Philips PageWriter TC70: troubleshooting and error codes

The ServiceForge knowledge base holds 115 documented symptoms for the Philips PageWriter TC70, with 142 ranked causes, 107 diagnostic checks with test points and expected values, and 38 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Screen is dark and shows no image

Display is blank or extremely dim with no visible content

Ranked causes

  1. 1. Cardiograph is in standby mode (common)
  2. 2. Cardiograph battery is extremely low (occasional)
  3. 3. AC Power not applied or AC Power indicator LED is dark (occasional)
  4. 4. Battery insertion or connection failure (occasional)
  5. 5. Display assembly failure (rare)
  6. 6. Main board failure (rare)
First check

Press On/Standby button

Test point
Power button
Expected
Cardiograph wakes up and display activates

3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer error: Out of paper, paper jam, or door open

Messages indicate paper supply exhausted, paper jammed, or door sensor triggered

Ranked causes

  1. 1. Paper drawer not fully closed (common)
  2. 2. Paper out of supply or loaded incorrectly (common)
  3. 3. Incompatible paper loaded (occasional)
  4. 4. Defective paper tray assembly (occasional)
  5. 5. Top-of-form sensor or door closed sensor malfunction (rare)
First check

Verify paper drawer fully closed

Test point
Paper drawer latch
Expected
Drawer fully closed and latched

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

One or more leads show leads off periodically patient safety

One or more ECG leads display red dashed line intermittently, signal interruptions occur

Ranked causes

  1. 1. Improper or incomplete patient and skin preparation (common)
  2. 2. Improperly attached electrodes or lead wires; expired gel electrodes (common)
  3. 3. Defib or high noise event (occasional)
  4. 4. PIM reset normal behavior (common)
First check

Verify patient skin preparation completed

Test point
Patient skin condition and electrode application area
Expected
Skin clean, dry, properly abraded; electrodes applied per protocol

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper does not advance

Paper feed mechanism fails to move paper through printer

Ranked causes

  1. 1. Application error (occasional)
  2. 2. Top-of-form sensor malfunction (occasional)
  3. 3. Defective print head assembly (occasional)
First check

Restart cardiograph

Test point
System startup
Expected
Paper advances normally

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM not detected message or PIM with red X icon appears patient safety

Display shows 'Patient Interface Module (PIM) not detected' message or PIM icon with red X on screen

Ranked causes

  1. 1. PIM not connected (common)
  2. 2. Faulty patient data cable (common)
  3. 3. Faulty PIM (occasional)
First check

Verify PIM patient data cable connection

Test point
PIM connector on rear of cardiograph
Expected
Cable firmly seated with no gaps

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

All leads show leads off continuously patient safety

Red dashed line displayed on all ECG leads continuously without clearing

Ranked causes

  1. 1. PIM communications lost due to USB patient data cable connection failure (common)
  2. 2. Defective leads connected to PIM (occasional)
First check

Check for PIM icon with red X on screen

Test point
Top right of display screen
Expected
PIM icon visible or not visible

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: PIM Test patient safety

Verification that Patient Interface Module (PIM) is communicating correctly with cardiograph and can acquire patient data when PIM error messages occur or data acquisition fails.

Ranked causes

  1. 1. PIM communication error or PIM patient data cable disconnection (common)
  2. 2. PIM hardware failure or defective PIM patient data cable (occasional)
First check

Verify PIM patient data cable connection

Test point
PIM connector on cardiograph rear panel, cable physical condition
Expected
Cable fully seated in connector with no gaps; connector barrel visible and accessible for grounding

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard is unresponsive

Keyboard input not registering or input focus lost

Ranked causes

  1. 1. Input focus lost (common)
  2. 2. Keyboard controller subsystem has failed (rare)
First check

Touch cardiograph screen

Test point
Display screen
Expected
Input focus restored and keyboard responds

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No matching ECG found on server

ECG with specified patient ID not found on selected TraceMaster server

Ranked causes

  1. 1. Incorrect patient ID specified (common)
  2. 2. ECG does not exist on server (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No reply received from server

Server did not respond during remote transfer; connection lost or server down

Ranked causes

  1. 1. LAN connection disconnected during transfer (occasional)
  2. 2. TraceMaster server is down (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper does not tear off cleanly or stops at wrong spot

Paper tearoff mechanism fails or paper stops at incorrect position

Ranked causes

  1. 1. Wrong paper size setting (common)
  2. 2. Wrong tearoff method used (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient Interface Module (PIM) communication failure

Cardiograph unable to acquire data from PIM or displays PIM error messages during operation

Ranked causes

  1. 1. PIM patient data cable not securely attached (common)
  2. 2. PIM hardware fault or cable defect (occasional)
First check

Verify PIM patient data cable is securely connected

Test point
PIM connector on rear of cardiograph
Expected
Cable is firmly seated and fully connected to PIM connector

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer not detected message

System fails to recognize or communicate with printer assembly

Ranked causes

  1. 1. Application error (occasional)
  2. 2. Defective print head assembly (occasional)
First check

Restart cardiograph

Test point
System startup
Expected
Printer detected and operational

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer unable to accept print job - busy state

Printer reports busy status or USB error after sending print job; printing disabled with low battery

Ranked causes

  1. 1. Printer cannot accept new print job (common)
  2. 2. Battery too low or missing for printing (common)
First check

Verify battery status before printing

Test point
Battery indicator on cardiograph
Expected
Two batteries installed and charged; AC power connected or batteries above minimum level

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch screen difficult to select or unresponsive

Items on touch screen require many attempts to select, specific screen areas are difficult to select, or part of screen appears unresponsive to touch input

Ranked causes

  1. 1. Touch screen requires recalibration due to usage environment or user change (common)
  2. 2. Touch screen calibration not improved after Force Calibration (occasional)
First check

Perform Force Calibration procedure

Test point
Touch screen with cross hair
Expected
Multiple cross hairs touched in sequence until procedure completes with acceptance message

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

USB memory stick not accessible message

System cannot detect or communicate with USB memory stick

Ranked causes

  1. 1. USB stick not fully inserted (common)
  2. 2. Unsupported software on USB stick (occasional)
First check

Verify USB stick insertion

Test point
USB connector on right side of cardiograph
Expected
USB stick fully inserted and recognized

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio system not emitting sound

Cardiograph produces no audible output during audio test or alarms are inaudible

Ranked causes

  1. 1. Audio system fault or speaker malfunction (rare)
First check

Run audio test and listen for chime tones

Test point
Audio Test button; speaker output
Expected
Audible chime tones are emitted from speaker during test

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Blank pages printed after ECG report

Extra blank pages output following completed ECG printout

Ranked causes

  1. 1. Wrong paper size setting (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery level indicator accuracy

Battery level indicator on Status Bar and Battery Status window may display inaccurate remaining runtime and power. Calibration enhances accuracy of battery indicators for reliable daily work.

Ranked causes

  1. 1. Battery calibration required due to clinical use factors (occasional)
First check

Attach AC power cord and verify connection

Test point
AC power indicator light next to power button
Expected
Lit

7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery status

Inaccurate battery status information displayed on cardiograph

Ranked causes

  1. 1. Battery calibration required (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Cardiograph sensitivity

Overall sensitivity test to verify cardiograph signal acquisition performance

Ranked causes

  1. 1. Yearly sensitivity verification (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Keyboard Test

Verification that all keyboard and hardware keys are properly recognized when pressed and functional for cardiograph operation.

Ranked causes

  1. 1. Need to verify keyboard functionality after maintenance or to diagnose key responsiveness issues (common)
First check

Execute Keyboard/Hardware Key Test

Test point
Service Utilities Keyboard Test screen, physical keys on cardiograph
Expected
Each key press is recognized and registered by test software; no missed or delayed key detections

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Lead Wire Tests patient safety

Verification of patient lead wire integrity through open and short circuit tests to detect breaks, continuity issues, or shorted connections.

Ranked causes

  1. 1. Need to verify lead wire continuity and integrity before patient use (common)
First check

Verify all lead wires continuity with open test

Test point
Each patient lead wire conductor
Expected
All leads show continuity; no open circuits detected; expected resistance within specification

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Power On Self-Test

Automatic self-tests performed during cardiograph boot sequence to verify software integrity, flash memory, PIM status, and USB connectivity.

Ranked causes

  1. 1. Need to verify system integrity after power cycle or maintenance (common)
First check

Observe software identification screen appearance

Test point
Cardiograph display
Expected
Software identification screen appears within approximately 20 seconds of power-on; audible beep heard

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Printer Test

Verification that cardiograph printer correctly prints test page and produces acceptable print quality without errors.

Ranked causes

  1. 1. Need to verify printer functionality or diagnose print quality issues (common)
First check

Printer test execution and paper output

Test point
Printer output
Expected
Test page prints successfully without paper jams, feed errors, or mechanical issues

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Simulated ECG Recording

Verification of ECG recording functionality using a 12-lead ECG simulator to confirm patient lead integrity, paper speed accuracy, and gain setting accuracy.

Ranked causes

  1. 1. Need to verify ECG signal acquisition and recording accuracy (common)
First check

Verify all lead wires connected to simulator

Test point
Display screen leads map
Expected
No red x marks next to any electrodes; all leads showing activity, flat lines initially

5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Touch screen

Touch screen requires frequent attempts to select items or selecting items in specific areas is difficult; may occur after relocation or with different user heights

Ranked causes

  1. 1. Touch screen drift or misalignment due to usage conditions (common)
First check

Initiate touch screen calibration

Test point
Keyboard Ctrl + Alt + C keys
Expected
White screen appears with cross hair at center

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Visual Inspection

External visual inspection of cardiograph for damage, wear, corrosion, contamination, and hardware integrity before performing functional tests.

Ranked causes

  1. 1. Equipment may have accumulated damage or wear during service (common)
First check

Inspect power cord condition

Test point
Power cord exterior
Expected
No visible wear, cracks, or damage to insulation; intact and secure connectors

10 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot access internal Compact Flash memory message

System unable to read or write to internal CF card

Ranked causes

  1. 1. Compact Flash card corruption (occasional)
First check

Run Internal Compact Flash test

Test point
Service Utilities
Expected
Test passes

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cannot store additional ECGs message

System prevents storing more ECGs due to storage capacity full

Ranked causes

  1. 1. CF card file system limitation reached (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cardiograph must restart to load custom settings

System requires restart after loading new configuration file from USB

Ranked causes

  1. 1. Configuration file loaded; restart needed to apply settings (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cardiograph unresponsive after rapid paper drawer cycling

System becomes unresponsive following repeated opening and closing of paper drawer

Ranked causes

  1. 1. Printer controller error from excessive drawer cycling (rare)
First check

Monitor system recovery

Test point
Cardiograph startup
Expected
System restarts and boots correctly; reports print correctly

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Check Printer message with no apparent problem

Check Printer message appears without observable printer malfunction

Ranked causes

  1. 1. Printer controller error (occasional)
First check

Reset paper drawer

Test point
Printer paper drawer
Expected
Message does not reoccur after drawer cycle

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Cleaning: Touch screen

Touch screen requires occasional cleaning to maintain visibility and responsiveness

Ranked causes

  1. 1. Dust, dirt or residue accumulation on touch screen surface (common)
First check

Prepare cleaning cloth

Test point
Soft cloth
Expected
Cloth dampened with water or isopropyl alcohol, excess moisture wrung out

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Command canceled

Operation was canceled by user or invalid information was specified

Ranked causes

  1. 1. User canceled action or provided invalid information (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Command cannot be completed

TraceMaster system is busy; operation request cannot be processed at this time

Ranked causes

  1. 1. TraceMasterVue system busy or resource contention (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Command timed-out message during archive operation

Error message 'Command timed-out' appears when attempting archive or transfer operation

Ranked causes

  1. 1. TraceMaster ECG Management system is busy (occasional)
First check

Retry archive or transfer operation

Test point
Network and TraceMaster server availability
Expected
Operation completes within timeout period

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Configuration file load error - invalid XML

Configuration file cannot be loaded due to invalid XML version, platform, or file corruption

Ranked causes

  1. 1. Corrupted or invalid configuration file (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECGs with incomplete patient information cannot be transferred message

Error message 'ECGs with incomplete required patient information fields cannot be transferred' displayed during archive

Ranked causes

  1. 1. Required patient information not entered for ECG record (common)
First check

Verify all required patient information fields completed

Test point
ECG record patient data fields
Expected
All required fields populated with patient information

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error accessing file from USB memory stick

Cannot copy configuration or log file from USB memory stick; file access error

Ranked causes

  1. 1. USB file access failure or file not present (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error opening corrupted file

Cannot open configuration or log file due to file corruption

Ranked causes

  1. 1. Configuration or log file is corrupted (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error retrieving ECG from archive - compact flash error

Cannot delete or retrieve selected ECGs from archive due to internal card error

Ranked causes

  1. 1. Internal compact flash card read/write error (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error: ECG from legacy TraceMaster system cannot be printed

ECG originated from older TraceMaster system with incompatible format

Ranked causes

  1. 1. ECG XML version incompatible from legacy system (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error: Lead system not supported on cardiograph

ECG XML version incompatible with cardiograph lead system configuration

Ranked causes

  1. 1. ECG XML version incompatible (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Failed to implement custom settings

Cardiograph automatically restarts and restores current settings, unable to load custom configuration

Ranked causes

  1. 1. Failure to load a configuration file (unknown)
First check

Restore factory default settings

Test point
Settings menu
Expected
Settings restored to factory defaults

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Fax transmission did not complete

Fax transmission failed to complete successfully

Ranked causes

  1. 1. Fax transmission failure (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

File already exists

Attempted to save log file to a name that already exists in the system

Ranked causes

  1. 1. Log file name already exists on system (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

File system is destroyed

Dialog box showing software version information with black background, indicating file system corruption

Ranked causes

  1. 1. File system corruption (occasional)
First check

Wait 2 minutes

Test point
System
Expected
System stabilizes

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Flash memory failure

Onboard Flash memory test fails or memory appears corrupted

Ranked causes

  1. 1. Onboard Flash memory fault (rare)
First check

Run onboard Flash memory test

Test point
Flash Memory Test button in Diagnostic Test screen
Expected
Test passes and confirms onboard Flash memory is working properly

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

FPGA communication or connection failure

Field Programmable Gate Array (FPGA) is not connected or incorrectly configured; internal error

Ranked causes

  1. 1. FPGA internal error or disconnection (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

ID or ECG button on keyboard is unresponsive

Specific button on keyboard not responding to input

Ranked causes

  1. 1. Input focus loss affecting button (common)
First check

Touch cardiograph screen to restore input focus

Test point
Display screen
Expected
Button becomes responsive after focus restored

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Insufficient internal storage for new orders

System reports not enough storage available to save downloaded or new orders

Ranked causes

  1. 1. Application storage full (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal Compact Flash failure

Internal Compact Flash test fails or card appears non-functional

Ranked causes

  1. 1. Internal Compact Flash card fault (rare)
First check

Run Internal Compact Flash test

Test point
Internal Compact Flash button in Diagnostic Test screen
Expected
Test passes and confirms Internal Compact Flash is working properly

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal CompactFlash card cannot be accessed message

System unable to communicate with or access internal CF storage

Ranked causes

  1. 1. Compact Flash card corruption (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal error with orders list

Orders list does not refresh; internal application error occurs

Ranked causes

  1. 1. Orders list refresh failure (rare)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Internal function error message

Error message 'Internal function error' displayed during operation

Ranked causes

  1. 1. Invalid file or file resides on unsupported media (occasional)
First check

Restart cardiograph

Test point
System power and boot sequence
Expected
System restarts without error messages

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid default gateway message

Cardiograph displays invalid default gateway error during wireless configuration

Ranked causes

  1. 1. Incorrect default gateway in TraceMasterVue configuration (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid default TraceMaster connection for OrderVue inbox

Incorrect TraceMasterVue server specified in configuration for selected OrderVue inbox

Ranked causes

  1. 1. TraceMasterVue server mismatch in OrderVue configuration (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid directory message

Cardiograph displays invalid directory error during wireless configuration

Ranked causes

  1. 1. Incorrect directory in TraceMasterVue configuration (common)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Invalid filename in save dialog

Invalid or missing file name in save logs dialog box

Ranked causes

  1. 1. File name field left blank in save logs dialog (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Invalid IP address message
  • Invalid path and file name specified
  • Invalid Primary DNS message
  • Invalid Primary WINS message
  • Invalid Report Handle message during preview, printing, or saving ECGs
  • Invalid subnet mask message
  • Invalid transfer destination selected
  • Invalid XML schema version message
  • Keyboard is typing incorrect characters
  • Keyboard or hardware key not recognized
  • LAN/WLAN connection not available
  • Maximum number of ECGs stored message
  • Missing fax recipient information
  • Missing network settings information
  • More than 200 ECGs on USB stick; only first 200 display
  • Network connection to server was reset
  • Network connectivity problem
  • No default TraceMaster server selected
  • No fax recipient available to edit
  • No filename specified
  • No order selected for ECG operation
  • No order source selected
  • No Report message during save or archive operation
  • No TraceMaster connections configured
  • No TraceMaster server to delete
  • Occasionally copying ECGs from archive to USB stick fails
  • Order application error - function failed
  • Order search range not specified
  • Order XML error
  • Orders limit reached (200 maximum)
  • Orders previously archived with ECGs cannot be downloaded
  • Patient ID field must be enabled for TraceMasterVue transfer
  • Print error: Unknown error
  • Printed report looks too small or compressed
  • Printer door is open message
  • Printer temperature too high
  • Printer test failed or poor print quality
  • Printer underflow message
  • Printer voltage high message
  • Printer voltage low message
  • Remote query could not be completed
  • Remote query get ECG error
  • Remote query get list error
  • Remote transfer error
  • Remote transfer schema error
  • Required patient information missing from current ECG
  • Selected ECG exceeds available space on TraceMaster message
  • Selection exceeds maximum number of outboxes
  • Touch screen calibration required
  • Unexpected reply received from server
  • Unknown error accessing storage capacity of archive
  • USB memory stick not detected
  • Waveform display no longer updating; system unresponsive
  • Worklist full; orders cannot be downloaded
  • XML version incompatible with TraceMaster

55 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
After pressing Archive, takes >40 secondsTakes longer than 40 seconds before Archive list is displayed with >130 ECGs storedAccessing Archive screen when main archive contains near-maximum number of stored ECGs (>130)warning
Archive function could not be completedArchive function could not be completed; please restart cardiograph and try againArchive process error or failureerror
Archive process was terminated due to an errorArchive process was terminated due to an error messageArchive process error during operationerror
Bad reply received from TraceMasterBad reply received from TraceMaster Remote site messageInvalid or malformed response received from TraceMaster server during archive operationerror
Cannot access internal Compact Flash memorySystem unable to read or write internal CF cardCF card corrupted, electrical connection failed, or hardware malfunctionerror
Cannot store additional ECGsCompactFlash card storage space is full or file system at capacityAvailable storage space on CF card exhaustedwarning
Check PrinterPrinter controller detected internal error conditionController detects communication error, timeout, or sensor malfunctionwarning
Command timed-outCommand timed-out message during archive or transfer operationTraceMaster system busy or network timeout during transfer operationwarning
ECGs incomplete required patient informationECGs with incomplete required patient information fields cannot be transferredRequired patient data fields missing when attempting archive or transfer to TraceMastererror
Error. Lead system not supportedECG XML file contains lead configuration not supported by this cardiograph modelECG transferred from TraceMasterVue with unsupported lead system or XML versionerror
Error. Legacy TraceMaster ECG cannot be printedECG originated from older TraceMaster system with incompatible XML formatECG file transferred from legacy TraceMaster systemerror
Internal CompactFlash card cannot be accessedSystem unable to communicate with or access internal CompactFlash storageCF card hardware failure, connection error, or file system corruptioncritical
Internal function errorInternal function error message displayedInvalid file or file resides on unsupported mediaerror
Invalid databaseArchive database may be corrupted messageArchive database access failure or corruption detected during archive operationerror
Invalid Report HandleReport handle is invalid when attempting to preview, print, or save ECGsInvalid file or corrupted report object mapping during ECG preview, print, or save operationerror
Maximum number of ECGs storedCompactFlash card archive has reached maximum ECG file count limitNumber of stored ECG files equals CF card file system limitinfo
More than 200 ECGs on USB stickUSB memory stick contains more than 200 ECG files; only first 200 will displayECG file count on USB stick exceeds 200 file display limitinfo
No reply from TraceMasterNo reply received from TraceMaster Remote site messageNo response from TraceMaster server during query or transfer operationerror
No ReportNo report message appears; software cannot map input report handle to internal report objectInvalid handle during save or archive operationerror
Paper does not advancePaper feed mechanism fails to move paper through printer during operationPaper motor not responding, sensor failure, or mechanical blockageerror
Patient ID field must be enabledAttempted ECG transfer to TraceMasterVue without Patient ID field enabled or populatedTransfer initiated with Patient ID field disabled or blankwarning
Printer Door is OpenPrinter door position sensor indicates door not fully closedDoor latch not engaged or sensor malfunctionwarning
Printer Error: Door OpenPrinter door position sensor indicates door not fully closedDoor latch not engaged or sensor malfunction during printingwarning
Printer Error: Out of PaperPaper supply is exhausted or not detected in trayTop-of-form sensor does not detect paper presence or paper runs out during printingerror
Printer Error: Paper JamPaper detected as stuck or jammed in paper pathMechanical jam detected or inconsistent paper feed during printingerror
Printer not detectedCardiograph software cannot communicate with or initialize printerPrinter interface initialization failure or print head not respondingerror
Print Error: Unknown ErrorUnidentified error condition detected during print operationApplication software encountered unexpected condition or timeouterror
Printer temperature is too highPrint head thermal sensor detected overtemperature conditionPrint head temperature exceeds safe operating limitwarning
Printer underflowCommunication dropout between cardiograph and printer during rhythm printingPIM interface loses signal or fails handshake during print operationerror
Printer voltage highMain board printer supply voltage exceeds maximum specificationPower supply output rises above acceptable upper thresholderror
Printer voltage lowMain board printer supply voltage falls below minimum specificationPower supply output drops below acceptable lower thresholderror
Requested Function could not be completedRequested function could not be completed message during transfer or query operationNetwork or modem phone line interrupted during operation; USB memory stick not fully insertederror
Required patient information is missingCurrent ECG lacks mandatory patient data needed for TraceMasterVue transferTransfer to TraceMasterVue attempted with incomplete patient informationwarning
Selected ECG exceeds available spaceSelected ECG(s) exceeds available space on TraceMaster Remote siteTraceMaster system disk full when attempting to transfer ECG dataerror
Transfer to USB fails with <200 ECGs on mediaTransfer of ECGs to USB memory stick fails with less than 200 ECGs on removable media deviceECG transfer to USB memory stick fails despite available space reportserror
Unexpected reply from TraceMasterUnexpected reply received from TraceMaster Remote Site messageUnexpected or invalid response received from TraceMaster server during archive operationerror
Unknown error accessing storage capacitySystem failed to determine available storage capacity of archive mediaStorage media access error or communication failure when querying capacityerror
USB memory stick not accessibleCardiograph cannot detect or access USB memory stickUSB stick not fully inserted or contains incompatible softwarewarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.