Philips IntelliVue MP90: troubleshooting and error codes
The ServiceForge knowledge base holds 128 documented symptoms for the Philips IntelliVue MP90, with 314 ranked causes, 233 diagnostic checks with test points and expected values, and 26 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Pneumatic system shows problem caution
Module diagnostic reports pneumatic system failure; pump weak, occlusion detected, or system leakage indicated
Ranked causes
- 1. Weak pump (common)
- 2. Occluded pneumatics (common)
- 3. Defective cables (occasional)
- 4. Solenoid or air reference filter problem (occasional)
- 5. Flowrate problem (occasional)
- 6. Flow transducer incorrectly connected to flow restrictor (rare)
- 7. Pump problem (occasional)
- 8. Defective power supply (rare)
- 9. Defective main PC board (rare)
13 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC LED does not light up caution
The AC power indicator LED fails to illuminate when AC mains are connected
Ranked causes
- 1. AC connection not OK (common)
- 2. LED defective (occasional)
- 3. Power Switch board not connected (occasional)
- 4. Power supply defective (occasional)
- 5. Secondary main board defective (rare)
- 6. Primary Video Board defective (rare)
- 7. Backplane Bottom defective (rare)
- 8. Primary Main Board defective (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green Power On LED and Red Error LED remain off caution
Both power and error LEDs fail to illuminate after pressing power button on power supply
Ranked causes
- 1. Remote devices connected (common)
- 2. Power Switch Micro Controller hung (common)
- 3. Power switch board not connected (occasional)
- 4. Power Switch Board defective (occasional)
- 5. I/O Board defective (rare)
- 6. Primary Video Board defective (rare)
- 7. Primary Main Board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operating temperature shows problem caution
Module diagnostic reports excessive operating temperature; fan failure, insufficient airflow, or environmental conditions exceeded
Ranked causes
- 1. Defective fan (common)
- 2. Insufficient air circulation due to fan working inefficiently (common)
- 3. Operating environment for module falls outside of specified limits (occasional)
- 4. Defective IR measurement assembly head (rare)
- 5. Defective agent ID head (rare)
- 6. Defective power supply (rare)
- 7. Defective main PC board (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Span calibration failed patient safety
Gas span calibration check failed for one or more gas channels
Ranked causes
- 1. Zero calibration failed for indicated channel (common)
- 2. Incorrect agent selection or calibration values (common)
- 3. Flow rate problem (occasional)
- 4. Leakage problem (occasional)
- 5. Wrong calibration gas applied (occasional)
- 6. Empty calibration reservoir or gas canister (occasional)
- 7. Measurement assembly problem (occasional)
Verify zero calibration passed
- Test point
- Gas Analyzer Calibration window
- Expected
- Zero calibration shows Passed for affected channel
3 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Zero calibration failed patient safety
AGM zero calibration check failed during startup or maintenance
Ranked causes
- 1. Solenoid or air reference filter problem (common)
- 2. Occluded pneumatics (common)
- 3. Flow rate problem (occasional)
- 4. Flow transducer incorrectly connected (occasional)
- 5. Pump problem (occasional)
- 6. IR measurement head problem (occasional)
- 7. Agent-ID problem (occasional)
Verify air filter and solenoid function
- Test point
- Room air filter inlet during zero calibration
- Expected
- Air being pulled in at room air filter
4 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no LED lights up
Alarm triggered but alarm LED does not illuminate
Ranked causes
- 1. Environmental Lighting too bright (common)
- 2. Remote Alarm Device not connected (common)
- 3. Remote Alarm Device defective (occasional)
- 4. Remote I/O board in wrong slot (occasional)
- 5. Remote I/O board defective (occasional)
- 6. Main Board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External device not receiving data
Data export from monitor to external device fails or produces no output
Ranked causes
- 1. MIB/RS232 port not configured for data export (common)
- 2. Wrong data export protocol driver configured (occasional)
- 3. Cable between device and monitor not connected or defective (common)
- 4. External device does not support data export protocol version (occasional)
- 5. Terminal concentrator does not support dynamic speed negotiation (rare)
- 6. MIB/RS232 board in wrong slot or defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 assembly shows problem caution
Module diagnostic reports O2 sensor assembly failure; jumpers incorrect, calibration failed, or sensor defective
Ranked causes
- 1. O2 jumpers incorrectly set (common)
- 2. O2 sensor failed zero/span calibration (common)
- 3. Defective O2 sensor (occasional)
- 4. Defective cables (occasional)
- 5. Defective power supply (rare)
- 6. Defective main PC board (rare)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED stays on continuously caution
The red error indicator LED illuminates and remains on without going off
Ranked causes
- 1. External connected device defective (common)
- 2. I/O Board defective (occasional)
- 3. Backplane top or secondary main board defective (occasional)
- 4. Primary Video Board defective (rare)
- 5. Backplane Bottom defective (rare)
- 6. Primary Main board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpeedPoint attached to Remote Extension Device not functioning
SpeedPoint input control attached to remote extension device does not work
Ranked causes
- 1. Remote Extension Device not connected to monitor (common)
- 2. SpeedPoint not connected properly (common)
- 3. SpeedPoint defective (occasional)
- 4. Remote Extension Device defective (occasional)
- 5. Remote Device I/O board in wrong slot (occasional)
- 6. Remote Device I/O board defective (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Agent ID assembly shows problem caution
Module diagnostic reports Agent ID assembly failure; software revision wrong, agent ID head defective, or power issue
Ranked causes
- 1. Wrong Agent-ID software revision (common)
- 2. Defective agent ID head (occasional)
- 3. Defective power supply (rare)
- 4. Defective cables (occasional)
- 5. Defective main PC board (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Agent identification missing patient safety
Anesthetic agent is not being identified or measured by AGM
Ranked causes
- 1. Sample line not connected (common)
- 2. Agent not being administered (occasional)
- 3. Agent-ID configuration incorrect (occasional)
- 4. Agent-ID cable connection issue (occasional)
- 5. Agent-ID assembly failure (rare)
Verify sample line connected
- Test point
- Sample line connection
- Expected
- Sample line connected between circuit and AGM
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication problem from beginning
LED is steadily on; monitor and AGM cannot establish communication at startup
Ranked causes
- 1. Anesthesia software bundle not installed (common)
- 2. Incorrect RS232 configuration or MIB port settings (common)
- 3. Physical cable connection problem (occasional)
- 4. Monitor hardware failure (rare)
- 5. AGM motherboard failure (rare)
Verify Anesthesia software bundle installed
- Test point
- Monitor software list
- Expected
- Anesthesia SW bundle #H30 present in Standby mode menu
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External display is blank
Monitor external video output not visible on connected display
Ranked causes
- 1. Video cable to external display not connected (common)
- 2. External display has no power (common)
- 3. External display is defective (occasional)
- 4. Video board defective (occasional)
- 5. Main board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External touch display not functioning
Touch screen input on external touch display does not respond to touch input
Ranked causes
- 1. External Touch cable not connected (common)
- 2. External Touch driver configuration incorrect (common)
- 3. MIB Board defective (occasional)
- 4. External touch defective (occasional)
- 5. Main board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Check Alarm Lamps issued
INOP 'Check Alarm Lamps' is issued
Ranked causes
- 1. Remote Alarm Device not connected (common)
- 2. Remote Alarm Device defective (occasional)
- 3. Remote Device I/O board in wrong slot (occasional)
- 4. Remote Device I/O board defective (occasional)
- 5. Main board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard or Mouse attached to Remote Extension Box not functioning
Keyboard or mouse connected through remote extension box does not work
Ranked causes
- 1. Remote Extension Box not connected or input device not connected (common)
- 2. Keyboard or Mouse defective (occasional)
- 3. Remote Extension Device defective (occasional)
- 4. Remote Device I/O board in wrong slot (occasional)
- 5. Remote Device I/O board defective (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No central monitoring possible patient safety
Monitor cannot connect to central monitoring station or other monitors on wireless network
Ranked causes
- 1. Monitor out of range of access point (common)
- 2. Wireless Adapter has no power (common)
- 3. Firmware revision in adapter is wrong (occasional)
- 4. Wireless adapter defective (occasional)
- 5. Wrong configuration in wireless adapter or access point (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AGM connected to RS232 port not functioning
AGM module fails to operate when connected to MIB/RS232 port
Ranked causes
- 1. MIB/RS232 port not configured for AGM (common)
- 2. Cable between AGM and monitor not connected or defective (common)
- 3. MIB/RS232 board in wrong slot (occasional)
- 4. MIB/RS232 board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no alarm sound caution
Alarm triggered but no audible alarm sound is produced
Ranked causes
- 1. Audible alarm indicators switched off (common)
- 2. Volume set to 0 (common)
- 3. Speaker defective (occasional)
- 4. Sound amplifier defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs on screen but no LED or alarm sound on alarm device
Alarm message displays on monitor screen but remote alarm device does not respond
Ranked causes
- 1. Cabling not connected (common)
- 2. Cabling defective (occasional)
- 3. I/O board defective (occasional)
- 4. Remote Alarm Device defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green Power On LED remains off after button press on remote device caution
Power LED fails to illuminate when power button pressed on remote extension or remote alarm device
Ranked causes
- 1. Cable to remote device incorrect (common)
- 2. Remote device defective (common)
- 3. Power switch board cable not connected (occasional)
- 4. Remote device interface board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard or Mouse not functioning
Keyboard or mouse input directly attached to monitor does not respond
Ranked causes
- 1. Keyboard or Mouse not connected properly (common)
- 2. Keyboard or Mouse defective (occasional)
- 3. PS/2 I/O board in wrong slot (occasional)
- 4. PS/2 I/O Board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement assembly shows problem caution
Module diagnostic reports IR measurement assembly failure; assembly head defective, power supply failed, or cable issue
Ranked causes
- 1. Defective IR measurement assembly head (common)
- 2. Defective power supply (rare)
- 3. Defective cables (occasional)
- 4. Defective main PC board (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarm sound but not Remote Device
Alarm sound on monitor but not on Remote Alarm Device
Ranked causes
- 1. Remote Alarm Device not connected (common)
- 2. Remote Alarm Device defective (occasional)
- 3. Remote I/O board in wrong slot (occasional)
- 4. Remote I/O board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarmed, Nurse Call did not activate patient safety
System triggers alarm but Nurse Call relay does not signal external system
Ranked causes
- 1. Incorrect configuration (Relay latency, Relay trigger) (common)
- 2. Cable connection to monitor or nurse call system not correct (common)
- 3. Nurse Call Relay board in wrong slot (occasional)
- 4. Nurse Call Relay board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity to PIC, no prompt or error message on monitor
Monitor has lost connection to Philips Information Center with no visible error indication
Ranked causes
- 1. Hardware defect - LAN cable not connected (common)
- 2. NGN Connector board in monitor defective (occasional)
- 3. Switch defect (occasional)
- 4. Configuration problem (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No marker pulse displayed on monitor or no ECG-OUT signal to Defib high voltage
ECG marker signal or defibrillator output signal not present despite data path integrity
Ranked causes
- 1. Cabling not connected (common)
- 2. ECG OUT board defective (LED does not switch on at power-on) (occasional)
- 3. ECG OUT board defective (LED remains on >20 seconds) (occasional)
- 4. Main Board defective (rare)
Disconnect MMS and Defib cables
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optical path shows problem caution
Module diagnostic reports optical measurement path failure; sample cell contaminated or IR measurement assembly defective
Ranked causes
- 1. Defective or contaminated sample cell (common)
- 2. Defective IR measurement assembly head (occasional)
- 3. Defective power supply (rare)
- 4. Defective main PC board (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply shows problem caution
Module diagnostic reports power supply voltage failure; pump weak, voltage rails out of specification, or defective supply
Ranked causes
- 1. Weak pump (occasional)
- 2. Defective power supply (common)
- 3. Defective cables (occasional)
- 4. Defective main PC board (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print device Local unavailable status message
Status 'Print device Local 1 (Local 2) unavailable' issued. Printer job stalled.
Ranked causes
- 1. Printer not switched on (common)
- 2. Printer paper tray empty (common)
- 3. Cabling not connected correctly (occasional)
- 4. I/O board defect (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printouts not as expected
Printed reports do not appear as expected
Ranked causes
- 1. Printer paper size not configured (common)
- 2. Printer resolution not configured (common)
- 3. Printer color support misconfigured (common)
- 4. Printer not compatible (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder not communicating with system
Module not establishing communication with front-end FMS or system not configured properly
Ranked causes
- 1. Poor connection to front-end FMS (common)
- 2. Only one recorder module may be used with each monitor (occasional)
- 3. System not configured properly (occasional)
- 4. Too many modules connected (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs illuminate but no alarm sound issued
Remote alarm device LEDs light up but no audio alarm generated
Ranked causes
- 1. Wrong I/O slot (occasional)
- 2. Speaker defective (occasional)
- 3. Remote device I/F defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK MONITOR TEMP error caution
Temperature inside monitor exceeds safe operating limits
Ranked causes
- 1. Monitor ventilation obstructed (common)
- 2. Check environment for possible causes (occasional)
- 3. Main Board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Disappearing waves in warmup mode
Gas measurement waveforms disappear during AGM warmup; may reappear in normal mode
Ranked causes
- 1. Pump or flow delivery issue (common)
- 2. Exhaust line blockage (occasional)
- 3. Instrument startup transient (common)
Disconnect exhaust tubing
- Test point
- Exhaust line
- Expected
- Exhaust tubing removed if present
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Frequent dropouts and network disconnects patient safety
Wireless connection frequently drops or disconnects during operation
Ranked causes
- 1. Excessive interference by other radio equipment or microwave ovens (common)
- 2. System capacity exceeded in coverage area (occasional)
- 3. Configuration problem (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas measurement accuracy problems patient safety
Measured gas values are inaccurate; calibration failures or sensor issues
Ranked causes
- 1. Accessory damage or misconfiguration (common)
- 2. Sample line not connected (common)
- 3. Calibration error (occasional)
Inspect accessories for damage
- Test point
- All tubing and connectors
- Expected
- No leaks, kinks, or visible damage
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Green On/Standby LED or Red Error LED remain off after button press caution
One or both of the power/error indicator LEDs fail to illuminate after pressing power button
Ranked causes
- 1. Power Switch Board not connected (common)
- 2. Power Switch board defective (occasional)
- 3. Main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP BAD SERVER LINK issued
INOP BAD SERVER LINK is displayed
Ranked causes
- 1. Unexpected data detected on MSL (common)
- 2. FMS or MMS incompatible software revision (common)
- 3. Communication between components not functioning (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP MsmtSrv not Supp message issued
INOP message 'MsmtSrv not Supp' is displayed
Ranked causes
- 1. Wrong Software Revision (common)
- 2. Too many measurement servers connected (common)
- 3. Unsupported type of measurement server (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Speaker Malfunction message
INOP 'SPEAKER MALFUNCTION' is displayed
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Unsupported LAN error
Monitor displays unsupported LAN network error indicating network connectivity issue
Ranked causes
- 1. Network failure (common)
- 2. Monitor connected to wrong network (occasional)
- 3. IP address conflict after infrastructure re-installation (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INTERNAL COMM.MALF. error
Problem with I2C Bus communication inside monitor
Ranked causes
- 1. External display communication issue (occasional)
- 2. Video board defective (occasional)
- 3. Main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 calibration failed patient safety
O2 zero or span calibration check failed
Ranked causes
- 1. Flow rate error (common)
- 2. O2 span issue (common)
- 3. O2 sensor defective (occasional)
Perform leakage check
- Test point
- O2 sensor flow path
- Expected
- No gas leakage detected
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Other Bed Overview not available patient safety
Monitor cannot display patient overview from other beds on network
Ranked causes
- 1. Configuration problem in PIC (common)
- 2. Switch configuration problem (occasional)
- 3. Feature not available for wireless beds (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED blinks (indicating cyclic reboots) caution
The red error LED blinks on and off repeatedly, indicating the monitor is repeatedly rebooting
Ranked causes
- 1. Hardware Failure (common)
- 2. Software Fault (occasional)
- 3. Hardware Failure (if software fix fails) (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SETTINGS MALFUNCTION error
Problem during cloning process or corrupted settings memory
Ranked causes
- 1. Problem during cloning process (occasional)
- 2. Memory space for settings corrupted (occasional)
- 3. Main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK MENU LABELS error
Monitor menu labels corrupted or not matching configured defaults after cold start
Ranked causes
- 1. Memory space for menu labels corrupted (occasional)
- 2. Main Board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
CHECK SETTINGS error
Monitor settings corrupted or not matching configured defaults after cold start
Ranked causes
- 1. Memory space for settings corrupted (occasional)
- 2. Main board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication problem during operation
LED flashing and INOP GA EQUIP MALFUNCT displayed; communication lost after startup
Ranked causes
- 1. Intermittent cable connection or internal monitor wiring issue (common)
- 2. Monitor motherboard failure (occasional)
Verify AGM software and settings
- Test point
- Monitor Standby mode menu and RS232/MIB settings
- Expected
- Anesthesia SW #H30 installed and RS232/MIB correctly configured
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Front End Measurement Module not recognized
Flexible Module Server LEDs ok but Front End Measurement Module not recognized
Ranked causes
- 1. Measurement Module or Connector defective (common)
- 2. MSL voltage out of range (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas measurement disturbed patient safety
Transient measurement disturbance or minor sensor issue occurring
Ranked causes
- 1. Minor transient IR head or O2 sensor problem (common)
- 2. Persistent IR head or O2 sensor issue (occasional)
Observe measurement stability
- Test point
- Gas display readings
- Expected
- Stable readings within 10 seconds
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement Server configuration not supported
Prompt message 'Measurement Server Configuration not supported' is issued
Ranked causes
- 1. Unsupported Measurement Server Extension connected (common)
- 2. Measurement Server defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement Server does not start but LEDs normal
LEDs are normal but server does not start up
Ranked causes
- 1. Communication lines broken in MSL cable (common)
- 2. MSL connector board defective (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No Test Sound issued
The monitor does not produce the expected test sound during boot sequence
Ranked causes
- 1. Speaker defective (common)
- 2. Main board defective (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 measurement missing patient safety
O2 is not being measured or displayed by the AGM
Ranked causes
- 1. Sample line not connected (common)
- 2. O2 assembly failure (occasional)
Verify sample line connected
- Test point
- Sample line connection
- Expected
- Sample line physically connected
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Paper not feeding properly
Recorder paper does not advance correctly during operation
Ranked causes
- 1. Paper roll off center (common)
- 2. Dirty roller (common)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor print quality
Recorder produces low-quality or unclear printouts
Ranked causes
- 1. Printhead dirty (common)
- 2. Printhead failure (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Print device Remote unavailable status
- Print job could not be queued message
- Prompt message no central assigned to this bed
- Sample cell misaligned or loose
- Agent ID head requires replacement
- Agent identification head removal
- Alarm on VueLink but no monitor sound
- Alarm sound issued but no LEDs light up
- Barometric pressure check performance test
- Buttons on Remote Extension Device not functioning
- Calibration: Anesthetic Gas Module
- Calibration: Anesthetic Gas Module flowrate
- Calibration: Anesthetic Gas Module safety and performance
- Calibration: barometric pressure baseline
- Calibration: Barometric pressure check and calibration
- Calibration: Span calibration check
- Calibration: Total flowrate in normal mode
- Calibration: Zero calibration
- CHECK INPUT DEVICES INOP message
- CHECK INTERNAL VOLTAGE error
- Diagnostic check performance test
- ECG Out/Marker In function does not function
- Fan check performance test
- Fan requires replacement
- Flat-cable connectors loose or damaged after sample cell replacement
- Flowrate check performance test
- INOP message CHECK NURSE CALL RELAY is issued
- INOP SERVERLINK MALF message displayed
- Leakage check performance test
- Main PC board removal or replacement
- Measurement does not appear on monitor
- Measurement Module currently ignored message
- Measurement Server does not start up
- Measurement Server not supported unplug message
- Module does not lock into FMS
- MSL Power High INOP issued
- MSL Power Off INOP issued
- MSL Power Overload INOP issued
- No connectivity in multi-access point coverage area
- Normal operation check performance test
- O2 sensor removal or replacement
- Other Bed Alarms not appearing
- Overview or Printing does not work
- Power fuses require replacement
- Power supply unit removal or replacement
- Power-on and self-test check
- Printing message but no report printed
- Pump requires replacement
- Pump test performance check
- Recorder won't run
- Red Error LED flashes
- Red Error LED stays on
- Remote input device attached to Remote Extension Device not functioning
- Safety verification after service event
- Sample cell pneumatic connections loose or damaged
- Solenoid valve #1 removal or replacement
- Solenoid valve #2 requires replacement
- Span calibration check performance test
- SpeedPoint Knob Rotation, Joystick Control or Selection control not functioning
- System thinks door is open when it is not
- System thinks recorder is out of paper when it is not
- Too many modules of same type connected message
- Top cover PC board requires replacement
- Touch position invalid
- Unrecognized Measurement Module in slot
- Visual inspection after repair
- Watertrap manifold and protector require replacement
- Zero calibration check performance test
68 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
AGENT IDENT MALF | Agent identification equipment malfunction | Serious Agent-ID assembly failure | error |
AGENT IDENT ZERO FAILED | Agent ID zero calibration failed | Agent identification sensor zero calibration failed | error |
AGT UNABLE TO MEASURE | Agent measurement unable to measure | Agent measurement failure or unavailable | error |
Bad Measurement Server | Measurement Server incompatible or defective | Incompatible server connected | error |
BAD SERVER LINK | Communication link to server bad | Unexpected data or incompatible software | error |
Check Alarm Lamps | Remote Alarm Device problem | Alarm device disconnected, defective, or misconfigured | error |
CO2 UNABLE TO MEASURE | CO2 sensor unable to measure | CO2 measurement failure or unavailable | error |
GA EQUIP MALF | Gas analyzer equipment malfunction | AGM-monitor connection problem or AGM hardware failure | critical |
GA INCOMPATIBLE | AGM incompatible with monitor | AGM software version incompatible with monitor version | error |
GA. NOT AVAILABLE | Gas analyzer not available or not responsive | AGM powered off or communication lost | error |
GAS AN ACCURACY ? | Gas analyzer accuracy suspect or unknown | Measurement accuracy check shows values out of specification | warning |
GAS OCCLUSION | Gas flow path occluded | Gas flow blocked internally or externally | error |
GA ZERO FAILED | Gas analyzer zero calibration failed | Zero calibration routine could not complete successfully | error |
INOP GA EQUIP MALFUNCT | Gas analyzer equipment malfunction | AGM communication lost or hardware fault during operation | error |
Measurement Server Configuration not supported | Unsupported Measurement Server Extension connected or server defective | Unsupported extension or defective server | error |
MSL Power High | Excessive power draw on MSL | Too many modules connected | warning |
MSL Power Off | MSL power shutdown due to overload | Excessive power for over 15 minutes | critical |
MSL Power Overload | Short circuit in MSL system | Electrical short detected | critical |
MsmtSrv not Supp | Measurement Server not supported | Wrong software or unsupported server type | error |
N2O UNABLE TO MEASURE | N2O sensor unable to measure | N2O measurement failure or unavailable | error |
O2 EQUIP MALF | O2 equipment malfunction | O2 sensor or span calibration failed | error |
O2 UNABLE TO MEASURE | O2 sensor unable to measure | O2 measurement failure or unavailable | error |
O2 ZERO FAILED | O2 sensor zero calibration failed | O2 zero calibration could not establish valid baseline | error |
SERVERLINK MALF | MSL hardware communication faulty | Hardware malfunction in interface | critical |
SPEAKER MALFUNCTION | Speaker or audio amplifier not functioning | Speaker cable disconnected or defective | error |
XXX MEAS DISTURBED | Gas measurement disturbed (N2O/CO2/agent/O2) | Transient measurement error or sensor instability | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI