The ServiceForge knowledge base holds 235 documented symptoms for the Philips IntelliVue MP60, with 524 ranked causes, 228 diagnostic checks with test points and expected values, and 39 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
Green Power On LED and Red Error LED remain off after pressing power button high voltage
Both power on and error indicator LEDs fail to illuminate when power button is pressed after boot attempt
Ranked causes
- 1. Remote devices causing power fault (common)
- 2. Power Switch Micro Controller hung (common)
- 3. Power switch board not connected to main board (occasional)
- 4. Power Switch Board defective (occasional)
- 5. I/O Board defective (occasional)
- 6. Touch controller, Flat Panel Adapter, ECG-Out or MSL-2 Board defective (occasional)
- 7. Measurement Server Mount defective (rare)
- 8. Integrated Module Slot defective (rare)
- 9. Video Board defective (rare)
- 10. Main Board defective (rare)
10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Integrated display blank or brightness reduced
Display is blank or has reduced brightness. May occur when room temperature exceeds 40°C or below 5°C.
Ranked causes
- 1. Display overheating from high room temperature (common)
- 2. Flat Panel Adapter cable disconnected (common)
- 3. Backlight Inverter Cable disconnected (common)
- 4. Backlight tubes defective (occasional)
- 5. Backlight Inverter board defective (occasional)
- 6. Panel Adapter board defective (occasional)
- 7. LCD Flat panel defective (occasional)
- 8. Video board defective (occasional)
- 9. Main board defective (rare)
Connect external XGA display
- Test point
- Video port
- Expected
- External display works or does not work
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Pneumatic system shows problem
Diagnostic message indicates pneumatic system failure or anomaly detected during operation
Ranked causes
- 1. Weak pump (occasional)
- 2. Occluded pneumatics (occasional)
- 3. Defective cables (rare)
- 4. Solenoid or air reference filter problem (occasional)
- 5. Flowrate problem (occasional)
- 6. Flow transducer incorrectly connected (rare)
- 7. Pump problem (occasional)
- 8. Defective power supply (rare)
- 9. Defective main PC board (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power-on LEDs do not illuminate caution
Green Power On LED and Red Error LED remain off after pressing power button
Ranked causes
- 1. Remote devices causing interference (common)
- 2. Power switch microcontroller hung or frozen (common)
- 3. Power switch board not connected to main board (occasional)
- 4. Power switch board defective (occasional)
- 5. I/O board defective (rare)
- 6. Touch controller, video, ECG-Out, or MSL-2 board defective (rare)
- 7. Measurement server mount defective (rare)
- 8. Video board defective (rare)
- 9. Main board defective (rare)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen not functioning
Touch panel does not respond to user input; touches are not registered or recognized by the system
Ranked causes
- 1. Touchscreen functionality temporarily disabled (common)
- 2. Touchscreen functionality permanently disabled (occasional)
- 3. Touch controller not connected (occasional)
- 4. Touch controller cable not connected (occasional)
- 5. Touch panel cable not connected (occasional)
- 6. Touch controller board defective (rare)
- 7. Touch screen defective (rare)
- 8. Main board defective (rare)
- 9. Touch not calibrated (occasional)
9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Touch Display not functioning
External touchscreen input device is not responding or not working.
Ranked causes
- 1. Touchscreen temporarily disabled (common)
- 2. Touchscreen permanently disabled (common)
- 3. External Touch cable not connected (common)
- 4. External Touch driver configuration error (occasional)
- 5. MIB Board defective (occasional)
- 6. External touch defective (occasional)
- 7. Main board defective (rare)
- 8. Touch not calibrated (common)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Integrated Touch Display not functioning
Touchscreen on integrated display is not responsive or not working.
Ranked causes
- 1. Touchscreen temporarily disabled (common)
- 2. Touchscreen permanently disabled (common)
- 3. Touch controller cable disconnected (occasional)
- 4. Touch panel cable disconnected (occasional)
- 5. Touch controller board defective (occasional)
- 6. Touch Sensor defective (occasional)
- 7. Main board defective (rare)
- 8. Touch not calibrated (common)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity to PIC, no prompt or error message on monitor
Monitor cannot communicate with Philips Information Center; network connection fails silently without error indication
Ranked causes
- 1. Monitor connected to wrong network (common)
- 2. IP address conflict after infrastructure re-installation (occasional)
- 3. LAN cable disconnected or defective (common)
- 4. NGN Connector board defective (occasional)
- 5. Hardware defect in network switch (rare)
- 6. Configuration problem in switch (common)
- 7. Monitor and PIC software not compatible (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Operating temperature shows problem
Diagnostic message indicates module operating temperature outside specified limits
Ranked causes
- 1. Defective fan (occasional)
- 2. Insufficient air circulation (occasional)
- 3. Operating environment outside limits (occasional)
- 4. Defective IR measurement assembly head (rare)
- 5. Defective agent ID head (rare)
- 6. Defective power supply (rare)
- 7. Defective main PC board (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Error LED stays on continuously caution
Red error indicator illuminated constantly after power-on
Ranked causes
- 1. External connected device causing fault signal (common)
- 2. I/O board defective (occasional)
- 3. Touch controller, flat panel adapter, ECG-Out, or MSL-2 board defective (rare)
- 4. Measurement server mount defective (rare)
- 5. Integrated module slot defective (rare)
- 6. Video board defective (rare)
- 7. Main board defective (rare)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs illuminate but no alarm sound issued patient safety
Remote alarm device shows LEDs but produces no audible alarm
Ranked causes
- 1. Remote alarm device in wrong I/O slot (common)
- 2. Speaker defective (occasional)
- 3. Remote device interface defective (occasional)
- 4. Cabling not connected (common)
- 5. Cabling defective (occasional)
- 6. I/O board defective (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anesthetic agent not identified or measured patient safety
Agent measurement missing; automatic agent identification fails; agent not detected; wrong agent identified
Ranked causes
- 1. Agent sample line not connected (common)
- 2. Agent parameter not enabled on monitor (common)
- 3. Incorrect agent configuration (auto vs manual mode) (occasional)
- 4. No anesthetic agent being administered (occasional)
- 5. Agent-ID assembly cable or connector failure (occasional)
- 6. Agent-ID assembly hardware failure (rare)
Verify agent sample line is connected
- Expected
- Sample line connected to AGM agent inlet
- Tools
- Visual inspection
5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
O2 Assembly shows problem
Diagnostic message indicates oxygen measurement assembly failure or measurement anomaly
Ranked causes
- 1. O2 jumpers incorrectly set (occasional)
- 2. O2 sensor failed calibration (occasional)
- 3. Defective O2 sensor (occasional)
- 4. Defective cables (rare)
- 5. Defective power supply (rare)
- 6. Defective main PC board (rare)
Measure O2 sensor output voltage
- Test point
- TP8
- Expected
- 1% O2 approx 10mV; 20.8% O2 approx 208mV; 50% O2 approx 500mV
- Tools
- Voltmeter
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speed point attached to remote extension box not functioning
Speed point device connected to remote extension box does not respond
Ranked causes
- 1. Remote extension device not connected to monitor (common)
- 2. Speed point not connected properly (common)
- 3. Speed point defective (occasional)
- 4. Remote extension device defective (occasional)
- 5. Remote device I/O board in wrong slot (occasional)
- 6. Remote device I/O board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SpeedPoint attached to Remote Extension Box not functioning
SpeedPoint input control on remote extension not responding
Ranked causes
- 1. Remote Extension Device not connected (common)
- 2. SpeedPoint not connected properly (common)
- 3. SpeedPoint defective (occasional)
- 4. Remote Extension Device defective (occasional)
- 5. Remote Device I/O board in wrong slot (rare)
- 6. Remote Device I/O board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Agent ID assembly shows problem
Diagnostic message indicates Agent ID assembly failure or anomaly
Ranked causes
- 1. Wrong Agent-ID software revision (occasional)
- 2. Defective agent ID head (occasional)
- 3. Defective power supply (rare)
- 4. Defective cables (rare)
- 5. Defective main PC board (rare)
Measure IR source voltage for agent ID head
- Test point
- TP11
- Expected
- 7.92V ±20mV, ±20mV
- Tools
- Voltmeter
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Anesthetic gas module does not communicate with monitor caution
LED steadily on or flashing; INOP GA EQUIP MALFUNCT displayed; module not responding to monitor
Ranked causes
- 1. Anesthesia software bundle not installed (common)
- 2. Incorrect RS232 or MIB port configuration (common)
- 3. Physical cable damage or loose connections (occasional)
- 4. AGM hardware failure (occasional)
- 5. Monitor motherboard failure (rare)
Verify Anesthesia software bundle is installed
- Expected
- IntelliVue Anesthesia #H30 present
- Tools
- Monitor
3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
External Display blank
External video display connected to monitor shows blank screen.
Ranked causes
- 1. Video cable not connected (common)
- 2. External display has no power (common)
- 3. External display defective (occasional)
- 4. Video board defective (occasional)
- 5. Main board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard or mouse on Remote Extension Box not functioning
Input devices connected to remote extension box not responding
Ranked causes
- 1. Remote Extension Box not connected (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. Remote Extension Device defective (occasional)
- 4. Remote Device I/O board in wrong slot (rare)
- 5. Remote Device I/O board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/mouse not functioning when attached to remote extension box
Keyboard or mouse connected to remote extension box does not respond
Ranked causes
- 1. Remote extension box not connected to monitor or device not connected to box (common)
- 2. Keyboard/mouse defective (occasional)
- 3. Remote extension device defective (occasional)
- 4. Remote device I/O board in wrong slot (occasional)
- 5. Remote device I/O board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG-OUT signal to the Defib high voltage
ECG output signal not transmitted to defibrillator device
Ranked causes
- 1. Cabling not connected (common)
- 2. ECG OUT board defective (red LED does not switch on for about 1 second at power on) (occasional)
- 3. Main Board defective (red LED switches on and remains on for more than 20 seconds) (occasional)
- 4. ECG OUT board defective (external electronic defect) (occasional)
- 5. Main Board defective (external electronic defect) (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless network: No central monitoring
Wireless connection failed; monitor out of range or adapter powerless
Ranked causes
- 1. Monitor out of access point range (common)
- 2. Wireless adapter has no power (occasional)
- 3. Wrong firmware revision in adapter (occasional)
- 4. Wireless adapter defective (occasional)
- 5. Wrong adapter or access point configuration (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC LED does not light up high voltage
Power indicator LED fails to illuminate when AC mains are connected and power button is pressed
Ranked causes
- 1. AC mains not powered or power cord defective (common)
- 2. Power Switch LED defective (occasional)
- 4. Power supply defective (occasional)
- 5. Integrated Module Slot defective (rare)
8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AGM connected to RS232 port not functioning
MIB/RS232 port communication failure with AGM device
Ranked causes
- 1. MIB/RS232 port not configured for AGM (common)
- 2. Cable between AGM and monitor not connected correctly or defective (common)
- 3. MIB/RS232 board in wrong slot (occasional)
- 4. MIB/RS232 board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AGM on RS232 port not functioning
AGM device connected to MIB/RS232 port is not operational
Ranked causes
- 1. MIB/RS232 port not configured for AGM (common)
- 2. Cable between AGM and monitor not connected or defective (occasional)
- 3. MIB/RS232 board in wrong slot after reconfiguration (occasional)
- 4. MIB/RS232 board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no alarm sound issued patient safety
Monitor displays alarm but produces no audible alert signal
Ranked causes
- 1. Audible alarm indicators switched off (common)
- 2. Volume set to zero (common)
- 3. Speaker defective (occasional)
- 4. Sound amplifier defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm on screen but no LED or sound on alarm device
Alarm is triggered and displayed on monitor screen, but remote alarm device produces no LED indication or sound.
Ranked causes
- 1. Cabling not connected (common)
- 2. Cabling defective (occasional)
- 3. I/O board defective (occasional)
- 4. Remote Alarm Device defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Gas flow performance
Annual check of gas delivery rate from 15210B calibration unit to ensure accuracy of calibration gas supply to tcpO2/tcpCO2 module
Ranked causes
- 1. Gas flow outside specification (unknown)
- 2. Gas flow less than minimum (occasional)
- 3. Gas flow greater than maximum (occasional)
- 4. Gas cylinder pressure depleted (common)
Verify cylinder has adequate gas pressure
- Test point
- Pressure indicator on front panel
- Expected
- Not in black zone
1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Gas measurement missing or inaccurate patient safety
Gas measurement display absent, incorrect, or waves disappearing; calibration failures; incorrect readings
Ranked causes
- 1. Sample line not connected (common)
- 2. Sample line kinked, blocked, or damaged (common)
- 3. Calibration failure or drift (occasional)
- 4. Gas parameter not enabled on monitor (occasional)
Verify sample line is physically connected to AGM
- Expected
- Sample line connected to AGM inlet
- Tools
- Visual inspection
3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
IR measurement assembly shows problem
Diagnostic message indicates IR measurement assembly failure or measurement anomaly
Ranked causes
- 1. Defective IR measurement assembly head (occasional)
- 2. Defective power supply (rare)
- 3. Defective cables (rare)
- 4. Defective main PC board (rare)
Measure IR source voltage for measurement assembly
- Test point
- TP6
- Expected
- 7.87V ±20mV, ±20mV
- Tools
- Voltmeter
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard or mouse attached to monitor not functioning
Input device directly connected to monitor not responding to user input
Ranked causes
- 1. Keyboard/Mouse not connected properly (common)
- 2. Keyboard/Mouse defective (occasional)
- 3. PS/2 I/O board in wrong slot (rare)
- 4. PS/2 I/O board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Keyboard/mouse not functioning when attached directly
Keyboard or mouse connected directly to monitor does not respond
Ranked causes
- 1. Keyboard/mouse not connected properly (common)
- 2. Keyboard/mouse defective (occasional)
- 3. PS/2 or USB I/O board in wrong slot (occasional)
- 4. PS/2 or USB I/O board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarmed but Nurse Call did not activate patient safety
Monitor alarm triggered but nurse call system not activated
Ranked causes
- 1. Incorrect configuration of relay latency or trigger (common)
- 2. Cable connection to monitor or nurse call system incorrect (common)
- 3. Nurse Call Relay board in wrong slot (occasional)
- 4. Nurse Call Relay board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor alarmed but Nurse Call not activated patient safety
Monitor generates alarm but nurse call relay does not trigger
Ranked causes
- 1. Incorrect configuration of relay latency or trigger (common)
- 2. Cable connection to monitor or nurse call system incorrect (occasional)
- 3. Nurse Call Relay board in wrong slot (occasional)
- 4. Nurse Call Relay board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity to PIC
No network connection to Philips Information Center; no error message displayed
Ranked causes
- 1. Hardware defect - LAN cable (common)
- 2. NGN Connector board defective (occasional)
- 3. Network switch problem (occasional)
- 4. Configuration problem (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
One LED next to USB port with connected device is off
USB port LED indicator off despite device connected to port
Ranked causes
- 1. Device not supported (occasional)
- 2. Printer not switched on (common)
- 3. Short-circuit (occasional)
- 4. Connected devices drawing too much electricity (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Optical path shows problem
Diagnostic message indicates optical path measurement assembly anomaly or failure
Ranked causes
- 1. Defective or contaminated sample cell (occasional)
- 2. Defective IR measurement assembly head (occasional)
- 3. Defective power supply (rare)
- 4. Defective main PC board (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power supply shows problem caution
Diagnostic message indicates power supply voltage anomaly or failure
Ranked causes
- 1. Weak pump (occasional)
- 2. Defective power supply (occasional)
- 3. Defective cables (rare)
- 4. Defective main PC board (rare)
Measure +12V HP supply voltage
- Test point
- TP14
- Expected
- +12V ±600mV, ±600mV
- Tools
- Voltmeter
4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Print device local unavailable, printer job stalled
Local printer status shows unavailable, print jobs cannot complete
Ranked causes
- 1. Printer not switched on (common)
- 2. Printer paper tray empty (common)
- 3. Cabling not connected correctly (occasional)
- 4. I/O board defect (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speed point attached directly to monitor not functioning
Speed point device connected directly to monitor does not respond
Ranked causes
- 1. Speed point not connected properly (common)
- 2. PS/2 or USB I/O board in wrong slot (common)
- 3. PS/2 or USB I/O board defective (occasional)
- 4. Speed point defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speed Point not functioning when directly attached
Speed Point input device connected directly to monitor does not respond.
Ranked causes
- 1. Speed Point not connected properly (common)
- 2. PS/2 I/O board in wrong slot (occasional)
- 3. PS/2 I/O Board defective (occasional)
- 4. SpeedPoint defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Status message 'Print device Local 1 (Local 2) unavailable' - printer job stalled
Local printer not available; printer off, out of paper, cabling issue, or I/O board defective
Ranked causes
- 1. Printer not switched on (common)
- 2. Printer paper tray empty (common)
- 3. Cabling not connected correctly (common)
- 4. PS/2 or USB I/O board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Wireless out of range message issued patient safety
INOP message WIRELESS OUT OF RANGE displayed; monitor lost connection to information center
Ranked causes
- 1. LAN cable disconnected (common)
- 2. Information center disconnected (common)
- 3. Network infrastructure failure (occasional)
- 4. Monitor out of wireless range (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Zero calibration failed
Gas analyzer zero calibration fails to complete or produces out-of-range results
Ranked causes
- 1. Solenoid or air reference filter problem (common)
- 2. Occluded pneumatics (common)
- 3. Flowrate problem (common)
- 4. IR measurement head problem (occasional)
Verify solenoid air draw
- Test point
- Room air filter
- Expected
- Air drawn at filter during calibration
- Tools
- Hand
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs illuminate but no alarm sound on Remote Alarm Device
LEDs on remote alarm device light up but speaker does not produce sound.
Ranked causes
- 1. Wrong I/O slot (common)
- 2. Speaker defective (occasional)
- 3. Remote device I/F defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Communication problem with IntelliVue 802.11 Bedside Adapter or low RSSI
Monitor cannot communicate with wireless bedside adapter or signal strength below acceptable level (RSSI < 30)
Ranked causes
- 1. Network infrastructure not functioning properly (common)
- 2. Antenna cable disconnected or defective (common)
- 3. Cable connection between adapter and system interface board defective (common)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG Out/Marker In function does not function
ECG output and/or marker input functionality not operational
Ranked causes
- 1. MSL cable hardware problem (common)
- 2. ECG Out hardware defect (occasional)
- 3. MSL connector defect in measurement server (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
ECG Out/Marker In function not functioning
ECG output or marker input feature not working
Ranked causes
- 1. Hardware problem - MSL cable (common)
- 2. ECG Out hardware defective in monitor (occasional)
- 3. MSL connector defective in measurement server (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External device not receiving data
Data export from monitor to external device not functioning
Ranked causes
- 2. Wrong data export protocol driver configured (common)
- 3. Cable between external device and monitor not connected or defective (common)
- 4. External device does not support data export protocol version (occasional)
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
General monitor INOP: CHECK MONITOR TEMP
Temperature inside the monitor is too high
Ranked causes
- 1. Monitor ventilation obstructed (common)
- 2. Environmental conditions cause overheating (occasional)
- 3. Main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
General monitor INOP: INTERNAL COMM. MALF.
Problem with I2C Bus communication in the monitor
Ranked causes
- 1. Video board defective or communication issue (occasional)
- 2. Video board defective (occasional)
- 3. Main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message 'Check Alarm Lamps' issued
Alarm LED board cable disconnected, board defective, or main board defective
Ranked causes
- 1. Alarm LED board cable disconnected (common)
- 2. Alarm LED board defective (common)
- 3. Main board defective (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message 'MsmtSrv not Supp' issued
System indicates measurement server is not supported by monitor; in-operative condition declared
Ranked causes
- 1. Wrong software revision on monitor or measurement server (common)
- 2. Too many measurement servers connected (occasional)
- 3. Unsupported type of measurement server connected (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Message SERVERLINK MALF displayed
Beep every two seconds audible indicator; FMS or MMS with incompatible software revision connected or hardware communication fault
Ranked causes
- 1. FMS or MMS with incompatible software revision connected (common)
- 2. Communication between components not functioning (common)
- 3. Hardware for communicating with Multi-Measurement Server is faulty (occasional)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message SPEAKER MALFUNCTION displayed caution
Speaker malfunction indicated. No alarm sound output capability.
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message: Bad Server Link
Unexpected data or communication failure in MSL connection
Ranked causes
- 1. Unexpected data on MSL (occasional)
- 2. Incompatible software revision (occasional)
- 3. Communication failure (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message: Check Alarm Lamps caution
Alarm LED system malfunction indicated
Ranked causes
- 1. Alarm LED board cable disconnected (common)
- 2. Alarm LED board defective (occasional)
- 3. Main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message: MsmtSrv not Supp
Measurement Server software incompatible with monitor software version
Ranked causes
- 1. Wrong software revision (common)
- 2. Too many measurement servers connected (occasional)
- 3. Unsupported type of measurement server (occasional)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP message: Unsupported LAN
Network interface error; may indicate network failure or configuration issue
Ranked causes
- 1. Network failure (occasional)
- 2. Monitor connected to wrong network (occasional)
- 3. IP address conflict (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IntelliVue 802.11 wireless adapter not connecting
Monitor with 802.11 adapter fails to authenticate or connect to access point; RSSI signal weak or connection status shows disconnected.
Ranked causes
- 1. WLAN configuration incorrect (common)
- 2. Access point distance or antenna orientation incorrect (common)
- 3. Access point incompatibility or failure (occasional)
Access WLAN Diagnostic service window
- Test point
- Service Mode > Main Setup > WLAN Diagnostic
- Expected
- Service window opens with Conn.Status field visible
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
More symptoms
- Measurement Server configuration not supported
- Medical electrical system non-compliance after installation
- Message 'Measurement Server Configuration not supported' issued
- Other Bed Overview not available
- Oxygen (O2) not measured
- Red Error LED stays on
- Touchscreen not responding to touch input
- Unit does not power on
- Wireless Adapter has no power
- Wireless network: Frequent dropouts and disconnects
- Alarm LEDs do not illuminate
- Alarm LEDs remain off
- Alarm occurs but no LED lights up
- Calibration: Anesthetic Gas Module
- Care and cleaning of external surfaces
- Disappearing waves in warmup mode
- Equipment leakage current excessive (normal condition)
- Equipment leakage current excessive (single fault condition)
- Flexible Module Rack not communicating with monitor
- Flowrate outside specification during performance test
- Frequent dropouts and network disconnects on wireless
- General monitor INOP: CHECK INTERNAL VOLTAGE
- General monitor INOP: SETTINGS MALFUNCTION
- Inserted module LEDs normal but module not recognized
- Leakage detected during performance test
- LEDs on IntelliVue 802.11 Bedside Adapter are off and remain off
- Measurement module not recognized despite LEDs normal
- Measurement Server does not start - LEDs normal
- Measurement Server does not start up; no LEDs active; no INOP or prompt displayed
- Monitor connected to LAN but not monitored at central
- Network prompt: No central assigned to bed
- Network related: NO CENTRAL MONITORING message
- No connectivity in multi-access-point coverage area
- No test sound issued
- Normal operation waves or numerics missing
- O2 measurement unable to measure
- Paper not feeding properly
- Parameter modules not detected or not functional
- Poor print quality
- Poor print quality from recorder
- Print device remote unavailable, printer job stalled
- Print job could not be queued, no print device found
- Prompt message 'Print job could not be queued'
- Protective earth resistance excessive
- Pump malfunction or failure
- Recorder not communicating with system
- Recorder paper not feeding properly
- Red Error LED blinks
- Red Error LED blinks (indicating cyclic reboots)
- Red Error LED flashes
- Sidestream CO2 flow rate out of tolerance
- Solenoid valve #2 not functioning
- Status message 'Print device Remote 1 (Remote 2, Remote 3) unavailable' - job stalled
- Top cover PC board malfunction or disconnection
- Watertrap manifold or protector damaged
- 'Other Bed' Alarms not appearing
- Agent gas measurement unable to measure
- Agent ID head damaged or misaligned
- Alarm occurs on screen but no LED or sound on alarm device
- Alarm on VueLink device but no sound on monitor
- Alarm sound issued but no LEDs light up
- All LEDs on USB board are off
- Applied part leakage current exceeds limit
- Applied part leakage current measurement (invasive connectors)
- Barometric pressure error outside tolerance
- Calibration: Anesthetic Gas Module flowrate measurement
- Calibration: Anesthetic Gas Module performance verification
- Calibration: Anesthetic Gas Module safety verification
- Calibration: Barometric pressure check and calibration
- Calibration: Microstream CO2
- Calibration: NBP
- Calibration: Noninvasive Blood Pressure Measurement
- Calibration: Safety testing after top cover removal
- Calibration: Span calibration check
- Calibration: Total flowrate check in normal mode
- Calibration: Touchscreen
- Calibration: Zero calibration
- CO2 measurement unable to measure
- CO2 scrubber replacement procedure
- Connector holder insertion into dovetail connection
- Cooling fan not operating smoothly
- Cooling fan not running or operating abnormally
- Diagnostic error in Gas Analyzer Diagnostic window
- Disposal: Empty calibration gas cylinder
- Dual link bar assembly with MSL flex connector
- ECG heart rate measurement inaccuracy
- ECG synchronization with defibrillator fails
- Equipment leakage current exceeds limits
- Final inspection of assembly
- Firmware revision in wireless adapter is wrong
- Flex cable connector insertion and alignment
- Flexible Module Rack LEDs ok but module not recognized
- FMS cables not properly connected
- General monitor INOP: MCC UNSUPPORTED
- Image sticking on LCD display
- INOP 'BAD SERVER LINK' issued
- INOP 'MSL Power High' issued
- INOP 'MSL Power Off' issued
- INOP 'MSL Power Overload' issued
- INOP EcgOut EQUIP MALF is issued
- INOP message CHECK NURSE CALL RELAY
- INOP message CHECK NURSE CALL RELAY is issued
- INOP message: CHECK INPUT DEVICES
- INOP message: MSL Power High
- INOP message: MSL Power Off
- INOP message: MSL Power Overload
- INOP message: Serverlink Malf
- Inserted module does not function
- Invasive pressure performance test failure
- IT equipment in patient vicinity leakage current non-compliance
- Mainstream CO2 accuracy test failure
- Measurement from server does not appear on monitor
- Measurement module currently ignored
- Measurement Module in Flexible Module Rack currently ignored
- Measurement Module in Integrated Module Slot currently ignored
- Measurement parameter not displayed on monitor
- Measurement Server connection error
- Measurement Server does not start - no power
- Measurement Server does not start but LEDs are normal
- Messages 'Measurement Server not supported, unplug device, switch monitor off/on' and INOP 'Bad Measurement Server' issued
- Module does not lock into FMS
- Monitor is out of range of wireless access point
- Monitor not progressing through boot sequence
- N2O measurement unable to measure
- NBP accuracy test failure
- NBP linearity test failure
- NBP system leakage detected
- NBP valve closure test failure
- Network connection lost
- Network related: UNSUPPORTED LAN message
- No central monitoring possible via wireless adapter
- No marker pulse displayed on the monitor
- No Network icon or Network icon flashes; no association to central station
- None of the connected USB devices are functioning
- Nurse Call Relay not functioning
- Overview and Printing do not work on wireless
- Power fuse failure or blown fuse
- Power Loss Alarm Buzzer not sounding
- Printing on remote device reported but no output
- Printouts are not as expected
- Prompt message 'Too many <label> modules connected'
- Prompt message: Too many modules of same type
- Protective earth leakage current from multiple socket outlet exceeds limit
- Pump filter maintenance required or failed
- Pump replacement and flow calibration
- Pump test failure
- Recorder module does not lock into FMS
- Remote extension device buttons not functioning
- Remote Extension Device buttons not responding
- Remote Extension Device input not functioning
- Replacement: Pump and scrubber
- Respiration rate measurement inaccuracy
- Routine maintenance: Gas cylinder replacement
- Sidestream CO2 accuracy test failure
- Span calibration check failure for gas channels
- Speed point knob rotation or joystick control not functioning
- SpeedPoint knob or joystick control not functioning
- SpO2 measurement out of expected range
- Status message 'Incompatible SW Revision versions' issued
- Status message 'Printing on device Remote 1...' but no report printed
- System thinks door is open when it is not
- System thinks recorder door is open when closed
- System thinks recorder is out of paper when it is not
- System thinks recorder out of paper when not
- Temperature measurement performance test failure
- Top cover refitting with MSL flex cable protection
- Unit setup before initial use
- Unrecognized Measurement Module in Flexible Module Rack slot
- Unrecognized Measurement Module in Integrated Module Slot
- Unrecognized measurement module in slot
- Wireless adapter defective
- Wireless monitor losing access point connection
- Wireless network: No connectivity with multiple access points
- Wireless network: Overview or printing not working
- Zero calibration check failure
175 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
16030 | Monitor spontaneously reboots | Spontaneous reboot without user interaction | critical |
16400 | Monitor spontaneously reboots | Spontaneous reboot without user interaction | critical |
16505 | Loss of configuration after system reboot | System reboot | critical |
17345 | MMS reboots unexpectedly | Spontaneous reboot of multi-measurement server | critical |
20005 | Loss of configuration after system reboot - monitor runs in default safe profile | System reboot or power cycle | error |
20212 | Monitor spontaneously reboots | Spontaneous reboot without user interaction | critical |
21002 | Monitor reboots upon removal of modules (M8048A) | Removal of I/O modules | error |
21002 | For M8048: Modules not recognized or monitor reboots upon module removal; For M3001: Monitor reboots after Temp or Pressure unused | Hardware module removal or unused module state | error |
21400 | MMS reboots unexpectedly, no waves or numerics displayed | Spontaneous reboot of multi-measurement server | critical |
30049 | Monitor spontaneously reboots | Spontaneous reboot without user interaction | critical |
32750 | Modules not recognized (M8048A) | Plug or removal of modules | error |
495 | Monitor spontaneously reboots | Spontaneous reboot without user interaction | critical |
500 | Monitor spontaneously reboots | Spontaneous reboot without user interaction | critical |
AGENT IDENT MALF | Agent identification sensor malfunction | Serious Agent-ID hardware problem | error |
AGENT IDENT ZERO FAILED | Agent identification sensor zero calibration failed | Solenoid 2 or Agent-ID problem | error |
CHECK INPUT DEVICES | Speed point or other input device defective | Input device malfunction detected during initialization or operation | warning |
CHECKINTERNALVOLTAGE | Problem with internal power supply voltages | Monitor detects abnormal 5V or 12V voltage levels | error |
CHECK MENU LABELS | Menu labels configuration error | Monitor performs cold start and attempts to reset menu labels to defaults | warning |
CHECK MONITOR FUNC | Alarm lamps, display or interfaces not functioning correctly | Detected when internal voltages are too low | error |
CHECK MONITOR TEMP | Temperature inside the monitor is too high | Temperature exceeds maximum operating limit | warning |
CHECK SETTINGS | Settings configuration error | Monitor performs cold start and settings default reset is detected | warning |
GA INCOMPATIBLE | Gas analyzer module version incompatible with monitor | Incompatible AGM version | error |
GA NOT AVAILABLE | Gas analyzer module not available to monitor | AGM not switched on or not properly connected | error |
GAS AN ACCURACY ? | Gas analyzer accuracy questionable | Measurement accuracy degraded | warning |
GAS EQUIP MALF | Gas equipment malfunction detected | AGM-monitor connection problem, subassembly failure, or Main PC Board issue | error |
GAS OCCLUSION | Gas flow occlusion detected | Blockage in gas flow path, external or internal | error |
GA ZERO FAILED | Gas analyzer zero calibration failed | Calibration procedure failed or produced out-of-range values | error |
INOP NBP Equipment Malfunction | Non-invasive blood pressure equipment malfunction preventing normal operation | Occurs in monitoring mode when NBP calibration is invalid or hardware fault detected | error |
INTERNAL COMM.MALF. | Internal I2C bus communication failure | Monitor detects I2C bus communication error | error |
MCC UNSUPPORTED | Unsupported MSL coupling cable connection detected | MSL coupling cable connected to incompatible device | warning |
MONITOR LABEL DUPLICATE | Another monitor on the network has the same equipment label | Duplicate equipment labels detected on network | warning |
NO CENTRAL MONITORING | Monitor is not assigned to sector or central connection lost | Monitor not assigned to monitoring sector or connection to information center lost | error |
O2 EQUIP MALF | Oxygen sensor equipment malfunction | O2 span calibration failed or sensor configuration error | error |
O2 ZERO FAILED | Oxygen sensor zero calibration failed | O2 sensor calibration failed or produced out-of-range constants | error |
SETTINGS MALFUNCTION | Configuration cloning failure | Problem occurs during configuration file cloning process | error |
SETTINGS MALFUNCTION | Problem during cloning process or memory corruption | Settings memory space corrupted or cloning process failed | error |
UNSUPPORTED LAN | Network failure detected after 1 minute of operation | Network infrastructure failure or extended communication loss | critical |
WIRELESS OUT OF RANGE | Monitor lost connection to the Information Center | Connection to information center lost or wireless access point unavailable | error |
XXX MEAS DISTURBED | Gas measurement disturbed (N2O, CO2, agent, or O2) | Transient measurement assembly problem | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI