Philips IntelliVue X2: troubleshooting and error codes
The ServiceForge knowledge base holds 49 documented symptoms for the Philips IntelliVue X2, with 122 ranked causes, 15 diagnostic checks with test points and expected values, and 19 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Combined Power On/Error LED remains off
Power On/Error LED does not light after pressing power on button
Ranked causes
- 1. Battery defective (common)
- 2. Flex cable to HIF board not connected or not positioned correctly (common)
- 3. Flex cable to HIF board defective (occasional)
- 4. HIF board defective (occasional)
- 5. IIT module defective (occasional)
- 6. NBP pump defective (occasional)
- 7. MSL Power Board defective (occasional)
- 8. Main Board defective (occasional)
Verify battery is present and has charge
- Test point
- Battery compartment
- Expected
- Battery inserted and charged
2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Power LED does not light up
External Power LED remains off when connected to external power or host monitor
Ranked causes
- 1. External Power Connection not ok (common)
- 2. External power supply or power cable defective (common)
- 3. MSL cable defective (occasional)
- 4. Flex cable to HIF board not connected correctly (common)
- 5. Flex cable to HIF board defective (occasional)
- 6. HIF board defective (occasional)
- 7. MSL Power board defective (occasional)
- 8. Main Board defective (occasional)
Verify external power sources are powered
- Test point
- AC outlet and power sources
- Expected
- Power sources operational
2 more checks with pass and fail actions, 8 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no LED lights up patient safety
Alarm triggered but alarm lamp does not illuminate; visible alarm indication missing
Ranked causes
- 1. Environmental lighting too bright (occasional)
- 2. Flex cable to HIF board not connected or mispositioned (common)
- 3. Flex cable to HIF board defective (occasional)
- 4. HIF board defective (occasional)
- 5. Main Board defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG Sync Pulse
ECG synchronization pulse signal absent or not detected
Ranked causes
- 1. Wrong or defective ECG cable (common)
- 2. No ECG pulse signal available (occasional)
- 3. ECG OUT connector not connected (occasional)
- 4. ECG connector defective (occasional)
- 5. Main board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
X2 does not enter companion mode when connected to host monitor
IntelliVue X2 fails to establish Companion Mode connection with host monitor
Ranked causes
- 1. Incompatible software revision of host monitor (common)
- 2. MSL connection not established correctly (occasional)
- 3. MSL power board defective (occasional)
- 4. Main board defective (occasional)
- 5. MSL on host monitor defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs remain off
Alarm indicator LEDs do not illuminate during operation or alarm conditions
Ranked causes
- 1. Flex cable to HIF board not connected or not positioned correctly (common)
- 2. HIF board defective (occasional)
- 3. Flex cable to HIF board defective (occasional)
- 4. Main board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no alarm sound issued patient safety
Alarm triggered but no audible tone produced; silent alarm condition
Ranked causes
- 1. Audible alarm indicators switched off (common)
- 2. Volume set to 0 (common)
- 3. Speaker defective (occasional)
- 4. Sound amplifier on main board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Unsupported LAN issued
INOP displayed with network connection icons showing unsupported LAN configuration
Ranked causes
- 1. Network failure (common)
- 2. Monitor connected to wrong network (occasional)
- 3. IP address conflict after infrastructure re-installation (occasional)
- 4. IIT installed but no IIT infrastructure available (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printouts not as expected
Printed output quality, format, or appearance incorrect or not matching report settings
Ranked causes
- 1. Printer paper size not correctly configured (common)
- 2. Printer resolution not correctly configured (common)
- 3. Printer color support misconfigured (occasional)
- 4. Printer not compatible (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Power On/Error LED stays on continuously
Power On/Error LED illuminates red and does not switch to green
Ranked causes
- 1. Main board defective (common)
- 2. IIT module defective (occasional)
- 3. NBP pump defective (occasional)
- 4. MSL Power Board defective (occasional)
Disconnect IIT module and test
- Test point
- IIT module connector
- Expected
- Red LED clears when module disconnected
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch Screen not functioning
Touchscreen does not respond to touch input or position is incorrect
Ranked causes
- 1. Touchscreen functionality has been temporarily disabled (common)
- 2. Previously stored touch calibration is lost (common)
- 3. Flex cable to display defective or not positioned/connected correctly (occasional)
- 4. Main board defective (occasional)
Perform touch calibration
- Test point
- Touch screen surface
- Expected
- Touch calibration completes successfully
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Monitor Temp INOP caution
INOP message indicates temperature inside monitor is too high
Ranked causes
- 1. Environmental heat exposure (common)
- 2. Battery temperature is too high (common)
- 3. Main Board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration problem using WPA Enterprise or WPA2 Enterprise
Device fails to authenticate with Enterprise WPA/WPA2 security; stuck in Scanning or Authenticating state
Ranked causes
- 1. Incorrect Mode, SSID, Country, or Security settings (common)
- 2. Authentication server or WLAN controller issues (occasional)
- 3. Device excluded from WLAN Controller (rare)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display is blank
Monitor screen shows no information; boot phase completed without error
Ranked causes
- 1. Flex cable to display not connected (common)
- 2. Display assembly defective (occasional)
- 3. Main board defective (occasional)
Verify flex cable connection to display
- Test point
- Display flex cable connector
- Expected
- Flex cable fully seated and secure
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message Check Alarm Lamps issued patient safety
Alarm lamp circuit failure or display communication error; alarm lamp malfunction detected
Ranked causes
- 1. Flex cable to HIF board not connected or mispositioned (common)
- 2. HIF board defective (occasional)
- 3. Main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message Speaker Malfunction displayed patient safety
Speaker circuit failure or audio amplifier malfunction detected; speaker system error reported
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker assembly defective (occasional)
- 3. Sound amplifier on main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity to IIC, no prompt or error message
Monitor has no connection to IntelliVue Information Center with no error messages displayed
Ranked causes
- 1. Hardware Defect (occasional)
- 2. Network switch issue (occasional)
- 3. Configuration problem (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Test Sound issued
Speaker does not produce test beep or alarm sounds during operation
Ranked causes
- 1. Speaker not connected (common)
- 2. Speaker defective (occasional)
- 3. Main board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Settings Malfunction INOP
INOP message indicates problem during cloning process or corrupted memory space storing settings
Ranked causes
- 1. Problem during cloning process (common)
- 2. Memory space in which the settings are stored has been corrupted (common)
- 3. Main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR communication aborted; SRR Interference or SRR Invalid Chan INOP issued patient safety
SRR communication fails completely; SRR Interference or Invalid Channel alarm triggered
Ranked causes
- 1. RF Interference present (common)
- 2. Too many SRR devices on one channel (occasional)
- 3. SRR device out of range (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Incompat INOP patient safety
INOP tone indicates incompatible battery cannot be used with this monitor
Ranked causes
- 1. Indicated battery cannot be used with this monitor (common)
- 2. Communication problem between power board and main board (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Monitor Func INOP patient safety
INOP message indicates problem with too low voltages (5V, 12V) causing alarm lamps, display or interfaces to malfunction
Ranked causes
- 1. Problem with too low voltages (5V, 12V) (common)
- 2. Main board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication dropouts or gaps in parameter waves; SRR Interference INOP may issue patient safety
Intermittent loss of SRR signal; parameter waveforms show dropouts or gaps; sporadic SRR Interference alarm
Ranked causes
- 1. RF Interference present (common)
- 2. Too many SRR devices on one channel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement selection icon does not change to SRR
Short Range Radio interface not available or not responding; measurement selection icon remains unassigned to SRR device
Ranked causes
- 1. SRR configuration incorrect or interface defective (common)
- 2. SRR interface incompatible or defective in telemetry transceiver (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No central assigned to this bed prompt message
Prompt message indicates no central assigned to bed or communication issue
Ranked causes
- 1. Monitor label not set in monitor (beds monitored-labeled in IIC) (common)
- 2. Problem with IIC to Switch communication (beds port-mapped in IIC) (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No network icon or icon flashes, no association to central station
Monitor fails to establish or maintain network connection to central station; network icon absent or flashing continuously
Ranked causes
- 1. Communication problem between monitor and adapter or low RSSI signal (common)
- 2. Incorrect RF Access Code or No IP Address (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prompt message Print job could not be queued; no print device found
Print job fails to queue; system unable to locate or communicate with printer
Ranked causes
- 1. Printer disabled in Setup Printers menu (common)
- 2. Paper size mismatch (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Power On/Error LED blinks (cyclic reboots)
Red LED on power button blinks repeatedly indicating continuous reboot cycle
Ranked causes
- 1. Software Fault (common)
- 2. Hardware Failure (occasional)
Connect Support Tool and scan for defective devices
- Test point
- Monitor via crossover cable to Support Tool
- Expected
- Support Tool detects device; Operating Mode shows Boot
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR icon changes to SRR but device assignment fails; SRR Interference INOP issued
SRR device detectable but communication fails or interference detected; INOP Interference alarm triggered
Ranked causes
- 1. RF Interference present (common)
- 2. Too many SRR devices on one channel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Status message Printer Remote unavailable; Print job stalled
Print device on remote IntelliVue Information Center unavailable; print job cannot be transmitted to printer
Ranked causes
- 1. Print error on IntelliVue Information Center (common)
- 2. Network connection to IIC not functioning (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Status message Printing on Remote device issued but no report printed
Print command sent to remote device but output does not appear; print job failed on remote end
Ranked causes
- 1. Print queue on IIC full; printer not switched on (common)
- 2. Printer paper tray empty (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
!!!Insert Battery INOP patient safety
INOP tone indicates monitor connected to AC mains without battery fully inserted
Ranked causes
- 1. Battery required for sealing battery compartment (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm sound too quiet patient safety
Audible alarm tone present but at insufficient volume level
Ranked causes
- 1. Speaker foil defective (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Empty INOP patient safety
Battery low warning with INOP tone and flashing battery LED; alarms cannot be paused or switched off
Ranked causes
- 1. Estimated remaining battery-powered operating time is 10 minutes (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Low INOP patient safety
INOP tone indicates estimated battery-powered operating time remaining is less than 20 minutes
Ranked causes
- 1. Low battery charge (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Batt Malf INOP patient safety
INOP tone and flashing battery LED indicate cannot determine battery status or communication problem; alarms cannot be paused/switched off if not on AC power
Ranked causes
- 1. Cannot determine battery status or communication problem between battery and main board (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger Malfunct INOP caution
INOP tone and possibly flashing battery LED indicate problem with battery charger in monitor
Ranked causes
- 1. Problem with battery charger in monitor (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check Batt Temp INOP caution
INOP tone indicates battery temperature is too high
Ranked causes
- 1. Battery exposed to heat (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration problem using WEP, WPA(PSK), WPA2(PSK)
Wireless connection fails due to incorrect security mode, SSID, Country, or security settings
Ranked causes
- 1. Mismatch between device and network settings (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disable touchscreen operation temporarily
Press and hold Main Screen key to disable touchscreen; message 'Touch disabled, press Main Screen' appears on display
Ranked causes
- 1. Operator needs to lock touchscreen to prevent accidental input (common)
Verify touchscreen is locked
- Test point
- Main Screen key
- Expected
- Message 'Touch disabled, press Main Screen' appears on screen
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
IntelliVue 802.11 adapter not yet operational
Wireless LAN menu shows 'Off', indicating adapter not operational or wireless LAN disabled
Ranked causes
- 1. Wireless LAN adapter not operational (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid Leadset INOP issued patient safety
Telemetry device leadset not detected or incompatible
Ranked causes
- 1. Leadset incompatible or not installed (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Re-enable touchscreen operation
Press and hold Main Screen key again to resume normal touchscreen input
Ranked causes
- 1. Operator needs to restore touchscreen functionality after temporary lock (common)
Verify touchscreen is unlocked
- Test point
- Main Screen key and display response
- Expected
- Touchscreen becomes responsive; lock message disappears
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace Battery T or Battery Low T INOP issued patient safety
No battery inserted in telemetry device or battery low
Ranked causes
- 1. No battery in device or battery low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tele Disconnected INOP issued patient safety
Telemetry device connection lost or disrupted; SRR link between device and monitor broken
Ranked causes
- 1. SRR link between telemetry device and monitor disrupted (common)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Tele Equip Malf INOP issued patient safety
No stable connection between telemetry device and monitor; telemetry equipment malfunction
Ranked causes
- 1. No stable connection between telemetry device and monitor (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Telemetry device using SRR not recognized by monitor patient safety
Monitor does not detect or register telemetry device connected via Short Range Radio interface
Ranked causes
- 1. Telemetry device not supported by SRR adapter (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch Position invalid
Touch input is recognized but position coordinates are incorrect or calibration is lost
Ranked causes
- 1. Touch not calibrated (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touchscreen operation disabled
Monitor displays 'Touch disabled, press Main Screen' message and touchscreen is not responsive
Ranked causes
- 1. Touchscreen temporarily locked via Main Screen key (common)
Verify touchscreen lock status message on display
- Test point
- Monitor screen
- Expected
- Touch disabled, press Main Screen message visible
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Batt Empty | Estimated remaining battery-powered operating time is 10 minutes | Battery charge falls below 10-minute threshold | critical |
Batt Incompat | Indicated battery cannot be used with this monitor | Wrong battery model installed | error |
Batt Low | Estimated battery-powered operating time remaining is less than 20 minutes | Battery charge falls below 20-minute threshold | warning |
Batt Malf | Monitor cannot determine battery status or communication problem between battery and main board | Battery status indeterminate or communication failure during startup or operation | critical |
Charger Malfunct | Problem with battery charger in monitor | Charger malfunction detected | warning |
Check Batt Temp | Battery temperature is too high | Excessive battery temperature detected | warning |
Check Monitor Func | Problem with too low voltages (5V, 12V) in the monitor affecting alarm lamps, display or interfaces | Low voltage condition detected during operation | warning |
Check Monitor Temp | Temperature inside monitor is too high | Excessive internal temperature detected | warning |
CO2 Equip Malf | CO2 measurement equipment malfunction | CO2 sensor or module failure detected | warning |
ECG Equip Malf | ECG measurement equipment malfunction | ECG sensor or module failure detected | warning |
!!!Insert Battery | Monitor connected to AC mains without battery fully inserted; battery required for sealing compartment | AC power connected and battery compartment not sealed | critical |
NBP Equip Malf | Non-invasive Blood Pressure measurement equipment malfunction | NBP sensor or module failure detected | warning |
<Pressure Label> Equip Malf | Pressure measurement (generic label) equipment malfunction | Pressure sensor or module failure detected | warning |
RESP Equip Malf | Respiration measurement equipment malfunction | Respiration sensor or module failure detected | warning |
Settings Malfunction | Problem during cloning process or corrupted memory storing settings | Configuration file error or memory corruption detected | error |
SpO2 Equip Malf | Pulse oximetry equipment malfunction | SpO2 sensor or module failure detected | warning |
SpO2 Transduc Malf | SpO2 transducer malfunction | SpO2 transducer sensor or module failure detected | warning |
<Temp Label> Equip Malf | Temperature measurement (generic label) equipment malfunction | Temperature sensor or module failure detected | warning |
UNSUPPORTED LAN | Network configuration or connectivity issue preventing central monitoring | Network failure, wrong network, IP conflict, IIT infrastructure unavailable | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI