Physio-Control LIFEPAK 15: troubleshooting and error codes

The ServiceForge knowledge base holds 166 documented symptoms for the Physio-Control LIFEPAK 15, with 230 ranked causes, 374 diagnostic checks with test points and expected values, and 162 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

No energy discharge from paddles high voltage

QUIK-COMBO or hard paddles fail to deliver electrical energy

Ranked causes

  1. 1. Test setup or procedure error (common)
  2. 2. Defective therapy cable or paddles (occasional)
  3. 3. W11 Therapy Connector Cable failure (occasional)
  4. 4. A04 Therapy PCB failure (rare)
  5. 5. A22 Biphasic Module PCB failure (rare)
  6. 6. A14 Inductive Resistor failure (rare)
  7. 7. A15 Energy Storage Capacitor failure (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 saturation reading missing or out of tolerance patient safety

SpO2 module displays XXX and Service LED OFF, indicates sensor/reading issue

Ranked causes

  1. 1. Test setup error (common)
  2. 2. Defective finger sensor (occasional)
  3. 3. Device configuration error (occasional)
  4. 4. A16 SpO2 Module failure (rare)
  5. 5. W22 SpO2 Connector Cable failure (rare)
  6. 6. A06 OEM PCB failure (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 fails calibration patient safety

CO2 module cannot complete calibration procedure

Ranked causes

  1. 1. Test setup error or empty calibration gas (common)
  2. 2. FilterLine disconnected (occasional)
  3. 3. Pinched or kinked tubing inside device (occasional)
  4. 4. A23 CO2 module failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Improper key response

Keypad keys do not respond properly to presses

Ranked causes

  1. 1. A09 Printer Control Keypad failure (occasional)
  2. 2. A10 Main Keypad failure (occasional)
  3. 3. A05 Interface PCB failure (rare)
  4. 4. A01 System PCB failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Missing dots in printed X pattern

Printer test pattern X has missing dots in output

Ranked causes

  1. 1. Improper printer paper in use (common)
  2. 2. Dirty printhead (common)
  3. 3. A12 Printer Assembly failure (occasional)
  4. 4. A01 System PCB failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Missing or broken characters in print output

Printed characters are incomplete or missing

Ranked causes

  1. 1. Improper printer paper (common)
  2. 2. Dirty printhead (common)
  3. 3. A12 Printer Assembly failure (occasional)
  4. 4. A01 System PCB failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No pacer output high voltage

Pacing function does not deliver pacing pulses

Ranked causes

  1. 1. Test setup error (common)
  2. 2. A13 Transfer Relay connection problem at P22 (occasional)
  3. 3. A13 Transfer Relay failure (occasional)
  4. 4. A04 Therapy PCB failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No power ON caution

Device will not power on when turned on

Ranked causes

  1. 1. Battery not fully charged or improperly maintained (common)
  2. 2. Loose or broken battery pin(s) (occasional)
  3. 3. W04 System/Interface PCB cable connection issue (occasional)
  4. 4. A03 Power PCB failure (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ABNORMAL ENERGY DELIVERY message appears high voltage

Device displays abnormal energy delivery alert

Ranked causes

  1. 1. A13 Transfer Relay connection problem at P22 (occasional)
  2. 2. A13 Transfer Relay failure (occasional)
  3. 3. A04 Therapy PCB failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CHECK PRINTER message appears

Device displays CHECK PRINTER alert message

Ranked causes

  1. 1. Paper sensor dirty (common)
  2. 2. Printer paper not correctly loaded (common)
  3. 3. A12 Printer Assembly failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 displays FilterLine Blockage message patient safety

CO2 module indicates FilterLine is blocked

Ranked causes

  1. 1. FilterLine blockage (common)
  2. 2. Occlusion or kinking of input tubing (occasional)
  3. 3. A23 CO2 module failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 displays XXX with Service LED OFF patient safety

CO2 module shows XXX reading and Service LED is OFF

Ranked causes

  1. 1. FilterLine blockage (common)
  2. 2. Occlusion or kinking of internal tubing (occasional)
  3. 3. A23 CO2 module failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failure to transfer energy at Sync mark high voltage

Energy discharge does not occur at sync mark during synchronized cardioversion

Ranked causes

  1. 1. Device in sync mode with discharge attempt needed (common)
  2. 2. A10 Main Keypad failure (occasional)
  3. 3. Defective therapy cables or paddles (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Improper LCD response

LCD display does not respond normally or shows degraded performance

Ranked causes

  1. 1. Display test failure indication (common)
  2. 2. A11 LCD Assembly failure (occasional)
  3. 3. A01 System PCB failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Inaudible or garbled audio patient safety

Audio output is not heard or is distorted

Ranked causes

  1. 1. W17 Speaker Assembly failure (common)
  2. 2. A05 Interface PCB failure (occasional)
  3. 3. A01 System PCB failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure P1/P2 channel reading out of tolerance patient safety

Mean Arterial Pressure (MAP) reading does not match simulated pressure input within tolerance window

Ranked causes

  1. 1. Pressure transducer requires zeroing or recalibration (common)
  2. 2. Invasive pressure connector loose or not properly seated (occasional)
  3. 3. Pressure transducer or cable fault (rare)
First check

Set simulator to ZERO output and zero pressure channel

Test point
P1/P2 channel display and MAP reading
Expected
-20 ±3 mmHg

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No Sync mark displayed high voltage

Synchronous cardioversion sync mark not visible on ECG waveform

Ranked causes

  1. 1. Test setup error (common)
  2. 2. ECG size adjustment needed (common)
  3. 3. A01 System PCB failure (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

3-Lead ECG leads off detection failure patient safety

Device does not display leads-off message or sound repeating priority 3 tone when individual ECG lead is disconnected during 3-lead test

Ranked causes

  1. 1. ECG cable connection loose or missing lead contact (common)
  2. 2. Leads-off detection circuit malfunction (occasional)
First check

Verify LL lead off detection at LEAD II

Test point
LL ECG lead connector
Expected
LL LEADS OFF message and repeating priority 3 tone
Tools
3-lead ECG cable, Impulse 7000DP simulator

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

5-Lead ECG leads off detection failure patient safety

Device does not display leads-off message or sound repeating priority 3 tone when individual ECG lead is disconnected during 5-lead test

Ranked causes

  1. 1. ECG cable connection loose or missing lead contact (common)
  2. 2. Leads-off detection circuit malfunction (occasional)
First check

Verify LL lead off detection at LEAD II

Test point
LL ECG lead connector
Expected
LL LEADS OFF message and repeating priority 3 tone
Tools
5-wire ECG cable, Impulse 7000DP simulator

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Batteries not fully charged or showing low battery warning patient safety

Lithium-ion batteries do not show more than two charge bars or LOW BATTERY message appears after powering on

Ranked causes

  1. 1. Batteries not fully charged before use (common)
  2. 2. Battery defective or worn out (occasional)
First check

Inspect charge bar display on batteries

Test point
Battery status indicators on installed batteries
Expected
More than two charge bars visible on each battery

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery connector pins damaged or loose high voltage

Battery connector pins are loose, cracked, or broken, preventing reliable power connection

Ranked causes

  1. 1. Loose battery connector pins (occasional)
  2. 2. Cracked or broken connector pin leaves (occasional)
First check

Visually inspect battery connector pins for tightness

Test point
Battery connector pins at bottom of device
Expected
Pins firmly seated, no movement

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery does not click into position in battery well patient safety

Lithium-ion battery fails to seat properly or make audible/tactile contact in battery compartment

Ranked causes

  1. 1. Battery contacts or well damaged or obstructed (occasional)
  2. 2. Battery case deformed or defective (rare)
First check

Attempt to insert battery into well and listen/feel for click

Test point
Battery well and battery case contact points
Expected
Audible or tactile click when battery seats fully

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery icons not visible when power adapter connected patient safety

Battery indicators do not appear on display when external AC/DC power adapter is connected to Auxiliary Connector

Ranked causes

  1. 1. Power adapter not properly connected or Auxiliary Connector issue (common)
  2. 2. Display or power management circuit malfunction (unknown)
First check

Verify power adapter cable is fully inserted into Auxiliary Connector

Test point
Auxiliary Connector at rear of device
Expected
Cable fully seated with no gaps

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery not fully charged or functional

Lithium-ion batteries do not show sufficient charge bars or return LOW BATTERY message when powered on

Ranked causes

  1. 1. Battery insufficient charge (common)
  2. 2. Battery defective or failed (occasional)
First check

Verify battery charge state

Test point
Battery status indicator on device display
Expected
More than two charge bars showing on both batteries

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bluetooth wireless pairing failure

Device fails to pair with computer via Bluetooth, no alert tone, LED not illuminated, or connection message does not appear.

Ranked causes

  1. 1. Bluetooth wireless disabled or not configured (common)
  2. 2. Passcode mismatch between device and computer (occasional)
First check

Verify CODE-STAT Reviewer application open on computer

Test point
Software setup
Expected
CODE-STAT Reviewer running, DOWNLOAD WIZARD accessible

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

CO2 FilterLine installed but no indication on display

FilterLine connected but device does not show CO2 is connected

Ranked causes

  1. 1. CO2 input connector dirty (common)
  2. 2. FilterLine defective (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Defibrillator sync function failure or delay out of range patient safety

SYNC LED not illuminating, R-wave markers not appearing, or sync delay exceeds 60 ms

Ranked causes

  1. 1. ECG cable connection or signal quality issue (common)
  2. 2. R-wave detection circuit malfunction or sync delay circuit degradation (occasional)
First check

Verify SYNC LED illuminates when SYNC button pressed

Test point
SYNC button and LED
Expected
SYNC LED on; R-wave markers visible on ECG
Tools
Defibrillator analyzer, main ECG cable

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Device shows no power indication after case reassembly

No indication that power is on after case is reassembled and device is turned ON

Ranked causes

  1. 1. Batteries are dead or not installed (common)
  2. 2. W04 cable connection not made during case reassembly (common)
First check

Verify batteries are installed and charged

Test point
Battery compartment
Expected
Fresh batteries installed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure P1 channel calibration and verification patient safety

Verification that P1 invasive pressure channel displays correct MAP values at multiple pressure points with proper zero offset

Ranked causes

  1. 1. Pressure transducer calibration drift or offset error (occasional)
  2. 2. Patient simulator not configured correctly (occasional)
First check

Connect invasive pressure cable to P1 connector with simulator at ZERO

Test point
P1 connector on parameter bezel
Expected
Connection established and display shows no error

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invasive pressure P1 channel does not read correctly patient safety

P1 pressure waveform or MAP display reads outside tolerance when simulated pressures applied, or channel will not zero

Ranked causes

  1. 1. P1 transducer calibration drift or sensor failure (common)
  2. 2. Invasive pressure cable disconnected or faulty (occasional)
First check

Verify cable connection to P1 connector

Test point
P1 connector on parameter bezel
Expected
Cable firmly seated

6 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keypad buttons not responding or stuck patient safety

Button does not highlight when pressed, or button remains highlighted at test start indicating short circuit

Ranked causes

  1. 1. Keypad has internal short circuit (occasional)
  2. 2. Button contact failure or defective key switch (occasional)
First check

Verify no buttons highlighted at start of test

Test point
Keypad button display
Expected
No buttons highlighted

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Loose or rattling hardware caution

Hardware inside device is loose, causing rattling noise when device is moved or inverted

Ranked causes

  1. 1. Fasteners not properly tightened (common)
  2. 2. Broken or missing fasteners (occasional)
First check

Invert device and listen for rattling

Test point
Entire device assembly
Expected
No rattling or loose hardware sounds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NIBP displays XXX with Service LED OFF patient safety

NIBP module displays XXX and Service LED is OFF, indicates measurement failure

Ranked causes

  1. 1. NIBP leakage or tubing blockage (common)
  2. 2. A21 NIBP module failure (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Therapy cable not connecting to therapy connector properly high voltage

QUIK-COMBO or standard paddles fail to connect securely to therapy connector

Ranked causes

  1. 1. Connector misaligned or contacts obstructed (common)
  2. 2. Connector contacts damaged or corroded (occasional)
First check

Verify therapy cable is fully seated in therapy connector

Test point
Therapy connector on device and cable connector
Expected
Cable fully inserted with no visible gaps or movement when gently pulled

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

12-lead ECG fails leads-off detection patient safety

Device does not display leads-off message or priority 3 tone does not sound when individual ECG lead is disconnected during acquisition

Ranked causes

  1. 1. Lead detection circuit failure or ECG cable fault (occasional)
First check

Connect main ECG cable to Impulse simulator

Test point
ECG cable limb and precordial lead attachments
Expected
All 10 ECG leads connected to simulator

5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

12-Lead ECG gain out of range

Printed ECG signal amplitude does not meet specification for one or more leads during 12-lead ECG gain test

Ranked causes

  1. 1. ECG signal chain gain mismatch or calibration drift (unknown)
First check

Measure 12-lead ECG gain amplitude for Lead I

Test point
ECG output
Expected
25 mm to 31 mm peak-to-peak, 25-31
Tools
ECG printer, Impulse 7000DP simulator

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

12-Lead ECG gain test

Verification that ECG signal amplitudes are correct for each lead at 4.0 ECG size setting

Ranked causes

  1. 1. ECG gain setting incorrect or amplifier circuit drift (occasional)
First check

Set simulator to 1 mV, 10 Hz sine wave and device ECG SIZE to 4.0

Test point
Device ECG SIZE setting and simulator output
Expected
Configuration complete for amplitude testing

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

12-Lead ECG leads-off detection test

Verification that device detects open or disconnected leads and displays appropriate alert messages with priority 3 tone

Ranked causes

  1. 1. ECG lead connection interrupted or cable damaged (common)
First check

Connect 12-lead ECG cable to patient simulator

Test point
Main ECG cable with limb and precordial lead attachments
Expected
All 10 leads connected to simulator

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

12-lead ECG signal amplitude out of range

Printed ECG signal amplitude does not meet specified peak-to-peak values on one or more leads when testing with Impulse 7000DP signal generator

Ranked causes

  1. 1. ECG signal gain incorrect or attenuated (common)
First check

Measure ECG Lead I printed signal amplitude

Test point
Printer output, Lead I
Expected
25 mm to 31 mm peak-to-peak
Tools
Impulse 7000DP, printer, ruler or calipers

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3-lead ECG gain incorrect

Printed ECG signal amplitude outside acceptable range for Leads I and II

Ranked causes

  1. 1. ECG amplifier gain miscalibrated (occasional)
First check

Set simulator to 1-mV, 10-Hz sine wave

Test point
Simulator output
Expected
1-mV, 10-Hz sine wave

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3-Lead ECG gain out of range

Printed ECG signal amplitude does not meet specification for Lead I or Lead II during 3-lead ECG gain test

Ranked causes

  1. 1. ECG signal chain gain mismatch or calibration drift (unknown)
First check

Measure 3-lead ECG gain amplitude for Lead I

Test point
ECG output
Expected
25 mm to 31 mm peak-to-peak, 25-31
Tools
ECG printer, Impulse 7000DP simulator

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3-lead ECG leads off detection not functioning patient safety

Device does not display correct LEADS OFF message or priority 3 tone does not sound when a 3-lead ECG cable lead is disconnected

Ranked causes

  1. 1. ECG lead connection loose or not properly seated (common)
First check

Verify LL LEADS OFF detection with 3-lead cable

Test point
ECG LL lead connection
Expected
LL LEADS OFF message displayed and priority 3 tone sounds

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3-lead ECG leads-off detection not alerting patient safety

Device should display specific LEADS OFF message and priority 3 tone when leads disconnected from patient simulator

Ranked causes

  1. 1. Lead not properly connected to patient simulator (common)
First check

Connect 3-lead ECG cable to device and simulator

Test point
3-lead cable connectors
Expected
Properly seated connections

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3-lead ECG signal amplitude out of range

Printed ECG signal amplitude does not meet specified peak-to-peak values when testing 3-lead cable with Impulse 7000DP signal generator

Ranked causes

  1. 1. ECG signal gain incorrect or signal path degraded (common)
First check

Measure 3-lead ECG Lead I printed signal amplitude

Test point
Printer output, Lead I
Expected
25 mm to 31 mm peak-to-peak
Tools
Impulse 7000DP, printer, ruler or calipers

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

5-lead ECG gain incorrect

Printed ECG signal amplitude outside acceptable range for Leads I and II

Ranked causes

  1. 1. ECG amplifier gain miscalibrated (occasional)
First check

Set patient simulator to 1-mV, 10-Hz sine wave

Test point
Patient simulator output
Expected
1-mV, 10-Hz sine wave

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

5-Lead ECG gain out of range

Printed ECG signal amplitude does not meet specification for Lead I or Lead II during 5-lead ECG gain test

Ranked causes

  1. 1. ECG signal chain gain mismatch or calibration drift (unknown)
First check

Measure 5-lead ECG gain amplitude for Lead I

Test point
ECG output
Expected
25 mm to 31 mm peak-to-peak, 25-31
Tools
ECG printer, Impulse 7000DP simulator

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

5-lead ECG leads off detection not functioning patient safety

Device does not display correct LEADS OFF message or priority 3 tone does not sound when an ECG lead is disconnected during 5-lead testing

Ranked causes

  1. 1. ECG lead connection loose or not properly seated (common)
First check

Verify LL LEADS OFF detection

Test point
ECG LL lead connection
Expected
LL LEADS OFF message displayed and priority 3 tone sounds

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

5-lead ECG leads-off detection not alerting patient safety

Device should display specific LEADS OFF message and priority 3 tone when individual leads disconnected from patient simulator

Ranked causes

  1. 1. Lead not properly connected to patient simulator (common)
First check

Remove RL lead and verify alert

Test point
RL lead connection
Expected
RL LEADS OFF message and priority 3 tone

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

5-Lead ECG leads-off detection test

Verification that device detects open or disconnected leads on 5-wire ECG cable and displays appropriate alerts

Ranked causes

  1. 1. 5-wire ECG lead connection interrupted or cable damaged (common)
First check

Connect 5-wire ECG cable to patient simulator

Test point
5-wire ECG cable connections
Expected
All 5 leads connected (RA, LA, RL, LL, V1)

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

5-lead ECG signal amplitude out of range

Printed ECG signal amplitude does not meet specified peak-to-peak values when testing 5-lead cable with Impulse 7000DP signal generator

Ranked causes

  1. 1. ECG signal gain incorrect or signal path degraded (common)
First check

Measure 5-lead ECG Lead I printed signal amplitude

Test point
Printer output, Lead I
Expected
25 mm to 31 mm peak-to-peak
Tools
Impulse 7000DP, printer, ruler or calipers

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Accessory cables or sensors expired or unsuitable patient safety

ECG, SpO2 sensors, CO2 tubing, NIBP tubing, temperature sensors, or related items are expired, damaged, or unsuitable for use

Ranked causes

  1. 1. Expired or damaged accessory items (common)
First check

Verify expiration dates and condition of all accessories

Test point
ECG cables, SpO2 sensors, CO2 tubing, NIBP tubing, temperature sensors
Expected
All items current (not expired), clean, and in good condition

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Active battery indicator not showing correctly patient safety

Device does not display which battery is in use via white battery number in black box, or priority selection (lowest charge first) is not functioning

Ranked causes

  1. 1. Battery charge states not properly detected (occasional)
First check

Active battery indicator display

Test point
Battery status display area
Expected
White battery number in black box indicating active battery

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Analog ECG output amplitude incorrect

Oscilloscope measurement of analog ECG output outside acceptable voltage range

Ranked causes

  1. 1. Analog output stage gain miscalibrated (occasional)
First check

Connect device, simulator, and oscilloscope

Test point
ECG cable and oscilloscope connections
Expected
All connections properly seated

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio output distorted or inaudible patient safety

Voice prompts not clearly audible, reproduced with distortion, or incorrect language

Ranked causes

  1. 1. Speaker defective or audio circuit fault (occasional)
First check

Play voice prompts via VOICE/TONE test

Test point
Device speaker output
Expected
Voice prompts clearly audible, no distortion, correct language

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio output unclear or distorted

Voice prompts from speaker are not clearly audible or contain distortion during audio test

Ranked causes

  1. 1. Speaker malfunction or degraded audio output (unknown)
First check

Confirm voice prompts are clear and undistorted

Test point
Speaker output
Expected
Voice prompts clearly audible without distortion

Audio test for voice prompts and tones

Verification that device produces clear, undistorted voice prompts and tones from speaker

Ranked causes

  1. 1. Speaker not functioning or audio signal distorted (unknown)
First check

Select and start voice/tone test

Test point
Service/Tests submenu VOICE/TONE option
Expected
Voice prompts play from speaker

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auxiliary power switching not functioning high voltage

Device does not properly switch between battery and auxiliary power, or battery icons do not indicate power source correctly

Ranked causes

  1. 1. Auxiliary power connector circuit failure (rare)
First check

Connect power adapter and verify battery icon display

Test point
Battery status indicators on display
Expected
Battery icons appear but neither is highlighted when adapter connected

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery icons not highlighted when power adapter unplugged patient safety

Battery indicator icons do not show highlight after external power adapter is disconnected from Auxiliary Connector

Ranked causes

  1. 1. Power switching circuit malfunction or battery detection failure (unknown)
First check

Unplug power adapter and observe battery icons on display

Test point
LCD display screen after disconnecting Auxiliary Connector cable
Expected
One device battery icon is highlighted

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery power source not correctly displayed or managed

Device does not correctly display battery status indicators, which battery is in use, charge state, or does not properly manage battery switching when batteries are installed or removed

Ranked causes

  1. 1. Battery detection or power management circuit issue (occasional)
First check

Turn device ON and verify battery status display

Test point
Battery status indicators on display
Expected
Indicators show presence of both batteries, charge state on each, which battery in use

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery status indicators not displaying correctly

Device fails to show battery presence, usage status, or charge state on display

Ranked causes

  1. 1. Battery status display malfunction (unknown)
First check

Power on device and observe battery status display

Test point
Display screen
Expected
Battery status indicators visible showing presence of batteries in Wells 1 and 2, which battery in use (white number in black box), and state of charge on each battery

2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Battery well 1 presence not detected
  • Battery well 2 presence not detected
  • Bluetooth wireless connection setup
  • Calibration: Analog ECG output
  • Calibration: CO2 Module
  • Calibration: Defibrillator ECG gain
  • Calibration: Defibrillator Energy
  • Calibration: Defibrillator Isolation
  • Calibration: Defibrillator Output Waveform
  • Calibration: Direct Applied Part Leakage Test - ECG
  • Calibration: Direct Applied Part Leakage Test - SpO2
  • Calibration: Direct Applied Part Leakage Test - Therapy
  • Calibration: Direct Equipment Leakage Test
  • Calibration: ECG Restore Test - Standard Paddles
  • Calibration: Leakage Current - AC Powered Setup
  • Calibration: Leakage Current - Battery Powered Setup
  • Calibration: Pacer LEADs-Off Detection
  • Calibration: Pacer output current
  • Calibration: Pacer Pulse Width
  • Calibration: Patient Impedance Sense
  • Calibration: Performance Inspection Procedures
  • Calibration: Printer at 25 mm
  • Calibration: Quik-Combo defibrillator delivered energy
  • Calibration: QUIK-COMBO ECG characteristics
  • Calibration: R-Wave Detection - Standard Paddles
  • Calibration: Standard paddles delivered energy
  • Calibration: Standard paddles ECG characteristics
  • Calibration: Standard paddles synchronous cardioversion
  • Calibration: Standard paddles user test
  • Calibration: Temperature
  • Carrying strap or mounts damaged or missing
  • Configure Safety Analyzer for IEC 62353 testing
  • Damaged keypad or label
  • Date/Time Incorrect on Printout
  • Defibrillation event signal recovery
  • Defibrillator charge time exceeds specification
  • Defibrillator delivered energy out of range
  • Defibrillator delivered energy outside acceptable range
  • Defibrillator shock marker or energy event not recorded on printout
  • Defibrillator synchronization delay exceeds limit
  • Defibrillator synchronous cardioversion not synchronizing
  • Delivered energy out of tolerance
  • Device does not complete power-on self-test within specification
  • Device does not detect positive R-waves on QUIK-COMBO paddles
  • Device does not use correct battery when two installed
  • Device preparation for Performance Inspection Procedures
  • Display screen does not match expected startup appearance
  • Display screen does not show expected startup screen
  • ECG amplitude does not restore to adequate level after defibrillation
  • ECG analog output signal amplitude out of range
  • ECG baseline does not restore to zero offset after defibrillation
  • ECG gain on QUIK-COMBO paddles incorrect
  • ECG signal amplitude outside specification during gain test
  • ECG signal does not restore to baseline after defibrillation
  • Excessive leakage current on ECG direct applied parts
  • Excessive leakage current on SpO₂ applied parts
  • Excessive leakage current on therapy (defibrillation) applied parts
  • Exterior physical damage or wear
  • Front panel LEDs do not flash during self-test
  • Invasive pressure P2 channel calibration and verification
  • Invasive pressure P2 channel does not read correctly
  • Keypad button does not respond
  • Keypad function test
  • Keypad or overlay damaged
  • Leakage current testing setup - AC powered
  • Leakage current testing setup - battery powered
  • Loose or broken hardware
  • MAINTENANCE DUE message appears
  • Monitor cable or sensor connectors damaged or contaminated
  • One or more horizontal lines missing or distorted
  • Pacer output current outside acceptable range
  • Pacer pulse width out of range
  • Pacing leads-off detection not triggered on electrode disconnect
  • Paddle leads off detection failure
  • Paddles lead-off detection malfunction
  • Positive R-wave not detected or recorded on ECG
  • Power LED not illuminated after self-test
  • Power-on LED does not remain illuminated after self-test
  • Power-on self-test fails or completes slowly
  • Printer door open alert displayed
  • Printer paper not installed or loaded incorrectly
  • Printer speed test at 12.5 mm/sec
  • Printer speed test at 12.5 mm/sec fails
  • Printer speed test at 25 mm/sec fails
  • Rubber feet damaged or missing
  • SERVICE indicator is ON after case reassembly
  • Service LED illuminated after self-test
  • Service LED recurrence after clearing and restart
  • Service LED remains illuminated after self-test
  • Service LED remains ON
  • Setup: TCP workstation configuration
  • SpCO reading outside acceptable range
  • Speaker does not emit clear test tone at power-on
  • Speaker does not emit test tone
  • SpMet reading outside acceptable range
  • SpO2 reading outside acceptable range
  • SpO2/SpCO/SpMet probe or sensor fault
  • SpO2/SpCO/SpMet sensor test
  • Standard Paddles defibrillator delivers energy with paddles in wells
  • Standard paddles paddle surfaces and wells preparation
  • Synchronization delay exceeds specification during synchronized shock
  • Therapy cable or connector damaged
  • Therapy connector spring button not functional
  • Transient error after EMI or ESD event
  • User Test Failure
  • Verify battery charge state

106 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1005UI_ERROR_DISPLAY_SELF_TEST; self-test failedDisplay self-test failure during power-on or diagnosticserror
1006UI_ERROR_ENERGY_FAULT; defib charge out of toleranceDefib charge deviates beyond 15% tolerance during manual modewarning
1007UI_ERROR_12LEAD_KEY_SEEN; 12-lead key detected in non-configured unitSoftware detected 12-lead key closure but unit not configured for 12-leaderror
1008UI_ERROR_ANALYZE_KEY_SEEN; AED mode key detected in non-AED unitSoftware detected AED key closure in unit not configured for AEDerror
100aUI_ERROR_NIBP_KEY_SEEN; NIBP key detected in non-NIBP unitSoftware detected NIBP key closure in non-NIBP configured uniterror
100bUI_ERROR_CURRENTUP_KEY_SEEN; pacing current-up key in non-pacing unitSoftware detected current-up key closure in non-pacing configured uniterror
100cUI_ERROR_CURRENTDOWN_KEY_SEEN; pacing current-down key in non-pacing unitSoftware detected current-down key closure in non-pacing configured uniterror
100dUI_ERROR_RATEUP_KEY_SEEN; pacing rate-up key in non-pacing unitSoftware detected rate-up key closure in non-pacing configured uniterror
100eUI_ERROR_RATEDOWN_KEY_SEEN; pacing rate-down key in non-pacing unitSoftware detected rate-down key closure in non-pacing configured uniterror
100fUI_ERROR_PACER_KEY_SEEN; pacer key in non-pacing unitSoftware detected pacer key closure in non-pacing configured uniterror
1010UI_ERROR_PAUSE_KEY_SEEN; pause key in non-pacing unitSoftware detected pause key closure in non-pacing configured uniterror
1037UI_ERROR_UNCONFIGURED_BUTTON; button test detected unexpected keyButton self-test detected key not in current hardware configurationerror
1038UI_ERROR_CPR_KEY_SEEN; CPR key in non-AED unitSoftware detected CPR key closure in unit not configured for AEDerror
103bUI_ERROR_MISSING_LANGUAGE; configured language unavailableSystem attempted to load configured language that is not availableerror
1fffUI_ERROR_EXTRA_INFORMATION; supplemental error code dataProvides additional context for preceding UI errorinfo
3005DM_ERROR_DATABASE_ERASE_ADJUST; unable to write new recordsCannot write new database records during cleanuperror
3006DM_ERROR_DATABASE_ERASE_FAILED; erase block failedFlash erase operation failed on specified blockerror
3007DM_ERROR_DATABASE_ERASE_VERIFY; erased block verification failedVerification of erased flash block returned invalid dataerror
3008DM_ERROR_DATABASE_FLASH_ERASE; database erase failedFlash erase operation failed during database managementerror
3fffDM_ERROR_EXTRA_INFORMATION; supplemental DM error dataProvides additional context for preceding data management errorinfo
4009SM_ERROR_RAM_FAILURE; RAM failure during self-test16-bit RAM test failure detected during power-on self-testcritical
400aSM_ERROR_BAD_CRC; program flash CRC failureCRC verification of program flash failed after power-oncritical
400bSM_ERROR_CRC_FAILURE; program contents failed CRC testProgram memory contents do not match computed CRC valuecritical
400cSM_ERROR_VOLTAGE_LOW; ADC voltage reading lowHardware voltage measurement below acceptable rangeerror
400dSM_ERROR_VOLTAGE_HIGH; ADC voltage reading highHardware voltage measurement above acceptable rangeerror
4010SM_ERROR_SERVICE_LED; service LED failureService LED expected to be illuminated but is notwarning
4011SM_ERROR_DEFIB_SERVICE_SYNC; defib sync failed after cold bootFailed to synchronize defib charge after cold power cycleerror
4012SM_ERROR_FONT_VOICE_CKSUM; invalid checksum in font/voiceChecksum validation of font/voice data failed at cold booterror
4013SM_ERROR_FONT_VOICE_CRC; invalid CRC in font/voice flashCRC validation of font/voice data failed at cold booterror
5002PC_ERROR_WATCHDOG_SHORT_FAILURE; watchdog short test failedMain watchdog short duration test failed during diagnosticserror
5003PC_ERROR_WATCHDOG_LONG_FAILURE; watchdog long test failedMain watchdog long duration test failed during diagnosticserror
5004PC_ERROR_WATCHDOG_UNEXPECTED; unexpected watchdog reset occurredProcessor detected unexpected watchdog reset during operationwarning
5005PC_ERROR_RAM_AT_BOOT; RAM error detected during bootRAM failure detected during power-on self-test sequencecritical
5006PC_ERROR_BAD_CHECKSUM; program checksum verification failedProgram contents do not match stored checksum valuecritical
500dPC_ERROR_CONFIG_VERSION; system configuration version mismatchSystem configuration version does not match expected versionerror
500ePC_ERROR_CONFIG_CRC; NVRAM system configuration CRC badConfiguration data stored in NVRAM fails CRC validationerror
5011PC_ERROR_METERS_VERSION; system meters version mismatchSystem meters/counters version does not match expected versionerror
5012PC_ERROR_METERS_CRC; NVRAM meters/counters CRC badMeters and counters data stored in NVRAM fails CRC validationerror
5013PC_ERROR_MFG_DATA_VERSION; manufacturing data version mismatchManufacturing data version does not match expected versionerror
5014PC_ERROR_MFG_DATA_CRC; NVRAM manufacturing data CRC badManufacturing data stored in NVRAM fails CRC validationerror
5015PC_ERROR_FORCED_RESET_FAILED; forced watchdog reset failedProcessor failed to reset after watchdog forced reset attemptcritical
5019PC_ERROR_RTC_BAD; RTC is not runningReal-time clock stopped or not responding during booterror
501aPC_ERROR_RTC_DRIFT; processor and RTC time out of syncProcessor system clock and RTC clock drift beyond tolerancewarning
5032PC_ERROR_RTC_NO_BATTERY; no coin cell battery for RTCCoin cell battery for real-time clock is missing or deaderror
5033PC_ERROR_RTC_READ; RTC read is not consistentRTC returns inconsistent data on successive read attemptserror
600cECG_ERROR_DSP_VOLTAGE; DSP preamp supply voltage out of rangeDSP preamplifier power supply voltage outside acceptable limitserror
600dECG_ERROR_PREAMP_CALIBRATION; calibration constants out of rangeNVRAM calibration constants found to be outside valid rangeerror
600eECG_ERROR_NVRAM_FAULT; NVRAM redundant value mismatchRedundant NVRAM values do not match during verificationerror
900ePPSP_ERROR_CONFIG; SpO2 module configuration mismatchSpO2 module detected but not in manufacturing configurationerror
902dPPSP_ERROR_MNC_CONFIG_B; SpO2 Nellcor configuration mismatchSpO2 module detected but config not set for Nellcor sensor supporterror
9119PPNI_ERROR_CONFIG_MISMATCH; NIBP module configuration mismatchNIBP module detected but unit not configured for iterror
9205PPCO_ERROR_WRITING_TO_MODULE; CO2 module configuration mismatchCO2 module detected but not in manufacturing configurationerror
9304PPTI_CONFLICT_IP_TEMPER_ERROR_CONFIG; conflicting module configurationIP and Temperature module config bits both set simultaneouslyerror
a002TH_ERROR_DEFIB_LINK_DOWN; lost defib processor communicationSerial communications link between main and defib processor failedcritical
a003TH_ERROR_PACER_LINK_DOWN; lost pacer processor communicationSerial communications link between main and pacer processor failedcritical
a004TH_ERROR_UNEXPECTED_ENERGY; unexpected energy in capacitorEnergy detected in capacitor when none should be presenterror
a005TH_ERROR_CAP_OVERCHARGED; over-charged capacitorCapacitor charge exceeded maximum safe limit during chargingerror
a008TH_ERROR_DEFIB_DISABLE; no communication with defib hardwareNo response from defib hardware module during initializationcritical
a00aTH_ERROR_DE_WRONG_ENG_SELECT; incorrect energy selectedEnergy selection does not match user request or stored configurationerror
a00bTH_ERROR_SHOCK_NOT_DELIVERED; shock delivery failedShock was not successfully delivered to patientcritical
a00dTH_ERROR_CHARGING_EXPIRED; charging time exceededCapacitor charging operation exceeded maximum time limiterror
a00fTH_ERROR_CAP_OUT_OF_RANGE; capacitor out of calibration rangeCapacitor voltage measurement outside acceptable calibration limitserror
a010TH_ERROR_PA_RATE_OUT_OF_RANGE; pacing rate out of tolerancePacer rate deviates beyond acceptable limits from selected rateerror
a011TH_ERROR_PA_CURRENT_OUT_OF_RANGE; pacing current out of tolerancePacing current deviates beyond acceptable limits from selected currenterror
a017TH_ERROR_DEFIB; defibrillation error reportGeneric defibrillation system error during therapy operationerror
a018TH_ERROR_PACER; pacing error reportGeneric pacing system error during therapy operationerror
a01aTH_ERROR_PACER_FAULT; pacing parameter limit exceededPacing rate, current, or pulse width exceeded safe operating limitserror
a01bTH_ERROR_DEFIB_WDT_DISABLE_FAIL; unable to turn off defib watchdogFailed to disable defib subsystem watchdog timer during shutdownerror
a020TH_ERROR_PACER_DISABLE; pacer disabled communication lostCannot communicate with pacer processor; pacer function disablederror
a021TH_ERROR_CAP_CHARGE_FAIL; capacitor stays zero while chargingCapacitor voltage remains at zero despite charging operationerror
a022TH_ERROR_CORRUPT_ENERGY_SELECT; energy selection corruptedEnergy selection value in memory does not match expected formaterror
a023TH_ERROR_XFER_ENABLE_ON; transfer-enable line high at startupDefib transfer-enable line not properly disabled during system startuperror
a024TH_ERROR_VCAP2_SATURATED; VCAP2 reading full scale continuouslyCapacitor voltage monitoring line 2 reading at maximum valueerror
a026TH_ERROR_ENERGY_RESIDUE; residual energy on capacitorEnergy remains on capacitor after discharge or dump operationerror
a027TH_ERROR_PA_RATE_CORRUPT; pacing rate storage corruptedStored pacing rate value does not match expected format or rangeerror
a028TH_ERROR_CAL_ENERGY_FAIL; calibrated voltage out of rangeCalibrated voltage value exceeds acceptable limits during calibrationerror
a029TH_ERROR_BTE_FAIL; BTE board error conditionBiphasic Therapy Electronics board reported fault or control errorerror
a02bTH_ERROR_DEFIB_CONFIG; conflicting defib type configurationDefib type in hardware does not match software configurationerror
a02cTH_ERROR_DUMP_LINE_FAIL; capacitor dump line failedCapacitor dump circuit failed during self-test or operationerror
a02dTH_ERROR_WRONG_DEFIB_TYPE; wrong defib type for softwareDefibrillator hardware type does not match loaded software versionerror
a02eTH_ERROR_ADC_READ; ADC read failure for cap chargeADC failed to return valid reading when measuring capacitor chargeerror
a101DE_ERROR_LONG_WDT; long watchdog test failedLong watchdog timer test failed or watchdog did not reset in timeerror
a102DE_ERROR_SHORT_WDT; short watchdog test failedShort watchdog timer test failed during diagnosticserror
a103DE_ERROR_SCI_RCV; serial port receiver errorSerial communication receiver detected error during data receptionerror
a104DE_ERROR_XFER_KEY; defib hardware errorTransfer key or enable line hardware fault detectederror
a106DE_ERROR_ENERGY_OUT_OF_BOUND; VCAP-1 over/under chargeCapacitor 1 voltage exceeds upper or falls below lower limitserror
a107DE_ERROR_HP_ENG_SELECT; cannot determine rotary energy settingEnergy selection rotary switch position cannot be readerror
a109DE_ERROR_CAL_CRC; calibration table CRC errorCRC verification of calibration table failederror
a10bDE_ERROR_CHG_TIME; charging takes too long to reach targetCapacitor charging operation exceeded acceptable time limiterror
a10cDE_ERROR_CHG_ENABLE; CHG_EN1 line stuck highCharge enable control line remains high when it should be lowerror
a10dDE_ERROR_DUMP_ENERGY; defib hardware error during dumpCapacitor dump operation failed or reported hardware errorerror
a10eDE_ERROR_RCV; defib hardware communication errorDefib hardware failed to respond to receiver commanderror
a111DE_ERROR_ENERGY_NOT_ZERO; unexpected energy during biphasic testCapacitor energy detected during high-pot test when none expectederror
a112DE_ERROR_TEST_XFER_ENABLE; transfer enable control problemTransfer enable line did not respond during main testerror
a113DE_ERROR_TEST_XFER_ENGAGE; transfer engage feedback errorDefib hardware failed to report transfer engage feedback statuserror
a114DE_ERROR_TEST_DUMP_RELAY; biphasic dump relay test failedBiphasic dump relay failed during self-test procedureerror
a115DE_ERROR_TEST_ADC; ADC hardware test failedAnalog-to-digital converter hardware test failederror
a116DE_ERROR_TEST_DAC; DAC hardware test failedDigital-to-analog converter hardware test failederror
a117DE_ERROR_TEST_HARDWARE; general defib hardware test failureGeneric defibrillator hardware test failederror
a118DE_ERROR_TEST_RAM; defib RAM test failedRAM test on defib processor failederror
a119DE_ERROR_TEST_ROM; defib ROM test failedROM test on defib processor failederror
a11aDE_ERROR_TEST_CPU; defib CPU test failedDefib processor CPU test failederror
a11dDE_ERROR_XFER_TIMEOUT; transfer operation timeout errorDefib transfer operation did not complete within time limiterror
a11eDE_ERROR_BUTTONS_UP; defib hardware button errorDefib hardware button test failederror
a11fDE_ERROR_SYNC_INTERRUPT; defib sync interrupt errorSynchronization interrupt signal from defib hardware invaliderror
a120DE_ERROR_SELF_TEST_INCOMPLETE; defib self-test failed to completeDefib hardware self-test sequence did not complete successfullyerror
a123DE_ERROR_CAL_RCV_CRC; CRC failed for new calibration dataCRC verification failed when receiving new calibration dataerror
a124DE_ERROR_CAL_NVM_CRC; calibration write to EEPROM failedCannot write energy table to nonvolatile memory due to CRC errorerror
a125DE_ERROR_DAC_WRITE; DAC byte write failed via SPIFailed to write calibration byte to DAC through SPI interfaceerror
a126Cannot read from ADCADC read operation failederror
a127Must be idle to switch to test modeAttempt to enter test mode while device not idlewarning
a129Defib hardware error - transfer cable issueDefib transfer cable failure detectederror
a12aDefib hardware error - transfer paddle issueDefib transfer paddle failure detectederror
a12cDefib hardware error - charge inhibit failureCharge inhibit circuit failure detectederror
a12dCharge enable feedback indicates not chargingCharging process initiated but feedback shows no charge occurringerror
a12eCedar BTE fault line state detectedBTE fault signal assertederror
a12fCedar BTE fault line state clearedBTE fault signal de-assertedinfo
a130Defib hardware error - BTE resetBTE (Biphasic Therapy Engine) reset signal issuederror
a131Biphasic hardware not foundSystem boot or initialization detects missing Biphasic modulecritical
a132Defib hardware error - BTE transfer failureData transfer to/from Biphasic module failederror
a133A/D high bit sticky - Cedar BTE A/D converter faultA/D converter high bit remains fixed during conversionerror
a201Long watchdog timer test failedWatchdog timer did not respond correctly during testerror
a202Short watchdog timer test failedShort watchdog timer did not respond during testerror
a203SCI received overrun, framing or parity errorSerial communication interface detects data corruptionwarning
a2042 ms ECG sampling overrunECG sampling processor unable to complete within 2 ms cycleerror
a208Received message incompleteCommunication message reception detected as incompleteerror
a209Received message size error or input buffer fullMessage size incorrect or input buffer capacity exceedederror
a20dPacing pulse process overrunPacing pulse generation process exceeds time budgeterror
a20ePacing pulse too longPacing pulse width exceeds maximum allowable durationerror
a20fInternal A/D conversion timeoutInternal analog-to-digital converter operation did not complete in timeerror
a210External A/D conversion timeoutExternal analog-to-digital converter operation did not complete in timeerror
a211SPI transfer timeoutSerial Peripheral Interface data transfer did not complete in timeerror
a212RAM test failedMemory self-test detected RAM failurecritical
a213ROM CRC test failedRead-only memory checksum verification failedcritical
a214Stack overrun occurredCPU stack pointer exceeded allocated stack memorycritical
a215Isolated +5 V reference out of rangeIsolated +5 V supply voltage measured outside specificationerror
a216High voltage present when not pacingHV detected on output when pacing not activecritical
a217+12 V voltage out of range+12 V supply voltage measured outside specificationerror
a218High voltage present when not pacingHV detected on output when pacing not activecritical
a219HVIS voltage out of rangeHVIS (High Voltage Isolated Sense) voltage measured outside specificationerror
a21aCurrent calibration failedPacer output current calibration procedure unsuccessfulerror
a21bImpedance 300 ohm calibration failedImpedance calibration at 300 ohm load unsuccessfulerror
a21cCurrent calibration table corruptedCRC check of current calibration data failederror
a21dImpedance calibration values corruptedCRC check of impedance calibration data failederror
a21eImpedance 0 ohm calibration failedImpedance calibration at 0 ohm load unsuccessfulerror
a21fPacing current present when not pacingOutput current detected when pacing mode inactivecritical
a220No high voltage present when pacingExpected HV output not detected during pacingcritical
a221External A/D test register reset failedA/D converter test register did not reset properly during diagnosticerror
a222No high voltage present before current calibrationHV not detected during calibration setup procedureerror
a223No Q4 sense when not pacing (Q4 shorted)Q4 sense signal missing when pacing inactiveerror
a224Q4 sense present when pacing (CR25 open)Q4 sense signal present when pacing activeerror
b001Printhead ADC reading too low; temperature errorPrinthead thermistor ADC value below thresholderror
b002Printhead ADC reading too high; temperature errorPrinthead thermistor ADC value above thresholderror
b003Printer self-test failed; invalid CRCHardware self-test detects incorrect CRC from printer moduleerror
b004Wrong printer type detected; 50/100 mismatchInstalled printer type does not match device configurationerror
c008Lost communications with power PCBSerial communication with power management board failederror
c009Power PCB running on internal oscillator; external clock failurePower PCB detects external clock failure during initializationwarning
d003SCI frame errorSerial communication interface detects frame synchronization errorwarning
d004SCI parity errorSerial communication parity check failswarning
d005SCI read buffer fullSerial communication input buffer capacity exceededwarning
d006SCI initialization failedSerial communication interface did not initialize during startuperror
d00dMessage queue full; queue overflowInternal message queue capacity exceedederror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI