Puritan Bennett 840: troubleshooting and error codes

The ServiceForge knowledge base holds 222 documented symptoms for the Puritan Bennett 840, with 568 ranked causes, 314 diagnostic checks with test points and expected values, and 571 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Exhalation valve calibration fails

Calibration procedure does not complete successfully; error messages displayed during Exp Valve Calibration function execution.

Ranked causes

  1. 1. System leak, particularly expiratory filter leaks (common)
  2. 2. Exhalation valve and/or analog interface PCB out of specification (occasional)
  3. 3. Analog Interface PCB out of specification (occasional)
  4. 4. Gold standard circuit not connected or expiratory filter not locked (occasional)
  5. 5. Exhalation valve temperature out of range (rare)
  6. 6. Expiratory filter leaks (common)
  7. 7. Flow sensor calibration table checksum error (rare)
  8. 8. No air connected to ventilator (occasional)
First check

Verify ventilator warm-up time

Test point
Service mode
Expected
Ventilator has been in service mode for at least 10 minutes

3 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen zero flow check failed caution

Q1 oxygen flow sensor reads > 0.05 L/min during zero-flow check when supplies disconnected

Ranked causes

  1. 1. Gas supplies not disconnected (common)
  2. 2. Compressor still running (common)
  3. 3. Q2 O-ring leakage (affects Q1) (occasional)
  4. 4. PSOL2 internal leak (affects O2 circuit) (occasional)
  5. 5. PSOL2 seating improper (occasional)
  6. 6. Q1 sensor calibration drift (occasional)
  7. 7. PSOL2 failure (rare)
  8. 8. Q2 sensor failure (rare)
First check

Read Q1 oxygen flow sensor during zero-flow condition

Test point
Q1 (oxygen flow sensor)
Expected
≤0.05 L/min, ≤
Tools
EST system flow sensor reading

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air flow cannot establish at 120 L/min patient safety

Flow controller unable to establish and control air flow at 120 L/min during EST

Ranked causes

  1. 1. Air supply disconnected, restricted, or low pressure (common)
  2. 2. Air PSOL defective (occasional)
  3. 3. System leak or occlusion (occasional)
  4. 4. Flow sensor calibration out of specification (occasional)
  5. 5. Q2 air flow sensor defective (rare)
  6. 6. Inspiratory Electronics PCB defective (rare)
  7. 7. AI PCB defective (rare)
First check

Verify air supply connected and unrestricted

Test point
air inlet
Expected
connected and clear

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air zero flow check failed caution

Q2 air flow sensor reads > 0.05 L/min during zero-flow check when supplies disconnected

Ranked causes

  1. 1. Gas supplies not disconnected (common)
  2. 2. Compressor still running (common)
  3. 3. Q2 O-ring leakage (occasional)
  4. 5. PSOL2 seating improper (occasional)
  5. 6. Q2 sensor calibration drift (occasional)
  6. 7. PSOL2 internal fault (rare)
  7. 8. Q2 sensor failure (rare)
First check

Read Q2 air flow sensor during zero-flow condition

Test point
Q2 (air flow sensor)
Expected
≤0.05 L/min, ≤
Tools
EST system flow sensor reading

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Flow Sensor

Automatic calibration of air, oxygen, and exhalation flow sensors with atmospheric pressure transducer zeroing using external barometer reference

Ranked causes

  1. 1. Air offset out of range during calibration (unknown)
  2. 2. O2 offset out of range during calibration (unknown)
  3. 3. Cannot achieve minimum air flow (60 L/min) (unknown)
  4. 4. Cannot achieve minimum O2 flow (60 L/min) (unknown)
  5. 5. No air connected (unknown)
  6. 6. No O2 connected (unknown)
  7. 7. AC power not connected (unknown)
First check

Enter Service Mode

Test point
Service Mode menu
Expected
Service Mode accessible

7 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen flow cannot establish at 120 L/min patient safety

Flow controller unable to establish and control oxygen flow at 120 L/min during EST

Ranked causes

  1. 1. Oxygen supply disconnected, restricted, or low pressure (common)
  2. 2. Oxygen PSOL defective (occasional)
  3. 3. System leak or occlusion (occasional)
  4. 4. Flow sensor calibration out of specification (occasional)
  5. 5. Q1 oxygen flow sensor defective (rare)
  6. 6. Inspiratory Electronics PCB defective (rare)
  7. 7. AI PCB defective (rare)
First check

Verify oxygen supply connected and unrestricted

Test point
oxygen inlet
Expected
connected and clear

2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air PSOL command current out of range patient safety

PSOL2 command current exceeds acceptable range during air flow verification at 120, 60, 5, or 1 L/min

Ranked causes

  1. 1. Air supply disconnected or restricted (common)
  2. 2. Air supply quality poor (occasional)
  3. 3. System leaks or occlusions (occasional)
  4. 4. PSOL2 defective (occasional)
  5. 5. Flow sensor calibration needed (occasional)
  6. 6. Q1 or Q2 flow sensor defective (rare)
First check

Verify air supply connected and unrestricted

Test point
air inlet
Expected
connected and clear

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not charging after AC power reconnection high voltage

BPS not charging after AC power was reconnected during EST Battery Test.

Ranked causes

  1. 1. AC power not properly reconnected or AC supply is bad (common)
  2. 2. Faulty BPS PCB (occasional)
  3. 3. Defective battery pack (occasional)
  4. 4. Faulty power supply (occasional)
  5. 5. Defective BD CPU (rare)
  6. 6. Faulty AI PCB (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 flow sensor cross check failed patient safety

Q1 and Q3 readings are too far apart during oxygen flow verification at 120, 60, 5, and 1 L/min

Ranked causes

  1. 1. Oxygen supply disconnected or restricted (common)
  2. 2. Oxygen supply quality poor (occasional)
  3. 3. System leaks or occlusions (occasional)
  4. 4. PSOL1 defective (occasional)
  5. 5. Flow sensor calibration out of specification (occasional)
  6. 6. Q1 oxygen flow sensor defective (occasional)
First check

Verify oxygen supply connected and unrestricted

Test point
oxygen inlet
Expected
connected and clear

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 PSOL command current out of range patient safety

PSOL1 command current exceeds acceptable range during oxygen flow verification at 120, 60, 5, and 1 L/min

Ranked causes

  1. 1. Oxygen supply disconnected or restricted (common)
  2. 2. Oxygen supply quality poor (occasional)
  3. 3. System leaks or occlusions (occasional)
  4. 4. PSOL1 defective (occasional)
  5. 5. Flow sensor calibration needed (occasional)
  6. 6. Q1 or Q2 flow sensor defective (rare)
First check

Verify oxygen supply connected and unrestricted

Test point
oxygen inlet
Expected
connected and clear

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen flow sensor cross-check failed patient safety

SST Flow Sensor Test failure: Q1 and Q3 readings are too far apart at oxygen flows of 120, 60, 5, and 1 L/min

Ranked causes

  1. 1. System leak (common)
  2. 2. Safety valve open (occasional)
  3. 3. Flow sensor calibration drift (occasional)
  4. 4. Q1 oxygen flow sensor malfunction (occasional)
  5. 5. Q3 exhalation flow sensor malfunction (occasional)
  6. 6. AI PCB malfunction (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air PSOL excessive pressure buildup 50-100 cmH2O caution

Air PSOL forward leak detected via excessive pressure in blocked inspiratory module, pressure between 50 and 100 cmH2O

Ranked causes

  1. 1. Leaking Q2 O-rings (common)
  2. 2. Air supply disconnected or restricted (occasional)
  3. 3. PSOL2 seating issue (occasional)
  4. 4. Air PSOL internal leak (occasional)
  5. 5. AI PCB fault (rare)
First check

Measure inspiratory pressure with safety valve open and then closed

Test point
Inspiratory pressure sensor
Expected
50-100 cmH2O (indicating forward leak)
Tools
Pressure transducer

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air PSOL forward leak detected patient safety

Excessive pressure buildup (> 50 cmH2O but < 100 cmH2O) detected in blocked inspiratory module with air PSOL enabled

Ranked causes

  1. 1. Q2 O-ring leaking (common)
  2. 2. Air supply not connected or restricted (common)
  3. 3. Air PSOL2 not fully seated (occasional)
  4. 4. Air PSOL2 defective (occasional)
  5. 5. AI PCB defective (rare)
First check

Check Q2 O-rings for leaks

Test point
Q2 O-ring seals
Expected
dry, no visible leakage

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backup power supply voltage out of range

BPS voltage dropped below accepted level or dropped too quickly during discharge test

Ranked causes

  1. 1. Battery pack not fully charged before EST (common)
  2. 2. Battery pack defective (occasional)
  3. 3. BPS PCB control circuit failure (occasional)
  4. 4. Power supply unit failure (occasional)
  5. 5. AI PCB failure (rare)
First check

Check BPS LED charge status before EST

Test point
BPS LED indicator
Expected
Green LED lit (fully charged)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery backup not charging

BPS (Battery Power Supply) not charging after AC power reconnected

Ranked causes

  1. 1. AC power not properly reconnected (common)
  2. 2. BPS PCB charging circuit failure (occasional)
  3. 3. Battery pack defective (occasional)
  4. 4. Power supply unit failure (occasional)
  5. 5. BD CPU or AI PCB failure (rare)
First check

Verify AC power supply and connection

Test point
AC power cord, outlet
Expected
AC power good and connected

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery voltage drops during backup power test high voltage

During EST Battery Test, backup power supply (BPS) voltage dropped below accepted level or too quickly while discharging.

Ranked causes

  1. 1. Battery pack not fully charged (common)
  2. 2. Defective battery pack (occasional)
  3. 3. Faulty BPS PCB (occasional)
  4. 4. Faulty AI PCB (occasional)
  5. 5. Defective power supply (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation zero flow check failed caution

Q3 exhalation flow sensor reads > 0.1 L/min during zero-flow check when supplies disconnected

Ranked causes

  1. 1. Gas supplies not disconnected (common)
  2. 2. Compressor still running (common)
  3. 3. Q3 O-ring leakage (occasional)
  4. 4. Q3 sensor calibration drift (occasional)
  5. 5. Q3 sensor failure (rare)
First check

Read Q3 exhalation flow sensor during zero-flow condition

Test point
Q3 (exhalation flow sensor)
Expected
≤0.1 L/min, ≤
Tools
EST system flow sensor reading

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory and expiratory pressure cross-check failed at 10 cmH2O patient safety

Inspiratory and expiratory pressure transducer readings diverge beyond acceptable tolerance at 10 cmH2O test pressure.

Ranked causes

  1. 1. Patient circuit leak or occlusion (common)
  2. 2. Ventilator system leak or occlusion (common)
  3. 3. Inspiratory pressure transducer out of range (occasional)
  4. 4. Expiratory pressure transducer out of range (occasional)
  5. 5. AI PCB electronics failure (rare)
First check

Verify patient circuit has no leaks or occlusions

Test point
Patient circuit system
Expected
No leaks or occlusions

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory and expiratory pressure cross-check failed at 50 and 100 cmH2O patient safety

Inspiratory and expiratory pressure transducer readings diverge beyond acceptable tolerance at high pressure levels (50 and 100 cmH2O).

Ranked causes

  1. 1. Expiratory bacteria filter clogged or restricted (common)
  2. 2. System leak or occlusion (common)
  3. 3. Inspiratory pressure transducer out of range (occasional)
  4. 4. Expiratory pressure transducer out of range (occasional)
  5. 5. AI PCB electronics failure (rare)
First check

Inspect expiratory bacteria filter

Test point
Expiratory filter
Expected
Filter clean and unobstructed

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 zero flow check failed patient safety

Oxygen flow exceeds 0.153 L/min when PSOL1 is commanded to deliver 0 L/min (closed)

Ranked causes

  1. 1. Q1 O-ring leaking (common)
  2. 2. PSOL1 leaking (common)
  3. 3. PSOL1 defective (occasional)
  4. 4. Flow sensor calibration out of specification (occasional)
  5. 5. Q1 oxygen flow sensor defective (rare)
First check

Verify no leaks at Q1 O-rings

Test point
Q1 O-ring seals
Expected
dry, no visible leakage

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen flow sensor cross-check failure patient safety

O2 flow sensor readings (Q1) and exhalation sensor readings (Q3) diverge beyond acceptable tolerance during oxygen flow test.

Ranked causes

  1. 1. System leak or occlusion (common)
  2. 2. Safety valve stuck open (occasional)
  3. 3. Q1 sensor malfunction (occasional)
  4. 4. Q3 sensor malfunction (occasional)
  5. 5. AI PCB electronics failure (rare)
First check

Check for system leak

Test point
System circuit
Expected
No leaks detected

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen PSOL current out of range patient safety

Oxygen proportional solenoid (PSOL1) command current is outside acceptable range during diagnostic test.

Ranked causes

  1. 1. System leak or occlusion (common)
  2. 2. Safety valve stuck open (occasional)
  3. 3. Oxygen regulator pressure out of specification (occasional)
  4. 4. PSOL1 solenoid valve malfunction (occasional)
  5. 5. AI PCB electronics failure (rare)
First check

Check for system leak

Test point
System circuit
Expected
No leaks detected

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen PSOL excessive pressure buildup 50-100 cmH2O caution

Oxygen PSOL forward leak detected via excessive pressure in blocked inspiratory module, pressure between 50 and 100 cmH2O

Ranked causes

  1. 1. Leaking Q1 O-rings (common)
  2. 2. Oxygen supply disconnected or restricted (occasional)
  3. 3. PSOL1 seating issue (occasional)
  4. 4. Oxygen PSOL internal leak (occasional)
  5. 5. AI PCB fault (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SST flow sensor test fails - unable to establish oxygen flow patient safety

Flow controller unable to establish and control oxygen flow at 120 L/min during SST test

Ranked causes

  1. 1. Oxygen supply not connected or unrestricted (common)
  2. 2. Patient circuit system has leaks or occlusions (common)
  3. 3. PSOL (proportional solenoid) malfunction (occasional)
  4. 4. Q1 or Q2 flow control valve malfunction (occasional)
  5. 5. Flow sensor calibration error (occasional)
First check

Verify oxygen supply connection and pressure

Test point
PS1 pressure switch, oxygen regulator output
Expected
9-12 psi at regulator; PS1 detects pressure

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Test circuit not connected caution

Pressure not detected on expiratory side during leak test; inspiratory/expiratory pressure transducer readings too far apart

Ranked causes

  1. 1. Test circuit not properly connected (common)
  2. 2. Expiratory bacteria filter blocked or defective (occasional)
  3. 3. Q3 flow sensor O-ring leakage (occasional)
  4. 4. Exhalation valve malfunction (occasional)
  5. 5. Expiratory pressure transducer PCB failure (rare)
First check

Compare inspiratory vs expiratory pressure readings during pressure hold

Test point
Inspiratory pressure sensor vs expiratory pressure sensor
Expected
Pressures closely match
Tools
Pressure transducers, EST system

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Test circuit not connected (pressure transducer readings far apart)

Pressure not detected on expiratory side; inspiratory/expiratory pressure transducer readings too far apart

Ranked causes

  1. 1. Test circuit not properly connected (common)
  2. 2. Expiratory bacteria filter clogged (occasional)
  3. 3. Leaks at Q3 flow sensor (occasional)
  4. 4. Exhalation valve malfunction (occasional)
  5. 5. Expiratory pressure transducer PCB failure (rare)
First check

Verify test circuit connection to expiratory side

Expected
Pressure detected on expiratory transducer

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to establish air flow at 120 L/min patient safety

Flow controller cannot establish and control air flow at 120 L/min during diagnostic test.

Ranked causes

  1. 1. Air supply disconnected or restricted (common)
  2. 2. Patient circuit leaks or occlusions (common)
  3. 3. Air regulator pressure out of range (occasional)
  4. 4. PSOL2 solenoid malfunction (occasional)
  5. 5. Q2 air sensor malfunction (occasional)
First check

Verify air supply connection and restriction

Test point
Air inlet
Expected
Connected and unrestricted

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to establish air flow at 60, 5, and 1 L/min patient safety

Flow controller cannot establish and control air flow at lower rates (60, 5, 1 L/min) during diagnostic test.

Ranked causes

  1. 1. Air supply disconnected or restricted (common)
  2. 2. System leaks or occlusions (common)
  3. 3. Air regulator pressure out of specification (occasional)
  4. 4. PSOL2 solenoid malfunction (occasional)
  5. 5. Q3 exhalation sensor malfunction (occasional)
First check

Verify air supply connection

Test point
Air inlet
Expected
Connected and unrestricted

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to establish oxygen flow patient safety

Flow controller cannot establish and control oxygen flow at 60, 5, and 1 L/min during diagnostic test.

Ranked causes

  1. 1. Oxygen supply disconnected or restricted (common)
  2. 2. System leaks or occlusions (common)
  3. 3. PSOL1 solenoid valve malfunction (occasional)
  4. 4. Q1 oxygen sensor malfunction (occasional)
  5. 5. Q3 exhalation sensor malfunction (occasional)
First check

Verify oxygen supply connection and restriction

Test point
Oxygen inlet
Expected
Connected and unrestricted

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to establish oxygen flow at 120 L/min

Flow controller unable to establish and control oxygen flow at 120 L/min during SST flow sensor test.

Ranked causes

  1. 1. Oxygen supply disconnected, restricted, or low pressure (common)
  2. 2. Patient circuit system has leaks or occlusions (common)
  3. 3. Oxygen inlet filter assembly blocked (occasional)
  4. 4. Defective flow control solenoid Q1 (occasional)
  5. 5. Defective PSOL1 oxygen solenoid (occasional)
First check

Verify oxygen supply is connected and flow is good

Test point
oxygen inlet
Expected
oxygen supply connected and unrestricted

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to establish oxygen flow at 60, 5, and 1 L/min patient safety

SST Flow Sensor Test failure: flow controller cannot establish and control oxygen flow at 60, 5, and 1 L/min (lower flows)

Ranked causes

  1. 1. Oxygen supply restricted or disconnected (common)
  2. 2. System leak or occlusion (occasional)
  3. 3. PSOL1 malfunction (occasional)
  4. 4. Flow sensor Q1 malfunction (occasional)
  5. 5. Exhalation sensor Q3 malfunction (rare)
First check

Verify oxygen supply

Test point
Oxygen supply connection and regulated pressure
Expected
Connected and unrestricted; adequate supply

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vent Inop Test fails

Service mode test procedure fails with 'Vent Inop Test failed' message on lower screen during BDU alarm, LED, or PSOL/valve inactivation checks

Ranked causes

  1. 1. BDU alarm failure (GUI vent inop test phase) (unknown)
  2. 2. Ventilator inoperative or safety valve open LED failure (unknown)
  3. 3. BDU alarm and LED failures combined (unknown)
  4. 4. PSOL/exhalation valve/safety valve inactivation failure (unknown)
  5. 5. Failure in Vent Inop A, B, or ten-second test phases (unknown)
First check

Verify BDU alarm operation in GUI vent inop test phase

Test point
BDU alarm
Expected
BDU alarm on (user prompt)

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air PSOL current out of range patient safety

Air proportional solenoid (PSOL2) command current is outside acceptable range during diagnostic test.

Ranked causes

  1. 1. Air supply disconnected or restricted (common)
  2. 3. Air supply pressure out of specification (occasional)
  3. 4. PSOL2 solenoid valve malfunction (occasional)
  4. 5. AI PCB electronics failure (rare)
First check

Verify air supply connection and restriction

Test point
Air inlet
Expected
Connected and unrestricted

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery backup not discharging

BPS not discharging when loaded during test phase

Ranked causes

  1. 1. AC power not disconnected during test (common)
  2. 2. BPS PCB discharge circuit failure (occasional)
  3. 3. Battery pack defective or depleted (occasional)
  4. 4. BD CPU or AI PCB failure (rare)
First check

Verify AC power is disconnected

Test point
AC power cord
Expected
AC power disconnected

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not fully charged

BPS not fully charged before EST begins

Ranked causes

  1. 1. Charging not allowed to complete (common)
  2. 2. BPS PCB charging circuit defective (occasional)
  3. 3. Battery pack defective (occasional)
  4. 4. BD CPU failure (rare)
First check

Verify AC power connected and BPS charging

Test point
AC power cord, BPS LED
Expected
AC connected, BPS LED indicates charging

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

BD alarm cable voltage out of range

Alarm cable voltage is < 3.5 V or > 5.05 V

Ranked causes

  1. 1. BD alarm cable not connected (common)
  2. 2. BD alarm malfunction (occasional)
  3. 3. AI PCB malfunction (occasional)
  4. 4. BD alarm cable damaged (occasional)
First check

Measure alarm cable voltage

Expected
3.5 to 5.05 V, 3.5-5.05 V
Tools
DMM

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Initialize Flow Sensor

Automatic copying of flow sensor calibration data from each sensor's serial EEPROM to BDU NOVRAM when data mismatch detected

Ranked causes

  1. 1. Unable to read air flow sensor serial EEPROM (unknown)
  2. 2. Unable to read O2 flow sensor serial EEPROM (unknown)
  3. 3. Unable to read exhalation flow sensor serial EEPROM (unknown)
  4. 4. Unable to program BDU NOVRAM (unknown)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compliance low (below 1.56 adult / 1.34 pediatric, second occurrence)

Low compliance falls below 1.56 mL/cmH2O (adult) or 1.34 mL/cmH2O (pediatric) but not below 1.05 mL/cmH2O during SST Compliance calibration test.

Ranked causes

  1. 1. Wrong patient tubing type specified in SST (common)
  2. 2. Defective patient circuit or filter (occasional)
  3. 3. Defective pressure transducer (PI or PE) (occasional)
  4. 4. Atmospheric pressure transducer out of calibration (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compliance low (below 1.56 adult / 1.34 pediatric)

Compliance falls below 1.56 mL/cmH2O (adult) or 1.34 mL/cmH2O (pediatric) but not lower than 1.05 mL/cmH2O during SST Compliance calibration test.

Ranked causes

  1. 1. Wrong patient tubing type specified in SST (common)
  2. 2. Defective patient circuit or filter (occasional)
  3. 3. Defective pressure transducer (PI or PE) (occasional)
  4. 4. Atmospheric pressure transducer out of calibration (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive compliance (>6 mL/cmH2O adult / 4.5 pediatric but <12/9)

Low compliance >6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric) but <12 mL/cmH2O (adult) or 9 mL/cmH2O (pediatric) during SST Compliance calibration test.

Ranked causes

  1. 1. Wrong patient tubing type specified in SST (common)
  2. 2. Defective patient circuit or filter (occasional)
  3. 3. Defective pressure transducer (PI or PE) (occasional)
  4. 4. Atmospheric pressure transducer out of calibration (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive system leak detected patient safety

System pressure drops excessively during leak test, indicating significant leak or occlusion in the patient circuit or system.

Ranked causes

  1. 1. Patient circuit not connected (common)
  2. 2. Patient circuit leak (common)
  3. 3. Exhalation valve malfunction (occasional)
  4. 4. Expiratory filter connection loose (occasional)
First check

Verify test circuit is connected

Test point
Test circuit connector
Expected
Circuit properly connected and seated

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Excessive system leak during leak test caution

System pressure dropped below 85 cmH2O or pressure drops ≥5 cmH2O within 10 seconds after reaching 85 cmH2O

Ranked causes

  1. 1. Test circuit not connected or leaking (common)
  2. 2. System internal leaks (common)
  3. 3. Expiratory filter connection loose (occasional)
  4. 4. Exhalation valve malfunction (occasional)
First check

Wait for inspiratory pressure decay from 90 cmH2O or wait 10 seconds

Test point
Inspiratory pressure sensor
Expected
Pressure drops to 85 cmH2O or stable
Tools
Pressure transducer, EST system

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation valve seal leaking caution

Exhalation valve seal test shows pressure differential too high, indicating valve diaphragm not sealing properly

Ranked causes

  1. 1. Exhalation valve diaphragm contaminated (common)
  2. 2. Valve calibration incorrect (common)
  3. 3. Flow sensor Q3 miscalibrated (occasional)
  4. 4. Exhalation valve diaphragm damaged (occasional)
First check

Clean exhalation valve diaphragm

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow Sensor Calibration

Calibration of exhalation flow sensor (Q3) offset table using gold standard test circuit and dual gas sources (air and oxygen)

Ranked causes

  1. 1. Test circuit not attached or expiratory filter not secure (common)
  2. 2. System leak or expiratory filter leak (common)
  3. 3. Ventilator not at operating temperature (occasional)
  4. 4. Ventilator failed EST leak test (occasional)
First check

Verify ventilator has reached operating temperature

Test point
Ventilator thermal status
Expected
Operating temperature reached after 10 minutes warm-up in service mode

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flow sensor calibration fails

Flow Sensor Calibration procedure does not complete; error messages displayed during calibration execution.

Ranked causes

  1. 1. System leak, particularly expiratory filter leaks (common)
  2. 2. Test circuit not attached or not secure (occasional)
  3. 3. Ventilator has not reached operating temperature (occasional)
  4. 4. Flow sensor not passing EST leak test (occasional)
First check

Verify ventilator warm-up time

Test point
Service mode operation
Expected
Ventilator has operated in service mode for at least 10 minutes

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Initialize Flow Sensor errors

Automatic flow sensor calibration data initialization fails with error messages during service mode startup

Ranked causes

  1. 1. Unable to read air flow sensor (Q2) EEPROM (unknown)
  2. 2. Unable to read oxygen flow sensor (Q1) EEPROM (unknown)
  3. 3. Unable to read exhalation flow sensor (Q3) EEPROM (unknown)
  4. 4. Unable to program BDU NOVRAM flash memory (unknown)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory and expiratory pressure transducer readings too far apart patient safety

Cross-check between inspiratory and expiratory pressure transducers at 50 and 100 cmH2O fails; readings diverge beyond acceptable tolerance

Ranked causes

  1. 1. Expired bacteria filter clogged or defective (common)
  2. 2. System leaks or occlusions (common)
  3. 3. Inspiratory or expiratory pressure transducer out of range (occasional)
  4. 4. AI PCB defective (occasional)
First check

Verify no system leaks or occlusions

Test point
circuit
Expected
no leaks or occlusions

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inspiratory limb resistance low

Inspiratory limb pressure <0.6 cmH2O (adult) or 5.5 cmH2O (pediatric) but >0.2 cmH2O during SST Circuit Resistance test.

Ranked causes

  1. 1. Inspiratory filter not installed (common)
  2. 2. Test procedure not followed correctly (occasional)
  3. 3. Defective inspiratory filter (occasional)
  4. 4. Defective patient circuit (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen PSOL forward leak detected (pressure 50-100 cmH2O) caution

Excessive pressure buildup (> 50 cmH2O but < 100 cmH2O) detected in blocked inspiratory module

Ranked causes

  1. 1. Oxygen PSOL forward leak (common)
  2. 2. Leaks around Q1 O-rings (occasional)
  3. 3. Restricted oxygen supply (occasional)
  4. 4. AI PCB malfunction (rare)
First check

Block inspiratory module and measure pressure

Expected
Pressure between 50-100 cmH2O indicates forward leak

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power failure capacitor voltage out of range caution

Initial power failure capacitor voltage is < 4.5 V or > 5.05 V

Ranked causes

  1. 1. BD alarm cable not connected (common)
  2. 2. BD alarm malfunction (occasional)
  3. 3. AI PCB malfunction (occasional)
  4. 4. Motherboard PCB failure (rare)
First check

Measure capacitor voltage

Expected
4.5 to 5.05 V, 4.5-5.05 V
Tools
DMM

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure transducer cross-check failed at 10 cmH2O patient safety

SST Circuit Pressure Test failure: inspiratory and expiratory pressure transducer readings are too far apart at 10 cmH2O

Ranked causes

  1. 1. Patient circuit leak or occlusion (common)
  2. 2. System leak or occlusion (occasional)
  3. 3. Pressure transducer out of range (occasional)
  4. 4. AI PCB malfunction (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressure transducer cross-check failed at 50 or 100 cmH2O patient safety

SST Circuit Pressure Test failure: inspiratory and expiratory pressure transducer readings are too far apart at 50 or 100 cmH2O

Ranked causes

  1. 1. Expiratory bacteria filter clogged (common)
  2. 2. System leak or occlusion (occasional)
  3. 3. Pressure transducer out of range (occasional)
  4. 4. AI PCB malfunction (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Safety valve pressure relief failed patient safety

System unable to establish stable flow or safety valve cracking pressure is out of specification

Ranked causes

  1. 1. To patient port not blocked during test (common)
  2. 2. Leak at O2 sensor, SOL1, or PI (occasional)
  3. 3. Safety valve defective (occasional)
  4. 4. PI pressure transducer defective (occasional)
First check

Verify To patient port is blocked

Test point
To patient port
Expected
blocked

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to establish leak test pressure caution

System cannot attain 90 cmH2O starting pressure during leak test within timeout period with either oxygen or air

Ranked causes

  1. 1. Test circuit not connected (common)
  2. 2. System leaks present (common)
  3. 3. Exhalation valve malfunction (occasional)
  4. 4. Expiratory filter connection loose (occasional)
First check

Pressurize system to 90 cmH2O using 2 L/min flow with oxygen or air

Test point
Inspiratory pressure sensor
Expected
≥90 cmH2O, ≥
Tools
Pressure transducer, EST system

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to establish test pressure

System cannot attain 90 cmH2O pressure with oxygen or air at 2 L/min flow within timeout

Ranked causes

  1. 1. Test circuit not connected (common)
  2. 2. System leaks present (common)
  3. 3. Exhalation valve not functioning (occasional)
  4. 4. Expiratory filter not securely connected (occasional)
First check

Verify test circuit connection

Expected
Test circuit properly connected to system

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vent Inop Test failure during GUI vent inop test phase

GUI vent inop test phase fails during Vent Inop Test execution; BDU alarm, LED indicators, or PSOLs/exhalation valve/safety valve may not function correctly

Ranked causes

  1. 1. BDU alarm failure (occasional)
  2. 2. Ventilator inoperative or SVO LED failure (occasional)
  3. 3. BDU-GUI cable or CPU PCB communication failure (occasional)
  4. 4. BD CPU PCB or AI PCB failure (occasional)
First check

Observe BDU alarm during GUI vent inop test phase

Test point
BDU alarm output
Expected
Alarm activates on user prompt

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power not connected for EST patient safety

EST (Equipment System Test) cannot run on battery power; AC power required

Ranked causes

  1. 1. Ventilator not plugged into AC outlet (common)
  2. 2. Battery Power Supply (BPS) active (occasional)
  3. 3. Power supply defective (rare)
First check

Verify AC power connection

Test point
Ventilator power inlet
Expected
AC power cord connected to facility outlet

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air flow cannot establish at 60, 5, and 1 L/min patient safety

Flow controller unable to establish and control air flow at lower rates during EST

Ranked causes

  1. 1. Air supply disconnected or low pressure (common)
  2. 2. PSOL2 or Q2 air flow sensor defective (occasional)
  3. 3. AI PCB defective (rare)
First check

Verify air supply connected

Test point
air inlet
Expected
connected

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air flow sensor cross-check failed patient safety

Air flow sensor (Q2) and exhalation sensor (Q3) readings diverge beyond acceptable tolerance during air flow test.

Ranked causes

  1. 2. Ventilator system leak or occlusion (common)
  2. 4. PSOL2 solenoid malfunction (occasional)
  3. 5. Q2 air sensor malfunction (occasional)
First check

Inspect patient circuit for leaks or occlusions

Test point
Patient circuit
Expected
No leaks or occlusions; no debris

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Air PSOL forward leak at high pressure patient safety

Excessive pressure buildup (> 100 cmH2O) detected in blocked inspiratory module with air PSOL enabled

Ranked causes

  1. 1. Q2 O-ring leaking (common)
  2. 2. Safety valve not opening at correct pressure (occasional)
  3. 3. Air PSOL2 defective (occasional)
First check

Verify safety valve cracks above 100 cmH2O

Test point
safety valve outlet
Expected
relief heard at pressure > 100 cmH2O on display

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Atmospheric Pressure Transducer calibration out of range
  • Battery pack system error
  • BD communication loss
  • Compressor load test failed
  • Compressor pressure not detected
  • Excessive circuit leak detected
  • Excessive leak in system (pressure drops >= 5 cmH2O in 10 seconds)
  • Excessive system leak
  • Exhalation valve temperature out of specification
  • Expiratory pressure transducer autozero out of range
  • Expiratory pressure transducer autozero solenoid SOL2 bad
  • Failed to reach test pressure
  • Failed to reach test pressure (10 cmH2O at 5 L/min)
  • GUI alarm sound not working
  • Inspiratory pressure transducer autozero out of range
  • Inspiratory pressure transducer autozero solenoid SOL1 bad
  • Inspiratory/expiratory pressure transducer cross-check failed
  • LCD controller PCB fault detected
  • Low circuit compliance
  • Low oxygen percentage (↓O2%)
  • O2 pressure detected with supply disconnected
  • Oxygen flow cannot establish at 60, 5, and 1 L/min
  • Oxygen pressure detected when disconnected
  • Power supply voltage error
  • Touch screen input failure
  • Touchscreen error occurred
  • Wall air pressure detected when disconnected
  • Wall air pressure detected with supply disconnected
  • AC power loss
  • Air PSOL excessive pressure buildup >100 cmH2O
  • Bad expiratory autozero solenoid
  • Bad inspiratory autozero solenoid
  • Cal Info Duplication errors
  • Calibration: Cal Info Duplication
  • Calibration: Power on self test (POST)
  • Calibration: Serial Number Setup
  • Circuit disconnect
  • Compressor leak detected
  • Compressor run time out of range
  • Excessive circuit compliance
  • Excessive compliance (high)
  • Expiratory autozero out of range
  • Failed to reach test pressure (50 or 100 cmH2O)
  • Failed to reach test pressure at 10 cmH2O
  • Failed to reach test pressure at 50 and 100 cmH2O
  • GUI 100% O2 LED fails
  • GUI 100% O2/CAL key fails
  • GUI Accept key fails
  • GUI Alarm Reset key fails
  • GUI Alarm Silence key fails
  • GUI Alarm Silence LED fails
  • GUI Alarm Volume key fails
  • GUI Battery Backup LED fails
  • GUI Clear key fails
  • GUI Compressor Operating LED fails
  • GUI Compressor Ready LED fails
  • GUI CPU not operational
  • GUI Expiratory Pause key fails
  • GUI High Alarm LED fails
  • GUI Info key fails
  • GUI Inspiratory Pause key fails
  • GUI keyboard key malfunction
  • GUI knob malfunction
  • GUI knob not responding
  • GUI LED indicator not illuminated
  • GUI Low Alarm LED fails
  • GUI Manual Inspiration key fails
  • GUI Medium Alarm LED fails
  • GUI Normal LED fails
  • GUI On Battery Power LED fails
  • GUI Screen brightness key fails
  • GUI Screen contrast key fails
  • GUI Screen lock key fails
  • GUI Screen Lock LED fails
  • High tidal volume (ΔVE TOT)
  • Low AC power
  • Low delivered tidal volume (ΔVE MAND)
  • Low expiratory filter pressure drop
  • Low inspiratory limb resistance
  • No air supply
  • Nurse call relay stuck off
  • Nurse call relay stuck on
  • O2 pressure not detected
  • Occluded exhalation compartment
  • Occluded expiratory compartment
  • Oxygen pressure not detected
  • Oxygen PSOL excessive pressure buildup >100 cmH2O
  • Oxygen PSOL forward leak detected (pressure > 100 cmH2O)
  • Wall air pressure not detected
  • Wall air pressure still detected after disconnect
  • Watchdog timer failure
  • AC power loss detected
  • Air pressure not detected via PS2 or PC
  • Air supply loss with O2 at 100%
  • Assertion boundary check failure in background diagnostics
  • Bad GUI serial port
  • Bus error or access fault
  • Calibration error during Flow Sensor Calibration
  • Calibration: Datakey Update
  • Calibration: Exhalation system
  • Calibration: Expiratory valve
  • Calibration: O2 sensor
  • Calibration: Oxygen sensor
  • Calibration: Performance verification
  • Calibration: Post-service verification
  • Calibration: Regulator calibration hose repair
  • Calibration: Remove or install 10.4-inch GUI
  • Calibration: Serial Loopback Test
  • Calibration: System Leak and Compliance Characterization (SST)
  • Cannot access service mode
  • Cannot program flash error
  • Compressor disabled mode time out of range
  • Compressor pressure detected during EST
  • Compressor pressure detected when depressurized
  • Compressor pressure detected when disabled
  • Compressor standby mode time out of range
  • Datakey not installed or not detected
  • Datakey update procedure
  • Electrical safety testing
  • Exhalation valve leak or EST failure
  • Expiratory limb resistance low
  • Gas not connected
  • GUI cable connection verification
  • High oxygen percentage (↑O2%)
  • Inspiratory autozero out of range
  • Loss of GUI LED fails
  • Low battery warning during AC loss
  • Low exhalation limb resistance
  • No O2 supply
  • No ventilation with safety valve open
  • O2 sensor out of calibration or failure
  • O2 supply loss with FiO2 at 21%
  • Occluded exhalation limb
  • Occluded expiratory filter
  • Occluded inspiratory limb
  • Oxygen pressure not detected via PS1
  • Performance verification testing required
  • Performance verification: calibration and testing
  • POST major fault detected; ventilator inoperative
  • POST minor fault detected during self-test
  • Procedure error during patient setup
  • Replace 10.4-inch GUI front housing
  • Replace 10.4-inch GUI keyboard assembly
  • Screen block or touch screen fault
  • Serial number does not match datakey
  • Service mode functions not executing
  • Severe occlusion in ventilator circuit
  • SST flow sensor test fails - air only available
  • SST flow sensor test fails - no gas available
  • SST flow sensor test fails - oxygen only available
  • SVO LED fails
  • System unable to establish pressure
  • Total loss of power
  • Troubleshooting ventilator diagnostic codes
  • Unable to establish stable expiratory filter flow
  • Unable to pressurize circuit during compliance calibration
  • Unable to reach minimum peak flow
  • Unable to reach minimum peak flow (alert threshold)
  • Unable to reach minimum peak flow (threshold 80-60 L/min)
  • Ventilator inoperative LED fails
  • Volume not delivered on VC+ or VS breaths
  • Wye not blocked during circuit resistance test

162 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0BAlarm Cable Voltage test failurePOST analog device test failureerror
0CO2 PSOL Current test failurePOST analog device test failureerror
0DAir PSOL Current test failurePOST analog device test failureerror
0ELow Voltage Reference test failurePOST analog device test failureerror
0FAtmospheric Pressure Transducer test failurePOST analog device test failureerror
10Exhalation Coil Temperature test failurePOST analog device test failureerror
11Exhalation Pressure Transducer test failurePOST analog device test failureerror
12O2 Sensor test failurePOST analog device test failureerror
12 Vdc Overvoltage12V power supply overvoltage (GUI only)Excessive 12V outputerror
12 Vdc Undervoltage12V power supply undervoltage (GUI only)Insufficient 12V outputerror
13GUI 5V dc Sentry test failurePOST analog device test failureerror
1412V dc Sentry test failurePOST analog device test failureerror
15Safety Valve Current test failurePOST analog device test failureerror
16+15V Sentry test failurePOST analog device test failureerror
17-15V Sentry test failurePOST analog device test failureerror
18Power Fail Cap Voltage test failurePOST analog device test failureerror
19Exhalation Manifold Heater Temperature test failurePOST analog device test failureerror
1ABPS Battery Voltage Signal test failurePOST analog device test failureerror
1B5V dc Venthead test failurePOST analog device test failureerror
1CBPS Battery Current test failurePOST analog device test failureerror
1Dac Line Voltage test failurePOST analog device test failureerror
1EExhalation Motor Current test failurePOST analog device test failureerror
1F10V Sentry test failurePOST analog device test failureerror
5 Vdc Overvoltage5V power supply overvoltage (GUI only)Excessive 5V outputerror
Access Fault/Bus ErrorMemory access fault or bus errorInvalid memory access during POSTerror
AC POWER LOSSOperating on battery; AC not availablePower switch on, AC unavailable, ventilator operating on BPSwarning
AC POWER LOSSAC mains power unavailableAC power disconnected or failed; ventilator switches to BPS battery powererror
AC POWER LOSSAC input power is unavailable or insufficient; ventilator switched to battery powerAC monitor voltage falls below equivalent of 80 V AC at bulk supply inputwarning
ac power not connectedAC power is not supplied to ventilator during exhalation valve calibrationExhalation valve calibration procedure detects no AC powercritical
AC power not connectedCalibration cannot proceed; AC power is not connected to ventilatorCalibration attempted without AC power supplyerror
AE0110Unable to establish oxygen flow at 120 L/minFlow controller unable to establish and control oxygen flow at 120 L/minerror
AE0111Unable to establish air flow at 120 L/minFlow controller unable to establish and control air flow at 120 L/minerror
AE0213Air PSOL leakEST Gas Supply/SV Testwarning
AE0213Air PSOL leak: Excessive pressure buildup 50-100 cmH2O in blocked inspiratory moduleEST disables compressor, bleeds pressure, reads inspiratory pressure; forward leak detected via buildup > 50 but < 100 cmH2Owarning
AE0214Oxygen PSOL forward leak detected, pressure 50-100 cmH2OExcessive pressure buildup in blocked inspiratory module during safety valve testwarning
AE0306Test circuit not properly connected or expiratory pressure not detectedInspiratory and expiratory pressure transducer readings too far apart during leak testwarning
AE0601GUI High Alarm LED failsCLEAR key pressed to indicate LED not onwarning
AE0602GUI Medium Alarm LED failsCLEAR key pressed to indicate LED not onwarning
AE0603GUI Low Alarm LED failsCLEAR key pressed to indicate LED not onwarning
AE0604GUI Normal LED failsCLEAR key pressed to indicate LED not onwarning
AE0605GUI Batt Backup LED failsCLEAR key pressed to indicate LED not onwarning
AE0606GUI On Batt Pwr LED failsCLEAR key pressed to indicate LED not onwarning
AE0607GUI Compressor Ready LED failsCLEAR key pressed to indicate LED not onwarning
AE0608GUI Compressor Operating LED failsCLEAR key pressed to indicate LED not onwarning
AE0609GUI 100% O2 LED failsCLEAR key pressed to indicate LED not onwarning
AE0610GUI Alarm Silence LED failsCLEAR key pressed to indicate LED not onwarning
AE0611GUI Screen Lock LED failsCLEAR key pressed to indicate LED not onwarning
AE0702Bad Vent inop LEDEST BD Lamp Testwarning
AE0703Bad SVO LEDEST BD Lamp Testwarning
AE0704Loss of GUI LED failsLoss of GUI LED not illuminating on backup display unitwarning
AE1001Air PSOL loopback current out of rangeAir PSOL (PSOL2) loopback current outside expected range at various drive currentserror
AE1002O2 PSOL loopback current out of rangeOxygen PSOL (PSOL1) loopback current outside expected range at various drive currentserror
AE1104Insp check valve test failedEST Safety System Testwarning
AE1201Exp valve loopback current out of rangeExhalation valve loopback current outside expected range of drive current at test openingswarning
AE1305Seal test failedExhalation valve seal test pressure delta above alert level but below failure levelwarning
AE1600Compressor Test - Not installedEST Compressor Testinfo
AE1601Wall air pressure detected when should be disconnectedPS2 pressure switch closed when should be open during compressor test setuperror
AE1602AC power not connected - compressor cannot runSystem running on battery power when AC power should be available; BPS charged/charging activewarning
AE1603Compressor pressure not detectedPressure transducer PC indicates no compressor air present although compressor motor is onerror
AE1604Run mode time OOREST Compressor Testwarning
AE1604Compressor run time out of rangeCalculated run elapsed time not within acceptable range after 15 s delayerror
AE1606Compressor pressure present after motor disabledCompressor pressure transducer indicates air present after motor disabled and accumulator drainederror
AE1607Compressor disabled time out of rangeCalculated disabled elapsed time not within acceptable range after 15 s delayerror
AE1608Unable to test compressor standby modeCompressor timer could not be verified as functional in prior run and disabled testserror
AE1609Compressor standby time out of rangeCalculated standby mode elapsed time not within acceptable range; compressor motor still running during standby test phaseerror
AE1610Unable to perform compressor load testEST Compressor Testwarning
AE1611Compressor load test failed - pressure not maintainedPressure transducer PC indicates compressor unable to maintain minimum pressure during worst-case breath delivery waveform (200 lpm at 45 cmH2O)error
AE1700Compressor Leak Test - Not installedEST Compressor Leak Testinfo
AE1701Unable to perform compressor leak testCompressor timer tests not passed during prior Compressor Testerror
AE1702Wall air pressure detected during leak test disable phasePressure switch PS2 closed when should be open after disconnecting wall airerror
AE1703Wall air pressure not detected during leak test enable phasePressure switch PS2 open when should be closed after reconnecting wall airerror
AE1704Compressor leak detectedCompressor pressure transducer PC indicates loss of compressor pressure during 1-minute observation interval in standby mode, or compressor elapsed time > 1 s in standbyerror
AE1901GUI touch: ErrorEST GUI Touch Testwarning
AE2001Bad GUI serial portEST GUI Serial Port Testwarning
AE2001Serial port loopback test failedIn loopback mode, received message did not match transmittederror
AE2101Backup power supply not fully chargedBPS not fully charged at start of testerror
AE2102Battery pack not dischargingBPS not discharging after AC power disconnectederror
AE2103Bad Backup Power Supply - voltage issueBPS voltage dropped below accepted level or dropped too quickly during dischargeerror
AE2104Battery not charging after AC reconnectionBPS not charging after AC power reconnected (BPS charged bit on and BPS charging bit off)error
AE2300GUI Nurse Call Test - Not installedUser pressed CLEAR to indicate nurse call device not installedinfo
Air offset out of rangeAir flow sensor calibration failed; measured offset exceeds acceptable rangeCalibration procedure detects invalid air sensor offset during Flow Sensor Calibrationerror
Analog Interface ErrorAnalog interface failure during POSTerror
APNEASet apnea interval elapsed without triggering a breathApnea interval elapsed without ventilator, patient, or operator triggering a breathwarning
APNEA (lockable)No breath triggered within apnea intervalSet apnea interval elapses without patient, ventilator, or operator triggering breathwarning
AS0010Unable to establish O2 flowFlow controller unable to establish and control oxygen flow at 120 L/min during SSTerror
AS0011Unable to establish air flowFlow controller unable to establish and control air flow at 120 L/min during SSTerror
AS0012Oxygen pressure not detectedOnly air detected during SST flow sensor testerror
AS0013Wall air pressure not detectedOnly oxygen detected during SST flow sensor testerror
AS0202Excessive leakPressure drops to 85 cmH2O in 10 s, then in 10 s more by ≥10 cmH2O during SST circuit leak testerror
AS0202Excessive system leak (patient circuit test)Pressure drops to 85 cmH2O in 10 s, then by ≥ 10 cmH2O in next 10 seconds during patient circuit leak testerror
AS0206Test circuit not connectedInspiratory and expiratory pressure transducer readings diverge excessively during leak test pressure measurementswarning
AS0305Occluded expiratory filterPressure drop across filter 2-3 cmH2O (alert threshold)warning
AS0308Occluded exhalation compartmentExhalation compartment pressure >3 cmH2O but <4 cmH2O during SST expiratory filter testerror
AS0308Occluded expiratory compartmentExhalation compartment pressure >3 but <4 cmH2Owarning
AS0311Low expiratory filter pressure dropPressure drop across expiratory filter <0.4 but >0.1 cmH2Owarning
AS0403Occluded inspiratory limbInspiratory limb pressure 8.5-12.5 cmH2O (adult) or 5.5-7.5 cmH2O (pediatric)warning
AS0406Occluded exhalation limbExpiratory limb pressure 8.5-12.5 cmH2O (adult) or 5.5-7.5 cmH2O (pediatric)warning
AS0407Peak flow below minimum during circuit testPeak flow < 80 L/min but > 60 L/min during characterizationwarning
AS0407Unable to reach minimum peak flowPeak flow 60-80 L/min (alert threshold) during inspiratory limb resistance testwarning
AS0411Unable to reach minimum peak flowPeak flow < alert threshold but > failure threshold during circuit resistance testwarning
AS0411Peak flow below alert thresholdPeak flow below alert threshold during circuit characterizationerror
AS0413Insp limb resistance lowInspiratory limb pressure <0.6 cmH2O (adult) or 5.5 cmH2O (pediatric) but >0.2 cmH2O during SST circuit resistance testwarning
AS0414Expiratory limb resistance lowExpiratory limb pressure 0.2-0.6 cmH2O (adult) or 0.2-0.5 cmH2O (pediatric)warning
AS0505Patient circuit compliance abnormally highCompliance > 6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric), < 12/9 mL/cmH2Oerror
AS0505Excessive complianceHigh compliance >6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric) but <12 mL/cmH2O (adult) or 9 mL/cmH2O (pediatric) during SST compliance calibration testerror
AS0507Compliance lowHigh compliance falls below 1.56 mL/cmH2O (adult) or 1.34 mL/cmH2O (pediatric) but not lower than 1.05 mL/cmH2O during SST compliance calibration testerror
AS0509Excessive complianceLow compliance >6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric) but <12 or 9 mL/cmH2O respectivelywarning
AS0509Patient circuit compliance abnormally highCompliance > 6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric), < 12/9 mL/cmH2Oerror
AS0511Compliance lowLow compliance <1.56 mL/cmH2O (adult) or 1.34 mL/cmH2O (pediatric) but not below 1.05 mL/cmH2Owarning
Bad exp valve loopback currentExhalation valve loopback current measurement is out of specification during calibrationExhalation valve calibration detects abnormal loopback currenterror
Both insp and exp autozero failedBoth inspiration and expiration pressure transducer autozero operations failed during calibrationAutozero calibration step fails for both transducerserror
Calibration failed: current limit exceededCurrent supplied to exhalation valve exceeded allowable limit during calibrationCurrent measurement during exhalation valve calibration exceeds specificationerror
Calibration failed: gain resolutionExhalation valve calibration gain adjustment resolution exceededDuring exhalation valve gain calibration steperror
Calibration failed: gain resolutionGain resolution calculation failed during exhalation valve calibrationCalibration algorithm detects gain resolution errorerror
Calibration failed: gain resolutionDAC gain or resolution measurement out of specificationCalibration gain characterization steperror
Calibration failed: Pressure build time-outVentilator unable to build required test pressure within expected time during calibrationPressure build step times out during exhalation valve calibrationerror
Cannot achieve minimum air flowSystem cannot maintain minimum 60 L/min air flow during calibrationFlow Sensor Calibration procedure unable to sustain minimum air flowerror
Cannot achieve minimum O2 flowSystem cannot maintain minimum 60 L/min oxygen flow during calibrationFlow Sensor Calibration procedure unable to sustain minimum oxygen flowerror
Cannot program flashBDU and/or GUI NOVRAM cannot be successfully programmedAttempted datakey or serial number programming in service modeerror
CIRCUIT DISCONNECTNo ventilation; circuit disconnectedUnintended power loss recovery with circuit disconnect detected; or circuit disconnect detected during operationcritical
CIRCUIT DISCONNECT (lockable)Patient circuit disconnected or sensing failureVentilator detects loss of circuit continuityerror
COMPLIANCE LIMITED VTCompliance volume exceeds maximum allowedCompliance volume required for volume-controlled breath exceeds maximum for 3 of last 4 breathswarning
COMPLIANCE LIMITED VT (lockable)Excessive compliance volume required for VC breath deliveryCompliance volume required exceeds maximum allowed for 3 of last 4 breathswarning
COMPRESSOR INOPERATIVECompressor not operating due to low ACCompressor unable to operate; low AC powerwarning
Data key is not installedDatakey not detected by ventilatorDatakey insertion or boot-up in service modeerror
DEVICE ALERTBackground check or POST detected a problemPower-On Self-Test (POST) or background diagnostic failurewarning
DRAM Parity ErrorParity check failure in DRAMMemory data corruptionerror
DT0002Bus error / Access faultFailure that results in a POST or a resetcritical
DxxxxxAssertion Background Checks (BD)Failure that results in a POST or a resetcritical
Ethernet Parity ErrorParity check failure in EthernetData corruption in Ethernet transmissionerror
Exhalation pressure transducer autozero failedExhalation pressure transducer DAC count at 0 cmH2O is out of range during autozeroAutozero calibration step fails for exhalation transducererror
Exhalation valve temperature out of rangeExhalation valve temperature outside acceptable operating range during calibrationTemperature sensor reading during calibrationerror
ExxxxxIntentional system status eventStatus message indicating intentionally caused eventinfo
ExxxxxStatus message from GUI Background ChecksStatus message indicating an intentionally caused eventinfo
FE0001Inspiratory autozero out of rangeEST Circuit Pressure Test: Inspiratory pressure transducer ADC count at 0 cmH2O out of rangeerror
FE0002Expiratory autozero out of rangeEST Circuit Pressure Test: Exhalation pressure sensor ADC count at 0 cmH2O out of rangeerror
FE0003Failed to reach test pressureUnable to build pressure (air) to 10 cmH2O at 5 L/min within time-out period during EST circuit pressure testerror
FE0004Circuit pressure cross-check failed at 10 cmH2OInspiratory and expiratory pressure transducer readings too far apart during 10 cmH2O testerror
FE0005Bad inspiratory autozero solenoidInspiratory pressure reading out of range (-0.60 to 0.60 cmH2O) when SOL1 openederror
FE0006Bad expiratory autozero solenoidExpiratory pressure reading out of range (-0.60 to 0.60 cmH2O) when SOL2 openederror
FE0007Cross-check failed: Inspiratory/expiratory pressure transducers too far apartTest reads inspiratory and expiratory pressure transducers at 50 and 100 cmH2O; readings exceed toleranceerror
FE0007Cross-check failedInspiratory/expiratory pressure transducer measurements at 50 or 100 cmH2O test pressure are too far aparterror
FE0007Circuit pressure cross-check failed at 50 and 100 cmH2OInspiratory and expiratory pressure transducer readings too far apart during 50 and 100 cmH2O testerror
FE0008Failed to reach test pressureUnable to build pressure to 50 or 100 cmH2O at 5 L/min within timeout perioderror
FE0010AC power not connectedEST Circuit Pressure Test: System running on battery after user prompted to connect AC powercritical
FE0101Oxygen flow sensor cross-check failedQ1 and Q3 flow sensor readings too far apart at oxygen flows of 120, 60, 5, and 1 L/minerror
FE0102Oxygen PSOL command current out of rangePSOL1 command current out of range at oxygen flows of 120, 60, 5, and 1 L/minerror
FE0103Air flow sensor cross-check failedQ2 and Q3 flow sensor readings too far apart at air flows of 120, 60, 5, and 1 L/minerror
FE0104Air PSOL command current out of rangePSOL2 command current out of range at air flows of 120, 60, 5, or 1 L/minerror
FE0106Unable to establish O2 flow: Flow controller cannot control oxygen at lower ratesEST test verifies oxygen flow at 60, 5, and 1 L/min; flow controller unable to establish and controlerror
FE0107Unable to establish air flow: Flow controller cannot control air at lower ratesEST test verifies air flow at 60, 5, and 1 L/min; flow controller unable to establish and controlerror
FE0108O2 zero flow check failed: Oxygen flows when commanded to closeEST commands oxygen PSOL (PSOL1) to 0 L/min (closed); measured oxygen flow > 0.153 L/minerror
FE0109Air zero flow check failed: Air flows when commanded to closeEST commands air PSOL (PSOL2) to 0 L/min (closed); measured air flow > 0.153 L/minerror
FE0204Wall air pressure not detectedACCEPT key pressed but pressure switch PS2 still open after prompting user to connect wall airerror
FE0205Oxygen pressure not detectedACCEPT key pressed but pressure switch PS1 still open after prompting user to connect oxygenerror
FE0206O2 pressure detected: PS1 remains closed when oxygen disconnectedEST prompts disconnect oxygen; PS1 should open but remains closederror
FE0207Air PSOL leak: Excessive pressure buildup > 100 cmH2O in blocked inspiratory moduleEST disables compressor, bleeds pressure, reads inspiratory pressure; forward leak detected via buildup > 100 cmH2Oerror
FE0207Air PSOL leakAir PSOL (PSOL2) forward leak detected via excessive pressure buildup (>100 cmH2O) in blocked inspiratory moduleerror
FE0208Wall air pressure detected: PS2 remains closed when air disconnectedEST disables compressor control and prompts disconnect air; PS2 should open but remains closederror
FE0209O2 PSOL leakOxygen PSOL (PSOL1) forward leak detected via excessive pressure buildup (>100 cmH2O) in blocked inspiratory moduleerror
FE0210Safety valve pressure relief failedMultiple test failures: (1) stable 1 L/min flow not established with patient port blocked; (2) safety valve cracking pressure out of range; (3) neither PSOL could deliver 100 L/min within prescribed count; (4) peak pressure too high during PSOL testerror
FE0210SV pressure relief failed: Safety valve cracking pressure out of range or flow not establishedEST tests safety valve cracking pressure and/or establishes flow to patient port; pressure or flow out of specificationerror
FE0211Oxygen pressure not detectedPS1 (oxygen pressure switch) reads open when oxygen should be connectederror
FE0212Compressor pressure detected during EST depressurization testCompressor pressure transducer indicates non-zero pressure when should be depressurizederror
FE0215Air zero flow check failedDuring zero-flow check, air flow sensor (Q2) reads > 0.05 L/minerror
FE0216O2 zero flow check failedDuring zero-flow check, oxygen flow sensor (Q1) reads > 0.05 L/minerror
FE0217Exhalation zero flow check failed, Q3 >0.1 L/minExhalation flow sensor reads excessive flow during zero-flow test with supplies disconnectederror
FE0301Excessive leak in EST Leak TestSystem pressure dropped below failure pressure level during leak testerror
FE0301Excessive system leak detected during leak testSystem pressure dropped below 85 cmH2O or drops ≥5 cmH2O within 10 secondserror
FE0305Unable to establish pressure in EST Leak TestSystem cannot attain leak test starting pressure within time-out perioderror
FE0401Accept key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0402Clear key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0403Insp. Pause key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0404Exp. Pause key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0405Man Insp key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0405Manual Inspiration key fails during EST keyboard testMan Insp key not pressed or wrong key pressed within 15 secondserror
FE0406100% O2/CAL key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0407Info key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0408Alarm Reset key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0409Alarm Silence key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0410Alarm Volume key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0411Screen brightness fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0412Screen contrast key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0413Screen lock key fails in GUI Keyboard TestWrong key pressed or key not pressed within 15 seconds during ESTwarning
FE0501Knob not respondingKnob not turned in prompted direction within 15 secondserror
FE0501Bad knobEST GUI Knob Test - knob not turned in direction as prompted within 15 swarning
FE0501Knob rotation not detected during EST knob testKnob not turned in commanded direction within 15 secondserror
FE0801SAAS (Safety Audible Alarm System) test failedGUI alarm sound not heard during diagnostic test or cable disconnectedcritical
FE0801SAAS (Safety Audible Alarm System) test failed in GUI Audio TestCLEAR key pressed to indicate GUI audio diagnostic failederror
FE0901Bad alarm cable in BDU Audio TestAlarm cable voltage out of range (< 3.5 or > 5.05 V)error
FE0902Bad power fail capacitor (initial) in BDU Audio TestPower failure capacitor initial voltage out of range (< 4.5 or > 5.05 V)error
FE0903Bad power fail capacitor (final) in BDU Audio TestPower failure capacitor final voltage out of range or RC constant < 60 serror
FE0904Bad BD audio in BDU Audio TestCLEAR key pressed to indicate user did not hear alarm although alarm was activeerror
FE0904BD alarm unit not sounding or respondingSelf-test sounds alarm; user presses CLEAR indicating no audible alarm hearderror
FE1101Safety valve occluded in Safety System TestExcessive safety valve back pressure when safety valve is opencritical
FE1102Bad safety valve driver or loopback in Safety System TestSafety valve loopback current out of range during timed test pointserror
FE1103Insp check valve test failed in Safety System TestTook too long to relieve excess pressure through open safety valve, indicating inspiratory check valve occluded or test circuit too largeerror
FE1105Unable to establish flow in Safety System TestFlow controller unable to establish and control gas flow at 60 L/minerror
FE1301Seal test failedExhalation valve seal test pressure delta exceeds failure thresholdcritical
FE1301Seal test failed in Exp Valve Seal TestSeal test delta pressure above failure levelerror
FE1302Exp valve temp OOR in Exp Valve Seal TestExhalation valve magnet temperature out of range (10 to 100 degrees C)warning
FE1302Exp valve temp out of rangeExhalation valve magnet temperature outside 10-100 °C range during warmup diagnosticwarning
FE1303Unable to establish exp flowExhalation flow controller cannot establish/maintain 5 L/min air flow with 45 cmH2O system pressurecritical
FE1304Exp valve not calibrated in Exp Valve Seal TestExhalation valve table checksum invalid or last calibration not completed successfullyerror
FE1401Bad calibrationMeasured system pressure at one or more exhalation test points is out of rangeerror
FE1401Bad calibration in Exp Valve TestMeasured system pressure at one or more test points out of rangeerror
FE1402Exp valve not calibrated in Exp Valve TestExhalation valve table checksum invalid or last calibration not completed successfullyerror
FE1403Unable to establish exp flow in Exp Valve TestFlow controller unable to establish and control air flow measured by exhalation flow sensor Q3error
FE1501Unable to establish air flow in Exp Heater TestFlow controller unable to establish and control 60 L/min air flowerror
FE1502Bad exp heaterEST Exp Heater Test failure - exhalation heater temperature did not rise sufficiently when heater turned onerror
FE1502Bad exp heater in Exp Heater TestExhalation heater temperature did not rise sufficiently when heater was turned onerror
FE1503Bad exp heater (cooling) in Exp Heater TestHeater temperature did not drop sufficiently during interval after heater turned offerror
FE1503Bad exhalation heater - insufficient temperature dropHeater temperature did not drop ≥ 2 °C from heater-on temperature during interval after heater turned offerror
FE2201Low exp ∆PEV Velocity Transducer Test - expiratory delta pressure too low, indicating velocity transducer not responding properlywarning
FE2201Low exp delta PEV Velocity Transducer Test failure - expiratory delta P too low; velocity transducer not responding properlyerror
FE2201Low exp delta P in EV Velocity Transducer TestExpiratory delta pressure too low, indicating velocity transducer not responding properlyerror
FE2301Nurse call stuck on in GUI Nurse Call TestCLEAR key pressed to indicate nurse call (remote alarm) relay stuck on when should be offerror
FE2302Nurse call stuck off in GUI Nurse Call TestCLEAR key pressed to indicate nurse call (remote alarm) relay stuck off when should be onerror
Flow sensor cross-check failedExhalation air flow out of specification relative to inspiration test flowFlow measurement comparison during calibration cross-checkerror
Flow sensor cross-check: unable to establish flowAir flow sensor unable to detect 5 L/min test flow during cross-checkFlow controller test during exhalation valve calibrationerror
Flow sensor info invalidFlow sensor calibration table checksum error during exhalation valve calibrationChecksum verification of flow sensor calibration table failserror
FS0001O2 flow sensor cross check failed in SST Flow Sensor TestOxygen flow sensor (Q1) cross-check failed during self-service testerror
FS0002O2 PSOL current out of range in SST Flow Sensor TestOxygen PSOL (PSOL1) command current out of range with respect to oxygen flow sensor Q1error
FS0003Air flow sensor cross check failed in SST Flow Sensor TestAir flow sensor (Q2) cross-check failed during self-service testerror
FS0004Air PSOL current out of range in SST Flow Sensor TestAir PSOL (PSOL2) command current out of range with respect to air flow sensor Q2error
FS0005Gas not connected in SST Flow Sensor TestNo gas available (neither PS1, PS2, nor PC detects pressure)warning
FS0006Unable to establish O2 flowSST Flow Sensor Test - flow controller unable to establish and control oxygen flow at 60, 5, and 1 L/mincritical
FS0007Unable to establish air flowSST Flow Sensor Test - flow controller unable to establish and control air flow at 60, 5, and 1 L/mincritical
FS0008Oxygen zero flow check failedOxygen flow exceeds 0.153 L/min when PSOL1 is commanded to deliver 0 L/min (closed)error
FS0009Air zero flow check failedAir flow > 0.153 L/min when PSOL2 commanded closederror
FS0101Inspiratory autozero out of rangeInspiratory pressure transducer ADC count at 0 cmH2O is outside acceptable range during autozero calibrationerror
FS0102Expiratory autozero out of rangeExpiratory pressure transducer ADC count at 0 cmH2O is outside acceptable range during autozero calibrationerror
FS0103Failed to reach test pressure at 10 cmH2OSystem unable to pressurize to 10 cmH2O at 5 L/min flow within timeout perioderror
FS0104Pressure cross-check failed at 10 cmH2OInspiratory and expiratory pressure transducer readings diverge beyond acceptable tolerance at 10 cmH2O test pressureerror
FS0105Bad inspiratory autozero solenoidInspiratory pressure reading outside range (-0.60 to 0.60 cmH2O) when autozero solenoid (SOL1) commanded openerror
FS0106Bad expiratory autozero solenoidExpiratory pressure reading outside range (-0.60 to 0.60 cmH2O) when autozero solenoid (SOL2) commanded openerror
FS0107Pressure cross-check failed at 50 and 100 cmH2OInspiratory and expiratory pressure transducer readings diverge beyond acceptable tolerance at high pressure levels (50 and 100 cmH2O)error
FS0108Failed to reach high test pressure (50 or 100 cmH2O)System unable to pressurize to 50 or 100 cmH2O at 5 L/min flow within timeout perioderror
FS0201Excessive system leak detectedSystem pressure drops below 85 cmH2O or drops by ≥ 30 cmH2O over 20 seconds during leak testerror
FS0205Unable to establish pressureUnable to attain leak test starting pressure within timeout period using oxygen or airerror
FS0301Unable to establish flowSystem cannot deliver stable 60 L/min gas flowerror
FS0303Occluded expiratory compartmentExhalation compartment pressure >4 cmH2Oerror
FS0304Occluded expiratory filterPressure drop across expiratory filter > 3 cmH2Oerror
FS0306Patient circuit not disconnectedExpiratory flow detected during circuit disconnection testerror
FS0307Patient circuit not reconnectedExpiratory flow reading indicates circuit not reconnected after testerror
FS0310Low expiratory filter pressure dropPressure drop across expiratory filter <0.1 cmH2Oerror
FS0401Unable to establish flowFlow controller cannot establish and maintain gas flow during inspiratory limb testerror
FS0402Occluded inspiratory limbInspiratory limb pressure > 12.5 cmH2O (adult) or 7.5 cmH2O (pediatric)error
FS0404Unable to reach minimum peak flowPeak flow < 60 L/min during inspiratory limb resistance testerror
FS0405Occluded exhalation limbExpiratory limb pressure > 12.5 cmH2O (adult) or 7.5 cmH2O (pediatric)error
FS0408Inspiratory limb resistance lowInspiratory limb pressure <0.2 cmH2Oerror
FS0409Expiratory limb resistance lowExpiratory limb pressure < 0.2 cmH2Oerror
FS0410Unable to reach minimum peak flowPeak flow falls below failure threshold during circuit resistance characterizationerror
FS0412Wye not blockedUnable to pressurize system to 100 cmH2O at 60 L/min within timeouterror
FS0501Unable to pressurize circuitSystem cannot reach one or more test pressure points during compliance measurementerror
FS0502Unable to pressurize circuitSystem cannot pressurize to test pressure points during high-flow compliance testerror
FS0503Compliance calculation failureOut-of-range pressure ratio calculation during compliance measurementerror
FS0504Excessive complianceHigh compliance >12 mL/cmH2O (adult) or 9 mL/cmH2O (pediatric)error
FS0506Compliance lowHigh compliance falls below 1.05 mL/cmH2Oerror
FS0508Excessive complianceLow compliance >12 mL/cmH2O (adult) or 9 mL/cmH2O (pediatric)error
FS0510Compliance lowLow compliance < 1.05 mL/cmH2Oerror
↑fTOTTotal respiratory rate at or above set limit (updated at each inspiration)Total respiratory rate ≥ set limit, updated at beginning of each inspirationwarning
↑fTOT (High Respiratory Rate Total)Total respiratory rate at or above set alarm limitMeasured respiratory rate ≥ set limitwarning
HN0001Non-maskable interrupt (NMI): DRAM Parity ErrorBDU reset triggered by DRAM parity errorcritical
HN0002NMI: Ethernet Parity ErrorBDU reset triggered by Ethernet parity errorcritical
HN0016NMI: Analog Interface Error (ADC Channel Sequencer Fault)Non-maskable interrupt from ADC channel sequencer faultcritical
HN0016_ATFNMI: Analog Interface Error - ADC Timing FaultBDU reset triggered by ADC timing faultcritical
HN0016_CSFNMI: Analog Interface Error - ADC Channel Sequencer FaultBDU reset triggered by ADC channel sequencer faultcritical
HN0016_HDFNMI: Analog Interface Error - Hamming Decode FaultBDU reset triggered by Hamming decode faultcritical
HN0080NMI: Analog Interface Error - ADC Timing FaultBDU reset triggered by ADC timing faultcritical
HN0129NMI: DRAM Parity ErrorNon-maskable interrupt from DRAM parity errorcritical
HN0130NMI: Ethernet Parity ErrorNon-maskable interrupt from Ethernet parity errorcritical
INOPERATIVE BATTERYBackup Power System (BPS) installed but not functioningBPS detected but fails initialization or operation testwarning
Inspiration autozero failedInspiration pressure transducer DAC count at 0 cmH2O is out of range during autozeroAutozero calibration step fails for inspiration transducererror
INSPIRATION TOO LONGInspiratory time for spontaneous breath exceeds IBW-based limitInspiratory time for spontaneous breath ≥ IBW-based limitwarning
INSPIRATION TOO LONG (lockable)Spontaneous breath inspiratory time exceeds ideal body weight (IBW) limitInspiratory time for spontaneous breath ≥IBW-based limitwarning
KB0001Bad safety valve switched sideMeasured voltage on safety valve switched side indicates valve is not in expected statecritical
KB0002Bad expiratory flowExhalation flow sensor (Q3) reading out of rangecritical
KB0003Bad O2 PSOL currentOxygen PSOL (PSOL1) current out of rangecritical
KB0004Bad air PSOL currentAir PSOL (PSOL2) current out of rangecritical
KB0005Bad exp motor currentCurrent to exhalation valve motor out of rangecritical
KB0007Bad exp pressurePE (expiratory pressure) reading out of rangecritical
KB0008Bad insp pressurePI (inspiratory pressure) reading out of rangecritical
KB0009Air flow out of range HIGHQ2 (air flow sensor) reading too highcritical
KB0012O2 flow out of range HIGHQ1 (oxygen flow sensor) reading too highcritical
KB0016Bad BD 10V supplyBDU +10 V supply out of rangecritical
KB0017Bad BD 12 V supplyBDU +12V supply out of rangecritical
KB0018Bad BD 15 V supplyBDU +15V supply out of rangecritical
KB0019Bad BD -15 V supplyBDU background check detects -15 V supply out of rangecritical
KB0022Bad BD 5 V supplyBDU background check detects +5 V supply out of rangecritical
KB0023O2 PSOL stuckOxygen PSOL (PSOL1) command current out of rangecritical
KB0024Air PSOL stuckAir PSOL (PSOL2) command current out of rangecritical
KB0025Air PSOL stuck openAir PSOL (PSOL2) command current reading indicates solenoid cannot closecritical
KB0026O2 PSOL stuck openOxygen PSOL (PSOL1) command current reading indicates solenoid cannot closecritical
KB0030Bad safety valve currentSafety valve solenoid current out of range during BDU checkcritical
KB0031Insp pressure stuckInspiratory pressure transducer (PI) reading stuck during BDU checkcritical
KB0032Exp pressure stuckExpiratory pressure transducer (PE) reading stuck during BDU checkcritical
KB0033Insp pressure autozero offset failedInspiratory pressure transducer (PI) reading following autozero out of rangecritical
KB0034Exp pressure autozero offset failedExpiratory pressure transducer (PE) reading following autozero out of rangecritical
KB0037Analog-Digital converter failed highAnalog-to-digital converter failure during BDU checkcritical
KB0038Analog-Digital converter failed lowAnalog-to-digital converter failure during BDU checkcritical
KB0039Analog-Digital loopback failedADC loopback test failure during BDU checkcritical
KB0044BD NOVRAM checksum errorBD NOVRAM (non-volatile RAM) checksum validation fails during BDU checkcritical
KB0053BD EEPROM checksum errorBD EEPROM (non-volatile memory) checksum validation fails during BDU checkcritical
KB0064LV Ref out of rangeLow-voltage reference circuit reading out of specification during BDU checkcritical
KB0065SV current out of rangeSafety valve solenoid current out of specification during BDU checkcritical
KB0076Task MonitorTask monitor function failure during BDU checkcritical
KB0079BK vent inop occurredBDU detects ventilator inoperative signal active from GUI; indicates GUI has declared vent inoperativecritical
KB0082Watchdog failure occurredBD CPU watchdog timer failed during BDU checkcritical
KP0001Processor Initialization test failedPhase 1 POST processor initialization failserror
KP0002Integer Unit test failedPhase 1 POST general CPU confidence test failserror
KP0003DRAM Refresh Timer TestPOST detects DRAM refresh circuit failurecritical
KP0004Kernel DRAM TestPOST detects kernel DRAM memory failurecritical
KP0005Boot EPROM Checksum TestPOST detects EPROM checksum validation failurecritical
KP0006POST Phase 2 InitializationPOST Phase 2 initialization sequence failurecritical
KP0007Addressing Mode TestPOST detects CPU addressing mode test failurecritical
KP0008Kernel NOVRAM TestPOST detects kernel NOVRAM test failurecritical
KP0009Rolling Thunder TestPOST has started but not completed three or more times for reasons unrelated to power failurecritical
KP0011Time of Day Clock Test failed (Major)Fails during Phase 1 kernel POSTcritical
KP0012Timer TestTimer failed comparison with real-time clock or failed to generate interruptcritical
KP0013Watchdog Timer TestPOST detects watchdog timer did not time out as expectedcritical
KP0015FLASH Memory Checksum TestPOST detects FLASH memory checksum validation failurecritical
KP0081Memory Management Unit TestPOST detects memory management unit (MMU) circuit failurecritical
KP0082Bus Timer TestLocal bus activity has ceased or bus time-out circuit did not generate expected interruptcritical
KP0083NMI Source Register TestPOST detects non-maskable interrupt (NMI) source register test failurecritical
KP0084DRAM Memory test failedPhase 2 POST DRAM data pattern verification failserror
KP0084POST DRAM TestPOST detects DRAM memory failure during POST sequencecritical
KP0085Ethernet Self-Test start failedPhase 2 POST Ethernet controller initial verification failserror
KP0087Unexpected Reset Umpire TestPOST has been invoked three times within 24 operational hours due to unexpected resetscritical
KP0089Floating Point Unit TestPOST detects floating-point math unit failurecritical
KP0090DRAM Parity Circuit TestPOST detects DRAM parity error detection circuit failurecritical
KP0112VENT INOP Test failedVentilator in VENT INOP statecritical
KP0113Analog Interface PCB TestPOST detects AI PCB circuit test failurecritical
KP0114ADC TestPOST detects ADC (analog-to-digital converter) data not available to CPU as expectedcritical
KP0115DAC TestPOST detects DAC (digital-to-analog converter) output data not as expectedcritical
KP0116BDU POST analog devices test failure; check specific channel per addressBDU POST analog devices test (Phase 2) failurecritical
KP0116BDU POST analog device test failureAnalog device failure during BDU POST Phase 2critical
KP0117BD Serial Device TestPOST detects serial communication device test failurecritical
KP0120Safe State System TestPOST detects failure to place PSOLs and exhalation valve into safe state or verify they cannot be commandedcritical
KP0128Download operating system (OS) BootPOST detects OS boot download failurecritical
KP0129Operating System Initialization failedOS fails to loadcritical
KP0129Application OS BootPOST detects application OS boot failurecritical
KP0130PB-MON BootPOST detects PB-MON (monitoring application) boot failurecritical
KP0131Application BootPOST detects main application boot failurecritical
LB0006Bad exhalation valve coil tempBackground check detects exhalation valve coil temperature out of rangewarning
LB0010Air flow out of range LOWBackground check detects air flow sensor (Q2) reading too lowwarning
LB0011Air flow sensor thermistor out of rangeBackground check detects Q2 thermistor reading out of rangeerror
LB0011Bad air flow temperatureAir flow sensor Q2 thermistor reading out of rangeunknown
LB0013O2 flow measurement below specificationOxygen flow sensor (Q1) reading too lowerror
LB0013Oxygen flow sensor reading too lowBackground check detects Q1 reading below rangeerror
LB0014Oxygen flow sensor thermistor out of rangeBackground check detects Q1 thermistor reading out of rangeerror
LB0014Bad O2 flow tempOxygen flow sensor Q1 thermistor reading out of rangeunknown
LB0015Expiratory flow sensor thermistor out of rangeBackground check detects Q3 thermistor reading out of rangeerror
LB0015Expiratory flow sensor thermistor reading out of rangeExpiratory flow sensor (Q3) thermistor malfunctionerror
LB0020GUI +12V supply out of rangeBackground check detects supply voltage out of rangeerror
LB0020GUI +12V supply voltage out of specificationGUI +12V power supply voltage out of rangeerror
LB0021GUI +5V supply out of rangeBackground check detects supply voltage out of rangeerror
LB0021GUI +5V supply voltage out of specificationGUI +5V power supply voltage out of rangeerror
LB0027Atmospheric pressure transducer reading out of specificationAtmospheric pressure sensor malfunctionerror
LB0027Atmospheric pressure transducer reading out of rangeBackground check detects pressure reading out of rangeerror
LB0028Bad O2 sensor OOROxygen sensor percentage reading out of rangeunknown
LB0028Oxygen sensor percentage reading out of rangeBackground check detects sensor reading outside limitserror
LB0029Oxygen sensor alert was resetBackground check status messageinfo
LB0035Power fail capacitor voltage out of rangeBackground check detects capacitor voltage outside limitserror
LB0036Alarm cable voltage out of rangeBackground check detects cable voltage anomalyerror
LB0039ADC loopback constant test failureBackground check detects ADC loopback failureerror
LB0043Power switch is in off state but power monitoring signal not detectedPower switch failure; switch appears off but power fail NMI not triggerederror
LB0043Power switch appears stuckBackground check detects switch in off state but power fail NMI did not occurerror
LB0045Real-time clock failureBackground check detects BD time of day failureerror
LB0048Backup power supply voltage out of rangeBackground check detects BPS voltage anomalyerror
LB0049Backup power supply current out of rangeBackground check detects BPS current anomalyerror
LB0050Backup power supply model information incorrectBackground check detects wrong BPS model dataerror
LB0051Exhalation heater thermistor reading out of rangeBackground check detects heater thermistor anomalyerror
LB0051Bad exp heaterExhalation heater thermistor reading out of rangeunknown
LB0056Compressor elapsed timer errorCompressor timer failureunknown
LB0057Compressor EPROM data corruptionBackground check detects unexpected EPROM dataerror
LB0058Loss of communication with GUIBackground check detects BD-GUI communication failureerror
LB0060Resume GUI communicationCommunication with GUI was resumedunknown
LB0060Communication with GUI has been restoredBackground check status messageinfo
LB0062EST required but not runBackground check detects user powered into normal mode without running ESTerror
LB0062EST (Equipment Self Test) required but not runUser powered on in normal mode instead of service mode when EST was needederror
LB0075Data key update failedBackground check detects data key update failureerror
LB0076Task monitor alertBackground check detects task monitoring anomalywarning
LB0080Breath delivery extended beyond expected timeBackground check detects breath delivery timing faultcritical
LB0083GUI communication initialized and resumedBackground check initialization status messageinfo
LB0083Initialization: GUI communication resumedGUI communication restored during power-up sequenceinfo
LB0085GUI communication lost during initializationBackground check initialization status messageinfo
LB0085Initialization: GUI communication lostGUI communication lost during power-up sequenceinfo
LB0087Compressor serial number updatedBackground check status messageinfo
LB008810000 hours stored for elapsed timeCompressor hours corrupted; 10000 hour limit reachedunknown
LB0088Compressor hours corrupted (10000 hours stored)Background check detects elapsed time corruptionerror
LB0089Cannot determine data key sizeBackground check detects data key read failureerror
LB0092Battery event status indicationBackground check detects battery state changewarning
LB0092Battery Event alert related to backup power supply statusBackup power supply event detectederror
LC1xxxCommunications error between BD and peripheral devicesSerial or network communication failure with external deviceserror
LCxxxxNetwork communications error between BDU and GUICommunications error detected with variable code suffixerror
LOSS OF POWERInsufficient AC and BPS power; independent audio alarm soundsPower switch on but insufficient AC and BPS power combinedcritical
LOW AC POWERAC power dropped below 80% of nominal for 1 sMains (AC) power <80% of nominal for ≥1 swarning
LOW BATTERYBattery Backup Power remaining < 2 minutesBPS operational time falls below approximately 2 minuteswarning
LOW BATTERYBattery has <2 minutes of operational time remainingBPS operational time <2 minuteswarning
LOW INSP PRESSUREInspiratory target pressure below minimum required (VC+ breath)Inspiratory target pressure <(PEEP + 5 cmH2O) when mandatory type is VC+warning
LOW INSP PRESSURE (lockable)Inspiratory target pressure below minimum safe levelInspiratory target pressure < (PEEP + 5 cmH2O) when breath type VC+warning
LP0008Non-critical NOVRAM data failure during startupNVRAM self-test detected non-critical data corruptionerror
LP0008Kernel NOVRAM data failure (noncritical)POST (BD) detects noncritical NOVRAM data anomalyerror
LP0010Interrupt controller test failurePOST (BD) detects interrupt controller faultcritical
LP0011Time of Day Clock test failed (minor)Phase 1 POST Time of Day clock functionality verification failswarning
LP0085Ethernet self-test startPOST (BD) ethernet diagnostics begincritical
LP0086Ethernet Self-Test end status failedPhase 2 POST Ethernet controller final status check failswarning
LP0087Unexpected reset detected (watchdog timeout)POST (BDU) invoked due to watchdog time-outerror
LP0088NOVRAM data failure detected during POSTNon-critical NVRAM data corruption during self-testerror
LP0088POST NOVRAM test failure (noncritical)POST (BDU) detects noncritical NOVRAM data anomalyerror
LP0112Ventilator INOP signal in unexpected statePOST (BDU) detects VENT INOP signal faultcritical
LP0114ADC (analog-to-digital converter) test failurePOST (BD) detects ADC faultcritical
LP0114ADC TestAnalog to Digital Converter test failureunknown
LP0115DAC (digital-to-analog converter) test failurePOST (BD) detects DAC faultcritical
LP0115DAC TestDigital to Analog Converter test failureunknown
LP0116BDU POST analog device test failureAnalog device failure during BDU POSTcritical
LP0116GUI analog devices test failure; check specific channel per addressAnalog devices test (Phase 2) failure at GUIcritical
LP0121Service switch appears to be stuck closedPOST (BDU) detects service switch stuckerror
LP0122AC Voltage Test failedAC monitor voltage outside specified rangewarning
LP0122AC input voltage outside nominal rangePOST (BDU) detects AC monitor voltage not within ±20% of set nominalerror
No air connectedAir supply pressure not detected during calibrationAir supply pressure switch check at calibration starterror
No air connectedWall and compressor air pressure switches detect no air connected to ventilatorAir pressure switches report no air supply during calibrationcritical
NO AIR SUPPLYAir inlet filter blocked or air source disconnected; compressor inoperativeAir pressure switch fails to detect supply pressure after system startupwarning
NO AIR SUPPLYAir supply unavailable; ventilator delivers 100% oxygen or only O2 availableAir inlet pressure loss; compressor inoperative; safety valve opencritical
No O2 connectedOxygen supply is not detected; calibration cannot proceed without oxygenOxygen pressure switch (PS1) detects no oxygen connected during calibration setuperror
NO O2 SUPPLYOxygen source lost or inlet filter blockedO2 pressure switch fails to detect oxygen pressurewarning
NO O2 SUPPLYOxygen supply unavailable; ventilator delivers 21% or only air availableOxygen inlet pressure loss; safety valve opencritical
↑O2%Measured O2% 7% or more above set valueO2% measured ≥7% above setting for ≥30 s (12% during first hour; +5% for 4 min after decrease)warning
↓O2%Measured O2% 7% or more below set value or below 18%O2% measured ≤7% below setting for ≥30 s (or ≤12% during first hour; +5% for 4 min after increase) or <18%error
↑O2% (High O2%)Measured oxygen percentage above set limitMeasured O2% ≥7% above set O2% for ≥30 s (or ≥12% in first hour)warning
↓O2% (Low O2%)Measured oxygen percentage below set limitMeasured O2% ≤7% below set O2% for ≥30 s, or below 18%error
O2 offset out of rangeOxygen flow sensor calibration failed; measured offset exceeds acceptable rangeCalibration procedure detects invalid oxygen sensor offset during Flow Sensor Calibrationerror
O2 SENSORO2 sensor out of calibration or sensor failureSensor output falls outside acceptable range or sensor not respondingwarning
Operating on batteryVentilator is powered by internal BPS battery; AC power unavailableAC input voltage drops below +80 V AC equivalentwarning
Operational time < 2 minutesLess than 2 minutes of battery power remains; critical low-battery conditionBPS discharge reaches point where only 2 minutes of powered operation remaincritical
↓PCOMP↓PPEAK detected but not ↓PCOMP; target pressure not metTarget pressure <(↓PPEAK - 5 cmH2O)warning
↑PMEANMeasured airway pressure at or above set limitMeasured airway pressure ≥ set limitwarning
↑PMEAN (lockable)Mean airway pressure at or above set limitMeasured airway pressure ≥ set limitwarning
POST Internal Software FailurePOST internal software errorSoftware logic failureerror
Power Fail NMIPower failure during POST (BDU only)Loss of AC power during POSTcritical
↑PPEAKPeak inspiratory pressure detected; not compliance↑PPEAK detected and NOT ↓PCOMPwarning
↑PPEAK (lockable)Peak inspiratory pressure at or above set limitPeak pressure detected exceeds set limitwarning
Pressure sensor cross-check alert: approaching spec limitPressure cross-check alert at one or more test pressures without failuresPressure measurement during five-point pressure cross-check (5, 25, 50, 75, 100 cmH2O)warning
Pressure sensor cross-check failedPressure transducer cross-check failed at one or more of five test pressuresPressure measurement comparison at 5, 25, 50, 75, or 100 cmH2O during calibrationerror
Pressure sensor cross-check: unable to build pressureSystem unable to pressurize to one or more of five test pressuresPressurization algorithm test sequence during calibrationerror
PROCEDURE ERRORPatient connected before setup procedure completePatient circuit connected while ventilator startup sequence incompletewarning
↑PVENTInspiratory pressure >100 cmH2O; VC breath typeInspiratory pressure >100 cmH2O and mandatory type = VCerror
↑PVENT (lockable)Inspiratory pressure exceeds 100 cmH2O with volume control breathInspiratory pressure > 100 cmH2O and breath type VCerror
SAAS NMIGUI alarm or alarm cabling interruptAlarm system anomalywarning
SCREEN BLOCKPossible blocked beam on display GUI or touch screen malfunctionDisplay not responding or sensor detecting obstructioninfo
Serial number(s) doesn't match the data keyInserted datakey does not match ventilator serial numbers; invalid datakeyDatakey insertion when serial numbers were initialized with different datakeywarning
SEVERE OCCLUSIONSevere obstruction in ventilator circuit preventing ventilationVentilator detects very low or no flow despite pressure generationcritical
SEVERE OCCLUSIONLittle or no ventilation; ventilator in occlusion status cycling (OSC)Severe circuit or patient airway occlusion detectedcritical
SxxxxxStatus message indicating intentional eventBackground check status codeinfo
SxxxxxStatusIntentionally caused eventunknown
U04004Assertion failure from intermittent digital I/O connectionIntermittent connection to digital I/O line detectederror
U04004Assertion failure (I/O intermittent fault)Background check GUI detects I/O connection anomalywarning
U05013Assertion failure (I/O intermittent fault)Background check GUI detects I/O connection anomalywarning
U05013Assertion failure from intermittent digital I/O connectionIntermittent connection to digital I/O line detectederror
U08002Serial data buffer overflow from host deviceBackground check BDU detects excessive serial errorserror
U08002Assertion failure from GUI data buffer overflowGUI overloaded with invalid data from serial port; host transmitting bad data causing over-run, parity, or framing errorserror
U08012Assertion failure generating soft resetBackground check GUI detects anomalywarning
U08012Assertion failure; GUI generates soft reset to resynchronizeGUI detects error condition and initiates soft reseterror
U14019Assertion failure generating soft resetBackground check GUI detects anomalywarning
U14019Assertion failure; GUI generates soft reset to resynchronizeGUI detects error and initiates soft reseterror
U14021Assertion Failure - Background check (GUI)Time stamps between GUI and BD are out of syncwarning
Unable to burn flashGUI NOVRAM cannot be successfully programmedCal Info Duplication function cannot write calibration data to GUI NOVRAMerror
Unable to establish flowFlow controller unable to establish 5 L/min test flow via exhalation flow sensorExhalation flow controller test during calibrationerror
Unable to program flashCalibration table data unable to write to BDU NOVRAMCalibration completion when storing results to non-volatile memoryerror
Unable to program flashBDU NOVRAM cannot be successfully programmedInitialize Flow Sensor function cannot write calibration data to BDU NOVRAMerror
Unable to program flashUnable to program calibration table into BDU NOVRAM memory during calibrationFlash programming step fails at end of calibration procedurecritical
Unable to read air flow sensorError reading air flow sensor (Q2) serial EEPROM dataInitialize Flow Sensor function detects Q2 EEPROM read failureerror
Unable to read O2 flow sensorError reading oxygen flow sensor (Q1) serial EEPROM dataInitialize Flow Sensor function detects Q1 EEPROM read failureerror
Unable to read the expiratory flow sensorError reading exhalation flow sensor (Q3) serial EEPROM dataInitialize Flow Sensor function detects Q3 EEPROM read failureerror
Unable to receive flash dataGUI did not receive all calibration data from BDU within 10 secondsCal Info Duplication function timeout waiting for BDU data transmissionerror
Unexpected InterruptUnexpected interrupt occurredUnidentified interrupterror
Unknown NMIUnknown non-maskable interruptCPU interrupt during POSTerror
UT0002LCD controller PCB fault - bus error or access fault in VGA controllerVGA controller failure on upper or lower LCD PCBerror
UT0002Assertion Failure - Background check (GUI)Hardware-trap detection triggered; processor exception encounteredwarning
UxxxxxAssertion failure (generic)Background check GUI detects assertionwarning
↑VE TOTExpiratory minute volume at or above set limitExpiratory minute volume ≥ set limitwarning
↓VE TOTTotal minute volume below set limit for ≤20 s, >20 s, or >40 sTotal minute volume ≤ set limit for specified durationwarning
↑VE TOT (lockable)Expiratory minute volume at or above set limitExhaled minute volume ≥ set limitinfo
↓VE TOT (Low Minute Volume Total)Total minute volume below set limitExhaled minute volume ≤ set limit for various durationswarning
VN0129NMI: DRAM Parity ErrorMemory parity error detected in GUI DRAM during operationerror
VN0130NMI: Ethernet Parity ErrorEthernet controller parity error detectederror
VN0132NMI: +5 V DC Overvoltage+5V power rail exceeds safe operating voltageerror
VN0136NMI: +12 V DC Overvoltage+12V power rail exceeds safe operating voltageerror
VN0144NMI: +12 V DC Undervoltage+12V power rail drops below minimum operating voltageerror
VOLUME NOT DELIVEREDVentilator unable to deliver target tidal volume on VC+ or VS breathsMeasured delivered volume below target; pressure reaches maximum limitwarning
↑VTEExhaled tidal volume at or above set limitExhaled tidal volume ≥ set limitwarning
↑VTE (lockable)Exhaled tidal volume at or above set limitExhaled tidal volume ≥ set limitinfo
↓VTE MANDExhaled mandatory tidal volume below set limit (2 of last 4 breaths)Exhaled mandatory tidal volume ≤ set limit for 2 of last 4 mandatory breathswarning
↓VTE MAND (Low Tidal Volume Mandatory)Exhaled mandatory tidal volume below set limitExhaled mandatory VT ≤ set limit for multiple breathswarning
↓VTE SPONTExhaled spontaneous tidal volume below set limitExhaled spontaneous tidal volume ≤ set limitwarning
↓VTE SPONT (Low Tidal Volume Spontaneous)Exhaled spontaneous tidal volume below set limitExhaled spontaneous VT ≤ set limit; alarm updated when recalculatedwarning
↑VTI MANDDelivered inspiratory volume exceeds high limit (VC+ breath)Delivered inspiratory volume ≥ high inspiratory volume limit (breath type = VC+)warning
↑VTI MAND (lockable)Delivered inspiratory volume exceeds high limit (VC+ breaths)Delivered inspiratory volume ≥ high inspiratory volume limit (breath type VC+)info
↑VTI SPONTDelivered inspiratory volume exceeds high limit (TC or VS breath)Delivered inspiratory volume ≥ high inspiratory volume limit (breath type = TC or VS)warning
↑VTI SPONT (lockable)Delivered inspiratory volume exceeds high limit (TC or VS breaths)Delivered inspiratory volume ≥ high inspiratory volume limit (breath type TC or VS)info
XB0066Monitor alarms failAlarm monitoring system background check failure during GUI operationcritical
XB0067Monitor apnea alarm failsApnea alarm monitoring background check failurecritical
XB0068Monitor apnea int failsApnea interval monitoring system failurecritical
XB0069Monitor high inspiratory pressure alarm test failureHigh pressure alarm circuit malfunctioncritical
XB0069Monitor hip failsHigh Inspiration Pressure (HIP) alarm monitoring failurecritical
XB0070Monitor insp time failsInspiration time monitoring background check failurecritical
XB0071Monitor no dataNo monitoring data received during background checkcritical
XB0072Monitor data corruptedMonitoring data integrity check failure; data corruption detectedcritical
XB0073Monitor O2 mixture failsO2 mixture monitoring background check failurecritical
XB0074Monitor breath time failsBreath timing monitoring system failureerror
XP0001Processor Initialization failed (GUI)Fails during Phase 1 kernel POSTcritical
XP0002Integer Unit Test failed (GUI)Fails during Phase 1 kernel POSTcritical
XP0003DRAM Refresh Timer Test failed (GUI)Fails during Phase 1 kernel POSTcritical
XP0004Kernel DRAM Test failed (GUI)Fails during Phase 1 kernel POSTcritical
XP0005Boot PROM Checksum Test failed (GUI)Fails during Phase 1 kernel POSTcritical
XP0006Phase 2 Initialization Start failed (GUI)Non-kernel POST portion fails to startcritical
XP0007Addressing Mode Test failed (GUI)Fails during Phase 1 kernel POSTcritical
XP0008Kernel NOVRAM Test failed (GUI)Fails during Phase 1 kernel POSTcritical
XP0010Interrupt Controller TestInterrupt handling circuit test failure during POSTcritical
XP0011Time of Day Clock Test failed (GUI, Major)Fails during Phase 1 kernel POSTcritical
XP0012Timer Test failed (GUI)Fails during Phase 1 kernel POSTcritical
XP0013Watchdog Timer TestWatchdog timer did not respond as expected or trigger reseterror
XP0015EEPROM Checksum Test failed (GUI)Fails during Phase 1 kernel POSTcritical
XP0081Memory Management Unit Test failed (GUI)MMU test failurecritical
XP0082Bus Timer Test failed (GUI)Bus activity monitoring failurecritical
XP0083NMI Source Register TestNon-maskable interrupt source register test failurecritical
XP0084DRAM Memory Test failed (GUI)DRAM data pattern verification failscritical
XP0084POST DRAM TestPOST phase DRAM memory test failurecritical
XP0085Ethernet Self-Test Start failed (GUI)Ethernet controller functionality errorcritical
XP0087Unexpected Reset Umpire TestPOST invoked three or more times in 24 operational hours due to unexpected resetserror
XP0089Floating Point Unit Test failed (GUI)FPU functionality test failurecritical
XP0090DRAM Parity Circuit TestDRAM parity checking circuit test failure during POSTcritical
XP0097SAAS Self-test Start failedAlarm circuitry error or no audio feedbackcritical
XP0098SAAS Self-test End failedAlarm circuitry errorcritical
XP0128Download OS BootOperating system download/initialization failurecritical
XP0129Application OS BootApplication operating system boot failurecritical
XP0129Operating System Initialization failed (GUI)OS fails to loadcritical
XP0130PB-MON BootPuritan Bennett monitoring system boot failurecritical
XP0131Application Initialization failed (GUI)Application software fails to loadcritical
XP0131Application BootMain application software boot failurecritical
ZB0040Touch screen failedTouch screen hardware not responding; no input detectionwarning
ZB0041Touch screen blockedTouch screen input blocked; screen obstruction detected or input errorunknown
ZB0042Touch screen resumedTouch screen communication or functionality restored after previous failureinfo
ZB0046GUI Time of Day failedReal-time clock (time of day) initialization or operation failurewarning
ZB0047GUI NOVRAM checksum errorNOVRAM data integrity check failure; noncritical data corruption detectedwarning
ZB0052GUI key stuckKeyboard key detected as continuously pressed or stuckwarning
ZB0054GUI EEPROM checksum errorEEPROM data integrity check failure during boot or operationwarning
ZB0055GUI SAAS communication failedSelf-Addressing Alarm System communication link failurewarning
ZB0059Loss of BD communicationCommunication lost between GUI and BD (Breathing Delivery) unitcritical
ZB0061Resume BD communicationBD communication link re-established after previous lossinfo
ZB0063GUI SAAS Audio failedSAAS audio output (alarm speaker) circuit failurewarning
ZB0076Task MonitorSystem task monitoring flag set; indicates potential system performance issueinfo
ZB0077GUI dropped a waveform packetWaveform data packet lost in transmission from BD to GUIinfo
ZB0084Init Resume BD communicationBD communication reestablished during initialization or after recoveryinfo
ZB0086Init Loss of BD communicationBD communication lost during initialization sequencecritical
ZB0090Touch Screen blockedTouch screen input blocked due to visible obstruction or contaminationunknown
ZB0091Touch Screen resumedTouch screen functionality restored; blocked error clearedinfo
ZC0xxxCommunications error (BDU)Communication error detected in BDU with BPS or external systemswarning
ZC0xxx_BDUCommunications error (BDU); battery power system conditionCommunications error (BDU); BPS charging or status issuewarning
ZC1xxxCommunications error (GUI)Additional GUI communication error variant; protocol or interface failurewarning
ZC2000DCI parity errorData Communication Interface parity check failurewarning
ZC2001DCI input buffer overflow errorCommunications error with GUIwarning
ZC2002DCI command error; undefined command receivedCommunications error (GUI) undefined command received from hostwarning
ZC2003DCI unknown errorUnspecified communications errorwarning
ZP0008Kernel NOVRAM Test failureNoncritical NOVRAM data failurewarning
ZP0011Time of Day Clock Test failed (GUI, Minor)Fails during Phase 1 kernel POSTwarning
ZP0085Ethernet Self-Test Start failureEthernet test failed during POSTerror
ZP0086Ethernet Self-test End (GUI)Ethernet test completionwarning
ZP0087Unexpected Reset Umpire TestPOST invoked after unexpected resetwarning
ZP0088Phase 2 NOVRAM Test failed (GUI)NOVRAM data out of valid rangewarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.