Philips IntelliVue MX40: troubleshooting and error codes
The ServiceForge knowledge base holds 57 documented symptoms for the Philips IntelliVue MX40, with 68 ranked causes, 7 diagnostic checks with test points and expected values, and 61 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Telemeter weak signal reception
Patient at outer range of radio coverage or receiving weak signal with high data loss from access point. May indicate AP issue.
Ranked causes
- 1. Patient outside optimal coverage area (common)
- 2. MX40 reception or RF hardware failure (occasional)
- 3. Access point coverage degraded in area (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display does not react to touch
MX40 touchscreen unresponsive to finger input
Ranked causes
- 1. Screen is locked (common)
- 2. User not using finger to touch screen (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Display does not turn on
MX40 screen remains blank when powered on
Ranked causes
- 1. AA Battery Tray inserted backwards (common)
- 2. User not holding Main Screen button long enough (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leadset unplugged patient safety
Patient cable unplugged from MX40 or incompatible leadset attached to patient cable
Ranked causes
- 1. Patient cable disconnected from monitor (common)
- 2. Incompatible leadset attached (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No central monitor connection patient safety
MX40 is out of range of network or Patient Sector at Information Center is in Standby
Ranked causes
- 1. MX40 out of network coverage range (common)
- 2. Information Center Patient Sector in Standby (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No signal at Information Center patient safety
MX40 is outside coverage area, has no batteries, or has failed; displayed as NO DATA PWM with IIC iX
Ranked causes
- 1. MX40 outside coverage area (common)
- 2. MX40 has dead batteries or failed (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 sensor poor signal quality patient safety
Although measurement is possible, accuracy may be reduced. Sensor receives poor signal quality during operation.
Ranked causes
- 1. Improper sensor application (common)
- 2. Sensor location suboptimal (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2T sensor not properly applied patient safety
Algorithm detected sensor is connected but not properly applied to patient. Sensor application is incorrect despite connection.
Ranked causes
- 1. Sensor improperly applied to patient (common)
- 2. Sensor location inadequate (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
WLAN connection status abnormal or offline patient safety
MX40 WLAN device fails to establish or maintain connection to Smart-hopping network or IntelliVue Information Center.
Ranked causes
- 1. MX40 not assigned to sector or INOP messages present (common)
- 2. MX40 connection status not Active (common)
Verify MX40 equipment sector assignment
- Test point
- IntelliVue Information Center status display
- Expected
- MX40 listed as assigned to a specific sector with no INOP messages
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
WLAN signal strength insufficient for clinical operation patient safety
MX40 WLAN device shows poor radio signal strength (RSSI) in patient care areas, affecting wireless network performance and patient monitoring continuity.
Ranked causes
- 1. Access point position or insufficient coverage (common)
- 2. RF interference in coverage area (occasional)
Verify MX40 setup for coverage test
- Test point
- Patient cable connection, RF auto-shutoff
- Expected
- Patient cable attached; RF auto-shutoff avoided by simulator, shorting ECG leads, or SpO2 measurement
4 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low patient safety
Less than 15 minutes of AA battery monitoring time remaining, or lithium-ion battery <10% or <30 minutes remaining
Ranked causes
- 1. AA or lithium-ion battery depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cableless NBP battery empty patient safety
CL NBP Pod battery completely depleted; monitoring is not possible
Ranked causes
- 1. CL NBP Pod battery completely discharged (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cableless NBP battery low patient safety
CL NBP Pod has weak battery condition and cannot perform measurements
Ranked causes
- 1. CL NBP Pod battery depleted or low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cableless NBP disconnected patient safety
CL NBP Pod is not connected with the MX40; out of range or excessive radio interference
Ranked causes
- 1. Radio range exceeded or interference present (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cableless SpO2 battery empty patient safety
CL SpO2 Pod battery completely depleted; monitoring is not possible
Ranked causes
- 1. CL SpO2 Pod battery completely discharged (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cableless SpO2 battery low patient safety
CL SpO2 Pod has weak battery condition; charging required
Ranked causes
- 1. CL SpO2 Pod battery low (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cableless SpO2 disconnected patient safety
CL SpO2 Pod is not connected with the MX40; out of range or excessive interference
Ranked causes
- 1. Radio range exceeded or interference present (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cannot analyze ECG patient safety
Arrhythmia algorithm cannot reliably analyze ECG data on any monitored leads
Ranked causes
- 1. Lead selection error or signal quality issue (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Check pairing failed
Problem with wireless device pairing; MX40 paired with X2 monitor has lost network connection
Ranked causes
- 1. Pairing configuration error or network connection lost (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuff not deflated patient safety
Cuff pressure has exceeded the specified safety limit during NBP measurement
Ranked causes
- 1. Cuff did not fully deflate during measurement cycle (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Cuff overpressure patient safety
Cuff pressure has increased above overpressure safety limits
Ranked causes
- 1. Cuff pressure exceeded safety threshold (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Device does not recognize patient cable type patient safety
MX40 unable to identify connected IntelliVue style leadset adapter cable
Ranked causes
- 1. IntelliVue style leadset adapter cable in use (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG leads off patient safety
Multiple ECG leads are disconnected from patient or not making proper contact
Ranked causes
- 1. ECG leads disconnected or loose contact with patient (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG/ARRH alarms off patient safety
ECG arrhythmia alarm monitoring has been disabled; no alarm generation for rhythm abnormalities
Ranked causes
- 1. ECG monitoring disabled by user or system (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Leadset life exceeded patient safety
Single-patient use leadset has exceeded its limit of 25 cycles and must be replaced
Ranked causes
- 1. Leadset cycle limit reached (25 cycles) (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Local audio off
No alarm audio notification when operating in Telemetry Mode; normal operation in Telemetry Mode
Ranked causes
- 1. Monitor in Telemetry Mode (no audio alarms) (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MX40 WLAN IP address unavailable
MX40 is connected to access point but cannot obtain IP address. Network communication fails.
Ranked causes
- 1. WLAN IP address assignment failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MX40 WLAN out of range
MX40 wireless connection is out of range of available access points. Radio link cannot be established.
Ranked causes
- 1. Device outside access point coverage area (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP equipment malfunction caution
Tubing may be obstructed, kinked, or hardware malfunction in NBP equipment
Ranked causes
- 1. Tubing obstruction/kink or NBP hardware failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurement failed patient safety
Measurement values cannot be derived from NBP Pod; obstructed tubing or hardware malfunction
Ranked causes
- 1. Tubing obstructed or kinked; hardware malfunction (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NBP measurement interrupted patient safety
Preset maximum time for total NBP measurement has been exceeded due to patient movement or cuff interaction
Ranked causes
- 1. Patient movement or cuff/tubing interaction during measurement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No alarm display patient safety
No local alarming at MX40 when networked with Information Center Release L or M
Ranked causes
- 1. Operating with older Information Center firmware (Release L or M) (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No host monitor connection
Paired MX40/bedside monitor is out of short-range radio range or experiencing excessive radio interference
Ranked causes
- 1. Radio range exceeded or radio interference present (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Replace battery required patient safety
Battery is dead and completely discharged; no monitoring occurring
Ranked causes
- 1. Battery completely discharged (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration equipment malfunction caution
Malfunction in respiration equipment or MX40 requires calibration
Ranked causes
- 1. Respiration equipment hardware failure or calibration needed (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Respiration lead off patient safety
Respiration lead is disconnected from patient; OR leadsets cannot be used for Resp monitoring
Ranked causes
- 1. Respiration lead disconnected or loose (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Single ECG lead off patient safety
One ECG lead is disconnected or not making proper contact with patient
Ranked causes
- 1. Single lead disconnected or loose contact (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Some ECG alarms off
Some ECG alarms have been turned off at the Information Center; informational message only
Ranked causes
- 1. ECG alarms disabled at Information Center (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 equipment malfunction caution
Malfunction in SpO2 equipment; sensor or transducer failure
Ranked causes
- 1. SpO2 equipment hardware failure (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 hardware upgrade in progress patient safety
SpO2 hardware is undergoing upgrade process. Patient monitoring is not possible during this time.
Ranked causes
- 1. SpO2 hardware upgrade in progress (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 interference caution
Ambient light or electrical interference too high; SpO2 sensor cannot measure SpO2 and pulse rate
Ranked causes
- 1. Excessive ambient light or electrical noise near sensor (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 low perfusion patient safety
Accuracy may be reduced due to low blood perfusion at sensor site; data displayed with ?
Ranked causes
- 1. Low blood perfusion at sensor site (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 measurement erratic patient safety
Erratic SpO2 measurements due to faulty sensor, invalid measurements, or incorrect transducer position
Ranked causes
- 1. Faulty sensor or incorrect sensor position on patient (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 no pulse detected patient safety
Pulse is too weak or not detectable; sensor has fallen off at patient or poor connection
Ranked causes
- 1. Pulse too weak or sensor disconnected from patient (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 no sensor attached patient safety
No SpO2 sensor is attached to SpO2 device; silencing alarm turns off SpO2 measurement
Ranked causes
- 1. SpO2 sensor not connected or unplugged (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 noisy signal caution
Excessive patient movements or electrical interference causing irregular pulse patterns in SpO2 data
Ranked causes
- 1. Patient movement or electrical noise during measurement (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 unsupported sensor type
Connected SpO2 sensor and/or adapter cable is not supported by installed hardware version. Incompatible equipment detected.
Ranked causes
- 1. Incompatible sensor or adapter cable for hardware version (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2 update period extended
Update period of SpO2 displayed values extended due to simultaneous NBP measurement or excessively noisy signal
Ranked causes
- 1. NBP measurement active on same limb or excessive signal noise (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2T searching for measurement
Patient signal is being analyzed but valid numeric value not yet available. System is still acquiring stable measurement.
Ranked causes
- 1. Measurement acquisition in progress (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpO2T sensor malfunction patient safety
Malfunction detected in SpO2 sensor or adapter cable. Hardware is defective and requires replacement.
Ranked causes
- 1. SpO2 sensor or adapter cable failure (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Telemeter battery charge cycle warning caution
Lithium-ion battery has less than 25 charge cycles remaining before reaching maximum limit. Battery nearing end of service life.
Ranked causes
- 1. Battery approaching charge cycle limit (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Telemeter battery critically low patient safety
Lithium-ion battery level is critically low with 10-minute countdown. MX40 will shut down if condition not cleared. Battery is depleted.
Ranked causes
- 1. Battery charge depleted below critical level (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Telemeter battery end of life reached caution
Lithium-ion battery has exceeded maximum charge cycle limit and reached end of useful life. Battery no longer functional.
Ranked causes
- 1. Battery maximum charge cycles exceeded (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Telemeter battery temperature critical removal required caution
Lithium-ion battery temperature exceeds 60°C. Battery must be removed immediately for safety.
Ranked causes
- 1. Battery temperature critically high (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Telemeter battery temperature out of range caution
Lithium-ion battery temperature is above 55°C or below -5°C. Battery thermal condition is unsafe.
Ranked causes
- 1. Battery temperature exceeds safe operating range (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Telemeter malfunction or self-test failure patient safety
MX40 malfunction detected or self-test failed. Device requires service replacement.
Ranked causes
- 1. MX40 hardware malfunction or failed self-test (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Transmitter RF auto shutoff triggered patient safety
RF transmitter automatically shut off after 10 minutes of all leads off and no SpO2 sensor connected. Transmission disabled.
Ranked causes
- 1. All leads detached and no SpO2 sensor for 10 minutes (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
BATTERY LOW T | Low battery level on AA or lithium-ion battery | AA batteries: less than 15 minutes remaining. Lithium-ion: level < 10% or <30 minutes remaining | warning |
BATTERY LOW T | Low battery condition with less than 15 minutes monitoring time remaining or lithium-ion <10% | Battery level drops below threshold: AA <15 min remaining or lithium-ion <10% or <30 min | warning |
CANNOT ANALYZE ECG | Arrhythmia algorithm unable to reliably analyze ECG data on any monitored leads | Arrhythmia algorithm detects unreliable or corrupted ECG signal quality on all leads | error |
CHECK PAIRING | Problem with wireless device pairing or network connectivity with paired monitor | Pairing configuration error or loss of network connection on wireless paired devices | warning |
cl NBP Batt Empty | Cableless NBP Pod battery is completely empty and depleted | CL NBP Pod battery completely discharged; cannot power device | critical |
cl NBP Batt Low | Cableless NBP Pod battery is in weak/low charge condition | CL NBP Pod battery voltage drops below operational threshold | error |
cl NBP DISCONNECT | Cableless NBP Pod is not connected with the MX40 monitor | Wireless communication link lost between CL NBP Pod and MX40 | error |
cl SpO2 Batt Empty | Cableless SpO2 Pod battery completely empty and depleted | CL SpO2 Pod battery completely discharged; cannot power device | critical |
cl SpO2 Batt Low | Cableless SpO2 Pod battery in weak/low charge condition | CL SpO2 Pod battery voltage drops below operational threshold | error |
cl SpO2 DISCONNECT | Cableless SpO2 Pod is not connected with the MX40 monitor | Wireless communication link lost between CL SpO2 Pod and MX40 | error |
!!!/!! CUFF NOT DEFLATED | Cuff pressure has exceeded specified safety limit during NBP measurement | Cuff pressure remains elevated beyond safety threshold after measurement cycle | critical |
!!!/!! CUFF OVERPRESS | Cuff pressure increased above overpressure safety limits during measurement | Cuff pressure exceeds maximum allowable overpressure threshold | critical |
ECG/ARRH ALARM OFF | ECG monitoring turned off | ECG function disabled by user | warning |
ECG/ARRH ALARM(S) OFF | ECG arrhythmia alarm monitoring is disabled; no alarm generation | ECG monitoring or arrhythmia alarm function disabled by user or system | warning |
ECG LEADS OFF | Multiple ECG leads are disconnected or not making proper contact with patient | One or more ECG electrode leads detected as open or loose | error |
<electrode> LEAD OFF | Single specified ECG lead is disconnected or not making proper contact | One ECG electrode detected as open or loose; MCL primary lead shows as V/C | error |
LEADSET LIFE | Single-patient use leadset has exceeded its limit of 25 cycles | Leadset cycle counter reaches 25 use cycles | warning |
LEADSET UNPLUGGED | Patient cable unplugged from MX40 or incompatible leadset attached to patient cable | Patient cable disconnected from monitor connector or wrong leadset type installed | error |
LOCAL AUDIO OFF | No alarm audio notification when operating in Telemetry Mode (normal) | MX40 operating in Telemetry Mode; audio alarms suppressed for patient privacy | info |
NBP EQUIP MALF | NBP equipment malfunction; tubing obstruction or hardware failure | NBP system detects tubing blockage, kink, or internal hardware malfunction | error |
NBP INTERRUPTED | NBP measurement time exceeded maximum preset duration | Total measurement time exceeded preset maximum | error |
NBP INTERRUPTED | NBP measurement interrupted; preset maximum measurement time exceeded | NBP measurement duration exceeds preset maximum time limit | error |
NBP MEASURE FAILED | NBP measurement values cannot be derived; measurement unsuccessful | NBP algorithm fails to calculate systolic/diastolic/mean pressure from cuff data | error |
NO ALARM DISPLAY | No local alarm display at MX40 when operating with Information Center Release L or M | Operating with Information Center Release L or M in networked or non-networked mode | warning |
NO CENTRAL MONITOR | MX40 is out of range of network or Information Center Patient Sector in Standby | Wireless network connection lost or Information Center Patient Sector in Standby mode | error |
NO HOST MONITOR | Paired MX40/bedside monitor is out of short-range radio range or experiencing interference | Wireless short-range link lost between MX40 and bedside monitor or host device | error |
NO HOST MONITORING | MX40 out of short-range radio range or excessive radio interference | Loss of short-range wireless connectivity or high RF interference | error |
NO SIGNAL | MX40 not detected at Information Center; displayed as NO DATA PWM with IIC iX | Information Center cannot receive signal from MX40 monitor | error |
NO SIGNAL | MX40 outside coverage area, no batteries, or device failed | MX40 not detected by Information Center due to range loss, power loss, or failure | error |
REPLACE BATTERY T | Dead battery with no monitoring occurring; battery must be replaced immediately | Battery completely discharged and unable to power the device | critical |
RESP EQUIP MALF | Malfunction in respiration measurement equipment or MX40 requires calibration | Respiration sensor detects malfunction or MX40 firmware detects need for calibration | error |
RESP LEAD OFF | Respiration lead disconnected from patient; OR leadsets cannot monitor Resp | Respiration electrode lead detected as open or loose | error |
SOME ECG ALARMS OFF | Some ECG alarms have been disabled at Information Center level | Selected ECG alarm thresholds disabled by Information Center configuration | info |
SpO2 POOR SIGNAL | Measurement possible but accuracy reduced due to poor signal quality | Low-amplitude or noisy SpO2 signal | warning |
SpO2T EQUIP MALF | Malfunction in SpO2 measurement equipment or transducer | SpO2 sensor detects internal hardware malfunction or communication error | error |
SpO2T ERRATIC | Erratic SpO2 measurements | Faulty sensor, invalid measurements, or incorrect transducer position | error |
SpO2T ERRATIC | Erratic SpO2 measurements due to faulty sensor or incorrect transducer position | SpO2 algorithm detects invalid or unstable measurement values | error |
SpO2T EXTD UPDATE | Update period of SpO2 displayed values extended due to NBP or signal noise | SpO2 algorithm detects high noise or simultaneous NBP measurement on same limb | warning |
SpO2T INTERFERENCE | Ambient light or electrical interference too high; SpO2 sensor cannot measure | SpO2 sensor detects excessive optical or electrical noise in measurement signal | error |
SpO2T LOW PERF | Accuracy reduced due to low perfusion at sensor site; data shown with ? | SpO2 algorithm detects insufficient blood flow at measurement site | warning |
%SpO2T NOISY SIGN | Excessive patient movement or electrical interference causing irregular pulse patterns | SpO2 algorithm detects high-frequency noise in pulse waveform | error |
SpO2T NO PULSE | Pulse is too weak or not detectable; sensor may have fallen off patient | SpO2 algorithm cannot detect pulse wave in measurement signal | error |
SpO2T NO SENSOR | No SpO2 sensor attached to SpO2 measurement device | SpO2 system detects missing or disconnected sensor | error |
SpO2T SEARCHING | SpO2 measurement acquisition in progress | Sensor connected and analyzing patient signal | info |
SpO2T SEARCHING | Patient signal analyzed but valid numeric not yet available | SpO2 measurement algorithm initializing or processing signal | info |
SpO2T SENSOR MALF | Malfunction of the SpO2 sensor or adapter cable | SpO2 sensor or adapter hardware failure | error |
SpO2T SENSOR OFF | Sensor connected but not properly applied to patient | Sensor not making proper contact with patient skin | error |
SpO2T UPGRADE | SpO2 hardware firmware upgrade in progress | Upgrade process initiated and running | warning |
SpO2 UNKN SENSOR | Connected SpO2 sensor and/or adapter cable not supported by hardware version | Incompatible sensor or adapter cable attached | error |
TELE BATT EMPTY | Lithium-ion battery critically low with 10-minute countdown to shutdown | Lithium-ion battery level critically depleted | error |
TELE BATTERY LOW | Lithium-ion battery level critically low | Lithium-ion battery level < 20% or < 30 minutes remaining | warning |
TELE BATTERY TEMP | Temperature of lithium-ion battery outside safe operating range | Battery temperature above 55°C or below -5°C | error |
TELE BATTERY TEMP | Telemeter battery temperature out of safe range | Battery temperature exceeded limits (>55°C or <-5°C) | error |
TELE CHECK BATT | Lithium-ion battery nearing maximum charge cycle limit | Battery has < 25 charge cycles remaining before reaching maximum | warning |
TELE MALFUNCTION | MX40 malfunction or self-test failure | Internal device failure or Power On Self Test failure detected | critical |
TELE REMOVE BATT | Battery temperature exceeds safe limit and must be removed immediately | Lithium-ion battery temperature > 60°C | critical |
TELE SERVICE BATT | Lithium-ion battery reached end of useful life | Battery has exceeded maximum charge cycle limit | error |
TELE WEAK SIGNAL | Weak wireless signal reception with high data loss | Patient at outer range of coverage area or MX40 receiving weak signal from AP | warning |
TRANSMITTER OFF | RF transmitter disabled after 10 minutes with all leads off and no SpO2 sensor | All ECG leads off and no SpO2 sensor connected for 10 minutes | error |
UNSUPPORTED LAN | MX40 cannot obtain IP address on WLAN | Device connected to AP but IP assignment failed | error |
WIFI OUT OF RANGE | MX40 WLAN out of access point coverage range | No access points detected or signal too weak | error |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI