Philips REMstar M: troubleshooting and error codes
The ServiceForge knowledge base holds 2 documented symptoms for the Philips REMstar M, with 2 ranked causes, 5 diagnostic checks with test points and expected values, and 129 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Pressure sensor auto null
Automatic calibration of pressure sensor zero point when blower is off and flow sensor signal is stable
Ranked causes
- 1. Auto null calibration procedure required during operation (common)
Verify blower is stopped and flow sensor signal is stable
- Test point
- Blower status, flow sensor output
- Expected
- Blower off, flow sensor signal stable
- Tools
- Visual inspection, oscilloscope
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pressure sensor for snore detection
Calibrate pressure sensor amplifier and signal chain for snore detection; involves applying known pressures and verifying DC offset and frequency response
Ranked causes
- 1. Sensor calibration required at manufacture and after board installation (common)
Verify sensor output voltage at calibrated pressures with 5.0V supply
- Test point
- Sensor output
- Expected
- 0.5V at -5 hPa (-5 cm H2O) and 4.5V at 40 hPa (40 cm H2O), Not specified
- Tools
- Voltmeter, pressure calibrator
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
001 | Failed destructive power on self test of external RAM address lines | Power-on self test execution | error |
002 | Failed destructive power on self test of external RAM data lines | Power-on self test execution | error |
003 | Failed non-destructive power on self test of internal RAM | Power-on self test execution | error |
004 | Failed non-destructive continuous self test of external RAM | Continuous self test during operation | error |
005 | ERR_RTC_DRIFT | Real-time clock time has drifted beyond acceptable tolerance | warning |
005 | RTC drift exceeds 4 seconds per minute | Continuous comparison between RTC and crystal oscillator | warning |
006 | Device improperly configured; pressure sensor required but absent | Device startup with sensor requirement mismatch | critical |
007 | Error in program flow | Program flow check failure | error |
008 | Non-volatile RAM storage unit corrupted upon power-up | Power-on NVRAM verification | error |
009 | ERR_FLOW_SENSOR_ABSENT | Flow sensor is not detected or not responding on sensor board | critical |
010 | Watchdog timeout after initialization completed | Watchdog timeout during normal operation | info |
011 | Watchdog timeout after initialization completed | Watchdog timeout during normal operation | error |
012 | Watchdog timeout before initialization completed | Watchdog timeout during startup sequence | error |
013 | Power on self test of watchdog timer failed; RAM flag test failed | Watchdog timer self test | error |
014 | Power on self test of watchdog timer failed | Watchdog timer self test | info |
015 | Power on self test of watchdog timer failed | Watchdog timer self test | error |
016 | DSP crystal fallen off board during operation | DSP clock monitoring | error |
017 | Therapy controller constant registers corrupted | Compiler optimization register verification | error |
018 | Stack memory overrun | Stack memory monitoring during operation | error |
019 | Therapy controller application software corrupted in Flash memory | Flash memory integrity check at startup | error |
019 | Device error after reflashing or Therapy PCA defect | Software reflash or hardware fault | error |
020 | Therapy controller vectored to undefined interrupt | Interrupt vector verification | error |
021 | Device attempting to access non-existent external RAM | External RAM access attempt | critical |
022 | PCA communication failure | Therapy PCA and Sensor PCA cannot communicate | error |
022 | ERR_SENSOR_TIMEOUT_RESET | Sensor board communication timeout; system performed reset | warning |
022 | 100 ms without pressure/flow message from Sensor PCA | Message timeout during sensor polling | warning |
023 | 100 ms without pressure/flow message after Sensor PCA reset | Persistent message timeout after reset attempt | error |
023 | ERR_SENSOR_TIMEOUT_REBOOT | Sensor board communication timeout; system rebooted | error |
023 | Sensor PCA or communication error | Sensor or PCA fault detected | error |
024 | ERR_SENSOR_STATUS_TIMEOUT_STOP | Sensor status response timeout; device stopped | critical |
024 | 3 seconds without TOD message from Sensor PCA | TOD message timeout | critical |
025 | DSP watchdog timeout occurred | DSP watchdog monitoring | error |
026 | ERR_RTC_VALUE | Real-time clock value is invalid or out of range | error |
026 | Sensor PCA battery may be dead; RTC never set | Sensor power/RTC status check | info |
027 | ERR_RTC_CHECKSUM | Real-time clock checksum verification failed | error |
028 | ERR_FLOW_SENSOR_PWM_REBOOT | Flow sensor PWM signal invalid; system rebooted | error |
028 | Sensor PCA flow signal processing error during motor spinup | Motor spinup signal processing check | error |
029 | Motor rotor not spinning or wires broken during spinup | Motor spinup verification | error |
030 | ERR_SENSOR_RX_REBOOT | Sensor board receive data error; system rebooted | error |
030 | Sensor PCA >30 communication errors within last minute | Communication error rate monitoring | error |
031 | DSP application software corrupted in Flash memory | Flash memory integrity check | error |
031 | Device error after reflashing or Therapy PCA defect | Software reflash or hardware fault | error |
032 | Sensor PCA cannot keep up with communication message rate | Sensor message processing rate check | error |
032 | ERR_SENSOR_REQUEST_OVERRUN | Sensor request queue overflow; data lost | warning |
033 | Humidifier plate not making heat | Humidifier temperature monitoring | warning |
034 | Humidifier plate temperature at maximum (183°F) | Humidifier overheat protection | warning |
035 | Sensor PCA flow offset error; offset exceeded ±40 lpm for 5 min | Continuous flow offset monitoring | critical |
035 | ERR_FLOWOFFSET_STOP | Flow offset calibration out of range; device stopped | critical |
036 | DSP boot software corrupted in Flash memory | Flash boot code integrity check | error |
037 | ERR_PRESSUREOFFSET_STOP | Pressure offset calibration out of range; device stopped | critical |
037 | Sensor PCA pressure offset error; offset exceeded ±3 cm H2O for 5 min | Continuous pressure offset monitoring | critical |
038 | RAM copy of NVRAM storage unit corrupted | NVRAM consistency check during operation | critical |
039 | Invalid key depressed at startup or held >30 seconds post-startup | Key input monitoring at startup | info |
040 | Blower commanded on with corrupted Sensor PCA calibration table | Blower start with calibration check | critical |
040 | ERR_SENSOR_TABLE_CHECKSUM | Sensor calibration table checksum verification failed | error |
041 | Therapy controller 10 ms periodic loop exceeded 10 ms | Periodic task timing monitoring | info |
042 | Therapy controller 10 ms periodic loop exceeded 10 ms | Periodic task timing monitoring | error |
043 | Daily Values data corruption; both NV Images corrupted and defaulted | Daily averaging data integrity check | info |
043 | Therapy PCA error | Post-reflash fault or hardware defect | warning |
044 | DSP not responding to software version request at startup | DSP boot communication check | error |
045 | Therapy controller not getting responses from DSP | DSP communication monitoring | info |
046 | Therapy controller not getting responses from DSP | DSP communication monitoring during operation | error |
047 | Therapy processor cannot shut drive down via hardware; power-on self test | Motor shutdown verification during self test | error |
048 | DSP reported fault unable to be translated to error code | Therapy controller receives untranslatable fault from DSP | error |
049 | DSP not receiving messages from therapy controller | Communication loss between DSP and therapy controller | info |
050 | DSP not receiving messages from therapy controller | Persistent communication loss between DSP and therapy controller | error |
051 | DSP detected hardware trip line activation | Hardware trip line becomes active | error |
052 | V bus voltage dropped below minimum threshold | Supply voltage drops below 7.08 VDC for 25 ms | critical |
052 | Noisy or improper power supply | Electrical noise or improper power input detected | warning |
053 | Over-voltage condition | DC power supply voltage too high | error |
053 | V bus voltage exceeded maximum threshold | Supply voltage rises above 17.3 VDC for 25 ms | critical |
054 | Motor control failure | Motor cannot be controlled or is not spinning | error |
055 | DSP application data corrupted in Flash memory | Firmware detects data corruption during self-test | critical |
055 | Device error after reflashing or Therapy PCA defect | Software reflash or hardware fault | error |
056 | Motor speed detected as too high; control failure | Motor speed exceeds safe operating range | error |
057 | Motor speed detected as negative; control failure | Motor speed measurement indicates reverse rotation or negative value | error |
058 | Motor thermistor open circuit | Temperature sensor reads open or absent | warning |
059 | Motor thermistor shorted | Temperature sensor short circuit detected | warning |
060 | Program flow error; likely programming defect | DSP detects unexpected code execution path | error |
061 | Motor parameter measurement not found within timeout | Motor parameter measurement fails to complete in 18 seconds | error |
062 | ERR_RTC_STOPPED | Real-time clock stopped or not running | critical |
062 | Real-time clock not incrementing properly | RTC does not increment within one-minute period | info |
063 | Motor drive calibration data corrupted | Calibration data fails integrity check | error |
064 | Communications to humidifier lost | Humidifier link lost for 5 seconds after establishment | info |
065 | Humidifier thermistor may be shorted | Temperature sensor reads short circuit condition | info |
066 | Humidifier thermistor may be open | Temperature sensor reads open circuit condition | info |
067 | Humidifier reported untranslatable fault | Humidifier sends fault code that controller cannot interpret | info |
069 | DSP vectored to undefined interrupt | Processor executes undefined interrupt handler | error |
070 | ERR_SENSOR_BOARD_ABSENT | Sensor board is not detected at power-up | critical |
070 | Sensor PCA software version or serial number unreadable at startup | Sensor module fails to respond with version/serial at boot | error |
070 | Sensor or communication error | Sensor or PCA fault detected | error |
071 | Motor drive detected development boot code | Motor drive firmware is in development/debug mode | error |
072 | Therapy controller detected DSP running in boot mode | DSP does not transition from boot to application mode | critical |
073 | Program flow error; likely programming defect | DSP detects unexpected code execution path | info |
074 | Non-destructive power-on self-test of DSP RAM failed | RAM self-test fails during initialization | critical |
075 | Motor startup control failure | Attempts to start motor have failed | error |
076 | Motor acceleration control failure | Motor fails to accelerate after startup attempt | error |
077 | Motor deceleration control failure | Motor fails to decelerate properly | error |
078 | DSP 20 kHz periodic loop execution time exceeded | 20 kHz loop execution exceeds 50 microseconds | error |
079 | DSP 100 Hz periodic loop execution time exceeded | 100 Hz loop execution exceeds 10 milliseconds | error |
080 | DSP stack memory overrun | Stack pointer exceeds allocated memory bounds | critical |
081 | Sensor PCA communication handler exceeds message period | Sensor communication handler execution exceeds 5 ms | info |
081 | ERR_SENSOR_HANDLER_OVERRUN | Sensor interrupt handler queue overflowed | warning |
082 | Therapy Event Queue in Auto device is full | Queue buffer reaches maximum capacity | info |
083 | Therapy or DSP software does not support hardware version | Software compatibility check fails at startup | error |
084 | Event Type and length do not match | Event data structure validation fails | info |
085 | Pressure sensor missing or incompatible for Auto mode | Auto titration mode active without compatible digital pressure sensor | error |
086 | Debug-only log of missing digital pressure sensor | Debugging enabled and digital pressure sensor absent | info |
087 | Flow sensor tubing occluded with water | Constant near-zero flow output detected in Auto mode after blower off | critical |
087 | Flow sensor water damage or malfunction | Water damage detected or sensor fault after reflash | error |
088 | Flow sensor water damage or malfunction | Water damage detected or sensor fault after reflash | error |
088 | Flow sensor tubing may be occluded with water (log-only) | Constant near-zero flow output detected when not in Auto mode | info |
089 | Background thread task stuck preventing execution | One background task blocks others from running | error |
090 | Background thread task stuck (log-only) | One background task blocks others from running | info |
091 | Exception Stack end location memory overrun | Exception stack pointer exceeds upper memory bound | critical |
092 | Exception Stack has written into reserve area | Exception stack usage exceeds normal range | info |
093 | Exception Stack starting location memory overrun | Exception stack pointer exceeds lower memory bound | critical |
094 | FIQ Stack end location memory overrun | Fast Interrupt Queue stack exceeds upper bound | critical |
095 | FIQ Stack has written into reserve area | FIQ stack usage exceeds normal range | info |
096 | FIQ Stack starting location memory overrun | FIQ stack pointer exceeds lower memory bound | critical |
097 | IRQ Stack end location memory overrun | Interrupt Request stack exceeds upper bound | critical |
098 | IRQ Stack has written into reserve area | IRQ stack usage exceeds normal range | info |
099 | IRQ Stack starting location memory overrun | IRQ stack pointer exceeds lower memory bound | critical |
100 | Service Stack end location memory overrun | Service routine stack exceeds upper bound | critical |
101 | Service Stack has written into reserve area | Service stack usage exceeds normal range | info |
102 | Service Stack starting location memory overrun | Service stack pointer exceeds lower memory bound | critical |
103 | Data written to an illegal address | Memory write operation to invalid address detected | critical |
104 | Jump to an illegal address | Code execution jumps to invalid memory address | critical |
105 | Illegal instruction detected | Processor executes undefined or invalid instruction | critical |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI