The ServiceForge knowledge base holds 41 documented symptoms for the Physio-Control LIFEPAK 35, with 76 ranked causes, 58 diagnostic checks with test points and expected values, and 313 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No power when turned on

Device does not power up; no indication of operation

Ranked causes

  1. 1. Low battery voltage (common)
  2. 2. Battery connector pin loose, covered with foreign substance, or damaged (common)
  3. 3. Defective battery (occasional)
  4. 4. Defective power cord (occasional)
  5. 5. Power adapter not properly connected (occasional)
  6. 6. Defective power adapter (occasional)
First check

Verify battery charge level

Test point
Battery voltage or charge indicator
Expected
Battery fully charged

4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP (blood pressure) function malfunction patient safety

Non-invasive blood pressure monitoring not measuring correctly

Ranked causes

  1. 1. NIBP calibration error (common)
  2. 2. NIBP hose or connector leak (common)
  3. 3. NIBP module cable connection loose (common)
  4. 4. NIBP pump motor failure (occasional)
  5. 5. NIBP module electronics failure (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery charging LED not illuminated

Battery charging indicator LED inactive; battery not charging or charging circuit issue

Ranked causes

  1. 1. Device not properly connected to AC power source (common)
  2. 2. Battery not properly inserted in battery well (common)
  3. 3. No batteries installed (common)
  4. 4. Defective battery (occasional)
First check

Verify AC adapter connection

Test point
Power adapter connection to device inlet
Expected
Securely connected to device

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio system malfunction caution

Speaker or audio output not functioning or output distorted

Ranked causes

  1. 1. System flex cable connection loose (common)
  2. 2. Speaker cable connection loose (common)
  3. 3. Speaker assembly failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auxiliary power LED not illuminated high voltage

AC power indicator LED dark; device not receiving AC power

Ranked causes

  1. 1. Device not properly connected to AC power source (common)
  2. 2. Defective power cord (occasional)
  3. 3. Defective power adapter (occasional)
First check

Verify AC power connection

Test point
Device power inlet and power cord connection to wall outlet
Expected
Securely connected to active AC outlet

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Capnography (CO2) monitoring malfunction patient safety

End-tidal carbon dioxide measurement not functioning

Ranked causes

  1. 1. CO2 sensor calibration drift (common)
  2. 2. CO2 module cable or hose issue (common)
  3. 3. CO2 module hardware failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pulse oximetry (SpO2) monitoring malfunction patient safety

Oxygen saturation measurement not functioning or reading inaccurately

Ranked causes

  1. 1. SpO2 module cable connection loose (common)
  2. 2. SpO2 probe connector malfunction (common)
  3. 3. SpO2 module electronics failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

System overheating or fan malfunction caution

System temperature elevated or cooling fan not operating

Ranked causes

  1. 1. System flex cable connection loose (common)
  2. 2. Fan cable connection loose (common)
  3. 3. Fan motor failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery will not charge patient safety

Lithium-ion battery fails to charge when connected to AC power or charger.

Ranked causes

  1. 1. Battery temperature out of range (common)
  2. 2. Battery damaged or leaking (occasional)
First check

Inspect battery for damage or leakage

Test point
Battery exterior
Expected
No cracks, punctures, or fluid leakage visible

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charged battery loses power during patient care patient safety

Device powered by battery unexpectedly loses power or shuts down during use

Ranked causes

  1. 1. Improperly maintained battery (occasional)
  2. 2. Incompatible battery or charger used (rare)
First check

Verify battery fully charged before use

Test point
Battery charge indicator on device
Expected
Full charge indicated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Energy storage capacitor issue high voltage

Device capacitor charge or discharge malfunctioning affecting defibrillator energy storage

Ranked causes

  1. 1. Capacitor connection loose or corroded (common)
  2. 2. Energy storage capacitor hardware failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

High voltage exposure during PCB servicing high voltage

Lethal voltages may be present in the device even without operator action. Energy storage capacitors retain dangerous charge after power-off.

Ranked causes

  1. 1. Device not powered off or still connected to power (common)
  2. 2. Energy storage capacitor not discharged (common)
First check

Verify device is powered off

Test point
Power switch and indicator lights
Expected
All power off, no LEDs illuminated

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad malfunction patient safety

Keypad does not respond to user input or has intermittent response

Ranked causes

  1. 1. Keypad connection loose or corroded (common)
  2. 2. Keypad hardware failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient parameter monitoring malfunction patient safety

Patient vital signs monitoring not functioning correctly

Ranked causes

  1. 1. Parameter module cable connection loose (common)
  2. 2. Patient parameter PCBA failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Possible power loss during patient care patient safety

Device may lose power without warning during operation, potentially delaying therapy.

Ranked causes

  1. 1. Improperly maintained battery (common)
  2. 2. Use of unauthorized third-party batteries or chargers (occasional)
First check

Verify battery type and origin

Test point
Battery
Expected
LIFEPAK Flex Lithium-ion battery, not from other manufacturers

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Service codes detected and reported

Device operating software detects improper operation or malfunction; four-digit hexadecimal service code written to prognostic log; Service LED indicator illuminates

Ranked causes

  1. 1. Internal program or process execution failure (unknown)
  2. 2. Transient electromagnetic interference (EMI) or electrostatic discharge (ESD) (occasional)
First check

Display current active service codes

Test point
Service log in PSST
Expected
View active four-digit hexadecimal codes

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Static discharge damage to electronic components caution

SSDs (static-sensitive devices) such as MOS ICs, FETs, optical isolators, and film resistors can be damaged or destroyed by static electricity discharge, often imperceptibly

Ranked causes

  1. 1. Operator static charge buildup from synthetic clothing (common)
  2. 2. PCBAs not handled with proper antistatic precautions (common)
First check

Verify wrist strap is worn and properly connected

Test point
Wrist strap connection to mat and ground
Expected
Snug contact against bare skin with conductive connection established

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Static discharge damage to PCBs during handling caution

Semiconductor devices such as MOS ICs, FETs, optical isolators and film resistors may be damaged or destroyed by static electricity discharge, often imperceptible to the person causing it.

Ranked causes

  1. 1. Improper PCB handling without static protection (common)
  2. 2. Improper PCB storage and transport (common)
First check

Verify static-dissipative mat is in use and properly grounded

Test point
Mat ground connection
Expected
Connected to earth ground

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy cable does not disconnect smoothly high voltage

Release button does not function or cable does not slide out easily

Ranked causes

  1. 1. Release button stuck or not fully pressed (common)
  2. 2. Connector stuck due to corrosion or contamination (occasional)
First check

Press release button on therapy cable connector

Test point
Release button on connector
Expected
Button fully depresses and resets

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy cable does not lock in place high voltage

Connector fails to fully seat or does not produce audible click when inserted into receptacle

Ranked causes

  1. 1. Misaligned connector or receptacle (common)
  2. 2. Damaged connector or receptacle (occasional)
First check

Verify connector and receptacle alignment

Test point
Red therapy cable connector, red receptacle
Expected
Red connector aligned with red receptacle, gray release button facing up

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy cable or connector damage from disconnection without release button caution

Therapy cable connector damaged, loose, or broken from improper disconnection attempt without pressing release button

Ranked causes

  1. 1. Therapy cable disconnected without pressing release button (occasional)
  2. 2. Connector left exposed when not in use (occasional)
First check

Inspect therapy cable connector for physical damage

Test point
Therapy cable connector pins and housing
Expected
No visible cracks, bent pins, or deformation

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: CO2 sensor

Annual CO2 calibration followed by performance inspection to maintain measurement accuracy

Ranked causes

  1. 1. Annual calibration interval reached (common)
First check

Schedule and perform CO2 calibration

Test point
CO2 sensor
Expected
Calibration completed per TCP procedures

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Maintenance prompt interval

Set or reset the MAINTENANCE DUE message interval to 3, 6, or 12 months; reset after scheduled maintenance via PSST

Ranked causes

  1. 1. Maintenance interval needs configuration or reset (common)
First check

MAINTENANCE DUE message appears continuously for 10 minutes after power-on when interval reached

Test point
Device display
Expected
Message appears or does not appear per configured interval

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Capacitor discharge verification high voltage

Neon lamp test of energy storage capacitor charge status before service work

Ranked causes

  1. 1. Capacitor remains charged after case disassembly (common)
First check

Observe neon lamp in discharge tool while probes contact discharge points

Test point
Capacitor discharge points as shown in diagram
Expected
Neon lamp lights if charge of approximately 90 volts is present

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Component damage from static electricity caution

PCBAs with static-sensitive devices (SSDs) can be damaged or destroyed by electrostatic discharge during service, maintenance, or repair procedures. Damage may be imperceptible at the time of discharge.

Ranked causes

  1. 1. Operator static charge buildup from synthetic clothing (common)
First check

Identify static-sensitive devices (SSDs) by warning symbol before handling

Test point
PCBAs
Expected
SSD warning symbol present

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Defib energy calibration error patient safety

Defibrillator output energy does not match programmed or expected value

Ranked causes

  1. 1. Energy calibration offset needs adjustment (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

High voltage shock hazard during service high voltage

Lethal voltages may be present in the device even without operator action, posing electrocution risk during maintenance or repair

Ranked causes

  1. 1. Energy storage capacitor charged (common)
First check

Verify device is powered off

Test point
Power switch
Expected
OFF position

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incompatible therapy cable installed patient safety

Therapy cable from LIFEPAK 15 or 20e device connected to LIFEPAK 35

Ranked causes

  1. 1. Wrong therapy cable model used (occasional)
First check

Verify therapy cable catalog number

Test point
Therapy cable label or packaging
Expected
Catalog number 11113-000008

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Interface PCBA malfunction patient safety

Main interface printed circuit board assembly not functioning properly

Ranked causes

  1. 1. Interface PCBA hardware failure (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

MAINTENANCE DUE message appears

Device displays MAINTENANCE DUE message on screen

Ranked causes

  1. 1. Maintenance prompt set to display every 3, 6, or 12 months (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Physical device inspection after repair

Visual and physical verification that device is intact and safe for return to service

Ranked causes

  1. 1. Repair work may have introduced damage or improper assembly (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Post-repair performance inspection setup

Restore setup configuration and execute performance tests after device repair

Ranked causes

  1. 1. Setup options lost after PCBA replacement (common)
First check

Verify device setup configuration restored

Test point
Device setup mode or LIFENET system
Expected
Setup options profile downloaded and active on device

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power LED illuminated but screen blank and device does not operate

LED indicates power present but display inactive and no function

Ranked causes

  1. 1. Device boot up has failed (common)
First check

Perform power reset

Test point
POWER button and LED indicator
Expected
LED turns off after ~5 seconds hold; device restarts and displays normally

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Readiness indicator beacon inoperative or damaged

Readiness indicator beacon requires replacement, necessitating keypad replacement as well

Ranked causes

  1. 1. Beacon failure or damage (unknown)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Service LED indicator illuminated caution

Service LED illuminates indicating Service codes are present and require investigation before device returns to active use

Ranked causes

  1. 1. Active service codes present (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Static discharge damage to components caution

Semiconductor devices such as MOS ICs, FETs, optical isolators, and film resistors damaged or destroyed by static electricity discharge during maintenance or repair

Ranked causes

  1. 1. Static charge buildup from synthetic clothing and touching objects (common)
First check

Identify static-sensitive devices on PCBAs

Test point
PCBA components
Expected
SSD warning symbol visible on device

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Static discharge damage to SSDs and PCBAs caution

Static electricity can damage or destroy static-sensitive semiconductor devices (MOS ICs, FETs, optical isolators, film resistors) on circuit board assemblies

Ranked causes

  1. 1. Static charge buildup from synthetic clothing or improper handling (common)
First check

Identify SSDs on PCBA

Test point
Component markings
Expected
SSD warning symbol present on static-sensitive devices

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy cable connector incompatibility patient safety

Therapy cable from LIFEPAK 15 or 20e device does not connect to LIFEPAK 35 or is damaged by incompatible connection attempt

Ranked causes

  1. 1. Wrong therapy cable model used (occasional)
First check

Verify therapy cable part number

Test point
Therapy cable label or documentation
Expected
Catalog number 11113-000008 for LIFEPAK 35

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy cable connector not locking to receptacle patient safety

Therapy cable connector does not lock into receptacle; no click sound heard; connector may fall out or be loose

Ranked causes

  1. 1. Connector not properly aligned with receptacle (common)
First check

Visually inspect connector and receptacle alignment

Test point
Red therapy cable connector and red receptacle
Expected
Red connector aligned with red receptacle, gray release button facing up

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wire damage or pinching during reassembly caution

Wires may be pinched and damaged during device reassembly, causing electrical faults or short circuits

Ranked causes

  1. 1. Improper wire routing during assembly (occasional)
First check

Inspect all wires for damage or pinching

Test point
Wire routing path
Expected
No visible damage, wires not trapped or compressed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wire pinching and damage during reassembly caution

Wires may be pinched and damaged during reassembly, causing equipment malfunction or electrical hazards.

Ranked causes

  1. 1. Failure to follow assembly instructions carefully (common)
First check

Visually inspect wires during reassembly at each step

Test point
All wire routing paths and connector areas
Expected
No wires pinched, routed clearly away from moving parts and sharp edges

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
18System PCBA or accessory issueDetected system PCBA or accessory failure conditionerror
19Shift check failures and accessory problemsShift check test failure detectederror
20Speed dial issueSpeed dial control malfunction detectedwarning
21Beacon issueBeacon malfunctionunknown
22Invasive pressure monitoring issueInvasive pressure monitoring system malfunctionwarning
23VSM cable authentication logic issueVSM cable authentication check failswarning
24Possible coin cell battery power lossCoin cell battery voltage low or depletedwarning
25Possible issue with therapy connector or cablingTherapy connector or cable malfunction detectedwarning
E010IMX6 patient application integrity check failedSystem integrity verification check on patient application during startup or operationcritical
E014IMX6 runtest application integrity check failedSystem integrity verification check on runtest application during startupcritical
E021IMX6 therapy-related software failureFailure in software module responsible for therapy delivery functionscritical
E022IMX6 non-therapy software failureFailure in non-critical software module not related to therapy deliverywarning
E023IMX6 NAND flash integrity check failedMemory integrity verification check on NAND flash storage during startup or operationwarning
E024System software verification errorGeneral system software verification check failure during startupcritical
E025IMX6 application software integrity check failedIntegrity verification of application software layer during startup or operationcritical
E026IMX6 watchdog reset occurredWatchdog timer expired; processor did not respond to watchdog within timeout periodwarning
E028IMX6 3.3V USB OTG rail voltage out of specificationVoltage monitoring during self-test detects improper USB OTG rail voltagewarning
E029IMX6 5.0V rail voltage out of specificationSoftware monitoring detects voltage deviation during power-up or operationwarning
E02AIMX6 3.3V to PMOD rail voltage out of specificationVoltage monitoring detects improper 3.3V PMOD rail voltagewarning
E02BIMX6 3.3V cable voltage out of specificationVoltage monitoring detects improper 3.3V on cable connectionerror
E02CIMX6 3.3V rail voltage out of specificationSelf-test detects 3.3V rail voltage outside acceptable limitserror
E02DIMX6 3.9V rail voltage out of specificationSoftware monitoring detects voltage deviation during power-up or operationerror
E030Setup options integrity check failureData integrity check fails during startup or self-testerror
E031Manufacturing configuration integrity check failureMFG configuration data corruption detected during startuperror
E032Touchscreen failure detectedSelf-test fails to communicate with touchscreen or detect responsewarning
E034Active code table access failureSystem cannot read active code table from memoryerror
E036IMX6 PAD connector attached but not properly recognizedConnector detection mismatch during hardware initializationerror
E037Printer power overcurrent detectedCurrent draw from printer exceeds safe limitwarning
E038Interface USB bottom port overcurrentOvercurrent detected on USB bottom portwarning
E039Interface USB top port overcurrent detectedCurrent draw from USB top port exceeds safe threshold during device connectionwarning
E03APPB module overvoltage to USB bottom portOvervoltage condition detected from PPB module to USB bottom portwarning
E03BPPB module overvoltage to USB top portVoltage on USB top port exceeds safe threshold during PPB module operationwarning
E03CUSB port 5.0V power good signal failure5.0V supply monitoring detects power supply deviation on USB portwarning
E03DNIBP 6.0V supply voltage out of specificationSoftware monitoring detects voltage deviation during NIBP module operationwarning
E03ECapnography 5.0V supply voltage out of specificationSoftware monitoring detects voltage deviation during capnography module operationwarning
E040–E07FPaddles processor error on System PCBAReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E040-E07FPaddles processor errorFailure in paddles processor subsystemcritical
E050Paddles self-test failureDefibrillation paddles self-test executed at startup or on demand failserror
E055Paddles microcontroller software integrity check failurePaddle firmware checksum verification failserror
E057Paddles eye-close test failurePaddle eye sensor fails to respond to open/close cycle during testwarning
E058Paddles communications timeoutMain processor fails to receive response from paddle microcontroller within timeout perioderror
E059Paddles microcontroller state mismatchPaddle state machine detects inconsistent or invalid state during operationerror
E060Paddles watchdog test failureWatchdog timer fails during paddles self-testerror
E061Paddles stack test failurePaddle microcontroller stack integrity test fails during self-testerror
E062Paddles RAM test failurePaddle microcontroller RAM test fails during self-test sequenceerror
E063Paddles flash CRC test failureFlash memory CRC checksum verification in paddles failserror
E064Paddles power monitor failurePaddle power supply monitoring detects voltage out of specification during self-testerror
E065Paddles message CRC failureMessage checksum verification between main processor and paddles failserror
E066Paddles message ID errorPaddle receives message with unrecognized or invalid message IDerror
E067Paddles message sequencing failurePaddle messages received out of sequence during communicationerror
E068Paddles impedance not calibratedPaddle impedance measurement falls outside acceptable calibrated rangeerror
E069Paddles impedance calibration CRC errorDetected during self-testerror
E06APaddles ADC timeoutADC communication timeout during paddle diagnosticserror
E06CPaddles over voltage resetOver-voltage condition detected on paddle circuiterror
E06DPaddles ECG self test failedECG self-test failure during paddle diagnosticserror
E06EPaddles Z1 self test failedImpedance Z1 self-test failureerror
E06FPaddles Z2 self test failedImpedance Z2 self-test failureerror
E070Paddles M1 self test failedM1 measurement self-test failureerror
E071Paddles M2 self test failedM2 measurement self-test failureerror
E080–E0FFInternal VSM module error on System PCBAReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E080-E0FFInternal VSM module errorReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspection procedurecritical
E0A1VSM microcontroller communication failureLoss of communication with VSM microcontrollererror
E0A2VSM microcontroller communication CRC errorCRC error in VSM communication dataerror
E0A5VSM software integrity check failedSoftware checksum or integrity verification failureerror
E0A7VSM microcontroller communication aperiodicAperiodic or irregular communication with VSMerror
E0A8VSM 3.3V rail voltage faultDetected during VSM power supply monitoringerror
E0A9VSM analog 3.3V rail voltage faultDetected during analog power supply monitoringerror
E0AAVSM 1.2V rail voltage faultDetected during VSM power supply monitoringerror
E0ABVSM 5.5V rail voltage faultDetected during VSM power supply monitoringerror
E0ACVSM -5.5V rail voltage faultDetected during VSM power supply monitoringerror
E0ADVSM 10V rail voltage faultDetected during VSM power supply monitoringerror
E0AEVSM -10V rail voltage faultDetected during VSM power supply monitoringerror
E0AFVSM leads channel I faultDetected during VSM lead monitoringerror
E0B0VSM leads channel II faultDetected during VSM lead monitoringerror
E0B1VSM leads channel III faultDetected during VSM lead monitoringerror
E0B2VSM chopper circuit faultDetected during chopper diagnosticserror
E0B3VSM Wilson central terminal calculation faultDetected during Wilson terminal calculationerror
E0B4VSM voltage measurement faultDetected during voltage measurement monitoringerror
E0B5VSM IBP calibration faultDetected during invasive blood pressure calibrationerror
E0B6VSM IBP monitor faultDetected during IBP signal monitoringerror
E0B7VSM patient temperature calibration faultDetected during temperature calibrationerror
E0BAVSM FPGA faultDetected during FPGA diagnosticserror
E0BBVSM cable authentication failureDetected during cable authentication checkerror
E0BDVSM SPI ready signal high faultSPI ready signal stuck high during communicationerror
E0BEVSM SPI ready signal low faultSPI ready signal stuck low during communicationerror
E0BFVSM QRS GPIO toggle faultQRS GPIO toggle failure during operationerror
E0XXMain processor or paddles processor or internal VSM module errorReal time error during device use or failure during power-on self-test, auto-test, user test, or performance inspection procedurecritical
E0XXMain processor errorReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspection procedurecritical
E0XX (E000-E03F)Main processor errorFailure in main processor or System PCBA detected during operation or self-testcritical
E0XX (E000–E03F)Main processor error on System PCBAReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E0XX (E040-E07F)Paddles processor errorFailure in paddles processor detected during operation or self-testcritical
E0XX (E080-E0FF)Internal VSM module errorFailure in internal VSM module detected during operation or self-testcritical
E100–E11FInternal communication devices error (WiFi)Real-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E100-E11FInternal communication devices errorReal-time error during device use or failure during self-test procedurescritical
E101WiFi power-on self test failedDetected during WiFi module startuperror
E120IMX6 I2C2 BUS ERRORIMX6 I2C2 bus communication failureunknown
E120–E12FSystem PCBA or IMX6 device error affecting access port, touch screen, audioReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E120-E12FSystem PCBA or IMX6 device error (access port, touch screen, audio)Real-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspection procedurecritical
E122Touch screen device failureTouch screen hardware not detected or respondingunknown
E123TOUCH SCREEN FUNCTION ERRORTouch screen functionality check failureunknown
E124Audio amplifier device failureAudio amplifier not detected or respondingunknown
E126Audio codec device failureAudio codec hardware not detected or respondingunknown
E127Audio codec function checkAudio codec self-test or functional check failureunknown
E128SYS-PCBA EEPROM device failureSystem EEPROM hardware not detected or respondingunknown
E129SYS-PCBA EEPROM function errorEEPROM read/write test failureunknown
E12AACCESS 15 Volt ADC device failure15V ADC converter not detected or respondingunknown
E12CTouch controller pin fault test failureTouch controller pin continuity or connectivity test failureunknown
E12ETouch controller AVDD test failureTouch controller analog supply voltage test failureunknown
E130IMX6 I2C3 ERRORIMX6 I2C3 bus communication failureunknown
E130–E13FIMX6 device error affecting usage counters or SOMReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E130-E13FIMX6 device error (access port, usage counters, SOM)Real-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspection procedurecritical
E132SOM user EEPROM device failureUser EEPROM on SOM not detected or respondingunknown
E133SOM user EEPROM function errorUser EEPROM read/write test failureunknown
E134SOM manufacturing EEPROM device failureManufacturing EEPROM on SOM not detected or respondingunknown
E135SOM manufacturing EEPROM function errorManufacturing EEPROM read/write test failureunknown
E136SOM temperature sensor 0 device failureTemperature sensor 0 on SOM not detected or respondingunknown
E138SOM temperature sensor 1 device failureTemperature sensor 1 on SOM not detected or respondingunknown
E13CHard paddles ADC device failureHard paddle ADC converter not detected or respondingunknown
E13DHard paddles ADC function errorHard paddle ADC read/conversion test failureunknown
E141Bluetooth power-on self testBluetooth module self-test failure during startupunknown
E180-E1FFIMX6 peripherals error (SD card, data logging)Real-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspection procedurecritical
E180–E1FFIMX6 peripherals error affecting SD card or data loggingReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E1XXInternal communication devices, system access, or IMX6 peripherals errorReal time error during device use or failure during power-on self-test, auto-test, user test, or performance inspection procedurecritical
E1XX (E100-E11F)Internal communication devices errorFailure in WiFi communication detected during operation or self-testerror
E1XX (E120-E12F)System access port or touch screen errorFailure in access port, touch screen, or audio functions detected during operation or self-testerror
E1XX (E130-E13F)IMX6 device or usage counter errorFailure in IMX6 device, access port, or usage counter detected during operation or self-testerror
E1XX (E180-E1FF)IMX6 peripherals error (SD card or data logging)Failure in SD card or data logging system detected during operation or self-testwarning
E200-E3FFPower and therapy delivery errorReal-time error during device use or failure during self-test procedurescritical
E200–E3FFPower and therapy delivery error on Energy delivery PCBAReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E225Power microcontroller software integrity checkPower microcontroller firmware checksum or integrity test failureerror
E226Power microcontroller watchdog resetPower microcontroller watchdog timer expiredwarning
E227Power microcontroller communications errorCommunication between power microcontroller and other modules failederror
E228Power microcontroller communications error with SOMCommunication between power microcontroller and SOM module failederror
E229Power microcontroller communications aperiodic message checkExpected periodic message from power microcontroller not receivederror
E22APower microcontroller EEPROM service read errorEEPROM read operation from power microcontroller failederror
E22CPower source failure checkPower source voltage or current test failureerror
E230Power microcontroller 15V bus rail15V power rail voltage out of specificationerror
E231Power microcontroller analog output 3.3V rail3.3V analog output power rail voltage out of specificationerror
E233Service LED failureService LED not illuminating or responding to control signalwarning
E238Power button stuckPower button switch stuck in pressed or released positionerror
E239RTC check failure indicating possible power lossReal-time clock verification fails during startup or operationwarning
E23BCoin cell battery check failureCoin cell voltage below acceptable thresholdwarning
E2A3Therapy microcontroller state mismatchTherapy module state does not match expected state during operationerror
E2A5Therapy microcontroller software integrity check failureCRC or checksum validation fails on therapy microcontroller firmwareerror
E2A6Therapy pacing errorPacing output circuit detects malfunction during pacing modeerror
E2A7Therapy pacing rate errorPacing rate generation circuit fails or generates incorrect frequencyerror
E2A8Therapy prepulse GPIO errorGeneral-purpose input/output prepulse control line malfunctionerror
E2A9Therapy QRS GPIO errorQRS detection GPIO control line fails or is in wrong stateerror
E2AACharge remains on capacitor after shock deliveryHigh-voltage capacitor does not fully discharge after defibrillation shockerror
E2ABCharge energy incorrectMeasured delivered energy does not match selected energy levelerror
E2ACCharge energy incorrect with internal paddlesDelivered energy incorrect when using internal paddle electrodeserror
E2ADCharge energy too high; output disabledCharging circuit produces voltage exceeding safe maximum thresholderror
E2AFTherapy communication timeoutMain processor fails to receive timely response from therapy microcontrollererror
E2C0Therapy microcontroller watchdog check failureWatchdog timer expires; microcontroller not responding within timeout perioderror
E2C1Therapy microcontroller stack check failureStack pointer corruption or stack overflow detectederror
E2C2Therapy microcontroller RAM test failureRAM memory self-test detects bad memory location or bit errorerror
E2C3Therapy flash CRC test failureFlash memory checksum verification failserror
E2C4Therapy 3.3V monitor check failure3.3V supply voltage out of spec or supply monitor circuit failureerror
E2C5Therapy 24V monitor check failure24V supply voltage out of spec or supply monitor circuit failureerror
E2C6Therapy ADC conversion timeoutAnalog-to-digital converter fails to complete conversion within timeouterror
E2C7Therapy oscillator clock errorMicrocontroller oscillator fails to generate correct clock frequencyerror
E2C8Therapy transfer open errorOutput contactor or relay fails to open after shock deliveryerror
E2C9Therapy transfer close errorOutput contactor or relay fails to close for shock deliveryerror
E2CATherapy microcontroller communication CRC errorCRC checksum failure on message from therapy microcontrollererror
E2CBTherapy microcontroller communication internal issueTherapy microcontroller reports internal communication protocol errorerror
E2CCTherapy microcontroller communication dropped messageMain processor detects missing expected message from therapy microcontrollererror
E2CDTherapy charge command errorTherapy microcontroller rejects or fails to execute charge command from main processorerror
E2CETherapy arm command errorTherapy microcontroller rejects or fails to execute arm commanderror
E2CFTherapy capacitor overvoltage conditionHigh-voltage capacitor charging circuit output exceeds safe maximum voltageerror
E2D0Therapy charging timeoutCapacitor fails to charge to target voltage within timeout perioderror
E2D1Therapy dump timeoutCapacitor fails to fully discharge within timeout period after shock or when commandederror
E2D2SC_TH_ED_ERROR_PACE_ON_CMD - pacing on command errorTherapy module fails to execute or acknowledge pacing enable commanderror
E2D2SC_TH_ED_ERROR_PACE_ON_CMDPacing command execution errorerror
E2D3Therapy pulse width errorPacing pulse width generation circuit produces incorrect pulse widtherror
E2D4Therapy pace charge timeoutPacing charge circuit fails to reach target within timeout perioderror
E2D5SC_TH_ED_ERROR_PULSE_START_CD_TOPulse start charge timeout during shock deliveryerror
E2D6SC_TH_ED_ERROR_PULSE_STOP_CD_TOPulse stop charge delivery timeouterror
E2D7SC_TH_ED_ERROR_PULSE_STARTTherapy pulse start error detectederror
E2D8SC_TH_ED_ERROR_UNEXPECTED_WATCHDOGWatchdog timer reset during therapy operationerror
E2D9Therapy H-bridge errorH-bridge circuit fault detected during defibrillation charge or deliveryerror
E2DATherapy defib calibration errorCalibration check of defibrillation circuit failserror
E2DBTherapy ADC compare errorAnalog-to-digital converter comparison mismatcherror
E2DCTherapy voltage on capCapacitor voltage exceeds safe threshold or is out of expected rangeerror
E2DDTherapy redundant variableRedundancy check fails on therapy control variableserror
E2DETherapy relay errorTherapy relay does not respond or feedback signal incorrecterror
E2DFTherapy over voltage errorCapacitor or output voltage exceeds maximum safe limiterror
E2XXPower and therapy delivery errorReal time error during device use or failure during power-on self-test, auto-test, user test, or performance inspection procedurecritical
E2XX (E200-E3FF)Power and therapy delivery errorFailure in power supply or energy delivery system detected during operation or self-testcritical
E2XX-E3FFPower and therapy delivery errorReal-time error during device use or test failurecritical
E400-E5FFInterface PCBA errorReal-time error during device use or failure during self-test procedurescritical
E400–E5FFInterface PCBA errorReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E421Interface uC communication failureInterface microcontroller communication with main system failserror
E425Interface uC software integrity checkSoftware checksum or integrity verification failederror
E427Interface uC communication errorInterface microcontroller detects communication protocol errorerror
E430Interface uC power LEDPower LED does not illuminate or control signal failswarning
E431Interface uC shock LEDShock LED does not illuminate or control signal failswarning
E432Interface uC red beaconRed beacon light does not illuminate or control signal failswarning
E433Interface uC yellow beaconYellow beacon light does not illuminate or control signal failswarning
E434Interface uC 15V rail15V power supply voltage is out of specificationerror
E435Interface uC AVDD 3.3V railAnalog supply 3.3V voltage is out of specificationwarning
E437Interface uC 1.8V rail1.8V core supply voltage is out of specificationwarning
E438Interface uC 3.3V rail3.3V digital supply voltage is out of specificationerror
E439Interface uC 5V rail5V digital supply voltage is out of specificationwarning
E43AInterface uC 12V rail12V supply voltage out of toleranceerror
E43BInternal fanFan speed sensor or motor control signal failserror
E43CInternal speakerSpeaker does not produce sound or control signal failswarning
E43DInterface PCBA EEPROM failureEEPROM read or write error detectedwarning
E43EAmbient light sensorAmbient light sensor does not respond or signal out of rangewarning
E4E1Audio damaged files errorAudio file integrity check fails during boot or playbackwarning
E4XXInterface PCBA errorInterface PCBA subsystem failureerror
E4XX (E400-E5FF)Interface PCBA errorFailure in Interface PCBA detected during operation or self-testerror
E4XX-E5FFInterface PCBA errorReal-time error during device use or test failurecritical
E600–E7FFInternal OEM modules error (patient parameter, CO2, SpO2, NIBP modules)Real-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E600-E7FFInternal OEM modules errorReal-time error during device use or failure during self-test procedurescritical
E660Capnography self-test failureCapnography module internal self-test failswarning
E661Capnography device failureCapnography sensor or module is not functioningwarning
E662Capnography communication failureCommunication between main system and capnography module failswarning
E670Capnography calibration failureCapnography module calibration verification failederror
E6A0NIBP self-test failureNIBP module self-test check failswarning
E6A1NIBP device failureNIBP module device malfunction detectedwarning
E6A2NIBP communication failureLoss of communication with NIBP modulewarning
E6E0Oximetry self-test failureOximetry module self-test check failswarning
E6E1Oximetry device failureOximetry module hardware failure detectedwarning
E6E2Oximetry communication failureCommunication lost with oximetry modulewarning
E6XXInternal OEM module errorReal time error during device use or failure during diagnostic testcritical
E6XX (E600-E7FF)Internal OEM module errorFailure in patient parameter module or CO2/SpO2/NIBP module detected during operation or self-testwarning
E6XX-E7FFInternal OEM modules errorReal-time error during device use or test failurecritical
E800Therapy cable validity check failureTherapy cable validation test failswarning
E800–E9FFExternal OEM devices errorReal-time error during device use or failure during power-on self-test, auto-test, user test, or performance inspectioncritical
E800-E9FFExternal OEM devices errorReal-time error during device use or failure during self-test procedurescritical
E8XXExternal OEM device errorExternal patient parameter module failurewarning
E8XX (E800-E9FF)External OEM device errorFailure in external patient parameter device detected during operation or self-testwarning
E8XX-E9FFExternal OEM devices errorReal-time error during device use or test failurecritical
EA00–EADFUser tests or non-ready conditionsFailure during user test or performance inspection procedurecritical
EA00-EADFUser tests errorReal-time error during device use or failure during self-test procedurescritical
EA02Shiftcheck charge button failureCharge button self-test failswarning
EA03Shiftcheck shock button failureShock button self-test failswarning
EA04Shiftcheck analyze button failureAnalyze button self-test failswarning
EA05Shiftcheck speed dial push failureSpeed dial push function fails self-testwarning
EA06Shiftcheck speed dial clockwise failureSpeed dial clockwise rotation fails self-testwarning
EA07Shiftcheck speed dial counter-clockwise failureSpeed dial counter-clockwise rotation self-test failswarning
EA08Shiftcheck power button LED failurePower button LED self-test failswarning
EA09Shiftcheck shock button LED failureShock button LED self-test failswarning
EA0AShiftcheck auxiliary power LED failureAuxiliary power LED self-test failswarning
EA0BShiftcheck battery power LED failureBattery power LED self-test failswarning
EA0CShiftcheck service LED failureService LED self-test failswarning
EA0DShiftcheck beacon patient red failureBeacon patient red light self-test failswarning
EA0EShiftcheck beacon patient yellow failureBeacon patient yellow light self-test failswarning
EA0FShiftcheck readiness LED yellow failureReadiness yellow LED self-test failswarning
EA10Shiftcheck touch screen press failureTouch screen press self-test failswarning
EA11Shiftcheck pixel display failureDisplay pixel test fails or display malfunction detectedwarning
EA12Shiftcheck microphone speaker failureAudio input or output device fails self-testwarning
EA13Shiftcheck device damageVisual inspection detects device damage during self-checkwarning
EA13Device damage shiftcheck failureVisual inspection self-test during startuperror
EA14Shiftcheck device substance contaminationSubstance contamination detected during self-checkwarning
EA15Shiftcheck power cord damagePower cord damage detected during inspectionwarning
EA16Printer paper shiftcheck failurePrinter paper status checkwarning
EA16Shiftcheck printer paper low or absentPrinter paper level low or absent during self-checkinfo
EA17Shiftcheck printer operation failurePrinter operation self-test failswarning
EA18Shiftcheck spare pads absentSpare defibrillation pads not found during self-checkinfo
EA19Shiftcheck pads date expiredDefibrillation pads have expired during self-checkinfo
EA1APaddles clean shiftcheck failurePaddle cleanliness inspection self-testwarning
EA1AShiftcheck paddles not cleanPaddle surfaces not clean during self-checkinfo
EA1BShiftcheck electrode package issueElectrode package problem detected during self-checkinfo
EA1CTherapy cable shiftcheck failureTherapy cable inspection self-testerror
EA1CShiftcheck therapy cable issueTherapy cable problem detected during self-checkwarning
EA1DShiftCheck ECG cableAutomatic equipment self-test during shift checkwarning
EA1EShiftCheck SpO2 cableAutomatic equipment self-test during shift checkwarning
EA1FShiftCheck NIBP accessAutomatic equipment self-test during shift checkwarning
EA20ShiftCheck AC powerAutomatic equipment self-test during shift checkwarning
EA21ShiftCheck battery powerAutomatic self-test during shift checkwarning
EA22ShiftCheck charged batteriesAutomatic equipment self-test during shift checkwarning
EA23ShiftCheck temperature cableAutomatic equipment self-test during shift checkwarning
EA24ShiftCheck IBP cableAutomatic equipment self-test during shift checkwarning
EA43Energy dial clockwiseEnergy dial rotation test during shift checkwarning
EA44Energy dial counterclockwiseEnergy dial rotation test during shift checkwarning
EA45Paddles charge buttonPaddle charge button test during shift checkwarning
EA46Paddle sternum buttonSternum paddle button pressed during functional testwarning
EA47Paddles apex buttonApex paddle button pressed during functional testwarning
EA48Paddles charge buttonPaddle charge button test during shift checkwarning
EA49Paddles print buttonPaddle print button pressed during functional testwarning
EA4APaddles surface inspectionVisual paddle surface inspection during shift checkwarning
EA4BPaddles cable inspectionVisual inspection of paddle cables during testwarning
EA4CPaddles connector inspectionPaddle connector visual inspection during shift checkwarning
EA4DPaddles connection checkPaddle electrical connection test during shift checkwarning
EA4EPaddles in wells checkPaddle position detection test during shift checkwarning
EA4FPaddles charge checkPaddle charging function test during shift checkwarning
EA50Paddles energy delivery checkPaddle energy delivery test during shift checkwarning
EA51Paddles test load detectedTest load connected to paddles during functional testwarning
EA52Energy test load detectedTest load not detected during energy delivery testwarning
EA53Energy test load charge checkTest of charging energy into test loadwarning
EA54Energy test load delivery checkTest load energy delivery verification during shift checkwarning
EA55Energy test load leads off detectedLeads-off condition detected during test load operationwarning
EA56Energy test load cable checkTest load cable integrity check during shift checkwarning
EAE0Auto-test service error checkAutomatic self-test detects internal service errorerror
EAXXUser tests or performance inspection errorFailure during user test or performance inspection procedureerror
EAXX (EA00-EADF)User test or performance inspection codeFailure or result during user test or performance inspection procedure (PIP)warning
EB00Battery 1 missingEquipment startup or battery detection checkerror
EB01Battery 1 faultBattery 1 detected with fault condition during testwarning
EB02Battery 1 not authorizedBattery authentication check during startuperror
EB03Battery 1 lowBattery charge level monitoring during operationwarning
EB04Battery 2 missingEquipment startup or battery detection checkerror
EB05Battery 2 faultBattery 2 detected with fault condition during testwarning
EB06Battery 2 not authorizedUnauthorized battery detectedwarning
EB07Battery 2 lowBattery 2 charge below operational thresholdwarning
EB08Auxiliary power not connectedAuxiliary power source disconnected or unavailablewarning
EB09Battery 1 end-of-lifeBattery 1 has reached end-of-life thresholdwarning
EB0ABattery 2 end-of-lifeBattery 2 has reached end-of-life thresholdwarning
EB13Electrodes test load detectedTest load detected on electrodesinfo
EB14Electrodes simulator detectedElectrodes simulator connected during normal operationinfo
EB20Therapy cable missingTherapy cable not detected or disconnectederror
E_UXXUser test codeFailure during user test executionwarning
MAINTENANCE DUEScheduled maintenance interval has been reachedDevice powered on after maintenance interval (3, 6, or 12 months) has elapsedwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI