Philips PICKER PQ: troubleshooting and error codes
The ServiceForge knowledge base holds 301 documented symptoms for the Philips PICKER PQ, with 441 ranked causes, 767 diagnostic checks with test points and expected values, and 103 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Unit does not tilt patient safety
Gantry tilt function not responding to operator commands at touch panel
Ranked causes
- 1. Tilt motor circuit breakers CB303 or CB302 off or tripped (common)
- 2. Loose or disconnected cable connectors (common)
- 3. Upper front gantry cover not closed (occasional)
- 4. Lift switches open or not reset (common)
- 5. Missing or incorrect DC power supply voltages (occasional)
- 6. 208 VAC supply missing or disconnected (occasional)
- 7. Right tilt motor no DC voltage (occasional)
- 8. Left tilt motor no DC voltage (occasional)
Verify circuit breaker CB303 position
- Test point
- CB303 on AC Control chassis
- Expected
- ON
9 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing or corrupted data in CT image acquisition
Data blocks missing from image; visible data patterns indicating detector or electronics failures; image quality degraded
Ranked causes
- 1. Analog switch malfunction in detector module (rare)
- 2. Single detector element failure (occasional)
- 3. Specific detector channel malfunction (occasional)
- 4. PSDC card failure or memory corruption (occasional)
- 5. Cable malfunction or poor connection between detector and PSDC (occasional)
- 6. VFSC card failure (rare)
- 7. Bad select cable causing multiple channel failures (occasional)
Perform front end background check scan using diagscans
- Test point
- Diagnostic scan software
- Expected
- Scan completes without errors; compare output to baseline
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry does not power on high voltage
Gantry power system non-responsive; LEDs on AC Control Panel not illuminated or circuit breakers tripped
Ranked causes
- 1. Gantry key switch not in NORMAL position (common)
- 2. Circuit breaker tripped or open (common)
- 3. Fuse failure (occasional)
- 4. E-STOP activated or disconnected (occasional)
- 5. Console and Patient Support interconnect cables disconnected (occasional)
- 6. AC Control Chassis fault (occasional)
Verify required LEDs illuminated on AC Control Panel
- Test point
- AC Control Panel LEDs
- Expected
- CONTROL POWER, SYSTEM ENABLED, DC OK, DRIVES all ON
6 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No vertical motion high voltage
Vertical positioning controls do not produce vertical motion on gantry
Ranked causes
- 1. Vertical touch panel malfunction or disconnection (common)
- 2. TILT enable command not reaching CTCC (common)
- 3. TVENB enable signal failure (common)
- 4. Vertical potentiometer open circuit (occasional)
- 5. CTCC card failure (occasional)
- 6. GPUC (GPU card) failure (rare)
Measure AC voltage on connector J415 pins 1-2, 1-3, 1-4 while pressing vertical controls
- Test point
- J415 pins 1-2, 1-3, 1-4
- Expected
- 120 VAC
- Tools
- DMM set to 300 VAC
7 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No vertical motion of patient support high voltage
Patient support does not move vertically; tilt control may also be affected
Ranked causes
- 1. Objects inhibiting vertical motion (common)
- 2. Tilt not at 0 degrees or support position too high/low (common)
- 3. Soft limit switch malfunction (occasional)
- 4. Connector or ribbon cable fault (occasional)
- 5. CTCC (gantry control card) failure (occasional)
- 6. Loss of 208 VAC power to vertical motor (occasional)
Inspect for obstructions around patient support
- Test point
- Entire vertical motion path
- Expected
- No objects present
9 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display not functioning or image not visible
Monitor shows no image or display is blank; keyboard and monitor connections may be affected
Ranked causes
- 1. Keyboard not plugged in or internal cable disconnected (common)
- 2. Contrast control incorrect setting (common)
- 3. Monitor power switch off (common)
- 4. Monitor fuse open or blown (occasional)
- 5. Power or signal cables disconnected (common)
Keyboard connection status
- Test point
- Keyboard connector
- Expected
- Plugged in and internal cable connected
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive Off signal HIGH - main drive disabled caution
Drive Off signal remains HIGH, preventing K306 energization and disabling MAIN DRIVE ON function
Ranked causes
- 1. GPU microprocessor not functional (common)
- 2. Overspeed condition detected (occasional)
- 3. Scan position measurement fault (occasional)
- 4. Power up reset signal LOW (occasional)
- 5. Over temperature condition (occasional)
Verify Drive Off signal state
- Test point
- B71, slot 19
- Expected
- LOW when normal operation intended
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Drive Off signal HIGH prevents main drive energization high voltage
K306 relay cannot be energized; MAIN DRIVE ON disabled due to Drive Off signal
Ranked causes
- 1. GPU Microprocessor not functional (rare)
- 2. Scan frame overspeed detected (occasional)
- 3. Scan Position Measurement Fault (occasional)
- 4. Power Up Reset signal LOW (rare)
- 5. Over Temperature condition (occasional)
Locate and monitor Drive Off signal at B71, slot 19
- Test point
- B71, slot 19
- Expected
- LOW (signal inactive; drive enabled)
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Erratic horizontal motion patient safety
Table top movement is unpredictable or erratic along horizontal axis during positioning
Ranked causes
- 1. Mechanical belt or motor misalignment (common)
- 2. Faulty potentiometer or resolver (common)
- 3. Horizontal Power Amp output failure (occasional)
- 4. Horizontal Servo Motor failure (occasional)
- 5. Voltage regulation or signal integrity issue (occasional)
Run diagnostics
- Test point
- System diagnostics
- Expected
- Diagnostics complete without errors
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No horizontal motion, local mode
Table or gantry will not move horizontally when using local mode touch panels; positioning system unresponsive
Ranked causes
- 1. Fuse F402 open (common)
- 2. Microswitch failure or misalignment (common)
- 3. Limit switch cable open or short (occasional)
- 4. Potentiometer failure or incorrect resistance (occasional)
- 5. Resolver-to-digital converter malfunction (occasional)
Run cocal and codd diagnostics; verify fuse F402
- Test point
- Fuse F402
- Expected
- Good
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vertical motion in one direction only patient safety
Support moves vertically in only one direction (up or down), not both
Ranked causes
- 1. Invalid VDRUP or VDRDN signal (common)
- 2. Faulty tilt panel (occasional)
- 3. Defective limit switch or cable (occasional)
- 4. Faulty relay K401 or K402 (rare)
- 5. Motor power issue or brake rubbing (common)
Verify VDRUP signal validity
- Test point
- SSR403 logic
- Expected
- Valid logic level
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
999 error in error logger with Command Shell error high voltage
999 error appears in erl error logger; Command Shell error condition; scan frame may not respond to commands
Ranked causes
- 1. No communication to GALIL motion chip (common)
- 2. Bad GALIL chip (common)
- 3. Bad communication between GPUC and PMDC (occasional)
- 4. Bad DUART (occasional)
Visually inspect scan frame and mechanical components
- Test point
- Scan frame assembly
- Expected
- No mechanical interference or damage
9 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Autovoice system not functioning
Speaker or intercom buttons (Local, Remote, Mute, Answer) not responding; associated LEDs not illuminating when buttons pressed
Ranked causes
- 1. Poor speaker connections (common)
- 2. Broken PCB to SYS MON/Autovoice cable #L7855 (occasional)
- 3. Broken PCB to Autovoice and Display cable #L7856 (occasional)
- 4. Autovoice PCB failure or LED circuit failure (rare)
CPU diagnostics test
- Test point
- System diagnostics
- Expected
- Diagnostics completed successfully
5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Scan drive offset and phase adjustment radiation
RDC offset and Motor Pole offset switches require adjustment to ensure X-ray beam alignment and smooth scan drive rotation without back EMF-induced stalling or backward rotation
Ranked causes
- 1. RDC offset switches incorrectly set (common)
- 2. Motor Pole offset switches incorrectly phased (common)
- 3. Resolver replaced or PMDC board replaced (occasional)
- 4. PMDC with resolver error correction circuitry installed (occasional)
Power removal and access preparation
- Test point
- System wallbox and gantry covers
- Expected
- System powered off; gantry covers open; cardfile access cover open; DRIVES switch OFF; service keyswitch NORMAL; three motor drive fuses removed
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows 99.9 on Tilt and 1904 on Horiz display caution
Tilt display reads 99.9 and Horizontal display shows 1904 simultaneously
Ranked causes
- 1. No communication to GALIL motion chip (common)
- 2. Bad GALIL chip (occasional)
- 3. Bad communication between GPUC and PMDC (occasional)
- 4. Bad DUART chip (occasional)
Check GALIL chip communication using oscilloscope
- Test point
- Communication signal lines
- Expected
- Stable communication signal present
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect display, vertical
Vertical display on patient positioning system is incorrect or not updating properly
Ranked causes
- 1. Vertical potentiometer (transducer) malfunction (common)
- 2. Cable BL7161 connection failure or open circuit (common)
- 3. CTCC (Control Card) failure (occasional)
- 4. GPUC (Graphics Processing Unit Card) failure (rare)
Measure vertical potentiometer resistance
- Test point
- Connector P405 sockets 1 and 2
- Expected
- 500 ohms
- Tools
- DMM
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser power supply fault high voltage
Laser does not output 120 VAC or input power is incorrect
Ranked causes
- 1. No 120 VAC input at laser connector (common)
- 2. Laser Power Supply Assembly output failure (occasional)
- 3. Laser cable open or damaged (rare)
- 4. No +5 VDC at laser input (common)
Measure 120 VAC at J2 input connector
- Test point
- J2 pins 5 (hot) and 1 (neutral)
- Expected
- 120 VAC
- Tools
- Multimeter
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No vertical motion on gantry high voltage
Gantry support does not move up, down, or tilt when vertical touch panel controls are pressed
Ranked causes
- 1. Relays K401 and K402 not seated properly in sockets (common)
- 2. No 208 VAC supply to vertical motor (common)
- 3. Invalid VDRUP or VDRDN signal (occasional)
- 4. No 208 VAC at motor terminals (occasional)
Verify relay K401 and K402 seating
- Test point
- K401 and K402 sockets
- Expected
- Relays recessed snugly in sockets
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vertical motion erratic high voltage
Support moves erratically up or down during patient positioning
Ranked causes
- 1. 208 VAC supply voltage absent or incorrect at motor (common)
- 3. 208 VAC supply voltage absent or incorrect at PSCC (common)
- 4. Relays K401 or K402 malfunction (occasional)
- 6. Bad limit switch cable or defective switch (occasional)
Verify relay seating
- Test point
- K401 and K402
- Expected
- Relays recessed in their sockets
12 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
3 KVA transformer replacement high voltage
Procedure to remove and replace the 3 KVA transformer (80 pounds) in the gantry assembly
Ranked causes
- 1. Transformer requires replacement (unknown)
- 2. Remove transformer from gantry (unknown)
- 3. Install replacement transformer (unknown)
Verify 230 VAC 3-phase power on L1, L2, L3
- Test point
- Terminal block on front of transformer
- Expected
- 230 vac 3-phase power present on l1, l2 and l3
- Tools
- DVM
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
512 error in error logger - GPU crash caution
System displays error code 512 in the error log (erl), indicating GPU processor failure or malfunction affecting scan drive operation
Ranked causes
- 2. Bad GALIL motion chip (occasional)
- 3. Bad DUART (occasional)
- 4. No communication to GALIL motion chip (occasional)
Check error log for rotating frame errors or ANB on GPU
- Test point
- erl error logger
- Expected
- No 512 error code present
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blower inoperative or running intermittently caution
Gantry cooling blower does not run continuously or fails to start; may cause thermal issues. Single blower failure or capacitor fault.
Ranked causes
- 1. Blower capacitor failed or capacitance out of range (common)
- 2. Fuse F315 or F317 open (occasional)
- 3. Blower motor winding or bearing failure (occasional)
Verify fuse F315 (blowers) continuity
- Test point
- Fuse F315 on AC Control Chassis
- Expected
- Continuity (less than 1 ohm)
- Tools
- Ohmmeter or continuity tester
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardfile circuit boards require replacement high voltage
DC voltages (+5, +15, -15V) absent at backplane buss points; suspected cardfile or power supply failure
Ranked causes
- 1. Cardfile circuit board failure or damage (unknown)
- 2. Cardfile connection errors (common)
- 3. DC power supply malfunction (common)
Measure DC voltages at cardfile backplane buss points
- Test point
- backplane buss points on top of cardfile
- Expected
- +5 V, +15 V, –15 V
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Collimator controls relay board +5V supply not operational high voltage
DS4 indicator does not light on drives interface board, indicating +5RS (SOURCE) is not available to relays and feedback circuits
Ranked causes
- 1. 120AUX AC power not reaching power supply (common)
- 2. Relay K3 PUP not latched or pulling in (occasional)
- 3. Regulated +5V output not reaching drives interface board (occasional)
Verify 120MOM is on
- Test point
- 120MOM input
- Expected
- On state
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC Power Supply Fault/Sense board missing 120 VAC input high voltage
No 120 VAC detected at T1 pins 1–4 on DC Power Supply Fault/Sense board; prevents proper power supply sensing and fault detection
Ranked causes
- 1. Open in 120 VAC wiring to DC Power Supply board (occasional)
- 2. Failure of 8 kVA transformer or upstream AC power path (occasional)
- 3. Circuit breaker CB 304 open or faulty (occasional)
Verify 8 kVA transformer output voltage
- Test point
- Output terminals of 8 kVA transformer
- Expected
- 114 VAC to 126 VAC line-to-neutral
- Tools
- Multimeter
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry DC power supply voltages out of tolerance high voltage
DC voltages at Cardfile backplane bus points do not meet specification; +5V, +15V, or -15V supplies outside acceptable range or ripple excessive.
Ranked causes
- 1. Input 120 VAC not present at DC power supply terminals (common)
- 2. Shorts or excessive loading on DC supply outputs (common)
- 3. Circuit breaker CB 304 open or defective (occasional)
Measure 120 VAC at DC power supply input terminals
- Test point
- Input terminals of each DC power supply
- Expected
- 120 VAC
- Tools
- DMM
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Hard drive test reports superblock corruption error
Running option 0 (Execute test) or option 5 (Magnetic media non-destructive write/read test) on newly formatted, virgin, or recently destructively tested disk produces superblock corruption error message
Ranked causes
- 1. Newly formatted or virgin disk without operating system (common)
- 2. Disk previously had destructive write/read test run (occasional)
- 3. Genuine defective media with actual corruption (rare)
Run option 6 (Magnetic media destructive write/read test)
- Test point
- Hard drive diagnostic menu
- Expected
- Test completes without errors
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser button does not respond high voltage
Soft Bank 4 LED does not flash when pushing Laser button
Ranked causes
- 1. Faulty GPUC circuit card or signal lines (unknown)
- 2. GCPMI signal absent on PMDC card (unknown)
- 3. Drives Interface Board or XSC board failure (occasional)
Measure EXTIRQ pulse on GPUC circuit card
- Test point
- U33 pin 13
- Expected
- Pulse present per schematic T61G-615
- Tools
- Oscilloscope
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No communication to GALIL motion chip
99.9 displayed on Tilt display and 1904 displayed on Horiz. display; GALIL motion control chip not responding.
Ranked causes
- 1. Bad GALIL chip (rare)
- 2. Bad DUART - communication interface fault (rare)
- 3. Communication failure between GPUC and PMDC (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No horizontal motion in local mode high voltage
Carbon table top does not move horizontally when in local mode during patient positioning
Ranked causes
- 1. Resolver input signal absent or incorrect (common)
- 2. Potentiometer on CTCC out of adjustment (common)
- 3. Collision plate switch malfunction (occasional)
Measure resolver input signal
- Test point
- Pin 19 or 20 at location 3-01 on the CTCC
- Expected
- 2 kHz sine wave at approximately 2 Vrms (5.65 V p-p) present and constant
- Tools
- Oscilloscope
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No output from laser radiation
Laser has no output or 'Check Printer' message displayed on LCDI board display unit.
Ranked causes
- 1. Connector pins bent or loose on LCDI board or back panel (common)
- 2. Back panel or camera cable connection issues (common)
- 3. LCDI board failure (occasional)
Deenergize and inspect LCDI board connectors P3 and P4
- Test point
- P3 and P4 connectors on LCDI board
- Expected
- No bent pins, connectors fully seated
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scan frame does not move high voltage
Scan frame cannot be rotated by hand or motorized movement fails; mechanical interference or drive system malfunction
Ranked causes
- 1. Mechanical interference in scan frame (common)
- 2. Main Drive relay K306 cannot be energized (occasional)
- 4. Bad DUART communication chip (rare)
Inspect for mechanical interference
- Test point
- Scan frame rotation area
- Expected
- Free movement with no grinding or binding
7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Superblock corruption error during hard drive diagnostic
Error message 'Test termination due to superblock corruption or attempting to run this test on the image disk' appears when running option 0 or 5 on newly formatted, virgin, or recently destructively tested disk
Ranked causes
- 1. Disk newly formatted or virgin (common)
- 2. Disk had destructive write/read test run on it (common)
- 3. Genuinely defective media (rare)
Run magnetic media destructive write/read test
- Test point
- Hard drive diagnostic option 6
- Expected
- Test completes without errors
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tilt and horizontal displays show abnormal values
Tilt display shows '99.9' and horizontal display shows '1904' indicating motion control communication loss
Ranked causes
- 1. No communication to GALIL motion chip (occasional)
- 2. Bad GALIL chip in PMDC (rare)
- 3. Bad DUART in GPUC (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tilt display shows 99.9 and Horiz display shows 1904 high voltage
Both displays show specific error values: 99.9 on tilt and 1904 on horizontal
Ranked causes
- 1. No communication to GALIL motion chip (common)
- 2. Bad GALIL chip (occasional)
- 3. Bad DUART communication device (occasional)
Check error log for rotating frame errors or ANB on GPU
- Test point
- Error log
- Expected
- No rotating frame errors or ANB codes
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tilt display shows 99.9 and horizontal display shows 1904 caution
Tilt display reads 99.9 and horizontal display reads 1904, indicating communication or sensor failure
Ranked causes
- 1. No communication to GALIL motion chip (common)
- 2. Bad GALIL chip (occasional)
- 3. Bad DUART (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit has no vertical motion high voltage
Gantry support does not move up, down, or tilt when vertical touch panel controls are pressed
Ranked causes
- 1. Relays K401 and K402 not seated properly (common)
- 2. No 208 VAC at vertical motor (common)
- 3. VDRUP or VDRDN logic signals absent or invalid (common)
Verify relays K401 and K402 seating
- Test point
- K401 and K402 sockets
- Expected
- Relays recessed in sockets snugly
6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Carbon top horizontal position
Horizontal positioning of carbon top requires calibration after resolver replacement
Ranked causes
- 1. Resolver replacement (common)
- 2. Resolver connector disconnection (unknown)
Verify resolver inserted into bracket opening
- Test point
- resolver bracket opening
- Expected
- resolver fully seated
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Center program execution radiation
Execute center program to calculate centering constants from centering rod and air scans for X-ray focal spot alignment
Ranked causes
- 1. Initial centering constant calculation required (common)
- 2. Global centering constant exceeds tolerance (occasional)
Scan centering rod without metal pin
- Test point
- X-ray detector
- Expected
- Image processes without error
5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Console Intercom/Speakerphone high voltage
System setup and verification of Intercom/Speakerphone panel operation on X-ray console
Ranked causes
- 1. Initial calibration required (unknown)
- 2. REMOTE button LED not illuminating (unknown)
Power state: Gantry and Console
- Test point
- Power controls
- Expected
- Gantry power removed, Console OFF
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gantry tilt A/D converter offset high voltage
Adjustment of tilt potentiometer output to 0.0 VDC at neutral position; part of gantry calibration procedure
Ranked causes
- 1. Tilt A/D converter requires offset calibration (common)
- 2. Limit switches require verification after calibration (common)
Connect DMM to tilt A/D input
- Test point
- Pin 16 IC (location 16–09), TP3 on CTCC foil positive; TP1 on CTCC foil negative (ground)
- Expected
- DMM connected and ready for measurement
- Tools
- Digital multi-meter
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gantry tilt limit switch high voltage
Tilt limit switch requires adjustment after replacement or maintenance
Ranked causes
- 1. Tilt limit switch replaced or requires adjustment (common)
- 2. Gantry covers removed during service (common)
Apply gantry power
- Test point
- Gantry power supply
- Expected
- Power applied successfully
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tilt switch verification and adjustment caution
Gantry tilt position display accuracy and limit switch operation; requires protractor measurement against scan frame and potentiometer adjustment on CTCC card to match display to physical position.
Ranked causes
- 1. Tilt switches require verification or adjustment after replacement (unknown)
- 2. Tilt switches require adjustment after replacement or repositioning (unknown)
Display reading at –10 degree tilt position
- Test point
- Gantry touch panel display
- Expected
- –10 degrees, matches protractor measurement
- Tools
- Protractor
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardfile installation verification high voltage
Voltage and continuity checks after reinstalling cardfile assembly before returning to service
Ranked causes
- 1. Connection error at cardfile connector (occasional)
- 2. DC power supply malfunction (occasional)
Check cardfile dc voltages at backplane buss points
- Test point
- backplane top buss points
- Expected
- +5, +15, –15 V
- Tools
- digital multimeter
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Console keyswitch OFF disconnects relay power supply high voltage
Rotating console keyswitch to OFF position or actuating E–Stop disconnects 120AUX from the +5 volt relay power supply
Ranked causes
- 1. Normal operation – keyswitch or E–Stop activated as intended (common)
- 2. System-generated power down signal active (occasional)
Verify console keyswitch position
- Test point
- Console keyswitch
- Expected
- ON position
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Console keyswitch troubleshooting
Diagnostics for Console keyswitch and related power control circuits
Ranked causes
- 1. Keyswitch failure or J530 connector issue (occasional)
- 2. Pilot Control connection failure (occasional)
Run SMB diagnostic
- Test point
- Diagnostic system
- Expected
- No keyswitch faults reported
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
DC Power Supply U1 voltages missing or out of specification high voltage
Voltages at U1 (Digital supply IC) pins 11, 12, 15, 4 are absent or outside tolerance; indicates power supply failure or internal short
Ranked causes
- 1. Internal failure of DC power supply or U1 IC (common)
- 2. Short circuit on digital supply outputs (occasional)
Measure +5V (Digital) at U1 pin 11
- Test point
- U1 pin 11 on DC Power Supply Fault/Sense board
- Expected
- +5 VDC, +/- 0.25V
- Tools
- DMM
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry DC supply voltages out of tolerance high voltage
DC voltages at Cardfile backplane bus points exceed specified ranges: +5V, +15V, or -15V outside tolerance bands or ripple exceeds limits
Ranked causes
- 1. Excessive loading on DC supply outputs (common)
- 2. Failure of Cardfile DC power supply (occasional)
Verify 120 VAC at DC power supply input terminals
- Test point
- Input terminals of each DC power supply
- Expected
- 120 VAC present
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gantry SSR control signal absent or stuck caution
Logic signal to Solid State Relay (SSR) not switching between +5VDC and 0VDC when Gantry power is applied; prevents SSR activation and power delivery to gantry
Ranked causes
- 1. AC Control Chassis control logic fault or component failure (occasional)
- 2. Disconnected or damaged signal cable from Console (occasional)
Verify cable L7156 connection and continuity
- Test point
- Cable L7156 at P593 (Console) and J103 (Gantry)
- Expected
- Cable fully seated; continuity on all pins
- Tools
- Visual inspection; continuity tester
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Missing 120 VAC at DC power supply input high voltage
120 VAC not present at input terminals of DC power supply
Ranked causes
- 1. Fault in 120 VAC path through MK 301 and CB 304 (common)
- 2. Defective DC power supply (occasional)
Verify 120 VAC present at DC power supply input terminals
- Test point
- DC power supply input
- Expected
- 120 VAC present
- Tools
- DMM
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No tilt response from controls patient safety
Patient positioning tilt function does not respond when FWD or REV tilt switches are pressed; no movement on gantry
Ranked causes
- 1. Missing or incorrect enable command from CTCC to tilt motor driver (unknown)
- 2. Faulty Hall sensor in tilt motor feedback chain (unknown)
Measure T-VENA enable line voltage at rest
- Test point
- TP304
- Expected
- +5 VDC
- Tools
- DMM
7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power brush block replacement high voltage
Removal and replacement of slip ring brush blocks for power and signal in X-ray gantry system
Ranked causes
- 1. Brush block maintenance or failure (unknown)
- 2. Reinstall brush block (unknown)
Power isolated at wallbox
- Test point
- wallbox disconnect
- Expected
- gantry power OFF
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply conversion for 220V-240V primary source high voltage
Systems requiring modifications to power supply and control unit stickers when 220V-240V is the primary power source and Q-Console or Voxel-Q 120 vac access is not available.
Ranked causes
- 1. Power supply configured for incorrect input voltage (common)
- 2. Fuse panel orientation incorrect (occasional)
Verify fuses F4 and F5 are 1 amp type
- Test point
- Fuses F4 and F5 inside power input/power cord connection
- Expected
- LITTLEFUSE part # 312 001 or BUSSMAN AGC-1A
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Quick mechanical checks for scan frame issues caution
Initial visual and manual inspection of scan frame mechanics before detailed electrical diagnostics
Ranked causes
- 1. Mechanical interference in scan frame (common)
- 2. Gull-wing or collision switch malfunction (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
RDC converter failure or input signal loss
Resolver-to-digital converter (RDC) not producing valid 10-bit digital output; converter or input signal fault condition detected
Ranked causes
- 1. Fine RDC excitation lost or winding open (occasional)
- 2. Table position RDC failure (occasional)
Verify 2.5 kHz reference signal present at fine RDC input
- Test point
- Fine RDC reference input
- Expected
- 2.5 kHz signal present and stable
- Tools
- Oscilloscope
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Resolver output signal loss or incorrect amplitude
Fine sine/cosine resolver outputs (fsin, fcos) or reference signal missing or out of specification at RDC input
Ranked causes
- 1. Primary excitation voltage out of specification (common)
- 2. Phase error between primary and secondary windings (occasional)
Measure sine and cosine winding output amplitude
- Test point
- fsin, fcos outputs
- Expected
- 2 Vrms (max)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scan frame does not rotate smoothly or has loose cables caution
Operator detects resistance, grinding, or visible loose cables during scan frame rotation; potential clearance issues between rotating and stationary frame components.
Ranked causes
- 1. Loose cables connecting rotating and stationary frame (common)
- 2. Slip ring brush damage from incorrect rotation direction (occasional)
Manually rotate scan frame and visually inspect cables for slack or improper routing
- Test point
- All rotating frame cable connections and stationary frame attachment points
- Expected
- Cables taut with no visible slack; frame rotates smoothly without resistance
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scan frame does not rotate when commanded
Gantry fails to rotate when upper front cover is closed and interlock switch is in normal position
Ranked causes
- 1. Interlock switch failure preventing rotation (unknown)
- 2. Gantry mechanical obstruction or malfunction (unknown)
Verify interlock switch returns to center position
- Test point
- Interlock switch
- Expected
- Switch centered after pull-check
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Scan frame drive command signal abnormal caution
Oscilloscope monitoring of drive command voltage shows insufficient or unstable signal output
Ranked causes
- 1. Drives Off switch malfunction (occasional)
- 2. Servo controller fault (occasional)
Measure drive command signal with oscilloscope
- Test point
- A26
- Expected
- Swinging voltage approximately +/- 10V when manually rotating scan frame, +/- 10V
- Tools
- oscilloscope
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Scan frame rotation inspection caution
Visual and mechanical inspection of scan frame rotation after repair work; check for loose cables and frame clearance
Ranked causes
- 1. Loose cables or obstructions from service work (common)
- 2. Cable interference during frame rotation (occasional)
Visually inspect all scan frame cables for looseness
- Test point
- cable connections and routing
- Expected
- all cables secure with no visible slack or damage
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Solid state relay or bridge rectifier replacement
- VDRUP signal invalid or absent
- Vertical transducer removal hazards
- Zero interrupter signal not detected
- Zero vane interrupter malfunction
- 24 VAC pilot power check
- 512 error in error log
- 8 kVA transformer input voltage out of tolerance
- 999 error in error log
- Access gantry cover for servicing
- Access gantry covers for servicing
- Access to gantry components
- Align patient positioning laser beams
- Autovoice button LED not illuminating
- Autovoice connection lost or unstable
- Bridge rectifier replacement
- Bus error test failed
- Calibrate gantry tilt angle measurement
- Calibration: 19" display monitor
- Calibration: 19" display monitor adjustment
- Calibration: 19" display monitor aspect ratio
- Calibration: AC power supply voltages
- Calibration: ACQPLAN option enabling
- Calibration: AcQSim laser alignment to gantry isocenter
- Calibration: AcQSim laser positioning initial setup
- Calibration: AcQSim laser positioning verification
- Calibration: AcQSim laser system verification
- Calibration: Align ACQSIM side lasers
- Calibration: Autovoice audio level adjustment
- Calibration: Barker laser alignment
- Calibration: Beam rotation angle
- Calibration: brush block alignment
- Calibration: Center calibration procedure
- Calibration: Coarse zero position reference
- Calibration: Cosine waveform amplitude
- Calibration: Couch
- Calibration: CTSIM offset configuration
- Calibration: DC power supply voltages
- Calibration: Enter lateral offset into AcQSim software
- Calibration: EXACT Therapy Top installation and alignment
- Calibration: Gantry and patient support verification
- Calibration: Gantry tilt angle switches
- Calibration: Gantry tilt angle verification
- Calibration: Gantry tilt scale potentiometer (CTCC)
- Calibration: Gantry X-ray field alignment
- Calibration: Global Centering Constant
- Calibration: Horizontal belt tension and position
- Calibration: Horizontal position linear adjustment
- Calibration: Horizontal position resolver feedback
- Calibration: Horizontal position resolver feedback amplitude
- Calibration: Horizontal resolver feedback signal adjustment
- Calibration: Horizontal resolver feedback signal amplitude
- Calibration: Horizontal sine input amplitude
- Calibration: Horizontal sine input verification
- Calibration: Horizontal sine input voltage
- Calibration: Large focal spot centering
- Calibration: Laser alignment verification
- Calibration: Laser assembly mounting
- Calibration: Lateral laser offset
- Calibration: Lateral laser offset correction
- Calibration: Lateral tube alignment
- Calibration: Main Drive
- Calibration: Main drive excitation level
- Calibration: Mark X and Z axis center of scan plane
- Calibration: Motor pole offset adjustment
- Calibration: Patient support cover installation
- Calibration: Patient support cover mounting
- Calibration: Patient Support Cover removal and replacement
- Calibration: Patient support cover removal/replacement
- Calibration: Patient Support vertical and horizontal positioning
- Calibration: Power supply 12 VDC reconfiguration
- Calibration: Resolver error correction circuitry
- Calibration: Rotating frame lateral alignment
- Calibration: Scan drive adjustment
- Calibration: scan drive using MADCAP
- Calibration: Scan frame rotation and laser alignment
- Calibration: Signal brush block alignment
- Calibration: Telescoping base cover removal and reinstallation
- Calibration: Telescoping base cover top section removal
- Calibration: Tighten Patient Support bolts
- Calibration: Tilt potentiometer and limit switches
- Calibration: Tilt switches verification and adjustment
- Calibration: Ultra Q floppy disk drive installation
- Calibration: vertical transducer
- Calibration: X-ray field alignment
- Calibration: X-ray tube axial alignment
- Calibration: X-ray tube axial mechanical alignment
- Calibration: X-ray tube lateral alignment
- Cannot control gantry from operator console during service
- Card cage in extended position during gantry tilt
- Card cage retraction requirement
- Ceiling-mounted LTU may interfere with injector arms
- Collision plate not functioning
- Collision plate not responding
- Collision Pleate activation check
- Cone gasket deterioration or uneven compression
- Configuration: AcQSim network and system settings
- Counter test failed
- Dangerous voltage exposure with covers removed
- Dangerous voltages exposed during gantry cover removal
- Dangerous voltages exposed with gantry covers removed
- DC voltages out of tolerance on Cardfile backplane
- Disk format timeout
- Drive Command signal absent or incorrect
- Drive verification using ring diagnostic
- DUART operation test
- DUART timer and interrupt test
- During removal/replacement operations on telescoping base covers
- Electrical hazard from exposed voltages
- Electrostatic damage to PCB components
- Equipment covers open safety risk
- Error code 512 in error log
- Error code 999 in error log
- Excessive DC supply output loading or shorts
- Exit madcap program
- Film and reference marker placement for calibration
- Final laser alignment verification
- Floppy drive not recognized after motherboard replacement
- Fluid leak into detector electronics
- Gantry 480 VAC 3-phase input voltage out of range
- Gantry cover and panel safety
- Gantry cover removal and securing
- Gantry covers exposed high voltage
- Gantry covers removed exposing dangerous voltages
- Gantry covers removed operation
- Gantry does not power down
- Gantry laser covers need securing before calibration
- Gantry lasers misaligned
- Gantry LOCAL START procedure
- Gantry operation verification after component replacement
- Gantry power and warm-up setup
- Gantry power application
- Gantry tilt limit switch replacement
- Gantry tilt switch replacement/adjustment procedure
- Gantry tilting motion verification
- Ground potential shift at AC Control Chassis
- Gull-wing, collision or drives switch malfunction
- High voltage exposure hazard during cover removal
- Horizontal drive motor replacement - vibration after installation
- Injury from sharp upper gull wing panels
- Injury risk when lifting LTU during installation
- Installation: Mount lateral laser assembly on column
- Intercom/Speakerphone panel LED not illuminating
- Intercom/Speakerphone REMOTE LED not illuminating
- Interlock switch does not prevent frame rotation
- Interlock switch does not prevent gantry rotation
- Interlock switch not preventing frame rotation
- Interlock switch not preventing scan frame rotation
- Laser beam adjustment interaction
- Laser beam alignment check
- Laser beam focus adjustment
- Laser beam misaligned or not operational
- Laser radiation and rotating machinery injury hazard
- Laser radiation hazard during positioning
- Limit switch replacement procedure
- Loose connector P402 at J402
- LTU mounting interference with injector arm or gull wing
- LTU mounting interferes with injector arm or gull wing
- Memory test failed above address 0x20000
- Memory test failed below address 0x20000
- Missing or low 480 VAC 3-phase at transformer input
- Monitor cannot achieve correct aspect ratio
- Monitor cannot achieve specified aspect ratio
- Monitor cannot be adjusted to specified aspect ratio
- MPC test failed
- Negative limit switch does not interrupt tilting motion
- Obstruction in X-ray beam path
- Opening gantry covers for scan drive control service
- Optical disk diagnostic fails due to unwritten media
- Over-temperature shutdown activated
- Patient support cover removal
- Patient support cover removal and replacement
- Patient support covers removal and replacement
- Patient support covers removal/replacement
- Patient support covers require removal and replacement
- Patient Support not centered in scan field
- Patient Support not perpendicular to gantry
- Patient Support not tracking properly during vertical travel
- Personnel injury from lifting sagittal laser
- Power-up reset signal abnormal
- Reconfigure 24 VDC power supply
- Remove and replace patient support covers
- Replace 3 KVA scan drive transformer
- Replace 3KVA scan drive transformer
- Replace bridge rectifier
- Replace bridge rectifier on scan drive control
- Replace gantry limit switch
- Replace limit switch assembly
- Replace limit switch on patient support
- Replace power brush block
- Replace solid state relay (SSR)
- Replace solid state relay on scan drive control
- Replace vertical limit switch
- Resolver replacement
- Rotating frame errors or ANB indication present
- Rotating frame hazard during operation
- Rotating frame inspection and rotation
- Rotating machinery hazard during operation
- S-Bus board pins damaged during installation
- Safety: High voltage and heavy equipment handling
- Sagittal laser position deviation
- Sagittal laser readout displays incorrect positive/negative numbers
- Scan drive overspeed condition
- Scan drive position error
- Scan frame does not begin CCW rotation after stop
- Scan frame does not decelerate to stop at zero position
- Scan frame home position rotation
- Scan frame motor initialization required
- Scan frame position not established at power-up
- Scan frame rotates with gantry cover open
- Scan frame rotates with upper front cover open
- Scan frame rotation causes slip ring brush damage
- Scan frame rotation hazard
- Scan frame will not move or has mechanical interference
- Scan position measurement fault condition
- Scissor-action mechanism hazard with support base covers removed
- Sharp edges on gantry covers hazard
- Signal brush block removal and replacement
- Solid state relay replacement
- Static electricity damage to circuit boards
- System does not operate with front covers open
- System verification after cardfile replacement
- Telescoping base cover replacement
- Telescoping base covers removal and replacement
- Transducer wire damage during connection/disconnection
- Transducer wire snap hazard during connection
- Transformer replacement procedure
- Two service persons required for cover procedures
- Unexpected exception
- Unexpected exception sitting in pROBE
- Unit does not start after power on
- VDRDN signal invalid or absent
- Vertical limit switch diagnostic check
- Vertical touch panel malfunction
- Vertical transducer operation verification
- Vertical transducer replacement and calibration
- Visual test execution on display board
- Work on gantry with upper panels open
- Work on X-ray gantry with covers removed
- X-ray tube axial mechanical alignment
- Zero interrupter not triggering
241 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0 | Did not boot; never made to beginning of code | Constant ACFAIL from RCB/DIB; bad software or hardware failure | critical |
1 | Failed RAM Address write test | Failure during RAM address write test | critical |
11 | DSP memory test failure | Acquisition time-out error or memory corruption during DSP diagnostics | error |
13 | Zero Reference ADCA Test failure | DSP power-up diagnostics detect zero reference error in analog-to-digital converter A | error |
14 | Positive 15 V rail test failure | DSP diagnostics detect positive 15 V supply out of specification | error |
14 | Unexpected PITR interrupt during testing | Unexpected PITR interrupt detected during diagnostic testing | error |
15 | Interrupt mask error | SR mask does not match expected mask value during interrupt testing | error |
15 | Negative 15 V rail test failure | DSP diagnostics detect negative 15 V supply out of specification | error |
16 | ADCB Reference Test failure | DSP power-up diagnostics detect reference error in analog-to-digital converter B | error |
16 | SCI TE Interrupt Timeout | Serial communication interface transmit enable interrupt failed to occur | error |
17 | 5 V rail test failure | DSP diagnostics detect 5 V supply out of specification | error |
17 | SCI TC Interrupt Timeout | Serial communication interface transmit complete interrupt failed to occur | error |
18 | SCI RDFRF Interrupt Timeout | Serial communication interface receive data register full interrupt failed to occur | error |
18 | ADCA Reference Test failure | DSP power-up diagnostics detect reference error in analog-to-digital converter A | error |
19 | Starter 10 V Reference Test failure | DSP diagnostics detect 10 V starter reference out of specification | error |
19 | SCI Mask Error | SCI interrupt mask does not match expected configuration | error |
1900 | Cannot setup ANB Driver | ANB Driver initialization fails during startup | critical |
1903 | Cannot initialize Frame Galil | Frame Galil motor controller fails to initialize | critical |
1904 | Cannot initialize Couch Galil | Couch Galil motor controller fails to initialize | critical |
1908 | Cannot start GPU Manager Process | GPU Manager process fails to start during boot | critical |
2 | Failed RAM 5555hex test | Bad bit detected during RAM 5555hex pattern test | critical |
20 | Temperature Test failure | DSP diagnostics detect out-of-range temperature sensor output | error |
20 | Soft Watchdog Error | Software watchdog timer detected error condition | error |
21 | Filament DAC Zero Output Test failure | Filament digital-to-analog converter does not output zero when commanded | error |
21 | Watchdog Intermit Error | Watchdog interrupt error during diagnostic test | error |
22 | Soft Watchdog Error Reset | Software watchdog triggered error and system reset | error |
22 | Filament DAC Full Scale Output Test failure | Filament DAC does not output full-scale voltage when commanded | error |
222 | Spurious Interrupt Error | Fatal error in gantry control system before GPMN process starts or within error process itself | critical |
23 | ROM Check Sum Error | ROM memory checksum verification failed during power-up diagnostics | critical |
23 | Filament DAC Linearity Test failure | Filament DAC output does not scale linearly with input commands | error |
24 | RSR Error | RSR (register) error detected during diagnostic test | error |
24 | MADAC Zero Output Test failure | MA (tube current) DAC does not output zero when commanded | error |
25 | Soft Watchdog Timeout Error | Software watchdog timer exceeded timeout threshold | error |
25 | MA DAC Full Scale Output Test failure | MA DAC does not output full-scale voltage when commanded | error |
26 | MADAC Linearity Test failure | MA DAC output does not scale linearly with input commands | error |
27 | KV DAC Zero Output Test failure | High voltage (kV) DAC does not output zero when commanded | error |
28 | KV DAC Full Scale Output Test failure | KV DAC does not output full-scale voltage when commanded | error |
29 | KV DAC Linearity Test failure | KV DAC output does not scale linearly with input commands | error |
3 | Failed RAM AAAAhex test | Bad RAM detected during AAAAhex pattern test | critical |
30 | Unexpected Exception | Unexpected exception condition detected during diagnostics | error |
333 | Watchdog Timer Timeout Error | Fatal error when watchdog timer expires in gantry control system before GPMN process starts or within error process itself | critical |
4 | Failed NOV RAM test | Failure during NOV RAM diagnostic test | critical |
444 | System exception error | Exception error detected during XSC power diagnostics | error |
5 | Failed ROM test | Failure during ROM integrity check | critical |
512 | GPU Crash in error logger | GPU microprocessor failure or overspeed condition during scan operation | critical |
512 | GPU crash - graphics processing unit or motion control processor failure | GPU microprocessor malfunction, bad GALIL motion chip, bad DUART, or communication loss to motion chip | critical |
555 | Bus address error | Bus address error detected during XSC power diagnostics | error |
6 | Failed CPUFAIL interrupt level 7 test | Constant ACFAIL from RCB/DIB; CPU interrupt level 7 failure | critical |
7 | Failed interrupt level 6 test | Fault circuitry failure detected during interrupt level 6 test | error |
8 | Failed timer Z48 test | Failure during timer Z48 diagnostic test | error |
9 | Failed sync test | Sync test failure during diagnostics | warning |
999 | Command Shell error - system command interpreter or communication failure | Command shell communication loss or command processing failure | error |
999 | Command Shell error in error logger | Loss of communication or command processing failure in scan drive control system | error |
A | Failed interrupt level 3 test | Interrupt level 3 failure during diagnostic test | error |
ACQSim functions not active | ACQSim software features are not enabled in the Voxel-Q system | ACQSim functions not activated after software installation | error |
B | Failed interrupt level 2 test | Interrupt level 2 failure during diagnostic test | error |
C | Failed interrupt level 1 test | Interrupt level 1 failure during diagnostic test | error |
can't identify plot on display | System unable to recognize or validate plot data on display output | Plot validation check after rendering | error |
can't start acceldma on heurikon | Unable to initiate acceldma process on Heurikon hardware | Scan start command issued | error |
can't start plot on display | Unable to initiate plot rendering on display module | Display update command | warning |
D | Failed DUART serial test | Serial port DUART chip test failure | error |
datpos | Drives at position | Scan frame and table have reached commanded positions | info |
dits counter not reset to 0 error | DITS counter failed to reset to zero at initialization | System startup or scan preparation sequence | warning |
divide by 0 error, plot file not updated | Mathematical division by zero encountered; plot file not written as a result | Plot calculation or normalization | error |
drvflt | Drives fault - directly shuts down hardware | Various drive-related faults detected | critical |
E | Failed bus address error test | Bus address error detection failure | error |
Error logged when power switch turned OFF with tube heat abo | Gantry shut down immediately due to unsafe shutdown procedure | Console power switch turned OFF when tube heat above 35% | warning |
error reading scan data | Scan data file could not be read or accessed | Data retrieval after scan completion | error |
F | Wouldn't boot to applications side | Failure to boot to applications side of system | critical |
fatal error message | Unrecoverable system error encountered | Critical software or hardware fault during operation | critical |
hdperr | Horizontal drive position error | Table actual position differs from commanded reference | error |
hdract | Horizontal drive active - status signal indicating proper operation | Horizontal drive is functioning normally | info |
hpmflt | Horizontal position measurement fault | Table position RDC input signal lost or RDC failure | critical |
illegal signal | Received signal does not conform to expected format or protocol | Signal reception and validation | warning |
lread error of bad.dat | Failed to read bad.dat file | Loading bad data records | error |
lwrite error bad.dat | Failed to write bad data records to bad.dat file | Recording out-of-range or invalid data points | error |
lwrite error to raw data file | Failed to write scan data to raw data file | Saving raw scan data to disk | error |
lwrite of to plot.dat error | Failed to write processed data to plot.dat file | Saving processed plot data to disk | error |
plot timeout | Plot generation or rendering did not complete within expected time | Plot rendering operation | warning |
scan positioning timeout | Scan positioning operation did not complete within expected timeframe | Attempted to position scan but system failed to respond within timeout period | error |
sdovsp | Scan drive overspeed | Scan frame rotation speed exceeds safe limit | warning |
sdperr | Scan drive position error | Scan frame actual position differs from commanded reference | error |
sdract | Scan drive active - status signal indicating proper operation | Scan drive is functioning normally | info |
spmflt | Scan position measurement fault | Fine RDC input signal lost or RDC failure; excitation or winding loss | critical |
Superblock corruption | Test termination due to superblock corruption or attempting to run test on image disk | Running option 0 or option 5 on newly formatted, virgin disk, or disk with previous destructive test run | warning |
Test 1 | SCI DATA communication error | SCI DATA timeout or communication integrity error during boot | error |
Test 10 | Level 4 PITR interrupt | Level 4 PITR interrupt not properly cleared | error |
Test 11 | PITR count out of range | PITR timer count value outside acceptable range | error |
Test 12 | Multiple interrupts in timer mode | More than one interrupt detected during timer mode test | warning |
Test 13 | Unexpected DUART interrupt | DUART status interrupt detected during test | warning |
Test 2 | SCI Data mismatch error | Transmitted and received SCI data do not match | error |
Test 3 | NOVRAM data error | NOVRAM data read/write failure or checksum mismatch | error |
Test 4 | MC68332 RAM error | MC68332 processor RAM test failure | error |
Test 5 | Static memory data error | Static memory write/read test failure | error |
Test 6 | Static RAM march memory error | Static RAM march pattern test failure | error |
Test 7 | Bus error | CPU system bus communication error | error |
Test 8 | DUART no level 2 interrupt | DUART level 2 interrupt signal not detected | error |
Test 9 | DUART counter interrupt status bit error | DUART counter interrupt status incorrect during test | error |
Test termination due to superblock corruption | Hard drive diagnostic terminated because superblock is corrupted or test attempted on image disk | Running diagnostic option 0 or 5 on newly formatted, virgin, or recently destructively tested disk | warning |
timeout on signal from acceldma done | Expected completion signal from acceldma module not received | Scan execution | error |
timeout on signal from acceldma to start scan | Expected start signal from acceldma module not received within timeout | Scan initiation sequence | error |
too many beginning points out of range | Initial data points at scan start exceed acceptable range | Data validation at scan beginning | warning |
too many points out of range | Excessive number of data points exceed acceptable value range | Data validation after scan or plot generation | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.