Physio-Control LIFEPAK 20: troubleshooting and error codes

The ServiceForge knowledge base holds 186 documented symptoms for the Physio-Control LIFEPAK 20, with 330 ranked causes, 230 diagnostic checks with test points and expected values, and 240 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Not printing caution

Printer does not produce output

Ranked causes

  1. 1. Printer test failure (common)
  2. 2. Paper not loaded or improper (common)
  3. 3. Printer power supply issue (occasional)
  4. 4. W14 Printer Flex Cable shorted (occasional)
  5. 5. A12 Printer Assembly defective (occasional)
  6. 6. A05 UI PCB faulty (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No lead channel sync mark patient safety

Synchronous cardioversion ECG lead sync mark not displayed

Ranked causes

  1. 1. Noisy ECG signal (common)
  2. 2. ECG cable defective (common)
  3. 3. W06 ECG connector defective (common)
  4. 4. A02 Patient Parameter PCB faulty (occasional)
  5. 5. A01 System PCB faulty (unknown)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No lead channel Sync mark displayed patient safety

Synchronous cardioversion Sync mark not displayed on ECG lead channel

Ranked causes

  1. 1. Noisy ECG signal (common)
  2. 2. Patient cable issue (occasional)
  3. 3. W06 ECG connector issue (occasional)
  4. 4. A02 Patient Parameter PCB failure (occasional)
  5. 5. A01 System PCB sync circuit failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No remote sync patient safety

Remote synchronization function not working

Ranked causes

  1. 1. Remote sync function disabled (common)
  2. 2. ECG Out/Sync In connector defective (common)
  3. 3. W09 Power Cable defective (occasional)
  4. 4. A01 System PCB faulty (occasional)
  5. 5. A03 Power Module faulty (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No remote Sync available

Remote sync function does not work; external sync signal not accepted

Ranked causes

  1. 1. Remote sync function disabled (common)
  2. 2. ECG Out/Sync In connector or power module issue (occasional)
  3. 3. W09 Power Cable issue (occasional)
  4. 4. A01 System PCB sync circuit failure (rare)
  5. 5. Power module sync circuit failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to complete auto test or user test patient safety

Device cannot successfully run built-in diagnostic tests

Ranked causes

  1. 1. Improper test load connection (common)
  2. 2. Test plug or shorting bar continuity issue (common)
  3. 3. QUIK-COMBO cable continuity issue (common)
  4. 4. W01 Therapy Connector defective (occasional)
  5. 5. A04 Therapy PCB faulty (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Damaged keypad or labels

Keypad or labels are damaged or worn

Ranked causes

  1. 1. Damaged elastomer keypad (common)
  2. 2. Damaged product identification label (occasional)
  3. 3. Damaged explosion/hazard label (occasional)
  4. 4. Damaged operating instruction label (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hard paddle buttons patient safety

Defibrillator paddle buttons are stiff

Ranked causes

  1. 1. Paddle button mechanical issue (common)
  2. 2. Hard paddles defective (common)
  3. 3. W01 Therapy Connector defective (unknown)
  4. 4. A04 Therapy PCB faulty (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaudible or garbled audio patient safety

Speaker output absent, quiet, or unintelligible

Ranked causes

  1. 1. Audio test failure (common)
  2. 2. Speaker connection loose (common)
  3. 3. W02 Speaker Assembly defective (common)
  4. 4. A01 System PCB faulty (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pacer does not turn on patient safety

Pacing mode cannot be activated or does not operate

Ranked causes

  1. 1. Pacer not enabled or keypad issue (common)
  2. 2. Elastomer keypad defective (occasional)
  3. 3. A05 UI PCB failure (occasional)
  4. 4. Pacer control circuit failure or error code (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer does not print

Printer produces no output, no paper movement, or no marks on paper

Ranked causes

  1. 1. No paper or paper improperly loaded (common)
  2. 2. W14 Printer Flex Cable voltage high (occasional)
  3. 3. A12 Printer Assembly failure (occasional)
  4. 4. A05 UI PCB printer control failure (rare)
First check

Run printer test

Test point
Printer
Expected
Printer test passes
Tools
Run diagnostic test

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Saturation reading missing or out of tolerance patient safety

SpO2 saturation reading absent or outside normal range

Ranked causes

  1. 1. SpO2 cable defective or disconnected (common)
  2. 2. SpO2 sensor defective (common)
  3. 3. W05 SpO2 Assembly defective (occasional)
  4. 4. A06 OEM PCB faulty (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unable to complete energy delivery auto test high voltage

Defibrillator cannot complete automatic delivered energy test

Ranked causes

  1. 1. Test load shorting bar or plug continuity issue (common)
  2. 2. QUIK-COMBO cable continuity issue (occasional)
  3. 3. W01 Therapy Connector issue (occasional)
  4. 4. A04 Therapy PCB failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormal energy delivery patient safety

Defibrillator delivers unexpected energy

Ranked causes

  1. 1. A14 Inductive Resistor faulty (common)
  2. 2. A04 Therapy PCB faulty (occasional)
  3. 3. A01 System PCB faulty (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Abnormal energy delivery (patient impedance channel broken) patient safety

Defibrillator delivers abnormal or incorrect shock energy to patient

Ranked causes

  1. 1. A14 Inductive Resistor open or high resistance (common)
  2. 2. A04 Therapy PCB failure (occasional)
  3. 3. A01 System PCB failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Computer-assisted module tests

Serial communication, USB interface, battery power, battery I2C, wireless and wireless switch functionality verification

Ranked causes

  1. 1. DUT does not communicate with test computer (occasional)
  2. 2. Serial to USB adapter misconfiguration or incompatibility (occasional)
  3. 3. Weak wireless signal interference (occasional)
First check

Verify wireless network availability

Test point
Test environment
Expected
Operational WPA/WPA2 enterprise or personal signal broadcasted

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display is blank

Display is completely blank with no backlighting or text visible

Ranked causes

  1. 1. Display backlight/display assembly failure (common)
  2. 2. A04 Therapy PCB voltage issue (occasional)
  3. 3. Power module failure (rare)
First check

Measure 12V on P21 to ground

Test point
P21 connector
Expected
12 V to ground
Tools
DMM

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display on but power on LED not illuminated

Display functions but power on indicator LED is not lit

Ranked causes

  1. 1. UI Flex Cable issue (common)
  2. 2. A05 UI PCB failure (occasional)
  3. 3. A04 Therapy PCB failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display on, no power-on LED

Display shows information but power-on LED not lit

Ranked causes

  1. 1. W18 UI Flex Cable defective (unknown)
  2. 2. A05 UI PCB faulty (unknown)
  3. 3. A04 Therapy PCB faulty (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display shows white or blue screen

Display is on but shows white or blue screen instead of normal operation

Ranked causes

  1. 1. A05 UI PCB voltage issue (common)
  2. 2. W18 UI Flex Cable disconnected (occasional)
  3. 3. A01 System PCB failure (occasional)
First check

Measure voltage on A05 UI PCB

Test point
C96 connector
Expected
3.3 V to ground
Tools
DMM

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG analog out missing or out of tolerance patient safety

ECG analog output signal absent or outside spec

Ranked causes

  1. 1. Simulator output verification needed (common)
  2. 2. ECG display not showing signal (common)
  3. 3. W11 ECG out connector defective (common)

ECG gain out of tolerance patient safety

ECG signal amplification outside specifications

Ranked causes

  1. 1. Simulator output verification needed (common)
  2. 2. ECG cable defective (common)
  3. 3. A02 Patient Parameter PCB faulty (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to transfer coincident with Sync mark patient safety

Defibrillator does not deliver therapy when Sync mark aligns properly with R-wave

Ranked causes

  1. 1. Sync marker not properly placed on R-wave (common)
  2. 2. Keypad button issue or UI PCB failure (occasional)
  3. 3. User test failure or error code (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Improper button response caution

Buttons do not respond as expected

Ranked causes

  1. 1. Keypad malfunction (common)
  2. 2. Elastomer keypad defective (common)
  3. 3. A05 UI PCB faulty (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Light print caution

Printer output is faint

Ranked causes

  1. 1. Improper paper type (common)
  2. 2. W14 Printer Flex Cable disconnected (common)
  3. 3. A05 UI PCB faulty (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Missing or broken characters in print caution

Printed output has missing or illegible characters

Ranked causes

  1. 1. Improper paper type (common)
  2. 2. Printhead dirty or clogged (common)
  3. 3. A12 Printer Assembly defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No ECG patient safety

No ECG signal displayed

Ranked causes

  1. 1. ECG cable defective (common)
  2. 2. W06 ECG Connector defective (common)
  3. 3. A02 Patient Parameter PCB faulty (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No ECG signal patient safety

ECG waveform not displayed; no cardiac rhythm detected

Ranked causes

  1. 1. ECG cable disconnected or failed (common)
  2. 2. ECG connector failure (occasional)
  3. 3. Patient Parameter PCB failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No ECG signal displayed

ECG trace is not visible on display or no waveform is present

Ranked causes

  1. 1. ECG cable disconnected or damaged (common)
  2. 2. W06 ECG connector issue (occasional)
  3. 3. A02 Patient Parameter PCB failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No paddles channel sync mark patient safety

Synchronous cardioversion paddle sync mark not displayed

Ranked causes

  1. 1. Therapy cable defective (common)
  2. 2. W01 Therapy Connector defective (common)
  3. 3. A01 System PCB faulty (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No paddles channel Sync mark displayed patient safety

Synchronous cardioversion Sync mark not displayed on paddles channel

Ranked causes

  1. 1. Therapy cable issue (common)
  2. 2. W01 Therapy Connector issue (occasional)
  3. 3. A01 System PCB sync circuit failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No power on patient safety

Device does not power up when turned on

Ranked causes

  1. 1. AC power not connected (common)
  2. 2. Therapy PCB failure (occasional)
  3. 3. Power module failure (occasional)
First check

Verify AC power connection

Test point
Power cord and outlet
Expected
Device plugged in and powered
Tools
None

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No SpO2 response (no cable detected)

SpO2 sensor not detected; no pulse oximetry signal available

Ranked causes

  1. 1. SpO2 cable disconnected or damaged (common)
  2. 2. SpO2 sensor defective (occasional)
  3. 3. W05 SpO2 Assembly issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer output is light

Printed output is faint, light, or barely visible

Ranked causes

  1. 1. Wrong paper type used (common)
  2. 2. W14 Printer Flex Cable connection issue (occasional)
  3. 3. A05 UI PCB printhead control issue (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 saturation reading missing or out of tolerance

SpO2 value not displayed or reading is outside expected physiological range

Ranked causes

  1. 1. SpO2 cable or sensor issue (common)
  2. 2. W05 SpO2 Assembly failure (occasional)
  3. 3. A06 OEM PCB failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy cable leads off high voltage

QUIK-COMBO therapy cable not properly connected

Ranked causes

  1. 1. QUIK-COMBO cable continuity broken (common)
  2. 2. W01 Therapy Connector defective (common)
  3. 3. A01 System PCB faulty (unknown)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy cable leads off (QUIK-COMBO only) patient safety

QUIK-COMBO therapy cable disconnects or lead continuity is broken

Ranked causes

  1. 1. QUIK-COMBO cable continuity broken (common)
  2. 2. W01 Therapy Connector disconnected (occasional)
  3. 3. A01 System PCB therapy circuit failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy ECG gain out of tolerance patient safety

Therapy paddle ECG signal amplification outside spec

Ranked causes

  1. 1. Simulator output verification needed (common)
  2. 2. Therapy cable defective (common)
  3. 3. A01 System PCB faulty (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on high voltage

Defibrillator does not turn on when power button is pressed

Ranked causes

  1. 1. Not connected to AC power or UI power issue (common)
  2. 2. UI flex cable issue (occasional)
  3. 3. Power module failure (occasional)
First check

Verify AC power connection

Test point
AC outlet
Expected
Power connected
Tools
Visual inspection

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AED mode not functioning patient safety

AED mode does not activate, voice prompts inaudible, or device fails to switch to manual mode when energy select pressed

Ranked causes

  1. 1. Hard paddles not properly installed or analyzer not connected (common)
  2. 2. Device speaker or voice prompt system malfunction (occasional)
First check

Connect therapy cable/paddles and press ANALYZE key to activate AED mode

Test point
Device ANALYZE key, display
Expected
AED mode activated

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio is inaudible or garbled

Speaker produces no sound or produces garbled, distorted sound

Ranked causes

  1. 1. Speaker disconnected or defective (common)
  2. 2. A01 System PCB audio circuit failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio output distorted or silent patient safety

Voice messages or tone alerts are not clear or not audible

Ranked causes

  1. 1. Speaker damage or audio circuit fault (rare)
  2. 2. Audio mute setting enabled (occasional)
First check

Trigger voice prompt and tone sequences

Test point
Speaker output
Expected
Clear, undistorted voice and tone messages

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Backup battery does not operate caution

Unit cannot run on backup battery; device powers off when AC is disconnected

Ranked causes

  1. 1. A07 Battery dead or missing (common)
  2. 2. Power module charging circuit failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Maintenance prompt interval

Set or reset the scheduled preventative maintenance prompt interval

Ranked causes

  1. 1. Configure maintenance reminder schedule (unknown)
  2. 2. Reset maintenance interval after performing maintenance (common)
First check

Display MAINT PROMPT menu and select INTERVAL option

Test point
Service screen - MAINT PROMPT option
Expected
Menu shows OFF, 3 MONTHS, 6 MONTHS, 12 MONTHS choices

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Chassis leakage current elevated high voltage

Leakage current from chassis to ground exceeds safe limits in any polarity/neutral configuration

Ranked causes

  1. 1. Insulation breakdown in power supply (rare)
  2. 2. Wet or contaminated PCB (occasional)
First check

Measure leakage current (Neutral Closed, Normal polarity, Lead-Chassis Normal)

Test point
Leakage path simulation
Expected
< 100 µA (medical equipment standard)
Tools
Leakage current tester

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Earth leakage current elevated high voltage

Leakage current from device to earth exceeds safe limits across all polarity/neutral states

Ranked causes

  1. 1. Insulation failure in patient circuit (rare)
  2. 2. Contamination or moisture in patient connections (occasional)
First check

Measure earth leakage (Neutral Closed, Normal polarity)

Test point
Earth connection path
Expected
< 100 µA
Tools
Leakage current tester

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG gain out of tolerance (3 or 5-lead)

ECG signal amplitude does not meet gain specifications on lead inputs

Ranked causes

  1. 1. ECG simulator or cable issue (common)
  2. 2. A02 Patient Parameter PCB failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG lead leakage current elevated patient safety

Leakage current between ECG leads or to ground exceeds safe limits across all test configurations

Ranked causes

  1. 1. ECG cable insulation damaged or wet (occasional)
  2. 2. ECG input isolation circuit failure (rare)
First check

Measure lead-to-ground leakage (Neutral Closed, Normal, Lead-Gnd Normal)

Test point
ECG lead connections
Expected
< 50 µA
Tools
Leakage current tester

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG lead signal amplitude out of range

5-lead ECG gain does not meet specification during acquisition

Ranked causes

  1. 1. Loose or damaged ECG cable (common)
  2. 2. ECG preamplifier circuit fault (rare)
First check

Record 5-lead ECG gain amplitude

Test point
ECG lead input
Expected
36 to 44 mm, 36 to 44 mm

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG signal has no amplitude

ECG trace displays flat line or zero amplitude signal

Ranked causes

  1. 1. ECG cable disconnected or damaged (common)
  2. 2. A02 Patient Parameter PCB failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fails ground resistance test high voltage

Unit fails ground continuity or grounding resistance safety test

Ranked causes

  1. 1. Power cord grounding issue (common)
  2. 2. Power module grounding circuit failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Frozen at power-on screen caution

Device stops at startup screen

Ranked causes

  1. 1. W18 UI Flex Cable defective (unknown)
  2. 2. A01 System PCB faulty (unknown)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ground resistance out of specification high voltage

Measured ground resistance does not meet safety requirements

Ranked causes

  1. 1. Loose or corroded ground connection (common)
  2. 2. Ground conductor broken or high-resistance path (occasional)
First check

Measure DC ground resistance

Test point
Chassis ground point to mains earth
Expected
< 0.1 ohms (typical for medical equipment)
Tools
Ohmmeter with 4-wire capability

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

HARD PADDLE key not responding patient safety

HARD PADDLE key does not respond when pressed

Ranked causes

  1. 1. Hard paddles damaged or connector issue (common)
  2. 2. A04 Therapy PCB failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad buttons not responding properly

Buttons do not respond to presses or respond inconsistently

Ranked causes

  1. 1. Elastomer keypad worn or damaged (common)
  2. 2. A05 UI PCB button circuit failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loose or broken hardware caution

Loose or broken components on the device

Ranked causes

  1. 1. Loose hardware (common)
  2. 2. Broken components (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low patient impedance

Patient impedance measurement lower than expected

Ranked causes

  1. 1. Incorrect test load used (common)
  2. 2. A01 System PCB faulty (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Low patient impedance during test

Patient impedance measurement is lower than expected during testing

Ranked causes

  1. 1. Wrong test plug used or test procedure issue (common)
  2. 2. A01 System PCB impedance measurement failure (occasional)
First check

Test into 50 ohm load during 3 A.M. test

Test point
Test load
Expected
Correct test plug used
Tools
50 ohm test load

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Maintenance Due message remains on screen

Maintenance alert persists on display

Ranked causes

  1. 1. Maintenance prompt interval not set (common)
  2. 2. Maintenance prompt interval needs reset (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No ac power reminder tone caution

AC power availability alert tone not sounding

Ranked causes

  1. 1. User configuration setting incorrect (common)
  2. 2. A03 Power module faulty (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • No AC power reminder tone or alert
  • No amplitude ECG
  • No backup battery operation
  • No display (white or blue screen)
  • Pacer output current out of tolerance
  • Printer characters are missing or broken
  • Printer malfunction or not installed
  • Printer test strip defects
  • QUIK-COMBO ECG analog output missing or out of tolerance
  • QUIK-COMBO ECG gain out of tolerance
  • Remote sync function not responding to synchronization signal
  • Remote sync LED not flashing
  • Service code logged, service LED on
  • Speed Dial not functioning
  • SpO2 leakage current elevated
  • SpO2 probe signal not detected or reading out of range
  • SpO2 reading out of specification
  • Standard Paddles ECG gain out of tolerance
  • Therapy delivered energy out of tolerance
  • Therapy ECG characteristic amplitude out of range
  • Therapy impedance sense malfunction
  • Therapy impedance sense not functioning
  • Therapy leakage current elevated
  • Unit frozen at power on screen
  • User test does not complete or report not printed
  • AC Mains indicator not on
  • AC power LED on LIFEPAK 20 device not illuminated
  • Accessory cables or sensors expired or unsuitable
  • Calibration: Audio test
  • Calibration: Chassis Leakage Current
  • Calibration: CO2 measurement
  • Calibration: Defibrillation
  • Calibration: Defibrillator isolation
  • Calibration: Earth leakage current normal condition
  • Calibration: Earth leakage current single fault condition
  • Calibration: Ground Resistance
  • Calibration: Keypad test
  • Calibration: LIFEPAK 20 TCP setup
  • Calibration: LIFEPAK 20/20e instrument calibration
  • Calibration: Pacer Leads Off Detection
  • Calibration: Pacer Option Characteristics
  • Calibration: Pacer Output Current
  • Calibration: Performance Inspection Procedure
  • Calibration: Performance Inspection Procedure
  • Calibration: Pixels test
  • Calibration: Printer test
  • Calibration: Protective earth resistance CodeManagement Module ground stud
  • Calibration: Protective earth resistance defibrillator ground stud
  • Calibration: Reset Maintenance Prompt
  • Calibration: Service Mode Passcode
  • Calibration: Standard Paddles Defibrillation Isolation
  • Calibration: Standard Paddles Delivered Energy
  • Calibration: Standard Paddles ECG Characteristics
  • Calibration: Standard Paddles Remote Synchronous Cardioversion
  • Calibration: Standard Paddles Synchronous Cardioversion R-Wave
  • Calibration: Standard Paddles User Test
  • Calibration: TCP - Resource Requirements
  • Calibration: TCP Defibrillator
  • Calibration: TCP Defibrillator Isolation
  • Calibration: TCP delivered energy
  • Calibration: TCP setup
  • Carrying strap or mounts damaged or missing
  • CHECK PRINTER message does not appear
  • Connectors damaged or contaminated
  • Continuous reset
  • Contrast Test
  • Date and time incorrect
  • Defibrillation capacitor charge/dump operation fails
  • Defibrillator capacitor charge/dump cycle fails during USER TEST
  • Defibrillator isolation failure
  • Delivered energy at 10 joules out of range
  • Delivered energy at 100 joules out of range
  • Delivered energy at 15 joules out of range
  • Delivered energy at 2 joules out of range
  • Delivered energy at 360 joules out of range
  • Delivered energy at 50 joules out of range
  • Delivered energy at 70 joules out of range
  • Delivered energy out of tolerance
  • Device beeps when turned off
  • Device exterior damage or wear
  • Dirt, fluids, or foreign objects on device
  • Dirt, fluids, or foreign objects present
  • Display does not illuminate
  • Display is distorted
  • Display pixel anomalies
  • Distorted display
  • ECG analog output amplitude out of range
  • ECG leads not detected
  • ECG signal amplitude out of specification
  • ECG signal is saturated
  • Fails chassis leakage current test
  • Fails chassis leakage test
  • Hidden switch between OPTIONS and PAUSE not functioning
  • Keypad button not responding
  • Keypad controls unresponsive
  • Keypad or overlays damaged
  • LCD contrast insufficient
  • LIFEPAK 20 Device Data Checksum Error
  • LIFEPAK 20 Device Serial Port Communication Failure
  • Loose or rattling hardware inside device
  • MAINTENANCE DUE indicator remains on
  • No CO2 Operation displayed on monitor
  • No wireless connection to LIFENET System
  • Non-volatile RAM Integrity Test Failure
  • Pacer leads off detection failed
  • Pacer pulse width out of specification
  • Pacing current or rate out of tolerance
  • Pacing current/rate out of tolerance
  • Paddles ECG signal amplitude out of specification
  • Paddles require cleaning
  • Pixel Test
  • Power on sequence incomplete
  • Real Time Clock Failure
  • Remote sync energy transfer failed
  • Rubber feet damaged or missing
  • Service indicator illuminated
  • Service indicator LED illuminates after power-on
  • Service indicator LED illuminates during self-test
  • SERVICE indicator lights during self-test
  • SERVICE indicator remains on
  • Synchronization delay excessive
  • Therapy delivered energy out of specification
  • Therapy user test failed
  • Unit continuously resets
  • User Test Failed
  • Voice prompts inaudible or distorted

126 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0002System Flash memory voltage errorFlash memory accesserror
0003Cannot erase system Flash memoryMemory erase operationerror
0004Cannot write to system Flash memoryMemory write operationerror
0006System ADC read errorADC read operationerror
0007System DAC not respondingDAC access during operationerror
0008ECG OUT DAC self-test failedECG output DAC self-test failscritical
000aSystem ADC background test failedAnalog-to-digital converter self-test failscritical
000bSystem ADC failed self-calibrationADC self-calibration procedure failscritical
000cSystem Flash memory ID unknownFlash memory chip identification failscritical
000dSystem hardware/software configuration lostConfiguration data becomes unavailablecritical
01Real Time Clock FailureReal Time Clock failure detected by CodeManagement Module self-testwarning
02Non-volatile RAM Integrity Test FailureNon-volatile RAM integrity test failure detectedwarning
03LIFEPAK 20 Device Data Checksum ErrorData checksum error detected in LIFEPAK 20 deviceerror
04LIFEPAK 20 Device Serial Port Communication FailureSerial port communication failure between CodeManagement Module and LIFEPAK 20error
100eUI – system communication lostUI-System communication failurecritical
100fDisplay update timer errorDisplay update timer malfunctionserror
1010Display update queue errorDisplay update queue processing failserror
1013System detected unexpected UI resetUI resets unexpectedly during operationerror
1014Voice prompt/audio watchdog failureAudio system watchdog detects failurewarning
1015USB data time-out errorUSB data transfer exceeds time limiterror
1016UI USB data errorUSB data corruption detected on UI interfaceerror
1c01UI FPGA programming errorField programmable gate array programming failserror
1c02UI FPGA verification errorFPGA verification fails after programmingerror
1c03UI FPGA program file errorFPGA program file is corrupted or missingerror
1c04UI ADC not functioningUser interface ADC converter failserror
1c05Printer ADC out of tolerance (temperature)Printer temperature sensor reading out of specwarning
1c06Printer ADC out of tolerance (voltage)Printer voltage sensor ADC reading exceeds limitswarning
1c07UI voltage out of tolerance (5 V)5V power supply reading out of specerror
1c08UI voltage out of tolerance (3.3 V)3.3V supply on UI PCB exceeds toleranceerror
1c09UI voltage out of tolerance (2.5 V)2.5V supply on UI PCB exceeds toleranceerror
1c0aUI voltage out of tolerance (35 V)35V supply on UI PCB exceeds toleranceerror
1c0bUI voltage out of tolerance (SW VBATT)Switched battery voltage on UI PCB exceeds toleranceerror
1c0cUI voltage out of tolerance (ground)Ground reference voltage on UI PCB out of rangeerror
1c0fUI hardware I.D. corruptedHardware identification data on UI PCB becomes corruptedcritical
1c10UI boot program corruptedUI boot program becomes corruptedcritical
1c11UI application Flash memory corruptedUI application code in Flash memory corruptedcritical
1c12UI font Flash memory corruptedFont data in UI Flash memory corruptedcritical
1c13UI FPGA Flash memory corruptedFPGA program in Flash memory corruptedcritical
1c16UI CPU RAM test failed on power-onUI RAM self-test fails during power-oncritical
1c17UI CPU RAM test failed during normal operationUI RAM test fails during normal operationcritical
1fffAdditional information related to prior error codeLogged before another error code for diagnostic contextinfo
2004System cannot initialize serial port (system connector)Serial port initialization failserror
2005System cannot initialize driver for serial portSerial port driver fails to initializeerror
3001System cannot use data management Flash memoryData management Flash memory becomes unavailableerror
3002System data management Flash memory corruptedData management Flash memory data becomes corruptedcritical
3005System cannot delete DM recordData management record deletion failserror
3007System cannot create new DM recordData management record creation failserror
3008System could not store DM recordData management record storage failserror
3009System could not erase oldest DM recordCannot erase oldest data management record when memory fullerror
300aSystem cannot clear DM memoryData management memory clear operation failserror
300bSystem error writing DM recordData management record write operation failserror
300cSystem cannot read archived DM recordCannot read archived data management recorderror
3010System error DM memory corruptData management memory corruption detectedcritical
3fffAdditional information related to prior error codeLogged before another error code for diagnostic contextinfo
4005System NVRAM error logNon-volatile RAM error log becomes corruptederror
4006Error log queue not functioningError logging queue malfunctionserror
4008Error log count corruptedError log counter becomes corruptederror
4009System RAM test failedSystem RAM self-test failscritical
400aSystem program Flash memory corruptedSystem program code in Flash memory corruptedcritical
400bProgram contents failureSystem program code verification failscritical
400cSystem ADC voltage lowSystem power supply voltage falls below minimumwarning
400dSystem ADC voltage highSystem power supply voltage exceeds maximumwarning
4010Service LED failedService indicator LED fails to illuminatewarning
4012Voice Flash memory corruptedVoice prompts in Flash memory corruptedwarning
4013Voice/font Flash memory invalidVoice and font data in Flash memory invalidwarning
4fffAdditional information related to prior error codeLogged before another error code for diagnostic contextinfo
5001System watchdog failedWatchdog timer failurecritical
5003System watchdog failedWatchdog timer failurecritical
5004System watchdog resetSystem watchdog forces reseterror
5005System CPU error during boot-upCPU encounters error during startupcritical
5006System RAM failed during boot-upRAM self-test fails during startupcritical
5007System checksum failure during boot-upBoot code checksum verification failscritical
5008System boot program failureBoot program execution failscritical
5009Real-time clock (RTC) access failedRTC becomes inaccessibleerror
500aCoin cell battery not detectedBackup battery missing or connection lostwarning
500bCannot use system NVRAMNon-volatile RAM becomes unavailableerror
500cSystem application start errorSystem application fails to starterror
500dSystem software initialization time outSystem initialization exceeds time limiterror
500eSystem application start errorSystem application fails to starterror
5010Configuration mismatchSystem configuration inconsistentwarning
5011NVRAM configuration data errorConfiguration data in NVRAM corruptederror
5012Configuration data errorConfiguration data becomes corruptederror
5013System meter initialization errorMeter initialization failswarning
5014System meter mismatchMeter configuration mismatch detectedwarning
5015NVRAM errorNon-volatile RAM errorerror
5016MFG data mismatchManufacturing data inconsistencyerror
5017NVRAM MFG data lostManufacturing data in NVRAM lostcritical
5018Watchdog reset failedWatchdog reset operation failscritical
5019NVRAM corruptedNon-volatile RAM data corruptedcritical
501aNVRAM corruptedNon-volatile RAM data corruptedcritical
501cRTC not runningReal-time clock stops runningwarning
501dRTC out of syncReal-time clock time synchronization lostwarning
501eSystem software execution errorSoftware execution error occurserror
501fSystem software read errorSoftware read operation failserror
5020System software write errorSoftware write operation failserror
5021System software errorSystem software encounters errorerror
5022System software exception codeSoftware exception occurserror
5023System software exception codeSoftware exception occurserror
5024System software exception codeSoftware exception occurserror
5026NVRAM low battery interruptBackup battery voltage falls below thresholdwarning
5027System USB did not initializeSystem USB interface fails to initializeerror
5028PP USB did not initializePatient module USB interface fails to initializeerror
5029Therapy USB did not initializeerror
502aUI USB did not initializeerror
502bUSB system failederror
502fDevice type invaliderror
5030PP USB disconnecterror
5031UI USB disconnecterror
5032Therapy USB disconnectUSB connection lost between therapy module and systemerror
5033PP USB download time outPP module fails to complete USB download within timeout windowerror
5036USB initialization failedUSB subsystem fails to initializeerror
5037USB unitization errorUSB unit fails to initialize properlyerror
5038Cannot clear interrupt sourceInterrupt source cannot be cleared during shutdown or reseterror
5039Bus error reported by USB PP channelPP module reports USB bus errorerror
503aBus error reported by USB UI channelUI module reports USB bus errorerror
503bBus error reported by USB therapy channelTherapy module reports USB bus errorerror
503cUSB host driver cannot create transfer descriptorUSB host driver fails to allocate transfer descriptorerror
503dUSB host driver received message larger than maximum allowedReceived USB message exceeds size limiterror
5040Inconsistent RTC timeReal-time clock value is inconsistent or corruptedwarning
5105Battery failed to reach charge in 2.5 hourserror
5106Power supply out of tolerancePower supply voltage outside acceptable rangeerror
5107Power module RAM errorRAM test failure in power moduleerror
5108Power module self-test service codePower module self-test failure detectederror
510aBattery <10volts after 20minutes of chargingBattery charge cycle fails to reach minimum voltagewarning
510bDoes not switch to battery powerDevice fails to switch to battery when AC power is removedcritical
510cDoes not detect ac disconnectAC disconnect detection circuit failurecritical
510dBattery powered when connected to ac powerDevice incorrectly uses battery while AC is connectederror
510eBattery <10volts after charge cycleBattery voltage remains low after chargingwarning
510fBattery charge cycle stoppedCharging process stopped prematurelyerror
5110AC isolation Diode Shortederror
5112Battery not detectedBattery presence or communication not detectedcritical
5115Battery thermistor <400 OhmsBattery thermistor resistance below 400 ohmserror
5116Charger reporting zero during battery chargingCharger circuit reports zero output during chargingerror
5117No 12C connection detectedI2C communication link lost (LIFEPAK 20e only)error
5118Power module connected to NiMH batteryPower module for lithium battery connected to NiMH battery (LIFEPAK 20 only)error
511aBattery failureBattery has reached end of life or is defectiveerror
5fffAdditional information related to prior error codeinfo
6002PP program corruptedPulsepoint processor program memory corruptederror
6003PP program not foundPulsepoint processor program not present in memoryerror
6004PP boot-up errorPulsepoint processor fails to booterror
6009No PP dataerror
600cSpO2 misconfiguredSpO2 module configuration is incorrectwarning
600ePP resetPulsepoint processor reset unexpectedlywarning
600fOEM configuration errorOEM-specific configuration is invalidwarning
6010PP initialization errorPulsepoint processor fails to initializeerror
6018Incorrect updated SpO2 imageSpO2 firmware image update is incorrect or corruptederror
6019Incorrect updated SpO2 image CRCSpO2 firmware image CRC check failed after updateerror
6801PP power supply out of tolerancePulsepoint processor power supply voltage out of rangeerror
6802PP pre-amp data invalidPulsepoint processor pre-amp output invaliderror
6804PP data RAM test errorPulsepoint processor data RAM test failureerror
6805PP RAM test errorPulsepoint processor RAM test failureerror
6806PP CRC test errorPulsepoint processor program CRC check failureerror
6807PP ECG test errorPulsepoint processor ECG test failureerror
680bSpO2 board errorSpO2 module hardware failure detectederror
6fffAdditional information related to prior error codeinfo
800aSystem DSP errorDigital signal processor firmware errorerror
8013Voice format errorVoice message format is invalid or corruptedwarning
8014No paddles dataPaddles firmware or data not availableerror
801bDSP did not receive USB SOF interruptDSP fails to receive USB start-of-frame interrupterror
8105Impedance channel out of calibrationImpedance measurement circuit calibration losterror
8108Paddles data out of syncPaddles firmware synchronization losterror
8109Paddles pre-amp user test failedPaddles pre-amp self-test failureerror
810bReal impedance <-30 ohms for one secondMeasured impedance is negative or out of rangeerror
8fffAdditional information related to prior error codeinfo
9004Unable to initialize therapy controlTherapy control circuit fails to initializecritical
9005Defib disabledDefibrillation circuit disabled by firmwarecritical
9007Shock not deliveredDefibrillation shock failed to delivercritical
9009Defib charge time expiredcritical
900bPacing rate out of tolerancePacing rate outside acceptable rangecritical
900fUnable to initialize therapy controlTherapy control stack initialization failurecritical
9011Pacer faultPacemaker circuit hardware failurecritical
9017Pacer disabledPacemaker circuit disabledwarning
901aPacer rate storage corruptedPacemaker rate data memory corruptederror
901bTherapy PCB communication lostCommunication link to therapy PCB lostcritical
901dTesting purpose onlyReserved test codeinfo
9021Impedance value indicates regulator failure in System PCB pre-ampSystem PCB pre-amp regulator detected as failederror
9c03Therapy processor unplanned resetTherapy processor reset unexpectedlyerror
9c04Therapy/system controller communication watchdogCommunication watchdog timeout between therapy and system controllerscritical
9c05CRC errorTherapy processor program CRC check failederror
9c06Calibration constant A out of rangeTherapy calibration constant A is outside acceptable limitserror
9c07Calibration constant B out of rangeTherapy calibration constant B is outside acceptable limitserror
9C08Therapy ROM Cyclic Redundancy Check (CRC) failederror
9c09Therapy RAM pattern write test failedTherapy processor RAM write test failureerror
9c0aTherapy relay idle coil voltage out of rangeTherapy relay idle state coil voltage out of specerror
9c0bTherapy relay enabled coil voltage out of range (5 ms)Therapy relay enabled state coil voltage at 5ms out of specerror
9c0cTherapy relay enabled coil voltage out of range (100 ms)Therapy relay enabled state coil voltage at 100ms out of specerror
9c0dTherapy relay drive enabled coil voltage out of rangeTherapy relay drive enable coil voltage out of specerror
9c0fTherapy/power assembly communication errorCommunication lost between therapy and power modulescritical
9c11Capacitor Dump failedHigh voltage capacitor discharge failedcritical
9C12Therapy pcba 5 volt out of rangeerror
9C13Therapy pcba 15 volt out of rangeerror
9c14Therapy pcba -15 volt out of rangeerror
9c15Therapy ADC time out errorAnalog-to-digital converter response timeouterror
9C183AM H bridge test: NE leg shortederror
9C193AM H bridge test: SE leg shortederror
9C1a3AM H bridge test: NW leg shortederror
9C1b3AM H bridge test: SW leg shortederror
9C1c3AM H bridge test: East side stuck openerror
9C1d3AM H bridge test: West side stuck openerror
9C1e3AM H bridge test: Charge time out of rangeerror
9c1f3:00 AM shorted paddles relay contact test: relay shortederror
9c20Shorted paddles relay contact test failure3:00 AM shorted paddles relay contact test detected relay shortederror
9c21Pace drive power supply stuck on3:00 AM pace drive test detected power supply stuck in on stateerror
9c22Pace drive power supply inoperable3:00 AM pace drive test found power supply not functioningerror
9c23Relay contacts shorted in pace drive3:00 AM pace drive test detected shorted relay contactserror
9c24Relay drive low side shorted3:00 AM pace drive test detected low side short circuiterror
9c25Relay drive high side shorted3:00 AM pace drive test detected high side short circuiterror
9c26Pace FET shorted3:00 AM pace drive test detected shorted field effect transistorerror
9c27Pace current path open3:00 AM pace drive test detected open circuit in current patherror
9c28Pace set point error3:00 AM pace drive test detected incorrect pace set point valueerror
9c29Redundant controls stuck on3:00 AM redundant controls test detected controls stuck in on stateerror
9c2aRedundant controls enable 2 stuck on3:00 AM redundant controls test detected enable 2 stuck onerror
9c2bRedundant controls enable 1 stuck on3:00 AM redundant controls test detected enable 1 stuck onerror
9c2cCapacitor voltage per pacing pulse too highPacing pulse capacitor voltage exceeds specificationerror
9c2dCapacitor current per pacing pulse too highPacing pulse capacitor current exceeds specificationerror
9c2eCap current per pacing pulse too highPacing pulse capacitor current exceeds specificationerror
9c2fPacing current and selected current out of rangeMeasured pacing current does not match programmed valueerror
9c30Pacing pulse width too shortPacing pulse width is shorter than specificationerror
9c31Pacing pulse width too longPacing pulse width is longer than specificationerror
9c32Capacitor voltage and predicted capacitor voltage mismatchMeasured capacitor voltage does not match predicted valueerror
9C33Capacitor charge time >25secondserror
9c35Therapy CPU instruction test failedCPU self-test detected instruction execution errorerror
9c36Therapy software stack overflowSoftware stack memory exceeded during operationerror
9c3aEnergy capacitor overvoltage errorEnergy capacitor voltage exceeds maximum safe levelcritical
9c3bRedundant controls charge rate stuck on3:00 AM redundant controls test detected charge rate control stuck onerror
9c3eTherapy software errorSoftware detected an unspecified error conditionerror
9c3fTherapy software errorSoftware detected an unspecified error conditionerror
A00bPrinter communication lostPrinter interface failureerror
a00ePrinter initialization errorPrinter fails to initialize during startuperror
b001Invalid state requestSystem receives invalid state commanderror
b011System behavior manager errorSystem behavior manager encounters error conditionerror
b012Energy cap charging time outDefibrillator capacitor fails to charge within timeout periodwarning
B013Shock advisory system errorAlgorithm or sensor failurecritical
b014Shock advisory system time outShock advisory analysis exceeds time limiterror
b016Motion detect timer errorMotion detection timer malfunctionswarning
b017Shock result time outShock delivery monitoring exceeds time limiterror
b018USB interrupt errorUSB interrupt handling failserror
MAINTENANCE DUEScheduled maintenance interval has passed and maintenance is dueDevice powered on after scheduled maintenance interval has expiredwarning
PASSCODE INCORRECT - TRY AGAINService mode passcode entry was invalidIncorrect service mode passcode entered when accessing SERVICE menuwarning
Service Log entrySelf-test failure detectedErrors detected during power-on or continuous self-test cycleserror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.