Philips PCR Compano Reader: troubleshooting and error codes
The ServiceForge knowledge base holds 19 documented symptoms for the Philips PCR Compano Reader, with 21 ranked causes, 46 diagnostic checks with test points and expected values, and 2 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Calibration: Sensitivity adjustment radiation
Procedure to adjust reader sensitivity using a measured dose exposure and verify pixel values and density are within tolerance after shading/polygon correction is applied.
Ranked causes
- 1. Sensitivity requires calibration adjustment (unknown)
- 2. Extreme differences detected in calibration values (rare)
Record initial HVCENTR value
- Test point
- HVCENTR in DISPLAY CURRENT DATA
- Expected
- Initial value e.g. 474
9 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software upgrade procedure for PCR Compano Reader caution
Step-by-step process to upgrade reader unit software from CD-ROM, including verification of upgrade completion via command prompt utility.
Ranked causes
- 1. Software upgrade required (common)
- 2. MAIN CPU IPL not upgraded to A02 (occasional)
Verify CR-IR346RU Application Software version
- Test point
- Command prompt mutl VERSION DISPLAY output
- Expected
- 114Y5436002A09
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sensitivity and shading correction radiation
Verify and adjust the sensitivity (S value) and pixel/density values of the PCR Compano Reader using a uniformly exposed imaging plate. Required when S value or pixel values fall outside specification after exposure.
Ranked causes
- 1. Sensitivity or shading correction requires adjustment (unknown)
Verify scanner unit cleaning and data backup
- Test point
- Scanning unit and service program PC-MUTL
- Expected
- Unit cleaned; data copied to floppy diskette
5 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Sensitivity check with double dose radiation
Alternative sensitivity check procedure using double dosage (2 mR) when standard 1mR dosage cannot be used.
Ranked causes
- 1. Standard sensitivity check needs alternative dosage (unknown)
Verify S value with double dose exposure
- Test point
- EasyVision display after double dose exposure
- Expected
- S value: 85 … 115 S units, 85 to 115
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check sensitivity setting radiation
Verification procedure to determine if sensitivity and shading correction adjustment is required. Performed after cleaning scanning unit, using uniform X-ray exposure of imaging plate.
Ranked causes
- 1. Need to verify current sensitivity before performing adjustment (common)
Expose imaging plate with uniform X-ray dosage
- Test point
- Dosimeter reading at IP location
- Expected
- 1 mR (8.7 µGy) at 80 kV; alternative: 2 mR (17.4 µGy) double dose, as specified per dose mode selected
- Tools
- Dosimeter, X-ray tube (min 1.8 m SID), collimator
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU board application software reload caution
Restore application software to CPU board when MUTL program fails but command prompt appears
Ranked causes
- 1. CPU board unable to execute MUTL program (unknown)
Verify command prompt appears
- Test point
- Serial or network connection
- Expected
- '->' prompt displayed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU board cannot execute application software
CPU board fails to run MUTL program; however prompt '->' still appears on display, indicating partial communication possible
Ranked causes
- 1. Application software (RU) damaged or corrupted (unknown)
Verify prompt '->' appears on display
- Test point
- Reader display
- Expected
- Prompt '->' visible
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU board cannot execute MUTL program
CPU board fails to execute the MUTL program; prompt '->' still appears on display
Ranked causes
- 1. CPU board software corruption or malfunction (unknown)
Verify CPU board prompt '->' appears on display
- Test point
- Reader display/console
- Expected
- Prompt '->' visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU board not executing MUTL program caution
CPU board unable to run MUTL program but prompt still appears; application software reload required
Ranked causes
- 1. CPU board application software corrupted or missing (rare)
Verify prompt '->' appears at connection
- Test point
- Serial or network terminal
- Expected
- Prompt '->' appears
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
CPU board software recovery via PC Maintenance Utility
Alternative method to reload CPU board application software using PC Maintenance Utility when network connection functional
Ranked causes
- 1. CPU board application software needs recovery via network (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disable automatic software updates
Configure AUTO UPDATE setting to OFF to prevent unexpected software uploads from Compano PC
Ranked causes
- 1. AUTO UPDATE enabled during normal operation (common)
Verify AUTO UPDATE configuration setting
- Test point
- Reader configuration
- Expected
- AUTO UPDATE = OFF
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Host does not work for FTP_SERV
FTP Server communication fails between Compano PC and reader
Ranked causes
- 1. IP name and IP address mismatch between Compano PC and reader (common)
Verify IP name and address match between Compano PC and reader
- Test point
- FTP_SERV configuration
- Expected
- IP settings identical in reader and Windows
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Manual Main IPL software upgrade required patient safety
Automatic upgrade has failed; reader requires manual software update via CD-ROM and batch file execution.
Ranked causes
- 1. Previous automatic upgrade failed (unknown)
Confirm application software version before upgrade
- Test point
- CR-IR346RU Application Software display
- Expected
- A05 or later
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reader Backup memory needs clearing
Reader requires FLASHROM clear and IP address reset to factory default 172.16.1.10
Ranked causes
- 1. Maintenance procedure: factory reset required (unknown)
Verify primary erasure indicator LED activation
- Test point
- Reader panel primary erasure indicator LED
- Expected
- LED comes on
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reader Backup memory needs clearing via PC utility
Alternative method to clear FLASHROM using PC Maintenance Utility instead of hardware switch
Ranked causes
- 1. Maintenance procedure: factory reset via software (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reader fails to communicate with Compano PC
Network communication between Reader and PC cannot be established during initial setup
Ranked causes
- 1. Reader IP address not compatible with hospital network (occasional)
Verify network communication with Reader
- Test point
- Default Reader IP address
- Expected
- 172.16.1.10
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reader LEDs illuminate and beeps sound during boot
All LEDs on RU panel light up with short beeps when AUTO UPDATE is ON and updated software or configuration is detected after boot process
Ranked causes
- 1. AUTO UPDATE enabled with new software or configuration detected (common)
Observe RU panel LEDs and beeps during boot
- Test point
- RU panel
- Expected
- All LEDs illuminated with short beeps
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Resetting reader to factory defaults
Procedure to clear backup memory (FLASHROM) and restore default IP address 172.16.1.10
Ranked causes
- 1. User requires factory reset (unknown)
Hold Erasure Process switch and power on reader
- Test point
- Reader power and Erasure Process switch
- Expected
- Primary erasure indicator LED illuminates
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Telnet/Batch program cannot access reader unit
Telnet or batch program fails to communicate with RU (reader unit)
Ranked causes
- 1. IP addresses and names in Flash ROM do not match Compano PC configuration (common)
Verify IP names and addresses match between reader Flash ROM and Compano PC
- Test point
- Reader configuration and Compano PC settings
- Expected
- IP names and addresses identical
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
12094 | Warning written to ERRORB.LOG file | PC sends termination command to Reader unit during shutdown or application software termination | warning |
12094 | Warning logged during reader termination command | PC sends termination command to Reader unit during shutdown or application software termination | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.