Physio-Control LIFEPAK 500: troubleshooting and error codes
The ServiceForge knowledge base holds 72 documented symptoms for the Physio-Control LIFEPAK 500, with 93 ranked causes, 107 diagnostic checks with test points and expected values, and 126 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit displays damage or wear patient safety
Visible defects on external surfaces preventing safe operation
Ranked causes
- 1. Damage to exterior surfaces (unknown)
- 2. Excessive wear (unknown)
- 3. Improper mechanical function (unknown)
- 4. Damaged connectors (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on patient safety
Device fails to charge or respond to operation
Ranked causes
- 1. Defective test set or setup issue (common)
- 2. Patient interface connector damaged (occasional)
- 3. Energy storage capacitor issue (occasional)
- 4. Main PCW defective (occasional)
Verify test set-up and reset
- Test point
- test setup
- Expected
- test setup correct
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fault codes recorded in LIFEPAK 500
LIFEPAK 500 AED has recorded one or more fault codes during internal self-test or routine use. Fault codes are stored in the Fault Code Log and may indicate transient electromagnetic interference or electrostatic discharge. Codes may not indicate permanent equipment failure.
Ranked causes
- 1. Transient electromagnetic interference (EMI) (occasional)
- 2. Electrostatic discharge (ESD) (occasional)
- 3. Permanent hardware error (rare)
Review recorded fault code(s) in Fault Code Log
- Test point
- Fault Code Log memory
- Expected
- Fault code present with date and time of occurrence
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: AED impedance patient safety
Verify AED output impedance matches specification using decade resistance box and test adapter
Ranked causes
- 1. Impedance calibration required per service procedure (common)
- 2. Impedance not within specification (unknown)
Verify AED impedance output at test post
- Test point
- QUIK-COMBO test post adapter
- Expected
- 225 OHMS Ω
- Tools
- decade resistance box set to 225 Ω, VP78-QUIK-COMBO test post adapter
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Energy meter baseline and component connection high voltage
Connect all components to LIFEPAK 500 patient connector and establish energy meter baseline
Ranked causes
- 1. Components must be connected to patient connector for testing (common)
- 2. Energy meter baseline must be established using LIFEPAK energy meter (common)
Unplug all components from LIFEPAK 500 patient connector
- Test point
- LIFEPAK 500 patient connector
- Expected
- all components disconnected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LIFEPAK 500 boot routine after battery replacement patient safety
Restore normal operation after battery has been replaced in the unit
Ranked causes
- 1. Battery replacement requires boot routine execution (common)
- 2. Verify battery status message (common)
Insert fully functional battery
- Test point
- Battery compartment
- Expected
- Battery inserted, does not return battery fault indications
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LIFEPAK 500 energy output patient safety
Verify defibrillator delivers specified energy level when PUSH TO SHOCK button is activated during test mode
Ranked causes
- 1. Energy calibration out of specification (unknown)
- 2. Calibration procedure execution (unknown)
Access energy calibration screen
- Test point
- LIFEPAK 500 user interface
- Expected
- CALIBRATE ENERGY screen displayed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot set time or date
Unit will not accept or retain time and date settings
Ranked causes
- 1. Time/Date clock battery dead or missing (common)
- 2. Key switch installation improper (occasional)
Verify Time/Date clock battery
- Test point
- Battery compartment
- Expected
- Battery present and functional
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charge time exceeds specification patient safety
AED takes longer than 15 seconds to charge to 360 J
Ranked causes
- 1. Battery voltage low or charger circuit slow (common)
- 2. High-voltage charging circuit degradation (occasional)
Measure charge time from power button to delivery
- Test point
- AED output; timer
- Expected
- 360 J in 15 seconds or less, ≤15 s
- Tools
- Stopwatch or analyzer with timing function
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defib output insufficient or weak patient safety
Defibrillator produces reduced or no therapeutic output
Ranked causes
- 1. Battery low or failing (unknown)
- 2. Electrode pads defective or expired (unknown)
Electrode impedance test fails patient safety
AED does not respond correctly to electrode impedance tests or readings are incorrect
Ranked causes
- 1. Electrode cable or connector defect (common)
- 2. Impedance measurement circuit failure (occasional)
Visually inspect electrode cables and connectors
- Test point
- Electrode connectors on AED
- Expected
- No visible damage, corrosion, or loose connections
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrode pads not detected patient safety
Electrodes fail to register or are not recognized
Ranked causes
- 1. Electrode pads not seated properly (unknown)
- 2. Low battery or weak battery (unknown)
Energy output and charge time deficient patient safety
Defibrillator will not charge or charges slowly
Ranked causes
- 1. Faulty functional battery (common)
- 2. Energy storage capacitor defective (occasional)
Verify battery is fully charged
- Test point
- battery status
- Expected
- fully functional battery
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Energy output incorrect patient safety
AED does not deliver specified energy levels to patient
Ranked causes
- 1. Capacitor charge or discharge circuit failure (common)
- 2. Energy calibration drift (occasional)
Measure energy output at 360 J setting
- Test point
- Defibrillator analyzer input
- Expected
- 342–378 J for monophasic; 342–378 J for biphasic shock 2, Monophasic 360 J ±5%; Biphasic 100 J ±5%, 360 J ±5%
- Tools
- Defibrillator analyzer
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Firmware or system error
Device firmware or internal system malfunction
Ranked causes
- 1. Firmware corruption or error (rare)
- 2. Internal voltage reference error (rare)
Run system diagnostics
- Test point
- Device self-test
- Expected
- All systems pass
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lead-to-lead leakage current excessive high voltage
Leakage current between electrode leads exceeds safe limit
Ranked causes
- 1. Electrode cable insulation breakdown or contamination (occasional)
- 2. High-voltage output circuit isolation fault (rare)
Measure lead-to-lead leakage current
- Test point
- Electrode output connectors
- Expected
- does not exceed 9 µA, ≤9 µA
- Tools
- Leakage current meter or digital multimeter on µA range
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AED fails to connect to analyzer
QUIK-COMBO cable not properly seated or defective
Ranked causes
- 1. QUIK-COMBO test post adapter not connected (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analyze LIFEPAK menu access
Menu navigation to adjust or verify defibrillator calibration settings
Ranked causes
- 1. Menu system access needed for parameter adjustment (unknown)
Confirm ANALYZE MENU key accessibility
- Test point
- LIFEPAK 500 keypad
- Expected
- Menu system responds to key press
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto battery test disabled patient safety
Automatic battery self-test mode is not enabled; unit cannot perform ongoing battery monitoring
Ranked causes
- 1. Auto battery test mode set to OFF (common)
Verify auto battery test mode setting
- Test point
- Service Mode menu
- Expected
- AUTO BATTERY TEST mode is ON
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Automatic advisory operation failed patient safety
AED does not complete automatic advisory operation diagnostic tests; CPR coaching or shock advisory may be non-functional
Ranked causes
- 1. Advisory algorithm or analysis circuit failure (unknown)
Verify automatic advisory operation test completion
- Test point
- AED analysis and advice module
- Expected
- AED completes all automatic advisory operation tests
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Automatic advisory operation fails patient safety
AED does not complete automatic advisory operation tests correctly
Ranked causes
- 1. Software or processing circuit fault (occasional)
Run automatic advisory operation diagnostic
- Test point
- Service Mode diagnostic menu
- Expected
- AED completes all automatic advisory operation tests without error
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Automatic test results fail patient safety
One or more automatic tests display failure instead of PASS in the Automatic Test Results window
Ranked causes
- 1. Test execution error or equipment malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery auto-test disabled patient safety
AUTO BATTERY TEST mode is not enabled
Ranked causes
- 1. Auto battery test setting disabled in software (occasional)
Verify AUTO BATTERY TEST mode setting
- Test point
- Service Mode menu
- Expected
- AUTO BATTERY TEST mode is set to ON
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery cartridge leaking or defective caution
Battery shows signs of leakage or does not function
Ranked causes
- 1. Battery cartridge defective (unknown)
Battery charger voltage out of specification
Battery charger output voltage is outside 9.78–10.00 V range; charger or power supply defective
Ranked causes
- 1. Battery charger circuit degradation or external power supply failure (occasional)
Measure battery charger output voltage
- Test point
- Charger output connector
- Expected
- 9.78 to 10.00 volts, 9.78–10.00 V
- Tools
- Digital multimeter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery connection fault
Battery not detected or connection issue
Ranked causes
- 1. Battery connector disconnected or loose (common)
Check battery connector seated properly on Main PCW
- Test point
- Battery connector
- Expected
- Connector fully inserted and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery error patient safety
Battery-related fault or alarm condition
Ranked causes
- 1. Low battery voltage (common)
Check battery voltage
- Test point
- Battery terminals
- Expected
- Nominal voltage for battery type
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery polarity incorrect caution
Battery installed with reversed polarity
Ranked causes
- 1. Battery installed backward (occasional)
Verify battery polarity matches markings
- Test point
- Battery compartment
- Expected
- Positive and negative terminals aligned correctly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage low or absent
Battery voltage reading below expected level
Ranked causes
- 1. Battery depleted or defective (common)
Measure battery voltage
- Test point
- Battery terminals
- Expected
- Within specification per battery type
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cables, connectors, or electrodes physically damaged patient safety
Visual or physical inspection reveals cracks, corrosion, loose connections, frayed wires, or other damage to cables, connectors, or electrode pads
Ranked causes
- 1. Mechanical wear, environmental exposure, or manufacturing defect (common)
Visual and tactile inspection of cables, connectors, electrodes
- Test point
- Entire cable assembly and electrode pads
- Expected
- No visible cracks, corrosion, fraying, or loose connections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery resistance check
Measure resistance of negative battery terminal to verify battery condition and connector integrity
Ranked causes
- 1. Excessive battery or connector resistance (unknown)
Measure resistance from negative battery terminal to Sternum Output Pin 1
- Test point
- negative battery terminal to Pin 1 (patient connector)
- Expected
- less than 8 Ω, < 8 Ω
- Tools
- DMM
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Defibrillation output energy measurement high voltage
Measure defibrillation output energy using LIFEPAK energy meter connected to analyzer
Ranked causes
- 1. AED must be in Clinical mode for defibrillation measurement (common)
Place AED in Clinical mode
- Test point
- AED control panel
- Expected
- AED in Clinical mode
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Energy output patient safety
Verification that the defibrillator energy output meets specified performance within acceptable tolerance
Ranked causes
- 1. Energy calibration required (unknown)
Set defibrillation analysis mode to ENERGY
- Test point
- ENERGY selection
- Expected
- @??? t 1range
6 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ESD test verification caution
Electrostatic discharge test procedure for LIFEPAK 500 using ESD10 device per IEC 61000-4-2 standard
Ranked causes
- 1. ESD immunity verification required (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LCD contrast adjustment
Adjustment procedure to set LCD contrast on LIFEPAK 500 AED to optimal display level
Ranked causes
- 1. LCD contrast requires calibration (unknown)
Access ADJUST LCD CONTRAST screen
- Test point
- Service menu
- Expected
- ADJUST LCD CONTRAST screen displays
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: LIFEPAK 500 boot test patient safety
Procedure to verify LIFEPAK 500 boots correctly into CONNECT ELECTRODES state after completing TimeOut/Max Battery Recovery procedure
Ranked causes
- 1. TimeOut/Max Battery Recovery procedure completion (common)
Insert insulated shunter cable between TR+XX and TR-abb as shown at the left
- Test point
- TR+XX and TR-abb
- Expected
- Insulated shunter cable connected
5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: LIFEPAK 500 defibrillator analysis function patient safety
Verify ANALYZE button and menu functions for proper operation and device response
Ranked causes
- 1. Device ANALYZE function requires calibration verification (unknown)
Power on device in HEART mode
- Test point
- Power switch
- Expected
- Device boots to HEART mode
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Oscilloscope defibrillation output high voltage
Set oscilloscope to measure defibrillation energy output with correct settings and energy meter baseline
Ranked causes
- 1. Oscilloscope requires proper configuration for energy measurement (common)
Connect oscilloscope to defibrillation analyzer scope output
- Test point
- scope output connection
- Expected
- oscilloscope connected
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance inspection patient safety
Computer-assisted automated tests for LIFEPAK 500 defibrillator including serial communications, battery sense, slow charge rate, 9 V monitor, motion detection, and PPE self-test functions
Ranked causes
- 1. Calibration procedure execution (common)
Serial port communication
- Test point
- COM1 or COM2
- Expected
- Unit under test connected to selected serial port
5 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance Inspection Procedure patient safety
Verification of LIFEPAK 500 defibrillator performance by measuring defibrillator impedance readings against specified tolerances using a decade resistance box and VPK-100 test adaptor.
Ranked causes
- 1. Routine calibration and performance verification (common)
Press ANALYZE button to initiate impedance measurement
- Test point
- ANALYZE button
- Expected
- Initiate Part A impedance test
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance Inspection Procedure - Fail condition patient safety
Action required when impedance measurements are outside specified tolerances.
Ranked causes
- 1. Impedance measurement outside tolerance (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: remaining energy levels patient safety
Verify remaining energy levels on display are within acceptable output ranges for monophasic and biphasic LIFEPAK 500 defibrillators
Ranked causes
- 1. Energy level output verification required (unknown)
Verify monophasic AED remaining energy output
- Test point
- LIFEPAK 500 Monophasic AED display
- Expected
- 200 ± and 360 ±
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Computer assisted tests fail caution
AED fails computer-assisted diagnostic tests in Service Mode
Ranked causes
- 1. Internal fault in tested circuit or software error (occasional)
Execute computer-assisted test suite in Service Mode
- Test point
- Service Mode test menu
- Expected
- All computer-assisted tests pass without errors
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Computer-assisted tests failed
AED does not pass computer-assisted diagnostic tests; internal system fault detected
Ranked causes
- 1. Microprocessor, memory, or internal system fault (unknown)
Run computer-assisted diagnostic test suite
- Test point
- AED microprocessor and internal circuits
- Expected
- All computer-assisted tests pass
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data or memory issue
Device memory or stored data error
Ranked causes
- 1. Data corruption in memory (rare)
Test memory integrity
- Test point
- Device memory
- Expected
- No error codes or CRC failures
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Data save not completed patient safety
Device powers up but data save operation did not complete
Ranked causes
- 1. Data save operation failed (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Date or time incorrect
AED internal clock displays wrong date or time after self-test
Ranked causes
- 1. Battery-backed real-time clock failure or user error during setup (occasional)
Verify time and date display
- Test point
- AED display after self-test
- Expected
- Current date and time displayed correctly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device selected to wrong mode patient safety
Unit operating in incorrect mode or configuration
Ranked causes
- 1. Mode switch set incorrectly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Electrode impedance test failure
AED does not respond properly to electrode impedance measurement tests; may indicate electrode or sensing circuit defect
Ranked causes
- 1. Defective electrodes or impedance sensing circuit (unknown)
Verify electrode impedance test response
- Test point
- Electrode connectors
- Expected
- AED properly responds to impedance tests
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Energy output out of specification patient safety
Actual delivered energy does not match nominal specification or tolerance range for any shock energy level
Ranked causes
- 1. High-voltage supply, charging circuit, or capacitor degradation (unknown)
Measure shock energy at each level
- Test point
- Defibrillator analyzer input
- Expected
- Monophasic: 200 J (190–210), 300 J (285–315), 360 J (342–378); Biphasic: 100 J (95–105), 360 J (342–378)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error message or alarm code displayed patient safety
Device displays error or alarm notification to operator
Ranked causes
- 1. Internal device error or alarm condition (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fault code does not clear after restart patient safety
Fault code persists in Fault Code Log after shutdown and restart of the AED, indicating possible permanent hardware error or recurring transient interference.
Ranked causes
- 1. Permanent hardware failure (rare)
Verify fault code persists after two restart cycles
- Test point
- Fault Code Log
- Expected
- Same fault code recorded after restart; does not clear
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improved power on sequence
Device powers on in unexpected sequence or manner
Ranked causes
- 1. Dime-of-Way clock battery defect (rare)
Verify Main PCW installation and operation
- Test point
- Main PCW
- Expected
- PCW installed and functioning properly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect time or date display caution
AED displays wrong time and date values
Ranked causes
- 1. Real-time clock battery depleted or circuit failure (occasional)
Verify displayed time and date are correct
- Test point
- AED display screen
- Expected
- Time and date match current values
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard does not respond caution
AED fails to enter Service Mode when keyboard is tested
Ranked causes
- 1. Keyboard membrane or circuit failure (occasional)
Attempt keyboard test in Service Mode
- Test point
- Keyboard input
- Expected
- AED successfully enters Service Mode
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad test failed
Keypad does not respond to input or shows error during test
Ranked causes
- 1. Key switch caps loose or damaged (occasional)
Verify keypad overlay is properly installed
- Test point
- Top cover keypad area
- Expected
- Cover properly installed and secure
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leakage current exceeds limit high voltage
Lead-to-lead leakage current exceeds 9 µA; electrical safety hazard to patient or operator
Ranked causes
- 1. Insulation breakdown or contamination on PCB or connectors (rare)
Measure lead-to-lead leakage current
- Test point
- Electrode connectors or internal test points
- Expected
- Leakage current ≤9 µA, ≤9 µA
- Tools
- Medical-grade leakage current meter
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main CPU error
Main CPU fault code detected during self-test
Ranked causes
- 1. Defective Main CPU (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pad or electrode issue patient safety
Problem with defibrillator pads or electrodes
Ranked causes
- 1. Pads expired or defective (common)
Verify pad expiration date
- Test point
- Pad package
- Expected
- Current date before expiration
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance Inspection patient safety
Verify LIFEPAK 500 AED functions correctly through systematic checks of power, indicators, and button operation
Ranked causes
- 1. Unit requires periodic performance verification (common)
Power on the LIFEPAK 500 AED
- Test point
- ON/OFF pad
- Expected
- Unit powers on
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Power-on sequence incomplete
- Review fault codes on display
- Serial number mismatch
- Service Mode entry failed
- Shock counter display review
- Test equipment verification
- Unit displays ANALYZING NOW—STAND CLEAR
- Unit displays ANALYZING NOW—STAND CLEAR—NO SHOCK ADVISED—CHECK FOR PULSE
- Unit displays CONNECT ELECTRODES
- Unit displays fault code
- Unit marked as out of use
- Yellow operating system light on
12 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
**?? | CPU data save volume error | CPU unable to access data volume | error |
**?` | CPU measured clock input out of range | Clock frequency outside specification | error |
**?@ | CPU SV voltage check failed | Voltage measurement out of range during test | error |
**?> | CPU clock usage flag not organized | Clock memory structure corrupted | error |
**?^ | CPU symbol stored flag high (repeated) | Symbol storage flag remains set | warning |
**?] | CPU symbol two stored flag high (repeated) | Secondary symbol flag remains set | warning |
**?a | CPU symbol two stored flag high | Secondary symbol flag set abnormally | warning |
**?b | CPU symbol stored flag high | Symbol storage flag set incorrectly | warning |
Battery Low Battery | Battery voltage is low but unit may still operate | Battery voltage drops below normal operating level | warning |
Battery Out of Service | Battery is no longer usable or not installed | Battery voltage below threshold or expired battery detected | error |
CCC 1 | Corrective action code indicating first fault route | Primary fault detection | unknown |
CCC 2 | Corrective action code indicating second fault route | Primary fault persists after first remedy | unknown |
D>?@ | Data save memory read error | Device unable to read from data save memory | error |
D>?? | Data save mount volume error | Device unable to access data save location | error |
D>?` | RAM or long-term battery discrepancy | Mismatch between RAM and backup battery status | error |
**?d | CPU measured alert input out of range | Alert signal voltage abnormal | error |
Data CRC Data Corrupted Data | Data integrity check failed; memory corruption detected | Device detects checksum mismatch in stored data | error |
DDMMMYY HH:MM | Fault code with date and time stamp | When fault codes are present in device memory | unknown |
E-. | Data memory test zero battery malfunction | Battery test failure | error |
E-$ | Data memory test zero zero battery malfunction | Battery test failure | error |
E-^?? | Control circuit power mismatch | Control board voltage regulation out of specification | error |
E-0@. | System waveform shape error | Output waveform deviates from specified shape | error |
E-0@] | Device low power, system status condition | Battery or power supply below operating threshold | warning |
E-0@$ | Device queue buffer overrun | Data buffer overflow condition | warning |
E-0@^ | System low amplitude error | Output signal amplitude below specification | error |
E-0@] | System telemetry function error | Data transmission or telemetry subsystem failure | error |
E-01 | Loose or incorrect pad application | Pad sensor detects abnormal impedance or contact | error |
E-01 | Connector, battery contacts defective | Battery contact or connector failure detected | error |
E-02 | Battery signal data corrupted | Battery data reading error | error |
E-02 | Loose or incorrect connector | Electrode connector not fully seated or damaged | error |
E-03 | Loose skin electrode, mode | Electrode contact lost during operation | warning |
E-04 | Battery rigging to internal power data | Battery rigging configuration error | error |
E-04 | Loose skin electrode, missing AED connector | Electrode signal intermittent or absent | error |
E-05 | Loose skeletal, to-I, mode | Skeletal muscle artifact or motion detected | warning |
E-06 | Loose heart, mode, off, more, or on | Electrode or lead contact problem during operation | error |
E-07 | Loose heart, mode, all 1 st, on or off | Intermittent electrode signal during defibrillation | error |
E-08 | Loose heart, receive | Electrode contact lost during analysis phase | error |
E-09 | Loose E15, do-I, receive, mode | E15 sensor or lead contact intermittent | warning |
E-0@a | System address error | Memory or communication address fault detected | error |
E-0@b | Device high power, system status condition | Battery or power supply above normal operating threshold | warning |
E-0@E | System configuration power mismatch | Configuration does not match actual power supply | error |
E-1 | Unit will not load or power on | Device fails to power up | critical |
E-1 | AED not ready—test abortion detected | Self-test failure | critical |
E-10 | Loose R15#, mode registered on subsystem | R15# sensor reporting abnormal state | warning |
E-10 | Control invalid (I/O E15, I14, HE) | Control processing error detected | error |
E-10 | Communication test defibrillator missing success | Unit fails communication diagnostic test | error |
E101 | E15 chip malfunction | Chip hardware failure | critical |
E102 | E15 chip not ready | Chip initialization failure | critical |
E103 | E15 chip sample timeout not stored | Timeout during sampling | error |
E104 | E15 chip queue full error | Queue overflow | error |
E105 | E15 chip shutdown data corrupted | Data corruption on shutdown | critical |
E106 | Control circuit current short detected anchor | Short circuit detected | critical |
E107 | Control circuit current short leads condition | Short circuit on leads | error |
E108 | Control circuit current short energy status | Energy circuit short | critical |
E-11 | Loose display, reset, mode | Display firmware or state error | warning |
E-11 | Startup ROM test failed, ROM not masked | ROM integrity check failure during startup | critical |
E-11 | Communication Y defibrillator missing success | Y-channel communication failure | error |
E-12 | Communication test X defibrillator missing success | X-channel communication failure | error |
E-12 | Control invalid, startup data error, data category mismatch | Startup configuration data error | error |
E-12 | Loose axon motor, bank, subsystem, mode | Motor control or feedback error | error |
E-13 | Anomaly in ROM test expired | ROM anomaly detection during test sequence | error |
E-13 | Communication Y X defibrillator missing success | Dual-channel communication failure | error |
E-13 | Loose extension, to-I, bank, subsystem, mode | Extension module connection or control error | warning |
E-14 | Control invalid, EEPROM abort object, ROM test expired | EEPROM or ROM integrity failure | critical |
E-14 | Loose bus, bank, subsystem, mode | Internal bus communication error | error |
E-14 | Communication test band error | Band frequency communication error detected | error |
E-15 | Communication Y band error | Y-band communication frequency error | error |
E-15 | Startup ROM anomaly offset, EEPROM not masked | ROM offset or EEPROM integrity issue during startup | error |
E-15 | Loose monitor, E15, bank, subsystem, mode | Monitor module or E15 interface malfunction | error |
E-16 | Loose electro, bank, subsystem, mode | Electrocardiogram processing error | error |
E-16 | Mohawk band HIS band error at test complex malfunction | HIS band frequency or modulation error | error |
E-16 | Startup sensor support, EEPROM not masked | Sensor configuration or EEPROM error at startup | error |
E-17 | Mohawk HIS complex at Y complex malfunction | HIS-Y complex signal integrity failure | error |
E-17 | Loose control, and, bank, subsystem, mode | Control system error or interface failure | error |
E-17 | Startup output signal, data flowfield, ROM not masked | Output signal or data flow error at startup | error |
E-18 | Loose non, bank, subsystem, mode | Non-standard or undefined subsystem error | warning |
E-18 | Startup clock check, ROM not masked | Clock circuit or ROM integrity failure | error |
E-18 | Check Fuses | Fuse integrity diagnostic failure | warning |
E-19 | Startup CPU, ROM not masked | CPU initialization or ROM failure | critical |
E-19 | Check the Battery | Battery diagnostic test failed | warning |
E-1b?? | Analog low voltage malfunction | Analog circuit supply voltage below minimum threshold | critical |
E-1b?@ | Non-rank waveform absent | Expected waveform signal not detected on reference channel | warning |
E-1d?@ | EEPROM memory not initialized | System detects uninitialized EEPROM on startup | error |
E-1d?` | EEPROM code download failed | Firmware upload to EEPROM does not complete successfully | error |
E-1d?? | EEPROM communication checksum failure | Data read from EEPROM fails checksum validation | error |
E-1d?d | EEPROM communications bad packet detected | Received data packet fails integrity check | error |
E-1d?Z | EEPROM no data received from timeout | EEPROM communication times out waiting for data | error |
E-2 | Unit will not charge or battery low | Battery charging failure detected | error |
E-20 | Startup EEPROM, ROM not masked | EEPROM access or ROM integrity failure | critical |
E-20 | Replace battery Time Mask Clock Battery | Clock or backup battery depleted | warning |
E-21 | Startup input abort, ROM not masked | Input circuit or ROM failure at startup | error |
E-22 | Startup input anomaly, ROM not masked | Input anomaly detection during startup | error |
E-23 | Startup output abort, ROM not masked | Output circuit or ROM failure at startup | error |
E-24 | Startup input abort, EEPROM, ROM not masked | Input circuit, EEPROM, or ROM failure | critical |
E-25 | Startup ROM anomaly, ROM not masked | ROM anomaly detected at startup | error |
E-3 | Battery is damaged or will not hold charge | Battery fails to maintain charge | error |
E-3 | AED not ready—dump memory error | Memory dump failed | critical |
E-4 | AED not ready—data memory error | Memory corruption | critical |
E-4 | User pads, electrodes, energy output unable to discharge | Defibrillation output failure | critical |
E-5 | AED not ready—memory directory system error | Directory system failure | critical |
E-5 | User pads, therapy button will not discharge | Therapy button fails to trigger discharge | error |
E-6 | AED not ready—memory data mismatch error | Data integrity check failed | critical |
E-6 | User pads, data will not discharge | Data transmission failure during discharge | warning |
E-7 | AED not ready—memory directory system error | Directory malfunction | critical |
E-7 | User pads, therapy button will not discharge will not discharge | Repeated therapy button discharge failure | error |
E-8 | User pads, capacitor malfunction | Capacitor charging or holding failure | error |
E-9 | Decomfibrillator ECG incomplete at ECG complex malfunction | ECG signal acquisition or analysis failure | error |
E-^?d | Control board signal update timeout | Control signal update cycle exceeds timeout threshold | warning |
EE-01 | Self-test failed too many times | Multiple consecutive self-test failures | critical |
EE-02 | Self-test abort not insured | Abort safety mechanism failure during self-test | critical |
EE-03 | Self-test runtime asserts | Runtime assertion detected during self-test | error |
EE-04 | Non-volatile dropped lead dead message | Lead-off signal loss during operation | error |
EE-05 | Non-volatile dropped low battery message | Battery voltage drops unexpectedly | error |
EE-06 | Non-volatile battery not continual | Battery continuity check failure | error |
EE-07 | Incorrect file of waveform data type | Waveform data format error detected | warning |
EE-08 | Configuration of battery used out of order function message | Battery configuration mismatch detected | error |
EE-09 | Non-verbose dropped lead dead message | Lead-off detection during normal operation | warning |
EE-10 | Control voltage jumps startup switch might | Control voltage fluctuation at startup | error |
EE-11 | Control battery high voltage | Battery voltage exceeds safe range | critical |
E-^?Z | Control board signal output lead status timeout | Control output signal fails to reach expected state | error |
Fault codes stored in Fault Code Log | AED has detected and recorded one or more internal faults during self-test or operation | Internal self-test program execution or routine AED operation | warning |
Pad Battery Out of Service Pad Battery | Defibrillator pads are expired or non-functional | Pads beyond expiration date or connection failure | error |
S?? | Unit data save volume error | Data volume access failure | error |
Sensor Reset Energy | Sensor or energy measurement system error | Device cannot measure or reset energy-related sensors | error |
System FPGA Firmware Voltage Reference | Internal system, FPGA, firmware, or voltage reference failure | Self-test detects failure in core system component | critical |
**?Z | CPU measured pump input out of range | Pump drive signal abnormal | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.