Philips SureSigns VM1: troubleshooting and error codes
The ServiceForge knowledge base holds 30 documented symptoms for the Philips SureSigns VM1, with 79 ranked causes, 27 diagnostic checks with test points and expected values, and 48 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Monitor does not turn on with AC power caution
Monitor powers on only with battery power, not AC
Ranked causes
- 1. AC power cord unplugged or broken (common)
- 2. Fuse is blown (common)
- 3. Power supply cable malfunction (occasional)
- 4. Power supply malfunction (occasional)
- 5. AC power connector malfunction (occasional)
- 6. Main board malfunction (occasional)
Check power supply output voltage
- Test point
- Power supply terminals
- Expected
- 15 V
- Tools
- DMM
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder does not print caution
Recorder not producing output on paper
Ranked causes
- 1. Recorder is out of paper (common)
- 2. Recorder door is open (common)
- 3. Improper paper type (common)
- 4. Recorder not recognized by system (occasional)
- 5. Recorder module malfunction (occasional)
- 6. Main board malfunction (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner cannot read barcodes caution
Scanner lights and sound work but cannot read barcodes
Ranked causes
- 1. Scanner held too close or too far from barcode (common)
- 2. Barcode format not recognized by system (occasional)
- 3. Scanner needs reset (occasional)
- 4. System does not recognize scanner (occasional)
- 5. Scanner mechanical problem (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner does not function caution
Scanner has no lights, no sound, cannot read barcodes
Ranked causes
- 1. Scanner not plugged into USB port (common)
- 2. Scanner window not clean or blocked (common)
- 3. System does not recognize scanner (occasional)
- 4. Scanner mechanical problem (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not respond to navigation wheel or front panel keys caution
Input controls not functioning
Ranked causes
- 1. Bad connection to main board (common)
- 2. Navigation wheel or keypad is broken (occasional)
- 3. Membrane switch assembly malfunction (occasional)
- 4. Main board malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor screen is blank
Power is on but display shows no content
Ranked causes
- 1. Monitor is in power save mode (common)
- 2. Bad LCD connection (common)
- 3. LCD is defective (occasional)
- 4. Main board malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No CO2 readings displayed
CO2 module not providing measurements
Ranked causes
- 1. CO2 module is initializing (common)
- 2. CO2 tube is kinked (common)
- 3. CO2 module malfunction (occasional)
- 4. Main board malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No SpO2 readings displayed
SpO2 module not providing measurements
Ranked causes
- 1. SpO2 module in learning phase (common)
- 2. SpO2 sensor malfunction (occasional)
- 3. SpO2 board malfunction (occasional)
- 4. Main board malfunction (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade fails or monitor becomes unresponsive during upgrade
Monitor may not complete software upgrade process, display indicates upgrade in progress but does not complete, or upgrade is interrupted causing data loss
Ranked causes
- 1. Monitor unplugged during upgrade process (occasional)
- 2. USB flash drive removed during upgrade (occasional)
- 3. Keys pressed during upgrade process (occasional)
- 4. Invalid software image on USB flash drive (occasional)
Verify valid software image is present on USB flash drive
- Test point
- USB flash drive
- Expected
- Monitor detects valid software image and displays updated image information in Upgrade Software window
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audible alarms do not sound patient safety
No sound output from alarm speakers
Ranked causes
- 1. Bad speaker connection (common)
- 2. Speaker is defective (occasional)
- 3. Main board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CO2 readings are unreliable patient safety
CO2 measurements inconsistent or inaccurate
Ranked causes
- 1. CO2 tube is kinked (common)
- 2. External problem with CO2 setup (occasional)
- 3. CO2 board not calibrated (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor cannot detect configuration file on USB flash drive during import
Monitor displays error message when Import Settings is selected; USB flash drive with exported settings is not recognized or configuration file is not properly formatted
Ranked causes
- 1. USB flash drive not inserted or not detected (common)
- 2. Configuration file not properly formatted or corrupted (occasional)
- 3. Hardware configuration or software version mismatch between monitors (occasional)
Verify USB flash drive is inserted into USB port
- Test point
- USB port on back of monitor
- Expected
- USB flash drive is fully inserted and recognized
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor displays random or distorted graphics with white background
Display shows garbled or distorted content
Ranked causes
- 1. LCD cable not attached correctly (common)
- 2. LCD is defective (occasional)
- 3. Main board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on
Monitor fails to respond when power switch is pressed, no display, no LEDs light
Ranked causes
- 1. Power switch not turned on (common)
- 2. Battery not adequately charged (common)
- 3. AC power cord not connected (common)
Verify power switch position
- Test point
- Power key
- Expected
- Power switch turned on
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not turn on with battery power caution
Monitor powers on only with AC power, not battery
Ranked causes
- 1. Battery has lost charge (common)
- 2. Battery connector is loose (common)
- 3. Main board malfunction (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder paper is jamming caution
Paper not feeding or stuck in recorder
Ranked causes
- 1. Paper not installed correctly (common)
- 2. Improper paper type (common)
- 3. Recorder mechanical problem (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Clear patient data operation interrupted or incomplete
Patient data not cleared from memory after Clear Data command executed; system does not restart after clearing data
Ranked causes
- 1. Incorrect maintenance password entered (common)
- 2. Clear Data operation not confirmed (occasional)
Navigate to Service menu and access Diagnostics
- Test point
- System Admin Menu
- Expected
- System Diagnostics menu displays with Maintenance option available
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder test shows poor print quality or no output
Optional recorder does not print test pattern or prints with voids, streaks, faded areas, or mechanical issues
Ranked causes
- 1. Faulty recorder print head (common)
- 2. Recorder motor malfunction (common)
Run recorder test and inspect print quality
- Test point
- Recorder output paper
- Expected
- Clean, continuous test pattern with no voids, streaks, or fading
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 readings are unreliable patient safety
SpO2 measurements inconsistent or inaccurate
Ranked causes
- 1. External problem with SpO2 setup (occasional)
- 2. Low perfusion (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner does not read all patient ID methods caution
Scanner only reads primary patient identification method
Ranked causes
- 1. Scanner configured for primary ID only (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery information unavailable or no data from battery
Battery Info window displays message 'No data from battery. Please see your Service Guide' or battery parameters cannot be read
Ranked causes
- 1. Battery not properly seated or connection loose (common)
Access Battery Info window in System Diagnostics
- Test point
- Monitor menu
- Expected
- Battery Info window displays parameters including manufacturer, chemistry, serial number, date, cycle count, relative charge percentage
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: IBP (Invasive Blood Pressure)
Perform IBP calibration via Maintenance menu options
Ranked causes
- 1. Routine calibration or maintenance requirement (unknown)
Access Maintenance options
- Test point
- System Diagnostics Maintenance button
- Expected
- Maintenance menu displays with IBP Calibration option
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Microstream CO2
Perform CO2 calibration using calibration gas and equipment at recommended intervals or after CO2 module replacement
Ranked causes
- 1. Routine preventive maintenance or CO2 module replacement (common)
Access System Diagnostics menu
- Test point
- System Admin Menu Service button
- Expected
- System Diagnostics menu appears with Errors: 0
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows no image or distorted patterns
Monitor screen does not display proper test patterns or shows garbled/missing bars, colors, or text during display self-test
Ranked causes
- 1. Faulty LCD display or video controller (common)
Verify vertical bar patterns display correctly
- Test point
- Monitor screen
- Expected
- Alternating black and white vertical bars starting with white; inverse pattern starting with black; red description text on top
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Key press does not access expected function caution
Button input not performing correct action or no click feedback
Ranked causes
- 1. Membrane switch assembly malfunction (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
LED does not function as expected during test
Battery or power LED does not light or flash in correct sequence; LED remains off or shows wrong colors during LED self-test
Ranked causes
- 1. Faulty battery LED or power LED (occasional)
Verify battery LED lights yellow
- Test point
- Battery LED indicator
- Expected
- Battery LED lights steady yellow for five seconds
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor display is blank or incorrect
LCD screen shows no content, garbled display, or incorrect colors after power on
Ranked causes
- 1. LCD display failure (occasional)
Verify proper boot sequence
- Test point
- Screen display
- Expected
- Color bars displayed for five seconds, then LCD off for three seconds, Philips screen appears with tone
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not turn on with either AC or battery power caution
No power from AC or battery sources
Ranked causes
- 1. LCD, cables, battery, power supply, or main board malfunction (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Navigation wheel or keys not responding
Monitor keys or rotary wheel do not respond to user input; icons in test window do not highlight or wheel dots do not change color when operated
Ranked causes
- 1. Faulty navigation wheel (common)
Test each monitor key individually
- Test point
- Monitor keys
- Expected
- Key icon lights with border for five seconds when pressed
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Opening screen or color bar hangs during startup
Monitor starts boot sequence but then freezes
Ranked causes
- 1. Main board is defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
1025 | System Error | Could not load ResourceDLL.dll. | error |
1026 | System Error | Could not load data from flash. | error |
1027 | System Error | Could not start front end. | error |
1028 | System Error | Could not stop front end. | error |
1029 | System Error | Could not allocate enough memory. | error |
1030 | System Error | An error occurred when system shut down front end devices. | error |
1031 | System Error | The monitoring and watchdog thread could not initialize. | error |
1032 | System Error | Could not open audio files. | error |
1033 | System Error | Unspecified error. | error |
1034 | System Error | Failed to load product info. | error |
1035 | System Error | Cannot initialize front end manager. | error |
1036 | System Error | Failed to start work thread. | error |
1037 | System Error | Cannot start polling thread. | error |
1042 | System Error | Failed to start up network driver. | error |
1043 | System Error | Front end software error. | error |
1044 | System Error | Front end software error. | error |
1045 | System error (code 1045) | Detected during self-test or operation; logged with timestamp | error |
1047 | System Error | Monitoring software error. | error |
1048 | System Error | An exception was detected in the alarming software. | error |
1050 | System Error | Failed to stop work thread. | error |
1052 | System Error | Memory error. | error |
1053 | System Error | Loss of monitoring. | info |
257 | System error (code 257) | Detected during self-test or operation; logged with timestamp | error |
261 | System error (code 261) | Detected during self-test or operation; logged with timestamp | error |
264 | System Error | Battery over temperature | error |
833 (1) | SpO2 Equip Malfunc (1) | SpO2 ROM checksum error | error |
833 (10) | SpO2 Equip Malfunc (10) | SpO2 host communication error | error |
833 (2) | SpO2 Equip Malfunc (2) | SpO2 RAM test error | error |
833 (20) | SpO2 Equip Malfunc (20) | SpO2 hardware defect | error |
833 (4) | SpO2 Equip Malfunc (4) | SpO2 analog signal path self-test failed | error |
833 (40) | SpO2 Equip Malfunc (40) | SpO2 non-volatile memory corrupted | error |
833 (6) | SpO2 Equip Malfunc (6) | SpO2 self-test timeout | error |
833 (8) | SpO2 Equip Malfunc (8) | SpO2 internal DSP communication error | error |
834 | SpO2 Sensor Malfunc | Malfunction of the SpO2 sensor or sensor cable | error |
835 | FPGA FIFO Overrun | SpO2 software error | info |
836 | SpO2 Communication Error | SpO2 communication error | info |
837 | SpO2 Error. The front end will now reset itself. | No data from SpO2 | error |
849 | FPGA FIFO Overrun | CO2 software error | info |
850 | CO2 Error. The front end will now reset itself. | No data from CO2 | error |
851 (1) | CO2 Equip Malfunc (1) | CO2 calibration error | error |
851 (16) | CO2 Equip Malfunc (16) | CO2 main board error | error |
851 (17) | CO2 Equip Malfunc (17) | CO2 module malfunction | error |
851 (18) | CO2 Equip Malfunc (18) | CO2 scrubber, pump error | error |
851 (19) | CO2 Equip Malfunc (19) | CO2 sensor malfunction | error |
851 (2) | CO2 Equip Malfunc (2) | CO2 air line malfunction | error |
851 (20) | CO2 Equip Malfunc (20) | CO2 15V voltage out of range | error |
851 (3) | CO2 Equip Malfunc (3) | Occlusion in CO2 input line | error |
851 (99) | CO2 Equip Malfunc (99) | No data from CO2 | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI