Philips PageWriter TC70: troubleshooting and error codes
The ServiceForge knowledge base holds 115 documented symptoms for the Philips PageWriter TC70, with 142 ranked causes, 107 diagnostic checks with test points and expected values, and 38 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Screen is dark and shows no image
Display is blank or extremely dim with no visible content
Ranked causes
- 1. Cardiograph is in standby mode (common)
- 2. Cardiograph battery is extremely low (occasional)
- 3. AC Power not applied or AC Power indicator LED is dark (occasional)
- 4. Battery insertion or connection failure (occasional)
- 5. Display assembly failure (rare)
- 6. Main board failure (rare)
Press On/Standby button
- Test point
- Power button
- Expected
- Cardiograph wakes up and display activates
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer error: Out of paper, paper jam, or door open
Messages indicate paper supply exhausted, paper jammed, or door sensor triggered
Ranked causes
- 1. Paper drawer not fully closed (common)
- 2. Paper out of supply or loaded incorrectly (common)
- 3. Incompatible paper loaded (occasional)
- 4. Defective paper tray assembly (occasional)
- 5. Top-of-form sensor or door closed sensor malfunction (rare)
Verify paper drawer fully closed
- Test point
- Paper drawer latch
- Expected
- Drawer fully closed and latched
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more leads show leads off periodically patient safety
One or more ECG leads display red dashed line intermittently, signal interruptions occur
Ranked causes
- 1. Improper or incomplete patient and skin preparation (common)
- 2. Improperly attached electrodes or lead wires; expired gel electrodes (common)
- 3. Defib or high noise event (occasional)
- 4. PIM reset normal behavior (common)
Verify patient skin preparation completed
- Test point
- Patient skin condition and electrode application area
- Expected
- Skin clean, dry, properly abraded; electrodes applied per protocol
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not advance
Paper feed mechanism fails to move paper through printer
Ranked causes
- 1. Application error (occasional)
- 2. Top-of-form sensor malfunction (occasional)
- 3. Defective print head assembly (occasional)
Restart cardiograph
- Test point
- System startup
- Expected
- Paper advances normally
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM not detected message or PIM with red X icon appears patient safety
Display shows 'Patient Interface Module (PIM) not detected' message or PIM icon with red X on screen
Ranked causes
- 1. PIM not connected (common)
- 2. Faulty patient data cable (common)
- 3. Faulty PIM (occasional)
Verify PIM patient data cable connection
- Test point
- PIM connector on rear of cardiograph
- Expected
- Cable firmly seated with no gaps
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
All leads show leads off continuously patient safety
Red dashed line displayed on all ECG leads continuously without clearing
Ranked causes
- 1. PIM communications lost due to USB patient data cable connection failure (common)
- 2. Defective leads connected to PIM (occasional)
Check for PIM icon with red X on screen
- Test point
- Top right of display screen
- Expected
- PIM icon visible or not visible
3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PIM Test patient safety
Verification that Patient Interface Module (PIM) is communicating correctly with cardiograph and can acquire patient data when PIM error messages occur or data acquisition fails.
Ranked causes
- 1. PIM communication error or PIM patient data cable disconnection (common)
- 2. PIM hardware failure or defective PIM patient data cable (occasional)
Verify PIM patient data cable connection
- Test point
- PIM connector on cardiograph rear panel, cable physical condition
- Expected
- Cable fully seated in connector with no gaps; connector barrel visible and accessible for grounding
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard is unresponsive
Keyboard input not registering or input focus lost
Ranked causes
- 1. Input focus lost (common)
- 2. Keyboard controller subsystem has failed (rare)
Touch cardiograph screen
- Test point
- Display screen
- Expected
- Input focus restored and keyboard responds
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No matching ECG found on server
ECG with specified patient ID not found on selected TraceMaster server
Ranked causes
- 1. Incorrect patient ID specified (common)
- 2. ECG does not exist on server (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No reply received from server
Server did not respond during remote transfer; connection lost or server down
Ranked causes
- 1. LAN connection disconnected during transfer (occasional)
- 2. TraceMaster server is down (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not tear off cleanly or stops at wrong spot
Paper tearoff mechanism fails or paper stops at incorrect position
Ranked causes
- 1. Wrong paper size setting (common)
- 2. Wrong tearoff method used (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient Interface Module (PIM) communication failure
Cardiograph unable to acquire data from PIM or displays PIM error messages during operation
Ranked causes
- 1. PIM patient data cable not securely attached (common)
- 2. PIM hardware fault or cable defect (occasional)
Verify PIM patient data cable is securely connected
- Test point
- PIM connector on rear of cardiograph
- Expected
- Cable is firmly seated and fully connected to PIM connector
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not detected message
System fails to recognize or communicate with printer assembly
Ranked causes
- 1. Application error (occasional)
- 2. Defective print head assembly (occasional)
Restart cardiograph
- Test point
- System startup
- Expected
- Printer detected and operational
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer unable to accept print job - busy state
Printer reports busy status or USB error after sending print job; printing disabled with low battery
Ranked causes
- 1. Printer cannot accept new print job (common)
- 2. Battery too low or missing for printing (common)
Verify battery status before printing
- Test point
- Battery indicator on cardiograph
- Expected
- Two batteries installed and charged; AC power connected or batteries above minimum level
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen difficult to select or unresponsive
Items on touch screen require many attempts to select, specific screen areas are difficult to select, or part of screen appears unresponsive to touch input
Ranked causes
- 1. Touch screen requires recalibration due to usage environment or user change (common)
- 2. Touch screen calibration not improved after Force Calibration (occasional)
Perform Force Calibration procedure
- Test point
- Touch screen with cross hair
- Expected
- Multiple cross hairs touched in sequence until procedure completes with acceptance message
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB memory stick not accessible message
System cannot detect or communicate with USB memory stick
Ranked causes
- 1. USB stick not fully inserted (common)
- 2. Unsupported software on USB stick (occasional)
Verify USB stick insertion
- Test point
- USB connector on right side of cardiograph
- Expected
- USB stick fully inserted and recognized
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio system not emitting sound
Cardiograph produces no audible output during audio test or alarms are inaudible
Ranked causes
- 1. Audio system fault or speaker malfunction (rare)
Run audio test and listen for chime tones
- Test point
- Audio Test button; speaker output
- Expected
- Audible chime tones are emitted from speaker during test
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Blank pages printed after ECG report
Extra blank pages output following completed ECG printout
Ranked causes
- 1. Wrong paper size setting (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery level indicator accuracy
Battery level indicator on Status Bar and Battery Status window may display inaccurate remaining runtime and power. Calibration enhances accuracy of battery indicators for reliable daily work.
Ranked causes
- 1. Battery calibration required due to clinical use factors (occasional)
Attach AC power cord and verify connection
- Test point
- AC power indicator light next to power button
- Expected
- Lit
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery status
Inaccurate battery status information displayed on cardiograph
Ranked causes
- 1. Battery calibration required (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cardiograph sensitivity
Overall sensitivity test to verify cardiograph signal acquisition performance
Ranked causes
- 1. Yearly sensitivity verification (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Keyboard Test
Verification that all keyboard and hardware keys are properly recognized when pressed and functional for cardiograph operation.
Ranked causes
- 1. Need to verify keyboard functionality after maintenance or to diagnose key responsiveness issues (common)
Execute Keyboard/Hardware Key Test
- Test point
- Service Utilities Keyboard Test screen, physical keys on cardiograph
- Expected
- Each key press is recognized and registered by test software; no missed or delayed key detections
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Lead Wire Tests patient safety
Verification of patient lead wire integrity through open and short circuit tests to detect breaks, continuity issues, or shorted connections.
Ranked causes
- 1. Need to verify lead wire continuity and integrity before patient use (common)
Verify all lead wires continuity with open test
- Test point
- Each patient lead wire conductor
- Expected
- All leads show continuity; no open circuits detected; expected resistance within specification
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Power On Self-Test
Automatic self-tests performed during cardiograph boot sequence to verify software integrity, flash memory, PIM status, and USB connectivity.
Ranked causes
- 1. Need to verify system integrity after power cycle or maintenance (common)
Observe software identification screen appearance
- Test point
- Cardiograph display
- Expected
- Software identification screen appears within approximately 20 seconds of power-on; audible beep heard
4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Printer Test
Verification that cardiograph printer correctly prints test page and produces acceptable print quality without errors.
Ranked causes
- 1. Need to verify printer functionality or diagnose print quality issues (common)
Printer test execution and paper output
- Test point
- Printer output
- Expected
- Test page prints successfully without paper jams, feed errors, or mechanical issues
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Simulated ECG Recording
Verification of ECG recording functionality using a 12-lead ECG simulator to confirm patient lead integrity, paper speed accuracy, and gain setting accuracy.
Ranked causes
- 1. Need to verify ECG signal acquisition and recording accuracy (common)
Verify all lead wires connected to simulator
- Test point
- Display screen leads map
- Expected
- No red x marks next to any electrodes; all leads showing activity, flat lines initially
5 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch screen
Touch screen requires frequent attempts to select items or selecting items in specific areas is difficult; may occur after relocation or with different user heights
Ranked causes
- 1. Touch screen drift or misalignment due to usage conditions (common)
Initiate touch screen calibration
- Test point
- Keyboard Ctrl + Alt + C keys
- Expected
- White screen appears with cross hair at center
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Visual Inspection
External visual inspection of cardiograph for damage, wear, corrosion, contamination, and hardware integrity before performing functional tests.
Ranked causes
- 1. Equipment may have accumulated damage or wear during service (common)
Inspect power cord condition
- Test point
- Power cord exterior
- Expected
- No visible wear, cracks, or damage to insulation; intact and secure connectors
10 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot access internal Compact Flash memory message
System unable to read or write to internal CF card
Ranked causes
- 1. Compact Flash card corruption (occasional)
Run Internal Compact Flash test
- Test point
- Service Utilities
- Expected
- Test passes
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot store additional ECGs message
System prevents storing more ECGs due to storage capacity full
Ranked causes
- 1. CF card file system limitation reached (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiograph must restart to load custom settings
System requires restart after loading new configuration file from USB
Ranked causes
- 1. Configuration file loaded; restart needed to apply settings (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cardiograph unresponsive after rapid paper drawer cycling
System becomes unresponsive following repeated opening and closing of paper drawer
Ranked causes
- 1. Printer controller error from excessive drawer cycling (rare)
Monitor system recovery
- Test point
- Cardiograph startup
- Expected
- System restarts and boots correctly; reports print correctly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Printer message with no apparent problem
Check Printer message appears without observable printer malfunction
Ranked causes
- 1. Printer controller error (occasional)
Reset paper drawer
- Test point
- Printer paper drawer
- Expected
- Message does not reoccur after drawer cycle
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cleaning: Touch screen
Touch screen requires occasional cleaning to maintain visibility and responsiveness
Ranked causes
- 1. Dust, dirt or residue accumulation on touch screen surface (common)
Prepare cleaning cloth
- Test point
- Soft cloth
- Expected
- Cloth dampened with water or isopropyl alcohol, excess moisture wrung out
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Command canceled
Operation was canceled by user or invalid information was specified
Ranked causes
- 1. User canceled action or provided invalid information (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Command cannot be completed
TraceMaster system is busy; operation request cannot be processed at this time
Ranked causes
- 1. TraceMasterVue system busy or resource contention (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Command timed-out message during archive operation
Error message 'Command timed-out' appears when attempting archive or transfer operation
Ranked causes
- 1. TraceMaster ECG Management system is busy (occasional)
Retry archive or transfer operation
- Test point
- Network and TraceMaster server availability
- Expected
- Operation completes within timeout period
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration file load error - invalid XML
Configuration file cannot be loaded due to invalid XML version, platform, or file corruption
Ranked causes
- 1. Corrupted or invalid configuration file (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECGs with incomplete patient information cannot be transferred message
Error message 'ECGs with incomplete required patient information fields cannot be transferred' displayed during archive
Ranked causes
- 1. Required patient information not entered for ECG record (common)
Verify all required patient information fields completed
- Test point
- ECG record patient data fields
- Expected
- All required fields populated with patient information
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error accessing file from USB memory stick
Cannot copy configuration or log file from USB memory stick; file access error
Ranked causes
- 1. USB file access failure or file not present (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error opening corrupted file
Cannot open configuration or log file due to file corruption
Ranked causes
- 1. Configuration or log file is corrupted (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error retrieving ECG from archive - compact flash error
Cannot delete or retrieve selected ECGs from archive due to internal card error
Ranked causes
- 1. Internal compact flash card read/write error (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error: ECG from legacy TraceMaster system cannot be printed
ECG originated from older TraceMaster system with incompatible format
Ranked causes
- 1. ECG XML version incompatible from legacy system (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error: Lead system not supported on cardiograph
ECG XML version incompatible with cardiograph lead system configuration
Ranked causes
- 1. ECG XML version incompatible (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Failed to implement custom settings
Cardiograph automatically restarts and restores current settings, unable to load custom configuration
Ranked causes
- 1. Failure to load a configuration file (unknown)
Restore factory default settings
- Test point
- Settings menu
- Expected
- Settings restored to factory defaults
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Fax transmission did not complete
Fax transmission failed to complete successfully
Ranked causes
- 1. Fax transmission failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
File already exists
Attempted to save log file to a name that already exists in the system
Ranked causes
- 1. Log file name already exists on system (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
File system is destroyed
Dialog box showing software version information with black background, indicating file system corruption
Ranked causes
- 1. File system corruption (occasional)
Wait 2 minutes
- Test point
- System
- Expected
- System stabilizes
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flash memory failure
Onboard Flash memory test fails or memory appears corrupted
Ranked causes
- 1. Onboard Flash memory fault (rare)
Run onboard Flash memory test
- Test point
- Flash Memory Test button in Diagnostic Test screen
- Expected
- Test passes and confirms onboard Flash memory is working properly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
FPGA communication or connection failure
Field Programmable Gate Array (FPGA) is not connected or incorrectly configured; internal error
Ranked causes
- 1. FPGA internal error or disconnection (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ID or ECG button on keyboard is unresponsive
Specific button on keyboard not responding to input
Ranked causes
- 1. Input focus loss affecting button (common)
Touch cardiograph screen to restore input focus
- Test point
- Display screen
- Expected
- Button becomes responsive after focus restored
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Insufficient internal storage for new orders
System reports not enough storage available to save downloaded or new orders
Ranked causes
- 1. Application storage full (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal Compact Flash failure
Internal Compact Flash test fails or card appears non-functional
Ranked causes
- 1. Internal Compact Flash card fault (rare)
Run Internal Compact Flash test
- Test point
- Internal Compact Flash button in Diagnostic Test screen
- Expected
- Test passes and confirms Internal Compact Flash is working properly
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal CompactFlash card cannot be accessed message
System unable to communicate with or access internal CF storage
Ranked causes
- 1. Compact Flash card corruption (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal error with orders list
Orders list does not refresh; internal application error occurs
Ranked causes
- 1. Orders list refresh failure (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal function error message
Error message 'Internal function error' displayed during operation
Ranked causes
- 1. Invalid file or file resides on unsupported media (occasional)
Restart cardiograph
- Test point
- System power and boot sequence
- Expected
- System restarts without error messages
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid default gateway message
Cardiograph displays invalid default gateway error during wireless configuration
Ranked causes
- 1. Incorrect default gateway in TraceMasterVue configuration (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid default TraceMaster connection for OrderVue inbox
Incorrect TraceMasterVue server specified in configuration for selected OrderVue inbox
Ranked causes
- 1. TraceMasterVue server mismatch in OrderVue configuration (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid directory message
Cardiograph displays invalid directory error during wireless configuration
Ranked causes
- 1. Incorrect directory in TraceMasterVue configuration (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid filename in save dialog
Invalid or missing file name in save logs dialog box
Ranked causes
- 1. File name field left blank in save logs dialog (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Invalid IP address message
- Invalid path and file name specified
- Invalid Primary DNS message
- Invalid Primary WINS message
- Invalid Report Handle message during preview, printing, or saving ECGs
- Invalid subnet mask message
- Invalid transfer destination selected
- Invalid XML schema version message
- Keyboard is typing incorrect characters
- Keyboard or hardware key not recognized
- LAN/WLAN connection not available
- Maximum number of ECGs stored message
- Missing fax recipient information
- Missing network settings information
- More than 200 ECGs on USB stick; only first 200 display
- Network connection to server was reset
- Network connectivity problem
- No default TraceMaster server selected
- No fax recipient available to edit
- No filename specified
- No order selected for ECG operation
- No order source selected
- No Report message during save or archive operation
- No TraceMaster connections configured
- No TraceMaster server to delete
- Occasionally copying ECGs from archive to USB stick fails
- Order application error - function failed
- Order search range not specified
- Order XML error
- Orders limit reached (200 maximum)
- Orders previously archived with ECGs cannot be downloaded
- Patient ID field must be enabled for TraceMasterVue transfer
- Print error: Unknown error
- Printed report looks too small or compressed
- Printer door is open message
- Printer temperature too high
- Printer test failed or poor print quality
- Printer underflow message
- Printer voltage high message
- Printer voltage low message
- Remote query could not be completed
- Remote query get ECG error
- Remote query get list error
- Remote transfer error
- Remote transfer schema error
- Required patient information missing from current ECG
- Selected ECG exceeds available space on TraceMaster message
- Selection exceeds maximum number of outboxes
- Touch screen calibration required
- Unexpected reply received from server
- Unknown error accessing storage capacity of archive
- USB memory stick not detected
- Waveform display no longer updating; system unresponsive
- Worklist full; orders cannot be downloaded
- XML version incompatible with TraceMaster
55 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
After pressing Archive, takes >40 seconds | Takes longer than 40 seconds before Archive list is displayed with >130 ECGs stored | Accessing Archive screen when main archive contains near-maximum number of stored ECGs (>130) | warning |
Archive function could not be completed | Archive function could not be completed; please restart cardiograph and try again | Archive process error or failure | error |
Archive process was terminated due to an error | Archive process was terminated due to an error message | Archive process error during operation | error |
Bad reply received from TraceMaster | Bad reply received from TraceMaster Remote site message | Invalid or malformed response received from TraceMaster server during archive operation | error |
Cannot access internal Compact Flash memory | System unable to read or write internal CF card | CF card corrupted, electrical connection failed, or hardware malfunction | error |
Cannot store additional ECGs | CompactFlash card storage space is full or file system at capacity | Available storage space on CF card exhausted | warning |
Check Printer | Printer controller detected internal error condition | Controller detects communication error, timeout, or sensor malfunction | warning |
Command timed-out | Command timed-out message during archive or transfer operation | TraceMaster system busy or network timeout during transfer operation | warning |
ECGs incomplete required patient information | ECGs with incomplete required patient information fields cannot be transferred | Required patient data fields missing when attempting archive or transfer to TraceMaster | error |
Error. Lead system not supported | ECG XML file contains lead configuration not supported by this cardiograph model | ECG transferred from TraceMasterVue with unsupported lead system or XML version | error |
Error. Legacy TraceMaster ECG cannot be printed | ECG originated from older TraceMaster system with incompatible XML format | ECG file transferred from legacy TraceMaster system | error |
Internal CompactFlash card cannot be accessed | System unable to communicate with or access internal CompactFlash storage | CF card hardware failure, connection error, or file system corruption | critical |
Internal function error | Internal function error message displayed | Invalid file or file resides on unsupported media | error |
Invalid database | Archive database may be corrupted message | Archive database access failure or corruption detected during archive operation | error |
Invalid Report Handle | Report handle is invalid when attempting to preview, print, or save ECGs | Invalid file or corrupted report object mapping during ECG preview, print, or save operation | error |
Maximum number of ECGs stored | CompactFlash card archive has reached maximum ECG file count limit | Number of stored ECG files equals CF card file system limit | info |
More than 200 ECGs on USB stick | USB memory stick contains more than 200 ECG files; only first 200 will display | ECG file count on USB stick exceeds 200 file display limit | info |
No reply from TraceMaster | No reply received from TraceMaster Remote site message | No response from TraceMaster server during query or transfer operation | error |
No Report | No report message appears; software cannot map input report handle to internal report object | Invalid handle during save or archive operation | error |
Paper does not advance | Paper feed mechanism fails to move paper through printer during operation | Paper motor not responding, sensor failure, or mechanical blockage | error |
Patient ID field must be enabled | Attempted ECG transfer to TraceMasterVue without Patient ID field enabled or populated | Transfer initiated with Patient ID field disabled or blank | warning |
Printer Door is Open | Printer door position sensor indicates door not fully closed | Door latch not engaged or sensor malfunction | warning |
Printer Error: Door Open | Printer door position sensor indicates door not fully closed | Door latch not engaged or sensor malfunction during printing | warning |
Printer Error: Out of Paper | Paper supply is exhausted or not detected in tray | Top-of-form sensor does not detect paper presence or paper runs out during printing | error |
Printer Error: Paper Jam | Paper detected as stuck or jammed in paper path | Mechanical jam detected or inconsistent paper feed during printing | error |
Printer not detected | Cardiograph software cannot communicate with or initialize printer | Printer interface initialization failure or print head not responding | error |
Print Error: Unknown Error | Unidentified error condition detected during print operation | Application software encountered unexpected condition or timeout | error |
Printer temperature is too high | Print head thermal sensor detected overtemperature condition | Print head temperature exceeds safe operating limit | warning |
Printer underflow | Communication dropout between cardiograph and printer during rhythm printing | PIM interface loses signal or fails handshake during print operation | error |
Printer voltage high | Main board printer supply voltage exceeds maximum specification | Power supply output rises above acceptable upper threshold | error |
Printer voltage low | Main board printer supply voltage falls below minimum specification | Power supply output drops below acceptable lower threshold | error |
Requested Function could not be completed | Requested function could not be completed message during transfer or query operation | Network or modem phone line interrupted during operation; USB memory stick not fully inserted | error |
Required patient information is missing | Current ECG lacks mandatory patient data needed for TraceMasterVue transfer | Transfer to TraceMasterVue attempted with incomplete patient information | warning |
Selected ECG exceeds available space | Selected ECG(s) exceeds available space on TraceMaster Remote site | TraceMaster system disk full when attempting to transfer ECG data | error |
Transfer to USB fails with <200 ECGs on media | Transfer of ECGs to USB memory stick fails with less than 200 ECGs on removable media device | ECG transfer to USB memory stick fails despite available space reports | error |
Unexpected reply from TraceMaster | Unexpected reply received from TraceMaster Remote Site message | Unexpected or invalid response received from TraceMaster server during archive operation | error |
Unknown error accessing storage capacity | System failed to determine available storage capacity of archive media | Storage media access error or communication failure when querying capacity | error |
USB memory stick not accessible | Cardiograph cannot detect or access USB memory stick | USB stick not fully inserted or contains incompatible software | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI