Philips Picker AcQsim: troubleshooting and error codes
The ServiceForge knowledge base holds 31 documented symptoms for the Philips Picker AcQsim, with 33 ranked causes, 63 diagnostic checks with test points and expected values, and 0 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Alignment of laser system - Non X-ray test
Daily QA procedure to verify lateral wall, sagittal wall, and ceiling lasers are correctly aligned to gantry and table for accurate isocenter marking
Ranked causes
- 1. Lateral wall laser misaligned with gantry and table (unknown)
- 2. Sagittal laser misaligned with gantry and table (unknown)
Zero sagittal laser at control panel
- Test point
- Sagittal laser
- Expected
- Zeroed
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: table reproducibility under gantry and console controls
Verify reproducibility of table position under both gantry and console control in longitudinal and vertical directions
Ranked causes
- 1. Table positioning mechanism drift or calibration loss (occasional)
- 2. Console control positioning error (occasional)
Measure table position reproducibility under gantry control in longitudinal direction
- Test point
- table
- Expected
- within 1.0 mm of specified position, ±1.0 mm
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alignment of localization lasers (X-ray and non-X-ray) caution
Congruence of localization lasers should be verified to ensure CT scanner isocenter and table center are accurately indicated; laser accuracy within specified tolerance
Ranked causes
- 1. Laser misalignment or drift from specified position (occasional)
Measure laser position relative to CT scanner isocenter
- Test point
- isocenter
- Expected
- within +1.0 mm of specified values, +1.0 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Horizontal table movements
Monthly verification of horizontal (lateral and longitudinal) table motion for patient positioning
Ranked causes
- 1. Horizontal table movements require monthly verification (common)
Test horizontal table movements
- Test point
- Horizontal table motion mechanisms
- Expected
- Smooth motion in both horizontal directions
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Laser alignment
Daily QA procedure for sagittal wall laser and lateral lasers alignment verification
Ranked causes
- 1. Lasers out of alignment (unknown)
Alignment of sagittal wall laser
- Test point
- Sagittal wall laser
- Expected
- Within alignment tolerance
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Laser alignment accuracy radiation
Verify congruence of CT scanner localization lasers (X-ray and non-X-ray) to ensure isocenter and table center are accurately indicated
Ranked causes
- 1. Laser misalignment from specified values (unknown)
Verify laser position accuracy
- Test point
- Isocenter and table center
- Expected
- within +1.0 mm of specified values, +/-1.0 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Laser alignment verification
Yearly comprehensive alignment of all laser systems for continued accuracy
Ranked causes
- 1. Laser systems require yearly alignment verification (common)
Verify all laser system alignments
- Test point
- All laser positioning references
- Expected
- All lasers properly aligned
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Laser and radiation field coincidence radiation
Yearly verification that laser positioning lights align with actual radiation field for scan accuracy
Ranked causes
- 1. Laser and radiation field require yearly coincidence verification (common)
Verify laser and radiation field alignment
- Test point
- Laser position and radiation field boundaries
- Expected
- Laser lights coincident with radiation field
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Laser localization alignment patient safety
Localization lasers may not accurately indicate CT scanner isocenter and table center; accuracy drift requires verification against specified values
Ranked causes
- 1. Laser alignment drift from specified position (occasional)
Measure laser position deviation from isocenter and table center
- Test point
- CT scanner isocenter and table center
- Expected
- Within +1.0 mm of specified values, +1.0 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Laser offset
Update external lateral laser offset distance between scanner isocenter and laser plane after alignment adjustments
Ranked causes
- 1. Offset value changed after alignment adjustments (common)
Measure offset value according to procedure
- Test point
- External lateral lasers vs scanner isocenter
- Expected
- Distance measurement in mm
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Laser system alignment
Daily QA procedure to verify lateral wall, sagittal wall, and ceiling lasers are correctly aligned with gantry and table for accurate isocenter marking.
Ranked causes
- 1. Lateral or sagittal laser misaligned with gantry and table (unknown)
Zero sagittal laser at control panel
- Test point
- control panel sagittal laser
- Expected
- zeroed
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Laser system alignment with X-ray radiation
Weekly X-ray test alignment of laser system to verify coincidence with radiation field
Ranked causes
- 1. Laser system requires weekly X-ray alignment verification (common)
Verify laser system alignment with X-ray exposure
- Test point
- Laser and radiation field alignment
- Expected
- Laser and X-ray field properly coincident
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Localization light accuracy
Monthly verification of positioning light accuracy for proper scan region localization
Ranked causes
- 1. Localization light accuracy requires monthly verification (common)
Verify localization light accuracy
- Test point
- Positioning light alignment
- Expected
- Light accurately marks scan region
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Pilot localization accuracy
Yearly verification of pilot light accuracy for initial patient positioning
Ranked causes
- 1. Pilot localization accuracy requires yearly verification (common)
Verify pilot localization light accuracy
- Test point
- Pilot light positioning
- Expected
- Pilot light marks correct anatomical position
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sagittal laser assembly
Periodic calibration of laser tracker unit (LTU) to correct distance measurements from CT data. Performed when measured values differ >1 mm from previous readings.
Ranked causes
- 1. Calibration values drift >1 mm from baseline (occasional)
Measure sagittal laser calibration data
- Test point
- LTU control unit
- Expected
- Compare to previously recorded values
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sagittal tracking laser
Verify alignment and calibration of sagittal laser tracking unit; establish relationship between distance moved at unit and distance at isocenter
Ranked causes
- 1. Sagittal laser drift or misalignment (occasional)
Verify mechanical accuracy at sagittal laser unit
- Test point
- Sagittal laser unit
- Expected
- in excess of 0.1 mm accuracy with stringent rotational limits, > 0.1 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: sagittal tracking laser alignment
Sagittal laser tracking unit requires periodic alignment and calibration to verify relationship between distance moved at unit and distance measured at isocenter; laser line projected over 3.5 meters or greater
Ranked causes
- 1. Calibration drift or mechanical disturbance of sagittal tracking laser (occasional)
Verify laser line projection distance
- Test point
- sagittal laser unit
- Expected
- 3.5 meters or greater
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sagittal tracking unit
Monthly calibration of sagittal tracking unit for accurate patient positioning along scanner axis
Ranked causes
- 1. Sagittal tracking unit requires monthly calibration (common)
Perform sagittal tracking unit calibration
- Test point
- Sagittal tracking mechanism
- Expected
- Tracking unit within specification
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: table increment and laser/radiation field coincidence patient safety
Verify accurate laser/radiation field coincidence and table increment to enable accurate correlation of slice thickness with patient's anatomical structures
Ranked causes
- 1. Laser and radiation field misalignment or drift (occasional)
Measure difference between laser and radiation field center
- Test point
- isocenter
- Expected
- not exceed +1.0 mm, ±1.0 mm
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Table position reproducibility
Verify table position reproducibility under both gantry and console controls in longitudinal and vertical directions
Ranked causes
- 1. Table positioning drift or mechanical wear (occasional)
Measure table position reproducibility under gantry control - longitudinal direction
- Test point
- Table position, in/out of scanner
- Expected
- within 1.0 mm of specified position, 1.0 mm
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Table reproducibility
Verify table position reproducibility under both gantry and console controls in longitudinal and vertical directions
Ranked causes
- 1. Table position drift or mechanical play in gantry or console controls (unknown)
Measure table position reproducibility - longitudinal direction
- Test point
- Table position under gantry control
- Expected
- within 1.0 mm of specified position, +/-1.0 mm
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Tracking unit positioning accuracy patient safety
Sagittal laser positioning must be accurate to 0.5mm for correct patient isocenter determination.
Ranked causes
- 1. Tracking unit misalignment (unknown)
Prepare calibration surface
- Test point
- Vertical film holder front face
- Expected
- Level surface with graph paper taped
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Horizontal table position out of tolerance
Table horizontal movement does not reach expected 300 mm displacement
Ranked causes
- 1. Gantry or console control horizontal drive mechanism out of adjustment (unknown)
Verify horizontal motion displacement gantry control positive
- Test point
- Horizontal axis into gantry
- Expected
- 300 mm
Laser offset requires updating in configuration
External lateral lasers no longer accurately mark isocenter plane after alignment adjustments; offset value has changed by millimeters
Ranked causes
- 1. Isocenter plane shifted after alignment adjustments (occasional)
Measure offset value between actual isocenter plane and external lateral laser plane
- Test point
- Isocenter plane vs. external lateral lasers
- Expected
- Recorded offset value from measurement procedure
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser/radiation field coincidence error radiation
Misalignment between radiation field and light field in CT simulator during quality assurance testing
Ranked causes
- 1. Laser and radiation field not aligned (unknown)
Measure error between radiation field and light field
- Test point
- Laser/radiation field
- Expected
- 2 mm slice: +/-2.0 mm; 5 mm slice: +2.0 mm; 10 mm slice: +2.0 mm, See expected by slice thickness
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Laser/radiation field misalignment radiation
Light field does not coincide with radiation field during CT simulation setup
Ranked causes
- 1. Laser alignment drift or mechanical misalignment (unknown)
Measure error between radiation field and light field at 2mm slice thickness
- Test point
- Laser and radiation field
- Expected
- +/-2.0 mm, +/-2.0 mm
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lasers misaligned beyond tolerance
Paper clip or phantom image offset from scanner isocenter exceeds 2 mm tolerance
Ranked causes
- 1. Lateral or sagittal laser drift or mechanical shift (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lateral lasers misalignment
Lateral lasers not properly aligned for CT simulator positioning
Ranked causes
- 1. Laser alignment drift or mechanical shift (unknown)
Verify lateral laser alignment
- Test point
- Lateral lasers
- Expected
- Properly aligned
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pilot localization accuracy error exceeds tolerance patient safety
Light field defined by laser line does not align with metallic marker within specified tolerance in X, Y, or Z dimensions
Ranked causes
- 1. Pilot localization misalignment (unknown)
Measure localization error in X dimension
- Test point
- Light field and metallic marker alignment
- Expected
- ≤ 0.5 mm maximum error; ≤ 1 mm all dimensions, ≤ 1 mm
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Sagittal wall laser misalignment
Sagittal wall laser not properly aligned for CT simulator positioning
Ranked causes
- 1. Laser alignment drift or mechanical shift (unknown)
Verify sagittal wall laser alignment
- Test point
- Sagittal wall laser
- Expected
- Properly aligned
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Vertical table position out of tolerance
Table vertical movement does not reach expected 150 mm displacement
Ranked causes
- 1. Gantry control vertical drive mechanism out of adjustment (unknown)
Verify vertical motion displacement gantry control positive
- Test point
- Vertical axis upwards
- Expected
- 150 mm
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.