Shimadzu MobileDaRt Evolution: troubleshooting and error codes
The ServiceForge knowledge base holds 48 documented symptoms for the Shimadzu MobileDaRt Evolution, with 57 ranked causes, 82 diagnostic checks with test points and expected values, and 21 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Part replacement notification code reference
Understanding the notification codes and part usage monitoring system for the Notification Function.
Ranked causes
- 1. Main battery on two year timer (common)
- 2. CPU backup battery on two year timer (common)
- 3. Emergency brake release battery on two year timer (common)
- 4. X-ray tube usage counter at 30,000 exposures (occasional)
- 5. Motor usage counter at 7,500 km (occasional)
Temporarily disable battery replacement notification for one month
- Test point
- mAs +/- buttons and key button
- Expected
- Code clears and 'clr yes' selection appears on display
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Notification alert displayed on power up
Red status light indicators lit, three digit alphanumeric code (starting with 'u' followed by 1-5) displayed on kV window upon power up. All functions locked except lock release button and tube arm motion.
Ranked causes
- 1. Battery replacement period reached (main battery, CPU backup battery, or emergency brake release battery) (common)
- 2. X-ray tube replacement period reached (30,000 cycles) (occasional)
- 3. Motor replacement period reached (7,500 km) (occasional)
Verify which part requires replacement by viewing notification code
- Test point
- kV window display
- Expected
- Three digit alphanumeric code starting with 'u' (u1-u5)
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Intermittent F47 errors during charging
User receives intermittent F47 error messages while the system is charging.
Ranked causes
- 1. Outdated software version prior to 1.22 (common)
- 2. Capacitor voltage monitoring during charge cycle (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC will not boot after charging while powered on patient safety
Operator connects charging cord to wall outlet while PC is still ON, then turns off key switch. PC fails to boot subsequently.
Ranked causes
- 1. Inrush current damages F1 on MUX Power 32K or F21 near DR components (common)
- 2. K22 relay welded together due to inrush current (occasional)
Verify PC power state before charging
- Test point
- PC power switch
- Expected
- PC must be OFF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Simulation of radiography cannot be executed
Radiography simulation mode fails; high voltage not generated for testing inverter control circuit
Ranked causes
- 1. Internal capacitor not discharged (occasional)
- 2. Jumper short-circuit at CP3 [CHECK1] to CP4 [CHECK2] on X-CONT2008 board (occasional)
Verify internal capacitor is discharged
- Test point
- X-CONT2008 board capacitor voltage
- Expected
- 0 V
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
70C image noise from line interference radiation
Line noise visible on 70C sensor images affecting image quality
Ranked causes
- 1. 70C sensor susceptibility to line noise interference (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Brake not released when driving with arm lock disengaged patient safety
Brake remains engaged when arm lock is not engaged and operator attempts to drive
Ranked causes
- 1. Software version prior to 1.20 (common)
Verify software version and driving controller operation
- Test point
- System firmware and brake controller status
- Expected
- Version 1.20 or later; brake releases when arm lock disengaged
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery voltage detection
Calibrate battery voltage detection circuit for accurate power monitoring
Ranked causes
- 1. System initialization or battery circuit modification (common)
Perform battery voltage detection calibration
- Test point
- Battery voltage detection circuit
- Expected
- Calibration completed successfully
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Date and time setting
Set system date and time for accurate error log tracking
Ranked causes
- 1. System initialization or time loss (common)
Verify system date and time
- Test point
- System clock display
- Expected
- Current date and time set correctly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Distance Indicator board gain correction
Board requires gain correction setup using SW1 switches at 1m SID to ensure accurate distance measurement display.
Ranked causes
- 1. Gain correction not applied or incorrect (unknown)
Measure SID from focal spot
- Test point
- Focal spot to surface below collimator
- Expected
- 1m (exactly at collimator measuring tape mark)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Distance Indicator Measure-2002 PCB
Factory-adjusted ultrasonic measurement board requires verification and adjustment of measurement and ultrasonic frequencies if distance readings appear incorrect.
Ranked causes
- 1. Measurement pulse frequency out of tolerance (rare)
Measure CP6 frequency
- Test point
- CP6
- Expected
- 34.65KHz±0.5KHz kHz, ±0.5kHz
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Distance Indicator Ultrasonic frequency
Factory-adjusted board requires adjustment of ultrasonic frequency if distance readings appear incorrect.
Ranked causes
- 1. Ultrasonic frequency out of tolerance (rare)
Measure CP7 frequency
- Test point
- CP7
- Expected
- 40KHz±0.5KHz kHz, ±0.5kHz
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Distance Offset and Gain values patient safety
Generator dose calculation requires setting Distance Offset values and adjusting Distance Gain value at 40-inch SID to ensure accurate distance measurement.
Ranked causes
- 1. Distance Offset and Gain values not calibrated (unknown)
Measure and set SID distance
- Test point
- Focal spot to floor surface below collimator
- Expected
- 40 inches
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dose Calculation Unit generator setup patient safety
Dose calculation option requires calibration on the generator console starting in Adjustment Mode with specific kV/mA/msec settings before clinical use.
Ranked causes
- 1. Dose calculation option not calibrated (unknown)
Enter Adjustment Mode
- Test point
- SW3-8 on NEXS-H board
- Expected
- ON; portable reset after change
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Dose Correction Factors radiation
Calibration requires measuring dose at 11 kV stations using ion chamber dosimeter with specific phantom setup to calculate ckV correction factors.
Ranked causes
- 1. Dose correction factors not calculated (unknown)
Verify phantom and collimator setup
- Test point
- Physical setup at X-ray station
- Expected
- Acrylic phantom 20cm square minimum, 19cm thickness; collimator field 20cm square; SOD 100cm
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Finalize dose calculation settings patient safety
Final step to register all dose calculation calibration settings and ensure no further manual adjustments are made to measurement boards.
Ranked causes
- 1. Calibration settings not finalized and registered (unknown)
Return to first radiography condition display
- Test point
- Generator console display
- Expected
- 40kV, 100mA, 20ms; collimator lamp button shows exposure time
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: FVR two-point adjustment
Perform two-point field voltage regulation calibration for small and large field sizes at 50mA and specified mA values
Ranked causes
- 1. Initial system setup or settings reset (common)
Record FVR calibration values
- Test point
- Adjustment mode FVR settings
- Expected
- Small: 80kV 50mA, Small: 80kV 200mA, Large: 80kV 50mA, Large: 80kV 400mA
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Handle calibration
Perform handle sensor calibration adjustment after handle cable addition
Ranked causes
- 1. Handle cable added to system (common)
Verify handle cable is installed
- Test point
- Handle connector and cable
- Expected
- Cable fully connected and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Software upgrade for MobileDaRt Evolution
Complete software installation with settings preservation, NEXSH initialization, and version verification
Ranked causes
- 1. Software upgrade process (common)
Verify keyless entry switch position
- Test point
- Keyless entry switch
- Expected
- OFF position
14 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CXDI RD testing: Verify DICOM dose output
Test CXDI RD operation by taking test exposure and verifying Entrance Dose tag 0040,8302 is present and correct in DICOM header.
Ranked causes
- 1. DICOM dose output not verified or incorrect (unknown)
Setup PACS simulator and local storage destination
- Test point
- CXDI System > Setup Menu > Destination > Storage
- Expected
- STORAGE1 configured; SimpleV24.bat accessible at C:\Program Files\SimpleStorage\
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
F12 error occurs during radiography at high tube current
Tube voltage waveform fails to start; F12 error generated when FVR exceeds 4-5% of 400mA tube current
Ranked causes
- 1. Software version prior to 1.12; FVR waveform implementation issue (common)
Verify software version
- Test point
- System firmware version
- Expected
- 1.12 or later
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
F47 error detected during charge cycle
Error F47 occurs when system is charging; detected during generator charge operation
Ranked causes
- 1. Software version prior to 1.22 (common)
Verify software version
- Test point
- System firmware version display
- Expected
- 1.22 or later
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
False D13 error detection on power-up
D13 error triggered incorrectly when key switch turned on after power plug connected during power-off state
Ranked causes
- 1. Software version prior to 1.04 (common)
Verify software version
- Test point
- System firmware version
- Expected
- 1.04 or later
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
False D7/D8 error detection on key-on
D7 or D8 error triggered incorrectly when key switch turned on after manufacturing tolerance of handle neutral position
Ranked causes
- 1. Software version prior to 1.03; handle neutral position tolerance not accounted for (common)
Verify software version
- Test point
- System firmware version
- Expected
- 1.03 or later
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Falsely detected D13 error at startup
System reports D13 motor diagnostics error during initialization which should not occur
Ranked causes
- 1. Motor diagnostics tolerance too narrow at startup (common)
Verify software version after upgrade
- Test point
- Super Maintenance Mode mAs screen or TeraTerm command prompt
- Expected
- Version 1.04
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Falsely detected D7 error at startup
System reports D7 error during initialization which should not occur
Ranked causes
- 1. FSR status tolerance too narrow at startup (common)
Verify software version after upgrade
- Test point
- Super Maintenance Mode mAs screen or TeraTerm command prompt
- Expected
- Version 1.04
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Falsely detected D8 error at startup
System reports D8 error during initialization which should not occur
Ranked causes
- 1. FSR status tolerance too narrow at startup (common)
Verify software version after upgrade
- Test point
- Super Maintenance Mode mAs screen or TeraTerm command prompt
- Expected
- Version 1.04
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FSR values not displayed in handle adjustment calibration
FSR (focus size regulation) values not enabled for display during handle adjustment calibration
Ranked causes
- 1. Software version 1.01 limitation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
GCOM communication fails after TeraTerm session
Communication to generator fails; serial port misconfiguration prevents system operation
Ranked causes
- 1. NEXSH DIP SW2:7 not reset to OFF (common)
Verify NEXSH DIP SW2:7 position
- Test point
- NEXSH DIP Switch 2, pin 7
- Expected
- OFF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Grid line noise bands on 70C images radiation
Grid-like noise bands or lines appear on images obtained with CXDI-70C sensor
Ranked causes
- 1. Incorrect InrBase Weight value in image processing setup (common)
Verify InrBase Weight parameter value in Line Noise Reduction settings
- Test point
- 70C > Radiography > Line Noise Reduction > InrBaseWeight
- Expected
- 1, normal range 0 to 1
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Handle moves slowly without push after brake release patient safety
Handle drifts or moves unintentionally when brake release bar is released; lacks required operator input
Ranked causes
- 1. Software version prior to 1.03; handle detection range too wide (common)
Verify software version
- Test point
- System firmware version
- Expected
- 1.03 or later
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate battery indicator on PC startup
Battery indicator displays inaccurate status when turning on the PC
Ranked causes
- 1. Software version 1.01 display issue (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate display of FVR in tube current adjustment
FVR (filament voltage regulation) values not displayed accurately in two-point adjustment mode
Ranked causes
- 1. Software version 1.01 display bug (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keep NE software version 1.40.0.11 to 1.40.3.0 without update patient safety
Customer chooses not to upgrade software to newer version or apply patch
Ranked causes
- 1. System operator chooses to retain older NE software version (occasional)
Disable sharpness setting
- Test point
- NE software console sharpness control
- Expected
- Sharpness setting disabled
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lost images after NE 1.40 software update
Images may be lost during operation with NE 1.40 software on MobileDaRt Evolution systems
Ranked causes
- 1. NE 1.40 software issue on CANON NE system (unknown)
Confirm MobileDaRt Evolution with NE 1.40 software is affected
- Test point
- Serial number verification against affected list
- Expected
- Serial number matches affected equipment list
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lost images during acquisition
Images are lost or not saved during X-ray acquisition with NE 1.40 software
Ranked causes
- 1. NE 1.40 software defect (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lost images in CANON NE software patient safety
Images may be lost when using CANON NE software revisions 1.40.0.11 through 1.40.3.0
Ranked causes
- 1. CANON NE software version 1.40.0.11 to 1.40.3.0 installed (common)
Verify CXDI NE software version
- Test point
- CXDI NE system
- Expected
- Check software version number
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MAIN> prompt does not appear on terminal
Serial port setup incomplete; terminal does not display expected command prompt
Ranked causes
- 1. Serial port not configured correctly (common)
Verify serial port configuration
- Test point
- TeraTerm serial port settings
- Expected
- Configured per Appendix A
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software configuration: CCR Console and GUI display settings
Configure CXDI console menu and GUI settings to enable Entrance Dose display and DICOM output.
Ranked causes
- 1. Console menu and GUI display settings not configured for dose output (unknown)
Configure CCR Console menu entrance dose DICOM setting
- Test point
- CCR Setup menu path: 1. Setup > 0. Normal > 5. Transmit Setup
- Expected
- Entrance Dose in mGy (0040,8302) configured as required
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software configuration: CXDI NE and G-com files
CXDI NE software requires correct version-specific G-com files to ensure Entrance Dose tag 0040,8302 is included in image header when sent to PACS.
Ranked causes
- 1. CXDI NE software G-com files incorrect or not installed (unknown)
Verify installed G-com file version and type
- Test point
- System software directories and version numbers
- Expected
- G-com file version matches CXDI NE version; correct file type (DAP, EntDose, or No Dose) installed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software configuration: CXDI RD installation and DICOM dose output
Installing Canon CXDI RD software (minimum version 7.20) and GCOM (minimum 1.50) and configuring DICOM header tag 0040,8302 to output Entrance Dose value.
Ranked causes
- 1. CXDI RD and GCOM software not installed or configured (unknown)
Install CXDI Host Software
- Test point
- System software version check
- Expected
- CXDI Host Software version 7.20 or higher installed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software configuration: Konica Aero DR generator model and overlay settings
Konica Aero DR integration requires correct generator model selection and overlay settings to add Entrance Dose in mGy to clinical image headers.
Ranked causes
- 1. Generator model or overlay settings not configured for Konica Aero DR (unknown)
Verify and set generator model type
- Test point
- System > Utility > X-ray > Settings > Model dropdown
- Expected
- Model set to 'B-40/L40(SESD-10)'
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software installation failure or incomplete upgrade
Equipment does not boot or radiographic condition is not displayed after software installation.
Ranked causes
- 1. Software installation incomplete or corrupted (rare)
Verify equipment boots after installation
- Test point
- System startup
- Expected
- Equipment boots and radiographic condition is displayed when arm lock is released
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade required
Mobile Dart Evolution systems require mandatory software update to version 1.02 to address display and setting registration issues
Ranked causes
- 1. Outdated software version 1.01 (common)
Set up unit for upgrade
- Test point
- Rotary switch, Dip Switch 4:2
- Expected
- Switch in off position, Dip Switch 4:2=ON
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software version F81 or F82 error codes added
New error codes F81 and F82 introduced in software version 1.10 from 1.04
Ranked causes
- 1. Software upgrade adds new error detection (unknown)
System does not function properly after software installation
Equipment fails to operate correctly and displays F1b error after software upgrade.
Ranked causes
- 1. Wrong software version installed (occasional)
Verify correct software version matches generator capacity
- Test point
- System information
- Expected
- MobileArt Evolution 12.5kW uses type 12; MobileArt Evolution 32kW uses type 32; MobileDaRt Evolution uses type 32
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tube mount assembly bolts loose patient safety
Fixing bolts for tube support shafts on X-ray tube can come loose, potentially affecting equipment stability and safety.
Ranked causes
- 1. Slightly shorter screws used during assembly (common)
Inspect tube mount assembly bolts for security
- Test point
- M5 hexagon socket head cap screws on both tube support shafts
- Expected
- Bolts secure, no visible looseness
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
User settings not retained
Individual user settings based on user password not retained between system power cycles
Ranked causes
- 1. Software version 1.01 limitation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
D0273 | GCOM disconnection issue | Communication loss between Canon software components in CXDI NE versions prior to v2.02.0.19 | warning |
D0332 | Improper cropping or masking in image processing | Image acquisition or processing with CXDI NE software versions prior to v2.02.0.19 | error |
D0347 | GCOM disconnection variant | Communication loss between Canon software components in CXDI NE versions prior to v2.02.0.19 | warning |
D13 | False error detected on power-up | Key switch turned on after power plug connected during power-off state | warning |
D13 | Motor diagnostics error | System startup diagnostics | warning |
D7 | False handle position error on key-on | Key switch turned on; handle neutral position tolerance manufacturi issue | warning |
D7 | FSR (Flat Spot Rejection) status error | System startup diagnostics | warning |
D8 | FSR (Flat Spot Rejection) status error | System startup diagnostics | warning |
D8 | False handle position error on key-on | Key switch turned on; handle neutral position tolerance manufacturing issue | warning |
F010151101 | LNK Tool error - COM port mismatch | After logging in, improper USB port used for LNK Module | error |
F12 | Tube voltage waveform failure during radiography | Radiography at 400 mA when FVR exceeds 4-5% of tube current | error |
F1b | Software version mismatch or installation error | After software installation with incorrect version for generator capacity | error |
F47 | False error during generator charge cycle | Generator capacitor charge operation | warning |
F47 | Charging error | During system charging cycle | warning |
F81 | Error code added in software version 1.10 | Unknown | unknown |
F82 | Error code added in software version 1.10 | Unknown | unknown |
u1 | First monitored part (typically main battery) replacement period reached | Two year timer expires from initial factory set date or engineer reset date | warning |
u2 | Second monitored part (typically CPU backup battery) replacement period reached | Two year timer expires from initial factory set date or engineer reset date | warning |
u3 | Third monitored part (typically emergency brake release battery) replacement period reached | Two year timer expires from initial factory set date or engineer reset date | warning |
u4 | X-ray tube replacement period reached (30,000 exposures) | X-ray tube exposure counter reaches 30,000 from initial factory set date or engineer reset date | warning |
u5 | Motor replacement period reached (7,500 km) | Motor usage counter reaches 7,500 km from initial factory set date or engineer reset date | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.