QuidelOrtho VITROS 5600: troubleshooting and error codes
The ServiceForge knowledge base holds 123 documented symptoms for the QuidelOrtho VITROS 5600, with 181 ranked causes, 347 diagnostic checks with test points and expected values, and 15 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
MicroSensor Check Fluid results excessively out of range
Results from processing MicroSensor Check Fluids are out of specification, indicating potential issues with sensor performance or optical components.
Ranked causes
- 1. Light leak into system or defective MICROSENSOR SUBSYSTEM (common)
- 2. Defective MICROSENSOR SUBSYSTEM (common)
- 3. Defective REFLECTOMETER LAMP (occasional)
- 4. Defective ILLUMINATION FIBER (occasional)
- 5. Incorrectly positioned CUVETIP LIFTER (occasional)
- 6. Defective SPECTROPHOTOMETER (occasional)
- 7. Excessive electrical noise (occasional)
Perform MicroSensor Static Reference Test
- Test point
- Diagnostics > Performance Tests > MICROSENSOR > Static Reference
- Expected
- SD and Mean graphs display with pixel identification
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Reflectometer Correction Factors
Optimize optical calibration of the reflectometer using White Reference Slides to ensure accurate density measurements. Procedure takes 12 minutes and generates Pass, Update, or Repeat status results.
Ranked causes
- 1. Correction Factor assay value typed incorrectly (occasional)
- 2. White Reference Slide mishandled or expired (occasional)
- 3. Expected optical drift over long period (>6 months) (common)
- 5. Expected change after maintenance (lamp or aperture) (occasional)
Move WASH FLUID METERING to Maintenance position
- Test point
- Diagnostics > MEDs > WF Metering > Move/Cycle
- Expected
- Move complete, top right cover opened
8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MicroSlide Incubator reassembly after maintenance caution
Install and verify proper assembly of incubator components including wear pads, rings, belt, and covers
Ranked causes
- 1. Routine maintenance completion requiring reassembly (common)
- 2. CM/RT RING installation and alignment (common)
- 3. Motor belt installation (common)
- 4. Incubator covers and external components reinstallation (common)
Install wear pads correctly on hot plate
- Test point
- All three WEAR PAD positions on HOT PLATE
- Expected
- PADS recessed and worn side visible
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MicroTip reagent ring jammed
MicroTip reagent ring does not rotate; jam in load area or due to stuck pack opener or lifter
Ranked causes
- 1. Pack opener stuck in lowered position (common)
- 2. Pack lifter stuck in raised position (occasional)
- 3. Obstructions or stuck reagent pack in supply (occasional)
- 4. Jam not cleared after initial checks (rare)
Verify pack opener position
- Test point
- MICROTIP PACK OPENER
- Expected
- Not stuck in lowered position
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Photometer Static Read Test fails
Performance test does not complete successfully after lamp replacement or when assay precision affected. May indicate improper lamp seating, electrical interference, or vibration.
Ranked causes
- 1. Lamp not seated properly in housing (common)
- 2. Power disturbances or electrical interference in laboratory (occasional)
- 3. External vibration affecting lamp filament alignment (occasional)
- 4. Lamp filament failure or defect (rare)
Select wavelength and run test
- Test point
- Diagnostics > Performance Tests > PHOTOMETER > Static
- Expected
- Test completes successfully; status prompt shows pass
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Slide jam in MicroSlide Incubator caution
Slides stuck or jammed in incubator ring, platens, tip locator, or discard chutes
Ranked causes
- 1. Slide stuck at PM RING entrance (common)
- 2. Slide stuck in slide supply platen or dispense blade path (common)
- 3. Slide stuck at tip locator (occasional)
- 4. Slide stuck in discard chute or WF shuttle (occasional)
Attempt automatic jam clearing via system initialization
- Test point
- System Status Console
- Expected
- No condition code appears after initialization
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tip jam in MICROWELL REAGENT METERING TIP DISCARD CHUTE caution
Tips have become blocked or jammed in the discard chute, preventing normal tip disposal.
Ranked causes
- 1. Tips not properly falling into DISCARD CHUTE or accumulating in CHUTE (common)
- 2. Blockage inside TIP DISCARD CHUTE from tip jam (common)
- 3. Tip waste overflow in WASTE CONTAINER A restricting CHUTE drainage (occasional)
- 4. Blockage at bottom of TIP DISCARD CHUTE near TIP DEFLECTOR (occasional)
Inspect TIP EJECT MECHANISM for trapped tip
- Test point
- TIP EJECT MECHANISM inside top left cover
- Expected
- No tip present
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuvette jammed in supply or retrieve area patient safety
Cuvettes stuck in cuvette supply or between supply and incubator, cuvettes loaded incorrectly or diagonally
Ranked causes
- 1. Cuvette loaded incorrectly or jammed diagonally in supply (common)
- 2. Cuvette jammed in retrieve area between supply and incubator (occasional)
- 3. Cuvette completely inside incubator (occasional)
Determine jam location in supply
- Test point
- Cuvette supply
- Expected
- Jam location identified (diagonal, retrieve area, or inside incubator)
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard does not respond to any keys
No response when pressing keys on the keyboard connected to the system
Ranked causes
- 1. Keyboard connection issue or system needs restart (common)
- 2. Keyboard hardware failure (occasional)
- 3. Keyboard requires replacement (occasional)
Restart system with keyboard connected
- Test point
- System startup
- Expected
- System completes startup and keyboard responds to key presses
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid waste bottle leak or improper connection patient safety
Fluid aspirates or dispenses incorrectly, condition codes occur, or results are lost due to improper waste bottle cap or quick disconnect attachment
Ranked causes
- 1. Liquid Waste Bottle Cap not securely tightened or damaged (occasional)
- 2. Quick Disconnect Cap not fully attached to Bottle Cap (occasional)
- 3. Tubing on Quick Disconnect Cap discolored, cracked, leaking, or crimped (occasional)
Visually inspect Liquid Waste Bottle Cap for damage
- Test point
- Cap surface
- Expected
- Cap intact, no cracks or damage, screw attached securely
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MicroSlide Incubator does not rotate smoothly caution
CM/RT RING or PM RING does not move smoothly after reassembly or during operation
Ranked causes
- 1. Wear pads not properly recessed or shifted out of position (common)
- 2. Belt not properly installed on motor spindle (occasional)
- 3. PM RING sensor contamination or obstruction (occasional)
Visually inspect wear pad positioning and recessing
- Test point
- Position 36, all three PADS on HOT PLATE
- Expected
- All three PADS recessed with part of HOT PLATE visible through center of each PAD
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer does not print test page
Test page does not print when Print Test Page is selected after printer installation or during troubleshooting
Ranked causes
- 1. Printer not properly connected or powered (common)
- 2. Printer not enabled in system configuration (common)
- 3. Printer hardware failure or configuration error (occasional)
Verify printer power
- Test point
- Power outlet and printer power switch
- Expected
- Power cord plugged in, printer powered on
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal Reagent not dispensed at specified volume
SR fluid dispensing volume is below specification or inconsistent, affecting test accuracy and requiring system cleaning.
Ranked causes
- 1. Crystallization on DISPENSE PROBES (common)
- 2. Signal Reagent tubing or pump contamination (common)
- 3. Quality Control drift or calibration failure (occasional)
Remove SR pack from Load position
- Test point
- SR CAROUSEL Load position
- Expected
- SR pack removed; discard if used, retain if unused
7 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wash fluid leakage or metering issues caution
WF METERING shows leaks or improper fluid delivery after maintenance tasks
Ranked causes
- 1. TIP SLEEVE LOCK not fully tightened on PROBOSCIS (common)
- 2. PROBOSCIS not fully dried before MicroTip installation (occasional)
- 3. Contaminated PROBOSCIS or TIP SLEEVE LOCK (occasional)
Verify TIP SLEEVE LOCK is fully seated on PROBOSCIS
- Test point
- TIP SLEEVE LOCK threads on PROBOSCIS
- Expected
- LOCK contacts top of PROBOSCIS with no gap
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration program not defined correctly
User-defined calibration program does not include desired assays or calibrator assignments are incorrect
Ranked causes
- 1. Incorrect assay selection during program definition (occasional)
- 2. Calibrator kit lot numbers not assigned (occasional)
Verify assay selection matches calibration requirements
- Test point
- Calibration dialog assay buttons
- Expected
- All required assays selected; highlighted assays show applicable kit numbers
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CM RING SENSOR requires cleaning
System indicates sensor cleaning required via condition code or Ortho Care notification
Ranked causes
- 1. CM RING SENSOR contaminated with residue or particles (common)
- 2. System reports sensor error or inconsistent readings (occasional)
Stop MICROSLIDE INCUBATOR rotation
- Test point
- MEDs Incubators and Transports CM/RT Ring Reset button
- Expected
- Incubator has stopped rotating
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERF Metering pump failure or disconnection caution
ERF METERING PUMP not functioning, producing no metering action, or harness connections loose
Ranked causes
- 1. ERF METERING PUMP internal failure or wear (occasional)
- 2. Harness connections loose or disconnected (occasional)
Inspect PUMP harness connections for seating
- Test point
- FB2J2, FS2J6, FC1J7 connectors
- Expected
- Connectors fully seated with locking tabs engaged
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERF Metering tubing obstruction, disconnection, or leak
Tubing not properly connected, kinked, or blocked, preventing electrolyte reference fluid flow
Ranked causes
- 1. Tubing disconnected or insufficiently seated on pump fitting (common)
- 2. Tubing internal obstruction or damage (occasional)
Inspect tubing connection at pump fitting
- Test point
- Pump fitting where tubing connects
- Expected
- Tubing fully seated with no gap, secure connection
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IWF reservoir depletion or contamination
IWF reservoir becomes empty, contaminated, or lot number expires during operation
Ranked causes
- 1. 72-hour service interval reached for IWF reservoir (common)
- 2. Contaminated RESERVOIR BASE, LID, or LEAK PAD (occasional)
Verify IWF reservoir age and lot number match system requirements
- Test point
- Reservoir label lot number vs. dialog display
- Expected
- Lot numbers match; reservoir not expired per label date
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MICROIMMUNOASSAY REAGENT SUPPLY jam cannot be cleared by reset
A jam occurs in the MICROIMMUNOASSAY REAGENT SUPPLY that persists after attempting reset in MEDs.
Ranked causes
- 1. MICROTIP REAGENT RING is jammed (unknown)
- 2. MICROWELL REAGENT RING is jammed or MICROWELL DISPENSER stuck in lowered position (unknown)
Open MICROWELL and MICROTIP LOAD/UNLOAD DOORS and identify which RING is jammed
- Test point
- MICROWELL LOAD/UNLOAD DOOR and MICROTIP REAGENT LOAD/UNLOAD DOOR
- Expected
- One or both rings can be rotated by hand, or one is stuck
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MicroSlide cartridge count out of specification
Slide count detected during plunger calibration test differs from actual cartridge count beyond acceptable tolerance
Ranked causes
- 1. Plunger calibration drift or error (occasional)
- 2. Cartridge contamination or mechanical wear (occasional)
Run Count test on Supply 1 or Supply 2 Plunger
- Test point
- Diagnostics > MEDs > Slide Supplies
- Expected
- Initial cartridge count recorded
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MicroSlide Incubator post-maintenance verification
Perform diagnostic tests and quality control checks after maintenance or repair
Ranked causes
- 1. Routine post-maintenance verification required after incubator service (common)
- 2. Quality control validation of all affected tests (common)
Perform pad reflectance test
- Test point
- MicroSlide Incubator pad reflectance measurement
- Expected
- Pad reflectance within acceptable range
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MicroWell pack opener jam unresolved
Jam persists after initial troubleshooting steps; cannot open or close MicroWell pack system
Ranked causes
- 1. Debris or blockage in MicroIA reagent supply system (occasional)
- 2. Persistent mechanical jam or component failure (rare)
Verify jam status after removal and inspection
- Test point
- MicroWell pack opener knob rotation
- Expected
- Knob turns freely and pack opens/closes
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MicroWell Wash does not aspirate properly after filter replacement
System fails to aspirate reagent into Prime/Purge Station after WELL WASH ASPIRATE FILTER installation
Ranked causes
- 1. FILTER not threaded correctly onto MANIFOLD FITTING (common)
- 2. ADAPTER FITTING not attached correctly to FILTER (common)
Verify FILTER installation thread security
- Test point
- MANIFOLD FITTING connection
- Expected
- FILTER screwed on until it stops with finger-tight force
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Photometer Dark Static Read Test fails
Test to diagnose light leaks into photometer or controller board interference. Run when assay performance issues suspected.
Ranked causes
- 1. Light leaking into photometer housing (occasional)
- 2. Controller board electrical interference (occasional)
Run Dark Static Read Test
- Test point
- Diagnostics > Performance Tests > PHOTOMETER > Dark Static
- Expected
- Test completes successfully
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Photometer Water Blank results do not pass
Water Blank procedure completes but results fail validation when examined on screen
Ranked causes
- 1. Photometer calibration drift or reference issue (unknown)
- 2. System or supply issue (unknown)
Verify supplies are loaded and available before procedure
- Test point
- MICROIMMUNOASSAY REAGENT SUPPLY, CUVETTE SUPPLY, MICROIMMUNOASSAY VERSATIP RING
- Expected
- At least 1800 µL water, at least 3 rows cuvettes, at least 18 VersaTips
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Primary Tip Sealer temperature not stabilizing
Temperature does not come into range and stabilize within approximately 5 minutes after Tip Sealer turned on
Ranked causes
- 1. AMBIENT/PRIMARY TIP SEALER TEMPERATURE TRANSDUCER BOARD failure or disconnection (common)
- 2. Environmental condition preventing normal temperature range (occasional)
Verify Primary Tip Sealer temperature reading
- Test point
- Diagnostics > System Information > Environmental Monitoring
- Expected
- Temperature should stabilize within approximately 5 minutes
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reagent pack inventoried incorrectly
New MicroWell, MicroTip, or MicroSlide cartridge not properly recognized during system inventory
Ranked causes
- 1. Pack not properly positioned or detected during inventory (occasional)
- 2. Metering assembly not properly seated (occasional)
Identify reagent from Condition Review code
- Test point
- Condition Review screen code description
- Expected
- Reagent type identified
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Shutdown does not complete
System does not reach final shutdown state; status messages indicate ongoing processing or dialog appears
Ranked causes
- 1. Assays currently in progress (occasional)
- 2. Metering or assay processing in progress (common)
Check system activity status
- Test point
- Confirm Shutdown dialog
- Expected
- Dialog displays current system activity status
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal Reagent Metering accuracy incorrect patient safety
Signal Reagent volume dispensed does not meet calibration tolerance
Ranked causes
- 1. Dispense probe positioning incorrect relative to volumetric flask (common)
- 2. Tubing kinked or obstructed (occasional)
Measure volume dispensed against calibration line
- Test point
- Meniscus level in Class A volumetric flask after dispense
- Expected
- Meniscus on or below calibration line
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tips jammed in Secondary Tip Sealer caution
Tips stuck in Dispose area or Strip area; Dispose area to WASTE CONTAINER B tubing may be blocked
Ranked causes
- 1. Tips accumulated in Dispose or Strip areas (common)
- 2. Damaged or missing DIVERTER (occasional)
Visually inspect Dispose and Strip areas for tips
- Test point
- Dispose area and Strip area
- Expected
- No tips visible; area clear
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VersaTip tube jam at coupling caution
Tips jammed in microimmunoassay versatip tube; cannot be cleared through load coupling
Ranked causes
- 1. Tips jammed in load area (common)
- 2. Tips jammed inside tube end or coupling (occasional)
Check load coupling position
- Test point
- TIP LOAD COUPLING
- Expected
- Unlocked position (0)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
WF Metering Leak and Hysteresis Test failure
Test indicates metering errors or leaks detected after MicroTip replacement or tubing work
Ranked causes
- 1. MicroTip damaged during test (occasional)
- 2. Tubing improperly installed or clogged (occasional)
Inspect MicroTip for visible damage after test completion
- Test point
- MicroTip visual inspection
- Expected
- MicroTip intact with no cracks, deformation, or contamination
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
WF Metering leak detected
System condition codes indicate Immuno-Wash Fluid leaks; TIP SLEEVE LOCK not tight enough or tubing damaged/pinched
Ranked causes
- 1. TIP SLEEVE LOCK not tightened completely (common)
- 2. WF METERING TUBING worn, pinched, or damaged (common)
Examine WF METERING TUBING for damage, blockage, or pinching
- Test point
- WF METERING PROBOSCIS tubing attachment
- Expected
- Tubing appears undamaged, clear, not pinched or blocked
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader not reading samples
System fails to scan sample trays or read barcodes during sample loading or metering
Ranked causes
- 1. Dirty barcode reader window (common)
Inspect barcode reader windows for dirt or contamination
- Test point
- STAT and routine barcode reader windows
- Expected
- Clean, clear windows without dirt or debris
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Barcode reader alignment check
Periodic verification that barcode readers are positioned and functioning correctly for reliable sample identification
Ranked causes
- 1. Barcode reader requires periodic maintenance and alignment check (occasional)
Position empty transport at barcode reader
- Test point
- Routine or STAT barcode reader position
- Expected
- Transport centered under reader for optimal scanning
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Luminometer IRS
Incubator Reference System calibration procedure for luminometer to ensure proper signal detection and measurement accuracy.
Ranked causes
- 1. Luminometer requires IRS calibration per maintenance schedule or condition code (common)
Open top left cover
- Test point
- top left cover
- Expected
- cover open and accessible
11 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MicroWell Wash Aspiration Rate
Verification and calibration of MicroWell Wash aspiration rate performance
Ranked causes
- 1. Aspiration rate out of specification (unknown)
Verify MICROWELL WASH pressure
- Test point
- 10 psi REGULATOR pressure gauge
- Expected
- 10 psi
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: MicroWell Wash Dispense Volume
Verification and calibration of MicroWell Wash dispense volume performance
Ranked causes
- 1. Dispense volume out of specification (unknown)
Verify MICROWELL WASH pressure
- Test point
- 10 psi REGULATOR pressure gauge
- Expected
- 10 psi
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Photometer wavelength selection for Static Read Test
Select correct wavelength for Photometer Static Read Test based on assay type being tested to ensure proper photometer performance validation.
Ranked causes
- 1. Need to validate photometer response at assay-specific wavelength (common)
Select wavelength 340 if new lamp installed
- Test point
- Filter pulldown in Static Read Test
- Expected
- 340 nm selected
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Slide dispense blade installation
Verification of proper blade orientation and seating during installation
Ranked causes
- 1. Incorrect blade orientation during installation (occasional)
Verify blade orientation: ridged side position
- Test point
- Blade slot
- Expected
- Ridged side facing upward
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: VERSATIP SUPPLY COMPRESSOR FILTER Replacement caution
Routine maintenance procedure to replace compressor filter every 6 months or as indicated by condition code. System components may be damaged if circuit breaker not powered off.
Ranked causes
- 1. Routine maintenance interval or condition code alert (common)
Power off Circuit Breaker 12
- Test point
- Circuit Breaker 12
- Expected
- Power off
6 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CM and RT DISCARD CHUTE blockage or contamination
Slides not advancing through discard chutes or debris accumulation in chutes
Ranked causes
- 1. Slide residue or debris in DISCARD CHUTES (occasional)
Inspect CM/RT DISCARD CHUTES for blocked slides or debris
- Test point
- Interior of CM DISCARD CHUTE and RT DISCARD CHUTE
- Expected
- Chutes clear and clean with no slides or debris
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CM/RT RING cleaning required
Incubator ring contamination affecting system performance per condition code or service guidance
Ranked causes
- 1. Slide residue or wear material accumulated on CM/RT RING and HOT PLATE (occasional)
Stop MICROSLIDE INCUBATOR rotation
- Test point
- CM/RT Ring status
- Expected
- Ring stopped rotating
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Condition codes displayed after sensor disconnect
System displays condition codes following disconnection of auxiliary wash or waste bottle sensors during bottle replacement
Ranked causes
- 1. Sensor harness disconnected during auxiliary bottle replacement (common)
Verify sensor connections reconnected
- Test point
- Auxiliary wash bottle sensor (JS label) and auxiliary waste bottle sensor (WS label) harness connectors
- Expected
- Both sensor harnesses fully seated and connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuvette jammed in incubator patient safety
Cuvettes stuck inside incubator rows or behind heater block, preventing normal operation
Ranked causes
- 1. Cuvette positioned incorrectly or debris in incubator (common)
Locate jam position
- Test point
- Incubator rows and space behind heater block
- Expected
- Jammed cuvette visible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Dispense blade replacement caution
Procedure to reset and retract dispense blade during maintenance cycle. Part of proboscis assembly maintenance workflow.
Ranked causes
- 1. Routine dispense blade maintenance (common)
BLADE retracts when motor knob is turned clockwise
- Test point
- DISPENSE BLADE MOTOR knob
- Expected
- BLADE retracts after 5-6 turns
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERF Metering condition codes persist
ERF METERING displays condition codes that do not clear after routine maintenance or component replacement attempts
Ranked causes
- 1. ERF CARRIER ASSEMBLY obstruction or failure (common)
Visually inspect CARRIER ASSEMBLY for obstructions
- Test point
- CARRIER in Z-TRACK ASSEMBLY
- Expected
- Clear, no visible debris or damage
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERF metering MicroTip bent or damaged
MicroTip appears bent or damaged during visual inspection or operation
Ranked causes
- 1. MicroTip wear, bending, or damage (common)
Loosen tip sleeve by turning left and unscrewing completely
- Test point
- ERF METERING TIP SLEEVE
- Expected
- Sleeve loosened and removed from retainer
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERF Metering MicroTip degradation or improper installation
MicroTip not dispensing or metering samples correctly, or tip seal compromised
Ranked causes
- 1. MicroTip worn, contaminated, or inserted incorrectly (common)
Inspect TIP SLEEVE for contamination or damage
- Test point
- TIP SLEEVE after removal from carrier
- Expected
- Sleeve clean, no visible residue or cracks
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERF metering tubing damaged, blocked, or pinched
Tubing appears damaged, blocked by foreign material, or pinched during visual inspection of ERF metering tubing path
Ranked causes
- 1. ERF metering tubing wear, pinch, or blockage (common)
Visually inspect ERF metering tubing at tip retainer fitting for damage, blockage, or pinching
- Test point
- ERF METERING TIP RETAINER fitting
- Expected
- Tubing attached securely, undamaged, not blocked or pinched
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ERF reservoir low or empty
Electrolyte Reference Fluid level in reservoir is low or depleted
Ranked causes
- 1. Normal fluid consumption or reservoir depletion (common)
Touch Reagents > View ERF/IWF in system menu
- Test point
- System user interface
- Expected
- View ERF/IWF Fluids screen displayed
11 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fluid aspiration or dispensing incorrect or condition codes after cap attachment caution
System exhibits fluid handling errors or displays condition codes after quick disconnect cap replacement or reattachment
Ranked causes
- 1. Quick disconnect cap not fully seated on bottle cap (common)
Verify quick disconnect cap is fully attached
- Test point
- Quick disconnect cap silver tabs on UWR or liquid waste bottle cap
- Expected
- Silver tabs fully extended and cap snapped into place with audible or tactile click
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heartbeat LED flashing faster than once per second
After installing a POWER BOARD, the heartbeat LED flashes faster than once per second indicating an error condition
Ranked causes
- 1. Error in POWER BOARD (occasional)
Press Reset button above heartbeat LED
- Test point
- Reset button above heartbeat LED
- Expected
- Error condition cleared
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Heartbeat LED flashing slower than once per second
After installing a POWER BOARD, the heartbeat LED (D7 on LOW POWER DSP BOARDS, D6 on HIGH POWER DSP BOARDS) flashes slower than once per second
Ranked causes
- 1. Memory for POWER BOARD not loaded (common)
Verify heartbeat LED flashing rate
- Test point
- Heartbeat LED (D7 on LOW POWER DSP BOARDS, D6 on HIGH POWER DSP BOARDS)
- Expected
- LED flashes once every second
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaccurate fluid level reported
System displays incorrect liquid waste bottle level readings in the UWR and Waste system
Ranked causes
- 1. Ultrasonic transducer wet after cleaning (common)
Verify transducer is completely dry
- Test point
- Ultrasonic transducer in liquid handling tray
- Expected
- Transducer surface dry to touch, no visible moisture
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IWF reservoir replacement caution
Wash fluid immuno reagent (IWF) reservoir is depleted or out and requires replacement during reagent management.
Ranked causes
- 1. IWF fluid depleted or low (common)
Schedule IWF reservoir load
- Test point
- Reagents > View ERF/IWF dialog
- Expected
- Schedule Load button touched; countdown timer displayed
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard does not respond to a particular key
Specific keys on the keyboard do not respond when pressed
Ranked causes
- 1. Debris or contamination under key (common)
Visual inspection and cleaning of affected key
- Test point
- Keyboard surface and key mechanism
- Expected
- Debris removed, key moves freely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Liquid waste bottle full or nearly full patient safety
System Status screen reports liquid waste bottle is full or nearly full, requires emptying
Ranked causes
- 1. Normal accumulation of sample fluid in waste bottle during operation (common)
Confirm system status indicates waste bottle is full or nearly full
- Test point
- System Status screen
- Expected
- Full or nearly full indicator
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Luminometer Performance Test
Diagnostic test to verify luminometer is operating properly and meeting performance specifications.
Ranked causes
- 1. Routine performance verification or condition code indicates need for luminometer test (occasional)
Verify luminometer is at proper temperature
- Test point
- luminometer internal temperature
- Expected
- stable at operating temperature
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Luminometer Read Incubator Reference
- Luminometer Read Normal Sequence
- Manual MicroTip pack loading with unreadable barcode
- Manual MICROTIP PACK OPENER cannot move with MEDs
- Manual MicroWell pack loading with unreadable barcode
- Metering Leak and Hysteresis Test failure
- Microimmunoassay reagent hatch installation
- Microimmunoassay reagent supply cooler filter needs replacement
- Microimmunoassay reagent supply interior contamination or moisture
- MicroSensor fiber optics contaminated with cotton fibers
- MicroSlide Incubator sensors require cleaning
- MICROSLIDE INCUBATOR temperature out of range
- MicroTip assay imprecision
- MicroTip bent or damaged
- MicroTip pack opener cap stuck on spindle
- MICROTIP PACK OPENER will not respond after manual raising and reset attempt
- MicroTip reagent pack cap too tight for automatic opener
- MicroWell pack cannot open or close despite knob turning
- MicroWell pack opener knob does not turn
- Outdated and empty reagents removal
- Pad Reflectance test fails
- Pad Reflectance Test Range or Grand Mean out of specification
- Photometer lamp requires replacement
- PISTON CAP clogged
- PISTON CAP loose or detached
- PM RING SENSOR requires cleaning
- Power board error after reset
- Power board memory not loaded
- Power board replacement needed
- Pressure release sound when removing wash bottle cap
- PRIMARY TIP SEALER replacement required
- Proboscis/Piston Assembly replacement needed
- PROBOSCIS/PISTON ASSEMBLY residue or clogging
- Quality Control precision data review
- Reagent pack bubbles detected
- Registration rails jam clearance
- Repeated slide dispense condition code on same cartridge
- Sample information lost during tray removal
- Secondary Tip Sealer requires cleaning
- Signal Reagent dispense probes not dispensing
- Slide dispense blade worn or defective
- Slide jam in CM or RT DISCARD CHUTE
- Slide jam in PM RING
- Slide jam in WF SHUTTLE
- SLIDE SUPPLY ACCESS DOOR PIVOT SPRINGS worn or failed
- Slide Transport Blade Sensor contamination
- Slides not visible in dispense window
- System does not power on
- System startup takes longer than expected
- TIP LOCATOR not seated properly
- Tips jammed in microimmunoassay VersaTip tube
- Tips jammed in Sampling Center VersaTip Ring Load Area
- Tips jammed in Sampling Center VersaTip Tube
- Tray transport stuck or misaligned
- Uncapped MicroTip reagent pack located and requires recapping
- VersaTip hopper level exceeds maximum capacity
- VersaTip or MicroTip not advancing during Scrap Run Test
- VersaTip tips overfill in carousel
- VersaTips jammed in escapement delivery tube
- VersaTips nested in tip hopper
- WF metering leak and hysteresis test
- WF metering MicroTip replacement
- WF metering pump requires replacement
63 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Barcode? | System could not read pack bar code label | Barcode reading failure during reagent loading | warning |
PW8-103 | Luminometer reading failure - repeated failures | 3 successive luminometer performance test failures on any reading type | critical |
PW8-400 | Luminometer reference photon reading failure | Reference Photon reading fails performance test criteria | error |
PW8-401 | Luminometer reading variability failure | Photon or analog reading coefficient of variation exceeds 0.61% | error |
PW8-402 | Luminometer reference analog reading failure | Reference Analog reading fails performance test criteria | error |
PW8-403 | Luminometer well reading failure | Well Photon or Well Analog reading fails criteria during performance test | error |
PW8-404 | Luminometer reading consistency failure | Photon or analog reading fails consistency or variability criteria | error |
PW8-407 | Luminometer well analog reading out of tolerance | Well Analog reading deviates from expected range during performance test | error |
PW8-408 | Luminometer reference analog dark reading failure | Reference Analog dark reading fails performance test criteria | error |
PW8-409 | Luminometer LR Delta photon out of tolerance | LR Deltas (Photon) difference between Reference and Well readings exceeds 1% | warning |
PW8-410 | Luminometer LR Delta analog out of tolerance | LR Deltas (Analog) difference between Reference and Well readings exceeds 1% | warning |
PW8-413 | Luminometer well analog performance degradation | Well Analog reading shows poor performance during test | error |
PW8-414 | Luminometer reference analog LRS bias failure | Reference Analog LRS %Bias fails during performance test | error |
TT9-304 | VERSATIP SUPPLY - No Tip Delivered to S VTip Ring Load Position | Escapement or tip feed tube jam in red-taped side preventing delivery to sampling center versatip ring | error |
TT9-324 | VERSATIP SUPPLY - No Tip Delivered to uIA VTip Ring Load Position | Escapement or tip feed tube jam in green-taped side preventing delivery to microimmunoassay versatip ring | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI