Philips M1026B: troubleshooting and error codes

The ServiceForge knowledge base holds 5 documented symptoms for the Philips M1026B, with 9 ranked causes, 7 diagnostic checks with test points and expected values, and 21 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Span Check failed

Span Check procedure did not pass; errors detected in module measurement

Ranked causes

  1. 1. Leaks in reservoir bag or tubing (common)
  2. 2. Reservoir bag emptied before procedure complete (common)
  3. 3. Incorrect gas values entered (occasional)
  4. 4. Insufficient warm-up time (occasional)
  5. 5. Module measurement error or component failure (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Anesthetic Gas Module patient safety

Periodic verification and calibration of M1026B to ensure operation within specified limits. Required every 12 months or after repair.

Ranked causes

  1. 1. Routine maintenance and compliance requirement (common)
First check

Perform leakage check

Test point
Gas system
Expected
No leaks detected

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Anesthetic Gas Module span caution

Verify span calibration of anesthetic gases using check gas and calibration equipment

Ranked causes

  1. 1. Span calibration drift or incorrect gas measurement (unknown)
First check

Verify no reported errors on display

Test point
AGM display
Expected
No error messages displayed

2 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pneumatic assembly replacement required

Pneumatic assembly needs to be replaced in the anesthetic gas module

Ranked causes

  1. 1. Worn or defective pneumatic assembly (unknown)
First check

Verify all tubing connections to replacement assembly

Test point
Five tubing connections per diagram
Expected
All tubing connected correctly to designated outlets and sensors

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Unit fails leak check patient safety

Leak check does not pass or shows error during process

Ranked causes

  1. 1. Blockage removed prematurely or not properly sealed (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
AGENT MIXTUREAgent mixture detected instead of pure agentMultiple agents or mixed gas detectedwarning
AGM ACCURACY / GAS AN ACCURACY ?Gas measurement accuracy questionableMeasurement variance detectedwarning
AGM EQUIP MALF / GAS AN. EQUIP MALFEquipment malfunction detectedHardware communication or subsystem failureerror
AGM NOT AVAIL. / GAS AN. NOT AVAIL.Anesthesia gas module not availableModule not switched on or not properly connectederror
AGM OCCLUSION / GAS AN. OCCLUSIONGas flow occlusion detectedBlocked gas inlet, outlet, or internal flow patherror
AGM SELFTEST / GAS AN: SELFTESTSelf-test in progressAutomatic self-test initiated at startupinfo
AGM UNABLE TO MEAS / GAS AN. UNABLE TO MEAS.Unable to measure gas concentrationSensor or measurement circuit failurewarning
AGM ZERO FAILED / GA ZERO FAILEDZero calibration failedFailure during automatic zero-point calibrationerror
AGMZERO RUNNING / GAS. AN. ZERO RUNNGAutozero calibration in progressAutomatic zero-point calibration sequence initiatedinfo
AGT CHANGE SCALE / AGT REDUCE SIZEAgent data cannot be displayed correctly due to scale configuration mismatchSelected display scale does not match configured agent rangeinfo
<AGT> UNABLE TO MEAS.Gas analyzer cannot currently measure the specified agentDuring gas measurement in Normal modeerror
AGT UNABLE TO MEASUREAgent (anesthetic gas) data not validGas analyzer unable to measure selected agenterror
AWRR OVERRANGERespiration rate exceeds measurement rangeMeasured respiration rate higher than device maximumwarning
CHECK AGENTSelected agent does not match agent detected by gas analyzerAgent mismatch between operator selection and analyzer detectionwarning
CO2 UNABLE TO MEASURECO2 measurement unavailableCO2 sensor or measurement circuit failurewarning
GAS CONTAMINANTGas contaminant has been detectedContaminant detected in gas samplewarning
INOPModule inoperative on IntelliVue or CMS/V24/V26 connectionCommunication or functional failure detectederror
N2O UNABLE TO MEASURENitrous oxide (N2O) data not validGas analyzer unable to measure N2Oerror
O2 EQUIP MALFO2 equipment malfunctionO2 module improperly configurederror
O2 UNABLE TO MEASUREO2 measurement unavailableO2 sensor or measurement circuit failurewarning
O2 ZERO FAILEDO2 zero calibration failedO2 sensor calibration out of rangeerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.