Pulmonetic Systems LTV 1000: troubleshooting and error codes
The ServiceForge knowledge base holds 141 documented symptoms for the Pulmonetic Systems LTV 1000, with 350 ranked causes, 165 diagnostic checks with test points and expected values, and 100 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Monitored volume is low with delivered volume low
Both monitored and delivered volumes are reading lower than expected
Ranked causes
- 1. Sense lines are reversed (occasional)
- 2. Leak Compensation is not on (common)
- 3. Failed auto zero (common)
- 4. VHome setting does not match flow valve (occasional)
- 5. Bypass flexible tube pinched or leaking (occasional)
- 6. Failed calibration (common)
- 7. Defective flow valve (occasional)
- 8. Defective turbine (occasional)
- 9. Circuit leak (occasional)
- 10. High or low side sense line or elbow at patient wye loose or leaking (common)
- 11. Exhalation drive line leaking or loose (common)
11 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated DISC/SENSE alarms patient safety
Disconnect or sense pressure alarms recur during normal operation; may include exhalation valve or pressure sensing faults
Ranked causes
- 1. Circuit disconnected from patient, wye or vent (common)
- 2. Exhalation valve disconnected from wye (common)
- 3. PEEP valve or peepless cap disconnected from wye (common)
- 4. High or low side sense lines disconnected or inappropriately attached (common)
- 5. High or low sense line or elbow at patient wye loose or leaking (occasional)
- 6. High or low sense lines are occluded (occasional)
- 7. High or low sense ports in the wye are occluded (occasional)
- 8. Exhalation drive line leaking or loose (occasional)
- 9. Defective transducer on analog board (occasional)
- 10. Defective auto zero solenoid (occasional)
Visually inspect all circuit connections at patient interface
- Test point
- Circuit, wye, exhalation valve, PEEP valve connections
- Expected
- All connections secure and intact; no visible disconnections
2 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivered pressure is high patient safety
Ventilator output pressure exceeds expected levels during operation
Ranked causes
- 1. Pressure sensing line issue (common)
- 2. Sense lines occluded (occasional)
- 3. High or low sense ports in wye occluded (occasional)
- 4. Failed auto zero (occasional)
- 5. VHome setting does not match flow valve (occasional)
- 6. Failed calibration or analog board defective (occasional)
- 7. Internal flexible tubing issues (occasional)
- 8. Defective flow valve (occasional)
- 9. Defective turbine (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator is auto cycling patient safety
Ventilator cycles or triggers breaths without patient effort
Ranked causes
- 1. Circuit leak (common)
- 2. High or low side sense line or elbow at patient wye loose or leaking (common)
- 3. High or low sense lines are occluded (occasional)
- 4. High or low sense ports in wye are occluded (occasional)
- 5. Bypass flexible tube pinched (occasional)
- 6. Pinhole leaks in bypass flexible tube (occasional)
- 7. Pressure Control or Pressure Support set below PEEP (occasional)
- 8. Failed auto zero (occasional)
- 9. Failed calibration (occasional)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitored pressure is low
System pressure transducer reading below expected levels
Ranked causes
- 1. Pressure sensing line or port issue (common)
- 2. High or low sense lines occluded (occasional)
- 3. High or low sense ports in wye occluded (occasional)
- 4. Failed auto zero (occasional)
- 5. VHome setting does not match flow valve (occasional)
- 6. Failed calibration or transducer drift (occasional)
- 7. Defective flow valve (occasional)
- 8. Defective turbine (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2% is high patient safety
Delivered oxygen percentage exceeds prescribed FIO2 level (LTV 1000 only)
Ranked causes
- 1. O2 inlet pressure too high (common)
- 2. O2 inlet flow too high (common)
- 3. Low O2 Source incorrectly selected (occasional)
- 4. Flow valve output is low (occasional)
- 5. Failed calibration or analog board defective (occasional)
- 6. Internal flexible tubing issues (occasional)
- 7. Defective or leaking solenoid manifold (rare)
- 8. Incorrect O2 blender solenoid output flow (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2% is low patient safety
Delivered oxygen percentage below prescribed FIO2 level (LTV 1000 only)
Ranked causes
- 1. O2 inlet flow too low (common)
- 2. Rolled or leaking flow valve inlet gasket (occasional)
- 3. Flow valve output is high (occasional)
- 4. Failed calibration or analog board defective (occasional)
- 5. Internal flexible tubing issues (occasional)
- 6. Defective or leaking solenoid manifold (rare)
- 7. Incorrect O2 blender solenoid output flow (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PEEP low or sags during exhalation patient safety
Monitored PEEP drops over long exhalation or does not hold steady pressure
Ranked causes
- 1. Circuit leak (common)
- 2. PEEP spring not installed in exhalation valve (occasional)
- 3. Diaphragm incorrectly seated or installed backwards in exhalation valve (occasional)
- 4. Worn PEEP spring (occasional)
- 5. High side sense line or elbow at patient wye loose or leaking (common)
- 6. Failed calibration (occasional)
- 7. Defective analog board (rare)
View monitored PEEP using RT XDCR DATA display
- Test point
- RT XDCR DATA
- Expected
- Monitored PEEP matches actual PEEP
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator does not operate from external power
Unit fails to run when connected to AC adapter power source
Ranked causes
- 1. Defective AC source (common)
- 2. AC adapter power cord loose (common)
- 3. Defective AC adapter (occasional)
- 4. Defective power board (occasional)
- 5. Internal battery depleted (common)
- 6. Internal battery not connected (occasional)
- 7. Defective internal battery (occasional)
Verify AC adapter is plugged into good AC source
- Test point
- AC power outlet
- Expected
- AC power available
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitored volume is high with delivered volume high
Both monitored and delivered volumes are reading higher than expected
Ranked causes
- 1. Very small ET tube connected directly to wye (rare)
- 2. Failed calibration (common)
- 3. Defective flow valve (occasional)
- 4. Defective turbine (occasional)
- 5. Defective or leaking solenoid manifold (occasional)
- 6. Circuit leak (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vent Inop LED is on and ventilator not ventilating patient safety
Ventilator powers on but Vent Inop LED illuminates and no ventilation occurs
Ranked causes
- 1. Ventilator ran on internal battery and battery became depleted (common)
- 2. Event trace shows reason for inop condition (common)
- 3. Memory board JP3 connector jumper not installed or mispositioned (occasional)
- 4. Defective memory board (occasional)
- 5. Defective power board (occasional)
- 6. Defective main board (rare)
Verify external power connection
- Test point
- AC power source and adapter
- Expected
- Ventilator connected to external power
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivered pressure is low with PEEP low and ventilator auto cycling patient safety
Delivered pressure is low, PEEP is low, and ventilator is auto cycling abnormally
Ranked causes
- 1. Leak Compensation is not on (common)
- 2. Sense lines are reversed (occasional)
- 3. High or low sense lines are occluded or at wye loose or leaking (common)
- 4. Exhalation drive line leaking or loose (common)
- 5. Exhalation valve leaking during inspiration (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on self-test failure patient safety
POST messages (CPU, SRAM, INT VECTOR, ROM CRC, EEPROM) do not complete or ventilation does not begin within 20 seconds of power-on
Ranked causes
- 1. CPU failure or malfunction (unknown)
- 2. SRAM failure or malfunction (unknown)
- 3. INT VECTOR failure (unknown)
- 4. ROM CRC error (unknown)
- 5. EEPROM failure or malfunction (unknown)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated DISC/SENSE alarms with XDCR FAULT also occurring patient safety
Combination of disconnect/sense and transducer fault alarms occurring repeatedly
Ranked causes
- 1. Pressure Control or Pressure Support set below PEEP (common)
- 2. Internal flexible tubing occluded or pinched (occasional)
- 3. Internal flexible tubing has pinhole leaks or leaking at connections (occasional)
- 4. Defective transducer on analog board (occasional)
- 5. Defective auto zero solenoid (occasional)
Verify Pressure Control and Pressure Support settings
- Test point
- Ventilator control panel settings display
- Expected
- PC and PS values set at or above PEEP value
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator does not power on
Unit fails to power up or turn on when power button is pressed
Ranked causes
- 1. Defective AC source, adapter, or depleted internal battery (common)
- 2. Memory board JP3 connector jumper not installed or mispositioned (common)
- 3. Defective memory board (occasional)
- 4. Defective power board (occasional)
- 5. Defective main board (rare)
Verify external AC power connection
- Test point
- AC adapter connection
- Expected
- Power cord fully seated to adapter and wall outlet
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator will not exhale, repeated HIGH PRES alarms patient safety
Ventilator stuck in inspiration with continuously rising pressure; exhalation function fails; turbine stops and auto-cycles to HIGH PRES
Ranked causes
- 1. Diaphragm installed backwards or incorrectly seated in exhalation valve (common)
- 2. Sense lines occluded, pinched or reversed (common)
- 3. Internal flexible tubing occluded or pinched (occasional)
- 4. Internal flexible tubing has pinhole leaks or leaking at connections (occasional)
- 5. Defective exhalation drive solenoid (occasional)
Visually inspect exhalation valve diaphragm orientation
- Test point
- Exhalation valve interior
- Expected
- Diaphragm and spring correctly seated per Chapter 4 diagram
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not sound during ALARM test
No audible alarm during alarm test checkout
Ranked causes
- 1. Defective power board (common)
- 2. Ventilator does not have power board P/N 15000 installed (occasional)
- 3. Audible alarm did not sound long enough before test termination (occasional)
- 4. Alarm sounder disconnected (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivered flow is high patient safety
Ventilator delivering higher than prescribed tidal or minute volume
Ranked causes
- 1. Circuit leak (common)
- 2. Sense lines reversed (occasional)
- 3. Pressure sense line or port issue (common)
- 4. Control settings incorrect (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivered flow is low patient safety
Ventilator delivering lower than prescribed tidal or minute volume
Ranked causes
- 1. Failed auto zero (common)
- 2. Leak Compensation not enabled (common)
- 3. Bypass flexible tube issues (occasional)
- 4. Failed calibration or transducer drift (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exhalation valve leaking during inspiration patient safety
Exhalation valve allows gas to escape to atmosphere during inspiratory phase instead of maintaining pressure
Ranked causes
- 1. Diaphragm installed incorrectly or spring misseated (common)
- 2. Exhalation diaphragm worn or defective (occasional)
- 3. PEEP spring worn or defective (occasional)
- 4. Exhalation valve body defective or damaged (occasional)
Verify diaphragm and spring correct orientation and seating
- Test point
- Exhalation valve interior
- Expected
- Diaphragm and spring seated per Chapter 4 diagram; no gaps
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leak test fails
Leak test checkout fails
Ranked causes
- 1. Circuit connections or accessories are leaking (common)
- 2. Internal flexible tubing occluded, pinched, or leaking (common)
- 3. Over pressure relief valve, subambient relief valve, or flow valve leaking (common)
- 4. VHome setting does not match flow valve (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated RESET alarms with CRC, STACK, POST, or RUNAWAY event log
RESET alarm recurs; Event Log shows CRC, STACK, POST, or RUNAWAY error codes
Ranked causes
- 1. Electrostatic discharge (ESD) (occasional)
- 2. Defective memory board (common)
- 3. Defective main board (common)
- 4. All other causes (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated XDCR FAULT alarms
Transducer fault alarms occur repeatedly; pressure measurement or sensing circuitry failure
Ranked causes
- 1. Internal flexible tubing occluded or pinched (common)
- 2. Internal flexible tubing has pinhole leaks or leaking at connections (common)
- 3. Defective transducer on analog board (occasional)
- 4. Defective auto zero solenoid (occasional)
Inspect internal flexible tubing for occlusions and kinks
- Test point
- Solenoid manifold and analog board tubing connections
- Expected
- All tubing clear of obstructions and not pinched
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sensitivity does not appear to be accurate patient safety
Ventilator trigger sensitivity setting does not match actual sensitivity response
Ranked causes
- 1. High or low sense lines occluded (occasional)
- 2. High or low sense ports in wye occluded (occasional)
- 3. Condensation in sense lines (occasional)
- 4. Failed calibration (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator does not operate from internal battery
Unit fails to run when on battery power or shuts off when external power removed
Ranked causes
- 1. Internal battery depleted (common)
- 2. Internal battery not connected (occasional)
- 3. Defective internal battery (occasional)
- 4. Defective power board (occasional)
Measure battery runtime at nominal settings
- Test point
- Ventilator output during battery operation
- Expected
- Runs for more than 40 minutes at nominal settings
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator is on, gas is not delivered and turbine is running patient safety
Ventilator powered and turbine spinning but no gas output to patient
Ranked causes
- 1. Defective or disconnected solenoids (occasional)
- 2. Defective or disconnected flow valve (occasional)
- 3. Defective turbine (occasional)
- 4. Defective purge solenoids (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator makes a high pitched noise when in Standby caution
Audible high-pitched sound emitted while ventilator in Standby mode
Ranked causes
- 1. Battery charge circuit running in bulk charge (common)
- 2. Defective or disconnected fan (occasional)
- 3. Defective power board (occasional)
- 4. Defective thermo conductive heat sink pads (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator won't trigger at sensitivity setting of 1 Lpm patient safety
Ventilator does not initiate breath when patient effort is 1 Lpm or below set sensitivity
Ranked causes
- 1. Patient effort inadequate (occasional)
- 2. Failed auto zero (occasional)
- 3. Leak Compensation is not on (occasional)
- 4. Failed calibration (occasional)
Perform auto zero calibration
- Test point
- XDCR ZERO menu
- Expected
- Auto zero completes without error
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
A control does not operate
Front panel control is unresponsive or non-functional
Ranked causes
- 1. Front panel ribbon cable not properly connected (common)
- 2. Front panel ribbon cable damaged (occasional)
- 3. Defective switch or main board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm does not sound at all patient safety
Audible alarm system completely silent; no alarm sounds for any alarm condition
Ranked causes
- 1. Ventilator did not run for more than 1 minute (common)
- 2. Defective or disconnected sounder (occasional)
- 3. Defective power board (occasional)
Run ventilator for minimum 1 minute
- Test point
- Ventilator operation
- Expected
- Unit runs continuously for at least 1 minute
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm too soft during ALARM test
Audible alarm volume is insufficient during alarm test checkout
Ranked causes
- 1. Alarm volume set too low (common)
- 2. Alarm sounder blocked (common)
- 3. Defective alarm sounder (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Charge Status LED is flashing amber
Charge Status LED flashes amber during battery evaluation or charging phase
Ranked causes
- 1. Internal battery charging normally (common)
- 2. Defective internal battery (occasional)
- 3. Defective power board (occasional)
Observe Charge Status LED behavior during charging
- Test point
- Charge Status LED on ventilator panel
- Expected
- Flashing amber for up to 1 hour if deeply discharged, then solid amber while charging
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery Charge Status LED is solid red
Charge Status LED displays solid red indicating internal battery disconnection or defect
Ranked causes
- 1. Internal battery not connected (common)
- 2. Defective internal battery (occasional)
- 3. Defective power board (occasional)
Verify internal battery is physically connected
- Test point
- Internal battery connector inside ventilator
- Expected
- Battery connector fully seated and secured
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not reach full charge
Internal battery fails to charge to full capacity or depletes too quickly during use
Ranked causes
- 1. Internal battery deeply discharged (common)
- 2. Defective internal battery (occasional)
- 3. Defective power board (occasional)
Charge battery for extended period if deeply discharged
- Test point
- Charge Status LED indicator
- Expected
- Solid amber LED after initial evaluation phase
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Control display flashing when setting a control
Control value display flashes while attempting to adjust a setting
Ranked causes
- 1. Control setting is limited by other control values (common)
- 2. Wrong model selected in VENT MTNCE (occasional)
- 3. LED misalignment (rare)
Identify which controls are flashing
- Test point
- Display screen during adjustment
- Expected
- Other controls flash to indicate dependency
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DEFAULTS alarm with invalid or out-of-range settings
DEFAULTS alarm occurs; some or all control settings were invalid or out of range on power up and were restored to defaults
Ranked causes
- 1. Electrostatic discharge (ESD) (occasional)
- 2. Defective main board (common)
- 3. Defective memory board (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display or LED does not illuminate during DISPLAY test
A display or LED fails to illuminate during display test checkout
Ranked causes
- 1. Misaligned LED (occasional)
- 2. Defective LED or display (common)
- 3. Defective main board (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
HIGH PRES alarm occurred but did not sound patient safety
High pressure condition detected and displayed but audible alarm did not activate
Ranked causes
- 1. Alarm silence was already active (common)
- 2. High pressure alarm delay is on (common)
- 3. Alarm automatically silenced after 3 seconds (occasional)
Check Silence/Reset LED status
- Test point
- Silence/Reset LED indicator
- Expected
- Green or off; not red
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect message displayed when rotary switch turned during CONTROL test
Correct message not displayed when rotary switch is turned, or incorrect message is displayed
Ranked causes
- 1. Front panel ribbon cable not properly connected (occasional)
- 2. Front panel ribbon cable damaged (occasional)
- 3. Defective main board (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Oxygen enrichment out of specification patient safety
External oxygen monitor does not read 55% to 65% O2 when O2% control is set to 60% with 0-80 PSIG oxygen supply
Ranked causes
- 1. Oxygen mixing valve malfunction or blockage (occasional)
- 2. Oxygen supply pressure out of range (occasional)
- 3. External oxygen monitor malfunction (occasional)
Connect oxygen supply and set O2% control to 60%
- Test point
- Ventilator O2% control
- Expected
- O2% control set to 60
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated HW FAULT alarms with delivered pressures and volumes off
Hardware fault alarms occurring with inaccurate pressure and volume delivery; multiple system failures
Ranked causes
- 1. Flow valve malfunction or disconnection (common)
- 2. Motor board defective or failed (occasional)
- 3. Event Log shows SYNC ER1 or HOME ER1 (occasional)
Check Event Trace for SYNC ER1 or HOME ER1 errors
- Test point
- Event Trace log in ventilator memory
- Expected
- No SYNC ER1 or HOME ER1 errors present
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated HW FAULT alarms with EEPROM event log
HW FAULT alarm recurs; Event Log shows EEPROM error
Ranked causes
- 1. Defective power board (common)
- 2. Electrostatic discharge (ESD) (occasional)
- 3. Defective main board (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Repeated HW FAULT alarms with fan not turning and excessive heat caution
Hardware fault alarm with fan failure; ventilator overheating due to non-functional cooling fan
Ranked causes
- 1. Defective or disconnected fan (common)
- 2. Fan housing screws over-tight (occasional)
- 3. Event Log shows FAN FLT1 (occasional)
Visually observe fan operation through side grill
- Test point
- Fan grill on side of ventilator case
- Expected
- Fan blades rotating smoothly at normal speed
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator auto cycling with very small monitored volumes patient safety
Ventilator cycles abnormally with low monitored volumes; RT XDCR DATA item shows negative flows during exhalation and positive flows during inspiration
Ranked causes
- 1. Sense lines are reversed (occasional)
- 2. Diaphragm installed backwards or incorrectly seated in exhalation valve (occasional)
- 3. Sense lines occluded, pinched or reversed (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ventilator does not work with LTM Graphics Monitor
Ventilator cannot communicate with or operate LTM Graphics Monitor
Ranked causes
- 1. Communications setting is not set to MONITOR mode (common)
- 2. Ventilator requires upgrades for LTM compatibility (occasional)
- 3. Defective connections between LTM Graphics Monitor and ventilator (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Volume mode operation out of specification patient safety
Exhaled tidal volume, I:E ratio, breath rate, or minute volume readings outside expected ranges during volume assist/control operation
Ranked causes
- 1. Incorrect ventilator settings (common)
- 2. Circuit leak or patient disconnect (common)
- 3. Test lung or patient compliance issue (occasional)
Verify ventilator mode and basic settings are correctly configured
- Test point
- Ventilator control panel
- Expected
- Mode: Volume Assist/Ctrl; BPM: 12; Tidal Volume: 500 ml; Insp. Time: 1.0 sec; PEEP: 5 cmH2O; O2%: 21; Sensitivity: 3; High Pressure Limit: 100; Low Pressure alarm: 5; Low Min Vol: 1.0 L
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AP transducer out of tolerance
AP transducer reading variance exceeds acceptable tolerance when tested with applied pressures
Ranked causes
- 1. Transducer requires recalibration (common)
- 2. Analog board defective (occasional)
Verify RISE TIME setting
- Test point
- VENT OP menu
- Expected
- Profile 5
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarm too soft during Vent Inop Alarm test patient safety
Vent Inop alarm audible tone is too soft during test
Ranked causes
- 1. Alarm sounder blocked (common)
- 2. Sounder fails to sound (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bypass Valve suspected faulty patient safety
Tidal volume variance between horizontal and upright positions exceeds 100 mL when measured over 8 consecutive breaths
Ranked causes
- 1. Leaking or damaged tubing from flow valve to turbine manifold (common)
- 2. Faulty turbine manifold (occasional)
Measure Vte in horizontal position
- Test point
- Display
- Expected
- average of 8 consecutive breaths
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Turbine flow valve
Turbine flow valve accuracy checkout and recalibration of flow output
Ranked causes
- 1. Flow valve output variance within recalibration range (common)
- 2. Flow valve defective or out of range (occasional)
Test at 10 Lpm and 4000 RPM
- Test point
- Calibrated flow meter output
- Expected
- 9.5 to 10.5 Lpm, 9.5 to 10.5
- Tools
- Calibrated flow meter
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Confirming audible chirp does not sound during Vent Inop Alarm test
No confirming chirp from Vent Inop alarm during test
Ranked causes
- 1. Audible alarm did not sound long enough before test termination (occasional)
- 2. Defective power board (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Controls are locked
LOCKED message displays when attempting to select or adjust a control
Ranked causes
- 1. Control Lock button engaged (common)
- 2. Control Lock button in HARD mode (occasional)
Press Control Lock button
- Test point
- Control Lock button
- Expected
- LOCKED message disappears; controls become active
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Date/time/usage meter not incrementing correctly
Date, time, or usage meter does not increment correctly after 1 hour of operation with test lung and power cycle
Ranked causes
- 1. Real-time clock malfunction or battery depleted (occasional)
- 2. Usage meter counter circuit malfunction (occasional)
Set ventilator date to current date; note date and format displayed
- Test point
- Ventilator date setting menu
- Expected
- Current date set in displayed format
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defective flow valve
Flow valve not calibrated or functioning correctly
Ranked causes
- 1. Flow valve requires calibration (common)
- 2. Flow valve defective or failed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivered and monitored volumes, pressures, and sensitivity are off patient safety
Displayed values for volume, pressure, or sensitivity do not match actual values
Ranked causes
- 1. Failed calibration (common)
- 2. VHome setting does not match flow valve (occasional)
Recalibrate ventilator
- Test point
- Calibration menu
- Expected
- All values match actual measured values
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Delivered pressure higher than set pressure on test lung
Pressure readings exceed the set value when testing with a test lung
Ranked causes
- 1. Pressure > 40 cmH2O used on small test lung (Pulmonetic Systems or Siemens 190) (common)
- 2. Test lung with small aperture connected directly to wye (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disc/Sense alarm does not activate patient safety
DISC/SENSE alarm fails to activate when high pressure sense line is disconnected from ventilator
Ranked causes
- 1. Pressure sense line sensor failure (occasional)
- 2. Sense line disconnection detection circuit failure (rare)
Disconnect high pressure sense line and observe alarm response on next breath
- Test point
- Ventilator high pressure sense port, alarm panel
- Expected
- DISC/SENSE alarm activates on next breath
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display or LED does not illuminate
A display element or LED is dark or non-functional
Ranked causes
- 1. Defective LED or display element (common)
- 2. Defective main board (occasional)
Perform display test during startup
- Test point
- All display elements
- Expected
- All displays illuminate and respond during test
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failed calibration
Calibration procedure fails to complete or pressure/flow readings remain inaccurate after recalibration
Ranked causes
- 1. Calibration procedure incomplete or incorrect (common)
- 2. Transducer drifting excessively (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
FDb (Inspired Volume) transducer out of tolerance
FDb transducer reading variance exceeds acceptable tolerance when tested at 30 cmH2O (LTM only)
Ranked causes
- 1. Transducer requires recalibration (common)
- 2. Analog board defective (occasional)
Apply 30 cmH2O to Flow Differential transducer low side port
- Test point
- FDb transducer low side port
- Expected
- FDb readout within +/- 0.5 cmH2O of 30, +/- 0.5 cmH2O
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- FDw (Monitored Volume) transducer out of tolerance
- Flow valve inaccurate
- FVd (Delivered Volume) transducer out of tolerance
- Hi and Low Flow transducer leak test fails
- High pressure alarm does not activate
- HW FAULT alarm with AD MMTCH error
- Internal inlet filter requires cleaning or replacement
- Keypad ribbon cable damage during main board replacement
- Low minute volume alarm does not activate
- Low O2 pressure alarm does not activate
- Low pressure alarm does not activate
- Message not displayed when control button pressed during CONTROL test
- O2 inlet pressure transducer out of tolerance
- O2% display flashing
- Over-pressure relief valve exceeds pressure limit
- Oxygen transducer calibration error
- Patient Assist port resistance out of specification
- Power board assembly requires replacement
- Power Lost alarm does not activate
- Pressure operation monitoring out of specification
- Set pressure not reached and turbine humming
- Set Value knob does not operate
- Vent Inop alarm does not sound continuously
- Vent Inop alarm does not sound during Vent Inop Alarm test
- Vent Inop alarm tone and/or LED fails test
- Ventilator does not reset when Watchdog test performed
- Audible alarm too loud during ALARM test
- Battery Charge Status LED is flashing red
- Bypass flexible tube pinched
- Calibration verification after software upgrade
- Cannot unlock controls
- Configuration settings check after software upgrade
- Confirming audible chirp does not sound during ALARM test
- Control button name not displayed or incorrect name displayed during CONTROL test
- Control is not selected
- Control not active in selected mode
- Controls are limited
- Defective analog board
- Defective power board
- Defective turbine
- Delivered pressure is low
- Delivered volume is high
- Displays are blanked while on battery power
- Failed auto zero
- High Pres Limit display flashing
- Internal battery pack requires replacement
- Internal flexible tubing has pinhole leaks or leaking at connections
- Internal flexible tubing occluded or leaking
- Internal flexible tubing occluded or pinched
- Leak test fails with LEAK --- FAIL message
- LED alignment check during board replacement
- LED portion of Vent Inop Alarm test fails
- Low Min Vol display flashing
- Low Pressure display flashing
- Main board assembly requires replacement
- Main board installation misalignment
- Memory board installation with JP3 jumper verification
- Monitored pressure is high
- Monitored volumes very high on test lung
- NO CAL DATA alarm with no calibration displayed
- O2 blender label check for software version 5.0 or greater on LTV 1000
- O2 blender label replacement for version 5.0 upgrade on LTV 1000
- O2 blender solenoid output flows out of specification
- O2 Leak Test fails
- OPRV test fails
- Oxygen transducer leaks
- PEEP is low
- Pinhole leaks in bypass flexible tube
- Power board replacement procedure
- Pressure Control display flashing
- Pressure Support display flashing
- Pressure Support or Pressure Control not functioning properly
- Repeated HW FAULT alarms with INTRRPT1 or INTRRPT2 event log
- RESET alarm occurs after POST tests
- Software version verification after memory board upgrade
- Vent in Standby
- Vent Inop alarm tone and/or LED fails 15-second duration test
- Ventilator gets excessively hot
- Ventilator running but displays are turned off
- Ventilator will not allow patient to exhale
- VHome setting does not match flow valve
81 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
01 | Power on | Ventilator powered on | info |
02 | Power off | Ventilator powered off | info |
03 | Set hour meter | Operating hours recorded | info |
04 | Set vent check | Entered VENT CHECK mode | info |
05 | Apnea mode entered | Apnea mode activated | warning |
06 | Apnea mode exited | Apnea mode deactivated | info |
08 | High side disconnect | High pressure sensor circuit opened | error |
09 | Low side disconnect | Low pressure sensor circuit opened | error |
10 | Circuit disconnect exited | Disconnect condition resolved | info |
11 | Internal battery empty occurred | Battery charge depleted | critical |
12 | Internal battery empty exited | Battery charge restored | info |
13 | Internal battery low occurred | Battery charge critically low | warning |
14 | Internal battery low exited | Battery charge above low threshold | info |
15 | External power lost occurred | AC power disconnected | critical |
16 | External power lost exited | AC power restored | info |
17 | External power low occurred | AC voltage below minimum | warning |
18 | External power low exited | AC voltage restored to normal | info |
19 | XDCR fault occurred | Transducer signal loss or fault detected | error |
20 | XDCR fault exited | Transducer fault resolved | info |
21 | O2 pressure low occurred | Oxygen supply pressure below minimum | critical |
22 | O2 pressure low exited | Oxygen supply pressure restored | info |
23 | O2 pressure high occurred | Oxygen supply pressure exceeds maximum | warning |
24 | O2 pressure high exited | Oxygen supply pressure normalized | info |
25 | Defaults set | System reset to factory defaults | warning |
26 | No calibration data found | System startup with missing calibration | error |
27 | Fan fault occurred | Cooling fan malfunction detected | error |
28 | Fan fault exited | Fan fault resolved | info |
31 | Spurious interrupt occurred ms | Microcontroller received unexpected interrupt | error |
32 | Spurious interrupt occurred ls | Microcontroller received unexpected interrupt | error |
33 | ADC mismatch | A/D converter channel mismatch detected | error |
34 | ADC mismatch occurred | A/D converter mismatch first detected | error |
35 | ADC mismatch cleared | A/D converter mismatch resolved | info |
36 | Stepper motor lost sync occurred | Stepper motor lost position synchronization | error |
37 | Stepper motor lost sync exited | Stepper motor sync recovered | info |
38 | Stepper motor home failure occurred | Motor unable to reach home position | error |
39 | Stepper motor home failure exited | Motor home sequence recovered | info |
40 | EEPROM degraded | EEPROM read/write errors detected | error |
41 | Memory CRC check failed | CRC mismatch on stored data | critical |
42 | High pressure occurred | Peak airway pressure exceeds alarm threshold | critical |
43 | High pressure exited | Peak airway pressure returned to normal | info |
44 | Turbine immediate stop occurred | Emergency stop activated on turbine | critical |
45 | Turbine set to zero flow occurred | Turbine flow reduced to zero | warning |
46 | Turbine emergency stop occurred | Turbine emergency stop activated | critical |
47 | Low minute volume occurred | Delivered minute volume below threshold | error |
48 | Low minute volume exited | Minute volume returned to acceptable range | info |
49 | Low peak pressure occurred | Peak inspiratory pressure below threshold | warning |
50 | Low peak pressure exited | Peak pressure returned to normal range | info |
51 | Event log cleared | Event history erased from memory | info |
52 | Control settings cleared | Operating settings reset | warning |
53 | Date set | System date configured | info |
54 | Time set | System time configured | info |
56 | Stack overflow detected | Microcontroller stack memory exceeded | critical |
57 | POST failure | Power-on self-test failed | critical |
58 | Code runaway detected | Software execution exceeded time limit | critical |
59 | Watchdog test run | Watchdog timer test executed | info |
60 | Calibration records cleared | Calibration data erased | error |
61 | Differential pressure transducer narrow channel fault | FDn transducer sensor failure | error |
62 | Differential pressure transducer wide channel fault | FDw transducer sensor failure | error |
63 | Differential pressure transducer bi-directional channel fault | FDb transducer sensor failure | error |
64 | Airway pressure transducer fault | Main airway pressure transducer signal error | error |
65 | AD mismatch primary channel fault value | A/D converter primary channel mismatch | error |
66 | Turbine Hold Stop occurred | Turbine hold stop activated | warning |
67 | Shutdown for other than pressing On/Standby button | Unexpected shutdown detected | error |
68 | EEPROM write error during shutdown | Data write failed to EEPROM at shutdown | critical |
68 | A problem is detected writing data to EEPROM during shutdown | Memory write failure at system shutdown | error |
69 | Problem detected with audible alarm circuitry | Primary or redundant alarm circuit malfunction | warning |
70 | Recovery from audible alarm circuitry problem | Alarm circuit fault resolved | info |
71 | Alarm sounder error | Speaker or sounder malfunction | warning |
72 | Recovery from alarm sounder error | Alarm sounder fault resolved | info |
73 | Reserved for future use | Not currently assigned | unknown |
74 | Reserved for future use | Not currently assigned | unknown |
75 | Reserved for future use | Not currently assigned | unknown |
76 | Reserved for future use | Not currently assigned | unknown |
77 | Reserved for future use | Not currently assigned | unknown |
78 | Reserved for future use | Not currently assigned | unknown |
79 | Reserved for future use | Not currently assigned | unknown |
80 | Reserved for future use | Not currently assigned | unknown |
81 | Reserved for future use | Not currently assigned | unknown |
82 | Reserved for future use | Not currently assigned | unknown |
83 | Reserved for future use | Not currently assigned | unknown |
84 | Reserved for future use | Not currently assigned | unknown |
85 | Reserved for future use | Not currently assigned | unknown |
86 | Reserved for future use | Not currently assigned | unknown |
87 | Reserved for future use | Not currently assigned | unknown |
88 | Reclaims all incorrectly recognized bad EEPROM records | Software 5.01 recovery routine executed | warning |
CAL FAIL | Calibration procedure failed or aborted | Error encountered during any calibration step (Valve Differential, O2 inlet, or Motor Drive) | error |
DEFAULTS | Control settings restored to default values | Invalid or out-of-range control settings detected on power up | warning |
DISC/SENSE | Disconnect or sense circuit failure alarm | Pressure sensing circuit fault or circuit disconnection detected | error |
HIGH PRES | High pressure condition detected | Airway pressure exceeds high pressure limit setting during ventilation | warning |
HW FAULT | Hardware fault detected; multiple potential causes | Internal hardware failure or circuit error detected during power-up self-test or operation | critical |
HW FAULT (EEPROM) | Hardware fault related to EEPROM memory | EEPROM error detected during operation | error |
HW FAULT (INTRRPT1/INTRRPT2) | Hardware fault related to interrupt errors | Interrupt errors detected during operation | error |
LMV LPPS OFF | Low Minute Volume alarm is turned off and the LPP ALARM has been set to VC/PC ONLY | LMV LPPS OFF message displayed on ventilator screen | info |
LMV OFF | Low Minute Volume alarm is turned off | LMV OFF message displayed on ventilator screen | info |
LOW O2 PRES | Low oxygen pressure condition | Oxygen inlet pressure falls below acceptable level during oxygen delivery | warning |
LPPS OFF | LPP ALARM has been set to VC/PC ONLY mode | LPPS OFF message displayed on ventilator screen | info |
NO CAL DATA | Missing or failed calibration data | Calibration data not found or invalid during operation | warning |
RESET | System reset alarm; ESD or internal reset event | Electrostatic discharge or internal watchdog timer reset during normal operation | warning |
RESET (CRC/POST/RUNAWAY/STACK) | Reset alarm with various error types (CRC, POST, RUNAWAY, STACK) | Occurs at conclusion of POST after tests that cause ventilator to be inoperative; can also recur repeatedly | warning |
XDCR FAULT | Transducer circuit fault alarm | Pressure transducer or analog measurement circuit failure detected | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI