Shimadzu MUX100D: troubleshooting and error codes
The ServiceForge knowledge base holds 4 documented symptoms for the Shimadzu MUX100D, with 4 ranked causes, 18 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Battery code requires update for Taiwan-made battery
When using new Taiwan-made PX-12090 battery, system battery code needs to be set to code 5
Ranked causes
- 1. Taiwan-made battery installed with PROM V1.13 (common)
Identify battery country of origin
- Test point
- Battery label on lower side
- Expected
- MADE IN TAIWAN
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exit long exposure time mode radiation
User needs to disable 3-second exposure capability and return to standard mode
Ranked causes
- 1. Long exposure mode enabled on current APR selection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Increase Canon CXDI-50G exposure time to 3 seconds radiation
User requests capability to extend maximum exposure time from default 1 sec to 3 sec on Mux-100D with Canon CXDI-50G detector
Ranked causes
- 1. Software version prior to v1.12 limits exposure to 1 second (common)
Back up X-ray tube calibration data
- Test point
- FVR data storage
- Expected
- FVR data saved before EEPROM replacement
13 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube voltage waveform does not reach set value at low battery high voltage
When battery level lowers to one indicator green lamp, tube voltage waveform may not reach the set value during operation
Ranked causes
- 1. PROM version earlier than V1.13 with Taiwan-made battery (common)
Verify PROM version on unit
- Test point
- System display or CPU board identification
- Expected
- PROM version 1.13 or later
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI