Physio-Control LIFEPAK 20: troubleshooting and error codes
The ServiceForge knowledge base holds 186 documented symptoms for the Physio-Control LIFEPAK 20, with 330 ranked causes, 230 diagnostic checks with test points and expected values, and 240 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Not printing caution
Printer does not produce output
Ranked causes
- 1. Printer test failure (common)
- 2. Paper not loaded or improper (common)
- 3. Printer power supply issue (occasional)
- 4. W14 Printer Flex Cable shorted (occasional)
- 5. A12 Printer Assembly defective (occasional)
- 6. A05 UI PCB faulty (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No lead channel sync mark patient safety
Synchronous cardioversion ECG lead sync mark not displayed
Ranked causes
- 1. Noisy ECG signal (common)
- 2. ECG cable defective (common)
- 3. W06 ECG connector defective (common)
- 4. A02 Patient Parameter PCB faulty (occasional)
- 5. A01 System PCB faulty (unknown)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No lead channel Sync mark displayed patient safety
Synchronous cardioversion Sync mark not displayed on ECG lead channel
Ranked causes
- 1. Noisy ECG signal (common)
- 2. Patient cable issue (occasional)
- 3. W06 ECG connector issue (occasional)
- 4. A02 Patient Parameter PCB failure (occasional)
- 5. A01 System PCB sync circuit failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No remote sync patient safety
Remote synchronization function not working
Ranked causes
- 1. Remote sync function disabled (common)
- 2. ECG Out/Sync In connector defective (common)
- 3. W09 Power Cable defective (occasional)
- 4. A01 System PCB faulty (occasional)
- 5. A03 Power Module faulty (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No remote Sync available
Remote sync function does not work; external sync signal not accepted
Ranked causes
- 1. Remote sync function disabled (common)
- 2. ECG Out/Sync In connector or power module issue (occasional)
- 3. W09 Power Cable issue (occasional)
- 4. A01 System PCB sync circuit failure (rare)
- 5. Power module sync circuit failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to complete auto test or user test patient safety
Device cannot successfully run built-in diagnostic tests
Ranked causes
- 1. Improper test load connection (common)
- 2. Test plug or shorting bar continuity issue (common)
- 3. QUIK-COMBO cable continuity issue (common)
- 4. W01 Therapy Connector defective (occasional)
- 5. A04 Therapy PCB faulty (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Damaged keypad or labels
Keypad or labels are damaged or worn
Ranked causes
- 1. Damaged elastomer keypad (common)
- 2. Damaged product identification label (occasional)
- 3. Damaged explosion/hazard label (occasional)
- 4. Damaged operating instruction label (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hard paddle buttons patient safety
Defibrillator paddle buttons are stiff
Ranked causes
- 1. Paddle button mechanical issue (common)
- 2. Hard paddles defective (common)
- 3. W01 Therapy Connector defective (unknown)
- 4. A04 Therapy PCB faulty (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaudible or garbled audio patient safety
Speaker output absent, quiet, or unintelligible
Ranked causes
- 1. Audio test failure (common)
- 2. Speaker connection loose (common)
- 3. W02 Speaker Assembly defective (common)
- 4. A01 System PCB faulty (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pacer does not turn on patient safety
Pacing mode cannot be activated or does not operate
Ranked causes
- 1. Pacer not enabled or keypad issue (common)
- 2. Elastomer keypad defective (occasional)
- 3. A05 UI PCB failure (occasional)
- 4. Pacer control circuit failure or error code (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer does not print
Printer produces no output, no paper movement, or no marks on paper
Ranked causes
- 1. No paper or paper improperly loaded (common)
- 2. W14 Printer Flex Cable voltage high (occasional)
- 3. A12 Printer Assembly failure (occasional)
- 4. A05 UI PCB printer control failure (rare)
Run printer test
- Test point
- Printer
- Expected
- Printer test passes
- Tools
- Run diagnostic test
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Saturation reading missing or out of tolerance patient safety
SpO2 saturation reading absent or outside normal range
Ranked causes
- 1. SpO2 cable defective or disconnected (common)
- 2. SpO2 sensor defective (common)
- 3. W05 SpO2 Assembly defective (occasional)
- 4. A06 OEM PCB faulty (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unable to complete energy delivery auto test high voltage
Defibrillator cannot complete automatic delivered energy test
Ranked causes
- 1. Test load shorting bar or plug continuity issue (common)
- 2. QUIK-COMBO cable continuity issue (occasional)
- 3. W01 Therapy Connector issue (occasional)
- 4. A04 Therapy PCB failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal energy delivery patient safety
Defibrillator delivers unexpected energy
Ranked causes
- 1. A14 Inductive Resistor faulty (common)
- 2. A04 Therapy PCB faulty (occasional)
- 3. A01 System PCB faulty (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal energy delivery (patient impedance channel broken) patient safety
Defibrillator delivers abnormal or incorrect shock energy to patient
Ranked causes
- 1. A14 Inductive Resistor open or high resistance (common)
- 2. A04 Therapy PCB failure (occasional)
- 3. A01 System PCB failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Computer-assisted module tests
Serial communication, USB interface, battery power, battery I2C, wireless and wireless switch functionality verification
Ranked causes
- 1. DUT does not communicate with test computer (occasional)
- 2. Serial to USB adapter misconfiguration or incompatibility (occasional)
- 3. Weak wireless signal interference (occasional)
Verify wireless network availability
- Test point
- Test environment
- Expected
- Operational WPA/WPA2 enterprise or personal signal broadcasted
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is blank
Display is completely blank with no backlighting or text visible
Ranked causes
- 1. Display backlight/display assembly failure (common)
- 2. A04 Therapy PCB voltage issue (occasional)
- 3. Power module failure (rare)
Measure 12V on P21 to ground
- Test point
- P21 connector
- Expected
- 12 V to ground
- Tools
- DMM
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display on but power on LED not illuminated
Display functions but power on indicator LED is not lit
Ranked causes
- 1. UI Flex Cable issue (common)
- 2. A05 UI PCB failure (occasional)
- 3. A04 Therapy PCB failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display on, no power-on LED
Display shows information but power-on LED not lit
Ranked causes
- 1. W18 UI Flex Cable defective (unknown)
- 2. A05 UI PCB faulty (unknown)
- 3. A04 Therapy PCB faulty (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows white or blue screen
Display is on but shows white or blue screen instead of normal operation
Ranked causes
- 1. A05 UI PCB voltage issue (common)
- 2. W18 UI Flex Cable disconnected (occasional)
- 3. A01 System PCB failure (occasional)
Measure voltage on A05 UI PCB
- Test point
- C96 connector
- Expected
- 3.3 V to ground
- Tools
- DMM
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG analog out missing or out of tolerance patient safety
ECG analog output signal absent or outside spec
Ranked causes
- 1. Simulator output verification needed (common)
- 2. ECG display not showing signal (common)
- 3. W11 ECG out connector defective (common)
ECG gain out of tolerance patient safety
ECG signal amplification outside specifications
Ranked causes
- 1. Simulator output verification needed (common)
- 2. ECG cable defective (common)
- 3. A02 Patient Parameter PCB faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to transfer coincident with Sync mark patient safety
Defibrillator does not deliver therapy when Sync mark aligns properly with R-wave
Ranked causes
- 1. Sync marker not properly placed on R-wave (common)
- 2. Keypad button issue or UI PCB failure (occasional)
- 3. User test failure or error code (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper button response caution
Buttons do not respond as expected
Ranked causes
- 1. Keypad malfunction (common)
- 2. Elastomer keypad defective (common)
- 3. A05 UI PCB faulty (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Light print caution
Printer output is faint
Ranked causes
- 1. Improper paper type (common)
- 2. W14 Printer Flex Cable disconnected (common)
- 3. A05 UI PCB faulty (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing or broken characters in print caution
Printed output has missing or illegible characters
Ranked causes
- 1. Improper paper type (common)
- 2. Printhead dirty or clogged (common)
- 3. A12 Printer Assembly defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG patient safety
No ECG signal displayed
Ranked causes
- 1. ECG cable defective (common)
- 2. W06 ECG Connector defective (common)
- 3. A02 Patient Parameter PCB faulty (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG signal patient safety
ECG waveform not displayed; no cardiac rhythm detected
Ranked causes
- 1. ECG cable disconnected or failed (common)
- 2. ECG connector failure (occasional)
- 3. Patient Parameter PCB failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG signal displayed
ECG trace is not visible on display or no waveform is present
Ranked causes
- 1. ECG cable disconnected or damaged (common)
- 2. W06 ECG connector issue (occasional)
- 3. A02 Patient Parameter PCB failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No paddles channel sync mark patient safety
Synchronous cardioversion paddle sync mark not displayed
Ranked causes
- 1. Therapy cable defective (common)
- 2. W01 Therapy Connector defective (common)
- 3. A01 System PCB faulty (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No paddles channel Sync mark displayed patient safety
Synchronous cardioversion Sync mark not displayed on paddles channel
Ranked causes
- 1. Therapy cable issue (common)
- 2. W01 Therapy Connector issue (occasional)
- 3. A01 System PCB sync circuit failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power on patient safety
Device does not power up when turned on
Ranked causes
- 1. AC power not connected (common)
- 2. Therapy PCB failure (occasional)
- 3. Power module failure (occasional)
Verify AC power connection
- Test point
- Power cord and outlet
- Expected
- Device plugged in and powered
- Tools
- None
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 response (no cable detected)
SpO2 sensor not detected; no pulse oximetry signal available
Ranked causes
- 1. SpO2 cable disconnected or damaged (common)
- 2. SpO2 sensor defective (occasional)
- 3. W05 SpO2 Assembly issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer output is light
Printed output is faint, light, or barely visible
Ranked causes
- 1. Wrong paper type used (common)
- 2. W14 Printer Flex Cable connection issue (occasional)
- 3. A05 UI PCB printhead control issue (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 saturation reading missing or out of tolerance
SpO2 value not displayed or reading is outside expected physiological range
Ranked causes
- 1. SpO2 cable or sensor issue (common)
- 2. W05 SpO2 Assembly failure (occasional)
- 3. A06 OEM PCB failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Therapy cable leads off high voltage
QUIK-COMBO therapy cable not properly connected
Ranked causes
- 1. QUIK-COMBO cable continuity broken (common)
- 2. W01 Therapy Connector defective (common)
- 3. A01 System PCB faulty (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Therapy cable leads off (QUIK-COMBO only) patient safety
QUIK-COMBO therapy cable disconnects or lead continuity is broken
Ranked causes
- 1. QUIK-COMBO cable continuity broken (common)
- 2. W01 Therapy Connector disconnected (occasional)
- 3. A01 System PCB therapy circuit failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Therapy ECG gain out of tolerance patient safety
Therapy paddle ECG signal amplification outside spec
Ranked causes
- 1. Simulator output verification needed (common)
- 2. Therapy cable defective (common)
- 3. A01 System PCB faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on high voltage
Defibrillator does not turn on when power button is pressed
Ranked causes
- 1. Not connected to AC power or UI power issue (common)
- 2. UI flex cable issue (occasional)
- 3. Power module failure (occasional)
Verify AC power connection
- Test point
- AC outlet
- Expected
- Power connected
- Tools
- Visual inspection
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AED mode not functioning patient safety
AED mode does not activate, voice prompts inaudible, or device fails to switch to manual mode when energy select pressed
Ranked causes
- 1. Hard paddles not properly installed or analyzer not connected (common)
- 2. Device speaker or voice prompt system malfunction (occasional)
Connect therapy cable/paddles and press ANALYZE key to activate AED mode
- Test point
- Device ANALYZE key, display
- Expected
- AED mode activated
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio is inaudible or garbled
Speaker produces no sound or produces garbled, distorted sound
Ranked causes
- 1. Speaker disconnected or defective (common)
- 2. A01 System PCB audio circuit failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio output distorted or silent patient safety
Voice messages or tone alerts are not clear or not audible
Ranked causes
- 1. Speaker damage or audio circuit fault (rare)
- 2. Audio mute setting enabled (occasional)
Trigger voice prompt and tone sequences
- Test point
- Speaker output
- Expected
- Clear, undistorted voice and tone messages
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Backup battery does not operate caution
Unit cannot run on backup battery; device powers off when AC is disconnected
Ranked causes
- 1. A07 Battery dead or missing (common)
- 2. Power module charging circuit failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Maintenance prompt interval
Set or reset the scheduled preventative maintenance prompt interval
Ranked causes
- 1. Configure maintenance reminder schedule (unknown)
- 2. Reset maintenance interval after performing maintenance (common)
Display MAINT PROMPT menu and select INTERVAL option
- Test point
- Service screen - MAINT PROMPT option
- Expected
- Menu shows OFF, 3 MONTHS, 6 MONTHS, 12 MONTHS choices
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Chassis leakage current elevated high voltage
Leakage current from chassis to ground exceeds safe limits in any polarity/neutral configuration
Ranked causes
- 1. Insulation breakdown in power supply (rare)
- 2. Wet or contaminated PCB (occasional)
Measure leakage current (Neutral Closed, Normal polarity, Lead-Chassis Normal)
- Test point
- Leakage path simulation
- Expected
- < 100 µA (medical equipment standard)
- Tools
- Leakage current tester
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Earth leakage current elevated high voltage
Leakage current from device to earth exceeds safe limits across all polarity/neutral states
Ranked causes
- 1. Insulation failure in patient circuit (rare)
- 2. Contamination or moisture in patient connections (occasional)
Measure earth leakage (Neutral Closed, Normal polarity)
- Test point
- Earth connection path
- Expected
- < 100 µA
- Tools
- Leakage current tester
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG gain out of tolerance (3 or 5-lead)
ECG signal amplitude does not meet gain specifications on lead inputs
Ranked causes
- 1. ECG simulator or cable issue (common)
- 2. A02 Patient Parameter PCB failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG lead leakage current elevated patient safety
Leakage current between ECG leads or to ground exceeds safe limits across all test configurations
Ranked causes
- 1. ECG cable insulation damaged or wet (occasional)
- 2. ECG input isolation circuit failure (rare)
Measure lead-to-ground leakage (Neutral Closed, Normal, Lead-Gnd Normal)
- Test point
- ECG lead connections
- Expected
- < 50 µA
- Tools
- Leakage current tester
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG lead signal amplitude out of range
5-lead ECG gain does not meet specification during acquisition
Ranked causes
- 1. Loose or damaged ECG cable (common)
- 2. ECG preamplifier circuit fault (rare)
Record 5-lead ECG gain amplitude
- Test point
- ECG lead input
- Expected
- 36 to 44 mm, 36 to 44 mm
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG signal has no amplitude
ECG trace displays flat line or zero amplitude signal
Ranked causes
- 1. ECG cable disconnected or damaged (common)
- 2. A02 Patient Parameter PCB failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fails ground resistance test high voltage
Unit fails ground continuity or grounding resistance safety test
Ranked causes
- 1. Power cord grounding issue (common)
- 2. Power module grounding circuit failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frozen at power-on screen caution
Device stops at startup screen
Ranked causes
- 1. W18 UI Flex Cable defective (unknown)
- 2. A01 System PCB faulty (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ground resistance out of specification high voltage
Measured ground resistance does not meet safety requirements
Ranked causes
- 1. Loose or corroded ground connection (common)
- 2. Ground conductor broken or high-resistance path (occasional)
Measure DC ground resistance
- Test point
- Chassis ground point to mains earth
- Expected
- < 0.1 ohms (typical for medical equipment)
- Tools
- Ohmmeter with 4-wire capability
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
HARD PADDLE key not responding patient safety
HARD PADDLE key does not respond when pressed
Ranked causes
- 1. Hard paddles damaged or connector issue (common)
- 2. A04 Therapy PCB failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad buttons not responding properly
Buttons do not respond to presses or respond inconsistently
Ranked causes
- 1. Elastomer keypad worn or damaged (common)
- 2. A05 UI PCB button circuit failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose or broken hardware caution
Loose or broken components on the device
Ranked causes
- 1. Loose hardware (common)
- 2. Broken components (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low patient impedance
Patient impedance measurement lower than expected
Ranked causes
- 1. Incorrect test load used (common)
- 2. A01 System PCB faulty (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Low patient impedance during test
Patient impedance measurement is lower than expected during testing
Ranked causes
- 1. Wrong test plug used or test procedure issue (common)
- 2. A01 System PCB impedance measurement failure (occasional)
Test into 50 ohm load during 3 A.M. test
- Test point
- Test load
- Expected
- Correct test plug used
- Tools
- 50 ohm test load
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Maintenance Due message remains on screen
Maintenance alert persists on display
Ranked causes
- 1. Maintenance prompt interval not set (common)
- 2. Maintenance prompt interval needs reset (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ac power reminder tone caution
AC power availability alert tone not sounding
Ranked causes
- 1. User configuration setting incorrect (common)
- 2. A03 Power module faulty (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- No AC power reminder tone or alert
- No amplitude ECG
- No backup battery operation
- No display (white or blue screen)
- Pacer output current out of tolerance
- Printer characters are missing or broken
- Printer malfunction or not installed
- Printer test strip defects
- QUIK-COMBO ECG analog output missing or out of tolerance
- QUIK-COMBO ECG gain out of tolerance
- Remote sync function not responding to synchronization signal
- Remote sync LED not flashing
- Service code logged, service LED on
- Speed Dial not functioning
- SpO2 leakage current elevated
- SpO2 probe signal not detected or reading out of range
- SpO2 reading out of specification
- Standard Paddles ECG gain out of tolerance
- Therapy delivered energy out of tolerance
- Therapy ECG characteristic amplitude out of range
- Therapy impedance sense malfunction
- Therapy impedance sense not functioning
- Therapy leakage current elevated
- Unit frozen at power on screen
- User test does not complete or report not printed
- AC Mains indicator not on
- AC power LED on LIFEPAK 20 device not illuminated
- Accessory cables or sensors expired or unsuitable
- Calibration: Audio test
- Calibration: Chassis Leakage Current
- Calibration: CO2 measurement
- Calibration: Defibrillation
- Calibration: Defibrillator isolation
- Calibration: Earth leakage current normal condition
- Calibration: Earth leakage current single fault condition
- Calibration: Ground Resistance
- Calibration: Keypad test
- Calibration: LIFEPAK 20 TCP setup
- Calibration: LIFEPAK 20/20e instrument calibration
- Calibration: Pacer Leads Off Detection
- Calibration: Pacer Option Characteristics
- Calibration: Pacer Output Current
- Calibration: Performance Inspection Procedure
- Calibration: Performance Inspection Procedure
- Calibration: Pixels test
- Calibration: Printer test
- Calibration: Protective earth resistance CodeManagement Module ground stud
- Calibration: Protective earth resistance defibrillator ground stud
- Calibration: Reset Maintenance Prompt
- Calibration: Service Mode Passcode
- Calibration: Standard Paddles Defibrillation Isolation
- Calibration: Standard Paddles Delivered Energy
- Calibration: Standard Paddles ECG Characteristics
- Calibration: Standard Paddles Remote Synchronous Cardioversion
- Calibration: Standard Paddles Synchronous Cardioversion R-Wave
- Calibration: Standard Paddles User Test
- Calibration: TCP - Resource Requirements
- Calibration: TCP Defibrillator
- Calibration: TCP Defibrillator Isolation
- Calibration: TCP delivered energy
- Calibration: TCP setup
- Carrying strap or mounts damaged or missing
- CHECK PRINTER message does not appear
- Connectors damaged or contaminated
- Continuous reset
- Contrast Test
- Date and time incorrect
- Defibrillation capacitor charge/dump operation fails
- Defibrillator capacitor charge/dump cycle fails during USER TEST
- Defibrillator isolation failure
- Delivered energy at 10 joules out of range
- Delivered energy at 100 joules out of range
- Delivered energy at 15 joules out of range
- Delivered energy at 2 joules out of range
- Delivered energy at 360 joules out of range
- Delivered energy at 50 joules out of range
- Delivered energy at 70 joules out of range
- Delivered energy out of tolerance
- Device beeps when turned off
- Device exterior damage or wear
- Dirt, fluids, or foreign objects on device
- Dirt, fluids, or foreign objects present
- Display does not illuminate
- Display is distorted
- Display pixel anomalies
- Distorted display
- ECG analog output amplitude out of range
- ECG leads not detected
- ECG signal amplitude out of specification
- ECG signal is saturated
- Fails chassis leakage current test
- Fails chassis leakage test
- Hidden switch between OPTIONS and PAUSE not functioning
- Keypad button not responding
- Keypad controls unresponsive
- Keypad or overlays damaged
- LCD contrast insufficient
- LIFEPAK 20 Device Data Checksum Error
- LIFEPAK 20 Device Serial Port Communication Failure
- Loose or rattling hardware inside device
- MAINTENANCE DUE indicator remains on
- No CO2 Operation displayed on monitor
- No wireless connection to LIFENET System
- Non-volatile RAM Integrity Test Failure
- Pacer leads off detection failed
- Pacer pulse width out of specification
- Pacing current or rate out of tolerance
- Pacing current/rate out of tolerance
- Paddles ECG signal amplitude out of specification
- Paddles require cleaning
- Pixel Test
- Power on sequence incomplete
- Real Time Clock Failure
- Remote sync energy transfer failed
- Rubber feet damaged or missing
- Service indicator illuminated
- Service indicator LED illuminates after power-on
- Service indicator LED illuminates during self-test
- SERVICE indicator lights during self-test
- SERVICE indicator remains on
- Synchronization delay excessive
- Therapy delivered energy out of specification
- Therapy user test failed
- Unit continuously resets
- User Test Failed
- Voice prompts inaudible or distorted
126 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0002 | System Flash memory voltage error | Flash memory access | error |
0003 | Cannot erase system Flash memory | Memory erase operation | error |
0004 | Cannot write to system Flash memory | Memory write operation | error |
0006 | System ADC read error | ADC read operation | error |
0007 | System DAC not responding | DAC access during operation | error |
0008 | ECG OUT DAC self-test failed | ECG output DAC self-test fails | critical |
000a | System ADC background test failed | Analog-to-digital converter self-test fails | critical |
000b | System ADC failed self-calibration | ADC self-calibration procedure fails | critical |
000c | System Flash memory ID unknown | Flash memory chip identification fails | critical |
000d | System hardware/software configuration lost | Configuration data becomes unavailable | critical |
01 | Real Time Clock Failure | Real Time Clock failure detected by CodeManagement Module self-test | warning |
02 | Non-volatile RAM Integrity Test Failure | Non-volatile RAM integrity test failure detected | warning |
03 | LIFEPAK 20 Device Data Checksum Error | Data checksum error detected in LIFEPAK 20 device | error |
04 | LIFEPAK 20 Device Serial Port Communication Failure | Serial port communication failure between CodeManagement Module and LIFEPAK 20 | error |
100e | UI – system communication lost | UI-System communication failure | critical |
100f | Display update timer error | Display update timer malfunctions | error |
1010 | Display update queue error | Display update queue processing fails | error |
1013 | System detected unexpected UI reset | UI resets unexpectedly during operation | error |
1014 | Voice prompt/audio watchdog failure | Audio system watchdog detects failure | warning |
1015 | USB data time-out error | USB data transfer exceeds time limit | error |
1016 | UI USB data error | USB data corruption detected on UI interface | error |
1c01 | UI FPGA programming error | Field programmable gate array programming fails | error |
1c02 | UI FPGA verification error | FPGA verification fails after programming | error |
1c03 | UI FPGA program file error | FPGA program file is corrupted or missing | error |
1c04 | UI ADC not functioning | User interface ADC converter fails | error |
1c05 | Printer ADC out of tolerance (temperature) | Printer temperature sensor reading out of spec | warning |
1c06 | Printer ADC out of tolerance (voltage) | Printer voltage sensor ADC reading exceeds limits | warning |
1c07 | UI voltage out of tolerance (5 V) | 5V power supply reading out of spec | error |
1c08 | UI voltage out of tolerance (3.3 V) | 3.3V supply on UI PCB exceeds tolerance | error |
1c09 | UI voltage out of tolerance (2.5 V) | 2.5V supply on UI PCB exceeds tolerance | error |
1c0a | UI voltage out of tolerance (35 V) | 35V supply on UI PCB exceeds tolerance | error |
1c0b | UI voltage out of tolerance (SW VBATT) | Switched battery voltage on UI PCB exceeds tolerance | error |
1c0c | UI voltage out of tolerance (ground) | Ground reference voltage on UI PCB out of range | error |
1c0f | UI hardware I.D. corrupted | Hardware identification data on UI PCB becomes corrupted | critical |
1c10 | UI boot program corrupted | UI boot program becomes corrupted | critical |
1c11 | UI application Flash memory corrupted | UI application code in Flash memory corrupted | critical |
1c12 | UI font Flash memory corrupted | Font data in UI Flash memory corrupted | critical |
1c13 | UI FPGA Flash memory corrupted | FPGA program in Flash memory corrupted | critical |
1c16 | UI CPU RAM test failed on power-on | UI RAM self-test fails during power-on | critical |
1c17 | UI CPU RAM test failed during normal operation | UI RAM test fails during normal operation | critical |
1fff | Additional information related to prior error code | Logged before another error code for diagnostic context | info |
2004 | System cannot initialize serial port (system connector) | Serial port initialization fails | error |
2005 | System cannot initialize driver for serial port | Serial port driver fails to initialize | error |
3001 | System cannot use data management Flash memory | Data management Flash memory becomes unavailable | error |
3002 | System data management Flash memory corrupted | Data management Flash memory data becomes corrupted | critical |
3005 | System cannot delete DM record | Data management record deletion fails | error |
3007 | System cannot create new DM record | Data management record creation fails | error |
3008 | System could not store DM record | Data management record storage fails | error |
3009 | System could not erase oldest DM record | Cannot erase oldest data management record when memory full | error |
300a | System cannot clear DM memory | Data management memory clear operation fails | error |
300b | System error writing DM record | Data management record write operation fails | error |
300c | System cannot read archived DM record | Cannot read archived data management record | error |
3010 | System error DM memory corrupt | Data management memory corruption detected | critical |
3fff | Additional information related to prior error code | Logged before another error code for diagnostic context | info |
4005 | System NVRAM error log | Non-volatile RAM error log becomes corrupted | error |
4006 | Error log queue not functioning | Error logging queue malfunctions | error |
4008 | Error log count corrupted | Error log counter becomes corrupted | error |
4009 | System RAM test failed | System RAM self-test fails | critical |
400a | System program Flash memory corrupted | System program code in Flash memory corrupted | critical |
400b | Program contents failure | System program code verification fails | critical |
400c | System ADC voltage low | System power supply voltage falls below minimum | warning |
400d | System ADC voltage high | System power supply voltage exceeds maximum | warning |
4010 | Service LED failed | Service indicator LED fails to illuminate | warning |
4012 | Voice Flash memory corrupted | Voice prompts in Flash memory corrupted | warning |
4013 | Voice/font Flash memory invalid | Voice and font data in Flash memory invalid | warning |
4fff | Additional information related to prior error code | Logged before another error code for diagnostic context | info |
5001 | System watchdog failed | Watchdog timer failure | critical |
5003 | System watchdog failed | Watchdog timer failure | critical |
5004 | System watchdog reset | System watchdog forces reset | error |
5005 | System CPU error during boot-up | CPU encounters error during startup | critical |
5006 | System RAM failed during boot-up | RAM self-test fails during startup | critical |
5007 | System checksum failure during boot-up | Boot code checksum verification fails | critical |
5008 | System boot program failure | Boot program execution fails | critical |
5009 | Real-time clock (RTC) access failed | RTC becomes inaccessible | error |
500a | Coin cell battery not detected | Backup battery missing or connection lost | warning |
500b | Cannot use system NVRAM | Non-volatile RAM becomes unavailable | error |
500c | System application start error | System application fails to start | error |
500d | System software initialization time out | System initialization exceeds time limit | error |
500e | System application start error | System application fails to start | error |
5010 | Configuration mismatch | System configuration inconsistent | warning |
5011 | NVRAM configuration data error | Configuration data in NVRAM corrupted | error |
5012 | Configuration data error | Configuration data becomes corrupted | error |
5013 | System meter initialization error | Meter initialization fails | warning |
5014 | System meter mismatch | Meter configuration mismatch detected | warning |
5015 | NVRAM error | Non-volatile RAM error | error |
5016 | MFG data mismatch | Manufacturing data inconsistency | error |
5017 | NVRAM MFG data lost | Manufacturing data in NVRAM lost | critical |
5018 | Watchdog reset failed | Watchdog reset operation fails | critical |
5019 | NVRAM corrupted | Non-volatile RAM data corrupted | critical |
501a | NVRAM corrupted | Non-volatile RAM data corrupted | critical |
501c | RTC not running | Real-time clock stops running | warning |
501d | RTC out of sync | Real-time clock time synchronization lost | warning |
501e | System software execution error | Software execution error occurs | error |
501f | System software read error | Software read operation fails | error |
5020 | System software write error | Software write operation fails | error |
5021 | System software error | System software encounters error | error |
5022 | System software exception code | Software exception occurs | error |
5023 | System software exception code | Software exception occurs | error |
5024 | System software exception code | Software exception occurs | error |
5026 | NVRAM low battery interrupt | Backup battery voltage falls below threshold | warning |
5027 | System USB did not initialize | System USB interface fails to initialize | error |
5028 | PP USB did not initialize | Patient module USB interface fails to initialize | error |
5029 | Therapy USB did not initialize | error | |
502a | UI USB did not initialize | error | |
502b | USB system failed | error | |
502f | Device type invalid | error | |
5030 | PP USB disconnect | error | |
5031 | UI USB disconnect | error | |
5032 | Therapy USB disconnect | USB connection lost between therapy module and system | error |
5033 | PP USB download time out | PP module fails to complete USB download within timeout window | error |
5036 | USB initialization failed | USB subsystem fails to initialize | error |
5037 | USB unitization error | USB unit fails to initialize properly | error |
5038 | Cannot clear interrupt source | Interrupt source cannot be cleared during shutdown or reset | error |
5039 | Bus error reported by USB PP channel | PP module reports USB bus error | error |
503a | Bus error reported by USB UI channel | UI module reports USB bus error | error |
503b | Bus error reported by USB therapy channel | Therapy module reports USB bus error | error |
503c | USB host driver cannot create transfer descriptor | USB host driver fails to allocate transfer descriptor | error |
503d | USB host driver received message larger than maximum allowed | Received USB message exceeds size limit | error |
5040 | Inconsistent RTC time | Real-time clock value is inconsistent or corrupted | warning |
5105 | Battery failed to reach charge in 2.5 hours | error | |
5106 | Power supply out of tolerance | Power supply voltage outside acceptable range | error |
5107 | Power module RAM error | RAM test failure in power module | error |
5108 | Power module self-test service code | Power module self-test failure detected | error |
510a | Battery <10volts after 20minutes of charging | Battery charge cycle fails to reach minimum voltage | warning |
510b | Does not switch to battery power | Device fails to switch to battery when AC power is removed | critical |
510c | Does not detect ac disconnect | AC disconnect detection circuit failure | critical |
510d | Battery powered when connected to ac power | Device incorrectly uses battery while AC is connected | error |
510e | Battery <10volts after charge cycle | Battery voltage remains low after charging | warning |
510f | Battery charge cycle stopped | Charging process stopped prematurely | error |
5110 | AC isolation Diode Shorted | error | |
5112 | Battery not detected | Battery presence or communication not detected | critical |
5115 | Battery thermistor <400 Ohms | Battery thermistor resistance below 400 ohms | error |
5116 | Charger reporting zero during battery charging | Charger circuit reports zero output during charging | error |
5117 | No 12C connection detected | I2C communication link lost (LIFEPAK 20e only) | error |
5118 | Power module connected to NiMH battery | Power module for lithium battery connected to NiMH battery (LIFEPAK 20 only) | error |
511a | Battery failure | Battery has reached end of life or is defective | error |
5fff | Additional information related to prior error code | info | |
6002 | PP program corrupted | Pulsepoint processor program memory corrupted | error |
6003 | PP program not found | Pulsepoint processor program not present in memory | error |
6004 | PP boot-up error | Pulsepoint processor fails to boot | error |
6009 | No PP data | error | |
600c | SpO2 misconfigured | SpO2 module configuration is incorrect | warning |
600e | PP reset | Pulsepoint processor reset unexpectedly | warning |
600f | OEM configuration error | OEM-specific configuration is invalid | warning |
6010 | PP initialization error | Pulsepoint processor fails to initialize | error |
6018 | Incorrect updated SpO2 image | SpO2 firmware image update is incorrect or corrupted | error |
6019 | Incorrect updated SpO2 image CRC | SpO2 firmware image CRC check failed after update | error |
6801 | PP power supply out of tolerance | Pulsepoint processor power supply voltage out of range | error |
6802 | PP pre-amp data invalid | Pulsepoint processor pre-amp output invalid | error |
6804 | PP data RAM test error | Pulsepoint processor data RAM test failure | error |
6805 | PP RAM test error | Pulsepoint processor RAM test failure | error |
6806 | PP CRC test error | Pulsepoint processor program CRC check failure | error |
6807 | PP ECG test error | Pulsepoint processor ECG test failure | error |
680b | SpO2 board error | SpO2 module hardware failure detected | error |
6fff | Additional information related to prior error code | info | |
800a | System DSP error | Digital signal processor firmware error | error |
8013 | Voice format error | Voice message format is invalid or corrupted | warning |
8014 | No paddles data | Paddles firmware or data not available | error |
801b | DSP did not receive USB SOF interrupt | DSP fails to receive USB start-of-frame interrupt | error |
8105 | Impedance channel out of calibration | Impedance measurement circuit calibration lost | error |
8108 | Paddles data out of sync | Paddles firmware synchronization lost | error |
8109 | Paddles pre-amp user test failed | Paddles pre-amp self-test failure | error |
810b | Real impedance <-30 ohms for one second | Measured impedance is negative or out of range | error |
8fff | Additional information related to prior error code | info | |
9004 | Unable to initialize therapy control | Therapy control circuit fails to initialize | critical |
9005 | Defib disabled | Defibrillation circuit disabled by firmware | critical |
9007 | Shock not delivered | Defibrillation shock failed to deliver | critical |
9009 | Defib charge time expired | critical | |
900b | Pacing rate out of tolerance | Pacing rate outside acceptable range | critical |
900f | Unable to initialize therapy control | Therapy control stack initialization failure | critical |
9011 | Pacer fault | Pacemaker circuit hardware failure | critical |
9017 | Pacer disabled | Pacemaker circuit disabled | warning |
901a | Pacer rate storage corrupted | Pacemaker rate data memory corrupted | error |
901b | Therapy PCB communication lost | Communication link to therapy PCB lost | critical |
901d | Testing purpose only | Reserved test code | info |
9021 | Impedance value indicates regulator failure in System PCB pre-amp | System PCB pre-amp regulator detected as failed | error |
9c03 | Therapy processor unplanned reset | Therapy processor reset unexpectedly | error |
9c04 | Therapy/system controller communication watchdog | Communication watchdog timeout between therapy and system controllers | critical |
9c05 | CRC error | Therapy processor program CRC check failed | error |
9c06 | Calibration constant A out of range | Therapy calibration constant A is outside acceptable limits | error |
9c07 | Calibration constant B out of range | Therapy calibration constant B is outside acceptable limits | error |
9C08 | Therapy ROM Cyclic Redundancy Check (CRC) failed | error | |
9c09 | Therapy RAM pattern write test failed | Therapy processor RAM write test failure | error |
9c0a | Therapy relay idle coil voltage out of range | Therapy relay idle state coil voltage out of spec | error |
9c0b | Therapy relay enabled coil voltage out of range (5 ms) | Therapy relay enabled state coil voltage at 5ms out of spec | error |
9c0c | Therapy relay enabled coil voltage out of range (100 ms) | Therapy relay enabled state coil voltage at 100ms out of spec | error |
9c0d | Therapy relay drive enabled coil voltage out of range | Therapy relay drive enable coil voltage out of spec | error |
9c0f | Therapy/power assembly communication error | Communication lost between therapy and power modules | critical |
9c11 | Capacitor Dump failed | High voltage capacitor discharge failed | critical |
9C12 | Therapy pcba 5 volt out of range | error | |
9C13 | Therapy pcba 15 volt out of range | error | |
9c14 | Therapy pcba -15 volt out of range | error | |
9c15 | Therapy ADC time out error | Analog-to-digital converter response timeout | error |
9C18 | 3AM H bridge test: NE leg shorted | error | |
9C19 | 3AM H bridge test: SE leg shorted | error | |
9C1a | 3AM H bridge test: NW leg shorted | error | |
9C1b | 3AM H bridge test: SW leg shorted | error | |
9C1c | 3AM H bridge test: East side stuck open | error | |
9C1d | 3AM H bridge test: West side stuck open | error | |
9C1e | 3AM H bridge test: Charge time out of range | error | |
9c1f | 3:00 AM shorted paddles relay contact test: relay shorted | error | |
9c20 | Shorted paddles relay contact test failure | 3:00 AM shorted paddles relay contact test detected relay shorted | error |
9c21 | Pace drive power supply stuck on | 3:00 AM pace drive test detected power supply stuck in on state | error |
9c22 | Pace drive power supply inoperable | 3:00 AM pace drive test found power supply not functioning | error |
9c23 | Relay contacts shorted in pace drive | 3:00 AM pace drive test detected shorted relay contacts | error |
9c24 | Relay drive low side shorted | 3:00 AM pace drive test detected low side short circuit | error |
9c25 | Relay drive high side shorted | 3:00 AM pace drive test detected high side short circuit | error |
9c26 | Pace FET shorted | 3:00 AM pace drive test detected shorted field effect transistor | error |
9c27 | Pace current path open | 3:00 AM pace drive test detected open circuit in current path | error |
9c28 | Pace set point error | 3:00 AM pace drive test detected incorrect pace set point value | error |
9c29 | Redundant controls stuck on | 3:00 AM redundant controls test detected controls stuck in on state | error |
9c2a | Redundant controls enable 2 stuck on | 3:00 AM redundant controls test detected enable 2 stuck on | error |
9c2b | Redundant controls enable 1 stuck on | 3:00 AM redundant controls test detected enable 1 stuck on | error |
9c2c | Capacitor voltage per pacing pulse too high | Pacing pulse capacitor voltage exceeds specification | error |
9c2d | Capacitor current per pacing pulse too high | Pacing pulse capacitor current exceeds specification | error |
9c2e | Cap current per pacing pulse too high | Pacing pulse capacitor current exceeds specification | error |
9c2f | Pacing current and selected current out of range | Measured pacing current does not match programmed value | error |
9c30 | Pacing pulse width too short | Pacing pulse width is shorter than specification | error |
9c31 | Pacing pulse width too long | Pacing pulse width is longer than specification | error |
9c32 | Capacitor voltage and predicted capacitor voltage mismatch | Measured capacitor voltage does not match predicted value | error |
9C33 | Capacitor charge time >25seconds | error | |
9c35 | Therapy CPU instruction test failed | CPU self-test detected instruction execution error | error |
9c36 | Therapy software stack overflow | Software stack memory exceeded during operation | error |
9c3a | Energy capacitor overvoltage error | Energy capacitor voltage exceeds maximum safe level | critical |
9c3b | Redundant controls charge rate stuck on | 3:00 AM redundant controls test detected charge rate control stuck on | error |
9c3e | Therapy software error | Software detected an unspecified error condition | error |
9c3f | Therapy software error | Software detected an unspecified error condition | error |
A00b | Printer communication lost | Printer interface failure | error |
a00e | Printer initialization error | Printer fails to initialize during startup | error |
b001 | Invalid state request | System receives invalid state command | error |
b011 | System behavior manager error | System behavior manager encounters error condition | error |
b012 | Energy cap charging time out | Defibrillator capacitor fails to charge within timeout period | warning |
B013 | Shock advisory system error | Algorithm or sensor failure | critical |
b014 | Shock advisory system time out | Shock advisory analysis exceeds time limit | error |
b016 | Motion detect timer error | Motion detection timer malfunctions | warning |
b017 | Shock result time out | Shock delivery monitoring exceeds time limit | error |
b018 | USB interrupt error | USB interrupt handling fails | error |
MAINTENANCE DUE | Scheduled maintenance interval has passed and maintenance is due | Device powered on after scheduled maintenance interval has expired | warning |
PASSCODE INCORRECT - TRY AGAIN | Service mode passcode entry was invalid | Incorrect service mode passcode entered when accessing SERVICE menu | warning |
Service Log entry | Self-test failure detected | Errors detected during power-on or continuous self-test cycles | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.