Philips VS4 QuickAlerts Configuration: troubleshooting and error codes
The ServiceForge knowledge base holds 65 documented symptoms for the Philips VS4 QuickAlerts Configuration, with 136 ranked causes, 123 diagnostic checks with test points and expected values, and 128 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Temperature measurements not displayed
Temperature reading does not appear on monitor screen
Ranked causes
- 1. Temperature probe disconnected from monitor (common)
- 2. Temperature probe misplaced (common)
- 3. Predictive temperature out of range (common)
- 4. Temperature probe malfunctioned (occasional)
- 5. Front end board malfunction (occasional)
- 6. Temperature module malfunction (occasional)
- 7. Main board malfunction (occasional)
Verify temperature probe connection and seating
- Test point
- Temperature probe connector and probe well
- Expected
- Probe firmly seated in probe well
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurements not displayed patient safety
NBP readings are not shown on monitor screen even after measurement attempt
Ranked causes
- 1. Wrong cuff size or incorrect cuff placement (common)
- 2. Initial pressure setting incorrect (common)
- 3. Air tube bent, twisted, or cuff has air leakage (common)
- 4. External problem with measurement conditions (common)
- 5. NBP pump filter is blocked (occasional)
- 6. NBP module error (occasional)
Verify cuff size and placement
- Test point
- Patient arm and cuff fit
- Expected
- Correct size cuff applied at correct location
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not turn on with AC power high voltage
Monitor powers on with battery but not when connected to AC outlet
Ranked causes
- 1. AC power cord unplugged or broken (common)
- 2. Fuse is blown (common)
- 3. Power supply cable malfunction (occasional)
- 4. Power supply output voltage incorrect (occasional)
- 5. AC power connector malfunction (occasional)
- 6. Main board malfunction (occasional)
Verify AC power cord connection
- Test point
- AC outlet and monitor connector
- Expected
- Power cord plugged in securely
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor power and basic operation troubleshooting
Initial diagnostic checks before detailed troubleshooting: power switch status, battery charge, AC connection, display function, and LED indicators
Ranked causes
- 1. Power switch not enabled (common)
- 2. Battery not adequately charged (common)
- 3. AC power not connected or outlet non-functional (common)
- 4. Display malfunction (occasional)
- 5. LED indicators not illuminated as expected (occasional)
Verify power switch is turned on
- Test point
- Power switch on front panel
- Expected
- Power switch in ON position
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP cuff does not inflate patient safety
Blood pressure cuff fails to inflate when NBP measurement is initiated
Ranked causes
- 1. Air tube or cuff bent or twisted (common)
- 2. Air leak in air tube or cuff (common)
- 3. NBP module malfunctioned (occasional)
- 4. NBP pump filter is blocked (occasional)
- 5. Main board malfunctioned (occasional)
Inspect air tube and cuff for kinks
- Test point
- Air tube and cuff pathway
- Expected
- Tube and cuff straight and unkindled
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call function does not signal alarm conditions patient safety
Nurse call system fails to alert when monitor alarms are triggered
Ranked causes
- 1. Minimum Nurse Call Alarm Priority set too high (common)
- 2. Alarms have been silenced (common)
- 3. Phono connector is loose or broken (occasional)
- 4. Communications board malfunctioned (occasional)
- 5. Main board malfunctioned (occasional)
Verify nurse call alarm priority threshold
- Test point
- Configuration menu alarm priority settings
- Expected
- Priority threshold set appropriately low
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not print
Recorder module fails to produce printed output
Ranked causes
- 1. Recorder is out of paper (common)
- 2. Recorder door is open (common)
- 3. System does not recognize recorder (occasional)
- 4. Recorder module malfunctioned (occasional)
- 5. Main board malfunctioned (occasional)
Verify paper supply
- Test point
- Recorder paper roll
- Expected
- Paper loaded in recorder
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurements not displayed
Oxygen saturation readings do not appear on monitor screen
Ranked causes
- 1. SpO2 module in learning phase (common)
- 2. SpO2 sensor malfunctioned (occasional)
- 3. SpO2 board has faulty connection to front end board (occasional)
- 4. SpO2 board malfunctioned (occasional)
- 5. Main board malfunctioned (occasional)
Wait for SpO2 module learning phase to complete
- Test point
- SpO2 module status indicator
- Expected
- Learning phase completed
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not sound patient safety
Audio alarm functionality is silent; monitor does not produce alarm tones
Ranked causes
- 1. Monitor in Audio Pause or Audio Off mode (common)
- 2. Speaker connection faulty (occasional)
- 3. Speaker malfunctioned (occasional)
- 4. Main board malfunction (occasional)
Verify audio alarm mode setting
- Test point
- System Menu audio settings
- Expected
- Audio mode set to normal (not Audio Off or Audio Pause)
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not respond to navigation wheel or keys
Front panel keys and navigation wheel do not produce any function
Ranked causes
- 1. Main board has faulty connection (occasional)
- 2. Front panel keypad or navigation wheel broken (occasional)
- 3. Main board malfunctioned (occasional)
- 4. Membrane switch assembly malfunction (occasional)
Inspect main board connector seating
- Test point
- Main board connectors
- Expected
- All connectors firmly seated
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not respond to touch screen
Touch screen input is not recognized or does not function
Ranked causes
- 1. Touch screen is locked (common)
- 2. Touch screen is not configured (occasional)
- 3. Touch screen is not calibrated (occasional)
- 4. Touch screen malfunctioned (occasional)
Verify touch screen is not locked
- Test point
- System Menu or lock status display
- Expected
- Touch screen unlocked status shown
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurements unreliable patient safety
Blood pressure readings are inconsistent, inaccurate, or vary widely between measurements
Ranked causes
- 1. Wrong cuff size or incorrect cuff placement (common)
- 2. Initial pressure setting incorrect (common)
- 3. External problem with measurement conditions (common)
- 4. NBP module needs calibration (occasional)
Verify cuff size and placement
- Test point
- Patient arm and cuff fit
- Expected
- Correct size cuff applied at correct location
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurements unreliable
Temperature readings are inconsistent, inaccurate, or vary between measurements
Ranked causes
- 1. Temperature probe malfunctioned (common)
- 2. Temperature module not calibrated (common)
- 3. Front end board malfunction (occasional)
- 4. Temperature module malfunctioned (occasional)
Verify temperature module calibration
- Test point
- Calibration status in diagnostics
- Expected
- Module calibrated and within tolerance
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not turn on with battery power
Monitor powers on with AC but not when using battery
Ranked causes
- 1. Battery not charged (common)
- 2. Battery or battery connector loose (common)
- 3. Battery board not connected to main board (occasional)
- 4. Main board malfunction (occasional)
Verify battery charge status
- Test point
- Battery charging indicator or battery gauge
- Expected
- Battery charged or charging LED lit
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB hub does not work
USB hub ports are non-functional or not recognized
Ranked causes
- 1. USB hub is disconnected from monitor (common)
- 2. Hub USB ports are blocked by dust or dirt (common)
- 3. USB hub is broken (occasional)
- 4. LAN board is broken or not properly connected (occasional)
Verify USB hub cable connection to monitor
- Test point
- USB hub power and data connectors
- Expected
- All USB plugs firmly connected to ports
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor screen is blank
Power is on but display shows no image or content
Ranked causes
- 1. Monitor in Standby mode (common)
- 2. LCD backlight connection faulty (occasional)
- 3. Main board malfunction (occasional)
Verify monitor is not in Standby mode
- Test point
- Monitor status indicator and display
- Expected
- Monitor active, display lit
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Random or distorted graphics with white background
Display shows garbled text and images with white background
Ranked causes
- 1. LCD signal cable not attached correctly (common)
- 2. LCD malfunctioned (occasional)
- 3. Main board malfunction (occasional)
Inspect and reseat LCD signal cable
- Test point
- LCD signal cable connectors
- Expected
- Cable seated fully on both connectors
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder paper is jamming
Recorder paper does not feed properly or becomes stuck
Ranked causes
- 1. Paper not installed correctly (common)
- 2. Wrong type of paper installed (common)
- 3. Recorder has mechanical problem (occasional)
Remove and inspect recorder paper installation
- Test point
- Recorder paper path and rollers
- Expected
- Paper properly threaded and fed straight
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade failure or interruption patient safety
Monitor enters failed state if software upgrade is interrupted or USB drive is removed during process
Ranked causes
- 1. Upgrade interrupted by unplugging monitor during process (rare)
- 2. Upgrade interrupted by removing USB flash drive during process (rare)
- 3. Upgrade interrupted by pressing keys during process (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on
No power to monitor; verify AC power cord and battery connected, check LED indicators
Ranked causes
- 1. No AC power or battery installed (common)
- 2. Power supply or charging issue (common)
Verify AC power cord or battery inserted
- Test point
- Power connector and battery slot
- Expected
- Power source connected to monitor
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
NBP initial pressure too high patient safety
Blood pressure cuff inflates to excessively high pressure on initial measurement
Ranked causes
- 1. Cuff is too tight (common)
- 2. Initial inflation pressure setting incorrect (common)
Inspect cuff fit on patient
- Test point
- Cuff tightness around arm
- Expected
- Cuff snug but not constricting circulation
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder test pattern abnormal
Optional recorder prints incorrect or defective test pattern during diagnostics
Ranked causes
- 1. Faulty print head (unknown)
- 2. Motor malfunction (unknown)
Run recorder test from System Diagnostics menu
- Test point
- Recorder Test button
- Expected
- Test pattern prints
Self test fails on NBP module
Self test displays Failed (red text) for NBP test during diagnostic run
Ranked causes
- 1. NBP module not detected or NBP initialization failure (common)
- 2. Main board failure (occasional)
Run self test
- Test point
- System Diagnostics menu, Self Test button
- Expected
- All test results display Passed or Not Supported
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurements unreliable
Oxygen saturation readings are inconsistent, inaccurate, or unstable
Ranked causes
- 1. External problem such as excessive patient movement (common)
- 2. Perfusion is low (occasional)
Verify patient positioning and movement
- Test point
- Patient arm position and body movement
- Expected
- Patient calm and arm stable
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error code 1051: Cannot initialize NBP
Self test reports system error code 1051 during startup or diagnostic run
Ranked causes
- 1. NBP module communication failure (common)
- 2. Main board fault (occasional)
Shutdown and restart
- Test point
- Power supply
- Expected
- Monitor boots without 1051 error
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Access System Admin Menu with password
Procedure to unlock System Admin Menu for configuring password-protected functions including Demo Mode, system diagnostics, and software upgrade
Ranked causes
- 1. Need to access System Admin functions (common)
Access System Menu and select System Admin
- Test point
- System Menu interface
- Expected
- System Admin button visible and selectable in System Menu
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Access System Diagnostics for verification testing
Open System Diagnostics menu from System Admin to perform verification tests; monitoring is suspended while menu is open
Ranked causes
- 1. Need to perform verification testing on monitor (common)
Navigate to Diagnostics in System Admin Menu
- Test point
- System Admin Menu display
- Expected
- Diagnostics button visible and available (not grayed out)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
AUDIO FAILED error appears
Flashing red and white box with AUDIO FAILED message displayed on screen
Ranked causes
- 1. Speaker malfunctioned or connection faulty (common)
Verify speaker connection
- Test point
- Speaker connector at main board
- Expected
- Speaker cable firmly seated
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto Set Alarm Limits button unavailable
Auto Set Alarm Limits function cannot be accessed or is greyed out
Ranked causes
- 1. Monitor not in NBP Interval or Continuous SpO2 mode (common)
Verify monitor measurement mode
- Test point
- Monitor operating mode display
- Expected
- Monitor in NBP Interval mode or Continuous SpO2 mode
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery LED test failure
Charging LED does not display expected yellow-green-yellow-green sequence during System Diagnostics LED test
Ranked causes
- 1. Battery charging circuit malfunction (common)
Battery LED test sequence
- Test point
- Charging LED indicator
- Expected
- Lights yellow 5s, lights green 5s, flashes yellow 5s, flashes green 5s
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery reconditioning required
Battery Max Error is 10% or greater, indicating degraded capacity
Ranked causes
- 1. Battery Max Error 10% or greater (occasional)
Measure battery Max Error
- Test point
- Battery system
- Expected
- Less than 10%
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery service required message displayed
Monitor displays Battery service required alert on screen
Ranked causes
- 1. Max Error value exceeds 8% or Full Capacity half of Design Capacity (common)
Check battery error and capacity values
- Test point
- Battery diagnostics in System Menu
- Expected
- Max Error < 8%, Full Capacity >= 50% of Design Capacity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Monitor internal repair verification patient safety
Performance verification testing after opening monitor or replacing internal parts
Ranked causes
- 1. Monitor opened or internal parts replaced (non-NBP) (common)
Power-on self test
- Test point
- monitor startup
- Expected
- Self test completes without errors
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP accuracy
Non-invasive blood pressure measurement accuracy verification using reference manometer and expansion chamber connected to NBP connector.
Ranked causes
- 1. NBP sensor drift or offset error requiring calibration (common)
Connect manometer pump and tubing to NBP connector
- Test point
- NBP connector
- Expected
- Connection established
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP module patient safety
Verify NBP (Non-invasive Blood Pressure) module calibration after replacement or maintenance
Ranked causes
- 1. NBP module or parts replaced (common)
Power-on self test
- Test point
- Monitor startup
- Expected
- Self test completes successfully
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP module verification patient safety
Performance verification and testing of NBP (non-invasive blood pressure) system after replacement or maintenance
Ranked causes
- 1. NBP module or parts replaced (common)
Power-on self test
- Test point
- monitor startup
- Expected
- Self test completes without errors
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: NBP pneumatic leakage
Pneumatic system and valve leakage test to verify pressure retention over time in the NBP module.
Ranked causes
- 1. Pneumatic tubing leak or valve malfunction (occasional)
Apply 280 mmHg pressure and record initial value (P1)
- Test point
- NBP Test window pressure display
- Expected
- 280 mmHg applied and stabilized
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Software upgrade verification patient safety
Verification testing after software upgrade to ensure system settings and functionality
Ranked causes
- 1. Software upgraded (occasional)
Power-on self test
- Test point
- monitor startup
- Expected
- Self test completes without errors
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: SpO2 performance verification patient safety
Oxygen saturation measurement accuracy requires periodic verification
Ranked causes
- 1. SpO2 measurement accuracy drift (occasional)
Perform SpO2 accuracy test
- Test point
- SpO2 measurement system
- Expected
- Within device specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tympanic temperature accuracy patient safety
Temperature measurement may drift out of specification
Ranked causes
- 1. Temperature measurement drift over time (common)
Verify predictive temperature accuracy is within specification
- Test point
- Temperature measurement system
- Expected
- Within device specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot initialize NBP patient safety
System cannot initialize non-invasive blood pressure module
Ranked causes
- 1. NBP module hardware fault or disconnection (unknown)
Verify NBP module connection
- Test point
- NBP module connector at main board
- Expected
- Connector fully seated, no visible damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charging LED test failure
Battery LED does not exhibit expected light sequence during LED test
Ranked causes
- 1. LED or battery charging circuit malfunction (unknown)
Run LED test from System Diagnostics menu
- Test point
- Charging LED indicator
- Expected
- Lights yellow 5 sec, green 5 sec, flashes yellow 5 sec, flashes green 5 sec; AC Power LED stays green
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clear patient data from monitor memory
Remove all stored patient information before returning monitor for service or maintenance; does not affect calibration or hardware configuration data
Ranked causes
- 1. Patient data retention in memory (common)
Open Maintenance options
- Test point
- System Diagnostics menu
- Expected
- Maintenance options accessible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Enable Demo Mode on vital signs monitor patient safety
Activate Demo Mode to demonstrate monitor without actual patient connection; demo simulates all patient parameters and generates alarms when thresholds exceeded
Ranked causes
- 1. Need to demonstrate monitor functionality without patient (common)
Open System Admin Menu
- Test point
- System Admin Menu access
- Expected
- System Admin Menu displays all configuration options
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse replacement required high voltage
AC power fuse in connector needs replacement due to blown or failed fuse
Ranked causes
- 1. Blown or failed AC power fuse (unknown)
Disconnect AC power cord
- Test point
- Rear of monitor
- Expected
- Power cord removed
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal parts replaced (non-NBP, non-SpO2) caution
Verify monitor function after replacement of internal components other than NBP or SpO2 modules
Ranked causes
- 1. Internal component replaced (occasional)
Power-on self test
- Test point
- Monitor startup and diagnostics
- Expected
- Self test completes successfully
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Key is collapsed or does not click when pressed
Front panel key is unresponsive or lacks tactile feedback
Ranked causes
- 1. Membrane switch assembly malfunctioned (occasional)
Inspect affected key physical condition
- Test point
- Key button surface and mechanical travel
- Expected
- Key has normal tactile feedback and click
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Loss of monitoring patient safety
Monitoring capability temporarily lost or not available
Ranked causes
- 1. Temporary system state or informational condition (unknown)
Memory error detected patient safety
System memory test failed during self-test or startup
Ranked causes
- 1. Memory module fault or unseated connection (unknown)
Note error code during startup
- Test point
- System startup display
- Expected
- No error code 1052 displayed
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display incorrect or absent
LCD not displaying correctly or shows no image after power-up
Ranked causes
- 1. Display initialization failure (unknown)
Verify color bar appears on startup
- Test point
- LCD display
- Expected
- Color bar visible for ~5 seconds
Monitor does not power up after power supply reconnection high voltage
Monitor fails to power on after reconnecting power supply connectors to main board
Ranked causes
- 1. Power supply connectors reversed during reconnection (occasional)
Verify power supply connector polarity
- Test point
- 6-pin and 2-pin connectors at main board
- Expected
- White 6-pin connector connected to power supply; black connector connected to main board
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor freezes on opening screen or color bar
Monitor starts bootup but freezes on splash screen or test pattern
Ranked causes
- 1. Main board malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor opened for any reason patient safety
Perform required tests after opening monitor enclosure for repairs or maintenance
Ranked causes
- 1. Monitor enclosure opened for service (common)
Power-on self test
- Test point
- Monitor startup
- Expected
- Self test completes successfully
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call relay function not operating
Nurse call alarm output relay (phone jack connector) does not engage or release contacts (tip/ring/sleeve) during alarm condition or normal operation.
Ranked causes
- 1. Nurse call relay contacts not functioning (tip normally open, ring normally closed, or common not connected) (unknown)
Verify tip contact (normally open relay)
- Test point
- Nurse call connector tip
- Expected
- Open at rest, closed when alarm active
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Only one charging LED color displayed
Charging LED shows only one color instead of expected indication
Ranked causes
- 1. Membrane switch assembly malfunction (occasional)
Inspect charging LED display
- Test point
- Charging indicator LED
- Expected
- Both LED colors visible during charge cycle
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient leakage current excessive with mains voltage applied high voltage
Single fault condition test with mains voltage on applied part; leakage current from applied part to earth measured at S5 open and closed, all polarity combinations tested via S2 and S6 for each measurement input.
Ranked causes
- 1. Leakage current exceeds 50 µA at 250 V under single fault condition (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressing a key does not access expected function
Front panel key performs incorrect action (e.g., pressing Print initiates NBP)
Ranked causes
- 1. Membrane switch assembly malfunctioned (occasional)
Test all front panel keys for correct function
- Test point
- Each key and expected function
- Expected
- Each key triggers correct action
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error code 1052: Memory error
Self test reports system error code 1052 indicating internal memory failure
Ranked causes
- 1. Main board memory failure (common)
Shutdown and restart
- Test point
- Power supply and system memory
- Expected
- Monitor boots without 1052 error
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
System error code 1053: Loss of monitoring patient safety
System reports error code 1053 indicating loss of monitoring capability
Ranked causes
- 1. Informational message; normal operation mode change (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test: Nurse call relay verification patient safety
Nurse call relay function requires periodic verification to ensure alarm notification
Ranked causes
- 1. Nurse call relay degradation over time (rare)
Verify nurse call relay operates correctly
- Test point
- Nurse call relay output
- Expected
- Relay activates and deactivates as commanded
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Tympanic temperature reading unusually high
- Tympanic temperature reading unusually low
- Unit does not turn on with either AC or battery power
- Upgrade monitor software
- Verify system after software upgrade
5 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1025 | System Error (1025) | Could not load ResourceDLL.dll | error |
1026 | System Error (1026) | Could not load data from flash | error |
1027 | System Error (1027) | Could not start the front end | error |
1028 | System Error (1028) | Could not stop the front end | error |
1029 | System Error (1029) | Could not allocate enough memory | error |
1030 | System Error (1030) | Error when system shut down front end | error |
1031 | System Error (1031) | Monitoring and watchdog thread could not initialize | error |
1032 | System Error | Could not open audio files | error |
1033 | System Error | Unspecified error | error |
1034 | System Error | Failed to load product info | error |
1035 | System Error | Cannot initialize the front-end manager | error |
1036 | System Error | Failed to start work thread | error |
1037 | System Error | Cannot start polling thread | error |
1039 | System Error | Cannot get front-end version | error |
1042 | System Error | Failed to start up network driver | error |
1043 | System Error | Work thread timed out | error |
1044 | System Error | Front end software error | error |
1045 | System Error | Printing software error | error |
1047 | System Error | Monitoring software error | error |
1048 | System Error | Exception detected in alarming software | error |
1050 | System Error | Failed to stop work thread | error |
1051 | Cannot initialize NBP module | NBP module initialization fails during startup or self test | error |
1052 | Memory error | Memory test fails during startup or self test execution | error |
1053 | Loss of monitoring | Monitoring capability temporarily lost | info |
257 | System Error - battery charger power failure | Battery charger unable to deliver power | error |
261 | System Error - battery overcharged | Battery charging circuit detects overcharge condition | error |
264 | System Error - battery over temperature | Battery temperature sensor detects excessive temperature | warning |
817 | FPGA FIFO Overrun | FPGA first-in-first-out buffer overflow during NBP operation | info |
818 (101) | NBP Equipment Malfunction - task overrun | NBP task scheduler unable to complete tasks in time | warning |
818 (102) | NBP Equipment Malfunction - interrupt stack overwritten | Interrupt handler stack buffer overflow | error |
818 (103) | NBP Equipment Malfunction - user stack overwritten | User program stack buffer overflow | error |
818 (104) | NBP Equipment Malfunction - data validity check failed | Data integrity verification fails during operation | error |
818 (106) | NBP Equipment Malfunction - unable to write init flag to non-volatile memory | Non-volatile memory write failure during initialization | error |
818 (107) | NBP Equipment Malfunction - unable to verify init flag data | Non-volatile memory verification fails after write | error |
818 (109) | NBP Equipment Malfunction - unable to write NBP configuration to non-volatile memory | Configuration data write failure | error |
818 (110) | NBP Equipment Malfunction - unable to write NBP configuration to non-volatile memory | Configuration data write failure | error |
818 (111) | NBP Equipment Malfunction - calibration data checksum error | Calibration data checksum verification fails | error |
818 (112) | NBP Equipment Malfunction - unable to write calibration data to non-volatile memory | Calibration data write failure | error |
818 (113) | NBP Equipment Malfunction - unable to verify calibration data | Calibration data verification fails after write | error |
818 (114) | NBP Equipment Malfunction - error log checksum error | Error log data checksum verification fails | error |
818 (115) | NBP Equipment Malfunction - unable to write error log to non-volatile memory | Error log data write failure | error |
818 (116) | NBP Equipment Malfunction - unable to verify error log data | Error log verification fails after write | error |
818 (117) | NBP Equipment Malfunction - inflation cycle count and runtime data checksum error | Cycle count data checksum verification fails | error |
818 (118) | NBP Equipment Malfunction - unable to write inflation cycle count and runtime data | Cycle count data write failure | error |
818 (119) | NBP Equipment Malfunction - unable to verify inflation cycle count and runtime data | Cycle count data verification fails after write | error |
818 (120) | NBP Equipment Malfunction - message sending conflict | New message queued before previous message completes transmission | warning |
818 (121) | NBP Equipment Malfunction - pressure sensor 2 data unavailable | Second pressure sensor fails to return data | error |
818 (122) | NBP Equipment Malfunction - 5-volt supply voltage out of range | Voltage regulation circuit detects improper 5V supply | error |
818 (123) | NBP Equipment Malfunction - 5-volt regulator error | 5-volt voltage regulator detects internal failure | error |
818 (124) | NBP Equipment Malfunction - 12-volt supply voltage out of range | Voltage regulation circuit detects improper 12V supply | error |
818 (128) | NBP Equipment Malfunction - pressure sensor 1 offset out of range | Pressure sensor 1 offset calibration fails validation | error |
818 (129) | NBP Equipment Malfunction - pressure sensor 2 offset out of range | Pressure sensor 2 offset calibration fails validation | error |
818 (130) | NBP Equipment Malfunction - pressure sensor 2 test signal response out of range | Pressure sensor 2 self-test fails validation | error |
818 (131) | NBP Equipment Malfunction - pressure channel difference too large | Difference between two pressure channels exceeds threshold | error |
818 (132) | NBP Equipment Malfunction - valve activated too long | Valve activation timer exceeds maximum allowed duration | error |
818 (133) | NBP Equipment Malfunction - calibration data CRC mismatch | Calibration data CRC verification fails | error |
818 (134) | NBP Equipment Malfunction - error log data CRC mismatch | Error log data CRC verification fails | error |
818 (135) | NBP Equipment Malfunction - inflation cycle count data CRC mismatch | Cycle count data CRC verification fails | error |
818 (136) | NBP Equipment Malfunction - pneumatic safety switch failure | Pneumatic safety switch detects malfunction | error |
818 (137) | NBP Equipment Malfunction - valve 2 defective or blocked | Valve 2 unable to operate or blocked | error |
818 (141) | NBP Equipment Malfunction - inflation time exceeded for selected patient type | NBP inflation cycle duration exceeds maximum for patient type | warning |
818 (15) | NBP Equipment Malfunction - internal firmware error | NBP detects unrecoverable system errors in firmware | critical |
818 (199) | NBP equipment malfunction | No data from NBP block | error |
818 (2) | NBP Equipment Malfunction - A/D converter or pressure offset failure | NBP power-on self-test fails | error |
818 (25) | NBP Equipment Malfunction - no data received | NBP hardware unable to receive data from pressure sensors | error |
819 | NBP equipment malfunction - not calibrated | NBP module not calibrated | warning |
819 (127) | NBP equipment malfunction | NBP module not calibrated | error |
820 (138) | NBP hose blocked | Problem in pneumatic system | warning |
821 (139) | NBP air leak | Air leak or valve defect in NBP system | warning |
833 (1) | SpO2 equipment malfunction | SpO2 ROM checksum error | error |
833 (10) | SpO2 equipment malfunction | SpO2 host communication error | error |
833 (2) | SpO2 equipment malfunction | SpO2 RAM test error | error |
833 (20) | SpO2 equipment malfunction | SpO2 hardware defect | error |
833 (4) | SpO2 equipment malfunction | SpO2 analog signal path self test failed | error |
833 (40) | SpO2 equipment malfunction | SpO2 non-volatile memory corrupted | error |
833 (6) | SpO2 equipment malfunction | SpO2 self test timeout | warning |
833 (8) | SpO2 equipment malfunction | SpO2 internal DSP communication error | error |
834 | SpO2 sensor malfunction | Malfunction of SpO2 sensor or sensor cable | warning |
835 | FPGA FIFO overrun | SpO2 software error | info |
836 | SpO2 communication error | SpO2 communication error | info |
837 | SpO2 error, front end reset | No data from SpO2 | error |
865 | FPGA FIFO overrun | Temperature software error | info |
866 (21) | Temperature probe error | Probe operating above 43.3°C (112°F) | warning |
866 (22) | Temperature probe error | Excessive heater energy | warning |
866 (32) | Temperature probe error | Thermistor pulled away or heater broken | warning |
866 (33) | Temperature probe error | Probe not responsive | warning |
866 (34) | Temperature probe error | Probe not characterized or calibrated | warning |
866 (45) | Temperature probe error | Measurement below allowable temperature values | warning |
866 (46) | Temperature probe error | Measurement above allowable temperature values | warning |
866 (52) | Temperature probe error | Probe not characterized or calibrated | warning |
866 (63) | Temperature probe error | Probe well missing or not installed properly | warning |
866 (70) | Temperature probe error | Error reading probe EEPROM | warning |
866 (75) | Temperature probe error | Error reading probe EEPROM or not factory tested | warning |
866 (78) | Temperature module malfunction | Probe non-volatile error detection mechanism triggered | error |
866 (79) | Temperature module malfunction | Module non-volatile error log memory error detected | error |
866 (80) | Temperature probe error | Software turning off heater but heater feedback still on | error |
866 (81) | Temperature probe error | Software turning on heater but heater feedback still off | error |
867 (101) | Temp Module Malfunc (101) | Tympanic temperature module checksum error | error |
867 (102) | Temp Module Malfunc (102) | Tympanic temperature module calibration error | error |
867 (103) | Temp Module Malfunc (103) | Ambient temperature below minimum operating limit | warning |
867 (104) | Temp Module Malfunc (104) | Ambient temperature above maximum operating limit | warning |
867 (24) | Temperature module malfunction | Ambient temperature too high | warning |
867 (25) | Temperature module malfunction | Ambient temperature too low | warning |
867 (27) | Temperature module malfunction | Battery or power supply voltage exceeds maximum | error |
867 (28) | Temperature module malfunction | Battery or power supply voltage below minimum | error |
867 (47) | Temperature module malfunction | Internal calibration resistor RCAL damaged | error |
867 (48) | Temperature module malfunction | Internal calibration resistor RCAL damaged | error |
867 (49) | Temperature module malfunction | Internal circuit validation resistor PTB damaged | error |
867 (50) | Temperature module malfunction | Internal circuit validation resistor PTB damaged | error |
867 (51) | Temperature module malfunction | A/D measurement timed out | error |
867 (59) | Temperature module malfunction | Battery or power supply voltage below maximum | error |
867 (60) | Temperature module malfunction | Battery or power supply voltage exceeds minimum | error |
867 (61) | Temperature module malfunction | Reference voltage circuit under voltage or unstable | error |
867 (62) | Temperature module malfunction | Module not calibrated | error |
867 (65) | Temp Module Malfunc (65) | Error saving to module EEPROM | warning |
867 (66) | Temp Module Malfunc (66) | Non-volatile calibration memory error detected | warning |
867 (67) | Temp Module Malfunc (67) | Error reading/writing EEPROM | warning |
867 (68) | Temp Module Malfunc (68) | Error reading/writing EEPROM | warning |
867 (69) | Temp Module Malfunc (69) | Error reading/writing EEPROM | warning |
867 (74) | Temp Module Malfunc (74) | Internal error during EEPROM initialization | warning |
867 (82) | Temp Module Malfunc (82) | HTR_Q on, HTRC off, but voltage present | warning |
867 (83) | Temp Module Malfunc (83) | HTR_Q tri-stated with HTRC enabled and heater power | warning |
867 (84) | Temp Module Malfunc (84) | Q&C on but heater voltage insufficient | warning |
867 (85) | Temp Module Malfunc (85) | Heater fail-safe did not turn off when expected | error |
868 | Temp Error. Front end will reset. | No data from temperature sensor | error |
881 | Speaker Malfunc (881) | Speaker failed | warning |
882 | Wireless Malfunc | Wireless module malfunction | error |
883 | Touch Screen Malfunc | Touch screen malfunction | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.