Philips IntelliVue MP60: troubleshooting and error codes

The ServiceForge knowledge base holds 235 documented symptoms for the Philips IntelliVue MP60, with 524 ranked causes, 228 diagnostic checks with test points and expected values, and 39 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI

Symptoms

Green Power On LED and Red Error LED remain off after pressing power button high voltage

Both power on and error indicator LEDs fail to illuminate when power button is pressed after boot attempt

Ranked causes

  1. 1. Remote devices causing power fault (common)
  2. 2. Power Switch Micro Controller hung (common)
  3. 3. Power switch board not connected to main board (occasional)
  4. 4. Power Switch Board defective (occasional)
  5. 5. I/O Board defective (occasional)
  6. 6. Touch controller, Flat Panel Adapter, ECG-Out or MSL-2 Board defective (occasional)
  7. 7. Measurement Server Mount defective (rare)
  8. 8. Integrated Module Slot defective (rare)
  9. 9. Video Board defective (rare)
  10. 10. Main Board defective (rare)

10 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Integrated display blank or brightness reduced

Display is blank or has reduced brightness. May occur when room temperature exceeds 40°C or below 5°C.

Ranked causes

  1. 1. Display overheating from high room temperature (common)
  2. 2. Flat Panel Adapter cable disconnected (common)
  3. 3. Backlight Inverter Cable disconnected (common)
  4. 4. Backlight tubes defective (occasional)
  5. 5. Backlight Inverter board defective (occasional)
  6. 6. Panel Adapter board defective (occasional)
  7. 7. LCD Flat panel defective (occasional)
  8. 8. Video board defective (occasional)
  9. 9. Main board defective (rare)
First check

Connect external XGA display

Test point
Video port
Expected
External display works or does not work

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Pneumatic system shows problem

Diagnostic message indicates pneumatic system failure or anomaly detected during operation

Ranked causes

  1. 1. Weak pump (occasional)
  2. 2. Occluded pneumatics (occasional)
  3. 3. Defective cables (rare)
  4. 4. Solenoid or air reference filter problem (occasional)
  5. 5. Flowrate problem (occasional)
  6. 6. Flow transducer incorrectly connected (rare)
  7. 7. Pump problem (occasional)
  8. 8. Defective power supply (rare)
  9. 9. Defective main PC board (rare)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power-on LEDs do not illuminate caution

Green Power On LED and Red Error LED remain off after pressing power button

Ranked causes

  1. 1. Remote devices causing interference (common)
  2. 2. Power switch microcontroller hung or frozen (common)
  3. 3. Power switch board not connected to main board (occasional)
  4. 4. Power switch board defective (occasional)
  5. 5. I/O board defective (rare)
  6. 6. Touch controller, video, ECG-Out, or MSL-2 board defective (rare)
  7. 7. Measurement server mount defective (rare)
  8. 8. Video board defective (rare)
  9. 9. Main board defective (rare)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Touch screen not functioning

Touch panel does not respond to user input; touches are not registered or recognized by the system

Ranked causes

  1. 1. Touchscreen functionality temporarily disabled (common)
  2. 2. Touchscreen functionality permanently disabled (occasional)
  3. 3. Touch controller not connected (occasional)
  4. 4. Touch controller cable not connected (occasional)
  5. 5. Touch panel cable not connected (occasional)
  6. 6. Touch controller board defective (rare)
  7. 7. Touch screen defective (rare)
  8. 8. Main board defective (rare)
  9. 9. Touch not calibrated (occasional)

9 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

External Touch Display not functioning

External touchscreen input device is not responding or not working.

Ranked causes

  1. 1. Touchscreen temporarily disabled (common)
  2. 2. Touchscreen permanently disabled (common)
  3. 3. External Touch cable not connected (common)
  4. 4. External Touch driver configuration error (occasional)
  5. 5. MIB Board defective (occasional)
  6. 6. External touch defective (occasional)
  7. 7. Main board defective (rare)
  8. 8. Touch not calibrated (common)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Integrated Touch Display not functioning

Touchscreen on integrated display is not responsive or not working.

Ranked causes

  1. 1. Touchscreen temporarily disabled (common)
  2. 2. Touchscreen permanently disabled (common)
  3. 3. Touch controller cable disconnected (occasional)
  4. 4. Touch panel cable disconnected (occasional)
  5. 5. Touch controller board defective (occasional)
  6. 6. Touch Sensor defective (occasional)
  7. 7. Main board defective (rare)
  8. 8. Touch not calibrated (common)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No connectivity to PIC, no prompt or error message on monitor

Monitor cannot communicate with Philips Information Center; network connection fails silently without error indication

Ranked causes

  1. 1. Monitor connected to wrong network (common)
  2. 2. IP address conflict after infrastructure re-installation (occasional)
  3. 3. LAN cable disconnected or defective (common)
  4. 4. NGN Connector board defective (occasional)
  5. 5. Hardware defect in network switch (rare)
  6. 6. Configuration problem in switch (common)
  7. 7. Monitor and PIC software not compatible (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Operating temperature shows problem

Diagnostic message indicates module operating temperature outside specified limits

Ranked causes

  1. 1. Defective fan (occasional)
  2. 2. Insufficient air circulation (occasional)
  3. 3. Operating environment outside limits (occasional)
  4. 4. Defective IR measurement assembly head (rare)
  5. 5. Defective agent ID head (rare)
  6. 6. Defective power supply (rare)
  7. 7. Defective main PC board (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Red Error LED stays on continuously caution

Red error indicator illuminated constantly after power-on

Ranked causes

  1. 1. External connected device causing fault signal (common)
  2. 2. I/O board defective (occasional)
  3. 3. Touch controller, flat panel adapter, ECG-Out, or MSL-2 board defective (rare)
  4. 4. Measurement server mount defective (rare)
  5. 5. Integrated module slot defective (rare)
  6. 6. Video board defective (rare)
  7. 7. Main board defective (rare)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LEDs illuminate but no alarm sound issued patient safety

Remote alarm device shows LEDs but produces no audible alarm

Ranked causes

  1. 1. Remote alarm device in wrong I/O slot (common)
  2. 2. Speaker defective (occasional)
  3. 3. Remote device interface defective (occasional)
  4. 4. Cabling not connected (common)
  5. 5. Cabling defective (occasional)
  6. 6. I/O board defective (occasional)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Anesthetic agent not identified or measured patient safety

Agent measurement missing; automatic agent identification fails; agent not detected; wrong agent identified

Ranked causes

  1. 1. Agent sample line not connected (common)
  2. 2. Agent parameter not enabled on monitor (common)
  3. 3. Incorrect agent configuration (auto vs manual mode) (occasional)
  4. 4. No anesthetic agent being administered (occasional)
  5. 5. Agent-ID assembly cable or connector failure (occasional)
  6. 6. Agent-ID assembly hardware failure (rare)
First check

Verify agent sample line is connected

Expected
Sample line connected to AGM agent inlet
Tools
Visual inspection

5 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

O2 Assembly shows problem

Diagnostic message indicates oxygen measurement assembly failure or measurement anomaly

Ranked causes

  1. 1. O2 jumpers incorrectly set (occasional)
  2. 2. O2 sensor failed calibration (occasional)
  3. 3. Defective O2 sensor (occasional)
  4. 4. Defective cables (rare)
  5. 5. Defective power supply (rare)
  6. 6. Defective main PC board (rare)
First check

Measure O2 sensor output voltage

Test point
TP8
Expected
1% O2 approx 10mV; 20.8% O2 approx 208mV; 50% O2 approx 500mV
Tools
Voltmeter

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Speed point attached to remote extension box not functioning

Speed point device connected to remote extension box does not respond

Ranked causes

  1. 1. Remote extension device not connected to monitor (common)
  2. 2. Speed point not connected properly (common)
  3. 3. Speed point defective (occasional)
  4. 4. Remote extension device defective (occasional)
  5. 5. Remote device I/O board in wrong slot (occasional)
  6. 6. Remote device I/O board defective (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

SpeedPoint attached to Remote Extension Box not functioning

SpeedPoint input control on remote extension not responding

Ranked causes

  1. 1. Remote Extension Device not connected (common)
  2. 2. SpeedPoint not connected properly (common)
  3. 3. SpeedPoint defective (occasional)
  4. 4. Remote Extension Device defective (occasional)
  5. 5. Remote Device I/O board in wrong slot (rare)
  6. 6. Remote Device I/O board defective (rare)

6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Agent ID assembly shows problem

Diagnostic message indicates Agent ID assembly failure or anomaly

Ranked causes

  1. 1. Wrong Agent-ID software revision (occasional)
  2. 2. Defective agent ID head (occasional)
  3. 3. Defective power supply (rare)
  4. 4. Defective cables (rare)
  5. 5. Defective main PC board (rare)
First check

Measure IR source voltage for agent ID head

Test point
TP11
Expected
7.92V ±20mV, ±20mV
Tools
Voltmeter

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Anesthetic gas module does not communicate with monitor caution

LED steadily on or flashing; INOP GA EQUIP MALFUNCT displayed; module not responding to monitor

Ranked causes

  1. 1. Anesthesia software bundle not installed (common)
  2. 2. Incorrect RS232 or MIB port configuration (common)
  3. 3. Physical cable damage or loose connections (occasional)
  4. 4. AGM hardware failure (occasional)
  5. 5. Monitor motherboard failure (rare)
First check

Verify Anesthesia software bundle is installed

Expected
IntelliVue Anesthesia #H30 present
Tools
Monitor

3 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

External Display blank

External video display connected to monitor shows blank screen.

Ranked causes

  1. 1. Video cable not connected (common)
  2. 2. External display has no power (common)
  3. 3. External display defective (occasional)
  4. 4. Video board defective (occasional)
  5. 5. Main board defective (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard or mouse on Remote Extension Box not functioning

Input devices connected to remote extension box not responding

Ranked causes

  1. 1. Remote Extension Box not connected (common)
  2. 2. Keyboard/Mouse defective (occasional)
  3. 3. Remote Extension Device defective (occasional)
  4. 4. Remote Device I/O board in wrong slot (rare)
  5. 5. Remote Device I/O board defective (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard/mouse not functioning when attached to remote extension box

Keyboard or mouse connected to remote extension box does not respond

Ranked causes

  1. 1. Remote extension box not connected to monitor or device not connected to box (common)
  2. 2. Keyboard/mouse defective (occasional)
  3. 3. Remote extension device defective (occasional)
  4. 4. Remote device I/O board in wrong slot (occasional)
  5. 5. Remote device I/O board defective (rare)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No ECG-OUT signal to the Defib high voltage

ECG output signal not transmitted to defibrillator device

Ranked causes

  1. 1. Cabling not connected (common)
  2. 2. ECG OUT board defective (red LED does not switch on for about 1 second at power on) (occasional)
  3. 3. Main Board defective (red LED switches on and remains on for more than 20 seconds) (occasional)
  4. 4. ECG OUT board defective (external electronic defect) (occasional)
  5. 5. Main Board defective (external electronic defect) (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wireless network: No central monitoring

Wireless connection failed; monitor out of range or adapter powerless

Ranked causes

  1. 1. Monitor out of access point range (common)
  2. 2. Wireless adapter has no power (occasional)
  3. 3. Wrong firmware revision in adapter (occasional)
  4. 4. Wireless adapter defective (occasional)
  5. 5. Wrong adapter or access point configuration (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AC LED does not light up high voltage

Power indicator LED fails to illuminate when AC mains are connected and power button is pressed

Ranked causes

  1. 1. AC mains not powered or power cord defective (common)
  2. 2. Power Switch LED defective (occasional)
  3. 4. Power supply defective (occasional)
  4. 5. Integrated Module Slot defective (rare)

8 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AGM connected to RS232 port not functioning

MIB/RS232 port communication failure with AGM device

Ranked causes

  1. 1. MIB/RS232 port not configured for AGM (common)
  2. 2. Cable between AGM and monitor not connected correctly or defective (common)
  3. 3. MIB/RS232 board in wrong slot (occasional)
  4. 4. MIB/RS232 board defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

AGM on RS232 port not functioning

AGM device connected to MIB/RS232 port is not operational

Ranked causes

  1. 1. MIB/RS232 port not configured for AGM (common)
  2. 2. Cable between AGM and monitor not connected or defective (occasional)
  3. 3. MIB/RS232 board in wrong slot after reconfiguration (occasional)
  4. 4. MIB/RS232 board defective (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm occurs but no alarm sound issued patient safety

Monitor displays alarm but produces no audible alert signal

Ranked causes

  1. 1. Audible alarm indicators switched off (common)
  2. 2. Volume set to zero (common)
  3. 3. Speaker defective (occasional)
  4. 4. Sound amplifier defective (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm on screen but no LED or sound on alarm device

Alarm is triggered and displayed on monitor screen, but remote alarm device produces no LED indication or sound.

Ranked causes

  1. 1. Cabling not connected (common)
  2. 2. Cabling defective (occasional)
  3. 3. I/O board defective (occasional)
  4. 4. Remote Alarm Device defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Calibration: Gas flow performance

Annual check of gas delivery rate from 15210B calibration unit to ensure accuracy of calibration gas supply to tcpO2/tcpCO2 module

Ranked causes

  1. 1. Gas flow outside specification (unknown)
  2. 2. Gas flow less than minimum (occasional)
  3. 3. Gas flow greater than maximum (occasional)
  4. 4. Gas cylinder pressure depleted (common)
First check

Verify cylinder has adequate gas pressure

Test point
Pressure indicator on front panel
Expected
Not in black zone

1 more check with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Gas measurement missing or inaccurate patient safety

Gas measurement display absent, incorrect, or waves disappearing; calibration failures; incorrect readings

Ranked causes

  1. 1. Sample line not connected (common)
  2. 2. Sample line kinked, blocked, or damaged (common)
  3. 3. Calibration failure or drift (occasional)
  4. 4. Gas parameter not enabled on monitor (occasional)
First check

Verify sample line is physically connected to AGM

Expected
Sample line connected to AGM inlet
Tools
Visual inspection

3 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

IR measurement assembly shows problem

Diagnostic message indicates IR measurement assembly failure or measurement anomaly

Ranked causes

  1. 1. Defective IR measurement assembly head (occasional)
  2. 2. Defective power supply (rare)
  3. 3. Defective cables (rare)
  4. 4. Defective main PC board (rare)
First check

Measure IR source voltage for measurement assembly

Test point
TP6
Expected
7.87V ±20mV, ±20mV
Tools
Voltmeter

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard or mouse attached to monitor not functioning

Input device directly connected to monitor not responding to user input

Ranked causes

  1. 1. Keyboard/Mouse not connected properly (common)
  2. 2. Keyboard/Mouse defective (occasional)
  3. 3. PS/2 I/O board in wrong slot (rare)
  4. 4. PS/2 I/O board defective (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Keyboard/mouse not functioning when attached directly

Keyboard or mouse connected directly to monitor does not respond

Ranked causes

  1. 1. Keyboard/mouse not connected properly (common)
  2. 2. Keyboard/mouse defective (occasional)
  3. 3. PS/2 or USB I/O board in wrong slot (occasional)
  4. 4. PS/2 or USB I/O board defective (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor alarmed but Nurse Call did not activate patient safety

Monitor alarm triggered but nurse call system not activated

Ranked causes

  1. 1. Incorrect configuration of relay latency or trigger (common)
  2. 2. Cable connection to monitor or nurse call system incorrect (common)
  3. 3. Nurse Call Relay board in wrong slot (occasional)
  4. 4. Nurse Call Relay board defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Monitor alarmed but Nurse Call not activated patient safety

Monitor generates alarm but nurse call relay does not trigger

Ranked causes

  1. 1. Incorrect configuration of relay latency or trigger (common)
  2. 2. Cable connection to monitor or nurse call system incorrect (occasional)
  3. 3. Nurse Call Relay board in wrong slot (occasional)
  4. 4. Nurse Call Relay board defective (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

No connectivity to PIC

No network connection to Philips Information Center; no error message displayed

Ranked causes

  1. 1. Hardware defect - LAN cable (common)
  2. 2. NGN Connector board defective (occasional)
  3. 3. Network switch problem (occasional)
  4. 4. Configuration problem (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

One LED next to USB port with connected device is off

USB port LED indicator off despite device connected to port

Ranked causes

  1. 1. Device not supported (occasional)
  2. 2. Printer not switched on (common)
  3. 3. Short-circuit (occasional)
  4. 4. Connected devices drawing too much electricity (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Optical path shows problem

Diagnostic message indicates optical path measurement assembly anomaly or failure

Ranked causes

  1. 1. Defective or contaminated sample cell (occasional)
  2. 2. Defective IR measurement assembly head (occasional)
  3. 3. Defective power supply (rare)
  4. 4. Defective main PC board (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Power supply shows problem caution

Diagnostic message indicates power supply voltage anomaly or failure

Ranked causes

  1. 1. Weak pump (occasional)
  2. 2. Defective power supply (occasional)
  3. 3. Defective cables (rare)
  4. 4. Defective main PC board (rare)
First check

Measure +12V HP supply voltage

Test point
TP14
Expected
+12V ±600mV, ±600mV
Tools
Voltmeter

4 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

Print device local unavailable, printer job stalled

Local printer status shows unavailable, print jobs cannot complete

Ranked causes

  1. 1. Printer not switched on (common)
  2. 2. Printer paper tray empty (common)
  3. 3. Cabling not connected correctly (occasional)
  4. 4. I/O board defect (rare)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Speed point attached directly to monitor not functioning

Speed point device connected directly to monitor does not respond

Ranked causes

  1. 1. Speed point not connected properly (common)
  2. 2. PS/2 or USB I/O board in wrong slot (common)
  3. 3. PS/2 or USB I/O board defective (occasional)
  4. 4. Speed point defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Speed Point not functioning when directly attached

Speed Point input device connected directly to monitor does not respond.

Ranked causes

  1. 1. Speed Point not connected properly (common)
  2. 2. PS/2 I/O board in wrong slot (occasional)
  3. 3. PS/2 I/O Board defective (occasional)
  4. 4. SpeedPoint defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Status message 'Print device Local 1 (Local 2) unavailable' - printer job stalled

Local printer not available; printer off, out of paper, cabling issue, or I/O board defective

Ranked causes

  1. 1. Printer not switched on (common)
  2. 2. Printer paper tray empty (common)
  3. 3. Cabling not connected correctly (common)
  4. 4. PS/2 or USB I/O board defective (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Wireless out of range message issued patient safety

INOP message WIRELESS OUT OF RANGE displayed; monitor lost connection to information center

Ranked causes

  1. 1. LAN cable disconnected (common)
  2. 2. Information center disconnected (common)
  3. 3. Network infrastructure failure (occasional)
  4. 4. Monitor out of wireless range (occasional)

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Zero calibration failed

Gas analyzer zero calibration fails to complete or produces out-of-range results

Ranked causes

  1. 1. Solenoid or air reference filter problem (common)
  2. 2. Occluded pneumatics (common)
  3. 3. Flowrate problem (common)
  4. 4. IR measurement head problem (occasional)
First check

Verify solenoid air draw

Test point
Room air filter
Expected
Air drawn at filter during calibration
Tools
Hand

4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Alarm LEDs illuminate but no alarm sound on Remote Alarm Device

LEDs on remote alarm device light up but speaker does not produce sound.

Ranked causes

  1. 1. Wrong I/O slot (common)
  2. 2. Speaker defective (occasional)
  3. 3. Remote device I/F defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

Communication problem with IntelliVue 802.11 Bedside Adapter or low RSSI

Monitor cannot communicate with wireless bedside adapter or signal strength below acceptable level (RSSI < 30)

Ranked causes

  1. 1. Network infrastructure not functioning properly (common)
  2. 2. Antenna cable disconnected or defective (common)
  3. 3. Cable connection between adapter and system interface board defective (common)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG Out/Marker In function does not function

ECG output and/or marker input functionality not operational

Ranked causes

  1. 1. MSL cable hardware problem (common)
  2. 2. ECG Out hardware defect (occasional)
  3. 3. MSL connector defect in measurement server (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

ECG Out/Marker In function not functioning

ECG output or marker input feature not working

Ranked causes

  1. 1. Hardware problem - MSL cable (common)
  2. 2. ECG Out hardware defective in monitor (occasional)
  3. 3. MSL connector defective in measurement server (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

External device not receiving data

Data export from monitor to external device not functioning

Ranked causes

  1. 2. Wrong data export protocol driver configured (common)
  2. 3. Cable between external device and monitor not connected or defective (common)
  3. 4. External device does not support data export protocol version (occasional)

7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

General monitor INOP: CHECK MONITOR TEMP

Temperature inside the monitor is too high

Ranked causes

  1. 1. Monitor ventilation obstructed (common)
  2. 2. Environmental conditions cause overheating (occasional)
  3. 3. Main board defective (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

General monitor INOP: INTERNAL COMM. MALF.

Problem with I2C Bus communication in the monitor

Ranked causes

  1. 1. Video board defective or communication issue (occasional)
  2. 2. Video board defective (occasional)
  3. 3. Main board defective (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP Message 'Check Alarm Lamps' issued

Alarm LED board cable disconnected, board defective, or main board defective

Ranked causes

  1. 1. Alarm LED board cable disconnected (common)
  2. 2. Alarm LED board defective (common)
  3. 3. Main board defective (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP message 'MsmtSrv not Supp' issued

System indicates measurement server is not supported by monitor; in-operative condition declared

Ranked causes

  1. 1. Wrong software revision on monitor or measurement server (common)
  2. 2. Too many measurement servers connected (occasional)
  3. 3. Unsupported type of measurement server connected (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP Message SERVERLINK MALF displayed

Beep every two seconds audible indicator; FMS or MMS with incompatible software revision connected or hardware communication fault

Ranked causes

  1. 1. FMS or MMS with incompatible software revision connected (common)
  2. 2. Communication between components not functioning (common)
  3. 3. Hardware for communicating with Multi-Measurement Server is faulty (occasional)

5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP message SPEAKER MALFUNCTION displayed caution

Speaker malfunction indicated. No alarm sound output capability.

Ranked causes

  1. 1. Speaker cable disconnected (common)
  2. 2. Speaker defective (occasional)
  3. 3. Sound amplifier on main board defective (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP message: Bad Server Link

Unexpected data or communication failure in MSL connection

Ranked causes

  1. 1. Unexpected data on MSL (occasional)
  2. 2. Incompatible software revision (occasional)
  3. 3. Communication failure (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP message: Check Alarm Lamps caution

Alarm LED system malfunction indicated

Ranked causes

  1. 1. Alarm LED board cable disconnected (common)
  2. 2. Alarm LED board defective (occasional)
  3. 3. Main board defective (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP message: MsmtSrv not Supp

Measurement Server software incompatible with monitor software version

Ranked causes

  1. 1. Wrong software revision (common)
  2. 2. Too many measurement servers connected (occasional)
  3. 3. Unsupported type of measurement server (occasional)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

INOP message: Unsupported LAN

Network interface error; may indicate network failure or configuration issue

Ranked causes

  1. 1. Network failure (occasional)
  2. 2. Monitor connected to wrong network (occasional)
  3. 3. IP address conflict (rare)

3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI

IntelliVue 802.11 wireless adapter not connecting

Monitor with 802.11 adapter fails to authenticate or connect to access point; RSSI signal weak or connection status shows disconnected.

Ranked causes

  1. 1. WLAN configuration incorrect (common)
  2. 2. Access point distance or antenna orientation incorrect (common)
  3. 3. Access point incompatibility or failure (occasional)
First check

Access WLAN Diagnostic service window

Test point
Service Mode > Main Setup > WLAN Diagnostic
Expected
Service window opens with Conn.Status field visible

2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI

More symptoms

  • Measurement Server configuration not supported
  • Medical electrical system non-compliance after installation
  • Message 'Measurement Server Configuration not supported' issued
  • Other Bed Overview not available
  • Oxygen (O2) not measured
  • Red Error LED stays on
  • Touchscreen not responding to touch input
  • Unit does not power on
  • Wireless Adapter has no power
  • Wireless network: Frequent dropouts and disconnects
  • Alarm LEDs do not illuminate
  • Alarm LEDs remain off
  • Alarm occurs but no LED lights up
  • Calibration: Anesthetic Gas Module
  • Care and cleaning of external surfaces
  • Disappearing waves in warmup mode
  • Equipment leakage current excessive (normal condition)
  • Equipment leakage current excessive (single fault condition)
  • Flexible Module Rack not communicating with monitor
  • Flowrate outside specification during performance test
  • Frequent dropouts and network disconnects on wireless
  • General monitor INOP: CHECK INTERNAL VOLTAGE
  • General monitor INOP: SETTINGS MALFUNCTION
  • Inserted module LEDs normal but module not recognized
  • Leakage detected during performance test
  • LEDs on IntelliVue 802.11 Bedside Adapter are off and remain off
  • Measurement module not recognized despite LEDs normal
  • Measurement Server does not start - LEDs normal
  • Measurement Server does not start up; no LEDs active; no INOP or prompt displayed
  • Monitor connected to LAN but not monitored at central
  • Network prompt: No central assigned to bed
  • Network related: NO CENTRAL MONITORING message
  • No connectivity in multi-access-point coverage area
  • No test sound issued
  • Normal operation waves or numerics missing
  • O2 measurement unable to measure
  • Paper not feeding properly
  • Parameter modules not detected or not functional
  • Poor print quality
  • Poor print quality from recorder
  • Print device remote unavailable, printer job stalled
  • Print job could not be queued, no print device found
  • Prompt message 'Print job could not be queued'
  • Protective earth resistance excessive
  • Pump malfunction or failure
  • Recorder not communicating with system
  • Recorder paper not feeding properly
  • Red Error LED blinks
  • Red Error LED blinks (indicating cyclic reboots)
  • Red Error LED flashes
  • Sidestream CO2 flow rate out of tolerance
  • Solenoid valve #2 not functioning
  • Status message 'Print device Remote 1 (Remote 2, Remote 3) unavailable' - job stalled
  • Top cover PC board malfunction or disconnection
  • Watertrap manifold or protector damaged
  • 'Other Bed' Alarms not appearing
  • Agent gas measurement unable to measure
  • Agent ID head damaged or misaligned
  • Alarm occurs on screen but no LED or sound on alarm device
  • Alarm on VueLink device but no sound on monitor
  • Alarm sound issued but no LEDs light up
  • All LEDs on USB board are off
  • Applied part leakage current exceeds limit
  • Applied part leakage current measurement (invasive connectors)
  • Barometric pressure error outside tolerance
  • Calibration: Anesthetic Gas Module flowrate measurement
  • Calibration: Anesthetic Gas Module performance verification
  • Calibration: Anesthetic Gas Module safety verification
  • Calibration: Barometric pressure check and calibration
  • Calibration: Microstream CO2
  • Calibration: NBP
  • Calibration: Noninvasive Blood Pressure Measurement
  • Calibration: Safety testing after top cover removal
  • Calibration: Span calibration check
  • Calibration: Total flowrate check in normal mode
  • Calibration: Touchscreen
  • Calibration: Zero calibration
  • CO2 measurement unable to measure
  • CO2 scrubber replacement procedure
  • Connector holder insertion into dovetail connection
  • Cooling fan not operating smoothly
  • Cooling fan not running or operating abnormally
  • Diagnostic error in Gas Analyzer Diagnostic window
  • Disposal: Empty calibration gas cylinder
  • Dual link bar assembly with MSL flex connector
  • ECG heart rate measurement inaccuracy
  • ECG synchronization with defibrillator fails
  • Equipment leakage current exceeds limits
  • Final inspection of assembly
  • Firmware revision in wireless adapter is wrong
  • Flex cable connector insertion and alignment
  • Flexible Module Rack LEDs ok but module not recognized
  • FMS cables not properly connected
  • General monitor INOP: MCC UNSUPPORTED
  • Image sticking on LCD display
  • INOP 'BAD SERVER LINK' issued
  • INOP 'MSL Power High' issued
  • INOP 'MSL Power Off' issued
  • INOP 'MSL Power Overload' issued
  • INOP EcgOut EQUIP MALF is issued
  • INOP message CHECK NURSE CALL RELAY
  • INOP message CHECK NURSE CALL RELAY is issued
  • INOP message: CHECK INPUT DEVICES
  • INOP message: MSL Power High
  • INOP message: MSL Power Off
  • INOP message: MSL Power Overload
  • INOP message: Serverlink Malf
  • Inserted module does not function
  • Invasive pressure performance test failure
  • IT equipment in patient vicinity leakage current non-compliance
  • Mainstream CO2 accuracy test failure
  • Measurement from server does not appear on monitor
  • Measurement module currently ignored
  • Measurement Module in Flexible Module Rack currently ignored
  • Measurement Module in Integrated Module Slot currently ignored
  • Measurement parameter not displayed on monitor
  • Measurement Server connection error
  • Measurement Server does not start - no power
  • Measurement Server does not start but LEDs are normal
  • Messages 'Measurement Server not supported, unplug device, switch monitor off/on' and INOP 'Bad Measurement Server' issued
  • Module does not lock into FMS
  • Monitor is out of range of wireless access point
  • Monitor not progressing through boot sequence
  • N2O measurement unable to measure
  • NBP accuracy test failure
  • NBP linearity test failure
  • NBP system leakage detected
  • NBP valve closure test failure
  • Network connection lost
  • Network related: UNSUPPORTED LAN message
  • No central monitoring possible via wireless adapter
  • No marker pulse displayed on the monitor
  • No Network icon or Network icon flashes; no association to central station
  • None of the connected USB devices are functioning
  • Nurse Call Relay not functioning
  • Overview and Printing do not work on wireless
  • Power fuse failure or blown fuse
  • Power Loss Alarm Buzzer not sounding
  • Printing on remote device reported but no output
  • Printouts are not as expected
  • Prompt message 'Too many <label> modules connected'
  • Prompt message: Too many modules of same type
  • Protective earth leakage current from multiple socket outlet exceeds limit
  • Pump filter maintenance required or failed
  • Pump replacement and flow calibration
  • Pump test failure
  • Recorder module does not lock into FMS
  • Remote extension device buttons not functioning
  • Remote Extension Device buttons not responding
  • Remote Extension Device input not functioning
  • Replacement: Pump and scrubber
  • Respiration rate measurement inaccuracy
  • Routine maintenance: Gas cylinder replacement
  • Sidestream CO2 accuracy test failure
  • Span calibration check failure for gas channels
  • Speed point knob rotation or joystick control not functioning
  • SpeedPoint knob or joystick control not functioning
  • SpO2 measurement out of expected range
  • Status message 'Incompatible SW Revision versions' issued
  • Status message 'Printing on device Remote 1...' but no report printed
  • System thinks door is open when it is not
  • System thinks recorder door is open when closed
  • System thinks recorder is out of paper when it is not
  • System thinks recorder out of paper when not
  • Temperature measurement performance test failure
  • Top cover refitting with MSL flex cable protection
  • Unit setup before initial use
  • Unrecognized Measurement Module in Flexible Module Rack slot
  • Unrecognized Measurement Module in Integrated Module Slot
  • Unrecognized measurement module in slot
  • Wireless adapter defective
  • Wireless monitor losing access point connection
  • Wireless network: No connectivity with multiple access points
  • Wireless network: Overview or printing not working
  • Zero calibration check failure

175 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI

Error codes

CodeMeaningTriggerSeverity
16030Monitor spontaneously rebootsSpontaneous reboot without user interactioncritical
16400Monitor spontaneously rebootsSpontaneous reboot without user interactioncritical
16505Loss of configuration after system rebootSystem rebootcritical
17345MMS reboots unexpectedlySpontaneous reboot of multi-measurement servercritical
20005Loss of configuration after system reboot - monitor runs in default safe profileSystem reboot or power cycleerror
20212Monitor spontaneously rebootsSpontaneous reboot without user interactioncritical
21002Monitor reboots upon removal of modules (M8048A)Removal of I/O moduleserror
21002For M8048: Modules not recognized or monitor reboots upon module removal; For M3001: Monitor reboots after Temp or Pressure unusedHardware module removal or unused module stateerror
21400MMS reboots unexpectedly, no waves or numerics displayedSpontaneous reboot of multi-measurement servercritical
30049Monitor spontaneously rebootsSpontaneous reboot without user interactioncritical
32750Modules not recognized (M8048A)Plug or removal of moduleserror
495Monitor spontaneously rebootsSpontaneous reboot without user interactioncritical
500Monitor spontaneously rebootsSpontaneous reboot without user interactioncritical
AGENT IDENT MALFAgent identification sensor malfunctionSerious Agent-ID hardware problemerror
AGENT IDENT ZERO FAILEDAgent identification sensor zero calibration failedSolenoid 2 or Agent-ID problemerror
CHECK INPUT DEVICESSpeed point or other input device defectiveInput device malfunction detected during initialization or operationwarning
CHECKINTERNALVOLTAGEProblem with internal power supply voltagesMonitor detects abnormal 5V or 12V voltage levelserror
CHECK MENU LABELSMenu labels configuration errorMonitor performs cold start and attempts to reset menu labels to defaultswarning
CHECK MONITOR FUNCAlarm lamps, display or interfaces not functioning correctlyDetected when internal voltages are too lowerror
CHECK MONITOR TEMPTemperature inside the monitor is too highTemperature exceeds maximum operating limitwarning
CHECK SETTINGSSettings configuration errorMonitor performs cold start and settings default reset is detectedwarning
GA INCOMPATIBLEGas analyzer module version incompatible with monitorIncompatible AGM versionerror
GA NOT AVAILABLEGas analyzer module not available to monitorAGM not switched on or not properly connectederror
GAS AN ACCURACY ?Gas analyzer accuracy questionableMeasurement accuracy degradedwarning
GAS EQUIP MALFGas equipment malfunction detectedAGM-monitor connection problem, subassembly failure, or Main PC Board issueerror
GAS OCCLUSIONGas flow occlusion detectedBlockage in gas flow path, external or internalerror
GA ZERO FAILEDGas analyzer zero calibration failedCalibration procedure failed or produced out-of-range valueserror
INOP NBP Equipment MalfunctionNon-invasive blood pressure equipment malfunction preventing normal operationOccurs in monitoring mode when NBP calibration is invalid or hardware fault detectederror
INTERNAL COMM.MALF.Internal I2C bus communication failureMonitor detects I2C bus communication errorerror
MCC UNSUPPORTEDUnsupported MSL coupling cable connection detectedMSL coupling cable connected to incompatible devicewarning
MONITOR LABEL DUPLICATEAnother monitor on the network has the same equipment labelDuplicate equipment labels detected on networkwarning
NO CENTRAL MONITORINGMonitor is not assigned to sector or central connection lostMonitor not assigned to monitoring sector or connection to information center losterror
O2 EQUIP MALFOxygen sensor equipment malfunctionO2 span calibration failed or sensor configuration errorerror
O2 ZERO FAILEDOxygen sensor zero calibration failedO2 sensor calibration failed or produced out-of-range constantserror
SETTINGS MALFUNCTIONConfiguration cloning failureProblem occurs during configuration file cloning processerror
SETTINGS MALFUNCTIONProblem during cloning process or memory corruptionSettings memory space corrupted or cloning process failederror
UNSUPPORTED LANNetwork failure detected after 1 minute of operationNetwork infrastructure failure or extended communication losscritical
WIRELESS OUT OF RANGEMonitor lost connection to the Information CenterConnection to information center lost or wireless access point unavailableerror
XXX MEAS DISTURBEDGas measurement disturbed (N2O, CO2, agent, or O2)Transient measurement assembly problemwarning

Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI