Philips SureSigns VSV: troubleshooting and error codes
The ServiceForge knowledge base holds 48 documented symptoms for the Philips SureSigns VSV, with 112 ranked causes, 95 diagnostic checks with test points and expected values, and 17 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Bedside monitor does not appear at VSV
Bedside monitor not visible or not appearing in VSV display
Ranked causes
- 1. Bedside monitor disconnected from network (common)
- 2. VSV disconnected from network (common)
- 3. Monitor not scanned or joined by VSV (occasional)
- 4. Monitor not placed in VSV layout (occasional)
- 5. LAN board broken or not connected (occasional)
- 6. Router broken (occasional)
- 7. Main board malfunction (occasional)
Bedside monitor network cable
- Expected
- Firmly connected
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot turn on VSV with AC power
VSV does not power on using AC adapter but powers on with battery
Ranked causes
- 1. AC power cord unplugged or broken (common)
- 2. Fuse is blown (common)
- 3. Power supply cable malfunction (occasional)
- 4. Power supply malfunction (occasional)
- 5. AC power connector malfunction (occasional)
- 6. VSV keypad worn (rare)
- 7. Main board malfunction (occasional)
AC power cord connection
- Expected
- Firmly plugged into outlet
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power on but VSV screen is blank
Power indicator on but display screen shows nothing
Ranked causes
- 1. VSV in Deep Sleep mode (common)
- 2. Speaker/fan cables disconnected (common)
- 3. Backlight LCD disconnected (common)
- 4. Backlight inverter failed (occasional)
- 5. LCD failure (occasional)
- 6. Main board malfunction (occasional)
Deep Sleep mode status
- Expected
- Not in Deep Sleep
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard does not work
Keyboard not responding to key presses or input
Ranked causes
- 1. Keyboard disconnected from USB hub (common)
- 2. Keyboard keys jammed or dirty (common)
- 3. Keyboard broken (occasional)
- 4. LAN board broken or not connected (occasional)
- 5. Main board malfunction (occasional)
USB keyboard connection
- Test point
- USB hub port
- Expected
- Firmly connected
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Mouse does not work
Mouse not responding or not tracking properly
Ranked causes
- 1. Mouse disconnected from USB (common)
- 2. Mouse optical sensor blocked (common)
- 3. Mouse broken (occasional)
- 4. LAN board broken or not connected (occasional)
- 5. Main board malfunction (occasional)
USB mouse connection
- Test point
- USB hub port
- Expected
- Firmly connected
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB hub does not work
USB hub not providing power or connectivity to USB devices
Ranked causes
- 1. USB hub disconnected from VSV (common)
- 2. USB hub ports blocked by dust (common)
- 3. USB hub broken (occasional)
- 4. LAN board broken or not connected (occasional)
- 5. Main board malfunction (occasional)
USB hub connection to VSV
- Test point
- VSV USB port
- Expected
- Firmly connected
1 more check with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot turn on VSV with battery power
VSV powers on with AC power but not with battery power
Ranked causes
- 1. Battery discharged (common)
- 2. Battery or connector loose (common)
- 3. Battery not connected to main board (occasional)
- 4. Main board malfunction (occasional)
Battery charge level
- Test point
- Battery
- Expected
- Charged
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External display does not work
External SVGA display not showing information from VSV or not powered on
Ranked causes
- 1. External display not turned on (common)
- 2. VGA cable disconnected from VSV (common)
- 3. External display not connected to AC power (common)
- 4. External display broken (occasional)
External display power switch
- Expected
- Display powered on
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not print
Recorder does not produce any printed output
Ranked causes
- 1. No paper or recorder door open (common)
- 2. System does not recognize recorder (common)
- 3. Recorder module malfunction (occasional)
- 4. Main board malfunction (occasional)
Paper supply level
- Expected
- Paper installed
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder paper jamming
Recorder paper fails to advance or jams during operation
Ranked causes
- 1. Paper not installed correctly (common)
- 2. Environmental conditions (temp/humidity) (occasional)
- 3. Improper recorder paper used (occasional)
- 4. Mechanical problem in recorder (occasional)
Paper installation
- Expected
- Paper correctly installed and threaded
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV does not export data to external server
Data export to external server/source fails or is not working
Ranked causes
- 1. Data Export disabled (common)
- 2. Server IP or Port incorrect (common)
- 3. LAN board broken or not connected (occasional)
- 4. Main board malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not sound caution
System does not produce audible alarm sound
Ranked causes
- 1. Bad connection to speaker (common)
- 2. Defective speaker (common)
- 3. Main board malfunction (occasional)
Speaker connection
- Test point
- Main PCB speaker connector
- Expected
- Firmly connected
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery damaged, leaking, or improperly functioning patient safety
Battery shows visible damage, leakage, low charge, or error codes preventing normal operation
Ranked causes
- 1. Battery cell damage or internal short circuit (occasional)
- 2. Battery not fully charged or low battery voltage during use (common)
- 3. Service error code indicates battery communication or health failure (occasional)
Inspect battery for visible damage or leakage
- Test point
- Battery exterior surface
- Expected
- No cracks, leaks, or swelling
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Random distorted graphics with white background
Display shows garbled or distorted graphics with white background
Ranked causes
- 1. LCD signal cable not attached correctly (common)
- 2. LCD has failed (occasional)
- 3. Main board malfunction (occasional)
LCD signal cable connection
- Test point
- LCD signal cable connector
- Expected
- Firmly seated
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade interrupted by user action caution
Upgrade fails if user unplugs VSV, removes USB drive, or presses keys during upgrade process.
Ranked causes
- 1. VSV unplugged during upgrade (occasional)
- 2. USB flash drive removed during upgrade (occasional)
- 3. Keys pressed during upgrade (occasional)
Monitor upgrade progress without interruption
- Test point
- Upgrade in Progress indicator and status messages
- Expected
- Indicator increments; messages 'Checking Memory CRC' and 'Upgrade Successful' appear
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV cannot turn on with AC or battery power
VSV completely unresponsive to power on attempts via AC or battery
Ranked causes
- 1. AC fuse failure (common)
- 2. No voltage on main PCB (common)
- 3. LCD/cables/battery/power supply malfunction (occasional)
AC power connection
- Expected
- Firmly connected
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV does not receive power
Power indicator is off, no display, unit unresponsive
Ranked causes
- 1. Power turned off (common)
- 2. Battery inadequately charged (common)
- 3. AC power cord not connected or plugged in (common)
Verify power switch is ON
- Test point
- VSV power switch
- Expected
- Power switch in ON position
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV unresponsive to navigation or buttons
Navigation wheel or front panel buttons do not respond to user input
Ranked causes
- 1. Navigation wheel disconnected (common)
- 2. Keypad or navigation wheel broken (common)
- 3. Main board malfunction (occasional)
Navigation wheel connector seating
- Test point
- Navigation wheel connectors on main board
- Expected
- Fully seated
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External display has static or noise
External display shows interference static or noise in image
Ranked causes
- 1. Electromagnetic interference on display (common)
- 2. External display broken (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External display not showing VSV information
External display monitor not receiving video output or displaying different content than VSV screen
Ranked causes
- 1. External display not powered on or video cable disconnected (common)
- 2. Video output port defective or display incompatible (occasional)
Verify external display is powered on
- Test point
- External display power indicator
- Expected
- Display powered and receiving video input
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Router LEDs do not light
Router 'System' and/or port LEDs not illuminated
Ranked causes
- 1. Router not connected to AC power (common)
- 2. Router broken (occasional)
Router AC power connection
- Expected
- Plugged into outlet and powered on
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Router power and network connectivity not normal
Router System light not illuminated or port LEDs do not light for connected VSV or bed network connections
Ranked causes
- 1. Router power off or not receiving power (common)
- 2. Network cable disconnected from router port (common)
Verify router System light is illuminated
- Test point
- Router System light LED
- Expected
- System light on (green or as specified)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB hub, keyboard, or mouse not operating normally
Mouse movement does not move cursor; scroll wheel does not work; keyboard arrow keys do not navigate; System icon click does not open menu
Ranked causes
- 1. USB peripheral not powered or not properly connected (common)
- 2. USB hub or peripheral hardware failure (occasional)
Verify VSV system is powered on
- Test point
- VSV power indicator
- Expected
- VSV powered on and displaying normal screen
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV cannot find valid software image on USB drive
Upgrade Software menu fails to detect software image; screen returns to System Admin Menu.
Ranked causes
- 1. Invalid or corrupted software image on USB drive (occasional)
- 2. Upgrade software accessed through USB hub instead of rear port (occasional)
Verify software image validity
- Test point
- USB flash drive contents
- Expected
- Valid software upgrade file present on drive
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bedside monitors display static or noise
One or more bedside monitors show distorted waveforms with static/noise
Ranked causes
- 1. Electromagnetic interference on bedside monitors (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clear patient data and user settings
Reset system to clear all memory, patient data, patient list entries, viewing pane layout, and monitor connections; resets all user settings to factory defaults
Ranked causes
- 1. User requests system reset to factory defaults (unknown)
Access Maintenance functions in System Diagnostics menu
- Test point
- System Diagnostics menu
- Expected
- Maintenance functions accessible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display test patterns not showing correctly
Display test sequence does not draw expected patterns or patterns do not display for correct duration on main screen or external display
Ranked causes
- 1. Display hardware malfunction or connection issue (unknown)
Access System Diagnostics Menu
- Test point
- Main screen menu
- Expected
- System Diagnostics Menu displays
8 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Downgrade clears all patient data and resets settings patient safety
Performing a software downgrade erases all persisted memory including patient list, settings, and logs.
Ranked causes
- 1. Software downgrade operation initiated (common)
Confirm software version on USB drive is lower than current VSV version
- Test point
- Upgrade Software menu showing Current Version and New Version
- Expected
- New Version number is lower than Current Version
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External display dark, dim, or distorted
External display has power but image is dark, characters dim, or distorted
Ranked causes
- 1. Display controls set improperly (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
External display does not shut off
External display power button does not turn off the display
Ranked causes
- 1. Power button not held long enough (common)
Power button press duration
- Expected
- Held for 5+ seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fuse value incorrect after replacement high voltage
Power connector fuse installed is wrong value or rating after fuse replacement procedure
Ranked causes
- 1. Incorrect fuse type installed during replacement (common)
Verify power connector fuse value
- Test point
- Fuse in power connector
- Expected
- Correct fuse value as per equipment manual specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Labeling or markings not legible patient safety
Equipment markings, labels, or safety labels worn, faded, illegible, or missing
Ranked causes
- 1. Wear, fading, or damage to labels and markings (occasional)
Inspect all equipment labels and markings
- Test point
- Equipment exterior, connectors, control panels
- Expected
- All labels legible; markings clear and complete
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Main screen fixed key not responding
Button press does not illuminate button icon in test window with grey border for five seconds
Ranked causes
- 1. Main screen fixed key malfunction (unknown)
Press main screen fixed key during Button Test
- Test point
- main screen button icon in test window
- Expected
- button icon lights with grey border for five seconds
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network monitoring suspended during maintenance access patient safety
Bedside monitor alarms are not reported at the VSV when System Diagnostics menu is open
Ranked causes
- 1. System Diagnostics menu opened (common)
Verify network monitoring status
- Test point
- System Diagnostics menu display
- Expected
- Network Monitoring Suspended message visible when menu is open
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call relay not operating correctly
Nurse call output relay tip and ring contacts do not switch as expected during alarm
Ranked causes
- 1. Relay electrical malfunction or contact failure (unknown)
Connect patient simulator to bedside monitor linked to VSV
- Test point
- Patient simulator and bedside monitor interface
- Expected
- Monitor connected to VSV via network
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reset error log counter
Reset the counter tracking the number of errors in the error log.
Ranked causes
- 1. Error log maintenance (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reset LCD hour counter
Reset the counter tracking the number of hours the LCD display has been on.
Ranked causes
- 1. LCD replacement maintenance (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reset serial number
Main PCB replacement requires re-entering VSV serial number from label on back panel; causes unit reboot
Ranked causes
- 1. Main PCB replaced (common)
Locate serial number label on back of VSV
- Test point
- Back panel label marked SN
- Expected
- Serial number visible on label
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Rotary wheel not responding
Wheel rotation does not draw red dots in test window in direction turned
Ranked causes
- 1. Rotary wheel malfunction (unknown)
Turn rotary wheel clockwise or counterclockwise during Button Test
- Test point
- wheel icon in test window
- Expected
- wheel icon draws red dots in the direction that you turn the wheel
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade fails or VSV cannot boot after power loss
If AC power is interrupted during software upgrade, the VSV cannot boot and becomes inoperable.
Ranked causes
- 1. Power loss during software upgrade process (occasional)
Verify whether power was lost during upgrade progress
- Test point
- VSV display and boot status
- Expected
- VSV boots normally or displays upgrade in progress indicator
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
System diagnostics needed for unknown issue
Troubleshooting table or error code table did not identify the problem; run system diagnostic software
Ranked causes
- 1. Unknown issue requires system diagnostic testing (unknown)
Run Display Test to verify LCD functionality
- Test point
- System Diagnostics menu
- Expected
- Display test executes and shows result
6 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not function after internal parts replacement
VSV fails to operate or shows errors following internal component replacement
Ranked causes
- 1. Incorrect reassembly after parts replacement (common)
Power-on test after internal parts replacement
- Test point
- VSV startup
- Expected
- Unit powers on, display shows normal boot sequence
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Visual system damage or obstruction caution
Obvious signs of physical damage such as cracks, cuts, or breakage; obstructed mechanical parts
Ranked causes
- 1. Physical impact or mishandling (unknown)
Inspect system exterior for physical damage
- Test point
- External casing and enclosure
- Expected
- No cracks, cuts, breakage, or obstructions visible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV display is incorrect or absent
Display blank, garbled, or not functioning while power is present
Ranked causes
- 1. Display malfunction or connection issue (unknown)
Verify power and display operation
- Test point
- VSV power indicator and display screen
- Expected
- Power LED on and display shows output
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV does not boot correctly
System fails to display expected boot sequence indicators; does not reach main operating screen
Ranked causes
- 1. +3.3V system board supply voltage not working properly (common)
Plug AC power and/or fully charged battery into unpowered VSV; press Power button
- Test point
- Power input, Power button
- Expected
- Screen displays color bars for ~5 seconds
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV hangs during startup
Opening screen/color bars appear but then hang, VSV does not complete boot
Ranked causes
- 1. Main board defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV reboots after serial number reset
Device automatically restarts when Reset S/N function is executed
Ranked causes
- 1. Serial number reset initiated via Maintenance functions (common)
Monitor reboot completion
- Test point
- VSV display and system status
- Expected
- Device restarts and returns to normal operating state
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
VSV self test fails or shows errors
Memory test detects errors during VSV resident self test; test window displays FAILED in red for failed items
Ranked causes
- 1. Memory error detected (unknown)
Open System Diagnostics Menu
- Test point
- System menu access
- Expected
- Diagnostics menu opens
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1025 | System Error - Could not load ResourceDLL.dll | Firmware unable to locate or load required DLL resource file | error |
1026 | System Error - Could not load data from flash | Flash memory read failure or corrupted data structure | error |
1029 | System Error - Could not allocate enough memory | Dynamic memory allocation failure; insufficient heap memory | error |
1031 | System Error - Monitoring and watchdog thread could not initialize | Operating system thread creation failure for monitor/watchdog task | error |
1032 | System Error - Could not open audio files | Firmware unable to access audio alarm files from storage | error |
1033 | System Error - Unspecified error | General fault with unknown root cause detected by firmware | error |
1034 | System Error - Failed load product info | Firmware unable to read product information from storage | error |
1036 | System Error - Failed to start work thread | Operating system cannot create worker thread | error |
1042 | System Error - Failed to start up network driver | Network interface driver initialization failure | error |
1043 | System Error - Software problem (thread reference number provided) | Software exception in specified thread | error |
1045 | System Error | Unspecified system error condition detected | error |
1050 | System Error - Failed to stop work thread | Operating system cannot terminate worker thread gracefully | error |
1052 | System Error - Heap Corrupt | Dynamic memory heap data structure corrupted | critical |
257 | System Error | Unspecified system error condition detected | error |
257 | System Error - Battery charger power failure | Battery charger unable to deliver power or charger circuit malfunction | error |
261 | System Error - Battery overcharged | Battery charging voltage or current exceeds limits | error |
264 | System Error - Battery over temperature | Battery temperature exceeds safe operating range during charging or use | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.