Philips VS3: troubleshooting and error codes
The ServiceForge knowledge base holds 205 documented symptoms for the Philips VS3, with 446 ranked causes, 281 diagnostic checks with test points and expected values, and 286 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Monitor will not power on with AC caution
Monitor powers on with battery but not when connected to AC outlet
Ranked causes
- 1. AC power cord unplugged (common)
- 2. AC power cord damaged (common)
- 3. Power supply fuse blown (occasional)
- 4. Power supply cable malfunction (occasional)
- 5. Power supply output voltage failure (occasional)
- 6. AC connector malfunction (rare)
- 7. Monitor keypad worn (rare)
- 8. Main board failure (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on with AC power, works with battery
Monitor cannot be turned on using AC power connection but powers on normally when using battery power
Ranked causes
- 1. AC power cord unplugged or broken (common)
- 2. Fuse is blown (common)
- 3. Power supply cable has malfunctioned (occasional)
- 4. Power supply has malfunctioned (occasional)
- 5. AC power connector has malfunctioned (occasional)
- 6. Monitor keypad is worn (rare)
- 7. Main board has malfunctioned (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor will not power on with AC power (battery works) caution
AC power button does not turn on monitor; battery power operates normally
Ranked causes
- 1. AC power cord unplugged or broken (common)
- 2. Fuse is blown (common)
- 3. Power supply cable malfunction (occasional)
- 4. Power supply output failure (occasional)
- 5. AC power connector malfunction (occasional)
- 6. Monitor keypad worn (rare)
- 7. Main board malfunction (rare)
Verify AC power cord is plugged in
- Expected
- Power cord connected to wall outlet
- Tools
- Visual inspection
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurements are not displayed
Temperature readings do not appear on monitor screen
Ranked causes
- 1. Probe is disconnected from monitor (common)
- 2. Probe is misplaced (common)
- 3. Predictive temperature is over range (occasional)
- 4. Probe malfunctioned (occasional)
- 5. Front end board malfunction (rare)
- 6. Temperature module malfunctioned (rare)
- 7. Main board malfunctioned (rare)
Verify probe is connected and seated
- Test point
- Probe connector and probe well
- Expected
- Probe connected and fully seated
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on with AC power high voltage
Monitor turns on with battery power but not with AC power supply
Ranked causes
- 1. AC power cord unplugged or disconnected (common)
- 2. AC power cord broken (common)
- 3. Fuse blown (common)
- 4. Power supply cable malfunctioned (occasional)
- 5. Power supply output voltage incorrect (occasional)
- 6. AC power connector malfunctioned (occasional)
- 7. Main board malfunctioned (rare)
Verify AC power cord is connected to outlet
- Test point
- AC power outlet
- Expected
- Power cord plugged in
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery: Verify charge level and health status patient safety
Check battery charge status via LED indicator, battery status pane, and alarm messages to assess battery condition and charge level.
Ranked causes
- 1. Low battery charge (common)
- 2. Battery nearly discharged (common)
- 3. Battery module malfunction or temperature out of range (occasional)
- 4. Battery charger power failure (rare)
- 5. Battery overcharged (occasional)
- 6. Battery over temperature (occasional)
Observe LED charging indicator color
- Test point
- front panel charging LED
- Expected
- Green=full/nearly full; Flashing green=medium; Yellow=low; Flashing yellow=nearly discharged
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor screen is blank
Display remains dark with no visible content when power is on
Ranked causes
- 1. Monitor in power save mode (common)
- 2. Speaker/fan cables disconnected (common)
- 3. Bad backlight LCD connection (occasional)
- 4. Bad backlight inverter (occasional)
- 5. Bad LCD (occasional)
- 6. Main PCB failure (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor screen is blank despite power on caution
Monitor power is on but display shows nothing; no image on screen
Ranked causes
- 1. Monitor in power save mode (common)
- 2. Speaker or fan cables disconnected (common)
- 3. Bad LCD backlight connection (occasional)
- 4. Bad backlight inverter (occasional)
- 5. LCD malfunction (occasional)
- 6. Main board malfunction (rare)
Verify monitor is not in power save mode
- Expected
- Screen displays active image
- Tools
- Power button
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor won't power on with AC but works on battery
Power button does not activate when AC supply connected, but device powers on using battery
Ranked causes
- 1. AC power cord unplugged or broken (common)
- 2. Power supply fuse blown (common)
- 3. Power supply cable malfunction (occasional)
- 4. Power supply output failure (occasional)
- 5. AC connector malfunction (occasional)
- 6. Main PCB failure (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurements are not displayed
Non-invasive blood pressure readings do not appear on screen
Ranked causes
- 1. Wrong cuff size or incorrect placement (common)
- 2. Initial pressure setting is incorrect (common)
- 3. Air tube bent, twisted, or kinked with air leakage (common)
- 4. External problem affecting measurement (occasional)
- 5. NBP pump filter is blocked (occasional)
- 6. NBP module error (rare)
Verify correct cuff size is selected and applied
- Test point
- Cuff selection and application site
- Expected
- Correct cuff size for patient used
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No NBP readings displayed patient safety
Blood pressure measurement shows no numeric values or dashes
Ranked causes
- 1. Incorrect cuff size or placement (common)
- 2. Incorrect initial pressure setting (common)
- 3. Kinked tube or air leakage (common)
- 4. External patient or environmental issue (occasional)
- 5. NBP pump filter blocked (occasional)
- 6. NBP module error (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No temperature readings
Temperature measurement displays no value or shows out-of-range error
Ranked causes
- 1. Temperature probe lead disconnected (common)
- 2. Temperature reading out of range (occasional)
- 3. Temperature probe malfunction (occasional)
- 4. Front end malfunction (occasional)
- 5. Temperature module malfunction (occasional)
- 6. Main board malfunction (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No temperature readings displayed patient safety
Temperature measurement missing from display; probe not recording or calculating temperature
Ranked causes
- 1. Probe not connected (common)
- 2. Temperature reading out of range (common)
- 3. Probe malfunction (occasional)
- 4. Front end malfunction (occasional)
- 5. Temperature module malfunction (occasional)
- 6. Main board malfunction (rare)
Verify temperature probe is physically connected
- Expected
- Probe cable securely connected to monitor
- Tools
- Visual inspection
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade fails or is interrupted patient safety
Upgrade process halted unexpectedly, main board data lost, monitor non-functional after attempted software update
Ranked causes
- 1. USB flash drive removed during upgrade (occasional)
- 2. Monitor unplugged during upgrade (occasional)
- 3. Keys pressed during upgrade (occasional)
- 4. Software upgrade through USB hub (rare)
- 5. Invalid software image on USB drive (occasional)
- 6. Battery not fully charged (common)
Verify battery charge level
- Test point
- Battery indicator on monitor display
- Expected
- Fully charged
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner cannot read barcodes (laser and sound work)
Barcode scanner powers on and responds but fails to read bar codes; laser visible but no decode
Ranked causes
- 1. Scanner held at wrong distance from barcode (common)
- 2. System does not recognize barcode format (occasional)
- 3. Unrecognized format but scanner functional (occasional)
- 4. System does not recognize scanner (rare)
- 5. Scanner mechanical failure (rare)
Verify scanner distance from barcode
- Expected
- Scanner held 6 to 12 inches from barcode
- Tools
- Ruler, visual alignment
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor power and basic operation troubleshooting
Initial diagnostic checks before detailed troubleshooting: power switch status, battery charge, AC connection, display function, and LED indicators
Ranked causes
- 1. Power switch not enabled (common)
- 2. Battery not adequately charged (common)
- 3. AC power not connected or outlet non-functional (common)
- 4. Display malfunction (occasional)
- 5. LED indicators not illuminated as expected (occasional)
Verify power switch is turned on
- Test point
- Power switch on front panel
- Expected
- Power switch in ON position
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP cuff does not inflate
Non-invasive blood pressure cuff fails to inflate during measurement
Ranked causes
- 1. Air tube or cuff bent or twisted (common)
- 2. Air leak in tube or cuff (common)
- 3. NBP module malfunctioned (occasional)
- 4. NBP pump filter is blocked (occasional)
- 5. Main board malfunctioned (rare)
Inspect air tube and cuff for kinks or bends
- Test point
- Air tube and cuff along full length
- Expected
- Tube and cuff straight with no twists or bends
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 readings patient safety
Pulse oximetry measurement fails to return values
Ranked causes
- 1. SpO2 module in learning phase (common)
- 2. SpO2 sensor failure (common)
- 3. SpO2 board connection loose (occasional)
- 4. SpO2 board failure (occasional)
- 5. Main board failure (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call does not signal alarms patient safety
Nurse call function fails to trigger alerts for alarm conditions
Ranked causes
- 1. Minimum Nurse Call Alarm Priority too high (common)
- 2. Alarms silenced (common)
- 3. Nurse call connector loose or broken (occasional)
- 4. Communications board failure (occasional)
- 5. Main board failure (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call function does not signal alarm conditions patient safety
Nurse call system does not alert nursing staff to alarm conditions
Ranked causes
- 1. Minimum Nurse Call Alarm Priority set too high (common)
- 2. Alarms have been silenced (common)
- 3. Phono connector is loose or broken (occasional)
- 4. Communications board malfunctioned (rare)
- 5. Main board malfunctioned (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not print
Recorder fails to produce paper output
Ranked causes
- 1. Recorder is out of paper (common)
- 2. Recorder door is open (common)
- 3. System does not recognize the recorder (occasional)
- 4. Recorder module malfunctioned (occasional)
- 5. Main board malfunctioned (rare)
Check recorder paper status
- Test point
- Paper compartment
- Expected
- Paper loaded and available
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade fails or monitor becomes unresponsive during upgrade patient safety
Monitor stops responding, upgrade process interrupts, or main board data is lost during software update
Ranked causes
- 1. Monitor unplugged during upgrade (common)
- 2. USB flash drive removed during upgrade (common)
- 3. Keys pressed during upgrade (common)
- 4. USB drive not fully inserted or clamp blocking insertion (occasional)
- 5. Upgrade interrupted, main board data lost (rare)
Verify AC power connection
- Test point
- Monitor AC power port
- Expected
- Monitor connected to AC mains and powered on
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurements are not displayed
Oxygen saturation readings do not appear on monitor screen
Ranked causes
- 1. SpO2 module in learning phase (common)
- 2. SpO2 sensor malfunctioned (occasional)
- 3. SpO2 board connection to front end board faulty (occasional)
- 4. SpO2 board malfunctioned (rare)
- 5. Main board malfunctioned (rare)
Monitor SpO2 module learning phase
- Test point
- SpO2 module status
- Expected
- Learning phase complete
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not sound patient safety
Audio alarms are silent or not producing sound
Ranked causes
- 1. Audio Pause or Audio Off mode enabled (common)
- 2. Connection to speaker is faulty (occasional)
- 3. Speaker malfunctioned (occasional)
- 4. Main board malfunctioned (rare)
Verify audio alarms are enabled
- Test point
- Audio settings menu
- Expected
- Audio mode is not Pause or Off
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner cannot read barcodes
Laser and sound function but barcode is not decoded
Ranked causes
- 1. Incorrect scanner distance from barcode (common)
- 2. Unrecognized barcode format (common)
- 3. Scanner needs reset (occasional)
- 5. Scanner mechanical problem (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner does not function (no laser, no sound)
Barcode scanner completely non-functional; no laser beam visible and no response sound
Ranked causes
- 1. Scanner not connected to USB port (common)
- 2. Scanner window not clean or blocked (common)
- 3. System does not recognize scanner (occasional)
- 4. Scanner mechanical failure (occasional)
Verify scanner USB connection
- Expected
- Scanner cable securely connected to USB port
- Tools
- Visual inspection
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner has no laser or sound
Scanner produces no audible tone and emits no laser beam
Ranked causes
- 1. Scanner not plugged into USB port (common)
- 2. Scanner window blocked or dirty (common)
- 3. System does not recognize scanner (occasional)
- 4. Scanner mechanical problem (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displays random or distorted graphics
Display shows corrupted visual content with white background instead of normal image
Ranked causes
- 1. LCD cable not attached correctly (common)
- 2. Missing or improper ferrite clamp on LCD signal cable (common)
- 3. LCD failure (occasional)
- 4. Main board malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on with battery power, works with AC
Monitor cannot be turned on using battery power but powers on normally when using AC power
Ranked causes
- 1. Battery has lost its charge (common)
- 2. Battery or connector is loose (common)
- 3. Battery does not have connection to main board (occasional)
- 4. Main board has malfunctioned (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not respond to navigation wheel or front panel buttons caution
Buttons and navigation wheel do not register input; monitor does not respond to user commands
Ranked causes
- 1. Bad connection to main board (common)
- 2. Keypad or navigation wheel broken (occasional)
- 3. Main board malfunction (rare)
- 4. Membrane switch assembly malfunction (occasional)
Check all connectors to main board
- Expected
- All connectors fully seated
- Tools
- Visual inspection
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not respond to navigation wheel or front panel keys
Navigation wheel and front panel keys do not provide input to the monitor
Ranked causes
- 1. Main board has faulty connection (occasional)
- 2. Front panel keypad or navigation wheel broken (common)
- 3. Main board malfunctioned (rare)
- 4. Membrane switch assembly malfunctioned (rare)
Check main board connector seating
- Test point
- Main board connectors
- Expected
- All connectors fully seated
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor unresponsive to controls
Navigation wheel or front panel buttons do not control device
Ranked causes
- 1. Bad connection to main board (common)
- 2. Broken keypad or navigation wheel (common)
- 3. Main PCB failure (occasional)
- 4. Membrane switch assembly malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor will not power on with battery caution
Monitor powers on with AC power but not on battery
Ranked causes
- 1. Battery discharged (common)
- 2. Battery or connector loose (common)
- 3. Battery board connection failure (occasional)
- 4. Main board failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor will not power on with battery power (AC works) caution
Battery power button does not turn on monitor; AC power operates normally
Ranked causes
- 1. Battery has lost charge (common)
- 2. Battery or connector loose (common)
- 3. Battery not connected to main board (occasional)
- 4. Main board malfunction (rare)
Check battery charge level
- Expected
- Battery fully charged
- Tools
- Battery charger
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor won't power on with battery but works on AC
Power button does not activate using battery, but device powers on when connected to AC supply
Ranked causes
- 1. Battery discharged (common)
- 2. Battery or connector loose (common)
- 3. Battery board disconnection (occasional)
- 4. Main PCB failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Navigation wheel or keys not responding caution
Monitor does not respond to navigation wheel or front panel key presses
Ranked causes
- 1. Main board connector loose (common)
- 2. Keypad or navigation wheel broken (common)
- 3. Membrane switch malfunction (common)
- 4. Main board failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurements are unreliable patient safety
Blood pressure readings are inconsistent or inaccurate
Ranked causes
- 1. Wrong cuff size or incorrect placement (common)
- 2. Initial pressure setting is incorrect (common)
- 3. External problem affecting measurement (occasional)
- 4. NBP module needs calibration (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP readings are unreliable or inconsistent patient safety
Blood pressure readings vary significantly or are outside expected range; inconsistent measurement results
Ranked causes
- 1. Wrong cuff size or incorrect placement (common)
- 2. Incorrect initial pressure setting (common)
- 3. External factor or patient movement (occasional)
- 4. Calibration drift (occasional)
Verify cuff size is appropriate and correctly placed
- Expected
- Proper cuff size applied to correct limb location
- Tools
- Visual inspection
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No NBP readings obtained patient safety
Blood pressure measurement fails to complete; no systolic, diastolic, or MAP reading displayed
Ranked causes
- 1. Wrong cuff size or incorrect placement (common)
- 2. Incorrect initial pressure setting (common)
- 3. Kinked tube or air leakage (occasional)
- 4. External factor or patient movement (occasional)
Verify cuff size is appropriate and correctly placed
- Expected
- Proper cuff size applied to correct limb location
- Tools
- Visual inspection
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 readings displayed patient safety
Pulse oximetry measurement missing from display; SpO2 sensor not recording oxygen saturation
Ranked causes
- 1. SpO2 module in learning phase (common)
- 2. SpO2 sensor malfunction (occasional)
- 3. SpO2 board malfunction (occasional)
- 4. Main board malfunction (rare)
Check if SpO2 module is in learning phase
- Expected
- Learning phase complete; SpO2 reading available
- Tools
- Monitor display
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not print
Recorder fails to produce output on paper
Ranked causes
- 1. Out of paper (common)
- 2. Recorder door open (common)
- 4. Recorder module failure (occasional)
- 5. Main board failure (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self test fails patient safety
Monitor resident self test reports FAILED result for one or more test items (Memory, SpO2, NBP, Temperature)
Ranked causes
- 1. Memory test failure (rare)
- 2. SpO2 module not detected or error (occasional)
- 3. NBP module not detected or error (occasional)
- 4. Temperature module not detected or error (occasional)
Press Self Test button to initiate monitor resident self test
- Test point
- Self Test button
- Expected
- Self Test window appears and displays test status
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurements are unreliable patient safety
Temperature readings are inconsistent or inaccurate
Ranked causes
- 1. Probe malfunctioned (common)
- 2. Temperature module not calibrated (common)
- 3. Front end board malfunctioned (occasional)
- 4. Temperature module malfunctioned (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature readings are unreliable or inconsistent patient safety
Temperature values vary significantly or outside expected range; erratic temperature measurement
Ranked causes
- 1. Probe malfunction (common)
- 2. Temperature module accuracy drift (occasional)
- 3. Front end malfunction (occasional)
- 4. Temperature module malfunction (occasional)
Perform temperature module accuracy verification
- Expected
- Module readings match reference thermometer within tolerance
- Tools
- Reference thermometer, Predictive Temperature Test
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on with battery power
Monitor turns on with AC power but not with battery power supply
Ranked causes
- 1. Battery not charged (common)
- 2. Battery or connector loose (common)
- 3. Battery board not connected to main board (occasional)
- 4. Main board malfunctioned (rare)
Verify battery charge status
- Test point
- Battery charging port
- Expected
- Battery charged
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unreliable NBP readings patient safety
Blood pressure measurements are inconsistent or fluctuate significantly
Ranked causes
- 1. Incorrect cuff size or placement (common)
- 2. Incorrect initial pressure setting (common)
- 3. External patient or environmental issue (occasional)
- 4. NBP module calibration out of spec (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB hub does not work
External USB hub connected to monitor is not functioning
Ranked causes
- 1. USB hub is disconnected from monitor (common)
- 2. Hub USB ports blocked by dust or dirt (common)
- 3. USB hub is broken (occasional)
- 4. LAN (I/O Communications) board broken or not properly connected (rare)
Verify USB plugs are firmly connected
- Test point
- USB ports on monitor and hub
- Expected
- All plugs seated firmly in ports
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarms not functioning patient safety
Visual and audio alarms do not trigger during NBP or SpO2 monitoring
Ranked causes
- 1. NBP alarm disabled or audio muted (common)
- 2. NBP module malfunction or improper connection (occasional)
- 3. SpO2 sensor malfunction or improper connection (occasional)
Verify alarms are enabled on monitor
- Test point
- Audio and alarm mode
- Expected
- Audio Pause or Audio Off mode not active
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not sound patient safety
Monitor does not produce audible alarm sounds
Ranked causes
- 1. Audio mode disabled (common)
- 2. Speaker connection loose (common)
- 4. Main board failure (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AUDIO FAILED message displayed with flashing red and white box patient safety
Flashing red and white box with AUDIO FAILED message appears on screen
Ranked causes
- 1. Speaker malfunctioned (common)
- 2. Connection to speaker is faulty (common)
- 3. Speaker is faulty (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner does not function
Scanner shows no laser, no sound; cannot read barcodes or function at all.
Ranked causes
- 1. Scanner not plugged into USB port (common)
- 2. System does not recognize scanner (occasional)
- 3. Scanner has mechanical problem (occasional)
Verify scanner physical connection to USB port
- Test point
- Monitor USB port
- Expected
- Scanner firmly connected
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display shows distorted graphics
Monitor displays random or distorted graphics with white background
Ranked causes
- 1. LCD signal cable not properly attached (common)
- 2. LCD failure (occasional)
- 3. Main board failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displays random or distorted graphics with white background caution
Screen shows garbled graphics, corrupted text, or white background with colored noise
Ranked causes
- 1. LCD signal cable not attached correctly (common)
- 2. LCD malfunction (occasional)
- 3. Main board malfunction (rare)
Reseat LCD signal cable
- Expected
- Graphics display normally after reseating
- Tools
- Visual inspection
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displays random or distorted graphics with white background
Screen shows corrupted display with random graphics and white background
Ranked causes
- 1. LCD signal cable not attached correctly (common)
- 2. LCD malfunctioned (occasional)
- 3. Main board malfunctioned (rare)
Inspect and reseat LCD signal cable
- Test point
- LCD signal cable connectors
- Expected
- Cable fully seated and not damaged
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on
No power indication; display inactive; LEDs not lit
Ranked causes
- 1. Power switch off (common)
- 2. Battery inadequately charged (common)
- 3. AC power cord disconnected or outlet non-functional (common)
Verify power switch position
- Test point
- Main power switch
- Expected
- Switch in ON position
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not respond to controls
Navigation wheel and front panel buttons do not control menu or operation
Ranked causes
- 1. Bad connection to main board (common)
- 2. Keypad or navigation wheel broken (occasional)
- 3. Main board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor screen is blank when powered on
Power is on but monitor screen displays nothing
Ranked causes
- 1. Monitor in Standby mode (common)
- 2. Backlight LCD connection faulty (occasional)
- 3. Main board malfunctioned (rare)
Verify Standby mode is not active
- Test point
- Power button status
- Expected
- Monitor in normal operating mode
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP cuff does not inflate caution
Blood pressure cuff fails to inflate when measurement initiated; no pressure buildup
Ranked causes
- 1. Tube or cuff is kinked (common)
- 3. NBP module malfunction (occasional)
- 5. Main board malfunction (rare)
Inspect tube and cuff for kinks
- Expected
- Tube and cuff straight and flexible
- Tools
- Visual inspection
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Random or distorted graphics displayed
Monitor shows random patterns or distorted image with white background
Ranked causes
- 1. LCD cable not properly attached (common)
- 2. Bad LCD (occasional)
- 3. Main PCB failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder paper is jamming
Recorder paper gets stuck or jams during operation
Ranked causes
- 1. Paper not installed correctly (common)
- 2. Wrong type of paper is installed (common)
- 3. Recorder has mechanical problem (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Software upgrade fails or monitor cannot boot
- Software upgrade interrupted or fails
- Unit does not power on
- Unreliable temperature readings
- USB hub does not work
- Audio failed alarm displayed
- Barcode scanner: Verify scanner accuracy and patient data input
- Battery service required message displayed
- Monitor boot sequence does not complete
- Monitor cannot find valid software image on USB flash drive
- Monitor does not power on
- Monitor has no power
- NBP alarm not functioning
- NBP initial pressure too high
- NBP pneumatic leakage exceeds limit
- Pneumatic leakage test failure
- Recorder malfunction
- Recorder paper is jamming
- Recorder test pattern abnormal
- Recorder test pattern abnormal
- Recorder test pattern indicates print head or motor fault
- SpO2 measurement out of range
- SpO2 measurements are unreliable
- SpO2 reading outside normal range
- SpO2 readings are unreliable or inconsistent
- SpO2 readings unreliable
- Access Maintenance options in System Diagnostics
- Access System Diagnostics menu
- Alarms testing
- Audio output not heard or incorrect
- Audio test not producing tones
- Auto Set Alarm Limits button unavailable
- Auto Set Alarm Limits button unavailable
- Barcode scanner cannot read all patient identification methods
- Barcode scanner programming reset
- Barcode scanner reads only primary ID method
- Battery capacity or type anomaly
- Battery charging LED displays only one color
- Battery charging LED malfunction
- Battery charging LED shows only one color
- Battery LED not behaving as expected during test
- Battery LED not functioning
- Battery LED not showing expected color sequence
- Battery LED test failure
- Button collapsed or not clicking
- Button does not access correct function
- Button is collapsed or does not click when pressed
- Button press accesses wrong function
- Button Test: rotary wheel not responding
- Button Test: rotary wheel not responding or drawing red dots
- Button Test: rotary wheel small movements not detected
- Button Test: system button not responding
- Button Test: system icon button not responding
- Calibration: Battery reconditioning
- Calibration: Internal parts replaced
- Calibration: Monitor opened for maintenance
- Calibration: NBP accuracy
- Calibration: NBP module
- Calibration: NBP overpressure valve
- Calibration: NBP pneumatic leakage
- Calibration: Performance verification after internal parts replacement
- Calibration: Performance verification after monitor installation
- Calibration: Performance verification after monitor opened
- Calibration: Performance verification after NBP module replacement
- Calibration: Performance verification after software upgrade
- Calibration: Reset serial number after main PCB replacement
- Calibration: Safety tests
- Calibration: Software upgrade
- Calibration: SpO2 accuracy
- Calibration: Temperature accuracy
- Calibration: Temperature measurement verification
- Cannot initialize NBP
- Display test pattern not showing or showing incorrectly
- Display test patterns not appearing
- Enclosure leakage current exceeds safety limit
- Failed to stop work thread
- Front panel LEDs not lit as expected
- Hardware incompatibility or system corruption after software downgrade
- Internal parts replaced (non-NBP, non-SpO2)
- Key collapsed or unresponsive to press
- Key is collapsed or does not click when pressed
- Key press activates wrong function
- LED test fails or unexpected LED behavior
- Loss of monitoring
- Memory error
- Monitor display inactive or malfunctioning
- Monitor display incorrect or absent
- Monitor display is blank or incorrect
- Monitor does not power on with either AC or battery power
- Monitor fails to boot
- Monitor freezes during startup
- Monitor hangs during startup
- Monitor in suspended mode does not respond to power button
- Monitor opened for service
- Monitor will not power on at all
- Monitor will not power on with AC or battery power
- Monitor will not power up after reassembly
- Navigation wheel or keys not responding
- NBP accuracy exceeds tolerance
- NBP cycle count requires reset after module replacement
- NBP equipment not calibrated
- NBP measurement exceeds tolerance
- NBP overpressure valve failure
- NBP overpressure valve not opening
- No startup tone during boot
- Nurse call relay does not operate
- Nurse call relay: Verify relay contact operation
- Only one charging LED color displayed
- Opening screen hangs
- Opening screen hangs or does not complete
- Opening screen or color bar hangs on startup
- Power connector fuse replacement needed
- Power-on self test
- Predictive Temp module test fails during self-test
- Predictive temperature measurement error
- Pressing a key does not access expected function
- Pump filter maintenance or replacement needed
- Recorder prints incorrectly
- Reset tracked component usage parameters
- Reset tracked parameter
- Rotary wheel or button not responding
- Safety test: Enclosure leakage current
- Safety test: Ground integrity
- Safety test: Patient leakage current
- Self test fails for module
- Self test fails on NBP module
- Self test in progress
- Software configuration settings lost after upgrade
- Software downgrade attempted
- Software upgrade completed
- Speaker and fan cable connections failing
- Speaker not producing audio
- System Admin Menu locks after use
- System exception in alarming software
- System settings and patient records lost after software upgrade
- Temperature measurement out of range
- Tracked parameters need resetting after part replacement
- Tympanic temperature reading is unusually high
- Tympanic temperature reading is unusually low
- Unable to access System Admin Menu
- Unit does not power on with AC or battery power
- Unreliable SpO2 readings
- USB flash drive does not fit in USB port
- Visual damage or obstruction present
- Visual inspection
145 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1025 | System Error (1025) | Could not load ResourceDLL.dll | critical |
1025 | System Error (1025) | Could not load ResourceDLL.dll | error |
1026 | System Error (1026) | Could not load data from flash | error |
1026 | System Error (1026) | Could not load data from flash | critical |
1027 | System Error (1027) | Could not start front-end | critical |
1027 | System Error (1027) | Could not start the front end | error |
1028 | System Error (1028) | Could not stop the front end | error |
1028 | System Error (1028) | Could not stop front-end | critical |
1029 | System Error (1029) | Could not allocate enough memory | critical |
1029 | System Error (1029) | Could not allocate enough memory | error |
1030 | System Error (1030) | An error occurred when the system shut down front end devices | error |
1030 | System Error (1030) | Error during system shutdown of front end devices | critical |
1031 | System Error (1031) | Monitoring and watchdog thread could not initialize | critical |
1031 | System Error (1031) | The monitoring and watchdog thread could not initialize | error |
1032 | System Error (1032) | Could not open audio files | error |
1032 | System Error (1032) | Could not open audio files | critical |
1033 | System Error (1033) | Unspecified error | critical |
1033 | System Error (1033) | Unspecified error | error |
1034 | System Error (1034) | Failed to load product info | error |
1034 | System Error (1034) | Failed to load product info | critical |
1035 | System Error (1035) | Cannot initialize front-end manager | critical |
1035 | System Error (1035) | Cannot initialize the front-end manager | error |
1036 | System Error (1036) | Failed to start work thread | error |
1036 | System Error (1036) | Failed to start work thread | critical |
1037 | System Error (1037) | Cannot start polling thread | critical |
1037 | System Error (1037) | Cannot start polling thread | error |
1039 | System Error (1039) | Cannot get front-end version | error |
1039 | System Error (1039) | Cannot get front-end version | critical |
1042 | System Error (1042) | Failed to start up network driver | critical |
1042 | System Error (1042) | Failed to start up network driver | error |
1043 | System Error (1043) | Work thread timed out | error |
1043 | System Error (1043) | Work thread timed out | critical |
1043 (1) | System Error (1) | Front end software problem | critical |
1043 (2) | System Error (2) | Alarm software problem | critical |
1044 | System Error (1044) | Front end software error | critical |
1044 | System Error (1044) | Front end software error | error |
1045 | System Error (1045) | Printing software error | error |
1045 | System Error (1045) | Printing software error | critical |
1047 | System Error (1047) | Monitoring software error | critical |
1047 | System Error (1047) | Monitoring software error | error |
1048 | System Error (1048) | An exception was detected in the alarming software | error |
1048 | System error - exception in alarming software | Exception detected in alarming software | error |
1048 | System Error (1048) | Exception detected in alarming software | critical |
1050 | System Error (1050) | Failed to stop work thread | critical |
1050 | System error - failed to stop work thread | Failed to stop work thread | error |
1050 | System Error (1050) | Failed to stop work thread | error |
1051 | System Error (1051) | Cannot initialize NBP | error |
1051 | System Error (1051) | Cannot initialize NBP | critical |
1052 | System Error (1052) | Memory error | critical |
1052 | System Error | Memory error detected during operation | critical |
1053 | System Error | Loss of monitoring signal detected | info |
1053 | System error - loss of monitoring | Loss of monitoring capability | info |
257 | System Error | Battery charger power failure | error |
257 | System Error | Battery charger power failure | error |
257 | System Error - Battery charger power failure | Battery charger power failure | error |
261 | Battery is overcharged | charging circuit fails to regulate battery voltage; battery remains connected to charger beyond full charge state | warning |
261 | System Error - Battery is overcharged | Battery is overcharged | error |
261 | System Error | Battery is overcharged | error |
264 | System Error | Battery is over temperature | error |
264 | System Error | Battery over temperature condition detected | error |
264 | System Error - Battery over temperature | Battery over temperature | warning |
264 temp | System Error | Battery temperature exceeds safe operating limit | error |
264 temp_value | Battery is over temperature | battery temperature exceeds safe operating range during charging or discharge | warning |
817 | FPGA FIFO Overrun | NBP software problem detected during operation | error |
817 | FPGA FIFO Overrun | NBP software error | info |
818 (101) | NBP Equip Malfunc (101) | Task overrun | error |
818 (101) | NBP Equip Malfunc (101) | Task overrun in NBP equipment | error |
818 (102) | NBP Equip Malfunc (102) | Interrupt stack overwritten | error |
818 (102) | NBP Equip Malfunc (102) | Interrupt stack overwritten | error |
818 (103) | NBP Equip Malfunc (103) | User stack overwritten | error |
818 (103) | NBP Equip Malfunc (103) | User stack overwritten | error |
818 (104) | NBP Equip Malfunc (104) | Data validity check failed | error |
818 (104) | NBP Equip Malfunc (104) | Data validity check failed | error |
818 (106) | NBP Equip Malfunc (106) | Unable to write init flag data to non-volatile memory | error |
818 (106) | NBP Equip Malfunc (106) | Unable to write init flag data to non-volatile memory | error |
818 (107) | NBP Equip Malfunc (107) | Unable to verify init flag data written to non-volatile memory | error |
818 (107) | NBP Equip Malfunc (107) | Unable to verify init flag data in non-volatile memory | error |
818 (109) | NBP Equip Malfunc (109) | Unable to write NBP configuration data | error |
818 (109) | NBP Equip Malfunc (109) | Unable to write NBP configuration data to non-volatile memory | error |
818 (110) | NBP Equip Malfunc (110) | Unable to write NBP configuration data to non-volatile memory | error |
818 (110) | NBP Equip Malfunc (110) | Unable to write NBP configuration data | error |
818 (111) | NBP Equip Malfunc (111) | Calibration data checksum error | error |
818 (111) | NBP Equip Malfunc (111) | Calibration data checksum error detected | error |
818 (112) | NBP Equip Malfunc (112) | Unable to write calibration data to non-volatile memory | error |
818 (112) | NBP Equip Malfunc (112) | Unable to write calibration data | error |
818 (113) | NBP Equip Malfunc (113) | Unable to verify calibration data | error |
818 (113) | NBP Equip Malfunc (113) | Unable to verify calibration data written to non-volatile memory | error |
818 (114) | NBP Equip Malfunc (114) | Error log checksum error detected | error |
818 (114) | NBP Equip Malfunc (114) | Error log checksum error | error |
818 (115) | NBP Equip Malfunc (115) | Unable to write error log to non-volatile memory | error |
818 (115) | NBP Equip Malfunc (115) | Unable to write error log to non-volatile memory | error |
818 (116) | NBP Equip Malfunc (116) | Unable to verify error log data written to non-volatile memory | error |
818 (116) | NBP Equip Malfunc (116) | Unable to verify error log data | error |
818 (117) | NBP Equip Malfunc (117) | Inflation cycle count and runtime data checksum error | error |
818 (117) | NBP Equip Malfunc (117) | Inflation cycle count and runtime data checksum error | error |
818 (118) | NBP Equip Malfunc (118) | Unable to write inflation cycle count and runtime data to non-volatile memory | error |
818 (118) | NBP Equip Malfunc (118) | Unable to write inflation cycle count data | error |
818 (119) | NBP Equip Malfunc (119) | Unable to verify inflation cycle count data | error |
818 (119) | NBP Equip Malfunc (119) | Unable to verify inflation cycle count and runtime data written to non-volatile memory | error |
818 (120) | NBP Equip Malfunc (120) | New message sent to host while previous message not finished | error |
818 (120) | NBP Equip Malfunc (120) | New message sent while previous message not finished | error |
818 (121) | NBP Equip Malfunc (121) | Pressure value from second sensor not available | error |
818 (121) | NBP Equip Malfunc (121) | Pressure value from second pressure sensor not available | error |
818 (122) | NBP Equip Malfunc (122) | 5-volt supply voltage value is too high or low | error |
818 (122) | NBP Equip Malfunc (122) | 5-volt supply voltage too high or low | error |
818 (123) | NBP Equip Malfunc (123) | Error from 5-volt regulator | error |
818 (123) | NBP Equip Malfunc (123) | Error from the 5-volt regulator | error |
818 (124) | NBP Equip Malfunc (124) | 12-volt supply voltage value is too high or low | error |
818 (124) | NBP Equip Malfunc (124) | 12-volt supply voltage too high or low | error |
818 (128) | NBP Equip Malfunc (128) | Pressure sensor 1 offset out of range | error |
818 (128) | NBP Equip Malfunc (128) | Offset of pressure sensor 1 is out of range | error |
818 (129) | NBP Equip Malfunc (129) | Offset of pressure sensor 2 is out of range | error |
818 (129) | NBP Equip Malfunc (129) | Pressure sensor 2 offset out of range | error |
818 (130) | NBP Equip Malfunc (130) | Pressure sensor 2 test signal response out of range | error |
818 (130) | NBP Equip Malfunc (130) | Response of pressure sensor 2 to test signal is out of range | error |
818 (131) | NBP Equip Malfunc (131) | Difference between two pressure channels is too large | error |
818 (131) | NBP Equip Malfunc (131) | Difference between two pressure channels too large | error |
818 (132) | NBP Equip Malfunc (132) | Valve activated for too long | error |
818 (132) | NBP Equip Malfunc (132) | One of the two valves was activated for too long | error |
818 (133) | NBP Equip Malfunc (133) | CRC mismatch in calibration data | error |
818 (133) | NBP Equip Malfunc (133) | CRC mismatch in calibration data | error |
818 (134) | NBP Equip Malfunc (134) | CRC mismatch in error log data | error |
818 (134) | NBP Equip Malfunc (134) | CRC mismatch in error log data | error |
818 (135) | NBP Equip Malfunc (135) | CRC mismatch in inflation cycle count and runtime data | error |
818 (135) | NBP Equip Malfunc (135) | CRC mismatch in inflation cycle count data | error |
818 (136) | NBP Equip Malfunc (136) | Pneumatic safety switch malfunction | error |
818 (136) | NBP Equip Malfunc (136) | Pneumatic safety switch is defective | error |
818 (137) | NBP Equip Malfunc (137) | Valve 2 is defective or blocked | error |
818 (137) | NBP Equip Malfunc (137) | Valve 2 is defective or blocked | error |
818 (141) | NBP Equip Malfunc (141) | Inflation time exceeded for selected patient type | warning |
818 (141) | NBP Equip Malfunc (141) | Inflation time for selected patient type exceeded | warning |
818 (15) | NBP Equip Malfunc (15) | Abnormal unrecoverable system errors with internal firmware | error |
818 (15) | NBP Equip Malfunc (15) - NBP abnormal, unrecoverable system errors with internal firmware | NBP abnormal, unrecoverable system errors with internal firmware | critical |
818 (15) | NBP Equip Malfunc (15) | NBP abnormal unrecoverable system errors | error |
818 (199) | NBP Equipment Malfunction | No data from NBP block | error |
818 (199) | NBP Equip Malfunc (199) | No data from NBP block | error |
818 (2) | NBP Equip Malfunc (2) | NBP power-on self test failed | error |
818 (2) | NBP Equip Malfunc (2) - A/D converter inoperative or pressure transducer offset too large | NBP power-on self test failed; A/D converter inoperative or pressure transducer offset too large | error |
818 (2) | NBP Equip Malfunc (2) | NBP power-on self test failed | error |
818 (25) | NBP Equip Malfunc (25) | NBP hardware malfunction with no data | error |
818 (25) | NBP Equip Malfunc (25) - NBP hardware has malfunctioned because of no data | NBP hardware has malfunctioned because of no data | error |
818 (25) | NBP Equip Malfunc (25) | NBP hardware malfunction due to no data | error |
819 | NBP Equipment Malfunction - Not Calibrated | NBP unit not calibrated | error |
819 | NBP Equipment Malfunction - Not Calibrated | NBP module not calibrated | error |
819 (127) | NBP Equipment Malfunction | NBP module not calibrated | error |
819 (127) | NBP Equip Malfunc (127) | NBP module is not calibrated | error |
819 (138) | NBP Hose Blocked | Problem in pneumatic system | error |
819 (139) | NBP Air Leak | Air leak or valve defect in NBP system | error |
820 (138) | NBP Hose Blocked | Problem in pneumatic system | error |
821 (139) | NBP Air Leak | Air leak or valve defect in pneumatic system | error |
833 | Temperature Probe Error | Temperature probe detection or malfunction | error |
833 (1) | SpO2 Equip Malfunc (1) - SpO2 ROM checksum error | SpO2 ROM checksum error | error |
833 (1) | SpO2 Equipment Malfunction | SpO2 ROM checksum error | error |
833 (10) | SpO2 Equipment Malfunction | SpO2 host communication error | error |
833 (10) | SpO2 Equip Malfunc (10) - SpO2 host communication error | SpO2 host communication error | error |
833 (2) | SpO2 Equip Malfunc (2) - SpO2 RAM test error | SpO2 RAM test error | error |
833 (2) | SpO2 Equipment Malfunction | SpO2 RAM test error | error |
833 (20) | SpO2 Equipment Malfunction | SpO2 hardware defect | error |
833 (20) | SpO2 Equip Malfunc (20) - SpO2 hardware defect | SpO2 hardware defect | error |
833 (4) | SpO2 Equipment Malfunction - Analog Signal Path Self Test Failed | SpO2 analog signal path self test failed | error |
833 (4) | SpO2 Equip Malfunc (4) - SpO2 analog signal path self test failed | SpO2 analog signal path self test failed | error |
833 (4) | SpO2 Equipment Malfunction | SpO2 analog signal path self test failed | error |
833 (40) | SpO2 Equipment Malfunction | SpO2 non-volatile memory corrupted | error |
833 (40) | SpO2 Equip Malfunc (40) - SpO2 non-volatile memory corrupted | SpO2 non-volatile memory corrupted | error |
833 (40) | SpO2 Equipment Malfunction - Non-Volatile Memory Corrupted | SpO2 non-volatile memory corruption detected | error |
833 (6) | SpO2 Equip Malfunc (6) - SpO2 self test timeout | SpO2 self test timeout | error |
833 (6) | SpO2 Equipment Malfunction | SpO2 self test timeout | error |
833 (8) | SpO2 Equipment Malfunction | SpO2 internal DSP communication error | error |
833 (8) | SpO2 Equip Malfunc (8) - SpO2 internal DSP communication error | SpO2 internal DSP communication error | error |
834 | SpO2 Sensor Malfunc - Malfunction of SpO2 sensor or sensor cable | Malfunction of the SpO2 sensor or sensor cable | warning |
834 | SpO2 Sensor Malfunction | Malfunction of SpO2 sensor or sensor cable | warning |
835 | FPGA FIFO Overrun | SpO2 software error | info |
835 | FPGA FIFO Overrun | SpO2 software problem | error |
836 | SpO2 Communication Error | SpO2 communication error between modules | error |
836 | SpO2 Communication Error | SpO2 communication error | warning |
837 | SpO2 Error | No data from SpO2, front end will reset | error |
837 | SpO2 Error - No data from SpO2 | No data from SpO2 | error |
865 | FPGA FIFO Overrun | Temperature software problem | error |
865 | FPGA FIFO Overrun | Temperature software error | info |
866 | SpO2 Equipment Malfunction | SpO2 module detection or communication failure | error |
866 (21) | Temp Probe Error (21) - Probe operating in temperature above 112°F/43.3°C | Probe is operating in temperature conditions above 112°F/43.3°C; probe tip is defective with heater too close to thermistor | warning |
866 (21) | Temperature Probe Error | Probe operating at temperature above 112°F/43.3°C or probe tip defective | unknown |
866 (21) | Temperature Probe Error | Probe operating above 43.3°C (112°F) | error |
866 (22) | Temperature Probe Error | Excessive heater energy | warning |
866 (22) | Temp Probe Error - Excessive Heater Energy | Excessive heater energy detected | error |
866 (22) | Temp Probe Error (22) - Excessive heater energy | Excessive heater energy; heater not working or took too long to predict | warning |
866 (32) | Temp Probe Error (32) - Thermistor pulled away or heater broken | Thermistor pulled away from the tip or heater broken | warning |
866 (32) | Temp Probe Error (32) | Thermistor pulled away from tip or heater broken | warning |
866 (32) | Temperature Probe Error | Thermistor pulled away from tip or heater broken | warning |
866 (33) | Temperature Probe Error | Probe not responsive | warning |
866 (33) | Temp Probe Error (33) | Probe not connected at startup or not responsive | error |
866 (33) | Temp Probe Error (33) - Probe not responsive or not calibrated | Probe not responsive; probe not characterized/calibrated | warning |
866 (34) | Temp Probe Error (34) | Probe not characterized or calibrated | error |
866 (34) | Temp Probe Error (34) - Probe not characterized or calibrated | Probe not characterized/calibrated | warning |
866 (34) | Temperature Probe Error | Probe not characterized or calibrated | warning |
866 (45) | Temperature Probe Error | Measurement below allowable temperature values | warning |
866 (45) | Temp Probe Error (45) | Measurement below allowable temperature values | error |
866 (45) | Temp Probe Error (45) - Measurement below allowable temperature values | Measurement below allowable temperature values and far below low ambient limits | warning |
866 (46) | Temp Probe Error (46) | Measurement above allowable temperature values | error |
866 (46) | Temp Probe Error (46) - Measurement above allowable temperature values | Measurement above allowable temperature values and far above high ambient and/or patient limits | warning |
866 (46) | Temperature Probe Error | Measurement above allowable temperature values | warning |
866 (52) | Temperature Probe Error | Probe not characterized or calibrated | error |
866 (52) | Temp Probe Error - Not Characterized | Probe not characterized or calibrated | error |
866 (52) | Temp Probe Error (52) - Probe not characterized or calibrated | Probe not characterized/calibrated | warning |
866 (63) | Temp Probe Error (63) | Probe well missing or not installed properly | error |
866 (63) | Temp Probe Error (63) - Probe well missing or not installed properly | Probe well missing or not installed properly | warning |
866 (63) | Temperature Probe Error | Probe well missing or not installed properly | error |
866 (70) | Temperature Probe Error | Error reading probe EEPROM | error |
866 (70) | Temp Probe Error (70) - Error reading the probe's EEPROM | Error reading the probe's EEPROM | warning |
866 (70) | Temp Probe Error - EEPROM Read Error | Error reading probe's EEPROM | error |
866 (75) | Temp Probe Error (75) - Error reading probe EEPROM or probe not factory tested | Error reading the probe's EEPROM correctly or the probe was not factory tested | warning |
866 (75) | Temp Probe Error (75) | Problem reading probe EEPROM correctly or probe not factory tested | error |
866 (75) | Temperature Probe Error | Error reading probe EEPROM or probe not factory tested | error |
866 (78) | Temperature Module Malfunction | Probe NVM error detection mechanism detected error | error |
866 (78) | Temp Module Malfunc (78) - Probe's non-volatile error detection mechanism detected error | The Probe's non-volatile error detection mechanism detected an error | error |
866 (78) | Temp Module Malfunction - Probe Non-Volatile Error | Probe non-volatile error detection mechanism triggered | error |
866 (79) | Temp Module Malfunction - Error Log Memory Error | Module non-volatile error log memory error detected | error |
866 (79) | Temp Module Malfunc (79) - Module's non-volatile error log memory error detection mechanism detected error | The module's non-volatile error log memory error detection mechanism detected an error | error |
866 (79) | Temperature Module Malfunction | Module NVM error log memory error detected | error |
866 (80) | Temperature Probe Error | Software trying to turn off probe heater but heater still on | error |
866 (80) | Temp Probe Error - Heater Stuck On | Software unable to turn off probe heater feedback still on | error |
866 (80) | Temp Probe Error (80) - Software trying to turn off heater but heater still on | Software is trying to turn off the probe heater, but the heater feedback signal says it is still on | error |
866 (81) | Temp Probe Error - Heater Stuck Off | Software unable to turn on probe heater feedback still off | error |
866 (81) | Temp Probe Error (81) - Software trying to turn on heater but heater still off | Software is trying to turn on the probe heater, but the heater feedback signal says it is still off | error |
866 (81) | Temperature Probe Error | Software trying to turn on probe heater but heater still off | error |
867 (101) | Temp Module Malfunc (101) | Tympanic temperature module checksum error | error |
867 (102) | Temp Module Malfunc (102) | Tympanic temperature module calibration error | error |
867 (103) | Temp Module Malfunc (103) | Ambient temperature is below the minimum operating temperature limit | warning |
867 (104) | Temp Module Malfunc (104) | Ambient temperature is above the maximum operating temperature limit | warning |
867 (24) | Temperature Module Malfunction | Ambient temperature too high | warning |
867 (24) | Temp Module Malfunc (24) - Ambient temperature too high | Ambient temperature too high | warning |
867 (25) | Temp Module Malfunction - Ambient Temperature Too Low | Ambient temperature below minimum operating range | warning |
867 (25) | Temperature Module Malfunction | Ambient temperature too low | warning |
867 (27) | Temperature Module Malfunction | Battery or power supply voltage exceeds maximum | error |
867 (27) | Temp Module Malfunction - Voltage Exceeds Maximum | Battery or power supply voltage exceeds maximum | error |
867 (28) | Temperature Module Malfunction | Battery or power supply voltage below minimum | error |
867 (28) | Temp Module Malfunction - Voltage Below Minimum | Battery or power supply voltage below minimum | error |
867 (47) | Temperature Module Malfunction | Internal calibration resistor RCAL damaged | error |
867 (47) | Temp Module Malfunc (47) | Internal calibration resistor (RCAL) damaged | error |
867 (47) | Temp Module Malfunction - Internal Calibration Resistor RCAL Damaged | Internal calibration resistor RCAL damaged or defective | error |
867 (48) | Temp Module Malfunction - Calibration Resistor RCAL Damaged | Internal calibration resistor RCAL damaged | error |
867 (48) | Temperature Module Malfunction | Internal calibration resistor RCAL damaged | error |
867 (49) | Temperature Module Malfunction | Internal circuit validation resistor PTB damaged | error |
867 (49) | Temp Module Malfunction - Validation Resistor PTB Damaged | Internal circuit validation resistor PTB damaged | error |
867 (50) | Temp Module Malfunction - Validation Resistor PTB Damaged | Internal circuit validation resistor PTB damaged | error |
867 (50) | Temperature Module Malfunction | Internal circuit validation resistor PTB damaged | error |
867 (51) | Temperature Module Malfunction | A/D measurement timeout | error |
867 (51) | Temp Module Malfunction - A/D Measurement Timeout | A/D converter measurement timed out | error |
867 (59) | Temp Module Malfunction - Voltage Below Maximum | Battery or power supply voltage below maximum | error |
867 (59) | Temperature Module Malfunction | Battery or power supply voltage below maximum | error |
867 (60) | Temperature Module Malfunction | Battery or power supply voltage exceeds minimum | error |
867 (60) | Temp Module Malfunction - Voltage Exceeds Minimum | Battery or power supply voltage exceeds minimum | error |
867 (61) | Temp Module Malfunction - Reference Voltage Unstable | Reference voltage circuit under voltage or unstable | error |
867 (61) | Temperature Module Malfunction | Reference voltage circuit under voltage or unstable | error |
867 (62) | Temperature Module Malfunction | Module not calibrated | error |
867 (62) | Temp Module Malfunction - Not Calibrated | Temperature module not calibrated | error |
867 (65) | Temp Module Malfunction - EEPROM Write Error | Error saving to module's EEPROM | error |
867 (65) | Temp Module Malfunc (65) | Error saving to the module's EEPROM | error |
867 (66) | Temp Module Malfunc (66) | Module's non-volatile calibration memory error detection mechanism detected an error | error |
867 (66) | Temp Module Malfunction - Non-Volatile Calibration Error | Non-volatile calibration memory error detected | error |
867 (67) | Temp Module Malfunc (67) | Error reading/writing to the module's EEPROM | error |
867 (67) | Temp Module Malfunc (67) | Error reading/writing to the module's EEPROM | error |
867 (68) | Temp Module Malfunc (68) | Error reading/writing to the module's EEPROM | error |
867 (68) | Temp Module Malfunc (68) | Error reading/writing to the module's EEPROM | error |
867 (69) | Temp Module Malfunc (69) | Error reading/writing to the module's EEPROM | error |
867 (69) | Temp Module Malfunc (69) | Error reading/writing to the module's EEPROM | error |
867 (74) | Temp Module Malfunc (74) | Internal error during module EEPROM initialization | error |
867 (74) | Temp Module Malfunc (74) | Internal error during EEPROM initialization | error |
867 (80) | Temp Module Malfunc (80) | Software trying to turn off probe heater but feedback signal shows still on | error |
867 (81) | Temp Module Malfunc (81) | Software trying to turn on probe heater but feedback signal shows still off | error |
867 (82) | Temp Module Malfunc (82) | Signal HTR_Q on, HTRC off, but voltage still present | error |
867 (82) | Temp Module Malfunc (82) | Signal HTR_Q is on and signal HTRC is off, but still have voltage | error |
867 (83) | Temp Module Malfunc (83) | Signal HTR_Q is tri-stated with signal HTRC enabled, and has heater power | error |
867 (83) | Temp Module Malfunc (83) | Signal HTR_Q tri-stated with HTRC enabled and heater power present | error |
867 (84) | Temp Module Malfunc (84) | Signal Q&C on with insufficient heater voltage | error |
867 (84) | Temp Module Malfunc (84) | Signal Q&C is turned on and the heater voltage is not high enough | error |
867 (85) | Temp Module Malfunc (85) | Heater hardware fail-safe should have turned off, but did not | error |
867 (85) | Temp Module Malfunc (85) | Heater fail-safe did not turn off as expected | error |
868 | Temp Error. The front end will now reset itself. | Temperature measurement failure or module communication error | error |
868 | Temperature error | No data from temperature module | error |
868 | Temp Error. The front end will now reset itself. | No data from temperature sensor | error |
881 | Speaker Malfunc (881) | The speaker failed | error |
881 | Speaker Malfunc (881) | Speaker hardware failure | warning |
Err3 | Load beam cable disconnection or defect | One or more load beam cables disconnected or damaged | error |
Upgrade interrupted - Main board data lost | Software upgrade process was interrupted; main board memory corrupted and cannot be recovered | USB flash drive removed, monitor unplugged, or keys pressed during upgrade process | critical |
Upgrade Software window displays no valid image | USB flash drive does not contain valid software upgrade image or image is corrupted | After inserting USB flash drive and selecting Upgrade Software from System Admin Menu, no software version information appears | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.