Shimadzu MUX-100: troubleshooting and error codes
The ServiceForge knowledge base holds 132 documented symptoms for the Shimadzu MUX-100, with 214 ranked causes, 327 diagnostic checks with test points and expected values, and 82 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
X-ray tube rotor not rotating (F61) radiation
Starter error; starter does not issue current detection confirmation signal; rotor does not rotate due to absent or low stator coil current
Ranked causes
- 1. Fuse F3 blown (common)
- 2. Stator coil current low or absent (common)
- 3. Low voltage cable poorly contacted (occasional)
- 4. INVERTER UNIT abnormal (occasional)
- 5. MUX CHARGE abnormal (occasional)
- 6. X-ray tube coil open or high resistance (occasional)
- 7. SW1A not set to ON (occasional)
- 8. X CONT or NEXSC abnormal (occasional)
Check fuse F3 on MUX POWER
- Test point
- Fuse F3
- Expected
- Not blown
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal motor movement patient safety
Arm or motor exhibits abnormal motion during operation, either unable to drive, restricted high-speed driving capability, or intermittent brake function.
Ranked causes
- 1. Arm catcher mechanism misalignment (common)
- 2. Motor malfunction (common)
- 3. FSR (Field Speed Regulator) malfunction (common)
- 4. Inch moving switch failure (occasional)
- 5. Operation panel light failure (occasional)
- 6. Photo sensor malfunction (occasional)
- 7. Brake intermittent function (occasional)
Verify driving is possible
- Test point
- Motor control circuit
- Expected
- Unit responds to drive command
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
F12 (KV UNDER) - tube voltage low high voltage
Measured tube voltage is 50% or more below the preset value, detected by software at 20ms after exposure and at each 100ms interval
Ranked causes
- 1. Glow discharge occurred in X-ray tube (common)
- 2. High voltage transformer defective (occasional)
- 3. High voltage cable defective (occasional)
- 4. X CONT abnormal (occasional)
- 5. NEXSC abnormal (occasional)
- 6. INVERTER UNIT abnormal (occasional)
Check CP9 (TKV) and CP12 (RKV) on X CONT
- Test point
- CP9 (TKV), CP12 (RKV)
- Expected
- Normal voltage
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
F12 (KV UNDER) error: tube voltage low high voltage
Measured tube voltage is 50% or more below preset value, detected by software at 20ms after exposure and every 100ms
Ranked causes
- 1. Glow discharge in X-ray tube (common)
- 3. High voltage cable defective (occasional)
- 4. X CONT (X-ray control unit) abnormal (occasional)
- 5. NEXSC unit abnormal (occasional)
- 7. FVR adjustment out of range (occasional)
- 8. KVFeedback abnormal (occasional)
Measure or monitor tube voltage at control points
- Test point
- CP9 (TKV) and CP12 (RKV) on X CONT
- Expected
- Normal within specification
3 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
F23 NO MA zero or very low tube current high voltage
Measured tube current is approximately 10 mA or less, detected by software at 20ms after exposure and each 100ms. Filament may be broken or high voltage path interrupted.
Ranked causes
- 1. Short-circuit bar missing or high voltage path open (common)
- 2. High voltage cable not conductive or broken (common)
- 3. Cable JFB between transformer and X CONT broken or loose (occasional)
- 4. X-ray tube filament broken (rare)
- 5. High voltage transformer filament winding or current sensing circuit fault (occasional)
- 6. Fuse F3 blown or short circuit in filament path (occasional)
Verify short-circuit bar on terminal block X1
- Test point
- Terminal block X1 on X CONT (mA+ and mA- terminals)
- Expected
- Short-circuit bar present and secure
- Tools
- Visual inspection
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Filament current low or absent (F33) radiation
Measured filament current approximately 1 A or less, detected every 800 ms during prep or 400 ms when ready
Ranked causes
- 1. High voltage cable poorly contacted (common)
- 2. INVERTER UNIT abnormal (occasional)
- 3. X CONT abnormal (occasional)
- 4. X-ray tube filament broken (occasional)
- 5. High voltage transformer defective (rare)
Verify C0 and CL waveforms normal
- Test point
- X CONT display
- Expected
- Normal waveforms
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage out of range (F51, F52, F54, F55, F56, F5b) high voltage
Battery voltage abnormal during charging: too high (>155V), too low (<100V), or imbalanced between plus and minus sides (>15% difference)
Ranked causes
- 1. Battery voltage out of specification (common)
- 2. MUX CHARGE unit abnormal (occasional)
- 3. Voltage detection circuit out of adjustment (occasional)
- 4. X CONT 2000 abnormal (occasional)
Check battery voltage at TER1-TER5
- Test point
- Battery terminals TER1-TER5
- Expected
- 100-155V, -55V to 0V (min 100V)
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
d1 error - Motor drawing excessive current
Drive motor exceeds 6A for more than 10 seconds; warning prevents Mobile Art travel
Ranked causes
- 1. DC motor brush wear with residual powder shorting rotor (common)
- 2. Left motor defective and drawing excessive current (common)
- 3. Right motor defective and drawing excessive current (common)
- 4. MU Driver PCB defective (occasional)
Measure left motor current draw at CP9 on MU Driver PCB
- Test point
- CP9
- Expected
- Less than 3.5V (less than 6A), below threshold
- Tools
- DMM
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
d1 warning error displayed
Warning error message d1 appears briefly on panel display and is logged in system error log. Indicates one of the DC drive motors is drawing excessive current.
Ranked causes
- 1. DC motor brush wear and powder accumulation on positive electrode (common)
- 2. Left drive motor failure (occasional)
- 3. Right drive motor failure (occasional)
- 4. MU Driver PCB failure (rare)
Measure left motor current draw at test point
- Test point
- CP9 on MU Driver PCB
- Expected
- Less than 3.5V (corresponds to less than 6A current draw); 2.5VDC offset, every 0.1V = 0.67A
- Tools
- DMM
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
F01/F02 supply voltage out of range high voltage
Power supply voltage is abnormal, either approximately 20% high (F01) or 30% low (F02), detected by software.
Ranked causes
- 1. Input voltage exceeds tolerance (F01) or falls below tolerance (F02) (common)
- 2. Transformer T1 primary winding open or power line fault (occasional)
- 3. MUX CHARGE 99 board malfunction (occasional)
- 4. X CONT or NEXSC board malfunction (occasional)
Measure CP23 (LV) on X CONT board
- Test point
- CP23 on X CONT-2000
- Expected
- Within range 2V+20%~-30% (nominal ~2V), +20% to -30%
- Tools
- Multimeter
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noise during unit movement and driving
Noise created during unit movement along the column, arm forward/backward movement, and arm lock operation
Ranked causes
- 1. Worn or misaligned brake for column up/down movement (common)
- 2. Noise in upper column section (occasional)
- 3. Noise in arm forward/backward movement (occasional)
- 4. Noise in arm lock section (occasional)
Remove arm cover
- Test point
- Arm cover fasteners
- Expected
- Cover removed cleanly
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit generates excessive driving noise during movement
Noise occurs when the unit moves (column up/down, arm forward/backward, or arm lock operation)
Ranked causes
- 1. Brake for column up/down movement requires modification (common)
- 2. Upper column section requires noise countermeasures (common)
- 3. Arm forward/backward movement section requires noise countermeasures (common)
- 4. Arm lock section requires noise countermeasures (common)
Remove arm cover
- Test point
- Arm assembly
- Expected
- Cover removed without damage
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charging behavior differs by EMC standard compliance
System charge control behavior varies based on EMC Standard edition (First vs Second) and software version, affecting when battery charging is possible relative to Key Switch position
Ranked causes
- 1. Software version mismatch with EMC standard compliance level (common)
- 2. NEXSC DIPSW2-6 setting incorrect for installed software version (occasional)
- 3. Battery charge prevented due to Key Switch position incompatibility (common)
Determine installed software version
- Test point
- System firmware version display
- Expected
- V1.50 (Second Edition only), V1.52 or later (both editions), or earlier than V1.50 (First Edition only)
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error F55 displayed during battery charging
Unit displays error code F55 indicating battery charged voltage of + side is more than 12V higher than – side; error appears during charging and when charging stops
Ranked causes
- 1. One battery block has significantly higher voltage than others (common)
- 2. Defective battery with excessive voltage drop (common)
- 3. Software version incompatibility (occasional)
Measure charged voltage of battery block BAT1
- Test point
- BAT1
- Expected
- Voltage approximately equal to BAT2, BAT3, BAT4
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exposure control circuit abnormal (F6F) radiation
KVT (kV rising signal) not returned within 250 ms after EXR sent; phototiming shutdown signal IPOF detected simultaneously with EXR, or setting error if phototimer not provided
Ranked causes
- 1. Phototiming conflict during exposure (occasional)
- 2. Incorrect initial settings without phototimer (occasional)
- 3. X CONT-2000 PCB abnormal (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
F12 KV under voltage radiation
Measured tube voltage is lower by 50% or more than preset value, detected by software at 20ms after exposure and each 100ms. Glow discharge likely in X-ray tube.
Ranked causes
- 1. X-ray tube glow discharge or filament circuit fault (common)
- 2. X-ray tube filament broken (rare)
- 3. High voltage transformer or cable defective (occasional)
Measure CP9 (TKV) and CP12 (RKV) on X CONT
- Test point
- Test points CP9 (TKV) and CP12 (RKV) on X CONT
- Expected
- Both normal or RKV abnormal
- Tools
- Multimeter
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
F13 NO KV zero or very low voltage high voltage
Measured tube voltage is approximately 10 kV or less, detected by software at 20ms after exposure and each 100ms.
Ranked causes
- 1. INVERTER UNIT fuse blown or IGBT damaged (common)
- 2. Tube voltage feedback or control circuit fault (occasional)
- 3. High voltage transformer defective (rare)
Verify fuse on INVERTER UNIT
- Test point
- Fuse location on INVERTER UNIT
- Expected
- Fuse intact and conductive
- Tools
- Visual inspection, continuity meter
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Switch stuck ON at power-on (D4-D11) patient safety
Any control switch detected as ON during power-up; D4 brake release, D5-D6 small movement forward/backward, D7-D8 drive handle left/right, D9 bumper, D10-D11 arm lock release switch and lever
Ranked causes
- 1. Switch stuck or jammed (common)
- 2. Electrical connection fault (common)
- 3. MUX DRIVE logic fault (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System veers left or right, difficult to drive straight patient safety
Unit drifts consistently in one direction despite operator input; no error messages displayed; symptom occurs with either open F4 or F5 fuses
Ranked causes
- 1. Fuse F4 (+120V) open (common)
- 2. Fuse F5 (-120V) open (common)
- 3. Shorted FETs Q6, Q7, Q8, Q9 on MU Driver board (occasional)
Measure resistance across source and drain of FETs Q6, Q7, Q8, Q9
- Test point
- Q6, Q7, Q8, Q9 on MU Driver board
- Expected
- ~0 ohms indicates short
- Tools
- Ohmmeter
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal encoder detected (D3) caution
Encoder signal abnormal or not detected
Ranked causes
- 1. MUX DRIVE unit abnormal (common)
- 2. Motor or encoder faulty (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arm lock and release not functioning caution
Arm cannot be easily locked or released from parked position, or drive motors not in high speed during parking
Ranked causes
- 1. Lock release button malfunction (unknown)
- 2. Drive motor speed not at high speed during parking (unknown)
Verify arm can be easily locked from parked position
- Test point
- lock release buttons
- Expected
- Arm locks smoothly
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery indicator shows 1-3 LEDs despite full charge
Battery indicator light displays 1-3 LEDs even though batteries are fully charged; condition persists after power recycling; all 5 LEDs light when charging is attempted again
Ranked causes
- 1. Power cable issue or charging connection problem (common)
- 2. Software version incompatibility (common)
Verify power cable is properly connected
- Test point
- Power cable connection
- Expected
- Cable properly seated and making contact
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
d3 error - Abnormal motor encoder detected
Encoder signal abnormal or absent for one or both motors
Ranked causes
- 1. Loose or damaged encoder cable connections (common)
- 2. Motor encoder defective (occasional)
Verify motor encoder cable connections on MU Driver PCB
- Test point
- Encoder connectors on MU Driver PCB
- Expected
- Cables fully seated and undamaged
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
d3 warning error displayed
Warning error message d3 appears on panel display. Indicates abnormal encoder detected for one of the drive motors.
Ranked causes
- 1. Defective motor encoder (common)
- 2. Encoder cable connection fault (occasional)
Verify encoder motor speeds in Adjustment Mode diagnostics
- Test point
- Motor speed check in Adjustment Mode
- Expected
- Left wheel speed and right wheel speed should be relatively close: around 1km/hr in slow speed and 5km/hr in high speed
- Tools
- Remote control, Adjustment Mode tool
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Determine MA setting for KV and MAS selection radiation
MA (milliamperage) is automatically determined by software based on operator-selected KV and MAS values and battery charge state. System operates in normal mode when green LEDs are lit without red LED, or low battery mode when red LED and one green LED are lit.
Ranked causes
- 1. Normal battery mode operation (common)
- 2. Low battery mode operation (occasional)
Verify battery status by checking control panel indicator LEDs
- Test point
- Battery indicator on control panel
- Expected
- Green LEDs lit with red LED off for normal mode, or red LED and one green LED lit for low battery mode
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
F11 KV over voltage high voltage
Measured tube voltage is higher by approximately 30 kV or more than preset value, detected by software at 20ms after exposure and each 100ms.
Ranked causes
- 1. Tube voltage feedback circuit malfunction (common)
- 2. X CONT, X-ray tube, or high voltage transformer defective (occasional)
Measure CP9 (TKV), CP12 (RKV), CP11 (TMA), CP13 (RMA) on X CONT
- Test point
- Test points CP9, CP12, CP11, CP13 on X CONT board
- Expected
- RKV and RMA normal feedback; TKV and TMA as programmed
- Tools
- Multimeter or oscilloscope
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
F14 KV brake abnormal high voltage system discharge radiation
Abnormal measured tube voltage or X-ray tube discharge detected by hardware. Error may be eliminated by aging; discharge in high voltage system suspected.
Ranked causes
- 1. X-ray tube discharge, high voltage cable, transformer, or control circuit fault (common)
- 2. False error message or low-probability PCB defect (rare)
Perform aging on X-ray tube
- Test point
- X-ray tube high voltage circuit
- Expected
- Aging completes without excessive discharge sounds or error recurrence
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
F31 IF over filament current high high voltage
Measured filament current is higher by 50% or more than preset value, detected by software at each 800ms during prep and 400ms during ready.
Ranked causes
- 1. Filament heating control or sensing circuit malfunction (common)
- 2. INVERTER UNIT or high voltage filament transformer fault (occasional)
Measure FVR (CP15) and IFV (CP7) on X CONT
- Test point
- Test points CP15 (FVR) and CP7 (IFV) on X CONT
- Expected
- Both normal, or only FVR abnormal
- Tools
- Multimeter
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Force Sensing Resistor (FSR) response degraded or absent
Handle does not respond predictably to pressure; driving characteristics inconsistent or no motor output when handle is pressed.
Ranked causes
- 1. FSR pads aged or degraded (common)
- 2. FSR cable connection loose or damaged (occasional)
Measure FSR input signal at right FSR test point with no pressure applied
- Test point
- CP11 on MU Driver PCB (right FSR)
- Expected
- 2.5VDC offset
- Tools
- DMM (DC voltage)
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse F4 or F5 blown repeatedly caution
Fast blow fuse opens immediately after power-on; slow blow fuse may overheat without opening initially; affects motor power supply
Ranked causes
- 1. Shorted FET on MU Driver board (common)
- 2. Short circuit in motor winding or motor power cables (occasional)
Measure FET source-drain resistance on MU Driver board
- Test point
- Q6, Q7, Q8, Q9 on MU Driver board
- Expected
- ~0 ohms indicates short; normal should be open circuit or high ohms
- Tools
- Ohmmeter
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
H3 X-ray tube assembly temperature high patient safety
Temperature of X-ray tube assembly is too high beyond safe operating threshold.
Ranked causes
- 1. X-ray tube thermal switch activated due to overheating (common)
- 2. Thermal switch signal fault (false alarm) (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Handle Force Sensing Resistor (FSR) not responding correctly
FSR pads fail to detect handle pressure or provide inconsistent input
Ranked causes
- 1. FSR pads aged or degraded (common)
- 2. FSR pads damaged or worn (occasional)
Measure right FSR signal at CP11 on MU Driver PCB
- Test point
- CP11
- Expected
- 2.5V offset with no pressure; normal range 1-4V with handle pressure
- Tools
- Voltmeter
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect motor output voltage for drive speed high voltage
Motor power output does not match expected voltage for selected drive speed direction
Ranked causes
- 1. Input voltage from MUX Power PCB incorrect (occasional)
- 2. MU Driver PCB not applying correct motor voltage (occasional)
Measure MUX Power PCB input voltage at CP15
- Test point
- CP15
- Expected
- 120V
- Tools
- Voltmeter
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Motor overcurrent detected (D1) caution
Motor current detected as 6A or greater for more than 10 seconds
Ranked causes
- 1. MUX DRIVE unit faulty (common)
- 2. Motor faulty (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Noises during travel caution
Unusual noises heard during unit movement
Ranked causes
- 1. Loose fastening hardware (common)
- 2. Worn or damaged cables (occasional)
Listen for unusual noises during travel
- Test point
- entire moving assembly
- Expected
- No unusual noises during extension, tilt, rotation, and pivot motion
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software version compatibility with EMC Standard editions
Specific software versions are required or restricted based on system EMC Standard compliance edition, with version-dependent DIPSW2-6 settings
Ranked causes
- 1. V1.50 software only compatible with EMC Second Edition systems (common)
- 2. V1.52 and later support both EMC Standard editions with proper DIPSW2-6 configuration (common)
Identify system EMC Standard edition
- Test point
- System specification or documentation
- Expected
- First Edition compliant or Second Edition compliant
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software version incompatibility with EMC standard compliance
Incorrect software version installed for system EMC standard edition
Ranked causes
- 1. V1.50 software on EMC First Edition compliant system (occasional)
- 2. Software version earlier than V1.50 on EMC Second Edition compliant system (occasional)
Verify software version and EMC standard compliance of system
- Test point
- System software version and EMC certification
- Expected
- For EMC First Edition: V1.50 or later (V1.52+ recommended). For EMC Second Edition: V1.50 or later with DIPSW2-6 OFF
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal sound or vibration during arm movement caution
Audible noise or unusual phenomenon when moving the gantry arm forward and backward
Ranked causes
- 1. Improper chain tension (common)
Listen for abnormal sound during arm forward and backward movement
- Test point
- Gantry arm
- Expected
- No abnormal sound or phenomenon
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arm not locked during retraction patient safety
X-ray tube unit may run off toward the front when arm section is retracted during normal moving
Ranked causes
- 1. Arm lock mechanism failed or not engaged (unknown)
Visually inspect arm lock status
- Test point
- Arm section during retraction
- Expected
- Arm is locked and X-ray tube unit does not move toward front
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Arm rotation exceeds ±180 degrees caution
Gantry arm cannot rotate properly around its axis, rotation range is insufficient
Ranked causes
- 1. Inappropriate cable wiring (common)
Verify arm rotation range around vertical axis
- Test point
- Gantry arm axis
- Expected
- Arm rotation of ±180 degrees
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charge current abnormal (F5a, F5E, F5F) high voltage
Charging current exceeds 3.8A, or measured current deviates >50% (F5E) or <30% (F5F) from specified value for >10 seconds
Ranked causes
- 1. Charge circuit problem (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery indicator light turned red after charging
Battery indicator light displays red after the battery has been charged
Ranked causes
- 1. Software version earlier than 1.53 (common)
Verify current software version
- Test point
- System display
- Expected
- Version 1.52 or earlier
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery indicator Red LED illuminates despite nearly full charge
Battery indicator Red LED lights on even though batteries are almost fully charged; occurs when battery charging has not completed within 10 hours
Ranked causes
- 1. Charging cycle did not complete within 10 hours (common)
Verify charging duration
- Test point
- Charging time
- Expected
- Charging completes within 10 hours
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery indicator shows insufficient charge despite full battery
Battery indicator displays only 4 green bars or fewer even when battery is fully charged
Ranked causes
- 1. Incorrect battery code set for old-type PXL12072 battery (common)
Identify battery type by serial number
- Test point
- Battery serial number stamp
- Expected
- 000703 or earlier for old type
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery level indicator inaccuracy
Battery level indicator does not accurately reflect actual battery voltage
Ranked causes
- 1. Outdated software V1.24 or earlier on NEXSC board (common)
Power OFF and remove right side cover
- Test point
- MUX-100 housing
- Expected
- Access to NEXSC board visible
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bumper touches floor improperly
Front bumper not touching floor or contact condition is incorrect
Ranked causes
- 1. Bumper pad worn or dirty (occasional)
Verify bumper touches floor
- Test point
- Front bumper contact point
- Expected
- Bumper touching the floor
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Buzzer does not sound during exposure
Alarm buzzer fails to activate when X-ray exposure occurs
Ranked causes
- 1. Buzzer alarm malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Arm retraction detector micro switch patient safety
Adjustment of micro switch position to ensure proper arm retraction detection and safety lock function
Ranked causes
- 1. Micro switch requires repositioning after locking hook replacement (common)
Place spacer washers and simulate cover thickness
- Test point
- Catch position frame
- Expected
- M6 flat washers placed; top-plate positioned on washers
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Arm rotation torque and x-ray tube alignment high voltage
Post-assembly adjustment of FU nut torque and x-ray tube tilt to ensure proper mechanical function and beam alignment
Ranked causes
- 1. Assembly completion requires torque and alignment verification (common)
Measure torque at collimator handle after FU nut adjustment
- Test point
- Collimator handle front end during arm rotation
- Expected
- 5 to 7 kg, 5-7 kg
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Brake plate replacement for noise suppression caution
Installation of noise-suppressing brake plate (t=3.2) in the electromagnetic brake assembly for up/down column movement
Ranked causes
- 1. Standard noise reduction modification (unknown)
Obtain required kit and tools
- Test point
- Kit 503-52206 contents, tool inventory
- Expected
- Kit 503-52206 with plate 503-52158, Phillips screwdriver, L-wrench M5, Lock-tight #242, small tie-wrap available
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cable and connector inspection high voltage
Inspection and lubrication of cables for damage, wear, and proper terminal condition on HV cable
Ranked causes
- 1. Cable damage, excessive wear, or dry terminals (unknown)
Inspect all cables visually
- Test point
- Cable insulation and routing
- Expected
- No damage, cracks, cuts, or excessive wear visible
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Drive handle and brake release operation caution
Verification that drive handle operation and brake release correspond correctly to unit motion
Ranked causes
- 1. Drive handle or brake control misalignment or wear (unknown)
Operate drive handle through full range
- Test point
- Drive handle control and motor response
- Expected
- Unit motion corresponds to handle position
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Drive handle fastening hardware verification caution
Verification that drive handle fastening hardware is properly secured
Ranked causes
- 1. Loose fastening hardware on drive handle (unknown)
Inspect drive handle fastening hardware
- Test point
- All drive handle mounting fasteners
- Expected
- All hardware firmly secured and tightened
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Drive motor and lock operation caution
Verification that arm lock and release functions work properly and drive motors operate at correct speed when tube arm is parked
Ranked causes
- 1. Lock mechanism wear or motor speed degradation (unknown)
Lock arm from parked position
- Test point
- Arm lock mechanism
- Expected
- Arm locks easily and securely
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Emergency stop switch operation patient safety
Verification that emergency stop button illuminates when pressed and all driving capability is suspended
Ranked causes
- 1. Emergency stop switch failure or electrical fault (unknown)
Verify button illumination
- Test point
- Emergency stop button indicator lamp
- Expected
- Button illuminates when pressed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Fast blow fuse upgrade caution
Procedure to change drive motor fuses from slow blow to fast blow type for improved protection against shorted FET damage
Ranked causes
- 1. Upgrade fuses for enhanced FET protection (common)
Identify and remove fuse F4 on MUX Power board
- Test point
- Fuse F4 location (+120V supply)
- Expected
- Current fuse is slow blow type 072-01659-83 (10A 250V)
- Tools
- Field Service Tools
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: FET protection with RTV rubber caution
Preventive application of RTV silicone rubber to FET circuit terminals during periodic maintenance to prevent short circuits from dust contamination
Ranked causes
- 1. Preventive maintenance procedure (common)
Inspect FET circuit area for contamination
- Test point
- Area around Q6, Q7, Q8, Q9 terminals
- Expected
- Clean, dust-free surface with no foreign matter
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Front bumper alert and motion stop patient safety
Verification that front bumper contact triggers audible alert and stops forward unit motion
Ranked causes
- 1. Bumper switch or alert circuit failure (unknown)
Engage front bumper
- Test point
- Front bumper switch
- Expected
- Audible alert sounds
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Handle adjustment
Adjust handle calibration per adjustment mode procedure
Ranked causes
- 1. Handle requires calibration (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Handle Force Sensing Resistor
Adjust FSR response to ensure consistent driving response as pads age or after replacement.
Ranked causes
- 1. FSR pad aging affects response consistency (common)
Enter Adjustment Mode and access Handle Calibration
- Test point
- Adjustment Mode menu
- Expected
- Handle Calibration option accessible
- Tools
- Remote control, MUX-100 system
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Input supply voltage setting
- Calibration: Lock release button operation
- Calibration: Mechanical motion verification
- Calibration: Motor replacement (old to new style)
- Calibration: MUX-100 exposure linearity
- Calibration: Waveform Adjustment
- Calibration: X-ray collimator field alignment
- Calibration: X-ray tube and collimator alignment
- Calibration: X-ray tube and fastening hardware verification
- Charging cable does not illuminate charging meter
- Charging cable does not retract and lock
- Charging circuit does not work normally with Key Switch ON
- Charging does not start with power plug connected
- Column ASSY roller base plate damaged
- Column leaning or misalignment
- Counterweight fixation before tube removal
- Counterweight not fixed before tube removal
- Drive handle or brake release operation mismatched to motion
- E battery backup life low
- EEE battery backup life expired
- Emergency circuit triggered (F6d)
- Emergency discharge recovery charging behavior change
- Emergency stop circuit malfunction
- Emergency stop switch not functional
- Emergency stop switch not functioning
- Error code F5b occurs
- Error code F5b triggered by battery behavior
- Excessive noise during unit movement and driving
- Exposure buzzer does not sound during X-ray exposure
- Exposure cutoff circuit abnormal (F62)
- Exposure did not terminate (F6E)
- F03 control circuit power supply abnormal
- F12 code troubleshooting flowchart
- F21 MA over current
- F22 MA under current
- F32 IF under filament current low
- Film box frame sagging
- Front bumper audible alert not functioning
- Front bumper misalignment
- H1 anode heat over tolerance
- H2 anode heat at tolerance limit
- High voltage cable and X-ray tube connection verification
- High voltage generated with disconnected cable or tube
- IGBT abnormal gate waveform
- IGBT gate waveform observed in charged status
- Improper arm rotation torque
- Incorrect MA displayed or applied during exposure
- Inspection: Arm catch alignment verification
- Inspection: Collimator alignment verification
- Inspection: Front bumper alignment verification
- Inspection: Unusual noises during travel
- Mobile Art mobility characteristics altered
- Motor abnormality not detected
- Motor power supply voltage incorrect or absent
- Motor travel does not respond to handle input or is unpredictable
- Operator receives static shock from unit
- Photo sensor wires disconnected on power assist handle
- Photo-sensor or FSR wiring breakage on motorized handle
- PROM memory data loss after firmware update
- Radiographic exposure linearity correction needed
- Short circuit between FET pins on MU DRIVER board
- Sliding arm cannot support X-ray tube
- Sliding arm support degradation
- Stopper Pin drops from Arm unit
- Tube arm does not park correctly
- Tube voltage drops with new-type PXL12072 battery installed
- Wheels and hubcaps not secure
- Wire rope corrosion or damage
- X-ray tube arm releases from latch and moves upward
- X-ray tube fastening hardware loose
- X-ray tube support arm deterioration
- X-ray tube tilting at vertical detent position
72 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
d0 | Motor abnormality detected | Motor malfunction detected during operation | error |
d0 | Drive motor error detected | Motor feedback irregularity detected; error appears briefly on panel display and is logged in system error log | warning |
d1 | DC drive motor excessive current draw | One of the DC drive motors draws more than 6A for more than 10 seconds | warning |
D10 | Arm lock release switch stuck ON at power-up | Arm lock release switch detected ON during startup | warning |
D11 | Arm lock release lever stuck ON at power-up | Arm lock release lever detected ON during system startup | warning |
d2 | Drive motor fault variant | Motor feedback sampling detects irregularity | warning |
d3 | Abnormal encoder detected for one drive motor | Encoder signal absent, abnormal, or out of specification detected for left or right motor | warning |
D4 | Brake release switch stuck ON at power-up | Brake release switch detected ON during system startup | warning |
D5 | Small movement forward switch stuck ON at power-up | Forward small-movement switch detected ON during startup | warning |
D6 | Small movement backward switch stuck ON at power-up | Backward small-movement switch detected ON during startup | warning |
D7 | Drive handle left switch stuck ON at power-up | Left drive handle switch detected ON during startup | warning |
D8 | Drive handle right switch stuck ON at power-up | Right drive handle switch detected ON during startup | warning |
D9 | Bumper switch stuck ON at power-up | Bumper switch detected ON during system startup | warning |
E | Battery backup life low | CPU memory backup lithium battery voltage below minimum threshold | warning |
E | CPU memory backup battery needs replacement | Battery voltage drops below acceptable level for memory backup | error |
EEE | Battery backup life expired | CPU memory backup lithium battery voltage at or below end-of-life threshold | error |
EEE | Data erasure confirmation message displayed during boot | Appears on display approximately 3 seconds after power-on following EPROM upgrade | info |
F01 | Supply voltage high | Input supply voltage approximately 20% or more above nominal value, detected by software | error |
F01, F02 | Supply voltage out of normal range | Input mains voltage or system supply voltage outside specifications | error |
F02 | Supply voltage low | Input supply voltage approximately 30% or more below nominal value, detected by software | error |
F03 | -15V control circuit power abnormal (hardware detected) | Monitoring detects -15V supply fault on X CONT | error |
F03 | -15V check signal is not supplied | Missing -15V reference signal at startup | error |
F03 | Power supply -15V circuit fault | Hardware detection of -15V power supply abnormality | error |
F11 | KV is detected to be more than 30KV higher than the set value | Tube voltage exceeds setpoint by >30KV | error |
F11 | Tube voltage over limit (KV OVER) | Measured tube voltage is 30 kV or more above preset value at 20ms after exposure start and each subsequent 100ms | error |
F11, F12, F13, F14 | X-ray tube voltage control fault | High voltage control circuit malfunction or voltage regulation failure | error |
F12 | Tube voltage under limit (KV UNDER) | Measured tube voltage is 50% or less of preset value at 20ms after exposure and each 100ms; glow discharge likely in tube | error |
F12 | Measured kV below minimum threshold of kV setting | Detected when measured kV falls below threshold | error |
F12 | KV UNDER - tube voltage is lower by 50% or more than preset value | Detected by software at 20ms after exposure and at each 100ms interval | error |
F12 | KV UNDER – measured tube voltage is 50% or more below preset value | Detected by software at 20ms after exposure and at each 100ms interval | error |
F13 | KV is lower than 10KV | Tube voltage drops below 10KV minimum | critical |
F13 | No tube voltage (NO KV) | Measured tube voltage is 10 kV or less at 20ms after exposure and each 100ms | critical |
F14 | KVBR signal is supplied | KV breakdown signal detected | critical |
F14 | KV brake - high voltage discharge detected | Abnormal measured tube voltage or X-ray tube discharge detected by hardware | error |
F21 | Tube current over limit (MA OVER) | Measured tube current is 50 mA or more above preset value at 20ms after exposure and each 100ms | error |
F21 | mA is more than 50mA higher than the set value | Tube current exceeds setpoint by >50mA | warning |
F21, F22, F23 | X-ray tube current control fault | Tube current (mA) control circuit malfunction or feedback error | error |
F22 | mA is more than 50mA lower than the set value | Tube current falls below setpoint by >50mA | warning |
F22 | Tube current under limit (MA UNDER) | Measured tube current is 50% or less of preset value at 20ms after exposure and each 100ms | error |
F23 | No tube current (NO MA) | Measured tube current is 10 mA or less at 20ms after exposure and each 100ms | critical |
F23 | mA is lower than 10mA | Tube current drops below 10mA minimum | error |
F31 | Filament current over limit (IF OVER) | Measured filament current is 50% or more above preset value at 800ms intervals during prep, 400ms during ready | error |
F31 | Heater voltage is more than 50% higher than the set value | Filament heater voltage exceeds 50% above nominal | warning |
F31, F32, F33 | Filament heating control fault | Filament heater circuit malfunction or temperature control failure | error |
F32 | Heater voltage is more than 50% lower than the set value | Filament heater voltage drops >50% below nominal | warning |
F32 | Filament current under limit (IF UNDER) | Measured filament current is 50% or more below preset value at 800ms intervals during prep, 400ms during ready | error |
F33 | Heater voltage is lower than 1V | Filament heater voltage below minimum 1V | error |
F33 | No filament current or current below 1A | Filament current <1A detected every 800ms during prep or 400ms when ready | error |
F51 | Battery imbalance detected | Voltage imbalance between battery sides during operation | warning |
F51 | Battery voltage during charging >155V | Charging cycle voltage exceeds 155V | warning |
F51, F52, F53, F54, F55, F56, F57, F58, F59, F5A, F5B, F5C, | Charge control fault | High voltage charging or capacitor charge control circuit malfunction | error |
F52 | Battery imbalance detected | Voltage imbalance between battery sides during operation | warning |
F52 | Battery voltage during charging <100V | Charging cycle voltage falls below 100V | warning |
F53 | Battery imbalance detected | Voltage imbalance between battery sides during operation | warning |
F54 | Battery imbalance detected | Voltage imbalance between battery sides during operation | warning |
F54 | Secondary battery voltage during charging outside range | Secondary battery voltage <100V or >155V | warning |
F55 | Battery charged voltage imbalance: + side voltage more than 12V higher than – side | Detected during charging and when charging stops | error |
F56 | Minus-side battery voltage >15% higher than plus-side | Voltage imbalance: minus side exceeds plus side by >15% | warning |
F5a | Signal CHBR1 is supplied (Charging current is higher than 3.8A) | Battery charging current exceeds 3.8A threshold | warning |
F5b | Battery voltage exceeds ±155V | BRON signal supplied; battery voltage >+155V or <-155V | error |
F5b | Signal CHBR2 is supplied (Battery voltage on plus or minus side during charging is higher than 155V) | Battery voltage exceeds 155V during charging cycle | error |
F5b | Battery level indicator misalignment with actual battery voltage | Discrepancy between battery level indicator and actual battery voltage | error |
F5b | Battery behavior related error | Incorrect battery voltage interpretation by pre-V1.25 software | warning |
F5b | Battery-related error triggered during recharge after prolonged idle period | Attempting to recharge MUX-100 after extended unused period (e.g., post-installation) | warning |
F5C | Check main breaker | Main breaker fault detected | error |
F5E | Charging current >50% above specified value for >10s | Measured charging current deviates >50% high for ≥10 seconds | error |
F5F | Charging current <30% of specified value for >10s | Measured charging current deviates >70% low for ≥10 seconds | error |
F61 | Starter fault | X-ray tube starter (trigger) circuit malfunction | error |
F61 | Rotor starter error; rotor does not rotate | Starter does not issue current detection confirmation signal; stator coil current absent or <1V | error |
F62 | Bucky unit operation abnormality | Automatic table or Bucky grid positioning mechanism malfunction | error |
F62 | Exposure control system not responding | XOK signal not returned within 1 second after HX1 input | warning |
F6d | Emergency circuit activated | Emergency status detected by control system | critical |
F6E | Exposure did not terminate as commanded | Exposure continued beyond preset timing or mAs setting | critical |
F6E, F6F | Irradiation control or photo timer fault | Exposure control or automatic exposure shutoff circuit malfunction | error |
F6F | Exposure control circuit abnormal | KVT (kV rising signal) not returned within 250ms after EXR sent; simultaneous IPOF signal or initial setting error | error |
H1 | Anode heat over tolerance | Accumulated heat on anode exceeds preset limit during CPU heat calculation | warning |
H1 | Accumulated heat capacity alarm | CPU calculates excessive accumulated heat from prior exposures | warning |
H2 | Accumulated heat capacity alarm | CPU calculates excessive accumulated heat from prior exposures | warning |
H2 | Anode heat at tolerance limit | Accumulated heat on anode reaches preset limit; further exposure will exceed limit | warning |
H3 | X-ray tube thermal switch activated | Thermal switch of X-ray tube activates | warning |
H3 | X-ray tube assembly temperature too high | X-ray tube thermal switch activated or thermal switch signal fault detected | warning |
H4 | X-ray tube assembly temperature too high | Temperature of X-ray tube assembly exceeds safe operating limit | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI