Q Core Sapphire: troubleshooting and error codes
The ServiceForge knowledge base holds 177 documented symptoms for the Q Core Sapphire, with 216 ranked causes, 117 diagnostic checks with test points and expected values, and 72 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Bolus button not responding patient safety
Pump does not respond when bolus button or handle is pressed
Ranked causes
- 1. Treatment does not allow bolus administration (occasional)
- 2. Bolus handle connected to cradle instead of pump (common)
- 3. Bolus handle, cable, or connector broken or disconnected (common)
- 4. Bolus administration not available in current treatment state (occasional)
- 5. Bolus handle defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump is not charging while attached to mini cradle
Pump fails to charge when placed in the mini cradle on the Mounting System
Ranked causes
- 1. Mounting System power supply is defective (common)
- 2. AC input cord plugged improperly or defective (common)
- 3. DC output connector plugged improperly or defective (common)
- 4. Splitter socket is defective (occasional)
Verify AC power cord connection
- Test point
- AC input to Mounting System power supply
- Expected
- Cord properly seated and Power LED ON (blue light)
- Tools
- Visual inspection
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accuracy could not be verified
Flow Accuracy test fails to verify delivery accuracy
Ranked causes
- 1. Set not properly primed (common)
- 2. Fluid bag is empty (occasional)
- 3. Kinks in administration set (occasional)
Verify set is properly primed
- Test point
- Administration set
- Expected
- All air removed
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration verification test failure analysis
Identifying which of the 3 calibration tests (Occlusion Sensor, Air Detector, Flow Accuracy) failed and taking corrective action
Ranked causes
- 1. Occlusion Sensor test failed (unknown)
- 2. Air Detector test failed (unknown)
- 3. Flow Accuracy test failed (unknown)
Review failed test results display
- Test point
- Results screen Figure 4.29
- Expected
- Failed test arrow and number displayed outside scale on relevant side
Failed to connect to server
Application cannot connect to Q Core server even with correct username and password entered
Ranked causes
- 1. PC not connected to internet (common)
- 2. Firewall blocking communication to server (common)
- 3. Server connectivity issue or configuration error (occasional)
Verify PC internet connectivity
- Test point
- Network connection
- Expected
- Active internet connection to external servers
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow Accuracy test completes too quickly patient safety
Flow Accuracy test finishes before expected completion time; error message appears
Ranked causes
- 1. Flow Accuracy Container not in upright position (common)
- 2. Hydrophobic filter wet or saturated (common)
- 3. Flow Accuracy Container not full (common)
Verify container position
- Test point
- Flow Accuracy Container orientation
- Expected
- Container upright with hanger in use
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow Accuracy test not completed within expected time patient safety
Flow Accuracy test fails to complete within expected timeframe; error message appears
Ranked causes
- 1. Administration set not connected to Flow Accuracy Container (common)
- 2. Loose connections (common)
- 3. Leaks in container or cap (common)
Verify administration set connection
- Test point
- Administration set to Flow Accuracy Container
- Expected
- Administration set securely connected to container
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump disconnected during configuration file loading patient safety
Pump loses communication or power during loading of configuration file; pump will not be operable until loading completes
Ranked causes
- 1. Communication cable disconnected during loading process (common)
- 2. Pump disconnected from power supply during loading (common)
- 3. Communication interrupted for other reason during loading (occasional)
Verify communication cable connections during loading
- Test point
- Pump connector and computer RS-232 port
- Expected
- Cable firmly seated, connection maintained during entire loading
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump does not detect rising pressure between marks patient safety
Occlusion Sensor test fails; pump cannot detect pressure rise between test marks
Ranked causes
- 1. Administration set damaged (common)
- 2. Extension Set not connected correctly (common)
- 3. Air not inserted into system (common)
Inspect administration set for damage
- Test point
- Administration set
- Expected
- Set is undamaged with no cracks, leaks or kinks
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen has no backlight
Display screen is dark and cannot be read
Ranked causes
- 1. Lock screen or patient lockout active (common)
- 2. Backlight disabled in running state (common)
- 3. Backlight hardware failure or battery issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACT test failed patient safety
Certification test does not pass; test failure indicated
Ranked causes
- 1. Incorrect test technique used (common)
- 2. Faulty test equipment (occasional)
Review test procedure and technique
- Test point
- ACT test steps
- Expected
- Technician follows all procedure steps correctly
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air not detected by Air Detector
Air Detector fails to detect air introduced into the administration set during test
Ranked causes
- 1. Air not properly inserted into set (common)
- 2. Cassette chamber or light sensor dirty (occasional)
Verify air has been properly inserted into set
- Test point
- Administration set
- Expected
- Air present
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air not detected within 30 seconds during Air Detector test
Air Detector test fails with error; air not detected within expected timeframe
Ranked causes
- 1. Air not inserted into administration set (common)
- 2. Air detector sensor malfunction (occasional)
Verify air insertion in administration set
- Test point
- Administration set air chamber
- Expected
- Air bubble visible and detectable in administration set
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bolus button does not appear in toolbar during continuous infusion patient safety
Bolus control button absent from toolbar when running continuous infusion
Ranked causes
- 1. Allow Bolus feature not enabled (common)
- 2. Drug profile not configured to support bolus (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Connection with pump lost
Communication cable disconnected or faulty; no samples received from pump in 4 seconds
Ranked causes
- 1. Communication cable not properly connected (common)
- 2. Communication cable is faulty (occasional)
Verify communication cable connections
- Test point
- Pump connector and PC connector
- Expected
- Cable firmly seated in both connectors
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous air in line alarms without visible bubbles patient safety
Air in line alarm triggers repeatedly despite no visible air bubbles in tubing
Ranked causes
- 1. Cassette not fitted properly or small bubbles trapped in silicone tube (common)
- 2. Pump internal damage affecting air detection (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Detected air exceeded expected amount
Occlusion sensor test detects more air than anticipated
Ranked causes
- 1. Set not properly primed (common)
- 2. Stuck bubbles inside cassette tube (occasional)
Verify set is properly primed
- Test point
- Administration set
- Expected
- All air removed, properly primed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion alarms triggered without actual occlusion patient safety
Occlusion alarm activates when no blockage exists in delivery system
Ranked causes
- 1. Cassette not fitted properly in position (common)
- 2. Defective administration set or tubing (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occlusion sensor cannot detect expected pressure increase
Occlusion sensor fails to detect the expected rise in pressure during test
Ranked causes
- 1. Administration set is damaged (occasional)
- 2. Administration set was not clamped (common)
Verify administration set is clamped
- Test point
- Administration set clamp
- Expected
- Set properly clamped
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump disconnected during configuration file creation patient safety
Pump lost communication or power during configuration file creation process
Ranked causes
- 1. Communication cable disconnected from pump during creation (common)
- 2. Pump disconnected from power supply during creation (common)
Verify communication cable properly connected
- Test point
- Pump connector and PC RS-232 port
- Expected
- Cable firmly seated and maintained connection
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump is not charging
Battery fails to charge when connected to charger
Ranked causes
- 1. Charger disconnected from mini cradle or main power supply (common)
- 2. Charger malfunction or not functioning (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pump Mechanism Error (PME) patient safety
Family of alarms related to problems with the pump mechanism; may indicate pump setup issue or low battery.
Ranked causes
- 1. Pump mechanism failure (occasional)
- 2. Pump setup issue or low battery voltage (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recurring air in line alarms patient safety
Air in line alarm repeats despite corrective actions
Ranked causes
- 1. Treatment near end or air detection settings too sensitive (common)
- 2. Air detection sensitivity requires adjustment (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen saver does not appear
Screen saver feature not activating on display
Ranked causes
- 1. Screen Saver option not enabled (common)
- 2. Pump in state that prevents screen saver activation (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test failure due to water residue or occluded clamp
Certification test fails; water contains visible residues or clamp area is occluded
Ranked causes
- 1. Visible residues in test water (occasional)
- 2. Clamp area occluded (occasional)
Inspect water in administration set for residues
- Test point
- Test water in tubing
- Expected
- Clear with no visible residues
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Treatment cannot be started patient safety
Pump rejects treatment initiation or will not deliver infusion
Ranked causes
- 1. Incorrect treatment parameters (common)
- 2. Pump in low lock level or locked state (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
.NET Framework 4.0 installation required for Pump Loader
Computer prompts requiring .NET Framework 4.0 full installation during Pump Loader installation
Ranked causes
- 1. .NET Framework 4.0 not installed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
30 min to battery depletion
User has approximately 30 minutes of pump operation at current rate before battery depletes and pump shuts off.
Ranked causes
- 1. Normal operation - low battery warning (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accumulated Air in line
Accumulated bubble volume has reached user-defined limit within 15 minutes (counts only bubbles meeting minimum size threshold).
Ranked causes
- 1. Normal operation - accumulated air threshold reached (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accumulated air in line over limit patient safety
Alarm displayed indicating excessive air has accumulated in the administration line during infusion
Ranked causes
- 1. Air in administration line (common)
Observe Air in Line alarm on display
- Test point
- Pump touchscreen
- Expected
- Alarm Title: Air in Line with message about accumulated air
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACT installation error: .NET Framework 4.0 not found
Installation fails with message requiring .NET Framework 4.0 full installation
Ranked causes
- 1. .NET Framework 4.0 not installed (common)
Cancel current installation
- Test point
- .NET Framework installation dialog
- Expected
- Installation cancelled and dialog closed
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACT installation error: unsupported operating system
Installation fails with message indicating unsupported Windows version
Ranked causes
- 1. Operating system not supported (common)
Verify operating system version
- Test point
- Computer system properties
- Expected
- Windows XP, Windows 7 or newer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACT software stops during test
Certification test software unexpectedly stops or crashes during execution
Ranked causes
- 1. Software crash or unexpected termination (occasional)
Restart ACT software
- Test point
- ACT software application
- Expected
- Software launches without errors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ACT unexpected termination recovery
Restoring pump settings after ACT terminates unexpectedly, with parameters potentially changed
Ranked causes
- 1. ACT terminated unexpectedly (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Administration cassette door cannot close after insertion
Cassette housing door fails to snap shut after cassette placement
Ranked causes
- 1. Cassette not properly seated in housing (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Administration cassette door cannot close without cassette
Door fails to snap closed when cassette is not inserted
Ranked causes
- 1. Door latch or hinge hardware failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air detector Faulty caution
Air detector unable to make correct readings.
Ranked causes
- 1. Air detector malfunction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air in Line
Air bubble equal to or larger than user-defined threshold detected by light sensor.
Ranked causes
- 1. Normal operation - air detection (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED is blinking
Alarm indicator LED flashing
Ranked causes
- 1. Alarm event occurred (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LED is on
Alarm indicator LED illuminated solid
Ranked causes
- 1. Alarm event occurred (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AS not loaded/misplaced
Administration Cassette not loaded successfully or not loaded at all.
Ranked causes
- 1. Administration Cassette loading issue (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Depleted caution
Battery voltage has dropped below 6.8V; insufficient power to operate pump.
Ranked causes
- 1. Battery voltage below 6.8V (unknown)
Measure battery voltage
- Test point
- Battery terminal
- Expected
- Above 6.8V, >
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Depleted alarm patient safety
Level 1 alarm occurs; infusion stops immediately, pump shuts down within 3 minutes; auditory alarm sounds; soft keys available: Mute, Shutdown
Ranked causes
- 1. Battery depleted or disconnected from AC power (common)
Verify AC power connection
- Test point
- Power inlet
- Expected
- Pump connected to AC supply and powered on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Depleted alarm occurs patient safety
Pump will automatically shut down in 3 minutes. Infusion stops immediately if running. Soft keys available: Mute and Shutdown.
Ranked causes
- 1. Battery depleted (common)
Verify AC power connection
- Test point
- AC power inlet
- Expected
- Pump connected to AC supply
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery icon shows red line when not on power supply
Battery icon displays one red line after upgrade despite battery not being empty, when pump not connected to power supply
Ranked causes
- 1. Pump firmware initialization not complete (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Issue error message patient safety
System displays Battery Issue error message
Ranked causes
- 1. Battery fault or low charge (unknown)
Contact Q Core Service department
- Test point
- Service support
- Expected
- Support team provides guidance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery life expires 2 days caution
Battery will reach end of 2-year lifespan in 2 days.
Ranked causes
- 1. Battery lifespan approaching expiration (2 days remaining) (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Temperature out of range caution
Battery temperature has reached or exceeded 60°C critical threshold.
Ranked causes
- 1. Battery overheating (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Black screen on pump after software update completion
Pump screen remains black and does not display content after update process completes
Ranked causes
- 1. Software LCM update error or communication loss (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cable connected and COM port selected but pump not appearing
Pump does not appear in SCM pump list even though cable is connected and COM port is selected
Ranked causes
- 1. Delay in pump detection or appearance in list (occasional)
Wait for pump detection in SCM list
- Test point
- Pump list in SCM application
- Expected
- Pump serial number appears in the available pumps list
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration Due 2 days
Working time or calendar time has ended and calibration is due in 2 days.
Ranked causes
- 1. Calibration due in 2 days (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration Due 2 weeks
Working time or calendar time has ended and calibration is due in 2 weeks.
Ranked causes
- 1. Calibration due in 2 weeks (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration Due now
Working time or calendar time has ended; calibration may be needed.
Ranked causes
- 1. Calibration interval expired (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Screen
Screen values and options do not match user input or selection
Ranked causes
- 1. Screen calibration required (common)
Access Tech. options screen
- Test point
- Tech. options menu
- Expected
- Screen calibration option visible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot connect to COM port during station setup
COM port connection fails when configuring stations in Sapphire Configuration Manager
Ranked causes
- 1. COM port configured for different station or in use by another application (common)
Verify COM port is not in use
- Test point
- COM port status
- Expected
- Selected COM port available and not assigned to another station or application
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot Start Pump error message during Air Detector test
Air Detector test displays 'Cannot Start Pump' error message
Ranked causes
- 1. Test software error or improper initialization (occasional)
Close and reopen test software
- Test point
- ACT software
- Expected
- Software closes cleanly and reopens without errors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot stop treatment when pressing Stop button patient safety
Pump continues infusion despite activating stop control
Ranked causes
- 1. Stop button malfunction or software issue (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cassette misplaced or not loaded patient safety
Administration cassette is not properly installed or positioned in the pump
Ranked causes
- 1. Cassette not inserted or not seated correctly (common)
Verify cassette insertion and flange position
- Test point
- Safety door and cassette flanges
- Expected
- Both cassette flanges fully seated inside safety door
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Certificate printing failed; no certificate image background
Certificate printing process fails; certificate information prints without background image
Ranked causes
- 1. Printer or print driver issue (common)
Attempt certificate download from service portal
- Test point
- https://service.qcore.com/ Device Maintenance tab
- Expected
- Certificate downloads successfully
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Certificate printing process failed
Certificate fails to print with background image, information only prints without certificate background
Ranked causes
- 1. Printer driver or system issue (rare)
Attempt to download certificate from service portal
- Test point
- https://service.qcore.com/
- Expected
- Certificate successfully downloaded
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Certification due date discrepancy between certification and pump menu
- Certification software version out of date
- Charge error
- Charging LED is blinking
- Charging LED is on
- Check for Occlusion
- Check for occlusion in line
- Computer sleep mode interrupts software update
- Configuration file creation failed retry
- Configuration file loading failed
- Configuration file loading pump software version mismatch
- Connection lost with pump during certification
- Corrupt configuration file
- Could not log in with username and password
- Critical alarm triggered during certification process
- Display resolution incompatible on Windows 10
- Display resolution incompatible on Windows 7 or 8
- Door opened
- Dose calculation option missing after revision upgrade
- Downstream occlusion detected
- Downstream/Upstream Occlusion
- Drug Library compatibility during software update
- End of Battery Life
- Error Code 7 appears briefly during software update
- Error message prompted on screen during upgrade
- Excessive environmental light
- External Communication Error
- External Watchdog
- Failed to connect to Q Core server with username and password
- Failed to create configuration
- FasTest PM installation: .NET Framework 4.0 required
- FasTest PM software stops during test
- File name contains unsupported characters
- File name required error
- Fingers not cycling on startup after upgrade
- Flow error detected
- Graphic information missing from pump screen during certification
- Hard keys not responding
- Inconsistent flow
- Incorrect pressure value reading
- Installation failed before software update starts
- Installation Failed prompt appears
- Internal Communication Error
- Internal Error alarm
- Internal Error alarm occurs
- Invalid Kit Number error during certification
- Low voltage for current rate
- MedNet icon is gray
- No audible sound for alarms
- No audible sound when pressing keys or touch screen
- No communication between pump and computer
- No pump attached to station error when double-clicking station
- No User Logged In error when starting certification
- Non-Authorized Battery
- Occlusion alarm appears during Air Detector & Flow test
- Occlusion alarm triggered immediately after infusion or bolus start
- Occlusion in administration line
- Occlusion Sensor test performed incorrectly
- On/Off hard key not responding
- Password validation error during login
- Please Wait screen during pump update with computer failure display
- Possible Occlusion
- Potential air in line detected
- Preliminary Inspections verification
- PreSet Programs option does not appear on Start screen
- Prime button disabled on Start Up screen after upload
- Process logs saved locally, uploading to server later
- Programming cannot be completed
- Pump automatically locked when infusion starts
- Pump cannot be turned off
- Pump cannot be turned on
- Pump does not appear after cable connection and COM port selection
- Pump has new Drug Library pending upload
- Pump Loader crashes during upgrade
- Pump Loader fails to connect to Q Core server
- Pump screen does not turn on after upgrade
- Pump shuts down immediately after disconnecting from external power
- Pump software not compatible with SCM
- Pump software version obsolete for certification
- Pump stopped unexpectedly
- Pump stopped without pressure buildup during Flow Accuracy test
- Pump unattended
- Recurring occlusion alarms
- Run LED is blinking
- Sapphire Configuration Manager version out of date
- SCM automatically shut down after pump disconnection during loading
- Screen header appears in red during upgrade
- Serial number not displayed in View System menu after upgrade
- Service Q Core Logger Server failed to start
- Service Q Core Logger Server fails to start during certification software installation
- Set Delay option does not appear on Start screen
- Software execution failure message appears
- Software execution failure message on pump
- Software update fails with Please Wait screen
- Software version incompatible with device hardware
- Software version upgrade required
- Station appears inactive despite proper configuration
- Stop button not responding
- System Error
- System Temperature out of range
- Test not completed within 30 seconds after pressing Mark 1
- Testing: Hard keys
- Touch screen not responding
- Treatment complete
- Treatment near end
- Unauthorized Pump message appears
- Unknown Occlusion
- Unknown User error after login attempt
- Unsupported operating system for Pump Loader
- Username or password is incorrect
- Username validation error during login
- Version error Code 7 after configuration file loading
- Version error Code 7 on pump screen
- View current software version
- Watchdog Timer Reset
- Wi-Fi disconnected and firmware update required in software 14.50.0
- WiFi icon is gray
117 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
30 min to battery depletion | Approximately 30 minutes of pump operation remaining at current rate before battery depletes. | Battery voltage drops to 30-minute reserve level | info |
Accumulated Air in line | Accumulated bubble volume reached user-defined limit within 15 minutes. | Accumulated air volume ≥user limit in 15 min | info |
Air detector Faulty | Air detector unable to make correct readings | Air detector sensor failure | warning |
Air in Line | Accumulated air in line has exceeded the allowable limit | Pump detects air volume in administration line exceeding preset threshold during infusion | warning |
Air in Line | Air bubble detected by light sensor equal to or larger than user-defined threshold | Detected air bubble equal to or exceeding user threshold | info |
Annual certification date is overdue | Scheduled maintenance certification has expired | Annual certification date passed | warning |
Annual certification due in 2 days | Scheduled maintenance certification expires in 2 days | Annual certification date approaching | warning |
Annual certification due in 2 weeks | Scheduled maintenance certification expires in 2 weeks | Annual certification date approaching | info |
AS not loaded/misplaced | Administration Cassette not loaded successfully or not loaded at all. | Cassette detection failure | info |
Battery Charging Error | Error detected during battery charging process | Charging circuit fault or power supply failure | error |
Battery Depleted | Battery voltage below 6.8V and can no longer supply sufficient power | Battery voltage drops below 6.8V | error |
Battery Depleted | Pump battery has been depleted below operational threshold | Battery charge level falls to critical level during operation | warning |
Battery Depleted | Battery power is exhausted or pump is disconnected from AC supply | Battery charge depleted below operational threshold or AC power disconnected | critical |
Battery Issue | Battery fault or low charge condition detected | Battery failure or critical low charge | warning |
Battery life expires 2 days | Battery will reach 2-year life limit in 2 days | Battery age approaches 2-year maximum | warning |
Battery life will expire in 2 days | Battery end-of-life approaching | Battery age or charge cycles exceed service life | warning |
Battery Reminder | End of battery life | Battery reaches end of service life | warning |
Battery Temperature out of range | Battery reached or exceeded critical temperature of 60°C | Battery temperature reaches or exceeds 60°C | warning |
Calibration Due 2 days | Working time or calendar time for calibration has ended; due in 2 days. | Calibration interval expiring in 2 days | warning |
Calibration Due 2 weeks | Working time or calendar time for calibration has ended; due in 2 weeks. | Calibration interval expiring in 2 weeks | warning |
Calibration Due now | Working time or calendar time for calibration has ended. | Calibration interval expired | warning |
Cassette Misplaced | Administration cassette is not loaded or not properly positioned | Pump detects missing cassette or improper seating during initialization or operation | error |
Charge error | Pump taking longer than usual to charge; triggered after 7 hours of charge time. | Charge time exceeds 7 hours | warning |
Check for Occlusion | Possible occlusion detected in the administration line | Pump detects pressure or flow anomaly indicating potential line occlusion | warning |
Check for Occlusion | Occlusion detected at beginning of treatment while pump determines reference. | Occlusion detected during reference determination | info |
Door open | Safety door is not closed properly | Administration cassette door not fully secured | warning |
Door opened | Administration set door is open or not fully closed | Door position switch indicates open or partially open state | info |
Downstream Occlusion | Occlusion detected in the portion of the line downstream from the cassette | Pump detects excessive backpressure or flow resistance beyond the cassette during infusion | error |
Downstream/Upstream Occlusion | Occlusion detected either from pump to patient (downstream) or reservoir to pump (upstream) | Infusion line or reservoir line blockage detected | info |
End of Battery Life | Battery exceeded usage life of 2 years or 500 charge cycles | Battery in use for more than 2 years or reached 500 charge cycles | warning |
Excessive environmental light | Excessive environmental light detected in addition to air in line conditions | High ambient light levels affecting sensor readings | info |
External Communication Error | CPU lost communication with LCD module, or LCD module lost communication with CPU. | CPU-LCD communication failure | error |
External Watchdog | External watchdog lost communication with internal watchdog CPU. | Watchdog communication failure | critical |
Flow Error | Pump detected an error in medication flow rate or delivery | Pump cannot achieve or maintain expected flow rate during infusion | error |
Flow Error | Flow measurement or delivery error | Unable to deliver fluid at programmed rate | error |
Inconsistent flow | Pump not running at stable flow rate | Flow rate instability detected during infusion | error |
Incorrect pressure value reading | Pump unable to obtain correct pressure reading. | Pressure sensor failure | error |
Incorrect pressure value reading | Pressure sensor providing incorrect readings | Pressure sensor malfunction or fault | error |
Infusion complete | Programmed infusion volume has been delivered | Infusion reaches programmed total volume | info |
Infusion near end | Infusion approaching programmed completion volume | Remaining volume drops to near-completion threshold | info |
Insufficient Battery | Battery voltage insufficient for current infusion rate | Low battery voltage during operation | error |
Internal Communication Error | Pump lost communication with battery | Loss of communication between pump CPU and battery | warning |
Internal Error | Unspecified internal hardware or software malfunction detected | Internal diagnostic check detected anomaly during operation or startup | critical |
Key stuck | Physical button or keypad detected in pressed state | Key contact closure not released | critical |
Low Battery | 30 minutes remaining before battery depletion | Battery charge drops to critical level | warning |
Low voltage for current rate | Pump cannot reach current infusion rate; PWM at 140 with insufficient battery voltage. | Requested rate exceeds PWM 140 limit at current battery voltage | warning |
Mechanism Error | A pump fault has occurred | Pump mechanism failure | critical |
Non-Authorized Battery | Battery not authorized by Q Core for use with pump | Non-authorized battery installed | error |
Occlusion | Occlusion detected in the administration line | Pump detects excessive pressure or reduced flow indicating line obstruction | error |
PME | Pump Mechanism Error - family of alarms related to pump mechanism problems | Mechanism failure detected during operation | error |
Possible Downstream Occlusion | Infusion line may be blocked downstream of pump | Back-pressure detected during infusion | warning |
Possible Occlusion | Infusion started with clamps closed | Infusion started while clamps are in closed position | info |
Potential Air in Line | Pump has detected possible air in the administration line | Sensor detects potential air presence or environmental light interference during infusion | warning |
Pump Fault | A pump fault has occurred | Pump malfunction | critical |
Pump inactive | Pump has not been operating for specified duration | No infusion activity for specified minutes | info |
Pump Stopped | Infusion pump has halted operation | Pump stop command or automatic halt due to error condition | warning |
Pump unattended | Pump has been idle for user-set amount of time | Pump idle duration exceeds user-configured threshold | info |
Software execution failure! Please re-install software | Pump firmware update encountered interference and did not complete successfully | During or immediately after software upload to pump | error |
System Error | Group of system errors; may not be written to Event Log depending on cause. | System malfunction | critical |
System Temperature out of range | System temperature reached or exceeded 70°C critical threshold. | System temperature ≥70°C | critical |
System temperature out of range | Internal pump temperature is outside operating limits | Temperature sensor reading exceeds safe operating range | error |
System Temperature out of range | System reached critical temperature of 70°C | System temperature reaches or exceeds 70°C | critical |
The battery could not be fully charged | Battery charging failed or incomplete | Charging circuit or battery fault detected | error |
Treatment complete | Total volume to be infused (VTBI) has been fully delivered. | Infused volume equals VTBI | info |
Treatment near end | Less than xx minutes remaining before end of treatment. | Time remaining until VTBI <xx minutes | info |
Unauthorized Pump | Pump is running unsupported software version | Software revision is not 15 | error |
Unknown Occlusion | Occlusion detected but pump unable to determine if upstream or downstream. | Occlusion with indeterminate location | info |
Upstream Occlusion | Infusion line is blocked upstream of the pump | Obstruction in administration line or connections | warning |
Version error Code: 7 | Pump software version mismatch or incompatibility during configuration file loading | Configuration file loading fails due to software version mismatch | error |
Watchdog Timer Reset | External watchdog lost communication with pump main processing unit; internal watchdog commands reset | Loss of communication between external watchdog and main processing unit | critical |
WiFi disconnected | Wireless network connection lost | WiFi signal lost or connection dropped | warning |
WiFi disconnected - contact technician | Wireless network connection lost with unresolved error | WiFi connection failed and cannot auto-recover | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.