Physio-Control LUCAS 3: troubleshooting and error codes
The ServiceForge knowledge base holds 102 documented symptoms for the Physio-Control LUCAS 3, with 119 ranked causes, 140 diagnostic checks with test points and expected values, and 73 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Linear sensor calibration failed patient safety
POT CAL value for Control exceeds 6% or Protective exceeds 3.1%, alarm sounds and error code displayed in Control CPU window
Ranked causes
- 1. Suction cup obstruction or interference (common)
- 2. Device not in upright position (common)
- 3. Suction cup not at correct starting position (occasional)
- 4. Compression module defect (rare)
Verify suction cup movement is unobstructed
- Test point
- suction cup
- Expected
- free movement without interference
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
30:2 compression mode malfunction patient safety
Device does not operate correctly in 30:2 mode; sound is unusual, LED does not confirm mode, or alerts do not function
Ranked causes
- 1. Software or mode configuration issue (occasional)
- 2. Sound generation or speaker malfunction (occasional)
- 3. LED or indicator malfunction (occasional)
Activate 30:2 mode; listen for compression cycles
- Test point
- Device speakers and compression mechanism
- Expected
- Compression sound is normal; no unusual noise
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bellows leaking or not sealing properly
Air leakage from bellows assembly or loss of pressure during compression cycles
Ranked causes
- 1. Bellows ring misaligned or improperly seated (common)
- 2. Bellows clamp bolts loose or undertightened (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth connection to PC not working
Device fails to establish or maintain Bluetooth connection with PC during Transmit mode
Ranked causes
- 1. Bluetooth module malfunction or firmware issue (unknown)
- 2. COM Board App not properly updated or configured (unknown)
Initiate Transmit mode and attempt Bluetooth connection to PC
- Test point
- Bluetooth module communication
- Expected
- Device successfully pairs and connects to PC via Bluetooth
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth connectivity failure in transmit mode
Device cannot establish Bluetooth connection to PC or connection is unstable in transmit mode
Ranked causes
- 1. Bluetooth module or driver issue (occasional)
- 2. Software version mismatch or compatibility issue (occasional)
Activate transmit mode; attempt Bluetooth connection to PC
- Test point
- Device Bluetooth module and PC Bluetooth interface
- Expected
- Device successfully pairs and connects to PC
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charging sequence does not start high voltage
Device fails to initiate charging sequence when PSU is connected
Ranked causes
- 1. Power supply not properly connected or faulty (unknown)
- 2. Device charging electronics malfunction (unknown)
Connect PSU to device and observe charging sequence initiation
- Test point
- Power supply connection and charging LED
- Expected
- Charging sequence starts; appropriate LED indicator illuminates
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Claw Mechanism does not lock securely patient safety
Release Ring exhibits play or Claw Mechanism does not remain locked when pressure is applied
Ranked causes
- 1. Claw Lock Mechanism requires lubrication or adjustment (unknown)
- 2. Release Ring has excessive play or wear (unknown)
Check Release Ring play with Claw Mechanism unlocked
- Test point
- Release Ring
- Expected
- Minimal or no play
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Claw mechanism unlocks under pressure patient safety
Claw mechanism does not remain locked when pressure is applied during compression
Ranked causes
- 1. Claw lock mechanism wear or lubrication issue (occasional)
- 2. Claw mechanism internal damage or misalignment (rare)
Engage claw mechanism; apply manual downward pressure on compressor
- Test point
- Claw mechanism lock
- Expected
- Claw remains locked without disengagement
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Compression module misaligned or making noise
Grinding, squealing, or knocking sounds during compression; module may be loose or out of position
Ranked causes
- 1. Compression module mounting screws loose or undertightened (common)
- 2. PCB bracket fasteners loose (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Decompression Spring does not move smoothly
Decompression Spring exhibits stiffness, grinding, or unusual noise during operation
Ranked causes
- 1. Decompression Spring requires lubrication or maintenance (unknown)
- 2. Decompression Spring mechanism fouled or damaged (unknown)
Manually operate Decompression Spring
- Test point
- Decompression Spring mechanism
- Expected
- Smooth motion without unusual noise
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive belt slipping or not driving motor
Motor does not turn or compression cycles are weak; belt may be slipping or misaligned
Ranked causes
- 1. Drive belt worn, damaged, or broken (occasional)
- 2. Belt cover screws loose (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electric motor not running or running poorly high voltage
Motor fails to start, runs intermittently, or produces weak output; compression cycles may be weak or absent
Ranked causes
- 1. Motor mounting screws loose or undertightened (common)
- 2. Motor terminal connector loose or corroded (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hood cover cables disconnected or damaged patient safety
Hood not powering up, no audio output, or communication issues; loose cable connections visible
Ranked causes
- 1. Cable connectors loose or not fully seated (common)
- 2. Cable damaged or pinched during hood installation (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal function test LED not green or test fails
Internal function test does not complete successfully or LED does not show green status
Ranked causes
- 1. Software or firmware issue (occasional)
- 2. LED module malfunction (occasional)
Run internal function test via device interface
- Test point
- Device control panel or software interface
- Expected
- LED displays green; test completes without error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
30:2 mode active LED not green
Active 30:2 LED does not display green during 30:2 mode operation
Ranked causes
- 1. LED failure or 30:2 mode not functioning (unknown)
30:2 mode active LED
- Test point
- LED indicator
- Expected
- Green
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
30:2 mode does not function correctly patient safety
Device produces unusual sound, LED indicators not illuminated correctly, or audible alert not functioning in 30:2 mode
Ranked causes
- 1. Electronics malfunction or LED/audio failure (unknown)
Operate device in 30:2 mode and listen for unusual sound
- Test point
- Device operation in 30:2 mode
- Expected
- Runs without unusual sound
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
30:2 mode does not operate correctly patient safety
Device produces unusual sound, active 30:2 LED does not illuminate green, or audible alert prior to ventilation pause is missing
Ranked causes
- 1. Electronic malfunction or software issue in 30:2 mode (unknown)
Listen for unusual sound during 30:2 mode operation
- Test point
- Device audio output
- Expected
- Device runs without unusual sound
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
30:2 mode LED not active or green
30:2 active LED fails to illuminate green during 30:2 mode operation
Ranked causes
- 1. LED failure or electronics malfunction (unknown)
30:2 active LED status
- Test point
- 30:2 active LED
- Expected
- Green
30:2 mode LED not green or audible alert missing patient safety
30:2 LED fails to illuminate green or audible alert does not sound prior to ventilation pause
Ranked causes
- 1. Electronics or firmware issue affecting 30:2 mode operation (unknown)
Run device in 30:2 mode and verify 30:2 LED illuminates green
- Test point
- 30:2 mode LED
- Expected
- LED illuminates green during compression cycles
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
30:2 mode malfunction patient safety
Device runs with unusual sound, 30:2 LED not green, or audible alert and intermittent LED before ventilation pause not working
Ranked causes
- 1. Software or electronic control issue in 30:2 mode (rare)
Test device operation in 30:2 mode
- Test point
- Device audio, compression rhythm, LED indicator
- Expected
- Runs without unusual sound; active 30:2 LED is green
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
30:2 mode operation abnormal patient safety
Device produces unusual sound, active 30:2 LED not green, or audible alert/LED missing before ventilation pause
Ranked causes
- 1. Electronics or mode control malfunction (unknown)
Run device in 30:2 mode and listen for unusual sounds
- Test point
- 30:2 mode operation
- Expected
- Device runs without unusual sound
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
30:2 mode produces unusual sound
Device produces unusual sound during 30:2 compression/ventilation mode operation
Ranked causes
- 1. Mechanical wear or lubrication issue (unknown)
30:2 mode audio inspection
- Test point
- Device audio output during 30:2 mode
- Expected
- No unusual sound
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Adjust mode suction cup operation abnormal
Suction cup does not run smoothly in adjust mode
Ranked causes
- 1. Suction cup wear or mechanism misalignment (unknown)
Enter adjust mode and observe suction cup operation
- Test point
- Suction cup
- Expected
- Suction cup runs smoothly
Audible alert absent before ventilation pause
Device does not produce audible alert prior to ventilation pause in 30:2 mode
Ranked causes
- 1. Buzzer failure or alert function disabled (unknown)
Audible alert before ventilation pause
- Test point
- Device speaker or buzzer
- Expected
- Audible alert and intermittent LED
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alert or intermittent LED absent before ventilation pause patient safety
Device fails to produce audible alert or intermittent LED prior to ventilation pause in 30:2 mode
Ranked causes
- 1. Electronics or audio circuit malfunction (unknown)
Audible alert and intermittent LED prior to ventilation pause
- Test point
- Audio and LED output
- Expected
- Audible alert and intermittent LED active before ventilation pause
Bellows or Suction Cup damaged caution
Bellows or Suction Cup exhibit damage, cracks, or deterioration
Ranked causes
- 1. Wear, impact damage, or material degradation (unknown)
Bellows and Suction Cup physical inspection
- Test point
- Bellows and Suction Cup
- Expected
- No damage or deterioration
Bellows or Suction Cup malfunction caution
Bellows or Suction Cup not functioning properly
Ranked causes
- 1. Wear, damage, or contamination of Bellows or Suction Cup (common)
Visually inspect Bellows and Suction Cup
- Test point
- Bellows and Suction Cup components
- Expected
- No cracks, deformation, or contamination
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth connection setup
Device must be placed in Bluetooth mode before wireless connection with Program Loader
Ranked causes
- 1. Device not in Bluetooth mode (common)
Device powered off
- Test point
- Device
- Expected
- No LEDs illuminated
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth connection to PC does not work in transmit mode
Device fails to establish Bluetooth connection to personal computer in transmit mode
Ranked causes
- 1. Bluetooth module malfunction or connection issue (unknown)
Enable transmit mode and attempt Bluetooth pairing with PC
- Test point
- Bluetooth module connection
- Expected
- Bluetooth connection to PC is established and working
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth connection to PC fails
Device cannot establish or maintain Bluetooth connection to personal computer in Transmit mode
Ranked causes
- 1. Bluetooth module malfunction or configuration issue (occasional)
Test Bluetooth connection in Transmit mode
- Test point
- Bluetooth radio and PC interface
- Expected
- Connection established and working
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Linear Sensor (POT CAL)
Procedure to calibrate the linear sensor potentiometer after software update or when POT CAL recalibration is needed
Ranked causes
- 1. Routine calibration after software update (common)
Device position setup
- Test point
- device
- Expected
- upright, suction cup ~20mm from top
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carry ball nut does not move smoothly
Carry ball nut produces unusual noise or moves with resistance during operation
Ranked causes
- 1. Carry ball nut requires lubrication (unknown)
Observe carry ball nut movement
- Test point
- Carry ball nut
- Expected
- Smooth movement without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carry Ball Nut does not run smoothly
Carry Ball Nut exhibits stiffness, grinding, or unusual noise during operation
Ranked causes
- 1. Carry Ball Screw requires lubrication (unknown)
Operate Carry Ball Nut through full range
- Test point
- Carry Ball Nut assembly
- Expected
- Smooth motion without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carry ball nut movement abnormal
Carry ball nut does not run smoothly or produces unusual noise
Ranked causes
- 1. Carry ball nut wear or insufficient lubrication (common)
Observe carry ball nut operation for smoothness and noise
- Test point
- Carry ball nut
- Expected
- Smooth operation without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carry ball nut moves with unusual noise
Carry ball nut operation is noisy or not smooth during mechanics test
Ranked causes
- 1. Carry ball nut mechanical wear or inadequate lubrication (unknown)
Carry ball nut operation
- Test point
- Carry ball nut
- Expected
- Runs smoothly without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carry Ball Nut moves with unusual noise or stiffness
Carry Ball Nut does not run smoothly or produces unusual noise during operation
Ranked causes
- 1. Carry Ball Screw requires lubrication (unknown)
Inspect and listen to Carry Ball Nut movement
- Test point
- Carry Ball Nut assembly
- Expected
- Runs smoothly without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carry Ball Nut noise or binding caution
Carry Ball Nut does not run smoothly or produces unusual noise
Ranked causes
- 1. Wear or contamination of ball nut bearing surfaces (occasional)
Test Carry Ball Nut movement
- Test point
- Carry Ball Nut on Ball Screw
- Expected
- Runs smoothly without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carry ball nut operation rough or noisy
Carry ball nut does not run smoothly or produces unusual sound during compression cycles
Ranked causes
- 1. Carry ball nut requires lubrication (common)
Manually cycle carry ball nut and listen for smooth operation
- Test point
- Carry ball nut on ball screw
- Expected
- Runs smoothly without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Carry Ball Nut runs with noise or stiffness
Carry Ball Nut does not run smoothly or produces unusual noise during device operation
Ranked causes
- 1. Carry Ball Nut requires lubrication or maintenance (unknown)
Carry Ball Nut movement inspection
- Test point
- Carry Ball Nut
- Expected
- Runs smoothly without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charge mode does not initiate charging sequence high voltage
Device does not begin charging when PSU is connected in charge mode
Ranked causes
- 1. PSU connection issue or faulty charging circuit (occasional)
Connect PSU to device in charge mode; observe charging indication
- Test point
- PSU connector and charging indicator
- Expected
- Charging sequence initiates; indicator shows charging status
Claw mechanism binding or not moving smoothly
Claw lock does not open/close smoothly or feels stiff during operation
Ranked causes
- 1. Insufficient lubrication on claw moving parts (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Claw Mechanism does not remain locked patient safety
Claw Mechanism fails to stay locked when pressure is applied
Ranked causes
- 1. Claw Lock Mechanism requires lubrication or adjustment (unknown)
Claw Mechanism lock test
- Test point
- Claw Mechanism
- Expected
- Remains locked when pressure is applied
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Claw mechanism does not remain locked under pressure patient safety
Claw mechanism unlocks or moves when pressure is applied during mechanics test
Ranked causes
- 1. Claw lock mechanism worn or inadequately lubricated (common)
Claw mechanism lock under pressure
- Test point
- Claw mechanism
- Expected
- Remains locked when pressure applied
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Claw mechanism release ring play excessive
Release ring exhibits excessive play when claw mechanism is unlocked
Ranked causes
- 1. Release ring wear or claw mechanism wear (unknown)
Inspect release ring for excessive play with claw mechanism unlocked
- Test point
- Release ring
- Expected
- No excessive play
Communication PCB replacement caution
Communication PCB on Controller PCB requires replacement; device needs reprogramming after installation.
Ranked causes
- 1. Communication PCB failure or degradation (unknown)
Visual inspection of Communication PCB for damage or corrosion
- Test point
- Communication PCB surface
- Expected
- No visible damage, corrosion, or component failure
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous compression mode malfunction patient safety
Device does not operate correctly in continuous mode; LED does not confirm mode or does not blink at correct interval
Ranked causes
- 1. Software or mode timing issue (occasional)
Activate continuous compression mode; observe LED indicator
- Test point
- Continuous mode LED
- Expected
- LED is green during continuous operation
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous mode does not function correctly patient safety
Continuous mode LED not illuminated or LED does not blink at correct interval
Ranked causes
- 1. Electronics malfunction or LED failure (unknown)
Activate continuous mode and verify LED illumination
- Test point
- Continuous mode LED indicator
- Expected
- LED is green
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Continuous mode does not operate correctly
Continuous mode LED does not illuminate green or does not blink every 10 strokes
Ranked causes
- 1. Electronic or software malfunction in continuous mode (unknown)
Verify continuous mode LED illumination
- Test point
- Continuous mode LED
- Expected
- LED is green
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous mode LED blink pattern incorrect
LED fails to blink every 10 strokes during continuous mode operation
Ranked causes
- 1. Electronics or timing circuit malfunction (unknown)
Continuous mode LED blink frequency
- Test point
- LED blink pattern
- Expected
- Blinks every 10 strokes
Continuous mode LED not active or green
Continuous mode LED fails to illuminate green during continuous mode operation
Ranked causes
- 1. LED failure or electronics malfunction (unknown)
Continuous mode LED status
- Test point
- Continuous LED
- Expected
- Green
Continuous mode LED not green
Continuous mode LED does not display green during continuous mode operation
Ranked causes
- 1. LED failure or continuous mode not functioning (unknown)
Continuous mode LED
- Test point
- LED indicator
- Expected
- Green
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous mode LED not illuminating or blink rate incorrect
Continuous mode LED fails to illuminate green or does not blink at correct 10-stroke interval
Ranked causes
- 1. Electronics or firmware issue affecting Continuous mode operation (unknown)
Run device in Continuous mode and verify LED illuminates green
- Test point
- Continuous mode LED
- Expected
- LED illuminates green during continuous compression
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous mode malfunction patient safety
Continuous mode LED not green or LED does not blink every 10 strokes
Ranked causes
- 1. Software or electronic issue in continuous compression mode (rare)
Verify continuous mode LED status
- Test point
- Continuous mode LED
- Expected
- Green
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Continuous mode operation abnormal
Continuous mode LED not green or LED does not blink every 10 strokes
Ranked causes
- 1. Electronics or mode control malfunction (unknown)
Verify continuous LED indicator in continuous mode
- Test point
- Continuous LED
- Expected
- LED is green
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Decompression spring movement abnormal
Spring does not move smoothly or produces unusual noise during operation
Ranked causes
- 1. Decompression spring wear or misalignment (unknown)
Observe decompression spring movement for smoothness and noise
- Test point
- Decompression spring
- Expected
- Smooth movement without unusual noise
Decompression spring moves with noise or resistance
Spring movement is not smooth or produces unusual sound during operation
Ranked causes
- 1. Decompression spring requires maintenance or lubrication (common)
Manually operate decompression spring and listen for noise
- Test point
- Decompression spring assembly
- Expected
- Moves smoothly without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Decompression Spring moves with noise or stiffness
Decompression Spring does not move smoothly or produces unusual noise during operation
Ranked causes
- 1. Decompression Spring requires lubrication or maintenance (unknown)
Decompression Spring movement inspection
- Test point
- Decompression Spring
- Expected
- Moves smoothly without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Decompression spring moves with unusual noise
Decompression spring operation is noisy or not smooth during mechanics test
Ranked causes
- 1. Decompression spring mechanical wear or misalignment (unknown)
Decompression spring operation
- Test point
- Decompression spring
- Expected
- Moves smoothly without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Decompression spring moves with unusual noise or stiffness
Decompression spring does not move smoothly or produces unusual noise during operation
Ranked causes
- 1. Decompression spring requires lubrication or has mechanical wear (unknown)
Observe decompression spring movement
- Test point
- Decompression spring
- Expected
- Smooth movement without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Decompression Spring stiffness or noise caution
Decompression Spring does not move smoothly or produces unusual noise
Ranked causes
- 1. Wear, dirt, or lack of lubrication in spring mechanism (occasional)
Test Decompression Spring movement
- Test point
- Decompression Spring
- Expected
- Moves smoothly without unusual noise
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Device does not charge when PSU is connected
- Device exterior condition is damaged or dirty
- Device exterior damage or bellows condition issues
- Device exterior damage or contamination
- Device exterior damaged or dirty
- Device exterior damaged or soiled
- Device produces unusual sound during 30:2 mode
- Device runs with unusual sound during 30:2 mode
- Electric fan and mesh grill require cleaning
- Electric Fan or Mesh Grill contaminated
- Fabric condition issues with straps or carrying bag
- Fabric straps or carrying bag damaged
- Fabrics are damaged or worn
- Fabrics worn or damaged
- Hood cannot be opened or removed
- Internal function test LED does not show green
- Internal function test LED not green
- Last service date not recorded on PC
- Last Service Date not set to current date
- Last service date not set to PC date
- Last Service Date not synchronized with PC Date
- LED does not blink at correct interval in continuous mode
- Main Body/Frame Assembly replacement
- Patient straps or carrying bag damage or deterioration
- Patient straps or stabilization strap degraded
- PAUSE LED does not illuminate green
- PAUSE LED not green
- PAUSE mode LED not green
- Power Inlet replacement
- Program Loader Bluetooth initialization
- Release ring has excessive play
- Release Ring play detected in Claw Mechanism
- Release ring play detected when claw unlocked
- Service date not set correctly
- Software installation required
- Software version and checksum not verified
- Software version mismatch or checksum error
- Software version requires verification or update
- Suction Cup does not run smoothly in Adjust mode
- Suction Cup operation rough in Adjust mode
- USB connection setup
- View device error codes and warnings
42 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
ALARM TYPE ROM TEST | Protective CPU internal ROM test failure at startup | ROM self-test fails during power-on | critical |
BATT_10_MIN_ALARM | Battery voltage drops below 25.3V | Battery voltage indicates low charge state | warning |
BATT_EMPTY_ALARM | Battery voltage drops below 25V | Battery depleted during operation | critical |
BATT_HIGH_TEMP_ALARM | Battery temperature exceeds 70°C | Battery temperature sensor detects >70°C | critical |
CHARGER_STOP | Charger fault detected with motor disabled | Charger system malfunction during operation | critical |
COM_B_TIMEOUT (0x55) | I2C battery charger communication timeout | No I2C communication with Charger system for >100 seconds | critical |
COM_I2C (0x56) | I2C communication faults excessive | I2C bus communication errors above threshold | warning |
COM_P_TIMEOUT (0x50) | I2C protective system communication timeout | No I2C communication with Protective system for >100 seconds | critical |
COM_RTC (0x53) | Real-time clock fault | RTC read failure at startup or SPI communication fault | warning |
COM_SD_MEM (0x54) | SD card communication fault | SD card not inserted or broken | warning |
CONTROL_LIFETICK | I2C communication fault with Control CPU (warning) | I2C communication failure at mode change | warning |
CONTROL_STOP | Control fault detected with motor disabled | Control system malfunction during operation | critical |
INT_1msCODE (0x3A) | 1ms code access timeout | Halted or disturbed Control CPU response | error |
INT_C_24V (0x32) | Control 24V supply less than 18V | Main power supply drops below 18V | error |
INT_C_5V (0x31) | Control 5V supply out of range | Control 5V less than 4.5V or greater than 5.5V | error |
INT_C_HALL (0x3C) | Motor Hall sensor fault | Hall sensor signal failure detected | warning |
INT_CODE (0x36) | Access of vital function code fault (10ms) | Control CPU hangs or timing violation detected | critical |
INT_C_POTCAL (0x3D) | Control potentiometer linearity fault | Control potentiometer response is too non-linear | critical |
INT_C_REF (0x33) | Control 2.5V reference out of range | Control 2.5V reference less than 2V or greater than 3V | error |
INT_C_TEMP (0x34) | Too high internal board temperature (>85°C) | Control board temperature exceeds 85°C limit | critical |
INT_CURRENT (0x38) | Motor current fault exceeds 20A | Mean current to motor greater than 20A | error |
INTERNAL TEMPERATURE | Protective board temperature exceeds 85°C | Temperature sensor on Protective board detects >85°C | critical |
INT_P_5V (0x30) | Protective 5V supply out of range | Protective 5V less than 4.5V or greater than 5.5V | error |
INT_POTHALL (0x37) | Potentiometer position and Hall sensor mismatch | Mismatch between potentiometer and Hall sensor readings after correct start | warning |
INT_P_POTCAL (0x3E) | Protective potentiometer linearity fault | Protective potentiometer response is too non-linear | critical |
INT_P_REBOOT (0x39) | Protective CPU reboot in active mode | Protective board resets during active operation | error |
PISTON_TIME_OUT | Protective potentiometer detects no 50% stroke movement for 10 seconds | Piston does not move 50% of stroke within 10 second window | critical |
RUN_13N (0x49) | 13N release spring force high (>10mm) | Release force displacement exceeds 10mm | warning |
RUN_ADJUST (0x46) | Incorrect motor power at adjust mode | Motor power greater than 25W during adjust servo operation | warning |
RUN_NO_POTCAL (0x4B) | Potentiometer not calibrated | System detects potentiometer has not been calibrated | critical |
RUN_POTCAL (0x4A) | Potentiometer calibration fault | Potentiometer not moved during calibration or read error detected | critical |
RUN_PROTECTIVE (0x45) | Halt or reverse by protective system during active moving mode (>1s) | Protective system disables active mode during operation | critical |
RUN_P_STOP (0x4C) | Protective system has stopped operation | Protective system error detected | critical |
RUN_RATIO (0x44) | Piston ratio fault (50 consecutive strokes) | Compression ratio deviates from preset ratio +/- 20% over 50 strokes | critical |
RUN_TIMEDOWN (0x48) | Compression stroke time too long or too shallow | 50 consecutive strokes with timing or depth out of range | critical |
RUN_TIMEOUT (0x40) | Piston not moved during active moving mode (>1s) | No motor movement detected during >1 second of active mode | critical |
RUN_TIMEUP (0x47) | Time up too long >0.15s (50 consecutive strokes) | Release phase exceeds 0.15 seconds during 50 strokes | critical |
RUN_TOO_DEEP (0x41) | Piston too deep (compression depth excessive) | Piston depth exceeds normal range during operation | critical |
RUN_TOO_SHALLOW (0x42) | Piston too shallow (50 consecutive strokes @ 100 bpm) | 50 consecutive strokes measure <target depth minus 10mm at 100bpm | critical |
T1_BATT_COM_ALARM | Battery communication fault | Battery communication error detected at startup | warning |
T1_BATT_COPYRIGHT_ALARM | Incorrect copyright string in battery | Battery copyright verification fails at startup | warning |
T1_BATT_CS_ALARM | Battery communication checksum fault | Battery communication checksum validation failure | warning |
T1_B_OFF (0x2A) | Battery charger did not respond to startup test | Charger CPU does not complete startup test | warning |
T1_CAL_EE (0x15) | Control potentiometer calibration data checksum fault | Calibration data corruption detected | critical |
T1_C_DOWN (0x2D) | Control not moving down before reverse test at startup | Motor does not move down during startup reverse test sequence | warning |
T1_C_EE (0x10) | Control EE-Prom checksum data error | CPU fault detected during initialization | critical |
T1_C_INITCURR (0x20) | High current detected at startup | High current read during initial startup | warning |
T1_C_LOWCURR (0x24) | Too low current detected at first current test | Motor current below expected value during startup test | warning |
T1_C_POT (0x28) | POT value incorrect or mismatch at start or no change during moving | Control/protective POT divergence >10mm at start or non-linear response | critical |
T1_C_POT (0x28) | Potentiometer value mismatch at warm start | Control and Protective potentiometer divergence >10mm at warm start | warning |
T1_C_RAM (0x11) | Control preset checksum data error | CPU or communication fault during startup | critical |
T1_C_RELEASE (0x22) | Current detected at release/disable test by control | Unexpected current during release phase | critical |
T1_C_TOTOP (0x2F) | Control timeout at move to top at startup | Motor cannot move to top position within timeout | warning |
T1_E2_PROM | Protective CPU internal E2PROM test failure at startup | EEPROM self-test fails during power-on | critical |
T1_E2_PROM_ALARM | Charger CPU internal E2PROM test failure at startup | EEPROM self-test fails during Charger power-on | critical |
T1_PB_FLASH (0x14) | Incompatible programs in protective board | Program mismatch detected after board change | critical |
T1_P_DISABLE (0x25) | Protective disable failure at current test startup | Protective system fails to disable at expected point | warning |
T1 PISTON RELEASE ERROR | Protective release line operation fault at startup | Current detected during disable test or moved too low during enable test | critical |
T1 PISTON REVERSE ERROR | Reverse test failure at startup | Motor reverse movement incorrect or fault current read at startup | critical |
T1_POWER_OFF_ALARM | Charger startup test incomplete | T1 test does not end correctly | critical |
T1_P_POT (0x27) | Read Protective potentiometer timeout at startup | Protective potentiometer read operation times out | warning |
T1_P_REW (0x2E) | Protective timeout at reverse test at startup | Motor reverse test does not complete within timeout | warning |
T1_P_STATE_1 (0x21) | Start of release test timeout from protective board | Protective board fails to complete release test | critical |
T1_P_STATE_3 (0x23) | Start of current test timeout or current detected by protective | Protective fails current test or detects unexpected current | critical |
T1_P_STATE_5 (0x29) | No end of Protective current test at startup | Protective current test does not complete within timeout | warning |
T1_RAM | Protective CPU internal RAM test failure at startup | RAM self-test fails during power-on | critical |
T1_RAM_ALARM | Charger CPU internal RAM test failure at startup | RAM self-test fails during Charger power-on | critical |
T1_ROM_ALARM | Charger CPU internal ROM test failure at startup | ROM self-test fails during Charger power-on | critical |
T1_START_COM (0x13) | No communication from Protective or Battery-Charger CPUs at startup | I2C communication failure at device startup | warning |
TOO_DEEP_COMPRESSION | Compression depth exceeds safe limits | Excessive compression depth detected during active mode | critical |
TOO_MANY_USES | Battery cycle count exceeds 200 uses | Battery use counter reaches threshold | warning |
TOO_OLD_BATTERY | Battery age exceeds recommended service life | Battery exceeds maximum age specification | warning |
TOO_SHALLOW_COMPRESSION | Compression depth below minimum after 50 strokes | Insufficient compression depth for 50 consecutive strokes | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.