Philips PICKER PQ: troubleshooting and error codes

The ServiceForge knowledge base holds 301 documented symptoms for the Philips PICKER PQ, with 441 ranked causes, 767 diagnostic checks with test points and expected values, and 103 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Unit does not tilt patient safety

Gantry tilt function not responding to operator commands at touch panel

Ranked causes

  1. 1. Tilt motor circuit breakers CB303 or CB302 off or tripped (common)
  2. 2. Loose or disconnected cable connectors (common)
  3. 3. Upper front gantry cover not closed (occasional)
  4. 4. Lift switches open or not reset (common)
  5. 5. Missing or incorrect DC power supply voltages (occasional)
  6. 6. 208 VAC supply missing or disconnected (occasional)
  7. 7. Right tilt motor no DC voltage (occasional)
  8. 8. Left tilt motor no DC voltage (occasional)
First check

Verify circuit breaker CB303 position

Test point
CB303 on AC Control chassis
Expected
ON

9 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Missing or corrupted data in CT image acquisition

Data blocks missing from image; visible data patterns indicating detector or electronics failures; image quality degraded

Ranked causes

  1. 1. Analog switch malfunction in detector module (rare)
  2. 2. Single detector element failure (occasional)
  3. 3. Specific detector channel malfunction (occasional)
  4. 4. PSDC card failure or memory corruption (occasional)
  5. 5. Cable malfunction or poor connection between detector and PSDC (occasional)
  6. 6. VFSC card failure (rare)
  7. 7. Bad select cable causing multiple channel failures (occasional)
First check

Perform front end background check scan using diagscans

Test point
Diagnostic scan software
Expected
Scan completes without errors; compare output to baseline

5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gantry does not power on high voltage

Gantry power system non-responsive; LEDs on AC Control Panel not illuminated or circuit breakers tripped

Ranked causes

  1. 1. Gantry key switch not in NORMAL position (common)
  2. 2. Circuit breaker tripped or open (common)
  3. 3. Fuse failure (occasional)
  4. 4. E-STOP activated or disconnected (occasional)
  5. 5. Console and Patient Support interconnect cables disconnected (occasional)
  6. 6. AC Control Chassis fault (occasional)
First check

Verify required LEDs illuminated on AC Control Panel

Test point
AC Control Panel LEDs
Expected
CONTROL POWER, SYSTEM ENABLED, DC OK, DRIVES all ON

6 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No vertical motion high voltage

Vertical positioning controls do not produce vertical motion on gantry

Ranked causes

  1. 1. Vertical touch panel malfunction or disconnection (common)
  2. 2. TILT enable command not reaching CTCC (common)
  3. 3. TVENB enable signal failure (common)
  4. 4. Vertical potentiometer open circuit (occasional)
  5. 5. CTCC card failure (occasional)
  6. 6. GPUC (GPU card) failure (rare)
First check

Measure AC voltage on connector J415 pins 1-2, 1-3, 1-4 while pressing vertical controls

Test point
J415 pins 1-2, 1-3, 1-4
Expected
120 VAC
Tools
DMM set to 300 VAC

7 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No vertical motion of patient support high voltage

Patient support does not move vertically; tilt control may also be affected

Ranked causes

  1. 1. Objects inhibiting vertical motion (common)
  2. 2. Tilt not at 0 degrees or support position too high/low (common)
  3. 3. Soft limit switch malfunction (occasional)
  4. 4. Connector or ribbon cable fault (occasional)
  5. 5. CTCC (gantry control card) failure (occasional)
  6. 6. Loss of 208 VAC power to vertical motor (occasional)
First check

Inspect for obstructions around patient support

Test point
Entire vertical motion path
Expected
No objects present

9 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display not functioning or image not visible

Monitor shows no image or display is blank; keyboard and monitor connections may be affected

Ranked causes

  1. 1. Keyboard not plugged in or internal cable disconnected (common)
  2. 2. Contrast control incorrect setting (common)
  3. 3. Monitor power switch off (common)
  4. 4. Monitor fuse open or blown (occasional)
  5. 5. Power or signal cables disconnected (common)
First check

Keyboard connection status

Test point
Keyboard connector
Expected
Plugged in and internal cable connected

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive Off signal HIGH - main drive disabled caution

Drive Off signal remains HIGH, preventing K306 energization and disabling MAIN DRIVE ON function

Ranked causes

  1. 1. GPU microprocessor not functional (common)
  2. 2. Overspeed condition detected (occasional)
  3. 3. Scan position measurement fault (occasional)
  4. 4. Power up reset signal LOW (occasional)
  5. 5. Over temperature condition (occasional)
First check

Verify Drive Off signal state

Test point
B71, slot 19
Expected
LOW when normal operation intended

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Drive Off signal HIGH prevents main drive energization high voltage

K306 relay cannot be energized; MAIN DRIVE ON disabled due to Drive Off signal

Ranked causes

  1. 1. GPU Microprocessor not functional (rare)
  2. 2. Scan frame overspeed detected (occasional)
  3. 3. Scan Position Measurement Fault (occasional)
  4. 4. Power Up Reset signal LOW (rare)
  5. 5. Over Temperature condition (occasional)
First check

Locate and monitor Drive Off signal at B71, slot 19

Test point
B71, slot 19
Expected
LOW (signal inactive; drive enabled)

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Erratic horizontal motion patient safety

Table top movement is unpredictable or erratic along horizontal axis during positioning

Ranked causes

  1. 1. Mechanical belt or motor misalignment (common)
  2. 2. Faulty potentiometer or resolver (common)
  3. 3. Horizontal Power Amp output failure (occasional)
  4. 4. Horizontal Servo Motor failure (occasional)
  5. 5. Voltage regulation or signal integrity issue (occasional)
First check

Run diagnostics

Test point
System diagnostics
Expected
Diagnostics complete without errors

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No horizontal motion, local mode

Table or gantry will not move horizontally when using local mode touch panels; positioning system unresponsive

Ranked causes

  1. 1. Fuse F402 open (common)
  2. 2. Microswitch failure or misalignment (common)
  3. 3. Limit switch cable open or short (occasional)
  4. 4. Potentiometer failure or incorrect resistance (occasional)
  5. 5. Resolver-to-digital converter malfunction (occasional)
First check

Run cocal and codd diagnostics; verify fuse F402

Test point
Fuse F402
Expected
Good

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vertical motion in one direction only patient safety

Support moves vertically in only one direction (up or down), not both

Ranked causes

  1. 1. Invalid VDRUP or VDRDN signal (common)
  2. 2. Faulty tilt panel (occasional)
  3. 3. Defective limit switch or cable (occasional)
  4. 4. Faulty relay K401 or K402 (rare)
  5. 5. Motor power issue or brake rubbing (common)
First check

Verify VDRUP signal validity

Test point
SSR403 logic
Expected
Valid logic level

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

999 error in error logger with Command Shell error high voltage

999 error appears in erl error logger; Command Shell error condition; scan frame may not respond to commands

Ranked causes

  1. 1. No communication to GALIL motion chip (common)
  2. 2. Bad GALIL chip (common)
  3. 3. Bad communication between GPUC and PMDC (occasional)
  4. 4. Bad DUART (occasional)
First check

Visually inspect scan frame and mechanical components

Test point
Scan frame assembly
Expected
No mechanical interference or damage

9 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Autovoice system not functioning

Speaker or intercom buttons (Local, Remote, Mute, Answer) not responding; associated LEDs not illuminating when buttons pressed

Ranked causes

  1. 1. Poor speaker connections (common)
  2. 2. Broken PCB to SYS MON/Autovoice cable #L7855 (occasional)
  3. 3. Broken PCB to Autovoice and Display cable #L7856 (occasional)
  4. 4. Autovoice PCB failure or LED circuit failure (rare)
First check

CPU diagnostics test

Test point
System diagnostics
Expected
Diagnostics completed successfully

5 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Scan drive offset and phase adjustment radiation

RDC offset and Motor Pole offset switches require adjustment to ensure X-ray beam alignment and smooth scan drive rotation without back EMF-induced stalling or backward rotation

Ranked causes

  1. 1. RDC offset switches incorrectly set (common)
  2. 2. Motor Pole offset switches incorrectly phased (common)
  3. 3. Resolver replaced or PMDC board replaced (occasional)
  4. 4. PMDC with resolver error correction circuitry installed (occasional)
First check

Power removal and access preparation

Test point
System wallbox and gantry covers
Expected
System powered off; gantry covers open; cardfile access cover open; DRIVES switch OFF; service keyswitch NORMAL; three motor drive fuses removed

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Display shows 99.9 on Tilt and 1904 on Horiz display caution

Tilt display reads 99.9 and Horizontal display shows 1904 simultaneously

Ranked causes

  1. 1. No communication to GALIL motion chip (common)
  2. 2. Bad GALIL chip (occasional)
  3. 3. Bad communication between GPUC and PMDC (occasional)
  4. 4. Bad DUART chip (occasional)
First check

Check GALIL chip communication using oscilloscope

Test point
Communication signal lines
Expected
Stable communication signal present

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect display, vertical

Vertical display on patient positioning system is incorrect or not updating properly

Ranked causes

  1. 1. Vertical potentiometer (transducer) malfunction (common)
  2. 2. Cable BL7161 connection failure or open circuit (common)
  3. 3. CTCC (Control Card) failure (occasional)
  4. 4. GPUC (Graphics Processing Unit Card) failure (rare)
First check

Measure vertical potentiometer resistance

Test point
Connector P405 sockets 1 and 2
Expected
500 ohms
Tools
DMM

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Laser power supply fault high voltage

Laser does not output 120 VAC or input power is incorrect

Ranked causes

  1. 1. No 120 VAC input at laser connector (common)
  2. 2. Laser Power Supply Assembly output failure (occasional)
  3. 3. Laser cable open or damaged (rare)
  4. 4. No +5 VDC at laser input (common)
First check

Measure 120 VAC at J2 input connector

Test point
J2 pins 5 (hot) and 1 (neutral)
Expected
120 VAC
Tools
Multimeter

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No vertical motion on gantry high voltage

Gantry support does not move up, down, or tilt when vertical touch panel controls are pressed

Ranked causes

  1. 1. Relays K401 and K402 not seated properly in sockets (common)
  2. 2. No 208 VAC supply to vertical motor (common)
  3. 3. Invalid VDRUP or VDRDN signal (occasional)
  4. 4. No 208 VAC at motor terminals (occasional)
First check

Verify relay K401 and K402 seating

Test point
K401 and K402 sockets
Expected
Relays recessed snugly in sockets

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Vertical motion erratic high voltage

Support moves erratically up or down during patient positioning

Ranked causes

  1. 1. 208 VAC supply voltage absent or incorrect at motor (common)
  2. 3. 208 VAC supply voltage absent or incorrect at PSCC (common)
  3. 4. Relays K401 or K402 malfunction (occasional)
  4. 6. Bad limit switch cable or defective switch (occasional)
First check

Verify relay seating

Test point
K401 and K402
Expected
Relays recessed in their sockets

12 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

3 KVA transformer replacement high voltage

Procedure to remove and replace the 3 KVA transformer (80 pounds) in the gantry assembly

Ranked causes

  1. 1. Transformer requires replacement (unknown)
  2. 2. Remove transformer from gantry (unknown)
  3. 3. Install replacement transformer (unknown)
First check

Verify 230 VAC 3-phase power on L1, L2, L3

Test point
Terminal block on front of transformer
Expected
230 vac 3-phase power present on l1, l2 and l3
Tools
DVM

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

512 error in error logger - GPU crash caution

System displays error code 512 in the error log (erl), indicating GPU processor failure or malfunction affecting scan drive operation

Ranked causes

  1. 2. Bad GALIL motion chip (occasional)
  2. 3. Bad DUART (occasional)
  3. 4. No communication to GALIL motion chip (occasional)
First check

Check error log for rotating frame errors or ANB on GPU

Test point
erl error logger
Expected
No 512 error code present

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blower inoperative or running intermittently caution

Gantry cooling blower does not run continuously or fails to start; may cause thermal issues. Single blower failure or capacitor fault.

Ranked causes

  1. 1. Blower capacitor failed or capacitance out of range (common)
  2. 2. Fuse F315 or F317 open (occasional)
  3. 3. Blower motor winding or bearing failure (occasional)
First check

Verify fuse F315 (blowers) continuity

Test point
Fuse F315 on AC Control Chassis
Expected
Continuity (less than 1 ohm)
Tools
Ohmmeter or continuity tester

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cardfile circuit boards require replacement high voltage

DC voltages (+5, +15, -15V) absent at backplane buss points; suspected cardfile or power supply failure

Ranked causes

  1. 1. Cardfile circuit board failure or damage (unknown)
  2. 2. Cardfile connection errors (common)
  3. 3. DC power supply malfunction (common)
First check

Measure DC voltages at cardfile backplane buss points

Test point
backplane buss points on top of cardfile
Expected
+5 V, +15 V, –15 V

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Collimator controls relay board +5V supply not operational high voltage

DS4 indicator does not light on drives interface board, indicating +5RS (SOURCE) is not available to relays and feedback circuits

Ranked causes

  1. 1. 120AUX AC power not reaching power supply (common)
  2. 2. Relay K3 PUP not latched or pulling in (occasional)
  3. 3. Regulated +5V output not reaching drives interface board (occasional)
First check

Verify 120MOM is on

Test point
120MOM input
Expected
On state

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

DC Power Supply Fault/Sense board missing 120 VAC input high voltage

No 120 VAC detected at T1 pins 1–4 on DC Power Supply Fault/Sense board; prevents proper power supply sensing and fault detection

Ranked causes

  1. 1. Open in 120 VAC wiring to DC Power Supply board (occasional)
  2. 2. Failure of 8 kVA transformer or upstream AC power path (occasional)
  3. 3. Circuit breaker CB 304 open or faulty (occasional)
First check

Verify 8 kVA transformer output voltage

Test point
Output terminals of 8 kVA transformer
Expected
114 VAC to 126 VAC line-to-neutral
Tools
Multimeter

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gantry DC power supply voltages out of tolerance high voltage

DC voltages at Cardfile backplane bus points do not meet specification; +5V, +15V, or -15V supplies outside acceptable range or ripple excessive.

Ranked causes

  1. 1. Input 120 VAC not present at DC power supply terminals (common)
  2. 2. Shorts or excessive loading on DC supply outputs (common)
  3. 3. Circuit breaker CB 304 open or defective (occasional)
First check

Measure 120 VAC at DC power supply input terminals

Test point
Input terminals of each DC power supply
Expected
120 VAC
Tools
DMM

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Hard drive test reports superblock corruption error

Running option 0 (Execute test) or option 5 (Magnetic media non-destructive write/read test) on newly formatted, virgin, or recently destructively tested disk produces superblock corruption error message

Ranked causes

  1. 1. Newly formatted or virgin disk without operating system (common)
  2. 2. Disk previously had destructive write/read test run (occasional)
  3. 3. Genuine defective media with actual corruption (rare)
First check

Run option 6 (Magnetic media destructive write/read test)

Test point
Hard drive diagnostic menu
Expected
Test completes without errors

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Laser button does not respond high voltage

Soft Bank 4 LED does not flash when pushing Laser button

Ranked causes

  1. 1. Faulty GPUC circuit card or signal lines (unknown)
  2. 2. GCPMI signal absent on PMDC card (unknown)
  3. 3. Drives Interface Board or XSC board failure (occasional)
First check

Measure EXTIRQ pulse on GPUC circuit card

Test point
U33 pin 13
Expected
Pulse present per schematic T61G-615
Tools
Oscilloscope

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No communication to GALIL motion chip

99.9 displayed on Tilt display and 1904 displayed on Horiz. display; GALIL motion control chip not responding.

Ranked causes

  1. 1. Bad GALIL chip (rare)
  2. 2. Bad DUART - communication interface fault (rare)
  3. 3. Communication failure between GPUC and PMDC (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No horizontal motion in local mode high voltage

Carbon table top does not move horizontally when in local mode during patient positioning

Ranked causes

  1. 1. Resolver input signal absent or incorrect (common)
  2. 2. Potentiometer on CTCC out of adjustment (common)
  3. 3. Collision plate switch malfunction (occasional)
First check

Measure resolver input signal

Test point
Pin 19 or 20 at location 3-01 on the CTCC
Expected
2 kHz sine wave at approximately 2 Vrms (5.65 V p-p) present and constant
Tools
Oscilloscope

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No output from laser radiation

Laser has no output or 'Check Printer' message displayed on LCDI board display unit.

Ranked causes

  1. 1. Connector pins bent or loose on LCDI board or back panel (common)
  2. 2. Back panel or camera cable connection issues (common)
  3. 3. LCDI board failure (occasional)
First check

Deenergize and inspect LCDI board connectors P3 and P4

Test point
P3 and P4 connectors on LCDI board
Expected
No bent pins, connectors fully seated

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scan frame does not move high voltage

Scan frame cannot be rotated by hand or motorized movement fails; mechanical interference or drive system malfunction

Ranked causes

  1. 1. Mechanical interference in scan frame (common)
  2. 2. Main Drive relay K306 cannot be energized (occasional)
  3. 4. Bad DUART communication chip (rare)
First check

Inspect for mechanical interference

Test point
Scan frame rotation area
Expected
Free movement with no grinding or binding

7 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Superblock corruption error during hard drive diagnostic

Error message 'Test termination due to superblock corruption or attempting to run this test on the image disk' appears when running option 0 or 5 on newly formatted, virgin, or recently destructively tested disk

Ranked causes

  1. 1. Disk newly formatted or virgin (common)
  2. 2. Disk had destructive write/read test run on it (common)
  3. 3. Genuinely defective media (rare)
First check

Run magnetic media destructive write/read test

Test point
Hard drive diagnostic option 6
Expected
Test completes without errors

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tilt and horizontal displays show abnormal values

Tilt display shows '99.9' and horizontal display shows '1904' indicating motion control communication loss

Ranked causes

  1. 1. No communication to GALIL motion chip (occasional)
  2. 2. Bad GALIL chip in PMDC (rare)
  3. 3. Bad DUART in GPUC (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tilt display shows 99.9 and Horiz display shows 1904 high voltage

Both displays show specific error values: 99.9 on tilt and 1904 on horizontal

Ranked causes

  1. 1. No communication to GALIL motion chip (common)
  2. 2. Bad GALIL chip (occasional)
  3. 3. Bad DUART communication device (occasional)
First check

Check error log for rotating frame errors or ANB on GPU

Test point
Error log
Expected
No rotating frame errors or ANB codes

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tilt display shows 99.9 and horizontal display shows 1904 caution

Tilt display reads 99.9 and horizontal display reads 1904, indicating communication or sensor failure

Ranked causes

  1. 1. No communication to GALIL motion chip (common)
  2. 2. Bad GALIL chip (occasional)
  3. 3. Bad DUART (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit has no vertical motion high voltage

Gantry support does not move up, down, or tilt when vertical touch panel controls are pressed

Ranked causes

  1. 1. Relays K401 and K402 not seated properly (common)
  2. 2. No 208 VAC at vertical motor (common)
  3. 3. VDRUP or VDRDN logic signals absent or invalid (common)
First check

Verify relays K401 and K402 seating

Test point
K401 and K402 sockets
Expected
Relays recessed in sockets snugly

6 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Carbon top horizontal position

Horizontal positioning of carbon top requires calibration after resolver replacement

Ranked causes

  1. 1. Resolver replacement (common)
  2. 2. Resolver connector disconnection (unknown)
First check

Verify resolver inserted into bracket opening

Test point
resolver bracket opening
Expected
resolver fully seated

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Center program execution radiation

Execute center program to calculate centering constants from centering rod and air scans for X-ray focal spot alignment

Ranked causes

  1. 1. Initial centering constant calculation required (common)
  2. 2. Global centering constant exceeds tolerance (occasional)
First check

Scan centering rod without metal pin

Test point
X-ray detector
Expected
Image processes without error

5 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Console Intercom/Speakerphone high voltage

System setup and verification of Intercom/Speakerphone panel operation on X-ray console

Ranked causes

  1. 1. Initial calibration required (unknown)
  2. 2. REMOTE button LED not illuminating (unknown)
First check

Power state: Gantry and Console

Test point
Power controls
Expected
Gantry power removed, Console OFF

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Gantry tilt A/D converter offset high voltage

Adjustment of tilt potentiometer output to 0.0 VDC at neutral position; part of gantry calibration procedure

Ranked causes

  1. 1. Tilt A/D converter requires offset calibration (common)
  2. 2. Limit switches require verification after calibration (common)
First check

Connect DMM to tilt A/D input

Test point
Pin 16 IC (location 16–09), TP3 on CTCC foil positive; TP1 on CTCC foil negative (ground)
Expected
DMM connected and ready for measurement
Tools
Digital multi-meter

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Gantry tilt limit switch high voltage

Tilt limit switch requires adjustment after replacement or maintenance

Ranked causes

  1. 1. Tilt limit switch replaced or requires adjustment (common)
  2. 2. Gantry covers removed during service (common)
First check

Apply gantry power

Test point
Gantry power supply
Expected
Power applied successfully

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tilt switch verification and adjustment caution

Gantry tilt position display accuracy and limit switch operation; requires protractor measurement against scan frame and potentiometer adjustment on CTCC card to match display to physical position.

Ranked causes

  1. 1. Tilt switches require verification or adjustment after replacement (unknown)
  2. 2. Tilt switches require adjustment after replacement or repositioning (unknown)
First check

Display reading at –10 degree tilt position

Test point
Gantry touch panel display
Expected
–10 degrees, matches protractor measurement
Tools
Protractor

4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cardfile installation verification high voltage

Voltage and continuity checks after reinstalling cardfile assembly before returning to service

Ranked causes

  1. 1. Connection error at cardfile connector (occasional)
  2. 2. DC power supply malfunction (occasional)
First check

Check cardfile dc voltages at backplane buss points

Test point
backplane top buss points
Expected
+5, +15, –15 V
Tools
digital multimeter

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Console keyswitch OFF disconnects relay power supply high voltage

Rotating console keyswitch to OFF position or actuating E–Stop disconnects 120AUX from the +5 volt relay power supply

Ranked causes

  1. 1. Normal operation – keyswitch or E–Stop activated as intended (common)
  2. 2. System-generated power down signal active (occasional)
First check

Verify console keyswitch position

Test point
Console keyswitch
Expected
ON position

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Console keyswitch troubleshooting

Diagnostics for Console keyswitch and related power control circuits

Ranked causes

  1. 1. Keyswitch failure or J530 connector issue (occasional)
  2. 2. Pilot Control connection failure (occasional)
First check

Run SMB diagnostic

Test point
Diagnostic system
Expected
No keyswitch faults reported

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

DC Power Supply U1 voltages missing or out of specification high voltage

Voltages at U1 (Digital supply IC) pins 11, 12, 15, 4 are absent or outside tolerance; indicates power supply failure or internal short

Ranked causes

  1. 1. Internal failure of DC power supply or U1 IC (common)
  2. 2. Short circuit on digital supply outputs (occasional)
First check

Measure +5V (Digital) at U1 pin 11

Test point
U1 pin 11 on DC Power Supply Fault/Sense board
Expected
+5 VDC, +/- 0.25V
Tools
DMM

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gantry DC supply voltages out of tolerance high voltage

DC voltages at Cardfile backplane bus points exceed specified ranges: +5V, +15V, or -15V outside tolerance bands or ripple exceeds limits

Ranked causes

  1. 1. Excessive loading on DC supply outputs (common)
  2. 2. Failure of Cardfile DC power supply (occasional)
First check

Verify 120 VAC at DC power supply input terminals

Test point
Input terminals of each DC power supply
Expected
120 VAC present

4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gantry SSR control signal absent or stuck caution

Logic signal to Solid State Relay (SSR) not switching between +5VDC and 0VDC when Gantry power is applied; prevents SSR activation and power delivery to gantry

Ranked causes

  1. 1. AC Control Chassis control logic fault or component failure (occasional)
  2. 2. Disconnected or damaged signal cable from Console (occasional)
First check

Verify cable L7156 connection and continuity

Test point
Cable L7156 at P593 (Console) and J103 (Gantry)
Expected
Cable fully seated; continuity on all pins
Tools
Visual inspection; continuity tester

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Missing 120 VAC at DC power supply input high voltage

120 VAC not present at input terminals of DC power supply

Ranked causes

  1. 1. Fault in 120 VAC path through MK 301 and CB 304 (common)
  2. 2. Defective DC power supply (occasional)
First check

Verify 120 VAC present at DC power supply input terminals

Test point
DC power supply input
Expected
120 VAC present
Tools
DMM

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No tilt response from controls patient safety

Patient positioning tilt function does not respond when FWD or REV tilt switches are pressed; no movement on gantry

Ranked causes

  1. 1. Missing or incorrect enable command from CTCC to tilt motor driver (unknown)
  2. 2. Faulty Hall sensor in tilt motor feedback chain (unknown)
First check

Measure T-VENA enable line voltage at rest

Test point
TP304
Expected
+5 VDC
Tools
DMM

7 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power brush block replacement high voltage

Removal and replacement of slip ring brush blocks for power and signal in X-ray gantry system

Ranked causes

  1. 1. Brush block maintenance or failure (unknown)
  2. 2. Reinstall brush block (unknown)
First check

Power isolated at wallbox

Test point
wallbox disconnect
Expected
gantry power OFF

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply conversion for 220V-240V primary source high voltage

Systems requiring modifications to power supply and control unit stickers when 220V-240V is the primary power source and Q-Console or Voxel-Q 120 vac access is not available.

Ranked causes

  1. 1. Power supply configured for incorrect input voltage (common)
  2. 2. Fuse panel orientation incorrect (occasional)
First check

Verify fuses F4 and F5 are 1 amp type

Test point
Fuses F4 and F5 inside power input/power cord connection
Expected
LITTLEFUSE part # 312 001 or BUSSMAN AGC-1A

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Quick mechanical checks for scan frame issues caution

Initial visual and manual inspection of scan frame mechanics before detailed electrical diagnostics

Ranked causes

  1. 1. Mechanical interference in scan frame (common)
  2. 2. Gull-wing or collision switch malfunction (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

RDC converter failure or input signal loss

Resolver-to-digital converter (RDC) not producing valid 10-bit digital output; converter or input signal fault condition detected

Ranked causes

  1. 1. Fine RDC excitation lost or winding open (occasional)
  2. 2. Table position RDC failure (occasional)
First check

Verify 2.5 kHz reference signal present at fine RDC input

Test point
Fine RDC reference input
Expected
2.5 kHz signal present and stable
Tools
Oscilloscope

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Resolver output signal loss or incorrect amplitude

Fine sine/cosine resolver outputs (fsin, fcos) or reference signal missing or out of specification at RDC input

Ranked causes

  1. 1. Primary excitation voltage out of specification (common)
  2. 2. Phase error between primary and secondary windings (occasional)
First check

Measure sine and cosine winding output amplitude

Test point
fsin, fcos outputs
Expected
2 Vrms (max)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scan frame does not rotate smoothly or has loose cables caution

Operator detects resistance, grinding, or visible loose cables during scan frame rotation; potential clearance issues between rotating and stationary frame components.

Ranked causes

  1. 1. Loose cables connecting rotating and stationary frame (common)
  2. 2. Slip ring brush damage from incorrect rotation direction (occasional)
First check

Manually rotate scan frame and visually inspect cables for slack or improper routing

Test point
All rotating frame cable connections and stationary frame attachment points
Expected
Cables taut with no visible slack; frame rotates smoothly without resistance

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scan frame does not rotate when commanded

Gantry fails to rotate when upper front cover is closed and interlock switch is in normal position

Ranked causes

  1. 1. Interlock switch failure preventing rotation (unknown)
  2. 2. Gantry mechanical obstruction or malfunction (unknown)
First check

Verify interlock switch returns to center position

Test point
Interlock switch
Expected
Switch centered after pull-check

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Scan frame drive command signal abnormal caution

Oscilloscope monitoring of drive command voltage shows insufficient or unstable signal output

Ranked causes

  1. 1. Drives Off switch malfunction (occasional)
  2. 2. Servo controller fault (occasional)
First check

Measure drive command signal with oscilloscope

Test point
A26
Expected
Swinging voltage approximately +/- 10V when manually rotating scan frame, +/- 10V
Tools
oscilloscope

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Scan frame rotation inspection caution

Visual and mechanical inspection of scan frame rotation after repair work; check for loose cables and frame clearance

Ranked causes

  1. 1. Loose cables or obstructions from service work (common)
  2. 2. Cable interference during frame rotation (occasional)
First check

Visually inspect all scan frame cables for looseness

Test point
cable connections and routing
Expected
all cables secure with no visible slack or damage

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Solid state relay or bridge rectifier replacement
  • VDRUP signal invalid or absent
  • Vertical transducer removal hazards
  • Zero interrupter signal not detected
  • Zero vane interrupter malfunction
  • 24 VAC pilot power check
  • 512 error in error log
  • 8 kVA transformer input voltage out of tolerance
  • 999 error in error log
  • Access gantry cover for servicing
  • Access gantry covers for servicing
  • Access to gantry components
  • Align patient positioning laser beams
  • Autovoice button LED not illuminating
  • Autovoice connection lost or unstable
  • Bridge rectifier replacement
  • Bus error test failed
  • Calibrate gantry tilt angle measurement
  • Calibration: 19" display monitor
  • Calibration: 19" display monitor adjustment
  • Calibration: 19" display monitor aspect ratio
  • Calibration: AC power supply voltages
  • Calibration: ACQPLAN option enabling
  • Calibration: AcQSim laser alignment to gantry isocenter
  • Calibration: AcQSim laser positioning initial setup
  • Calibration: AcQSim laser positioning verification
  • Calibration: AcQSim laser system verification
  • Calibration: Align ACQSIM side lasers
  • Calibration: Autovoice audio level adjustment
  • Calibration: Barker laser alignment
  • Calibration: Beam rotation angle
  • Calibration: brush block alignment
  • Calibration: Center calibration procedure
  • Calibration: Coarse zero position reference
  • Calibration: Cosine waveform amplitude
  • Calibration: Couch
  • Calibration: CTSIM offset configuration
  • Calibration: DC power supply voltages
  • Calibration: Enter lateral offset into AcQSim software
  • Calibration: EXACT Therapy Top installation and alignment
  • Calibration: Gantry and patient support verification
  • Calibration: Gantry tilt angle switches
  • Calibration: Gantry tilt angle verification
  • Calibration: Gantry tilt scale potentiometer (CTCC)
  • Calibration: Gantry X-ray field alignment
  • Calibration: Global Centering Constant
  • Calibration: Horizontal belt tension and position
  • Calibration: Horizontal position linear adjustment
  • Calibration: Horizontal position resolver feedback
  • Calibration: Horizontal position resolver feedback amplitude
  • Calibration: Horizontal resolver feedback signal adjustment
  • Calibration: Horizontal resolver feedback signal amplitude
  • Calibration: Horizontal sine input amplitude
  • Calibration: Horizontal sine input verification
  • Calibration: Horizontal sine input voltage
  • Calibration: Large focal spot centering
  • Calibration: Laser alignment verification
  • Calibration: Laser assembly mounting
  • Calibration: Lateral laser offset
  • Calibration: Lateral laser offset correction
  • Calibration: Lateral tube alignment
  • Calibration: Main Drive
  • Calibration: Main drive excitation level
  • Calibration: Mark X and Z axis center of scan plane
  • Calibration: Motor pole offset adjustment
  • Calibration: Patient support cover installation
  • Calibration: Patient support cover mounting
  • Calibration: Patient Support Cover removal and replacement
  • Calibration: Patient support cover removal/replacement
  • Calibration: Patient Support vertical and horizontal positioning
  • Calibration: Power supply 12 VDC reconfiguration
  • Calibration: Resolver error correction circuitry
  • Calibration: Rotating frame lateral alignment
  • Calibration: Scan drive adjustment
  • Calibration: scan drive using MADCAP
  • Calibration: Scan frame rotation and laser alignment
  • Calibration: Signal brush block alignment
  • Calibration: Telescoping base cover removal and reinstallation
  • Calibration: Telescoping base cover top section removal
  • Calibration: Tighten Patient Support bolts
  • Calibration: Tilt potentiometer and limit switches
  • Calibration: Tilt switches verification and adjustment
  • Calibration: Ultra Q floppy disk drive installation
  • Calibration: vertical transducer
  • Calibration: X-ray field alignment
  • Calibration: X-ray tube axial alignment
  • Calibration: X-ray tube axial mechanical alignment
  • Calibration: X-ray tube lateral alignment
  • Cannot control gantry from operator console during service
  • Card cage in extended position during gantry tilt
  • Card cage retraction requirement
  • Ceiling-mounted LTU may interfere with injector arms
  • Collision plate not functioning
  • Collision plate not responding
  • Collision Pleate activation check
  • Cone gasket deterioration or uneven compression
  • Configuration: AcQSim network and system settings
  • Counter test failed
  • Dangerous voltage exposure with covers removed
  • Dangerous voltages exposed during gantry cover removal
  • Dangerous voltages exposed with gantry covers removed
  • DC voltages out of tolerance on Cardfile backplane
  • Disk format timeout
  • Drive Command signal absent or incorrect
  • Drive verification using ring diagnostic
  • DUART operation test
  • DUART timer and interrupt test
  • During removal/replacement operations on telescoping base covers
  • Electrical hazard from exposed voltages
  • Electrostatic damage to PCB components
  • Equipment covers open safety risk
  • Error code 512 in error log
  • Error code 999 in error log
  • Excessive DC supply output loading or shorts
  • Exit madcap program
  • Film and reference marker placement for calibration
  • Final laser alignment verification
  • Floppy drive not recognized after motherboard replacement
  • Fluid leak into detector electronics
  • Gantry 480 VAC 3-phase input voltage out of range
  • Gantry cover and panel safety
  • Gantry cover removal and securing
  • Gantry covers exposed high voltage
  • Gantry covers removed exposing dangerous voltages
  • Gantry covers removed operation
  • Gantry does not power down
  • Gantry laser covers need securing before calibration
  • Gantry lasers misaligned
  • Gantry LOCAL START procedure
  • Gantry operation verification after component replacement
  • Gantry power and warm-up setup
  • Gantry power application
  • Gantry tilt limit switch replacement
  • Gantry tilt switch replacement/adjustment procedure
  • Gantry tilting motion verification
  • Ground potential shift at AC Control Chassis
  • Gull-wing, collision or drives switch malfunction
  • High voltage exposure hazard during cover removal
  • Horizontal drive motor replacement - vibration after installation
  • Injury from sharp upper gull wing panels
  • Injury risk when lifting LTU during installation
  • Installation: Mount lateral laser assembly on column
  • Intercom/Speakerphone panel LED not illuminating
  • Intercom/Speakerphone REMOTE LED not illuminating
  • Interlock switch does not prevent frame rotation
  • Interlock switch does not prevent gantry rotation
  • Interlock switch not preventing frame rotation
  • Interlock switch not preventing scan frame rotation
  • Laser beam adjustment interaction
  • Laser beam alignment check
  • Laser beam focus adjustment
  • Laser beam misaligned or not operational
  • Laser radiation and rotating machinery injury hazard
  • Laser radiation hazard during positioning
  • Limit switch replacement procedure
  • Loose connector P402 at J402
  • LTU mounting interference with injector arm or gull wing
  • LTU mounting interferes with injector arm or gull wing
  • Memory test failed above address 0x20000
  • Memory test failed below address 0x20000
  • Missing or low 480 VAC 3-phase at transformer input
  • Monitor cannot achieve correct aspect ratio
  • Monitor cannot achieve specified aspect ratio
  • Monitor cannot be adjusted to specified aspect ratio
  • MPC test failed
  • Negative limit switch does not interrupt tilting motion
  • Obstruction in X-ray beam path
  • Opening gantry covers for scan drive control service
  • Optical disk diagnostic fails due to unwritten media
  • Over-temperature shutdown activated
  • Patient support cover removal
  • Patient support cover removal and replacement
  • Patient support covers removal and replacement
  • Patient support covers removal/replacement
  • Patient support covers require removal and replacement
  • Patient Support not centered in scan field
  • Patient Support not perpendicular to gantry
  • Patient Support not tracking properly during vertical travel
  • Personnel injury from lifting sagittal laser
  • Power-up reset signal abnormal
  • Reconfigure 24 VDC power supply
  • Remove and replace patient support covers
  • Replace 3 KVA scan drive transformer
  • Replace 3KVA scan drive transformer
  • Replace bridge rectifier
  • Replace bridge rectifier on scan drive control
  • Replace gantry limit switch
  • Replace limit switch assembly
  • Replace limit switch on patient support
  • Replace power brush block
  • Replace solid state relay (SSR)
  • Replace solid state relay on scan drive control
  • Replace vertical limit switch
  • Resolver replacement
  • Rotating frame errors or ANB indication present
  • Rotating frame hazard during operation
  • Rotating frame inspection and rotation
  • Rotating machinery hazard during operation
  • S-Bus board pins damaged during installation
  • Safety: High voltage and heavy equipment handling
  • Sagittal laser position deviation
  • Sagittal laser readout displays incorrect positive/negative numbers
  • Scan drive overspeed condition
  • Scan drive position error
  • Scan frame does not begin CCW rotation after stop
  • Scan frame does not decelerate to stop at zero position
  • Scan frame home position rotation
  • Scan frame motor initialization required
  • Scan frame position not established at power-up
  • Scan frame rotates with gantry cover open
  • Scan frame rotates with upper front cover open
  • Scan frame rotation causes slip ring brush damage
  • Scan frame rotation hazard
  • Scan frame will not move or has mechanical interference
  • Scan position measurement fault condition
  • Scissor-action mechanism hazard with support base covers removed
  • Sharp edges on gantry covers hazard
  • Signal brush block removal and replacement
  • Solid state relay replacement
  • Static electricity damage to circuit boards
  • System does not operate with front covers open
  • System verification after cardfile replacement
  • Telescoping base cover replacement
  • Telescoping base covers removal and replacement
  • Transducer wire damage during connection/disconnection
  • Transducer wire snap hazard during connection
  • Transformer replacement procedure
  • Two service persons required for cover procedures
  • Unexpected exception
  • Unexpected exception sitting in pROBE
  • Unit does not start after power on
  • VDRDN signal invalid or absent
  • Vertical limit switch diagnostic check
  • Vertical touch panel malfunction
  • Vertical transducer operation verification
  • Vertical transducer replacement and calibration
  • Visual test execution on display board
  • Work on gantry with upper panels open
  • Work on X-ray gantry with covers removed
  • X-ray tube axial mechanical alignment
  • Zero interrupter not triggering

241 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0Did not boot; never made to beginning of codeConstant ACFAIL from RCB/DIB; bad software or hardware failurecritical
1Failed RAM Address write testFailure during RAM address write testcritical
11DSP memory test failureAcquisition time-out error or memory corruption during DSP diagnosticserror
13Zero Reference ADCA Test failureDSP power-up diagnostics detect zero reference error in analog-to-digital converter Aerror
14Positive 15 V rail test failureDSP diagnostics detect positive 15 V supply out of specificationerror
14Unexpected PITR interrupt during testingUnexpected PITR interrupt detected during diagnostic testingerror
15Interrupt mask errorSR mask does not match expected mask value during interrupt testingerror
15Negative 15 V rail test failureDSP diagnostics detect negative 15 V supply out of specificationerror
16ADCB Reference Test failureDSP power-up diagnostics detect reference error in analog-to-digital converter Berror
16SCI TE Interrupt TimeoutSerial communication interface transmit enable interrupt failed to occurerror
175 V rail test failureDSP diagnostics detect 5 V supply out of specificationerror
17SCI TC Interrupt TimeoutSerial communication interface transmit complete interrupt failed to occurerror
18SCI RDFRF Interrupt TimeoutSerial communication interface receive data register full interrupt failed to occurerror
18ADCA Reference Test failureDSP power-up diagnostics detect reference error in analog-to-digital converter Aerror
19Starter 10 V Reference Test failureDSP diagnostics detect 10 V starter reference out of specificationerror
19SCI Mask ErrorSCI interrupt mask does not match expected configurationerror
1900Cannot setup ANB DriverANB Driver initialization fails during startupcritical
1903Cannot initialize Frame GalilFrame Galil motor controller fails to initializecritical
1904Cannot initialize Couch GalilCouch Galil motor controller fails to initializecritical
1908Cannot start GPU Manager ProcessGPU Manager process fails to start during bootcritical
2Failed RAM 5555hex testBad bit detected during RAM 5555hex pattern testcritical
20Temperature Test failureDSP diagnostics detect out-of-range temperature sensor outputerror
20Soft Watchdog ErrorSoftware watchdog timer detected error conditionerror
21Filament DAC Zero Output Test failureFilament digital-to-analog converter does not output zero when commandederror
21Watchdog Intermit ErrorWatchdog interrupt error during diagnostic testerror
22Soft Watchdog Error ResetSoftware watchdog triggered error and system reseterror
22Filament DAC Full Scale Output Test failureFilament DAC does not output full-scale voltage when commandederror
222Spurious Interrupt ErrorFatal error in gantry control system before GPMN process starts or within error process itselfcritical
23ROM Check Sum ErrorROM memory checksum verification failed during power-up diagnosticscritical
23Filament DAC Linearity Test failureFilament DAC output does not scale linearly with input commandserror
24RSR ErrorRSR (register) error detected during diagnostic testerror
24MADAC Zero Output Test failureMA (tube current) DAC does not output zero when commandederror
25Soft Watchdog Timeout ErrorSoftware watchdog timer exceeded timeout thresholderror
25MA DAC Full Scale Output Test failureMA DAC does not output full-scale voltage when commandederror
26MADAC Linearity Test failureMA DAC output does not scale linearly with input commandserror
27KV DAC Zero Output Test failureHigh voltage (kV) DAC does not output zero when commandederror
28KV DAC Full Scale Output Test failureKV DAC does not output full-scale voltage when commandederror
29KV DAC Linearity Test failureKV DAC output does not scale linearly with input commandserror
3Failed RAM AAAAhex testBad RAM detected during AAAAhex pattern testcritical
30Unexpected ExceptionUnexpected exception condition detected during diagnosticserror
333Watchdog Timer Timeout ErrorFatal error when watchdog timer expires in gantry control system before GPMN process starts or within error process itselfcritical
4Failed NOV RAM testFailure during NOV RAM diagnostic testcritical
444System exception errorException error detected during XSC power diagnosticserror
5Failed ROM testFailure during ROM integrity checkcritical
512GPU Crash in error loggerGPU microprocessor failure or overspeed condition during scan operationcritical
512GPU crash - graphics processing unit or motion control processor failureGPU microprocessor malfunction, bad GALIL motion chip, bad DUART, or communication loss to motion chipcritical
555Bus address errorBus address error detected during XSC power diagnosticserror
6Failed CPUFAIL interrupt level 7 testConstant ACFAIL from RCB/DIB; CPU interrupt level 7 failurecritical
7Failed interrupt level 6 testFault circuitry failure detected during interrupt level 6 testerror
8Failed timer Z48 testFailure during timer Z48 diagnostic testerror
9Failed sync testSync test failure during diagnosticswarning
999Command Shell error - system command interpreter or communication failureCommand shell communication loss or command processing failureerror
999Command Shell error in error loggerLoss of communication or command processing failure in scan drive control systemerror
AFailed interrupt level 3 testInterrupt level 3 failure during diagnostic testerror
ACQSim functions not activeACQSim software features are not enabled in the Voxel-Q systemACQSim functions not activated after software installationerror
BFailed interrupt level 2 testInterrupt level 2 failure during diagnostic testerror
CFailed interrupt level 1 testInterrupt level 1 failure during diagnostic testerror
can't identify plot on displaySystem unable to recognize or validate plot data on display outputPlot validation check after renderingerror
can't start acceldma on heurikonUnable to initiate acceldma process on Heurikon hardwareScan start command issuederror
can't start plot on displayUnable to initiate plot rendering on display moduleDisplay update commandwarning
DFailed DUART serial testSerial port DUART chip test failureerror
datposDrives at positionScan frame and table have reached commanded positionsinfo
dits counter not reset to 0 errorDITS counter failed to reset to zero at initializationSystem startup or scan preparation sequencewarning
divide by 0 error, plot file not updatedMathematical division by zero encountered; plot file not written as a resultPlot calculation or normalizationerror
drvfltDrives fault - directly shuts down hardwareVarious drive-related faults detectedcritical
EFailed bus address error testBus address error detection failureerror
Error logged when power switch turned OFF with tube heat aboGantry shut down immediately due to unsafe shutdown procedureConsole power switch turned OFF when tube heat above 35%warning
error reading scan dataScan data file could not be read or accessedData retrieval after scan completionerror
FWouldn't boot to applications sideFailure to boot to applications side of systemcritical
fatal error messageUnrecoverable system error encounteredCritical software or hardware fault during operationcritical
hdperrHorizontal drive position errorTable actual position differs from commanded referenceerror
hdractHorizontal drive active - status signal indicating proper operationHorizontal drive is functioning normallyinfo
hpmfltHorizontal position measurement faultTable position RDC input signal lost or RDC failurecritical
illegal signalReceived signal does not conform to expected format or protocolSignal reception and validationwarning
lread error of bad.datFailed to read bad.dat fileLoading bad data recordserror
lwrite error bad.datFailed to write bad data records to bad.dat fileRecording out-of-range or invalid data pointserror
lwrite error to raw data fileFailed to write scan data to raw data fileSaving raw scan data to diskerror
lwrite of to plot.dat errorFailed to write processed data to plot.dat fileSaving processed plot data to diskerror
plot timeoutPlot generation or rendering did not complete within expected timePlot rendering operationwarning
scan positioning timeoutScan positioning operation did not complete within expected timeframeAttempted to position scan but system failed to respond within timeout perioderror
sdovspScan drive overspeedScan frame rotation speed exceeds safe limitwarning
sdperrScan drive position errorScan frame actual position differs from commanded referenceerror
sdractScan drive active - status signal indicating proper operationScan drive is functioning normallyinfo
spmfltScan position measurement faultFine RDC input signal lost or RDC failure; excitation or winding losscritical
Superblock corruptionTest termination due to superblock corruption or attempting to run test on image diskRunning option 0 or option 5 on newly formatted, virgin disk, or disk with previous destructive test runwarning
Test 1SCI DATA communication errorSCI DATA timeout or communication integrity error during booterror
Test 10Level 4 PITR interruptLevel 4 PITR interrupt not properly clearederror
Test 11PITR count out of rangePITR timer count value outside acceptable rangeerror
Test 12Multiple interrupts in timer modeMore than one interrupt detected during timer mode testwarning
Test 13Unexpected DUART interruptDUART status interrupt detected during testwarning
Test 2SCI Data mismatch errorTransmitted and received SCI data do not matcherror
Test 3NOVRAM data errorNOVRAM data read/write failure or checksum mismatcherror
Test 4MC68332 RAM errorMC68332 processor RAM test failureerror
Test 5Static memory data errorStatic memory write/read test failureerror
Test 6Static RAM march memory errorStatic RAM march pattern test failureerror
Test 7Bus errorCPU system bus communication errorerror
Test 8DUART no level 2 interruptDUART level 2 interrupt signal not detectederror
Test 9DUART counter interrupt status bit errorDUART counter interrupt status incorrect during testerror
Test termination due to superblock corruptionHard drive diagnostic terminated because superblock is corrupted or test attempted on image diskRunning diagnostic option 0 or 5 on newly formatted, virgin, or recently destructively tested diskwarning
timeout on signal from acceldma doneExpected completion signal from acceldma module not receivedScan executionerror
timeout on signal from acceldma to start scanExpected start signal from acceldma module not received within timeoutScan initiation sequenceerror
too many beginning points out of rangeInitial data points at scan start exceed acceptable rangeData validation at scan beginningwarning
too many points out of rangeExcessive number of data points exceed acceptable value rangeData validation after scan or plot generationwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.