Shimadzu Mobile Art: troubleshooting and error codes
The ServiceForge knowledge base holds 15 documented symptoms for the Shimadzu Mobile Art, with 20 ranked causes, 34 diagnostic checks with test points and expected values, and 7 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
TeraTerm communication not establishing
MAIN> command prompt not appearing when connecting via TeraTerm serial terminal
Ranked causes
- 1. NEX-SH DIP Switch 2:7 not set to OFF (common)
- 2. TeraTerm serial port configuration incorrect (common)
- 3. Terminal settings not properly configured (common)
- 4. Generator in Auto Power Off (APO) mode (occasional)
Verify DIP Switch 2:7 position on NEX-SH PCB
- Test point
- NEX-SH PCB DIP Switch 2, position 7
- Expected
- OFF
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator leaf coming loose or falling off patient safety
R-20C collimator leaves become loose and may fall off during operation
Ranked causes
- 1. Inadequate washer diameter allowing leaf to shift (common)
- 2. Leaves not adequately secured by standard fastening method (common)
Inspect all 16 collimator leaf washers
- Test point
- Each leaf fastening point
- Expected
- All washers are wide diameter non-reflective black type
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data upload/download failure during system restoration
File upload or download of system settings and calibrations fails when using MAINTENANCE command in TeraTerm
Ranked causes
- 1. TeraTerm version older than 4.57 (common)
- 2. Serial port or terminal settings not properly configured (common)
Verify TeraTerm version installed
- Test point
- TeraTerm application version info
- Expected
- Version 4.57 or later
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charging capacity degradation
Mobile X-ray system battery loses charging capacity over time, reducing operational availability
Ranked causes
- 1. Premature Capacity Loss (PCL) from frequent shallow charge cycles (common)
Monitor battery remaining charge LED status during daily operation
- Test point
- Battery LED indicator
- Expected
- At least four LEDs unlit, leaving one LED lit or blinking before initiating charge
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charging capacity drops
Mobile system battery no longer charges to full capacity; premature capacity loss (PCL) occurs when nonconductive barrier layer forms on positive electrode
Ranked causes
- 1. Frequent charging and discharging cycles (common)
Verify number of remaining charge LEDs lit before charging
- Test point
- Battery remaining charge indicator LEDs
- Expected
- One LED lit or blinking (four unlit)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery charging capacity drops suddenly
Mobile Art or MUX-100/100H/100D system battery loses charging capacity abruptly; system runtime decreases despite full charge cycles
Ranked causes
- 1. Premature Capacity Loss (PCL) from frequent shallow charge cycles (common)
Observe battery remaining charge LED indicators
- Test point
- Battery LED panel
- Expected
- At least four LEDs unlit, one LED lit or blinking
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery replacement procedure for MobileArt/DaRt Evolution and MUX-100 series
New battery model (PXL-12090, P/N 074-71014-11) replaces discontinued PX-12090 (P/N 074-71014-01). Software and battery code settings must be updated to ensure system operates normally.
Ranked causes
- 1. Battery model discontinued; new PXL-12090 (P/N 074-71014-11) requires software and code updates (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration and system settings require restoration after initialization
Multiple adjustment settings and calibrations need to be manually configured after system initialization or data restore
Ranked causes
- 1. System initialization or EPROM reset performed (DIPSW 2-8/3-1 ON) (common)
Verify FVR calibration values restored
- Test point
- FVR settings in equipment
- Expected
- Original recorded FVR values
5 more checks with pass and fail actions, 12 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator blades coming loose or falling off high voltage
Collimator blades have become loose and in some cases fallen off the mobile X-ray system
Ranked causes
- 1. Small washer inadequate for securing collimator blade screws (common)
Inspect collimator blade attachment for loose or missing washers
- Test point
- Collimator blade screw fasteners
- Expected
- Washers present and secure
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuses blow during charging high voltage
Fuses fail when the system attempts to charge the high-voltage generator or capacitor bank during startup or operation.
Ranked causes
- 1. Charging circuit fault or component failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuses blowing during charging high voltage
Fuses F21, F1, or F10 blow during system charging. In worst cases, PS12 (24V DC) power supply fails or relays K22 and K21 weld together.
Ranked causes
- 1. Inadequate fuse protection on charging circuit (common)
Capacitor discharge via discharge key-switch
- Test point
- Discharge key-switch on right side next to MUX CAPACITOR board
- Expected
- Position set to DISCHARGE
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
GCOM communication fails after TeraTerm session
Generator communication (GCOM) fails or is broken after using TeraTerm terminal session on MobileDaRt
Ranked causes
- 1. NEX-SH DIP Switch 2:7 not reset to ON after TeraTerm session (common)
Verify NEX-SH DIP Switch 2:7 position
- Test point
- NEX-SH PCB DIP Switch 2, position 7
- Expected
- ON (for MobileDaRt; OFF only during TeraTerm use)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MobileArt/DaRt Evolution battery code setting verification
Confirm battery code is set to '6' after software update and system operates normally with new battery installed.
Ranked causes
- 1. Battery code must be updated from '5' to '6' for compatibility with new PXL-12090 battery (common)
Verify system operates normally after software update and battery code change
- Test point
- System operation
- Expected
- System operates normally; settings retained in system
Software not displaying correctly after installation
Equipment fails to boot or radiographic condition is not displayed when releasing arm lock after software installation
Ranked causes
- 1. Software installation incomplete or failed (common)
Verify software version is displayed correctly
- Test point
- Super Maintenance Mode mAs screen or TeraTerm command prompt
- Expected
- Software version displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System displays F1b error after software upgrade
F1b error appears on display after applying incorrect software version for generator capacity
Ranked causes
- 1. Incorrect software version installed for generator capacity (common)
Verify generator type and required software version
- Test point
- System display or generator specifications
- Expected
- Software Type 12 for MobileArt Evolution 12.5kW; Software Type 32 for MobileArt Evolution 32kW or MobileDaRt Evolution
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
D13 | Power initialization error | False detection when key switch switched ON after power plug connected under power OFF status, or when key switch switched ON under power OFF status | warning |
D7/D8 | Handle detection error | Detected when key switch turned ON, false detection occurs with handle neutral position allowance manufacturing tolerance | error |
F12 | Tube voltage start waveform failure | When FVR value set exceeding tube current of 4-5 percent of 400mA during radiography at 400mA | error |
F1b | Incorrect software version applied to system | Wrong software version loaded for generator capacity | error |
F47 | Charge cycle detection error | Detected during equipment charge cycle in software versions prior to 1.22 | warning |
F81 | Error code F81 (specific cause not detailed) | Introduced in software version 1.10 | unknown |
F82 | Error code F82 (specific cause not detailed) | Introduced in software version 1.10 | unknown |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI