Puritan Bennett 7200 Ventilator: troubleshooting and error codes
The ServiceForge knowledge base holds 93 documented symptoms for the Puritan Bennett 7200 Ventilator, with 237 ranked causes, 135 diagnostic checks with test points and expected values, and 5 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Power disconnect alarm test failed patient safety
Backup power alarm system does not activate when main power lost
Ranked causes
- 1. Low or defective battery voltage (unknown)
- 2. Battery charging circuit defective (unknown)
- 3. Diode CR2 blown on motherboard (unknown)
- 4. Power switch defective (unknown)
- 5. Audio alarm harness open or short circuit (unknown)
- 6. Display cable assembly open or short circuit (unknown)
- 7. Conversion PCB interconnection cable open or short circuit (unknown)
- 8. Power switch continuity incorrect (unknown)
- 9. Audio alarm assembly defective (unknown)
- 10. Display controller or PCBs defective (unknown)
- 11. Conversion PCB defective (unknown)
11 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio alarm does not operate properly patient safety
Audible alarm system not functioning or malfunctioning
Ranked causes
- 1. Utility panel harness fuse blown (unknown)
- 2. Audio alarm assembly defective (unknown)
- 3. Power switch defective (unknown)
- 4. Diode CR2 blown on motherboard (unknown)
- 5. Interface PCB defective (unknown)
- 6. Display controller defective (unknown)
- 7. Front panel display PCB defective (unknown)
- 8. Utility panel harness defective (unknown)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leakage current out of specification high voltage
Electrical leakage current exceeds safety limits
Ranked causes
- 1. Compressor connected with leakage problem (unknown)
- 2. Humidifier connected with leakage problem (unknown)
- 3. Power cord damaged (unknown)
- 4. Defective compressor shock mounts (unknown)
- 5. Compressor defective (unknown)
- 6. Power supply defective (unknown)
- 7. Surge suppressor defective (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen percentages out of range patient safety
Delivered oxygen concentration not meeting specification
Ranked causes
- 1. Oxygen analyzer not properly calibrated (unknown)
- 2. Oxygen analyzer gel or membrane cap worn (unknown)
- 3. Oxygen sensor defective (unknown)
- 4. Oxygen analyzer instrument defective (unknown)
- 5. Proportional solenoid valves defective (unknown)
- 6. Conversion PCB defective (unknown)
- 7. Interface PCB defective (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulated compressor pressure out of range
Compressor regulator pressure outside 9.9 to 11.0 psig (68.26 to 75.85 kPa)
Ranked causes
- 1. Compressor inlet or outlet filter clogged (unknown)
- 2. Leaks in pulsation damper or plastic tees (unknown)
- 3. Obstruction in CV4 (unknown)
- 4. REG3 requires adjustment or is defective (unknown)
- 5. Compressor defective (unknown)
- 6. Leaky CV2 (unknown)
- 7. SOL9 not closing after compressor startup (unknown)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog meter or bar graph display PEEP reading out of range patient safety
Positive end-expiratory pressure reading on meter or display incorrect
Ranked causes
- 1. Tubing leak or obstruction in PEEP circuit (unknown)
- 2. Exhalation valve area ratio nonlinear (unknown)
- 3. Conversion PCB defective (unknown)
- 4. Exhalation valve defective (unknown)
- 5. Pressure transducer PCB defective (unknown)
- 6. Interface PCB defective (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor pressure less than 7.4 psig at high flow patient safety
Compressor discharge pressure insufficient at maximum flow
Ranked causes
- 1. Flow coming out of air inlet port (unknown)
- 2. Leaks in pulsation damper and plastic tees (unknown)
- 3. Obstruction in compressor circuit (unknown)
- 4. SOL9 valve leaking (unknown)
- 5. REG3 defective (unknown)
- 6. Compressor defective (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows all data on one line
Display output is compressed to a single line instead of multiple lines
Ranked causes
- 1. Display cable not connected properly to PCB assembly connectors (common)
- 2. Rear switch or jumpers on controller PCB set incorrectly (common)
- 3. Screen template EPROM (U17) not seated properly (occasional)
- 4. Screen template EPROM (U17) defective (occasional)
- 5. Electroluminescent or power supply PCB defective (occasional)
- 6. Controller PCB assembly defective (rare)
Display cable connection to PCB assembly connectors
- Test point
- PCB assembly connectors
- Expected
- Cable connected properly with all pins inserted in connector shell
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lamps, indicators, displays, or audio alarm not active during lamp test
Visual or audible output not functioning during self-test
Ranked causes
- 1. Defective audio alarm or diode CR2 (unknown)
- 2. Defective audio alarm harness or analog meter harness (unknown)
- 3. Front panel display PCB defective (unknown)
- 4. Display controller or display cable defective (unknown)
- 5. Conversion PCB, interface PCB, or interconnecting cables defective (unknown)
- 6. Analog meter defective (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No cursor appears on display high voltage
Display shows no cursor; troubleshooting procedure to verify ventilator configuration, DISPLAY port, DCI settings, cable connections, test patterns, and power supply voltages.
Ranked causes
- 1. Ventilator not properly equipped with DISPLAY port or incorrect hardware/software (common)
- 2. DISPLAY port not functioning (common)
- 3. DCI port configuration incorrect (common)
- 4. Display cable disconnected or improperly connected (common)
- 5. Test screen pattern generation failure (occasional)
- 6. Display power supply voltage out of specification (occasional)
Verify ventilator model and DISPLAY port hardware
- Test point
- Memory PCB (7200a) or DCI/display controller PCB (7200ae) switches S1, S2, S3
- Expected
- Per Section 3.1.2
13 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulated air pressure out of range
Air regulator pressure outside 9.9 to 11.0 psig (68.26 to 75.85 kPa)
Ranked causes
- 1. Source pressure or flow insufficient (unknown)
- 2. Filters clogged (unknown)
- 3. Obstruction in CV2 (unknown)
- 4. Leaky compressor check valve CV4 (unknown)
- 5. REG2 requires adjustment (unknown)
- 6. REG2 defective (unknown)
Verify regulated air pressure
- Expected
- 9.9 to 11.0 psig, ±0.55 psig
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensitivity problems patient safety
Patient trigger sensitivity not working correctly or baseline pressure incorrect
Ranked causes
- 1. Analog meter reading incorrect or pressure trapped in circuit (unknown)
- 2. Check valve CV5 not functioning (unknown)
- 3. Improper patient pressure connections (unknown)
- 4. Analog meter defective (unknown)
- 5. Pressure transducer PCB defective (unknown)
- 6. Exhalation valve defective (unknown)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator does not communicate with display
Ventilator fails to communicate with 7202 Display but works properly with another display or printer
Ranked causes
- 1. Display rear switch in incorrect position (common)
- 2. Display jumpers set incorrectly (common)
- 3. Improper cable or connector connections (common)
- 4. Display cable damaged or faulty (occasional)
- 5. Electroluminescent or power supply PCB faulty (occasional)
- 6. Controller PCB assembly faulty (rare)
Verify display rear switch position
- Test point
- rear switch on display
- Expected
- correct position per Figure 20-28
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivered gas volumes out of range patient safety
Tidal volume or minute ventilation delivered to patient outside specification
Ranked causes
- 1. Flow measurement instrument not in correct mode or not zeroed (unknown)
- 2. Tubing leak or obstruction in gas delivery circuit (unknown)
- 3. Defective oxygen or air flow sensor (unknown)
- 4. Interface PCB defective (unknown)
- 5. Conversion PCB defective (unknown)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen display otherwise flawed
Display shows snow, missing segments, bright or dim lines or spots, ghost images, or fluctuating displays
Ranked causes
- 1. Display cable improperly connected to PCB assembly connectors (common)
- 2. Ribbon cables or interconnect cable improperly connected (common)
- 3. Electroluminescent PCB assembly failure (occasional)
- 4. Controller PCB assembly failure (occasional)
- 5. Flat toroids not transferred to new controller PCB (occasional)
Verify display cable connection to PCB assembly connectors
- Test point
- PCB assembly connectors
- Expected
- Display cable properly connected with all pins inserted in connector shell
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog meter or patient pressure display reading incorrect patient safety
Pressure measurement displays incorrect value or high pressure alarm not activating
Ranked causes
- 1. Tubing leak in pressure monitoring circuit (unknown)
- 2. Analog meter requires adjustment (unknown)
- 3. Pressure transducer PCB defective (unknown)
- 4. Analog meter defective (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor air flow less than 110 lpm with maximum PEEP patient safety
Compressor air output insufficient for high PEEP ventilation
Ranked causes
- 1. Obstruction in compressor circuit (unknown)
- 2. Flow coming out of air inlet port (unknown)
- 3. REG3 requires adjustment (unknown)
- 4. Compressor defective (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display works but printer does not
Printer does not produce output or function when connected to ventilator with working display
Ranked causes
- 1. Printer out of paper or cartridge (common)
- 2. Printer not powered or on-line (common)
- 3. Printer cable not properly connected (common)
- 5. Printer hardware malfunction (occasional)
Printer has paper and cartridge
- Test point
- printer paper and cartridge compartments
- Expected
- paper present and cartridge installed
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard test failed
Keyboard input not functioning properly during system test
Ranked causes
- 1. Audio alarm not functioning (unknown)
- 2. Keyboard defective (unknown)
- 3. Front panel display PCB defective (unknown)
- 4. Display controller defective (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Regulated oxygen pressure out of range
Oxygen regulator pressure outside 9.9 to 11.0 psig (68.26 to 75.85 kPa)
Ranked causes
- 1. Source pressure or flow insufficient (unknown)
- 2. Filters clogged (unknown)
- 3. REG1 requires adjustment (unknown)
- 4. REG1 defective (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Voltage levels out of specification at analog output connector
Analog output voltage signals outside expected range
Ranked causes
- 1. Open or short circuits in utility panel harness (unknown)
- 2. Interface or conversion PCBs not properly seated (unknown)
- 3. Interface PCB defective (unknown)
- 4. Conversion PCB defective (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Current draw out of specification caution
Ventilator electrical current consumption exceeds normal range
Ranked causes
- 1. Humidifier drawing excessive current (unknown)
- 2. Compressor drawing excessive current (unknown)
- 3. Power supply defective (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
EST failure after power-on patient safety
Extended Self-Test fails with error codes displayed; ventilator will not pass EST until error conditions corrected
Ranked causes
- 1. Pneumatic system failure (common)
- 2. Transducer autozero failure (occasional)
- 3. Front panel control or indicator failure (occasional)
Press EST button and observe POST execution
- Test point
- EST button on utility panel; 20-character display
- Expected
- POST completes with no critical errors; EST begins automatically
- Tools
- Visual inspection
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
EST test fails with error display patient safety
During EST execution, ventilator displays 'xxyz ERR' indicating hardware malfunction detected by self-test
Ranked causes
- 1. Hardware malfunction (unknown)
- 2. Operator response timeout (occasional)
- 3. PEEP circuit cannot achieve desired setting (occasional)
Verify operator response to EST prompts within time limit
- Test point
- Display prompt
- Expected
- Operator presses required key within allotted time
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation check valve assembly twisted or misaligned patient safety
Check valve assembly twists during door closure or operation, increasing exhalation resistance and potentially causing patient breathing obstruction
Ranked causes
- 1. Exhalation bacteria filter retaining ring not tight (common)
- 2. Check valve assembly not properly secured to Q3/T3 (common)
- 3. Front-loading compartment door not properly latched (occasional)
Verify bacteria filter retaining ring is tight
- Test point
- Exhalation bacteria filter retaining ring
- Expected
- Filter cannot turn by hand
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance verification test execution troubleshooting patient safety
Guidance for interpreting test results and determining when to proceed to repair procedures.
Ranked causes
- 1. Incorrect procedure execution (common)
- 2. Improper data entry on ventilator keyboard (common)
- 3. Accessory performance not meeting specifications (occasional)
Verify procedure execution is correct
- Test point
- Test setup and sequence
- Expected
- All settings and steps match documented procedure exactly
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
POST error codes following power interruption
Self-test errors appear after loss of power to ventilator
Ranked causes
- 1. Low battery voltage from power interruption (unknown)
- 2. Defective diode CR1 on motherboard (unknown)
- 3. Error codes 1401 or 1501 persist (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
POST failure during power-on patient safety
LED indicators on CPU PCB and/or 20-character display show error code continuously during Power-On Self-Test; ventilator cannot resume normal operation
Ranked causes
- 1. CPU kernel failure (rare)
- 2. CPU PCB test failure (occasional)
- 3. Offboard circuitry or transducer failure (occasional)
Observe LED indicators on CPU PCB during power-on
- Test point
- CPU PCB front LEDs (80188) or back LEDs (8088)
- Expected
- No error codes displayed; code 7 (POST passed) or progress through test sections 0-5
- Tools
- Visual inspection
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Q3/T3 flow sensor damage from debris or contamination patient safety
Sensor fails to provide accurate mass flow or temperature data; debris contacting sensing element destroys sensor; condensation from heated wires damages components
Ranked causes
- 1. Debris entering exhalation system or contacting sensor (common)
- 2. Condensation from heated exhalation limb wires (occasional)
- 3. Inadequate filtration allowing moisture or debris ingress (occasional)
Visually inspect exhalation compartment and Q3/T3 sensor for debris or contamination
- Test point
- Q3/T3 sensing element and exhalation valve exit port
- Expected
- Clean, dry, no visible debris or liquid droplets
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test instrument PEEP reading out of range
Pressure measurement device reading incorrect PEEP level
Ranked causes
- 1. Tubing leak or obstruction in PEEP circuit (unknown)
- 2. Test instrument requires zeroing (unknown)
- 3. Pressure gauge defective (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test instrument PEEP reading out of range (alternate cause)
Pressure measurement device reading incorrect PEEP level (SOL7 related)
Ranked causes
- 1. SOL7 valve defective (unknown)
- 2. PCB contacts dirty or oxidized (unknown)
- 3. Poor connection between pneumatic harness and motherboard (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test instrument pressure reading does not match digital display
Pressure gauge reading differs from ventilator digital pressure reading
Ranked causes
- 1. Tubing leak between test points (unknown)
- 2. Test instrument not zeroed properly (unknown)
- 3. Pressure transducer PCB defective (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator settings changed due to power interruption patient safety
Operator settings lost or altered after power loss event
Ranked causes
- 1. Low or disconnected battery from power interruption (unknown)
- 2. Defective 80188 CPU PCB or memory PCB (unknown)
- 3. Motherboard defective (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Water trap/filter assembly leaks
Pneumatic system leaks detected after water trap/filter assembly disassembly and reassembly
Ranked causes
- 1. O-rings, gaskets, or filter elements not properly seated during reassembly (common)
- 2. Bowl threading damaged or bowls interchanged between versions (occasional)
- 3. Filter element F1 or F3 not replaced when required (occasional)
Verify filter element replacement
- Test point
- Filter F1 or F3 position
- Expected
- New filter element installed in plastic head assemblies with spring-loaded drain valves
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Air flow less than 162 lpm patient safety
Air gas flow output below minimum specification of 162 liters per minute
Ranked causes
- 1. Flow reading error when EST test differed from performance test (unknown)
- 2. Ventilator unable to reach required flow (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Back-up ventilation rates incorrect patient safety
Backup ventilation tidal volumes or rates not meeting specification
Ranked causes
- 1. BUV tidal volumes require calibration (unknown)
- 2. Interface PCB defective (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor and ventilator cooling fans not operating properly caution
Cooling fan systems fail to run or operate at incorrect speed
Ranked causes
- 1. Insufficient AC voltage to fan (unknown)
- 2. Fan motor defective (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Condensation in exhalation compartment patient safety
Liquid droplets visible on exit port of exhalation valve; moisture present in exhalation system
Ranked causes
- 1. Heated wires on exhalation limb causing condensation (common)
- 2. Moisture in exhalation compartment damaging sensor or valve (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electroluminescent or controller PCB failure caution
After all troubleshooting steps exhausted, display PCBs require replacement as matched sets.
Ranked causes
- 1. Electroluminescent or power supply PCB defective (rare)
- 2. Controller or piggyback PCB defective (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation bacteria filter not warm caution
Exhalation filter heater not functioning or not reaching operating temperature
Ranked causes
- 1. No power to heater element (unknown)
- 2. Heater assembly defective (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation valve not opening during exhalation patient safety
Internal exhalation valve fails to open sufficiently during exhalation to maintain operator-selected PEEP/CPAP pressure
Ranked causes
- 1. Internal exhalation valve malfunction or obstruction (unknown)
- 2. Internal exhalation valve malfunction or obstruction (front-loading) (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ground line resistance exceeds 0.1 ohm high voltage
Safety ground circuit resistance too high, indicating potential electrical hazard
Ranked causes
- 1. Short circuits or damage to power cord (unknown)
- 2. Faulty AC outlet (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen flow less than 162 lpm patient safety
Oxygen gas flow output below minimum specification of 162 liters per minute
Ranked causes
- 1. Flow reading error when EST test differed from performance test (unknown)
- 2. Ventilator unable to reach required flow (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply voltages out of specification caution
Internal power supply output voltages outside normal range
Ranked causes
- 1. Power supply requires voltage adjustment (unknown)
- 2. Power supply or module defective (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Proportional valve group defective patient safety
Oxygen, air, or nebulizer solenoid valves not mixing or controlling gases for delivery to patient
Ranked causes
- 1. PSOL1 or PSOL2 solenoid defective (unknown)
- 2. SOL1 or SOL2 solenoid defective (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Q3/T3 flow sensor damaged patient safety
Exhalation flow/temperature sensor not functioning or providing incorrect data to CPU; sensor destroyed by debris or contamination
Ranked causes
- 1. Debris contamination of sensing element (common)
- 2. Attempting to clean contaminated sensor with alcohol or water (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Resistance across nurse's call relay not infinite when alarm off
Nurse call signal circuit not open when alarm inactive
Ranked causes
- 1. Short circuit in NURSE'S CALL circuitry wiring (unknown)
- 2. Interface PCB defective (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Resistance across nurse's call relay out of range
Nurse call signal circuit impedance outside specifications
Ranked causes
- 1. Short circuit in utility panel harness (unknown)
- 2. Interface PCB defective (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SOL1 or SOL2 coil resistance out of spec
Coil resistance measurement indicates electrical failure, possible open or short circuit in solenoid coil
Ranked causes
- 1. Solenoid coil opened (occasional)
- 2. Solenoid coil shorted (occasional)
Measure solenoid coil resistance with DMM
- Test point
- Coil leads
- Expected
- 18 Ω for Allied coils or 19.4 Ω for Predyne coils
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SOL1/SOL2 solenoid does not actuate
No distinct clapping noise heard when voltage applied to solenoid coil leads; valve does not respond to electrical command.
Ranked causes
- 1. Coil opened or shorted (common)
- 2. Valve poppet or diaphragm sticking or contaminated (occasional)
Measure coil resistance
- Test point
- Solenoid coil leads
- Expected
- Approximately 18 Ω for Allied coils; 19.4 Ω for Predyne coils, ±1 Ω typical
- Tools
- DMM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Too much resistance across nurse's call relay during alarm
Nurse call signal circuit impedance excessive when alarm activated
Ranked causes
- 1. High resistance or short circuits in test connector wiring (unknown)
- 2. Interface PCB defective (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: BUV air volume patient safety
Back-up ventilator air volume potentiometer (R85) adjustment to ensure RT-200 displays 500 mL ± 10 mL
Ranked causes
- 1. R85 potentiometer requires adjustment (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: BUV air volume (R85)
Back-up ventilator air volume potentiometer on interface PCB requires adjustment to 500 mL ± 10 mL
Ranked causes
- 1. R85 potentiometer out of calibration (unknown)
Switch test equipment to air gas type and zero
- Test point
- PTS-2000 or RT-200 display
- Expected
- Volume Air setting selected, equipment zeroed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: BUV oxygen volume patient safety
Back-up ventilator oxygen volume potentiometer (R86) adjustment to ensure RT-200 displays 500 mL ± 10 mL
Ranked causes
- 1. R86 potentiometer requires adjustment (unknown)
Place ventilator into BUV mode using Method One or Method Two
- Test point
- Ventilator operating mode
- Expected
- Ventilator in BUV operating mode
8 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: BUV oxygen volume (R86)
Back-up ventilator oxygen volume potentiometer on interface PCB requires adjustment to 500 mL ± 10 mL
Ranked causes
- 1. R86 potentiometer out of calibration (unknown)
Place ventilator into BUV mode using Method One or Two
- Test point
- Ventilator control panel
- Expected
- Ventilator operates in BUV mode
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: BUV volumes (R85 and R86) patient safety
Back-up ventilator oxygen and air volume potentiometers on interface PCB require calibration to ensure correct delivery volumes
Ranked causes
- 1. BUV volume potentiometers R85 and R86 require calibration (unknown)
Select VOLUME screen and BTPS attribute on PTS-2000
- Test point
- PTS-2000 display
- Expected
- Volume screen with BTPS, L, corrected selected, threshold 10
10 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor volume
Metabolic monitor volume calibration required after battery power disconnection from motherboard
Ranked causes
- 1. Battery power disconnected from motherboard connector J11 (common)
Verify battery reconnected to motherboard connector J11
- Test point
- Connector J11
- Expected
- Battery power connected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PEEP System Test patient safety
Verify PEEP system performance by measuring pressure at various settings (10, 30, and maximum cmH2O) using ventilator display and external test instrument.
Ranked causes
- 1. PEEP system requires calibration verification (common)
Configure ventilator settings
- Test point
- Ventilator controls
- Expected
- CMV mode, 0.5L tidal volume, 1 bpm rate, 20 cmH2O sensitivity, 120 cmH2O high pressure limit (100 cmH2O for German ventilators), 3 cmH2O low inspiration pressure, 0 cmH2O low PEEP/CPAP pressure
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check valve assembly twisting risk patient safety
Risk of increased expiratory resistance if check valve assembly twists due to loose retaining ring
Ranked causes
- 1. Filter retaining ring not tight (common)
Verify filter does not turn when retaining ring is engaged
- Test point
- Bacteria filter
- Expected
- Filter stationary and cannot be rotated by hand
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compressor and cooling fans not operating properly caution
Compressor or ventilator cooling fans fail to operate or operate intermittently.
Ranked causes
- 1. Insufficient ac voltage to fan (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Cooling fan operation failure
- Debris protection for flow sensors during gas supply assembly service
- Electrical safety test failure: current draw out of range
- Electrical safety test failure: ground resistance out of range
- Electrical safety test failure: leakage current out of range
- Electroluminescent PCB and power supply PCB replaced without matched set
- Electronics compartment fan filter requires cleaning or replacement
- EST power loss during execution
- Exhalation bacteria filter heater not warm
- Exhalation bacteria filter heater removal required
- Exhalation check valve (CV5) assembly dysfunction
- Ground line resistance exceeds 0.1 Ω
- High pressure leak in pneumatic chassis
- Lamps, indicators, displays, and alarm not activating in test sequence
- Moisture condensation in exhalation bacteria filter or Q3/T3
- Moisture condensing in exhalation filter or Q3/T3
- Oxygen and air flow sensor assembly replacement required
- Oxygen delivery percentage out of range
- PCB assembly cable connection check
- POST and Total EST not passed
- POST error codes displayed after power cycle
- Power disconnect alarm not sounding
- Pressure transducer PCB not responding
- Printer self-test malfunction
- REG4 cracking pressure and full back pressure too low
- Replace audio alarm
- Safety valve back pressure regulator (REG4) requires removal or replacement
- Settings lost or changed after power cycle
- SOL1 or SOL2 valve does not actuate
- Temperature sensor integral to flow sensor assembly Q1/T1 and Q2/T2
- Test REG1 or REG2 outlet pressure
- Test regulator REG1 or REG2 outlet pressure
- Water trap/filter assembly air/oxygen cross-connection
33 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1401 | Battery disconnection or calendar-clock reset event | After battery reconnection to Q1/T1 or Q2/T2 sensor assembly | info |
INVALID KEY | Operator pressed incorrect key in response to EST prompt | Wrong key pressed twice in response to prompt | warning |
OPERATOR RESPONSE TIMED OUT. PLEASE RESPOND TO THE PROMPT. | Operator failed to respond to EST prompt within allotted time | No key entry received before timeout interval expires | warning |
TIME OUT | Test timeout; operator response delayed or ventilator unable to respond to action | During PEEP setting, PEEP circuit cannot achieve desired level | error |
xxyz ERR | EST hardware error with specific error code xxyz | EST detects hardware malfunction during test execution | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI