Physio-Control LIFEPAK CR+: troubleshooting and error codes
The ServiceForge knowledge base holds 11 documented symptoms for the Physio-Control LIFEPAK CR+, with 18 ranked causes, 37 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Battery charger malfunction indicated patient safety
BATTERY indicator displayed on device; diagnostic tone codes provided when charger reinserted
Ranked causes
- 1. Internal charger error (occasional)
- 2. Patient load previously connected to charger (occasional)
- 3. Cumulative ON time exceeds 30 minutes (common)
- 4. Battery charger expiration date passed (common)
- 5. Battery charger incorrectly homed (occasional)
Turn device OFF and wait at least 60 seconds
- Test point
- Device power state
- Expected
- Device powered down for minimum 60 seconds
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Install replacement kit and test cables
Installation of Medtronic replacement kit and FAST-PATCH cable kit required before PIP
Ranked causes
- 1. Medtronic replacement kit MIN 3201616-000 required (common)
- 2. Medtronic FAST-PATCH cable kit required (common)
- 3. Electrode Test Post Adapter (2 each) required (common)
Verify replacement kit presence
- Test point
- Equipment inventory
- Expected
- Medtronic MIN 3201616-000
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Final device status verification patient safety
Confirm overall device status after component replacement and verify readiness for active service
Ranked causes
- 1. Internal battery partially depleted (ATTENTION indicator visible) (common)
- 2. WRENCH symbol displayed (service required) (occasional)
Turn device on and wait for self-test to complete
- Test point
- Device power button; display
- Expected
- Self-test completes without errors
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Decade resistance box setup
Performance Inspection Procedure requires decade resistance box with 0-500 Ω range and 1% accuracy
Ranked causes
- 1. IET Model RS-200 decade resistance box required (common)
Verify resistance box range
- Test point
- Decade box specification
- Expected
- 0-500 Ω
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Defibrillator analyzer setup
Performance Inspection Procedure requires defibrillator analyzer with specific power range, load resistance, accuracy and waveform support
Ranked causes
- 1. Fluke QED-6 analyzer required for PIP (common)
Verify defibrillator analyzer power range
- Test point
- Analyzer specification
- Expected
- 0-450 J
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Defibrillator energy output high voltage
Verify defibrillator-delivered energy levels at 100 J and 360 J meet acceptable output ranges using analyzer measurement
Ranked causes
- 1. Energy output requires calibration verification (common)
Unplug all components connected to defibrillator patient connector
- Test point
- Patient connector
- Expected
- All components removed
12 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Install PIP software on laptop
Laptop computer with Medtronic PIP and TCP software required to perform Performance Inspection Procedure
Ranked causes
- 1. Medtronic PIP & TCP software required on laptop (common)
Verify laptop has Medtronic PIP & TCP software installed
- Test point
- Laptop software inventory
- Expected
- Medtronic PIP & TCP software MIN 3202957
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Safety analyzer setup high voltage
Performance Inspection Procedure requires safety analyzer with line voltage input and microampere current measurement capability
Ranked causes
- 1. Dale Model 600/601 or 600E/601E safety analyzer required (common)
Verify safety analyzer line voltage capability
- Test point
- Analyzer specification
- Expected
- 110 or 220 Vac
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device initialization displays only ATTENTION symbol
Only ATTENTION symbol visible in Readiness Display during initialization test
Ranked causes
- 1. Internal device battery recharging from CHARGE-PAK (common)
Observe Readiness Display and wait for ATTENTION symbol to clear
- Test point
- Readiness Display
- Expected
- ATTENTION symbol extinguishes; OK symbol appears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device initialization displays WRENCH symbol patient safety
WRENCH symbol illuminated in Readiness Display during initialization test
Ranked causes
- 1. Device requires service due to internal fault (unknown)
Observe Readiness Display during initialization
- Test point
- Readiness Display
- Expected
- OK symbol should be visible; not WRENCH
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unknown battery charger tone code (8-10 tones) patient safety
Battery charger produces 8 to 10 beep tones indicating undefined condition
Ranked causes
- 1. Undefined charger fault condition (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI