Philips IntelliVue MX700: troubleshooting and error codes
The ServiceForge knowledge base holds 103 documented symptoms for the Philips IntelliVue MX700, with 237 ranked causes, 25 diagnostic checks with test points and expected values, and 13 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
AC LED does not light up caution
AC Power LED is not illuminated when power is connected
Ranked causes
- 1. AC Connection not ok (common)
- 2. LED defective (occasional)
- 3. Power Switch board not connected to main board (occasional)
- 4. Power supply defective (occasional)
- 5. iPC defective (rare)
- 6. I/O board defective (rare)
- 7. Connector board defective (rare)
- 8. Panel adapter board defective (rare)
- 9. Mixer board defective (rare)
- 10. Main Board defective (rare)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Touch Display not functioning
External touchscreen does not respond to touch input
Ranked causes
- 1. Touchscreen temporarily disabled (common)
- 2. Touchscreen permanently disabled (occasional)
- 3. External Touch cable not connected (occasional)
- 4. External Touch driver configuration issue (occasional)
- 5. Connector board defective (rare)
- 6. External touch defective (rare)
- 7. Main board defective (rare)
- 8. Touch not calibrated (occasional)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External device not receiving data via USB
USB-connected external device not receiving monitor data
Ranked causes
- 1. Port not configured for data export (common)
- 2. Wrong protocol driver configured (common)
- 3. Cable not connected or defective (common)
- 4. Device does not support protocol version (occasional)
- 5. Terminal concentrator incompatible (occasional)
- 6. I/O board in wrong slot (occasional)
- 7. MIB/RS232 board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green Power On LED and Red Error LED remain off after pressing power on button caution
Neither Green Power On LED nor Red Error LED illuminate after power button is pressed
Ranked causes
- 1. Power Switch Micro Controller hung (common)
- 2. Power switch board not connected to main board (occasional)
- 3. Power Switch Board defective (occasional)
- 4. I/O Board defective (occasional)
- 5. Connector board defective (rare)
- 6. Mixer board, IIT, WLAN, panel adapter or Alarm LED board defective (rare)
- 7. Main board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Integrated display is blank or brightness is reduced caution
Display shows no content or display brightness is lower than normal, assuming boot completed without error
Ranked causes
- 1. Display brightness reduced by overheating (common)
- 2. Backlight Inverter Cable not connected (occasional)
- 3. Backlight tubes defective (occasional)
- 4. Backlight Inverter board defective (occasional)
- 5. Panel Adapter board defective (rare)
- 6. LCD Flat panel defective (rare)
- 7. Main board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch Screen not functioning
Integrated touchscreen does not respond to touch input
Ranked causes
- 1. Touchscreen temporarily disabled (common)
- 2. Touchscreen permanently disabled (occasional)
- 3. Touch screen not connected (occasional)
- 4. Panel adapter board defective (rare)
- 5. Touch screen defective (rare)
- 6. Main board defective (rare)
- 7. Touch not calibrated (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Display is blank caution
Slave Display shows no content or no picture output
Ranked causes
- 1. Integrated display also blank (common)
- 2. Video cable to external display not connected (occasional)
- 3. External display has no power (occasional)
- 4. External display is defective (occasional)
- 5. Connector Board defective (rare)
- 6. Main board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Display is blank (iPC connected) caution
External display connected to iPC shows no content
Ranked causes
- 1. Integrated display also blank (common)
- 2. Video cable to external display not connected (occasional)
- 3. External display has no power (occasional)
- 4. External display is defective (occasional)
- 5. iPC defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MMS Extension prompt or INOP message - unsupported configuration
Prompt message 'Measurement Server Configuration not supported' or INOP 'MsmtSrv not Supp' displayed
Ranked causes
- 1. Unsupported MMS extension connected (common)
- 2. MMS extension defective (occasional)
- 3. Measurement server defective (occasional)
- 4. Wrong software revision or too many servers connected (occasional)
- 5. Unsupported measurement server type (e.g. M3000A) (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED stays on continuously caution
Red Error LED does not turn off after power-on sequence
Ranked causes
- 1. External connected device defective (common)
- 2. I/O Board defective (occasional)
- 3. Connector board defective (rare)
- 4. Mixer board, IIT, WLAN, panel adapter or Alarm LED board defective (rare)
- 5. Main board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AGM connected to RS232 not functioning
AGM device connected to RS232 port not working
Ranked causes
- 1. RS232 port not configured for AGM (common)
- 2. Cable not connected or defective (common)
- 3. MIB/RS232 board in wrong slot (occasional)
- 4. MIB/RS232 board or connector defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm sounds but no LED indication
Alarm occurs but LED does not light up
Ranked causes
- 1. Environment lighting too bright (common)
- 2. Alarm LED board cable disconnected (occasional)
- 3. Alarm LED board defective (occasional)
- 4. Main board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm sounds without audio output
Alarm occurs but no alarm sound is issued
Ranked causes
- 1. Audible alarm indicators switched off (common)
- 2. Volume set to zero (common)
- 3. Speaker defective (occasional)
- 4. Sound amplifier on main board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Unsupported LAN message
Monitor displays 'Unsupported LAN' INOP message
Ranked causes
- 1. Network failure (common)
- 2. Monitor connected to wrong network (occasional)
- 3. IP address conflict after infrastructure re-installation (occasional)
- 4. IIT is enabled but no IIT infrastructure can be found (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer unavailable (Local)
Status message 'Print device Local 1 (Local 2) unavailable'; printer job stalled
Ranked causes
- 1. Printer not powered on (common)
- 2. Printer paper tray empty (common)
- 3. Cabling not connected correctly (occasional)
- 4. Connector board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printouts not as expected
Printer output quality or appearance issues
Ranked causes
- 1. Paper size not correctly configured (common)
- 2. Printer resolution not correctly configured (common)
- 3. Color support incorrectly configured (occasional)
- 4. Printer not compatible (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder not communicating
Recorder not communicating with system
Ranked causes
- 1. Poor connection to FMS (common)
- 2. Too many recorder modules (occasional)
- 3. System not configured properly (occasional)
- 4. Too many modules connected (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm Lamp INOP message
Check Alarm Lamps INOP message displayed
Ranked causes
- 1. Alarm LED board cable disconnected (common)
- 2. Alarm LED board defective (occasional)
- 3. Main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Configuration problem - WPA Enterprise or WPA2 Enterprise
Monitor unable to authenticate using WPA Enterprise or WPA2 Enterprise security; device shows 'Scanning' or 'Authenticating' state
Ranked causes
- 1. Mode, SSID, Country or Security settings mismatch during Scanning state (common)
- 2. Authentication server or WLAN controller error during Authenticating state (occasional)
- 3. Device excluded from WLAN controller or invalid certificates (rare)
Check connection status for Scanning state
- Test point
- Main Setup -> Network -> WLAN Diagnostics -> Conn.Status
- Expected
- State not 'Scanning'
4 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG Out function not operating caution
ECG output function does not work; signal may not reach defibrillator
Ranked causes
- 1. MSL cable issue (common)
- 2. Power Switch/ECG OUT board faulty (occasional)
- 3. MSL connector in measurement server faulty (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Bad Server Link
INOP Bad Server Link message displayed; indicates communication or compatibility issue between monitor and FMS/MMS modules
Ranked causes
- 1. Unexpected data detected on MSL bus or communication failure (common)
- 2. FMS or MMS with incompatible software revision connected (occasional)
- 3. Communication between components not functioning (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/Mouse not functioning (direct attachment)
Keyboard or mouse connected directly to monitor is unresponsive
Ranked causes
- 1. Keyboard/Mouse not connected properly (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. Connector board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/Mouse not functioning (iPC connection)
Keyboard or mouse connected to iPC is unresponsive
Ranked causes
- 1. Keyboard/Mouse not connected properly (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. iPC defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Module not recognized with LEDs functional
Front End Measurement Module not recognized; FMS LEDs are normal
Ranked causes
- 1. Module or connector defective (common)
- 2. MSL voltage out of range (occasional)
- 3. Mother board or connector faulty (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarmed but Nurse Call not activated
Monitor alarm occurs but Nurse Call relay does not activate
Ranked causes
- 1. Relay configuration incorrect (common)
- 2. Cable connection not correct (common)
- 3. Connector or Relay board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity to PIC, no prompt or error message
Monitor has no connection to Philips Information Center with no error displayed
Ranked causes
- 1. Hardware Defect (occasional)
- 2. NGN Connector board defective (occasional)
- 3. Switch defective or misconfigured (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No sounds from PC
PC audio is not being output through monitor speakers
Ranked causes
- 1. PC sounds disabled (common)
- 2. Flat ribbon cable to iPC not connected correctly (common)
- 3. Windows audio configuration or drivers not correct (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Test Sound issued
Test sound is not played during boot phase
Ranked causes
- 1. INOP condition present (common)
- 2. Speaker defective (occasional)
- 3. Main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Other Bed Overview not available
Other Bed Overview feature is not accessible on monitor
Ranked causes
- 1. Configuration Problem (common)
- 2. Switch multicast filters not configured (occasional)
- 3. Function not available for IntelliVue Instrument Telemetry or early IIC with WLAN (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC does not start automatically and/or cannot be started manually
iPC fails to power on or start automatically
Ranked causes
- 1. HW setting incorrect (common)
- 2. Flat ribbon cable to iPC not connected correctly (common)
- 3. iPC defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC HW Malfunction prompt message
Monitor displays 'PC HW Malfunction' error message
Ranked causes
- 1. Battery empty (common)
- 2. PC temperature is too high (common)
- 3. PC defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED stays on
Red Error LED continuously lit on Flexible Module Rack
Ranked causes
- 1. MSL cable defective (common)
- 2. Flex connector to MSL defective (occasional)
- 3. CPU module defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote Control not recognized by monitor
Wired remote control does not respond or is not detected
Ranked causes
- 1. Connector Board defective (common)
- 2. Remote Control not plugged according to USB rules (occasional)
- 3. Remote Control defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker Malfunction INOP message
INOP Speaker Malfunction message displayed
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR communication aborted or Invalid Chan INOP
Short Range Radio communication aborted; SRR Interference or SRR Invalid Chan INOP message displayed
Ranked causes
- 1. RF interference in environment (common)
- 2. Too many SRR devices allocated to one channel (occasional)
- 3. SRR device out of range (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs remain off caution
Alarm LED indicators do not light up during boot or operation
Ranked causes
- 1. Alarm LED board is defective (common)
- 2. Main board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IIT adapter - no network icon or flashing
Monitor does not display network indicator or indicator flashes; no association to central station via IIT adapter
Ranked causes
- 1. Communication problem or MAC Instr. Tele. field is 0000 0000 0000 (common)
- 2. Incorrect RF Access Code or no IP address (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
LAN connection not functioning
Network connection is not operational
Ranked causes
- 1. Flat ribbon cable to iPC not connected correctly (common)
- 2. iPC defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement Server starts but no LEDs active
Measurement server powers up but LED indicators do not illuminate
Ranked causes
- 1. MSL cable or connector communication lines broken (common)
- 2. Connector board or main board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Navigation Point not functioning
Navigation Point input device does not respond
Ranked causes
- 1. Navigation Point not connected properly (common)
- 2. Navigation Point defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Central Monitoring message issued patient safety
Monitor displays 'No Central Monitoring' INOP message
Ranked causes
- 1. Monitor is not assigned to a sector (common)
- 2. Another monitor on network has same Equipment Label (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No network icon or network icon flashes
Monitor does not display network connectivity indicator, or indicator flashes intermittently, with no association to central station
Ranked causes
- 1. Communication problem or RSSI value below 30 (common)
- 2. Wireless adapter not yet operational (occasional)
Verify RSSI indicator is rotating
- Test point
- Main Setup -> Network -> WLAN Diagnostic -> RSSI
- Expected
- Indicator rotating
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse Call Relay INOP
INOP message 'Check Nurse Call Relay' displayed
Ranked causes
- 1. Connector board defective (common)
- 2. Flexible Nurse Call Relay board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
PC Operating System does not start up
Operating system fails to boot after power on
Ranked causes
- 1. Cable connection to hard disk loose (common)
- 2. Hard disk defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print job queuing error
Prompt message 'Print job could not be queued'; no printer found
Ranked causes
- 1. Printer disabled in setup (common)
- 2. Printer paper size mismatch (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer unavailable (Remote)
Status message 'Print device Remote 1/2/3 unavailable'; printer job stalled
Ranked causes
- 1. Print error on Philips Information Center (common)
- 2. Network connection not functioning (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Prompt message 'no central assigned to this bed' patient safety
Monitor displays 'no central assigned to this bed' message
Ranked causes
- 1. Monitor label is not set in monitor (common)
- 2. Problem with Philips Information Center to Switch communication (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder paper not feeding properly
Paper feed problem in recorder
Ranked causes
- 1. Paper roll off center (common)
- 2. Dirty roller (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder poor print quality
Print quality from recorder is poor
Ranked causes
- 1. Printhead dirty (common)
- 2. Printhead failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED blinks (cyclic reboots) caution
Red Error LED blinks indicating device is rebooting repeatedly
Ranked causes
- 1. Hardware Failure (common)
- 2. Software Fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Remote Control not functioning
Remote control does not respond to user input
Ranked causes
- 1. Low battery (common)
- 2. Remote Control defective (occasional)
Verify battery presence and condition
- Test point
- battery compartment
- Expected
- battery installed and not depleted
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Serverlink hardware failure
INOP Message 'Serverlink Malf' displayed with audible beep every two seconds
Ranked causes
- 1. MSL cable faulty (common)
- 2. Connector board faulty (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR communication dropouts or gaps in parameter waves patient safety
Measurement data from SRR connection exhibits dropouts or gaps; SRR Interference INOP may be issued
Ranked causes
- 1. RF interference in environment (common)
- 2. Too many SRR devices allocated to one channel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR icon changes to SRR but assignment fails or SRR Interference INOP
Monitor measurement selection icon changes to SRR but device assignment fails; SRR Interference INOP message displayed
Ranked causes
- 1. RF interference in environment (common)
- 2. Too many SRR devices on one channel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR measurement selection icon does not change to SRR
Monitor unable to assign Short Range Radio (SRR) device; measurement selection icon does not switch to SRR mode
Ranked causes
- 1. SRR configuration settings incorrect or interface not installed (common)
- 2. SRR interface defective or incompatible (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration unit contains air before first use
Initial air in unit before first operation causes inaccurate calibration results
Ranked causes
- 1. Air trapped in new calibration unit (common)
Verify gas pressure indicator reading before calibration
- Test point
- Pressure indicator on front panel
- Expected
- Not in black zone
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Basic performance assurance patient safety
Basic system performance test to verify connected parameter modules display values in demo mode
Ranked causes
- 1. Routine performance verification (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Cover removal
Procedural steps to remove calibration unit cover for access to internal flow restriction components
Ranked causes
- 1. Cover access needed for internal adjustment (unknown)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: ECG and respiration performance patient safety
Detailed ECG and respiration measurement performance test using patient simulator to verify signal quality and heart rate accuracy
Ranked causes
- 1. Performance assurance testing required (common)
Connect patient simulator to ECG/Resp connector
- Test point
- ECG/Resp connector on MMS/IntelliVue X2
- Expected
- Patient simulator connected
3 more checks with pass and fail actions, 1 corrective action, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gas flow adjustment
Gas flow rate exceeds permitted maximum on Philips 15210B calibration unit; requires restriction adjustment and screw tuning
Ranked causes
- 1. Gas flow restriction insufficient (unknown)
Gas flow performance check
- Test point
- Calibration unit outlet
- Expected
- Within permitted maximum
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Calibration: Gas Flow Performance Check
- Calibration: Microstream CO2 measurement
- Calibration: Noninvasive blood pressure measurement
- Calibration: Regulator/control block removal
- Calibration: Timer control knob removal
- Calibration: Touchscreen
- Configuration problem - WEP, WPA(PSK), WPA2(PSK)
- Date and Time not displaying correctly
- Display Resolution incorrect
- DVI, VGA or USB connections not functioning
- ECG Out equipment malfunction
- Gas outlet blockage or deposits
- Gas pressure indicator in black zone
- INOP Message Check Input Devices issued
- iPC menu not available
- LCD display image sticking
- Measurement module ignored in slot
- Measurement module label conflict
- Measurement Server does not start - no LEDs active
- MSL Power High INOP
- MSL Power Off INOP
- MSL Power Overload INOP
- Navigation Point Knob Rotation or Selection control not functioning
- No connected USB devices functioning
- No ECG-OUT signal to defibrillator
- Operating System crashes during startup
- Other Bed Alarms not appearing
- PC is unstable or shows colors incorrectly
- Printing on remote device but no output
- Recorder door switch issue
- Recorder module does not lock into FMS
- Recorder paper-out sensor issue
- Recorder won't run
- Red Error LED flashes
- Remote Control cannot be assigned to a monitor
- Status Message Incompatible SW Revision versions
- Telemetry device not recognized by monitor via SRR
- Too many modules of same type connected
- Touchscreen operation disabled permanently
- Touchscreen operation disabled temporarily
- Undesired PC sounds from internal speaker
- Unrecognized measurement module in slot
- VueLink alarm without audio on monitor
43 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
Bad Server Link INOP | Communication failure between monitor and measurement server (FMS or MMS) | Data error on MSL bus or software version mismatch | error |
Check internvoltage | Problem with too low voltages in monitor (5V, 12V) | Monitor internal voltage check fails | error |
Check Monitor Func | Alarm lamps, display or interfaces may not function correctly | Too low internal voltages detected | error |
Check Monitor Temp | Temperature inside monitor is too high | Internal temperature exceeds safe operating range | critical |
Internal Comm.Malf. | Problem with the I2C Bus communication in the monitor | I2C Bus communication failure detected | error |
MCC Unsupported | MSL coupling cable connected to device which does not support MSL coupling | MSL coupling cable connected to unsupported device | warning |
Measurement Server Configuration not supported | Unsupported MMS extension or configuration has been connected | Incompatible MMS extension module detected at startup | warning |
Measurement Server not supported, unplug device, switch moni | Critical incompatibility with connected measurement server module | M3000A Revision A detected or parameter board failure | critical |
MSL Power High INOP | Module supply line (MSL) bus is drawing excessive current but below shutdown threshold | Total current draw from connected FMS and MMS exceeds design limit but monitor remains powered | warning |
MSL Power Off INOP | Module supply line (MSL) power has been disabled by monitor to protect system from overload | Excessive current draw on MSL for sustained period (>15 minutes) | error |
MSL Power Overload INOP | Short circuit or dangerous condition detected on MSL bus | Shorted connector or internal fault in connected module | error |
MsmtSrv not Supp | Unsupported measurement server type or too many servers connected | M3000A or other incompatible model connected; >1 measurement server with monitor | warning |
Settings Malfunction | Problem during configuration cloning process or corrupted settings memory | Settings memory corrupted or clone process failed | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI
Community
0 fixes logged by technicians through their AI, with outcomes, for this model.
ServiceForge knowledge base record. Field-unverified unless technicians confirmed it. A teaching summary of the record, not a service manual. Export 2026-10-14.