SEDECAL SG-80: troubleshooting and error codes
The ServiceForge knowledge base holds 23 documented symptoms for the SEDECAL SG-80, with 31 ranked causes, 33 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Bucky vertical movement not possible patient safety
Bucky cannot move vertically; may be obstructed, unlocked, or power issue
Ranked causes
- 1. Column stand obstructed by element (common)
- 2. Power supply is OFF (common)
- 3. Vertical lock engaged (occasional)
- 4. Switch malfunction (occasional)
- 5. Broken chains (rare)
Inspect for obstructions blocking bucky movement
- Test point
- Column stand path
- Expected
- Clear of obstructions
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky does not tilt patient safety
Bucky cannot tilt; power or mechanical issue
Ranked causes
- 1. Power supply is OFF (common)
- 2. No power in coil (occasional)
- 3. Broken cables (occasional)
Verify power supply status
- Test point
- Wall Stand switch
- Expected
- ON
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto-tracking does not work during vertical movement
Auto-tracking feature fails while bucky moves vertically
Ranked causes
- 1. Potentiometer broken (occasional)
- 2. Service Tool not working properly (occasional)
Test potentiometer condition (automatic and auto-tracking models)
- Test point
- Potentiometer
- Expected
- Not broken
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto-tracking does not work when tilting
Auto-tracking feature fails during bucky tilt operation
Ranked causes
- 1. Potentiometer broken (occasional)
- 2. Service Tool not working properly (occasional)
Test potentiometer condition
- Test point
- Potentiometer
- Expected
- Not broken
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Angulation lock does not hold or binds (SG120 only)
Angulation lock fails to engage or binds after release, preventing secure angular positioning
Ranked causes
- 1. Lock hardware misalignment, wear, or damage (occasional)
Release angulation lock button and verify lock engages
- Test point
- Angulation lock mechanism
- Expected
- Lock engages and prevents further angulation
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Angulation lock does not release (SG120 only)
Bucky angulation lock remains engaged after pressing release button behind front panel
Ranked causes
- 1. Lock board malfunction (unknown)
Press angulation lock push button and verify lock releases
- Test point
- Angulation lock mechanism
- Expected
- Lock releases and Bucky tilts freely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky carriage does not move easily vertically patient safety
Bucky carriage movement becomes difficult or hard during vertical displacement with power on and electromechanical brake released
Ranked causes
- 1. Counterweight and bucky carriage not well balanced (unknown)
Verify counterweight and bucky carriage balance
- Test point
- Bucky carriage vertical motion
- Expected
- Carriage moves easily in vertical direction
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky grid does not oscillate
Grid does not oscillate during vertical bucky stand exposure; no audible oscillation sound or vibration
Ranked causes
- 1. Grid motor or drive mechanism failure (unknown)
Take exposure with vertical bucky stand selected
- Test point
- Bucky grid assembly
- Expected
- Grid oscillates with distinct audible sound and visible vibration
Bucky grid shading or objectionable grid lines in film
Radiographic image exhibits grid shading or objectionable grid line artifacts
Ranked causes
- 1. Improper grid alignment or focal distance setting (common)
Observe exposed film for quality
- Test point
- Radiographic image
- Expected
- No shading or objectionable grid line patterns
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky rotation direction incorrect (SG120 only) patient safety
Bucky rotates in wrong direction compared to standard configuration (left or right hand dependent)
Ranked causes
- 1. Rotation limit screw positioned at wrong microswitch (occasional)
Check bucky rotation direction against Illustration 4-1
- Test point
- Bucky rotation
- Expected
- Rotation direction matches system configuration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky rotation lock does not engage (SG120 only)
Rotation lock lever fails to prevent Bucky rotation after lever is applied
Ranked causes
- 1. Rotation lock hardware misalignment, wear, or damage (occasional)
Release rotation lock lever and verify Bucky rotates 0° to 180°
- Test point
- Rotation lock lever
- Expected
- Bucky rotates easily from 0° (upright) to 180° (upside down)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bucky voltage mismatch with generator high voltage
Bucky control board voltage setting does not match generator voltage supply
Ranked causes
- 1. Bucky factory set to 115VAC, generator supplies different voltage (occasional)
Verify Bucky voltage setting matches generator
- Test point
- Bucky control board configuration
- Expected
- Bucky voltage matches generator voltage (115VAC or 230VAC)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carriage motion not smooth or jerky
Vertical carriage does not move smoothly, exhibits stiff or binding motion
Ranked causes
- 1. Incorrect counterweight installation or imbalance (common)
Inspect counterweight installation
- Test point
- Counterweight assembly
- Expected
- Counterweights correctly installed and trimcovers in place
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carriage remains locked when electromechanical lock released
Vertical carriage does not move freely after releasing the vertical lock handle with power on
Ranked causes
- 1. Lock board not feeding electromechanical lock correctly (unknown)
Verify carriage moves freely and smoothly over entire range
- Test point
- Carriage travel
- Expected
- Free and smooth motion with minimum side-play
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carriage suddenly becomes difficult or hard to move vertically caution
Vertical carriage motion abruptly becomes stiff or requires excessive force during operation
Ranked causes
- 1. Unknown mechanical binding or failure (unknown)
Monitor carriage movement difficulty
- Test point
- Carriage vertical motion
- Expected
- Smooth consistent movement without sudden resistance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Equipment instability during standing or movement patient safety
Risk of tipping or unbalancing the SG80/120 stand during setup or repositioning
Ranked causes
- 1. Insufficient personnel for equipment handling (common)
Verify personnel count
- Test point
- standing area
- Expected
- minimum two people present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microswitch S1 not actuated when Bucky within 10° vertical
Microswitch S1 fails to trigger when Bucky is within 10° of vertical position
Ranked causes
- 1. Microswitch S1 misalignment or mechanical failure (occasional)
Position Bucky within 10° of vertical and verify S1 actuation
- Test point
- Microswitch S1
- Expected
- Microswitch S1 actuated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microswitch S2-1 not actuated at Bucky 0° position (SG120 only)
Microswitch S2-1 does not trigger when Bucky is at 0° (upright) position
Ranked causes
- 1. Microswitch S2-1 misalignment or mechanical failure (occasional)
Position Bucky at 0° and verify microswitch S2-1 actuation
- Test point
- Microswitch S2-1
- Expected
- Microswitch S2-1 actuated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microswitch S2-2 not actuated at Bucky 180° position (SG120 only)
Microswitch S2-2 does not trigger when Bucky is at 180° (inverted) position
Ranked causes
- 1. Microswitch S2-2 misalignment or mechanical failure (occasional)
Position Bucky at 180° and verify microswitch S2-2 actuation
- Test point
- Microswitch S2-2
- Expected
- Microswitch S2-2 actuated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microswitch S3 not actuated when Bucky within 10° horizontal
Microswitch S3 fails to trigger when Bucky is within 10° of horizontal position
Ranked causes
- 1. Microswitch S3 misalignment or mechanical failure (occasional)
Position Bucky within 10° of horizontal and verify S3 actuation
- Test point
- Microswitch S3
- Expected
- Microswitch S3 actuated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microswitch S4 not actuated at near vertical SID point (SG120 only)
Microswitch S4 does not trigger when Bucky is raised to near vertical SID point
Ranked causes
- 1. Microswitch S4 misalignment or mechanical failure (occasional)
Raise Bucky and verify S4 actuation at near vertical SID point
- Test point
- Microswitch S4
- Expected
- Microswitch S4 actuated when Bucky front panel is at table height
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Microswitch S5 not actuated at far vertical SID point (SG120 only)
Microswitch S5 does not trigger when Bucky is at far vertical SID point after rotating to 0° and angulating 90°
Ranked causes
- 1. Microswitch S5 misalignment or mechanical failure (occasional)
Rotate Bucky to 0° and angulate 90° (horizontal), verify S5 actuation
- Test point
- Microswitch S5
- Expected
- Microswitch S5 actuated at far vertical SID point
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pulley chains off sprockets after unpacking caution
Top cover removed during setup reveals pulley chains not seated on sprockets
Ranked causes
- 1. Chains displaced during shipping (occasional)
Inspect pulley chain seating
- Test point
- pulley sprockets
- Expected
- chains seated fully on sprockets
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI