Philips REMstar M: troubleshooting and error codes

The ServiceForge knowledge base holds 2 documented symptoms for the Philips REMstar M, with 2 ranked causes, 5 diagnostic checks with test points and expected values, and 129 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Calibration: Pressure sensor auto null

Automatic calibration of pressure sensor zero point when blower is off and flow sensor signal is stable

Ranked causes

  1. 1. Auto null calibration procedure required during operation (common)
First check

Verify blower is stopped and flow sensor signal is stable

Test point
Blower status, flow sensor output
Expected
Blower off, flow sensor signal stable
Tools
Visual inspection, oscilloscope

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure sensor for snore detection

Calibrate pressure sensor amplifier and signal chain for snore detection; involves applying known pressures and verifying DC offset and frequency response

Ranked causes

  1. 1. Sensor calibration required at manufacture and after board installation (common)
First check

Verify sensor output voltage at calibrated pressures with 5.0V supply

Test point
Sensor output
Expected
0.5V at -5 hPa (-5 cm H2O) and 4.5V at 40 hPa (40 cm H2O), Not specified
Tools
Voltmeter, pressure calibrator

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
001Failed destructive power on self test of external RAM address linesPower-on self test executionerror
002Failed destructive power on self test of external RAM data linesPower-on self test executionerror
003Failed non-destructive power on self test of internal RAMPower-on self test executionerror
004Failed non-destructive continuous self test of external RAMContinuous self test during operationerror
005ERR_RTC_DRIFTReal-time clock time has drifted beyond acceptable tolerancewarning
005RTC drift exceeds 4 seconds per minuteContinuous comparison between RTC and crystal oscillatorwarning
006Device improperly configured; pressure sensor required but absentDevice startup with sensor requirement mismatchcritical
007Error in program flowProgram flow check failureerror
008Non-volatile RAM storage unit corrupted upon power-upPower-on NVRAM verificationerror
009ERR_FLOW_SENSOR_ABSENTFlow sensor is not detected or not responding on sensor boardcritical
010Watchdog timeout after initialization completedWatchdog timeout during normal operationinfo
011Watchdog timeout after initialization completedWatchdog timeout during normal operationerror
012Watchdog timeout before initialization completedWatchdog timeout during startup sequenceerror
013Power on self test of watchdog timer failed; RAM flag test failedWatchdog timer self testerror
014Power on self test of watchdog timer failedWatchdog timer self testinfo
015Power on self test of watchdog timer failedWatchdog timer self testerror
016DSP crystal fallen off board during operationDSP clock monitoringerror
017Therapy controller constant registers corruptedCompiler optimization register verificationerror
018Stack memory overrunStack memory monitoring during operationerror
019Therapy controller application software corrupted in Flash memoryFlash memory integrity check at startuperror
019Device error after reflashing or Therapy PCA defectSoftware reflash or hardware faulterror
020Therapy controller vectored to undefined interruptInterrupt vector verificationerror
021Device attempting to access non-existent external RAMExternal RAM access attemptcritical
022PCA communication failureTherapy PCA and Sensor PCA cannot communicateerror
022ERR_SENSOR_TIMEOUT_RESETSensor board communication timeout; system performed resetwarning
022100 ms without pressure/flow message from Sensor PCAMessage timeout during sensor pollingwarning
023100 ms without pressure/flow message after Sensor PCA resetPersistent message timeout after reset attempterror
023ERR_SENSOR_TIMEOUT_REBOOTSensor board communication timeout; system rebootederror
023Sensor PCA or communication errorSensor or PCA fault detectederror
024ERR_SENSOR_STATUS_TIMEOUT_STOPSensor status response timeout; device stoppedcritical
0243 seconds without TOD message from Sensor PCATOD message timeoutcritical
025DSP watchdog timeout occurredDSP watchdog monitoringerror
026ERR_RTC_VALUEReal-time clock value is invalid or out of rangeerror
026Sensor PCA battery may be dead; RTC never setSensor power/RTC status checkinfo
027ERR_RTC_CHECKSUMReal-time clock checksum verification failederror
028ERR_FLOW_SENSOR_PWM_REBOOTFlow sensor PWM signal invalid; system rebootederror
028Sensor PCA flow signal processing error during motor spinupMotor spinup signal processing checkerror
029Motor rotor not spinning or wires broken during spinupMotor spinup verificationerror
030ERR_SENSOR_RX_REBOOTSensor board receive data error; system rebootederror
030Sensor PCA >30 communication errors within last minuteCommunication error rate monitoringerror
031DSP application software corrupted in Flash memoryFlash memory integrity checkerror
031Device error after reflashing or Therapy PCA defectSoftware reflash or hardware faulterror
032Sensor PCA cannot keep up with communication message rateSensor message processing rate checkerror
032ERR_SENSOR_REQUEST_OVERRUNSensor request queue overflow; data lostwarning
033Humidifier plate not making heatHumidifier temperature monitoringwarning
034Humidifier plate temperature at maximum (183°F)Humidifier overheat protectionwarning
035Sensor PCA flow offset error; offset exceeded ±40 lpm for 5 minContinuous flow offset monitoringcritical
035ERR_FLOWOFFSET_STOPFlow offset calibration out of range; device stoppedcritical
036DSP boot software corrupted in Flash memoryFlash boot code integrity checkerror
037ERR_PRESSUREOFFSET_STOPPressure offset calibration out of range; device stoppedcritical
037Sensor PCA pressure offset error; offset exceeded ±3 cm H2O for 5 minContinuous pressure offset monitoringcritical
038RAM copy of NVRAM storage unit corruptedNVRAM consistency check during operationcritical
039Invalid key depressed at startup or held >30 seconds post-startupKey input monitoring at startupinfo
040Blower commanded on with corrupted Sensor PCA calibration tableBlower start with calibration checkcritical
040ERR_SENSOR_TABLE_CHECKSUMSensor calibration table checksum verification failederror
041Therapy controller 10 ms periodic loop exceeded 10 msPeriodic task timing monitoringinfo
042Therapy controller 10 ms periodic loop exceeded 10 msPeriodic task timing monitoringerror
043Daily Values data corruption; both NV Images corrupted and defaultedDaily averaging data integrity checkinfo
043Therapy PCA errorPost-reflash fault or hardware defectwarning
044DSP not responding to software version request at startupDSP boot communication checkerror
045Therapy controller not getting responses from DSPDSP communication monitoringinfo
046Therapy controller not getting responses from DSPDSP communication monitoring during operationerror
047Therapy processor cannot shut drive down via hardware; power-on self testMotor shutdown verification during self testerror
048DSP reported fault unable to be translated to error codeTherapy controller receives untranslatable fault from DSPerror
049DSP not receiving messages from therapy controllerCommunication loss between DSP and therapy controllerinfo
050DSP not receiving messages from therapy controllerPersistent communication loss between DSP and therapy controllererror
051DSP detected hardware trip line activationHardware trip line becomes activeerror
052V bus voltage dropped below minimum thresholdSupply voltage drops below 7.08 VDC for 25 mscritical
052Noisy or improper power supplyElectrical noise or improper power input detectedwarning
053Over-voltage conditionDC power supply voltage too higherror
053V bus voltage exceeded maximum thresholdSupply voltage rises above 17.3 VDC for 25 mscritical
054Motor control failureMotor cannot be controlled or is not spinningerror
055DSP application data corrupted in Flash memoryFirmware detects data corruption during self-testcritical
055Device error after reflashing or Therapy PCA defectSoftware reflash or hardware faulterror
056Motor speed detected as too high; control failureMotor speed exceeds safe operating rangeerror
057Motor speed detected as negative; control failureMotor speed measurement indicates reverse rotation or negative valueerror
058Motor thermistor open circuitTemperature sensor reads open or absentwarning
059Motor thermistor shortedTemperature sensor short circuit detectedwarning
060Program flow error; likely programming defectDSP detects unexpected code execution patherror
061Motor parameter measurement not found within timeoutMotor parameter measurement fails to complete in 18 secondserror
062ERR_RTC_STOPPEDReal-time clock stopped or not runningcritical
062Real-time clock not incrementing properlyRTC does not increment within one-minute periodinfo
063Motor drive calibration data corruptedCalibration data fails integrity checkerror
064Communications to humidifier lostHumidifier link lost for 5 seconds after establishmentinfo
065Humidifier thermistor may be shortedTemperature sensor reads short circuit conditioninfo
066Humidifier thermistor may be openTemperature sensor reads open circuit conditioninfo
067Humidifier reported untranslatable faultHumidifier sends fault code that controller cannot interpretinfo
069DSP vectored to undefined interruptProcessor executes undefined interrupt handlererror
070ERR_SENSOR_BOARD_ABSENTSensor board is not detected at power-upcritical
070Sensor PCA software version or serial number unreadable at startupSensor module fails to respond with version/serial at booterror
070Sensor or communication errorSensor or PCA fault detectederror
071Motor drive detected development boot codeMotor drive firmware is in development/debug modeerror
072Therapy controller detected DSP running in boot modeDSP does not transition from boot to application modecritical
073Program flow error; likely programming defectDSP detects unexpected code execution pathinfo
074Non-destructive power-on self-test of DSP RAM failedRAM self-test fails during initializationcritical
075Motor startup control failureAttempts to start motor have failederror
076Motor acceleration control failureMotor fails to accelerate after startup attempterror
077Motor deceleration control failureMotor fails to decelerate properlyerror
078DSP 20 kHz periodic loop execution time exceeded20 kHz loop execution exceeds 50 microsecondserror
079DSP 100 Hz periodic loop execution time exceeded100 Hz loop execution exceeds 10 millisecondserror
080DSP stack memory overrunStack pointer exceeds allocated memory boundscritical
081Sensor PCA communication handler exceeds message periodSensor communication handler execution exceeds 5 msinfo
081ERR_SENSOR_HANDLER_OVERRUNSensor interrupt handler queue overflowedwarning
082Therapy Event Queue in Auto device is fullQueue buffer reaches maximum capacityinfo
083Therapy or DSP software does not support hardware versionSoftware compatibility check fails at startuperror
084Event Type and length do not matchEvent data structure validation failsinfo
085Pressure sensor missing or incompatible for Auto modeAuto titration mode active without compatible digital pressure sensorerror
086Debug-only log of missing digital pressure sensorDebugging enabled and digital pressure sensor absentinfo
087Flow sensor tubing occluded with waterConstant near-zero flow output detected in Auto mode after blower offcritical
087Flow sensor water damage or malfunctionWater damage detected or sensor fault after reflasherror
088Flow sensor water damage or malfunctionWater damage detected or sensor fault after reflasherror
088Flow sensor tubing may be occluded with water (log-only)Constant near-zero flow output detected when not in Auto modeinfo
089Background thread task stuck preventing executionOne background task blocks others from runningerror
090Background thread task stuck (log-only)One background task blocks others from runninginfo
091Exception Stack end location memory overrunException stack pointer exceeds upper memory boundcritical
092Exception Stack has written into reserve areaException stack usage exceeds normal rangeinfo
093Exception Stack starting location memory overrunException stack pointer exceeds lower memory boundcritical
094FIQ Stack end location memory overrunFast Interrupt Queue stack exceeds upper boundcritical
095FIQ Stack has written into reserve areaFIQ stack usage exceeds normal rangeinfo
096FIQ Stack starting location memory overrunFIQ stack pointer exceeds lower memory boundcritical
097IRQ Stack end location memory overrunInterrupt Request stack exceeds upper boundcritical
098IRQ Stack has written into reserve areaIRQ stack usage exceeds normal rangeinfo
099IRQ Stack starting location memory overrunIRQ stack pointer exceeds lower memory boundcritical
100Service Stack end location memory overrunService routine stack exceeds upper boundcritical
101Service Stack has written into reserve areaService stack usage exceeds normal rangeinfo
102Service Stack starting location memory overrunService stack pointer exceeds lower memory boundcritical
103Data written to an illegal addressMemory write operation to invalid address detectedcritical
104Jump to an illegal addressCode execution jumps to invalid memory addresscritical
105Illegal instruction detectedProcessor executes undefined or invalid instructioncritical

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI