Pulmonetic Systems LTV 1000: troubleshooting and error codes

The ServiceForge knowledge base holds 141 documented symptoms for the Pulmonetic Systems LTV 1000, with 350 ranked causes, 165 diagnostic checks with test points and expected values, and 100 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Monitored volume is low with delivered volume low

Both monitored and delivered volumes are reading lower than expected

Ranked causes

  1. 1. Sense lines are reversed (occasional)
  2. 2. Leak Compensation is not on (common)
  3. 3. Failed auto zero (common)
  4. 4. VHome setting does not match flow valve (occasional)
  5. 5. Bypass flexible tube pinched or leaking (occasional)
  6. 6. Failed calibration (common)
  7. 7. Defective flow valve (occasional)
  8. 8. Defective turbine (occasional)
  9. 9. Circuit leak (occasional)
  10. 10. High or low side sense line or elbow at patient wye loose or leaking (common)
  11. 11. Exhalation drive line leaking or loose (common)

11 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Repeated DISC/SENSE alarms patient safety

Disconnect or sense pressure alarms recur during normal operation; may include exhalation valve or pressure sensing faults

Ranked causes

  1. 1. Circuit disconnected from patient, wye or vent (common)
  2. 2. Exhalation valve disconnected from wye (common)
  3. 3. PEEP valve or peepless cap disconnected from wye (common)
  4. 4. High or low side sense lines disconnected or inappropriately attached (common)
  5. 5. High or low sense line or elbow at patient wye loose or leaking (occasional)
  6. 6. High or low sense lines are occluded (occasional)
  7. 7. High or low sense ports in the wye are occluded (occasional)
  8. 8. Exhalation drive line leaking or loose (occasional)
  9. 9. Defective transducer on analog board (occasional)
  10. 10. Defective auto zero solenoid (occasional)
First check

Visually inspect all circuit connections at patient interface

Test point
Circuit, wye, exhalation valve, PEEP valve connections
Expected
All connections secure and intact; no visible disconnections

2 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivered pressure is high patient safety

Ventilator output pressure exceeds expected levels during operation

Ranked causes

  1. 1. Pressure sensing line issue (common)
  2. 2. Sense lines occluded (occasional)
  3. 3. High or low sense ports in wye occluded (occasional)
  4. 4. Failed auto zero (occasional)
  5. 5. VHome setting does not match flow valve (occasional)
  6. 6. Failed calibration or analog board defective (occasional)
  7. 7. Internal flexible tubing issues (occasional)
  8. 8. Defective flow valve (occasional)
  9. 9. Defective turbine (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator is auto cycling patient safety

Ventilator cycles or triggers breaths without patient effort

Ranked causes

  1. 1. Circuit leak (common)
  2. 2. High or low side sense line or elbow at patient wye loose or leaking (common)
  3. 3. High or low sense lines are occluded (occasional)
  4. 4. High or low sense ports in wye are occluded (occasional)
  5. 5. Bypass flexible tube pinched (occasional)
  6. 6. Pinhole leaks in bypass flexible tube (occasional)
  7. 7. Pressure Control or Pressure Support set below PEEP (occasional)
  8. 8. Failed auto zero (occasional)
  9. 9. Failed calibration (occasional)

10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitored pressure is low

System pressure transducer reading below expected levels

Ranked causes

  1. 1. Pressure sensing line or port issue (common)
  2. 2. High or low sense lines occluded (occasional)
  3. 3. High or low sense ports in wye occluded (occasional)
  4. 4. Failed auto zero (occasional)
  5. 5. VHome setting does not match flow valve (occasional)
  6. 6. Failed calibration or transducer drift (occasional)
  7. 7. Defective flow valve (occasional)
  8. 8. Defective turbine (rare)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2% is high patient safety

Delivered oxygen percentage exceeds prescribed FIO2 level (LTV 1000 only)

Ranked causes

  1. 1. O2 inlet pressure too high (common)
  2. 2. O2 inlet flow too high (common)
  3. 3. Low O2 Source incorrectly selected (occasional)
  4. 4. Flow valve output is low (occasional)
  5. 5. Failed calibration or analog board defective (occasional)
  6. 6. Internal flexible tubing issues (occasional)
  7. 7. Defective or leaking solenoid manifold (rare)
  8. 8. Incorrect O2 blender solenoid output flow (rare)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2% is low patient safety

Delivered oxygen percentage below prescribed FIO2 level (LTV 1000 only)

Ranked causes

  1. 1. O2 inlet flow too low (common)
  2. 2. Rolled or leaking flow valve inlet gasket (occasional)
  3. 3. Flow valve output is high (occasional)
  4. 4. Failed calibration or analog board defective (occasional)
  5. 5. Internal flexible tubing issues (occasional)
  6. 6. Defective or leaking solenoid manifold (rare)
  7. 7. Incorrect O2 blender solenoid output flow (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PEEP low or sags during exhalation patient safety

Monitored PEEP drops over long exhalation or does not hold steady pressure

Ranked causes

  1. 1. Circuit leak (common)
  2. 2. PEEP spring not installed in exhalation valve (occasional)
  3. 3. Diaphragm incorrectly seated or installed backwards in exhalation valve (occasional)
  4. 4. Worn PEEP spring (occasional)
  5. 5. High side sense line or elbow at patient wye loose or leaking (common)
  6. 6. Failed calibration (occasional)
  7. 7. Defective analog board (rare)
First check

View monitored PEEP using RT XDCR DATA display

Test point
RT XDCR DATA
Expected
Monitored PEEP matches actual PEEP

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator does not operate from external power

Unit fails to run when connected to AC adapter power source

Ranked causes

  1. 1. Defective AC source (common)
  2. 2. AC adapter power cord loose (common)
  3. 3. Defective AC adapter (occasional)
  4. 4. Defective power board (occasional)
  5. 5. Internal battery depleted (common)
  6. 6. Internal battery not connected (occasional)
  7. 7. Defective internal battery (occasional)
First check

Verify AC adapter is plugged into good AC source

Test point
AC power outlet
Expected
AC power available

1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitored volume is high with delivered volume high

Both monitored and delivered volumes are reading higher than expected

Ranked causes

  1. 1. Very small ET tube connected directly to wye (rare)
  2. 2. Failed calibration (common)
  3. 3. Defective flow valve (occasional)
  4. 4. Defective turbine (occasional)
  5. 5. Defective or leaking solenoid manifold (occasional)
  6. 6. Circuit leak (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vent Inop LED is on and ventilator not ventilating patient safety

Ventilator powers on but Vent Inop LED illuminates and no ventilation occurs

Ranked causes

  1. 1. Ventilator ran on internal battery and battery became depleted (common)
  2. 2. Event trace shows reason for inop condition (common)
  3. 3. Memory board JP3 connector jumper not installed or mispositioned (occasional)
  4. 4. Defective memory board (occasional)
  5. 5. Defective power board (occasional)
  6. 6. Defective main board (rare)
First check

Verify external power connection

Test point
AC power source and adapter
Expected
Ventilator connected to external power

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivered pressure is low with PEEP low and ventilator auto cycling patient safety

Delivered pressure is low, PEEP is low, and ventilator is auto cycling abnormally

Ranked causes

  1. 1. Leak Compensation is not on (common)
  2. 2. Sense lines are reversed (occasional)
  3. 3. High or low sense lines are occluded or at wye loose or leaking (common)
  4. 4. Exhalation drive line leaking or loose (common)
  5. 5. Exhalation valve leaking during inspiration (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power-on self-test failure patient safety

POST messages (CPU, SRAM, INT VECTOR, ROM CRC, EEPROM) do not complete or ventilation does not begin within 20 seconds of power-on

Ranked causes

  1. 1. CPU failure or malfunction (unknown)
  2. 2. SRAM failure or malfunction (unknown)
  3. 3. INT VECTOR failure (unknown)
  4. 4. ROM CRC error (unknown)
  5. 5. EEPROM failure or malfunction (unknown)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Repeated DISC/SENSE alarms with XDCR FAULT also occurring patient safety

Combination of disconnect/sense and transducer fault alarms occurring repeatedly

Ranked causes

  1. 1. Pressure Control or Pressure Support set below PEEP (common)
  2. 2. Internal flexible tubing occluded or pinched (occasional)
  3. 3. Internal flexible tubing has pinhole leaks or leaking at connections (occasional)
  4. 4. Defective transducer on analog board (occasional)
  5. 5. Defective auto zero solenoid (occasional)
First check

Verify Pressure Control and Pressure Support settings

Test point
Ventilator control panel settings display
Expected
PC and PS values set at or above PEEP value

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator does not power on

Unit fails to power up or turn on when power button is pressed

Ranked causes

  1. 1. Defective AC source, adapter, or depleted internal battery (common)
  2. 2. Memory board JP3 connector jumper not installed or mispositioned (common)
  3. 3. Defective memory board (occasional)
  4. 4. Defective power board (occasional)
  5. 5. Defective main board (rare)
First check

Verify external AC power connection

Test point
AC adapter connection
Expected
Power cord fully seated to adapter and wall outlet

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator will not exhale, repeated HIGH PRES alarms patient safety

Ventilator stuck in inspiration with continuously rising pressure; exhalation function fails; turbine stops and auto-cycles to HIGH PRES

Ranked causes

  1. 1. Diaphragm installed backwards or incorrectly seated in exhalation valve (common)
  2. 2. Sense lines occluded, pinched or reversed (common)
  3. 3. Internal flexible tubing occluded or pinched (occasional)
  4. 4. Internal flexible tubing has pinhole leaks or leaking at connections (occasional)
  5. 5. Defective exhalation drive solenoid (occasional)
First check

Visually inspect exhalation valve diaphragm orientation

Test point
Exhalation valve interior
Expected
Diaphragm and spring correctly seated per Chapter 4 diagram

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm does not sound during ALARM test

No audible alarm during alarm test checkout

Ranked causes

  1. 1. Defective power board (common)
  2. 2. Ventilator does not have power board P/N 15000 installed (occasional)
  3. 3. Audible alarm did not sound long enough before test termination (occasional)
  4. 4. Alarm sounder disconnected (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivered flow is high patient safety

Ventilator delivering higher than prescribed tidal or minute volume

Ranked causes

  1. 1. Circuit leak (common)
  2. 2. Sense lines reversed (occasional)
  3. 3. Pressure sense line or port issue (common)
  4. 4. Control settings incorrect (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivered flow is low patient safety

Ventilator delivering lower than prescribed tidal or minute volume

Ranked causes

  1. 1. Failed auto zero (common)
  2. 2. Leak Compensation not enabled (common)
  3. 3. Bypass flexible tube issues (occasional)
  4. 4. Failed calibration or transducer drift (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Exhalation valve leaking during inspiration patient safety

Exhalation valve allows gas to escape to atmosphere during inspiratory phase instead of maintaining pressure

Ranked causes

  1. 1. Diaphragm installed incorrectly or spring misseated (common)
  2. 2. Exhalation diaphragm worn or defective (occasional)
  3. 3. PEEP spring worn or defective (occasional)
  4. 4. Exhalation valve body defective or damaged (occasional)
First check

Verify diaphragm and spring correct orientation and seating

Test point
Exhalation valve interior
Expected
Diaphragm and spring seated per Chapter 4 diagram; no gaps

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Leak test fails

Leak test checkout fails

Ranked causes

  1. 1. Circuit connections or accessories are leaking (common)
  2. 2. Internal flexible tubing occluded, pinched, or leaking (common)
  3. 3. Over pressure relief valve, subambient relief valve, or flow valve leaking (common)
  4. 4. VHome setting does not match flow valve (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Repeated RESET alarms with CRC, STACK, POST, or RUNAWAY event log

RESET alarm recurs; Event Log shows CRC, STACK, POST, or RUNAWAY error codes

Ranked causes

  1. 1. Electrostatic discharge (ESD) (occasional)
  2. 2. Defective memory board (common)
  3. 3. Defective main board (common)
  4. 4. All other causes (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Repeated XDCR FAULT alarms

Transducer fault alarms occur repeatedly; pressure measurement or sensing circuitry failure

Ranked causes

  1. 1. Internal flexible tubing occluded or pinched (common)
  2. 2. Internal flexible tubing has pinhole leaks or leaking at connections (common)
  3. 3. Defective transducer on analog board (occasional)
  4. 4. Defective auto zero solenoid (occasional)
First check

Inspect internal flexible tubing for occlusions and kinks

Test point
Solenoid manifold and analog board tubing connections
Expected
All tubing clear of obstructions and not pinched

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Sensitivity does not appear to be accurate patient safety

Ventilator trigger sensitivity setting does not match actual sensitivity response

Ranked causes

  1. 1. High or low sense lines occluded (occasional)
  2. 2. High or low sense ports in wye occluded (occasional)
  3. 3. Condensation in sense lines (occasional)
  4. 4. Failed calibration (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator does not operate from internal battery

Unit fails to run when on battery power or shuts off when external power removed

Ranked causes

  1. 1. Internal battery depleted (common)
  2. 2. Internal battery not connected (occasional)
  3. 3. Defective internal battery (occasional)
  4. 4. Defective power board (occasional)
First check

Measure battery runtime at nominal settings

Test point
Ventilator output during battery operation
Expected
Runs for more than 40 minutes at nominal settings

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator is on, gas is not delivered and turbine is running patient safety

Ventilator powered and turbine spinning but no gas output to patient

Ranked causes

  1. 1. Defective or disconnected solenoids (occasional)
  2. 2. Defective or disconnected flow valve (occasional)
  3. 3. Defective turbine (occasional)
  4. 4. Defective purge solenoids (rare)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator makes a high pitched noise when in Standby caution

Audible high-pitched sound emitted while ventilator in Standby mode

Ranked causes

  1. 1. Battery charge circuit running in bulk charge (common)
  2. 2. Defective or disconnected fan (occasional)
  3. 3. Defective power board (occasional)
  4. 4. Defective thermo conductive heat sink pads (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator won't trigger at sensitivity setting of 1 Lpm patient safety

Ventilator does not initiate breath when patient effort is 1 Lpm or below set sensitivity

Ranked causes

  1. 1. Patient effort inadequate (occasional)
  2. 2. Failed auto zero (occasional)
  3. 3. Leak Compensation is not on (occasional)
  4. 4. Failed calibration (occasional)
First check

Perform auto zero calibration

Test point
XDCR ZERO menu
Expected
Auto zero completes without error

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

A control does not operate

Front panel control is unresponsive or non-functional

Ranked causes

  1. 1. Front panel ribbon cable not properly connected (common)
  2. 2. Front panel ribbon cable damaged (occasional)
  3. 3. Defective switch or main board (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm does not sound at all patient safety

Audible alarm system completely silent; no alarm sounds for any alarm condition

Ranked causes

  1. 1. Ventilator did not run for more than 1 minute (common)
  2. 2. Defective or disconnected sounder (occasional)
  3. 3. Defective power board (occasional)
First check

Run ventilator for minimum 1 minute

Test point
Ventilator operation
Expected
Unit runs continuously for at least 1 minute

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm too soft during ALARM test

Audible alarm volume is insufficient during alarm test checkout

Ranked causes

  1. 1. Alarm volume set too low (common)
  2. 2. Alarm sounder blocked (common)
  3. 3. Defective alarm sounder (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Charge Status LED is flashing amber

Charge Status LED flashes amber during battery evaluation or charging phase

Ranked causes

  1. 1. Internal battery charging normally (common)
  2. 2. Defective internal battery (occasional)
  3. 3. Defective power board (occasional)
First check

Observe Charge Status LED behavior during charging

Test point
Charge Status LED on ventilator panel
Expected
Flashing amber for up to 1 hour if deeply discharged, then solid amber while charging

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery Charge Status LED is solid red

Charge Status LED displays solid red indicating internal battery disconnection or defect

Ranked causes

  1. 1. Internal battery not connected (common)
  2. 2. Defective internal battery (occasional)
  3. 3. Defective power board (occasional)
First check

Verify internal battery is physically connected

Test point
Internal battery connector inside ventilator
Expected
Battery connector fully seated and secured

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery does not reach full charge

Internal battery fails to charge to full capacity or depletes too quickly during use

Ranked causes

  1. 1. Internal battery deeply discharged (common)
  2. 2. Defective internal battery (occasional)
  3. 3. Defective power board (occasional)
First check

Charge battery for extended period if deeply discharged

Test point
Charge Status LED indicator
Expected
Solid amber LED after initial evaluation phase

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Control display flashing when setting a control

Control value display flashes while attempting to adjust a setting

Ranked causes

  1. 1. Control setting is limited by other control values (common)
  2. 2. Wrong model selected in VENT MTNCE (occasional)
  3. 3. LED misalignment (rare)
First check

Identify which controls are flashing

Test point
Display screen during adjustment
Expected
Other controls flash to indicate dependency

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

DEFAULTS alarm with invalid or out-of-range settings

DEFAULTS alarm occurs; some or all control settings were invalid or out of range on power up and were restored to defaults

Ranked causes

  1. 1. Electrostatic discharge (ESD) (occasional)
  2. 2. Defective main board (common)
  3. 3. Defective memory board (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display or LED does not illuminate during DISPLAY test

A display or LED fails to illuminate during display test checkout

Ranked causes

  1. 1. Misaligned LED (occasional)
  2. 2. Defective LED or display (common)
  3. 3. Defective main board (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

HIGH PRES alarm occurred but did not sound patient safety

High pressure condition detected and displayed but audible alarm did not activate

Ranked causes

  1. 1. Alarm silence was already active (common)
  2. 2. High pressure alarm delay is on (common)
  3. 3. Alarm automatically silenced after 3 seconds (occasional)
First check

Check Silence/Reset LED status

Test point
Silence/Reset LED indicator
Expected
Green or off; not red

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect message displayed when rotary switch turned during CONTROL test

Correct message not displayed when rotary switch is turned, or incorrect message is displayed

Ranked causes

  1. 1. Front panel ribbon cable not properly connected (occasional)
  2. 2. Front panel ribbon cable damaged (occasional)
  3. 3. Defective main board (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Oxygen enrichment out of specification patient safety

External oxygen monitor does not read 55% to 65% O2 when O2% control is set to 60% with 0-80 PSIG oxygen supply

Ranked causes

  1. 1. Oxygen mixing valve malfunction or blockage (occasional)
  2. 2. Oxygen supply pressure out of range (occasional)
  3. 3. External oxygen monitor malfunction (occasional)
First check

Connect oxygen supply and set O2% control to 60%

Test point
Ventilator O2% control
Expected
O2% control set to 60

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Repeated HW FAULT alarms with delivered pressures and volumes off

Hardware fault alarms occurring with inaccurate pressure and volume delivery; multiple system failures

Ranked causes

  1. 1. Flow valve malfunction or disconnection (common)
  2. 2. Motor board defective or failed (occasional)
  3. 3. Event Log shows SYNC ER1 or HOME ER1 (occasional)
First check

Check Event Trace for SYNC ER1 or HOME ER1 errors

Test point
Event Trace log in ventilator memory
Expected
No SYNC ER1 or HOME ER1 errors present

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Repeated HW FAULT alarms with EEPROM event log

HW FAULT alarm recurs; Event Log shows EEPROM error

Ranked causes

  1. 1. Defective power board (common)
  2. 2. Electrostatic discharge (ESD) (occasional)
  3. 3. Defective main board (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Repeated HW FAULT alarms with fan not turning and excessive heat caution

Hardware fault alarm with fan failure; ventilator overheating due to non-functional cooling fan

Ranked causes

  1. 1. Defective or disconnected fan (common)
  2. 2. Fan housing screws over-tight (occasional)
  3. 3. Event Log shows FAN FLT1 (occasional)
First check

Visually observe fan operation through side grill

Test point
Fan grill on side of ventilator case
Expected
Fan blades rotating smoothly at normal speed

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator auto cycling with very small monitored volumes patient safety

Ventilator cycles abnormally with low monitored volumes; RT XDCR DATA item shows negative flows during exhalation and positive flows during inspiration

Ranked causes

  1. 1. Sense lines are reversed (occasional)
  2. 2. Diaphragm installed backwards or incorrectly seated in exhalation valve (occasional)
  3. 3. Sense lines occluded, pinched or reversed (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ventilator does not work with LTM Graphics Monitor

Ventilator cannot communicate with or operate LTM Graphics Monitor

Ranked causes

  1. 1. Communications setting is not set to MONITOR mode (common)
  2. 2. Ventilator requires upgrades for LTM compatibility (occasional)
  3. 3. Defective connections between LTM Graphics Monitor and ventilator (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Volume mode operation out of specification patient safety

Exhaled tidal volume, I:E ratio, breath rate, or minute volume readings outside expected ranges during volume assist/control operation

Ranked causes

  1. 1. Incorrect ventilator settings (common)
  2. 2. Circuit leak or patient disconnect (common)
  3. 3. Test lung or patient compliance issue (occasional)
First check

Verify ventilator mode and basic settings are correctly configured

Test point
Ventilator control panel
Expected
Mode: Volume Assist/Ctrl; BPM: 12; Tidal Volume: 500 ml; Insp. Time: 1.0 sec; PEEP: 5 cmH2O; O2%: 21; Sensitivity: 3; High Pressure Limit: 100; Low Pressure alarm: 5; Low Min Vol: 1.0 L

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AP transducer out of tolerance

AP transducer reading variance exceeds acceptable tolerance when tested with applied pressures

Ranked causes

  1. 1. Transducer requires recalibration (common)
  2. 2. Analog board defective (occasional)
First check

Verify RISE TIME setting

Test point
VENT OP menu
Expected
Profile 5

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarm too soft during Vent Inop Alarm test patient safety

Vent Inop alarm audible tone is too soft during test

Ranked causes

  1. 1. Alarm sounder blocked (common)
  2. 2. Sounder fails to sound (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bypass Valve suspected faulty patient safety

Tidal volume variance between horizontal and upright positions exceeds 100 mL when measured over 8 consecutive breaths

Ranked causes

  1. 1. Leaking or damaged tubing from flow valve to turbine manifold (common)
  2. 2. Faulty turbine manifold (occasional)
First check

Measure Vte in horizontal position

Test point
Display
Expected
average of 8 consecutive breaths

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Turbine flow valve

Turbine flow valve accuracy checkout and recalibration of flow output

Ranked causes

  1. 1. Flow valve output variance within recalibration range (common)
  2. 2. Flow valve defective or out of range (occasional)
First check

Test at 10 Lpm and 4000 RPM

Test point
Calibrated flow meter output
Expected
9.5 to 10.5 Lpm, 9.5 to 10.5
Tools
Calibrated flow meter

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Confirming audible chirp does not sound during Vent Inop Alarm test

No confirming chirp from Vent Inop alarm during test

Ranked causes

  1. 1. Audible alarm did not sound long enough before test termination (occasional)
  2. 2. Defective power board (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Controls are locked

LOCKED message displays when attempting to select or adjust a control

Ranked causes

  1. 1. Control Lock button engaged (common)
  2. 2. Control Lock button in HARD mode (occasional)
First check

Press Control Lock button

Test point
Control Lock button
Expected
LOCKED message disappears; controls become active

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Date/time/usage meter not incrementing correctly

Date, time, or usage meter does not increment correctly after 1 hour of operation with test lung and power cycle

Ranked causes

  1. 1. Real-time clock malfunction or battery depleted (occasional)
  2. 2. Usage meter counter circuit malfunction (occasional)
First check

Set ventilator date to current date; note date and format displayed

Test point
Ventilator date setting menu
Expected
Current date set in displayed format

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Defective flow valve

Flow valve not calibrated or functioning correctly

Ranked causes

  1. 1. Flow valve requires calibration (common)
  2. 2. Flow valve defective or failed (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivered and monitored volumes, pressures, and sensitivity are off patient safety

Displayed values for volume, pressure, or sensitivity do not match actual values

Ranked causes

  1. 1. Failed calibration (common)
  2. 2. VHome setting does not match flow valve (occasional)
First check

Recalibrate ventilator

Test point
Calibration menu
Expected
All values match actual measured values

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Delivered pressure higher than set pressure on test lung

Pressure readings exceed the set value when testing with a test lung

Ranked causes

  1. 1. Pressure > 40 cmH2O used on small test lung (Pulmonetic Systems or Siemens 190) (common)
  2. 2. Test lung with small aperture connected directly to wye (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Disc/Sense alarm does not activate patient safety

DISC/SENSE alarm fails to activate when high pressure sense line is disconnected from ventilator

Ranked causes

  1. 1. Pressure sense line sensor failure (occasional)
  2. 2. Sense line disconnection detection circuit failure (rare)
First check

Disconnect high pressure sense line and observe alarm response on next breath

Test point
Ventilator high pressure sense port, alarm panel
Expected
DISC/SENSE alarm activates on next breath

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display or LED does not illuminate

A display element or LED is dark or non-functional

Ranked causes

  1. 1. Defective LED or display element (common)
  2. 2. Defective main board (occasional)
First check

Perform display test during startup

Test point
All display elements
Expected
All displays illuminate and respond during test

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failed calibration

Calibration procedure fails to complete or pressure/flow readings remain inaccurate after recalibration

Ranked causes

  1. 1. Calibration procedure incomplete or incorrect (common)
  2. 2. Transducer drifting excessively (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

FDb (Inspired Volume) transducer out of tolerance

FDb transducer reading variance exceeds acceptable tolerance when tested at 30 cmH2O (LTM only)

Ranked causes

  1. 1. Transducer requires recalibration (common)
  2. 2. Analog board defective (occasional)
First check

Apply 30 cmH2O to Flow Differential transducer low side port

Test point
FDb transducer low side port
Expected
FDb readout within +/- 0.5 cmH2O of 30, +/- 0.5 cmH2O

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • FDw (Monitored Volume) transducer out of tolerance
  • Flow valve inaccurate
  • FVd (Delivered Volume) transducer out of tolerance
  • Hi and Low Flow transducer leak test fails
  • High pressure alarm does not activate
  • HW FAULT alarm with AD MMTCH error
  • Internal inlet filter requires cleaning or replacement
  • Keypad ribbon cable damage during main board replacement
  • Low minute volume alarm does not activate
  • Low O2 pressure alarm does not activate
  • Low pressure alarm does not activate
  • Message not displayed when control button pressed during CONTROL test
  • O2 inlet pressure transducer out of tolerance
  • O2% display flashing
  • Over-pressure relief valve exceeds pressure limit
  • Oxygen transducer calibration error
  • Patient Assist port resistance out of specification
  • Power board assembly requires replacement
  • Power Lost alarm does not activate
  • Pressure operation monitoring out of specification
  • Set pressure not reached and turbine humming
  • Set Value knob does not operate
  • Vent Inop alarm does not sound continuously
  • Vent Inop alarm does not sound during Vent Inop Alarm test
  • Vent Inop alarm tone and/or LED fails test
  • Ventilator does not reset when Watchdog test performed
  • Audible alarm too loud during ALARM test
  • Battery Charge Status LED is flashing red
  • Bypass flexible tube pinched
  • Calibration verification after software upgrade
  • Cannot unlock controls
  • Configuration settings check after software upgrade
  • Confirming audible chirp does not sound during ALARM test
  • Control button name not displayed or incorrect name displayed during CONTROL test
  • Control is not selected
  • Control not active in selected mode
  • Controls are limited
  • Defective analog board
  • Defective power board
  • Defective turbine
  • Delivered pressure is low
  • Delivered volume is high
  • Displays are blanked while on battery power
  • Failed auto zero
  • High Pres Limit display flashing
  • Internal battery pack requires replacement
  • Internal flexible tubing has pinhole leaks or leaking at connections
  • Internal flexible tubing occluded or leaking
  • Internal flexible tubing occluded or pinched
  • Leak test fails with LEAK --- FAIL message
  • LED alignment check during board replacement
  • LED portion of Vent Inop Alarm test fails
  • Low Min Vol display flashing
  • Low Pressure display flashing
  • Main board assembly requires replacement
  • Main board installation misalignment
  • Memory board installation with JP3 jumper verification
  • Monitored pressure is high
  • Monitored volumes very high on test lung
  • NO CAL DATA alarm with no calibration displayed
  • O2 blender label check for software version 5.0 or greater on LTV 1000
  • O2 blender label replacement for version 5.0 upgrade on LTV 1000
  • O2 blender solenoid output flows out of specification
  • O2 Leak Test fails
  • OPRV test fails
  • Oxygen transducer leaks
  • PEEP is low
  • Pinhole leaks in bypass flexible tube
  • Power board replacement procedure
  • Pressure Control display flashing
  • Pressure Support display flashing
  • Pressure Support or Pressure Control not functioning properly
  • Repeated HW FAULT alarms with INTRRPT1 or INTRRPT2 event log
  • RESET alarm occurs after POST tests
  • Software version verification after memory board upgrade
  • Vent in Standby
  • Vent Inop alarm tone and/or LED fails 15-second duration test
  • Ventilator gets excessively hot
  • Ventilator running but displays are turned off
  • Ventilator will not allow patient to exhale
  • VHome setting does not match flow valve

81 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
01Power onVentilator powered oninfo
02Power offVentilator powered offinfo
03Set hour meterOperating hours recordedinfo
04Set vent checkEntered VENT CHECK modeinfo
05Apnea mode enteredApnea mode activatedwarning
06Apnea mode exitedApnea mode deactivatedinfo
08High side disconnectHigh pressure sensor circuit openederror
09Low side disconnectLow pressure sensor circuit openederror
10Circuit disconnect exitedDisconnect condition resolvedinfo
11Internal battery empty occurredBattery charge depletedcritical
12Internal battery empty exitedBattery charge restoredinfo
13Internal battery low occurredBattery charge critically lowwarning
14Internal battery low exitedBattery charge above low thresholdinfo
15External power lost occurredAC power disconnectedcritical
16External power lost exitedAC power restoredinfo
17External power low occurredAC voltage below minimumwarning
18External power low exitedAC voltage restored to normalinfo
19XDCR fault occurredTransducer signal loss or fault detectederror
20XDCR fault exitedTransducer fault resolvedinfo
21O2 pressure low occurredOxygen supply pressure below minimumcritical
22O2 pressure low exitedOxygen supply pressure restoredinfo
23O2 pressure high occurredOxygen supply pressure exceeds maximumwarning
24O2 pressure high exitedOxygen supply pressure normalizedinfo
25Defaults setSystem reset to factory defaultswarning
26No calibration data foundSystem startup with missing calibrationerror
27Fan fault occurredCooling fan malfunction detectederror
28Fan fault exitedFan fault resolvedinfo
31Spurious interrupt occurred msMicrocontroller received unexpected interrupterror
32Spurious interrupt occurred lsMicrocontroller received unexpected interrupterror
33ADC mismatchA/D converter channel mismatch detectederror
34ADC mismatch occurredA/D converter mismatch first detectederror
35ADC mismatch clearedA/D converter mismatch resolvedinfo
36Stepper motor lost sync occurredStepper motor lost position synchronizationerror
37Stepper motor lost sync exitedStepper motor sync recoveredinfo
38Stepper motor home failure occurredMotor unable to reach home positionerror
39Stepper motor home failure exitedMotor home sequence recoveredinfo
40EEPROM degradedEEPROM read/write errors detectederror
41Memory CRC check failedCRC mismatch on stored datacritical
42High pressure occurredPeak airway pressure exceeds alarm thresholdcritical
43High pressure exitedPeak airway pressure returned to normalinfo
44Turbine immediate stop occurredEmergency stop activated on turbinecritical
45Turbine set to zero flow occurredTurbine flow reduced to zerowarning
46Turbine emergency stop occurredTurbine emergency stop activatedcritical
47Low minute volume occurredDelivered minute volume below thresholderror
48Low minute volume exitedMinute volume returned to acceptable rangeinfo
49Low peak pressure occurredPeak inspiratory pressure below thresholdwarning
50Low peak pressure exitedPeak pressure returned to normal rangeinfo
51Event log clearedEvent history erased from memoryinfo
52Control settings clearedOperating settings resetwarning
53Date setSystem date configuredinfo
54Time setSystem time configuredinfo
56Stack overflow detectedMicrocontroller stack memory exceededcritical
57POST failurePower-on self-test failedcritical
58Code runaway detectedSoftware execution exceeded time limitcritical
59Watchdog test runWatchdog timer test executedinfo
60Calibration records clearedCalibration data erasederror
61Differential pressure transducer narrow channel faultFDn transducer sensor failureerror
62Differential pressure transducer wide channel faultFDw transducer sensor failureerror
63Differential pressure transducer bi-directional channel faultFDb transducer sensor failureerror
64Airway pressure transducer faultMain airway pressure transducer signal errorerror
65AD mismatch primary channel fault valueA/D converter primary channel mismatcherror
66Turbine Hold Stop occurredTurbine hold stop activatedwarning
67Shutdown for other than pressing On/Standby buttonUnexpected shutdown detectederror
68EEPROM write error during shutdownData write failed to EEPROM at shutdowncritical
68A problem is detected writing data to EEPROM during shutdownMemory write failure at system shutdownerror
69Problem detected with audible alarm circuitryPrimary or redundant alarm circuit malfunctionwarning
70Recovery from audible alarm circuitry problemAlarm circuit fault resolvedinfo
71Alarm sounder errorSpeaker or sounder malfunctionwarning
72Recovery from alarm sounder errorAlarm sounder fault resolvedinfo
73Reserved for future useNot currently assignedunknown
74Reserved for future useNot currently assignedunknown
75Reserved for future useNot currently assignedunknown
76Reserved for future useNot currently assignedunknown
77Reserved for future useNot currently assignedunknown
78Reserved for future useNot currently assignedunknown
79Reserved for future useNot currently assignedunknown
80Reserved for future useNot currently assignedunknown
81Reserved for future useNot currently assignedunknown
82Reserved for future useNot currently assignedunknown
83Reserved for future useNot currently assignedunknown
84Reserved for future useNot currently assignedunknown
85Reserved for future useNot currently assignedunknown
86Reserved for future useNot currently assignedunknown
87Reserved for future useNot currently assignedunknown
88Reclaims all incorrectly recognized bad EEPROM recordsSoftware 5.01 recovery routine executedwarning
CAL FAILCalibration procedure failed or abortedError encountered during any calibration step (Valve Differential, O2 inlet, or Motor Drive)error
DEFAULTSControl settings restored to default valuesInvalid or out-of-range control settings detected on power upwarning
DISC/SENSEDisconnect or sense circuit failure alarmPressure sensing circuit fault or circuit disconnection detectederror
HIGH PRESHigh pressure condition detectedAirway pressure exceeds high pressure limit setting during ventilationwarning
HW FAULTHardware fault detected; multiple potential causesInternal hardware failure or circuit error detected during power-up self-test or operationcritical
HW FAULT (EEPROM)Hardware fault related to EEPROM memoryEEPROM error detected during operationerror
HW FAULT (INTRRPT1/INTRRPT2)Hardware fault related to interrupt errorsInterrupt errors detected during operationerror
LMV LPPS OFFLow Minute Volume alarm is turned off and the LPP ALARM has been set to VC/PC ONLYLMV LPPS OFF message displayed on ventilator screeninfo
LMV OFFLow Minute Volume alarm is turned offLMV OFF message displayed on ventilator screeninfo
LOW O2 PRESLow oxygen pressure conditionOxygen inlet pressure falls below acceptable level during oxygen deliverywarning
LPPS OFFLPP ALARM has been set to VC/PC ONLY modeLPPS OFF message displayed on ventilator screeninfo
NO CAL DATAMissing or failed calibration dataCalibration data not found or invalid during operationwarning
RESETSystem reset alarm; ESD or internal reset eventElectrostatic discharge or internal watchdog timer reset during normal operationwarning
RESET (CRC/POST/RUNAWAY/STACK)Reset alarm with various error types (CRC, POST, RUNAWAY, STACK)Occurs at conclusion of POST after tests that cause ventilator to be inoperative; can also recur repeatedlywarning
XDCR FAULTTransducer circuit fault alarmPressure transducer or analog measurement circuit failure detectederror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI