Philips OPTIMUS RAD: troubleshooting and error codes

The ServiceForge knowledge base holds 25 documented symptoms for the Philips OPTIMUS RAD, with 61 ranked causes, 77 diagnostic checks with test points and expected values, and 558 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Switch-ON not possible high voltage

H1 indicator on PCB EN100 is not illuminated; unit cannot be powered on

Ranked causes

  1. 1. ENF1 was released (common)
  2. 2. ENF1 is not switched ON (common)
  3. 3. Mains voltage absent, especially phase L3 (common)
  4. 4. ENF2 was released (occasional)
  5. 5. ENF2 is not switched ON (occasional)
  6. 6. PCB EN100 or connections faulty (occasional)
First check

Verify low-voltage supply

Test point
Low-voltage supply output
Expected
Supply within specification

4 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tube adaptation radiation

Automatic process measuring mA offset, filament current, emission characteristic, and dynamic boost behavior after conditioning

Ranked causes

  1. 1. New conditioned tube requires adaptation (common)
  2. 2. Tube temperature too high during adaptation (occasional)
  3. 3. Adaptation stalls or error message flashes briefly (occasional)
  4. 4. Adaptation does not start or stalls mid-process for 30+ seconds (occasional)
  5. 5. Constant READY displayed during PREP and EXP activation (rare)
First check

Verify tube conditioning completed before starting adaptation

Test point
Tube conditioning procedure status
Expected
Conditioning procedure completed with no unresolved arcing

12 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit cannot be switched ON high voltage

Generator fails to switch on; ENF1 or ENF2 contactors not activated

Ranked causes

  1. 1. ENF1 not switched ON (common)
  2. 2. ENF1 released due to fault condition (common)
  3. 3. ENF2 not switched ON (common)
  4. 4. ENF2 released by low-voltage supply or filament circuit fault (occasional)
  5. 5. ON circuit EN100 defective (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC faulty exposure detection strategy verification radiation

Understanding when AEC faulty exposure detection prevents unnecessary radiation by checking 4% dose value at 10% of APR backup time; determines exposure termination limits for APR100, APR100*, APR800, APR800* configurations.

Ranked causes

  1. 1. AEC faulty exposure detection ON - APR100 (common)
  2. 2. AEC faulty exposure detection ON - APR800 (common)
  3. 3. AEC faulty exposure detection ON - APR100* and APR800* (common)
  4. 4. AEC faulty exposure detection OFF (common)
First check

Verify 10% backup time value for APR100

Test point
APR100 backup time
Expected
> 10ms

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator cannot reach READY state after switch-ON attempt high voltage

No desk display; low-voltage supply or contact closure problems

Ranked causes

  1. 1. Low-voltage supply fault (common)
  2. 2. ENF1 released by ground fault or short circuit (occasional)
  3. 3. Intermediate circuit voltage missing or intermediate circuit capacitor fault (occasional)
  4. 4. Intermediate circuit capacitor overvoltage due to high impedance or symmetry resistor tolerance (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator in error state with displayed code high voltage

Operating panel displays error code during initialization or adaptation phase

Ranked causes

  1. 1. No tube data loaded yet (common)
  2. 2. No RGDVs programmed yet (common)
  3. 3. No communication between desk and central unit (occasional)
  4. 4. Adaptation failure after READY signal (occasional)
First check

Verify tube data loaded

Test point
Operating panel display
Expected
Tube data displayed, not 'No tube data loaded yet'

2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC signal error 00PL radiation

AEC signal disturbance detected; may present as warning or error depending on signal quality

Ranked causes

  1. 1. Measuring chamber shielding connected to system ground at chamber or interconnection (common)
  2. 2. Missing ground connection in measuring chamber cable (common)
  3. 3. Defective measuring chamber (rare)
First check

Verify shielding-to-system-ground connection absence

Test point
Sub-D connector shielding or 3+ pin M to system ground
Expected
Connection must not be present

3 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Generator fails to initialize high voltage

Generator does not reach READY state; selftest or internal communication fails

Ranked causes

  1. 1. Selftest failure in control desk C, EZ139, EZ130, EZ119, EZ150, EY, or EWA/B 102 (unknown)
  2. 2. CAN bus communication failure between central unit and functional units (unknown)
  3. 3. External ready condition not met (unknown)
First check

Observe selftest LED status during startup

Test point
Associated PCB red status LED
Expected
Red status LED illuminates briefly then blinks

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

No startup high voltage

Generator receives continuous reset via reset signal; all red LEDs illuminated

Ranked causes

  1. 1. EN100 V1 is defective (common)
  2. 2. No boot PROM present: EZ139 D3 (occasional)
  3. 3. Flash PROMs EZ139 D4/D5 not correctly loaded (occasional)
First check

Observe LED status

Test point
Generator LEDs
Expected
All red LEDs illuminated

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC dose detection does not trigger switch-OFF at 4% threshold radiation

Measuring chamber fails to detect minimum 4% of desired dose at 10% backup time; exposure continues when it should stop to prevent low-density image

Ranked causes

  1. 1. Film/screen combination sensitivity exceeds 400 speed (occasional)
  2. 2. Measuring chamber not detecting dose properly (unknown)
First check

Verify film/screen combination speed rating

Test point
FSC film/screen combination parameters
Expected
Speed ≤ 400 for 4% dose detection to be active

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC exposure exceeds 9.5× typical mAs radiation

Generator does not stop exposure at expected dose limit; exposure continues beyond 9.5 times the organ-dependent background mAs value stored in APR program

Ranked causes

  1. 1. Technical problem occurred during exposure (unknown)
  2. 2. Patient size does not match selected APR (common)
First check

Verify AEC exposure terminates at 9.5× background mAs limit

Test point
Dose rate control supervision 1 setting
Expected
Exposure stops when 9.5 x mAsbackup is reached

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AEC exposure reaches maximum 4000 ms timeout without proper termination radiation

Exposure duration approaches or exceeds 4000 ms backup time limit without appropriate dose detection response

Ranked causes

  1. 1. Dose rate control backup time calculation error (rare)
  2. 2. Supervision switches manually overridden (occasional)
First check

Verify DRC backup time calculation parameters

Test point
DRC kV and filament load settings in control system
Expected
Backup time = 9.5 × mAsbackup ÷ (dose rate function of kV and filament)

2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

AGenT software menu limited or Faultfind unavailable

AGenT main menu appears but not all menu items available; Faultfind option missing

Ranked causes

  1. 1. PMSSec reader not installed or not recognized (common)
  2. 2. PMSSec reader password not entered or authentication cancelled (occasional)
First check

Verify PMSSec reader software is installed

Test point
System software installed programs
Expected
PMSSec reader version 2.307 or higher installed

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Tube conditioning radiation

New or non-adapted tube requires conditioning procedure with graduated exposure parameters to prepare tube for adaptation

Ranked causes

  1. 1. New tube installed (common)
  2. 2. Tube arcing during conditioning (occasional)
First check

Select large focus only and begin conditioning sequence

Test point
Desk indicator
Expected
READY state displayed

3 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

High voltage measurement drift or emission current issues during adaptation high voltage

kV actual value problems or arcing entries in error log suggest high voltage circuit or duty cycle factor misalignment after adaptation failure

Ranked causes

  1. 1. Arcing entries 02WG or 02WH in error log (occasional)
  2. 2. kV actual value problems 02HG or 02HH in error log (occasional)
First check

Check error log for arcing entries after adaptation failure

Test point
Error log index entry codes
Expected
No arcing entries 02WG or 02WH present

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Service PC connection not established

Unable to connect service PC to generator for fault finding; hardware key may be required for full access

Ranked causes

  1. 1. Serial cable not properly connected or incorrect cable type (common)
  2. 2. Hardware key not installed or not recognized (occasional)
First check

Verify serial cable connection from PC COM1 to generator X5

Test point
EZ139 X5 connector and PC COM1 port
Expected
Cable plugged into correct connectors with proper pin alignment

1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Switch-ON not possible, H1 illuminated high voltage

H1 indicator on PCB EN100 is illuminated but switch-ON cannot proceed

Ranked causes

  1. 1. EMERGENCY OFF circuit is open (common)
  2. 2. Operating desk not connected (common)

Calibration: AEC faulty exposure detection configuration radiation

Procedure to configure and verify AEC faulty exposure detection settings for patient safety during automatic exposures.

Ranked causes

  1. 1. Configuration selection: detection ON vs OFF (common)
First check

Confirm detection ON for APR100 termination at 4000ms

Test point
Backup time setting after parameter override
Expected
4000ms maximum

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Area dose calculator radiation

Verify that area dose calculator measurements match displayed values within acceptable tolerance. Procedure involves measuring dose at specified kV settings with calibrated dosemeter and comparing calculated vs. displayed area dose products.

Ranked causes

  1. 1. Default yield curve or filter correction factors require adjustment (occasional)
First check

Place lead plate and measuring cell on table in central radiation beam; set 1m SMD, free cassette, kV-mAs-s technique, 10 mAs, 0.1s, 10×10 cm collimation, no filter

Test point
Measuring cell position
Expected
1m source-measuring distance, collimation 10×10 cm at height of measuring cell
Tools
Calibrated dosemeter (DALI with cell 77334 or PMX3), 1mm lead plate, ruler

10 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: kV function unit alignment after component replacement high voltage

After replacing PCB kV-control, Converter, H.V. tank, or PCB Central unit, kV output must be realigned to ensure proper rise time and prevent break-in or overshoot. Requires oscilloscope measurement of UCOMP signal and adjustment of duty cycle factor via AGenT software.

Ranked causes

  1. 1. Component replacement affecting kV control loop calibration (common)
First check

Verify oscilloscope configuration and probe connections

Test point
EZ130 X3, X26, cabinet drilling holes
Expected
Probes connected to correct pins, grounds at drilling holes only, time base 5 or 10ms/div

8 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

LED indicator dark on low-voltage power supply EZ102 high voltage

One or more of LEDs H2 to H5 on PCB EZ102 is not illuminated, indicating absence of supply voltage

Ranked causes

  1. 1. External consumer short circuit (common)
First check

Verify which LED is dark

Test point
LEDs H2 to H5 on PCB EZ102
Expected
LED illuminated (supply voltage present)

1 more check with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Phase L1 missing from mains supply high voltage

Mains contactors ENK2 and ENK1 cannot be activated due to missing L1 phase

Ranked causes

  1. 1. Phase L1 missing from supply (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Phase L2 missing from mains supply high voltage

Generator switches ON but not READY; filament circuit supply missing; kV function error message displayed

Ranked causes

  1. 1. Phase L2 missing from mains supply (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Phase L3 missing from mains supply high voltage

ON circuit has no supply voltage; generator cannot switch ON

Ranked causes

  1. 1. Phase L3 missing from mains supply (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Tube arcing during kV alignment patient safety

X-ray tube arcing occurs during high kV exposure adjustment, indicating potential tube damage or conditioning requirement.

Ranked causes

  1. 1. Tube requires conditioning after component replacement or long storage (occasional)
First check

Monitor for tube arcing during high kV exposure

Test point
X-ray tube visual/auditory observation and oscilloscope signals
Expected
No arcing, stable UCOMP and AV HT signals

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
00B0CPU: Error in application data service interfaceApplication data service interface failureerror
00B3Initialization errorAppears during startup or adaptation phaseerror
00B3NVRAM: Data language selector is invalidInvalid language selector stored in non-volatile RAMerror
00B6Initialization errorAppears during startup phaseerror
00B6NVRAM: FU adap data for DI are invalidFunctional unit adapter data corruptederror
00B7CPU: Message cannot be sentMessage transmission failure in processorerror
00B8NVRAM: Tomo mode switch cannot be enabledTomography mode enable failureerror
00BGAPR: No more lowest level menus availableMenu system resource exhaustionerror
00BHAPR: Display position collisionAPR configuration conflict in display layouterror
00BIAPR: Menu / APR mismatch in same levelMenu structure configuration errorerror
00BJAPR: Menu name not foundMenu name lookup failureerror
00BKAPR: APR is assigned to a different RGDVAPR assignment conflicterror
00BLAPR: Menu name already existsDuplicate menu name in configurationerror
00BMAPR: Max. display position reachedDisplay position limit exceedederror
00BNAPR: APR not found in this menuAPR lookup failure in menu treeerror
00BQCPU: APR cannot be modifiedAttempt to modify read-only APRerror
00BRCPU: APR is not assigned to an RGDVAPR assignment missingerror
00BSAPR: The RGDV of the APR is not ready for operationRGDV initialization failureerror
00BUAdaptation phase errorOccurs after adaptation mode initiated following READY signalerror
00BVAdaptation phase errorOccurs during adaptation phaseerror
00BVWarning: Temperature threshold exceededSupervised temperature level exceeds specified limit during high power operation or adaptationwarning
00BWAPR: APR not accepted by general calculationAPR validation failure during calculationerror
00DACAN: No CPU access to CAN-chipCAN bus controller interface failureerror
00DECAN: Unexpected CAN-chip interrupt pointerInvalid interrupt from CAN controllererror
00DICAN-chip state bus OFFCAN bus offline condition detectederror
00E0iRMX exception occurredReal-time kernel exceptionerror
00G0Variable in case statement has undefined valueSoftware logic error in case statementwarning
00G1Condition_code <> OK after CALL to sendSend operation returned error statuserror
00I4One FU has a WD-error, scantime_TV not programmed correctlyWatchdog error in functional uniterror
00L1GC: Checksum errorData checksum validation failureerror
00L2GC: Data access errorGenerator control data read/write failureerror
00L3GC: Limit data errorGenerator control limit values invaliderror
00L4GC: Limits inconsistentGenerator control limit values contradicterror
00L5GC: Calculation errorGenerator control calculation failederror
00L6GC: Function not implementedGenerator control function called but not availableerror
00M0Unable to initialise FU(s)Functional unit initialization failureerror
00M1Configuration key is missing or defectiveConfiguration key not detectederror
00M3No response at all from FU(s)Functional unit communication timeouterror
00PDPC comm: Unknown TDL proc IDInvalid procedure ID in PC communicationerror
00PLAEC signal errorAutomatic Exposure Control signal disturbance detectedwarning
00S0PCcomm: Tube programming errorX-ray tube configuration failureerror
00SAPCcomm: Not enough space at destination segmentMemory buffer overflow during communicationerror
00SCPCcomm: Value too largeParameter value exceeds maximumerror
00SDPCcomm: Terminator not foundMessage format errorerror
00SEPCcomm: Error in descriptionInvalid description data formaterror
00SFPCcomm: Item type unknownUnrecognized item type in configurationerror
00SGPCcomm: Internal type unknownUnknown internal data typeerror
00SHPCcomm: Value negativeNegative value where positive requirederror
00SJPCcomm: Syntax wrongCommand syntax errorerror
00SKPCcomm: String too longString parameter exceeds length limiterror
00SNPCcomm: FU Reference Table fullFunctional unit reference table exhaustederror
00SOPCcomm: Node ID unknownUnknown CAN node identifiererror
00SPPCcomm: FU Code unknownUnrecognized functional unit codeerror
00SQPCcomm: Syntax error in node IDInvalid node ID formaterror
00SSPCcomm: Request not performedCommand execution failederror
00STPCcomm: RMX errorReal-time kernel error in PC communicationerror
00SVPCcomm: Mail corruptedInter-process message corruption detectederror
00SWPCcomm: Procedure ID unknownUnknown procedure identifiererror
00SXPCcomm: FU mA incompatibleIncompatible mA control uniterror
00SYPCcomm: FU Off request failedFailure to power off functional uniterror
00SZPCcomm: Wrong responseUnexpected response from functional uniterror
00TATTS: Received message unknownUnrecognized TTS message receivederror
00TBTTS: Tube supervision error from FU kVTube supervision fault detected during operationerror
00TCTTS: Internal TTS errorInternal tube supervision system malfunctionerror
00TDTTS: Tube number unknownSystem cannot identify installed tube during initializationerror
00TETTS: NVRAM checksum errorNVRAM data integrity check fails during startuperror
00TFTTS: NVRAM unavailableSystem cannot access NVRAM during initializationerror
00TGTTS: Tube overheatedTube temperature exceeds safe operating limiterror
00THTTS: Load data supply inconsistentTube load data does not match supply voltagewarning
00X0CPU wrong timer IDCPU timer reference is invalid during operationerror
00X1CPU wrong timer modeCPU timer mode register contains invalid valueerror
00X2CPU wrong message typeCAN or IIM message has unrecognized type identifiererror
00X3CPU DWORD does not fit into BYTE3Data size mismatch in memory operationwarning
00X4Timeout of X-ray backup timerX-ray backup timer expires without expected eventwarning
00X5Timeout of X-ray rotation timerAnode rotation timer exceeds safe limitwarning
00X6Adaptation errorOccurs after adaptation mode initiatederror
00X6Timeout setting FUs, response missingFunctional Unit fails to respond to command within timeoutwarning
00X7CPU curve token is NO_TOKENInvalid curve token encountered during operationwarning
00XANVRAM switch table invalidSwitch table in NVRAM fails checksum or format validationunknown
00XBNVRAM tube data rotation invalidTube rotation data in NVRAM fails validationwarning
00XCNVRAM watch dog invalidWatchdog timer data in NVRAM is corruptedwarning
00XDNVRAM configuration table invalidSystem configuration table in NVRAM fails validationwarning
00XENVRAM test data invalidTest data stored in NVRAM fails validation checkwarning
00XFNVRAM RoCo data invalidRotation control data in NVRAM fails validationwarning
00XGCPU received IIM is unknownUnrecognized IIM message received by CPUunknown
00XHCPU received FU-type is unknownUnrecognized Functional Unit type in CAN messageunknown
00XIInit with FU-RoCo not OKFunctional Unit rotation control initialization failserror
00XJExposure time too shortRequested exposure duration below minimum thresholdwarning
00XKCPU FUmA refuses set dataFunctional Unit mA control rejects commandwarning
00XLNVRAM tube yield table invalidTube yield calibration table in NVRAM fails validationwarning
00XMNVRAM add filter corr table invalidAdditional filter correction data in NVRAM fails validationwarning
00XNNVRAM wedge filter corr table invalidWedge filter correction table in NVRAM fails validationwarning
00XOExposure time too longRequested exposure duration exceeds maximum thresholderror
00XPExposure time too longExposure duration exceeds warning thresholdwarning
00XQNVRAM tube statistic data invalidTube usage statistics in NVRAM fail validationwarning
00XRNVRAM gsta data invalidSystem statistics data in NVRAM fails validationwarning
00XSTube no. in CU and FUkV differentTube identification mismatch between Control Unit and kV modulewarning
00XTRotation in CU and FURoCo FUCIE differentAnode rotation parameter mismatch between moduleswarning
00XUTransition endless loopState machine transitions endlessly without reaching target stateerror
00XVNVRAM HW test flags invalidHardware test flag data in NVRAM fails validationwarning
00XWEN_X active in startupX-ray enable signal remains active during system startuperror
00XXRD_PR_X stays active after prepX-ray prepare signal does not deactivate after preparation phaseerror
02ABProcedure called with wrong parameterFunction receives out-of-range or invalid parameter valuewarning
02ACWrong index for table accessArray or table index out of bounds during accesserror
02ADWrong do case entryCase statement receives unhandled valueerror
02AEUnknown IIM receivedUnrecognized Inter-Module-Interface message arriveswarning
02AFIIM parameter out of rangeIIM message contains parameter outside safe limitswarning
02CAError in CASE selectorCAN message has invalid CASE selector valuewarning
02CBCAN message with wrong IIM-no receivedCAN frame addressed to non-existent modulewarning
02CCMultiple reception of same CAN frameCAN transmitter sends duplicate framewarning
02CEUnexpected signal value in CAN rx taskCAN message contains unexpected signal valuewarning
02CFCAN bus timeout while domain transmissionCAN communication times out during data domain transferwarning
02CGToken of CAN response mailbox is not a mailbox tokenCAN mailbox contains invalid token typewarning
02CXMultiple rx of same CAN last/only frameCAN receiver duplicates final framewarning
02CYAborted CAN domain receiveCAN message reception aborted due to timeout or errorwarning
02CZUnexpected CAN domain frame receivedCAN frame arrives outside expected IIM reception windowwarning
02DANo CPU access to CAN controllerCPU cannot communicate with CAN controllerwarning
02DBReset or release of CAN controller not acknowledgedCAN controller does not respond to reset commandwarning
02DDCheck of CAN controller DPRAM failedDPRAM test in CAN controller fails validationwarning
02DEUnexpected interrupt pointer in CAN controllerCAN controller interrupt handler address is invalidwarning
02DFCAN controller state undefinedCAN controller enters unknown state during operationwarning
02DGCAN controller state ERROR ACTIVE after ERROR PASSIVECAN bus transitions from passive to active error statewarning
02DHCAN controller state ERROR PASSIVECAN bus enters error passive statewarning
02DICAN controller state BUS OFFCAN controller goes offline due to excessive errorswarning
02DJCAN controller state DPRAM ERRORCAN controller DPRAM test failswarning
02DKCAN controller state DPRAM ERROR and ERROR PASSIVECAN controller has both DPRAM fault and bus errorswarning
02EAInterrupt 0: Divide by zeroCPU attempts division by zeroerror
02EBInterrupt 1: Single stepCPU single-step interrupt activatederror
02ECInterrupt 2: NMINon-maskable interrupt triggered in CPUerror
02EDInterrupt 3: BreakpointCPU breakpoint instruction executederror
02EEInterrupt 4: Overflow exceptionCPU arithmetic overflow detectederror
02EFInterrupt 5: Array bounds exceptionArray index exceeds bounds during accesserror
02EGInterrupt 6: Unused opcodeCPU executes undefined instruction codeerror
02EHInterrupt 7: ESC opcodeCPU executes escape/coprocessor instructionerror
02EICAN connection to CU lostCommunication with Control Unit via CAN is losterror
02GAInterpolation not possibleData interpolation algorithm fails to completewarning
02HAkV nominal value out of rangeRequested kV deviates > 4% + 1kV, 3 detections in 30mswarning
02HBkV nominal value out of rangeRequested kV outside limits 0-170 kVerror
02HCZ nominal value out of rangeDuty cycle deviates > 1% ± 0.2, 3 in 30ms, DC 3-30%warning
02HDZ nominal value out of rangeDuty cycle outside limits 0-50%error
02HEkV value during standby too largeStandby kV > 3kV for > 400ms after PREPwarning
02HFkV value during standby too largeStandby kV > 4kV for > 400ms after PREPerror
02HGkV actual value out of tolerance during adaptationMeasured X-ray tube voltage deviates from expected value during adaptation or operationerror
02HHkV feedback or measurement circuit fault during adaptationHigh voltage feedback signal absent or out of range during adaptation cycleerror
02HHkV actual value out of rangeMeasured kV outside 20-170kV, 3 detections in 30mserror
02HIE value during standby out of rangeStandby filament voltage 470-780V, 3 in 30mswarning
02HJE value during standby out of rangeStandby filament voltage outside 450-800Verror
02HKE value during high tension out of rangeOperating filament voltage 400-780V, 3 in 30mswarning
02HLE value during high tension out of rangeOperating filament voltage outside 350-800Verror
02HMConverter 1 temperature out of rangeTemperature 0-85°C, 3 detections in 30mswarning
02HNConverter 1 temperature out of rangeTemperature outside 0-90°C rangeerror
02HOConverter 2 temperature out of rangeTemperature 0-85°C, 3 detections in 30mswarning
02HPConverter 2 temperature out of rangeTemperature outside 0-90°C rangeerror
02HQHigh tension tank temperature out of rangeTank temperature 0-80°C, 3 detections in 30mswarning
02HRHigh tension tank temperature out of rangeTank temperature outside 0-85°C rangeerror
02HSDivider test cathode out of rangeCathode divider voltage 45.5-50.5kV, 3 in 30mswarning
02HTDivider test cathode out of rangeCathode divider voltage outside 43-53kVerror
02HUDivider test anode out of rangeAnode divider voltage 45.5-50.5kV, 3 in 30mswarning
02HVDivider test anode out of rangeAnode divider voltage outside 43-53kVerror
02HWkV asymmetricalkV phase asymmetry > 15%, 2 detections in 20mswarning
02HXkV asymmetricalkV phase asymmetry > 15%, 3 detections in 30mserror
02MAState request not accepted because of grid modeInvalid state transition requested during grid mode operationerror
02MBState request not accepted because of error stateSystem in error state preventing state transitionerror
02MCState requested by CU unknownControl Unit requests undefined system statewarning
02OARMX error: TimeoutReal-time multitasking executive timeouterror
02OBRMX error: MemoryReal-time OS memory allocation failserror
02OCRMX error: BusyReal-time OS resource is busyerror
02OERMX error: LimitReal-time OS limit exceedederror
02OFRMX error: ContextReal-time OS context errorerror
02OGRMX error: ExistReal-time OS resource already existserror
02OHRMX error: StateReal-time OS invalid state transitionerror
02OIRMX error: Not configuredReal-time OS resource not configurederror
02OJRMX error: Interrupt saturationToo many interrupts pending in real-time OSerror
02OKRMX error: Interrupt overflowReal-time OS interrupt queue overflowserror
02OLRMX error: TransmissionReal-time OS message transmission failserror
02OMRMX error: Divide by zeroReal-time OS divide by zero exceptionerror
02ONRMX error: OverflowReal-time OS arithmetic overflowerror
02OORMX error: TypeReal-time OS type errorerror
02OPRMX error: ParameterReal-time OS invalid parametererror
02OQRMX error: Bad callReal-time OS invalid function callerror
02ORRMX error: Array boundReal-time OS array bounds exceedederror
02OSRMX error: NDP errorReal-time OS NDP (numeric data processor) errorerror
02OTRMX error: Illegal opcodeReal-time OS encounters illegal instructionerror
02OURMX error: Emulator trapReal-time OS emulator trap activatederror
02OVRMX error: Interrupt table limitReal-time OS interrupt table exceeds limiterror
02OWRMX error: CPU xfer data limitReal-time OS CPU data transfer limit exceedederror
02OXRMX error: Wrap aroundReal-time OS buffer wrap-around errorerror
02OYRMX error: Check exceptionReal-time OS check exception raisederror
02OZRMX error: UnknownReal-time OS unknown error conditionerror
02RAGrid mode changeover requested during prepOperator requests grid mode change while preparing exposurewarning
02RBTube switch requested during preparationOperator requests tube change during prep phasewarning
02RCRequested P out of rangePower parameter P outside valid operating rangewarning
02SANot enough space at destinationData buffer at destination has insufficient spacewarning
02SBBase out of rangeBase address or reference out of valid rangewarning
02SCPC comm.: Value too largeValue in PC communication exceeds maximumwarning
02SDTerminator not foundMessage terminator missing in communicationwarning
02SEPC comm.: Error in descriptionInvalid description in PC communication messagewarning
02SFPC comm.: Item type unknownUnknown item type in PC communicationwarning
02SGPC comm.: Internal type unknownInternal data type unrecognized in PC communicationwarning
02SHPC comm.: Value negativeNegative value received when positive expectedwarning
02SIPC comm.: No space at dest. bufferDestination buffer full in PC communicationwarning
02SJPC comm.: Syntax wrongCommand syntax error in PC communicationwarning
02SKPC comm.: String too longString in PC communication exceeds maximum lengthwarning
02SLPC comm.: String truncatedString in PC communication was truncatedwarning
02SOPC comm.: Unknown table ID receivedUnrecognized table identifier in PC communicationwarning
02SPPC comm.: Access level to lowUser access level insufficient for operationwarning
02SQPC comm.: Unknown action requestedUnrecognized action in PC communication commandwarning
02SRPC comm.: Routing or message corruptMessage routing error or corruption detectedwarning
02SSSource buffer too small for incoming messageSource buffer cannot accommodate incoming datawarning
02STCAN buffer too small for outgoing messageCAN transmission buffer cannot hold messagewarning
02SUPC comm.: Access level is N/AAccess level undefined or unavailablewarning
02UAHW configuration identifier wrongHardware configuration ID does not match expectederror
02UBSet Up request received during preparationSetup command received while system is preparingwarning
02WAWrong tube selectedSelected tube differs from installed tubewarning
02WBWrong tube selectedTube selection error prevents operationerror
02WCEN X C signal faultyX-ray enable control signal is abnormalwarning
02WDEN X C signal faultyX-ray enable control signal failure prevents operationerror
02WEWrong grid mode selectedSelected grid mode incompatible with operationwarning
02WFWrong grid mode selectedGrid mode selection prevents exposure generationerror
02WGTube arcing detectedElectrical arcing inside tube detected by sensorswarning
02WGArcing detected in high voltage circuit during adaptationElectrical arc between tube electrodes during adaptation exposures or conditioningerror
02WHArcing or electrical fault detected during adaptationElectrical arc or fault condition during high voltage operation in adaptation sequenceerror
02WHTube arcing detectedTube arcing causes shutdown to prevent damageerror
02WIkV over voltage detectedMeasured kV exceeds safe maximum during operationwarning
02WJkV over voltage detectedExcessive kV prevents safe operationerror
02WKMeasuring not stableMeasurement signals show instability or driftwarning
02WLTube supervision errorTube monitoring system detects an anomalyerror
02WMTube supervision errorTube supervision system critical failureerror
03DICAN controller state BUS OFFCAN controller enters bus off statewarning
03DJCAN controller state DPRAM ERRORDual-port RAM error detected in CAN controllerwarning
03DKCAN controller state DPRAM ERROR and ERROR PASSIVEDual-port RAM error with passive error modewarning
03EACPU interrupt 0CPU interrupt 0 faulterror
03EBCPU interrupt 1CPU interrupt 1 faulterror
03EDCPU interrupt 3CPU interrupt 3 faulterror
03EECPU interrupt 4CPU interrupt 4 faulterror
03EFCPU interrupt 5CPU interrupt 5 faulterror
03EGCPU interrupt 6CPU interrupt 6 faulterror
03EHCPU interrupt 7CPU interrupt 7 faulterror
03EICAN is unable to send an error to CUCAN error reporting failureerror
03FANVRAM: Invalid checksumNVRAM data integrity check failswarning
03FBNVRAM: Standby filament not foundStandby filament data missing from NVRAMwarning
03FCNo NVRAM plugged inNVRAM module not detectederror
03FDNVRAM empty; battery?NVRAM data missing; battery backup suspected failedwarning
03GALimit errorParameter exceeds limiterror
03GBReal math error: Real underflowFloating-point calculation underflowwarning
03GCReal math error: Real overflowFloating-point calculation overflowwarning
03GDReal math error: Dword overflow32-bit integer calculation overflowwarning
03GEReal math error: Integer overflowInteger calculation overflowwarning
03GFReal math error: Word overflow16-bit integer calculation overflowwarning
03GGSingular matrixMatrix calculation with no inversewarning
03HAUnknown hardwareHardware not recognized during initializationerror
03HBIntermediate circuit voltage < 200VDC bus voltage falls below 200Vwarning
03HFUndefined analog input channelAnalog input channel not configuredwarning
03HGIf--actual out of toleranceFilament current measurement exceeds tolerancewarning
03HHIf setpoint to largeFilament current setpoint exceeds maximumerror
03HIIf--actual out of toleranceFilament current out of tolerance during operationerror
03HJIf--actual out of toleranceFilament current exceeds limiterror
03HKIf--nominal out of toleranceNominal filament current set too highwarning
03HLIf--nominal out of toleranceFilament current nominal value invaliderror
03HMIf--nominal out of toleranceNominal filament current parameter out of rangeerror
03HNNo retrigger received from CUControl unit does not send retrigger signalerror
03IAAdaptation cannot be completedFilament adaptation procedure failswarning
03ICNo Ie--adaptation measurement valuesExposure time adaptation missing measurement datawarning
03IDIe--adaptation values not evaluableExposure time adaptation data cannot be processedwarning
03KACondi.--X--Ray mode without mAs parameterConditioning mode requested without mAs valuewarning
03MAUndefined statusSystem status unknownwarning
03MBStatus change not allowedState transition requested but not permittedwarning
03MCFU init data not expectedInitialization data received unexpectedlywarning
03OARMX exception: E$TIMETimeout in RTOS kernelerror
03OBRMX exception: E$MEMMemory allocation failureerror
03OCRMX exception: E$BUSYResource is busyerror
03ODRMX exception: E$LIMITResource limit exceedederror
03OERMX exception: E$CONTEXTInvalid execution contexterror
03OFRMX exception: E$EXISTResource already existserror
03OGRMX exception: E$STATEInvalid object stateerror
03OHRMX exception: E$NOT$CONFIGUREDResource not configurederror
03OIRMX exception: E$INTERRUPT$SATURATIONInterrupt handler queue overflowerror
03OJRMX exception: E$INTERRUPT$OVERFLOWInterrupt service routine overflowerror
03OKRMX exception: E$TRANSMISSIONCommunication transmission errorerror
03OLRMX exception: E$ZERO$DIVIDEDivide by zero in kernelerror
03OMRMX exception: E$OVERFLOWOverflow in kernel operationerror
03ONRMX exception: E$TYPEType mismatch in kernelerror
03OORMX exception: E$PARAMInvalid parameter to kernel functionerror
03OPRMX exception: E$BAD$CALLInvalid kernel function callerror
03OQRMX exception: E$ARRAY$BOUNDArray index out of boundserror
03ORRMX exception: E$NDP$ERRORFloating-point coprocessor errorerror
03OSRMX exception: E$ILLEGAL$OPCODEIllegal CPU instruction executederror
03OTRMX exception: E$EMULATOR$TRAPCPU emulation trap triggerederror
03OURMX exception: E$INTERRUPT$TABLE$LIMITInterrupt table size exceedederror
03OVRMX exception: E$CPUXFER$DATA$LIMITCPU transfer data limit exceedederror
03OWRMX exception: E$SEG$WRAP$AROUNDMemory segment wrap-arounderror
03OXRMX exception: E$CHECK$EXCEPTIONCheck exception in kernelerror
03OYUnknown RMX exceptionUnrecognized kernel exceptionerror
03PAIe zero measuredExposure time current measurement reads zeroerror
03PBIe out of tolerance: 10% (Ie > 5mA, exp. time ≤44ms) or 3% (Ie > 5mA, exp. time > 44ms)Exposure time current deviates from tolerancewarning
03PCIe out of tolerance: 30% (Ie > 5mA, exp. time > 44ms)Exposure time current exceeds 30% toleranceerror
03PDSetpoint for Ie--regulation incorrectExposure time current setpoint invalidwarning
03PEEmergency OFF! Grid not closed!Emergency stop triggered; grid not conductingerror
03PFNo kV discharged due to missing IeHigh voltage not applied because exposure time current absenterror
03SCPC comm.: Value too largeParameter value exceeds maximum during PC communicationwarning
03SEPC comm.: Error in descriptionParameter description invalidwarning
03SFPC comm.: Item type unknownUnknown parameter type receivedwarning
03SGPC comm.: Internal type unknownInternal parameter type not recognizedwarning
03SHPC comm.: Value negativeNegative value received where positive requiredwarning
03SIPC comm.: No space at dest. bufferReceive buffer fullwarning
03SJPC comm.: Syntax wrongInvalid message syntax from PCwarning
03SKPC comm.: String too longString parameter exceeds maximum lengthwarning
03SLPC comm.: String truncatedString parameter truncated during receptionwarning
03SOPC comm.: Unknown table ID receivedInvalid table identifier from PCwarning
03SPPC comm.: Access level to lowInsufficient access rights for operationwarning
03SQPC comm.: Unknown action requestedInvalid action command from PCwarning
03SRPC comm.: Routing or message corruptMessage routing or content corruptedwarning
03SUPC comm.: Access level is N/AAccess level attribute not availablewarning
07CACAN: Case--selector errorCase selector value invaliderror
07CBCAN: Invalid CAN ID %uCAN message with unrecognized identifierwarning
07CCCAN: Frame rep. overflow IIM%uRepeated CAN frame buffer overflowerror
07CDCAN: No RTR from CURemote transmission request not receivederror
07CECAN: rx signal conflict IIM%uReceived signal conflicts with expectederror
07CFCAN: tx timeoutCAN transmit operation times outerror
07CICAN: IMPOSSIBLE ERRORUndocumented CAN error statewarning
07CPCAN: CPU: PXerr %d %s(%d)PXROS CPU error with detailswarning
07CRCAN: CPU: Message request failMessage request operation failedwarning
07CSCAN: CPU: Message send errorCAN message transmission failedwarning
07CYCAN: rx abort IIM%uCAN reception abortederror
07CZCAN: Unexpected frame (IIM%u)Received CAN frame not expectedwarning
07DACAN: Chip access errorCAN chip memory access failserror
07DBCAN: Chip reset errorCAN chip reset failserror
07DCCAN: Chip reset release errorCAN chip fails to exit reseterror
07DECAN: Illegal interrupt pointerInvalid CAN interrupt handler pointerwarning
07DFCAN: Chip state undefinedCAN chip in undefined stateerror
07DGCAN: Chip err act. after pass.CAN chip error state active after passivewarning
07DHCAN: Chip state error passiveCAN chip in error passive statewarning
07DICAN: Chip state bus OFFCAN chip bus OFF statewarning
07DJCAN: Chip DPRAM errorDual-port RAM error in CAN chiperror
07DKCAN: Chip DPRAM error & passiveDPRAM error with passive error statewarning
07DLCAN: Unexpected interruptUnexpected CAN interrupt receivedwarning
07LACV received IIM unknownConversion received from unknown modulewarning
07LBRC stator number out of rangeRotor/stator number exceeds valid rangewarning
07LCRC stator not availableConfigured rotor stator not installedwarning
07LDRC stator 1 readback failedRotor stator 1 feedback missingerror
07LERC stator 2 readback failedRotor stator 2 feedback missingerror
07LFRC stator 3 readback failedRotor stator 3 feedback missingerror
07LGRC speed value out of rangeRotor speed parameter invalidwarning
07LHRC speed set timeoutRotor speed change times outerror
07LIRC maximal stator load exceededRotor torque exceeds maximumwarning
07LJRC maximal rotation time exceededRotor rotation duration exceeds limiterror
07LKAM amplimat chamber number out of rangeAmplimat chamber number invalidwarning
07LLAM amplimat field number out of rangeAmplimat field number invalidwarning
07LMAM amplimat delay value out of rangeAmplimat delay parameter invalidwarning
08CACAN: Case--selector errorCase selector value invalid on display moduleerror
08CBCAN: Invalid CAN ID %uCAN message with unrecognized identifierwarning
08CCCAN: Frame rep. overflow IIM%uRepeated CAN frame buffer overflowerror
08CDCAN: No RTR from CURemote transmission request not receivederror
08CECAN: rx signal conflict IIM%uReceived signal conflicts with expectederror
08CFCAN: tx timeoutCAN transmit operation times outerror
08CICAN: IMPOSSIBLE ERRORUndocumented CAN error statewarning
08CPCAN: CPU: PXerr %d %s(%d)PXROS CPU error with detailswarning
08CRCAN: CPU: message request failMessage request operation failedwarning
08CSCAN: CPU: message send errorCAN message transmission failedwarning
08CYCAN: rx abort IIM%uCAN reception abortederror
08CZCAN: Unexpected frame (IIM%u)Received CAN frame not expectedwarning
08DACAN: Chip access errorCAN chip memory access failserror
08DBCAN: Chip reset errorCAN chip reset failserror
08DCCAN: Chip reset release errorCAN chip fails to exit reseterror
08DECAN: Illegal interrupt pointerInvalid CAN interrupt handler pointerwarning
08DFCAN: Chip state undefinedCAN chip in undefined stateerror
08DGCAN: Chip err act. after pass.CAN chip error state active after passivewarning
08DHCAN: Chip state error passiveCAN chip in error passive statewarning
08DICAN: Chip state bus OFFCAN chip bus OFF statewarning
08DJCAN: Chip DPRAM errorDual-port RAM error in CAN chiperror
08DKCAN: Chip DPRAM error & passiveDPRAM error with passive error statewarning
08DLCAN: Unexpected interruptUnexpected CAN interrupt receivedwarning
08HANo message receive displaytaskDisplay task receives no messageerror
08HBNo message release displaytaskDisplay task release message failserror
08HCAPR not foundAlarm presentation resource not founderror
08HDOffset in menu structure out of rangeMenu offset exceeds valid rangeerror
08HFNo message request for test taskTest task receives no message requesterror
08HGNo message send for test taskTest task fails to send messageerror
08HHAPR buffer fullAlarm presentation buffer overflowserror
08HINo message send for test taskTest task message transmission failserror
08HJNo send message to CU from ODDMessage transmission from display to control unit failserror
08HKData error in CAN messageCAN message contains invalid dataerror
08HLNo message send for service taskService task fails to transmit messageerror
08IEWrong setup IIMInvalid setup for interface moduleerror
08SANo request domtxtask when scanningDomain text task request fails during scanerror
08SBNo request domtxtask when testingDomain text task request fails during testerror
08SCNo send message to task2_scMessage to scanner task 2 failserror
10CACAN: Case--selector errorCase selector value invalid on rotor controlerror
10CBCAN: Invalid CAN ID %uCAN message with unrecognized identifierwarning
10CCCAN: Frame rep. overflow IIM%uRepeated CAN frame buffer overflowerror
10CDCAN: No RTR from CURemote transmission request not receivederror
10CECAN: rx signal conflict IIM%uReceived signal conflicts with expectederror
10CFCAN: tx timeoutCAN transmit operation times outerror
10CICAN: IMPOSSIBLE ERRORUndocumented CAN error statewarning
10CPCAN: CPU: PXerr %d %s(%d)PXROS CPU error with detailswarning
10CRCAN: CPU: Message request failMessage request operation failedwarning
10CSCAN: CPU: Message send errorCAN message transmission failedwarning
10CYCAN: rx abort IIM%uCAN reception abortederror
10CZCAN: Unexpected frame (IIM%u)Received CAN frame not expectedwarning
10DACAN: Chip access errorCAN chip memory access failserror
10DBCAN: Chip reset errorCAN chip reset failserror
10DCCAN: Chip reset release errorCAN chip fails to exit reseterror
10DECAN: Illegal interrupt pointerInvalid CAN interrupt handler pointerwarning
10DFCAN: Chip state undefinedCAN chip in undefined stateerror
10DGCAN: Chip err act. after pass.CAN chip error state active after passivewarning
10DHCAN: Chip state error passiveCAN chip in error passive statewarning
10DICAN: Chip state bus OFFCAN chip bus OFF statewarning
10DJCAN: Chip DPRAM errorDual-port RAM error in CAN chiperror
10DKCAN: Chip DPRAM error & passiveDPRAM error with passive error statewarning
10DLCAN: Unexpected interruptUnexpected CAN interrupt receivedwarning
10FBShort circuit detectedElectrical short detected on rotor controlerror
10FTOvercurrent detectedRotor control motor current exceeds limitwarning
10IFInitialization failedRotor control initialization unsuccessfulwarning
10LAAcceleration count limit exceededRotor acceleration attempts exceed limitwarning
10LHPhase current %u mA (>%u)Motor phase current exceeds maximumerror
10LLPhase current %u mA (<%u)Motor phase current below minimumerror
10LOIntermediate voltage %u V (>%u)DC bus voltage exceeds maximumwarning
10LTTemperature limit exceededRotor control temperature too higherror
10LUIntermediate voltage %u V (<%u)DC bus voltage below minimumwarning
10LZTemperature sensor failureTemperature sensor not respondingerror
10OECPU: PXROS error %dPXROS kernel error on rotor controlwarning
10OFCPU: PXROS error %d %s(%d)PXROS kernel error with detailswarning
10RCRotation check failedRotor does not rotate as expectederror
10RIInvalid rotation request : %uInvalid rotor speed or direction requesterror
10RTRotation request timeoutTube rotation request does not complete within timeout perioderror
10TDInvalid data for tubeInvalid tube data received or detectederror
10TEStator hardware errorStator hardware malfunction detectederror
10TFStator switching failedStator switching command failed to executeerror
10TIInvalid stator requestInvalid or out-of-range stator request issuederror
10TRStator change with rotating anodeStator change attempted while anode is rotatingerror
10UIUnknown message from CUUnrecognized message received from control unitwarning
10UMUnexpected message from CUValid but unexpected message from control unitwarning
10WTCPU watchdog timeoutCPU watchdog timer expireswarning
10XXImpossible errorError condition marked as impossible occurredwarning
13CACAN case-selector errorCase selector communication failure on CAN buserror
13CBCAN invalid CAN IDInvalid CAN identifier receivedwarning
13CCCAN frame repetition overflowFrame repetition buffer overflow on CANerror
13CDCAN no remote transmission request from CUExpected RTR from control unit not receivederror
13CECAN receive signal conflictSignal conflict detected during CAN receptionerror
13CFCAN transmit timeoutCAN transmit operation times outerror
13CICAN impossible errorImpossible error condition on CANwarning
13CPCAN CPU process execution errorCPU process execution error detectedwarning
13CRCAN CPU message request failedMessage request failed to executewarning
13CSCAN CPU message send errorError sending message from CPUwarning
13CYCAN receive abortCAN reception abortederror
13CZCAN unexpected frameUnexpected frame received on CANwarning
13DACAN chip access errorError accessing CAN chip registerserror
13DBCAN chip reset errorCAN chip reset failederror
13DCCAN chip reset release errorError releasing CAN chip from reseterror
13DECAN illegal interrupt pointerInvalid interrupt pointer detectedwarning
13DFCAN chip state undefinedCAN chip in undefined stateerror
13DGCAN chip error active after passiveChip transitioned from error passive to error activewarning
13DHCAN chip state error passiveCAN chip in error passive statewarning
13DICAN chip state bus offCAN chip bus-off (offline)warning
13DJCAN chip DPRAM errorDual-port RAM error in CAN chiperror
13DKCAN chip DPRAM error and error passiveDPRAM error concurrent with error passive statewarning
13DLCAN unexpected interruptUnexpected interrupt from CAN chipwarning
13LACAN voltage CV received unknownVoltage message from unknown source receivedwarning
13LBCAN IO wrong bidirectional lines output valueIO bidirectional line has incorrect output valuewarning
13LCCAN TR TOMO value out of rangeTomography value for K5-K12 outside valid rangewarning
13LDCAN TR RGDV value out of rangeRGDV (readback gain/drive value) outside rangewarning
13LECAN TR RGDV readback failedFailed to read RGDV feedbackerror
13LFCAN TR wrong sync contact valueSync contact has incorrect valuewarning
13LGCAN TR wrong handswitch enable valueHand switch enable signal incorrectwarning
13LHCAN PR S1/S2 switch active during startupS1 or S2 safety switch active when system startserror
14CACAN case-selector error (secondary)Case selector communication failure on secondary CANerror
14CBCAN invalid CAN ID (secondary)Invalid CAN identifier on secondary channelwarning
14CCCAN frame repetition overflow (secondary)Secondary CAN frame repetition overflowerror
14CDCAN no RTR from CU (secondary)RTR not received from CU on secondaryerror
14CECAN receive signal conflict (secondary)Signal conflict on secondary CAN receptionerror
14CFCAN transmit timeout (secondary)Secondary CAN transmit times outerror
14CICAN impossible error (secondary)Impossible error on secondary channelwarning
14CPCAN CPU process execution error (secondary)CPU process error on secondarywarning
14CRCAN CPU message request failed (secondary)Secondary CPU message request failedwarning
14CSCAN CPU message send error (secondary)Secondary CPU message send failedwarning
14CYCAN receive abort (secondary)Secondary CAN reception abortederror
14CZCAN unexpected frame (secondary)Unexpected frame on secondary CANwarning
14DACAN chip access error (secondary)Secondary CAN chip register access errorerror
14DBCAN chip reset error (secondary)Secondary CAN chip reset failederror
14DCCAN chip reset release error (secondary)Secondary CAN chip reset release failederror
14DECAN illegal interrupt pointer (secondary)Secondary illegal interrupt pointerwarning
14DFCAN chip state undefined (secondary)Secondary CAN chip in undefined stateerror
14DGCAN chip error active after passive (secondary)Secondary chip error passive to active transitionwarning
14DHCAN chip state error passive (secondary)Secondary CAN chip error passivewarning
14DICAN chip state bus off (secondary)Secondary CAN chip bus-offwarning
14DJCAN chip DPRAM error (secondary)Secondary CAN chip DPRAM errorerror
14DKCAN chip DPRAM error and error passive (secondary)Secondary DPRAM error with error passivewarning
14DLCAN unexpected interrupt (secondary)Secondary unexpected interruptwarning
14LACAN voltage CV received unknown (secondary)Secondary unknown voltage messagewarning
14LBCAN IO wrong bidirectional lines (secondary)Secondary IO bidirectional line errorwarning
14LCCAN TR TOMO value out of range (secondary)Secondary tomography value out of rangewarning
14LDCAN TR RGDV value out of range (secondary)Secondary RGDV out of rangewarning
14LECAN TR RGDV readback failed (secondary)Secondary RGDV readback failederror
14LFCAN TR wrong sync contact value (secondary)Secondary sync contact wrongwarning
14LGCAN TR wrong handswitch enable (secondary)Secondary hand switch enable wrongwarning
14LHCAN PR S1/S2 switch active during startup (secondary)Secondary S1/S2 active at startuperror
15CACAN case-selector error (tertiary)Tertiary case selector communication failureerror
15CBCAN invalid CAN ID (tertiary)Tertiary invalid CAN IDwarning
15CCCAN frame repetition overflow (tertiary)Tertiary frame overflowerror
15CDCAN no RTR from CU (tertiary)Tertiary RTR not receivederror
15CECAN receive signal conflict (tertiary)Tertiary signal conflicterror
15CFCAN transmit timeout (tertiary)Tertiary transmit timeouterror
15CICAN impossible error (tertiary)Tertiary impossible errorwarning
15CPCAN CPU process error (tertiary)Tertiary CPU process errorwarning
15CRCAN CPU message request failed (tertiary)Tertiary message request failedwarning
15CSCAN CPU message send error (tertiary)Tertiary message send errorwarning
15CYCAN receive abort (tertiary)Tertiary reception abortederror
15CZCAN unexpected frame (tertiary)Tertiary unexpected framewarning
15DACAN chip access error (tertiary)Tertiary chip access errorerror
15DBCAN chip reset error (tertiary)Tertiary chip reset failederror
15DCCAN chip reset release error (tertiary)Tertiary reset release failederror
15DECAN illegal interrupt pointer (tertiary)Tertiary illegal interruptwarning
15DFCAN chip state undefined (tertiary)Tertiary chip state undefinederror
15DGCAN chip error active after passive (tertiary)Tertiary error passive to activewarning
15DHCAN chip state error passive (tertiary)Tertiary chip error passivewarning
15DICAN chip state bus off (tertiary)Tertiary chip bus-offwarning
15DJCAN chip DPRAM error (tertiary)Tertiary DPRAM errorerror
15DKCAN chip DPRAM error and error passive (tertiary)Tertiary DPRAM error with passivewarning
15DLCAN unexpected interrupt (tertiary)Tertiary unexpected interruptwarning
15LACAN voltage CV received unknown (tertiary)Tertiary unknown voltage messagewarning
15LBCAN IO wrong bidirectional lines (tertiary)Tertiary IO bidirectional errorwarning
15LCCAN TR TOMO value out of range (tertiary)Tertiary tomography value out of rangewarning
15LDCAN TR RGDV value out of range (tertiary)Tertiary RGDV out of rangewarning
15LECAN TR RGDV readback failed (tertiary)Tertiary RGDV readback failederror
15LFCAN TR wrong sync contact value (tertiary)Tertiary sync contact wrongwarning
15LGCAN TR wrong handswitch enable (tertiary)Tertiary hand switch enable wrongwarning
15LHCAN PR S1/S2 switch active during startup (tertiary)Tertiary S1/S2 active at startuperror
16CACAN case-selector error (quaternary)Quaternary case selector communication failureerror
16CBCAN invalid CAN ID (quaternary)Quaternary invalid CAN IDwarning
16CCCAN frame repetition overflow (quaternary)Quaternary frame overflowerror
16CDCAN no RTR from CU (quaternary)Quaternary RTR not receivederror
16CECAN receive signal conflict (quaternary)Quaternary signal conflicterror
16CFCAN transmit timeout (quaternary)Quaternary transmit timeouterror
16CICAN impossible error (quaternary)Quaternary impossible errorwarning
16CPCAN CPU process error (quaternary)Quaternary CPU process errorwarning
16CRCAN CPU message request failed (quaternary)Quaternary message request failedwarning
16CSCAN CPU message send error (quaternary)Quaternary message send errorwarning
16CYCAN receive abort (quaternary)Quaternary reception abortederror
16CZCAN unexpected frame (quaternary)Quaternary unexpected framewarning
16DACAN chip access error (quaternary)Quaternary chip access errorerror
16DBCAN chip reset error (quaternary)Quaternary chip reset failederror
16DCCAN chip reset release error (quaternary)Quaternary reset release failederror
16DECAN illegal interrupt pointer (quaternary)Quaternary illegal interruptwarning
16DFCAN chip state undefined (quaternary)Quaternary chip state undefinederror
16DGCAN chip error active after passive (quaternary)Quaternary error passive to activewarning
16DHCAN chip state error passive (quaternary)Quaternary chip error passivewarning
16DICAN chip state bus off (quaternary)Quaternary chip bus-offwarning
16DJCAN chip DPRAM error (quaternary)Quaternary DPRAM errorerror
16DKCAN chip DPRAM error and error passive (quaternary)Quaternary DPRAM error with passivewarning
16DLCAN unexpected interrupt (quaternary)Quaternary unexpected interruptwarning
16LACAN voltage CV received unknown (quaternary)Quaternary unknown voltage messagewarning
16LBCAN IO wrong bidirectional lines (quaternary)Quaternary IO bidirectional errorwarning
16LCCAN TR TOMO value out of range (quaternary)Quaternary tomography value out of rangewarning
16LDCAN TR RGDV value out of range (quaternary)Quaternary RGDV out of rangewarning
16LECAN TR RGDV readback failed (quaternary)Quaternary RGDV readback failederror
16LFCAN TR wrong sync contact value (quaternary)Quaternary sync contact wrongwarning
16LGCAN TR wrong handswitch enable (quaternary)Quaternary hand switch enable wrongwarning
16LHCAN PR S1/S2 switch active during startup (quaternary)Quaternary S1/S2 active at startuperror