Philips IntelliVue: troubleshooting and error codes
The ServiceForge knowledge base holds 153 documented symptoms for the Philips IntelliVue, with 317 ranked causes, 35 diagnostic checks with test points and expected values, and 51 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
AC LED does not light up
AC power indicator LED remains off after pressing power on button
Ranked causes
- 1. AC Connection not ok (common)
- 2. LED defective (occasional)
- 3. Power Switch board not connected to main board (occasional)
- 4. Power supply defective (occasional)
- 5. Secondary main board defective (rare)
- 6. Primary Video Board defective (rare)
- 7. Backplane Bottom defective (rare)
- 8. Primary Main Board defective (rare)
Verify AC mains power available
- Test point
- AC power inlet
- Expected
- AC mains powered and connected
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External device not receiving data
External device does not receive data from monitor export
Ranked causes
- 1. MIB/RS232 port not configured for data export (common)
- 2. Wrong data export protocol driver configured (common)
- 3. Cable not connected or defective (common)
- 4. External device does not support protocol version (occasional)
- 5. Terminal concentrator incompatible with dynamic speed negotiation (rare)
- 6. MIB/RS232 board in wrong slot (occasional)
- 7. MIB/RS232 board defective (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green Power On LED and Red Error LED remain off after pressing power on button on power supply
Both green Power On and red Error LEDs fail to illuminate when power button is pressed on power supply
Ranked causes
- 1. Remote Devices connected (occasional)
- 2. Power Switch Micro Controller hung (occasional)
- 3. Power switch board not connected to main board (occasional)
- 4. Power Switch Board defective (occasional)
- 5. I/O Board defective (rare)
- 6. Primary Video Board defective (rare)
- 7. Primary Main Board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch Screen not functioning
Touch screen input is not responsive to touch input
Ranked causes
- 1. Touchscreen functionality has been temporarily disabled (common)
- 2. Touchscreen functionality has been permanently disabled (occasional)
- 3. External Touch cable not connected (occasional)
- 4. External Touch driver configuration (occasional)
- 5. Connector board defective (rare)
- 6. External touch defective (rare)
- 7. Main board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Indep. Dsp. Malfunc. INOP issued
INOP (Inoperative) message Indep. Dsp. Malfunc. displayed, indicating external touch display malfunction
Ranked causes
- 1. Video cable to external display not connected (common)
- 2. External display has no power (common)
- 3. External display is defective (occasional)
- 4. Video board defective (rare)
- 5. Main board defective (rare)
- 6. Second main display defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED stays on continuously
Red Error LED remains illuminated continuously, indicating a persistent system error
Ranked causes
- 1. External connected device defective (common)
- 2. I/O Board defective (occasional)
- 3. Backplane top defective or secondary main board defective (occasional)
- 4. Primary Video Board defective (rare)
- 5. Backplane Bottom defective (rare)
- 6. Primary Main board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpeedPoint not functioning on remote extension
SpeedPoint knob or joystick attached to remote extension device inoperative
Ranked causes
- 1. Remote extension device not connected to monitor (common)
- 2. SpeedPoint not connected properly (common)
- 3. SpeedPoint defective (occasional)
- 4. Remote extension device defective (occasional)
- 5. Remote device I/O board in wrong slot (occasional)
- 6. Remote device I/O board defective (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner cannot read barcodes
Scanner lights and sounds function but cannot decode barcodes
Ranked causes
- 1. Scanner held at wrong distance from barcode (common)
- 2. Scanner window dirty or blocked (common)
- 3. Barcode format not recognized (common)
- 4. Scanner not recognized by system (occasional)
- 5. Scanner mechanical problem (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Display is blank
Connected external display screen shows no video output
Ranked causes
- 1. Video cable to external display not connected (common)
- 2. External display has no power (common)
- 3. External display is defective (occasional)
- 4. Video board defective (rare)
- 5. Main board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard or mouse on remote extension box not functioning
Keyboard/mouse attached to remote extension box does not respond
Ranked causes
- 1. Remote extension box disconnected from monitor (common)
- 2. Keyboard/mouse defective (common)
- 3. Remote extension device defective (occasional)
- 4. Remote device I/O board in wrong slot (occasional)
- 5. Remote device I/O board defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No central monitoring possible
Monitor cannot connect to central station; wireless adapter yellow sync LED not on steady
Ranked causes
- 1. Monitor out of range of access point (common)
- 2. Wireless adapter has no power (common)
- 3. Firmware revision in adapter is wrong (occasional)
- 4. Wireless adapter defective (occasional)
- 5. Wrong configuration in wireless adapter or access point (common)
Verify monitor is within coverage area
- Test point
- Wireless adapter yellow sync LED
- Expected
- LED on steady
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Third main display not functioning properly (D80 Intelligent Display)
Third display (D80) is not working properly; INOP messages issued such as Intell Dsp Malf, Intell Dsp Missing, or Intell Dsp Unsupp
Ranked causes
- 1. D80 not connected properly or defective (common)
- 2. Monitor software incompatible with D80 (occasional)
- 3. Monitor cannot communicate with D80 (Check MCC) (occasional)
- 4. MSL Coupling cable is reversed (occasional)
- 5. MSL coupling cable connected to unsupported device (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AGM connected to RS232 port not functioning
AGM device connected to MIB/RS232 port does not operate
Ranked causes
- 1. MIB/RS232 port not configured for AGM (common)
- 2. Cable between AGM and monitor not connected or defective (common)
- 3. MIB/RS232 board in wrong slot (occasional)
- 4. MIB/RS232 board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no alarm sound issued
Alarm condition triggered but no audible alarm tone is heard
Ranked causes
- 1. Audible alarm indicators switched off (common)
- 2. Volume set to 0 (common)
- 3. Speaker defective (occasional)
- 4. Sound amplifier on main board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no LED lights up
Alarm is triggered but alarm LED board does not illuminate
Ranked causes
- 1. Environmental lighting too bright (common)
- 2. Alarm LED board cable disconnected (common)
- 3. Alarm LED board defective (occasional)
- 4. Main Board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm on screen but no remote device LED or sound patient safety
Alarm message displayed on monitor but remote alarm device inactive
Ranked causes
- 1. Alarm device cabling not connected (common)
- 2. Alarm device cabling defective (common)
- 3. I/O board defective (occasional)
- 4. Remote alarm device defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green Power On LED and Red Error LED remain off after pressing power on button on remote extension device
Both green Power On and red Error LEDs fail to illuminate when power button is pressed on remote extension or alarm device
Ranked causes
- 1. Cable to remote device defective or disconnected (common)
- 2. Remote device defective (occasional)
- 3. Power switch board not connected to main board (occasional)
- 4. Remote device interface board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard or mouse directly attached not functioning
USB or PS/2 keyboard/mouse connected directly to monitor not responding
Ranked causes
- 1. Keyboard/mouse not connected properly (common)
- 2. Keyboard/mouse defective (common)
- 3. PS/2 or USB I/O board in wrong slot (occasional)
- 4. PS/2 or USB I/O board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarmed but Nurse Call did not activate patient safety
System alarmed but Nurse Call Relay did not function
Ranked causes
- 1. Incorrect configuration of relay latency or trigger (common)
- 2. Cable connection incorrect (common)
- 3. Nurse Call Relay board in wrong slot (occasional)
- 4. Nurse Call Relay board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor lost connection to information center patient safety
Monitor disconnected from central information center system
Ranked causes
- 1. LAN cable disconnected (common)
- 2. Information center disconnected (occasional)
- 3. Network infrastructure failure (occasional)
- 4. Wireless monitor out of range (occasional)
No connectivity to PIC without error message patient safety
Monitor has no connection to Philips Information Center with no error message displayed
Ranked causes
- 1. Hardware defect (LAN cable or connector) (common)
- 2. NGN connector board defective (occasional)
- 3. Switch defective (occasional)
- 4. Network configuration problem (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG-OUT signal to defibrillator patient safety
ECG OUT board not sending signal to defibrillator; red LED behavior diagnostic indicator
Ranked causes
- 1. Cabling not connected (common)
- 2. MSL cable or connectors defective (occasional)
- 3. Device not connected to first MSL port (occasional)
- 4. ECG OUT board defective (occasional)
Observe red LED in ECG OUT section after power cycle
- Test point
- ECG OUT section red LED (visible when housing bottom removed)
- Expected
- Red LED switches on for about 1 second at power on
- Tools
- Housing bottom removal
1 more check with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print device local unavailable
Status message indicates Local 1 or Local 2 printer unavailable; printer job stalled
Ranked causes
- 1. Printer not switched on (common)
- 2. Printer paper tray empty (common)
- 3. Cabling not connected correctly (occasional)
- 4. Connector board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printouts not as expected
Print output does not match expected quality or appearance
Ranked causes
- 1. Printer paper size not correctly configured (common)
- 2. Printer resolution not correctly configured (common)
- 3. Printer color support configured incorrectly (common)
- 4. Printer not compatible (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder M1116C: not printing or communicating
Recorder M1116C does not print or communicate with system
Ranked causes
- 1. Poor connection to rack (common)
- 2. Only one recorder module may be used per monitor (common)
- 3. Monitor not configured properly (occasional)
- 4. Recorder defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder not communicating with system
Recorder module does not communicate with monitor system
Ranked causes
- 1. Poor connection to front-end FMS (common)
- 2. Only one recorder module may be used per monitor (common)
- 3. System not configured properly (occasional)
- 4. Too many modules connected (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs illuminate but no sound patient safety
Remote alarm device shows LED indication without audio output
Ranked causes
- 1. Wrong I/O slot assignment (common)
- 2. Speaker defective (common)
- 3. Remote device interface defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Barcode scanner not functioning
Scanner completely inactive with no lights, sounds or barcode reading capability
Ranked causes
- 1. Scanner not plugged into USB port (common)
- 2. Scanner not recognized by system (common)
- 3. Scanner mechanical problem (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gas flow performance
Annual verification that gas flow rate from 15210B calibration unit meets specification
Ranked causes
- 1. Gas flow outside specification (unknown)
- 2. Gas flow less than minimum specification (occasional)
- 3. Gas flow greater than maximum specification (occasional)
Pressure indicator reading
- Test point
- Front panel pressure indicator
- Expected
- Not in black zone
1 more check with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration problem with WPA Enterprise or WPA2 Enterprise security
Monitor cannot authenticate with WPA Enterprise or WPA2 Enterprise security
Ranked causes
- 1. Connection status shows 'Scanning'; Mode, SSID, Country or Security settings incorrect (common)
- 2. Connection status shows 'Authenticating'; device fails to authenticate despite correct settings (occasional)
- 3. Device on exclude list of WLAN Controller due to wrong credentials (rare)
Check connection status
- Test point
- Main Setup -> Network -> WLAN Diagnostics -> Conn.Status
- Expected
- Status shows 'Connected' or 'Authenticating'
2 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG Out/Marker In function not operating patient safety
ECG Out or Marker In function not working; hardware problem affecting data flow
Ranked causes
- 1. MSL cable defect (occasional)
- 2. ECG Out hardware malfunction in monitor (occasional)
- 3. MSL connector in measurement server defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frequent dropouts and network disconnects
Monitor loses network connection intermittently
Ranked causes
- 1. Excessive interference by radio equipment or microwave ovens (common)
- 2. System capacity exceeded in coverage area (occasional)
- 3. Configuration problem (common)
Read wireless adapter statistics
- Test point
- Wireless adapter RS232 or PIC logging
- Expected
- Normal signal strength and interference levels
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Bad Server Link issued
Unexpected data detected on MSL or incompatible MMS software revision connected to monitor
Ranked causes
- 1. Unexpected data on MSL (unknown)
- 2. Incompatible MMS software revision (unknown)
- 3. Communication between components not functioning (unknown)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message 'MsmtSrv not Supp' issued
System displays INOP message indicating Measurement Server not supported
Ranked causes
- 1. Wrong Software Revision (common)
- 2. Too many measurement servers connected (common)
- 3. Unsupported type of measurement server connected (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message Check Alarm Lamps issued
Alarm LED lamp check message displayed, alarm LEDs not functioning properly
Ranked causes
- 1. Alarm LED board cable disconnected (common)
- 2. Alarm LED board defective (occasional)
- 3. Main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message Speaker Malfunction displayed
Speaker malfunction message displayed; alarm speaker not functioning
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Internal communication failure caution
I2C bus communication malfunction in monitor
Ranked causes
- 1. External display defective or incompatible (common)
- 2. Video board defective (occasional)
- 3. Main board defective (occasional)
Test with alternative external display
- Test point
- Display connector
- Expected
- Monitor communicates with replacement display
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement module not recognized at front end
LEDs OK but measurement module or connector not recognized, no prompt or INOP message
Ranked causes
- 1. Measurement module or connector defective (common)
- 2. MSL voltage from monitor too high or too low (occasional)
- 3. Mother board or connector on mother board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Other bed overview not available
Bed overview function for other beds not accessible
Ranked causes
- 1. Configuration problem in PIC (common)
- 2. Switch multicast filter configuration (occasional)
- 3. Wireless configuration limitation (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power not applied or LEDs not illuminated high voltage
No lights on front panel when monitor is switched on and AC power connected.
Ranked causes
- 1. AC power cord not connected or plugged in (common)
- 2. Power switch not turned on (common)
- 3. AC Power LED should be green when powered (occasional)
Verify AC power cord connection
- Test point
- AC outlet and instrument connector
- Expected
- AC power cord firmly connected
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prompt 'Measurement Server Configuration not supported' message
System displays prompt message about unsupported MMS Extension configuration
Ranked causes
- 1. Unsupported MMS Extension connected (common)
- 2. MMS Extension is defective (occasional)
- 3. Measurement Server defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote control not recognized
Wired remote control not detected by monitor
Ranked causes
- 1. Remote control not plugged according to USB rules (common)
- 2. Connector board defective (occasional)
- 3. Remote control defective (occasional)
Verify remote control USB connection
- Test point
- USB connector on monitor
- Expected
- Remote control recognized by monitor
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Unsupported LAN INOP message patient safety
Unsupported LAN INOP message issued; network connectivity problem
Ranked causes
- 1. Network infrastructure failure (common)
- 2. Monitor connected to wrong network (occasional)
- 3. IP address conflict after reconfig (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
USB device LED off
LED next to USB port with connected device is off
Ranked causes
- 1. Device connected to port is not supported (common)
- 2. Printer connected to port not switched on (common)
- 3. Short-circuit condition (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
All USB board LEDs off
All LEDs on USB board are off even though devices are connected
Ranked causes
- 1. Connected devices drawing too much power (common)
- 2. Unsupported combination of devices (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP EcgOut EQUIP MALF error patient safety
ECG OUT equipment malfunction alarm; communication problem or board defect
Ranked causes
- 1. Communication problem or ECG OUT board defective (common)
- 2. Main Board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement malfunction INOP patient safety
Measurement system inoperative after measurement block or cable reassembly
Ranked causes
- 1. Measurement ribbon cable inserted incorrectly (common)
- 2. Measurement block inserted with gap (common)
Verify measurement ribbon cable orientation and connection
- Test point
- Measurement ribbon cable connector
- Expected
- Cable inserted in correct orientation per picture, fully seated
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement Server does not start; LEDs are normal
Measurement Server not starting up despite normal LED illumination
Ranked causes
- 1. Communication lines in MSL cable or MSL connector broken (common)
- 2. MSL connector board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MMS or parameter modules not recognized
MMS Extension or parameter modules not communicating with monitor after insertion.
Ranked causes
- 1. MMS or MMS Extension not inserted correctly (common)
- 2. Parameter modules not plugged in correctly to FMS (common)
Verify MMS/MMS Extension insertion
- Test point
- MMS and MMS Extension connectors
- Expected
- Modules fully seated with no gaps
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor internal voltage too low caution
Alarm lamps, display or interfaces not functioning correctly due to low 5V or 12V supply voltages
Ranked causes
- 1. Defective I/O board (common)
- 2. Main board defective (occasional)
Remove and reinsert I/O boards one at a time
- Test point
- I/O board slots
- Expected
- Monitor functions restored
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor overheating caution
Temperature inside monitor exceeds safe operating limits
Ranked causes
- 1. Monitor ventilation obstructed (common)
- 2. Main board defective (occasional)
Check environment and monitor ventilation
- Test point
- Monitor vents and surrounding area
- Expected
- Ventilation clear of obstructions
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MSL Flex creasing or damage during reassembly caution
MSL Flex cable creased or damaged during installation
Ranked causes
- 1. MSL Flex cable creased during 180-degree bending (occasional)
- 2. MSL Flex damaged when inserting frame connector (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No central assigned to bed prompt message patient safety
Prompt message no central assigned to this bed displayed on monitor
Ranked causes
- 1. Monitor label not set (label-based bed assignment) (common)
- 2. PIC to switch communication problem (port-mapped) (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No central monitoring active patient safety
Monitor connected to LAN with IP address but bed not monitored at central station
Ranked causes
- 1. Monitor not assigned to sector (common)
- 2. Duplicate equipment label on network (occasional)
No network LAN connection
Monitor has no wireless or wired network connection; unable to communicate
Ranked causes
- 1. Wireless monitor cannot find access point (common)
- 2. Wired monitor no LAN traffic detected (common)
No Test Sound issued
Test sound (beep) is not heard during boot sequence when expected
Ranked causes
- 1. Speaker defective (common)
- 2. Main board defective (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print device remote unavailable
Status message indicates Remote 1, 2 or 3 printer unavailable; printer job stalled
Ranked causes
- 1. Print error on Philips Information Center (occasional)
- 2. Network connection to Philips Information Center not functioning (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print job could not be queued
Prompt message issued indicating no print device found
Ranked causes
- 1. Printer disabled in Setup Printers menu (common)
- 2. Printer paper size does not match report setting (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prompt 'Measurement Server not supported' and INOP 'Bad Measurement Server'
System displays prompt message and INOP error for unsupported or defective measurement server
Ranked causes
- 1. M3000A Measurement Server Revision A plugged; not compatible with IntelliVue (rare)
- 2. Parameter board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder M1116C: paper not feeding properly
Recorder M1116C paper does not advance correctly
Ranked causes
- 1. Paper roll off center (common)
- 2. Defective roller (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Recorder M1116C: poor print quality
- Recorder: paper not feeding properly
- Recorder: poor print quality
- Red Error LED blinks (indicating cyclic reboots)
- Settings malfunction
- Touchscreen calibration required
- Touchscreen operation disabled
- Tympanic Thermometer does not function
- Unit does not calibrate accurately
- Unsupported LAN network detected
- A measurement supported by server does not appear on monitor
- Agent Calculating INOP message
- Agent Change Scale message
- Agent Mixture detected
- Agent Overrange INOP
- Agent Unable To Meas INOP
- AGT Meas Malfunct INOP
- Alarm occurs on VueLink device but no sound on monitor
- Alarm sound issued but no remote LEDs light
- awRR Overrange INOP
- Check Watertrap INOP
- CO2 Change Scale message
- CO2 Overrange INOP
- CO2 Unable To Meas INOP
- Configuration problem with WEP/WPA/WPA2 security
- Connected USB devices not functioning
- Detailed protocol problem
- ECG Out/Marker In function does not work
- FMS cable connection problem
- GM Accuracy? INOP message
- GM Alarm Suppress message
- GM Align Watertrap message
- GM Component Malf INOP
- GM Incompatible INOP
- GM Malfunction INOP
- GM No Breath message
- GM Occlusion INOP
- GM Standby message
- GM Switched Off message
- GM Unplugged message
- GM Warmup message
- GM Zero Failed INOP
- GM Zero Running message
- Image sticking on LCD display
- Indep DSP. NotSupp. INOP issued
- INOP Message Serverlink Malf displayed
- INOP message: check Nurse Call Relay
- INOP SpO2 label Unkn. Sensor issued
- INOP SpO2 label Upgrade issued
- Input devices INOP message issued
- IntelliVue 802.11 Bedside Adapter not yet operational
- LEDs on IntelliVue 802.11 Bedside Adapter are off and remain off
- MAC Check Sources INOP
- MAC Correction? message
- Measurement malfunction INOP after main board refitting
- Measurement module ignored in slot
- Measurement Server does not start up; no LEDs active
- MSL coupling cable unsupported
- MSL Power High INOP issued
- MSL Power Off INOP issued
- MSL Power Overload INOP issued
- N2O Change Scale message
- N2O Overrange INOP
- N2O Unable To Meas INOP
- NBP pump malfunction after replacement
- No connectivity in multi-access point coverage area
- No marker pulse displayed on monitor
- No network icon or flashing network icon
- O2 Overrange INOP
- O2 Unable To Meas INOP
- O2 wave clipped on display
- O2 Zero Failed INOP
- Other bed alarms not appearing
- Printing does not work on wireless network
- Printing on remote device but no output
- Prompt message 'Too many modules connected'
- Recorder M1116C: content not as expected
- Recorder M1116C: does not lock into rack
- Recorder M1116C: door open prompt when not open
- Recorder M1116C: out of paper prompt when not empty
- Recorder module does not lock into FMS
- Recorder won't run
- Recorder: system thinks door is open
- Recorder: system thinks out of paper
- Red error LED flashes
- Remote extension device buttons not functioning
- Remote input device not functioning
- SpeedPoint rotation or joystick not functioning
- Temperature reading unusually high
- Temperature reading unusually low
- Touch position invalid (D80 Intelligent Display)
- Unrecognized measurement module in slot
- Wireless monitor out of range
93 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
16400 | Monitor spontaneously reboots without user interaction | Unexpected system reboot | error |
16505 | Loss of configuration after system reboot | Monitor reboots; configuration data lost | error |
17345 | MMS reboots unexpectedly; no waves or numerics from MMS | Unexpected MMS reboot during monitoring | error |
32750 | Modules not recognized or monitor reboots upon module removal | Module removal or system boot with module not detected | error |
Agent Calculating | Gas analyzer calculating agent concentration; measurement in progress | Gas analyzer performing agent concentration calculation | info |
Agent Mixture | Multiple anesthetic agents detected in sample | Gas analyzer detects two or more anesthetic agents | warning |
Agent Overrange | Agent gas value exceeds maximum measurement range | Agent concentration reading above sensor maximum range | warning |
Agent Unable To Meas | Gas analyzer cannot measure agent gas | Sensor failure, occlusion, or calibration failure | error |
AGT Meas Malfunct | Agent measurement hardware malfunction | Agent sensor or measurement circuit failure | error |
awRR Overrange | Automatic respiration rate exceeds maximum measurable range | Measured respiratory rate above sensor maximum | warning |
Check Watertrap | Gas analyzer water trap full or disconnected | Water trap level high or sample line/water trap connection lost | warning |
CO2 Overrange | CO2 value exceeds maximum measurement range | CO2 concentration reading above sensor maximum range | warning |
CO2 Unable To Meas | Gas analyzer cannot measure CO2 | CO2 sensor failure or measurement circuit failure | error |
GM Accuracy? | Gas analyzer measurement accuracy may be reduced | Obstruction detected in gas inlet, outlet, or water trap tubing | warning |
GM Alarm Suppress | Gas analyzer alarms paused pending breath detection | Gas analyzer configured to suppress alarms until breathing detected | info |
GM Align Watertrap | Water trap not in correct vertical position | Water trap orientation incorrect | critical |
GM Component Malf | Gas analyzer internal component malfunction | Component hardware failure detected during operation | error |
GM Incompatible | Gas analyzer software version not supported by monitor | Incompatible gas analyzer firmware detected | critical |
GM Malfunction | Gas analyzer hardware malfunction detected | Gas analyzer hardware failure during operation | critical |
GM No Breath | No breathing activity detected by gas analyzer | No inspiratory or expiratory gas flow detected | warning |
GM Occlusion | Sample or exhaust line kinked or blocked | Obstruction detected in sample or exhaust gas flow path | error |
GM Standby | Gas analyzer in standby mode; monitoring suspended | Gas analyzer placed in standby by user or monitor | info |
GM Switched Off | Gas analyzer shut down due to overheating; all internal components disabled | Gas analyzer internal temperature exceeds safe operating limit | critical |
GM Unplugged | Gas analyzer module removed from monitor | Gas analyzer module physically disconnected | warning |
GM Warmup | Gas analyzer warming up; measurement accuracy not yet at specification | Gas analyzer power-on; sensor not yet stabilized | info |
GM Zero Failed | Gas analyzer zero calibration failed | Automatic or manual zero calibration did not complete successfully | error |
GM Zero Running | Gas analyzer automatic zero calibration in progress | Gas analyzer performing automatic zero calibration routine | info |
INOP Bad Server Link | Unexpected data on MSL or incompatible MMS software revision | Unexpected data detected on MSL line or incompatible MMS connected | error |
INOP Message Check Alarm Lamps | Alarm LED lamp monitoring circuit has failed | Alarm LED board communication loss or board failure | error |
INOP Message Serverlink Malf | Multi-Measurement Server communication hardware is faulty | Hardware failure in MSL communication path | critical |
INOP Message Speaker Malfunction | Speaker or speaker amplifier is not functioning | Speaker hardware failure or connection loss | error |
MAC Check Sources | Not all required measurements available for MAC calculation | One or more measurement sources missing, invalid, or questionable | warning |
MAC Correction? | Enhanced MAC correction enabled but required patient data not available | Enhanced MAC correction active but patient age and/or temperature not entered | warning |
N2O Overrange | Problem with N2O measurement; value may exceed range | N2O sensor or measurement circuit malfunction | error |
N2O Unable To Meas | Gas analyzer cannot measure N2O | N2O sensor failure or measurement circuit failure | error |
Noninvasive Blood Pressure Accuracy Test | Verify blood pressure reading accuracy within tolerance | During performance testing | warning |
Noninvasive Blood Pressure Leakage Test | Verify blood pressure cuff leakage within limits | During performance testing | warning |
Noninvasive Blood Pressure Linearity Test | Verify linear response across blood pressure measurement range | During performance testing | warning |
Noninvasive Blood Pressure Valve Test | Verify blood pressure valve performance | During performance testing | warning |
O2 Overrange | Problem with O2 measurement; value may exceed range | O2 sensor or measurement circuit malfunction | error |
O2 Unable To Meas | Gas analyzer cannot measure O2 | O2 sensor failure or measurement circuit failure | error |
O2 Zero Failed | O2 zero calibration failed | O2 sensor zero calibration did not complete successfully | error |
Power On Test | Device self-test completion verification | Power on the unit | error |
Safety Test (1): Protective Earth Resistance | Verify protective earth connection integrity | During safety testing | critical |
Safety Test (2): Equipment Leakage Current - Normal Conditio | Verify equipment leakage current within safe limits during normal operation | During safety testing | critical |
Safety Test (3): Equipment Leakage Current - Single Fault Co | Verify equipment leakage current within safe limits when protective earth is open | During safety testing with earth disconnected | critical |
Safety Test (4): Applied Part Leakage Current - Single Fault | Verify applied part leakage current within safe limits when mains is on applied part | During safety testing with simulated fault | critical |
System Test (Sys 1-2): Equipment Leakage Current | Verify system equipment leakage current in normal and single fault conditions | During system testing per IEC/EN 60601-1-1 or IEC/EN 60601-1+A1 Ed.3 clause 16 | critical |
System Test (Sys 3): Protective Earth Leakage Current | Verify protective earth leakage current in multi-component systems | When medical electrical system components connected to same Multiple Portable Socket Outlet | critical |
Temperature Performance Test | Verify temperature measurement accuracy at specified temperature | During performance testing | warning |
Visual Inspection | Initial device inspection for physical damage or defects | Test before putting into service, recurrent test, or after repair | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI