Philips PageWriter Touch: troubleshooting and error codes
The ServiceForge knowledge base holds 88 documented symptoms for the Philips PageWriter Touch, with 136 ranked causes, 161 diagnostic checks with test points and expected values, and 17 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Wireless adapter not associating with access point
Wireless adapter fails to establish association with access point; LED indicators show lack of association or connection failure
Ranked causes
- 1. Weak wireless signal reception (common)
- 2. Incorrect wireless profile or configuration selected (occasional)
- 3. Access point not powered on or not operating (occasional)
- 4. MAC or IP filtering blocking adapter connection (occasional)
- 5. WEP key mismatch or incorrect index (occasional)
- 6. Wired Ethernet DHCP taking precedence over wireless (occasional)
- 7. Wireless adapter not properly seated or not powered (rare)
Check wireless adapter LED indicators
- Test point
- Red and Green LEDs on wireless adapter
- Expected
- LED state indicates association status per Table 5-14
7 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen is dark and no image displays
Display shows no content when powered on
Ranked causes
- 1. AC power cord not connected or damaged (common)
- 2. Power supply failure (occasional)
- 3. Cardiograph in Standby or QuickStart mode (common)
- 4. Main Controller Board failure (rare)
- 5. Battery power depleted or not seated (occasional)
Verify AC power cord connection
- Test point
- rear AC power inlet
- Expected
- AC power cord connected to grounded outlet
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No reply received from TraceMasterVue Remote Site server
Error message displayed when attempting ECG transmission; cardiograph fails to receive response from server within specified timeout
Ranked causes
- 1. Wireless connection to access point does not exist (common)
- 2. Network connection to TraceMasterVue Remote Site server not configured properly (common)
- 3. No response from server within timeout period (occasional)
- 4. Remote Site name contains a space character (rare)
Verify wireless adapter LED status
- Test point
- Wireless adapter rear panel LEDs
- Expected
- Green LED blinking slowly or off (indicating associated to access point); not red LED blinking quickly with green LED blinking quickly
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer Error Out of Paper/Paper Jam/Door Open
Error message displays indicating paper out, jam, or door open condition
Ranked causes
- 1. Paper tray not fully closed (common)
- 2. Out of paper or incorrectly loaded (common)
- 3. Defective TOF sensor assembly (rare)
- 4. Defective paper tray assembly (rare)
Verify paper tray is fully closed
- Test point
- Paper tray latch
- Expected
- Tray fully closed and latched
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard typing incorrect characters
Keys produce wrong characters on display; case may be incorrect or characters scrambled.
Ranked causes
- 1. Keyboard has Caps Lock feature enabled (common)
- 2. Keyboard controller subsystem has experienced a failure (occasional)
- 3. Known defect with software version A.0x (rare)
Inspect character case; press Caps Lock key if only case is incorrect
- Test point
- Keyboard display and Caps Lock key
- Expected
- Characters appear in correct case after pressing Caps Lock
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not advance
Paper tray does not advance paper forward during printing operation
Ranked causes
- 1. Application error (occasional)
- 2. Defective TOF sensor assembly (rare)
- 3. Defective printer gearbox assembly (rare)
Press reset button on cardiograph rear
- Test point
- Reset button
- Expected
- Cardiograph restarts; paper advances normally
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not detected message appears
Message displays indicating printer is not detected by cardiograph
Ranked causes
- 1. Application error (occasional)
- 2. Main harness assembly loose or faulty (occasional)
- 3. Printer control board failure (rare)
Press reset button on cardiograph rear
- Test point
- Reset button
- Expected
- Cardiograph restarts and detects printer
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal Acquisition Failure message; Check Patient Input Module Connection patient safety
Message indicates signal acquisition failure; Check Patient Input Module Connection appears, indicating PIM communication issue
Ranked causes
- 1. PIM disconnected or damaged patient data cable (occasional)
- 2. Mechanical failure of PIM (rare)
- 3. Static discharge or unstable AC power (occasional)
Wait for automatic recovery if message not yet displayed
- Test point
- Message status on display
- Expected
- Acquisition Failure message disappears and communication resumes
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Signal acquisition failure or PIM communication error patient safety
Patient data acquisition fails; system displays error requesting check of patient input module connection; communication may be interrupted or cable damaged
Ranked causes
- 1. PIM communications temporarily interrupted or lost, or patient data cable damage (common)
- 2. Static discharge or unstable AC power condition (occasional)
- 3. Mechanical failure of PIM (rare)
Observe message and wait for automatic recovery
- Test point
- Cardiograph display
- Expected
- Check Patient Input Module Connection message does not display (normal recovery)
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unidentified PC card adapter error
Error 'Unidentified PC Card Adapter: Enter the name of the PC Card driver' displayed on system startup.
Ranked causes
- 1. Wireless adapter inserted incorrectly (common)
- 2. Adapter inserted during active operation (occasional)
- 3. Unsupported wireless adapter used (rare)
Close error dialog
- Test point
- Error message
- Expected
- Dialog dismissed
3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batteries do not operate for sufficient time after full recharge
Battery operation time is less than expected after full charge
Ranked causes
- 1. Batteries require calibration (common)
- 2. Batteries require replacement (occasional)
Perform battery calibration
- Test point
- Battery charging system
- Expected
- Battery operates for acceptable time after full recharge
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance Verification Tests patient safety
Complete set of Performance Verification tests including visual inspection and electrical safety checks required after service.
Ranked causes
- 1. Service maintenance completion verification (common)
- 2. Electrical Safety Testing (common)
Place unit in Standby mode and disconnect power
- Test point
- On/Standby button, AC power switch, power cord
- Expected
- Unit in Standby, switch off, cord unplugged
13 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiograph printer making clicking noise every 6 seconds
Printer emits a click sound at regular 6-second intervals
Ranked causes
- 1. Standby failure (occasional)
- 2. Main Controller Board failure (occasional)
Attempt unit restart
- Test point
- Reset button on rear
- Expected
- Clicking stops and unit restarts normally
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cisco wireless LAN adapter not found
Aironet Client Utility displays error 'Cisco Wireless LAN Adapter Not Found!' or 'Client Adapter Not Found'.
Ranked causes
- 1. Wireless adapter not inserted or ejected (common)
- 2. Aironet Client Utility launched prematurely (occasional)
Close error message
- Test point
- Error dialog
- Expected
- Message dismissed
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Copying ECGs from Archive to PC card fails occasionally
ECG copy operation from Archive to PC card intermittently fails; transfers sometimes succeed on retry
Ranked causes
- 1. PC card subsystem time-out failure (occasional)
- 2. Defective PC card (rare)
PC card subsystem response
- Test point
- PC card interface
- Expected
- Transfer operation completes successfully on first or second attempt
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Diskette drive read/write failure
Diskette drive cannot write to or read from inserted diskette
Ranked causes
- 1. Diskette is write-protected (common)
- 2. No diskette in drive (common)
Verify diskette is inserted and not write-protected
- Test point
- Diskette drive
- Expected
- Valid non-write-protected diskette in drive
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard unresponsive
Keyboard does not respond to key presses; no input registered on display.
Ranked causes
- 1. Keyboard controller has experienced a failure (common)
- 2. Keyboard controller subsystem has experienced a failure (occasional)
Press Reset button and launch Service Utility; run Keyboard test
- Test point
- Cardiograph rear reset button; Service Utility keyboard test
- Expected
- Keyboard becomes functional after reset and test completion
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No batteries detected
System reports no batteries detected; batteries completely depleted or not inserted
Ranked causes
- 1. Batteries completely depleted of power (common)
- 2. Batteries removed from cardiograph (occasional)
Open battery door compartment and inspect battery condition
- Test point
- Battery compartment
- Expected
- Batteries present and in good condition
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more leads display as leads off (red dashed line) continuously patient safety
ECG display shows red dashed line on one or more lead channels continuously during operation, indicating loss of signal
Ranked causes
- 1. Problem with electrodes or leads (common)
- 2. Lead wire or patient cable connection failure (occasional)
PIM lead wire connector seating
- Test point
- Lead wire connectors in PIM housing
- Expected
- All lead wires fully seated and secure
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more leads display as leads off periodically patient safety
Lead signal shows red dashed line intermittently; leads appear disconnected or signal is lost during operation
Ranked causes
- 1. Defib or high noise event occurred (common)
- 2. Problem with electrodes or leads (common)
Observe leads off event timing relative to movement and defib
- Test point
- Lead status display and patient movement
- Expected
- Leads off event coincides with electrode movement or defib event
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Orders Error: Command Timed Out
Error message appears when attempting to load or manage orders on cardiograph
Ranked causes
- 1. Invalid Order file loaded (common)
- 2. Pending Orders list too large (occasional)
Verify order files are valid
- Test point
- Removable media (diskette, PC card, USB)
- Expected
- Order files are valid format
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not tear off cleanly or does not advance to perforation
Paper either tears improperly or does not advance to perforated edge for clean separation
Ranked causes
- 1. Incorrect paper size configured in cardiograph (common)
- 2. Incorrect paper tear-off method used by operator (common)
Verify paper size matches cardiograph configuration
- Test point
- Configuration settings and loaded paper
- Expected
- Paper size in configuration matches physical paper loaded
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM not communicating with cardiograph
Patient Interface Module fails to establish data connection after installation or cable replacement
Ranked causes
- 1. PIM data cable connected to wrong port (LAN or modem card dongle instead of RJ-11) (common)
- 2. PIM software not loaded after hardware replacement (occasional)
Verify PIM data cable is connected to correct port
- Test point
- RJ-11 receptacle on right side of cardiograph
- Expected
- PIM data cable firmly seated in RJ-11 receptacle (not in LAN port or modem card dongle)
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen shows faint solid color with no image
Display illuminates with a single solid color but no image or content appears
Ranked causes
- 1. Software installation failure (occasional)
- 2. CompactFlash card failure (rare)
Perform reset and observe boot sequence
- Test point
- Reset button on rear
- Expected
- Display turns black, clicking sound heard, display returns to solid color
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test: barcode reader accuracy
Barcode reader scans incorrectly or Patient ID data not entered properly
Ranked causes
- 1. Barcode reader alignment or optics need calibration (common)
- 2. Barcode reader hardware failure (rare)
Barcode reader scan test
- Test point
- Barcode in Barcode Test window
- Expected
- Barcode data displays correctly next to Barcode Field
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test: magnetic card reader accuracy
Magnetic card reader not reading data correctly or Patient ID not entered properly
Ranked causes
- 1. Magnetic card reader requires test or data not reading (common)
- 2. Magnetic card reader hardware failure (rare)
Magnetic card read test
- Test point
- Magnetic card in slot (Figure 1-4, page 1-9)
- Expected
- Magnetic card data displays correctly next to Magcard Track 1
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Test: Patient Interface Module communication
Cardiograph displays PIM error messages or PIM patient data cable connection not recognized
Ranked causes
- 1. PIM data cable not securely attached (common)
- 2. PIM hardware failure or data cable defect (occasional)
PIM communication test
- Test point
- PIM via RJ-11 connection
- Expected
- PIM Test Passed message
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen unresponsive or taps register in wrong position patient safety
Touch screen does not respond to taps or input registers at different location than actual contact point
Ranked causes
- 1. Touch screen requires calibration (common)
- 2. Touch screen hardware failure (rare)
Remove obstructions from touch screen
- Test point
- touch screen surface
- Expected
- Nothing pressing or leaning on screen
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unformatted disk message on eject attempt
Message 'An unformatted disk has been inserted' appears when attempting to eject diskette from drive
Ranked causes
- 1. Eject button only partially pressed (common)
- 2. Diskette drive damaged (rare)
Press eject button firmly
- Test point
- Diskette eject button
- Expected
- Diskette ejects with no message
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveforms not refreshing; cardiograph not responsive
Real-time ECG waveforms on screen do not update and cardiograph is unresponsive to operator input
Ranked causes
- 1. Application error has occurred (occasional)
- 2. Main controller board failure (rare)
Wait for automatic reset or manual reset
- Test point
- Reset button on cardiograph rear
- Expected
- Cardiograph restarts and boots without error
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC power indicator light does not illuminate when plugged into AC power high voltage
AC power indicator light on front of cardiograph is dark when AC powered
Ranked causes
- 1. AC fuse is non-functional (common)
Visual inspection of AC indicator light
- Test point
- Front panel AC indicator
- Expected
- AC indicator light illuminates when cardiograph plugged into AC power
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Access Service Utility for diagnostics
Launch Service Utility to check unit status, software revisions, voltages, battery, diagnostics, and logs
Ranked causes
- 1. Routine diagnostic access required (common)
Put cardiograph in Standby mode
- Test point
- On/Standby button
- Expected
- Cardiograph enters Standby
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode data scans in incorrect case regardless of Caps Lock setting
Barcode data scans in lowercase when Caps Lock enabled or uppercase when disabled
Ranked causes
- 1. Barcode reader not mirroring Caps Lock setting (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader calibration verification
Barcode reader initial calibration after first setup or maintenance
Ranked causes
- 1. Barcode reader requires initial calibration procedure (common)
Barcode reader angle and button press
- Test point
- Barcode reader top button
- Expected
- Barcode reader ready for scan
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode reader does not correctly read scanned barcode
Scanned barcode data not correctly interpreted by barcode reader
Ranked causes
- 1. Barcode reader requires reconfiguration (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batch ECG copy to diskette fails with insufficient space error
Transfer of 5 to 6 ECGs from Archive to diskette fails despite adequate remaining diskette space
Ranked causes
- 1. Application overestimates diskette space requirement (common)
Verify available diskette space
- Test point
- Free space on diskette
- Expected
- At least 4-6 ECG capacity available
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery level indicator inaccurate
Battery level indicator on status bar does not accurately reflect power remaining
Ranked causes
- 1. Batteries require calibration (common)
Double-tap battery icon on Status Bar
- Test point
- Status Bar battery icon
- Expected
- Calibration recommendation displayed
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank pages printed after ECG report
Extra blank pages are printed following the ECG report output
Ranked causes
- 1. Cardiograph configured for wrong paper size (common)
Verify paper size in configuration
- Test point
- Configuration settings menu
- Expected
- Paper size setting matches loaded paper size
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: barcode reader
Barcode reader not scanning accurately or entering Patient ID data incorrectly
Ranked causes
- 1. Barcode reader requires calibration (common)
Barcode reader scans barcode correctly
- Test point
- Barcode in Figure 3-10
- Expected
- Three beeps heard and barcode data displays correctly
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery full cycle
Battery requires up to 12 hours to complete full calibration cycle consisting of two full charges and discharges. Display dims to reduce charge time. Process must restart if disrupted but remembers progress per battery.
Ranked causes
- 1. Battery calibration cycle required (common)
Allow first battery complete charge-discharge cycle
- Test point
- Battery status indicator
- Expected
- Full charge to full discharge
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery level indicator
Battery Level Indicator on Status Bar becomes inaccurate after multiple charge/discharge cycles; wrench icon displays to indicate calibration needed
Ranked causes
- 1. Battery accuracy degradation from charge/discharge cycles (common)
Verify wrench icon displays on Status Bar
- Test point
- Status Bar display
- Expected
- Wrench icon present
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance Verification after External Repair
Verification testing required after repair or replacement of external components (PIM, patient leads, PIM data cable, labels, batteries, paper tray, AC power cord, main fuse, PCMCIA card, PCMCIA modem, barcode reader, cart, external cables) without opening the cardiograph case
Ranked causes
- 1. External component repair or replacement performed (common)
Perform required test blocks from Table 4-1
- Test point
- Cardiograph system
- Expected
- All tests pass per Table 4-1 requirements
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Performance Verification after Internal Repair
Verification testing required after opening the cardiograph case regardless of the type of repair performed
Ranked causes
- 1. Internal repair performed (case opened) (common)
Perform all required test blocks from Table 4-1
- Test point
- Cardiograph system
- Expected
- All tests pass per Table 4-1 requirements for internal repairs
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power On Test
Verify cardiograph boots and displays software identification and real-time ECG screen without error codes
Ranked causes
- 1. Cardiograph does not reboot after Reset button press (unknown)
Press Reset button with paper clip
- Test point
- Reset button next to USB port on back
- Expected
- Cardiograph reboots within 40 seconds
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot access Windows CE desktop
Unable to reach Windows CE desktop for advanced troubleshooting and diagnostics
Ranked causes
- 1. Incorrect key combination during splash screen (common)
Cardiograph in Standby mode
- Test point
- Unit display
- Expected
- Standby mode active
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiograph not responding to reset or restarting
Touch screen does not shut off or system not restarting after reset button press; device unresponsive
Ranked causes
- 1. Reset button press ineffective or power not cycling (occasional)
Observe touch screen response after reset button press
- Test point
- Touch screen display
- Expected
- Touch screen shuts off and cardiograph restarts
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiograph slow or unresponsive when touching buttons
System responsiveness slow on any screen or Timed ECG acquisition fails; buttons not responding quickly to touches
Ranked causes
- 1. Main Archive full or close to capacity (common)
Verify Main Archive capacity status
- Test point
- Main Archive
- Expected
- Maintain at lower capacity to maximize performance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiograph unresponsive after Floppy Unmount Failed
Unit becomes unresponsive and frozen after touching Cancel button in response to 'Floppy Unmount Failed' message
Ranked causes
- 1. Application error when diskette removed prior to unmounting (occasional)
Verify diskette is in drive before unmount
- Test point
- Diskette presence in drive
- Expected
- Diskette present and seated
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Printer message appears with no printer problem
Message displays but printer appears to be functioning normally
Ranked causes
- 1. Printer controller false error, common with A4 paper (occasional)
Open and firmly close printer paper tray
- Test point
- Paper tray assembly
- Expected
- Tray latches securely
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Printer message persists after drawer closure
Message Printer Drawer is Open displays, then after closing drawer the Check Printer message continuously appears
Ranked causes
- 1. Paper drawer opened partially or not fully closed (common)
Pull paper tray out completely and close firmly
- Test point
- Paper tray drawer
- Expected
- Tray fully open then fully seated and latched
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display shows black or solid color screen
LCD display shows only black or solid color with no text or graphics; complete display failure
Ranked causes
- 1. Internal CF card or main controller board failure (rare)
Verify cardiograph power status and attempt reset
- Test point
- Power indicator and Reset button
- Expected
- Power on; reset button responsive
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect lead labeling standard appears on printed ECG
Printed ECG shows AHA or IEC lead standard that does not match cardiograph default setting
Ranked causes
- 1. Original ECG setting differs from cardiograph default lead standard (common)
Compare ECG lead standard with cardiograph default standard
- Test point
- cardiograph settings and retrieved ECG metadata
- Expected
- Both using same lead standard (AHA or IEC)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid Database Entry Detected
Error message when deleting multiple orders is canceled before completion
Ranked causes
- 1. Order deletion canceled before completion (common)
Retry deletion of orders
- Test point
- Orders delete function
- Expected
- Deletion completes successfully
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid WEP key length for 128-bit encryption
Error 'WEP Key X Must Be 26 Hex Digits!' displayed when attempting to configure 128-bit wireless security.
Ranked causes
- 1. Incorrect number of WEP key characters for 128-bit (common)
Verify WEP key character count
- Test point
- WEP key entry field
- Expected
- 26 hexadecimal characters
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid WEP key length for 40-bit encryption
Error 'WEP Key X Must Be 10 Hex Digits!' displayed when attempting to configure wireless security.
Ranked causes
- 1. Incorrect number of WEP key characters entered (common)
Verify WEP key character count
- Test point
- WEP key entry field
- Expected
- 10 hexadecimal characters
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Modem not functioning after card removal and reinsertion
Modem card fails to function after being removed from PC card slot and reinserted
Ranked causes
- 1. Main controller error or state mismatch from hot swap (common)
Modem functionality after reset
- Test point
- Modem card operation
- Expected
- Modem functions normally after reset
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
New PIM detected during active use patient safety
System detects PIM disconnection and reconnection or different PIM connected while cardiograph is in use; error message displayed
Ranked causes
- 1. PIM disconnected and reconnected or different PIM connected during active cardiograph use (common)
Touch Continue button on error message
- Test point
- Error message window
- Expected
- Error message disappears and cardiograph restarts
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No WEP key entered
Error 'You must enter a WEP Key!' displayed when attempting to enable security without providing key.
Ranked causes
- 1. WEP key field left empty (common)
Verify WEP key is entered
- Test point
- WEP keys window
- Expected
- Key field contains characters
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more leads display as leads off (red dotted line) when printing patient safety
ECG display shows red dotted line on one or more lead channels only during printing operations, indicating intermittent signal loss
Ranked causes
- 1. Faulty ground connection to printhead static brush (common)
Leads off condition during continuous rhythm printing
- Test point
- ECG display during print operation
- Expected
- Red dotted line disappears during continuous rhythm printing if brush contact is restored
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper tray opened during printing
ECG report appears corrupted or printed on multiple pages when paper tray is opened while printing
Ranked causes
- 1. Printer controller error from tray interruption (occasional)
Ensure printer door is firmly closed
- Test point
- Paper tray door
- Expected
- Door completely seated and latched
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Patient ID dialog missing required fields error after barcode scan
- Patient Interface Module dirty
- Patient lead damage
- PIM connection verification
- PIM cover difficult to open or will not seal
- PIM not found
- Printed ECGs appear different from other ECGs
- Printer Door is Open message appears and cardiograph inoperative
- Printer is OK message appears
- Printer test page did not print correctly
- Repair subassembly replacement protocol
- Screen Test: color display quality verification
- Selecting items in specific screen area is difficult
- Service Utility does not display current data
- Signal acquisition error or loss of PIM signal during printing
- Software version earlier than B.01
- Test: printer output quality
- Touch screen calibration after usage environment change
- Touch screen calibration needed after environment or user change
- Touch screen requires many attempts to select items
- Touch screen selection difficult in specific screen area
- Unable to copy ECG from Archive to PC card with less than 100 ECGs saved
- Unexpected reply received from Remote Site server
- Wireless adapter not associated to access point
- Wireless LAN unable to associate to access point after Ethernet cable disconnect
- Worn printer drive gear
- Worn printer roller
- Write operations occasionally fail to PC card
28 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
350 Series Radio is Not Associated | Wireless adapter not connected to access point | Loss of association with wireless access point | error |
Archive Error: Command Timed Out | ECG transmission operation did not complete within timeout period | TraceMaster ECG Management system busy or network/server capacity exceeded | warning |
Archive Error: Requested Function Failed | Archive function request failed to complete | Remote Site name contains spaces, modem disconnected, phone line busy, incompatible ECG format, transfer limit exceeded, or fax timeout | error |
Cannot access selected archive | LAN connection to TraceMasterVue Remote Site not accessible | Network not properly configured or network permission denied | error |
Cannot Access Selected Archive. Verify floppy disk is not wr | Diskette drive cannot access selected archive | No diskette inserted, diskette not fully inserted, or diskette write-protected | error |
Cannot read ECG. Only ECGs in XML schema format 1.04 can be | Selected ECG file uses XML schema version not compatible with installed software | Attempt to import ECG in XML version 1.03 to cardiograph with software version C.01.02 or higher | info |
Cisco Wireless LAN Adapter Not Found | Aironet wireless adapter not detected by system | ACU launched without adapter inserted or adapter ejected during operation | error |
Client Adapter Not Found | Wireless client adapter not recognized | ACU launched without adapter inserted or adapter ejected | error |
Invalid Database Entry Detected | Database error during order operations | Deletion of multiple orders canceled before command completion | warning |
No reply from Remote Site Server | No response received from remote server during transfer operation | Network resources burdened, LAN cable disconnected, network/modem connection interrupted, damaged cabling, or conflicting IP/DHCP settings | error |
No Reply Received From Remote Site Server | Cardiograph cannot establish or maintain communication with TraceMasterVue Remote Site server | Wireless connection to access point lost or network connection to server does not exist or not properly configured | error |
Orders Error: Command Timed Out | Orders failed to load within timeout period | Invalid order file loaded or Pending Orders list capacity exceeded | error |
Unexpected Reply Received from Remote Site | Connection exists to TraceMasterVue Remote Site server but server not properly acknowledging packet data receipt | Connection established to server but server response is not proper | warning |
Unidentified PC Card Adapter | System cannot recognize wireless adapter driver | Adapter inserted incorrectly, inserted during active operation, or unsupported model | error |
WEP Key X Must Be 10 Hex Digits | 40-bit WEP key character count or format invalid | Incorrect number of characters entered or non-hex characters in 40-bit key field | warning |
WEP Key X Must Be 26 Hex Digits | 128-bit WEP key character count or format invalid | Incorrect number of characters entered or non-hex characters in 128-bit key field | warning |
You must enter a WEP Key | WEP key field empty when security is enabled | Attempt to enable WEP without entering a key value | warning |