Philips PageWriter Trim I: troubleshooting and error codes
The ServiceForge knowledge base holds 135 documented symptoms for the Philips PageWriter Trim I, with 216 ranked causes, 202 diagnostic checks with test points and expected values, and 20 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Screen is dark and shows no image
Display remains blank or very faint with no visible content
Ranked causes
- 1. Cardiograph is in standby mode (common)
- 2. Contrast adjusted to lowest degree (occasional)
- 3. AC power cord disconnected or damaged (occasional)
- 4. Battery depleted (occasional)
- 5. Power supply failure (rare)
Verify cardiograph power state
- Test point
- front display
- Expected
- cardiograph wakes up
4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
All leads show leads off continuously patient safety
Red dashed line displayed for all ECG leads indicating no signal detected
Ranked causes
- 1. PIM communications lost or defective leads (common)
- 2. Precordial lead defect (occasional)
- 3. RL/R lead defect (occasional)
- 4. Internal cardiograph hardware failure (rare)
Verify PIM cable seating
- Test point
- PIM rear connector
- Expected
- cable firmly seated
4 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error message during software installation: incorrect model, token, or invalid key
Software install fails with error about incorrect product model, invalid token, or key mismatch
Ranked causes
- 1. PC card not fully inserted before power-up (common)
- 2. Software files on card are for wrong product model (occasional)
- 3. Invalid token key (occasional)
- 4. Invalid cardiograph serial number (occasional)
Verify PC card fully inserted before boot
- Test point
- PC card slot
- Expected
- Card seat fully with no gaps; no bent or broken pins visible on mainboard
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer error - out of paper, jam, or door open
Combined printer error condition indicating multiple possible causes on Trim II/III/Rx
Ranked causes
- 1. Paper drawer not fully closed (common)
- 2. Out of paper or incorrectly loaded (common)
- 3. Defective TOF sensor assembly (occasional)
- 4. Defective paper tray assembly (occasional)
Close drawer firmly
- Test point
- Printer drawer
- Expected
- Drawer fully closed
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer error code 0204 caution
Printer error 0204 displayed; paper jam, incorrect paper loading, defective TOF sensor, or defective paper tray assembly
Ranked causes
- 1. Paper jam (common)
- 2. Paper is not loaded correctly (common)
- 3. Defective TOF sensor assembly (occasional)
- 4. Defective paper tray assembly (occasional)
Inspect paper path for jam
- Test point
- Printer interior
- Expected
- Paper jam present or paper path clear
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer Error: Out of Paper/Paper Jam/Door Open
Error message indicating paper drawer, paper feed, or sensor issue
Ranked causes
- 1. Printer drawer not fully closed (common)
- 2. Out of paper or paper loaded incorrectly (common)
- 3. Defective TOF sensor assembly (occasional)
- 4. Defective paper tray assembly (occasional)
Inspect and close printer drawer
- Test point
- Drawer position and latch
- Expected
- Drawer fully seated and latched
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless adapter not associated to access point
350 Series Radio is Not Associated error; adapter cannot connect to wireless network
Ranked causes
- 1. Weak signal or out of range (common)
- 2. Access point not configured or enabled (occasional)
- 3. Wireless adapter settings mismatch (occasional)
- 4. LEAP authentication credentials not enabled (occasional)
Check LED indicators on wireless adapter
- Test point
- Wireless adapter LEDs
- Expected
- Red LED blinking slowly or off; Green LED blinking slowly or off
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
After attempting Archive screen operation, error message appears: Requested Function Failed
Error dialog displays Requested Function Failed after Archive screen operation; transfer or query incomplete
Ranked causes
- 1. Network or modem phone line was interrupted during operation (common)
- 2. Selected external archive media was not available (occasional)
- 3. An attempt to transfer an unsupported ECG report type was made to TraceMaster (occasional)
Verify network/phone line integrity
- Test point
- Network or modem connection
- Expected
- Cabling intact and connection stable
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader does not scan or returns inaccurate data
Barcode reader attached to rear port fails to scan test barcode or scans barcode data that does not match expected data
Ranked causes
- 1. Barcode reader scanning angle incorrect (common)
- 2. Barcode reader calibration drift (occasional)
- 3. Barcode reader connection loose or faulty (occasional)
Verify barcode reader is attached to correct port
- Test point
- Barcode reader port on rear of cardiograph
- Expected
- Barcode reader securely seated in port before power-on
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 0201: No battery or dead battery caution
Printer reports no battery or battery insufficient power via error code 0201
Ranked causes
- 1. Battery missing or depleted (common)
- 2. Motherboard failure (rare)
- 3. Power supply cord defective (occasional)
Verify battery presence and condition
- Test point
- Battery compartment
- Expected
- Battery present and charged
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 0203: Paper drawer issue or out of paper caution
Printer reports paper drawer not closed, out of paper, or paper loaded incorrectly
Ranked causes
- 1. Paper drawer not fully closed (common)
- 2. Out of paper or incorrectly loaded (common)
- 3. Application error (occasional)
Close printer drawer
- Test point
- Printer drawer
- Expected
- Drawer fully closed and latched
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error 0204: Paper jam or sensor failure caution
Printer reports paper jam, incorrect loading, defective sensor or tray assembly
Ranked causes
- 1. Paper jam or incorrect loading (common)
- 2. Defective TOF sensor assembly (occasional)
- 3. Defective paper tray assembly (occasional)
Clear paper jam and verify loading
- Test point
- Printer paper path
- Expected
- Paper path clear, paper properly loaded
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Magnetic card reader does not read or returns inaccurate data
Magnetic card reader fails to read data from magnetic card or reads data that does not match expected card contents
Ranked causes
- 1. Magnetic card not inserted correctly or left in slot too briefly (common)
- 2. Magnetic card reader connection loose (occasional)
- 3. Magnetic card reader malfunction (rare)
Verify magnetic card reader is connected
- Test point
- Magnetic card reader connector on rear of cardiograph
- Expected
- Magnetic card reader securely attached to port before power-on
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network test shows no connection or incorrect configuration caution
Network Connection window does not display configured network information or displays incorrect network data during network test
Ranked causes
- 1. LAN card not fully inserted in PC Card slot (common)
- 2. LAN cable not fully inserted into LAN card connector (common)
- 3. Incompatible or damaged LAN cable (occasional)
Verify LAN card is fully inserted in PC Card slot
- Test point
- LAN card in PC Card slot on rear of cardiograph
- Expected
- LAN card fully seated in slot with no visible gaps or loose contacts
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No reply from Remote Site Server
No Reply Received From Remote Site Server error; ECG transmission connection failure
Ranked causes
- 1. Wireless connection to access point lost (common)
- 2. Network connection to TraceMasterVue server not configured (common)
- 3. Cardiograph did not receive response within timeout period (occasional)
Verify wireless adapter LED status
- Test point
- Wireless adapter LEDs
- Expected
- Green LED associated with access point
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more leads show leads off periodically patient safety
Red dashed line shown on leads display, periodic signal loss reported by cardiograph
Ranked causes
- 1. PIM patient data cable connection failure (common)
- 2. Defective lead wires (common)
- 3. PIM PCA set failure (occasional)
Verify PIM cable connection
- Test point
- PIM connector on cardiograph rear
- Expected
- Cable firmly seated
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer error code 0201
Printer error 0201 displayed; no battery detected or battery is dead; printer cannot operate without power
Ranked causes
- 1. No battery or dead battery (common)
- 2. Motherboard failure (occasional)
- 3. Power supply cord failure (occasional)
Verify battery presence and condition
- Test point
- Battery compartment
- Expected
- Battery present and functional
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer error code 0203
Printer error 0203 displayed; paper drawer not fully closed, out of paper, incorrectly loaded paper, or application error
Ranked causes
- 1. Paper drawer is not fully closed (common)
- 2. Out of paper or paper loaded incorrectly (common)
- 3. Application error (occasional)
Verify printer drawer is fully closed
- Test point
- Printer drawer latch
- Expected
- Drawer fully closed and latched
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer paper jam or out of paper
Error code 0204 indicates paper jam, paper depletion, or paper feed failure
Ranked causes
- 1. Paper jam in printer mechanism (common)
- 2. Defective TOF sensor assembly (occasional)
- 3. Defective paper tray assembly (occasional)
Clear paper jam and verify paper loading
- Test point
- Printer paper mechanism
- Expected
- Paper properly loaded, no jammed paper visible
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unidentified PC card adapter detected
Unidentified PC card Adapter error with dialog to enter driver name
Ranked causes
- 1. Wireless adapter inserted incorrectly into PC card slot (common)
- 2. Adapter inserted during active use (not in Standby) (occasional)
- 3. Unsupported wireless adapter installed (rare)
Verify wireless adapter is properly seated
- Test point
- PC card slot on rear of cardiograph
- Expected
- Adapter fully inserted and secure
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB printer - paper jam or out of paper
Printer reports paper jam condition or no paper detected
Ranked causes
- 1. Paper drawer not closed or paper misaligned (common)
- 2. Defective TOF sensor assembly (occasional)
- 3. Defective paper tray assembly (occasional)
Close printer drawer firmly
- Test point
- Drawer latch
- Expected
- Drawer fully seated
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless adapter not properly inserted or detected
PC card adapter insertion issue; RED LED off and GREEN LED off
Ranked causes
- 1. Wireless adapter not securely inserted in PC card slot (common)
- 2. Cardiograph power cycle required (common)
- 3. Wireless adapter hardware failure (rare)
Verify wireless adapter seating in PC card slot
- Test point
- PC card slot on rear of cardiograph
- Expected
- Adapter fully and securely inserted
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader data not scanning accurately
Barcode reader fails to accurately scan barcode data or data appears incorrect after scanning
Ranked causes
- 1. Barcode reader not properly attached to port (common)
- 2. Barcode reader requires calibration (occasional)
Verify barcode reader attachment
- Test point
- Barcode reader port on rear of cardiograph
- Expected
- Barcode reader securely connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader not functioning
Barcode reader fails to read barcode or does not display correct barcode data
Ranked causes
- 1. Barcode reader not attached to rear port (common)
- 2. Barcode reader requires calibration (occasional)
Scan barcode at 45 degree angle and verify data display
- Test point
- Barcode reader sensor
- Expected
- Barcode data displayed correctly in Bar Code Field
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode Reader not reading barcode
Barcode reader fails to capture or display barcode data correctly
Ranked causes
- 1. Barcode Reader hardware failure or improper scanning technique (unknown)
- 2. Barcode Reader not installed or not attached to rear port (unknown)
Verify barcode reader attached to rear port
- Test point
- Rear port connection
- Expected
- Reader physically attached before power-on
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fax/modem card not recognized or not functioning
Fax/modem card fails recognition test or ROM/firmware verification
Ranked causes
- 1. Fax/modem card not inserted or incompatible (common)
- 2. ROM or firmware verification failure (occasional)
Verify modem recognition
- Test point
- PC card slot
- Expected
- Modem properly recognized by system
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fax/Modem card not recognized or responding
Fax/Modem card fails to recognize, respond to AT commands, or fails ROM/firmware verification
Ranked causes
- 1. Modem card not properly inserted or incompatible (unknown)
- 2. ROM or firmware verification failure (unknown)
Verify modem card properly inserted
- Test point
- PC card slot on rear
- Expected
- Approved modem card seated fully
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard is typing incorrect characters
Keyboard input produces wrong characters on display
Ranked causes
- 1. Keyboard Caps Lock is enabled (common)
- 3. Wrong keyboard version selected during software installation (rare)
Inspect character case
- Test point
- display
- Expected
- only character case incorrect if Caps Lock enabled
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Magnetic card reader data not reading correctly
Magnetic card reader fails to read data from magnetic card or data appears incorrect
Ranked causes
- 1. Magnetic card reader not properly attached (common)
- 2. Magnetic card reader malfunction or card issue (occasional)
Verify magnetic card reader attachment
- Test point
- Magnetic card reader port on rear of cardiograph
- Expected
- Reader securely connected
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Magnetic card reader not functioning
Magnetic card reader fails to read card data or does not display correct track information
Ranked causes
- 1. Magnetic card reader not attached to rear port (common)
- 2. Magnetic card reader hardware failure (occasional)
Insert magnetic card and verify track 1 data
- Test point
- Magnetic card reader slot on front
- Expected
- Magnetic card track 1 data displayed correctly
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Magnetic Card Reader not reading data
Magnetic card reader fails to display card data or displays incorrect data
Ranked causes
- 1. Magnetic Card Reader hardware failure or misalignment (unknown)
- 2. Magnetic Card Reader not installed or not attached to rear port (unknown)
Verify magnetic card reader attached to rear port
- Test point
- Rear port connection
- Expected
- Reader physically attached before power-on
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Network connection not displaying or incorrect caution
Network Connection window does not display network data or displays incorrect network configuration information
Ranked causes
- 1. LAN card not properly inserted or seated (common)
- 2. Non-approved or damaged LAN cable (occasional)
Verify LAN card insertion in PC Card slot
- Test point
- PC Card slot on rear of cardiograph
- Expected
- LAN card fully inserted
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Occasionally copying ECGs from archive to PC card/USB stick fails
Write operation to PC card or USB Memory Stick fails intermittently during ECG archive transfer
Ranked causes
- 1. PC card or USB memory stick subsystem time-out failure (occasional)
- 2. Defective or worn PC card/USB Memory Stick (occasional)
Retry ECG copy to media
- Test point
- Transfer operation status
- Expected
- Transfer completes without error
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not tear off cleanly or stops at wrong spot
Paper tearing at print end is not clean or paper feed stops at incorrect location
Ranked causes
- 1. Wrong paper size setting in configuration (common)
- 2. Incorrect paper tear-off method used (common)
Verify paper size setting in configuration menu
- Test point
- Configuration display
- Expected
- Paper size matches physical stock loaded (e.g., Letter or A4)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper tear off failure or incorrect advance
Paper does not tear off cleanly or stops at wrong position
Ranked causes
- 1. Wrong paper size setting in configuration (common)
- 2. Incorrect method used to tear paper (common)
Verify paper size configuration
- Test point
- Cardiograph configuration menu
- Expected
- Paper size correctly set
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC Card not detected or reports error
PCMCIA card not detected or error reported when polled during PC Card test
Ranked causes
- 1. PCMCIA card not properly inserted (common)
- 2. PCMCIA card hardware failure (occasional)
Insert blank PCMCIA card with adapter in slot
- Test point
- PCMCIA slot on rear of cardiograph
- Expected
- Card detected by system
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC card or USB transfer failure
Copying ECGs from archive to Flash card or USB Memory Stick fails occasionally
Ranked causes
- 1. PC card subsystem time-out failure (occasional)
- 2. Flash card defect (rare)
Retry transfer operation
- Test point
- PC card interface
- Expected
- Operation succeeds without error
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM communication failure
Cardiograph unable to acquire data from Patient Interface Module or displays PIM error messages
Ranked causes
- 1. PIM patient data cable not securely attached (common)
- 2. PIM hardware malfunction (occasional)
Verify PIM patient data cable connection
- Test point
- PIM connector on rear of cardiograph
- Expected
- Cable securely attached
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM not communicating with cardiograph patient safety
PIM test fails indicating Patient Interface Module is not communicating or unable to acquire data from PIM
Ranked causes
- 1. PIM patient data cable loose or disconnected (common)
- 2. PIM hardware failure or communication bus error (occasional)
Verify PIM patient data cable is firmly connected
- Test point
- PIM connector on rear of cardiograph
- Expected
- Cable fully seated and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM test fails or PIM communication error
Cardiograph displays PIM error messages or is unable to acquire data from the Patient Interface Module during normal operation or when PIM Test is executed
Ranked causes
- 1. PIM patient data cable not securely attached or disconnected (common)
- 2. PIM hardware malfunction (occasional)
Verify PIM patient data cable connection
- Test point
- PIM connector on rear of cardiograph
- Expected
- Cable securely attached with no visible damage or corrosion
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not detected message
Cardiograph displays Printer Not Detected message; affects Trim II/III/Rx
Ranked causes
- 1. Unseated or defective main harness assembly (occasional)
- 2. Defective printer control board (occasional)
Reset unit
- Test point
- Unit control
- Expected
- Message clears
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer test page does not print or displays incorrectly
Printer test page fails to print or output does not match test points specification
Ranked causes
- 1. Printer hardware malfunction (common)
- 2. Printer not properly connected or configured (occasional)
Select Printer Test and Start from Biomed Service Utility
- Test point
- Printer output
- Expected
- Test page prints with all test points visible
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer test page has quality defects or fails inspection
Printer test page prints with distortion, missing segments, incorrect spacing, breaks in lines, or illegible characters
Ranked causes
- 1. Printer hardware malfunction or wear (common)
- 2. Printer supply consumables depleted or degraded (occasional)
Inspect printer test page at point A for bar quality
- Test point
- Stepped bars on test page (Point A)
- Expected
- Stepped bars are sharp edged and printed cleanly without distortion or missing segments
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen shows CRC check failure
Unit stops at COPYRIGHT screen with warning dialog displaying CRC failed message
Ranked causes
- 1. Application software is damaged (common)
- 2. File system is destroyed (occasional)
Observe boot screen and error message
- Test point
- Display
- Expected
- Unit stops at COPYRIGHT screen with CRC failed dialog
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Smart Card reader not functioning
Smart Card reader fails to read card data or does not display card information correctly
Ranked causes
- 1. Smart Card reader not connected (common)
- 2. Smart Card reader hardware failure or card defect (occasional)
Insert Smart Card and verify data display
- Test point
- Smart Card reader slot
- Expected
- Smart Card data appears and is correct
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Smart Card Reader not reading data
Smart Card reader fails to display card data or displays incorrect data
Ranked causes
- 1. Smart Card Reader hardware failure or misalignment (unknown)
- 2. Smart Card Reader not installed (unknown)
Verify Smart Card reader installed and functional
- Test point
- Smart Card slot on front
- Expected
- Smart Card data displays correctly on screen
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Software version levels inconsistent
Version test fails indicating version levels of Application, Kernel, Loader, PIM Kernel, or Lead Set are inconsistent
Ranked causes
- 1. PIM not connected to cardiograph (common)
- 2. Software version mismatch after upgrade (occasional)
From Biomed Service Program select Test, Test Version, Start
- Test point
- Software version information
- Expected
- Application, Kernel, Loader, PIM Kernel, Lead Set versions all consistent
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unexpected reply from Remote Site Server
Unexpected Reply Received from Remote Site error; server responding but not properly acknowledging
Ranked causes
- 1. Remote Site Server URL incorrectly formatted (common)
- 2. TraceMasterVue server requires reset (occasional)
Verify Remote Site Server URL format
- Test point
- Configuration screen Remote Sites tab
- Expected
- URL format: http://<ip or hostname>/path/ with trailing slash
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not power on or show software identification screen
Display does not show PageWriter Trim software identification screen or R/T ECG screen after power on; no audible beep
Ranked causes
- 1. Power cord or AC connection issue (common)
- 2. Reset button needs activation (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit does not restart after Power On test attempt
Cardiograph fails to display software identification screen or R/T ECG screen after pressing On/Standby button or Reset button
Ranked causes
- 1. Power button not held long enough (common)
- 2. Reset button procedure not performed correctly (occasional)
Verify correct Power On button operation
- Test point
- On/Standby button
- Expected
- Display shows PageWriter Trim software identification screen after ~20 seconds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Weak wireless signal or signal quality issues
RED LED blinking quickly and GREEN LED blinking quickly; weak wireless signal strength
Ranked causes
- 1. Cardiograph located in weak signal area (common)
- 2. Adapter still not associated after relocation (occasional)
Move cardiograph to area with stronger wireless signal
- Test point
- RF signal strength indicator on ACU
- Expected
- Green or yellow bars appear in Signal Strength field
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
After attempting Archive screen operation, error message appears: Command timed-out
Error dialog displays Command timed-out after user tries Archive operation; transfer or query to remote system fails
Ranked causes
- 1. TraceMaster system is busy (common)
Retry Archive operation
- Test point
- Archive screen command
- Expected
- Operation succeeds on subsequent try
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
After pressing the Archive button, it takes longer than 40 seconds before the Archive list is displayed
Archive list display is delayed beyond 40 seconds when main archive contains more than 130 stored ECGs; slow system response
Ranked causes
- 1. System slow-down from full ECG database (common)
Check number of ECGs in main archive
- Test point
- Archive storage
- Expected
- More than 130 ECGs stored; known performance impact near maximum 200 ECGs
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive save to removable media fails
Error when attempting to save ECGs to PC card or USB memory stick
Ranked causes
- 1. Removable media not fully inserted (common)
Verify media insertion
- Test point
- PC card or USB slot
- Expected
- Media fully inserted and recognized
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive screen 'Requested Function Failed' error
Error message 'Requested Function Failed' appears during Archive operations
Ranked causes
- 1. Network or modem connection interrupted (common)
Retry operation
- Test point
- Network connection
- Expected
- Operation succeeds on retry
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive screen command timed-out error
Error message 'Command timed-out' appears during Archive screen operations
Ranked causes
- 1. TraceMaster ECG system busy (occasional)
Retry Archive operation
- Test point
- TraceMaster interface
- Expected
- Operation succeeds on retry
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive screen slow with >130 ECGs stored caution
Archive list display takes longer than 40 seconds to appear when main archive near full
Ranked causes
- 1. Full ECG database causing system slowdown (common)
Delete ECGs from main archive
- Test point
- Main Archive storage
- Expected
- Number of ECGs reduced below 130
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio not emitting from cardiograph
Cardiograph does not emit audible beeping tone during audio test
Ranked causes
- 1. Audio system malfunction (unknown)
Verify audio output during Audio Test
- Test point
- Cardiograph speaker
- Expected
- Audible beeping tone
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio test fails
Cardiograph does not emit audible beeping tones during audio diagnostic test
Ranked causes
- 1. Audio system malfunction or disconnection (unknown)
Verify audible tones are functioning on Display Control Unit and injector head
- Test point
- Display Control Unit and injector head speakers
- Expected
- Audible beeping tones present
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery voltage out of limits
Battery voltage test shows voltage is below or above acceptable voltage limits
Ranked causes
- 1. Battery discharged or degraded (common)
From Biomed Service Program select Test, Test Battery, Start
- Test point
- Battery voltage
- Expected
- Voltage within acceptable limits
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Blank pages print after report
- Blank pages printed after ECG report
- Calibration: Barcode reader
- Can not enter the configuration window
- Cannot enter configuration window
- Cardiograph operating outside acceptable voltage limits
- Cardiograph unresponsive after rapid drawer cycling
- Check Printer message with no printer problem present
- Chinese characters missing in test report
- Compact Flash archive storage not functioning
- Compact Flash storage not responding
- Compressed or small printed report
- Configuration options inconsistent or improperly configured
- Corrupted print image after drawer operation
- CRC startup failure
- Dialog shows software version information with black background
- Display shows software version with black background
- ECG transfer to removable media fails with <200 files
- Error 0202: Printer busy, cannot accept new print job
- Error 0205: No paper in drawer
- Fax transmission 'Requested Function Failure' message persists
- Invalid WEP key format (10 hex digits)
- Invalid WEP key format (26 hex digits)
- Keyboard is unresponsive
- Keyboard keys not recognized during test
- Keyboard keys not responding or not recognized
- Keyboard test fails
- LCD screen characters not displaying correctly
- Modem unavailable after card reinsertion
- Modem unavailable after PC card removal/reinsertion or USB reconnection
- Multiple ECG transfer interrupted, files not deleted
- No PIM or defective PIM at startup
- On/Standby button does not function
- Onboard Flash storage not responding
- Onboard Flash test fails
- Paper does not advance
- Print image corrupted after drawer opened during printing
- Print image corrupted and paper does not stop feeding
- Printed report looks too small or compressed
- Printer controller error message
- Printer Door is Open message and cardiograph unresponsive
- Printer door open - unresponsive unit
- Printer drawer open
- Printer error code 0202
- Printer error code 0205
- Printer error code 0206
- Printer test page has quality errors
- Printing continues on second page unexpectedly
- RAM memory test fails
- Retrieved ECGs look different from TraceMaster or XLi printouts
- Screen display fails color or pattern verification
- Some ECGs retrieved from TraceMaster and printed on the cardiograph look different from TraceMaster or XLi printouts
- Start Up CRC Failure
- System running slowly after boot-up
- System seems to be running slowly after boot-up
- Temperature shutdown
- Test report missing Chinese characters on Chinese cardiograph
- The message Requested Function Failure appears after fax ECG transmission has completed
- Transfer of an ECG to a PC card flash archive fails and there are still less than 200 ECGs on the flash card
- Trim Knob does not move or responds incorrectly
- Unit unresponsive after rapid drawer operations
- Unknown error
- Unknown error on startup
- Unsupported ECG report type fails to transfer to TraceMaster
- USB printer - door open
- USB printer - No USB connection
- USB printer - running (busy)
- USB printer busy
- USB printer not detected
- Waveform display is no longer updating, system is unresponsive
- Waveform display not updating, system unresponsive
- WEP key not entered
- When an error occurs during transfer of multiple ECGs to a TraceMaster system, the transfer process does not continue
- Wireless adapter not found
- Wireless LAN transfer fails without LEAP credentials
75 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
0101 | Start Up CRC Failure - firmware integrity check failed | System detects corrupted or missing firmware during boot sequence | critical |
0201 | USB printer - no USB connection detected | Cardiograph cannot communicate with USB printer device | error |
0201 | No battery or dead battery | Printer detects missing or depleted battery during power-up | error |
0202 | USB printer - printer is busy or stalled | Printer is in busy state and not accepting commands; print job may be stalled | warning |
0202 | UPR running - printer busy | New print job sent while printer processing previous job | warning |
0203 | USB printer - door open | Printer drawer open during operation or not fully closed | warning |
0203 | USB printer - paper drawer is open | Drawer open sensor activated or not fully latched | warning |
0204 | USB printer - paper jam or out of paper | Printer detects missing paper or jammed paper in feed path | error |
0205 | No paper in drawer | Printer detects empty paper drawer during operation | info |
0206 | Temperature shutdown - print head overheated | Print head temperature exceeds safe operating threshold | warning |
0207 | USB printer - unknown error | Unidentified printer error condition | critical |
0207 | Unknown error or no PIM detected | System firmware error, corrupted data, or PIM module not connected/detected at boot | error |
350 Series Radio is Not Associated | Wireless adapter cannot connect to any access point | Wireless adapter powered on but no successful association with access point | warning |
Cisco Wireless LAN Adapter Not Found | Wireless adapter cannot be detected by cardiograph | Aironet Client Utility launched without adapter inserted or adapter was ejected during operation | error |
No Reply Received From Remote Site Server | Cardiograph cannot reach or communicate with TraceMasterVue ECG server | ECG transmission attempt when wireless connection, network path, or server response unavailable | error |
Unexpected Reply Received from Remote Site | Server received transmission but did not properly acknowledge or respond to packet data | ECG transmission to TraceMasterVue server when server is online but sending malformed acknowledgment | warning |
Unidentified PC card Adapter | Cardiograph cannot identify or load driver for wireless adapter | Unsupported or incompletely inserted wireless adapter detected during boot or driver load sequence | error |
WEP Key X Must Be 10 Hex Digits | 40-bit WEP key has invalid character count or format | User enters WEP key with incorrect number of characters or invalid hex characters for 40-bit key | error |
WEP Key X Must Be 26 Hex Digits | 128-bit WEP key has invalid character count or format | User enters WEP key with incorrect number of characters or invalid hex characters for 128-bit key | error |
You must enter a WEP Key | WEP security enabled but no encryption key provided | WEP authentication configured but WEP key field left blank on WEP keys window | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI