Philips SureSigns VS2+: troubleshooting and error codes

The ServiceForge knowledge base holds 39 documented symptoms for the Philips SureSigns VS2+, with 106 ranked causes, 101 diagnostic checks with test points and expected values, and 122 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Monitor does not power up with AC power high voltage

Monitor powers on with battery but not with AC power supply connected

Ranked causes

  1. 1. Power cord is unplugged (common)
  2. 2. Power cord is broken (occasional)
  3. 3. Fuse is blown (common)
  4. 4. Power supply cable malfunctioned (occasional)
  5. 5. Power supply malfunctioned (occasional)
  6. 6. AC power connector malfunctioned (occasional)
  7. 7. Monitor keypad is worn (rare)
  8. 8. Main board malfunctioned (rare)
First check

Verify power cord connection

Test point
AC outlet
Expected
Power cord plugged in

5 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature measurements are not displayed patient safety

Temperature module fails to display or output measurement readings

Ranked causes

  1. 1. Probe is misplaced (common)
  2. 2. Probe cable is not connected (common)
  3. 3. Monitor has not completed start-up sequence (common)
  4. 4. Probe malfunctioned (occasional)
  5. 5. Temperature is over range (occasional)
  6. 6. Temperature module malfunctioned (rare)
  7. 7. Main board malfunctioned (rare)
First check

Verify probe placement at measurement site

Test point
Temperature probe location
Expected
Probe at proper anatomical site (oral, axillary, rectal per protocol)

3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP measurements are not displayed patient safety

Blood pressure module does not output or display measurement readings

Ranked causes

  1. 1. Wrong cuff size or incorrect cuff placement (common)
  2. 2. Tube is bent/twisted or air leakage in tube/cuff (common)
  3. 3. Problem external to monitor occurred (occasional)
  4. 4. Initial inflation pressure setting is incorrect (occasional)
  5. 5. NBP module error occurred (rare)
  6. 6. Main board malfunctioned (rare)
First check

Verify correct cuff size selected

Test point
Cuff size selector or patient configuration
Expected
Cuff size matches patient arm circumference per manual

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder does not print

Optional recorder unit installed and selected but produces no output on paper

Ranked causes

  1. 1. Optional recorder is not installed (common)
  2. 2. Recorder is out of paper (common)
  3. 3. Recorder door is open (common)
  4. 4. System does not recognize recorder (occasional)
  5. 5. Recorder module malfunctioned (occasional)
  6. 6. Main board malfunctioned (rare)
First check

Verify recorder module is physically installed

Test point
Recorder module slot or connector
Expected
Recorder module inserted and seated

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Nurse call function does not signal alarm conditions patient safety

Nurse call system does not activate or transmit alarm status to call station

Ranked causes

  1. 1. Minimum Nurse Call Alarm Priority is set too high (common)
  2. 2. Alarms have been silenced (common)
  3. 3. Phono connector is loose or broken (occasional)
  4. 4. Communications board malfunctioned (occasional)
  5. 5. Main board malfunctioned (rare)
First check

Verify Minimum Nurse Call Alarm Priority setting

Test point
Monitor configuration menu
Expected
Priority threshold set to appropriate level for facility

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audible alarms do not sound patient safety

Unit does not produce audio output when alarm conditions are triggered

Ranked causes

  1. 1. Monitor is in Audio Pause or Audio Off mode (common)
  2. 2. Connection to speaker is faulty (common)
  3. 3. Speaker is faulty (occasional)
  4. 4. Main board malfunctioned (rare)
First check

Verify audio mode setting

Test point
Monitor audio control menu
Expected
Audio mode set to normal (not Audio Off or Audio Pause)

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not respond to navigation wheel or keys

Front panel controls unresponsive; navigation wheel or buttons do not trigger expected functions

Ranked causes

  1. 1. Main board has faulty connection (common)
  2. 2. Monitor keypad or navigation wheel is broken (common)
  3. 3. Main board malfunctioned (rare)
  4. 4. Membrane switch assembly malfunctioned (rare)
First check

Inspect all main board connector seats

Test point
Main board connectors for controls and wheel
Expected
All connectors fully inserted and locked

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not turn on with AC or battery power high voltage

Monitor completely non-responsive to power, no indicators or display activity

Ranked causes

  1. 1. Power cord unplugged and battery not installed or charged (common)
  2. 2. Cables are disconnected (common)
  3. 3. Main board malfunctioned (rare)
  4. 4. Membrane switch assembly malfunctioned (rare)
First check

Verify power cord is plugged in

Test point
AC outlet and monitor
Expected
Power cord firmly connected

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor turns on with AC power but not battery power

Monitor powers on with AC power but fails to power on when using battery only

Ranked causes

  1. 1. Battery is not charged (common)
  2. 2. Battery connector is loose (common)
  3. 3. Battery not connected to main board (occasional)
  4. 4. Main board malfunctioned (rare)
First check

Check battery charge level

Test point
Battery status indicator
Expected
Battery fully charged

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP cuff does not inflate patient safety

Blood pressure measurement cuff fails to inflate during NBP measurement attempt

Ranked causes

  1. 1. Tube or cuff is bent or twisted (common)
  2. 2. Tube or cuff is leaking air (common)
  3. 3. NBP module malfunctioned (occasional)
  4. 4. Main board malfunctioned (rare)
First check

Inspect NBP tubing path

Test point
Cuff tube entire length
Expected
Tube straight and unkinked

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP measurements are unreliable patient safety

Blood pressure readings are inconsistent, erratic, or outside expected range

Ranked causes

  1. 1. Wrong cuff size or incorrect cuff placement (common)
  2. 2. External problem occurred (common)
  3. 3. Initial inflation pressure setting is incorrect (occasional)
  4. 4. Calibration is unreliable (occasional)
First check

Verify cuff size matches patient

Test point
Patient arm circumference vs cuff size
Expected
Correct cuff size for patient

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 measurements are not displayed patient safety

Pulse oximetry module fails to display or output saturation readings

Ranked causes

  1. 1. SpO2 module is in learning phase (common)
  2. 2. SpO2 sensor malfunctioned (common)
  3. 3. SpO2 board malfunctioned (occasional)
  4. 4. Main board malfunctioned (rare)
First check

Verify SpO2 module learning phase status

Test point
Module startup status indicator
Expected
Learning phase complete; module ready

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

USB hub does not work

USB hub connected to monitor is non-responsive; external USB devices not recognized or accessible

Ranked causes

  1. 1. USB hub is disconnected from monitor (common)
  2. 2. USB hub ports are blocked by dust or dirt (common)
  3. 3. USB hub is faulty (occasional)
  4. 4. Communications board is broken or not properly connected (occasional)
First check

Verify USB hub physical connection to monitor

Test point
USB connector at monitor and hub
Expected
USB plugs firmly inserted in ports

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AUDIO FAILED error message displayed caution

Flashing red/white box with AUDIO FAILED message appears on screen

Ranked causes

  1. 1. Speaker malfunctioned (occasional)
  2. 2. Connection to speaker is faulty (common)
  3. 3. Speaker is faulty (occasional)
First check

Verify speaker cable connection

Test point
Speaker connectors
Expected
Cable firmly connected to speaker and audio circuit

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor displays random or distorted graphics with white background

Display shows garbled or corrupt image with white background color

Ranked causes

  1. 1. LCD cable is not attached correctly (common)
  2. 2. LCD malfunctioned (occasional)
  3. 3. Main board malfunctioned (rare)
First check

Inspect and reseat LCD signal cable

Test point
LCD cable connectors at main board and display
Expected
Cable firmly seated in both connectors

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor screen blank with power on

Power is on and unit appears functional but display shows no image

Ranked causes

  1. 1. Monitor is in Standby mode (common)
  2. 2. Main board malfunctioned (rare)
  3. 3. LCD is faulty (occasional)
First check

Check Standby mode status

Test point
Monitor power state
Expected
Monitor not in Standby

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Recorder paper is jamming

Paper in optional recorder becomes stuck, preventing print operation

Ranked causes

  1. 1. Paper is not installed correctly (common)
  2. 2. Wrong type of paper is installed (common)
  3. 3. Recorder has mechanical problem (occasional)
First check

Verify paper installation procedure

Test point
Recorder paper path and guides
Expected
Paper properly threaded and seated in paper holder

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Temperature measurements are unreliable patient safety

Temperature readings are inconsistent, erratic, or suspect

Ranked causes

  1. 1. Probe malfunctioned (common)
  2. 2. Module is not calibrated (common)
  3. 3. Temperature module malfunctioned (occasional)
First check

Replace temperature probe

Test point
Temperature measurement
Expected
Consistent readings from probe

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on

Monitor shows no power indicators, display does not function

Ranked causes

  1. 1. Power cord not connected or AC outlet inactive (common)
  2. 2. Power switch not activated (common)
  3. 3. Battery not adequately charged (occasional)
First check

Verify AC outlet power

Test point
AC outlet
Expected
Power available

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery service required message displayed patient safety

Monitor displays 'Battery service required' message on screen

Ranked causes

  1. 1. Max Error value exceeds 8% (common)
  2. 2. Full Capacity value is half of Design Capacity value (common)
First check

Check battery error values

Test point
Battery diagnostics
Expected
Max Error less than 8%, Full Capacity at least 50% of Design Capacity

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor does not power on

Unit not responding to power button, LEDs not lit, no display output when AC power cord or battery connected

Ranked causes

  1. 1. AC power not supplied (common)
  2. 2. Battery inadequately charged (common)
First check

Verify power supply

Test point
AC outlet and power cord connection
Expected
Power cord seated in monitor; outlet providing power

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

NBP initial pressure too high patient safety

Blood pressure cuff reaches excessive initial inflation pressure during measurement cycle

Ranked causes

  1. 1. Cuff is too tight (common)
  2. 2. Initial inflation pressure setting is incorrect (common)
First check

Inspect cuff fit on patient arm

Test point
Cuff tightness and position
Expected
Cuff snug but not cutting off circulation; one finger fits under cuff

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpO2 readings are unreliable patient safety

Pulse oximetry measurements are inconsistent, erratic, or missing data

Ranked causes

  1. 1. Problem external to monitor occurred (common)
  2. 2. Perfusion is low (occasional)
First check

Verify SpO2 sensor is compatible and correctly placed

Test point
Sensor type and patient application site
Expected
Compatible sensor properly attached with good perfusion

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio test or speaker verification needed

Operator or technician needs to verify speaker function by listening for tones at different decibel levels

Ranked causes

  1. 1. Routine speaker verification procedure (common)
First check

Listen for audio test tones

Test point
Monitor speaker output
Expected
Three short tones audible at different Db levels

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Performance verification testing patient safety

Post-installation, post-repair and post-maintenance verification procedures to ensure monitor meets performance specifications

Ranked causes

  1. 1. Monitor requires verification after installation, repair, or software upgrade (common)
First check

Perform Power-on self test

Test point
Monitor startup
Expected
Self-test completes without error

5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pneumatic leakage test

Check pneumatic system and valve performance by measuring pressure drop over time

Ranked causes

  1. 1. Pneumatic system leakage exceeds acceptable level (unknown)
First check

Select Start Static Pressure Test in NBP Test menu

Test point
NBP Test menu
Expected
Static Pressure Test started

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Reset tracked parameters

Monitor tracks LCD usage hours, NBP cycle count, and error count. Reset procedure required after replacing associated parts.

Ranked causes

  1. 1. Parameter reset required after part replacement (common)
First check

Access Maintenance options

Test point
Monitor menu
Expected
Maintenance menu appears

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Temperature measurement

Verify accuracy of temperature probe and measurement system using calibration key

Ranked causes

  1. 1. Temperature measurement accuracy needs verification (unknown)
First check

Connect temperature probe to monitor

Test point
Temperature probe connector
Expected
Probe connected

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Charging LED does not function as expected

LED Test initiated but Charging LED does not cycle through yellow/green states or flash patterns as specified

Ranked causes

  1. 1. Charging LED hardware failure or driver circuit fault (common)
First check

Run LED Test in System Diagnostics and observe Charging LED behavior

Test point
Charging LED on monitor front panel; AC Power LED
Expected
Yellow 5 sec → Green 5 sec → Yellow flash 5 sec → Green flash 5 sec; AC Power LED stays green

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display not functioning or incorrect

LCD screen shows no output, blank display, or unexpected content after power-on sequence

Ranked causes

  1. 1. Display hardware failure (unknown)
First check

Verify monitor power and display status

Test point
LCD screen
Expected
Display functional with correct content

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Front panel LEDs not lit as expected

Indicator lights on front panel not illuminated or lit incorrectly during operation

Ranked causes

  1. 1. Delayed LED response after power state change (occasional)
First check

Observe AC Power LED timing

Test point
Front panel AC Power LED
Expected
LED response within several seconds of power cord connection or disconnection

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Key is collapsed or does not click

Front panel key physically damaged, depressed, or does not provide tactile feedback when pressed

Ranked causes

  1. 1. Membrane switch assembly malfunctioned (common)
First check

Inspect physical condition of all front panel keys

Test point
Each key dome and button face
Expected
All keys firm and provide click when pressed

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor turns on, opening screen starts, then freezes

Unit powers on and boots partially but display freezes during startup sequence

Ranked causes

  1. 1. Main board malfunctioned (common)
First check

Attempt to reset monitor

Test point
Power button
Expected
System completes startup sequence

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

No display or incorrect display

LCD does not show expected startup sequence or display is blank/garbled after power applied

Ranked causes

  1. 1. LCD failure or connection issue (unknown)
First check

Verify front panel LEDs

Test point
Front panel LEDs
Expected
LEDs lit as expected during startup

Nurse call relay not responding to alarms

Nurse call alarm output relay (phone jack connector) fails to operate correctly during alarm conditions

Ranked causes

  1. 1. Relay contacts not closing/opening as expected (unknown)
First check

Measure continuity at Tip contact (Relay Normally Open) during alarm condition

Test point
Nurse Call connector Tip
Expected
Closed

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Only one Battery Charging LED color displayed

Battery charging indicator shows only one color instead of expected color sequence

Ranked causes

  1. 1. Membrane switch assembly malfunctioned (common)
First check

Observe LED color display during charging

Test point
Battery charging LED
Expected
Color sequence should change during charge cycle

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pressing key does not access expected function caution

Front panel keys trigger incorrect functions; for example Print key initiates NBP measurement instead

Ranked causes

  1. 1. Membrane switch assembly malfunctioned (common)
First check

Verify key function mapping is correct

Test point
Each front panel key
Expected
Each key triggers its labeled function

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Replace fuse in AC power connector high voltage

Fuse replacement procedure for AC power connector on rear of monitor

Ranked causes

  1. 1. Fuse failure (unknown)
First check

Remove AC power cord from rear of monitor

Test point
AC power connector rear
Expected
Power cord disconnected

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Self test fails to detect module or reports error patient safety

Running the Self Test detects missing module or test error occurs; affects SpO2, NBP, or Predictive Temperature modules

Ranked causes

  1. 1. Module hardware not connected or malfunctioning (unknown)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
1025System ErrorCould not load ResourceDLL.dllerror
1026System ErrorCould not load data from flash memoryerror
1027System ErrorCould not start front enderror
1028System ErrorCould not stop front enderror
1029System ErrorCould not allocate enough memoryerror
1030System ErrorError shutting down front end deviceserror
1031System ErrorMonitoring and watchdog thread failed to initializeerror
1032System ErrorCould not open audio fileserror
1033System ErrorUnspecified system errorerror
1034System ErrorFailed to load product informationerror
1035System ErrorCannot initialize front end managererror
1036System ErrorFailed to start work threaderror
1037System ErrorCannot start polling threaderror
1042System ErrorFailed to start network drivererror
1043System ErrorFront end software errorerror
1044System ErrorFront end software errorerror
1045System ErrorPrinting software errorerror
1047System ErrorMonitoring software errorerror
1048System ErrorException in alarming softwareerror
1050System ErrorFailed to stop work threaderror
1051System ErrorCannot initialize NBPerror
1052System ErrorMemory error detectederror
1053System ErrorLoss of monitoringinfo
257System ErrorBattery charger power failureerror
261System ErrorBattery is overchargedwarning
264 temp valueSystem ErrorBattery over temperaturewarning
817FPGA FIFO OverrunNBP software errorinfo
818 (101)NBP Equip Malfunc (101)Task overrunerror
818 (102)NBP Equip Malfunc (102)Interrupt stack overwrittenerror
818 (103)NBP Equip Malfunc (103)User stack overwrittenerror
818 (104)NBP Equip Malfunc (104)Data validity check failederror
818 (106)NBP Equip Malfunc (106)Unable to write init flag data to non-volatile memoryerror
818 (107)NBP Equip Malfunc (107)Unable to verify init flag data written to non-volatile memoryerror
818 (109)NBP Equip Malfunc (109)Unable to write NBP configuration data to non-volatile memoryerror
818 (110)NBP Equip Malfunc (110)Unable to write NBP configuration data to non-volatile memoryerror
818 (111)NBP Equip Malfunc (111)Calibration data checksum errorerror
818 (112)NBP Equip Malfunc (112)Unable to write calibration data to non-volatile memoryerror
818 (113)NBP Equip Malfunc (113)Unable to verify calibration data written to non-volatile memoryerror
818 (114)NBP Equip Malfunc (114)Error log checksum errorerror
818 (115)NBP Equip Malfunc (115)Unable to write error log to non-volatile memoryerror
818 (116)NBP Equip Malfunc (116)Unable to verify error log data written to non-volatile memoryerror
818 (117)NBP Equip Malfunc (117)Inflation cycle count and runtime data checksum errorerror
818 (118)NBP Equip Malfunc (118)Unable to write inflation cycle count and runtime data to non-volatile memoryerror
818 (119)NBP Equip Malfunc (119)Unable to verify inflation cycle count and runtime data written to non-volatile memoryerror
818 (120)NBP Equip Malfunc (120)New message queuing before previous message finishederror
818 (121)NBP Equip Malfunc (121)Pressure value from second pressure sensor not availableerror
818 (122)NBP Equip Malfunc (122)5-volt supply voltage value is too high or lowerror
818 (123)NBP Equip Malfunc (123)Error from 5-volt regulatorerror
818 (124)NBP Equip Malfunc (124)12-volt supply voltage value is too high or lowerror
818 (128)NBP Equip Malfunc (128)Offset of pressure sensor 1 is out of rangeerror
818 (129)NBP Equip Malfunc (129)Offset of pressure sensor 2 is out of rangeerror
818 (130)NBP Equip Malfunc (130)Response of pressure sensor 2 to test signal is out of rangeerror
818 (131)NBP Equip Malfunc (131)Difference between two pressure channels is too largeerror
818 (132)NBP Equip Malfunc (132)One of the two valves was activated for too longerror
818 (133)NBP Equip Malfunc (133)CRC mismatch in calibration dataerror
818 (134)NBP Equip Malfunc (134)CRC mismatch in error log dataerror
818 (135)NBP Equip Malfunc (135)CRC mismatch in inflation cycle count and runtime dataerror
818 (136)NBP Equip Malfunc (136)Pneumatic safety switch is defectiveerror
818 (137)NBP Equip Malfunc (137)Valve 2 is defective or blockederror
818 (141)NBP Equip Malfunc (141)Inflation time for selected patient type exceededwarning
818 (15)NBP Equip Malfunc (15)NBP abnormal unrecoverable system errors with internal firmwareerror
818 (199)NBP Equip Malfunc (199)No data from NBP blockerror
818 (2)NBP Equip Malfunc (2)NBP power-on self test failed; A/D converter inoperative or pressure transducer offset too largeerror
818 (25)NBP Equip Malfunc (25)NBP hardware has malfunctioned because there is no dataerror
819NBP Equip Malfunc - Not CalibratedNBP not calibratedwarning
819 (127)NBP Equip Malfunc (127)NBP module is not calibratedwarning
819 (138)NBP Hose Blocked (138)Problem in pneumatic systemerror
819 (139)NBP Air Leak (139)Air leak or valve defecterror
833 (1)SpO2 Equip Malfunc (1)SpO2 ROM checksum errorerror
833 (10)SpO2 Equip Malfunc (10)SpO2 host communication errorerror
833 (2)SpO2 Equip Malfunc (2)SpO2 RAM test errorerror
833 (20)SpO2 Equip Malfunc (20)SpO2 hardware defecterror
833 (4)SpO2 Equip Malfunc (4)SpO2 analog signal path self-test failederror
833 (40)SpO2 Equip Malfunc (40)SpO2 non-volatile memory corruptederror
833 (6)SpO2 Equip Malfunc (6)SpO2 self-test timeouterror
833 (8)SpO2 Equip Malfunc (8)SpO2 internal DSP communication errorerror
834SpO2 Sensor MalfuncMalfunction of SpO2 sensor or sensor cableerror
835FPGA FIFO OverrunSpO2 software errorinfo
836SpO2 Communication ErrorSpO2 communication errorinfo
837SpO2 Error. Front end will now reset itselfNo data from SpO2warning
865FPGA FIFO OverrunTemperature software errorinfo
866 (21)Temp Probe Error (21)Probe operating above 112F/43.3C; heater too close to thermistorerror
866 (22)Temp Probe Error (22)Excessive heater energy; heater not working or slow responseerror
866 (32)Temp Probe Error (32)Thermistor pulled from tip or heater brokenerror
866 (33)Temp Probe Error (33)Probe not responsive; not characterized/calibratederror
866 (34)Temp Probe Error (34)Probe not characterized/calibratederror
866 (45)Temp Probe Error (45)Measurement below allowable values and far below low ambient limitswarning
866 (46)Temp Probe Error (46)Measurement above allowable values and far above high ambient limitswarning
866 (52)Temp Probe Error (52)Temperature probe not characterized/calibratederror
866 (63)Temp Probe Error (63)Probe well missing or not installed properlyerror
866 (70)Temp Probe Error (70)Problem reading temperature probe EEPROMerror
866 (75)Temp Probe Error (75)Problem reading probe EEPROM or probe not factory testederror
866 (78)Temp Probe Error (78)Probe errorerror
866 (79)Temp Probe Error (79)Module memory errorerror
866 (80)Temp Probe Error (80)Software trying to turn off probe heater but feedback says still onerror
866 (81)Temp Probe Error (81)Software trying to turn on probe heater but feedback says still offerror
867 (24)Temp Module Malfunc (24)Ambient temperature too highwarning
867 (25)Temp Module Malfunc (25)Ambient temperature too lowwarning
867 (27)Temp Module Malfunc (27)Battery or power supply voltage exceeds maximum valueerror
867 (28)Temp Module Malfunc (28)Battery or power supply voltage below minimum valueerror
867 (47)Temp Module Malfunc (47)Internal calibration resistor (RCAL) on board is damagederror
867 (48)Temp Module Malfunc (48)Internal calibration resistor (RCAL) is damagederror
867 (49)Temp Module Malfunc (49)Internal circuit validation resistor (PTB) is damagederror
867 (50)Temp Module Malfunc (50)Internal circuit validation resistor (PTB) is damagederror
867 (51)Temp Module Malfunc (51)A/D measurement timed out; did not finish in allowed time sloterror
867 (59)Temp Module Malfunc (59)Battery or power supply voltage below maximum valuewarning
867 (60)Temp Module Malfunc (60)Battery or power supply voltage exceeds minimum valuewarning
867 (61)Temp Module Malfunc (61)Reference voltage circuit under voltage or unstableerror
867 (62)Temp Module Malfunc (62)The module is not calibratedwarning
867 (65)Temp Module Malfunc (65)Problem saving to module EEPROMerror
867 (66)Temp Module Malfunc (66)Module non-volatile calibration memory error detectionerror
867 (67)Temp Module Malfunc (67)Problem reading/writing to module EEPROMerror
867 (68)Temp Module MalfunctionProblem reading/writing to module EEPROMerror
867 (69)Temp Module MalfunctionProblem reading/writing to module EEPROMerror
867 (74)Temp Module MalfunctionModule initializing EEPROM for first timewarning
867 (82)Temp Module MalfunctionHTR_Q signal on but HTRC signal off with voltage presenterror
867 (83)Temp Module MalfunctionHTR_Q tri-stated with HTRC enabled and heater powererror
867 (84)Temp Module MalfunctionQ&C turned on with insufficient heater voltageerror
867 (85)Temp Module MalfunctionHeater fail-safe should have turned off but did notcritical
868Temperature ErrorNo data from temperature sensorwarning
881Speaker MalfunctionSpeaker hardware failureunknown
882Wireless MalfunctionWireless firmware cannot loaderror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI