The ServiceForge knowledge base holds 102 documented symptoms for the Physio-Control LUCAS 3, with 119 ranked causes, 140 diagnostic checks with test points and expected values, and 73 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Linear sensor calibration failed patient safety

POT CAL value for Control exceeds 6% or Protective exceeds 3.1%, alarm sounds and error code displayed in Control CPU window

Ranked causes

  1. 1. Suction cup obstruction or interference (common)
  2. 2. Device not in upright position (common)
  3. 3. Suction cup not at correct starting position (occasional)
  4. 4. Compression module defect (rare)
First check

Verify suction cup movement is unobstructed

Test point
suction cup
Expected
free movement without interference

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

30:2 compression mode malfunction patient safety

Device does not operate correctly in 30:2 mode; sound is unusual, LED does not confirm mode, or alerts do not function

Ranked causes

  1. 1. Software or mode configuration issue (occasional)
  2. 2. Sound generation or speaker malfunction (occasional)
  3. 3. LED or indicator malfunction (occasional)
First check

Activate 30:2 mode; listen for compression cycles

Test point
Device speakers and compression mechanism
Expected
Compression sound is normal; no unusual noise

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bellows leaking or not sealing properly

Air leakage from bellows assembly or loss of pressure during compression cycles

Ranked causes

  1. 1. Bellows ring misaligned or improperly seated (common)
  2. 2. Bellows clamp bolts loose or undertightened (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth connection to PC not working

Device fails to establish or maintain Bluetooth connection with PC during Transmit mode

Ranked causes

  1. 1. Bluetooth module malfunction or firmware issue (unknown)
  2. 2. COM Board App not properly updated or configured (unknown)
First check

Initiate Transmit mode and attempt Bluetooth connection to PC

Test point
Bluetooth module communication
Expected
Device successfully pairs and connects to PC via Bluetooth

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth connectivity failure in transmit mode

Device cannot establish Bluetooth connection to PC or connection is unstable in transmit mode

Ranked causes

  1. 1. Bluetooth module or driver issue (occasional)
  2. 2. Software version mismatch or compatibility issue (occasional)
First check

Activate transmit mode; attempt Bluetooth connection to PC

Test point
Device Bluetooth module and PC Bluetooth interface
Expected
Device successfully pairs and connects to PC

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charging sequence does not start high voltage

Device fails to initiate charging sequence when PSU is connected

Ranked causes

  1. 1. Power supply not properly connected or faulty (unknown)
  2. 2. Device charging electronics malfunction (unknown)
First check

Connect PSU to device and observe charging sequence initiation

Test point
Power supply connection and charging LED
Expected
Charging sequence starts; appropriate LED indicator illuminates

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Claw Mechanism does not lock securely patient safety

Release Ring exhibits play or Claw Mechanism does not remain locked when pressure is applied

Ranked causes

  1. 1. Claw Lock Mechanism requires lubrication or adjustment (unknown)
  2. 2. Release Ring has excessive play or wear (unknown)
First check

Check Release Ring play with Claw Mechanism unlocked

Test point
Release Ring
Expected
Minimal or no play

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Claw mechanism unlocks under pressure patient safety

Claw mechanism does not remain locked when pressure is applied during compression

Ranked causes

  1. 1. Claw lock mechanism wear or lubrication issue (occasional)
  2. 2. Claw mechanism internal damage or misalignment (rare)
First check

Engage claw mechanism; apply manual downward pressure on compressor

Test point
Claw mechanism lock
Expected
Claw remains locked without disengagement

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Compression module misaligned or making noise

Grinding, squealing, or knocking sounds during compression; module may be loose or out of position

Ranked causes

  1. 1. Compression module mounting screws loose or undertightened (common)
  2. 2. PCB bracket fasteners loose (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Decompression Spring does not move smoothly

Decompression Spring exhibits stiffness, grinding, or unusual noise during operation

Ranked causes

  1. 1. Decompression Spring requires lubrication or maintenance (unknown)
  2. 2. Decompression Spring mechanism fouled or damaged (unknown)
First check

Manually operate Decompression Spring

Test point
Decompression Spring mechanism
Expected
Smooth motion without unusual noise

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive belt slipping or not driving motor

Motor does not turn or compression cycles are weak; belt may be slipping or misaligned

Ranked causes

  1. 1. Drive belt worn, damaged, or broken (occasional)
  2. 2. Belt cover screws loose (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Electric motor not running or running poorly high voltage

Motor fails to start, runs intermittently, or produces weak output; compression cycles may be weak or absent

Ranked causes

  1. 1. Motor mounting screws loose or undertightened (common)
  2. 2. Motor terminal connector loose or corroded (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hood cover cables disconnected or damaged patient safety

Hood not powering up, no audio output, or communication issues; loose cable connections visible

Ranked causes

  1. 1. Cable connectors loose or not fully seated (common)
  2. 2. Cable damaged or pinched during hood installation (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal function test LED not green or test fails

Internal function test does not complete successfully or LED does not show green status

Ranked causes

  1. 1. Software or firmware issue (occasional)
  2. 2. LED module malfunction (occasional)
First check

Run internal function test via device interface

Test point
Device control panel or software interface
Expected
LED displays green; test completes without error

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

30:2 mode active LED not green

Active 30:2 LED does not display green during 30:2 mode operation

Ranked causes

  1. 1. LED failure or 30:2 mode not functioning (unknown)
First check

30:2 mode active LED

Test point
LED indicator
Expected
Green

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

30:2 mode does not function correctly patient safety

Device produces unusual sound, LED indicators not illuminated correctly, or audible alert not functioning in 30:2 mode

Ranked causes

  1. 1. Electronics malfunction or LED/audio failure (unknown)
First check

Operate device in 30:2 mode and listen for unusual sound

Test point
Device operation in 30:2 mode
Expected
Runs without unusual sound

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

30:2 mode does not operate correctly patient safety

Device produces unusual sound, active 30:2 LED does not illuminate green, or audible alert prior to ventilation pause is missing

Ranked causes

  1. 1. Electronic malfunction or software issue in 30:2 mode (unknown)
First check

Listen for unusual sound during 30:2 mode operation

Test point
Device audio output
Expected
Device runs without unusual sound

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

30:2 mode LED not active or green

30:2 active LED fails to illuminate green during 30:2 mode operation

Ranked causes

  1. 1. LED failure or electronics malfunction (unknown)
First check

30:2 active LED status

Test point
30:2 active LED
Expected
Green

30:2 mode LED not green or audible alert missing patient safety

30:2 LED fails to illuminate green or audible alert does not sound prior to ventilation pause

Ranked causes

  1. 1. Electronics or firmware issue affecting 30:2 mode operation (unknown)
First check

Run device in 30:2 mode and verify 30:2 LED illuminates green

Test point
30:2 mode LED
Expected
LED illuminates green during compression cycles

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

30:2 mode malfunction patient safety

Device runs with unusual sound, 30:2 LED not green, or audible alert and intermittent LED before ventilation pause not working

Ranked causes

  1. 1. Software or electronic control issue in 30:2 mode (rare)
First check

Test device operation in 30:2 mode

Test point
Device audio, compression rhythm, LED indicator
Expected
Runs without unusual sound; active 30:2 LED is green

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

30:2 mode operation abnormal patient safety

Device produces unusual sound, active 30:2 LED not green, or audible alert/LED missing before ventilation pause

Ranked causes

  1. 1. Electronics or mode control malfunction (unknown)
First check

Run device in 30:2 mode and listen for unusual sounds

Test point
30:2 mode operation
Expected
Device runs without unusual sound

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

30:2 mode produces unusual sound

Device produces unusual sound during 30:2 compression/ventilation mode operation

Ranked causes

  1. 1. Mechanical wear or lubrication issue (unknown)
First check

30:2 mode audio inspection

Test point
Device audio output during 30:2 mode
Expected
No unusual sound

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Adjust mode suction cup operation abnormal

Suction cup does not run smoothly in adjust mode

Ranked causes

  1. 1. Suction cup wear or mechanism misalignment (unknown)
First check

Enter adjust mode and observe suction cup operation

Test point
Suction cup
Expected
Suction cup runs smoothly

Audible alert absent before ventilation pause

Device does not produce audible alert prior to ventilation pause in 30:2 mode

Ranked causes

  1. 1. Buzzer failure or alert function disabled (unknown)
First check

Audible alert before ventilation pause

Test point
Device speaker or buzzer
Expected
Audible alert and intermittent LED

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alert or intermittent LED absent before ventilation pause patient safety

Device fails to produce audible alert or intermittent LED prior to ventilation pause in 30:2 mode

Ranked causes

  1. 1. Electronics or audio circuit malfunction (unknown)
First check

Audible alert and intermittent LED prior to ventilation pause

Test point
Audio and LED output
Expected
Audible alert and intermittent LED active before ventilation pause

Bellows or Suction Cup damaged caution

Bellows or Suction Cup exhibit damage, cracks, or deterioration

Ranked causes

  1. 1. Wear, impact damage, or material degradation (unknown)
First check

Bellows and Suction Cup physical inspection

Test point
Bellows and Suction Cup
Expected
No damage or deterioration

Bellows or Suction Cup malfunction caution

Bellows or Suction Cup not functioning properly

Ranked causes

  1. 1. Wear, damage, or contamination of Bellows or Suction Cup (common)
First check

Visually inspect Bellows and Suction Cup

Test point
Bellows and Suction Cup components
Expected
No cracks, deformation, or contamination

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth connection setup

Device must be placed in Bluetooth mode before wireless connection with Program Loader

Ranked causes

  1. 1. Device not in Bluetooth mode (common)
First check

Device powered off

Test point
Device
Expected
No LEDs illuminated

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth connection to PC does not work in transmit mode

Device fails to establish Bluetooth connection to personal computer in transmit mode

Ranked causes

  1. 1. Bluetooth module malfunction or connection issue (unknown)
First check

Enable transmit mode and attempt Bluetooth pairing with PC

Test point
Bluetooth module connection
Expected
Bluetooth connection to PC is established and working

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth connection to PC fails

Device cannot establish or maintain Bluetooth connection to personal computer in Transmit mode

Ranked causes

  1. 1. Bluetooth module malfunction or configuration issue (occasional)
First check

Test Bluetooth connection in Transmit mode

Test point
Bluetooth radio and PC interface
Expected
Connection established and working

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Linear Sensor (POT CAL)

Procedure to calibrate the linear sensor potentiometer after software update or when POT CAL recalibration is needed

Ranked causes

  1. 1. Routine calibration after software update (common)
First check

Device position setup

Test point
device
Expected
upright, suction cup ~20mm from top

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Carry ball nut does not move smoothly

Carry ball nut produces unusual noise or moves with resistance during operation

Ranked causes

  1. 1. Carry ball nut requires lubrication (unknown)
First check

Observe carry ball nut movement

Test point
Carry ball nut
Expected
Smooth movement without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Carry Ball Nut does not run smoothly

Carry Ball Nut exhibits stiffness, grinding, or unusual noise during operation

Ranked causes

  1. 1. Carry Ball Screw requires lubrication (unknown)
First check

Operate Carry Ball Nut through full range

Test point
Carry Ball Nut assembly
Expected
Smooth motion without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Carry ball nut movement abnormal

Carry ball nut does not run smoothly or produces unusual noise

Ranked causes

  1. 1. Carry ball nut wear or insufficient lubrication (common)
First check

Observe carry ball nut operation for smoothness and noise

Test point
Carry ball nut
Expected
Smooth operation without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Carry ball nut moves with unusual noise

Carry ball nut operation is noisy or not smooth during mechanics test

Ranked causes

  1. 1. Carry ball nut mechanical wear or inadequate lubrication (unknown)
First check

Carry ball nut operation

Test point
Carry ball nut
Expected
Runs smoothly without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Carry Ball Nut moves with unusual noise or stiffness

Carry Ball Nut does not run smoothly or produces unusual noise during operation

Ranked causes

  1. 1. Carry Ball Screw requires lubrication (unknown)
First check

Inspect and listen to Carry Ball Nut movement

Test point
Carry Ball Nut assembly
Expected
Runs smoothly without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Carry Ball Nut noise or binding caution

Carry Ball Nut does not run smoothly or produces unusual noise

Ranked causes

  1. 1. Wear or contamination of ball nut bearing surfaces (occasional)
First check

Test Carry Ball Nut movement

Test point
Carry Ball Nut on Ball Screw
Expected
Runs smoothly without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Carry ball nut operation rough or noisy

Carry ball nut does not run smoothly or produces unusual sound during compression cycles

Ranked causes

  1. 1. Carry ball nut requires lubrication (common)
First check

Manually cycle carry ball nut and listen for smooth operation

Test point
Carry ball nut on ball screw
Expected
Runs smoothly without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Carry Ball Nut runs with noise or stiffness

Carry Ball Nut does not run smoothly or produces unusual noise during device operation

Ranked causes

  1. 1. Carry Ball Nut requires lubrication or maintenance (unknown)
First check

Carry Ball Nut movement inspection

Test point
Carry Ball Nut
Expected
Runs smoothly without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charge mode does not initiate charging sequence high voltage

Device does not begin charging when PSU is connected in charge mode

Ranked causes

  1. 1. PSU connection issue or faulty charging circuit (occasional)
First check

Connect PSU to device in charge mode; observe charging indication

Test point
PSU connector and charging indicator
Expected
Charging sequence initiates; indicator shows charging status

Claw mechanism binding or not moving smoothly

Claw lock does not open/close smoothly or feels stiff during operation

Ranked causes

  1. 1. Insufficient lubrication on claw moving parts (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Claw Mechanism does not remain locked patient safety

Claw Mechanism fails to stay locked when pressure is applied

Ranked causes

  1. 1. Claw Lock Mechanism requires lubrication or adjustment (unknown)
First check

Claw Mechanism lock test

Test point
Claw Mechanism
Expected
Remains locked when pressure is applied

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Claw mechanism does not remain locked under pressure patient safety

Claw mechanism unlocks or moves when pressure is applied during mechanics test

Ranked causes

  1. 1. Claw lock mechanism worn or inadequately lubricated (common)
First check

Claw mechanism lock under pressure

Test point
Claw mechanism
Expected
Remains locked when pressure applied

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Claw mechanism release ring play excessive

Release ring exhibits excessive play when claw mechanism is unlocked

Ranked causes

  1. 1. Release ring wear or claw mechanism wear (unknown)
First check

Inspect release ring for excessive play with claw mechanism unlocked

Test point
Release ring
Expected
No excessive play

Communication PCB replacement caution

Communication PCB on Controller PCB requires replacement; device needs reprogramming after installation.

Ranked causes

  1. 1. Communication PCB failure or degradation (unknown)
First check

Visual inspection of Communication PCB for damage or corrosion

Test point
Communication PCB surface
Expected
No visible damage, corrosion, or component failure

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous compression mode malfunction patient safety

Device does not operate correctly in continuous mode; LED does not confirm mode or does not blink at correct interval

Ranked causes

  1. 1. Software or mode timing issue (occasional)
First check

Activate continuous compression mode; observe LED indicator

Test point
Continuous mode LED
Expected
LED is green during continuous operation

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous mode does not function correctly patient safety

Continuous mode LED not illuminated or LED does not blink at correct interval

Ranked causes

  1. 1. Electronics malfunction or LED failure (unknown)
First check

Activate continuous mode and verify LED illumination

Test point
Continuous mode LED indicator
Expected
LED is green

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Continuous mode does not operate correctly

Continuous mode LED does not illuminate green or does not blink every 10 strokes

Ranked causes

  1. 1. Electronic or software malfunction in continuous mode (unknown)
First check

Verify continuous mode LED illumination

Test point
Continuous mode LED
Expected
LED is green

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous mode LED blink pattern incorrect

LED fails to blink every 10 strokes during continuous mode operation

Ranked causes

  1. 1. Electronics or timing circuit malfunction (unknown)
First check

Continuous mode LED blink frequency

Test point
LED blink pattern
Expected
Blinks every 10 strokes

Continuous mode LED not active or green

Continuous mode LED fails to illuminate green during continuous mode operation

Ranked causes

  1. 1. LED failure or electronics malfunction (unknown)
First check

Continuous mode LED status

Test point
Continuous LED
Expected
Green

Continuous mode LED not green

Continuous mode LED does not display green during continuous mode operation

Ranked causes

  1. 1. LED failure or continuous mode not functioning (unknown)
First check

Continuous mode LED

Test point
LED indicator
Expected
Green

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous mode LED not illuminating or blink rate incorrect

Continuous mode LED fails to illuminate green or does not blink at correct 10-stroke interval

Ranked causes

  1. 1. Electronics or firmware issue affecting Continuous mode operation (unknown)
First check

Run device in Continuous mode and verify LED illuminates green

Test point
Continuous mode LED
Expected
LED illuminates green during continuous compression

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous mode malfunction patient safety

Continuous mode LED not green or LED does not blink every 10 strokes

Ranked causes

  1. 1. Software or electronic issue in continuous compression mode (rare)
First check

Verify continuous mode LED status

Test point
Continuous mode LED
Expected
Green

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Continuous mode operation abnormal

Continuous mode LED not green or LED does not blink every 10 strokes

Ranked causes

  1. 1. Electronics or mode control malfunction (unknown)
First check

Verify continuous LED indicator in continuous mode

Test point
Continuous LED
Expected
LED is green

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Decompression spring movement abnormal

Spring does not move smoothly or produces unusual noise during operation

Ranked causes

  1. 1. Decompression spring wear or misalignment (unknown)
First check

Observe decompression spring movement for smoothness and noise

Test point
Decompression spring
Expected
Smooth movement without unusual noise

Decompression spring moves with noise or resistance

Spring movement is not smooth or produces unusual sound during operation

Ranked causes

  1. 1. Decompression spring requires maintenance or lubrication (common)
First check

Manually operate decompression spring and listen for noise

Test point
Decompression spring assembly
Expected
Moves smoothly without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Decompression Spring moves with noise or stiffness

Decompression Spring does not move smoothly or produces unusual noise during operation

Ranked causes

  1. 1. Decompression Spring requires lubrication or maintenance (unknown)
First check

Decompression Spring movement inspection

Test point
Decompression Spring
Expected
Moves smoothly without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Decompression spring moves with unusual noise

Decompression spring operation is noisy or not smooth during mechanics test

Ranked causes

  1. 1. Decompression spring mechanical wear or misalignment (unknown)
First check

Decompression spring operation

Test point
Decompression spring
Expected
Moves smoothly without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Decompression spring moves with unusual noise or stiffness

Decompression spring does not move smoothly or produces unusual noise during operation

Ranked causes

  1. 1. Decompression spring requires lubrication or has mechanical wear (unknown)
First check

Observe decompression spring movement

Test point
Decompression spring
Expected
Smooth movement without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Decompression Spring stiffness or noise caution

Decompression Spring does not move smoothly or produces unusual noise

Ranked causes

  1. 1. Wear, dirt, or lack of lubrication in spring mechanism (occasional)
First check

Test Decompression Spring movement

Test point
Decompression Spring
Expected
Moves smoothly without unusual noise

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Device does not charge when PSU is connected
  • Device exterior condition is damaged or dirty
  • Device exterior damage or bellows condition issues
  • Device exterior damage or contamination
  • Device exterior damaged or dirty
  • Device exterior damaged or soiled
  • Device produces unusual sound during 30:2 mode
  • Device runs with unusual sound during 30:2 mode
  • Electric fan and mesh grill require cleaning
  • Electric Fan or Mesh Grill contaminated
  • Fabric condition issues with straps or carrying bag
  • Fabric straps or carrying bag damaged
  • Fabrics are damaged or worn
  • Fabrics worn or damaged
  • Hood cannot be opened or removed
  • Internal function test LED does not show green
  • Internal function test LED not green
  • Last service date not recorded on PC
  • Last Service Date not set to current date
  • Last service date not set to PC date
  • Last Service Date not synchronized with PC Date
  • LED does not blink at correct interval in continuous mode
  • Main Body/Frame Assembly replacement
  • Patient straps or carrying bag damage or deterioration
  • Patient straps or stabilization strap degraded
  • PAUSE LED does not illuminate green
  • PAUSE LED not green
  • PAUSE mode LED not green
  • Power Inlet replacement
  • Program Loader Bluetooth initialization
  • Release ring has excessive play
  • Release Ring play detected in Claw Mechanism
  • Release ring play detected when claw unlocked
  • Service date not set correctly
  • Software installation required
  • Software version and checksum not verified
  • Software version mismatch or checksum error
  • Software version requires verification or update
  • Suction Cup does not run smoothly in Adjust mode
  • Suction Cup operation rough in Adjust mode
  • USB connection setup
  • View device error codes and warnings

42 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
ALARM TYPE ROM TESTProtective CPU internal ROM test failure at startupROM self-test fails during power-oncritical
BATT_10_MIN_ALARMBattery voltage drops below 25.3VBattery voltage indicates low charge statewarning
BATT_EMPTY_ALARMBattery voltage drops below 25VBattery depleted during operationcritical
BATT_HIGH_TEMP_ALARMBattery temperature exceeds 70°CBattery temperature sensor detects >70°Ccritical
CHARGER_STOPCharger fault detected with motor disabledCharger system malfunction during operationcritical
COM_B_TIMEOUT (0x55)I2C battery charger communication timeoutNo I2C communication with Charger system for >100 secondscritical
COM_I2C (0x56)I2C communication faults excessiveI2C bus communication errors above thresholdwarning
COM_P_TIMEOUT (0x50)I2C protective system communication timeoutNo I2C communication with Protective system for >100 secondscritical
COM_RTC (0x53)Real-time clock faultRTC read failure at startup or SPI communication faultwarning
COM_SD_MEM (0x54)SD card communication faultSD card not inserted or brokenwarning
CONTROL_LIFETICKI2C communication fault with Control CPU (warning)I2C communication failure at mode changewarning
CONTROL_STOPControl fault detected with motor disabledControl system malfunction during operationcritical
INT_1msCODE (0x3A)1ms code access timeoutHalted or disturbed Control CPU responseerror
INT_C_24V (0x32)Control 24V supply less than 18VMain power supply drops below 18Verror
INT_C_5V (0x31)Control 5V supply out of rangeControl 5V less than 4.5V or greater than 5.5Verror
INT_C_HALL (0x3C)Motor Hall sensor faultHall sensor signal failure detectedwarning
INT_CODE (0x36)Access of vital function code fault (10ms)Control CPU hangs or timing violation detectedcritical
INT_C_POTCAL (0x3D)Control potentiometer linearity faultControl potentiometer response is too non-linearcritical
INT_C_REF (0x33)Control 2.5V reference out of rangeControl 2.5V reference less than 2V or greater than 3Verror
INT_C_TEMP (0x34)Too high internal board temperature (>85°C)Control board temperature exceeds 85°C limitcritical
INT_CURRENT (0x38)Motor current fault exceeds 20AMean current to motor greater than 20Aerror
INTERNAL TEMPERATUREProtective board temperature exceeds 85°CTemperature sensor on Protective board detects >85°Ccritical
INT_P_5V (0x30)Protective 5V supply out of rangeProtective 5V less than 4.5V or greater than 5.5Verror
INT_POTHALL (0x37)Potentiometer position and Hall sensor mismatchMismatch between potentiometer and Hall sensor readings after correct startwarning
INT_P_POTCAL (0x3E)Protective potentiometer linearity faultProtective potentiometer response is too non-linearcritical
INT_P_REBOOT (0x39)Protective CPU reboot in active modeProtective board resets during active operationerror
PISTON_TIME_OUTProtective potentiometer detects no 50% stroke movement for 10 secondsPiston does not move 50% of stroke within 10 second windowcritical
RUN_13N (0x49)13N release spring force high (>10mm)Release force displacement exceeds 10mmwarning
RUN_ADJUST (0x46)Incorrect motor power at adjust modeMotor power greater than 25W during adjust servo operationwarning
RUN_NO_POTCAL (0x4B)Potentiometer not calibratedSystem detects potentiometer has not been calibratedcritical
RUN_POTCAL (0x4A)Potentiometer calibration faultPotentiometer not moved during calibration or read error detectedcritical
RUN_PROTECTIVE (0x45)Halt or reverse by protective system during active moving mode (>1s)Protective system disables active mode during operationcritical
RUN_P_STOP (0x4C)Protective system has stopped operationProtective system error detectedcritical
RUN_RATIO (0x44)Piston ratio fault (50 consecutive strokes)Compression ratio deviates from preset ratio +/- 20% over 50 strokescritical
RUN_TIMEDOWN (0x48)Compression stroke time too long or too shallow50 consecutive strokes with timing or depth out of rangecritical
RUN_TIMEOUT (0x40)Piston not moved during active moving mode (>1s)No motor movement detected during >1 second of active modecritical
RUN_TIMEUP (0x47)Time up too long >0.15s (50 consecutive strokes)Release phase exceeds 0.15 seconds during 50 strokescritical
RUN_TOO_DEEP (0x41)Piston too deep (compression depth excessive)Piston depth exceeds normal range during operationcritical
RUN_TOO_SHALLOW (0x42)Piston too shallow (50 consecutive strokes @ 100 bpm)50 consecutive strokes measure <target depth minus 10mm at 100bpmcritical
T1_BATT_COM_ALARMBattery communication faultBattery communication error detected at startupwarning
T1_BATT_COPYRIGHT_ALARMIncorrect copyright string in batteryBattery copyright verification fails at startupwarning
T1_BATT_CS_ALARMBattery communication checksum faultBattery communication checksum validation failurewarning
T1_B_OFF (0x2A)Battery charger did not respond to startup testCharger CPU does not complete startup testwarning
T1_CAL_EE (0x15)Control potentiometer calibration data checksum faultCalibration data corruption detectedcritical
T1_C_DOWN (0x2D)Control not moving down before reverse test at startupMotor does not move down during startup reverse test sequencewarning
T1_C_EE (0x10)Control EE-Prom checksum data errorCPU fault detected during initializationcritical
T1_C_INITCURR (0x20)High current detected at startupHigh current read during initial startupwarning
T1_C_LOWCURR (0x24)Too low current detected at first current testMotor current below expected value during startup testwarning
T1_C_POT (0x28)POT value incorrect or mismatch at start or no change during movingControl/protective POT divergence >10mm at start or non-linear responsecritical
T1_C_POT (0x28)Potentiometer value mismatch at warm startControl and Protective potentiometer divergence >10mm at warm startwarning
T1_C_RAM (0x11)Control preset checksum data errorCPU or communication fault during startupcritical
T1_C_RELEASE (0x22)Current detected at release/disable test by controlUnexpected current during release phasecritical
T1_C_TOTOP (0x2F)Control timeout at move to top at startupMotor cannot move to top position within timeoutwarning
T1_E2_PROMProtective CPU internal E2PROM test failure at startupEEPROM self-test fails during power-oncritical
T1_E2_PROM_ALARMCharger CPU internal E2PROM test failure at startupEEPROM self-test fails during Charger power-oncritical
T1_PB_FLASH (0x14)Incompatible programs in protective boardProgram mismatch detected after board changecritical
T1_P_DISABLE (0x25)Protective disable failure at current test startupProtective system fails to disable at expected pointwarning
T1 PISTON RELEASE ERRORProtective release line operation fault at startupCurrent detected during disable test or moved too low during enable testcritical
T1 PISTON REVERSE ERRORReverse test failure at startupMotor reverse movement incorrect or fault current read at startupcritical
T1_POWER_OFF_ALARMCharger startup test incompleteT1 test does not end correctlycritical
T1_P_POT (0x27)Read Protective potentiometer timeout at startupProtective potentiometer read operation times outwarning
T1_P_REW (0x2E)Protective timeout at reverse test at startupMotor reverse test does not complete within timeoutwarning
T1_P_STATE_1 (0x21)Start of release test timeout from protective boardProtective board fails to complete release testcritical
T1_P_STATE_3 (0x23)Start of current test timeout or current detected by protectiveProtective fails current test or detects unexpected currentcritical
T1_P_STATE_5 (0x29)No end of Protective current test at startupProtective current test does not complete within timeoutwarning
T1_RAMProtective CPU internal RAM test failure at startupRAM self-test fails during power-oncritical
T1_RAM_ALARMCharger CPU internal RAM test failure at startupRAM self-test fails during Charger power-oncritical
T1_ROM_ALARMCharger CPU internal ROM test failure at startupROM self-test fails during Charger power-oncritical
T1_START_COM (0x13)No communication from Protective or Battery-Charger CPUs at startupI2C communication failure at device startupwarning
TOO_DEEP_COMPRESSIONCompression depth exceeds safe limitsExcessive compression depth detected during active modecritical
TOO_MANY_USESBattery cycle count exceeds 200 usesBattery use counter reaches thresholdwarning
TOO_OLD_BATTERYBattery age exceeds recommended service lifeBattery exceeds maximum age specificationwarning
TOO_SHALLOW_COMPRESSIONCompression depth below minimum after 50 strokesInsufficient compression depth for 50 consecutive strokescritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI