Philips VS3: troubleshooting and error codes

The ServiceForge knowledge base holds 205 documented symptoms for the Philips VS3, with 446 ranked causes, 281 diagnostic checks with test points and expected values, and 286 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Monitor will not power on with AC caution

Monitor powers on with battery but not when connected to AC outlet

Ranked causes

  1. 1. AC power cord unplugged (common)
  2. 2. AC power cord damaged (common)
  3. 3. Power supply fuse blown (occasional)
  4. 4. Power supply cable malfunction (occasional)
  5. 5. Power supply output voltage failure (occasional)
  6. 6. AC connector malfunction (rare)
  7. 7. Monitor keypad worn (rare)
  8. 8. Main board failure (occasional)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not power on with AC power, works with battery

Monitor cannot be turned on using AC power connection but powers on normally when using battery power

Ranked causes

  1. 1. AC power cord unplugged or broken (common)
  2. 2. Fuse is blown (common)
  3. 3. Power supply cable has malfunctioned (occasional)
  4. 4. Power supply has malfunctioned (occasional)
  5. 5. AC power connector has malfunctioned (occasional)
  6. 6. Monitor keypad is worn (rare)
  7. 7. Main board has malfunctioned (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor will not power on with AC power (battery works) caution

AC power button does not turn on monitor; battery power operates normally

Ranked causes

  1. 1. AC power cord unplugged or broken (common)
  2. 2. Fuse is blown (common)
  3. 3. Power supply cable malfunction (occasional)
  4. 4. Power supply output failure (occasional)
  5. 5. AC power connector malfunction (occasional)
  6. 6. Monitor keypad worn (rare)
  7. 7. Main board malfunction (rare)
First check

Verify AC power cord is plugged in

Expected
Power cord connected to wall outlet
Tools
Visual inspection

2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature measurements are not displayed

Temperature readings do not appear on monitor screen

Ranked causes

  1. 1. Probe is disconnected from monitor (common)
  2. 2. Probe is misplaced (common)
  3. 3. Predictive temperature is over range (occasional)
  4. 4. Probe malfunctioned (occasional)
  5. 5. Front end board malfunction (rare)
  6. 6. Temperature module malfunctioned (rare)
  7. 7. Main board malfunctioned (rare)
First check

Verify probe is connected and seated

Test point
Probe connector and probe well
Expected
Probe connected and fully seated

1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on with AC power high voltage

Monitor turns on with battery power but not with AC power supply

Ranked causes

  1. 1. AC power cord unplugged or disconnected (common)
  2. 2. AC power cord broken (common)
  3. 3. Fuse blown (common)
  4. 4. Power supply cable malfunctioned (occasional)
  5. 5. Power supply output voltage incorrect (occasional)
  6. 6. AC power connector malfunctioned (occasional)
  7. 7. Main board malfunctioned (rare)
First check

Verify AC power cord is connected to outlet

Test point
AC power outlet
Expected
Power cord plugged in

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery: Verify charge level and health status patient safety

Check battery charge status via LED indicator, battery status pane, and alarm messages to assess battery condition and charge level.

Ranked causes

  1. 1. Low battery charge (common)
  2. 2. Battery nearly discharged (common)
  3. 3. Battery module malfunction or temperature out of range (occasional)
  4. 4. Battery charger power failure (rare)
  5. 5. Battery overcharged (occasional)
  6. 6. Battery over temperature (occasional)
First check

Observe LED charging indicator color

Test point
front panel charging LED
Expected
Green=full/nearly full; Flashing green=medium; Yellow=low; Flashing yellow=nearly discharged

2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor screen is blank

Display remains dark with no visible content when power is on

Ranked causes

  1. 1. Monitor in power save mode (common)
  2. 2. Speaker/fan cables disconnected (common)
  3. 3. Bad backlight LCD connection (occasional)
  4. 4. Bad backlight inverter (occasional)
  5. 5. Bad LCD (occasional)
  6. 6. Main PCB failure (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor screen is blank despite power on caution

Monitor power is on but display shows nothing; no image on screen

Ranked causes

  1. 1. Monitor in power save mode (common)
  2. 2. Speaker or fan cables disconnected (common)
  3. 3. Bad LCD backlight connection (occasional)
  4. 4. Bad backlight inverter (occasional)
  5. 5. LCD malfunction (occasional)
  6. 6. Main board malfunction (rare)
First check

Verify monitor is not in power save mode

Expected
Screen displays active image
Tools
Power button

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor won't power on with AC but works on battery

Power button does not activate when AC supply connected, but device powers on using battery

Ranked causes

  1. 1. AC power cord unplugged or broken (common)
  2. 2. Power supply fuse blown (common)
  3. 3. Power supply cable malfunction (occasional)
  4. 4. Power supply output failure (occasional)
  5. 5. AC connector malfunction (occasional)
  6. 6. Main PCB failure (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP measurements are not displayed

Non-invasive blood pressure readings do not appear on screen

Ranked causes

  1. 1. Wrong cuff size or incorrect placement (common)
  2. 2. Initial pressure setting is incorrect (common)
  3. 3. Air tube bent, twisted, or kinked with air leakage (common)
  4. 4. External problem affecting measurement (occasional)
  5. 5. NBP pump filter is blocked (occasional)
  6. 6. NBP module error (rare)
First check

Verify correct cuff size is selected and applied

Test point
Cuff selection and application site
Expected
Correct cuff size for patient used

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No NBP readings displayed patient safety

Blood pressure measurement shows no numeric values or dashes

Ranked causes

  1. 1. Incorrect cuff size or placement (common)
  2. 2. Incorrect initial pressure setting (common)
  3. 3. Kinked tube or air leakage (common)
  4. 4. External patient or environmental issue (occasional)
  5. 5. NBP pump filter blocked (occasional)
  6. 6. NBP module error (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No temperature readings

Temperature measurement displays no value or shows out-of-range error

Ranked causes

  1. 1. Temperature probe lead disconnected (common)
  2. 2. Temperature reading out of range (occasional)
  3. 3. Temperature probe malfunction (occasional)
  4. 4. Front end malfunction (occasional)
  5. 5. Temperature module malfunction (occasional)
  6. 6. Main board malfunction (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No temperature readings displayed patient safety

Temperature measurement missing from display; probe not recording or calculating temperature

Ranked causes

  1. 1. Probe not connected (common)
  2. 2. Temperature reading out of range (common)
  3. 3. Probe malfunction (occasional)
  4. 4. Front end malfunction (occasional)
  5. 5. Temperature module malfunction (occasional)
  6. 6. Main board malfunction (rare)
First check

Verify temperature probe is physically connected

Expected
Probe cable securely connected to monitor
Tools
Visual inspection

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software upgrade fails or is interrupted patient safety

Upgrade process halted unexpectedly, main board data lost, monitor non-functional after attempted software update

Ranked causes

  1. 1. USB flash drive removed during upgrade (occasional)
  2. 2. Monitor unplugged during upgrade (occasional)
  3. 3. Keys pressed during upgrade (occasional)
  4. 4. Software upgrade through USB hub (rare)
  5. 5. Invalid software image on USB drive (occasional)
  6. 6. Battery not fully charged (common)
First check

Verify battery charge level

Test point
Battery indicator on monitor display
Expected
Fully charged

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode scanner cannot read barcodes (laser and sound work)

Barcode scanner powers on and responds but fails to read bar codes; laser visible but no decode

Ranked causes

  1. 1. Scanner held at wrong distance from barcode (common)
  2. 2. System does not recognize barcode format (occasional)
  3. 3. Unrecognized format but scanner functional (occasional)
  4. 4. System does not recognize scanner (rare)
  5. 5. Scanner mechanical failure (rare)
First check

Verify scanner distance from barcode

Expected
Scanner held 6 to 12 inches from barcode
Tools
Ruler, visual alignment

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor power and basic operation troubleshooting

Initial diagnostic checks before detailed troubleshooting: power switch status, battery charge, AC connection, display function, and LED indicators

Ranked causes

  1. 1. Power switch not enabled (common)
  2. 2. Battery not adequately charged (common)
  3. 3. AC power not connected or outlet non-functional (common)
  4. 4. Display malfunction (occasional)
  5. 5. LED indicators not illuminated as expected (occasional)
First check

Verify power switch is turned on

Test point
Power switch on front panel
Expected
Power switch in ON position

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP cuff does not inflate

Non-invasive blood pressure cuff fails to inflate during measurement

Ranked causes

  1. 1. Air tube or cuff bent or twisted (common)
  2. 2. Air leak in tube or cuff (common)
  3. 3. NBP module malfunctioned (occasional)
  4. 4. NBP pump filter is blocked (occasional)
  5. 5. Main board malfunctioned (rare)
First check

Inspect air tube and cuff for kinks or bends

Test point
Air tube and cuff along full length
Expected
Tube and cuff straight with no twists or bends

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No SpO2 readings patient safety

Pulse oximetry measurement fails to return values

Ranked causes

  1. 1. SpO2 module in learning phase (common)
  2. 2. SpO2 sensor failure (common)
  3. 3. SpO2 board connection loose (occasional)
  4. 4. SpO2 board failure (occasional)
  5. 5. Main board failure (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nurse call does not signal alarms patient safety

Nurse call function fails to trigger alerts for alarm conditions

Ranked causes

  1. 1. Minimum Nurse Call Alarm Priority too high (common)
  2. 2. Alarms silenced (common)
  3. 3. Nurse call connector loose or broken (occasional)
  4. 4. Communications board failure (occasional)
  5. 5. Main board failure (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nurse call function does not signal alarm conditions patient safety

Nurse call system does not alert nursing staff to alarm conditions

Ranked causes

  1. 1. Minimum Nurse Call Alarm Priority set too high (common)
  2. 2. Alarms have been silenced (common)
  3. 3. Phono connector is loose or broken (occasional)
  4. 4. Communications board malfunctioned (rare)
  5. 5. Main board malfunctioned (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder does not print

Recorder fails to produce paper output

Ranked causes

  1. 1. Recorder is out of paper (common)
  2. 2. Recorder door is open (common)
  3. 3. System does not recognize the recorder (occasional)
  4. 4. Recorder module malfunctioned (occasional)
  5. 5. Main board malfunctioned (rare)
First check

Check recorder paper status

Test point
Paper compartment
Expected
Paper loaded and available

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software upgrade fails or monitor becomes unresponsive during upgrade patient safety

Monitor stops responding, upgrade process interrupts, or main board data is lost during software update

Ranked causes

  1. 1. Monitor unplugged during upgrade (common)
  2. 2. USB flash drive removed during upgrade (common)
  3. 3. Keys pressed during upgrade (common)
  4. 4. USB drive not fully inserted or clamp blocking insertion (occasional)
  5. 5. Upgrade interrupted, main board data lost (rare)
First check

Verify AC power connection

Test point
Monitor AC power port
Expected
Monitor connected to AC mains and powered on

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 measurements are not displayed

Oxygen saturation readings do not appear on monitor screen

Ranked causes

  1. 1. SpO2 module in learning phase (common)
  2. 2. SpO2 sensor malfunctioned (occasional)
  3. 3. SpO2 board connection to front end board faulty (occasional)
  4. 4. SpO2 board malfunctioned (rare)
  5. 5. Main board malfunctioned (rare)
First check

Monitor SpO2 module learning phase

Test point
SpO2 module status
Expected
Learning phase complete

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarms do not sound patient safety

Audio alarms are silent or not producing sound

Ranked causes

  1. 1. Audio Pause or Audio Off mode enabled (common)
  2. 2. Connection to speaker is faulty (occasional)
  3. 3. Speaker malfunctioned (occasional)
  4. 4. Main board malfunctioned (rare)
First check

Verify audio alarms are enabled

Test point
Audio settings menu
Expected
Audio mode is not Pause or Off

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode scanner cannot read barcodes

Laser and sound function but barcode is not decoded

Ranked causes

  1. 1. Incorrect scanner distance from barcode (common)
  2. 2. Unrecognized barcode format (common)
  3. 3. Scanner needs reset (occasional)
  4. 5. Scanner mechanical problem (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode scanner does not function (no laser, no sound)

Barcode scanner completely non-functional; no laser beam visible and no response sound

Ranked causes

  1. 1. Scanner not connected to USB port (common)
  2. 2. Scanner window not clean or blocked (common)
  3. 3. System does not recognize scanner (occasional)
  4. 4. Scanner mechanical failure (occasional)
First check

Verify scanner USB connection

Expected
Scanner cable securely connected to USB port
Tools
Visual inspection

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode scanner has no laser or sound

Scanner produces no audible tone and emits no laser beam

Ranked causes

  1. 1. Scanner not plugged into USB port (common)
  2. 2. Scanner window blocked or dirty (common)
  3. 3. System does not recognize scanner (occasional)
  4. 4. Scanner mechanical problem (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor displays random or distorted graphics

Display shows corrupted visual content with white background instead of normal image

Ranked causes

  1. 1. LCD cable not attached correctly (common)
  2. 2. Missing or improper ferrite clamp on LCD signal cable (common)
  3. 3. LCD failure (occasional)
  4. 4. Main board malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not power on with battery power, works with AC

Monitor cannot be turned on using battery power but powers on normally when using AC power

Ranked causes

  1. 1. Battery has lost its charge (common)
  2. 2. Battery or connector is loose (common)
  3. 3. Battery does not have connection to main board (occasional)
  4. 4. Main board has malfunctioned (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not respond to navigation wheel or front panel buttons caution

Buttons and navigation wheel do not register input; monitor does not respond to user commands

Ranked causes

  1. 1. Bad connection to main board (common)
  2. 2. Keypad or navigation wheel broken (occasional)
  3. 3. Main board malfunction (rare)
  4. 4. Membrane switch assembly malfunction (occasional)
First check

Check all connectors to main board

Expected
All connectors fully seated
Tools
Visual inspection

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not respond to navigation wheel or front panel keys

Navigation wheel and front panel keys do not provide input to the monitor

Ranked causes

  1. 1. Main board has faulty connection (occasional)
  2. 2. Front panel keypad or navigation wheel broken (common)
  3. 3. Main board malfunctioned (rare)
  4. 4. Membrane switch assembly malfunctioned (rare)
First check

Check main board connector seating

Test point
Main board connectors
Expected
All connectors fully seated

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor unresponsive to controls

Navigation wheel or front panel buttons do not control device

Ranked causes

  1. 1. Bad connection to main board (common)
  2. 2. Broken keypad or navigation wheel (common)
  3. 3. Main PCB failure (occasional)
  4. 4. Membrane switch assembly malfunction (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor will not power on with battery caution

Monitor powers on with AC power but not on battery

Ranked causes

  1. 1. Battery discharged (common)
  2. 2. Battery or connector loose (common)
  3. 3. Battery board connection failure (occasional)
  4. 4. Main board failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor will not power on with battery power (AC works) caution

Battery power button does not turn on monitor; AC power operates normally

Ranked causes

  1. 1. Battery has lost charge (common)
  2. 2. Battery or connector loose (common)
  3. 3. Battery not connected to main board (occasional)
  4. 4. Main board malfunction (rare)
First check

Check battery charge level

Expected
Battery fully charged
Tools
Battery charger

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor won't power on with battery but works on AC

Power button does not activate using battery, but device powers on when connected to AC supply

Ranked causes

  1. 1. Battery discharged (common)
  2. 2. Battery or connector loose (common)
  3. 3. Battery board disconnection (occasional)
  4. 4. Main PCB failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Navigation wheel or keys not responding caution

Monitor does not respond to navigation wheel or front panel key presses

Ranked causes

  1. 1. Main board connector loose (common)
  2. 2. Keypad or navigation wheel broken (common)
  3. 3. Membrane switch malfunction (common)
  4. 4. Main board failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP measurements are unreliable patient safety

Blood pressure readings are inconsistent or inaccurate

Ranked causes

  1. 1. Wrong cuff size or incorrect placement (common)
  2. 2. Initial pressure setting is incorrect (common)
  3. 3. External problem affecting measurement (occasional)
  4. 4. NBP module needs calibration (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP readings are unreliable or inconsistent patient safety

Blood pressure readings vary significantly or are outside expected range; inconsistent measurement results

Ranked causes

  1. 1. Wrong cuff size or incorrect placement (common)
  2. 2. Incorrect initial pressure setting (common)
  3. 3. External factor or patient movement (occasional)
  4. 4. Calibration drift (occasional)
First check

Verify cuff size is appropriate and correctly placed

Expected
Proper cuff size applied to correct limb location
Tools
Visual inspection

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No NBP readings obtained patient safety

Blood pressure measurement fails to complete; no systolic, diastolic, or MAP reading displayed

Ranked causes

  1. 1. Wrong cuff size or incorrect placement (common)
  2. 2. Incorrect initial pressure setting (common)
  3. 3. Kinked tube or air leakage (occasional)
  4. 4. External factor or patient movement (occasional)
First check

Verify cuff size is appropriate and correctly placed

Expected
Proper cuff size applied to correct limb location
Tools
Visual inspection

1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No SpO2 readings displayed patient safety

Pulse oximetry measurement missing from display; SpO2 sensor not recording oxygen saturation

Ranked causes

  1. 1. SpO2 module in learning phase (common)
  2. 2. SpO2 sensor malfunction (occasional)
  3. 3. SpO2 board malfunction (occasional)
  4. 4. Main board malfunction (rare)
First check

Check if SpO2 module is in learning phase

Expected
Learning phase complete; SpO2 reading available
Tools
Monitor display

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder does not print

Recorder fails to produce output on paper

Ranked causes

  1. 1. Out of paper (common)
  2. 2. Recorder door open (common)
  3. 4. Recorder module failure (occasional)
  4. 5. Main board failure (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test fails patient safety

Monitor resident self test reports FAILED result for one or more test items (Memory, SpO2, NBP, Temperature)

Ranked causes

  1. 1. Memory test failure (rare)
  2. 2. SpO2 module not detected or error (occasional)
  3. 3. NBP module not detected or error (occasional)
  4. 4. Temperature module not detected or error (occasional)
First check

Press Self Test button to initiate monitor resident self test

Test point
Self Test button
Expected
Self Test window appears and displays test status

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature measurements are unreliable patient safety

Temperature readings are inconsistent or inaccurate

Ranked causes

  1. 1. Probe malfunctioned (common)
  2. 2. Temperature module not calibrated (common)
  3. 3. Front end board malfunctioned (occasional)
  4. 4. Temperature module malfunctioned (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature readings are unreliable or inconsistent patient safety

Temperature values vary significantly or outside expected range; erratic temperature measurement

Ranked causes

  1. 1. Probe malfunction (common)
  2. 2. Temperature module accuracy drift (occasional)
  3. 3. Front end malfunction (occasional)
  4. 4. Temperature module malfunction (occasional)
First check

Perform temperature module accuracy verification

Expected
Module readings match reference thermometer within tolerance
Tools
Reference thermometer, Predictive Temperature Test

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on with battery power

Monitor turns on with AC power but not with battery power supply

Ranked causes

  1. 1. Battery not charged (common)
  2. 2. Battery or connector loose (common)
  3. 3. Battery board not connected to main board (occasional)
  4. 4. Main board malfunctioned (rare)
First check

Verify battery charge status

Test point
Battery charging port
Expected
Battery charged

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unreliable NBP readings patient safety

Blood pressure measurements are inconsistent or fluctuate significantly

Ranked causes

  1. 1. Incorrect cuff size or placement (common)
  2. 2. Incorrect initial pressure setting (common)
  3. 3. External patient or environmental issue (occasional)
  4. 4. NBP module calibration out of spec (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

USB hub does not work

External USB hub connected to monitor is not functioning

Ranked causes

  1. 1. USB hub is disconnected from monitor (common)
  2. 2. Hub USB ports blocked by dust or dirt (common)
  3. 3. USB hub is broken (occasional)
  4. 4. LAN (I/O Communications) board broken or not properly connected (rare)
First check

Verify USB plugs are firmly connected

Test point
USB ports on monitor and hub
Expected
All plugs seated firmly in ports

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarms not functioning patient safety

Visual and audio alarms do not trigger during NBP or SpO2 monitoring

Ranked causes

  1. 1. NBP alarm disabled or audio muted (common)
  2. 2. NBP module malfunction or improper connection (occasional)
  3. 3. SpO2 sensor malfunction or improper connection (occasional)
First check

Verify alarms are enabled on monitor

Test point
Audio and alarm mode
Expected
Audio Pause or Audio Off mode not active

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarms do not sound patient safety

Monitor does not produce audible alarm sounds

Ranked causes

  1. 1. Audio mode disabled (common)
  2. 2. Speaker connection loose (common)
  3. 4. Main board failure (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AUDIO FAILED message displayed with flashing red and white box patient safety

Flashing red and white box with AUDIO FAILED message appears on screen

Ranked causes

  1. 1. Speaker malfunctioned (common)
  2. 2. Connection to speaker is faulty (common)
  3. 3. Speaker is faulty (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode scanner does not function

Scanner shows no laser, no sound; cannot read barcodes or function at all.

Ranked causes

  1. 1. Scanner not plugged into USB port (common)
  2. 2. System does not recognize scanner (occasional)
  3. 3. Scanner has mechanical problem (occasional)
First check

Verify scanner physical connection to USB port

Test point
Monitor USB port
Expected
Scanner firmly connected

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor display shows distorted graphics

Monitor displays random or distorted graphics with white background

Ranked causes

  1. 1. LCD signal cable not properly attached (common)
  2. 2. LCD failure (occasional)
  3. 3. Main board failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor displays random or distorted graphics with white background caution

Screen shows garbled graphics, corrupted text, or white background with colored noise

Ranked causes

  1. 1. LCD signal cable not attached correctly (common)
  2. 2. LCD malfunction (occasional)
  3. 3. Main board malfunction (rare)
First check

Reseat LCD signal cable

Expected
Graphics display normally after reseating
Tools
Visual inspection

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor displays random or distorted graphics with white background

Screen shows corrupted display with random graphics and white background

Ranked causes

  1. 1. LCD signal cable not attached correctly (common)
  2. 2. LCD malfunctioned (occasional)
  3. 3. Main board malfunctioned (rare)
First check

Inspect and reseat LCD signal cable

Test point
LCD signal cable connectors
Expected
Cable fully seated and not damaged

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not power on

No power indication; display inactive; LEDs not lit

Ranked causes

  1. 1. Power switch off (common)
  2. 2. Battery inadequately charged (common)
  3. 3. AC power cord disconnected or outlet non-functional (common)
First check

Verify power switch position

Test point
Main power switch
Expected
Switch in ON position

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not respond to controls

Navigation wheel and front panel buttons do not control menu or operation

Ranked causes

  1. 1. Bad connection to main board (common)
  2. 2. Keypad or navigation wheel broken (occasional)
  3. 3. Main board malfunction (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor screen is blank when powered on

Power is on but monitor screen displays nothing

Ranked causes

  1. 1. Monitor in Standby mode (common)
  2. 2. Backlight LCD connection faulty (occasional)
  3. 3. Main board malfunctioned (rare)
First check

Verify Standby mode is not active

Test point
Power button status
Expected
Monitor in normal operating mode

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP cuff does not inflate caution

Blood pressure cuff fails to inflate when measurement initiated; no pressure buildup

Ranked causes

  1. 1. Tube or cuff is kinked (common)
  2. 3. NBP module malfunction (occasional)
  3. 5. Main board malfunction (rare)
First check

Inspect tube and cuff for kinks

Expected
Tube and cuff straight and flexible
Tools
Visual inspection

1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Random or distorted graphics displayed

Monitor shows random patterns or distorted image with white background

Ranked causes

  1. 1. LCD cable not properly attached (common)
  2. 2. Bad LCD (occasional)
  3. 3. Main PCB failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder paper is jamming

Recorder paper gets stuck or jams during operation

Ranked causes

  1. 1. Paper not installed correctly (common)
  2. 2. Wrong type of paper is installed (common)
  3. 3. Recorder has mechanical problem (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Software upgrade fails or monitor cannot boot
  • Software upgrade interrupted or fails
  • Unit does not power on
  • Unreliable temperature readings
  • USB hub does not work
  • Audio failed alarm displayed
  • Barcode scanner: Verify scanner accuracy and patient data input
  • Battery service required message displayed
  • Monitor boot sequence does not complete
  • Monitor cannot find valid software image on USB flash drive
  • Monitor does not power on
  • Monitor has no power
  • NBP alarm not functioning
  • NBP initial pressure too high
  • NBP pneumatic leakage exceeds limit
  • Pneumatic leakage test failure
  • Recorder malfunction
  • Recorder paper is jamming
  • Recorder test pattern abnormal
  • Recorder test pattern abnormal
  • Recorder test pattern indicates print head or motor fault
  • SpO2 measurement out of range
  • SpO2 measurements are unreliable
  • SpO2 reading outside normal range
  • SpO2 readings are unreliable or inconsistent
  • SpO2 readings unreliable
  • Access Maintenance options in System Diagnostics
  • Access System Diagnostics menu
  • Alarms testing
  • Audio output not heard or incorrect
  • Audio test not producing tones
  • Auto Set Alarm Limits button unavailable
  • Auto Set Alarm Limits button unavailable
  • Barcode scanner cannot read all patient identification methods
  • Barcode scanner programming reset
  • Barcode scanner reads only primary ID method
  • Battery capacity or type anomaly
  • Battery charging LED displays only one color
  • Battery charging LED malfunction
  • Battery charging LED shows only one color
  • Battery LED not behaving as expected during test
  • Battery LED not functioning
  • Battery LED not showing expected color sequence
  • Battery LED test failure
  • Button collapsed or not clicking
  • Button does not access correct function
  • Button is collapsed or does not click when pressed
  • Button press accesses wrong function
  • Button Test: rotary wheel not responding
  • Button Test: rotary wheel not responding or drawing red dots
  • Button Test: rotary wheel small movements not detected
  • Button Test: system button not responding
  • Button Test: system icon button not responding
  • Calibration: Battery reconditioning
  • Calibration: Internal parts replaced
  • Calibration: Monitor opened for maintenance
  • Calibration: NBP accuracy
  • Calibration: NBP module
  • Calibration: NBP overpressure valve
  • Calibration: NBP pneumatic leakage
  • Calibration: Performance verification after internal parts replacement
  • Calibration: Performance verification after monitor installation
  • Calibration: Performance verification after monitor opened
  • Calibration: Performance verification after NBP module replacement
  • Calibration: Performance verification after software upgrade
  • Calibration: Reset serial number after main PCB replacement
  • Calibration: Safety tests
  • Calibration: Software upgrade
  • Calibration: SpO2 accuracy
  • Calibration: Temperature accuracy
  • Calibration: Temperature measurement verification
  • Cannot initialize NBP
  • Display test pattern not showing or showing incorrectly
  • Display test patterns not appearing
  • Enclosure leakage current exceeds safety limit
  • Failed to stop work thread
  • Front panel LEDs not lit as expected
  • Hardware incompatibility or system corruption after software downgrade
  • Internal parts replaced (non-NBP, non-SpO2)
  • Key collapsed or unresponsive to press
  • Key is collapsed or does not click when pressed
  • Key press activates wrong function
  • LED test fails or unexpected LED behavior
  • Loss of monitoring
  • Memory error
  • Monitor display inactive or malfunctioning
  • Monitor display incorrect or absent
  • Monitor display is blank or incorrect
  • Monitor does not power on with either AC or battery power
  • Monitor fails to boot
  • Monitor freezes during startup
  • Monitor hangs during startup
  • Monitor in suspended mode does not respond to power button
  • Monitor opened for service
  • Monitor will not power on at all
  • Monitor will not power on with AC or battery power
  • Monitor will not power up after reassembly
  • Navigation wheel or keys not responding
  • NBP accuracy exceeds tolerance
  • NBP cycle count requires reset after module replacement
  • NBP equipment not calibrated
  • NBP measurement exceeds tolerance
  • NBP overpressure valve failure
  • NBP overpressure valve not opening
  • No startup tone during boot
  • Nurse call relay does not operate
  • Nurse call relay: Verify relay contact operation
  • Only one charging LED color displayed
  • Opening screen hangs
  • Opening screen hangs or does not complete
  • Opening screen or color bar hangs on startup
  • Power connector fuse replacement needed
  • Power-on self test
  • Predictive Temp module test fails during self-test
  • Predictive temperature measurement error
  • Pressing a key does not access expected function
  • Pump filter maintenance or replacement needed
  • Recorder prints incorrectly
  • Reset tracked component usage parameters
  • Reset tracked parameter
  • Rotary wheel or button not responding
  • Safety test: Enclosure leakage current
  • Safety test: Ground integrity
  • Safety test: Patient leakage current
  • Self test fails for module
  • Self test fails on NBP module
  • Self test in progress
  • Software configuration settings lost after upgrade
  • Software downgrade attempted
  • Software upgrade completed
  • Speaker and fan cable connections failing
  • Speaker not producing audio
  • System Admin Menu locks after use
  • System exception in alarming software
  • System settings and patient records lost after software upgrade
  • Temperature measurement out of range
  • Tracked parameters need resetting after part replacement
  • Tympanic temperature reading is unusually high
  • Tympanic temperature reading is unusually low
  • Unable to access System Admin Menu
  • Unit does not power on with AC or battery power
  • Unreliable SpO2 readings
  • USB flash drive does not fit in USB port
  • Visual damage or obstruction present
  • Visual inspection

145 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1025System Error (1025)Could not load ResourceDLL.dllcritical
1025System Error (1025)Could not load ResourceDLL.dllerror
1026System Error (1026)Could not load data from flasherror
1026System Error (1026)Could not load data from flashcritical
1027System Error (1027)Could not start front-endcritical
1027System Error (1027)Could not start the front enderror
1028System Error (1028)Could not stop the front enderror
1028System Error (1028)Could not stop front-endcritical
1029System Error (1029)Could not allocate enough memorycritical
1029System Error (1029)Could not allocate enough memoryerror
1030System Error (1030)An error occurred when the system shut down front end deviceserror
1030System Error (1030)Error during system shutdown of front end devicescritical
1031System Error (1031)Monitoring and watchdog thread could not initializecritical
1031System Error (1031)The monitoring and watchdog thread could not initializeerror
1032System Error (1032)Could not open audio fileserror
1032System Error (1032)Could not open audio filescritical
1033System Error (1033)Unspecified errorcritical
1033System Error (1033)Unspecified errorerror
1034System Error (1034)Failed to load product infoerror
1034System Error (1034)Failed to load product infocritical
1035System Error (1035)Cannot initialize front-end managercritical
1035System Error (1035)Cannot initialize the front-end managererror
1036System Error (1036)Failed to start work threaderror
1036System Error (1036)Failed to start work threadcritical
1037System Error (1037)Cannot start polling threadcritical
1037System Error (1037)Cannot start polling threaderror
1039System Error (1039)Cannot get front-end versionerror
1039System Error (1039)Cannot get front-end versioncritical
1042System Error (1042)Failed to start up network drivercritical
1042System Error (1042)Failed to start up network drivererror
1043System Error (1043)Work thread timed outerror
1043System Error (1043)Work thread timed outcritical
1043 (1)System Error (1)Front end software problemcritical
1043 (2)System Error (2)Alarm software problemcritical
1044System Error (1044)Front end software errorcritical
1044System Error (1044)Front end software errorerror
1045System Error (1045)Printing software errorerror
1045System Error (1045)Printing software errorcritical
1047System Error (1047)Monitoring software errorcritical
1047System Error (1047)Monitoring software errorerror
1048System Error (1048)An exception was detected in the alarming softwareerror
1048System error - exception in alarming softwareException detected in alarming softwareerror
1048System Error (1048)Exception detected in alarming softwarecritical
1050System Error (1050)Failed to stop work threadcritical
1050System error - failed to stop work threadFailed to stop work threaderror
1050System Error (1050)Failed to stop work threaderror
1051System Error (1051)Cannot initialize NBPerror
1051System Error (1051)Cannot initialize NBPcritical
1052System Error (1052)Memory errorcritical
1052System ErrorMemory error detected during operationcritical
1053System ErrorLoss of monitoring signal detectedinfo
1053System error - loss of monitoringLoss of monitoring capabilityinfo
257System ErrorBattery charger power failureerror
257System ErrorBattery charger power failureerror
257System Error - Battery charger power failureBattery charger power failureerror
261Battery is overchargedcharging circuit fails to regulate battery voltage; battery remains connected to charger beyond full charge statewarning
261System Error - Battery is overchargedBattery is overchargederror
261System ErrorBattery is overchargederror
264System ErrorBattery is over temperatureerror
264System ErrorBattery over temperature condition detectederror
264System Error - Battery over temperatureBattery over temperaturewarning
264 tempSystem ErrorBattery temperature exceeds safe operating limiterror
264 temp_valueBattery is over temperaturebattery temperature exceeds safe operating range during charging or dischargewarning
817FPGA FIFO OverrunNBP software problem detected during operationerror
817FPGA FIFO OverrunNBP software errorinfo
818 (101)NBP Equip Malfunc (101)Task overrunerror
818 (101)NBP Equip Malfunc (101)Task overrun in NBP equipmenterror
818 (102)NBP Equip Malfunc (102)Interrupt stack overwrittenerror
818 (102)NBP Equip Malfunc (102)Interrupt stack overwrittenerror
818 (103)NBP Equip Malfunc (103)User stack overwrittenerror
818 (103)NBP Equip Malfunc (103)User stack overwrittenerror
818 (104)NBP Equip Malfunc (104)Data validity check failederror
818 (104)NBP Equip Malfunc (104)Data validity check failederror
818 (106)NBP Equip Malfunc (106)Unable to write init flag data to non-volatile memoryerror
818 (106)NBP Equip Malfunc (106)Unable to write init flag data to non-volatile memoryerror
818 (107)NBP Equip Malfunc (107)Unable to verify init flag data written to non-volatile memoryerror
818 (107)NBP Equip Malfunc (107)Unable to verify init flag data in non-volatile memoryerror
818 (109)NBP Equip Malfunc (109)Unable to write NBP configuration dataerror
818 (109)NBP Equip Malfunc (109)Unable to write NBP configuration data to non-volatile memoryerror
818 (110)NBP Equip Malfunc (110)Unable to write NBP configuration data to non-volatile memoryerror
818 (110)NBP Equip Malfunc (110)Unable to write NBP configuration dataerror
818 (111)NBP Equip Malfunc (111)Calibration data checksum errorerror
818 (111)NBP Equip Malfunc (111)Calibration data checksum error detectederror
818 (112)NBP Equip Malfunc (112)Unable to write calibration data to non-volatile memoryerror
818 (112)NBP Equip Malfunc (112)Unable to write calibration dataerror
818 (113)NBP Equip Malfunc (113)Unable to verify calibration dataerror
818 (113)NBP Equip Malfunc (113)Unable to verify calibration data written to non-volatile memoryerror
818 (114)NBP Equip Malfunc (114)Error log checksum error detectederror
818 (114)NBP Equip Malfunc (114)Error log checksum errorerror
818 (115)NBP Equip Malfunc (115)Unable to write error log to non-volatile memoryerror
818 (115)NBP Equip Malfunc (115)Unable to write error log to non-volatile memoryerror
818 (116)NBP Equip Malfunc (116)Unable to verify error log data written to non-volatile memoryerror
818 (116)NBP Equip Malfunc (116)Unable to verify error log dataerror
818 (117)NBP Equip Malfunc (117)Inflation cycle count and runtime data checksum errorerror
818 (117)NBP Equip Malfunc (117)Inflation cycle count and runtime data checksum errorerror
818 (118)NBP Equip Malfunc (118)Unable to write inflation cycle count and runtime data to non-volatile memoryerror
818 (118)NBP Equip Malfunc (118)Unable to write inflation cycle count dataerror
818 (119)NBP Equip Malfunc (119)Unable to verify inflation cycle count dataerror
818 (119)NBP Equip Malfunc (119)Unable to verify inflation cycle count and runtime data written to non-volatile memoryerror
818 (120)NBP Equip Malfunc (120)New message sent to host while previous message not finishederror
818 (120)NBP Equip Malfunc (120)New message sent while previous message not finishederror
818 (121)NBP Equip Malfunc (121)Pressure value from second sensor not availableerror
818 (121)NBP Equip Malfunc (121)Pressure value from second pressure sensor not availableerror
818 (122)NBP Equip Malfunc (122)5-volt supply voltage value is too high or lowerror
818 (122)NBP Equip Malfunc (122)5-volt supply voltage too high or lowerror
818 (123)NBP Equip Malfunc (123)Error from 5-volt regulatorerror
818 (123)NBP Equip Malfunc (123)Error from the 5-volt regulatorerror
818 (124)NBP Equip Malfunc (124)12-volt supply voltage value is too high or lowerror
818 (124)NBP Equip Malfunc (124)12-volt supply voltage too high or lowerror
818 (128)NBP Equip Malfunc (128)Pressure sensor 1 offset out of rangeerror
818 (128)NBP Equip Malfunc (128)Offset of pressure sensor 1 is out of rangeerror
818 (129)NBP Equip Malfunc (129)Offset of pressure sensor 2 is out of rangeerror
818 (129)NBP Equip Malfunc (129)Pressure sensor 2 offset out of rangeerror
818 (130)NBP Equip Malfunc (130)Pressure sensor 2 test signal response out of rangeerror
818 (130)NBP Equip Malfunc (130)Response of pressure sensor 2 to test signal is out of rangeerror
818 (131)NBP Equip Malfunc (131)Difference between two pressure channels is too largeerror
818 (131)NBP Equip Malfunc (131)Difference between two pressure channels too largeerror
818 (132)NBP Equip Malfunc (132)Valve activated for too longerror
818 (132)NBP Equip Malfunc (132)One of the two valves was activated for too longerror
818 (133)NBP Equip Malfunc (133)CRC mismatch in calibration dataerror
818 (133)NBP Equip Malfunc (133)CRC mismatch in calibration dataerror
818 (134)NBP Equip Malfunc (134)CRC mismatch in error log dataerror
818 (134)NBP Equip Malfunc (134)CRC mismatch in error log dataerror
818 (135)NBP Equip Malfunc (135)CRC mismatch in inflation cycle count and runtime dataerror
818 (135)NBP Equip Malfunc (135)CRC mismatch in inflation cycle count dataerror
818 (136)NBP Equip Malfunc (136)Pneumatic safety switch malfunctionerror
818 (136)NBP Equip Malfunc (136)Pneumatic safety switch is defectiveerror
818 (137)NBP Equip Malfunc (137)Valve 2 is defective or blockederror
818 (137)NBP Equip Malfunc (137)Valve 2 is defective or blockederror
818 (141)NBP Equip Malfunc (141)Inflation time exceeded for selected patient typewarning
818 (141)NBP Equip Malfunc (141)Inflation time for selected patient type exceededwarning
818 (15)NBP Equip Malfunc (15)Abnormal unrecoverable system errors with internal firmwareerror
818 (15)NBP Equip Malfunc (15) - NBP abnormal, unrecoverable system errors with internal firmwareNBP abnormal, unrecoverable system errors with internal firmwarecritical
818 (15)NBP Equip Malfunc (15)NBP abnormal unrecoverable system errorserror
818 (199)NBP Equipment MalfunctionNo data from NBP blockerror
818 (199)NBP Equip Malfunc (199)No data from NBP blockerror
818 (2)NBP Equip Malfunc (2)NBP power-on self test failederror
818 (2)NBP Equip Malfunc (2) - A/D converter inoperative or pressure transducer offset too largeNBP power-on self test failed; A/D converter inoperative or pressure transducer offset too largeerror
818 (2)NBP Equip Malfunc (2)NBP power-on self test failederror
818 (25)NBP Equip Malfunc (25)NBP hardware malfunction with no dataerror
818 (25)NBP Equip Malfunc (25) - NBP hardware has malfunctioned because of no dataNBP hardware has malfunctioned because of no dataerror
818 (25)NBP Equip Malfunc (25)NBP hardware malfunction due to no dataerror
819NBP Equipment Malfunction - Not CalibratedNBP unit not calibratederror
819NBP Equipment Malfunction - Not CalibratedNBP module not calibratederror
819 (127)NBP Equipment MalfunctionNBP module not calibratederror
819 (127)NBP Equip Malfunc (127)NBP module is not calibratederror
819 (138)NBP Hose BlockedProblem in pneumatic systemerror
819 (139)NBP Air LeakAir leak or valve defect in NBP systemerror
820 (138)NBP Hose BlockedProblem in pneumatic systemerror
821 (139)NBP Air LeakAir leak or valve defect in pneumatic systemerror
833Temperature Probe ErrorTemperature probe detection or malfunctionerror
833 (1)SpO2 Equip Malfunc (1) - SpO2 ROM checksum errorSpO2 ROM checksum errorerror
833 (1)SpO2 Equipment MalfunctionSpO2 ROM checksum errorerror
833 (10)SpO2 Equipment MalfunctionSpO2 host communication errorerror
833 (10)SpO2 Equip Malfunc (10) - SpO2 host communication errorSpO2 host communication errorerror
833 (2)SpO2 Equip Malfunc (2) - SpO2 RAM test errorSpO2 RAM test errorerror
833 (2)SpO2 Equipment MalfunctionSpO2 RAM test errorerror
833 (20)SpO2 Equipment MalfunctionSpO2 hardware defecterror
833 (20)SpO2 Equip Malfunc (20) - SpO2 hardware defectSpO2 hardware defecterror
833 (4)SpO2 Equipment Malfunction - Analog Signal Path Self Test FailedSpO2 analog signal path self test failederror
833 (4)SpO2 Equip Malfunc (4) - SpO2 analog signal path self test failedSpO2 analog signal path self test failederror
833 (4)SpO2 Equipment MalfunctionSpO2 analog signal path self test failederror
833 (40)SpO2 Equipment MalfunctionSpO2 non-volatile memory corruptederror
833 (40)SpO2 Equip Malfunc (40) - SpO2 non-volatile memory corruptedSpO2 non-volatile memory corruptederror
833 (40)SpO2 Equipment Malfunction - Non-Volatile Memory CorruptedSpO2 non-volatile memory corruption detectederror
833 (6)SpO2 Equip Malfunc (6) - SpO2 self test timeoutSpO2 self test timeouterror
833 (6)SpO2 Equipment MalfunctionSpO2 self test timeouterror
833 (8)SpO2 Equipment MalfunctionSpO2 internal DSP communication errorerror
833 (8)SpO2 Equip Malfunc (8) - SpO2 internal DSP communication errorSpO2 internal DSP communication errorerror
834SpO2 Sensor Malfunc - Malfunction of SpO2 sensor or sensor cableMalfunction of the SpO2 sensor or sensor cablewarning
834SpO2 Sensor MalfunctionMalfunction of SpO2 sensor or sensor cablewarning
835FPGA FIFO OverrunSpO2 software errorinfo
835FPGA FIFO OverrunSpO2 software problemerror
836SpO2 Communication ErrorSpO2 communication error between moduleserror
836SpO2 Communication ErrorSpO2 communication errorwarning
837SpO2 ErrorNo data from SpO2, front end will reseterror
837SpO2 Error - No data from SpO2No data from SpO2error
865FPGA FIFO OverrunTemperature software problemerror
865FPGA FIFO OverrunTemperature software errorinfo
866SpO2 Equipment MalfunctionSpO2 module detection or communication failureerror
866 (21)Temp Probe Error (21) - Probe operating in temperature above 112°F/43.3°CProbe is operating in temperature conditions above 112°F/43.3°C; probe tip is defective with heater too close to thermistorwarning
866 (21)Temperature Probe ErrorProbe operating at temperature above 112°F/43.3°C or probe tip defectiveunknown
866 (21)Temperature Probe ErrorProbe operating above 43.3°C (112°F)error
866 (22)Temperature Probe ErrorExcessive heater energywarning
866 (22)Temp Probe Error - Excessive Heater EnergyExcessive heater energy detectederror
866 (22)Temp Probe Error (22) - Excessive heater energyExcessive heater energy; heater not working or took too long to predictwarning
866 (32)Temp Probe Error (32) - Thermistor pulled away or heater brokenThermistor pulled away from the tip or heater brokenwarning
866 (32)Temp Probe Error (32)Thermistor pulled away from tip or heater brokenwarning
866 (32)Temperature Probe ErrorThermistor pulled away from tip or heater brokenwarning
866 (33)Temperature Probe ErrorProbe not responsivewarning
866 (33)Temp Probe Error (33)Probe not connected at startup or not responsiveerror
866 (33)Temp Probe Error (33) - Probe not responsive or not calibratedProbe not responsive; probe not characterized/calibratedwarning
866 (34)Temp Probe Error (34)Probe not characterized or calibratederror
866 (34)Temp Probe Error (34) - Probe not characterized or calibratedProbe not characterized/calibratedwarning
866 (34)Temperature Probe ErrorProbe not characterized or calibratedwarning
866 (45)Temperature Probe ErrorMeasurement below allowable temperature valueswarning
866 (45)Temp Probe Error (45)Measurement below allowable temperature valueserror
866 (45)Temp Probe Error (45) - Measurement below allowable temperature valuesMeasurement below allowable temperature values and far below low ambient limitswarning
866 (46)Temp Probe Error (46)Measurement above allowable temperature valueserror
866 (46)Temp Probe Error (46) - Measurement above allowable temperature valuesMeasurement above allowable temperature values and far above high ambient and/or patient limitswarning
866 (46)Temperature Probe ErrorMeasurement above allowable temperature valueswarning
866 (52)Temperature Probe ErrorProbe not characterized or calibratederror
866 (52)Temp Probe Error - Not CharacterizedProbe not characterized or calibratederror
866 (52)Temp Probe Error (52) - Probe not characterized or calibratedProbe not characterized/calibratedwarning
866 (63)Temp Probe Error (63)Probe well missing or not installed properlyerror
866 (63)Temp Probe Error (63) - Probe well missing or not installed properlyProbe well missing or not installed properlywarning
866 (63)Temperature Probe ErrorProbe well missing or not installed properlyerror
866 (70)Temperature Probe ErrorError reading probe EEPROMerror
866 (70)Temp Probe Error (70) - Error reading the probe's EEPROMError reading the probe's EEPROMwarning
866 (70)Temp Probe Error - EEPROM Read ErrorError reading probe's EEPROMerror
866 (75)Temp Probe Error (75) - Error reading probe EEPROM or probe not factory testedError reading the probe's EEPROM correctly or the probe was not factory testedwarning
866 (75)Temp Probe Error (75)Problem reading probe EEPROM correctly or probe not factory testederror
866 (75)Temperature Probe ErrorError reading probe EEPROM or probe not factory testederror
866 (78)Temperature Module MalfunctionProbe NVM error detection mechanism detected errorerror
866 (78)Temp Module Malfunc (78) - Probe's non-volatile error detection mechanism detected errorThe Probe's non-volatile error detection mechanism detected an errorerror
866 (78)Temp Module Malfunction - Probe Non-Volatile ErrorProbe non-volatile error detection mechanism triggerederror
866 (79)Temp Module Malfunction - Error Log Memory ErrorModule non-volatile error log memory error detectederror
866 (79)Temp Module Malfunc (79) - Module's non-volatile error log memory error detection mechanism detected errorThe module's non-volatile error log memory error detection mechanism detected an errorerror
866 (79)Temperature Module MalfunctionModule NVM error log memory error detectederror
866 (80)Temperature Probe ErrorSoftware trying to turn off probe heater but heater still onerror
866 (80)Temp Probe Error - Heater Stuck OnSoftware unable to turn off probe heater feedback still onerror
866 (80)Temp Probe Error (80) - Software trying to turn off heater but heater still onSoftware is trying to turn off the probe heater, but the heater feedback signal says it is still onerror
866 (81)Temp Probe Error - Heater Stuck OffSoftware unable to turn on probe heater feedback still offerror
866 (81)Temp Probe Error (81) - Software trying to turn on heater but heater still offSoftware is trying to turn on the probe heater, but the heater feedback signal says it is still offerror
866 (81)Temperature Probe ErrorSoftware trying to turn on probe heater but heater still offerror
867 (101)Temp Module Malfunc (101)Tympanic temperature module checksum errorerror
867 (102)Temp Module Malfunc (102)Tympanic temperature module calibration errorerror
867 (103)Temp Module Malfunc (103)Ambient temperature is below the minimum operating temperature limitwarning
867 (104)Temp Module Malfunc (104)Ambient temperature is above the maximum operating temperature limitwarning
867 (24)Temperature Module MalfunctionAmbient temperature too highwarning
867 (24)Temp Module Malfunc (24) - Ambient temperature too highAmbient temperature too highwarning
867 (25)Temp Module Malfunction - Ambient Temperature Too LowAmbient temperature below minimum operating rangewarning
867 (25)Temperature Module MalfunctionAmbient temperature too lowwarning
867 (27)Temperature Module MalfunctionBattery or power supply voltage exceeds maximumerror
867 (27)Temp Module Malfunction - Voltage Exceeds MaximumBattery or power supply voltage exceeds maximumerror
867 (28)Temperature Module MalfunctionBattery or power supply voltage below minimumerror
867 (28)Temp Module Malfunction - Voltage Below MinimumBattery or power supply voltage below minimumerror
867 (47)Temperature Module MalfunctionInternal calibration resistor RCAL damagederror
867 (47)Temp Module Malfunc (47)Internal calibration resistor (RCAL) damagederror
867 (47)Temp Module Malfunction - Internal Calibration Resistor RCAL DamagedInternal calibration resistor RCAL damaged or defectiveerror
867 (48)Temp Module Malfunction - Calibration Resistor RCAL DamagedInternal calibration resistor RCAL damagederror
867 (48)Temperature Module MalfunctionInternal calibration resistor RCAL damagederror
867 (49)Temperature Module MalfunctionInternal circuit validation resistor PTB damagederror
867 (49)Temp Module Malfunction - Validation Resistor PTB DamagedInternal circuit validation resistor PTB damagederror
867 (50)Temp Module Malfunction - Validation Resistor PTB DamagedInternal circuit validation resistor PTB damagederror
867 (50)Temperature Module MalfunctionInternal circuit validation resistor PTB damagederror
867 (51)Temperature Module MalfunctionA/D measurement timeouterror
867 (51)Temp Module Malfunction - A/D Measurement TimeoutA/D converter measurement timed outerror
867 (59)Temp Module Malfunction - Voltage Below MaximumBattery or power supply voltage below maximumerror
867 (59)Temperature Module MalfunctionBattery or power supply voltage below maximumerror
867 (60)Temperature Module MalfunctionBattery or power supply voltage exceeds minimumerror
867 (60)Temp Module Malfunction - Voltage Exceeds MinimumBattery or power supply voltage exceeds minimumerror
867 (61)Temp Module Malfunction - Reference Voltage UnstableReference voltage circuit under voltage or unstableerror
867 (61)Temperature Module MalfunctionReference voltage circuit under voltage or unstableerror
867 (62)Temperature Module MalfunctionModule not calibratederror
867 (62)Temp Module Malfunction - Not CalibratedTemperature module not calibratederror
867 (65)Temp Module Malfunction - EEPROM Write ErrorError saving to module's EEPROMerror
867 (65)Temp Module Malfunc (65)Error saving to the module's EEPROMerror
867 (66)Temp Module Malfunc (66)Module's non-volatile calibration memory error detection mechanism detected an errorerror
867 (66)Temp Module Malfunction - Non-Volatile Calibration ErrorNon-volatile calibration memory error detectederror
867 (67)Temp Module Malfunc (67)Error reading/writing to the module's EEPROMerror
867 (67)Temp Module Malfunc (67)Error reading/writing to the module's EEPROMerror
867 (68)Temp Module Malfunc (68)Error reading/writing to the module's EEPROMerror
867 (68)Temp Module Malfunc (68)Error reading/writing to the module's EEPROMerror
867 (69)Temp Module Malfunc (69)Error reading/writing to the module's EEPROMerror
867 (69)Temp Module Malfunc (69)Error reading/writing to the module's EEPROMerror
867 (74)Temp Module Malfunc (74)Internal error during module EEPROM initializationerror
867 (74)Temp Module Malfunc (74)Internal error during EEPROM initializationerror
867 (80)Temp Module Malfunc (80)Software trying to turn off probe heater but feedback signal shows still onerror
867 (81)Temp Module Malfunc (81)Software trying to turn on probe heater but feedback signal shows still offerror
867 (82)Temp Module Malfunc (82)Signal HTR_Q on, HTRC off, but voltage still presenterror
867 (82)Temp Module Malfunc (82)Signal HTR_Q is on and signal HTRC is off, but still have voltageerror
867 (83)Temp Module Malfunc (83)Signal HTR_Q is tri-stated with signal HTRC enabled, and has heater powererror
867 (83)Temp Module Malfunc (83)Signal HTR_Q tri-stated with HTRC enabled and heater power presenterror
867 (84)Temp Module Malfunc (84)Signal Q&C on with insufficient heater voltageerror
867 (84)Temp Module Malfunc (84)Signal Q&C is turned on and the heater voltage is not high enougherror
867 (85)Temp Module Malfunc (85)Heater hardware fail-safe should have turned off, but did noterror
867 (85)Temp Module Malfunc (85)Heater fail-safe did not turn off as expectederror
868Temp Error. The front end will now reset itself.Temperature measurement failure or module communication errorerror
868Temperature errorNo data from temperature moduleerror
868Temp Error. The front end will now reset itself.No data from temperature sensorerror
881Speaker Malfunc (881)The speaker failederror
881Speaker Malfunc (881)Speaker hardware failurewarning
Err3Load beam cable disconnection or defectOne or more load beam cables disconnected or damagederror
Upgrade interrupted - Main board data lostSoftware upgrade process was interrupted; main board memory corrupted and cannot be recoveredUSB flash drive removed, monitor unplugged, or keys pressed during upgrade processcritical
Upgrade Software window displays no valid imageUSB flash drive does not contain valid software upgrade image or image is corruptedAfter inserting USB flash drive and selecting Upgrade Software from System Admin Menu, no software version information appearserror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.