Physio-Control LIFEPAK 15: troubleshooting and error codes
The ServiceForge knowledge base holds 166 documented symptoms for the Physio-Control LIFEPAK 15, with 230 ranked causes, 374 diagnostic checks with test points and expected values, and 162 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No energy discharge from paddles high voltage
QUIK-COMBO or hard paddles fail to deliver electrical energy
Ranked causes
- 1. Test setup or procedure error (common)
- 2. Defective therapy cable or paddles (occasional)
- 3. W11 Therapy Connector Cable failure (occasional)
- 4. A04 Therapy PCB failure (rare)
- 5. A22 Biphasic Module PCB failure (rare)
- 6. A14 Inductive Resistor failure (rare)
- 7. A15 Energy Storage Capacitor failure (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 saturation reading missing or out of tolerance patient safety
SpO2 module displays XXX and Service LED OFF, indicates sensor/reading issue
Ranked causes
- 1. Test setup error (common)
- 2. Defective finger sensor (occasional)
- 3. Device configuration error (occasional)
- 4. A16 SpO2 Module failure (rare)
- 5. W22 SpO2 Connector Cable failure (rare)
- 6. A06 OEM PCB failure (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 fails calibration patient safety
CO2 module cannot complete calibration procedure
Ranked causes
- 1. Test setup error or empty calibration gas (common)
- 2. FilterLine disconnected (occasional)
- 3. Pinched or kinked tubing inside device (occasional)
- 4. A23 CO2 module failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper key response
Keypad keys do not respond properly to presses
Ranked causes
- 1. A09 Printer Control Keypad failure (occasional)
- 2. A10 Main Keypad failure (occasional)
- 3. A05 Interface PCB failure (rare)
- 4. A01 System PCB failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing dots in printed X pattern
Printer test pattern X has missing dots in output
Ranked causes
- 1. Improper printer paper in use (common)
- 2. Dirty printhead (common)
- 3. A12 Printer Assembly failure (occasional)
- 4. A01 System PCB failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing or broken characters in print output
Printed characters are incomplete or missing
Ranked causes
- 1. Improper printer paper (common)
- 2. Dirty printhead (common)
- 3. A12 Printer Assembly failure (occasional)
- 4. A01 System PCB failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No pacer output high voltage
Pacing function does not deliver pacing pulses
Ranked causes
- 1. Test setup error (common)
- 2. A13 Transfer Relay connection problem at P22 (occasional)
- 3. A13 Transfer Relay failure (occasional)
- 4. A04 Therapy PCB failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power ON caution
Device will not power on when turned on
Ranked causes
- 1. Battery not fully charged or improperly maintained (common)
- 2. Loose or broken battery pin(s) (occasional)
- 3. W04 System/Interface PCB cable connection issue (occasional)
- 4. A03 Power PCB failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ABNORMAL ENERGY DELIVERY message appears high voltage
Device displays abnormal energy delivery alert
Ranked causes
- 1. A13 Transfer Relay connection problem at P22 (occasional)
- 2. A13 Transfer Relay failure (occasional)
- 3. A04 Therapy PCB failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK PRINTER message appears
Device displays CHECK PRINTER alert message
Ranked causes
- 1. Paper sensor dirty (common)
- 2. Printer paper not correctly loaded (common)
- 3. A12 Printer Assembly failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 displays FilterLine Blockage message patient safety
CO2 module indicates FilterLine is blocked
Ranked causes
- 1. FilterLine blockage (common)
- 2. Occlusion or kinking of input tubing (occasional)
- 3. A23 CO2 module failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 displays XXX with Service LED OFF patient safety
CO2 module shows XXX reading and Service LED is OFF
Ranked causes
- 1. FilterLine blockage (common)
- 2. Occlusion or kinking of internal tubing (occasional)
- 3. A23 CO2 module failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to transfer energy at Sync mark high voltage
Energy discharge does not occur at sync mark during synchronized cardioversion
Ranked causes
- 1. Device in sync mode with discharge attempt needed (common)
- 2. A10 Main Keypad failure (occasional)
- 3. Defective therapy cables or paddles (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper LCD response
LCD display does not respond normally or shows degraded performance
Ranked causes
- 1. Display test failure indication (common)
- 2. A11 LCD Assembly failure (occasional)
- 3. A01 System PCB failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaudible or garbled audio patient safety
Audio output is not heard or is distorted
Ranked causes
- 1. W17 Speaker Assembly failure (common)
- 2. A05 Interface PCB failure (occasional)
- 3. A01 System PCB failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure P1/P2 channel reading out of tolerance patient safety
Mean Arterial Pressure (MAP) reading does not match simulated pressure input within tolerance window
Ranked causes
- 1. Pressure transducer requires zeroing or recalibration (common)
- 2. Invasive pressure connector loose or not properly seated (occasional)
- 3. Pressure transducer or cable fault (rare)
Set simulator to ZERO output and zero pressure channel
- Test point
- P1/P2 channel display and MAP reading
- Expected
- -20 ±3 mmHg
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Sync mark displayed high voltage
Synchronous cardioversion sync mark not visible on ECG waveform
Ranked causes
- 1. Test setup error (common)
- 2. ECG size adjustment needed (common)
- 3. A01 System PCB failure (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-Lead ECG leads off detection failure patient safety
Device does not display leads-off message or sound repeating priority 3 tone when individual ECG lead is disconnected during 3-lead test
Ranked causes
- 1. ECG cable connection loose or missing lead contact (common)
- 2. Leads-off detection circuit malfunction (occasional)
Verify LL lead off detection at LEAD II
- Test point
- LL ECG lead connector
- Expected
- LL LEADS OFF message and repeating priority 3 tone
- Tools
- 3-lead ECG cable, Impulse 7000DP simulator
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
5-Lead ECG leads off detection failure patient safety
Device does not display leads-off message or sound repeating priority 3 tone when individual ECG lead is disconnected during 5-lead test
Ranked causes
- 1. ECG cable connection loose or missing lead contact (common)
- 2. Leads-off detection circuit malfunction (occasional)
Verify LL lead off detection at LEAD II
- Test point
- LL ECG lead connector
- Expected
- LL LEADS OFF message and repeating priority 3 tone
- Tools
- 5-wire ECG cable, Impulse 7000DP simulator
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batteries not fully charged or showing low battery warning patient safety
Lithium-ion batteries do not show more than two charge bars or LOW BATTERY message appears after powering on
Ranked causes
- 1. Batteries not fully charged before use (common)
- 2. Battery defective or worn out (occasional)
Inspect charge bar display on batteries
- Test point
- Battery status indicators on installed batteries
- Expected
- More than two charge bars visible on each battery
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery connector pins damaged or loose high voltage
Battery connector pins are loose, cracked, or broken, preventing reliable power connection
Ranked causes
- 1. Loose battery connector pins (occasional)
- 2. Cracked or broken connector pin leaves (occasional)
Visually inspect battery connector pins for tightness
- Test point
- Battery connector pins at bottom of device
- Expected
- Pins firmly seated, no movement
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery does not click into position in battery well patient safety
Lithium-ion battery fails to seat properly or make audible/tactile contact in battery compartment
Ranked causes
- 1. Battery contacts or well damaged or obstructed (occasional)
- 2. Battery case deformed or defective (rare)
Attempt to insert battery into well and listen/feel for click
- Test point
- Battery well and battery case contact points
- Expected
- Audible or tactile click when battery seats fully
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery icons not visible when power adapter connected patient safety
Battery indicators do not appear on display when external AC/DC power adapter is connected to Auxiliary Connector
Ranked causes
- 1. Power adapter not properly connected or Auxiliary Connector issue (common)
- 2. Display or power management circuit malfunction (unknown)
Verify power adapter cable is fully inserted into Auxiliary Connector
- Test point
- Auxiliary Connector at rear of device
- Expected
- Cable fully seated with no gaps
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not fully charged or functional
Lithium-ion batteries do not show sufficient charge bars or return LOW BATTERY message when powered on
Ranked causes
- 1. Battery insufficient charge (common)
- 2. Battery defective or failed (occasional)
Verify battery charge state
- Test point
- Battery status indicator on device display
- Expected
- More than two charge bars showing on both batteries
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bluetooth wireless pairing failure
Device fails to pair with computer via Bluetooth, no alert tone, LED not illuminated, or connection message does not appear.
Ranked causes
- 1. Bluetooth wireless disabled or not configured (common)
- 2. Passcode mismatch between device and computer (occasional)
Verify CODE-STAT Reviewer application open on computer
- Test point
- Software setup
- Expected
- CODE-STAT Reviewer running, DOWNLOAD WIZARD accessible
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 FilterLine installed but no indication on display
FilterLine connected but device does not show CO2 is connected
Ranked causes
- 1. CO2 input connector dirty (common)
- 2. FilterLine defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defibrillator sync function failure or delay out of range patient safety
SYNC LED not illuminating, R-wave markers not appearing, or sync delay exceeds 60 ms
Ranked causes
- 1. ECG cable connection or signal quality issue (common)
- 2. R-wave detection circuit malfunction or sync delay circuit degradation (occasional)
Verify SYNC LED illuminates when SYNC button pressed
- Test point
- SYNC button and LED
- Expected
- SYNC LED on; R-wave markers visible on ECG
- Tools
- Defibrillator analyzer, main ECG cable
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device shows no power indication after case reassembly
No indication that power is on after case is reassembled and device is turned ON
Ranked causes
- 1. Batteries are dead or not installed (common)
- 2. W04 cable connection not made during case reassembly (common)
Verify batteries are installed and charged
- Test point
- Battery compartment
- Expected
- Fresh batteries installed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure P1 channel calibration and verification patient safety
Verification that P1 invasive pressure channel displays correct MAP values at multiple pressure points with proper zero offset
Ranked causes
- 1. Pressure transducer calibration drift or offset error (occasional)
- 2. Patient simulator not configured correctly (occasional)
Connect invasive pressure cable to P1 connector with simulator at ZERO
- Test point
- P1 connector on parameter bezel
- Expected
- Connection established and display shows no error
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invasive pressure P1 channel does not read correctly patient safety
P1 pressure waveform or MAP display reads outside tolerance when simulated pressures applied, or channel will not zero
Ranked causes
- 1. P1 transducer calibration drift or sensor failure (common)
- 2. Invasive pressure cable disconnected or faulty (occasional)
Verify cable connection to P1 connector
- Test point
- P1 connector on parameter bezel
- Expected
- Cable firmly seated
6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad buttons not responding or stuck patient safety
Button does not highlight when pressed, or button remains highlighted at test start indicating short circuit
Ranked causes
- 1. Keypad has internal short circuit (occasional)
- 2. Button contact failure or defective key switch (occasional)
Verify no buttons highlighted at start of test
- Test point
- Keypad button display
- Expected
- No buttons highlighted
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loose or rattling hardware caution
Hardware inside device is loose, causing rattling noise when device is moved or inverted
Ranked causes
- 1. Fasteners not properly tightened (common)
- 2. Broken or missing fasteners (occasional)
Invert device and listen for rattling
- Test point
- Entire device assembly
- Expected
- No rattling or loose hardware sounds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP displays XXX with Service LED OFF patient safety
NIBP module displays XXX and Service LED is OFF, indicates measurement failure
Ranked causes
- 1. NIBP leakage or tubing blockage (common)
- 2. A21 NIBP module failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Therapy cable not connecting to therapy connector properly high voltage
QUIK-COMBO or standard paddles fail to connect securely to therapy connector
Ranked causes
- 1. Connector misaligned or contacts obstructed (common)
- 2. Connector contacts damaged or corroded (occasional)
Verify therapy cable is fully seated in therapy connector
- Test point
- Therapy connector on device and cable connector
- Expected
- Cable fully inserted with no visible gaps or movement when gently pulled
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-lead ECG fails leads-off detection patient safety
Device does not display leads-off message or priority 3 tone does not sound when individual ECG lead is disconnected during acquisition
Ranked causes
- 1. Lead detection circuit failure or ECG cable fault (occasional)
Connect main ECG cable to Impulse simulator
- Test point
- ECG cable limb and precordial lead attachments
- Expected
- All 10 ECG leads connected to simulator
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-Lead ECG gain out of range
Printed ECG signal amplitude does not meet specification for one or more leads during 12-lead ECG gain test
Ranked causes
- 1. ECG signal chain gain mismatch or calibration drift (unknown)
Measure 12-lead ECG gain amplitude for Lead I
- Test point
- ECG output
- Expected
- 25 mm to 31 mm peak-to-peak, 25-31
- Tools
- ECG printer, Impulse 7000DP simulator
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-Lead ECG gain test
Verification that ECG signal amplitudes are correct for each lead at 4.0 ECG size setting
Ranked causes
- 1. ECG gain setting incorrect or amplifier circuit drift (occasional)
Set simulator to 1 mV, 10 Hz sine wave and device ECG SIZE to 4.0
- Test point
- Device ECG SIZE setting and simulator output
- Expected
- Configuration complete for amplitude testing
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-Lead ECG leads-off detection test
Verification that device detects open or disconnected leads and displays appropriate alert messages with priority 3 tone
Ranked causes
- 1. ECG lead connection interrupted or cable damaged (common)
Connect 12-lead ECG cable to patient simulator
- Test point
- Main ECG cable with limb and precordial lead attachments
- Expected
- All 10 leads connected to simulator
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-lead ECG signal amplitude out of range
Printed ECG signal amplitude does not meet specified peak-to-peak values on one or more leads when testing with Impulse 7000DP signal generator
Ranked causes
- 1. ECG signal gain incorrect or attenuated (common)
Measure ECG Lead I printed signal amplitude
- Test point
- Printer output, Lead I
- Expected
- 25 mm to 31 mm peak-to-peak
- Tools
- Impulse 7000DP, printer, ruler or calipers
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-lead ECG gain incorrect
Printed ECG signal amplitude outside acceptable range for Leads I and II
Ranked causes
- 1. ECG amplifier gain miscalibrated (occasional)
Set simulator to 1-mV, 10-Hz sine wave
- Test point
- Simulator output
- Expected
- 1-mV, 10-Hz sine wave
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-Lead ECG gain out of range
Printed ECG signal amplitude does not meet specification for Lead I or Lead II during 3-lead ECG gain test
Ranked causes
- 1. ECG signal chain gain mismatch or calibration drift (unknown)
Measure 3-lead ECG gain amplitude for Lead I
- Test point
- ECG output
- Expected
- 25 mm to 31 mm peak-to-peak, 25-31
- Tools
- ECG printer, Impulse 7000DP simulator
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-lead ECG leads off detection not functioning patient safety
Device does not display correct LEADS OFF message or priority 3 tone does not sound when a 3-lead ECG cable lead is disconnected
Ranked causes
- 1. ECG lead connection loose or not properly seated (common)
Verify LL LEADS OFF detection with 3-lead cable
- Test point
- ECG LL lead connection
- Expected
- LL LEADS OFF message displayed and priority 3 tone sounds
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-lead ECG leads-off detection not alerting patient safety
Device should display specific LEADS OFF message and priority 3 tone when leads disconnected from patient simulator
Ranked causes
- 1. Lead not properly connected to patient simulator (common)
Connect 3-lead ECG cable to device and simulator
- Test point
- 3-lead cable connectors
- Expected
- Properly seated connections
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-lead ECG signal amplitude out of range
Printed ECG signal amplitude does not meet specified peak-to-peak values when testing 3-lead cable with Impulse 7000DP signal generator
Ranked causes
- 1. ECG signal gain incorrect or signal path degraded (common)
Measure 3-lead ECG Lead I printed signal amplitude
- Test point
- Printer output, Lead I
- Expected
- 25 mm to 31 mm peak-to-peak
- Tools
- Impulse 7000DP, printer, ruler or calipers
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
5-lead ECG gain incorrect
Printed ECG signal amplitude outside acceptable range for Leads I and II
Ranked causes
- 1. ECG amplifier gain miscalibrated (occasional)
Set patient simulator to 1-mV, 10-Hz sine wave
- Test point
- Patient simulator output
- Expected
- 1-mV, 10-Hz sine wave
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
5-Lead ECG gain out of range
Printed ECG signal amplitude does not meet specification for Lead I or Lead II during 5-lead ECG gain test
Ranked causes
- 1. ECG signal chain gain mismatch or calibration drift (unknown)
Measure 5-lead ECG gain amplitude for Lead I
- Test point
- ECG output
- Expected
- 25 mm to 31 mm peak-to-peak, 25-31
- Tools
- ECG printer, Impulse 7000DP simulator
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
5-lead ECG leads off detection not functioning patient safety
Device does not display correct LEADS OFF message or priority 3 tone does not sound when an ECG lead is disconnected during 5-lead testing
Ranked causes
- 1. ECG lead connection loose or not properly seated (common)
Verify LL LEADS OFF detection
- Test point
- ECG LL lead connection
- Expected
- LL LEADS OFF message displayed and priority 3 tone sounds
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
5-lead ECG leads-off detection not alerting patient safety
Device should display specific LEADS OFF message and priority 3 tone when individual leads disconnected from patient simulator
Ranked causes
- 1. Lead not properly connected to patient simulator (common)
Remove RL lead and verify alert
- Test point
- RL lead connection
- Expected
- RL LEADS OFF message and priority 3 tone
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
5-Lead ECG leads-off detection test
Verification that device detects open or disconnected leads on 5-wire ECG cable and displays appropriate alerts
Ranked causes
- 1. 5-wire ECG lead connection interrupted or cable damaged (common)
Connect 5-wire ECG cable to patient simulator
- Test point
- 5-wire ECG cable connections
- Expected
- All 5 leads connected (RA, LA, RL, LL, V1)
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
5-lead ECG signal amplitude out of range
Printed ECG signal amplitude does not meet specified peak-to-peak values when testing 5-lead cable with Impulse 7000DP signal generator
Ranked causes
- 1. ECG signal gain incorrect or signal path degraded (common)
Measure 5-lead ECG Lead I printed signal amplitude
- Test point
- Printer output, Lead I
- Expected
- 25 mm to 31 mm peak-to-peak
- Tools
- Impulse 7000DP, printer, ruler or calipers
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Accessory cables or sensors expired or unsuitable patient safety
ECG, SpO2 sensors, CO2 tubing, NIBP tubing, temperature sensors, or related items are expired, damaged, or unsuitable for use
Ranked causes
- 1. Expired or damaged accessory items (common)
Verify expiration dates and condition of all accessories
- Test point
- ECG cables, SpO2 sensors, CO2 tubing, NIBP tubing, temperature sensors
- Expected
- All items current (not expired), clean, and in good condition
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Active battery indicator not showing correctly patient safety
Device does not display which battery is in use via white battery number in black box, or priority selection (lowest charge first) is not functioning
Ranked causes
- 1. Battery charge states not properly detected (occasional)
Active battery indicator display
- Test point
- Battery status display area
- Expected
- White battery number in black box indicating active battery
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Analog ECG output amplitude incorrect
Oscilloscope measurement of analog ECG output outside acceptable voltage range
Ranked causes
- 1. Analog output stage gain miscalibrated (occasional)
Connect device, simulator, and oscilloscope
- Test point
- ECG cable and oscilloscope connections
- Expected
- All connections properly seated
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio output distorted or inaudible patient safety
Voice prompts not clearly audible, reproduced with distortion, or incorrect language
Ranked causes
- 1. Speaker defective or audio circuit fault (occasional)
Play voice prompts via VOICE/TONE test
- Test point
- Device speaker output
- Expected
- Voice prompts clearly audible, no distortion, correct language
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio output unclear or distorted
Voice prompts from speaker are not clearly audible or contain distortion during audio test
Ranked causes
- 1. Speaker malfunction or degraded audio output (unknown)
Confirm voice prompts are clear and undistorted
- Test point
- Speaker output
- Expected
- Voice prompts clearly audible without distortion
Audio test for voice prompts and tones
Verification that device produces clear, undistorted voice prompts and tones from speaker
Ranked causes
- 1. Speaker not functioning or audio signal distorted (unknown)
Select and start voice/tone test
- Test point
- Service/Tests submenu VOICE/TONE option
- Expected
- Voice prompts play from speaker
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary power switching not functioning high voltage
Device does not properly switch between battery and auxiliary power, or battery icons do not indicate power source correctly
Ranked causes
- 1. Auxiliary power connector circuit failure (rare)
Connect power adapter and verify battery icon display
- Test point
- Battery status indicators on display
- Expected
- Battery icons appear but neither is highlighted when adapter connected
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery icons not highlighted when power adapter unplugged patient safety
Battery indicator icons do not show highlight after external power adapter is disconnected from Auxiliary Connector
Ranked causes
- 1. Power switching circuit malfunction or battery detection failure (unknown)
Unplug power adapter and observe battery icons on display
- Test point
- LCD display screen after disconnecting Auxiliary Connector cable
- Expected
- One device battery icon is highlighted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery power source not correctly displayed or managed
Device does not correctly display battery status indicators, which battery is in use, charge state, or does not properly manage battery switching when batteries are installed or removed
Ranked causes
- 1. Battery detection or power management circuit issue (occasional)
Turn device ON and verify battery status display
- Test point
- Battery status indicators on display
- Expected
- Indicators show presence of both batteries, charge state on each, which battery in use
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery status indicators not displaying correctly
Device fails to show battery presence, usage status, or charge state on display
Ranked causes
- 1. Battery status display malfunction (unknown)
Power on device and observe battery status display
- Test point
- Display screen
- Expected
- Battery status indicators visible showing presence of batteries in Wells 1 and 2, which battery in use (white number in black box), and state of charge on each battery
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Battery well 1 presence not detected
- Battery well 2 presence not detected
- Bluetooth wireless connection setup
- Calibration: Analog ECG output
- Calibration: CO2 Module
- Calibration: Defibrillator ECG gain
- Calibration: Defibrillator Energy
- Calibration: Defibrillator Isolation
- Calibration: Defibrillator Output Waveform
- Calibration: Direct Applied Part Leakage Test - ECG
- Calibration: Direct Applied Part Leakage Test - SpO2
- Calibration: Direct Applied Part Leakage Test - Therapy
- Calibration: Direct Equipment Leakage Test
- Calibration: ECG Restore Test - Standard Paddles
- Calibration: Leakage Current - AC Powered Setup
- Calibration: Leakage Current - Battery Powered Setup
- Calibration: Pacer LEADs-Off Detection
- Calibration: Pacer output current
- Calibration: Pacer Pulse Width
- Calibration: Patient Impedance Sense
- Calibration: Performance Inspection Procedures
- Calibration: Printer at 25 mm
- Calibration: Quik-Combo defibrillator delivered energy
- Calibration: QUIK-COMBO ECG characteristics
- Calibration: R-Wave Detection - Standard Paddles
- Calibration: Standard paddles delivered energy
- Calibration: Standard paddles ECG characteristics
- Calibration: Standard paddles synchronous cardioversion
- Calibration: Standard paddles user test
- Calibration: Temperature
- Carrying strap or mounts damaged or missing
- Configure Safety Analyzer for IEC 62353 testing
- Damaged keypad or label
- Date/Time Incorrect on Printout
- Defibrillation event signal recovery
- Defibrillator charge time exceeds specification
- Defibrillator delivered energy out of range
- Defibrillator delivered energy outside acceptable range
- Defibrillator shock marker or energy event not recorded on printout
- Defibrillator synchronization delay exceeds limit
- Defibrillator synchronous cardioversion not synchronizing
- Delivered energy out of tolerance
- Device does not complete power-on self-test within specification
- Device does not detect positive R-waves on QUIK-COMBO paddles
- Device does not use correct battery when two installed
- Device preparation for Performance Inspection Procedures
- Display screen does not match expected startup appearance
- Display screen does not show expected startup screen
- ECG amplitude does not restore to adequate level after defibrillation
- ECG analog output signal amplitude out of range
- ECG baseline does not restore to zero offset after defibrillation
- ECG gain on QUIK-COMBO paddles incorrect
- ECG signal amplitude outside specification during gain test
- ECG signal does not restore to baseline after defibrillation
- Excessive leakage current on ECG direct applied parts
- Excessive leakage current on SpO₂ applied parts
- Excessive leakage current on therapy (defibrillation) applied parts
- Exterior physical damage or wear
- Front panel LEDs do not flash during self-test
- Invasive pressure P2 channel calibration and verification
- Invasive pressure P2 channel does not read correctly
- Keypad button does not respond
- Keypad function test
- Keypad or overlay damaged
- Leakage current testing setup - AC powered
- Leakage current testing setup - battery powered
- Loose or broken hardware
- MAINTENANCE DUE message appears
- Monitor cable or sensor connectors damaged or contaminated
- One or more horizontal lines missing or distorted
- Pacer output current outside acceptable range
- Pacer pulse width out of range
- Pacing leads-off detection not triggered on electrode disconnect
- Paddle leads off detection failure
- Paddles lead-off detection malfunction
- Positive R-wave not detected or recorded on ECG
- Power LED not illuminated after self-test
- Power-on LED does not remain illuminated after self-test
- Power-on self-test fails or completes slowly
- Printer door open alert displayed
- Printer paper not installed or loaded incorrectly
- Printer speed test at 12.5 mm/sec
- Printer speed test at 12.5 mm/sec fails
- Printer speed test at 25 mm/sec fails
- Rubber feet damaged or missing
- SERVICE indicator is ON after case reassembly
- Service LED illuminated after self-test
- Service LED recurrence after clearing and restart
- Service LED remains illuminated after self-test
- Service LED remains ON
- Setup: TCP workstation configuration
- SpCO reading outside acceptable range
- Speaker does not emit clear test tone at power-on
- Speaker does not emit test tone
- SpMet reading outside acceptable range
- SpO2 reading outside acceptable range
- SpO2/SpCO/SpMet probe or sensor fault
- SpO2/SpCO/SpMet sensor test
- Standard Paddles defibrillator delivers energy with paddles in wells
- Standard paddles paddle surfaces and wells preparation
- Synchronization delay exceeds specification during synchronized shock
- Therapy cable or connector damaged
- Therapy connector spring button not functional
- Transient error after EMI or ESD event
- User Test Failure
- Verify battery charge state
106 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1005 | UI_ERROR_DISPLAY_SELF_TEST; self-test failed | Display self-test failure during power-on or diagnostics | error |
1006 | UI_ERROR_ENERGY_FAULT; defib charge out of tolerance | Defib charge deviates beyond 15% tolerance during manual mode | warning |
1007 | UI_ERROR_12LEAD_KEY_SEEN; 12-lead key detected in non-configured unit | Software detected 12-lead key closure but unit not configured for 12-lead | error |
1008 | UI_ERROR_ANALYZE_KEY_SEEN; AED mode key detected in non-AED unit | Software detected AED key closure in unit not configured for AED | error |
100a | UI_ERROR_NIBP_KEY_SEEN; NIBP key detected in non-NIBP unit | Software detected NIBP key closure in non-NIBP configured unit | error |
100b | UI_ERROR_CURRENTUP_KEY_SEEN; pacing current-up key in non-pacing unit | Software detected current-up key closure in non-pacing configured unit | error |
100c | UI_ERROR_CURRENTDOWN_KEY_SEEN; pacing current-down key in non-pacing unit | Software detected current-down key closure in non-pacing configured unit | error |
100d | UI_ERROR_RATEUP_KEY_SEEN; pacing rate-up key in non-pacing unit | Software detected rate-up key closure in non-pacing configured unit | error |
100e | UI_ERROR_RATEDOWN_KEY_SEEN; pacing rate-down key in non-pacing unit | Software detected rate-down key closure in non-pacing configured unit | error |
100f | UI_ERROR_PACER_KEY_SEEN; pacer key in non-pacing unit | Software detected pacer key closure in non-pacing configured unit | error |
1010 | UI_ERROR_PAUSE_KEY_SEEN; pause key in non-pacing unit | Software detected pause key closure in non-pacing configured unit | error |
1037 | UI_ERROR_UNCONFIGURED_BUTTON; button test detected unexpected key | Button self-test detected key not in current hardware configuration | error |
1038 | UI_ERROR_CPR_KEY_SEEN; CPR key in non-AED unit | Software detected CPR key closure in unit not configured for AED | error |
103b | UI_ERROR_MISSING_LANGUAGE; configured language unavailable | System attempted to load configured language that is not available | error |
1fff | UI_ERROR_EXTRA_INFORMATION; supplemental error code data | Provides additional context for preceding UI error | info |
3005 | DM_ERROR_DATABASE_ERASE_ADJUST; unable to write new records | Cannot write new database records during cleanup | error |
3006 | DM_ERROR_DATABASE_ERASE_FAILED; erase block failed | Flash erase operation failed on specified block | error |
3007 | DM_ERROR_DATABASE_ERASE_VERIFY; erased block verification failed | Verification of erased flash block returned invalid data | error |
3008 | DM_ERROR_DATABASE_FLASH_ERASE; database erase failed | Flash erase operation failed during database management | error |
3fff | DM_ERROR_EXTRA_INFORMATION; supplemental DM error data | Provides additional context for preceding data management error | info |
4009 | SM_ERROR_RAM_FAILURE; RAM failure during self-test | 16-bit RAM test failure detected during power-on self-test | critical |
400a | SM_ERROR_BAD_CRC; program flash CRC failure | CRC verification of program flash failed after power-on | critical |
400b | SM_ERROR_CRC_FAILURE; program contents failed CRC test | Program memory contents do not match computed CRC value | critical |
400c | SM_ERROR_VOLTAGE_LOW; ADC voltage reading low | Hardware voltage measurement below acceptable range | error |
400d | SM_ERROR_VOLTAGE_HIGH; ADC voltage reading high | Hardware voltage measurement above acceptable range | error |
4010 | SM_ERROR_SERVICE_LED; service LED failure | Service LED expected to be illuminated but is not | warning |
4011 | SM_ERROR_DEFIB_SERVICE_SYNC; defib sync failed after cold boot | Failed to synchronize defib charge after cold power cycle | error |
4012 | SM_ERROR_FONT_VOICE_CKSUM; invalid checksum in font/voice | Checksum validation of font/voice data failed at cold boot | error |
4013 | SM_ERROR_FONT_VOICE_CRC; invalid CRC in font/voice flash | CRC validation of font/voice data failed at cold boot | error |
5002 | PC_ERROR_WATCHDOG_SHORT_FAILURE; watchdog short test failed | Main watchdog short duration test failed during diagnostics | error |
5003 | PC_ERROR_WATCHDOG_LONG_FAILURE; watchdog long test failed | Main watchdog long duration test failed during diagnostics | error |
5004 | PC_ERROR_WATCHDOG_UNEXPECTED; unexpected watchdog reset occurred | Processor detected unexpected watchdog reset during operation | warning |
5005 | PC_ERROR_RAM_AT_BOOT; RAM error detected during boot | RAM failure detected during power-on self-test sequence | critical |
5006 | PC_ERROR_BAD_CHECKSUM; program checksum verification failed | Program contents do not match stored checksum value | critical |
500d | PC_ERROR_CONFIG_VERSION; system configuration version mismatch | System configuration version does not match expected version | error |
500e | PC_ERROR_CONFIG_CRC; NVRAM system configuration CRC bad | Configuration data stored in NVRAM fails CRC validation | error |
5011 | PC_ERROR_METERS_VERSION; system meters version mismatch | System meters/counters version does not match expected version | error |
5012 | PC_ERROR_METERS_CRC; NVRAM meters/counters CRC bad | Meters and counters data stored in NVRAM fails CRC validation | error |
5013 | PC_ERROR_MFG_DATA_VERSION; manufacturing data version mismatch | Manufacturing data version does not match expected version | error |
5014 | PC_ERROR_MFG_DATA_CRC; NVRAM manufacturing data CRC bad | Manufacturing data stored in NVRAM fails CRC validation | error |
5015 | PC_ERROR_FORCED_RESET_FAILED; forced watchdog reset failed | Processor failed to reset after watchdog forced reset attempt | critical |
5019 | PC_ERROR_RTC_BAD; RTC is not running | Real-time clock stopped or not responding during boot | error |
501a | PC_ERROR_RTC_DRIFT; processor and RTC time out of sync | Processor system clock and RTC clock drift beyond tolerance | warning |
5032 | PC_ERROR_RTC_NO_BATTERY; no coin cell battery for RTC | Coin cell battery for real-time clock is missing or dead | error |
5033 | PC_ERROR_RTC_READ; RTC read is not consistent | RTC returns inconsistent data on successive read attempts | error |
600c | ECG_ERROR_DSP_VOLTAGE; DSP preamp supply voltage out of range | DSP preamplifier power supply voltage outside acceptable limits | error |
600d | ECG_ERROR_PREAMP_CALIBRATION; calibration constants out of range | NVRAM calibration constants found to be outside valid range | error |
600e | ECG_ERROR_NVRAM_FAULT; NVRAM redundant value mismatch | Redundant NVRAM values do not match during verification | error |
900e | PPSP_ERROR_CONFIG; SpO2 module configuration mismatch | SpO2 module detected but not in manufacturing configuration | error |
902d | PPSP_ERROR_MNC_CONFIG_B; SpO2 Nellcor configuration mismatch | SpO2 module detected but config not set for Nellcor sensor support | error |
9119 | PPNI_ERROR_CONFIG_MISMATCH; NIBP module configuration mismatch | NIBP module detected but unit not configured for it | error |
9205 | PPCO_ERROR_WRITING_TO_MODULE; CO2 module configuration mismatch | CO2 module detected but not in manufacturing configuration | error |
9304 | PPTI_CONFLICT_IP_TEMPER_ERROR_CONFIG; conflicting module configuration | IP and Temperature module config bits both set simultaneously | error |
a002 | TH_ERROR_DEFIB_LINK_DOWN; lost defib processor communication | Serial communications link between main and defib processor failed | critical |
a003 | TH_ERROR_PACER_LINK_DOWN; lost pacer processor communication | Serial communications link between main and pacer processor failed | critical |
a004 | TH_ERROR_UNEXPECTED_ENERGY; unexpected energy in capacitor | Energy detected in capacitor when none should be present | error |
a005 | TH_ERROR_CAP_OVERCHARGED; over-charged capacitor | Capacitor charge exceeded maximum safe limit during charging | error |
a008 | TH_ERROR_DEFIB_DISABLE; no communication with defib hardware | No response from defib hardware module during initialization | critical |
a00a | TH_ERROR_DE_WRONG_ENG_SELECT; incorrect energy selected | Energy selection does not match user request or stored configuration | error |
a00b | TH_ERROR_SHOCK_NOT_DELIVERED; shock delivery failed | Shock was not successfully delivered to patient | critical |
a00d | TH_ERROR_CHARGING_EXPIRED; charging time exceeded | Capacitor charging operation exceeded maximum time limit | error |
a00f | TH_ERROR_CAP_OUT_OF_RANGE; capacitor out of calibration range | Capacitor voltage measurement outside acceptable calibration limits | error |
a010 | TH_ERROR_PA_RATE_OUT_OF_RANGE; pacing rate out of tolerance | Pacer rate deviates beyond acceptable limits from selected rate | error |
a011 | TH_ERROR_PA_CURRENT_OUT_OF_RANGE; pacing current out of tolerance | Pacing current deviates beyond acceptable limits from selected current | error |
a017 | TH_ERROR_DEFIB; defibrillation error report | Generic defibrillation system error during therapy operation | error |
a018 | TH_ERROR_PACER; pacing error report | Generic pacing system error during therapy operation | error |
a01a | TH_ERROR_PACER_FAULT; pacing parameter limit exceeded | Pacing rate, current, or pulse width exceeded safe operating limits | error |
a01b | TH_ERROR_DEFIB_WDT_DISABLE_FAIL; unable to turn off defib watchdog | Failed to disable defib subsystem watchdog timer during shutdown | error |
a020 | TH_ERROR_PACER_DISABLE; pacer disabled communication lost | Cannot communicate with pacer processor; pacer function disabled | error |
a021 | TH_ERROR_CAP_CHARGE_FAIL; capacitor stays zero while charging | Capacitor voltage remains at zero despite charging operation | error |
a022 | TH_ERROR_CORRUPT_ENERGY_SELECT; energy selection corrupted | Energy selection value in memory does not match expected format | error |
a023 | TH_ERROR_XFER_ENABLE_ON; transfer-enable line high at startup | Defib transfer-enable line not properly disabled during system startup | error |
a024 | TH_ERROR_VCAP2_SATURATED; VCAP2 reading full scale continuously | Capacitor voltage monitoring line 2 reading at maximum value | error |
a026 | TH_ERROR_ENERGY_RESIDUE; residual energy on capacitor | Energy remains on capacitor after discharge or dump operation | error |
a027 | TH_ERROR_PA_RATE_CORRUPT; pacing rate storage corrupted | Stored pacing rate value does not match expected format or range | error |
a028 | TH_ERROR_CAL_ENERGY_FAIL; calibrated voltage out of range | Calibrated voltage value exceeds acceptable limits during calibration | error |
a029 | TH_ERROR_BTE_FAIL; BTE board error condition | Biphasic Therapy Electronics board reported fault or control error | error |
a02b | TH_ERROR_DEFIB_CONFIG; conflicting defib type configuration | Defib type in hardware does not match software configuration | error |
a02c | TH_ERROR_DUMP_LINE_FAIL; capacitor dump line failed | Capacitor dump circuit failed during self-test or operation | error |
a02d | TH_ERROR_WRONG_DEFIB_TYPE; wrong defib type for software | Defibrillator hardware type does not match loaded software version | error |
a02e | TH_ERROR_ADC_READ; ADC read failure for cap charge | ADC failed to return valid reading when measuring capacitor charge | error |
a101 | DE_ERROR_LONG_WDT; long watchdog test failed | Long watchdog timer test failed or watchdog did not reset in time | error |
a102 | DE_ERROR_SHORT_WDT; short watchdog test failed | Short watchdog timer test failed during diagnostics | error |
a103 | DE_ERROR_SCI_RCV; serial port receiver error | Serial communication receiver detected error during data reception | error |
a104 | DE_ERROR_XFER_KEY; defib hardware error | Transfer key or enable line hardware fault detected | error |
a106 | DE_ERROR_ENERGY_OUT_OF_BOUND; VCAP-1 over/under charge | Capacitor 1 voltage exceeds upper or falls below lower limits | error |
a107 | DE_ERROR_HP_ENG_SELECT; cannot determine rotary energy setting | Energy selection rotary switch position cannot be read | error |
a109 | DE_ERROR_CAL_CRC; calibration table CRC error | CRC verification of calibration table failed | error |
a10b | DE_ERROR_CHG_TIME; charging takes too long to reach target | Capacitor charging operation exceeded acceptable time limit | error |
a10c | DE_ERROR_CHG_ENABLE; CHG_EN1 line stuck high | Charge enable control line remains high when it should be low | error |
a10d | DE_ERROR_DUMP_ENERGY; defib hardware error during dump | Capacitor dump operation failed or reported hardware error | error |
a10e | DE_ERROR_RCV; defib hardware communication error | Defib hardware failed to respond to receiver command | error |
a111 | DE_ERROR_ENERGY_NOT_ZERO; unexpected energy during biphasic test | Capacitor energy detected during high-pot test when none expected | error |
a112 | DE_ERROR_TEST_XFER_ENABLE; transfer enable control problem | Transfer enable line did not respond during main test | error |
a113 | DE_ERROR_TEST_XFER_ENGAGE; transfer engage feedback error | Defib hardware failed to report transfer engage feedback status | error |
a114 | DE_ERROR_TEST_DUMP_RELAY; biphasic dump relay test failed | Biphasic dump relay failed during self-test procedure | error |
a115 | DE_ERROR_TEST_ADC; ADC hardware test failed | Analog-to-digital converter hardware test failed | error |
a116 | DE_ERROR_TEST_DAC; DAC hardware test failed | Digital-to-analog converter hardware test failed | error |
a117 | DE_ERROR_TEST_HARDWARE; general defib hardware test failure | Generic defibrillator hardware test failed | error |
a118 | DE_ERROR_TEST_RAM; defib RAM test failed | RAM test on defib processor failed | error |
a119 | DE_ERROR_TEST_ROM; defib ROM test failed | ROM test on defib processor failed | error |
a11a | DE_ERROR_TEST_CPU; defib CPU test failed | Defib processor CPU test failed | error |
a11d | DE_ERROR_XFER_TIMEOUT; transfer operation timeout error | Defib transfer operation did not complete within time limit | error |
a11e | DE_ERROR_BUTTONS_UP; defib hardware button error | Defib hardware button test failed | error |
a11f | DE_ERROR_SYNC_INTERRUPT; defib sync interrupt error | Synchronization interrupt signal from defib hardware invalid | error |
a120 | DE_ERROR_SELF_TEST_INCOMPLETE; defib self-test failed to complete | Defib hardware self-test sequence did not complete successfully | error |
a123 | DE_ERROR_CAL_RCV_CRC; CRC failed for new calibration data | CRC verification failed when receiving new calibration data | error |
a124 | DE_ERROR_CAL_NVM_CRC; calibration write to EEPROM failed | Cannot write energy table to nonvolatile memory due to CRC error | error |
a125 | DE_ERROR_DAC_WRITE; DAC byte write failed via SPI | Failed to write calibration byte to DAC through SPI interface | error |
a126 | Cannot read from ADC | ADC read operation failed | error |
a127 | Must be idle to switch to test mode | Attempt to enter test mode while device not idle | warning |
a129 | Defib hardware error - transfer cable issue | Defib transfer cable failure detected | error |
a12a | Defib hardware error - transfer paddle issue | Defib transfer paddle failure detected | error |
a12c | Defib hardware error - charge inhibit failure | Charge inhibit circuit failure detected | error |
a12d | Charge enable feedback indicates not charging | Charging process initiated but feedback shows no charge occurring | error |
a12e | Cedar BTE fault line state detected | BTE fault signal asserted | error |
a12f | Cedar BTE fault line state cleared | BTE fault signal de-asserted | info |
a130 | Defib hardware error - BTE reset | BTE (Biphasic Therapy Engine) reset signal issued | error |
a131 | Biphasic hardware not found | System boot or initialization detects missing Biphasic module | critical |
a132 | Defib hardware error - BTE transfer failure | Data transfer to/from Biphasic module failed | error |
a133 | A/D high bit sticky - Cedar BTE A/D converter fault | A/D converter high bit remains fixed during conversion | error |
a201 | Long watchdog timer test failed | Watchdog timer did not respond correctly during test | error |
a202 | Short watchdog timer test failed | Short watchdog timer did not respond during test | error |
a203 | SCI received overrun, framing or parity error | Serial communication interface detects data corruption | warning |
a204 | 2 ms ECG sampling overrun | ECG sampling processor unable to complete within 2 ms cycle | error |
a208 | Received message incomplete | Communication message reception detected as incomplete | error |
a209 | Received message size error or input buffer full | Message size incorrect or input buffer capacity exceeded | error |
a20d | Pacing pulse process overrun | Pacing pulse generation process exceeds time budget | error |
a20e | Pacing pulse too long | Pacing pulse width exceeds maximum allowable duration | error |
a20f | Internal A/D conversion timeout | Internal analog-to-digital converter operation did not complete in time | error |
a210 | External A/D conversion timeout | External analog-to-digital converter operation did not complete in time | error |
a211 | SPI transfer timeout | Serial Peripheral Interface data transfer did not complete in time | error |
a212 | RAM test failed | Memory self-test detected RAM failure | critical |
a213 | ROM CRC test failed | Read-only memory checksum verification failed | critical |
a214 | Stack overrun occurred | CPU stack pointer exceeded allocated stack memory | critical |
a215 | Isolated +5 V reference out of range | Isolated +5 V supply voltage measured outside specification | error |
a216 | High voltage present when not pacing | HV detected on output when pacing not active | critical |
a217 | +12 V voltage out of range | +12 V supply voltage measured outside specification | error |
a218 | High voltage present when not pacing | HV detected on output when pacing not active | critical |
a219 | HVIS voltage out of range | HVIS (High Voltage Isolated Sense) voltage measured outside specification | error |
a21a | Current calibration failed | Pacer output current calibration procedure unsuccessful | error |
a21b | Impedance 300 ohm calibration failed | Impedance calibration at 300 ohm load unsuccessful | error |
a21c | Current calibration table corrupted | CRC check of current calibration data failed | error |
a21d | Impedance calibration values corrupted | CRC check of impedance calibration data failed | error |
a21e | Impedance 0 ohm calibration failed | Impedance calibration at 0 ohm load unsuccessful | error |
a21f | Pacing current present when not pacing | Output current detected when pacing mode inactive | critical |
a220 | No high voltage present when pacing | Expected HV output not detected during pacing | critical |
a221 | External A/D test register reset failed | A/D converter test register did not reset properly during diagnostic | error |
a222 | No high voltage present before current calibration | HV not detected during calibration setup procedure | error |
a223 | No Q4 sense when not pacing (Q4 shorted) | Q4 sense signal missing when pacing inactive | error |
a224 | Q4 sense present when pacing (CR25 open) | Q4 sense signal present when pacing active | error |
b001 | Printhead ADC reading too low; temperature error | Printhead thermistor ADC value below threshold | error |
b002 | Printhead ADC reading too high; temperature error | Printhead thermistor ADC value above threshold | error |
b003 | Printer self-test failed; invalid CRC | Hardware self-test detects incorrect CRC from printer module | error |
b004 | Wrong printer type detected; 50/100 mismatch | Installed printer type does not match device configuration | error |
c008 | Lost communications with power PCB | Serial communication with power management board failed | error |
c009 | Power PCB running on internal oscillator; external clock failure | Power PCB detects external clock failure during initialization | warning |
d003 | SCI frame error | Serial communication interface detects frame synchronization error | warning |
d004 | SCI parity error | Serial communication parity check fails | warning |
d005 | SCI read buffer full | Serial communication input buffer capacity exceeded | warning |
d006 | SCI initialization failed | Serial communication interface did not initialize during startup | error |
d00d | Message queue full; queue overflow | Internal message queue capacity exceeded | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.