Shimadzu MUX100H: troubleshooting and error codes
The ServiceForge knowledge base holds 9 documented symptoms for the Shimadzu MUX100H, with 9 ranked causes, 18 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Abnormal activity due to NEX-SH
System exhibits abnormal behavior related to NEX-SH component
Ranked causes
- 1. NEX-SH PCB malfunction or failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
APR setting does not function as configured
APR parameter set to 0 but behaves as if set to 1
Ranked causes
- 1. Software bug in version prior to V1.52 (common)
Verify APR setting responds correctly to configuration
- Test point
- APR parameter setting
- Expected
- APR operates at configured value (0 or 1, as set)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not charge when unit is plugged in
Unit fails to initiate charging after power connection
Ranked causes
- 1. Software version prior to V1.52 (common)
Verify software version is V1.52 or higher
- Test point
- System firmware
- Expected
- V1.52 or later
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage indicator inaccurate
Battery indicator display does not correspond to actual battery voltage
Ranked causes
- 1. Software version prior to V1.52 (common)
Recalibrate battery voltage detection
- Test point
- Battery voltage detection calibration
- Expected
- As specified in Appendix B of Mux-100H installation manual M503-E312
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor power cuts off prematurely on startup
CPU cuts power to motor too quickly after power-on when Q6, Q7, Q8, Q9 on Mu-Driver 2 pcb are shorted
Ranked causes
- 1. Q6, Q7, Q8, Q9 shorted on Mu-Driver 2 pcb (rare)
Measure motor runtime after power-on
- Test point
- Motor operation at startup
- Expected
- Motor runs continuously during normal operation seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NEX-SH PCB replacement procedure for MUX-100H and Evolution models
Complete procedure for exchanging NEX-SH PCB on affected systems including preparation, exchange, initialization, date/time adjustment and data restoration
Ranked causes
- 1. Service bulletin XR-201112 requirement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remove pulley and shaft assembly from column caution
Procedure to safely remove pulley, shaft, collar and bearing assembly from upper part of column during roller support replacement.
Ranked causes
- 1. Required disassembly step for roller support replacement (unknown)
Undo press screw fixing collar
- Test point
- Collar press screw
- Expected
- Screw removed, collar loose
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace roller supports in column assembly caution
Roller supports in the Mux-100(H) column assembly require replacement. Procedure involves tilting column down, removing weight and pulley assembly, detaching old roller supports, and installing new upper and lower roller support assemblies with proper alignment and cushioning.
Ranked causes
- 1. Worn or damaged roller supports requiring replacement (unknown)
Verify shaft can be smoothly inserted into bearing of pulley before reassembly
- Test point
- Pulley bearing
- Expected
- Shaft inserts smoothly without resistance
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray tube anode continues rotation abnormally patient safety
Anode fails to stop rotating after X-ray exposure; insulating oil temperature rises, causing potential oil leakage from tube assembly and risk of operator or patient burns
Ranked causes
- 1. Abnormal NEX-SH PCB (common)
Record current DIP switch and jumper settings on installed NEX-SH PCB
- Test point
- NEX-SH PCB
- Expected
- All switches and jumpers documented
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI