Philips MX100: troubleshooting and error codes
The ServiceForge knowledge base holds 60 documented symptoms for the Philips MX100, with 131 ranked causes, 92 diagnostic checks with test points and expected values, and 12 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Combined power on/error LED remains off after pressing power on button patient safety
No LED illumination occurs when power button is pressed; monitor does not respond to power-on command
Ranked causes
- 1. Battery defective (common)
- 2. HIF board not connected or positioned correctly (common)
- 3. HIF board connector defective (occasional)
- 4. IIT board defective (occasional)
- 5. WLAN board defective (occasional)
- 6. Integrated measurements defective (occasional)
- 7. NBP pump defective (occasional)
- 8. Front-End Adapter board defective (occasional)
- 9. MSL board to mainboard connection issue (occasional)
- 10. MSL board defective (rare)
- 11. Mainboard defective (rare)
Check battery status and condition
- Test point
- Battery
- Expected
- Battery charged and functional
1 more check with pass and fail actions, 11 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External power LED does not light patient safety
LED remains off when external power source or host monitor is connected; battery may or may not be present
Ranked causes
- 1. External power connection not OK (common)
- 2. MSL cable defective (common)
- 3. IntelliVue Dock, external power supply, or power cord defective (occasional)
- 4. MSL board defective (occasional)
- 5. MSL board to mainboard connection issue (occasional)
- 6. HIF board not connected correctly (occasional)
- 7. HIF board connector defective (rare)
- 8. HIF board defective (rare)
- 9. Mainboard defective (rare)
Verify external power source is powered and connected
- Test point
- power cord, MSL cable
- Expected
- Connected and powered
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red power on/error LED stays on continuously patient safety
Power on/error LED illuminates red and remains red without cycling or blinking during operation
Ranked causes
- 1. IIT board defective (common)
- 2. WLAN module defective (common)
- 3. Integrated measurements defective (common)
- 4. NBP pump defective (common)
- 5. Front-End Adapter board defective (common)
- 6. Mainboard defective (occasional)
Disconnect IIT board
- Test point
- IIT board connector
- Expected
- Disconnected
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
X3 does not enter companion mode when connected to host monitor
X3 fails to establish companion mode connection with host monitor despite being connected
Ranked causes
- 1. Host monitor has incompatible software revision (common)
- 2. MSL connection not established correctly (common)
- 3. MSL board defective (occasional)
- 4. Connection between MSL board and mainboard (occasional)
- 5. X3 mainboard defective (occasional)
- 6. Host monitor MSL interface defective (occasional)
Verify host monitor software revision
- Test point
- Host monitor software version display
- Expected
- Software revision M.0 or higher
2 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No alarm speaker sound
Alarm occurs but no alarm sound issued
Ranked causes
- 1. Audible alarm indicators switched off (common)
- 2. Volume set to zero (common)
- 3. Speaker defective (occasional)
- 4. Sound amplifier on mainboard defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print output does not match expectations
Printouts are not as expected; incorrect formatting, resolution, or color in printed report
Ranked causes
- 1. Printer paper size not correctly configured (common)
- 2. Printer resolution not correctly configured (common)
- 3. Printer color support misconfigured; printer does not support color (occasional)
- 4. Printer not compatible with monitor (occasional)
Verify printer paper size configuration
- Test point
- Printer settings paper size parameter
- Expected
- Paper size matches inserted media type
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen not functioning patient safety
Touch input does not respond; touchscreen does not register or act upon finger or stylus contact; may be disabled
Ranked causes
- 1. Touchscreen functionality temporarily disabled (common)
- 2. Flex cable to touch defective or not positioned correctly (occasional)
- 3. Mainboard defective (rare)
- 4. Touch display assembly defective (rare)
Observe screen for disable message
- Test point
- Display screen
- Expected
- No 'Touch screen disabled' message visible
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm speaker malfunction message displayed
Display shows 'INOP Message Speaker Malfunction'
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker assembly defective (occasional)
- 3. Sound amplifier on mainboard defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Monitor Temp INOP message caution
Message indicates temperature inside monitor is too high; may indicate thermal stress or sensor malfunction
Ranked causes
- 1. Environment too hot (common)
- 2. Battery temperature too high (occasional)
- 3. Mainboard defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is blank patient safety
LCD display shows no image, content, or illumination even after boot completion; touchscreen area dark
Ranked causes
- 1. Flex cable to touch or display not connected (common)
- 2. Touch display assembly defective (occasional)
- 3. Mainboard defective (rare)
Inspect flex cable to display and touch
- Test point
- Flex cable connector at display and mainboard
- Expected
- Cable connected and seated properly
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MON Msmt Malf / X3 Msmt Malf INOP message patient safety
Measurement system malfunction; integrated measurements not functioning; may indicate sensor or interface problem
Ranked causes
- 1. Contact problem or defective cables (common)
- 2. Adapter board defective (occasional)
- 3. Mainboard defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor does not power on or respond
Device unresponsive after power-up, battery low or not charged, AC power issues
Ranked causes
- 1. Battery not adequately charged (common)
- 2. AC power cord not connected or AC outlet non-functional (common)
- 3. Device not properly connected to external power supply or host monitor (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor not charging on IntelliVue Dock; power LED green
MX100 or X3 not charging despite green power LED on IntelliVue Dock
Ranked causes
- 1. MSL connector defective (common)
- 2. IntelliVue Dock MSL board defective (occasional)
- 3. IntelliVue Dock defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No flexible sync out signal
System reports missing flexible sync out signal
Ranked causes
- 1. Flexible sync out not configured correctly (common)
- 2. Cable wrong or defective (occasional)
- 3. IntelliVue Dock defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No network connectivity; INOP 'Unsupported LAN' issued
No network icon or network icon flashes; no association to central station; IIT-related network problem
Ranked causes
- 1. WLAN and IIT both enabled simultaneously (common)
- 2. Incorrect RF access code; no IP address assigned (common)
- 3. Communication failure between monitor and IIT adapter (occasional)
Verify WLAN mode setting when IIT is installed
- Test point
- Network settings > WLAN Mode
- Expected
- WLAN Mode set to None if IIT is enabled
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No test sound issued patient safety
Speaker does not produce test tone during boot sequence; audio output absent or inaudible
Ranked causes
- 1. Speaker not connected (common)
- 2. Speaker defective (occasional)
- 3. Mainboard defective (rare)
Verify speaker connector is seated
- Test point
- Speaker connector on mainboard
- Expected
- Connector fully inserted and latched
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Parameter INOP message issued (CO2, ECG, NBP, Pressure, Resp, SpO2, Temperature Equip Malf) patient safety
Parameter equipment malfunction INOP messages indicate measurement block connection or hardware failure
Ranked causes
- 1. Loose or faulty connection between measurement block and mainboard (common)
- 2. Measurement block defective (occasional)
- 3. Mainboard or front end adapter board defective (occasional)
Verify measurement block connection to mainboard
- Test point
- Measurement block connector at mainboard
- Expected
- Connector properly seated and secure
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Settings Malfunction INOP message patient safety
Message indicates problem occurred during cloning process or settings storage corruption; configuration lost
Ranked causes
- 1. Cloning process error (common)
- 2. Settings memory corrupted (occasional)
- 3. Mainboard defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm lamp not lighting
Alarm occurs but LED does not light up
Ranked causes
- 1. Environmental lighting too bright (common)
- 2. Mainboard defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery incompatible with monitor
Monitor displays 'Batt Incompat' message with INOP tone
Ranked causes
- 1. Incorrect battery installed (common)
- 2. Communication problem between battery connection board and mainboard (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Monitor Func INOP message patient safety
Message indicates problem with too low voltages in monitor; alarms, display, or interfaces may not function correctly
Ranked causes
- 1. Low voltage on auxiliary boards (common)
- 2. Mainboard defective (occasional)
Disconnect and reconnect Front-End Adapter board cable
- Test point
- Front-End Adapter board connector
- Expected
- INOP message clears after reconnection
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device fails to authenticate to WLAN access point
Device shows 'Authenticating' state but fails to complete authentication; incorrect SSID, mode, country, or security settings; authentication server or WLAN controller errors
Ranked causes
- 1. Incorrect authentication credentials or certificate settings (common)
- 2. Device on WLAN controller exclude list due to failed credentials (occasional)
Verify connection status and SSID/mode/country/security settings
- Test point
- WLAN connection status display
- Expected
- Device shows 'Authenticating' state with correct SSID, Mode, Country, and Security settings
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor not charging on IntelliVue Dock; power LED off
MX100 or X3 not charging; power LED on IntelliVue Dock is off
Ranked causes
- 1. Power cord defective (common)
- 2. IntelliVue Dock defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Central Monitoring message; not assigned to sector
Monitor connected to LAN with IP address but not being monitored at central
Ranked causes
- 1. Monitor not assigned to sector (common)
- 2. Duplicate equipment label on network (occasional)
Performance test: IIT communication
Verify IIT (Instrument Telemetry) board installation and wireless connectivity to access point; test signal strength at distance
Ranked causes
- 1. IIT board not installed properly or RF access code incorrect (occasional)
- 2. IIT communication link defective or antenna damaged (rare)
Verify monitor is not on LAN before starting test
- Test point
- IntelliVue Dock/External Power Supply/host monitor connections
- Expected
- No LAN connection via dock, power supply or host monitor
8 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Print device shows unavailable status; printer job stalled
Status message 'Print device Remote 1/2/3 unavailable' issued; printer job is stalled or not printing expected output
Ranked causes
- 1. Print error on Information Center (PIIC/PIC iX) (common)
- 2. Network connection to Information Center not functioning (common)
Run test print on Information Center
- Test point
- Information Center print function
- Expected
- Test report prints successfully
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print job fails or cannot be queued
Print queue error or message 'Print job could not be queued' with no print device found
Ranked causes
- 1. Printer disabled in setup (common)
- 2. Paper size mismatch between printer and report (common)
Verify printer is enabled in setup
- Test point
- Setup Printers menu
- Expected
- Correct printer is enabled
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer power off or paper tray empty
Printing queued on Information Center but no report is printed; printer not switched on or paper tray is empty
Ranked causes
- 1. Printer power is off (common)
- 2. Printer paper tray is empty (common)
Verify printer power switch is on
- Test point
- Printer power switch and indicator light
- Expected
- Printer power on; ready indicator light active
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red power on/error LED blinks (cyclic reboots) patient safety
Power on/error LED flashes in cycling pattern indicating repeated reboot attempts; system does not stabilize
Ranked causes
- 1. Software failure (common)
- 2. Hardware failure (occasional)
Connect Support Tool and scan for devices
- Test point
- Host monitor MSL port
- Expected
- Device detected and Operating Mode shows Boot
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
WLAN and IIT both enabled; no network association
Monitor with WLAN enabled shows no network icon or flashing icon; no central station association
Ranked causes
- 1. WLAN and IIT both enabled simultaneously (common)
- 2. Monitor-to-WLAN board communication problem or low RSSI (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs remain off patient safety
Alarm indicator LEDs do not illuminate during boot sequence or operation; no visual alarm indications
Ranked causes
- 1. Mainboard defective (common)
Verify alarm LED illumination during boot
- Test point
- Alarm LEDs (blue, yellow, red)
- Expected
- LEDs test in sequence: blue on-off, yellow on-off, red on-off
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm sound too quiet
Alarm sounds but volume is lower than expected
Ranked causes
- 1. Speaker foil defective (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Assurance test: MSL connector condition caution
Visually inspect all MSL connector sockets on monitor, cables, modules and racks for physical damage
Ranked causes
- 1. MSL connector pins damaged, bent or jolted (occasional)
Inspect all MSL connector pins for jolting or deformation
- Test point
- cable/monitor/Module Rack/Multi-Measurement Module/X2/X3 connectors
- Expected
- No pins jolted, no pins bent inward or outward
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Empty INOP tone with flashing battery LED patient safety
Estimated remaining battery-powered operating time ≤10 minutes; alarms cannot be paused or switched off during this condition
Ranked causes
- 1. Battery nearly depleted (common)
Verify battery charge level
- Test point
- Battery LED and display status
- Expected
- Battery charge > 10 minutes operating time
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low
Monitor displays 'Batt Low' message with INOP tone; estimated operating time less than 20 minutes
Ranked causes
- 1. Battery charge insufficient (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery malfunction patient safety
Monitor displays 'Batt Malf' with INOP tone and battery LED flashing; alarms cannot be paused or switched off without AC power
Ranked causes
- 1. Battery status cannot be determined or communication failure (common)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not inserted
Monitor displays '!!!Insert Battery' with INOP tone; monitor connected to external power without battery fully inserted
Ranked causes
- 1. No battery in battery compartment (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery temperature too high
Monitor displays 'Check Batt Temp' message with INOP tone
Ranked causes
- 1. Monitor exposed to excessive heat (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration verification: CO2 measurement
Confirm CO2 calibration remained stable after calibration procedure using same test gases and apparatus
Ranked causes
- 1. Measurement extension values drift after calibration (unknown)
Reopen 5% calibration gas valve
- Test point
- flow regulator valve
- Expected
- 5% CO2 gas flowing, value stabilized
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: CO2 measurement
Verify CO2 gas measurement accuracy using 5% and 10% calibration gases; ensure displayed values match calculated mmHg values within tolerance
Ranked causes
- 1. Measurement extension not properly calibrated (unknown)
Wait after switching measurement extension on
- Test point
- extension power
- Expected
- At least 20 minutes elapsed
7 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger malfunction
Monitor displays 'Charger Malfunct' with INOP tone and battery LED flashing
Ranked causes
- 1. Battery charger malfunction (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Passwords INOP message patient safety
Message indicates technical or network defect has caused loss of password settings block; authentication system compromised
Ranked causes
- 1. Password settings block lost (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image sticking on LCD display
Static image displayed for long time on LCD screen causes temporarily retained ghost image on display
Ranked causes
- 1. Static image displayed for prolonged period (common)
Observe LCD display for ghost image or retained image
- Test point
- LCD display visual inspection
- Expected
- Display shows current screen content without ghost images
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor has no LAN connection when docked
Docked MX100/X3 shows no LAN connectivity
Ranked causes
- 1. LAN connector or MSL board defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG-OUT signal to defibrillator patient safety
System shows no ECG output signal being sent to defibrillator
Ranked causes
- 1. IntelliVue Dock defective or cable issue (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No network connection (no icon)
Monitor displays no network icon; no LAN connection or wireless AP access
Ranked causes
- 1. Network cable disconnected or no wireless access point (common)
Performance test: Cardiac output measurement
Verify cardiac output measurement accuracy using patient simulator with specified injection temperature and flow rate
Ranked causes
- 1. Cardiac output module measurement out of specification (unknown)
Connect patient simulator to C.O. module
- Test point
- patient cable connector
- Expected
- Patient simulator properly connected
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Performance test: Cardiac output service tool procedure version 1
Verify cardiac output module blood temperature measurement using service tool M1012-61601 without option C10
Ranked causes
- 1. Blood temperature sensor reading out of specification (unknown)
Connect C.O. interface cable
- Test point
- C.O. module connector
- Expected
- Interface cable connected in monitoring mode
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Performance test: Cardiac output service tool procedure version 2
Verify cardiac output module blood and injection temperature measurements using service tool M1012-61601 with option C10; verify transpulmonary method and catheter constant
Ranked causes
- 1. Temperature sensor or measurement method configuration out of specification (unknown)
Connect C.O. interface cable
- Test point
- C.O. module connector
- Expected
- Interface cable connected in monitoring mode
5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Performance test: IntelliVue Dock ECG synchronization patient safety
Verify ECG signal synchronization output to connected medical device when MX100/X3 connected to IntelliVue Dock
Ranked causes
- 1. ECG synchronization not functioning or synchronization output defective (unknown)
Connect patient simulator and medical device
- Test point
- ECG connector and flexible sync output
- Expected
- Patient simulator on ECG connector, medical device on flexible sync output
2 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Performance test: LAN communication
Verify LAN connectivity when MX100/X3 connected to IntelliVue Dock or External Power Supply with active LAN cable
Ranked causes
- 1. LAN network settings incorrect or DHCP/BOOTP server unavailable (occasional)
Verify LAN hardware connections
- Test point
- IntelliVue Dock or External Power Supply LAN port
- Expected
- MX100/X3 connected to dock or power supply with LAN cable connected
4 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Performance test: Lithium ion battery
Verify battery age and charge cycle count to assess remaining useful life; replace if approaching limits
Ranked causes
- 1. Battery age or cycle count exceeds service life limit (rare)
Access battery status information
- Test point
- monitor patient display battery symbol
- Expected
- Battery symbol accessible on screen
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Performance test: Microstream CO2 flow rate
Verify CO2 measurement extension flow rate matches specification using flow measurement display
Ranked causes
- 1. Flow rate out of specification (unknown)
Verify CO2 flow rate
- Test point
- flow measurement display (x2)
- Expected
- 50.0 ml/min (50+15/-7.5 ml/min tolerance), +15/-7.5 ml/min
Performance test: Microstream CO2 leakage
Verify CO2 measurement extension leakage is within acceptable limits using digital barometer readings
Ranked causes
- 1. Measurement extension has internal or external leak (unknown)
Monitor digital barometer reading change
- Test point
- digital barometer display (x1)
- Expected
- Reading change less than 20 mbar for 30 seconds, < 20 mbar
Performance test: Microstream CO2 noise
Verify CO2 measurement extension noise index is within acceptable limits
Ranked causes
- 1. Measurement extension noise level excessive (unknown)
Verify CO2 noise index
- Test point
- noise index display (x3)
- Expected
- Less than 3.0, < 3.0
Performance test: WLAN communication
Verify WLAN board installation and connectivity to wireless access point using configured parameters
Ranked causes
- 1. WLAN board not installed properly or network configuration incorrect (occasional)
Verify monitor is not on LAN before starting test
- Test point
- IntelliVue Dock/External Power Supply/host monitor connections
- Expected
- No LAN connection via dock, power supply or host monitor
5 more checks with pass and fail actions are in ServiceForge. Connect ServiceForge to your AI
Unsupported LAN message; no IP address
Monitor has LAN connection but cannot obtain IP address assignment
Ranked causes
- 1. DHCP or IP assignment failure (occasional)
WLAN configuration mismatch with WEP/WPA/WPA2
WLAN unable to connect with PSK security configuration mismatch
Ranked causes
- 1. Mode, SSID, Country or Security settings incorrect (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
WLAN Enterprise authentication failure
WLAN configuration problem using WPA or WPA2 Enterprise security
Ranked causes
- 1. Connection status or authentication settings mismatch (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
WLAN not yet operational
WLAN setup menu shows 'Off'; wireless adapter not operational
Ranked causes
- 1. WLAN hardware initialization in progress or connection issues (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
CO2 Equip Malf | CO2 measurement equipment malfunction | Connection failure between CO2 measurement block and mainboard or measurement block defect | error |
ECG Equip Malf | ECG measurement equipment malfunction | Connection failure between ECG measurement block and mainboard or measurement block defect | error |
NBP Equip Malf | Non-invasive blood pressure measurement equipment malfunction | Connection failure between NBP measurement block and mainboard or measurement block defect | error |
NBP Measure Failed | Non-invasive blood pressure measurement failed | NBP pump defective or measurement error during blood pressure acquisition | error |
<Pressure Label> Equip Malf | Pressure measurement equipment malfunction for configured pressure label | Connection failure between pressure measurement block and mainboard or measurement block defect | error |
RESP Equip Malf | Respiration measurement equipment malfunction | Connection failure between respiration measurement block and mainboard or measurement block defect | error |
SpO2 Equip Malf | Oxygen saturation measurement equipment malfunction | Connection failure between SpO2 measurement block and mainboard or measurement block defect | error |
<SpO2 label> Unkn. Sensor INOP | SpO2 sensor unknown; demo tool connected but monitor not in Demo Mode | Demo tool is connected but monitor is not in Demo Mode | error |
<SpO2 label> Upgrade INOP | SpO2 module firmware upgrade in progress but monitor exited Service Mode | Upgrade tool connected and firmware upgrade started in Service Mode, but monitor left Service Mode during upgrade | error |
SpO2 Transduc Malf | SpO2 transducer malfunction | Transducer connection or hardware failure | error |
<Temp Label> Equip Malf | Temperature measurement equipment malfunction for configured temperature label | Connection failure between temperature measurement block and mainboard or measurement block defect | error |
Unsupported LAN | Unsupported local area network configuration detected | Both WLAN and IIT modules are enabled simultaneously; incorrect RF access code; network setup failure | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.