Philips MX800: troubleshooting and error codes
The ServiceForge knowledge base holds 153 documented symptoms for the Philips MX800, with 327 ranked causes, 129 diagnostic checks with test points and expected values, and 16 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
AC LED does not light up high voltage
AC Power LED on front panel does not illuminate when AC mains connected and power switch pressed
Ranked causes
- 1. AC Connection not OK (common)
- 2. LED defective (occasional)
- 3. Power Switch board not connected (occasional)
- 4. Power supply defective (occasional)
- 5. iPC defective (rare)
- 6. I/O board defective (rare)
- 7. Connector board defective (rare)
- 8. Panel adapter board defective (rare)
- 9. Mixer board defective (rare)
- 10. Main Board defective (rare)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
iPC does not power up or shows PC HW Malfunction
Unit fails to boot or displays PC HW Malfunction message during power-on test
Ranked causes
- 1. Power is off or embedded controller failed (common)
- 2. System holds off due to over temperature or other error (common)
- 3. Voltage error (blink code 1) (occasional)
- 4. Temperature error (blink code 2) (occasional)
- 5. Fan error (blink code 3) (occasional)
- 6. CMOS battery low (blink code 4) (occasional)
- 7. Abnormal reset (blink code 5) (occasional)
- 8. Input voltage error (blink code 6) (occasional)
- 9. Embedded controller internal error (blink code 7) (rare)
Observe status LED color and blink pattern
- Test point
- Status LED behind monitor rear housing
- Expected
- Off = power off or embedded controller failed; Red = system holds off or error; Orange = Sleep/Hibernate/Soft Off; Green = fully functional
Green Power On LED and Red Error LED remain off high voltage
After pressing power on button, neither the green Power On/Standby LED nor the red Error LED illuminates
Ranked causes
- 1. Power Switch Micro Controller hung (common)
- 2. Power switch board not connected to main board (occasional)
- 3. Power Switch Board defective (occasional)
- 4. I/O Board defective (occasional)
- 5. Connector board defective (rare)
- 6. Mixer, MSL2, WLAN, Panel adapter or Alarm LED board defective (rare)
- 7. Main board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Integrated display is blank or brightness reduced caution
Built-in LCD display shows no image or brightness is noticeably reduced
Ranked causes
- 1. Thermal issue - room temperature too high or instrument nearby (common)
- 2. Backlight Inverter Cable not connected (occasional)
- 3. Backlight tubes defective (occasional)
- 4. Backlight Inverter board defective (occasional)
- 5. Panel Adapter board defective (rare)
- 6. LCD Flat panel defective (rare)
- 7. Main board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen not functioning
Touch input does not respond to user interaction
Ranked causes
- 1. Touchscreen functionality temporarily disabled (common)
- 2. Touchscreen functionality permanently disabled (occasional)
- 3. External Touch cable not connected (common)
- 4. External Touch driver configuration incorrect (occasional)
- 5. Connector board defective (rare)
- 6. External touch defective (occasional)
- 7. Main board defective (rare)
Touch calibration
- Test point
- touchscreen surface
- Expected
- calibration performed successfully
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External device not receiving data from monitor
Connected external equipment receiving no data from monitor; data export or transmission fails
Ranked causes
- 1. MIB/RS232 port not configured for data export (common)
- 2. Wrong data export protocol driver configured (common)
- 3. Cable between monitor and external device defective or disconnected (common)
- 4. External device does not support monitor's data export protocol version (occasional)
- 5. Terminal concentrator does not support dynamic baud rate negotiation (rare)
- 6. MIB/RS232 board in wrong slot or defective (occasional)
Verify MIB/RS232 port is configured for data export
- Test point
- Monitor configuration menu
- Expected
- Data export enabled on target port
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen not functioning caution
Touchscreen input does not respond to touch input or calibration is invalid
Ranked causes
- 1. Touchscreen temporarily disabled (common)
- 2. Touchscreen permanently disabled in service mode (occasional)
- 3. Touch screen not connected (occasional)
- 4. Panel adapter board defective (occasional)
- 5. Touch screen defective (occasional)
- 6. Main board defective (rare)
Check if touchscreen disabled
- Test point
- Main Screen key
- Expected
- Padlock symbol visible on Main Screen key if disabled
- Tools
- Visual inspection
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner cannot read barcodes
Scanner lights and sounds function but barcode reads fail
Ranked causes
- 1. Scanner held too close or too far from barcode (common)
- 2. Scanner window not clean or blocked (common)
- 3. System does not recognize barcode format (occasional)
- 4. System does not recognize scanner (occasional)
- 5. Scanner has mechanical problem (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External display connected to Independent Display Interface is blank caution
Display connected to Independent Display Interface port shows no video output
Ranked causes
- 1. Check integrated display first (common)
- 2. Video cable to external display not connected (common)
- 3. External display has no power (common)
- 4. External display is defective (occasional)
- 5. Independent Display Interface defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED stays on continuously high voltage
After power-on, the red Error LED remains illuminated and does not change to green
Ranked causes
- 1. External connected device defective (common)
- 2. I/O Board defective (occasional)
- 3. Connector board defective (rare)
- 4. Mixer, MSL2, WLAN, Panel adapter or Alarm LED board defective (rare)
- 5. Main board defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
802.11 Bedside Adapter configuration problem (WPA/WPA2 Enterprise) caution
Wireless adapter configuration issue with WPA or WPA2 Enterprise security authentication
Ranked causes
- 1. SSID, Mode, Country, or Security settings incorrect (common)
- 2. Authentication server or WLAN controller error (occasional)
- 3. Certificate validation failure or invalid configuration (occasional)
- 4. Invalid or incompatible CA certificate installed (rare)
Check connection status and SSID/Mode settings
- Test point
- Main Setup > Network > WLAN Diagnostics > Conn.Status
- Expected
- State shows 'Authenticating' or 'Connected'; Mode/SSID/Country match infrastructure
4 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no alarm sound is issued patient safety
Alarm condition triggered but no audible alarm tone produced by speaker
Ranked causes
- 1. Audible alarm indicators have been switched off (common)
- 2. Volume set to 0 (common)
- 3. Speaker defective (occasional)
- 4. Sound amplifier on main board defective (occasional)
Verify audible alarm indicators are enabled
- Test point
- System settings menu
- Expected
- Audible alarm indicators enabled
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no LED lights up caution
Alarm is triggered and audible but no LED indication is visible
Ranked causes
- 1. Environmental lighting too bright (occasional)
- 2. Alarm LED board cable disconnected (common)
- 3. Alarm LED board defective (occasional)
- 4. Main board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm on screen but no LED or sound on alarm device patient safety
Alarm displays on monitor but remote alarm device silent and dark
Ranked causes
- 1. Cabling not connected (common)
- 2. Cabling defective (occasional)
- 3. I/O board defective (occasional)
- 4. Remote Alarm Device defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External display is blank caution
Video output to external slave display is not visible
Ranked causes
- 1. Check integrated display first (common)
- 2. Video cable not connected (common)
- 5. Connector Board defective (occasional)
- 6. Main board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message Bad MSL issued
Monitor displays Bad MSL connection INOP message
Ranked causes
- 1. Measurement Extension defective (common)
- 2. Too many Multi-Measurement Modules connected (occasional)
- 3. Incorrect connections or cable damage (common)
- 4. Parameter board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message MSL Device Unsupp issued
Monitor displays unsupported MSL device INOP message
Ranked causes
- 1. Wrong software revision (common)
- 2. Too many Multi-Measurement Modules connected (common)
- 3. Unsupported type of Multi-Measurement Module connected (e.g. M3000A) (occasional)
- 4. Module Rack connected to incompatible monitor (e.g. MX400-550) (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leakage between calibration line inlet and pump inlet detected
Flowmeter reading exceeds 4 ml/min or cell pressure does not drop to 300-500 mmHg when calibration line inlet is blocked during leakage check part 2
Ranked causes
- 1. Poor calibration line connections (common)
- 2. Poor tubing connections at pump inlet or mCO2 gas inlet (common)
- 3. Defective calibration line (occasional)
- 4. Leakage in tubing of M3015A or mCO2 assembly (occasional)
Inspect calibration line connections
- Test point
- calibration line connection points
- Expected
- all connections tight with no visible damage
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Local printer unavailable caution
Status message issued; printer job stalled; device not responding
Ranked causes
- 1. Printer not switched on (common)
- 2. Printer paper tray empty (common)
- 3. Cabling not connected correctly (common)
- 4. Connector board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor lost connection to Information Center
Central icon shown as outline only with broken connection line; monitor was being monitored but lost connection to Information Center
Ranked causes
- 1. LAN cable disconnected (common)
- 2. Information Center disconnected (occasional)
- 3. Network infrastructure failure (occasional)
- 4. Wireless monitor out of range (occasional)
Check LAN cable connection
- Test point
- Network cable at monitor
- Expected
- Cable firmly seated and intact
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printouts not as expected caution
Print output quality or format incorrect; document appearance wrong
Ranked causes
- 1. Printer paper size not correctly configured (common)
- 2. Printer resolution not correctly configured (common)
- 3. Printer color support misconfigured (common)
- 4. Printer not compatible (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder M1116C not communicating caution
Recorder not responding; connection issue; no printing occurring
Ranked causes
- 1. Poor connection to rack (common)
- 2. Only one recorder module may be used (occasional)
- 3. Monitor not configured properly (occasional)
- 4. Recorder defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder not communicating with system caution
Recorder not responding to monitor commands; connection issue
Ranked causes
- 1. Poor connection to module rack (common)
- 2. Only one recorder module may be used (occasional)
- 3. System not configured properly (occasional)
- 4. Too many modules connected (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unsupported LAN message displayed
Monitor displays 'Unsupported LAN' message after 1 minute, indicating LAN connection exists but no IP address assigned
Ranked causes
- 1. Network failure (common)
- 2. Monitor connected to wrong network (occasional)
- 3. IP address conflict after infrastructure reinstallation (occasional)
- 4. IIT enabled but no infrastructure found (occasional)
Verify network connectivity
- Test point
- LAN port, network cable
- Expected
- Active network connection detected
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner does not function
Scanner produces no lights, sound, or barcode reads
Ranked causes
- 1. Scanner not plugged into USB port (common)
- 2. System does not recognize scanner (occasional)
- 3. Scanner has mechanical problem (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Monitor Temp INOP message caution
Internal monitor temperature exceeds safe operating limits
Ranked causes
- 1. Temperature inside monitor too high (common)
- 2. Monitor ventilation obstructed (common)
- 3. Main Board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG OUT signal missing to defibrillator patient safety
No ECG output signal transmitted to external defibrillator; ECG data not available to defib unit
Ranked causes
- 1. ECG OUT cable defective or disconnected; MSL cable/connectors defective (common)
- 2. Power Switch/ECG OUT board defective (occasional)
- 3. Main board defective (rare)
Verify ECG OUT cable and MSL connections integrity
- Test point
- ECG OUT cable to defib and MSL cable/connectors
- Expected
- Cables properly connected and not defective
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Flow rate out of tolerance
Measured pump flow not within 50 +15/-7.5 ml/min range during flow rate check
Ranked causes
- 1. Flow setting requires adjustment (common)
- 2. Defective pump (occasional)
- 3. Fault in measurement extension (occasional)
Measure pump flow and adjust if needed
- Test point
- flowmeter display
- Expected
- 50 +15/-7.5 ml/min (i.e., 42.5 to 65 ml/min), +15/-7.5 ml/min
- Tools
- Electronic flowmeter M1026-60144 or 453564178121 Mass Flowmeter
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Bad Server Link issued caution
Monitor displays Bad Server Link alert indicating unexpected data on MSL
Ranked causes
- 1. Unexpected data detected on MSL (common)
- 2. Module Rack or Multi-Measurement Module with incompatible software revision connected (occasional)
- 3. Communication between components not functioning (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message Check Alarm Lamps issued caution
Monitor displays INOP alert for alarm lamp system
Ranked causes
- 1. Alarm LED board cable disconnected (common)
- 2. Alarm LED board defective (occasional)
- 3. Main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message Msmt Ext. Unsupp issued
Monitor displays unsupported measurement extension INOP message
Ranked causes
- 1. Unsupported Measurement Extension connected (common)
- 2. Measurement Extension is defective (occasional)
- 3. Multi-Measurement Module defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message Speaker Malfunction displayed
Speaker malfunction indicator shown on display without specific audio issue described
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on main board defective (occasional)
Verify speaker cable connection
- Test point
- Speaker cable connector
- Expected
- Cable fully seated and connected
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal Comm.Malf. INOP message
I2C bus communication failure in monitor
Ranked causes
- 1. Problem with I2C Bus communication (occasional)
- 2. Connector board defective (occasional)
- 3. Main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
iPC does not show on monitor after start
iPC boots but output not visible on patient monitor screen
Ranked causes
- 1. Power cable/connection defective (common)
- 2. iPC to Video Mixer Board cable not working (occasional)
- 3. Video Mixer Board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/Mouse attached to monitor not functioning
Keyboard or mouse connected directly to monitor not responding
Ranked causes
- 1. Keyboard/Mouse not connected properly (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. Connector board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/Mouse connected to iPC not functioning
Keyboard or mouse connected to iPC not responding
Ranked causes
- 1. Keyboard/Mouse not connected properly (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. iPC defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement module not recognized at front end caution
Module Rack LEDs ok but measurement module not recognized; no prompt or INOP
Ranked causes
- 1. Measurement module or connector defective (common)
- 2. MSL voltage out of range (occasional)
- 3. Motherboard or motherboard connector defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module Rack red error LED stays on caution
Red error LED continuously lit; unrecoverable hardware selftest error
Ranked causes
- 1. MSL cable defective (common)
- 2. Flex connector main board to MSL defective (occasional)
- 3. CPU module defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarmed but Nurse Call did not activate patient safety
Monitor generates alarm condition but Nurse Call system is not triggered or activated by the alarm
Ranked causes
- 1. Incorrect Nurse Call Relay configuration (latency, trigger settings) (common)
- 2. Cable connection to monitor or nurse call system incorrect (common)
- 3. Connector board or Flexible Nurse Call Relay I/O board defective (occasional)
Verify Nurse Call Relay configuration settings
- Test point
- Monitor configuration menu - Relay latency and trigger settings
- Expected
- Configuration matches Nurse Call system requirements
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity to PIC without error message
Monitor has no connection to Philips Information Center; no prompt or error message appears on monitor display
Ranked causes
- 1. Hardware defect (LAN cable or NGN board) (occasional)
- 2. Network switch failure (occasional)
- 3. Configuration problem (occasional)
Inspect LAN cable
- Test point
- Cable ends, connectors
- Expected
- No visible damage, corrosion, or loose pins
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No sounds from PC
iPC produces no audio output to patient monitor
Ranked causes
- 1. PC sounds disabled (common)
- 2. Flat ribbon cable not connected correctly (common)
- 3. Windows audio drivers incorrect or disabled (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Test Sound issued caution
Audio test tone does not sound during boot sequence when expected
Ranked causes
- 1. INOPs condition (common)
- 2. Speaker defective (occasional)
- 3. Main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Other bed overview not available
Other bed overview function is not available for remote monitoring of adjacent beds
Ranked causes
- 1. Configuration problem in PIC (common)
- 2. Switch multicast filter configuration issue (occasional)
- 3. Feature incompatible with IntelliVue Instrument Telemetry or legacy WLAN (occasional)
Verify PIC configuration for care groups
- Test point
- PIC management interface
- Expected
- Monitor and adjacent monitors assigned to same care group
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC does not start automatically or manually
iPC fails to boot or start when powered
Ranked causes
- 1. HW setting incorrect (common)
- 2. Flat ribbon cable not connected correctly (common)
- 3. iPC defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC HW Malfunction prompt issued
Hardware malfunction detected in iPC during operation
Ranked causes
- 1. Battery empty (common)
- 2. PC temperature too high (common)
- 3. PC defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power Loss Alarm Buzzer not working or test fails patient safety
Power loss alarm buzzer does not sound or fails self-test
Ranked causes
- 1. Power loss sound is disabled (common)
- 2. Cable connections between power supply and main board loose or disconnected (occasional)
- 3. Main board defective (occasional)
Verify power loss sound is enabled
- Test point
- Main Setup -> Global Settings -> Power Loss Sound
- Expected
- Power loss sound enabled
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED blinks (cyclic reboots) high voltage
Red Error LED blinks repeatedly, indicating the monitor is cycling through reboot attempts
Ranked causes
- 1. Hardware or Software Failure (common)
- 2. Software Fault detected by Support Tool (occasional)
- 3. Hardware Failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote Control not recognized
Wired remote control not detected by monitor
Ranked causes
- 1. Connector Board defective (rare)
- 2. Remote Control not connected per USB rules (common)
- 3. Remote Control defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Settings Malfunction INOP message
Configuration settings corrupted or cloning process failed
Ranked causes
- 1. Problem during cloning process (occasional)
- 2. Memory space corrupted (occasional)
- 3. Main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless Remote Control not functioning
Wireless remote does not respond or control monitor
Ranked causes
- 1. SRR option not enabled (common)
- 2. Low battery (common)
- 3. Remote Control defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AGM connected to RS232 port not functioning
External Alarm Group Module device connected to MIB/RS232 port is not operational or unresponsive
Ranked causes
- 1. MIB/RS232 board in wrong slot after software configuration change (common)
- 2. MIB/RS232 board or connector board defective (occasional)
Verify I/O board slot placement matches current software configuration
- Test point
- MIB/RS232 board position in monitor
- Expected
- Board in correct slot per current configuration
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs illuminate but no alarm sound patient safety
Remote alarm device displays LEDs but produces no audio
Ranked causes
- 1. Speaker defective (common)
- 2. Remote device I/F defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs remain off patient safety
Alarm indicator LEDs on front panel do not illuminate during boot sequence
Ranked causes
- 1. Alarm LED board is defective (common)
- 2. Main board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication dropouts or gaps in parameter waves; SRR Interference INOP may be issued patient safety
Intermittent communication loss or waveform display gaps; possible SRR Interference INOP
Ranked causes
- 1. RF Interferences present (common)
- 2. Too many SRR devices allocated to one SRR channel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message: Check Nurse Call Relay issued patient safety
Monitor displays INOP (inoperable) status message for Nurse Call Relay; relay functionality disabled or flagged as failed
Ranked causes
- 1. Connector board defective (common)
- 2. Flexible Nurse Call Relay I/O board defective (common)
Verify Nurse Call Relay board type and check for defects
- Test point
- Connector board or Flexible Nurse Call Relay I/O board
- Expected
- Board intact and functional
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IntelliVue 802.11 Bedside Adapter not yet operational caution
Wireless adapter shows as 'Off' in Wireless LAN menu but WLAN has not been disabled by setting
Ranked causes
- 1. Adapter initializing or not fully activated (occasional)
- 2. Wired LAN cable installed, disabling wireless (occasional)
Verify wireless LAN menu status
- Test point
- Main Setup > Network > WLAN Diagnostic > Wireless LAN
- Expected
- Status is 'On' (not 'Off')
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
LAN connection not functioning
Network interface on iPC not working
Ranked causes
- 1. Flat ribbon cable not connected correctly (common)
- 2. iPC defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leakage between pump outlet and mCO2 outlet detected
Flowmeter reading exceeds 4 ml/min when mCO2 outlet is blocked during leakage check part 1
Ranked causes
- 1. Poor tubing connections at pump outlet or extension gas outlet (common)
- 2. Leakage in tubing (occasional)
Inspect tubing connections at pump outlet and extension gas outlet
- Test point
- pump outlet connection, extension gas outlet connection
- Expected
- connections tight with no visible damage or disconnection
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement selection icon changes to SRR but assignment fails; SRR Interference INOP issued
SRR device assignment fails after icon changes to SRR; Interference INOP message displayed
Ranked causes
- 1. RF Interferences present (common)
- 2. Too many SRR devices allocated to one SRR channel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Multi-Measurement Module does not start but LEDs are normal caution
Module powers on with normal LED status but fails to initialize or communicate
Ranked causes
- 1. Communication lines in MSL cable or MSL connector broken (common)
- 2. Connector board, MSL2 board or main board defective (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- No Central Monitoring message (monitor not assigned)
- No network icon displayed
- No network icon or network icon flashes
- PC Operating System does not start up
- Print job could not be queued
- Recorder M1116C paper not feeding
- Recorder M1116C poor print quality
- Recorder paper not feeding properly
- Recorder poor print quality
- Remote printer unavailable
- SRR communication aborted; SRR Interference or SRR Invalid Chan INOP may be issued
- Troubleshooting INOP ECG Out Equipment Malfunction
- Windows operating system corrupted
- 802.11 Bedside Adapter configuration problem (WEP, WPA/WPA2 PSK)
- Alarm sound issued but no LEDs light up
- Applied part leakage current test and performance testing
- Calibration verification: CO2 measurement
- Calibration: Cardiac output measurement
- Calibration: CO2 gas measurement
- Calibration: Spirometry barometer check
- Calibration: Spirometry flow measurement
- Calibration: Spirometry leak test
- Calibration: Touchscreen
- Check Monitor Func INOP message
- Check Passwords INOP message
- ChkAudioSettings INOP issued
- Date and Time not displaying correctly
- DC/DC board refitting
- Display Resolution incorrect
- DVI, VGA, DisplayPort or USB not functioning
- ECG Out function does not function
- Final inspection after reassembly
- Front bezel refitting
- Gas analyzer module switched off
- Gas analyzer warmup in progress
- Gas module unplugged
- Housing top cover refitting
- Incompatible software version message
- INOP Message Serverlink Malf displayed; audible indicator beep every two seconds
- iPC menu not available
- Keypad flexible connection refitting
- Keypad rigid connection refitting
- MAC correction patient data missing
- MAC correction source error
- Main board refitting and connection
- MCC Unsupported INOP message
- Measurement module ignored
- Measurement selection icon does not change to SRR
- Measurement supported by server does not come up on monitor
- Module Rack red error LED flashes
- MSL Flex assembly reassembly
- MSL Power High INOP issued
- MSL Power Off INOP issued
- MSL Power Overload INOP issued
- Multi-Measurement Module does not start up; no LEDs active, no INOP or prompt displayed
- N2O measurement overrange
- N2O unable to measure
- N2O waveform clipped
- NBP pump replacement and reassembly
- No Central Monitoring with Management Association available
- No network icon or network icon flashes with no central association
- No Time Service detected
- Noise index exceeds tolerance
- O2 measurement overrange
- O2 unable to measure
- O2 waveform clipped
- O2 zero calibration failed
- Operating System crashes during startup
- Other bed alarms not appearing
- PC is unstable or shows colors incorrectly
- Prompt message Too many <label> modules connected issued
- Re-enable certificate check after troubleshooting
- Recorder does not lock into module rack
- Recorder M1116C does not lock into rack
- Recorder M1116C door open error
- Recorder M1116C paper out error
- Recorder M1116C recording content unexpected
- Recorder thinks door is open
- Recorder thinks out of paper
- Recorder will not run
- Remote Control cannot be assigned to monitor
- Remote printer reports but does not print
- Support Tool unavailable for remote calibration initiation
- Telemetry device using SRR not recognized by monitor
- Touch position invalid
- Touchscreen calibration initiated on multiple display system
- Touchscreen deactivated after power loss during calibration
- Touchscreen mis-calibrated due to incorrect point selection
- Undesired PC sounds from internal speaker
- Unrecognized measurement module
- Wireless out of range
- Zero calibration failure altitude-dependent error
- Zero calibration in progress
93 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Agent Label> Overrange | Anesthetic agent concentration exceeds measurement range | Agent concentration higher than maximum measurable value | warning |
Agent Label> Unable To Meas | Gas analyzer unable to measure anesthetic agent | Agent measurement failure | error |
Check Watertrap | Water trap requires inspection or emptying | Water trap full or sample/exhaust line disconnected | warning |
CO2 Equip Malf | CO2 parameter module malfunction | CO2 module failure or disconnection | error |
ECG Equip Malf | ECG parameter module malfunction | ECG module failure or disconnection | error |
Gas Analyzer> Occlusion | Gas sample line or exhaust line obstruction | Restricted or blocked gas flow through analyzer | error |
Gas Analyzer> Zero Failed | Gas analyzer zero calibration failed | Unable to complete zero calibration procedure | warning |
GM Accuracy? | Gas analyzer measurement accuracy may be reduced | Potential sensor or calibration accuracy degradation | warning |
GM Component Malf | Gas analyzer component malfunction | Internal gas analyzer component failure | error |
GM Malfunction | Gas analyzer hardware malfunction | General gas analyzer hardware failure | critical |
INOP Check Nurse Call Relay | Nurse Call Relay system inoperable; relay not functioning | Relay board defective, connector problem, or configuration error | warning |
INOP EcgOut Equip Malf | ECG OUT equipment malfunction; ECG output system not operational | ECG OUT cable disconnected, MSL connection failure, or Power Switch/ECG OUT board defective | error |
NBP Equip Malf | Non-invasive blood pressure module malfunction | NBP module failure or disconnection | error |
RESP Equip Malf | Respiration parameter module malfunction | Respiration module failure or disconnection | error |
SpO2 Equip Malf | SpO2 parameter module malfunction | SpO2 module failure or disconnection | error |
SpO2 Transduc Malf | SpO2 transducer malfunction | SpO2 transducer failure or disconnection | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI