Philips PageWriter TC: troubleshooting and error codes

The ServiceForge knowledge base holds 246 documented symptoms for the Philips PageWriter TC, with 328 ranked causes, 363 diagnostic checks with test points and expected values, and 108 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

One or more leads show leads off periodically patient safety

Red dashed line appears on lead display intermittently during signal acquisition

Ranked causes

  1. 1. Improper or incomplete patient and skin preparation (common)
  2. 2. Improperly attached electrodes or lead wires; expired disposable gel electrodes (common)
  3. 3. Defib or high noise event (occasional)
  4. 4. PIM reset (up to ten second duration) (common)
  5. 5. PIM communications lost due to USB patient data cable failure or defective leads (occasional)
  6. 6. Specific electrode or cable failure (TC20 models) (occasional)
First check

Review patient preparation documentation

Test point
Patient and skin preparation area
Expected
Good patient preparation and skin preparation techniques followed per section 3-10

9 more checks with pass and fail actions, 14 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer Error: Out of Paper / Paper Jam / Door Open

Error message displayed for out of paper, paper jam, or door open; may appear despite correct paper loading and closed door

Ranked causes

  1. 1. Paper drawer not fully closed (common)
  2. 2. Out of paper or incorrectly loaded paper (common)
  3. 3. Incompatible paper loaded (occasional)
  4. 4. Defective paper tray assembly (rare)
  5. 5. Top-of-form sensor or door closed sensor failure (rare)
First check

Verify paper drawer is fully closed

Test point
Drawer latch
Expected
Drawer latches securely in fully closed position

4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer error: Out of Paper, Paper Jam, or Door Open messages caution

Printer displays error messages for out of paper, paper jam, or door open when actual condition may not exist.

Ranked causes

  1. 1. Paper drawer not fully closed (common)
  2. 2. Out of paper or incorrectly loaded (common)
  3. 3. Incompatible paper loaded (occasional)
  4. 4. Defective paper tray assembly (occasional)
  5. 5. Top-of-form or door closed sensor fault (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Screen is dark and shows no image

Display remains blank or dark; no visual output; backlight may be off

Ranked causes

  1. 1. Cardiograph in Standby mode (common)
  2. 2. Battery capacity insufficient (below 1080 mAh) (common)
  3. 3. AC power supply failure (occasional)
  4. 4. Battery insertion or charge issue (occasional)
  5. 5. File system corrupted (rare)
First check

Check if cardiograph is in Standby mode

Test point
Display state and power indicator
Expected
Display should show active screen when not in Standby

2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery cycle count exceeded patient safety

Battery has exceeded charge cycle limit (300 or 400); message indicates immediate or soon replacement required; battery continues to charge but degraded

Ranked causes

  1. 1. Battery exceeded 300 charge cycles (common)
  2. 2. Battery exceeded 400 charge cycles (common)
  3. 3. Battery state of health (SOH) <= 80% (common)
  4. 4. Battery state of health (SOH) <= 75% (common)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

One or more leads show low amplitude patient safety

ECG waveform appears with reduced signal strength or height on affected leads

Ranked causes

  1. 1. Improper or incomplete patient and skin preparation (common)
  2. 2. Lead placement not standard (common)
  3. 3. Gooey adapters; electrodes or lead wires not properly attached (common)
  4. 4. Faulty lead wires (occasional)
First check

Verify patient preparation technique

Test point
Patient skin preparation area
Expected
Proper skin prep completed per Instructions for Use

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM not detected message or disconnected icon appears patient safety

PIM not detected message displays or PIM disconnected icon appears on screen, indicating communication loss between cardiograph and PIM.

Ranked causes

  1. 1. PIM not connected to cardiograph (common)
  2. 2. Faulty patient data cable (common)
  3. 3. Faulty PIM (occasional)
  4. 4. Class A PIM is connected (rare)
First check

Verify Class B PIM is connected

Test point
PIM model identification
Expected
Class B PIM attached

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM not detected or PIM disconnected icon appears patient safety

Patient Interface Module (PIM) is not recognized by cardiograph; disconnect icon visible on screen

Ranked causes

  1. 1. PIM not connected (common)
  2. 2. Faulty patient data cable (common)
  3. 3. Faulty PIM (occasional)
  4. 4. Class A PIM is connected to the cardiograph (rare)
First check

Verify Class B PIM is connected and cable is firmly seated

Test point
PIM connector on rear panel
Expected
PIM firmly connected

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM reset with temporary leads off display

All leads display as red dashed lines for up to ten seconds and PIM disconnected icon appears during normal PIM reset operation.

Ranked causes

  1. 1. Normal PIM reset during power on/wakeup (common)
  2. 2. PIM reset during screen changes (common)
  3. 3. PIM reset during USB device insertion or removal (common)
  4. 4. PIM reset during electrostatic discharge (occasional)
First check

Verify cardiograph is powering on or waking up

Test point
System power state
Expected
Power on or wakeup event in progress

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

All leads show leads off (red dashed line) continuously patient safety

No ECG signal detected on any lead; red dashed lines displayed on screen instead of waveforms

Ranked causes

  1. 1. PIM communications lost due to USB patient data cable connection failure (common)
  2. 2. Defective lead wires connected to PIM (common)
  3. 3. Class A PIM is connected to the cardiograph (rare)
First check

Verify RL/N lead and electrode firmly attached

Test point
RL/N lead connection to patient
Expected
Firmly attached

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

All leads show leads off continuously patient safety

All leads display as red dashed lines continuously on the cardiograph screen, indicating no signal acquisition.

Ranked causes

  1. 1. USB patient data cable connection failure (common)
  2. 2. Defective leads connected to the PIM (common)
  3. 3. Class A PIM is connected to the cardiograph (occasional)
First check

Check if PIM disconnected icon appears on Status Bar

Test point
Status Bar top right of screen
Expected
PIM disconnected icon not visible

5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery ID not approved or not recognized caution

Battery ID on About the Cardiograph screen does not match approved list; risk of battery overheating, shortened life, or cardiograph damage

Ranked causes

  1. 1. Unapproved battery installed (unknown)
  2. 2. Unsupported battery installed (unknown)
  3. 3. Legacy battery PN 989803160981 installed (occasional)
First check

Verify battery ID matches approved list

Test point
About the Cardiograph screen
Expected
P508; P510; 202CJC; E013RJ; T508; or PH

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard is typing incorrect characters

Keyboard output does not match keys pressed; characters may be shifted or wrong language selected

Ranked causes

  1. 1. Caps Lock feature is enabled (common)
  2. 2. Incorrect keyboard option installed (rare)
  3. 3. Main board recently replaced; keyboard setting reset to incorrect option (occasional)
First check

Check Caps Lock status

Test point
Status Bar on display
Expected
Cap Lock icon absent or disabled

1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper does not advance

Paper fails to feed or advance through printer mechanism

Ranked causes

  1. 1. Application error (occasional)
  2. 2. Top-of-form sensor failure (rare)
  3. 3. Defective print head assembly (rare)
First check

Restart cardiograph

Test point
Printer paper feed mechanism
Expected
Paper advances normally on next print job

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Ping test failed

Network connectivity test fails or returns error

Ranked causes

  1. 1. LAN cable not securely attached (common)
  2. 2. Wireless connection not associated or signal lost (occasional)
  3. 3. Invalid IP address entered (occasional)
First check

Verify LAN cable is securely attached or wireless connection is active

Test point
LAN connector on rear or wireless Status Bar indicator
Expected
LAN cable fully seated OR green bars on Status Bar indicating live wireless connection

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Remote Query get ECG error

TraceMaster returned wrong ECG file; server or configuration issue

Ranked causes

  1. 1. TraceMaster configuration mismatch (common)
  2. 2. LAN/WLAN connection issue (common)
  3. 3. TraceMaster server offline (occasional)
First check

Check TraceMaster configuration

Test point
TraceMaster settings
Expected
Settings correct for server

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Requested Function could not be completed

Message indicates the requested function could not be completed due to network, modem, or media issues.

Ranked causes

  1. 1. Network or modem phone line interrupted during operation (common)
  2. 2. USB memory stick not fully inserted (common)
  3. 3. Attempt to transfer unsupported ECG report type (occasional)
First check

Retry operation

Test point
Network connectivity
Expected
Operation completes successfully

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Requested Function could not be completed message

Archive or transfer operation failed to complete

Ranked causes

  1. 1. Network interrupted during operation (occasional)
  2. 2. USB memory stick not fully inserted (common)
  3. 3. Attempt to transfer unsupported ECG report type (occasional)
First check

Check network connection

Test point
Network cable and connectivity
Expected
Network cable firmly connected and active

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unrecognized battery or battery mismatch patient safety

Battery status message indicates unsupported battery, different part numbers, or charging suspension; cardiograph will not operate or charging halted

Ranked causes

  1. 1. One or more installed batteries are not supported or not Philips-approved (common)
  2. 2. Two batteries with different part numbers installed (common)
  3. 3. Battery overcharged or non-Philips battery installed (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

USB memory stick not accessible

System cannot access USB memory stick. Message indicates stick not fully inserted or contains unsupported software.

Ranked causes

  1. 1. USB stick not fully inserted into USB connector (common)
  2. 2. USB stick contains unsupported software (occasional)
  3. 3. Defective or incompatible USB memory stick (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

USB memory stick not accessible or not fully inserted

Message displayed: 'The USB memory stick is not accessible. Please ensure that the USB memory stick is fully inserted into the USB connector (right side of cardiograph).'

Ranked causes

  1. 1. USB memory stick not fully inserted into slot (common)
  2. 2. USB memory stick contains unsupported software (occasional)
  3. 3. USB memory stick not fully functioning or incompatible (occasional)
First check

Verify USB memory stick insertion

Test point
USB connector on right side of cardiograph
Expected
USB memory stick fully inserted and recognized

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

A single key press produces repeated characters on screen

Keyboard input malfunction; one key press results in multiple character repetitions displayed

Ranked causes

  1. 1. Failure of keyboard key (common)
  2. 2. PIM patient data cable failure (occasional)
First check

Perform Keyboard/HW Key Test

Test point
Keyboard hardware
Expected
Test passes

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery over temperature patient safety

Battery has reached excessive temperature; display shows temperature warning and automatic shutdown imminent, or charging suspended pending cool-down

Ranked causes

  1. 1. Battery temperature reached 50°C or over temperature detected (common)
  2. 2. Battery temperature reached or exceeded 60°C (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery over temperature warning caution

Temperature warning displayed; cardiograph will shut down

Ranked causes

  1. 1. Battery temperature between 50°C and 59°C (occasional)
  2. 2. Battery temperature at or above 60°C (rare)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery performance degradation

Battery shows reduced runtime, slow charging, or capacity loss

Ranked causes

  1. 1. Storage temperature outside recommended range (common)
  2. 2. Operating temperature outside specification (occasional)
First check

Verify battery storage temperature

Test point
Storage area
Expected
25°C +/- 5°C (77°F +/- 10°F), +/- 5°C

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery remains in Standby instead of automatic shutdown

Cardiograph stays in Standby mode and does not proceed to automatic shutdown when conditions are met.

Ranked causes

  1. 1. Battery cycle count 600 or less with AC power connected (occasional)
  2. 2. Battery full charge capacity is 50% or greater with AC power connected (occasional)
First check

Check battery cycle count

Test point
Battery status or settings menu
Expected
Cycle count 600 or less

1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI

Calibration: Battery status accuracy

Battery status information on screen may be inaccurate; Expected Max Error (%) and Full Capacity (mAh) values need verification after battery depletion cycle.

Ranked causes

  1. 1. Battery calibration procedure required (common)
  2. 2. Expected Max Error field did not reset to 2% (occasional)
First check

Verify Expected Max Error (%) and Full Capacity (mAh) values changed from printed report values

Test point
Battery Status column on Service Utility screen
Expected
Values different than those printed on previous report

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: PIM communication verification

Patient Interface Module (PIM) not communicating with cardiograph or PIM error messages displayed. Test confirms PIM is communicating properly.

Ranked causes

  1. 1. PIM patient data cable loose or disconnected (common)
  2. 2. PIM hardware or patient data cable fault (occasional)
First check

Verify PIM patient data cable securely attached

Test point
PIM connector on rear of cardiograph
Expected
Cable firmly seated; PIM disconnected icon not visible on Status Bar

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG signal loss, noise, or reduced amplitude on leads patient safety

Intermittent ECG signal dropout, increased noise, or weaker than normal signal strength on one or more leads

Ranked causes

  1. 1. Lead wire shortage (common)
  2. 2. Cracked or broken cable insulation on lead wire (occasional)
First check

Run Lead Wire Performance Test to identify shortages

Test point
All lead wires
Expected
No shortages reported

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Incorrect language appears on cardiograph software

Software displays text in wrong language after main board replacement

Ranked causes

  1. 1. Main board recently replaced; software localization reset to wrong setting (common)
  2. 2. Persistent language setting issue after software reinstall (rare)
First check

Verify main board was recently replaced

Test point
Service records and timeline
Expected
Main board replacement documented within last service call

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard is unresponsive

Keyboard does not respond to user input; may be caused by input focus loss or controller subsystem failure

Ranked causes

  1. 1. Input focus loss (common)
  2. 2. Keyboard controller subsystem has failed (rare)
First check

Touch screen to restore input focus

Test point
Touch screen
Expected
Keyboard responds to input

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No matching ECG found on server

Patient ID incorrect or ECG does not exist on selected server

Ranked causes

  1. 1. Incorrect Patient ID (common)
  2. 2. ECG does not exist on server (occasional)
First check

Verify Patient ID

Test point
Patient ID field
Expected
Correct Patient ID entered

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No reply received from Remote site message

Archive transfer fails because remote server does not respond

Ranked causes

  1. 1. Server is unavailable (occasional)
  2. 2. Configuration error in server address or URL (occasional)
First check

Verify server availability

Test point
Remote server status
Expected
Server accessible and responding

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No reply received from server

LAN connection broken during transfer or server offline

Ranked causes

  1. 1. LAN connection broken during transfer (occasional)
  2. 2. TraceMaster server offline (occasional)
First check

Verify LAN connection

Test point
LAN link
Expected
Connection active

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Order error - requested function failed

Message displayed: 'Order Error: Requested function failed message'. Order application error.

Ranked causes

  1. 1. Order application function error (occasional)
  2. 2. Persistent order application error (rare)
First check

Restart the cardiograph

Test point
System boot sequence
Expected
Cardiograph restarts successfully

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper does not tear cleanly or stops at wrong spot caution

Paper tear mechanism fails or paper stops at unintended position during tear-off.

Ranked causes

  1. 1. Wrong paper size setting (common)
  2. 2. Wrong tear-off method (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Paper does not tear off cleanly or stops at wrong spot

Printed paper does not tear cleanly at expected location or stops tearing at incorrect position

Ranked causes

  1. 1. Wrong paper size setting configured (common)
  2. 2. Wrong method used to tear off paper (occasional)
First check

Verify paper size setting

Test point
Cardiograph configuration
Expected
Paper size setting matches loaded paper size

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Patient ID field must be enabled to transfer ECGs to TraceMaster

Message displayed: 'The Patient ID field must be enabled in order to transfer ECGs to TraceMaster. Touch the [ON] button underneath the Enabled column to activate the Patient ID field.' ECG transfer attempted without Patient ID.

Ranked causes

  1. 1. Patient ID field not enabled (common)
  2. 2. No Patient ID entered for ECG (common)
First check

Enable Patient ID field

Test point
Setup menu - Patient ID section
Expected
Patient ID field set to ON

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM Test failed

Error message appears when attempting to run tests

Ranked causes

  1. 1. PIM patient data cable not securely attached (common)
  2. 2. Cable or connector damaged (occasional)
First check

Verify PIM patient data cable is securely attached

Test point
PIM connector on rear of cardiograph
Expected
Cable fully seated with no gaps

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

PIM test failed or PIM communication error patient safety

PIM error messages displayed or cardiograph unable to acquire data from PIM; PIM disconnected icon appears on status bar.

Ranked causes

  1. 1. PIM patient data cable loose or disconnected (common)
  2. 2. PIM or PIM patient data cable defective (occasional)
First check

Verify PIM patient data cable connection

Test point
PIM connector on rear of cardiograph
Expected
Cable securely attached; PIM disconnected icon not present on status bar

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer not detected

Message appears indicating printer is not detected by the system

Ranked causes

  1. 1. Application error (occasional)
  2. 2. Defective print head assembly (rare)
First check

Restart cardiograph and check printer detection

Test point
Printer interface
Expected
Printer detected and operational

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Printer not detected message caution

Printer hardware is not recognized by the system.

Ranked causes

  1. 1. Application error (common)
  2. 2. Defective print head assembly (occasional)

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Remote Query get list error

TraceMaster returned wrong ECG list; configuration or network mismatch

Ranked causes

  1. 1. Incorrect TraceMaster configuration (common)
  2. 2. LAN/WLAN configuration error (common)
First check

Verify TraceMaster configuration

Test point
TraceMaster settings
Expected
Configuration matches expected server

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Transfer of ECGs to USB memory stick fails with less than 200 ECGs on media

ECG transfer fails even though fewer than 200 files exist on USB media

Ranked causes

  1. 1. USB memory stick file system limitation (occasional)
  2. 2. Not enough free space on USB stick to save selected ECGs (common)
First check

Check available space on USB stick

Test point
USB memory stick free space
Expected
Free space greater than total size of ECGs to transfer

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unknown error accessing storage capacity with archive

Message displayed: 'An unknown error occurred when accessing the storage capacity of the selected archive message'. Abnormal error occurs when getting storage capacity of selected storage media.

Ranked causes

  1. 1. Storage capacity access error during archive operation (rare)
  2. 2. Persistent storage media error (rare)
First check

Restart the cardiograph

Test point
System boot sequence
Expected
Cardiograph restarts successfully

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unknown error accessing USB memory stick storage capacity

System error occurs when querying available storage capacity on USB memory stick.

Ranked causes

  1. 1. Abnormal error getting storage capacity of selected media (rare)
  2. 2. Storage media abnormality (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Waveform display not updating; system unresponsive

ECG waveform freezes on display or entire system becomes unresponsive to user input

Ranked causes

  1. 1. Software is frozen (common)
  2. 2. Main board failure (occasional)
First check

Attempt system restart

Test point
Power/reset control
Expected
System powers down and restarts successfully; waveform updates resume

2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC power indicator light does not illuminate high voltage

Cardiograph is plugged into AC power but the green AC power indicator light does not illuminate, indicating no AC power is being supplied to the device

Ranked causes

  1. 1. AC fuse is open or blown (common)
First check

Verify AC fuse is blown or open

Test point
AC fuse located directly below AC power connector
Expected
Fuse element intact or blown

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Archive folder displays incorrect or corrupted data

Archive screen shows wrong reports or messages; archive list cannot be accessed or displays errors

Ranked causes

  1. 1. Archive corrupted or needs reset to factory defaults (occasional)
First check

Verify all archived reports transferred before reset

Test point
Archive folder contents
Expected
All critical reports backed up to remote site or USB

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Archive function could not be completed message

Archive operation fails with error message

Ranked causes

  1. 1. Archive error (occasional)
First check

Restart cardiograph

Test point
Power control
Expected
System restarts successfully and archive function available

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Archive list takes longer than 40 seconds to display with 130+ ECGs stored

System performance degradation when Main Archive contains near maximum ECG count

Ranked causes

  1. 1. System slowdown from full ECG database (occasional)
First check

Count ECGs in Main Archive

Test point
Archive database status
Expected
ECG count below 130 to restore normal performance

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Archive process was terminated due to an error message

Archive operation terminated with error before completion

Ranked causes

  1. 1. Archive error (occasional)
First check

Restart cardiograph

Test point
Power control
Expected
System restarts and archive function available

1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Archive transfer interrupted with multiple ECGs

When an error occurs during transfer of multiple ECGs to a TraceMaster system, the transfer process does not continue and transferred files are not deleted

Ranked causes

  1. 1. Known application limitation in ECG transfer (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio output not functioning

No audible sound emitted during audio test or patient does not hear chime tones

Ranked causes

  1. 1. Speaker or audio amplifier defect (occasional)
First check

Run Audio Test from Diagnostic and Test Utilities

Test point
Main screen > Setup > Service Utilities > Diagnostic and Test Utilities
Expected
Audible chime tones emitted

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Audio test fails or no sound emitted

Cardiograph does not emit audible chime tones during audio diagnostic test

Ranked causes

  1. 1. Audio system defective or speaker failed (common)
First check

Touch Audio button and listen for audible chime tones

Test point
speaker/audio system
Expected
Audible chime tones emitted

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Auto Shutdown not occurring as expected

Cardiograph Auto Shutdown feature does not trigger when configured, preventing automatic power-off

Ranked causes

  1. 1. Battery charge below 90% capacity (common)
First check

Check battery charge level via display

Test point
Battery status display
Expected
Battery charge at or above 90% capacity

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Bad reply received from Remote site message

Archive transfer fails with invalid response from remote server

Ranked causes

  1. 1. Archive error; bad reply from server (occasional)
First check

Restart cardiograph

Test point
Power control
Expected
System restarts successfully

2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Batteries not charging when cardiograph powered off

Batteries remain uncharged when cardiograph is in powered-off state despite AC power connection

Ranked causes

  1. 1. Cardiograph powered off during charging attempt (common)
First check

Verify cardiograph power state

Test point
Cardiograph power switch or display
Expected
Cardiograph in Run or Standby mode

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery calibration procedure

Deplete batteries fully while ID screen is open, then recharge to recalibrate battery monitoring. Used after battery replacement or to reset battery parameters.

Ranked causes

  1. 1. Battery calibration required after replacement or parameter reset (common)
First check

Verify AC power disconnected

Test point
AC power indicator light next to power button
Expected
Not lit

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Battery cycle count 600 or greater patient safety

Battery with 600+ cycles detected; batteries will not charge and should be replaced immediately

Ranked causes

  1. 1. Battery aged beyond safe operational limit (common)
First check

Check battery cycle count in Battery Status window

Test point
Battery Status window
Expected
Cycle count less than 600

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Battery cycle count 600 or greater or state of health 50% or less
  • Battery cycle count exceeds 300
  • Battery cycle count exceeds 400
  • Battery cycle count exceeds recommended limit
  • Battery damage or leakage detected
  • Battery damaged or leaking
  • Battery disposal required
  • Battery does not charge when cardiograph is shut down
  • Battery left in unused cardiograph longer than 2 months
  • Battery not charged to full capacity after shutdown with AC power connected
  • Battery not charged when device powered off
  • Battery not charging after cardiograph shutdown
  • Battery not charging when device powered off
  • Battery part number mismatch warning
  • Battery state of health 50% or less
  • Battery state of health at or below 75%
  • Battery state of health at or below 80%
  • Battery status shows Recommend Calibration
  • Blank pages printed after ECG report
  • Calibration: Battery level indicator accuracy
  • Calibration: ECG simulation with 12-lead simulator
  • Calibration: Patient Interface Module (PIM) Test
  • Calibration: Performance verification after external repair
  • Calibration: Performance verification after external repair with patient module access
  • Calibration: Performance verification after internal repair
  • Calibration: Performance verification after software upgrade
  • Calibration: Printer test
  • Calibration: Touch screen
  • Calibration: Touch Screen Display
  • Calibration: Touch screen force calibration
  • Calibration: Touch screen touch calibration test
  • Calibration: Verification of gain settings accuracy
  • Calibration: Verification of paper speed accuracy
  • Cannot access internal Compact Flash memory
  • Cannot disconnect modem connection from TraceMaster Remote Site
  • Cannot store additional ECGs message
  • Cannot store additional ECGs on CF card
  • Cardiograph does not restart
  • Cardiograph not to be used for extended period
  • Cardiograph return to manufacturer without battery removal
  • Cardiograph unresponsive after rapid paper drawer cycling
  • Check Printer message with no apparent problem
  • Command timed-out message
  • Connection to server was reset
  • Copying ECGs from Archive to USB Memory Stick fails
  • Copying ECGs to USB memory stick fails occasionally
  • Earth resistance test
  • ECG XML version not supported by TraceMaster
  • ECGs with incomplete patient information cannot transfer
  • ECGs with incomplete required patient information fields cannot be transferred message
  • Equipment enclosure leakage current exceeds limits
  • Error accessing external XML file message
  • Error message displayed with mismatched battery part numbers
  • Error occurred when deleting selected ECGs message
  • Error occurred when transferring selected ECGs message
  • Error occurred when trying to transmit PDF ECGs
  • Error XML file did not transform message
  • Error XML file is too large message
  • Error: ECG from legacy TraceMaster system cannot be printed
  • Error: ECG lead system not supported on cardiograph
  • Failed to implement custom settings
  • Failed to implement custom settings on startup
  • Failed to initialize configuration; factory default init required
  • Fatal Application Error on startup
  • Fax transmission did not complete
  • Flash memory self-test fails
  • Flash memory test fails
  • FPGA communication incorrectly configured
  • FPGA is not connected
  • Front/Back battery overcharged warning
  • General wireless connectivity issues
  • ID or ECG button on keyboard is unresponsive
  • Incorrect language displayed on cardiograph
  • Initial battery charge before first use
  • Install replacement batteries
  • Installing or replacing batteries with mismatched part numbers
  • Insufficient internal storage for new orders
  • Insufficient internal storage to download orders
  • Insufficient storage available message on USB memory stick transfer
  • Insufficient storage on USB memory stick
  • Internal Compact Flash self-test fails
  • Internal Compact Flash test fails
  • Internal CompactFlash card cannot be accessed
  • Internal function error message
  • Invalid database message
  • Invalid Default Gateway
  • Invalid default TraceMaster connection for OrderVue inbox
  • Invalid directory
  • Invalid IP Address
  • Invalid Primary DNS
  • Invalid Primary WINS
  • Invalid Report Handle message
  • Invalid Subnet Mask
  • Invalid transfer destination selected
  • Invalid XML schema version message
  • Keyboard key not recognized
  • Keyboard key recognition failure
  • Keyboard test incomplete or failed
  • LCD display flickers or shows corruption
  • Lead wire cable insulation damage
  • Lead wire cable insulation damaged
  • Lead wire intermittent ECG signal loss or noise
  • Lead wire performance test reminder icon displayed
  • Lead wire shortage detected
  • Lead wire shortage suspected
  • Lead wire shortages suspected
  • Low battery charge during operation
  • Low battery warning; printing disabled
  • Maximum number of ECGs stored message
  • Maximum number of ECGs stored on CF card
  • Message Busy appears after sending print job
  • Message USB appears when attempting to print rhythm strip
  • Missing network settings information
  • More than 200 ECGs on USB memory stick
  • New battery requires charging before first use
  • No default TraceMaster server selected
  • No order source selected
  • No Report message
  • No TraceMaster connections configured
  • No TraceMaster server to delete
  • Operating temperature outside specification
  • Order error - XML error
  • Orders cannot be downloaded - previously archived with ECGs
  • Orders number reached limit of 200
  • Patient leakage current exceeds limits
  • Patient leakage current with mains voltage on applied parts exceeds limits
  • Preparation for Philips service return
  • Print Error: Unknown Error
  • Printed report looks too small or compressed
  • Printed report too small or compressed
  • Printer door is open
  • Printer temperature too high
  • Printer test failed after passing display message
  • Printer test page has character distortion or illegibility
  • Printer test page has diagonal line distortion
  • Printer test page has distorted or missing segments
  • Printer test page has vertical line spacing error
  • Printer test page print quality errors
  • Printer underflow
  • Printer voltage high
  • Printer voltage low
  • Printing disabled with one battery only (PageWriter TC70)
  • Remote Query could not be completed
  • Remote transfer schema error
  • Required patient information missing from ECG
  • Safety Test S1: Earth resistance test
  • Safety Test S2: Equipment leakage current
  • Safety Test S3: Leads leakage current (CF applied parts)
  • Selected ECG exceeds available Archive space on USB Memory Stick
  • Selected ECG exceeds available TraceMaster space
  • Selected ECG(s) exceeds available space on Remote site message
  • Selected ECGs exceed available Archive space on USB
  • Some ECGs retrieved from ECG management display differently when printed
  • TC30 cardiograph cannot operate with mismatched battery part numbers
  • TC30 cardiograph does not operate with two installed batteries
  • TC30 cardiograph with mismatched battery part numbers
  • TC50 cardiograph does not operate with two installed batteries
  • TC50 cardiograph with mismatched battery part numbers
  • TC50 or TC70 cardiograph cannot operate with mismatched battery part numbers
  • TC70 cardiograph with single battery installed
  • There is no default server available for Auto Time Sync message
  • Timed ECG number exceeds available number in Main Archive
  • Touch screen cleaning required
  • Touch screen performs poorly in different usage settings
  • Touch screen requires calibration
  • Touch screen requires cleaning
  • Touch screen requires multiple attempts to select items
  • Touch screen responsiveness issues
  • TraceMaster internal error
  • TraceMaster operation canceled
  • TraceMaster operation cannot be completed
  • Two batteries with different part numbers installed
  • Unapproved battery warning
  • Unexpected reply received from Remote Site message
  • Unexpected reply received from server
  • Unknown error accessing storage capacity of selected archive
  • Unrecognized battery warning
  • Unsupported batteries detected after software upgrade
  • Unsupported XML version message
  • USB memory stick file system limitation with over 200 ECGs
  • USB write operations fail occasionally
  • Verification: ECG trace integrity from patient leads
  • When error occurs during multiple ECG transfer, process does not continue and transferred files not deleted
  • Worklist full - not all orders can be downloaded
  • XML error message
  • XML unicode encoding is not UTF-16 message

186 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
Archive function could not be completed messageArchive operation failed and cannot proceedArchive error occurred during archive processerror
Archive process terminated due to error messageArchive operation was stopped due to an errorAn error message occurred during archive process executionerror
Bad reply received from TraceMaster messageInvalid response received from TraceMaster serverArchive error with bad reply from servererror
Battery cycle count > 300Battery has exceeded 300 charge cycles; degraded capacity, replacement advised soonBattery usage monitoring detects > 300 complete charge cycleswarning
Battery cycle count > 400Battery has exceeded 400 charge cycles; critical degradation, immediate replacement requiredBattery usage monitoring detects > 400 complete charge cyclescritical
Battery mismatchTwo batteries with different part numbers installed, charging suspendedCardiograph detects two batteries with different Philips part numbers installederror
Battery overchargedFront/Back battery is overcharged; charger or battery faultBattery reaches overcharge condition during chargingwarning
Battery part number mismatchTwo installed batteries have different part numbers and are incompatibleTwo batteries with mismatched part numbers detectederror
Battery SOH <= 75%State of Health (SOH) degraded to 75% of design capacity; critical degradationBattery full charge capacity falls to 75% or below design capacitycritical
Battery SOH <= 80%Battery state of health has degraded to 80% of design capacityBattery full charge capacity reaches 80% of design capacitywarning
Battery temperature too high. Shut down in...Battery temperature reached or exceeded 60°C (140°F); auto-shutdown imminentBattery reaches or exceeds 60°C during operationcritical
Cannot access internal Compact Flash memoryCompact Flash memory card is inaccessible or corruptedAttempt to read or write to internal Compact Flash storageerror
Command canceledUser canceled the operation or provided invalid information to commandUser cancels operation or system detects invalid input parametersinfo
Command cannot be completedTraceMaster operation failed; system is currently busy processing another requestAttempt to execute TraceMaster command while system is processing requestswarning
Duplicate Profile name messageCannot save exam profile with duplicate nameAttempting to save exam profile under an existing namewarning
E-139Signal Acquisition Issue - Leads Off DisplayOne or more leads display red dashed line indicating loss of signalerror
E-140Signal Acquisition Issue - Low Amplitude LeadsOne or more leads display with reduced signal amplitudewarning
E-142aMain Screen Issue - Incorrect Language DisplayCardiograph software displays text in wrong language after main board replacementwarning
E-142bMain Screen Issue - Waveform Display FrozenECG waveform display stops updating or entire system becomes unresponsiveerror
E-142cReport Handle Error - Invalid ReportSystem displays Invalid Report Handle message when previewing, printing, or saving ECGserror
E-143aArchive Issue - No Report FoundSystem displays No Report message when accessing archived ECGerror
E-143bArchive Issue - Internal Function ErrorSystem displays Internal function error message when accessing archiveerror
E-143cArchive Issue - Command Timed OutSystem displays Command timed-out message during ECG archive operationswarning
E-143dArchive Issue - Incomplete Patient InformationSystem displays ECGs with incomplete required patient information cannot be transferrederror
E-143eArchive Issue - Remote Site Disk Space ExceededSystem displays Selected ECG(s) exceed available space on Remote site messageerror
E-144aArchive Issue - USB Memory Stick Storage FullSystem displays Insufficient storage available message when transferring to USB memory stickwarning
E-144bArchive Issue - Invalid DatabaseSystem displays Invalid database message when accessing archiveerror
E-144cArchive Issue - Multiple ECG Transfer ErrorError occurs during transfer of multiple ECGs; transfer stops and transferred files not deletedwarning
E-144dArchive Issue - PDF Export FailedSystem displays error when trying to transmit PDF ECGs to networkerror
E-144eArchive Performance - Slow Archive List DisplayArchive list takes longer than 40 seconds to display when Main Archive contains 130+ ECGswarning
E-145aArchive Error - Function Not CompletedSystem displays Archive function could not be completed. Please restart cardiograph messageerror
E-145bArchive Error - Process TerminatedSystem displays Archive process was terminated due to an error messageerror
E-146aArchive Error - Requested Function FailedSystem displays Requested Function could not be completed messageerror
E-146bArchive Error - USB Transfer Failed at Less Than 200 FilesECG transfer to USB memory stick fails despite fewer than 200 files on mediaerror
E-146cArchive Error - No Remote Site ReplySystem displays No reply received from Remote site message during transfererror
E-147aArchive Error - Bad Remote Site ReplySystem displays Bad reply received from Remote site messageerror
E-147bArchive Error - Unexpected Remote Site ReplySystem displays Unexpected reply received from Remote Site messageerror
E-147cXML Error - Invalid Schema VersionSystem displays Invalid XML schema version message during transfererror
E-147dXML Error - Unsupported XML VersionSystem displays Unsupported XML version message when attempting transferwarning
E-147eXML Error - File CorruptedSystem displays XML error message during ECG transfererror
E-148aXML Error - Invalid Unicode EncodingSystem displays XML unicode encoding is not UTF-16 messageerror
E-148bXML Error - File Access FailedSystem displays Error accessing external XML file messageerror
E-148cXML Error - File Too LargeSystem displays Error XML file is too large messageerror
E-148dXML Error - Transform FailedSystem displays Error XML file did not transform messageerror
E-148eArchive Error - No Time Sync Server ConfiguredSystem displays There is no default server available for Auto Time Sync messagewarning
E-149aArchive Error - Delete FailedSystem displays Error occurred when deleting selected ECGs messageerror
E-149bArchive Error - Transfer FailedSystem displays Error occurred when transferring selected ECGs messageerror
ECG lead system not supportedLead system configuration in ECG file is not supported by cardiographAttempt to display or print ECG with unsupported lead configurationerror
ECG XML version not supported by TraceMasterECG file XML version is incompatible with configured TraceMaster server versionAttempt to transfer or access ECG with unsupported XML version on TraceMastererror
Error accessing external XML file messageCannot access or read XML fileFile I/O error during XML file accesserror
Error occurred when deleting selected ECGs messageFailed to delete ECGs from storageInternal Compact Flash card read/write error during delete operationerror
Error occurred when transferring selected ECGs messageECG transfer operation failedTransmission error during ECG transfererror
Error XML file did not transform messageXML file transformation failedFile I/O error during XML transformationerror
Error XML file is too large messageXML file exceeds size limitXML file size exceeds system capacityerror
Failed to implement custom settings messageCorrupted configuration prevented loading custom settingsCorrupted configuration file detected when loadingerror
Fax transmission did not complete messageFax transmission operation failed to completeFax transmission interrupted or failederror
Insufficient storage for new ordersInternal cardiograph storage is full; cannot save additional ordersAttempt to save or download new orders to full storageerror
Insufficient storage to download ordersInternal cardiograph storage is full; cannot download ordersAttempt to download orders from TraceMaster to full storageerror
Internal CompactFlash card inaccessibleInternal Compact Flash card cannot be accessed and requires replacementAttempt to access internal Compact Flash storage for ECG datacritical
Internal errorInternal application error occurred; orders list failed to refreshAttempt to refresh or access orders listerror
Invalid TraceMaster for OrderVue inboxDefault TraceMaster server for OrderVue inbox is incorrect or unavailableAttempt to use OrderVue inbox with incorrectly configured TraceMaster servererror
Invalid transfer destination selectedSelected TraceMaster server is invalid or unavailable for transferAttempt to transfer ECG to an invalid or unreachable TraceMaster servererror
Invalid XML file detected messageConfiguration file is not valid XMLInvalid configuration file detected when loadingerror
Invalid XML platform detected messageConfiguration file contains invalid platform specificationInvalid platform detected when loading configuration fileerror
Invalid XML schema version messageXML schema version is not supportedArchive error with unsupported XML schemaerror
Invalid XML version detected messageConfiguration file contains invalid XML versionInvalid version detected when loading configuration fileerror
Legacy TraceMaster ECG cannot printECG from legacy TraceMaster system version is incompatible with cardiographAttempt to print ECG that originated from legacy TraceMaster systemerror
Missing network settings informationRequired network configuration parameters are incomplete for TraceMaster setupAttempt to save or apply TraceMaster configuration with incomplete network infoerror
No default TraceMaster server selectedDefault TraceMaster server has not been designated in configurationAttempt to use TraceMaster services when no default server is seterror
No order source selectedOrder retrieval attempted without specifying source server or locationAttempt to retrieve orders without selecting source TraceMaster serverinfo
No reply received from TraceMaster messageNo response from TraceMaster serverRemote site communication timeout or server unavailableerror
No TraceMaster connections configuredNo TraceMaster server connections have been set up in cardiograph configurationAttempt to transfer ECG or access TraceMaster services without configured servererror
No TraceMaster server to deleteDelete command executed without first selecting a TraceMaster serverAttempt to delete TraceMaster server configuration without selecting oneinfo
Order error: Function failedOrder application function returned error; requested operation failedExecution of order operation resulted in function failureerror
Order error: XML errorOrder contains malformed or invalid XML; cannot be processedParsing order data with invalid or malformed XML structureerror
Orders limit 200 reachedOrders list has reached maximum capacity of 200 orders; cannot append all new ordersAttempt to append new orders when list at 200-order limitwarning
Orders previously archived with ECGsRequested orders cannot be downloaded because they are already archived with ECGsAttempt to download orders that have existing associated ECGs in Main Archivewarning
Over TemperatureBattery temperature has exceeded safe operating rangeBattery reaches 50–59°C or ≥60°Cerror
Password is missing messageConfiguration screen password protection requires password entryAccessing password-protected Configuration screen without entering passwordwarning
Password is not confirmed messageConfiguration password was not confirmedEnabling configuration password without confirming itwarning
Patient ID field not enabledPatient ID field is disabled; cannot transfer ECGs without this required fieldAttempt to transfer ECG to TraceMaster without Patient ID presentwarning
PIM Test FailedPIM (Patient Interface Module) connectivity or communication test failedPIM patient data cable disconnected or not properly attached; persistent cable connection issueerror
Please select a TraceMaster server as the default connectionNo default TraceMaster server selectedAttempting to set default TraceMaster server without first selecting itwarning
Recommend CalibrationBattery level indicator accuracy affectedBattery power indicators not functioning usefully in daily work environmentwarning
Recommend ReplacementBatteries have reached end of useful lifeAppears in Battery Status window after calibration procedure completionerror
Required patient information missingECG transfer failed due to incomplete required patient informationAttempt to transfer ECG to TraceMaster without all required patient dataerror
Selected ECG exceeds USB archive spaceECG file is too large to fit in available USB memory stick storageAttempt to archive or save ECG to USB memory stickerror
Server connection resetNetwork connection to TraceMaster server was unexpectedly closed during operationServer closed connection during orders or ECG transfer operationerror
The cardiograph must restart to display simulated waveform dCardiograph restart required to enable simulation modeSimulation mode turned on; device requires restart to display simulated waveformsinfo
The cardiograph must restart to return to normal operating mCardiograph restart required to exit simulation modeSimulation mode turned off; device requires restart to return to normal operationinfo
The Front/Back Battery is overchargedBattery has been overcharged by a cardiograph running older softwareBattery charged by cardiograph without software version A.07.07warning
The maximum number of fax profiles has been configured messaCannot add more fax profiles; limit reachedAttempting to add fax profile when maximum is configuredwarning
The maximum number of OrderVue inboxes has been created messCannot create more OrderVue inboxes; limit reachedAttempting to create OrderVue inbox when maximum is configuredwarning
The maximum number of TraceMaster connections has been confiCannot add more TraceMaster connections; limit reachedAttempting to add TraceMaster connection when maximum is configuredwarning
There are more than 200 ECGs on USB stickUSB memory stick contains more than maximum supported ECG filesAttempt to display or manage ECG files from USB stick with >200 ECGswarning
There is no default TraceMaster server available for Auto TiNo default TraceMaster server configured for automatic time synchronizationTraceMaster settings not configured for Auto Time Syncwarning
Timed ECG exceeds Main Archive limitNew timed ECG cannot be saved as Main Archive has reached capacity limitAttempt to save a timed ECG to Main Archiveerror
Unapproved batteryBattery is not approved for use despite meeting Philips specificationsUnapproved battery detected at power-up, unlock, user login, or exit from sleep modewarning
Unexpected reply received from TraceMaster messageUnexpected response format received from TraceMaster serverArchive error with unexpected reply from servererror
Unknown error accessing archive storage capacityError occurred while querying storage capacity of archive mediaAttempt to check storage space on USB stick or internal Compact Flasherror
Unrecognized batteryInstalled battery is not supported and could damage the cardiographUnsupported battery detected at power-upcritical
Unsaved changes will be deleted messageConfiguration changes not saved and will be lostExiting configuration screens without saving changeswarning
Unsupported XML version messageXML version of selected ECG is not supported by TraceMaster serverXML schema of selected ECG not compatible with TraceMasterwarning
USB memory stick not accessibleUSB memory stick not recognized or not fully inserted into connectorAttempt to access USB memory stick for reading or writing ECGserror
W-149cArchive Display Difference - Non-PageWriter Source ECGsSome ECGs retrieved from ECG management print with visual differencesinfo
Worklist full - orders cannot downloadWorklist has reached maximum capacity; cannot download additional ordersAttempt to download orders when Worklist at full capacity limiterror
XML error messageXML ECG file is corruptedFile corruption detected during XML processingerror
XML unicode encoding is not UTF-16 messageXML file has invalid character encodingXML file encoding is not UTF-16error

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.