another

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Acoustic warning signal not functioning radiation

Compulsory radiation shutdown alarm or fluoroscopy timer audible warning does not activate

Ranked causes

  1. 1. Acoustic buzzer failure or configuration disabled (occasional)
First check

Boot system and access local service configuration

Test point
System PC local service interface
Expected
System boots and service menu accessible

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Acquisition image chain hardware fatal error radiation

Genesis board hardware failure detected during image acquisition

Ranked causes

  1. 1. Genesis graphic board failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Acquisition image chain software warning with detailed message radiation

Software warning requiring review of extended message text for diagnosis

Ranked causes

  1. 1. Persistent software issue combined with framegrabbing error (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Angulation brake braking force insufficient patient safety

Angulation brake does not reach required braking value when engaged

Ranked causes

  1. 1. Brake spring or mechanism wear (occasional)
First check

Measure angulation brake braking force with spring scale

Test point
Angulation brake
Expected
Braking value as specified in Replacement of Parts manual for ARCADIS Orbic

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Angulation brake insufficient braking force caution

Angulation brake does not meet specified braking value when engaged

Ranked causes

  1. 1. Brake spring loss or wear (unknown)
First check

Measure angulation brake force with spring scale

Test point
Angulation brake
Expected
Braking value per ARCADIS Orbic Replacement of Parts Manual
Tools
Spring scale

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Belt tension abnormal on toothed belt for orbital movement caution

Toothed belt tension for motor-controlled orbital movement requires verification during maintenance

Ranked causes

  1. 1. Toothed belt wear or improper tension adjustment (unknown)
First check

Inspect toothed belt tension for motor-controlled orbital movement

Test point
Toothed belt
Expected
Tension within manufacturer specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

C-arm bearing play or noise detected caution

During C-arm movements, technician observes play in bearings or audible bearing noises

Ranked causes

  1. 1. Bearing wear or damage (unknown)
First check

Perform C-arm movements

Test point
C-arm bearings
Expected
No play, no bearing noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

C-arm movement excessive play or noise caution

Bearing slackness or bearing noises detected during C-arm movement

Ranked causes

  1. 1. Bearing wear or contamination (occasional)
First check

Perform all C-arm movements and listen for noise

Test point
C-arm bearings
Expected
Smooth, quiet movement with no slackness

Cable deflector integrity caution

Cable deflectors on stand or monitor trolley show damage or degradation

Ranked causes

  1. 1. Physical damage or wear to cable deflectors (unknown)
First check

Visually inspect cable deflectors on stand and monitor trolley

Test point
All cable deflector installations
Expected
No visible damage, cracks, or improper seating

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration does not start radiation

Calibration process fails to initiate

Ranked causes

  1. 1. Navigation system faulty (rare)
First check

Verify navigation system operational status

Test point
navigation system
Expected
system ready to calibrate

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 3D calibration without navigation attachment radiation

Perform 3D calibration procedure using calibration phantom; system displays success message after first step but procedure must be completed fully per description window

Ranked causes

  1. 1. Initial calibration setup required for Orbic 3D system (common)
First check

Position 3D calibration phantom

Test point
Equipment workspace
Expected
Phantom ready per factory specification

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 3D image quality setup radiation

Adjustment of calibration phantom and image intensifier light localizer for Orbic 3D system to ensure optimum 3D image quality

Ranked causes

  1. 1. Initial 3D system installation or recalibration required (common)
First check

Verify calibration phantom is positioned correctly

Test point
Calibration phantom
Expected
Positioned per Fig. 29

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: 3D phantom positioning radiation

Adjustment of calibration phantom for 3D imaging to ensure optimal image quality

Ranked causes

  1. 1. Calibration phantom misalignment (common)
First check

Fluoroscopy position monitoring

Test point
C-arm position relative to calibration phantom
Expected
C-arm stable and aligned with phantom markings

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: ADR control curve for fluoroscopy mode radiation

Verify automatic dose rate (ADR) control settings match factory baseline for S2 and HC2 exam sets in fluoroscopy mode.

Ranked causes

  1. 1. ADR control curve deviation from factory setpoints (unknown)
First check

Measure S2 exam set tube voltage

Test point
Operating panel display
Expected
66 - 72 kV
Tools
Operating panel

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Local printer function check

Verify that the local Sony UPD970/UPD990 printer produces acceptable hardcopy output with proper brightness and contrast for diagnostic image review.

Ranked causes

  1. 1. Printer brightness/contrast not optimized (common)
First check

Verify printer connection and readiness

Test point
Local printer
Expected
Printer connected and ready to operate

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Save image for export

Copy patient image from patient browser to ImageExport folder for backup to CD-ROM via ServiceLogFilesExport

Ranked causes

  1. 1. Image export configuration required (common)
First check

Open patient browser

Test point
Application menu
Expected
Patient browser window opens

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Conductive rubber on stand or monitor trolley damaged or dirty high voltage

Voltage discharge rubber on stand base or monitor trolley wheels shows damage, contamination, or missing material

Ranked causes

  1. 1. Wear, contamination, or damage to conductive rubber (common)
First check

Inspect conductive rubber on stand and monitor trolley wheels

Test point
Voltage discharge rubber on stand base and wheel assemblies
Expected
Rubber intact, clean, without damage or contamination

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cover panel mechanical damage caution

Cover panels on stand or monitor trolley show signs of mechanical damage

Ranked causes

  1. 1. Physical damage to cover panels (unknown)
First check

Remove and inspect cover panels from stand and monitor trolley

Test point
All cover panels
Expected
No cracks, breaks, or deformation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Emergency stop switch not functioning patient safety

Emergency stop switch does not prevent lifting column movement when activated

Ranked causes

  1. 1. Emergency stop switch electrical failure (rare)
First check

Activate emergency stop switch

Test point
Emergency stop switch on main unit control panel
Expected
Lifting column does not move when raise/lower buttons pressed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exterior housing damage caution

Visual inspection reveals damage to housing or paint of the system

Ranked causes

  1. 1. Physical impact or wear to housing or paint (unknown)
First check

Visually inspect entire system exterior

Test point
Housing and painted surfaces
Expected
No visible damage, dents, or paint degradation

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fluoroscopy timer function abnormal radiation

Fluoroscopy timer does not properly limit exposure time or warning signal does not activate

Ranked causes

  1. 1. Timer circuit failure or software misconfiguration (occasional)
First check

Verify fluoroscopy timer settings in system configuration

Test point
Blocktime settings in Configuration menu
Expected
Timer settings match national regulations

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Foot brake braking effect insufficient patient safety

Foot brake on stand or monitor cart does not provide adequate braking on flat surface

Ranked causes

  1. 1. Brake pad wear or contamination (common)
First check

Test foot brake holding on flat surface

Test point
Foot brake pedal on stand and monitor cart
Expected
Equipment does not move when brake engaged

Foot brake ineffective patient safety

Foot brakes on ARCADIS Orbic stand and trolley do not provide adequate braking on flat surface

Ranked causes

  1. 1. Brake wear or malfunction (unknown)
First check

Test foot brake function

Test point
Foot brake on stand and trolley
Expected
Equipment remains stationary on flat surface when brakes applied

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Horizontal lift brake braking force insufficient patient safety

Horizontal lift brake does not reach required braking value when engaged

Ranked causes

  1. 1. Brake spring or mechanism wear (occasional)
First check

Measure horizontal lift brake braking force with spring scale

Test point
Horizontal lift brake
Expected
Braking value as specified in Replacement of Parts manual for ARCADIS Orbic

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Horizontal lift brake insufficient braking force caution

Horizontal lift brake does not meet specified braking value when engaged

Ranked causes

  1. 1. Brake spring loss or wear (unknown)
First check

Measure horizontal lift brake force with spring scale

Test point
Horizontal lift brake
Expected
Braking value per ARCADIS Orbic Replacement of Parts Manual
Tools
Spring scale

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

I.I. laser light localizer mechanical integrity

Integrated or stand-alone I.I. laser light localizer shows mechanical damage or improper seating

Ranked causes

  1. 1. Physical damage to laser light localizer assembly (unknown)
First check

Visually inspect laser light localizer for mechanical damage

Test point
I.I. laser light localizer assembly
Expected
No visible cracks, breaks, or deformation

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

I.I. laser light localizer mounting and locking

Mounted I.I. laser light localizer is loose, improperly seated, or tension band is not properly closed

Ranked causes

  1. 1. Improper mounting or locking of laser light localizer (unknown)
First check

Mount laser light localizer on I.I. assembly

Test point
Localizer mounting interface
Expected
Localizer fully seated with no gaps

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ID labels missing or illegible

Required ID labels are missing, damaged, or difficult to read

Ranked causes

  1. 1. Label fading, peeling, or loss (common)
First check

Verify all required ID labels present and legible

Test point
ID labels on stand and monitor trolley
Expected
All required labels affixed and clearly readable

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image disk transfer warning radiation

Warning during image disk transfer operation

Ranked causes

  1. 1. Unknown disk transfer issue (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image distorted radiation

Calibration produces distorted image output

Ranked causes

  1. 1. Lead balls missing in calibration phantom (occasional)
First check

Inspect calibration phantom for lead balls

Test point
calibration phantom
Expected
all lead balls present

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Image quality quick test execution required

Routine image quality testing with optional external monitor and 3D calibration checks

Ranked causes

  1. 1. Routine maintenance requirement (common)
First check

Perform image quality quick test

Test point
System imaging and display
Expected
Results per ARCADIS Orbic image quality quick test specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Imaging PC BIOS battery requires replacement

BIOS battery in Imaging PC needs preventive replacement at scheduled intervals

Ranked causes

  1. 1. Scheduled preventive maintenance interval reached (common)
First check

Determine Imaging PC model type to identify correct replacement manual

Test point
PC type designation
Expected
Gen2 (M450, M460, M470) or Gen1 (M420, M430, M450)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Iris collimator aperture monitor display misalignment patient safety

Monitor display of iris collimator aperture diameter and position does not match actual physical iris collimator position

Ranked causes

  1. 1. Iris collimator position sensor or display calibration error (occasional)
First check

Verify iris collimator display in I.I. full format with collimator closed

Test point
Monitor display and iris collimator (X-iris)
Expected
Displayed diameter and position must match actual iris collimator aperture when fluoroscopy released

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Iris collimator blade visibility incorrect radiation

Collimator blades not clearly visible at image edges in full format or zoom modes

Ranked causes

  1. 1. Collimator blade misalignment (occasional)
First check

Select I.I. full format and check collimator blade visibility

Test point
Image intensifier and collimator blades
Expected
Collimator blades clearly visible at edges of image

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Lifting column lubrication required

Lifting column spindle and slide cylinders require lubrication maintenance

Ranked causes

  1. 1. Normal wear, routine maintenance (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Lifting column visible damage caution

Visual damage detected on exterior or interior of lifting column

Ranked causes

  1. 1. Impact or wear damage to column structure (occasional)
First check

Visual inspection of lifting column

Test point
Basic unit lifting column exterior and interior surfaces
Expected
No cracks, dents, or visible damage

Local printer SMPTE test image quality unacceptable

5% and 95% fields on printed SMPTE test image not discernible; brightness and contrast require adjustment

Ranked causes

  1. 1. Printer brightness/contrast miscalibrated (common)
First check

Print SMPTE test image and verify field visibility

Test point
Printed SMPTE test image
Expected
5% and 95% fields discernible

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Markers not detectable by navigation system radiation

Markers on the image intensifier, test body, or pointer are not visible to the navigation system during 3D exposure and accuracy check.

Ranked causes

  1. 1. Markers blocked or obscured during 3D exposure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor-controlled orbital belt tension check required caution

Routine inspection of toothed belt for motor-controlled orbital movement

Ranked causes

  1. 1. Routine maintenance requirement (common)
First check

Inspect motor-controlled orbital belt tension

Test point
Toothed belt for motor-controlled orbital movement
Expected
Belt tension within specification and no visible wear

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motor-controlled orbital movement starts during emergency stop patient safety

3D scan initiated while emergency stop switch is pressed causes motor-controlled orbital movement to start, which must not occur

Ranked causes

  1. 1. Emergency stop switch malfunction or improper engagement (unknown)
First check

Verify emergency stop switch halts orbital movement

Test point
Emergency stop switch
Expected
Motor-controlled orbital movement must not start during 3D scan with emergency stop pressed

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Motorized monitor trolley lifting column visible damage caution

Visual damage detected on motorized monitor trolley lifting column

Ranked causes

  1. 1. Impact or wear damage to column (occasional)
First check

Visual inspection of motorized monitor trolley lifting column

Test point
Monitor trolley lifting column exterior and interior
Expected
No cracks, dents, or visible damage

No registration possible during calibration with marker ring radiation

Marker ring registration fails to initiate

Ranked causes

  1. 1. Markers not detected in starting position (occasional)
First check

Verify marker visibility and position

Test point
marker ring starting position
Expected
markers detected by system

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Orbital brake braking force insufficient patient safety

Orbital brake does not reach required braking value when engaged

Ranked causes

  1. 1. Brake spring or mechanism wear (occasional)
First check

Measure orbital brake braking force with spring scale

Test point
Orbital brake
Expected
Braking value as specified in Replacement of Parts manual for ARCADIS Orbic

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Orbital brake insufficient braking force caution

Orbital brake does not meet specified braking value when engaged

Ranked causes

  1. 1. Brake spring loss or wear (unknown)
First check

Measure orbital brake force with spring scale

Test point
Orbital brake
Expected
Braking value per ARCADIS Orbic Replacement of Parts Manual
Tools
Spring scale

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Poor image quality after calibration radiation

Image quality remains poor following calibration completion

Ranked causes

  1. 1. Calibration process deficient (occasional)
First check

Verify 3D limiting resolution

Test point
3D calibration parameters
Expected
within specification

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Radiation indicator not illuminating
  • SIEMENS logo not securely affixed or damaged
  • Slot diaphragm position monitor display misalignment
  • Swing brake braking force insufficient
  • Swing brake insufficient braking force
  • System cables or plugs damaged
  • System reboots multiple times daily
  • System requires second startup after power cycle
  • TFT monitor damage or improper attachment
  • UPS battery requires replacement
  • Warning labels missing or illegible
  • Wheels or wheel mounts damaged or misaligned

12 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
ASP_BIC 143Error from disk fullDisk storage capacity exceedederror
ASP_DIS 34Acquisition image chain hardware fatal errorGenesis board hardware failure during acquisitioncritical
ASP_DIS 40Image disk transfer warningError during image disk transferwarning
ASP_DIS 51Acquisition image chain software warning during framegrabbingError during image framegrabbingwarning
ASP_DIS 54Acquisition image chain software warningSoftware issue in image chainwarning
CAP_PR_32Shutdown cancelledSystem shutdown sequence was cancelledinfo

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI