Picker Q: troubleshooting and error codes
The ServiceForge knowledge base holds 16 documented symptoms for the Picker Q, with 20 ranked causes, 7 diagnostic checks with test points and expected values, and 123 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Couch is not moving patient safety
CT scanner table fails to respond to movement commands in servo mode
Ranked causes
- 1. Programmed position interrupt not enabled or failed to occur within expected time (common)
- 2. Data cable cut or DREQ signal grounded (occasional)
Verify exposure armed signal
- Test point
- XSC STAT register 5th nibble, Bit 1
- Expected
- Bit 1 = 1 (exposure armed)
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data acquisition did not complete properly
Either MPM or GPU failed to complete data acquisition correctly. Possible DAVG bit stuck on DPPC or faulty GPU.
Ranked causes
- 1. DAVG bit stuck on DPPC (unknown)
- 2. GPU data acquisition malfunction (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid GPU command received caution
Graphics Processing Unit cannot process command from touchpanel; couch or scanner function rejected
Ranked causes
- 1. GPU not programmed to understand command or GPU in incompatible state (occasional)
- 2. Crash pad engaged during scan or pilot (rare)
Verify GPU state
- Test point
- GPU status register
- Expected
- GPU not in active scan or pilot mode when new command issued
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SIF FIFO status error during or after scan
Incorrect SIF FIFO status detected. FIFO overflow bit should never be set during or after scan; FIFO empty bit should be set after scan completes.
Ranked causes
- 1. FIFO overflow during scan or improper status during acquisition (unknown)
- 2. FIFO empty bit not set after scan (unknown)
Monitor FIFO status during scan
- Test point
- FIFO status register
- Expected
- Overflow bit=0 during scan, Empty bit=1 after scan completion
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ANB error during scan communication
Error in communication with GPU or XSC. Max retries indicates loss of communications; max errors indicate faulty communication quality.
Ranked causes
- 1. XSC communication fault via Slip Rings 9, 10, 11, 12 (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad function parameter from GPU or XSC
Parameter returned from GPU or XSC does not match parameter sent by SIF
Ranked causes
- 1. GPU or XSC command echo mismatch (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
DMA channel timeout waiting for data
Data Acquisition Manager waiting for data on DMA channel but timeout occurred
Ranked causes
- 1. DMA data source not providing data or DMA channel inactive (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Exposure delay request not executed before next exposure
Normal delay request by XSC was not completed before the next exposure request was issued
Ranked causes
- 1. XSC delay request timeout or queue overflow (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
GPU device state error
GPU in unexpected device state. ASCII message contains device in error and expected state; actual state is in SIF error buffer.
Ranked causes
- 1. GPU state machine error or initialization failure (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
GPU/XSC command reply does not match request
Command sent by SIF does not match the reply received from GPU or XSC
Ranked causes
- 1. GPU or XSC processing error or response error (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Illegal SIF command for current state
Computer attempted command that is illegal given SIF's current operational state
Ranked causes
- 1. Software command sequencing error or state machine logic error (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser pilot positioning error during scan frame calibration radiation
Laser is turned on while scan frame is positioned for pilot scan, generating error condition
Ranked causes
- 1. Operator activates laser during pilot scan frame position (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SIF fatal error: cannot boot
SIF cannot boot and is inoperable. System cannot function without operational SIF.
Ranked causes
- 1. SIF board hardware fault or firmware corruption (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SIF state machine error during action
Action started by SIF is incompatible with SIF's current operational state
Ranked causes
- 1. SIF state machine logic error or state transition fault (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
X-ray system interlock preventing exposure high voltage
X-ray main contactor not energized or X-ray interlock condition active
Ranked causes
- 1. X-ray main contactor fault or interlock active (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
XSC device state error
XSC in unexpected device state. ASCII message contains device in error and expected state; actual state is in SIF error buffer. Check XSC STAT 2nd LSB nibble Bit 4 for boot status.
Ranked causes
- 1. XSC boot failure or improper initialization (unknown)
Check XSC STAT register bit 4 (boot status indicator)
- Test point
- XSC STAT register, 2nd LSB nibble, Bit 4
- Expected
- Bit 4 = 0 (boot successful)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1 | Create exchange failure | Failure to create message exchange object in operating system | error |
10 | Cannot ident process | Failure to identify or locate process | warning |
100 | Cannot get a Regulus Server | System attempts to obtain Regulus Server connection | error |
1000-1099 | Scan errors reported by the XSC | Scan acquisition or XSC operation | error |
100-120 | RpIOS errors | RpIOS operating system operation | error |
1002 | High voltage interface driver error | High voltage interface failure | critical |
1003 | Low speed rotor software error | Low speed rotor control failure | error |
1004 | Small drives error | Small drive system failure | error |
1005 | Filament software error | Filament control system failure | error |
1006 | X-ray software error | X-ray generation failure | critical |
1007 | Kill error - XSC rebooted software | XSC software reset or crash | warning |
1008 | X-ray overtime error | X-ray exposure time exceeded | warning |
1009 | Heat calculator error | Heat calculation system failure | warning |
101 | Return/release a Regulus Server | System attempts to return or release Regulus Server | error |
1010 | Low speed rotor hardware error | Low speed rotor mechanical or electrical failure | error |
1011 | HVPSI hardware error | High voltage power supply interface hardware failure | critical |
1012 | CRX Generator Error | Generator control fault detected | error |
1013 | XSC Fault Latch Error | XSC fault latch activated | error |
1014 | Heat Calculator Error | Heat calculation subsystem failure | error |
1016 | High Speed Starter Fault Latch | Starter fault latch activated | error |
1017 | High Speed Starter Driver Error | Starter driver fault detected | error |
1018 | Improper ESTOP Sequencing | Emergency stop sequence violation | warning |
102 | Cannot receive message from regulus queue | System attempts to receive message from Regulus queue | error |
103 | Cannot send message to regulus queue | System attempts to send message to Regulus queue | error |
104 | Unable to get message from regulus queue | System attempts to retrieve message from Regulus queue | error |
105 | cannot open a file | File I/O operation attempted | error |
106 | cannot close a file | File close operation attempted | error |
107 | cannot read a file | File read operation attempted | error |
108 | cannot write to a file | File write operation attempted | error |
109 | cannot lseek in a file | File seek operation attempted | error |
11 | Cannot delete process | Failure to terminate or delete process | error |
110 | cannot change/set mode of file | File mode change operation attempted | error |
1-10 | Scan and pipp setup errors | System startup or scan initialization | error |
1100-1199 | DAQMAN Errors | Data acquisition manager subsystem error | error |
111 | cannot restart an agent | Agent restart requested | error |
112 | cannot download a task into an agent | Task download to agent attempted | error |
11-25 | File system errors | File system operation or access | error |
113 | cannot reassign ownership of a task | Task ownership change requested | error |
114 | rplwrite error (Regulus to PSOS write) | Regulus to PSOS write operation | error |
116 | cannot find board | Board detection or access attempted | error |
12 | Cannot suspend process | System attempts to suspend a process | error |
1200-1299 | Calibration Former Errors | Calibration subsystem error | error |
13 | Cannot resume process | System attempts to resume a suspended process | error |
14 | Cannot change priority | System attempts to change process priority | error |
15 | Cannot change mode | System attempts to change process mode | error |
1501-1536 | Scheduler Errors | Image reconstruction request queue or execution failure | error |
150-199 | Regulus errors | Regulus module operation | error |
151 | message control call fails | Message control system called | error |
152 | PIPE function fails | PIPE function called | error |
153 | cannot read pipe | Pipe read operation attempted | error |
154 | cannot write pipe | Pipe write operation attempted | error |
1540-1544 | Optable Manager Error | Optable buffer management fault | error |
1545-1546 | Internal Communications Error | Communication fault between Control processes | error |
1547-1575 | Optable Manager Error | Optable buffer management fault | error |
155 | cannot close pipe | Pipe close operation attempted | error |
156 | cannot fork | Process fork operation attempted | error |
157 | cannot activate task | Task activation requested | error |
1576-1596 | Data Manager Error | Data buffer management failure | error |
1599 | Boot Control Error | High level boot or initialization failure | critical |
16 | Cannot change to supervisor mode | System attempts to elevate to supervisor mode | error |
1601-1609 | Memory Manager Error | Memory usage tracking failure | error |
17 | Cannot allocate segment | System attempts to allocate memory segment | error |
18 | Cannot free segment | System attempts to deallocate memory segment | error |
19 | Cannot assign segment | System attempts to assign segment to process | error |
2 | Attach to exchange error | Failure to attach to existing message exchange | error |
20 | Cannot grab segment | System attempts to acquire segment lock | error |
2002 | Data Acquisition error during scan | DINT count fails to increase from zero at start, does not reach expected value at completion, or stops increasing mid-scan | error |
200-299 | Database errors | Database operation or access | error |
2003 | Invalid System Identifier | System switch setting on CTCC board is not IQ, HP, PGS1, or PGS2 | error |
2004 | Door Open during scan | Scan room door is open while scan is in progress | error |
2006 | Invalid GPU Command | GPU receives command from touchpanel that it cannot process due to incompatible state or unknown command type | error |
2007 | Interference Limit exceeded | Couch positioned too close to gantry during horizontal or vertical servo motion or gantry tilt operation | warning |
2008 | Register Load error | GPU parameter register contents do not match values loaded after write and read verification check | critical |
2009 | Vertical Motion error | Couch vertical position changes by more than 3 mm during 50 msec scan interval | error |
21 | Cannot create partition | System attempts to create new partition | error |
22 | Cannot delete partition | System attempts to delete partition | error |
23 | Cannot get buffer | System attempts to allocate buffer | error |
24 | Cannot return buffer | System attempts to deallocate buffer | error |
25 | Cannot send signal | System attempts to send interprocess signal | error |
26 | No event received | System waits for expected event | warning |
26-99 | Program errors | Software execution or operation | warning |
27 | Same as 26 (no event received) | System waits for expected event | warning |
28 | Pause P error | System attempts pause operation | error |
29 | Announce T error | System attempts announcement operation | error |
3 | reqx function failure or SIF problem | Request exchange operation failed; possibly due to SIF hardware issue | error |
30 | Time and date are not set | System checks time/date configuration on startup | warning |
300-324 | System Monitor Board or Autovoice errors | System monitor or autovoice operation | error |
31 | Error setting time and date | System attempts to set time and date | error |
32 | Cannot initialize device | System attempts device initialization | error |
33 | Cannot open device for I/O | System attempts to open device file | error |
34 | Cannot close device after I/O | System attempts to close device after operation | error |
35 | Cannot read from device | System attempts to read data from device | error |
36 | Cannot write to device | System attempts to write data to device | error |
37 | Device control error | System attempts device control operation | error |
38 | Exception error (Task Died, Reboot!) | System encounters unhandled exception in task | critical |
4 | sendx function failure or SIF problem | Send exchange operation failed; possibly due to SIF hardware issue | error |
400-449 | Opcom errors | Opcom module operation | error |
450-499 | Display errors | Display subsystem operation | error |
5 | Cannot delete exchange | Failure to delete message exchange object | error |
50 | Cannot read (PPLREAD) from another agent | System attempts PPLREAD from remote agent | error |
500-599 | Scan errors reported by the SIF | Scan acquisition or SIF operation | error |
51 | PPCREATEX - creating exchange on another PSOS board fails | System attempts to create exchange on remote PSOS board | error |
512 | ANB error - Communication failure with GPU or XSC | Max retries on communication to GPU or XSC exceeded | error |
513 | Bad function parameter - GPU/XSC parameter mismatch | Parameter returned from GPU/XSC does not match parameter sent by SIF | error |
514 | X-ray error - Exposure prevention | X-ray main contactor not energized or safety interlock enabled | error |
515 | Bad function returned parameter - Command/reply mismatch | SIF command does not match GPU/XSC reply | error |
516 | Exposure request error - Timing violation | Exposure request issued before XSC delay request completed | error |
517 | Bad SIF state error - Incompatible state transition | SIF action initiated that is invalid for current SIF state | error |
518 | Bad SIF command error - Illegal command for state | Computer attempts command illegal for current SIF state | error |
519 | Data acquisition not complete - Acquisition failure | MPM or GPU did not complete data acquisition properly | error |
520 | Bad FIFO status error - FIFO overflow or status error | FIFO overflow detected during scan or FIFO not empty after scan | error |
523 | DMA channel timeout - Data acquisition timeout | Data Acquisition Manager waiting for DMA data; timeout occurred | error |
524 | XSC state error - XSC device state fault | XSC in unexpected state during operation | error |
525 | GPU state error - GPU device state fault | GPU in unexpected state during operation | error |
526 | SIF fatal error - SIF inoperable | SIF cannot boot; system is non-functional | critical |
527 | Laser at pilot start error - Laser positioning conflict | Laser activated while scan frame positioned for pilot scan | warning |
6 | Cannot free process | Failure to deallocate process resources | error |
600-699 | Scareman errors | Scareman module operation | error |
7 | Cannot send a message | Failure to send message on exchange | error |
700-799 | Network errors | Network communication or operation | error |
8 | Cannot spawn task | Failure to create new process/task | error |
800-899 | Auto_Cal errors | Auto_Cal calibration operation | error |
9 | Cannot activate process | Failure to activate or resume suspended process | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI