Puritan Bennett 840: troubleshooting and error codes
The ServiceForge knowledge base holds 222 documented symptoms for the Puritan Bennett 840, with 568 ranked causes, 314 diagnostic checks with test points and expected values, and 571 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Exhalation valve calibration fails
Calibration procedure does not complete successfully; error messages displayed during Exp Valve Calibration function execution.
Ranked causes
- 1. System leak, particularly expiratory filter leaks (common)
- 2. Exhalation valve and/or analog interface PCB out of specification (occasional)
- 3. Analog Interface PCB out of specification (occasional)
- 4. Gold standard circuit not connected or expiratory filter not locked (occasional)
- 5. Exhalation valve temperature out of range (rare)
- 6. Expiratory filter leaks (common)
- 7. Flow sensor calibration table checksum error (rare)
- 8. No air connected to ventilator (occasional)
Verify ventilator warm-up time
- Test point
- Service mode
- Expected
- Ventilator has been in service mode for at least 10 minutes
3 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen zero flow check failed caution
Q1 oxygen flow sensor reads > 0.05 L/min during zero-flow check when supplies disconnected
Ranked causes
- 1. Gas supplies not disconnected (common)
- 2. Compressor still running (common)
- 3. Q2 O-ring leakage (affects Q1) (occasional)
- 4. PSOL2 internal leak (affects O2 circuit) (occasional)
- 5. PSOL2 seating improper (occasional)
- 6. Q1 sensor calibration drift (occasional)
- 7. PSOL2 failure (rare)
- 8. Q2 sensor failure (rare)
Read Q1 oxygen flow sensor during zero-flow condition
- Test point
- Q1 (oxygen flow sensor)
- Expected
- ≤0.05 L/min, ≤
- Tools
- EST system flow sensor reading
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air flow cannot establish at 120 L/min patient safety
Flow controller unable to establish and control air flow at 120 L/min during EST
Ranked causes
- 1. Air supply disconnected, restricted, or low pressure (common)
- 2. Air PSOL defective (occasional)
- 3. System leak or occlusion (occasional)
- 4. Flow sensor calibration out of specification (occasional)
- 5. Q2 air flow sensor defective (rare)
- 6. Inspiratory Electronics PCB defective (rare)
- 7. AI PCB defective (rare)
Verify air supply connected and unrestricted
- Test point
- air inlet
- Expected
- connected and clear
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air zero flow check failed caution
Q2 air flow sensor reads > 0.05 L/min during zero-flow check when supplies disconnected
Ranked causes
- 1. Gas supplies not disconnected (common)
- 2. Compressor still running (common)
- 3. Q2 O-ring leakage (occasional)
- 5. PSOL2 seating improper (occasional)
- 6. Q2 sensor calibration drift (occasional)
- 7. PSOL2 internal fault (rare)
- 8. Q2 sensor failure (rare)
Read Q2 air flow sensor during zero-flow condition
- Test point
- Q2 (air flow sensor)
- Expected
- ≤0.05 L/min, ≤
- Tools
- EST system flow sensor reading
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Flow Sensor
Automatic calibration of air, oxygen, and exhalation flow sensors with atmospheric pressure transducer zeroing using external barometer reference
Ranked causes
- 1. Air offset out of range during calibration (unknown)
- 2. O2 offset out of range during calibration (unknown)
- 3. Cannot achieve minimum air flow (60 L/min) (unknown)
- 4. Cannot achieve minimum O2 flow (60 L/min) (unknown)
- 5. No air connected (unknown)
- 6. No O2 connected (unknown)
- 7. AC power not connected (unknown)
Enter Service Mode
- Test point
- Service Mode menu
- Expected
- Service Mode accessible
7 more checks with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen flow cannot establish at 120 L/min patient safety
Flow controller unable to establish and control oxygen flow at 120 L/min during EST
Ranked causes
- 1. Oxygen supply disconnected, restricted, or low pressure (common)
- 2. Oxygen PSOL defective (occasional)
- 3. System leak or occlusion (occasional)
- 4. Flow sensor calibration out of specification (occasional)
- 5. Q1 oxygen flow sensor defective (rare)
- 6. Inspiratory Electronics PCB defective (rare)
- 7. AI PCB defective (rare)
Verify oxygen supply connected and unrestricted
- Test point
- oxygen inlet
- Expected
- connected and clear
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air PSOL command current out of range patient safety
PSOL2 command current exceeds acceptable range during air flow verification at 120, 60, 5, or 1 L/min
Ranked causes
- 1. Air supply disconnected or restricted (common)
- 2. Air supply quality poor (occasional)
- 3. System leaks or occlusions (occasional)
- 4. PSOL2 defective (occasional)
- 5. Flow sensor calibration needed (occasional)
- 6. Q1 or Q2 flow sensor defective (rare)
Verify air supply connected and unrestricted
- Test point
- air inlet
- Expected
- connected and clear
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not charging after AC power reconnection high voltage
BPS not charging after AC power was reconnected during EST Battery Test.
Ranked causes
- 1. AC power not properly reconnected or AC supply is bad (common)
- 2. Faulty BPS PCB (occasional)
- 3. Defective battery pack (occasional)
- 4. Faulty power supply (occasional)
- 5. Defective BD CPU (rare)
- 6. Faulty AI PCB (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 flow sensor cross check failed patient safety
Q1 and Q3 readings are too far apart during oxygen flow verification at 120, 60, 5, and 1 L/min
Ranked causes
- 1. Oxygen supply disconnected or restricted (common)
- 2. Oxygen supply quality poor (occasional)
- 3. System leaks or occlusions (occasional)
- 4. PSOL1 defective (occasional)
- 5. Flow sensor calibration out of specification (occasional)
- 6. Q1 oxygen flow sensor defective (occasional)
Verify oxygen supply connected and unrestricted
- Test point
- oxygen inlet
- Expected
- connected and clear
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 PSOL command current out of range patient safety
PSOL1 command current exceeds acceptable range during oxygen flow verification at 120, 60, 5, and 1 L/min
Ranked causes
- 1. Oxygen supply disconnected or restricted (common)
- 2. Oxygen supply quality poor (occasional)
- 3. System leaks or occlusions (occasional)
- 4. PSOL1 defective (occasional)
- 5. Flow sensor calibration needed (occasional)
- 6. Q1 or Q2 flow sensor defective (rare)
Verify oxygen supply connected and unrestricted
- Test point
- oxygen inlet
- Expected
- connected and clear
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen flow sensor cross-check failed patient safety
SST Flow Sensor Test failure: Q1 and Q3 readings are too far apart at oxygen flows of 120, 60, 5, and 1 L/min
Ranked causes
- 1. System leak (common)
- 2. Safety valve open (occasional)
- 3. Flow sensor calibration drift (occasional)
- 4. Q1 oxygen flow sensor malfunction (occasional)
- 5. Q3 exhalation flow sensor malfunction (occasional)
- 6. AI PCB malfunction (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air PSOL excessive pressure buildup 50-100 cmH2O caution
Air PSOL forward leak detected via excessive pressure in blocked inspiratory module, pressure between 50 and 100 cmH2O
Ranked causes
- 1. Leaking Q2 O-rings (common)
- 2. Air supply disconnected or restricted (occasional)
- 3. PSOL2 seating issue (occasional)
- 4. Air PSOL internal leak (occasional)
- 5. AI PCB fault (rare)
Measure inspiratory pressure with safety valve open and then closed
- Test point
- Inspiratory pressure sensor
- Expected
- 50-100 cmH2O (indicating forward leak)
- Tools
- Pressure transducer
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air PSOL forward leak detected patient safety
Excessive pressure buildup (> 50 cmH2O but < 100 cmH2O) detected in blocked inspiratory module with air PSOL enabled
Ranked causes
- 1. Q2 O-ring leaking (common)
- 2. Air supply not connected or restricted (common)
- 3. Air PSOL2 not fully seated (occasional)
- 4. Air PSOL2 defective (occasional)
- 5. AI PCB defective (rare)
Check Q2 O-rings for leaks
- Test point
- Q2 O-ring seals
- Expected
- dry, no visible leakage
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backup power supply voltage out of range
BPS voltage dropped below accepted level or dropped too quickly during discharge test
Ranked causes
- 1. Battery pack not fully charged before EST (common)
- 2. Battery pack defective (occasional)
- 3. BPS PCB control circuit failure (occasional)
- 4. Power supply unit failure (occasional)
- 5. AI PCB failure (rare)
Check BPS LED charge status before EST
- Test point
- BPS LED indicator
- Expected
- Green LED lit (fully charged)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backup not charging
BPS (Battery Power Supply) not charging after AC power reconnected
Ranked causes
- 1. AC power not properly reconnected (common)
- 2. BPS PCB charging circuit failure (occasional)
- 3. Battery pack defective (occasional)
- 4. Power supply unit failure (occasional)
- 5. BD CPU or AI PCB failure (rare)
Verify AC power supply and connection
- Test point
- AC power cord, outlet
- Expected
- AC power good and connected
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage drops during backup power test high voltage
During EST Battery Test, backup power supply (BPS) voltage dropped below accepted level or too quickly while discharging.
Ranked causes
- 1. Battery pack not fully charged (common)
- 2. Defective battery pack (occasional)
- 3. Faulty BPS PCB (occasional)
- 4. Faulty AI PCB (occasional)
- 5. Defective power supply (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation zero flow check failed caution
Q3 exhalation flow sensor reads > 0.1 L/min during zero-flow check when supplies disconnected
Ranked causes
- 1. Gas supplies not disconnected (common)
- 2. Compressor still running (common)
- 3. Q3 O-ring leakage (occasional)
- 4. Q3 sensor calibration drift (occasional)
- 5. Q3 sensor failure (rare)
Read Q3 exhalation flow sensor during zero-flow condition
- Test point
- Q3 (exhalation flow sensor)
- Expected
- ≤0.1 L/min, ≤
- Tools
- EST system flow sensor reading
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory and expiratory pressure cross-check failed at 10 cmH2O patient safety
Inspiratory and expiratory pressure transducer readings diverge beyond acceptable tolerance at 10 cmH2O test pressure.
Ranked causes
- 1. Patient circuit leak or occlusion (common)
- 2. Ventilator system leak or occlusion (common)
- 3. Inspiratory pressure transducer out of range (occasional)
- 4. Expiratory pressure transducer out of range (occasional)
- 5. AI PCB electronics failure (rare)
Verify patient circuit has no leaks or occlusions
- Test point
- Patient circuit system
- Expected
- No leaks or occlusions
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory and expiratory pressure cross-check failed at 50 and 100 cmH2O patient safety
Inspiratory and expiratory pressure transducer readings diverge beyond acceptable tolerance at high pressure levels (50 and 100 cmH2O).
Ranked causes
- 1. Expiratory bacteria filter clogged or restricted (common)
- 2. System leak or occlusion (common)
- 3. Inspiratory pressure transducer out of range (occasional)
- 4. Expiratory pressure transducer out of range (occasional)
- 5. AI PCB electronics failure (rare)
Inspect expiratory bacteria filter
- Test point
- Expiratory filter
- Expected
- Filter clean and unobstructed
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 zero flow check failed patient safety
Oxygen flow exceeds 0.153 L/min when PSOL1 is commanded to deliver 0 L/min (closed)
Ranked causes
- 1. Q1 O-ring leaking (common)
- 2. PSOL1 leaking (common)
- 3. PSOL1 defective (occasional)
- 4. Flow sensor calibration out of specification (occasional)
- 5. Q1 oxygen flow sensor defective (rare)
Verify no leaks at Q1 O-rings
- Test point
- Q1 O-ring seals
- Expected
- dry, no visible leakage
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen flow sensor cross-check failure patient safety
O2 flow sensor readings (Q1) and exhalation sensor readings (Q3) diverge beyond acceptable tolerance during oxygen flow test.
Ranked causes
- 1. System leak or occlusion (common)
- 2. Safety valve stuck open (occasional)
- 3. Q1 sensor malfunction (occasional)
- 4. Q3 sensor malfunction (occasional)
- 5. AI PCB electronics failure (rare)
Check for system leak
- Test point
- System circuit
- Expected
- No leaks detected
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen PSOL current out of range patient safety
Oxygen proportional solenoid (PSOL1) command current is outside acceptable range during diagnostic test.
Ranked causes
- 1. System leak or occlusion (common)
- 2. Safety valve stuck open (occasional)
- 3. Oxygen regulator pressure out of specification (occasional)
- 4. PSOL1 solenoid valve malfunction (occasional)
- 5. AI PCB electronics failure (rare)
Check for system leak
- Test point
- System circuit
- Expected
- No leaks detected
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen PSOL excessive pressure buildup 50-100 cmH2O caution
Oxygen PSOL forward leak detected via excessive pressure in blocked inspiratory module, pressure between 50 and 100 cmH2O
Ranked causes
- 1. Leaking Q1 O-rings (common)
- 2. Oxygen supply disconnected or restricted (occasional)
- 3. PSOL1 seating issue (occasional)
- 4. Oxygen PSOL internal leak (occasional)
- 5. AI PCB fault (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SST flow sensor test fails - unable to establish oxygen flow patient safety
Flow controller unable to establish and control oxygen flow at 120 L/min during SST test
Ranked causes
- 1. Oxygen supply not connected or unrestricted (common)
- 2. Patient circuit system has leaks or occlusions (common)
- 3. PSOL (proportional solenoid) malfunction (occasional)
- 4. Q1 or Q2 flow control valve malfunction (occasional)
- 5. Flow sensor calibration error (occasional)
Verify oxygen supply connection and pressure
- Test point
- PS1 pressure switch, oxygen regulator output
- Expected
- 9-12 psi at regulator; PS1 detects pressure
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test circuit not connected caution
Pressure not detected on expiratory side during leak test; inspiratory/expiratory pressure transducer readings too far apart
Ranked causes
- 1. Test circuit not properly connected (common)
- 2. Expiratory bacteria filter blocked or defective (occasional)
- 3. Q3 flow sensor O-ring leakage (occasional)
- 4. Exhalation valve malfunction (occasional)
- 5. Expiratory pressure transducer PCB failure (rare)
Compare inspiratory vs expiratory pressure readings during pressure hold
- Test point
- Inspiratory pressure sensor vs expiratory pressure sensor
- Expected
- Pressures closely match
- Tools
- Pressure transducers, EST system
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test circuit not connected (pressure transducer readings far apart)
Pressure not detected on expiratory side; inspiratory/expiratory pressure transducer readings too far apart
Ranked causes
- 1. Test circuit not properly connected (common)
- 2. Expiratory bacteria filter clogged (occasional)
- 3. Leaks at Q3 flow sensor (occasional)
- 4. Exhalation valve malfunction (occasional)
- 5. Expiratory pressure transducer PCB failure (rare)
Verify test circuit connection to expiratory side
- Expected
- Pressure detected on expiratory transducer
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to establish air flow at 120 L/min patient safety
Flow controller cannot establish and control air flow at 120 L/min during diagnostic test.
Ranked causes
- 1. Air supply disconnected or restricted (common)
- 2. Patient circuit leaks or occlusions (common)
- 3. Air regulator pressure out of range (occasional)
- 4. PSOL2 solenoid malfunction (occasional)
- 5. Q2 air sensor malfunction (occasional)
Verify air supply connection and restriction
- Test point
- Air inlet
- Expected
- Connected and unrestricted
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to establish air flow at 60, 5, and 1 L/min patient safety
Flow controller cannot establish and control air flow at lower rates (60, 5, 1 L/min) during diagnostic test.
Ranked causes
- 1. Air supply disconnected or restricted (common)
- 2. System leaks or occlusions (common)
- 3. Air regulator pressure out of specification (occasional)
- 4. PSOL2 solenoid malfunction (occasional)
- 5. Q3 exhalation sensor malfunction (occasional)
Verify air supply connection
- Test point
- Air inlet
- Expected
- Connected and unrestricted
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to establish oxygen flow patient safety
Flow controller cannot establish and control oxygen flow at 60, 5, and 1 L/min during diagnostic test.
Ranked causes
- 1. Oxygen supply disconnected or restricted (common)
- 2. System leaks or occlusions (common)
- 3. PSOL1 solenoid valve malfunction (occasional)
- 4. Q1 oxygen sensor malfunction (occasional)
- 5. Q3 exhalation sensor malfunction (occasional)
Verify oxygen supply connection and restriction
- Test point
- Oxygen inlet
- Expected
- Connected and unrestricted
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to establish oxygen flow at 120 L/min
Flow controller unable to establish and control oxygen flow at 120 L/min during SST flow sensor test.
Ranked causes
- 1. Oxygen supply disconnected, restricted, or low pressure (common)
- 2. Patient circuit system has leaks or occlusions (common)
- 3. Oxygen inlet filter assembly blocked (occasional)
- 4. Defective flow control solenoid Q1 (occasional)
- 5. Defective PSOL1 oxygen solenoid (occasional)
Verify oxygen supply is connected and flow is good
- Test point
- oxygen inlet
- Expected
- oxygen supply connected and unrestricted
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to establish oxygen flow at 60, 5, and 1 L/min patient safety
SST Flow Sensor Test failure: flow controller cannot establish and control oxygen flow at 60, 5, and 1 L/min (lower flows)
Ranked causes
- 1. Oxygen supply restricted or disconnected (common)
- 2. System leak or occlusion (occasional)
- 3. PSOL1 malfunction (occasional)
- 4. Flow sensor Q1 malfunction (occasional)
- 5. Exhalation sensor Q3 malfunction (rare)
Verify oxygen supply
- Test point
- Oxygen supply connection and regulated pressure
- Expected
- Connected and unrestricted; adequate supply
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vent Inop Test fails
Service mode test procedure fails with 'Vent Inop Test failed' message on lower screen during BDU alarm, LED, or PSOL/valve inactivation checks
Ranked causes
- 1. BDU alarm failure (GUI vent inop test phase) (unknown)
- 2. Ventilator inoperative or safety valve open LED failure (unknown)
- 3. BDU alarm and LED failures combined (unknown)
- 4. PSOL/exhalation valve/safety valve inactivation failure (unknown)
- 5. Failure in Vent Inop A, B, or ten-second test phases (unknown)
Verify BDU alarm operation in GUI vent inop test phase
- Test point
- BDU alarm
- Expected
- BDU alarm on (user prompt)
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air PSOL current out of range patient safety
Air proportional solenoid (PSOL2) command current is outside acceptable range during diagnostic test.
Ranked causes
- 1. Air supply disconnected or restricted (common)
- 3. Air supply pressure out of specification (occasional)
- 4. PSOL2 solenoid valve malfunction (occasional)
- 5. AI PCB electronics failure (rare)
Verify air supply connection and restriction
- Test point
- Air inlet
- Expected
- Connected and unrestricted
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery backup not discharging
BPS not discharging when loaded during test phase
Ranked causes
- 1. AC power not disconnected during test (common)
- 2. BPS PCB discharge circuit failure (occasional)
- 3. Battery pack defective or depleted (occasional)
- 4. BD CPU or AI PCB failure (rare)
Verify AC power is disconnected
- Test point
- AC power cord
- Expected
- AC power disconnected
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not fully charged
BPS not fully charged before EST begins
Ranked causes
- 1. Charging not allowed to complete (common)
- 2. BPS PCB charging circuit defective (occasional)
- 3. Battery pack defective (occasional)
- 4. BD CPU failure (rare)
Verify AC power connected and BPS charging
- Test point
- AC power cord, BPS LED
- Expected
- AC connected, BPS LED indicates charging
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
BD alarm cable voltage out of range
Alarm cable voltage is < 3.5 V or > 5.05 V
Ranked causes
- 1. BD alarm cable not connected (common)
- 2. BD alarm malfunction (occasional)
- 3. AI PCB malfunction (occasional)
- 4. BD alarm cable damaged (occasional)
Measure alarm cable voltage
- Expected
- 3.5 to 5.05 V, 3.5-5.05 V
- Tools
- DMM
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Initialize Flow Sensor
Automatic copying of flow sensor calibration data from each sensor's serial EEPROM to BDU NOVRAM when data mismatch detected
Ranked causes
- 1. Unable to read air flow sensor serial EEPROM (unknown)
- 2. Unable to read O2 flow sensor serial EEPROM (unknown)
- 3. Unable to read exhalation flow sensor serial EEPROM (unknown)
- 4. Unable to program BDU NOVRAM (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compliance low (below 1.56 adult / 1.34 pediatric, second occurrence)
Low compliance falls below 1.56 mL/cmH2O (adult) or 1.34 mL/cmH2O (pediatric) but not below 1.05 mL/cmH2O during SST Compliance calibration test.
Ranked causes
- 1. Wrong patient tubing type specified in SST (common)
- 2. Defective patient circuit or filter (occasional)
- 3. Defective pressure transducer (PI or PE) (occasional)
- 4. Atmospheric pressure transducer out of calibration (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compliance low (below 1.56 adult / 1.34 pediatric)
Compliance falls below 1.56 mL/cmH2O (adult) or 1.34 mL/cmH2O (pediatric) but not lower than 1.05 mL/cmH2O during SST Compliance calibration test.
Ranked causes
- 1. Wrong patient tubing type specified in SST (common)
- 2. Defective patient circuit or filter (occasional)
- 3. Defective pressure transducer (PI or PE) (occasional)
- 4. Atmospheric pressure transducer out of calibration (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive compliance (>6 mL/cmH2O adult / 4.5 pediatric but <12/9)
Low compliance >6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric) but <12 mL/cmH2O (adult) or 9 mL/cmH2O (pediatric) during SST Compliance calibration test.
Ranked causes
- 1. Wrong patient tubing type specified in SST (common)
- 2. Defective patient circuit or filter (occasional)
- 3. Defective pressure transducer (PI or PE) (occasional)
- 4. Atmospheric pressure transducer out of calibration (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive system leak detected patient safety
System pressure drops excessively during leak test, indicating significant leak or occlusion in the patient circuit or system.
Ranked causes
- 1. Patient circuit not connected (common)
- 2. Patient circuit leak (common)
- 3. Exhalation valve malfunction (occasional)
- 4. Expiratory filter connection loose (occasional)
Verify test circuit is connected
- Test point
- Test circuit connector
- Expected
- Circuit properly connected and seated
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Excessive system leak during leak test caution
System pressure dropped below 85 cmH2O or pressure drops ≥5 cmH2O within 10 seconds after reaching 85 cmH2O
Ranked causes
- 1. Test circuit not connected or leaking (common)
- 2. System internal leaks (common)
- 3. Expiratory filter connection loose (occasional)
- 4. Exhalation valve malfunction (occasional)
Wait for inspiratory pressure decay from 90 cmH2O or wait 10 seconds
- Test point
- Inspiratory pressure sensor
- Expected
- Pressure drops to 85 cmH2O or stable
- Tools
- Pressure transducer, EST system
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation valve seal leaking caution
Exhalation valve seal test shows pressure differential too high, indicating valve diaphragm not sealing properly
Ranked causes
- 1. Exhalation valve diaphragm contaminated (common)
- 2. Valve calibration incorrect (common)
- 3. Flow sensor Q3 miscalibrated (occasional)
- 4. Exhalation valve diaphragm damaged (occasional)
Clean exhalation valve diaphragm
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow Sensor Calibration
Calibration of exhalation flow sensor (Q3) offset table using gold standard test circuit and dual gas sources (air and oxygen)
Ranked causes
- 1. Test circuit not attached or expiratory filter not secure (common)
- 2. System leak or expiratory filter leak (common)
- 3. Ventilator not at operating temperature (occasional)
- 4. Ventilator failed EST leak test (occasional)
Verify ventilator has reached operating temperature
- Test point
- Ventilator thermal status
- Expected
- Operating temperature reached after 10 minutes warm-up in service mode
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow sensor calibration fails
Flow Sensor Calibration procedure does not complete; error messages displayed during calibration execution.
Ranked causes
- 1. System leak, particularly expiratory filter leaks (common)
- 2. Test circuit not attached or not secure (occasional)
- 3. Ventilator has not reached operating temperature (occasional)
- 4. Flow sensor not passing EST leak test (occasional)
Verify ventilator warm-up time
- Test point
- Service mode operation
- Expected
- Ventilator has operated in service mode for at least 10 minutes
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Initialize Flow Sensor errors
Automatic flow sensor calibration data initialization fails with error messages during service mode startup
Ranked causes
- 1. Unable to read air flow sensor (Q2) EEPROM (unknown)
- 2. Unable to read oxygen flow sensor (Q1) EEPROM (unknown)
- 3. Unable to read exhalation flow sensor (Q3) EEPROM (unknown)
- 4. Unable to program BDU NOVRAM flash memory (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory and expiratory pressure transducer readings too far apart patient safety
Cross-check between inspiratory and expiratory pressure transducers at 50 and 100 cmH2O fails; readings diverge beyond acceptable tolerance
Ranked causes
- 1. Expired bacteria filter clogged or defective (common)
- 2. System leaks or occlusions (common)
- 3. Inspiratory or expiratory pressure transducer out of range (occasional)
- 4. AI PCB defective (occasional)
Verify no system leaks or occlusions
- Test point
- circuit
- Expected
- no leaks or occlusions
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inspiratory limb resistance low
Inspiratory limb pressure <0.6 cmH2O (adult) or 5.5 cmH2O (pediatric) but >0.2 cmH2O during SST Circuit Resistance test.
Ranked causes
- 1. Inspiratory filter not installed (common)
- 2. Test procedure not followed correctly (occasional)
- 3. Defective inspiratory filter (occasional)
- 4. Defective patient circuit (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen PSOL forward leak detected (pressure 50-100 cmH2O) caution
Excessive pressure buildup (> 50 cmH2O but < 100 cmH2O) detected in blocked inspiratory module
Ranked causes
- 1. Oxygen PSOL forward leak (common)
- 2. Leaks around Q1 O-rings (occasional)
- 3. Restricted oxygen supply (occasional)
- 4. AI PCB malfunction (rare)
Block inspiratory module and measure pressure
- Expected
- Pressure between 50-100 cmH2O indicates forward leak
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power failure capacitor voltage out of range caution
Initial power failure capacitor voltage is < 4.5 V or > 5.05 V
Ranked causes
- 1. BD alarm cable not connected (common)
- 2. BD alarm malfunction (occasional)
- 3. AI PCB malfunction (occasional)
- 4. Motherboard PCB failure (rare)
Measure capacitor voltage
- Expected
- 4.5 to 5.05 V, 4.5-5.05 V
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure transducer cross-check failed at 10 cmH2O patient safety
SST Circuit Pressure Test failure: inspiratory and expiratory pressure transducer readings are too far apart at 10 cmH2O
Ranked causes
- 1. Patient circuit leak or occlusion (common)
- 2. System leak or occlusion (occasional)
- 3. Pressure transducer out of range (occasional)
- 4. AI PCB malfunction (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressure transducer cross-check failed at 50 or 100 cmH2O patient safety
SST Circuit Pressure Test failure: inspiratory and expiratory pressure transducer readings are too far apart at 50 or 100 cmH2O
Ranked causes
- 1. Expiratory bacteria filter clogged (common)
- 2. System leak or occlusion (occasional)
- 3. Pressure transducer out of range (occasional)
- 4. AI PCB malfunction (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Safety valve pressure relief failed patient safety
System unable to establish stable flow or safety valve cracking pressure is out of specification
Ranked causes
- 1. To patient port not blocked during test (common)
- 2. Leak at O2 sensor, SOL1, or PI (occasional)
- 3. Safety valve defective (occasional)
- 4. PI pressure transducer defective (occasional)
Verify To patient port is blocked
- Test point
- To patient port
- Expected
- blocked
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to establish leak test pressure caution
System cannot attain 90 cmH2O starting pressure during leak test within timeout period with either oxygen or air
Ranked causes
- 1. Test circuit not connected (common)
- 2. System leaks present (common)
- 3. Exhalation valve malfunction (occasional)
- 4. Expiratory filter connection loose (occasional)
Pressurize system to 90 cmH2O using 2 L/min flow with oxygen or air
- Test point
- Inspiratory pressure sensor
- Expected
- ≥90 cmH2O, ≥
- Tools
- Pressure transducer, EST system
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to establish test pressure
System cannot attain 90 cmH2O pressure with oxygen or air at 2 L/min flow within timeout
Ranked causes
- 1. Test circuit not connected (common)
- 2. System leaks present (common)
- 3. Exhalation valve not functioning (occasional)
- 4. Expiratory filter not securely connected (occasional)
Verify test circuit connection
- Expected
- Test circuit properly connected to system
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vent Inop Test failure during GUI vent inop test phase
GUI vent inop test phase fails during Vent Inop Test execution; BDU alarm, LED indicators, or PSOLs/exhalation valve/safety valve may not function correctly
Ranked causes
- 1. BDU alarm failure (occasional)
- 2. Ventilator inoperative or SVO LED failure (occasional)
- 3. BDU-GUI cable or CPU PCB communication failure (occasional)
- 4. BD CPU PCB or AI PCB failure (occasional)
Observe BDU alarm during GUI vent inop test phase
- Test point
- BDU alarm output
- Expected
- Alarm activates on user prompt
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC power not connected for EST patient safety
EST (Equipment System Test) cannot run on battery power; AC power required
Ranked causes
- 1. Ventilator not plugged into AC outlet (common)
- 2. Battery Power Supply (BPS) active (occasional)
- 3. Power supply defective (rare)
Verify AC power connection
- Test point
- Ventilator power inlet
- Expected
- AC power cord connected to facility outlet
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air flow cannot establish at 60, 5, and 1 L/min patient safety
Flow controller unable to establish and control air flow at lower rates during EST
Ranked causes
- 1. Air supply disconnected or low pressure (common)
- 2. PSOL2 or Q2 air flow sensor defective (occasional)
- 3. AI PCB defective (rare)
Verify air supply connected
- Test point
- air inlet
- Expected
- connected
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air flow sensor cross-check failed patient safety
Air flow sensor (Q2) and exhalation sensor (Q3) readings diverge beyond acceptable tolerance during air flow test.
Ranked causes
- 2. Ventilator system leak or occlusion (common)
- 4. PSOL2 solenoid malfunction (occasional)
- 5. Q2 air sensor malfunction (occasional)
Inspect patient circuit for leaks or occlusions
- Test point
- Patient circuit
- Expected
- No leaks or occlusions; no debris
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air PSOL forward leak at high pressure patient safety
Excessive pressure buildup (> 100 cmH2O) detected in blocked inspiratory module with air PSOL enabled
Ranked causes
- 1. Q2 O-ring leaking (common)
- 2. Safety valve not opening at correct pressure (occasional)
- 3. Air PSOL2 defective (occasional)
Verify safety valve cracks above 100 cmH2O
- Test point
- safety valve outlet
- Expected
- relief heard at pressure > 100 cmH2O on display
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Atmospheric Pressure Transducer calibration out of range
- Battery pack system error
- BD communication loss
- Compressor load test failed
- Compressor pressure not detected
- Excessive circuit leak detected
- Excessive leak in system (pressure drops >= 5 cmH2O in 10 seconds)
- Excessive system leak
- Exhalation valve temperature out of specification
- Expiratory pressure transducer autozero out of range
- Expiratory pressure transducer autozero solenoid SOL2 bad
- Failed to reach test pressure
- Failed to reach test pressure (10 cmH2O at 5 L/min)
- GUI alarm sound not working
- Inspiratory pressure transducer autozero out of range
- Inspiratory pressure transducer autozero solenoid SOL1 bad
- Inspiratory/expiratory pressure transducer cross-check failed
- LCD controller PCB fault detected
- Low circuit compliance
- Low oxygen percentage (↓O2%)
- O2 pressure detected with supply disconnected
- Oxygen flow cannot establish at 60, 5, and 1 L/min
- Oxygen pressure detected when disconnected
- Power supply voltage error
- Touch screen input failure
- Touchscreen error occurred
- Wall air pressure detected when disconnected
- Wall air pressure detected with supply disconnected
- AC power loss
- Air PSOL excessive pressure buildup >100 cmH2O
- Bad expiratory autozero solenoid
- Bad inspiratory autozero solenoid
- Cal Info Duplication errors
- Calibration: Cal Info Duplication
- Calibration: Power on self test (POST)
- Calibration: Serial Number Setup
- Circuit disconnect
- Compressor leak detected
- Compressor run time out of range
- Excessive circuit compliance
- Excessive compliance (high)
- Expiratory autozero out of range
- Failed to reach test pressure (50 or 100 cmH2O)
- Failed to reach test pressure at 10 cmH2O
- Failed to reach test pressure at 50 and 100 cmH2O
- GUI 100% O2 LED fails
- GUI 100% O2/CAL key fails
- GUI Accept key fails
- GUI Alarm Reset key fails
- GUI Alarm Silence key fails
- GUI Alarm Silence LED fails
- GUI Alarm Volume key fails
- GUI Battery Backup LED fails
- GUI Clear key fails
- GUI Compressor Operating LED fails
- GUI Compressor Ready LED fails
- GUI CPU not operational
- GUI Expiratory Pause key fails
- GUI High Alarm LED fails
- GUI Info key fails
- GUI Inspiratory Pause key fails
- GUI keyboard key malfunction
- GUI knob malfunction
- GUI knob not responding
- GUI LED indicator not illuminated
- GUI Low Alarm LED fails
- GUI Manual Inspiration key fails
- GUI Medium Alarm LED fails
- GUI Normal LED fails
- GUI On Battery Power LED fails
- GUI Screen brightness key fails
- GUI Screen contrast key fails
- GUI Screen lock key fails
- GUI Screen Lock LED fails
- High tidal volume (ΔVE TOT)
- Low AC power
- Low delivered tidal volume (ΔVE MAND)
- Low expiratory filter pressure drop
- Low inspiratory limb resistance
- No air supply
- Nurse call relay stuck off
- Nurse call relay stuck on
- O2 pressure not detected
- Occluded exhalation compartment
- Occluded expiratory compartment
- Oxygen pressure not detected
- Oxygen PSOL excessive pressure buildup >100 cmH2O
- Oxygen PSOL forward leak detected (pressure > 100 cmH2O)
- Wall air pressure not detected
- Wall air pressure still detected after disconnect
- Watchdog timer failure
- AC power loss detected
- Air pressure not detected via PS2 or PC
- Air supply loss with O2 at 100%
- Assertion boundary check failure in background diagnostics
- Bad GUI serial port
- Bus error or access fault
- Calibration error during Flow Sensor Calibration
- Calibration: Datakey Update
- Calibration: Exhalation system
- Calibration: Expiratory valve
- Calibration: O2 sensor
- Calibration: Oxygen sensor
- Calibration: Performance verification
- Calibration: Post-service verification
- Calibration: Regulator calibration hose repair
- Calibration: Remove or install 10.4-inch GUI
- Calibration: Serial Loopback Test
- Calibration: System Leak and Compliance Characterization (SST)
- Cannot access service mode
- Cannot program flash error
- Compressor disabled mode time out of range
- Compressor pressure detected during EST
- Compressor pressure detected when depressurized
- Compressor pressure detected when disabled
- Compressor standby mode time out of range
- Datakey not installed or not detected
- Datakey update procedure
- Electrical safety testing
- Exhalation valve leak or EST failure
- Expiratory limb resistance low
- Gas not connected
- GUI cable connection verification
- High oxygen percentage (↑O2%)
- Inspiratory autozero out of range
- Loss of GUI LED fails
- Low battery warning during AC loss
- Low exhalation limb resistance
- No O2 supply
- No ventilation with safety valve open
- O2 sensor out of calibration or failure
- O2 supply loss with FiO2 at 21%
- Occluded exhalation limb
- Occluded expiratory filter
- Occluded inspiratory limb
- Oxygen pressure not detected via PS1
- Performance verification testing required
- Performance verification: calibration and testing
- POST major fault detected; ventilator inoperative
- POST minor fault detected during self-test
- Procedure error during patient setup
- Replace 10.4-inch GUI front housing
- Replace 10.4-inch GUI keyboard assembly
- Screen block or touch screen fault
- Serial number does not match datakey
- Service mode functions not executing
- Severe occlusion in ventilator circuit
- SST flow sensor test fails - air only available
- SST flow sensor test fails - no gas available
- SST flow sensor test fails - oxygen only available
- SVO LED fails
- System unable to establish pressure
- Total loss of power
- Troubleshooting ventilator diagnostic codes
- Unable to establish stable expiratory filter flow
- Unable to pressurize circuit during compliance calibration
- Unable to reach minimum peak flow
- Unable to reach minimum peak flow (alert threshold)
- Unable to reach minimum peak flow (threshold 80-60 L/min)
- Ventilator inoperative LED fails
- Volume not delivered on VC+ or VS breaths
- Wye not blocked during circuit resistance test
162 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0B | Alarm Cable Voltage test failure | POST analog device test failure | error |
0C | O2 PSOL Current test failure | POST analog device test failure | error |
0D | Air PSOL Current test failure | POST analog device test failure | error |
0E | Low Voltage Reference test failure | POST analog device test failure | error |
0F | Atmospheric Pressure Transducer test failure | POST analog device test failure | error |
10 | Exhalation Coil Temperature test failure | POST analog device test failure | error |
11 | Exhalation Pressure Transducer test failure | POST analog device test failure | error |
12 | O2 Sensor test failure | POST analog device test failure | error |
12 Vdc Overvoltage | 12V power supply overvoltage (GUI only) | Excessive 12V output | error |
12 Vdc Undervoltage | 12V power supply undervoltage (GUI only) | Insufficient 12V output | error |
13 | GUI 5V dc Sentry test failure | POST analog device test failure | error |
14 | 12V dc Sentry test failure | POST analog device test failure | error |
15 | Safety Valve Current test failure | POST analog device test failure | error |
16 | +15V Sentry test failure | POST analog device test failure | error |
17 | -15V Sentry test failure | POST analog device test failure | error |
18 | Power Fail Cap Voltage test failure | POST analog device test failure | error |
19 | Exhalation Manifold Heater Temperature test failure | POST analog device test failure | error |
1A | BPS Battery Voltage Signal test failure | POST analog device test failure | error |
1B | 5V dc Venthead test failure | POST analog device test failure | error |
1C | BPS Battery Current test failure | POST analog device test failure | error |
1D | ac Line Voltage test failure | POST analog device test failure | error |
1E | Exhalation Motor Current test failure | POST analog device test failure | error |
1F | 10V Sentry test failure | POST analog device test failure | error |
5 Vdc Overvoltage | 5V power supply overvoltage (GUI only) | Excessive 5V output | error |
Access Fault/Bus Error | Memory access fault or bus error | Invalid memory access during POST | error |
AC POWER LOSS | Operating on battery; AC not available | Power switch on, AC unavailable, ventilator operating on BPS | warning |
AC POWER LOSS | AC mains power unavailable | AC power disconnected or failed; ventilator switches to BPS battery power | error |
AC POWER LOSS | AC input power is unavailable or insufficient; ventilator switched to battery power | AC monitor voltage falls below equivalent of 80 V AC at bulk supply input | warning |
ac power not connected | AC power is not supplied to ventilator during exhalation valve calibration | Exhalation valve calibration procedure detects no AC power | critical |
AC power not connected | Calibration cannot proceed; AC power is not connected to ventilator | Calibration attempted without AC power supply | error |
AE0110 | Unable to establish oxygen flow at 120 L/min | Flow controller unable to establish and control oxygen flow at 120 L/min | error |
AE0111 | Unable to establish air flow at 120 L/min | Flow controller unable to establish and control air flow at 120 L/min | error |
AE0213 | Air PSOL leak | EST Gas Supply/SV Test | warning |
AE0213 | Air PSOL leak: Excessive pressure buildup 50-100 cmH2O in blocked inspiratory module | EST disables compressor, bleeds pressure, reads inspiratory pressure; forward leak detected via buildup > 50 but < 100 cmH2O | warning |
AE0214 | Oxygen PSOL forward leak detected, pressure 50-100 cmH2O | Excessive pressure buildup in blocked inspiratory module during safety valve test | warning |
AE0306 | Test circuit not properly connected or expiratory pressure not detected | Inspiratory and expiratory pressure transducer readings too far apart during leak test | warning |
AE0601 | GUI High Alarm LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0602 | GUI Medium Alarm LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0603 | GUI Low Alarm LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0604 | GUI Normal LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0605 | GUI Batt Backup LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0606 | GUI On Batt Pwr LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0607 | GUI Compressor Ready LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0608 | GUI Compressor Operating LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0609 | GUI 100% O2 LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0610 | GUI Alarm Silence LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0611 | GUI Screen Lock LED fails | CLEAR key pressed to indicate LED not on | warning |
AE0702 | Bad Vent inop LED | EST BD Lamp Test | warning |
AE0703 | Bad SVO LED | EST BD Lamp Test | warning |
AE0704 | Loss of GUI LED fails | Loss of GUI LED not illuminating on backup display unit | warning |
AE1001 | Air PSOL loopback current out of range | Air PSOL (PSOL2) loopback current outside expected range at various drive currents | error |
AE1002 | O2 PSOL loopback current out of range | Oxygen PSOL (PSOL1) loopback current outside expected range at various drive currents | error |
AE1104 | Insp check valve test failed | EST Safety System Test | warning |
AE1201 | Exp valve loopback current out of range | Exhalation valve loopback current outside expected range of drive current at test openings | warning |
AE1305 | Seal test failed | Exhalation valve seal test pressure delta above alert level but below failure level | warning |
AE1600 | Compressor Test - Not installed | EST Compressor Test | info |
AE1601 | Wall air pressure detected when should be disconnected | PS2 pressure switch closed when should be open during compressor test setup | error |
AE1602 | AC power not connected - compressor cannot run | System running on battery power when AC power should be available; BPS charged/charging active | warning |
AE1603 | Compressor pressure not detected | Pressure transducer PC indicates no compressor air present although compressor motor is on | error |
AE1604 | Run mode time OOR | EST Compressor Test | warning |
AE1604 | Compressor run time out of range | Calculated run elapsed time not within acceptable range after 15 s delay | error |
AE1606 | Compressor pressure present after motor disabled | Compressor pressure transducer indicates air present after motor disabled and accumulator drained | error |
AE1607 | Compressor disabled time out of range | Calculated disabled elapsed time not within acceptable range after 15 s delay | error |
AE1608 | Unable to test compressor standby mode | Compressor timer could not be verified as functional in prior run and disabled tests | error |
AE1609 | Compressor standby time out of range | Calculated standby mode elapsed time not within acceptable range; compressor motor still running during standby test phase | error |
AE1610 | Unable to perform compressor load test | EST Compressor Test | warning |
AE1611 | Compressor load test failed - pressure not maintained | Pressure transducer PC indicates compressor unable to maintain minimum pressure during worst-case breath delivery waveform (200 lpm at 45 cmH2O) | error |
AE1700 | Compressor Leak Test - Not installed | EST Compressor Leak Test | info |
AE1701 | Unable to perform compressor leak test | Compressor timer tests not passed during prior Compressor Test | error |
AE1702 | Wall air pressure detected during leak test disable phase | Pressure switch PS2 closed when should be open after disconnecting wall air | error |
AE1703 | Wall air pressure not detected during leak test enable phase | Pressure switch PS2 open when should be closed after reconnecting wall air | error |
AE1704 | Compressor leak detected | Compressor pressure transducer PC indicates loss of compressor pressure during 1-minute observation interval in standby mode, or compressor elapsed time > 1 s in standby | error |
AE1901 | GUI touch: Error | EST GUI Touch Test | warning |
AE2001 | Bad GUI serial port | EST GUI Serial Port Test | warning |
AE2001 | Serial port loopback test failed | In loopback mode, received message did not match transmitted | error |
AE2101 | Backup power supply not fully charged | BPS not fully charged at start of test | error |
AE2102 | Battery pack not discharging | BPS not discharging after AC power disconnected | error |
AE2103 | Bad Backup Power Supply - voltage issue | BPS voltage dropped below accepted level or dropped too quickly during discharge | error |
AE2104 | Battery not charging after AC reconnection | BPS not charging after AC power reconnected (BPS charged bit on and BPS charging bit off) | error |
AE2300 | GUI Nurse Call Test - Not installed | User pressed CLEAR to indicate nurse call device not installed | info |
Air offset out of range | Air flow sensor calibration failed; measured offset exceeds acceptable range | Calibration procedure detects invalid air sensor offset during Flow Sensor Calibration | error |
Analog Interface Error | Analog interface failure during POST | error | |
APNEA | Set apnea interval elapsed without triggering a breath | Apnea interval elapsed without ventilator, patient, or operator triggering a breath | warning |
APNEA (lockable) | No breath triggered within apnea interval | Set apnea interval elapses without patient, ventilator, or operator triggering breath | warning |
AS0010 | Unable to establish O2 flow | Flow controller unable to establish and control oxygen flow at 120 L/min during SST | error |
AS0011 | Unable to establish air flow | Flow controller unable to establish and control air flow at 120 L/min during SST | error |
AS0012 | Oxygen pressure not detected | Only air detected during SST flow sensor test | error |
AS0013 | Wall air pressure not detected | Only oxygen detected during SST flow sensor test | error |
AS0202 | Excessive leak | Pressure drops to 85 cmH2O in 10 s, then in 10 s more by ≥10 cmH2O during SST circuit leak test | error |
AS0202 | Excessive system leak (patient circuit test) | Pressure drops to 85 cmH2O in 10 s, then by ≥ 10 cmH2O in next 10 seconds during patient circuit leak test | error |
AS0206 | Test circuit not connected | Inspiratory and expiratory pressure transducer readings diverge excessively during leak test pressure measurements | warning |
AS0305 | Occluded expiratory filter | Pressure drop across filter 2-3 cmH2O (alert threshold) | warning |
AS0308 | Occluded exhalation compartment | Exhalation compartment pressure >3 cmH2O but <4 cmH2O during SST expiratory filter test | error |
AS0308 | Occluded expiratory compartment | Exhalation compartment pressure >3 but <4 cmH2O | warning |
AS0311 | Low expiratory filter pressure drop | Pressure drop across expiratory filter <0.4 but >0.1 cmH2O | warning |
AS0403 | Occluded inspiratory limb | Inspiratory limb pressure 8.5-12.5 cmH2O (adult) or 5.5-7.5 cmH2O (pediatric) | warning |
AS0406 | Occluded exhalation limb | Expiratory limb pressure 8.5-12.5 cmH2O (adult) or 5.5-7.5 cmH2O (pediatric) | warning |
AS0407 | Peak flow below minimum during circuit test | Peak flow < 80 L/min but > 60 L/min during characterization | warning |
AS0407 | Unable to reach minimum peak flow | Peak flow 60-80 L/min (alert threshold) during inspiratory limb resistance test | warning |
AS0411 | Unable to reach minimum peak flow | Peak flow < alert threshold but > failure threshold during circuit resistance test | warning |
AS0411 | Peak flow below alert threshold | Peak flow below alert threshold during circuit characterization | error |
AS0413 | Insp limb resistance low | Inspiratory limb pressure <0.6 cmH2O (adult) or 5.5 cmH2O (pediatric) but >0.2 cmH2O during SST circuit resistance test | warning |
AS0414 | Expiratory limb resistance low | Expiratory limb pressure 0.2-0.6 cmH2O (adult) or 0.2-0.5 cmH2O (pediatric) | warning |
AS0505 | Patient circuit compliance abnormally high | Compliance > 6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric), < 12/9 mL/cmH2O | error |
AS0505 | Excessive compliance | High compliance >6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric) but <12 mL/cmH2O (adult) or 9 mL/cmH2O (pediatric) during SST compliance calibration test | error |
AS0507 | Compliance low | High compliance falls below 1.56 mL/cmH2O (adult) or 1.34 mL/cmH2O (pediatric) but not lower than 1.05 mL/cmH2O during SST compliance calibration test | error |
AS0509 | Excessive compliance | Low compliance >6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric) but <12 or 9 mL/cmH2O respectively | warning |
AS0509 | Patient circuit compliance abnormally high | Compliance > 6 mL/cmH2O (adult) or 4.5 mL/cmH2O (pediatric), < 12/9 mL/cmH2O | error |
AS0511 | Compliance low | Low compliance <1.56 mL/cmH2O (adult) or 1.34 mL/cmH2O (pediatric) but not below 1.05 mL/cmH2O | warning |
Bad exp valve loopback current | Exhalation valve loopback current measurement is out of specification during calibration | Exhalation valve calibration detects abnormal loopback current | error |
Both insp and exp autozero failed | Both inspiration and expiration pressure transducer autozero operations failed during calibration | Autozero calibration step fails for both transducers | error |
Calibration failed: current limit exceeded | Current supplied to exhalation valve exceeded allowable limit during calibration | Current measurement during exhalation valve calibration exceeds specification | error |
Calibration failed: gain resolution | Exhalation valve calibration gain adjustment resolution exceeded | During exhalation valve gain calibration step | error |
Calibration failed: gain resolution | Gain resolution calculation failed during exhalation valve calibration | Calibration algorithm detects gain resolution error | error |
Calibration failed: gain resolution | DAC gain or resolution measurement out of specification | Calibration gain characterization step | error |
Calibration failed: Pressure build time-out | Ventilator unable to build required test pressure within expected time during calibration | Pressure build step times out during exhalation valve calibration | error |
Cannot achieve minimum air flow | System cannot maintain minimum 60 L/min air flow during calibration | Flow Sensor Calibration procedure unable to sustain minimum air flow | error |
Cannot achieve minimum O2 flow | System cannot maintain minimum 60 L/min oxygen flow during calibration | Flow Sensor Calibration procedure unable to sustain minimum oxygen flow | error |
Cannot program flash | BDU and/or GUI NOVRAM cannot be successfully programmed | Attempted datakey or serial number programming in service mode | error |
CIRCUIT DISCONNECT | No ventilation; circuit disconnected | Unintended power loss recovery with circuit disconnect detected; or circuit disconnect detected during operation | critical |
CIRCUIT DISCONNECT (lockable) | Patient circuit disconnected or sensing failure | Ventilator detects loss of circuit continuity | error |
COMPLIANCE LIMITED VT | Compliance volume exceeds maximum allowed | Compliance volume required for volume-controlled breath exceeds maximum for 3 of last 4 breaths | warning |
COMPLIANCE LIMITED VT (lockable) | Excessive compliance volume required for VC breath delivery | Compliance volume required exceeds maximum allowed for 3 of last 4 breaths | warning |
COMPRESSOR INOPERATIVE | Compressor not operating due to low AC | Compressor unable to operate; low AC power | warning |
Data key is not installed | Datakey not detected by ventilator | Datakey insertion or boot-up in service mode | error |
DEVICE ALERT | Background check or POST detected a problem | Power-On Self-Test (POST) or background diagnostic failure | warning |
DRAM Parity Error | Parity check failure in DRAM | Memory data corruption | error |
DT0002 | Bus error / Access fault | Failure that results in a POST or a reset | critical |
Dxxxxx | Assertion Background Checks (BD) | Failure that results in a POST or a reset | critical |
Ethernet Parity Error | Parity check failure in Ethernet | Data corruption in Ethernet transmission | error |
Exhalation pressure transducer autozero failed | Exhalation pressure transducer DAC count at 0 cmH2O is out of range during autozero | Autozero calibration step fails for exhalation transducer | error |
Exhalation valve temperature out of range | Exhalation valve temperature outside acceptable operating range during calibration | Temperature sensor reading during calibration | error |
Exxxxx | Intentional system status event | Status message indicating intentionally caused event | info |
Exxxxx | Status message from GUI Background Checks | Status message indicating an intentionally caused event | info |
FE0001 | Inspiratory autozero out of range | EST Circuit Pressure Test: Inspiratory pressure transducer ADC count at 0 cmH2O out of range | error |
FE0002 | Expiratory autozero out of range | EST Circuit Pressure Test: Exhalation pressure sensor ADC count at 0 cmH2O out of range | error |
FE0003 | Failed to reach test pressure | Unable to build pressure (air) to 10 cmH2O at 5 L/min within time-out period during EST circuit pressure test | error |
FE0004 | Circuit pressure cross-check failed at 10 cmH2O | Inspiratory and expiratory pressure transducer readings too far apart during 10 cmH2O test | error |
FE0005 | Bad inspiratory autozero solenoid | Inspiratory pressure reading out of range (-0.60 to 0.60 cmH2O) when SOL1 opened | error |
FE0006 | Bad expiratory autozero solenoid | Expiratory pressure reading out of range (-0.60 to 0.60 cmH2O) when SOL2 opened | error |
FE0007 | Cross-check failed: Inspiratory/expiratory pressure transducers too far apart | Test reads inspiratory and expiratory pressure transducers at 50 and 100 cmH2O; readings exceed tolerance | error |
FE0007 | Cross-check failed | Inspiratory/expiratory pressure transducer measurements at 50 or 100 cmH2O test pressure are too far apart | error |
FE0007 | Circuit pressure cross-check failed at 50 and 100 cmH2O | Inspiratory and expiratory pressure transducer readings too far apart during 50 and 100 cmH2O test | error |
FE0008 | Failed to reach test pressure | Unable to build pressure to 50 or 100 cmH2O at 5 L/min within timeout period | error |
FE0010 | AC power not connected | EST Circuit Pressure Test: System running on battery after user prompted to connect AC power | critical |
FE0101 | Oxygen flow sensor cross-check failed | Q1 and Q3 flow sensor readings too far apart at oxygen flows of 120, 60, 5, and 1 L/min | error |
FE0102 | Oxygen PSOL command current out of range | PSOL1 command current out of range at oxygen flows of 120, 60, 5, and 1 L/min | error |
FE0103 | Air flow sensor cross-check failed | Q2 and Q3 flow sensor readings too far apart at air flows of 120, 60, 5, and 1 L/min | error |
FE0104 | Air PSOL command current out of range | PSOL2 command current out of range at air flows of 120, 60, 5, or 1 L/min | error |
FE0106 | Unable to establish O2 flow: Flow controller cannot control oxygen at lower rates | EST test verifies oxygen flow at 60, 5, and 1 L/min; flow controller unable to establish and control | error |
FE0107 | Unable to establish air flow: Flow controller cannot control air at lower rates | EST test verifies air flow at 60, 5, and 1 L/min; flow controller unable to establish and control | error |
FE0108 | O2 zero flow check failed: Oxygen flows when commanded to close | EST commands oxygen PSOL (PSOL1) to 0 L/min (closed); measured oxygen flow > 0.153 L/min | error |
FE0109 | Air zero flow check failed: Air flows when commanded to close | EST commands air PSOL (PSOL2) to 0 L/min (closed); measured air flow > 0.153 L/min | error |
FE0204 | Wall air pressure not detected | ACCEPT key pressed but pressure switch PS2 still open after prompting user to connect wall air | error |
FE0205 | Oxygen pressure not detected | ACCEPT key pressed but pressure switch PS1 still open after prompting user to connect oxygen | error |
FE0206 | O2 pressure detected: PS1 remains closed when oxygen disconnected | EST prompts disconnect oxygen; PS1 should open but remains closed | error |
FE0207 | Air PSOL leak: Excessive pressure buildup > 100 cmH2O in blocked inspiratory module | EST disables compressor, bleeds pressure, reads inspiratory pressure; forward leak detected via buildup > 100 cmH2O | error |
FE0207 | Air PSOL leak | Air PSOL (PSOL2) forward leak detected via excessive pressure buildup (>100 cmH2O) in blocked inspiratory module | error |
FE0208 | Wall air pressure detected: PS2 remains closed when air disconnected | EST disables compressor control and prompts disconnect air; PS2 should open but remains closed | error |
FE0209 | O2 PSOL leak | Oxygen PSOL (PSOL1) forward leak detected via excessive pressure buildup (>100 cmH2O) in blocked inspiratory module | error |
FE0210 | Safety valve pressure relief failed | Multiple test failures: (1) stable 1 L/min flow not established with patient port blocked; (2) safety valve cracking pressure out of range; (3) neither PSOL could deliver 100 L/min within prescribed count; (4) peak pressure too high during PSOL test | error |
FE0210 | SV pressure relief failed: Safety valve cracking pressure out of range or flow not established | EST tests safety valve cracking pressure and/or establishes flow to patient port; pressure or flow out of specification | error |
FE0211 | Oxygen pressure not detected | PS1 (oxygen pressure switch) reads open when oxygen should be connected | error |
FE0212 | Compressor pressure detected during EST depressurization test | Compressor pressure transducer indicates non-zero pressure when should be depressurized | error |
FE0215 | Air zero flow check failed | During zero-flow check, air flow sensor (Q2) reads > 0.05 L/min | error |
FE0216 | O2 zero flow check failed | During zero-flow check, oxygen flow sensor (Q1) reads > 0.05 L/min | error |
FE0217 | Exhalation zero flow check failed, Q3 >0.1 L/min | Exhalation flow sensor reads excessive flow during zero-flow test with supplies disconnected | error |
FE0301 | Excessive leak in EST Leak Test | System pressure dropped below failure pressure level during leak test | error |
FE0301 | Excessive system leak detected during leak test | System pressure dropped below 85 cmH2O or drops ≥5 cmH2O within 10 seconds | error |
FE0305 | Unable to establish pressure in EST Leak Test | System cannot attain leak test starting pressure within time-out period | error |
FE0401 | Accept key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0402 | Clear key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0403 | Insp. Pause key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0404 | Exp. Pause key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0405 | Man Insp key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0405 | Manual Inspiration key fails during EST keyboard test | Man Insp key not pressed or wrong key pressed within 15 seconds | error |
FE0406 | 100% O2/CAL key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0407 | Info key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0408 | Alarm Reset key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0409 | Alarm Silence key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0410 | Alarm Volume key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0411 | Screen brightness fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0412 | Screen contrast key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0413 | Screen lock key fails in GUI Keyboard Test | Wrong key pressed or key not pressed within 15 seconds during EST | warning |
FE0501 | Knob not responding | Knob not turned in prompted direction within 15 seconds | error |
FE0501 | Bad knob | EST GUI Knob Test - knob not turned in direction as prompted within 15 s | warning |
FE0501 | Knob rotation not detected during EST knob test | Knob not turned in commanded direction within 15 seconds | error |
FE0801 | SAAS (Safety Audible Alarm System) test failed | GUI alarm sound not heard during diagnostic test or cable disconnected | critical |
FE0801 | SAAS (Safety Audible Alarm System) test failed in GUI Audio Test | CLEAR key pressed to indicate GUI audio diagnostic failed | error |
FE0901 | Bad alarm cable in BDU Audio Test | Alarm cable voltage out of range (< 3.5 or > 5.05 V) | error |
FE0902 | Bad power fail capacitor (initial) in BDU Audio Test | Power failure capacitor initial voltage out of range (< 4.5 or > 5.05 V) | error |
FE0903 | Bad power fail capacitor (final) in BDU Audio Test | Power failure capacitor final voltage out of range or RC constant < 60 s | error |
FE0904 | Bad BD audio in BDU Audio Test | CLEAR key pressed to indicate user did not hear alarm although alarm was active | error |
FE0904 | BD alarm unit not sounding or responding | Self-test sounds alarm; user presses CLEAR indicating no audible alarm heard | error |
FE1101 | Safety valve occluded in Safety System Test | Excessive safety valve back pressure when safety valve is open | critical |
FE1102 | Bad safety valve driver or loopback in Safety System Test | Safety valve loopback current out of range during timed test points | error |
FE1103 | Insp check valve test failed in Safety System Test | Took too long to relieve excess pressure through open safety valve, indicating inspiratory check valve occluded or test circuit too large | error |
FE1105 | Unable to establish flow in Safety System Test | Flow controller unable to establish and control gas flow at 60 L/min | error |
FE1301 | Seal test failed | Exhalation valve seal test pressure delta exceeds failure threshold | critical |
FE1301 | Seal test failed in Exp Valve Seal Test | Seal test delta pressure above failure level | error |
FE1302 | Exp valve temp OOR in Exp Valve Seal Test | Exhalation valve magnet temperature out of range (10 to 100 degrees C) | warning |
FE1302 | Exp valve temp out of range | Exhalation valve magnet temperature outside 10-100 °C range during warmup diagnostic | warning |
FE1303 | Unable to establish exp flow | Exhalation flow controller cannot establish/maintain 5 L/min air flow with 45 cmH2O system pressure | critical |
FE1304 | Exp valve not calibrated in Exp Valve Seal Test | Exhalation valve table checksum invalid or last calibration not completed successfully | error |
FE1401 | Bad calibration | Measured system pressure at one or more exhalation test points is out of range | error |
FE1401 | Bad calibration in Exp Valve Test | Measured system pressure at one or more test points out of range | error |
FE1402 | Exp valve not calibrated in Exp Valve Test | Exhalation valve table checksum invalid or last calibration not completed successfully | error |
FE1403 | Unable to establish exp flow in Exp Valve Test | Flow controller unable to establish and control air flow measured by exhalation flow sensor Q3 | error |
FE1501 | Unable to establish air flow in Exp Heater Test | Flow controller unable to establish and control 60 L/min air flow | error |
FE1502 | Bad exp heater | EST Exp Heater Test failure - exhalation heater temperature did not rise sufficiently when heater turned on | error |
FE1502 | Bad exp heater in Exp Heater Test | Exhalation heater temperature did not rise sufficiently when heater was turned on | error |
FE1503 | Bad exp heater (cooling) in Exp Heater Test | Heater temperature did not drop sufficiently during interval after heater turned off | error |
FE1503 | Bad exhalation heater - insufficient temperature drop | Heater temperature did not drop ≥ 2 °C from heater-on temperature during interval after heater turned off | error |
FE2201 | Low exp ∆P | EV Velocity Transducer Test - expiratory delta pressure too low, indicating velocity transducer not responding properly | warning |
FE2201 | Low exp delta P | EV Velocity Transducer Test failure - expiratory delta P too low; velocity transducer not responding properly | error |
FE2201 | Low exp delta P in EV Velocity Transducer Test | Expiratory delta pressure too low, indicating velocity transducer not responding properly | error |
FE2301 | Nurse call stuck on in GUI Nurse Call Test | CLEAR key pressed to indicate nurse call (remote alarm) relay stuck on when should be off | error |
FE2302 | Nurse call stuck off in GUI Nurse Call Test | CLEAR key pressed to indicate nurse call (remote alarm) relay stuck off when should be on | error |
Flow sensor cross-check failed | Exhalation air flow out of specification relative to inspiration test flow | Flow measurement comparison during calibration cross-check | error |
Flow sensor cross-check: unable to establish flow | Air flow sensor unable to detect 5 L/min test flow during cross-check | Flow controller test during exhalation valve calibration | error |
Flow sensor info invalid | Flow sensor calibration table checksum error during exhalation valve calibration | Checksum verification of flow sensor calibration table fails | error |
FS0001 | O2 flow sensor cross check failed in SST Flow Sensor Test | Oxygen flow sensor (Q1) cross-check failed during self-service test | error |
FS0002 | O2 PSOL current out of range in SST Flow Sensor Test | Oxygen PSOL (PSOL1) command current out of range with respect to oxygen flow sensor Q1 | error |
FS0003 | Air flow sensor cross check failed in SST Flow Sensor Test | Air flow sensor (Q2) cross-check failed during self-service test | error |
FS0004 | Air PSOL current out of range in SST Flow Sensor Test | Air PSOL (PSOL2) command current out of range with respect to air flow sensor Q2 | error |
FS0005 | Gas not connected in SST Flow Sensor Test | No gas available (neither PS1, PS2, nor PC detects pressure) | warning |
FS0006 | Unable to establish O2 flow | SST Flow Sensor Test - flow controller unable to establish and control oxygen flow at 60, 5, and 1 L/min | critical |
FS0007 | Unable to establish air flow | SST Flow Sensor Test - flow controller unable to establish and control air flow at 60, 5, and 1 L/min | critical |
FS0008 | Oxygen zero flow check failed | Oxygen flow exceeds 0.153 L/min when PSOL1 is commanded to deliver 0 L/min (closed) | error |
FS0009 | Air zero flow check failed | Air flow > 0.153 L/min when PSOL2 commanded closed | error |
FS0101 | Inspiratory autozero out of range | Inspiratory pressure transducer ADC count at 0 cmH2O is outside acceptable range during autozero calibration | error |
FS0102 | Expiratory autozero out of range | Expiratory pressure transducer ADC count at 0 cmH2O is outside acceptable range during autozero calibration | error |
FS0103 | Failed to reach test pressure at 10 cmH2O | System unable to pressurize to 10 cmH2O at 5 L/min flow within timeout period | error |
FS0104 | Pressure cross-check failed at 10 cmH2O | Inspiratory and expiratory pressure transducer readings diverge beyond acceptable tolerance at 10 cmH2O test pressure | error |
FS0105 | Bad inspiratory autozero solenoid | Inspiratory pressure reading outside range (-0.60 to 0.60 cmH2O) when autozero solenoid (SOL1) commanded open | error |
FS0106 | Bad expiratory autozero solenoid | Expiratory pressure reading outside range (-0.60 to 0.60 cmH2O) when autozero solenoid (SOL2) commanded open | error |
FS0107 | Pressure cross-check failed at 50 and 100 cmH2O | Inspiratory and expiratory pressure transducer readings diverge beyond acceptable tolerance at high pressure levels (50 and 100 cmH2O) | error |
FS0108 | Failed to reach high test pressure (50 or 100 cmH2O) | System unable to pressurize to 50 or 100 cmH2O at 5 L/min flow within timeout period | error |
FS0201 | Excessive system leak detected | System pressure drops below 85 cmH2O or drops by ≥ 30 cmH2O over 20 seconds during leak test | error |
FS0205 | Unable to establish pressure | Unable to attain leak test starting pressure within timeout period using oxygen or air | error |
FS0301 | Unable to establish flow | System cannot deliver stable 60 L/min gas flow | error |
FS0303 | Occluded expiratory compartment | Exhalation compartment pressure >4 cmH2O | error |
FS0304 | Occluded expiratory filter | Pressure drop across expiratory filter > 3 cmH2O | error |
FS0306 | Patient circuit not disconnected | Expiratory flow detected during circuit disconnection test | error |
FS0307 | Patient circuit not reconnected | Expiratory flow reading indicates circuit not reconnected after test | error |
FS0310 | Low expiratory filter pressure drop | Pressure drop across expiratory filter <0.1 cmH2O | error |
FS0401 | Unable to establish flow | Flow controller cannot establish and maintain gas flow during inspiratory limb test | error |
FS0402 | Occluded inspiratory limb | Inspiratory limb pressure > 12.5 cmH2O (adult) or 7.5 cmH2O (pediatric) | error |
FS0404 | Unable to reach minimum peak flow | Peak flow < 60 L/min during inspiratory limb resistance test | error |
FS0405 | Occluded exhalation limb | Expiratory limb pressure > 12.5 cmH2O (adult) or 7.5 cmH2O (pediatric) | error |
FS0408 | Inspiratory limb resistance low | Inspiratory limb pressure <0.2 cmH2O | error |
FS0409 | Expiratory limb resistance low | Expiratory limb pressure < 0.2 cmH2O | error |
FS0410 | Unable to reach minimum peak flow | Peak flow falls below failure threshold during circuit resistance characterization | error |
FS0412 | Wye not blocked | Unable to pressurize system to 100 cmH2O at 60 L/min within timeout | error |
FS0501 | Unable to pressurize circuit | System cannot reach one or more test pressure points during compliance measurement | error |
FS0502 | Unable to pressurize circuit | System cannot pressurize to test pressure points during high-flow compliance test | error |
FS0503 | Compliance calculation failure | Out-of-range pressure ratio calculation during compliance measurement | error |
FS0504 | Excessive compliance | High compliance >12 mL/cmH2O (adult) or 9 mL/cmH2O (pediatric) | error |
FS0506 | Compliance low | High compliance falls below 1.05 mL/cmH2O | error |
FS0508 | Excessive compliance | Low compliance >12 mL/cmH2O (adult) or 9 mL/cmH2O (pediatric) | error |
FS0510 | Compliance low | Low compliance < 1.05 mL/cmH2O | error |
↑fTOT | Total respiratory rate at or above set limit (updated at each inspiration) | Total respiratory rate ≥ set limit, updated at beginning of each inspiration | warning |
↑fTOT (High Respiratory Rate Total) | Total respiratory rate at or above set alarm limit | Measured respiratory rate ≥ set limit | warning |
HN0001 | Non-maskable interrupt (NMI): DRAM Parity Error | BDU reset triggered by DRAM parity error | critical |
HN0002 | NMI: Ethernet Parity Error | BDU reset triggered by Ethernet parity error | critical |
HN0016 | NMI: Analog Interface Error (ADC Channel Sequencer Fault) | Non-maskable interrupt from ADC channel sequencer fault | critical |
HN0016_ATF | NMI: Analog Interface Error - ADC Timing Fault | BDU reset triggered by ADC timing fault | critical |
HN0016_CSF | NMI: Analog Interface Error - ADC Channel Sequencer Fault | BDU reset triggered by ADC channel sequencer fault | critical |
HN0016_HDF | NMI: Analog Interface Error - Hamming Decode Fault | BDU reset triggered by Hamming decode fault | critical |
HN0080 | NMI: Analog Interface Error - ADC Timing Fault | BDU reset triggered by ADC timing fault | critical |
HN0129 | NMI: DRAM Parity Error | Non-maskable interrupt from DRAM parity error | critical |
HN0130 | NMI: Ethernet Parity Error | Non-maskable interrupt from Ethernet parity error | critical |
INOPERATIVE BATTERY | Backup Power System (BPS) installed but not functioning | BPS detected but fails initialization or operation test | warning |
Inspiration autozero failed | Inspiration pressure transducer DAC count at 0 cmH2O is out of range during autozero | Autozero calibration step fails for inspiration transducer | error |
INSPIRATION TOO LONG | Inspiratory time for spontaneous breath exceeds IBW-based limit | Inspiratory time for spontaneous breath ≥ IBW-based limit | warning |
INSPIRATION TOO LONG (lockable) | Spontaneous breath inspiratory time exceeds ideal body weight (IBW) limit | Inspiratory time for spontaneous breath ≥IBW-based limit | warning |
KB0001 | Bad safety valve switched side | Measured voltage on safety valve switched side indicates valve is not in expected state | critical |
KB0002 | Bad expiratory flow | Exhalation flow sensor (Q3) reading out of range | critical |
KB0003 | Bad O2 PSOL current | Oxygen PSOL (PSOL1) current out of range | critical |
KB0004 | Bad air PSOL current | Air PSOL (PSOL2) current out of range | critical |
KB0005 | Bad exp motor current | Current to exhalation valve motor out of range | critical |
KB0007 | Bad exp pressure | PE (expiratory pressure) reading out of range | critical |
KB0008 | Bad insp pressure | PI (inspiratory pressure) reading out of range | critical |
KB0009 | Air flow out of range HIGH | Q2 (air flow sensor) reading too high | critical |
KB0012 | O2 flow out of range HIGH | Q1 (oxygen flow sensor) reading too high | critical |
KB0016 | Bad BD 10V supply | BDU +10 V supply out of range | critical |
KB0017 | Bad BD 12 V supply | BDU +12V supply out of range | critical |
KB0018 | Bad BD 15 V supply | BDU +15V supply out of range | critical |
KB0019 | Bad BD -15 V supply | BDU background check detects -15 V supply out of range | critical |
KB0022 | Bad BD 5 V supply | BDU background check detects +5 V supply out of range | critical |
KB0023 | O2 PSOL stuck | Oxygen PSOL (PSOL1) command current out of range | critical |
KB0024 | Air PSOL stuck | Air PSOL (PSOL2) command current out of range | critical |
KB0025 | Air PSOL stuck open | Air PSOL (PSOL2) command current reading indicates solenoid cannot close | critical |
KB0026 | O2 PSOL stuck open | Oxygen PSOL (PSOL1) command current reading indicates solenoid cannot close | critical |
KB0030 | Bad safety valve current | Safety valve solenoid current out of range during BDU check | critical |
KB0031 | Insp pressure stuck | Inspiratory pressure transducer (PI) reading stuck during BDU check | critical |
KB0032 | Exp pressure stuck | Expiratory pressure transducer (PE) reading stuck during BDU check | critical |
KB0033 | Insp pressure autozero offset failed | Inspiratory pressure transducer (PI) reading following autozero out of range | critical |
KB0034 | Exp pressure autozero offset failed | Expiratory pressure transducer (PE) reading following autozero out of range | critical |
KB0037 | Analog-Digital converter failed high | Analog-to-digital converter failure during BDU check | critical |
KB0038 | Analog-Digital converter failed low | Analog-to-digital converter failure during BDU check | critical |
KB0039 | Analog-Digital loopback failed | ADC loopback test failure during BDU check | critical |
KB0044 | BD NOVRAM checksum error | BD NOVRAM (non-volatile RAM) checksum validation fails during BDU check | critical |
KB0053 | BD EEPROM checksum error | BD EEPROM (non-volatile memory) checksum validation fails during BDU check | critical |
KB0064 | LV Ref out of range | Low-voltage reference circuit reading out of specification during BDU check | critical |
KB0065 | SV current out of range | Safety valve solenoid current out of specification during BDU check | critical |
KB0076 | Task Monitor | Task monitor function failure during BDU check | critical |
KB0079 | BK vent inop occurred | BDU detects ventilator inoperative signal active from GUI; indicates GUI has declared vent inoperative | critical |
KB0082 | Watchdog failure occurred | BD CPU watchdog timer failed during BDU check | critical |
KP0001 | Processor Initialization test failed | Phase 1 POST processor initialization fails | error |
KP0002 | Integer Unit test failed | Phase 1 POST general CPU confidence test fails | error |
KP0003 | DRAM Refresh Timer Test | POST detects DRAM refresh circuit failure | critical |
KP0004 | Kernel DRAM Test | POST detects kernel DRAM memory failure | critical |
KP0005 | Boot EPROM Checksum Test | POST detects EPROM checksum validation failure | critical |
KP0006 | POST Phase 2 Initialization | POST Phase 2 initialization sequence failure | critical |
KP0007 | Addressing Mode Test | POST detects CPU addressing mode test failure | critical |
KP0008 | Kernel NOVRAM Test | POST detects kernel NOVRAM test failure | critical |
KP0009 | Rolling Thunder Test | POST has started but not completed three or more times for reasons unrelated to power failure | critical |
KP0011 | Time of Day Clock Test failed (Major) | Fails during Phase 1 kernel POST | critical |
KP0012 | Timer Test | Timer failed comparison with real-time clock or failed to generate interrupt | critical |
KP0013 | Watchdog Timer Test | POST detects watchdog timer did not time out as expected | critical |
KP0015 | FLASH Memory Checksum Test | POST detects FLASH memory checksum validation failure | critical |
KP0081 | Memory Management Unit Test | POST detects memory management unit (MMU) circuit failure | critical |
KP0082 | Bus Timer Test | Local bus activity has ceased or bus time-out circuit did not generate expected interrupt | critical |
KP0083 | NMI Source Register Test | POST detects non-maskable interrupt (NMI) source register test failure | critical |
KP0084 | DRAM Memory test failed | Phase 2 POST DRAM data pattern verification fails | error |
KP0084 | POST DRAM Test | POST detects DRAM memory failure during POST sequence | critical |
KP0085 | Ethernet Self-Test start failed | Phase 2 POST Ethernet controller initial verification fails | error |
KP0087 | Unexpected Reset Umpire Test | POST has been invoked three times within 24 operational hours due to unexpected resets | critical |
KP0089 | Floating Point Unit Test | POST detects floating-point math unit failure | critical |
KP0090 | DRAM Parity Circuit Test | POST detects DRAM parity error detection circuit failure | critical |
KP0112 | VENT INOP Test failed | Ventilator in VENT INOP state | critical |
KP0113 | Analog Interface PCB Test | POST detects AI PCB circuit test failure | critical |
KP0114 | ADC Test | POST detects ADC (analog-to-digital converter) data not available to CPU as expected | critical |
KP0115 | DAC Test | POST detects DAC (digital-to-analog converter) output data not as expected | critical |
KP0116 | BDU POST analog devices test failure; check specific channel per address | BDU POST analog devices test (Phase 2) failure | critical |
KP0116 | BDU POST analog device test failure | Analog device failure during BDU POST Phase 2 | critical |
KP0117 | BD Serial Device Test | POST detects serial communication device test failure | critical |
KP0120 | Safe State System Test | POST detects failure to place PSOLs and exhalation valve into safe state or verify they cannot be commanded | critical |
KP0128 | Download operating system (OS) Boot | POST detects OS boot download failure | critical |
KP0129 | Operating System Initialization failed | OS fails to load | critical |
KP0129 | Application OS Boot | POST detects application OS boot failure | critical |
KP0130 | PB-MON Boot | POST detects PB-MON (monitoring application) boot failure | critical |
KP0131 | Application Boot | POST detects main application boot failure | critical |
LB0006 | Bad exhalation valve coil temp | Background check detects exhalation valve coil temperature out of range | warning |
LB0010 | Air flow out of range LOW | Background check detects air flow sensor (Q2) reading too low | warning |
LB0011 | Air flow sensor thermistor out of range | Background check detects Q2 thermistor reading out of range | error |
LB0011 | Bad air flow temperature | Air flow sensor Q2 thermistor reading out of range | unknown |
LB0013 | O2 flow measurement below specification | Oxygen flow sensor (Q1) reading too low | error |
LB0013 | Oxygen flow sensor reading too low | Background check detects Q1 reading below range | error |
LB0014 | Oxygen flow sensor thermistor out of range | Background check detects Q1 thermistor reading out of range | error |
LB0014 | Bad O2 flow temp | Oxygen flow sensor Q1 thermistor reading out of range | unknown |
LB0015 | Expiratory flow sensor thermistor out of range | Background check detects Q3 thermistor reading out of range | error |
LB0015 | Expiratory flow sensor thermistor reading out of range | Expiratory flow sensor (Q3) thermistor malfunction | error |
LB0020 | GUI +12V supply out of range | Background check detects supply voltage out of range | error |
LB0020 | GUI +12V supply voltage out of specification | GUI +12V power supply voltage out of range | error |
LB0021 | GUI +5V supply out of range | Background check detects supply voltage out of range | error |
LB0021 | GUI +5V supply voltage out of specification | GUI +5V power supply voltage out of range | error |
LB0027 | Atmospheric pressure transducer reading out of specification | Atmospheric pressure sensor malfunction | error |
LB0027 | Atmospheric pressure transducer reading out of range | Background check detects pressure reading out of range | error |
LB0028 | Bad O2 sensor OOR | Oxygen sensor percentage reading out of range | unknown |
LB0028 | Oxygen sensor percentage reading out of range | Background check detects sensor reading outside limits | error |
LB0029 | Oxygen sensor alert was reset | Background check status message | info |
LB0035 | Power fail capacitor voltage out of range | Background check detects capacitor voltage outside limits | error |
LB0036 | Alarm cable voltage out of range | Background check detects cable voltage anomaly | error |
LB0039 | ADC loopback constant test failure | Background check detects ADC loopback failure | error |
LB0043 | Power switch is in off state but power monitoring signal not detected | Power switch failure; switch appears off but power fail NMI not triggered | error |
LB0043 | Power switch appears stuck | Background check detects switch in off state but power fail NMI did not occur | error |
LB0045 | Real-time clock failure | Background check detects BD time of day failure | error |
LB0048 | Backup power supply voltage out of range | Background check detects BPS voltage anomaly | error |
LB0049 | Backup power supply current out of range | Background check detects BPS current anomaly | error |
LB0050 | Backup power supply model information incorrect | Background check detects wrong BPS model data | error |
LB0051 | Exhalation heater thermistor reading out of range | Background check detects heater thermistor anomaly | error |
LB0051 | Bad exp heater | Exhalation heater thermistor reading out of range | unknown |
LB0056 | Compressor elapsed timer error | Compressor timer failure | unknown |
LB0057 | Compressor EPROM data corruption | Background check detects unexpected EPROM data | error |
LB0058 | Loss of communication with GUI | Background check detects BD-GUI communication failure | error |
LB0060 | Resume GUI communication | Communication with GUI was resumed | unknown |
LB0060 | Communication with GUI has been restored | Background check status message | info |
LB0062 | EST required but not run | Background check detects user powered into normal mode without running EST | error |
LB0062 | EST (Equipment Self Test) required but not run | User powered on in normal mode instead of service mode when EST was needed | error |
LB0075 | Data key update failed | Background check detects data key update failure | error |
LB0076 | Task monitor alert | Background check detects task monitoring anomaly | warning |
LB0080 | Breath delivery extended beyond expected time | Background check detects breath delivery timing fault | critical |
LB0083 | GUI communication initialized and resumed | Background check initialization status message | info |
LB0083 | Initialization: GUI communication resumed | GUI communication restored during power-up sequence | info |
LB0085 | GUI communication lost during initialization | Background check initialization status message | info |
LB0085 | Initialization: GUI communication lost | GUI communication lost during power-up sequence | info |
LB0087 | Compressor serial number updated | Background check status message | info |
LB0088 | 10000 hours stored for elapsed time | Compressor hours corrupted; 10000 hour limit reached | unknown |
LB0088 | Compressor hours corrupted (10000 hours stored) | Background check detects elapsed time corruption | error |
LB0089 | Cannot determine data key size | Background check detects data key read failure | error |
LB0092 | Battery event status indication | Background check detects battery state change | warning |
LB0092 | Battery Event alert related to backup power supply status | Backup power supply event detected | error |
LC1xxx | Communications error between BD and peripheral devices | Serial or network communication failure with external devices | error |
LCxxxx | Network communications error between BDU and GUI | Communications error detected with variable code suffix | error |
LOSS OF POWER | Insufficient AC and BPS power; independent audio alarm sounds | Power switch on but insufficient AC and BPS power combined | critical |
LOW AC POWER | AC power dropped below 80% of nominal for 1 s | Mains (AC) power <80% of nominal for ≥1 s | warning |
LOW BATTERY | Battery Backup Power remaining < 2 minutes | BPS operational time falls below approximately 2 minutes | warning |
LOW BATTERY | Battery has <2 minutes of operational time remaining | BPS operational time <2 minutes | warning |
LOW INSP PRESSURE | Inspiratory target pressure below minimum required (VC+ breath) | Inspiratory target pressure <(PEEP + 5 cmH2O) when mandatory type is VC+ | warning |
LOW INSP PRESSURE (lockable) | Inspiratory target pressure below minimum safe level | Inspiratory target pressure < (PEEP + 5 cmH2O) when breath type VC+ | warning |
LP0008 | Non-critical NOVRAM data failure during startup | NVRAM self-test detected non-critical data corruption | error |
LP0008 | Kernel NOVRAM data failure (noncritical) | POST (BD) detects noncritical NOVRAM data anomaly | error |
LP0010 | Interrupt controller test failure | POST (BD) detects interrupt controller fault | critical |
LP0011 | Time of Day Clock test failed (minor) | Phase 1 POST Time of Day clock functionality verification fails | warning |
LP0085 | Ethernet self-test start | POST (BD) ethernet diagnostics begin | critical |
LP0086 | Ethernet Self-Test end status failed | Phase 2 POST Ethernet controller final status check fails | warning |
LP0087 | Unexpected reset detected (watchdog timeout) | POST (BDU) invoked due to watchdog time-out | error |
LP0088 | NOVRAM data failure detected during POST | Non-critical NVRAM data corruption during self-test | error |
LP0088 | POST NOVRAM test failure (noncritical) | POST (BDU) detects noncritical NOVRAM data anomaly | error |
LP0112 | Ventilator INOP signal in unexpected state | POST (BDU) detects VENT INOP signal fault | critical |
LP0114 | ADC (analog-to-digital converter) test failure | POST (BD) detects ADC fault | critical |
LP0114 | ADC Test | Analog to Digital Converter test failure | unknown |
LP0115 | DAC (digital-to-analog converter) test failure | POST (BD) detects DAC fault | critical |
LP0115 | DAC Test | Digital to Analog Converter test failure | unknown |
LP0116 | BDU POST analog device test failure | Analog device failure during BDU POST | critical |
LP0116 | GUI analog devices test failure; check specific channel per address | Analog devices test (Phase 2) failure at GUI | critical |
LP0121 | Service switch appears to be stuck closed | POST (BDU) detects service switch stuck | error |
LP0122 | AC Voltage Test failed | AC monitor voltage outside specified range | warning |
LP0122 | AC input voltage outside nominal range | POST (BDU) detects AC monitor voltage not within ±20% of set nominal | error |
No air connected | Air supply pressure not detected during calibration | Air supply pressure switch check at calibration start | error |
No air connected | Wall and compressor air pressure switches detect no air connected to ventilator | Air pressure switches report no air supply during calibration | critical |
NO AIR SUPPLY | Air inlet filter blocked or air source disconnected; compressor inoperative | Air pressure switch fails to detect supply pressure after system startup | warning |
NO AIR SUPPLY | Air supply unavailable; ventilator delivers 100% oxygen or only O2 available | Air inlet pressure loss; compressor inoperative; safety valve open | critical |
No O2 connected | Oxygen supply is not detected; calibration cannot proceed without oxygen | Oxygen pressure switch (PS1) detects no oxygen connected during calibration setup | error |
NO O2 SUPPLY | Oxygen source lost or inlet filter blocked | O2 pressure switch fails to detect oxygen pressure | warning |
NO O2 SUPPLY | Oxygen supply unavailable; ventilator delivers 21% or only air available | Oxygen inlet pressure loss; safety valve open | critical |
↑O2% | Measured O2% 7% or more above set value | O2% measured ≥7% above setting for ≥30 s (12% during first hour; +5% for 4 min after decrease) | warning |
↓O2% | Measured O2% 7% or more below set value or below 18% | O2% measured ≤7% below setting for ≥30 s (or ≤12% during first hour; +5% for 4 min after increase) or <18% | error |
↑O2% (High O2%) | Measured oxygen percentage above set limit | Measured O2% ≥7% above set O2% for ≥30 s (or ≥12% in first hour) | warning |
↓O2% (Low O2%) | Measured oxygen percentage below set limit | Measured O2% ≤7% below set O2% for ≥30 s, or below 18% | error |
O2 offset out of range | Oxygen flow sensor calibration failed; measured offset exceeds acceptable range | Calibration procedure detects invalid oxygen sensor offset during Flow Sensor Calibration | error |
O2 SENSOR | O2 sensor out of calibration or sensor failure | Sensor output falls outside acceptable range or sensor not responding | warning |
Operating on battery | Ventilator is powered by internal BPS battery; AC power unavailable | AC input voltage drops below +80 V AC equivalent | warning |
Operational time < 2 minutes | Less than 2 minutes of battery power remains; critical low-battery condition | BPS discharge reaches point where only 2 minutes of powered operation remain | critical |
↓PCOMP | ↓PPEAK detected but not ↓PCOMP; target pressure not met | Target pressure <(↓PPEAK - 5 cmH2O) | warning |
↑PMEAN | Measured airway pressure at or above set limit | Measured airway pressure ≥ set limit | warning |
↑PMEAN (lockable) | Mean airway pressure at or above set limit | Measured airway pressure ≥ set limit | warning |
POST Internal Software Failure | POST internal software error | Software logic failure | error |
Power Fail NMI | Power failure during POST (BDU only) | Loss of AC power during POST | critical |
↑PPEAK | Peak inspiratory pressure detected; not compliance | ↑PPEAK detected and NOT ↓PCOMP | warning |
↑PPEAK (lockable) | Peak inspiratory pressure at or above set limit | Peak pressure detected exceeds set limit | warning |
Pressure sensor cross-check alert: approaching spec limit | Pressure cross-check alert at one or more test pressures without failures | Pressure measurement during five-point pressure cross-check (5, 25, 50, 75, 100 cmH2O) | warning |
Pressure sensor cross-check failed | Pressure transducer cross-check failed at one or more of five test pressures | Pressure measurement comparison at 5, 25, 50, 75, or 100 cmH2O during calibration | error |
Pressure sensor cross-check: unable to build pressure | System unable to pressurize to one or more of five test pressures | Pressurization algorithm test sequence during calibration | error |
PROCEDURE ERROR | Patient connected before setup procedure complete | Patient circuit connected while ventilator startup sequence incomplete | warning |
↑PVENT | Inspiratory pressure >100 cmH2O; VC breath type | Inspiratory pressure >100 cmH2O and mandatory type = VC | error |
↑PVENT (lockable) | Inspiratory pressure exceeds 100 cmH2O with volume control breath | Inspiratory pressure > 100 cmH2O and breath type VC | error |
SAAS NMI | GUI alarm or alarm cabling interrupt | Alarm system anomaly | warning |
SCREEN BLOCK | Possible blocked beam on display GUI or touch screen malfunction | Display not responding or sensor detecting obstruction | info |
Serial number(s) doesn't match the data key | Inserted datakey does not match ventilator serial numbers; invalid datakey | Datakey insertion when serial numbers were initialized with different datakey | warning |
SEVERE OCCLUSION | Severe obstruction in ventilator circuit preventing ventilation | Ventilator detects very low or no flow despite pressure generation | critical |
SEVERE OCCLUSION | Little or no ventilation; ventilator in occlusion status cycling (OSC) | Severe circuit or patient airway occlusion detected | critical |
Sxxxxx | Status message indicating intentional event | Background check status code | info |
Sxxxxx | Status | Intentionally caused event | unknown |
U04004 | Assertion failure from intermittent digital I/O connection | Intermittent connection to digital I/O line detected | error |
U04004 | Assertion failure (I/O intermittent fault) | Background check GUI detects I/O connection anomaly | warning |
U05013 | Assertion failure (I/O intermittent fault) | Background check GUI detects I/O connection anomaly | warning |
U05013 | Assertion failure from intermittent digital I/O connection | Intermittent connection to digital I/O line detected | error |
U08002 | Serial data buffer overflow from host device | Background check BDU detects excessive serial errors | error |
U08002 | Assertion failure from GUI data buffer overflow | GUI overloaded with invalid data from serial port; host transmitting bad data causing over-run, parity, or framing errors | error |
U08012 | Assertion failure generating soft reset | Background check GUI detects anomaly | warning |
U08012 | Assertion failure; GUI generates soft reset to resynchronize | GUI detects error condition and initiates soft reset | error |
U14019 | Assertion failure generating soft reset | Background check GUI detects anomaly | warning |
U14019 | Assertion failure; GUI generates soft reset to resynchronize | GUI detects error and initiates soft reset | error |
U14021 | Assertion Failure - Background check (GUI) | Time stamps between GUI and BD are out of sync | warning |
Unable to burn flash | GUI NOVRAM cannot be successfully programmed | Cal Info Duplication function cannot write calibration data to GUI NOVRAM | error |
Unable to establish flow | Flow controller unable to establish 5 L/min test flow via exhalation flow sensor | Exhalation flow controller test during calibration | error |
Unable to program flash | Calibration table data unable to write to BDU NOVRAM | Calibration completion when storing results to non-volatile memory | error |
Unable to program flash | BDU NOVRAM cannot be successfully programmed | Initialize Flow Sensor function cannot write calibration data to BDU NOVRAM | error |
Unable to program flash | Unable to program calibration table into BDU NOVRAM memory during calibration | Flash programming step fails at end of calibration procedure | critical |
Unable to read air flow sensor | Error reading air flow sensor (Q2) serial EEPROM data | Initialize Flow Sensor function detects Q2 EEPROM read failure | error |
Unable to read O2 flow sensor | Error reading oxygen flow sensor (Q1) serial EEPROM data | Initialize Flow Sensor function detects Q1 EEPROM read failure | error |
Unable to read the expiratory flow sensor | Error reading exhalation flow sensor (Q3) serial EEPROM data | Initialize Flow Sensor function detects Q3 EEPROM read failure | error |
Unable to receive flash data | GUI did not receive all calibration data from BDU within 10 seconds | Cal Info Duplication function timeout waiting for BDU data transmission | error |
Unexpected Interrupt | Unexpected interrupt occurred | Unidentified interrupt | error |
Unknown NMI | Unknown non-maskable interrupt | CPU interrupt during POST | error |
UT0002 | LCD controller PCB fault - bus error or access fault in VGA controller | VGA controller failure on upper or lower LCD PCB | error |
UT0002 | Assertion Failure - Background check (GUI) | Hardware-trap detection triggered; processor exception encountered | warning |
Uxxxxx | Assertion failure (generic) | Background check GUI detects assertion | warning |
↑VE TOT | Expiratory minute volume at or above set limit | Expiratory minute volume ≥ set limit | warning |
↓VE TOT | Total minute volume below set limit for ≤20 s, >20 s, or >40 s | Total minute volume ≤ set limit for specified duration | warning |
↑VE TOT (lockable) | Expiratory minute volume at or above set limit | Exhaled minute volume ≥ set limit | info |
↓VE TOT (Low Minute Volume Total) | Total minute volume below set limit | Exhaled minute volume ≤ set limit for various durations | warning |
VN0129 | NMI: DRAM Parity Error | Memory parity error detected in GUI DRAM during operation | error |
VN0130 | NMI: Ethernet Parity Error | Ethernet controller parity error detected | error |
VN0132 | NMI: +5 V DC Overvoltage | +5V power rail exceeds safe operating voltage | error |
VN0136 | NMI: +12 V DC Overvoltage | +12V power rail exceeds safe operating voltage | error |
VN0144 | NMI: +12 V DC Undervoltage | +12V power rail drops below minimum operating voltage | error |
VOLUME NOT DELIVERED | Ventilator unable to deliver target tidal volume on VC+ or VS breaths | Measured delivered volume below target; pressure reaches maximum limit | warning |
↑VTE | Exhaled tidal volume at or above set limit | Exhaled tidal volume ≥ set limit | warning |
↑VTE (lockable) | Exhaled tidal volume at or above set limit | Exhaled tidal volume ≥ set limit | info |
↓VTE MAND | Exhaled mandatory tidal volume below set limit (2 of last 4 breaths) | Exhaled mandatory tidal volume ≤ set limit for 2 of last 4 mandatory breaths | warning |
↓VTE MAND (Low Tidal Volume Mandatory) | Exhaled mandatory tidal volume below set limit | Exhaled mandatory VT ≤ set limit for multiple breaths | warning |
↓VTE SPONT | Exhaled spontaneous tidal volume below set limit | Exhaled spontaneous tidal volume ≤ set limit | warning |
↓VTE SPONT (Low Tidal Volume Spontaneous) | Exhaled spontaneous tidal volume below set limit | Exhaled spontaneous VT ≤ set limit; alarm updated when recalculated | warning |
↑VTI MAND | Delivered inspiratory volume exceeds high limit (VC+ breath) | Delivered inspiratory volume ≥ high inspiratory volume limit (breath type = VC+) | warning |
↑VTI MAND (lockable) | Delivered inspiratory volume exceeds high limit (VC+ breaths) | Delivered inspiratory volume ≥ high inspiratory volume limit (breath type VC+) | info |
↑VTI SPONT | Delivered inspiratory volume exceeds high limit (TC or VS breath) | Delivered inspiratory volume ≥ high inspiratory volume limit (breath type = TC or VS) | warning |
↑VTI SPONT (lockable) | Delivered inspiratory volume exceeds high limit (TC or VS breaths) | Delivered inspiratory volume ≥ high inspiratory volume limit (breath type TC or VS) | info |
XB0066 | Monitor alarms fail | Alarm monitoring system background check failure during GUI operation | critical |
XB0067 | Monitor apnea alarm fails | Apnea alarm monitoring background check failure | critical |
XB0068 | Monitor apnea int fails | Apnea interval monitoring system failure | critical |
XB0069 | Monitor high inspiratory pressure alarm test failure | High pressure alarm circuit malfunction | critical |
XB0069 | Monitor hip fails | High Inspiration Pressure (HIP) alarm monitoring failure | critical |
XB0070 | Monitor insp time fails | Inspiration time monitoring background check failure | critical |
XB0071 | Monitor no data | No monitoring data received during background check | critical |
XB0072 | Monitor data corrupted | Monitoring data integrity check failure; data corruption detected | critical |
XB0073 | Monitor O2 mixture fails | O2 mixture monitoring background check failure | critical |
XB0074 | Monitor breath time fails | Breath timing monitoring system failure | error |
XP0001 | Processor Initialization failed (GUI) | Fails during Phase 1 kernel POST | critical |
XP0002 | Integer Unit Test failed (GUI) | Fails during Phase 1 kernel POST | critical |
XP0003 | DRAM Refresh Timer Test failed (GUI) | Fails during Phase 1 kernel POST | critical |
XP0004 | Kernel DRAM Test failed (GUI) | Fails during Phase 1 kernel POST | critical |
XP0005 | Boot PROM Checksum Test failed (GUI) | Fails during Phase 1 kernel POST | critical |
XP0006 | Phase 2 Initialization Start failed (GUI) | Non-kernel POST portion fails to start | critical |
XP0007 | Addressing Mode Test failed (GUI) | Fails during Phase 1 kernel POST | critical |
XP0008 | Kernel NOVRAM Test failed (GUI) | Fails during Phase 1 kernel POST | critical |
XP0010 | Interrupt Controller Test | Interrupt handling circuit test failure during POST | critical |
XP0011 | Time of Day Clock Test failed (GUI, Major) | Fails during Phase 1 kernel POST | critical |
XP0012 | Timer Test failed (GUI) | Fails during Phase 1 kernel POST | critical |
XP0013 | Watchdog Timer Test | Watchdog timer did not respond as expected or trigger reset | error |
XP0015 | EEPROM Checksum Test failed (GUI) | Fails during Phase 1 kernel POST | critical |
XP0081 | Memory Management Unit Test failed (GUI) | MMU test failure | critical |
XP0082 | Bus Timer Test failed (GUI) | Bus activity monitoring failure | critical |
XP0083 | NMI Source Register Test | Non-maskable interrupt source register test failure | critical |
XP0084 | DRAM Memory Test failed (GUI) | DRAM data pattern verification fails | critical |
XP0084 | POST DRAM Test | POST phase DRAM memory test failure | critical |
XP0085 | Ethernet Self-Test Start failed (GUI) | Ethernet controller functionality error | critical |
XP0087 | Unexpected Reset Umpire Test | POST invoked three or more times in 24 operational hours due to unexpected resets | error |
XP0089 | Floating Point Unit Test failed (GUI) | FPU functionality test failure | critical |
XP0090 | DRAM Parity Circuit Test | DRAM parity checking circuit test failure during POST | critical |
XP0097 | SAAS Self-test Start failed | Alarm circuitry error or no audio feedback | critical |
XP0098 | SAAS Self-test End failed | Alarm circuitry error | critical |
XP0128 | Download OS Boot | Operating system download/initialization failure | critical |
XP0129 | Application OS Boot | Application operating system boot failure | critical |
XP0129 | Operating System Initialization failed (GUI) | OS fails to load | critical |
XP0130 | PB-MON Boot | Puritan Bennett monitoring system boot failure | critical |
XP0131 | Application Initialization failed (GUI) | Application software fails to load | critical |
XP0131 | Application Boot | Main application software boot failure | critical |
ZB0040 | Touch screen failed | Touch screen hardware not responding; no input detection | warning |
ZB0041 | Touch screen blocked | Touch screen input blocked; screen obstruction detected or input error | unknown |
ZB0042 | Touch screen resumed | Touch screen communication or functionality restored after previous failure | info |
ZB0046 | GUI Time of Day failed | Real-time clock (time of day) initialization or operation failure | warning |
ZB0047 | GUI NOVRAM checksum error | NOVRAM data integrity check failure; noncritical data corruption detected | warning |
ZB0052 | GUI key stuck | Keyboard key detected as continuously pressed or stuck | warning |
ZB0054 | GUI EEPROM checksum error | EEPROM data integrity check failure during boot or operation | warning |
ZB0055 | GUI SAAS communication failed | Self-Addressing Alarm System communication link failure | warning |
ZB0059 | Loss of BD communication | Communication lost between GUI and BD (Breathing Delivery) unit | critical |
ZB0061 | Resume BD communication | BD communication link re-established after previous loss | info |
ZB0063 | GUI SAAS Audio failed | SAAS audio output (alarm speaker) circuit failure | warning |
ZB0076 | Task Monitor | System task monitoring flag set; indicates potential system performance issue | info |
ZB0077 | GUI dropped a waveform packet | Waveform data packet lost in transmission from BD to GUI | info |
ZB0084 | Init Resume BD communication | BD communication reestablished during initialization or after recovery | info |
ZB0086 | Init Loss of BD communication | BD communication lost during initialization sequence | critical |
ZB0090 | Touch Screen blocked | Touch screen input blocked due to visible obstruction or contamination | unknown |
ZB0091 | Touch Screen resumed | Touch screen functionality restored; blocked error cleared | info |
ZC0xxx | Communications error (BDU) | Communication error detected in BDU with BPS or external systems | warning |
ZC0xxx_BDU | Communications error (BDU); battery power system condition | Communications error (BDU); BPS charging or status issue | warning |
ZC1xxx | Communications error (GUI) | Additional GUI communication error variant; protocol or interface failure | warning |
ZC2000 | DCI parity error | Data Communication Interface parity check failure | warning |
ZC2001 | DCI input buffer overflow error | Communications error with GUI | warning |
ZC2002 | DCI command error; undefined command received | Communications error (GUI) undefined command received from host | warning |
ZC2003 | DCI unknown error | Unspecified communications error | warning |
ZP0008 | Kernel NOVRAM Test failure | Noncritical NOVRAM data failure | warning |
ZP0011 | Time of Day Clock Test failed (GUI, Minor) | Fails during Phase 1 kernel POST | warning |
ZP0085 | Ethernet Self-Test Start failure | Ethernet test failed during POST | error |
ZP0086 | Ethernet Self-test End (GUI) | Ethernet test completion | warning |
ZP0087 | Unexpected Reset Umpire Test | POST invoked after unexpected reset | warning |
ZP0088 | Phase 2 NOVRAM Test failed (GUI) | NOVRAM data out of valid range | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI