Philips PageWriter Trim I: troubleshooting and error codes

The ServiceForge knowledge base holds 135 documented symptoms for the Philips PageWriter Trim I, with 216 ranked causes, 202 diagnostic checks with test points and expected values, and 20 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Screen is dark and shows no image

Display remains blank or very faint with no visible content

Ranked causes

  1. 1. Cardiograph is in standby mode (common)
  2. 2. Contrast adjusted to lowest degree (occasional)
  3. 3. AC power cord disconnected or damaged (occasional)
  4. 4. Battery depleted (occasional)
  5. 5. Power supply failure (rare)
First check

Verify cardiograph power state

Test point
front display
Expected
cardiograph wakes up

4 more checks with pass and fail actions, 9 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

All leads show leads off continuously patient safety

Red dashed line displayed for all ECG leads indicating no signal detected

Ranked causes

  1. 1. PIM communications lost or defective leads (common)
  2. 2. Precordial lead defect (occasional)
  3. 3. RL/R lead defect (occasional)
  4. 4. Internal cardiograph hardware failure (rare)
First check

Verify PIM cable seating

Test point
PIM rear connector
Expected
cable firmly seated

4 more checks with pass and fail actions, 10 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error message during software installation: incorrect model, token, or invalid key

Software install fails with error about incorrect product model, invalid token, or key mismatch

Ranked causes

  1. 1. PC card not fully inserted before power-up (common)
  2. 2. Software files on card are for wrong product model (occasional)
  3. 3. Invalid token key (occasional)
  4. 4. Invalid cardiograph serial number (occasional)
First check

Verify PC card fully inserted before boot

Test point
PC card slot
Expected
Card seat fully with no gaps; no bent or broken pins visible on mainboard

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer error - out of paper, jam, or door open

Combined printer error condition indicating multiple possible causes on Trim II/III/Rx

Ranked causes

  1. 1. Paper drawer not fully closed (common)
  2. 2. Out of paper or incorrectly loaded (common)
  3. 3. Defective TOF sensor assembly (occasional)
  4. 4. Defective paper tray assembly (occasional)
First check

Close drawer firmly

Test point
Printer drawer
Expected
Drawer fully closed

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer error code 0204 caution

Printer error 0204 displayed; paper jam, incorrect paper loading, defective TOF sensor, or defective paper tray assembly

Ranked causes

  1. 1. Paper jam (common)
  2. 2. Paper is not loaded correctly (common)
  3. 3. Defective TOF sensor assembly (occasional)
  4. 4. Defective paper tray assembly (occasional)
First check

Inspect paper path for jam

Test point
Printer interior
Expected
Paper jam present or paper path clear

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer Error: Out of Paper/Paper Jam/Door Open

Error message indicating paper drawer, paper feed, or sensor issue

Ranked causes

  1. 1. Printer drawer not fully closed (common)
  2. 2. Out of paper or paper loaded incorrectly (common)
  3. 3. Defective TOF sensor assembly (occasional)
  4. 4. Defective paper tray assembly (occasional)
First check

Inspect and close printer drawer

Test point
Drawer position and latch
Expected
Drawer fully seated and latched

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wireless adapter not associated to access point

350 Series Radio is Not Associated error; adapter cannot connect to wireless network

Ranked causes

  1. 1. Weak signal or out of range (common)
  2. 2. Access point not configured or enabled (occasional)
  3. 3. Wireless adapter settings mismatch (occasional)
  4. 4. LEAP authentication credentials not enabled (occasional)
First check

Check LED indicators on wireless adapter

Test point
Wireless adapter LEDs
Expected
Red LED blinking slowly or off; Green LED blinking slowly or off

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

After attempting Archive screen operation, error message appears: Requested Function Failed

Error dialog displays Requested Function Failed after Archive screen operation; transfer or query incomplete

Ranked causes

  1. 1. Network or modem phone line was interrupted during operation (common)
  2. 2. Selected external archive media was not available (occasional)
  3. 3. An attempt to transfer an unsupported ECG report type was made to TraceMaster (occasional)
First check

Verify network/phone line integrity

Test point
Network or modem connection
Expected
Cabling intact and connection stable

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode reader does not scan or returns inaccurate data

Barcode reader attached to rear port fails to scan test barcode or scans barcode data that does not match expected data

Ranked causes

  1. 1. Barcode reader scanning angle incorrect (common)
  2. 2. Barcode reader calibration drift (occasional)
  3. 3. Barcode reader connection loose or faulty (occasional)
First check

Verify barcode reader is attached to correct port

Test point
Barcode reader port on rear of cardiograph
Expected
Barcode reader securely seated in port before power-on

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 0201: No battery or dead battery caution

Printer reports no battery or battery insufficient power via error code 0201

Ranked causes

  1. 1. Battery missing or depleted (common)
  2. 2. Motherboard failure (rare)
  3. 3. Power supply cord defective (occasional)
First check

Verify battery presence and condition

Test point
Battery compartment
Expected
Battery present and charged

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 0203: Paper drawer issue or out of paper caution

Printer reports paper drawer not closed, out of paper, or paper loaded incorrectly

Ranked causes

  1. 1. Paper drawer not fully closed (common)
  2. 2. Out of paper or incorrectly loaded (common)
  3. 3. Application error (occasional)
First check

Close printer drawer

Test point
Printer drawer
Expected
Drawer fully closed and latched

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error 0204: Paper jam or sensor failure caution

Printer reports paper jam, incorrect loading, defective sensor or tray assembly

Ranked causes

  1. 1. Paper jam or incorrect loading (common)
  2. 2. Defective TOF sensor assembly (occasional)
  3. 3. Defective paper tray assembly (occasional)
First check

Clear paper jam and verify loading

Test point
Printer paper path
Expected
Paper path clear, paper properly loaded

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Magnetic card reader does not read or returns inaccurate data

Magnetic card reader fails to read data from magnetic card or reads data that does not match expected card contents

Ranked causes

  1. 1. Magnetic card not inserted correctly or left in slot too briefly (common)
  2. 2. Magnetic card reader connection loose (occasional)
  3. 3. Magnetic card reader malfunction (rare)
First check

Verify magnetic card reader is connected

Test point
Magnetic card reader connector on rear of cardiograph
Expected
Magnetic card reader securely attached to port before power-on

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Network test shows no connection or incorrect configuration caution

Network Connection window does not display configured network information or displays incorrect network data during network test

Ranked causes

  1. 1. LAN card not fully inserted in PC Card slot (common)
  2. 2. LAN cable not fully inserted into LAN card connector (common)
  3. 3. Incompatible or damaged LAN cable (occasional)
First check

Verify LAN card is fully inserted in PC Card slot

Test point
LAN card in PC Card slot on rear of cardiograph
Expected
LAN card fully seated in slot with no visible gaps or loose contacts

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No reply from Remote Site Server

No Reply Received From Remote Site Server error; ECG transmission connection failure

Ranked causes

  1. 1. Wireless connection to access point lost (common)
  2. 2. Network connection to TraceMasterVue server not configured (common)
  3. 3. Cardiograph did not receive response within timeout period (occasional)
First check

Verify wireless adapter LED status

Test point
Wireless adapter LEDs
Expected
Green LED associated with access point

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

One or more leads show leads off periodically patient safety

Red dashed line shown on leads display, periodic signal loss reported by cardiograph

Ranked causes

  1. 1. PIM patient data cable connection failure (common)
  2. 2. Defective lead wires (common)
  3. 3. PIM PCA set failure (occasional)
First check

Verify PIM cable connection

Test point
PIM connector on cardiograph rear
Expected
Cable firmly seated

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer error code 0201

Printer error 0201 displayed; no battery detected or battery is dead; printer cannot operate without power

Ranked causes

  1. 1. No battery or dead battery (common)
  2. 2. Motherboard failure (occasional)
  3. 3. Power supply cord failure (occasional)
First check

Verify battery presence and condition

Test point
Battery compartment
Expected
Battery present and functional

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer error code 0203

Printer error 0203 displayed; paper drawer not fully closed, out of paper, incorrectly loaded paper, or application error

Ranked causes

  1. 1. Paper drawer is not fully closed (common)
  2. 2. Out of paper or paper loaded incorrectly (common)
  3. 3. Application error (occasional)
First check

Verify printer drawer is fully closed

Test point
Printer drawer latch
Expected
Drawer fully closed and latched

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer paper jam or out of paper

Error code 0204 indicates paper jam, paper depletion, or paper feed failure

Ranked causes

  1. 1. Paper jam in printer mechanism (common)
  2. 2. Defective TOF sensor assembly (occasional)
  3. 3. Defective paper tray assembly (occasional)
First check

Clear paper jam and verify paper loading

Test point
Printer paper mechanism
Expected
Paper properly loaded, no jammed paper visible

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unidentified PC card adapter detected

Unidentified PC card Adapter error with dialog to enter driver name

Ranked causes

  1. 1. Wireless adapter inserted incorrectly into PC card slot (common)
  2. 2. Adapter inserted during active use (not in Standby) (occasional)
  3. 3. Unsupported wireless adapter installed (rare)
First check

Verify wireless adapter is properly seated

Test point
PC card slot on rear of cardiograph
Expected
Adapter fully inserted and secure

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

USB printer - paper jam or out of paper

Printer reports paper jam condition or no paper detected

Ranked causes

  1. 1. Paper drawer not closed or paper misaligned (common)
  2. 2. Defective TOF sensor assembly (occasional)
  3. 3. Defective paper tray assembly (occasional)
First check

Close printer drawer firmly

Test point
Drawer latch
Expected
Drawer fully seated

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wireless adapter not properly inserted or detected

PC card adapter insertion issue; RED LED off and GREEN LED off

Ranked causes

  1. 1. Wireless adapter not securely inserted in PC card slot (common)
  2. 2. Cardiograph power cycle required (common)
  3. 3. Wireless adapter hardware failure (rare)
First check

Verify wireless adapter seating in PC card slot

Test point
PC card slot on rear of cardiograph
Expected
Adapter fully and securely inserted

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode reader data not scanning accurately

Barcode reader fails to accurately scan barcode data or data appears incorrect after scanning

Ranked causes

  1. 1. Barcode reader not properly attached to port (common)
  2. 2. Barcode reader requires calibration (occasional)
First check

Verify barcode reader attachment

Test point
Barcode reader port on rear of cardiograph
Expected
Barcode reader securely connected

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode reader not functioning

Barcode reader fails to read barcode or does not display correct barcode data

Ranked causes

  1. 1. Barcode reader not attached to rear port (common)
  2. 2. Barcode reader requires calibration (occasional)
First check

Scan barcode at 45 degree angle and verify data display

Test point
Barcode reader sensor
Expected
Barcode data displayed correctly in Bar Code Field

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Barcode Reader not reading barcode

Barcode reader fails to capture or display barcode data correctly

Ranked causes

  1. 1. Barcode Reader hardware failure or improper scanning technique (unknown)
  2. 2. Barcode Reader not installed or not attached to rear port (unknown)
First check

Verify barcode reader attached to rear port

Test point
Rear port connection
Expected
Reader physically attached before power-on

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fax/modem card not recognized or not functioning

Fax/modem card fails recognition test or ROM/firmware verification

Ranked causes

  1. 1. Fax/modem card not inserted or incompatible (common)
  2. 2. ROM or firmware verification failure (occasional)
First check

Verify modem recognition

Test point
PC card slot
Expected
Modem properly recognized by system

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fax/Modem card not recognized or responding

Fax/Modem card fails to recognize, respond to AT commands, or fails ROM/firmware verification

Ranked causes

  1. 1. Modem card not properly inserted or incompatible (unknown)
  2. 2. ROM or firmware verification failure (unknown)
First check

Verify modem card properly inserted

Test point
PC card slot on rear
Expected
Approved modem card seated fully

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard is typing incorrect characters

Keyboard input produces wrong characters on display

Ranked causes

  1. 1. Keyboard Caps Lock is enabled (common)
  2. 3. Wrong keyboard version selected during software installation (rare)
First check

Inspect character case

Test point
display
Expected
only character case incorrect if Caps Lock enabled

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Magnetic card reader data not reading correctly

Magnetic card reader fails to read data from magnetic card or data appears incorrect

Ranked causes

  1. 1. Magnetic card reader not properly attached (common)
  2. 2. Magnetic card reader malfunction or card issue (occasional)
First check

Verify magnetic card reader attachment

Test point
Magnetic card reader port on rear of cardiograph
Expected
Reader securely connected

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Magnetic card reader not functioning

Magnetic card reader fails to read card data or does not display correct track information

Ranked causes

  1. 1. Magnetic card reader not attached to rear port (common)
  2. 2. Magnetic card reader hardware failure (occasional)
First check

Insert magnetic card and verify track 1 data

Test point
Magnetic card reader slot on front
Expected
Magnetic card track 1 data displayed correctly

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Magnetic Card Reader not reading data

Magnetic card reader fails to display card data or displays incorrect data

Ranked causes

  1. 1. Magnetic Card Reader hardware failure or misalignment (unknown)
  2. 2. Magnetic Card Reader not installed or not attached to rear port (unknown)
First check

Verify magnetic card reader attached to rear port

Test point
Rear port connection
Expected
Reader physically attached before power-on

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Network connection not displaying or incorrect caution

Network Connection window does not display network data or displays incorrect network configuration information

Ranked causes

  1. 1. LAN card not properly inserted or seated (common)
  2. 2. Non-approved or damaged LAN cable (occasional)
First check

Verify LAN card insertion in PC Card slot

Test point
PC Card slot on rear of cardiograph
Expected
LAN card fully inserted

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Occasionally copying ECGs from archive to PC card/USB stick fails

Write operation to PC card or USB Memory Stick fails intermittently during ECG archive transfer

Ranked causes

  1. 1. PC card or USB memory stick subsystem time-out failure (occasional)
  2. 2. Defective or worn PC card/USB Memory Stick (occasional)
First check

Retry ECG copy to media

Test point
Transfer operation status
Expected
Transfer completes without error

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper does not tear off cleanly or stops at wrong spot

Paper tearing at print end is not clean or paper feed stops at incorrect location

Ranked causes

  1. 1. Wrong paper size setting in configuration (common)
  2. 2. Incorrect paper tear-off method used (common)
First check

Verify paper size setting in configuration menu

Test point
Configuration display
Expected
Paper size matches physical stock loaded (e.g., Letter or A4)

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper tear off failure or incorrect advance

Paper does not tear off cleanly or stops at wrong position

Ranked causes

  1. 1. Wrong paper size setting in configuration (common)
  2. 2. Incorrect method used to tear paper (common)
First check

Verify paper size configuration

Test point
Cardiograph configuration menu
Expected
Paper size correctly set

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PC Card not detected or reports error

PCMCIA card not detected or error reported when polled during PC Card test

Ranked causes

  1. 1. PCMCIA card not properly inserted (common)
  2. 2. PCMCIA card hardware failure (occasional)
First check

Insert blank PCMCIA card with adapter in slot

Test point
PCMCIA slot on rear of cardiograph
Expected
Card detected by system

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PC card or USB transfer failure

Copying ECGs from archive to Flash card or USB Memory Stick fails occasionally

Ranked causes

  1. 1. PC card subsystem time-out failure (occasional)
  2. 2. Flash card defect (rare)
First check

Retry transfer operation

Test point
PC card interface
Expected
Operation succeeds without error

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM communication failure

Cardiograph unable to acquire data from Patient Interface Module or displays PIM error messages

Ranked causes

  1. 1. PIM patient data cable not securely attached (common)
  2. 2. PIM hardware malfunction (occasional)
First check

Verify PIM patient data cable connection

Test point
PIM connector on rear of cardiograph
Expected
Cable securely attached

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM not communicating with cardiograph patient safety

PIM test fails indicating Patient Interface Module is not communicating or unable to acquire data from PIM

Ranked causes

  1. 1. PIM patient data cable loose or disconnected (common)
  2. 2. PIM hardware failure or communication bus error (occasional)
First check

Verify PIM patient data cable is firmly connected

Test point
PIM connector on rear of cardiograph
Expected
Cable fully seated and secure

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM test fails or PIM communication error

Cardiograph displays PIM error messages or is unable to acquire data from the Patient Interface Module during normal operation or when PIM Test is executed

Ranked causes

  1. 1. PIM patient data cable not securely attached or disconnected (common)
  2. 2. PIM hardware malfunction (occasional)
First check

Verify PIM patient data cable connection

Test point
PIM connector on rear of cardiograph
Expected
Cable securely attached with no visible damage or corrosion

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer not detected message

Cardiograph displays Printer Not Detected message; affects Trim II/III/Rx

Ranked causes

  1. 1. Unseated or defective main harness assembly (occasional)
  2. 2. Defective printer control board (occasional)
First check

Reset unit

Test point
Unit control
Expected
Message clears

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer test page does not print or displays incorrectly

Printer test page fails to print or output does not match test points specification

Ranked causes

  1. 1. Printer hardware malfunction (common)
  2. 2. Printer not properly connected or configured (occasional)
First check

Select Printer Test and Start from Biomed Service Utility

Test point
Printer output
Expected
Test page prints with all test points visible

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer test page has quality defects or fails inspection

Printer test page prints with distortion, missing segments, incorrect spacing, breaks in lines, or illegible characters

Ranked causes

  1. 1. Printer hardware malfunction or wear (common)
  2. 2. Printer supply consumables depleted or degraded (occasional)
First check

Inspect printer test page at point A for bar quality

Test point
Stepped bars on test page (Point A)
Expected
Stepped bars are sharp edged and printed cleanly without distortion or missing segments

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Screen shows CRC check failure

Unit stops at COPYRIGHT screen with warning dialog displaying CRC failed message

Ranked causes

  1. 1. Application software is damaged (common)
  2. 2. File system is destroyed (occasional)
First check

Observe boot screen and error message

Test point
Display
Expected
Unit stops at COPYRIGHT screen with CRC failed dialog

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Smart Card reader not functioning

Smart Card reader fails to read card data or does not display card information correctly

Ranked causes

  1. 1. Smart Card reader not connected (common)
  2. 2. Smart Card reader hardware failure or card defect (occasional)
First check

Insert Smart Card and verify data display

Test point
Smart Card reader slot
Expected
Smart Card data appears and is correct

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Smart Card Reader not reading data

Smart Card reader fails to display card data or displays incorrect data

Ranked causes

  1. 1. Smart Card Reader hardware failure or misalignment (unknown)
  2. 2. Smart Card Reader not installed (unknown)
First check

Verify Smart Card reader installed and functional

Test point
Smart Card slot on front
Expected
Smart Card data displays correctly on screen

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Software version levels inconsistent

Version test fails indicating version levels of Application, Kernel, Loader, PIM Kernel, or Lead Set are inconsistent

Ranked causes

  1. 1. PIM not connected to cardiograph (common)
  2. 2. Software version mismatch after upgrade (occasional)
First check

From Biomed Service Program select Test, Test Version, Start

Test point
Software version information
Expected
Application, Kernel, Loader, PIM Kernel, Lead Set versions all consistent

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unexpected reply from Remote Site Server

Unexpected Reply Received from Remote Site error; server responding but not properly acknowledging

Ranked causes

  1. 1. Remote Site Server URL incorrectly formatted (common)
  2. 2. TraceMasterVue server requires reset (occasional)
First check

Verify Remote Site Server URL format

Test point
Configuration screen Remote Sites tab
Expected
URL format: http://<ip or hostname>/path/ with trailing slash

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not power on or show software identification screen

Display does not show PageWriter Trim software identification screen or R/T ECG screen after power on; no audible beep

Ranked causes

  1. 1. Power cord or AC connection issue (common)
  2. 2. Reset button needs activation (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit does not restart after Power On test attempt

Cardiograph fails to display software identification screen or R/T ECG screen after pressing On/Standby button or Reset button

Ranked causes

  1. 1. Power button not held long enough (common)
  2. 2. Reset button procedure not performed correctly (occasional)
First check

Verify correct Power On button operation

Test point
On/Standby button
Expected
Display shows PageWriter Trim software identification screen after ~20 seconds

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Weak wireless signal or signal quality issues

RED LED blinking quickly and GREEN LED blinking quickly; weak wireless signal strength

Ranked causes

  1. 1. Cardiograph located in weak signal area (common)
  2. 2. Adapter still not associated after relocation (occasional)
First check

Move cardiograph to area with stronger wireless signal

Test point
RF signal strength indicator on ACU
Expected
Green or yellow bars appear in Signal Strength field

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

After attempting Archive screen operation, error message appears: Command timed-out

Error dialog displays Command timed-out after user tries Archive operation; transfer or query to remote system fails

Ranked causes

  1. 1. TraceMaster system is busy (common)
First check

Retry Archive operation

Test point
Archive screen command
Expected
Operation succeeds on subsequent try

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

After pressing the Archive button, it takes longer than 40 seconds before the Archive list is displayed

Archive list display is delayed beyond 40 seconds when main archive contains more than 130 stored ECGs; slow system response

Ranked causes

  1. 1. System slow-down from full ECG database (common)
First check

Check number of ECGs in main archive

Test point
Archive storage
Expected
More than 130 ECGs stored; known performance impact near maximum 200 ECGs

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Archive save to removable media fails

Error when attempting to save ECGs to PC card or USB memory stick

Ranked causes

  1. 1. Removable media not fully inserted (common)
First check

Verify media insertion

Test point
PC card or USB slot
Expected
Media fully inserted and recognized

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Archive screen 'Requested Function Failed' error

Error message 'Requested Function Failed' appears during Archive operations

Ranked causes

  1. 1. Network or modem connection interrupted (common)
First check

Retry operation

Test point
Network connection
Expected
Operation succeeds on retry

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Archive screen command timed-out error

Error message 'Command timed-out' appears during Archive screen operations

Ranked causes

  1. 1. TraceMaster ECG system busy (occasional)
First check

Retry Archive operation

Test point
TraceMaster interface
Expected
Operation succeeds on retry

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Archive screen slow with >130 ECGs stored caution

Archive list display takes longer than 40 seconds to appear when main archive near full

Ranked causes

  1. 1. Full ECG database causing system slowdown (common)
First check

Delete ECGs from main archive

Test point
Main Archive storage
Expected
Number of ECGs reduced below 130

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio not emitting from cardiograph

Cardiograph does not emit audible beeping tone during audio test

Ranked causes

  1. 1. Audio system malfunction (unknown)
First check

Verify audio output during Audio Test

Test point
Cardiograph speaker
Expected
Audible beeping tone

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio test fails

Cardiograph does not emit audible beeping tones during audio diagnostic test

Ranked causes

  1. 1. Audio system malfunction or disconnection (unknown)
First check

Verify audible tones are functioning on Display Control Unit and injector head

Test point
Display Control Unit and injector head speakers
Expected
Audible beeping tones present

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery voltage out of limits

Battery voltage test shows voltage is below or above acceptable voltage limits

Ranked causes

  1. 1. Battery discharged or degraded (common)
First check

From Biomed Service Program select Test, Test Battery, Start

Test point
Battery voltage
Expected
Voltage within acceptable limits

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Blank pages print after report
  • Blank pages printed after ECG report
  • Calibration: Barcode reader
  • Can not enter the configuration window
  • Cannot enter configuration window
  • Cardiograph operating outside acceptable voltage limits
  • Cardiograph unresponsive after rapid drawer cycling
  • Check Printer message with no printer problem present
  • Chinese characters missing in test report
  • Compact Flash archive storage not functioning
  • Compact Flash storage not responding
  • Compressed or small printed report
  • Configuration options inconsistent or improperly configured
  • Corrupted print image after drawer operation
  • CRC startup failure
  • Dialog shows software version information with black background
  • Display shows software version with black background
  • ECG transfer to removable media fails with <200 files
  • Error 0202: Printer busy, cannot accept new print job
  • Error 0205: No paper in drawer
  • Fax transmission 'Requested Function Failure' message persists
  • Invalid WEP key format (10 hex digits)
  • Invalid WEP key format (26 hex digits)
  • Keyboard is unresponsive
  • Keyboard keys not recognized during test
  • Keyboard keys not responding or not recognized
  • Keyboard test fails
  • LCD screen characters not displaying correctly
  • Modem unavailable after card reinsertion
  • Modem unavailable after PC card removal/reinsertion or USB reconnection
  • Multiple ECG transfer interrupted, files not deleted
  • No PIM or defective PIM at startup
  • On/Standby button does not function
  • Onboard Flash storage not responding
  • Onboard Flash test fails
  • Paper does not advance
  • Print image corrupted after drawer opened during printing
  • Print image corrupted and paper does not stop feeding
  • Printed report looks too small or compressed
  • Printer controller error message
  • Printer Door is Open message and cardiograph unresponsive
  • Printer door open - unresponsive unit
  • Printer drawer open
  • Printer error code 0202
  • Printer error code 0205
  • Printer error code 0206
  • Printer test page has quality errors
  • Printing continues on second page unexpectedly
  • RAM memory test fails
  • Retrieved ECGs look different from TraceMaster or XLi printouts
  • Screen display fails color or pattern verification
  • Some ECGs retrieved from TraceMaster and printed on the cardiograph look different from TraceMaster or XLi printouts
  • Start Up CRC Failure
  • System running slowly after boot-up
  • System seems to be running slowly after boot-up
  • Temperature shutdown
  • Test report missing Chinese characters on Chinese cardiograph
  • The message Requested Function Failure appears after fax ECG transmission has completed
  • Transfer of an ECG to a PC card flash archive fails and there are still less than 200 ECGs on the flash card
  • Trim Knob does not move or responds incorrectly
  • Unit unresponsive after rapid drawer operations
  • Unknown error
  • Unknown error on startup
  • Unsupported ECG report type fails to transfer to TraceMaster
  • USB printer - door open
  • USB printer - No USB connection
  • USB printer - running (busy)
  • USB printer busy
  • USB printer not detected
  • Waveform display is no longer updating, system is unresponsive
  • Waveform display not updating, system unresponsive
  • WEP key not entered
  • When an error occurs during transfer of multiple ECGs to a TraceMaster system, the transfer process does not continue
  • Wireless adapter not found
  • Wireless LAN transfer fails without LEAP credentials

75 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
0101Start Up CRC Failure - firmware integrity check failedSystem detects corrupted or missing firmware during boot sequencecritical
0201USB printer - no USB connection detectedCardiograph cannot communicate with USB printer deviceerror
0201No battery or dead batteryPrinter detects missing or depleted battery during power-uperror
0202USB printer - printer is busy or stalledPrinter is in busy state and not accepting commands; print job may be stalledwarning
0202UPR running - printer busyNew print job sent while printer processing previous jobwarning
0203USB printer - door openPrinter drawer open during operation or not fully closedwarning
0203USB printer - paper drawer is openDrawer open sensor activated or not fully latchedwarning
0204USB printer - paper jam or out of paperPrinter detects missing paper or jammed paper in feed patherror
0205No paper in drawerPrinter detects empty paper drawer during operationinfo
0206Temperature shutdown - print head overheatedPrint head temperature exceeds safe operating thresholdwarning
0207USB printer - unknown errorUnidentified printer error conditioncritical
0207Unknown error or no PIM detectedSystem firmware error, corrupted data, or PIM module not connected/detected at booterror
350 Series Radio is Not AssociatedWireless adapter cannot connect to any access pointWireless adapter powered on but no successful association with access pointwarning
Cisco Wireless LAN Adapter Not FoundWireless adapter cannot be detected by cardiographAironet Client Utility launched without adapter inserted or adapter was ejected during operationerror
No Reply Received From Remote Site ServerCardiograph cannot reach or communicate with TraceMasterVue ECG serverECG transmission attempt when wireless connection, network path, or server response unavailableerror
Unexpected Reply Received from Remote SiteServer received transmission but did not properly acknowledge or respond to packet dataECG transmission to TraceMasterVue server when server is online but sending malformed acknowledgmentwarning
Unidentified PC card AdapterCardiograph cannot identify or load driver for wireless adapterUnsupported or incompletely inserted wireless adapter detected during boot or driver load sequenceerror
WEP Key X Must Be 10 Hex Digits40-bit WEP key has invalid character count or formatUser enters WEP key with incorrect number of characters or invalid hex characters for 40-bit keyerror
WEP Key X Must Be 26 Hex Digits128-bit WEP key has invalid character count or formatUser enters WEP key with incorrect number of characters or invalid hex characters for 128-bit keyerror
You must enter a WEP KeyWEP security enabled but no encryption key providedWEP authentication configured but WEP key field left blank on WEP keys windowerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.