Philips M1013A: troubleshooting and error codes

The ServiceForge knowledge base holds 13 documented symptoms for the Philips M1013A, with 21 ranked causes, 42 diagnostic checks with test points and expected values, and 34 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Measurement accuracy in doubt or INOP indication

Gas analyzer measurement accuracy suspect or INOP warning triggered; requires systematic troubleshooting via seven ordered tests

Ranked causes

  1. 1. Zero calibration drift or failure (common)
  2. 2. Leak in measurement circuit (common)
  3. 3. Pump malfunction (occasional)
  4. 4. Flow rate out of specification (occasional)
  5. 5. Pressure sensor failure (rare)
  6. 6. Component status failure (rare)
  7. 7. Gas calibration failure (rare)
First check

Zero calibration

Test point
Zero reference measurement
Expected
Successful zero calibration completion

6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Component Status Check fails

Status tab shows red indicators for Module INOP or Any Component Fail instead of green

Ranked causes

  1. 1. Pump malfunction (common)
  2. 2. Flowrate out of specification (common)
  3. 3. Gas analyzer hardware failure (rare)
First check

Wait for automatic zero calibration to complete

Test point
VISIA tool Status tab
Expected
First automatic zero calibration completes

4 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Annual checks

Annual service procedure for gas analyzer including pneumatic and gas calibration tests

Ranked causes

  1. 1. Routine annual maintenance required (common)
First check

Leak check

Test point
Pneumatic system
Expected
No leaks detected

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Electrical Safety Tests high voltage

Perform electrical safety checks according to IEC 60601-1/UL60601-1 standards using Metron Safety tester to verify instrument compliance and patient safety.

Ranked causes

  1. 1. Installation or maintenance service event requires safety validation (common)
First check

System Enclosure Leakage Current - Normal Condition

Test point
Exposed metal parts and patient environment connections
Expected
≤ 100mA, maximum
Tools
Metron Safety tester

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Gas analyzer accuracy verification caution

Verify measured gas concentrations match calibration gas canister values within tolerance limits using M2211A Flow Regulator and calibration gas bottle

Ranked causes

  1. 1. Gas analyzer readings outside tolerance limits (unknown)
First check

Set Breath Detection to Disabled

Test point
General Tab
Expected
Breath Detection = Disabled

14 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Gas analyzer flow rate caution

Adjust gas analyzer flow rate to within tolerance limits using flowmeter at inlet.

Ranked causes

  1. 1. Flow rate outside tolerance limits (unknown)
First check

Connect sample line to watertrap; check pressure difference

Test point
Sample cell and ambient
Expected
Pressure difference 80–100 mbar, 80–100 mbar
Tools
VISIA tool

3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure sensor amplification

Verify pressure sensor accuracy under negative pressure (amplification test) using digital pressure indicator.

Ranked causes

  1. 1. Pressure sensor offset or calibration drift (unknown)
First check

Amplification pressure test: apply negative pressure

Test point
Sample cell
Expected
Deviation < 10 mbar between pressure indicator and VISIA tool reading, < 10 mbar
Tools
Digital pressure indicator, leakage test kit M1013-9302A, VISIA tool

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Pressure sensor zero offset

Verify pressure sensor accuracy at zero (ambient pressure) using independent barometer.

Ranked causes

  1. 1. Pressure sensor offset or calibration drift (unknown)
First check

Zero pressure test: measure ambient pressure

Test point
Gas analyzer inlet
Expected
Deviation < 10 mbar between barometer and VISIA tool reading, < 10 mbar
Tools
Independent digital barometer, VISIA tool

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 CHANGE SCALE message displayed

O2 waveform clipped on display due to inappropriate scale selection

Ranked causes

  1. 1. Wave scale set too low for O2 signal (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 OVERRANGE - value exceeds measurement range

O2 numerics replaced by -?- and INOP tone; O2 value higher than maximum measurement range

Ranked causes

  1. 1. O2 measurement exceeds upper limit or sensor error (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 UNABLE TO MEAS - cannot measure O2

O2 numerics replaced by -?- and INOP tone; gas analyzer unable to measure O2

Ranked causes

  1. 1. O2 measurement circuit failure or sensor malfunction (occasional)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 ZERO FAILED - zero calibration failed

O2 numerics replaced by -?- and INOP tone; zero calibration attempt failed

Ranked causes

  1. 1. O2 sensor or calibration circuit failure (common)

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

VISIA tool connection issue

Serial communication not established automatically between PC and gas analyzer

Ranked causes

  1. 1. Incorrect COM port settings (common)
First check

Verify baud rate setting

Test point
COM port configuration
Expected
19200 bps

1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
AGENT CALCULATINGGas analyzer is calculating the agent concentrationAgent concentration measurement initiatedinfo
AGENT MIXTUREMore than one agent detected in gas sample; numerics shown with ? or -?-Multiple anesthetic agents presentwarning
<AGT> CHANGE SCALEWave of the agent shown is clipped (DES/ENF/HAL/SEV/ISO)Agent waveform exceeds display scalewarning
AGT MEAS MALFUNCTIONProblem with agent measurement; numerics replaced by -?-, INOP toneAgent measurement circuit failureerror
<AGT> OVERRANGEAgent value higher than measurement range; numerics replaced by -?-, INOP toneMeasured agent concentration exceeds maximum rangeerror
<AGT> UNABLE TO MEASGas Analyzer cannot measure the agent shown (DES/ENF/HAL/SEV/ISO); numerics replaced by -?-, INOP toneRequested agent measurement not configured or availableerror
AWRR OVERRANGEMeasured respiration rate exceeds maximum range; numerics shown with -?-, INOP toneRespiration rate measurement exceeds limiterror
CO2 CHANGE SCALECO2 waveform is clippedCO2 waveform exceeds display scalewarning
CO2 OVERRANGECO2 value higher than measurement range; numerics replaced by -?-, INOP toneMeasured CO2 concentration exceeds maximum rangeerror
CO2 UNABLE TO MEASGas Analyzer cannot measure CO2; numeric replaced by -?-, INOP toneCO2 measurement circuit failure or not availableerror
GM ACCURACY?Gas analyzer measurement accuracy may be reducedZero is suspendedwarning
GM ALARM SUPPRESSGas Analyzer alarms will be suppressed until breathing activity is first detectedSystem initialization or standby modeinfo
GM CAL RUNNINGInternal calibration of the gas analyzer is runningCalibration cycle initiatedinfo
GM CHECK WATERTRAPThe watertrap is fullWatertrap fluid level exceeds capacitywarning
GM COMPONENT MALFGas analyzer component malfunction; numerics shown with -?-Internal component failure detectederror
GM INCOMPATIBLEThis version of the Gas Analyzer is not supportedIncompatible firmware or hardware version detectederror
GM MALFUNCTIONProblem with the Gas Analyzer hardware; numerics replaced by -?-, INOP tone, Setup LED may blinkHardware fault detectederror
GM NO BREATHNo breath detected; et and in numerics show the same valueNo breathing activity detected at patientwarning
GM NOT AVAILABLEThe Gas Analyzer is either disconnected or switched off; INOP toneGas Analyzer offline or not connectedwarning
GM OCCLUSIONSample line or exhaust line tubing kinked or blocked; numerics replaced by -?-, INOP toneGas flow obstruction detectederror
GM STANDBYGas monitoring is in standby; alarms not activeStandby mode selected or during warmup phaseinfo
GM SWITCHED OFFGas analyzer switched off all internal components due to overheating; INOP toneInternal temperature exceeds safe limitcritical
GM WARMUPGas Analyzer has not reached operating temperature; numerics shown with -?-Device powered on, temperature not yet stabilizedinfo
GM ZERO FAILEDGas Analyzer zero calibration failed; numerics shown with -?-Zero calibration cycle failed to completeerror
GM ZERO RUNNINGAutozero in progress; numerics may show ? or -?- with INOP tone on retryZero calibration cycle initiatedinfo
MAC CHECK SOURCESNot all required measurements/values available for MAC calculation; INOP tone may appearRequired measurement source unavailable or invalidwarning
MAC CORRECTION?Enhanced MAC correction is on but patient age and/or temperature not availableMAC calculation requested without required parameterswarning
N2O CHANGE SCALEN2O waveform is clippedN2O waveform exceeds display scalewarning
N2O OVERRANGEN2O value higher than measurement range; numerics replaced by -?-, INOP toneMeasured N2O concentration exceeds maximum rangeerror
N2O UNABLE TO MEAS.Gas Analyzer cannot measure N2O; numerics replaced by -?-, INOP toneN2O measurement circuit failure or not availableerror
O2 CHANGE SCALEO2 waveform clipped due to display scale settingWave scale too small to display full O2 signal amplitudeinfo
O2 OVERRANGEO2 value exceeds maximum measurement rangeO2 concentration higher than instrument maximum specificationwarning
O2 UNABLE TO MEASGas analyzer cannot measure O2 concentrationO2 measurement circuit failure or sensor malfunctionerror
O2 ZERO FAILEDO2 zero calibration procedure failedO2 sensor or calibration circuit unable to establish zero referenceerror

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI

Community

0 fixes logged by technicians through their AI, with outcomes, for this model.

ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.