Philips M1026B: troubleshooting and error codes
The ServiceForge knowledge base holds 5 documented symptoms for the Philips M1026B, with 9 ranked causes, 7 diagnostic checks with test points and expected values, and 21 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Span Check failed
Span Check procedure did not pass; errors detected in module measurement
Ranked causes
- 1. Leaks in reservoir bag or tubing (common)
- 2. Reservoir bag emptied before procedure complete (common)
- 3. Incorrect gas values entered (occasional)
- 4. Insufficient warm-up time (occasional)
- 5. Module measurement error or component failure (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Anesthetic Gas Module patient safety
Periodic verification and calibration of M1026B to ensure operation within specified limits. Required every 12 months or after repair.
Ranked causes
- 1. Routine maintenance and compliance requirement (common)
Perform leakage check
- Test point
- Gas system
- Expected
- No leaks detected
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Anesthetic Gas Module span caution
Verify span calibration of anesthetic gases using check gas and calibration equipment
Ranked causes
- 1. Span calibration drift or incorrect gas measurement (unknown)
Verify no reported errors on display
- Test point
- AGM display
- Expected
- No error messages displayed
2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic assembly replacement required
Pneumatic assembly needs to be replaced in the anesthetic gas module
Ranked causes
- 1. Worn or defective pneumatic assembly (unknown)
Verify all tubing connections to replacement assembly
- Test point
- Five tubing connections per diagram
- Expected
- All tubing connected correctly to designated outlets and sensors
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unit fails leak check patient safety
Leak check does not pass or shows error during process
Ranked causes
- 1. Blockage removed prematurely or not properly sealed (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
AGENT MIXTURE | Agent mixture detected instead of pure agent | Multiple agents or mixed gas detected | warning |
AGM ACCURACY / GAS AN ACCURACY ? | Gas measurement accuracy questionable | Measurement variance detected | warning |
AGM EQUIP MALF / GAS AN. EQUIP MALF | Equipment malfunction detected | Hardware communication or subsystem failure | error |
AGM NOT AVAIL. / GAS AN. NOT AVAIL. | Anesthesia gas module not available | Module not switched on or not properly connected | error |
AGM OCCLUSION / GAS AN. OCCLUSION | Gas flow occlusion detected | Blocked gas inlet, outlet, or internal flow path | error |
AGM SELFTEST / GAS AN: SELFTEST | Self-test in progress | Automatic self-test initiated at startup | info |
AGM UNABLE TO MEAS / GAS AN. UNABLE TO MEAS. | Unable to measure gas concentration | Sensor or measurement circuit failure | warning |
AGM ZERO FAILED / GA ZERO FAILED | Zero calibration failed | Failure during automatic zero-point calibration | error |
AGMZERO RUNNING / GAS. AN. ZERO RUNNG | Autozero calibration in progress | Automatic zero-point calibration sequence initiated | info |
AGT CHANGE SCALE / AGT REDUCE SIZE | Agent data cannot be displayed correctly due to scale configuration mismatch | Selected display scale does not match configured agent range | info |
<AGT> UNABLE TO MEAS. | Gas analyzer cannot currently measure the specified agent | During gas measurement in Normal mode | error |
AGT UNABLE TO MEASURE | Agent (anesthetic gas) data not valid | Gas analyzer unable to measure selected agent | error |
AWRR OVERRANGE | Respiration rate exceeds measurement range | Measured respiration rate higher than device maximum | warning |
CHECK AGENT | Selected agent does not match agent detected by gas analyzer | Agent mismatch between operator selection and analyzer detection | warning |
CO2 UNABLE TO MEASURE | CO2 measurement unavailable | CO2 sensor or measurement circuit failure | warning |
GAS CONTAMINANT | Gas contaminant has been detected | Contaminant detected in gas sample | warning |
INOP | Module inoperative on IntelliVue or CMS/V24/V26 connection | Communication or functional failure detected | error |
N2O UNABLE TO MEASURE | Nitrous oxide (N2O) data not valid | Gas analyzer unable to measure N2O | error |
O2 EQUIP MALF | O2 equipment malfunction | O2 module improperly configured | error |
O2 UNABLE TO MEASURE | O2 measurement unavailable | O2 sensor or measurement circuit failure | warning |
O2 ZERO FAILED | O2 zero calibration failed | O2 sensor calibration out of range | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.