Philips PageWriter TC: troubleshooting and error codes
The ServiceForge knowledge base holds 246 documented symptoms for the Philips PageWriter TC, with 328 ranked causes, 363 diagnostic checks with test points and expected values, and 108 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
One or more leads show leads off periodically patient safety
Red dashed line appears on lead display intermittently during signal acquisition
Ranked causes
- 1. Improper or incomplete patient and skin preparation (common)
- 2. Improperly attached electrodes or lead wires; expired disposable gel electrodes (common)
- 3. Defib or high noise event (occasional)
- 4. PIM reset (up to ten second duration) (common)
- 5. PIM communications lost due to USB patient data cable failure or defective leads (occasional)
- 6. Specific electrode or cable failure (TC20 models) (occasional)
Review patient preparation documentation
- Test point
- Patient and skin preparation area
- Expected
- Good patient preparation and skin preparation techniques followed per section 3-10
9 more checks with pass and fail actions, 14 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer Error: Out of Paper / Paper Jam / Door Open
Error message displayed for out of paper, paper jam, or door open; may appear despite correct paper loading and closed door
Ranked causes
- 1. Paper drawer not fully closed (common)
- 2. Out of paper or incorrectly loaded paper (common)
- 3. Incompatible paper loaded (occasional)
- 4. Defective paper tray assembly (rare)
- 5. Top-of-form sensor or door closed sensor failure (rare)
Verify paper drawer is fully closed
- Test point
- Drawer latch
- Expected
- Drawer latches securely in fully closed position
4 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer error: Out of Paper, Paper Jam, or Door Open messages caution
Printer displays error messages for out of paper, paper jam, or door open when actual condition may not exist.
Ranked causes
- 1. Paper drawer not fully closed (common)
- 2. Out of paper or incorrectly loaded (common)
- 3. Incompatible paper loaded (occasional)
- 4. Defective paper tray assembly (occasional)
- 5. Top-of-form or door closed sensor fault (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Screen is dark and shows no image
Display remains blank or dark; no visual output; backlight may be off
Ranked causes
- 1. Cardiograph in Standby mode (common)
- 2. Battery capacity insufficient (below 1080 mAh) (common)
- 3. AC power supply failure (occasional)
- 4. Battery insertion or charge issue (occasional)
- 5. File system corrupted (rare)
Check if cardiograph is in Standby mode
- Test point
- Display state and power indicator
- Expected
- Display should show active screen when not in Standby
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery cycle count exceeded patient safety
Battery has exceeded charge cycle limit (300 or 400); message indicates immediate or soon replacement required; battery continues to charge but degraded
Ranked causes
- 1. Battery exceeded 300 charge cycles (common)
- 2. Battery exceeded 400 charge cycles (common)
- 3. Battery state of health (SOH) <= 80% (common)
- 4. Battery state of health (SOH) <= 75% (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
One or more leads show low amplitude patient safety
ECG waveform appears with reduced signal strength or height on affected leads
Ranked causes
- 1. Improper or incomplete patient and skin preparation (common)
- 2. Lead placement not standard (common)
- 3. Gooey adapters; electrodes or lead wires not properly attached (common)
- 4. Faulty lead wires (occasional)
Verify patient preparation technique
- Test point
- Patient skin preparation area
- Expected
- Proper skin prep completed per Instructions for Use
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM not detected message or disconnected icon appears patient safety
PIM not detected message displays or PIM disconnected icon appears on screen, indicating communication loss between cardiograph and PIM.
Ranked causes
- 1. PIM not connected to cardiograph (common)
- 2. Faulty patient data cable (common)
- 3. Faulty PIM (occasional)
- 4. Class A PIM is connected (rare)
Verify Class B PIM is connected
- Test point
- PIM model identification
- Expected
- Class B PIM attached
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM not detected or PIM disconnected icon appears patient safety
Patient Interface Module (PIM) is not recognized by cardiograph; disconnect icon visible on screen
Ranked causes
- 1. PIM not connected (common)
- 2. Faulty patient data cable (common)
- 3. Faulty PIM (occasional)
- 4. Class A PIM is connected to the cardiograph (rare)
Verify Class B PIM is connected and cable is firmly seated
- Test point
- PIM connector on rear panel
- Expected
- PIM firmly connected
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM reset with temporary leads off display
All leads display as red dashed lines for up to ten seconds and PIM disconnected icon appears during normal PIM reset operation.
Ranked causes
- 1. Normal PIM reset during power on/wakeup (common)
- 2. PIM reset during screen changes (common)
- 3. PIM reset during USB device insertion or removal (common)
- 4. PIM reset during electrostatic discharge (occasional)
Verify cardiograph is powering on or waking up
- Test point
- System power state
- Expected
- Power on or wakeup event in progress
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
All leads show leads off (red dashed line) continuously patient safety
No ECG signal detected on any lead; red dashed lines displayed on screen instead of waveforms
Ranked causes
- 1. PIM communications lost due to USB patient data cable connection failure (common)
- 2. Defective lead wires connected to PIM (common)
- 3. Class A PIM is connected to the cardiograph (rare)
Verify RL/N lead and electrode firmly attached
- Test point
- RL/N lead connection to patient
- Expected
- Firmly attached
4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
All leads show leads off continuously patient safety
All leads display as red dashed lines continuously on the cardiograph screen, indicating no signal acquisition.
Ranked causes
- 1. USB patient data cable connection failure (common)
- 2. Defective leads connected to the PIM (common)
- 3. Class A PIM is connected to the cardiograph (occasional)
Check if PIM disconnected icon appears on Status Bar
- Test point
- Status Bar top right of screen
- Expected
- PIM disconnected icon not visible
5 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery ID not approved or not recognized caution
Battery ID on About the Cardiograph screen does not match approved list; risk of battery overheating, shortened life, or cardiograph damage
Ranked causes
- 1. Unapproved battery installed (unknown)
- 2. Unsupported battery installed (unknown)
- 3. Legacy battery PN 989803160981 installed (occasional)
Verify battery ID matches approved list
- Test point
- About the Cardiograph screen
- Expected
- P508; P510; 202CJC; E013RJ; T508; or PH
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard is typing incorrect characters
Keyboard output does not match keys pressed; characters may be shifted or wrong language selected
Ranked causes
- 1. Caps Lock feature is enabled (common)
- 2. Incorrect keyboard option installed (rare)
- 3. Main board recently replaced; keyboard setting reset to incorrect option (occasional)
Check Caps Lock status
- Test point
- Status Bar on display
- Expected
- Cap Lock icon absent or disabled
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not advance
Paper fails to feed or advance through printer mechanism
Ranked causes
- 1. Application error (occasional)
- 2. Top-of-form sensor failure (rare)
- 3. Defective print head assembly (rare)
Restart cardiograph
- Test point
- Printer paper feed mechanism
- Expected
- Paper advances normally on next print job
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Ping test failed
Network connectivity test fails or returns error
Ranked causes
- 1. LAN cable not securely attached (common)
- 2. Wireless connection not associated or signal lost (occasional)
- 3. Invalid IP address entered (occasional)
Verify LAN cable is securely attached or wireless connection is active
- Test point
- LAN connector on rear or wireless Status Bar indicator
- Expected
- LAN cable fully seated OR green bars on Status Bar indicating live wireless connection
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote Query get ECG error
TraceMaster returned wrong ECG file; server or configuration issue
Ranked causes
- 1. TraceMaster configuration mismatch (common)
- 2. LAN/WLAN connection issue (common)
- 3. TraceMaster server offline (occasional)
Check TraceMaster configuration
- Test point
- TraceMaster settings
- Expected
- Settings correct for server
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Requested Function could not be completed
Message indicates the requested function could not be completed due to network, modem, or media issues.
Ranked causes
- 1. Network or modem phone line interrupted during operation (common)
- 2. USB memory stick not fully inserted (common)
- 3. Attempt to transfer unsupported ECG report type (occasional)
Retry operation
- Test point
- Network connectivity
- Expected
- Operation completes successfully
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Requested Function could not be completed message
Archive or transfer operation failed to complete
Ranked causes
- 1. Network interrupted during operation (occasional)
- 2. USB memory stick not fully inserted (common)
- 3. Attempt to transfer unsupported ECG report type (occasional)
Check network connection
- Test point
- Network cable and connectivity
- Expected
- Network cable firmly connected and active
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unrecognized battery or battery mismatch patient safety
Battery status message indicates unsupported battery, different part numbers, or charging suspension; cardiograph will not operate or charging halted
Ranked causes
- 1. One or more installed batteries are not supported or not Philips-approved (common)
- 2. Two batteries with different part numbers installed (common)
- 3. Battery overcharged or non-Philips battery installed (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB memory stick not accessible
System cannot access USB memory stick. Message indicates stick not fully inserted or contains unsupported software.
Ranked causes
- 1. USB stick not fully inserted into USB connector (common)
- 2. USB stick contains unsupported software (occasional)
- 3. Defective or incompatible USB memory stick (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB memory stick not accessible or not fully inserted
Message displayed: 'The USB memory stick is not accessible. Please ensure that the USB memory stick is fully inserted into the USB connector (right side of cardiograph).'
Ranked causes
- 1. USB memory stick not fully inserted into slot (common)
- 2. USB memory stick contains unsupported software (occasional)
- 3. USB memory stick not fully functioning or incompatible (occasional)
Verify USB memory stick insertion
- Test point
- USB connector on right side of cardiograph
- Expected
- USB memory stick fully inserted and recognized
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
A single key press produces repeated characters on screen
Keyboard input malfunction; one key press results in multiple character repetitions displayed
Ranked causes
- 1. Failure of keyboard key (common)
- 2. PIM patient data cable failure (occasional)
Perform Keyboard/HW Key Test
- Test point
- Keyboard hardware
- Expected
- Test passes
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery over temperature patient safety
Battery has reached excessive temperature; display shows temperature warning and automatic shutdown imminent, or charging suspended pending cool-down
Ranked causes
- 1. Battery temperature reached 50°C or over temperature detected (common)
- 2. Battery temperature reached or exceeded 60°C (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery over temperature warning caution
Temperature warning displayed; cardiograph will shut down
Ranked causes
- 1. Battery temperature between 50°C and 59°C (occasional)
- 2. Battery temperature at or above 60°C (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery performance degradation
Battery shows reduced runtime, slow charging, or capacity loss
Ranked causes
- 1. Storage temperature outside recommended range (common)
- 2. Operating temperature outside specification (occasional)
Verify battery storage temperature
- Test point
- Storage area
- Expected
- 25°C +/- 5°C (77°F +/- 10°F), +/- 5°C
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery remains in Standby instead of automatic shutdown
Cardiograph stays in Standby mode and does not proceed to automatic shutdown when conditions are met.
Ranked causes
- 1. Battery cycle count 600 or less with AC power connected (occasional)
- 2. Battery full charge capacity is 50% or greater with AC power connected (occasional)
Check battery cycle count
- Test point
- Battery status or settings menu
- Expected
- Cycle count 600 or less
1 more check with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Calibration: Battery status accuracy
Battery status information on screen may be inaccurate; Expected Max Error (%) and Full Capacity (mAh) values need verification after battery depletion cycle.
Ranked causes
- 1. Battery calibration procedure required (common)
- 2. Expected Max Error field did not reset to 2% (occasional)
Verify Expected Max Error (%) and Full Capacity (mAh) values changed from printed report values
- Test point
- Battery Status column on Service Utility screen
- Expected
- Values different than those printed on previous report
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: PIM communication verification
Patient Interface Module (PIM) not communicating with cardiograph or PIM error messages displayed. Test confirms PIM is communicating properly.
Ranked causes
- 1. PIM patient data cable loose or disconnected (common)
- 2. PIM hardware or patient data cable fault (occasional)
Verify PIM patient data cable securely attached
- Test point
- PIM connector on rear of cardiograph
- Expected
- Cable firmly seated; PIM disconnected icon not visible on Status Bar
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG signal loss, noise, or reduced amplitude on leads patient safety
Intermittent ECG signal dropout, increased noise, or weaker than normal signal strength on one or more leads
Ranked causes
- 1. Lead wire shortage (common)
- 2. Cracked or broken cable insulation on lead wire (occasional)
Run Lead Wire Performance Test to identify shortages
- Test point
- All lead wires
- Expected
- No shortages reported
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect language appears on cardiograph software
Software displays text in wrong language after main board replacement
Ranked causes
- 1. Main board recently replaced; software localization reset to wrong setting (common)
- 2. Persistent language setting issue after software reinstall (rare)
Verify main board was recently replaced
- Test point
- Service records and timeline
- Expected
- Main board replacement documented within last service call
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard is unresponsive
Keyboard does not respond to user input; may be caused by input focus loss or controller subsystem failure
Ranked causes
- 1. Input focus loss (common)
- 2. Keyboard controller subsystem has failed (rare)
Touch screen to restore input focus
- Test point
- Touch screen
- Expected
- Keyboard responds to input
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No matching ECG found on server
Patient ID incorrect or ECG does not exist on selected server
Ranked causes
- 1. Incorrect Patient ID (common)
- 2. ECG does not exist on server (occasional)
Verify Patient ID
- Test point
- Patient ID field
- Expected
- Correct Patient ID entered
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No reply received from Remote site message
Archive transfer fails because remote server does not respond
Ranked causes
- 1. Server is unavailable (occasional)
- 2. Configuration error in server address or URL (occasional)
Verify server availability
- Test point
- Remote server status
- Expected
- Server accessible and responding
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No reply received from server
LAN connection broken during transfer or server offline
Ranked causes
- 1. LAN connection broken during transfer (occasional)
- 2. TraceMaster server offline (occasional)
Verify LAN connection
- Test point
- LAN link
- Expected
- Connection active
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Order error - requested function failed
Message displayed: 'Order Error: Requested function failed message'. Order application error.
Ranked causes
- 1. Order application function error (occasional)
- 2. Persistent order application error (rare)
Restart the cardiograph
- Test point
- System boot sequence
- Expected
- Cardiograph restarts successfully
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not tear cleanly or stops at wrong spot caution
Paper tear mechanism fails or paper stops at unintended position during tear-off.
Ranked causes
- 1. Wrong paper size setting (common)
- 2. Wrong tear-off method (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper does not tear off cleanly or stops at wrong spot
Printed paper does not tear cleanly at expected location or stops tearing at incorrect position
Ranked causes
- 1. Wrong paper size setting configured (common)
- 2. Wrong method used to tear off paper (occasional)
Verify paper size setting
- Test point
- Cardiograph configuration
- Expected
- Paper size setting matches loaded paper size
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Patient ID field must be enabled to transfer ECGs to TraceMaster
Message displayed: 'The Patient ID field must be enabled in order to transfer ECGs to TraceMaster. Touch the [ON] button underneath the Enabled column to activate the Patient ID field.' ECG transfer attempted without Patient ID.
Ranked causes
- 1. Patient ID field not enabled (common)
- 2. No Patient ID entered for ECG (common)
Enable Patient ID field
- Test point
- Setup menu - Patient ID section
- Expected
- Patient ID field set to ON
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM Test failed
Error message appears when attempting to run tests
Ranked causes
- 1. PIM patient data cable not securely attached (common)
- 2. Cable or connector damaged (occasional)
Verify PIM patient data cable is securely attached
- Test point
- PIM connector on rear of cardiograph
- Expected
- Cable fully seated with no gaps
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PIM test failed or PIM communication error patient safety
PIM error messages displayed or cardiograph unable to acquire data from PIM; PIM disconnected icon appears on status bar.
Ranked causes
- 1. PIM patient data cable loose or disconnected (common)
- 2. PIM or PIM patient data cable defective (occasional)
Verify PIM patient data cable connection
- Test point
- PIM connector on rear of cardiograph
- Expected
- Cable securely attached; PIM disconnected icon not present on status bar
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not detected
Message appears indicating printer is not detected by the system
Ranked causes
- 1. Application error (occasional)
- 2. Defective print head assembly (rare)
Restart cardiograph and check printer detection
- Test point
- Printer interface
- Expected
- Printer detected and operational
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not detected message caution
Printer hardware is not recognized by the system.
Ranked causes
- 1. Application error (common)
- 2. Defective print head assembly (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote Query get list error
TraceMaster returned wrong ECG list; configuration or network mismatch
Ranked causes
- 1. Incorrect TraceMaster configuration (common)
- 2. LAN/WLAN configuration error (common)
Verify TraceMaster configuration
- Test point
- TraceMaster settings
- Expected
- Configuration matches expected server
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transfer of ECGs to USB memory stick fails with less than 200 ECGs on media
ECG transfer fails even though fewer than 200 files exist on USB media
Ranked causes
- 1. USB memory stick file system limitation (occasional)
- 2. Not enough free space on USB stick to save selected ECGs (common)
Check available space on USB stick
- Test point
- USB memory stick free space
- Expected
- Free space greater than total size of ECGs to transfer
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unknown error accessing storage capacity with archive
Message displayed: 'An unknown error occurred when accessing the storage capacity of the selected archive message'. Abnormal error occurs when getting storage capacity of selected storage media.
Ranked causes
- 1. Storage capacity access error during archive operation (rare)
- 2. Persistent storage media error (rare)
Restart the cardiograph
- Test point
- System boot sequence
- Expected
- Cardiograph restarts successfully
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unknown error accessing USB memory stick storage capacity
System error occurs when querying available storage capacity on USB memory stick.
Ranked causes
- 1. Abnormal error getting storage capacity of selected media (rare)
- 2. Storage media abnormality (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Waveform display not updating; system unresponsive
ECG waveform freezes on display or entire system becomes unresponsive to user input
Ranked causes
- 1. Software is frozen (common)
- 2. Main board failure (occasional)
Attempt system restart
- Test point
- Power/reset control
- Expected
- System powers down and restarts successfully; waveform updates resume
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC power indicator light does not illuminate high voltage
Cardiograph is plugged into AC power but the green AC power indicator light does not illuminate, indicating no AC power is being supplied to the device
Ranked causes
- 1. AC fuse is open or blown (common)
Verify AC fuse is blown or open
- Test point
- AC fuse located directly below AC power connector
- Expected
- Fuse element intact or blown
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive folder displays incorrect or corrupted data
Archive screen shows wrong reports or messages; archive list cannot be accessed or displays errors
Ranked causes
- 1. Archive corrupted or needs reset to factory defaults (occasional)
Verify all archived reports transferred before reset
- Test point
- Archive folder contents
- Expected
- All critical reports backed up to remote site or USB
1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive function could not be completed message
Archive operation fails with error message
Ranked causes
- 1. Archive error (occasional)
Restart cardiograph
- Test point
- Power control
- Expected
- System restarts successfully and archive function available
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive list takes longer than 40 seconds to display with 130+ ECGs stored
System performance degradation when Main Archive contains near maximum ECG count
Ranked causes
- 1. System slowdown from full ECG database (occasional)
Count ECGs in Main Archive
- Test point
- Archive database status
- Expected
- ECG count below 130 to restore normal performance
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive process was terminated due to an error message
Archive operation terminated with error before completion
Ranked causes
- 1. Archive error (occasional)
Restart cardiograph
- Test point
- Power control
- Expected
- System restarts and archive function available
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Archive transfer interrupted with multiple ECGs
When an error occurs during transfer of multiple ECGs to a TraceMaster system, the transfer process does not continue and transferred files are not deleted
Ranked causes
- 1. Known application limitation in ECG transfer (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio output not functioning
No audible sound emitted during audio test or patient does not hear chime tones
Ranked causes
- 1. Speaker or audio amplifier defect (occasional)
Run Audio Test from Diagnostic and Test Utilities
- Test point
- Main screen > Setup > Service Utilities > Diagnostic and Test Utilities
- Expected
- Audible chime tones emitted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Audio test fails or no sound emitted
Cardiograph does not emit audible chime tones during audio diagnostic test
Ranked causes
- 1. Audio system defective or speaker failed (common)
Touch Audio button and listen for audible chime tones
- Test point
- speaker/audio system
- Expected
- Audible chime tones emitted
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Auto Shutdown not occurring as expected
Cardiograph Auto Shutdown feature does not trigger when configured, preventing automatic power-off
Ranked causes
- 1. Battery charge below 90% capacity (common)
Check battery charge level via display
- Test point
- Battery status display
- Expected
- Battery charge at or above 90% capacity
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Bad reply received from Remote site message
Archive transfer fails with invalid response from remote server
Ranked causes
- 1. Archive error; bad reply from server (occasional)
Restart cardiograph
- Test point
- Power control
- Expected
- System restarts successfully
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batteries not charging when cardiograph powered off
Batteries remain uncharged when cardiograph is in powered-off state despite AC power connection
Ranked causes
- 1. Cardiograph powered off during charging attempt (common)
Verify cardiograph power state
- Test point
- Cardiograph power switch or display
- Expected
- Cardiograph in Run or Standby mode
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery calibration procedure
Deplete batteries fully while ID screen is open, then recharge to recalibrate battery monitoring. Used after battery replacement or to reset battery parameters.
Ranked causes
- 1. Battery calibration required after replacement or parameter reset (common)
Verify AC power disconnected
- Test point
- AC power indicator light next to power button
- Expected
- Not lit
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery cycle count 600 or greater patient safety
Battery with 600+ cycles detected; batteries will not charge and should be replaced immediately
Ranked causes
- 1. Battery aged beyond safe operational limit (common)
Check battery cycle count in Battery Status window
- Test point
- Battery Status window
- Expected
- Cycle count less than 600
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Battery cycle count 600 or greater or state of health 50% or less
- Battery cycle count exceeds 300
- Battery cycle count exceeds 400
- Battery cycle count exceeds recommended limit
- Battery damage or leakage detected
- Battery damaged or leaking
- Battery disposal required
- Battery does not charge when cardiograph is shut down
- Battery left in unused cardiograph longer than 2 months
- Battery not charged to full capacity after shutdown with AC power connected
- Battery not charged when device powered off
- Battery not charging after cardiograph shutdown
- Battery not charging when device powered off
- Battery part number mismatch warning
- Battery state of health 50% or less
- Battery state of health at or below 75%
- Battery state of health at or below 80%
- Battery status shows Recommend Calibration
- Blank pages printed after ECG report
- Calibration: Battery level indicator accuracy
- Calibration: ECG simulation with 12-lead simulator
- Calibration: Patient Interface Module (PIM) Test
- Calibration: Performance verification after external repair
- Calibration: Performance verification after external repair with patient module access
- Calibration: Performance verification after internal repair
- Calibration: Performance verification after software upgrade
- Calibration: Printer test
- Calibration: Touch screen
- Calibration: Touch Screen Display
- Calibration: Touch screen force calibration
- Calibration: Touch screen touch calibration test
- Calibration: Verification of gain settings accuracy
- Calibration: Verification of paper speed accuracy
- Cannot access internal Compact Flash memory
- Cannot disconnect modem connection from TraceMaster Remote Site
- Cannot store additional ECGs message
- Cannot store additional ECGs on CF card
- Cardiograph does not restart
- Cardiograph not to be used for extended period
- Cardiograph return to manufacturer without battery removal
- Cardiograph unresponsive after rapid paper drawer cycling
- Check Printer message with no apparent problem
- Command timed-out message
- Connection to server was reset
- Copying ECGs from Archive to USB Memory Stick fails
- Copying ECGs to USB memory stick fails occasionally
- Earth resistance test
- ECG XML version not supported by TraceMaster
- ECGs with incomplete patient information cannot transfer
- ECGs with incomplete required patient information fields cannot be transferred message
- Equipment enclosure leakage current exceeds limits
- Error accessing external XML file message
- Error message displayed with mismatched battery part numbers
- Error occurred when deleting selected ECGs message
- Error occurred when transferring selected ECGs message
- Error occurred when trying to transmit PDF ECGs
- Error XML file did not transform message
- Error XML file is too large message
- Error: ECG from legacy TraceMaster system cannot be printed
- Error: ECG lead system not supported on cardiograph
- Failed to implement custom settings
- Failed to implement custom settings on startup
- Failed to initialize configuration; factory default init required
- Fatal Application Error on startup
- Fax transmission did not complete
- Flash memory self-test fails
- Flash memory test fails
- FPGA communication incorrectly configured
- FPGA is not connected
- Front/Back battery overcharged warning
- General wireless connectivity issues
- ID or ECG button on keyboard is unresponsive
- Incorrect language displayed on cardiograph
- Initial battery charge before first use
- Install replacement batteries
- Installing or replacing batteries with mismatched part numbers
- Insufficient internal storage for new orders
- Insufficient internal storage to download orders
- Insufficient storage available message on USB memory stick transfer
- Insufficient storage on USB memory stick
- Internal Compact Flash self-test fails
- Internal Compact Flash test fails
- Internal CompactFlash card cannot be accessed
- Internal function error message
- Invalid database message
- Invalid Default Gateway
- Invalid default TraceMaster connection for OrderVue inbox
- Invalid directory
- Invalid IP Address
- Invalid Primary DNS
- Invalid Primary WINS
- Invalid Report Handle message
- Invalid Subnet Mask
- Invalid transfer destination selected
- Invalid XML schema version message
- Keyboard key not recognized
- Keyboard key recognition failure
- Keyboard test incomplete or failed
- LCD display flickers or shows corruption
- Lead wire cable insulation damage
- Lead wire cable insulation damaged
- Lead wire intermittent ECG signal loss or noise
- Lead wire performance test reminder icon displayed
- Lead wire shortage detected
- Lead wire shortage suspected
- Lead wire shortages suspected
- Low battery charge during operation
- Low battery warning; printing disabled
- Maximum number of ECGs stored message
- Maximum number of ECGs stored on CF card
- Message Busy appears after sending print job
- Message USB appears when attempting to print rhythm strip
- Missing network settings information
- More than 200 ECGs on USB memory stick
- New battery requires charging before first use
- No default TraceMaster server selected
- No order source selected
- No Report message
- No TraceMaster connections configured
- No TraceMaster server to delete
- Operating temperature outside specification
- Order error - XML error
- Orders cannot be downloaded - previously archived with ECGs
- Orders number reached limit of 200
- Patient leakage current exceeds limits
- Patient leakage current with mains voltage on applied parts exceeds limits
- Preparation for Philips service return
- Print Error: Unknown Error
- Printed report looks too small or compressed
- Printed report too small or compressed
- Printer door is open
- Printer temperature too high
- Printer test failed after passing display message
- Printer test page has character distortion or illegibility
- Printer test page has diagonal line distortion
- Printer test page has distorted or missing segments
- Printer test page has vertical line spacing error
- Printer test page print quality errors
- Printer underflow
- Printer voltage high
- Printer voltage low
- Printing disabled with one battery only (PageWriter TC70)
- Remote Query could not be completed
- Remote transfer schema error
- Required patient information missing from ECG
- Safety Test S1: Earth resistance test
- Safety Test S2: Equipment leakage current
- Safety Test S3: Leads leakage current (CF applied parts)
- Selected ECG exceeds available Archive space on USB Memory Stick
- Selected ECG exceeds available TraceMaster space
- Selected ECG(s) exceeds available space on Remote site message
- Selected ECGs exceed available Archive space on USB
- Some ECGs retrieved from ECG management display differently when printed
- TC30 cardiograph cannot operate with mismatched battery part numbers
- TC30 cardiograph does not operate with two installed batteries
- TC30 cardiograph with mismatched battery part numbers
- TC50 cardiograph does not operate with two installed batteries
- TC50 cardiograph with mismatched battery part numbers
- TC50 or TC70 cardiograph cannot operate with mismatched battery part numbers
- TC70 cardiograph with single battery installed
- There is no default server available for Auto Time Sync message
- Timed ECG number exceeds available number in Main Archive
- Touch screen cleaning required
- Touch screen performs poorly in different usage settings
- Touch screen requires calibration
- Touch screen requires cleaning
- Touch screen requires multiple attempts to select items
- Touch screen responsiveness issues
- TraceMaster internal error
- TraceMaster operation canceled
- TraceMaster operation cannot be completed
- Two batteries with different part numbers installed
- Unapproved battery warning
- Unexpected reply received from Remote Site message
- Unexpected reply received from server
- Unknown error accessing storage capacity of selected archive
- Unrecognized battery warning
- Unsupported batteries detected after software upgrade
- Unsupported XML version message
- USB memory stick file system limitation with over 200 ECGs
- USB write operations fail occasionally
- Verification: ECG trace integrity from patient leads
- When error occurs during multiple ECG transfer, process does not continue and transferred files not deleted
- Worklist full - not all orders can be downloaded
- XML error message
- XML unicode encoding is not UTF-16 message
186 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Archive function could not be completed message | Archive operation failed and cannot proceed | Archive error occurred during archive process | error |
Archive process terminated due to error message | Archive operation was stopped due to an error | An error message occurred during archive process execution | error |
Bad reply received from TraceMaster message | Invalid response received from TraceMaster server | Archive error with bad reply from server | error |
Battery cycle count > 300 | Battery has exceeded 300 charge cycles; degraded capacity, replacement advised soon | Battery usage monitoring detects > 300 complete charge cycles | warning |
Battery cycle count > 400 | Battery has exceeded 400 charge cycles; critical degradation, immediate replacement required | Battery usage monitoring detects > 400 complete charge cycles | critical |
Battery mismatch | Two batteries with different part numbers installed, charging suspended | Cardiograph detects two batteries with different Philips part numbers installed | error |
Battery overcharged | Front/Back battery is overcharged; charger or battery fault | Battery reaches overcharge condition during charging | warning |
Battery part number mismatch | Two installed batteries have different part numbers and are incompatible | Two batteries with mismatched part numbers detected | error |
Battery SOH <= 75% | State of Health (SOH) degraded to 75% of design capacity; critical degradation | Battery full charge capacity falls to 75% or below design capacity | critical |
Battery SOH <= 80% | Battery state of health has degraded to 80% of design capacity | Battery full charge capacity reaches 80% of design capacity | warning |
Battery temperature too high. Shut down in... | Battery temperature reached or exceeded 60°C (140°F); auto-shutdown imminent | Battery reaches or exceeds 60°C during operation | critical |
Cannot access internal Compact Flash memory | Compact Flash memory card is inaccessible or corrupted | Attempt to read or write to internal Compact Flash storage | error |
Command canceled | User canceled the operation or provided invalid information to command | User cancels operation or system detects invalid input parameters | info |
Command cannot be completed | TraceMaster operation failed; system is currently busy processing another request | Attempt to execute TraceMaster command while system is processing requests | warning |
Duplicate Profile name message | Cannot save exam profile with duplicate name | Attempting to save exam profile under an existing name | warning |
E-139 | Signal Acquisition Issue - Leads Off Display | One or more leads display red dashed line indicating loss of signal | error |
E-140 | Signal Acquisition Issue - Low Amplitude Leads | One or more leads display with reduced signal amplitude | warning |
E-142a | Main Screen Issue - Incorrect Language Display | Cardiograph software displays text in wrong language after main board replacement | warning |
E-142b | Main Screen Issue - Waveform Display Frozen | ECG waveform display stops updating or entire system becomes unresponsive | error |
E-142c | Report Handle Error - Invalid Report | System displays Invalid Report Handle message when previewing, printing, or saving ECGs | error |
E-143a | Archive Issue - No Report Found | System displays No Report message when accessing archived ECG | error |
E-143b | Archive Issue - Internal Function Error | System displays Internal function error message when accessing archive | error |
E-143c | Archive Issue - Command Timed Out | System displays Command timed-out message during ECG archive operations | warning |
E-143d | Archive Issue - Incomplete Patient Information | System displays ECGs with incomplete required patient information cannot be transferred | error |
E-143e | Archive Issue - Remote Site Disk Space Exceeded | System displays Selected ECG(s) exceed available space on Remote site message | error |
E-144a | Archive Issue - USB Memory Stick Storage Full | System displays Insufficient storage available message when transferring to USB memory stick | warning |
E-144b | Archive Issue - Invalid Database | System displays Invalid database message when accessing archive | error |
E-144c | Archive Issue - Multiple ECG Transfer Error | Error occurs during transfer of multiple ECGs; transfer stops and transferred files not deleted | warning |
E-144d | Archive Issue - PDF Export Failed | System displays error when trying to transmit PDF ECGs to network | error |
E-144e | Archive Performance - Slow Archive List Display | Archive list takes longer than 40 seconds to display when Main Archive contains 130+ ECGs | warning |
E-145a | Archive Error - Function Not Completed | System displays Archive function could not be completed. Please restart cardiograph message | error |
E-145b | Archive Error - Process Terminated | System displays Archive process was terminated due to an error message | error |
E-146a | Archive Error - Requested Function Failed | System displays Requested Function could not be completed message | error |
E-146b | Archive Error - USB Transfer Failed at Less Than 200 Files | ECG transfer to USB memory stick fails despite fewer than 200 files on media | error |
E-146c | Archive Error - No Remote Site Reply | System displays No reply received from Remote site message during transfer | error |
E-147a | Archive Error - Bad Remote Site Reply | System displays Bad reply received from Remote site message | error |
E-147b | Archive Error - Unexpected Remote Site Reply | System displays Unexpected reply received from Remote Site message | error |
E-147c | XML Error - Invalid Schema Version | System displays Invalid XML schema version message during transfer | error |
E-147d | XML Error - Unsupported XML Version | System displays Unsupported XML version message when attempting transfer | warning |
E-147e | XML Error - File Corrupted | System displays XML error message during ECG transfer | error |
E-148a | XML Error - Invalid Unicode Encoding | System displays XML unicode encoding is not UTF-16 message | error |
E-148b | XML Error - File Access Failed | System displays Error accessing external XML file message | error |
E-148c | XML Error - File Too Large | System displays Error XML file is too large message | error |
E-148d | XML Error - Transform Failed | System displays Error XML file did not transform message | error |
E-148e | Archive Error - No Time Sync Server Configured | System displays There is no default server available for Auto Time Sync message | warning |
E-149a | Archive Error - Delete Failed | System displays Error occurred when deleting selected ECGs message | error |
E-149b | Archive Error - Transfer Failed | System displays Error occurred when transferring selected ECGs message | error |
ECG lead system not supported | Lead system configuration in ECG file is not supported by cardiograph | Attempt to display or print ECG with unsupported lead configuration | error |
ECG XML version not supported by TraceMaster | ECG file XML version is incompatible with configured TraceMaster server version | Attempt to transfer or access ECG with unsupported XML version on TraceMaster | error |
Error accessing external XML file message | Cannot access or read XML file | File I/O error during XML file access | error |
Error occurred when deleting selected ECGs message | Failed to delete ECGs from storage | Internal Compact Flash card read/write error during delete operation | error |
Error occurred when transferring selected ECGs message | ECG transfer operation failed | Transmission error during ECG transfer | error |
Error XML file did not transform message | XML file transformation failed | File I/O error during XML transformation | error |
Error XML file is too large message | XML file exceeds size limit | XML file size exceeds system capacity | error |
Failed to implement custom settings message | Corrupted configuration prevented loading custom settings | Corrupted configuration file detected when loading | error |
Fax transmission did not complete message | Fax transmission operation failed to complete | Fax transmission interrupted or failed | error |
Insufficient storage for new orders | Internal cardiograph storage is full; cannot save additional orders | Attempt to save or download new orders to full storage | error |
Insufficient storage to download orders | Internal cardiograph storage is full; cannot download orders | Attempt to download orders from TraceMaster to full storage | error |
Internal CompactFlash card inaccessible | Internal Compact Flash card cannot be accessed and requires replacement | Attempt to access internal Compact Flash storage for ECG data | critical |
Internal error | Internal application error occurred; orders list failed to refresh | Attempt to refresh or access orders list | error |
Invalid TraceMaster for OrderVue inbox | Default TraceMaster server for OrderVue inbox is incorrect or unavailable | Attempt to use OrderVue inbox with incorrectly configured TraceMaster server | error |
Invalid transfer destination selected | Selected TraceMaster server is invalid or unavailable for transfer | Attempt to transfer ECG to an invalid or unreachable TraceMaster server | error |
Invalid XML file detected message | Configuration file is not valid XML | Invalid configuration file detected when loading | error |
Invalid XML platform detected message | Configuration file contains invalid platform specification | Invalid platform detected when loading configuration file | error |
Invalid XML schema version message | XML schema version is not supported | Archive error with unsupported XML schema | error |
Invalid XML version detected message | Configuration file contains invalid XML version | Invalid version detected when loading configuration file | error |
Legacy TraceMaster ECG cannot print | ECG from legacy TraceMaster system version is incompatible with cardiograph | Attempt to print ECG that originated from legacy TraceMaster system | error |
Missing network settings information | Required network configuration parameters are incomplete for TraceMaster setup | Attempt to save or apply TraceMaster configuration with incomplete network info | error |
No default TraceMaster server selected | Default TraceMaster server has not been designated in configuration | Attempt to use TraceMaster services when no default server is set | error |
No order source selected | Order retrieval attempted without specifying source server or location | Attempt to retrieve orders without selecting source TraceMaster server | info |
No reply received from TraceMaster message | No response from TraceMaster server | Remote site communication timeout or server unavailable | error |
No TraceMaster connections configured | No TraceMaster server connections have been set up in cardiograph configuration | Attempt to transfer ECG or access TraceMaster services without configured server | error |
No TraceMaster server to delete | Delete command executed without first selecting a TraceMaster server | Attempt to delete TraceMaster server configuration without selecting one | info |
Order error: Function failed | Order application function returned error; requested operation failed | Execution of order operation resulted in function failure | error |
Order error: XML error | Order contains malformed or invalid XML; cannot be processed | Parsing order data with invalid or malformed XML structure | error |
Orders limit 200 reached | Orders list has reached maximum capacity of 200 orders; cannot append all new orders | Attempt to append new orders when list at 200-order limit | warning |
Orders previously archived with ECGs | Requested orders cannot be downloaded because they are already archived with ECGs | Attempt to download orders that have existing associated ECGs in Main Archive | warning |
Over Temperature | Battery temperature has exceeded safe operating range | Battery reaches 50–59°C or ≥60°C | error |
Password is missing message | Configuration screen password protection requires password entry | Accessing password-protected Configuration screen without entering password | warning |
Password is not confirmed message | Configuration password was not confirmed | Enabling configuration password without confirming it | warning |
Patient ID field not enabled | Patient ID field is disabled; cannot transfer ECGs without this required field | Attempt to transfer ECG to TraceMaster without Patient ID present | warning |
PIM Test Failed | PIM (Patient Interface Module) connectivity or communication test failed | PIM patient data cable disconnected or not properly attached; persistent cable connection issue | error |
Please select a TraceMaster server as the default connection | No default TraceMaster server selected | Attempting to set default TraceMaster server without first selecting it | warning |
Recommend Calibration | Battery level indicator accuracy affected | Battery power indicators not functioning usefully in daily work environment | warning |
Recommend Replacement | Batteries have reached end of useful life | Appears in Battery Status window after calibration procedure completion | error |
Required patient information missing | ECG transfer failed due to incomplete required patient information | Attempt to transfer ECG to TraceMaster without all required patient data | error |
Selected ECG exceeds USB archive space | ECG file is too large to fit in available USB memory stick storage | Attempt to archive or save ECG to USB memory stick | error |
Server connection reset | Network connection to TraceMaster server was unexpectedly closed during operation | Server closed connection during orders or ECG transfer operation | error |
The cardiograph must restart to display simulated waveform d | Cardiograph restart required to enable simulation mode | Simulation mode turned on; device requires restart to display simulated waveforms | info |
The cardiograph must restart to return to normal operating m | Cardiograph restart required to exit simulation mode | Simulation mode turned off; device requires restart to return to normal operation | info |
The Front/Back Battery is overcharged | Battery has been overcharged by a cardiograph running older software | Battery charged by cardiograph without software version A.07.07 | warning |
The maximum number of fax profiles has been configured messa | Cannot add more fax profiles; limit reached | Attempting to add fax profile when maximum is configured | warning |
The maximum number of OrderVue inboxes has been created mess | Cannot create more OrderVue inboxes; limit reached | Attempting to create OrderVue inbox when maximum is configured | warning |
The maximum number of TraceMaster connections has been confi | Cannot add more TraceMaster connections; limit reached | Attempting to add TraceMaster connection when maximum is configured | warning |
There are more than 200 ECGs on USB stick | USB memory stick contains more than maximum supported ECG files | Attempt to display or manage ECG files from USB stick with >200 ECGs | warning |
There is no default TraceMaster server available for Auto Ti | No default TraceMaster server configured for automatic time synchronization | TraceMaster settings not configured for Auto Time Sync | warning |
Timed ECG exceeds Main Archive limit | New timed ECG cannot be saved as Main Archive has reached capacity limit | Attempt to save a timed ECG to Main Archive | error |
Unapproved battery | Battery is not approved for use despite meeting Philips specifications | Unapproved battery detected at power-up, unlock, user login, or exit from sleep mode | warning |
Unexpected reply received from TraceMaster message | Unexpected response format received from TraceMaster server | Archive error with unexpected reply from server | error |
Unknown error accessing archive storage capacity | Error occurred while querying storage capacity of archive media | Attempt to check storage space on USB stick or internal Compact Flash | error |
Unrecognized battery | Installed battery is not supported and could damage the cardiograph | Unsupported battery detected at power-up | critical |
Unsaved changes will be deleted message | Configuration changes not saved and will be lost | Exiting configuration screens without saving changes | warning |
Unsupported XML version message | XML version of selected ECG is not supported by TraceMaster server | XML schema of selected ECG not compatible with TraceMaster | warning |
USB memory stick not accessible | USB memory stick not recognized or not fully inserted into connector | Attempt to access USB memory stick for reading or writing ECGs | error |
W-149c | Archive Display Difference - Non-PageWriter Source ECGs | Some ECGs retrieved from ECG management print with visual differences | info |
Worklist full - orders cannot download | Worklist has reached maximum capacity; cannot download additional orders | Attempt to download orders when Worklist at full capacity limit | error |
XML error message | XML ECG file is corrupted | File corruption detected during XML processing | error |
XML unicode encoding is not UTF-16 message | XML file has invalid character encoding | XML file encoding is not UTF-16 | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI