Physio-Control LIFEPAK 12: troubleshooting and error codes
The ServiceForge knowledge base holds 154 documented symptoms for the Physio-Control LIFEPAK 12, with 239 ranked causes, 243 diagnostic checks with test points and expected values, and 684 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
No energy discharge from paddles patient safety
QUIK-COMBO or standard therapy paddles deliver no energy during test; defibrillator therapy circuit not operational
Ranked causes
- 1. Therapy cable or paddle set failure (common)
- 2. Therapy connector cable failure (common)
- 3. Therapy PCB failure (occasional)
- 4. Transfer relay assembly failure (rare)
- 5. Waveshaping inductor failure (rare)
- 6. Energy storage capacitor failure (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power source malfunction caution
Power source management issue; battery or AC power failure
Ranked causes
- 1. Battery failure or low charge (common)
- 2. Battery pin or connector issue (common)
- 3. Test battery fault (occasional)
- 4. AC power adapter failure (occasional)
- 5. Power PCB failure (rare)
- 6. Power PCB fuse failure (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-Lead ECG inappropriate screen message response patient safety
Unit displays incorrect or unexpected message during 12-lead ECG test
Ranked causes
- 1. Test setup configuration issue (common)
- 2. ECG Cable failure (common)
- 3. ECG Connector Cable failure (occasional)
- 4. Small Keypad failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-Lead ECG gain out of tolerance patient safety
ECG signal gain measurement exceeds acceptable range during 3-lead test
Ranked causes
- 1. Test setup configuration issue (common)
- 2. ECG Cable failure (common)
- 3. Therapy PCB failure (occasional)
- 4. System PCB failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-Lead ECG inappropriate screen message response patient safety
Unit displays incorrect or unexpected message during 3-lead ECG test
Ranked causes
- 1. Test setup configuration issue (common)
- 2. ECG Cable failure (common)
- 3. ECG Connector Cable failure (occasional)
- 4. Small Keypad failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG gain out of tolerance
QUIK-COMBO ECG gain measurement outside acceptable range
Ranked causes
- 1. Test setup error (common)
- 2. Therapy Cable faulty (occasional)
- 3. A04 Therapy PCB defective (occasional)
- 4. A01 System PCB defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper key response caution
Keypanel does not register key presses or responds inconsistently
Ranked causes
- 1. Small Keypad defective (common)
- 2. Large Keypad defective (common)
- 3. Interface PCB defective (occasional)
- 4. System PCB defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inappropriate screen message or alarm response patient safety
Pacer option displays incorrect message or alarm during testing
Ranked causes
- 1. Test setup error (common)
- 2. Therapy or ECG cable failure (occasional)
- 3. Connector cable or keypad failure (occasional)
- 4. A04 Therapy PCB failure (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inappropriate screen message response patient safety
QUIK-COMBO impedance sense displays incorrect or unexpected message on screen
Ranked causes
- 1. Test setup error or transient condition (common)
- 2. Therapy Cable fault (occasional)
- 3. W11 Therapy Connector Cable fault (occasional)
- 4. A04 Therapy PCB failure (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing or broken characters caution
Printer output shows missing or broken character representations in printed text
Ranked causes
- 1. Improper paper (common)
- 2. Printhead dirty or clogged (common)
- 3. Printer Assembly defective (occasional)
- 4. System PCB defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails chassis leakage test high voltage
Device fails leakage current safety testing for chassis leakage path during service inspection
Ranked causes
- 1. Test setup error or defective test leads (common)
- 2. AC line power supply issue (common)
- 3. AC Power Adapter defective or faulty (occasional)
- 4. Internal wire routing incorrect or damaged (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails earth leakage test high voltage
Device fails leakage current safety testing for earth/ground leakage path during service inspection
Ranked causes
- 1. Test setup error or defective test leads (common)
- 2. AC line power supply issue (common)
- 3. AC Power Adapter defective or faulty (occasional)
- 4. Internal wire routing incorrect or damaged (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails sink leakage test high voltage
Device fails leakage current safety testing for sink-side leakage during service inspection
Ranked causes
- 1. Test setup error or defective test leads (common)
- 2. AC line power supply issue (common)
- 3. AC Power Adapter defective or faulty (occasional)
- 4. Internal wire routing incorrect or damaged (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails source leakage test high voltage
Device fails leakage current safety testing for source-side leakage during service inspection
Ranked causes
- 1. Test setup error or defective test leads (common)
- 2. AC line power supply issue (common)
- 3. AC Power Adapter defective or faulty (occasional)
- 4. Internal wire routing incorrect or damaged (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Verify device configuration data integrity
Review Device Log entries to confirm serial number, manufacturing date, and software revision are correctly stored; detect data loss or discrepancies
Ranked causes
- 1. Serial number blank or missing from device memory (rare)
- 2. Manufacturing date shows 01 Jan 1970; configuration data lost or early manufacture (occasional)
- 3. Stored serial number does not match label on Battery Well 1 (occasional)
- 4. Defibrillation capacitor value below calibration threshold (occasional)
Display Device Log and review serial number field
- Test point
- SERVICE/STATUS/DEVICE LOG serial number entry
- Expected
- Serial number populated and matches label on Battery Well 1
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-Lead ECG gain out of tolerance patient safety
ECG signal gain measurement exceeds acceptable range during 12-lead test
Ranked causes
- 1. Test setup configuration issue (common)
- 2. ECG Cable failure (common)
- 3. System PCB failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 monitor fails calibration patient safety
EtCO2 (end-tidal CO2) monitor cannot complete calibration sequence
Ranked causes
- 1. Test setup error or depleted calibration gas (common)
- 2. Disconnected or kinked FilterLine (occasional)
- 3. CO2 module failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device powered on after case reassembly patient safety
Checking status after case reassembly to determine next troubleshooting step based on SERVICE indicator state
Ranked causes
- 1. SERVICE indicator off - device operating normally (common)
- 2. SERVICE indicator on - error codes present (occasional)
- 3. No power indication - battery or connection issue (occasional)
Observe SERVICE indicator state
- Test point
- SERVICE indicator LED on device
- Expected
- LED off indicates normal operation
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG fast restore out of tolerance
QUIK-COMBO ECG fast restore measurement outside acceptable range
Ranked causes
- 1. Test setup error (common)
- 2. Therapy cable faulty (occasional)
- 3. A01 System PCB defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failure to transfer coincident with Sync mark patient safety
Shock fails to deliver at the moment the sync mark appears on ECG during synchronous cardioversion
Ranked causes
- 1. Sync mode not properly enabled or firmware issue (common)
- 2. Keypad control malfunction (occasional)
- 3. Standard paddles defect (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Inaudible or garbled audio
Audio output from speaker is not audible or is garbled/distorted
Ranked causes
- 1. Speaker assembly failure (common)
- 2. Interface PCB failure (occasional)
- 3. System PCB failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keypad Error
Transient error condition detected by the device; may clear on retry.
Ranked causes
- 1. Transient keypad fault (occasional)
- 2. Small Keypad A09 defective (common)
- 3. Large Keypad A10 defective (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measured energy exceeds 18J high voltage
Standard paddles defibrillation isolation measurement exceeds 18 joules threshold
Ranked causes
- 1. Test setup error or measurement transient (common)
- 2. Internal high voltage wire routing incorrect (occasional)
- 3. High voltage wire connections loose or damaged (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing dots in printed X caution
Printer output shows missing dots forming part of printed characters, specifically in X pattern
Ranked causes
- 1. Improper printer paper (common)
- 2. Printhead dirty or clogged (common)
- 3. Printer Assembly defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No energy discharge from defibrillator paddles high voltage
Therapy energy does not deliver to patient pads or paddles
Ranked causes
- 1. Test setup error or equipment malfunction (common)
- 2. Therapy cable or paddle failure (occasional)
- 3. Internal therapy system component failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power on patient safety
Device does not power up or provide power
Ranked causes
- 1. Discharged or missing battery (common)
- 2. Damaged battery pins (occasional)
- 3. Power PCB failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No power on during self-test
Unit shows no signs of life during power-on sequence
Ranked causes
- 1. Batteries not installed, discharged, or incompatible (common)
- 2. Damaged or corroded battery pins (common)
- 3. Power PCB failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Sync mark patient safety
Synchronous cardioversion mode fails to display synchronization mark on ECG, preventing synchronized shock delivery
Ranked causes
- 1. Test setup error or ECG signal marginal (common)
- 2. ECG display size too small to show sync mark (occasional)
- 3. A01 System PCB failure or firmware issue (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer Error
Transient error condition detected in printer assembly; may clear on retry.
Ranked causes
- 1. Transient printer fault (occasional)
- 2. Incorrect Printer Assembly A12 installed (occasional)
- 3. Printer Assembly A12 defective (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 Error
Transient error condition detected in pulse oximetry module; may clear on retry.
Ranked causes
- 1. Transient SpO2 module fault (occasional)
- 2. SpO2 Module A16 defective (common)
- 3. OEM PCB A06 defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: TCP Printer
Adjust printer speed to produce correctly-spaced output marks at 25mm intervals
Ranked causes
- 1. Printer speed requires calibration after printhead installation or replacement (common)
- 2. Existing printer requires speed adjustment verification (common)
Measure spacing between printed tick marks
- Test point
- Printed output
- Expected
- 25mm interval between marks
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: TCP Printer speed and printhead
Printer produces tick marks at incorrect spacing or printhead resistance needs adjustment after replacement
Ranked causes
- 1. Printer speed not calibrated to 25 mm/sec spacing (common)
- 2. Printhead resistance value not set after printer replacement (occasional)
Determine if new printer is installed
- Test point
- Device configuration
- Expected
- New printer installed or existing printer
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Defibrillation Capacitor Error high voltage
Transient error condition detected in energy storage capacitor; may clear on retry.
Ranked causes
- 1. Transient capacitor fault (occasional)
- 2. Energy Storage Capacitor A15 defective (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Improper LCD response caution
LCD display does not respond properly to input or display commands
Ranked causes
- 1. LCD Assembly defective (common)
- 2. System PCB defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NIBP monitor displays XXX with SERVICE indicator OFF patient safety
NIBP (blood pressure) reading shown as XXX while service indicator is not active
Ranked causes
- 1. NIBP module leakage or electrical issue (common)
- 2. NIBP module failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Processing error code logged patient safety
SERVICE indicator lights; error code recorded in Error Code Log with date, time, and extension
Ranked causes
- 1. Internal program or process failure (common)
- 2. Transient EMI or ESD (occasional)
Display Error Code Log
- Test point
- Error Code Log display
- Expected
- Error code with date, time, and error extension visible
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-Lead ECG gain out of specification patient safety
ECG waveform amplitude does not fall within expected range when recorded
Ranked causes
- 1. ECG amplifier gain miscalibration or component failure (unknown)
Record 12-lead ECG test signal and measure peak-to-peak amplitude
- Test point
- ECG display screen
- Expected
- 36 to 44 mm, 36 to 44 mm
12-Lead ECG leads off detection failure patient safety
Device does not display leads off screen message when 12-Lead ECG leads are disconnected
Ranked causes
- 1. Lead detection circuit malfunction or software failure (unknown)
Disconnect 12-Lead ECG leads and monitor display
- Test point
- Display message output
- Expected
- Leads off screen message displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
12-Lead/3-Lead ECG fast restore time exceeds limit patient safety
DC step functions appear on screen and printout; flat line return time longer than specification
Ranked causes
- 1. Fast restore time performance degradation (unknown)
Verify test wire LA connected to test fixture 250mV terminal
- Test point
- Test fixture 250mV terminal
- Expected
- LA wire connected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
3-Lead ECG gain out of specification patient safety
3-lead ECG waveform amplitude does not fall within expected range when recorded
Ranked causes
- 1. ECG amplifier gain miscalibration or component failure (unknown)
Record 3-lead ECG test signal and measure peak-to-peak amplitude
- Test point
- ECG display screen
- Expected
- 36 to 44 mm, 36 to 44 mm
3-Lead ECG leads off detection failure patient safety
Device does not display leads off screen message when 3-Lead ECG leads are disconnected
Ranked causes
- 1. Lead detection circuit malfunction or software failure (unknown)
Disconnect 3-Lead ECG leads and monitor display
- Test point
- Display message output
- Expected
- Leads off screen message displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
A13 Transfer Relay Assembly replacement patient safety
Reinstallation of A13 Transfer Relay Assembly after removal, requiring calibration afterwards
Ranked causes
- 1. A13 Transfer Relay Assembly replaced (common)
Verify coil end positioning in case recess
- Test point
- A13 Transfer Relay Assembly coil end
- Expected
- Seated in case recess
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio malfunction caution
Voice messages and tones are distorted or unclear
Ranked causes
- 1. Speaker or audio amplifier failure (unknown)
Verify voice messages and tones during self-test
- Test point
- Audio output speaker
- Expected
- Clear, intelligible voice messages and audible tones without distortion
Audio output distorted or absent
Voice messages and tones are not clear or are distorted
Ranked causes
- 1. Unknown cause; diagnostic procedure required (unknown)
Play audio messages and tones during test
- Test point
- Speaker output
- Expected
- Voice messages and tones clear and undistorted
Audio system distorted or muted patient safety
Voice messages or alert tones are unclear, distorted, or absent
Ranked causes
- 1. Speaker malfunction, audio amplifier failure, or software issue (unknown)
Verify audio clarity during voice message and tone playback
- Test point
- Speaker output
- Expected
- Voice messages and tones clear and not distorted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auxiliary power switching failure caution
Auxiliary power source does not engage or disengage properly
Ranked causes
- 1. Auxiliary power circuit malfunction or control signal failure (unknown)
Verify auxiliary power switching response
- Test point
- Power source selection or indicator
- Expected
- Auxiliary power switches on and off as commanded
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery switching failure patient safety
Device does not properly switch between battery and AC power sources
Ranked causes
- 1. Power supply circuit failure or battery contact issue (unknown)
Test battery to AC power transition
- Test point
- Power source indicator or operational continuity
- Expected
- Seamless switching between power sources
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: A01 System PCB replacement patient safety
After replacing the A01 System PCB on LIFEPAK 12, defibrillator calibration constants are invalidated and must be recalibrated.
Ranked causes
- 1. A01 System PCB replacement performed (common)
Remove A02 Memory PCB standoffs from old A01 System PCB
- Test point
- A02 Memory PCB standoffs (251)
- Expected
- Standoffs removed and available for reuse
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Advisory mode key function
Verify ADVISORY key LED illuminates and ADVISORY MODE - MONITORING message displays when pressed
Ranked causes
- 1. ADVISORY key or LED malfunction (unknown)
Connect device to analyzer using QUIK-COMBO therapy cable and test post adapter
- Test point
- Therapy cable connectors
- Expected
- Secure connection established
3 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2
Device CO2 measurement requires calibration via service mode
Ranked causes
- 1. Calibration procedure required as part of service maintenance (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Coin battery replacement caution
Device configuration, calibration data, and real-time clock powered by coin battery; battery depletes over time and requires periodic replacement
Ranked causes
- 1. Coin battery depleted or removed (common)
Obtain new coin battery
- Test point
- Battery specification
- Expected
- CR2032 type
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Defibrillation patient safety
Device defibrillation output requires calibration via service mode to ensure proper energy delivery
Ranked causes
- 1. Calibration procedure required as part of service maintenance (unknown)
Enter service mode
- Test point
- Device startup with OPTIONS and EVENT keys held
- Expected
- Passcode overlay appears on SETUP screen
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Defibrillation capacitor self-test high voltage
TCP - Defibrillation Self Calibration procedure to verify defibrillation capacitor value; device calculates capacitor value during this procedure
Ranked causes
- 1. Defibrillation capacitor value below calibration levels (rare)
Complete TCP - Defibrillation Self Calibration procedure
- Test point
- Device screen
- Expected
- Defib Storage Cap Value recorded as approximately 52uF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Defibrillation output high voltage
Verify defibrillator energy output accuracy for safety and efficacy
Ranked causes
- 1. Defibrillation output requires periodic calibration verification (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Defibrillator delivered energy high voltage
Verify defibrillator output energy at 10 J, 200 J, and 360 J levels falls within acceptable limits using defibrillator analyzer
Ranked causes
- 1. Delivered energy outside acceptable range (unknown)
Measure delivered energy at 10 J
- Test point
- Defibrillator analyzer output
- Expected
- 9.0 to 11.0 J, 9.0 to 11.0 J
- Tools
- Defibrillator analyzer
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Defibrillator self calibration high voltage
Routine calibration procedure for LIFEPAK 12 defibrillator to verify energy output accuracy
Ranked causes
- 1. Routine maintenance calibration required (common)
Disconnect all front panel cables
- Test point
- UUT front panel
- Expected
- All cables disconnected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Delivered Energy measurement patient safety
Verify defibrillator output energy at specified levels meets acceptable range
Ranked causes
- 1. Delivered energy outside acceptable output range (unknown)
Measure delivered energy at specified energy levels
- Test point
- QUIK-COMBO output
- Expected
- Within acceptable output range per table
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Energy Storage Capacitor capacitance high voltage
Determine actual capacitance of A15 Energy Storage Capacitor via TCP – Defibrillation Calibration procedure; nominal value 50µF
Ranked causes
- 1. Capacitor capacitance verification required (unknown)
Verify A15 Energy Storage Capacitor capacitance via TCP calibration
- Test point
- Service/Status/Device Log screen
- Expected
- 50µF nominal
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive Pressure
Device invasive pressure measurement requires calibration via service mode
Ranked causes
- 1. Calibration procedure required as part of service maintenance (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Invasive pressure P1 verification
Static pressure accuracy test for P1 invasive pressure channel using simulator (DNI Nevada 217A or equivalent) at multiple pressure points: 250, 200, 100, 40, 20 mmHg
Ranked causes
- 1. P1 invasive pressure measurement out of specification (unknown)
Turn device ON and connect invasive pressure simulator to P1 connector
- Test point
- P1 connector on parameter bezel
- Expected
- Simulator connected and powered on
10 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Invasive pressure P2 verification
- Calibration: NIBP
- Calibration: NIBP leakage verification
- Calibration: NIBP pressure verification
- Calibration: No advisory mode function verification
- Calibration: Pacer leads off detection
- Calibration: Pacing
- Calibration: Pacing output
- Calibration: Paddles ECG fast restore recovery time
- Calibration: Paddles ECG gain and waveform amplitude
- Calibration: Performance Inspection Procedure (PIP)
- Calibration: Printer
- Calibration: Printer output
- Calibration: Printer TCP calibration
- Calibration: Printhead alignment
- Calibration: Printhead up/down alignment
- Calibration: Standard paddles user discharge test
- Calibration: Synchronous cardioversion sync delay
- Calibration: TCP Computer-Assisted Energy after A14 Waveshaping Inductor replacement
- Calibration: TCP delivered energy output
- Calibration: TCP Printer - New Printhead Setup
- Calibration: TCP Printer preparation
- Damaged battery pin(s)
- Damaged keypad or label
- Defibrillation capacitor fails calibration
- Defibrillator charge current abnormal with auxiliary power
- Defibrillator charge current anomaly
- Defibrillator charge current anomaly with auxiliary power
- Delivered energy out of specification at 10J
- Delivered energy out of specification at 200J
- Delivered energy out of specification at 2J
- Delivered energy out of specification at 360J
- Delivered energy out of specification at 70J
- Delivered energy out of tolerance
- Delivered therapy energy out of tolerance
- Door assembly installation problem
- Earth leakage current exceeds limits
- ECG analog output waveform missing or out of tolerance
- Error codes logged in system
- Evidence of dirt, fluids, or foreign objects
- F1 fuse replacement
- F2 fuse replacement
- F4 fuse replacement
- Fast-Restore duration exceeds maximum
- Improper 25mm marker spacing
- Inappropriate screen message response (No Pacer Option)
- Keypad controls not responsive
- Keypad malfunction
- Keypads not responding
- LIFENET BLUE wireless communication verification
- Loose or broken hardware
- Loose or broken hardware visible
- Motor assembly malfunction
- Motor mounting screws over-tightened
- Motor tubing disconnection
- One or more horizontal lines missing or distorted
- Oximeter saturation reading missing or out of tolerance
- Pacer out of calibration
- Pacer pulse width out of tolerance
- Paddles source leakage current exceeds limit
- Paper strippers not fitting correctly
- PC Card communications test
- PC Card inappropriate screen message response
- PC Card removal and reinstallation
- Peak current levels out of tolerance
- Physical inspection failure
- Power On sequence incomplete
- Printer Assembly type mismatch
- Printer malfunction
- QUIK-COMBO delivered energy out of tolerance at 10J
- QUIK-COMBO delivered energy out of tolerance at 200J
- QUIK-COMBO delivered energy out of tolerance at 2J
- QUIK-COMBO delivered energy out of tolerance at 360J
- QUIK-COMBO delivered energy out of tolerance at 70J
- QUIK-COMBO ECG gain out of specification
- QUIK-COMBO ECG gain out of tolerance
- QUIK-COMBO Fast-Restore duration exceeds maximum
- QUIK-COMBO fast-restore duration out of specification
- QUIK-COMBO impedance sense failure
- QUIK-COMBO impedance sense false negative
- QUIK-COMBO impedance sense false positive
- QUIK-COMBO impedance sense malfunction
- Record Operating Data incomplete
- Service indicator illuminated
- Service indicator illuminated on startup
- SERVICE indicator remains on
- Sync discharge time exceeds 60ms
- Syringe technique for NIBP system inflation
- User selected manual mode not confirmed
- W04 System PCB/Interface PCB cable disconnected or loose
- W09 Auxiliary Connector Cable replacement
- W10 Battery Pins/Power PCB Cable replacement
- W10 Battery Pins/Power PCB Cable requires replacement
- W22 SpO2 Connector Cable replacement
94 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
00000001 | PA_ERR_LONG_WDT | Long watchdog timer expiration in pacer module | critical |
00000002 | PA_ERR_SHORT_WDT | Short watchdog timer expiration in pacer module | error |
00000003 | PA_ERR_SCI_RCV | Serial communication receive error in pacer module | error |
00000004 | PA_ERR_2MS_OVERRUN | 2 millisecond task overrun in pacer module | error |
00000005 | PA_ERR_TICK_OVERRUN | Tick timer task overrun in pacer module | error |
00000006 | PA_ERR_MSG_CHKSUM | Message checksum error in pacer communication | error |
00000007 | PA_ERR_MSG_ID | Invalid message ID received by pacer module | error |
00000008 | PA_ERR_MSG_RESYNC | Message resynchronization error in pacer communication | error |
00000009 | PA_ERR_MSG_SIZE | Invalid message size received by pacer module | error |
0000000a | PA_ERR_SEQUENCING | Message sequencing error in pacer communication | error |
0000000b | PA_ERR_UNKNOWN_RATE | Pacing rate parameter unrecognized or out of valid range | error |
0000000c | PA_ERR_UNKNOWN_CURRENT | Pacing current parameter unrecognized or out of valid range | error |
0000000d | PA_ERR_PACE_OVERRUN | Pacing operation exceeds timing boundaries | error |
0000000e | DE_ERR_RCV | Unknown | unknown |
0000000e | PA_ERR_PULSE_WIDTH | Pacing pulse width out of acceptable range | error |
0000000f | PA_ERR_A2D_INT | Analog-to-digital converter internal error | critical |
00000010 | PA_ERR_A2D_EXT | Analog-to-digital converter external error | critical |
00000010 | DE_ERR_IGNORE_CHARGE | Unknown | unknown |
00000011 | DE_ERR_ENERGY_NOT_ZERO | Unknown | unknown |
00000011 | PA_ERR_SPI | Serial Peripheral Interface communication failure | critical |
00000012 | DE_ERR_TEST_XFER_ENABLE | Unknown | unknown |
00000012 | PA_ERR_RAM_TEST | RAM self-test failure on power-up or during operation | critical |
00000013 | PA_ERR_ROM_TEST | ROM self-test failure on power-up or during operation | critical |
00000013 | DE_ERR_TEST_XFER_ENGAGE | Unknown | unknown |
00000014 | DE_ERR_TEST_DUMP_RELAY | Unknown | unknown |
00000014 | PA_ERR_CPU_TEST | CPU self-test failure detected | critical |
00000015 | PA_ERR_STACK_CHECK | Stack overflow or corruption detected | critical |
00000016 | DE_ERR_TEST_DAC | Unknown | unknown |
00000016 | PA_ERR_V_ISO_MON | Isolation voltage monitoring error | critical |
00000017 | DE_ERR_TEST_HARDWARE | Unknown | unknown |
00000017 | PA_ERR_V_12V_MON | 12V power supply monitoring error | critical |
00000018 | DE_ERR_TEST_RAM | Unknown | unknown |
00000018 | PA_ERR_V_HVIS_SENSE | High voltage sense circuit error | critical |
00000019 | PA_ERR_V_HVIS | High voltage supply monitoring error | critical |
00000019 | DE_ERR_TEST_ROM | Unknown | unknown |
0000001a | DE_ERR_TEST_CPU | Unknown | unknown |
0000001a | PA_ERR_CAL_CURRENT | Current calibration error detected | error |
0000001b | PA_ERR_CAL_Z_300 | Calibration at 300 ohm impedance failed | error |
0000001b | DE_ERR_TEST_STACK | Unknown | unknown |
0000001c | PA_ERR_CAL_CURRENT_CRC | Current calibration data CRC check failed | error |
0000001d | PA_ERR_CAL_IMPEDANCE_CRC | Impedance calibration data CRC check failed | error |
0000001d | DE_ERR_XFER_TIMEOUT | Unknown | unknown |
0000001e | DE_ERR_BUTTONS_UP | Unknown | unknown |
0000001e | PA_ERR_CAL_Z_0 | Calibration at 0 ohm impedance failed | error |
0000001f | PA_ERR_PACE_I | Pacing current measurement or delivery error | error |
0000001f | DE_ERR_SYNC_INTERRUPT | Unknown | unknown |
00000020 | PA_ERR_NO_HVIS_SENSE | High-frequency current sensing failure | unknown |
00000020 | DE_ERR_SELF_TEST_INCOMPLETE | Unknown | unknown |
00000021 | PA_ERR_EXT_A2D_TEST | A2D converter external test failure | unknown |
00000021 | DE_ERR_MESSAGE_ABORTED | Unknown | unknown |
00000022 | PA_ERR_NO_CAL_HVIS_SENSE | Uncalibrated high-frequency current sensing | unknown |
00000022 | DE_ERR_MESSAGE_CHECKSUM | Unknown | unknown |
00000023 | DE_ERR_CAL_RCV_CRC | Calibration data receive CRC error | error |
00000024 | DE_ERR_CAL_NVM_CRC | Calibration non-volatile memory CRC error | error |
00000025 | DE_ERR_DAC_WRITE | DAC write operation failure | error |
00000026 | DE_ERR_ADC_READ | ADC read operation failure | error |
00000027 | DE_ERR_TEST_MODE | Device in test mode or test mode error | warning |
00000028 | DE_ERR_TEST_COMMAND | Invalid or unexpected test command received | warning |
00000029 | DE_ERR_XFER_CABLE | Data transfer cable error or disconnection | error |
0000002a | DE_ERR_XFER_PADDLE | Paddle/electrode connection error during data transfer | error |
0000002b | DE_ERR_BACKGROUND_IDLE | Background task or idle mode error | warning |
0000002c | DE_ERR_CHG_INHIBIT | Charge inhibit function error or charging blocked | error |
0000002d | DE_ERR_CHG_ENABLE_FAIL | Charge enable failure during charging sequence | error |
0001 | UT_ERROR_FLASH_ADDRESS | Flash memory address error during operation | error |
0002 | Error during flash block erase (valid for all flash sizes) | Flash memory block erase operation fails | error |
0002 | UT_ERROR_FLASH_VPP | Flash memory VPP (programming voltage) supply failure | error |
0003 | UT_ERROR_FLASH_ERASE | Flash memory erase operation failed | error |
0003 | Flash memory block erase failure (valid for all flash sizes) | Flash memory block erase operation fails | error |
0004 | UT_ERROR_FLASH_8BIT_WRITE | 8-bit flash write operation failure | error |
0005 | UT_ERROR_FLASH_16BIT_WRITE | 16-bit flash write operation failure | error |
0006 | UT_ERROR_FLASH_PAGE_WRITE | Flash page write operation failed | error |
0007 | UT_ERROR_ADC_CHANNEL | ADC channel selection or configuration error | error |
0008 | UT_ERROR_ADC_READ | ADC read operation failure | error |
0009 | UT_ERROR_DAC_ACCESS | DAC access or communication failure | error |
000a | UT_ERROR_DAC_FAILURE | DAC operational failure | error |
000a | ECG DAC self-test failed or ECG DAC failure after cold boot | DAC self-test fails or DAC does not respond after power-on | error |
000b | UT_ERROR_AIO_SEM_NOT_CREATED | Analog I/O semaphore not created in system initialization | error |
000c | UT_ERROR_ADC_TEST_REG | ADC test register error | error |
000d | UT_ERROR_ADC_CAL_NOT_COMPLETE | ADC calibration not completed during startup | warning |
000e | Unknown flash device ID for voice/printer flash | Voice or printer flash device identification fails | error |
000f | Unknown flash device ID for data/program flash | Data or program flash device identification fails | error |
1 | System Communications or System Processing Error | Communications or processing failure detected | error |
10 | Defibrillator out of calibration | Defibrillator calibration lost or invalid | error |
1001 | UI_ERROR_TIMEOUT_TABLE_FULL | Timeout table reached maximum capacity | error |
1002 | UI_ERROR_NO_TIMEOUT_SEMAPHORE | Timeout semaphore not available or not created | error |
1003 | UI_ERROR_BAD_MANUAL_ACCESS | Invalid manual mode access or command received | warning |
1004 | UI_ERROR_PADDLE_SHOCK_SEEN | Shock delivered via manual paddles during operation | info |
1005 | UI_ERROR_DISPLAY_SELF_TEST | Display self-test failure or timeout | error |
1006 | UI_ERROR_ENERGY_FAULT | Defibrillation charge output outside 15% tolerance during manual mode operation | critical |
1007 | UI_ERROR_12LEAD_KEY_SEEN | 12-lead ECG function key pressed but unit not configured to support 12-lead mode | warning |
1008 | UI_ERROR_ANALYZE_KEY_SEEN | AED Analyze mode key pressed but unit not configured for AED mode | warning |
1009 | UI_ERROR_ADVISORY_KEY_SEEN | AED Advisory mode key pressed but unit not configured for AED | warning |
100a | UI_ERROR_NIBP_KEY_SEEN | NIBP (blood pressure) function key pressed but NIBP not enabled on unit | warning |
100b | UI_ERROR_CURRENTUP_KEY_SEEN | Pacing current increase key pressed but pacing not supported | warning |
100c | UI_ERROR_CURRENTDOWN_KEY_SEEN | Pacing current decrease key pressed but pacing not supported | warning |
100d | UI_ERROR_RATEUP_KEY_SEEN | Pacing rate increase key pressed but pacing not configured | warning |
100e | UI_ERROR_RATEDOWN_KEY_SEEN | Pacing rate decrease key pressed but pacing not enabled | warning |
100f | UI_ERROR_PACER_KEY_SEEN | Pacer mode key pressed but pacing not supported on unit | warning |
1010 | UI_ERROR_PAUSE_KEY_SEEN | Pause key pressed but pacing not configured on unit | warning |
1011 | UI_ERROR_INVALID_CPR_TIME | Unknown | info |
1012 | UI_ERROR_INVALID_AED_STATE | Unknown | info |
1013 | UI_ERROR_INVALID_SAS_STATUS | Unknown | info |
1014 | UI_ERROR_INVALID_SILENCE | Unknown | info |
1015 | UI_ERROR_INVALID_CORNER | Unknown | info |
1016 | UI_ERROR_INVALID_FILL_COLOR | Unknown | info |
1017 | UI_ERROR_INVALID_TH_STATUS | Unknown | info |
1018 | UI_ERROR_INVALID_SEQUENCE_COUNT | Unknown | info |
1019 | UI_ERROR_INVALID_MENU_CALLBACK | Unknown | info |
101a | UI_ERROR_INVALID_KNOB_STATE | Unknown | info |
101b | UI_ERROR_INVALID_PADDLES_DEFAULT | Invalid paddle defaults | unknown |
101c | UI_ERROR_INVALID_STYLE | Unknown | info |
101d | UI_ERROR_INVALID_LED_STATE | Unknown | info |
101e | UI_ERROR_INVALID_LED | Unknown | info |
101f | UI_ERROR_INVALID_CHANNEL | Unknown | info |
1020 | UI_ERROR_INVALID_WAVE_2 | Unknown | info |
1021 | UI_ERROR_INVALID_WAVE_3 | Unknown | info |
1022 | UI_ERROR_INVALID_MENU_MODE | Unknown | info |
1023 | UI_ERROR_INVALID_DATA_SIZE | Unknown | info |
1024 | UI_ERROR_INVALID_SUB_FIELD | Unknown | info |
1025 | UI_ERROR_INVALID_REPORT_TYPE | Unknown | info |
1026 | UI_ERROR_INVALID_PACER_MODE | Unknown | info |
1027 | UI_ERROR_INVALID_ALARM_STATE | Unknown | info |
1028 | UI_ERROR_INVALID_QRS_STATE | Unknown | info |
1029 | UI_ERROR_INVALID_PP_STATUS | Unknown | info |
102a | UI_ERROR_INVALID_PR_STATUS | Unknown | info |
102b | UI_ERROR_INVALID_SERVICE_STATE | Unknown | info |
102c | UI_ERROR_INVALID_KEY_STYLE | Unknown | info |
102d | UI_ERROR_INVALID_TONE | Unknown | info |
102e | UI_ERROR_INVALID_MENU_TIMEOUT | Unknown | info |
102f | UI_ERROR_INVALID_BATTERY_CAPACITY | Unknown | info |
1030 | UI_ERROR_INVALID_DC_STATUS | Unknown | info |
1031 | UI_ERROR_INVALID_L12_STATUS | Unknown | info |
1032 | UI_ERROR_INVALID_WARNING_TONE | Unknown | info |
1033 | UI_ERROR_INVALID_WARNING_TYPE | Unknown | info |
1034 | UI_ERROR_CREATE_TIMEOUT_SEMAPHORE | Failed to create timeout semaphore | error |
1035 | UI_ERROR_INVALID_TIMEOUT_SEMAPHORE | Unknown | info |
1036 | UI_ERROR_CREATE_EVENT_QUEUE | Failed to create event queue | error |
11 | Keypad Error | Keypad malfunction detected | error |
12 | SpO2 Error | SpO2 module or measurement failure | error |
13 | Printer Error | Printer malfunction or configuration issue | error |
14 | Defibrillation Capacitor Error | Capacitor malfunction detected | error |
15 | Pacer out of calibration | Pacing output calibration lost or invalid | error |
16 | Pacer output circuit damaged | Pacing output circuit failure | error |
17 | Biphasic (BTE) fault or unknown energy | Biphasic waveform generation failure or energy mismatch | error |
18 | Defibrillator configuration error | Configuration mismatch detected | error |
19 | Dump error code (a02c) | System dump/core error detected | critical |
1fff | UI_ERROR_EXTRA_INFORMATION | Additional diagnostic information follows a preceding error code | info |
2 | Memory Error | Memory malfunction detected | error |
20 | ECG error | ECG acquisition or processing failure | error |
2001 | DC_ERROR_UNKNOWN | Unknown | info |
2002 | DC_ERROR_DC_MSG_QUEUE_CREATE | Unknown | info |
2003 | DC_ERROR_RING_BUFFER_CREATE | Unknown | info |
2004 | DC_ERROR_UNABLE_TO_INITIATE_SERIAL_PORT | Unknown | info |
2005 | DC_ERROR_UNABLE_TO_INITIATE_PCMCIA | Unknown | info |
2006 | DC_ERROR_UNABLE_CLOSE_PC_CARD_FD | Failed to close PC Card file descriptor | error |
2007 | DC_ERROR_UNABLE_TO_ACCESS_MSG_Q | Unknown | info |
2008 | DC_ERROR_UNSUCCESSFUL_DEVICE_READ | Device read operation failed | error |
2009 | DC_ERROR_UNSUCCESSFUL_RING_BUFFER_READ | Unknown | info |
200a | DC_ERROR_UNSUCCESSFUL_RING_BUFFER_WRITE | Unknown | info |
200b | DC_ERROR_UNSUCCESSFUL_DEVICE_WRITE | Device write operation failed | error |
200c | DC_ERROR_UNKNOWN_FD_DURING_QUERY | Unknown file descriptor encountered during query | error |
200d | DC_ERROR_UNKNOWN_FD_DURING_WRITE | Unknown file descriptor encountered during write | error |
200e | DC_ERROR_UNABLE_TO_COMPLETE_WRITE | Write operation could not be completed | error |
200f | DC_ERROR_UNKNOWN_ACCESSOR | Unknown | info |
2010 | DC_ERROR_UP_TO_APP_MSG_QUEUE_WRITE | Failed to write message queue for application layer | error |
2011 | Unable to initialize UART for direct transfer | Direct data transfer attempt failed | error |
2012 | DC_ERROR_UNABLE_TO_RESET_UART_FOR_DIRECT_XFER | Unknown | info |
2013 | Unable to initialize UART for external transfer | External data transfer attempt failed | error |
2014 | Unable to reset UART for external transfer | UART reset during external transfer failed | error |
2015 | Unable to get DM report | Data module report retrieval failed | error |
2016 | DC_ERROR_DOWN_TO_SESSION_MSG_QUEUE_WRITE | Unknown | info |
21 | Unknown error 21 | Error 21 detected | error |
22 | NIBP system error | Non-invasive blood pressure measurement failure | error |
23 | CO2 system error | End-tidal CO2 measurement failure | error |
24 | Unknown error 24 | Error 24 detected | error |
25 | Unknown error 25 | Error 25 detected | error |
26 | Therapy hardware/software mismatch | Hardware and software version incompatibility | error |
27 | Unknown error 27 | Error 27 detected | error |
28 | System Software upgrade is recommended | Software version notification | info |
29 | Possible transient from input power | Power supply transient detected | warning |
3 | System Overload | Simultaneous operations exceed capacity | warning |
30 | Possible therapy processor reset condition | Therapy processor reset detected | warning |
3001 | DM_ERROR_UNKNOWN | Unknown | info |
3002 | Data module main thread creation semaphore error | Semaphore creation failed during main initialization | error |
3003 | Data module main boot error | Boot sequence failed in data module | critical |
3004 | Data module main queue full | Message queue overflow during main operation | warning |
3005 | DM_ERROR_DATABASE_ERASE_ADJUST | Not able to write new links for new oldest record | error |
3006 | DM_ERROR_DATABASE_ERASE_FAILED | Erase block failed during database operation | error |
3007 | DM_ERROR_DATABASE_ERASE_VERIFY | Verification of erased block failed | error |
3008 | DM_ERROR_DATABASE_FLASH_ERASE | Database erase operation failed | error |
3009 | DM_ERROR_DATABASE_CREATE_SEM | Unknown | info |
300a | Database episode end error | Episode record termination failed | error |
300b | DM_ERROR_DATABASE_FLASH_BAD | Unknown | info |
300c | Database write header error | Header write to database failed | error |
300d | Database write content error | Content write to database failed | error |
300e | Database rewrite header error | Header rewrite operation failed | error |
300f | Database rewrite waveform error | Waveform rewrite operation failed | error |
3010 | Database rewrite vital signs error | Vital signs rewrite operation failed | error |
3011 | DM_ERROR_DATABASE_REWRITE_CONT | Unknown | info |
3012 | DM_ERROR_DATABASE_REWRITE_DELTA | Unknown | info |
3013 | DM_ERROR_DATABASE_VERIFY_HEADER | Unknown | info |
3013 | Database header verification error | Database header validation failure | error |
3014 | Database content verification error | Database content validation failure | error |
3014 | DM_ERROR_DATABASE_VERIFY_CONTENT | Unknown | info |
3015 | DM_ERROR_DATABASE_READ_CRC | Unknown | info |
3016 | DM_ERROR_BUFFER_ZEROED | Unknown | info |
3017 | DM_ERROR_BUFFER_FALLING_BEHIND | Unknown | info |
3018 | DM_ERROR_SCP_INIT_FAIL | Unknown | info |
3019 | DM_ERROR_SCP_BUFFER_OVERRUN | Unknown | info |
301a | DM_ERROR_BAD_LINKED_LIST | Unknown | info |
301b | DM_ERROR_SCP_MEM_OVERRUN | Unknown | info |
301c | DM_ERROR_SCP_MM_SEM_FAILED | Unknown | info |
301d | DM_ERROR_CHECK_NO_TYPE | Unknown | info |
301e | DM_ERROR_CHECK_BAD_TYPE | Unknown | info |
301f | DM_ERROR_CHECK_BAD_CRC | Unknown | info |
3020 | DM_ERROR_CHECK_NEWEST_FIND | Unknown | info |
3021 | DM_ERROR_CHECK_OLDEST_FIND | Unknown | info |
3022 | DM_ERROR_CHECK_EVENT_ID | Unknown | info |
3023 | DM_ERROR_CHECK_REST_BLOCK | Unknown | info |
3024 | DM_ERROR_MAIN_PKT_SEM | Unknown | info |
3025 | DM_ERROR_MAIN_CBUFFER_SEM | Unknown | info |
3026 | DM_ERROR_CHECK_OLDEST_TOTAL | Unknown | info |
3027 | DM_ERROR_CHECK_NEWEST_TOTAL | Unknown | info |
3028 | DM_ERROR_DELETE_EPISODE | Unknown | info |
3029 | DM_ERROR_STORE_VITAL_SIGNS | Unknown | info |
302a | DM_ERROR_STORE_WRITE | Unknown | info |
302b | DM_ERROR_DATABASE_RESET | Unknown | info |
302c | DM_ERROR_CHECK_FIRST_RECORD | Unknown | info |
302d | DM_ERROR_SCP_LEAD_ERROR | Unknown | info |
302e | DM_ERROR_CHECK_EPISODE | Unknown | info |
302f | DM_ERROR_STORAGE_OPEN | Unknown | info |
3030 | DM_ERROR_STORAGE_CLOSE | Unknown | info |
3031 | DM_ERROR_STORAGE_ERASE | Unknown | info |
3032 | DM_ERROR_STOARGE_CLEAR | Unknown | info |
3033 | DM_ERROR_STORAGE_WRITE | Unknown | info |
3034 | DM_ERROR_STORAGE_READ | Unknown | info |
3035 | DM_ERROR_STORAGE_CRC | Unknown | info |
3036 | DM_ERROR_STORE_WFSEM_TIMEOUT | Unknown | info |
3037 | DM_ERROR_STORE_SEM_TIMEOUT | Unknown | info |
3038 | DM_ERROR_DATABASE_SEM_TIMEOUT | Unknown | info |
3039 | DM_ERROR_FLASH_SEM_TIMEOUT | Unknown | info |
303a | DM_ERROR_CONTECG_SEM_TIMEOUT | Unknown | info |
303b | DM_ERROR_BUFFER_SEM_TIMEOUT | Unknown | info |
303C | DM_ERROR_BUFFER_SEM_CREATE | Unknown | info |
3040 | DM_ERROR_CONTECG_SEM_TIMEOUT | Unknown | info |
3041 | DM_ERROR_BUFFER_SEM_TIMEOUT | Unknown | info |
3042 | DM_ERROR_BUFFER_SEM_CREATE | Unknown | info |
3043 | DM_ERROR_FLASH_BLOCK | Unknown | info |
3044 | DM_ERROR_FLASH_OFFSET | Unknown | info |
3045 | DM_ERROR_WR_HD_VERIFY | Unknown | info |
31 | Clear Data Management memory | Data management memory error | warning |
32 | Replace PCMCIA modem card | Modem card failure | error |
33 | SpO2 sensor or connection error | SpO2 sensor malfunction or disconnection | error |
34 | Check NIBP calibration | NIBP calibration required or failed | warning |
35 | Perform CO2 calibration | CO2 calibration required or failed | warning |
3ffe | DM_DIAG_EXTRA_INFORMATION | Unknown | info |
3fff | DM_ERROR_EXTRA_INFORMATION | Extra error code information follows for a preceding error | info |
4 | System Error | Abnormal user activity detected | warning |
4001 | System manager init error with unknown task ID | Initialization failure with unrecognized task | critical |
4002 | SM_ERROR_UNKNOWN_TASK_ID | Unknown | info |
4003 | SM_ERROR_TASK_SUSPENDED | Unknown | info |
4004 | SM_ERROR_TASK_NOT_FOUND | Unknown | info |
4005 | SM_ERROR_LOG_CRC | Unknown | info |
4006 | SM_ERROR_LOG_MSGQ_NOT_CREATED | Unknown | info |
4007 | SM_ERROR_LOG_MSGQ_ERROR | Unknown | info |
4008 | SM_ERROR_LOG_LENGTH | Unknown | info |
4009 | SM_ERROR_RAM_FAILURE | RAM failure during selftest; 16-bit RAM test failure | error |
400a | SM_ERROR_BAD_CRC | CRC in program flash bad; Program test failure | error |
400b | SM_ERROR_CRC_FAILURE | Program contents failed CRC test | error |
400c | SM_ERROR_VOLTAGE_LOW | ADC voltage reading low; HW voltage low condition | warning |
400d | SM_ERROR_VOLTAGE_HIGH | ADC voltage reading high; HW voltage high condition | warning |
400e | SM_ERROR_STACK_LOW | Unknown | info |
400f | SM_ERROR_TASK_INFO_GET | Unknown | info |
4010 | SM_ERROR_SERVICE_LED | Service LED failure; LED expected to be on and it is not | warning |
4011 | SM_ERROR_DEFIB_SERVICE_SYNC | Failed to synchronize the defib charge after cold boot | error |
4012 | SM_ERROR_FONT_VOICE_CKSUM | Invalid checksum in font/voice after cold boot | error |
4013 | SM_ERROR_FONT_VOICE_CRC | Invalid CRC in font/voice flash memory | error |
4014 | SM_ERROR_DIAG_CRC | Unknown | info |
4015 | SM_ERROR_DIAG_LENGTH | Unknown | info |
5 | Test Interface Input Error | Invalid user input or action | warning |
5001 | PC_ERROR_SYNC_SEM_NOT_CREATED | Unknown | info |
5002 | Main watchdog short test failure | Processor watchdog circuit fails short test during self-check | error |
5003 | Main watchdog long test failure | Processor watchdog circuit fails long test during self-check | error |
5004 | Watchdog timer triggered unexpectedly | Unexpected watchdog timeout during normal operation | critical |
5004 | Unexpected main watchdog reset | Processor resets unexpectedly due to watchdog timeout | error |
5005 | RAM error detected during boot | Processor detects memory error during system initialization | error |
5006 | Program checksum failure | Processor program code fails checksum verification test | error |
5007 | Real-time clock semaphore not created | System initialization failure | error |
5008 | Non-volatile memory semaphore not created | System initialization failure | error |
5009 | PC_ERROR_TASK_NOT_SPAWNED | Unknown | info |
500a | Task initialization timeout | Task failed to initialize within timeout period | error |
500b | Task initialization semaphore error | Semaphore creation failed during task setup | error |
500c | Task delinquent | Required task missed deadline or is not responding | warning |
500d | System configuration version mismatch | Processor detects configuration data version incompatibility during boot | error |
500e | NVRAM system configuration CRC bad | Processor detects corrupted system configuration in non-volatile RAM | error |
500f | PC_ERROR_CONFIG_LIMIT | Unknown | info |
500f | Configuration parameter out of limit | Stored configuration value exceeds valid range | warning |
5010 | Watchdog creation failed | System unable to create watchdog timer during boot | error |
5010 | PC_ERROR_CREATE_WATCHDOG | Unknown | info |
5011 | System meters version mismatch | Processor detects version incompatibility in stored meter data | error |
5012 | NVRAM system meters/counters CRC bad | Processor detects corrupted meter or counter data in non-volatile RAM | error |
5012 | Meters data checksum error | Meters data failed integrity verification | warning |
5013 | Manufacturing data version mismatch | Processor detects version conflict in manufacturing data | error |
5014 | NVRAM manufacturing data CRC bad | Processor detects corrupted manufacturing data in non-volatile RAM | error |
5015 | Forced watchdog reset failed | Processor fails to reset when forced watchdog reset is triggered | critical |
5016 | Invalid shutdown time recorded | Shutdown timestamp outside valid range | warning |
5017 | Invalid absolute shutdown time | Absolute shutdown timestamp is invalid | warning |
5018 | Pacer power control error | Pacer power output not responding correctly | error |
5019 | PC_ERROR_RTC_BAD | Real time clock error detected | unknown |
5019 | RTC is not running | Real-time clock (RTC) circuit not functioning or stopped | error |
501a | PC_ERROR_RTC_DRIFT | Real time clock drift detected | unknown |
501a | Processor and RTC time out of sync | System time differs significantly from RTC time | warning |
501b | PC_ERROR_EXC_UNKNOWN | Unknown exception error | unknown |
501b | Unknown processor fault | Processor detects unknown fault type; reported twice with fault and thread IDs | error |
501c | Processor parallel fault | Processor detects parallel execution fault; reported twice with IDs | error |
501d | Processor instruction trace fault | Processor detects instruction execution trace error; reported twice with IDs | error |
501e | Processor branch trace fault | Processor detects branch prediction or execution fault; reported twice with IDs | error |
501f | Processor call trace fault | Processor detects call instruction execution fault; reported twice with IDs | error |
5020 | Processor return trace fault | Processor detects return instruction fault; reported twice with IDs | error |
5021 | Processor prereturn trace fault | Processor detects prereturn instruction fault; reported twice with IDs | error |
5022 | Processor supervisor trace fault | Processor detects supervisor mode execution fault; reported twice with IDs | error |
5023 | Processor mark trace fault | Processor detects mark instruction fault; reported twice with IDs | error |
5024 | Processor trace fault unknown subtype | Processor detects unknown trace fault type; reported twice with IDs | error |
5025 | Processor invalid opcode operation fault | Processor executes invalid or undefined opcode; reported twice with IDs | error |
5026 | Processor unimplemented operation fault | Processor attempts unimplemented operation; reported twice with IDs | error |
5027 | PC_ERROR_EXC_OP_UNALIGNED | Unaligned operation | unknown |
5027 | Processor unaligned operation fault | Processor detects misaligned memory access; reported twice with IDs | error |
5028 | Processor invalid operand operation fault | Processor detects invalid operand in operation; reported twice with IDs | error |
5029 | Processor operation fault unknown subtype | Processor detects unknown operation fault; reported twice with IDs | error |
502a | PC_ERROR_EXC_ARITH_OVERFLOW | Arithmetic overflow | unknown |
502a | Processor integer overflow arithmetic fault | Processor detects integer overflow in arithmetic operation; reported twice with IDs | error |
502b | PC_ERROR_EXC_ARITH_ZERO_DIVIDE | Division by zero | unknown |
502b | Processor zero-divide arithmetic fault | Processor detects division by zero; reported twice with IDs | error |
502c | Processor arithmetic fault unknown subtype | Processor detects unknown arithmetic fault; reported twice with IDs | error |
502d | Processor constraint fault | Processor detects constraint violation; reported twice with IDs | error |
502e | Processor protection fault | Processor detects memory protection violation; reported twice with IDs | error |
502f | Processor type fault | Processor detects type mismatch or type violation; reported twice with IDs | error |
502f | PC_ERROR_EXC_TYPE | Exception type error in software | error |
5030 | PC_ERROR_CRITICAL_DATA_CORRUPT | Critical data corruption detected | critical |
5031 | PC_ERROR_RESET_UNKNOWN | Unknown | info |
5100 | PC_ERROR_UT_DELINQUENT | Unknown | info |
5101 | PC_ERROR_UI_DELINQUENT | Unknown | info |
5102 | PC_ERROR_DC_DELINQUENT | Unknown | info |
5103 | PC_ERROR_DM_DELINQUENT | Unknown | info |
5104 | PC_ERROR_SM_DELINQUENT | Unknown | info |
5105 | PC_ERROR_PC_DELINQUENT | Unknown | info |
5106 | PC_ERROR_ECG_DELINQUENT | Unknown | info |
5107 | PC_ERROR_SAS_DELINQUENT | Unknown | info |
5108 | PC_ERROR_L12_DELINQUENT | Unknown | info |
5109 | PC_ERROR_PP_DELINQUENT | Unknown | info |
510a | PC_ERROR_TH_DELINQUENT | Unknown | info |
510b | PC_ERROR_PR_DELINQUENT | Unknown | info |
510c | PC_ERROR_PM_DELINQUENT | Unknown | info |
510d | PC_ERROR_SC_DELINQUENT | Unknown | info |
6 | Therapy Processor Error | Therapy processing failure | error |
6001 | ECG_ERROR_X | Unknown | info |
6002 | ECG_ERROR_MSG_QUEUE_CREATE | Message queue creation failure in ECG system | error |
6003 | ECG_ERROR_MSG_QUEUE_SEND | Message queue send failure in ECG system | error |
6004 | ECG_ERROR_MSG_QUEUE_RECEIVE | Message queue receive failure in ECG system | error |
6005 | ECG_ERROR_CONNECT_SYS_INT | System interface connection failure in ECG | error |
6006 | ECG_ERROR_CONNECT_PKT_INT | Packet interface connection failure in ECG | error |
6007 | ECG_ERROR_SAS_MOTION_QUEUE_SEND | Motion queue message send failure from ECG to SAS | error |
6008 | ECG_ERROR_SAS_ECG_QUEUE_SEND | ECG queue message send failure to SAS | error |
6009 | ECG_ERROR_CREATE_WATCHDOG | Watchdog timer creation failure in ECG | error |
600a | ECG_ERROR_SEMAPHORE_CREATE | Semaphore creation failure in ECG system | error |
600b | ECG_ERROR_CMD_QUEUE_FULL | ECG command queue is full | warning |
600c | ECG DSP preamp supply voltage out of range | DSP preamp supply voltage detected outside acceptable range | error |
600c | ECG_ERROR_DSP_VOLTAGE | DSP voltage error in ECG system | error |
600d | ECG preamp calibration error | NVRAM calibration constants out of range or HW unit calibration error | error |
600e | ECG NVRAM fault | NVRAM redundant value mismatch detected indicating memory corruption | error |
7 | Power Processor Error | Power processing failure | error |
7001 | SAS_ERROR_X | SAS (Sensor/Acquisition System) error | error |
7002 | SAS_ERROR_WDOG_STATUS | Watchdog status error in SAS | error |
7003 | SAS_ERROR_WDOG_CREATE | Watchdog creation failure in SAS | error |
7004 | SAS_ERROR_INVALID_CMND | Invalid command received by SAS | error |
7005 | SAS_ERROR_INVALID_STATE | Invalid state in SAS system | error |
7006 | SAS_ERROR_LIMIT_CTR_PTR | Limit counter pointer error in SAS | error |
7007 | SAS_ERROR_INVALID_TI_COMMAND | Unknown | info |
7008 | SAS_ERROR_TI_COMMAND_NOT_IDLE | Unknown | info |
7009 | SAS_ERROR_TI_REQUEST_WRITE | Unknown | info |
700a | SAS_ERROR_TI_MSGQ_NOT_CREATED | Unknown | info |
700b | SAS_ERROR_TI_QUEUE_READ | Unknown | info |
700c | SAS_ERROR_UI_QUEUE_READ | Unknown | info |
700d | SAS_ERROR_UI_COMMAND_NOT_IDLE | Unknown | info |
700e | SAS_ERROR_INVALID_UI_COMMAND | Unknown | info |
700f | SAS_ERROR_UI_QUEUE_WRITE | Unknown | info |
7010 | SAS_ERROR_ECG_FALLING_BEHIND | Unknown | info |
7011 | SAS_ERROR_IMP_FALLING_BEHIND | Unknown | info |
7012 | SAS_ERROR_UI_MSGQ_NOT_CREATED | Unknown | info |
7013 | SAS_ERROR_COULD_NOT_INIT | Unknown | info |
8 | Nonvolatile RAM lost coin battery power | At system startup | error |
8001 | L12_ERROR_UNKNOWN | Unknown condition in 12-lead ECG analysis | error |
8002 | L12_ERROR_ANALYSIS_SEMAPHORE | Semaphore error in ECG analysis thread | error |
8003 | L12_ERROR_STATE_SEMAPHORE | Semaphore error in ECG state management | error |
8004 | L12_ERROR_QUEUING_COMMANDS | Error queuing ECG analysis commands | error |
8005 | L12_ERROR_CREATING_QUEUE | Failure to create command queue | error |
8006 | L12_ERROR_UNKNOWN_START | Unknown start condition in ECG processing | error |
8007 | L12_ERROR_UNKNOWN_COMMAND | Unrecognized command in ECG processing | error |
8008 | L12_ERROR_BAD_STATE | Invalid state detected in ECG processing | error |
8009 | L12_ERROR_FALLING_BEHIND | System falling behind real-time processing | unknown |
9 | Additional error extension | Error indicator for extended error information | info |
9001 | PPSP_ERROR_OPEN_DEVICE | Patient parameters subsystem failed to open device | error |
9002 | PPSP_ERROR_MODULE | Module hardware or firmware error detected | unknown |
9002 | Masimo SpO2 module error | SpO2 module reports error (error code 63 may be followed by diagnostic code) | error |
9003 | PPSP_ERROR_PLETHYSMOGRAPH_INVALID | Invalid plethysmograph waveform or data detected | unknown |
9004 | PPSP_ERROR_INVALID_FAST_MSG | Malformed fast-rate message from device | unknown |
9005 | PPSP_ERROR_VERSION_MSG_LENGTH | Version message length incorrect or invalid | unknown |
9006 | PPSP_ERROR_INVALID_SLOW_MSG | Malformed slow-rate message from device | unknown |
9007 | PPSP_ERROR_SLOW_DATA_LENGTH | Slow-rate data message length invalid | unknown |
9008 | PPSP_ERROR_SLOW_MSG_LENGTH | Slow message frame length out of specification | unknown |
9009 | PPSP_ERROR_SLOW_MSG_CHECKSUM | Checksum validation failed on slow message | unknown |
900a | SpO2 module communication failure | Failed to communicate with SpO2 module after 5 seconds | error |
900b | PPSP_ERROR_MFG_ID | Manufacturing ID invalid or not recognized | unknown |
900c | SpO2 module write command failure | Failure to write setup commands to SpO2 module | error |
900c | PPSP_ERROR_WRITING_TO_MODULE | Failed to write data or firmware to module | unknown |
900d | PPSP_ERROR_SPO2_VALUE | Invalid or out-of-range SpO2 measurement value | unknown |
900d | Invalid SpO2 value or pulse rate | SpO2 module reports invalid saturation or pulse rate | warning |
900e | SpO2 module manufacturing configuration error | SpO2 module detected but not in manufacturing configuration | error |
900e | PPSP_ERROR_CONFIG | Module configuration error or parameter invalid | unknown |
900f | PPSP_ERROR_RAW_IOCTL_FAILED | Raw I/O control command failed | unknown |
9010 | PPSP_ERROR_PARITY_IOCTL_FAILED | Parity I/O control command failed | unknown |
9011 | PPSP_ERROR_ENABLE_IOCTL_FAILED | Enable I/O control command failed | unknown |
9012 | PPSP_ERROR_FLUSH_IOCTL_FAILED | Flush buffer I/O control command failed | unknown |
9013 | SpO2 device poll failure | Unable to poll SpO2 device; cable disconnected | error |
9014 | PPSP_ERROR_READ_FAILED | Failed to read data from device or module | unknown |
9015 | SpO2 write operation failed | write() call failed to SpO2 module | error |
9016 | PPSP error select failed | Select operation failure in PPSP | error |
9017 | PPSP error wrong length written | Incorrect data length written to PPSP | error |
9017 | SpO2 write returned incorrect length | Short write to SpO2 module returned fewer bytes than requested | error |
9018 | SpO2 sensor over current error | SpO2 device reports over current condition | error |
9019 | PPSP error RAM | RAM fault detected | error |
9019 | SpO2 module RAM error | SpO2 device reports RAM error | error |
901a | PPSP error ROM | ROM fault detected | error |
901a | SpO2 module ROM error | SpO2 device reports ROM error | error |
901b | PPSP error unknown | Unknown PPSP error condition | error |
9028 | PPSP_DIAG_OVER_CURRENT | Over-current detected in patient parameters subsystem | error |
9029 | PPSP_DIAG_RAM | RAM diagnostic failure | error |
902a | PPSP_DIAG_ROM | ROM diagnostic failure | error |
9101 | PPNI_ERROR_OPEN_DEVICE | Non-invasive blood pressure module failed to open device | error |
9102 | PPNI_ERROR_CHECKSUM | Message checksum failure in NIBP module | error |
9103 | NIBP module communication failure | Failed to communicate with NIBP blood pressure module | error |
9104 | PPNI_ERROR_MFG_ID | Manufacturing ID invalid in NIBP module | error |
9105 | PPNI_ERROR_WRITING_TO_MODULE | Failed to write data to NIBP module | error |
9106 | PPNI_ERROR_CONFIG | Configuration error in NIBP module | error |
9107 | NIBP module self-test failure | NIBP module failed self-test with specific fault bits | error |
9108 | NIBP ioctl options set failure | SW ioctl() call failed for options set | error |
9109 | NIBP ioctl parity set failure | SW ioctl() call failed for parity set | error |
910a | NIBP ioctl IO enable set failure | SW ioctl() call failed for IO enable set | error |
910b | NIBP ioctl IO flush failure | SW ioctl() call failed for IO flush | error |
910c | NIBP device poll failure | SW ioctl() call failed for device poll | error |
910d | NIBP serial read operation failed | read() call failed from NIBP module | error |
910e | NIBP write operation failed | write() call failed to NIBP module | error |
910f | PPNI_ERROR_SELECT_FAILED | Failed to select NIBP module | error |
9110 | NIBP write returned incorrect length | Short write to NIBP module returned fewer bytes than requested | error |
9111 | PPNI_ERROR_RING_BUFF_CREATE_FAILED | Failed to create ring buffer in NIBP module | error |
9112 | PPNI_ERROR_RING_BUFFER_WRITE_FAILED | Failed to write to ring buffer in NIBP module | error |
9113 | PPNI_ERROR_INCORRECT_RESULTS_STATUS | Incorrect blood pressure results status | error |
9114 | PPNI_ERROR_BAD_PRESSURE_SELECTED | Invalid pressure selection in NIBP measurement | error |
9115 | PPNI_ERROR_BAD_INTERVAL_SELECTED | Invalid measurement interval selected | error |
9116 | NIBP module reset detected | NIBP module reset detected outside first 10 seconds of reading | error |
9117 | NIBP module not responding | NIBP module not responding to start reading command | error |
9118 | PPNI_ERROR_LEAKAGE_TEST_FAILED | Leakage test failed during NIBP operation | error |
9119 | NIBP module configuration mismatch | Device configured without NIBP module but module detected or vice versa | error |
9201 | PPCO_ERROR_OPEN_DEVICE | Carbon monoxide monitor module failed to open device | error |
9202 | PPCO_ERROR_CHECKSUM | Message checksum failure in CO monitor module | error |
9203 | PPCO_ERROR_COMMUNICATION | Communication failure with CO monitor module | error |
9204 | CO2 module MFG ID communication failure | Failure to communicate with CO2 module or receive MFG ID | error |
9205 | CO2 module manufacturing configuration error | CO2 module detected but not in manufacturing configuration | error |
9206 | CO2 module service code error | CO2 module service code reported; likely hardware fault | error |
9207 | PPCO_ERROR_RAW_IOCTL_FAILED | Raw I/O control operation failed | error |
9208 | CO2 calibration parity failure | co2 calibration failure with AZ fail reported by module | error |
9209 | CO2 calibration enable failure | co2 calibration failure reported by module on enable | error |
920a | PPCO_ERROR_FLUSH_IOCTL_FAILED | Buffer flush operation failed | error |
920b | CO2 calibration poll failure | CO2 calibration failed during poll; unknown failure in service mode | error |
920c | CO2 waveform message cycle counter out of sync | CO2 waveform message cycle counter desynchronized | error |
920d | CO2 write operation failed (obsolete) | Obsolete error code | unknown |
920e | CO2 select operation failed (obsolete) | Obsolete error code | unknown |
920f | CO2 write returned incorrect length (obsolete) | Obsolete error code | unknown |
a001 | TH_ERROR_ERROR_UNKNOWN | Unknown error in therapy subsystem | error |
a002 | Lost contact with defib processor. Serial communications link between main and defib processor not functioning. | Defib processor communication failure | critical |
a002 | TH error defibrillator link down | Defibrillator communication loss | error |
a003 | Lost contact with pacer processor. Serial communications link between main and pacer processor not functioning. | Pacer processor communication failure | critical |
a003 | TH error pacer link down | Pacer communication loss | error |
a004 | Unexpected energy in the capacitor. No charge delivered. | Capacitor energy present unexpectedly | error |
a004 | TH error unexpected energy | Unexpected energy detected | error |
a005 | TH error capacitor overcharged | Capacitor voltage exceeds limit | error |
a006 | TH error wrong defibrillator mode | Incorrect defibrillator mode detected | error |
a007 | TH error wrong defibrillator state | Incorrect defibrillator state | error |
a008 | No communication with Defib HW. | Defib hardware communication loss | critical |
a008 | TH error defibrillator disable | Defibrillator disabled condition | error |
a009 | TH error unknown defibrillator response | Unexpected response from defibrillator | error |
a00a | Incorrect energy select. | Wrong energy selection | warning |
a00a | TH error defibrillator wrong engine select | Incorrect engine selected | error |
a00b | TH error shock not delivered | Defibrillator shock failed to deliver | critical |
a00c | TH error wrong transfer request | Invalid transfer request received | error |
a00d | TH error charging expired | Charging timeout or expiration | error |
a00e | TH error wrong charge request | Invalid charge request received | error |
a00f | Capacitor is out of range. Calibration Failure | Capacitor voltage out of specification range | error |
a00f | TH_ERROR_CAP_OUT_OF_RANGE | Capacitor voltage reading falls outside specified range during operation or self-test | error |
a010 | TH_ERROR_PA_RATE_OUT_OF_RANGE | Pacing rate parameter measured outside acceptable limits | error |
a011 | TH_ERROR_PA_CURRENT_OUT_OF_RANGE | Pacing current output measured outside acceptable limits | error |
a012 | TH_ERROR_PA_MISS_RATE | Pacing pulse missing from output or timing irregularity detected | error |
a013 | TH_ERROR_PA_MISS_CURRENT | Pacing current pulse not detected or insufficient current output | error |
a014 | TH_ERROR_DE_MISS_ENERGY | Defibrillation energy not detected or insufficient discharge | critical |
a015 | TH_ERROR_MSGQ_NOT_CREATED | Message queue initialization failure during power-on self-test | error |
a016 | TH_ERROR_MSG_CHECKSUM | Message checksum validation failure during inter-module communication | error |
a017 | TH_ERROR_DEFIB with multiple extension codes | Defibrillation subsystem error with specific fault code extension | critical |
a017 | Obsolete error codes remapped to a101 through a133. | Legacy defib error condition | unknown |
a018 | Obsolete error codes remapped to a201 through a224. | Legacy pacer error condition | unknown |
a018 | TH_ERROR_PACER | Pacer subsystem error detected | error |
a019 | TH_ERROR_XFER_HOLD_TEST | Transfer hold test failure | unknown |
a01a | TH_ERROR_PACER_FAULT | Pacing circuit fault detected | unknown |
a01a | Pacing fault condition occurred (rate, current, or pulse width limit exceeded). | Pacing parameter exceeds safe limit | error |
a01b | TH_ERROR_DEFIB_WDT_DISABLE_FAIL | Defibrillation watchdog timer disable failed | unknown |
a01b | Unable to turn off defib WDT. | Defib watchdog timer disable failure | error |
a01c | TH_ERROR_UNKWN_ADC_READ_REQUEST | Unknown ADC read request received | unknown |
a01d | TH_ERROR_UNKWN_DEFIB_STATE | Unknown defibrillation state detected | unknown |
a01e | TH_ERROR_UNKWN_ERR_ACTION | Unknown error action in therapy | error |
a01f | TH_ERROR_UNKWN_TEST_RESULT | Unknown test result received | unknown |
a020 | TH_ERROR_PACER_DISABLE | Pacer disable command failure | unknown |
a020 | Pacer disable. Cannot communicate with processor | Pacer processor communication loss during operation | critical |
a021 | Cap stays zero while charging. No charge. | Capacitor fails to charge | critical |
a021 | TH_ERROR_CAP_CHARGE_FAIL | Capacitor charge failure | unknown |
a022 | TH_ERROR_CORRUPT_ENERGY_SELECT | Energy selection parameter corrupted | unknown |
a023 | TH_ERROR_XFER_ENABLE_ON | Transfer enable remains on | unknown |
a023 | Defib xfer enable line high unexpected. Transfer-enable line was not off during startup. | Transfer enable line not disabled at startup | error |
a024 | TH_ERROR_VCAP2_SATURATED | Secondary capacitor voltage saturated | unknown |
a024 | VCAP2 reading full scale all the time. Reading not processed. | VCAP2 measurement stuck at maximum | error |
a025 | TH_ERROR_CORRUPT_PACER_STAT | Pacer status parameter corrupted | unknown |
a026 | There is still energy on the cap | Residual capacitor charge detected | critical |
a026 | TH_ERROR_ENERGY_RESIDUE | Residual energy detected in capacitors | unknown |
a027 | Pacing rate storage corrupted. | Pacing rate memory corruption | error |
a028 | Calibrated voltage is out of range. | Energy calibration voltage exceeds limits | error |
a029 | Error condition with BTE board. | BTE (biphasic therapy engine) board fault | error |
a02a | TH_ERROR_BTE_UNKWN_ENERGY | Unknown battery energy in therapy | error |
a02b | Conflict defib type. Cold boot if test bit set else disable biphasic. | Defib type conflict detected | warning |
a02c | One of dump line failed. | Dump circuit line failure during energy discharge | error |
a02d | ADC read failure when getting cap charge | ADC cannot read capacitor voltage | error |
a101 | Long Watchdog test failed or Watchdog didn't reset in time. | Defib watchdog timer test failure | error |
a102 | Short Watchdog test failed. | Defib short watchdog timer failure | error |
a103 | Serial port receiver error | Defib serial port communication error | error |
a104 | Defib HW error | Defib hardware fault detected | critical |
a105 | DE_ERROR_REDUN_MEM | Redundant memory error in defibrillator | error |
a106 | VCAP-1 over/under charge. | VCAP-1 voltage out of specification | error |
a107 | Cannot determine the rotary setting | Energy rotary selector position unreadable | error |
a108 | DE_ERROR_INVALID_ENERGY | Invalid defibrillation energy value | error |
a109 | Calibration Table CRC error. | Calibration data corruption detected | error |
a10a | DE_ERROR_UKWN_STATE | Unknown defibrillator state | error |
a10b | Takes too long to reach charge. | Charging time exceeds specification | error |
a10c | CHG_EN1 is stuck high | Charge enable signal remains high | error |
a10d | Defib HW error | Dump energy failure | critical |
a10e | Defib HW error | Receive operation failure | critical |
a10f | DE_ERROR_BUF_OUT | Buffer overflow in defibrillator subsystem | error |
a110 | DE_ERROR_IGNORE_CHARGE | Ignoring charge command in defibrillator | error |
a111 | Unexpected energy while biphasic is in hi-pot test. | Residual capacitor energy during test | error |
a112 | Problem with xfer turn on at main. | Transfer enable signal failure at startup | error |
a113 | Error reading the transfer engage feedback. | Transfer engage feedback signal unreadable | error |
a114 | Biphasic Error | Dump relay test failure | error |
a115 | Defib HW error | ADC test failure | critical |
a116 | Defib HW error | DAC test failure | critical |
a117 | Defib HW error | Hardware self-test failure | critical |
a118 | Defib HW error | RAM test failure | critical |
a119 | Defib HW error | ROM test failure | critical |
a11a | Defib HW error | CPU test failure | critical |
a11b | DE_ERROR_TEST_STACK | Stack test failure in defibrillator | error |
a11c | DE_ERROR_UNKNOWN_COMMAND | Unknown command in defibrillator | error |
a11d | Defib HW error | Transfer operation timeout | critical |
a11e | Defib HW error | Button interface error | critical |
a11f | Defib HW error | Sync interrupt failure | critical |
a120 | Defib HW error | Self-test incomplete | critical |
a121 | DE_ERROR_MESSAGE_ABORTED | Message aborted in defibrillator subsystem | error |
a122 | DE_ERROR_MESSAGE_CHECKSUM | Message checksum failure in defibrillator | error |
a123 | CRC failed for new calibration data. | Calibration data CRC error | error |
a124 | Cannot write energy table to EEPROM. | EEPROM write failure for calibration | critical |
a125 | Byte could not be written to the DAC through the SPI interface. | DAC SPI write failure | error |
a126 | Cannot read from ADC. | ADC read failure | error |
a127 | Must be idle to switch to test mode. | Test mode request while unit not idle | warning |
a128 | DE_ERROR_TEST_COMMAND | Test command failure in defibrillator | error |
a129 | Defib HW error | Transfer cable fault | critical |
a12a | Defib HW error | Transfer paddle fault | critical |
a12b | PA_ERROR_BACKGROUND_IDLE | Background idle error in pacer subsystem | error |
a12c | Defib HW error | Charging inhibit fault | critical |
a12d | Charge enable feedback indicates not charging. | Charge enable feedback signal lost | error |
a12e | Cedar BTE Fault Line State | BTE fault line detected | error |
a12f | Cedar BTE Fault Line State cleared | BTE fault line cleared after fault | info |
a130 | Defib HW error | BTE reset failure | critical |
a131 | Biphasic HW not found. | BTE hardware not detected during initialization | critical |
a132 | Defib HW error | BTE transfer fault | critical |
a133 | A/D high bit sticky. | ADC stuck bit detected | error |
a201 | Long watchdog timer test failed | Pacer long watchdog timer test failure | error |
a202 | Short watchdog timer test failed | Pacer short watchdog timer failure | error |
a203 | SCI rcv overrun, framing or parity. Unit used near high EMI causing these issues. | Serial communication error from EMI or malfunction | error |
a204 | 2ms ECG sampling overrun | ECG sampling rate exceeded | error |
a205 | PA_ERROR_TICK_OVERRUN | Timer tick overrun in pacer subsystem | error |
a206 | PA_ERROR_MSG_CHKSUM | Message checksum failure in pacer | error |
a207 | PA_ERROR_MSG_ID | Invalid message ID in pacer subsystem | error |
a208 | Rcvd message incomplete | Message reception incomplete or timeout | error |
a209 | Rcvd msg size error/input buff full | Message size error or input buffer overflow | error |
a20a | PA_ERROR_SEQUENCING | Sequencing error in pacer subsystem | error |
a20b | PA_ERROR_UNKNOWN_RATE | Unknown pacing rate in pacer subsystem | error |
a20c | PA_ERROR_UNKNOWN_CURRENT | Unknown pacing current in pacer subsystem | error |
a20d | Pacing pulse process overrun. Set current to zero. | Pacing pulse processing rate exceeded | error |
a20e | Pacing pulse too long | Pacing pulse width exceeds specification | error |
a20f | Internal A/D conversion timeout. Set current to zero. | ADC internal conversion timeout | error |
a210 | External A/D conversion timeout. Current set to zero | ADC external conversion timeout | error |
a211 | SPI transfer timeout. Current set to zero. | SPI interface communication timeout | error |
a212 | RAM test failed. Reset Pacer Processor | Pacer processor RAM test failure | critical |
a213 | ROM crc test failed | Pacer ROM CRC error | critical |
a214 | Stack overrun occured | Pacer processor stack overflow | critical |
a215 | Isolated +5V ref. out of range | Isolated 5V reference voltage out of specification | error |
a216 | HV present when not pacing | High voltage detected outside pacing cycle | critical |
a217 | +12V voltage out of range | 12V supply voltage out of specification | error |
a218 | HV present when not pacing | High voltage detected outside pacing cycle (HVIS sense) | critical |
a219 | HVIS voltage out of range | Pacing high voltage sense signal out of specification | error |
a21a | Current calibration failed | Pacing current calibration procedure failure | error |
a21b | Impedance 300 calibration failed | 300 ohm impedance calibration failure | error |
a21c | Current cal table corrupt | Current calibration table corruption detected | error |
a21d | Impedance cal values corrupt | Impedance calibration value corruption | error |
a21e | Impedance 0 calibration failed | Zero ohm impedance calibration failure | error |
a21f | Current present when not pacing | Pacing current detected outside pacing cycle | critical |
a220 | No HV present when pacing | Pacing high voltage not present during pacing cycle | critical |
a221 | External A/D test register reset failed | External ADC test register reset failure | error |
a222 | No HV present before current cal | High voltage absent during current calibration | error |
a223 | No Q4 sense when not pacing (Q4 shorted) | Q4 output shorted to ground | critical |
a224 | Q4 sense present when pacing (CR25 open) | Q4 feedback lost during pacing (CR25 open circuit) | critical |
b001 | Print head ADC reading too low | Printer power-on or operation | error |
b002 | Print head ADC reading too high | Printer power-on or operation | error |
b003 | Self-test failed with invalid CRC from hardware | Printer self-test at power-on | error |
b003 | PR_ERROR_SELF_TEST_FAIL | Power-on self-test failure | error |
b004 | PR_ERROR_WRONG_PRINTER_TYPE | Wrong printer type detected or configured | error |
b004 | Printer type mismatch detected | Printer initialization | error |
b005 | PR_ERROR_FALLING_BEHIND | Printer falling behind in processing | warning |
b006 | PR_ERROR_WAITING_FOR_DATA | Printer waiting for incoming data | info |
b007 | PR_ERROR_INVALID_ANNOTATION_STATE | Invalid annotation state encountered | error |
b008 | PR_ERROR_INVALID_LEAD_TO_STATUS | Invalid lead status detected | error |
b009 | PR_ERROR_INVALID_ALARM_EVENT | Invalid alarm event detected | error |
b00a | PR_ERROR_INVALID_PACER_EVENT_TYPE | Invalid pacer event type encountered | error |
b00b | PR_ERROR_INVALID_STROBE_STATE | Invalid strobe state detected | error |
b00c | PR_ERROR_INVALID_EVENT | Invalid event encountered | error |
b00d | PR_ERROR_INVALID_L12_LEAD | Invalid L12 lead configuration | error |
b00e | PR_ERROR_INVALID_GAIN | Invalid gain value detected | error |
be00 | PRFAX_ERROR_RING_BUFFER_CREATE | Failed to create ring buffer for FAX printer | error |
be01 | PRFAX_ERROR_ACCESS_RING_BUFFER | Failed to access ring buffer for FAX printer | error |
be02 | PRFAX_ERROR_4DEBUG | Debug-related error in FAX printer | error |
c001 | PM_ERROR_MSGQ_ERROR | Message queue error in power management | error |
c002 | PM_ERROR_NO_RESPONSE: No response from power board | Power management background check detecting communication failure | error |
c003 | PM_ERROR_MSG_UNKNOWN | Unknown message received | error |
c004 | PM_ERROR_BAD_CHECKSUM | Message checksum error | error |
c005 | PM_ERROR_MSG_SIZE | Message size error | error |
c006 | PM_ERROR_POWER_PCB_FAULT | Power PCB fault detected | critical |
c006 | PM_ERROR_POWER_PCB_FAULT: power PCB fault reported; fault status indicates PCB, battery, auxiliary power or communication issue | Power management background check | error |
CALIBRATION FAILED | EtCO2 module calibration did not complete successfully | Calibration process terminated without successful completion | error |
d001 | SC_ERROR_DRIVER_INSTALL | Serial communication driver installation failed | error |
d002 | SC_ERROR_DEVICE_CREATE | Device creation error | error |
d003 | SCI frame error | Serial communication frame error detected | error |
d004 | SCI parity error | Parity check failed during serial communication | error |
d005 | SCI read buffer full | Serial input buffer overflow, unable to store incoming data | warning |
d005 | Serial communication error: read full | Serial communications subsystem cannot complete full read operation | error |
d006 | Serial communication error: SCI initialize failed | Serial communications interface initialization fails during startup | error |
d006 | SCI did not initialize | Serial communication interface failed to initialize at startup | critical |
d007 | Device not supported by system | Detected device is not in system supported device list | error |
d008 | SC_ERROR_COULD_NOT_OPEN | Failed to open serial communication device | error |
d008 | Could not open serial communication | System cannot establish communication channel to device | error |
d009 | SC_ERROR_SCI_FIOSETOPTIONS | Failed to set serial I/O options | error |
d009 | Serial port option configuration failed | System cannot set required serial port options during initialization | error |
d00a | Message size error in communication | Received or sent message exceeds defined size limits | error |
d00a | SC_ERROR_MSG_SIZE | Invalid message size in serial communication | error |
d00b | SC_ERROR_SCI_WRITE | Serial communication write operation failed | error |
d00b | Serial communication write operation failed | System cannot transmit data to device via serial port | error |
d00c | SC_ERROR_SELECT | Device selection failed in serial communication | error |
d00c | Select operation failed in communication subsystem | System select call fails during port polling or data waiting | error |
d00d | Queue Full (message type: 0=pacer, 1=Defib, 2=power) | Message queue overflow for specified message type | error |
d00d | Message queue error in communication subsystem | Message queue operation fails during data handling | error |
d00e | Message queue could not be created | System cannot allocate resources to create message queue at startup | error |
d00f | Receive semaphore could not be created | System cannot create synchronization semaphore during initialization | error |
d010 | SC_ERROR_IOCTL_FAILED | I/O control operation failed in serial communication | error |
d010 | IOCTL (I/O control) operation failed | System I/O control command returns failure | error |
d011 | Paddles communication checksum error | Checksum validation fails on message from paddle device | error |
d011 | SC_ERROR_PADDLES_MSG_CKSUM | Paddle message checksum failure in serial communication | error |
d012 | SC_ERROR_PADDLES_MSG_SIZE | Invalid paddle message size in serial communication | error |
d012 | Paddles communication message size error | Received paddle message size exceeds or falls below expected length | error |
d013 | Paddles communication message ID error | Received message from paddles contains unrecognized message ID | error |
d013 | SC_ERROR_PADDLES_MSG_ID | Invalid paddle message ID in serial communication | error |
Error codes (general) | Internal program or process failure | Internal program or process fails to execute properly; EMI or ESD may cause transient errors | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI