Philips SureSigns VS2+: troubleshooting and error codes
The ServiceForge knowledge base holds 39 documented symptoms for the Philips SureSigns VS2+, with 106 ranked causes, 101 diagnostic checks with test points and expected values, and 122 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Monitor does not power up with AC power high voltage
Monitor powers on with battery but not with AC power supply connected
Ranked causes
- 1. Power cord is unplugged (common)
- 2. Power cord is broken (occasional)
- 3. Fuse is blown (common)
- 4. Power supply cable malfunctioned (occasional)
- 5. Power supply malfunctioned (occasional)
- 6. AC power connector malfunctioned (occasional)
- 7. Monitor keypad is worn (rare)
- 8. Main board malfunctioned (rare)
Verify power cord connection
- Test point
- AC outlet
- Expected
- Power cord plugged in
5 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurements are not displayed patient safety
Temperature module fails to display or output measurement readings
Ranked causes
- 1. Probe is misplaced (common)
- 2. Probe cable is not connected (common)
- 3. Monitor has not completed start-up sequence (common)
- 4. Probe malfunctioned (occasional)
- 5. Temperature is over range (occasional)
- 6. Temperature module malfunctioned (rare)
- 7. Main board malfunctioned (rare)
Verify probe placement at measurement site
- Test point
- Temperature probe location
- Expected
- Probe at proper anatomical site (oral, axillary, rectal per protocol)
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurements are not displayed patient safety
Blood pressure module does not output or display measurement readings
Ranked causes
- 1. Wrong cuff size or incorrect cuff placement (common)
- 2. Tube is bent/twisted or air leakage in tube/cuff (common)
- 3. Problem external to monitor occurred (occasional)
- 4. Initial inflation pressure setting is incorrect (occasional)
- 5. NBP module error occurred (rare)
- 6. Main board malfunctioned (rare)
Verify correct cuff size selected
- Test point
- Cuff size selector or patient configuration
- Expected
- Cuff size matches patient arm circumference per manual
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not print
Optional recorder unit installed and selected but produces no output on paper
Ranked causes
- 1. Optional recorder is not installed (common)
- 2. Recorder is out of paper (common)
- 3. Recorder door is open (common)
- 4. System does not recognize recorder (occasional)
- 5. Recorder module malfunctioned (occasional)
- 6. Main board malfunctioned (rare)
Verify recorder module is physically installed
- Test point
- Recorder module slot or connector
- Expected
- Recorder module inserted and seated
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call function does not signal alarm conditions patient safety
Nurse call system does not activate or transmit alarm status to call station
Ranked causes
- 1. Minimum Nurse Call Alarm Priority is set too high (common)
- 2. Alarms have been silenced (common)
- 3. Phono connector is loose or broken (occasional)
- 4. Communications board malfunctioned (occasional)
- 5. Main board malfunctioned (rare)
Verify Minimum Nurse Call Alarm Priority setting
- Test point
- Monitor configuration menu
- Expected
- Priority threshold set to appropriate level for facility
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not sound patient safety
Unit does not produce audio output when alarm conditions are triggered
Ranked causes
- 1. Monitor is in Audio Pause or Audio Off mode (common)
- 2. Connection to speaker is faulty (common)
- 3. Speaker is faulty (occasional)
- 4. Main board malfunctioned (rare)
Verify audio mode setting
- Test point
- Monitor audio control menu
- Expected
- Audio mode set to normal (not Audio Off or Audio Pause)
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not respond to navigation wheel or keys
Front panel controls unresponsive; navigation wheel or buttons do not trigger expected functions
Ranked causes
- 1. Main board has faulty connection (common)
- 2. Monitor keypad or navigation wheel is broken (common)
- 3. Main board malfunctioned (rare)
- 4. Membrane switch assembly malfunctioned (rare)
Inspect all main board connector seats
- Test point
- Main board connectors for controls and wheel
- Expected
- All connectors fully inserted and locked
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not turn on with AC or battery power high voltage
Monitor completely non-responsive to power, no indicators or display activity
Ranked causes
- 1. Power cord unplugged and battery not installed or charged (common)
- 2. Cables are disconnected (common)
- 3. Main board malfunctioned (rare)
- 4. Membrane switch assembly malfunctioned (rare)
Verify power cord is plugged in
- Test point
- AC outlet and monitor
- Expected
- Power cord firmly connected
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor turns on with AC power but not battery power
Monitor powers on with AC power but fails to power on when using battery only
Ranked causes
- 1. Battery is not charged (common)
- 2. Battery connector is loose (common)
- 3. Battery not connected to main board (occasional)
- 4. Main board malfunctioned (rare)
Check battery charge level
- Test point
- Battery status indicator
- Expected
- Battery fully charged
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP cuff does not inflate patient safety
Blood pressure measurement cuff fails to inflate during NBP measurement attempt
Ranked causes
- 1. Tube or cuff is bent or twisted (common)
- 2. Tube or cuff is leaking air (common)
- 3. NBP module malfunctioned (occasional)
- 4. Main board malfunctioned (rare)
Inspect NBP tubing path
- Test point
- Cuff tube entire length
- Expected
- Tube straight and unkinked
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurements are unreliable patient safety
Blood pressure readings are inconsistent, erratic, or outside expected range
Ranked causes
- 1. Wrong cuff size or incorrect cuff placement (common)
- 2. External problem occurred (common)
- 3. Initial inflation pressure setting is incorrect (occasional)
- 4. Calibration is unreliable (occasional)
Verify cuff size matches patient
- Test point
- Patient arm circumference vs cuff size
- Expected
- Correct cuff size for patient
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurements are not displayed patient safety
Pulse oximetry module fails to display or output saturation readings
Ranked causes
- 1. SpO2 module is in learning phase (common)
- 2. SpO2 sensor malfunctioned (common)
- 3. SpO2 board malfunctioned (occasional)
- 4. Main board malfunctioned (rare)
Verify SpO2 module learning phase status
- Test point
- Module startup status indicator
- Expected
- Learning phase complete; module ready
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB hub does not work
USB hub connected to monitor is non-responsive; external USB devices not recognized or accessible
Ranked causes
- 1. USB hub is disconnected from monitor (common)
- 2. USB hub ports are blocked by dust or dirt (common)
- 3. USB hub is faulty (occasional)
- 4. Communications board is broken or not properly connected (occasional)
Verify USB hub physical connection to monitor
- Test point
- USB connector at monitor and hub
- Expected
- USB plugs firmly inserted in ports
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AUDIO FAILED error message displayed caution
Flashing red/white box with AUDIO FAILED message appears on screen
Ranked causes
- 1. Speaker malfunctioned (occasional)
- 2. Connection to speaker is faulty (common)
- 3. Speaker is faulty (occasional)
Verify speaker cable connection
- Test point
- Speaker connectors
- Expected
- Cable firmly connected to speaker and audio circuit
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displays random or distorted graphics with white background
Display shows garbled or corrupt image with white background color
Ranked causes
- 1. LCD cable is not attached correctly (common)
- 2. LCD malfunctioned (occasional)
- 3. Main board malfunctioned (rare)
Inspect and reseat LCD signal cable
- Test point
- LCD cable connectors at main board and display
- Expected
- Cable firmly seated in both connectors
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor screen blank with power on
Power is on and unit appears functional but display shows no image
Ranked causes
- 1. Monitor is in Standby mode (common)
- 2. Main board malfunctioned (rare)
- 3. LCD is faulty (occasional)
Check Standby mode status
- Test point
- Monitor power state
- Expected
- Monitor not in Standby
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder paper is jamming
Paper in optional recorder becomes stuck, preventing print operation
Ranked causes
- 1. Paper is not installed correctly (common)
- 2. Wrong type of paper is installed (common)
- 3. Recorder has mechanical problem (occasional)
Verify paper installation procedure
- Test point
- Recorder paper path and guides
- Expected
- Paper properly threaded and seated in paper holder
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Temperature measurements are unreliable patient safety
Temperature readings are inconsistent, erratic, or suspect
Ranked causes
- 1. Probe malfunctioned (common)
- 2. Module is not calibrated (common)
- 3. Temperature module malfunctioned (occasional)
Replace temperature probe
- Test point
- Temperature measurement
- Expected
- Consistent readings from probe
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on
Monitor shows no power indicators, display does not function
Ranked causes
- 1. Power cord not connected or AC outlet inactive (common)
- 2. Power switch not activated (common)
- 3. Battery not adequately charged (occasional)
Verify AC outlet power
- Test point
- AC outlet
- Expected
- Power available
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery service required message displayed patient safety
Monitor displays 'Battery service required' message on screen
Ranked causes
- 1. Max Error value exceeds 8% (common)
- 2. Full Capacity value is half of Design Capacity value (common)
Check battery error values
- Test point
- Battery diagnostics
- Expected
- Max Error less than 8%, Full Capacity at least 50% of Design Capacity
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on
Unit not responding to power button, LEDs not lit, no display output when AC power cord or battery connected
Ranked causes
- 1. AC power not supplied (common)
- 2. Battery inadequately charged (common)
Verify power supply
- Test point
- AC outlet and power cord connection
- Expected
- Power cord seated in monitor; outlet providing power
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP initial pressure too high patient safety
Blood pressure cuff reaches excessive initial inflation pressure during measurement cycle
Ranked causes
- 1. Cuff is too tight (common)
- 2. Initial inflation pressure setting is incorrect (common)
Inspect cuff fit on patient arm
- Test point
- Cuff tightness and position
- Expected
- Cuff snug but not cutting off circulation; one finger fits under cuff
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 readings are unreliable patient safety
Pulse oximetry measurements are inconsistent, erratic, or missing data
Ranked causes
- 1. Problem external to monitor occurred (common)
- 2. Perfusion is low (occasional)
Verify SpO2 sensor is compatible and correctly placed
- Test point
- Sensor type and patient application site
- Expected
- Compatible sensor properly attached with good perfusion
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio test or speaker verification needed
Operator or technician needs to verify speaker function by listening for tones at different decibel levels
Ranked causes
- 1. Routine speaker verification procedure (common)
Listen for audio test tones
- Test point
- Monitor speaker output
- Expected
- Three short tones audible at different Db levels
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance verification testing patient safety
Post-installation, post-repair and post-maintenance verification procedures to ensure monitor meets performance specifications
Ranked causes
- 1. Monitor requires verification after installation, repair, or software upgrade (common)
Perform Power-on self test
- Test point
- Monitor startup
- Expected
- Self-test completes without error
5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pneumatic leakage test
Check pneumatic system and valve performance by measuring pressure drop over time
Ranked causes
- 1. Pneumatic system leakage exceeds acceptable level (unknown)
Select Start Static Pressure Test in NBP Test menu
- Test point
- NBP Test menu
- Expected
- Static Pressure Test started
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Reset tracked parameters
Monitor tracks LCD usage hours, NBP cycle count, and error count. Reset procedure required after replacing associated parts.
Ranked causes
- 1. Parameter reset required after part replacement (common)
Access Maintenance options
- Test point
- Monitor menu
- Expected
- Maintenance menu appears
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Temperature measurement
Verify accuracy of temperature probe and measurement system using calibration key
Ranked causes
- 1. Temperature measurement accuracy needs verification (unknown)
Connect temperature probe to monitor
- Test point
- Temperature probe connector
- Expected
- Probe connected
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charging LED does not function as expected
LED Test initiated but Charging LED does not cycle through yellow/green states or flash patterns as specified
Ranked causes
- 1. Charging LED hardware failure or driver circuit fault (common)
Run LED Test in System Diagnostics and observe Charging LED behavior
- Test point
- Charging LED on monitor front panel; AC Power LED
- Expected
- Yellow 5 sec → Green 5 sec → Yellow flash 5 sec → Green flash 5 sec; AC Power LED stays green
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display not functioning or incorrect
LCD screen shows no output, blank display, or unexpected content after power-on sequence
Ranked causes
- 1. Display hardware failure (unknown)
Verify monitor power and display status
- Test point
- LCD screen
- Expected
- Display functional with correct content
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Front panel LEDs not lit as expected
Indicator lights on front panel not illuminated or lit incorrectly during operation
Ranked causes
- 1. Delayed LED response after power state change (occasional)
Observe AC Power LED timing
- Test point
- Front panel AC Power LED
- Expected
- LED response within several seconds of power cord connection or disconnection
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Key is collapsed or does not click
Front panel key physically damaged, depressed, or does not provide tactile feedback when pressed
Ranked causes
- 1. Membrane switch assembly malfunctioned (common)
Inspect physical condition of all front panel keys
- Test point
- Each key dome and button face
- Expected
- All keys firm and provide click when pressed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor turns on, opening screen starts, then freezes
Unit powers on and boots partially but display freezes during startup sequence
Ranked causes
- 1. Main board malfunctioned (common)
Attempt to reset monitor
- Test point
- Power button
- Expected
- System completes startup sequence
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No display or incorrect display
LCD does not show expected startup sequence or display is blank/garbled after power applied
Ranked causes
- 1. LCD failure or connection issue (unknown)
Verify front panel LEDs
- Test point
- Front panel LEDs
- Expected
- LEDs lit as expected during startup
Nurse call relay not responding to alarms
Nurse call alarm output relay (phone jack connector) fails to operate correctly during alarm conditions
Ranked causes
- 1. Relay contacts not closing/opening as expected (unknown)
Measure continuity at Tip contact (Relay Normally Open) during alarm condition
- Test point
- Nurse Call connector Tip
- Expected
- Closed
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Only one Battery Charging LED color displayed
Battery charging indicator shows only one color instead of expected color sequence
Ranked causes
- 1. Membrane switch assembly malfunctioned (common)
Observe LED color display during charging
- Test point
- Battery charging LED
- Expected
- Color sequence should change during charge cycle
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pressing key does not access expected function caution
Front panel keys trigger incorrect functions; for example Print key initiates NBP measurement instead
Ranked causes
- 1. Membrane switch assembly malfunctioned (common)
Verify key function mapping is correct
- Test point
- Each front panel key
- Expected
- Each key triggers its labeled function
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace fuse in AC power connector high voltage
Fuse replacement procedure for AC power connector on rear of monitor
Ranked causes
- 1. Fuse failure (unknown)
Remove AC power cord from rear of monitor
- Test point
- AC power connector rear
- Expected
- Power cord disconnected
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Self test fails to detect module or reports error patient safety
Running the Self Test detects missing module or test error occurs; affects SpO2, NBP, or Predictive Temperature modules
Ranked causes
- 1. Module hardware not connected or malfunctioning (unknown)
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1025 | System Error | Could not load ResourceDLL.dll | error |
1026 | System Error | Could not load data from flash memory | error |
1027 | System Error | Could not start front end | error |
1028 | System Error | Could not stop front end | error |
1029 | System Error | Could not allocate enough memory | error |
1030 | System Error | Error shutting down front end devices | error |
1031 | System Error | Monitoring and watchdog thread failed to initialize | error |
1032 | System Error | Could not open audio files | error |
1033 | System Error | Unspecified system error | error |
1034 | System Error | Failed to load product information | error |
1035 | System Error | Cannot initialize front end manager | error |
1036 | System Error | Failed to start work thread | error |
1037 | System Error | Cannot start polling thread | error |
1042 | System Error | Failed to start network driver | error |
1043 | System Error | Front end software error | error |
1044 | System Error | Front end software error | error |
1045 | System Error | Printing software error | error |
1047 | System Error | Monitoring software error | error |
1048 | System Error | Exception in alarming software | error |
1050 | System Error | Failed to stop work thread | error |
1051 | System Error | Cannot initialize NBP | error |
1052 | System Error | Memory error detected | error |
1053 | System Error | Loss of monitoring | info |
257 | System Error | Battery charger power failure | error |
261 | System Error | Battery is overcharged | warning |
264 temp value | System Error | Battery over temperature | warning |
817 | FPGA FIFO Overrun | NBP software error | info |
818 (101) | NBP Equip Malfunc (101) | Task overrun | error |
818 (102) | NBP Equip Malfunc (102) | Interrupt stack overwritten | error |
818 (103) | NBP Equip Malfunc (103) | User stack overwritten | error |
818 (104) | NBP Equip Malfunc (104) | Data validity check failed | error |
818 (106) | NBP Equip Malfunc (106) | Unable to write init flag data to non-volatile memory | error |
818 (107) | NBP Equip Malfunc (107) | Unable to verify init flag data written to non-volatile memory | error |
818 (109) | NBP Equip Malfunc (109) | Unable to write NBP configuration data to non-volatile memory | error |
818 (110) | NBP Equip Malfunc (110) | Unable to write NBP configuration data to non-volatile memory | error |
818 (111) | NBP Equip Malfunc (111) | Calibration data checksum error | error |
818 (112) | NBP Equip Malfunc (112) | Unable to write calibration data to non-volatile memory | error |
818 (113) | NBP Equip Malfunc (113) | Unable to verify calibration data written to non-volatile memory | error |
818 (114) | NBP Equip Malfunc (114) | Error log checksum error | error |
818 (115) | NBP Equip Malfunc (115) | Unable to write error log to non-volatile memory | error |
818 (116) | NBP Equip Malfunc (116) | Unable to verify error log data written to non-volatile memory | error |
818 (117) | NBP Equip Malfunc (117) | Inflation cycle count and runtime data checksum error | error |
818 (118) | NBP Equip Malfunc (118) | Unable to write inflation cycle count and runtime data to non-volatile memory | error |
818 (119) | NBP Equip Malfunc (119) | Unable to verify inflation cycle count and runtime data written to non-volatile memory | error |
818 (120) | NBP Equip Malfunc (120) | New message queuing before previous message finished | error |
818 (121) | NBP Equip Malfunc (121) | Pressure value from second pressure sensor not available | error |
818 (122) | NBP Equip Malfunc (122) | 5-volt supply voltage value is too high or low | error |
818 (123) | NBP Equip Malfunc (123) | Error from 5-volt regulator | error |
818 (124) | NBP Equip Malfunc (124) | 12-volt supply voltage value is too high or low | error |
818 (128) | NBP Equip Malfunc (128) | Offset of pressure sensor 1 is out of range | error |
818 (129) | NBP Equip Malfunc (129) | Offset of pressure sensor 2 is out of range | error |
818 (130) | NBP Equip Malfunc (130) | Response of pressure sensor 2 to test signal is out of range | error |
818 (131) | NBP Equip Malfunc (131) | Difference between two pressure channels is too large | error |
818 (132) | NBP Equip Malfunc (132) | One of the two valves was activated for too long | error |
818 (133) | NBP Equip Malfunc (133) | CRC mismatch in calibration data | error |
818 (134) | NBP Equip Malfunc (134) | CRC mismatch in error log data | error |
818 (135) | NBP Equip Malfunc (135) | CRC mismatch in inflation cycle count and runtime data | error |
818 (136) | NBP Equip Malfunc (136) | Pneumatic safety switch is defective | error |
818 (137) | NBP Equip Malfunc (137) | Valve 2 is defective or blocked | error |
818 (141) | NBP Equip Malfunc (141) | Inflation time for selected patient type exceeded | warning |
818 (15) | NBP Equip Malfunc (15) | NBP abnormal unrecoverable system errors with internal firmware | error |
818 (199) | NBP Equip Malfunc (199) | No data from NBP block | error |
818 (2) | NBP Equip Malfunc (2) | NBP power-on self test failed; A/D converter inoperative or pressure transducer offset too large | error |
818 (25) | NBP Equip Malfunc (25) | NBP hardware has malfunctioned because there is no data | error |
819 | NBP Equip Malfunc - Not Calibrated | NBP not calibrated | warning |
819 (127) | NBP Equip Malfunc (127) | NBP module is not calibrated | warning |
819 (138) | NBP Hose Blocked (138) | Problem in pneumatic system | error |
819 (139) | NBP Air Leak (139) | Air leak or valve defect | error |
833 (1) | SpO2 Equip Malfunc (1) | SpO2 ROM checksum error | error |
833 (10) | SpO2 Equip Malfunc (10) | SpO2 host communication error | error |
833 (2) | SpO2 Equip Malfunc (2) | SpO2 RAM test error | error |
833 (20) | SpO2 Equip Malfunc (20) | SpO2 hardware defect | error |
833 (4) | SpO2 Equip Malfunc (4) | SpO2 analog signal path self-test failed | error |
833 (40) | SpO2 Equip Malfunc (40) | SpO2 non-volatile memory corrupted | error |
833 (6) | SpO2 Equip Malfunc (6) | SpO2 self-test timeout | error |
833 (8) | SpO2 Equip Malfunc (8) | SpO2 internal DSP communication error | error |
834 | SpO2 Sensor Malfunc | Malfunction of SpO2 sensor or sensor cable | error |
835 | FPGA FIFO Overrun | SpO2 software error | info |
836 | SpO2 Communication Error | SpO2 communication error | info |
837 | SpO2 Error. Front end will now reset itself | No data from SpO2 | warning |
865 | FPGA FIFO Overrun | Temperature software error | info |
866 (21) | Temp Probe Error (21) | Probe operating above 112F/43.3C; heater too close to thermistor | error |
866 (22) | Temp Probe Error (22) | Excessive heater energy; heater not working or slow response | error |
866 (32) | Temp Probe Error (32) | Thermistor pulled from tip or heater broken | error |
866 (33) | Temp Probe Error (33) | Probe not responsive; not characterized/calibrated | error |
866 (34) | Temp Probe Error (34) | Probe not characterized/calibrated | error |
866 (45) | Temp Probe Error (45) | Measurement below allowable values and far below low ambient limits | warning |
866 (46) | Temp Probe Error (46) | Measurement above allowable values and far above high ambient limits | warning |
866 (52) | Temp Probe Error (52) | Temperature probe not characterized/calibrated | error |
866 (63) | Temp Probe Error (63) | Probe well missing or not installed properly | error |
866 (70) | Temp Probe Error (70) | Problem reading temperature probe EEPROM | error |
866 (75) | Temp Probe Error (75) | Problem reading probe EEPROM or probe not factory tested | error |
866 (78) | Temp Probe Error (78) | Probe error | error |
866 (79) | Temp Probe Error (79) | Module memory error | error |
866 (80) | Temp Probe Error (80) | Software trying to turn off probe heater but feedback says still on | error |
866 (81) | Temp Probe Error (81) | Software trying to turn on probe heater but feedback says still off | error |
867 (24) | Temp Module Malfunc (24) | Ambient temperature too high | warning |
867 (25) | Temp Module Malfunc (25) | Ambient temperature too low | warning |
867 (27) | Temp Module Malfunc (27) | Battery or power supply voltage exceeds maximum value | error |
867 (28) | Temp Module Malfunc (28) | Battery or power supply voltage below minimum value | error |
867 (47) | Temp Module Malfunc (47) | Internal calibration resistor (RCAL) on board is damaged | error |
867 (48) | Temp Module Malfunc (48) | Internal calibration resistor (RCAL) is damaged | error |
867 (49) | Temp Module Malfunc (49) | Internal circuit validation resistor (PTB) is damaged | error |
867 (50) | Temp Module Malfunc (50) | Internal circuit validation resistor (PTB) is damaged | error |
867 (51) | Temp Module Malfunc (51) | A/D measurement timed out; did not finish in allowed time slot | error |
867 (59) | Temp Module Malfunc (59) | Battery or power supply voltage below maximum value | warning |
867 (60) | Temp Module Malfunc (60) | Battery or power supply voltage exceeds minimum value | warning |
867 (61) | Temp Module Malfunc (61) | Reference voltage circuit under voltage or unstable | error |
867 (62) | Temp Module Malfunc (62) | The module is not calibrated | warning |
867 (65) | Temp Module Malfunc (65) | Problem saving to module EEPROM | error |
867 (66) | Temp Module Malfunc (66) | Module non-volatile calibration memory error detection | error |
867 (67) | Temp Module Malfunc (67) | Problem reading/writing to module EEPROM | error |
867 (68) | Temp Module Malfunction | Problem reading/writing to module EEPROM | error |
867 (69) | Temp Module Malfunction | Problem reading/writing to module EEPROM | error |
867 (74) | Temp Module Malfunction | Module initializing EEPROM for first time | warning |
867 (82) | Temp Module Malfunction | HTR_Q signal on but HTRC signal off with voltage present | error |
867 (83) | Temp Module Malfunction | HTR_Q tri-stated with HTRC enabled and heater power | error |
867 (84) | Temp Module Malfunction | Q&C turned on with insufficient heater voltage | error |
867 (85) | Temp Module Malfunction | Heater fail-safe should have turned off but did not | critical |
868 | Temperature Error | No data from temperature sensor | warning |
881 | Speaker Malfunction | Speaker hardware failure | unknown |
882 | Wireless Malfunction | Wireless firmware cannot load | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI