Shimadzu All portable x-ray: troubleshooting and error codes

The ServiceForge knowledge base holds 4 documented symptoms for the Shimadzu All portable x-ray, with 5 ranked causes, 8 diagnostic checks with test points and expected values, and 14 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

System sluggishness or abnormal shutdown after prolonged screensaver operation

CCS application delayed startup, abnormal PC shutdown, or display of multiple error codes after screensaver runs for extended period

Ranked causes

  1. 1. Original screensaver consumes excessive CPU resources causing system instability (common)
  2. 2. Screensaver left enabled increases overall PC power consumption (occasional)
First check

Verify screensaver setting after upgrade

Test point
Operating system screensaver configuration
Expected
Screensaver enabled with 15-minute timeout or monitor shutdown after 15 minutes

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CXDI NE v2.12.0.13 software upgrade

Complete upgrade procedure for Canon CXDI NE software to version 2.12.0.13 on MobileDaRt Evolution systems

Ranked causes

  1. 1. Software upgrade required for new detector compatibility and system optimization (common)
First check

Determine which upgrade method applies to system

Test point
C:\Shimadzu\Version\Version.txt second line Image Version
Expected
EVNE2.0SP or higher = EVNE24 Utility; prior to EVNE2.0SP = SSUM

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error F040300054 displayed when clinical software executes

Error appears when running clinical software application after upgrade

Ranked causes

  1. 1. Detector firmware not updated to required version (common)
First check

Verify detector firmware version matches CXDI NE v2.12.0.13 requirements

Test point
Detector identification and firmware location
Expected
Detector firmware version compatible per Appendix A matrix

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

mAs value discrepancies between generator and CCS display

mAs values displayed at generator differ from CCS display due to msec rounding errors

Ranked causes

  1. 1. Standard GCOM file rounding errors in msec value calculation (common)
First check

Verify GCOM mAs feature file loaded after upgrade

Test point
GCOM file version in system configuration
Expected
GCOM v2.10-NE_EVO with mAs designation matching dose configuration

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
F03010000DCCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operation consuming high CPU resourceserror
F040200001CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F040300054Detector firmware incompatible or outdatedClinical software application execution after upgrade to CXDI NE v2.12.0.13 with incompatible detector firmwareerror
F04050100DCCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F04050100ECCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F040600003CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F040600004CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F040600005CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F040800001CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F040A00001CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F040D00001CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F040F00001CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F041000003CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror
F041100003CCS application error from screensaver CPU resource issueSystem startup after prolonged screensaver operationerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.