Philips IntelliVue MP5: troubleshooting and error codes
The ServiceForge knowledge base holds 79 documented symptoms for the Philips IntelliVue MP5, with 175 ranked causes, 77 diagnostic checks with test points and expected values, and 17 error codes. Each symptom below shows its causes and the first check; the rest is one connection away. Connect ServiceForge to your AI
Symptoms
External device not receiving data
Data export to external device is not working; no data communication from monitor to connected device via MIB/RS232 port
Ranked causes
- 1. MIB/RS232 port not configured for data export (common)
- 2. Wrong data export protocol driver configured (common)
- 3. Wrong or defective cable (occasional)
- 4. Cable not connected properly or is defective (occasional)
- 5. External device does not support monitor data export protocol version (occasional)
- 6. Terminal concentrator incompatibility with dynamic speed negotiation (rare)
- 7. System interface board defective (rare)
Verify port configured for data export
- Test point
- Monitor MIB/RS232 configuration settings
- Expected
- Port configured for data export
6 more checks with pass and fail actions, 7 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Integrated display is blank
Built-in display shows no image or text. (Only valid if boot phase completed without error)
Ranked causes
- 1. Flexible Display cable not connected (common)
- 2. Backlight Inverter Cable not connected (common)
- 3. Backlight tube cable not connected (common)
- 4. Backlight tubes defective (occasional)
- 5. Backlight Inverter board defective (occasional)
- 6. LCD Flat panel defective (rare)
- 7. Main board defective (rare)
Connect external display to video port
- Test point
- video port
- Expected
- external display shows image
- Tools
- external display
7 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Combined Power On/Error LED remains off after power button pressed
No LED indication (Power On/Error LED stays off) when power button pressed on monitor
Ranked causes
- 1. Remote devices connected causing boot failure (common)
- 2. Recorder PCA defective (occasional)
- 3. IIT module defective (occasional)
- 4. Flat Panel or Backlight Inverter defective (occasional)
- 5. System Interface Board or Main Board defective (occasional)
- 6. Main Board defective (occasional)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Incorrect predictive temperature values patient safety
Predictive temperature (pTemp) readings are inaccurate or do not match expected values; values from different body locations compared directly
Ranked causes
- 1. Wrong label chosen (e.g., pTaxial vs pToral) (common)
- 2. Wrong patient category selected (common)
- 3. Incorrect probe application (common)
- 4. Wrong probe covers used (occasional)
- 5. Probe defective (occasional)
- 6. Predictive temperature assembly defective (rare)
Verify correct temperature label is selected
- Test point
- Monitor configuration menu
- Expected
- Label matches measurement location (pToral, pTaxial, etc.)
3 more checks with pass and fail actions, 6 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
INOP Unsupported LAN issued
INOP message UNSUPPORTED LAN displayed; network icon shows disconnection or conflict
Ranked causes
- 1. Monitor not set up with correct label (common)
- 2. IIC to Switch communication problem (occasional)
- 3. Network failure or infrastructure down (occasional)
- 4. Monitor connected to wrong network (occasional)
- 5. IP address conflict after infrastructure change (occasional)
- 6. Hardware defect (system interface board or switch) (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Touch screen not responding to input
Touchscreen does not respond to touches or touch input is non-functional
Ranked causes
- 1. Touchscreen temporarily disabled via padlock (common)
- 2. Touch calibration lost after main board exchange (occasional)
- 3. Touch panel cable not connected (occasional)
- 4. Backlight tube cable positioned too close to touch sensor (rare)
- 5. Touch sensor defective (occasional)
- 6. Main board defective (rare)
6 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
External display is blank (Slave Display)
Secondary or external video display shows no image or text
Ranked causes
- 1. Video cable to external display not connected or defective (common)
- 2. External display has no power (common)
- 3. External display is defective (occasional)
- 4. System Interface board defective (rare)
- 5. Main board defective (rare)
5 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder communication or printing failure
Recorder is not communicating with monitor or is not printing; no output is produced
Ranked causes
- 1. Loose recorder connector (common)
- 2. Loose flat cable between recorder and main board (common)
- 3. Recorder defective (occasional)
- 4. Recorder interface board defective (occasional)
- 5. Main board defective (rare)
Inspect and reinsert recorder connector
- Test point
- Recorder connector socket on monitor
- Expected
- Recorder fully seated with audible click or firm connection
2 more checks with pass and fail actions, 5 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC Power LED does not illuminate
AC mains LED indicator does not turn on green when AC power cord connected
Ranked causes
- 1. AC connection not present or not okay (common)
- 2. Power supply defective (common)
- 3. System Interface Board defective (occasional)
- 4. Main Board defective (occasional)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no sound issued
Monitor triggers alarm condition but no audible alarm tone is produced by the system
Ranked causes
- 1. Audible alarm indicators switched off (common)
- 2. Volume set to 0 (common)
- 3. Speaker defective (occasional)
- 4. Sound amplifier on main board defective (occasional)
Verify audible alarm indicators are enabled
- Test point
- Monitor settings menu
- Expected
- Audible alarm indicators switched on
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Lost connection to Information Center
Monitor lost connection to central monitoring system; INOP message NO CENTRAL MONITORING issued
Ranked causes
- 1. LAN cable disconnected (common)
- 2. Information Center disconnected or offline (occasional)
- 3. Network infrastructure failure (occasional)
- 4. Wireless monitor out of range (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
MP5 does not start up when connected via MSL to host monitor
Companion-mode MP5 monitor fails to power on or initialize when connected via MSL cable to host monitor
Ranked causes
- 1. MP5 not receiving AC power or battery not charged (common)
- 2. Feature 'Power up on MSL power' not functioning (occasional)
- 3. Internal flat ribbon cable between MSL board and system interface board defective (occasional)
- 4. MSL board, system interface board, or main board defective (rare)
4 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printout quality not as expected
Printed reports have incorrect appearance, formatting, or visual quality compared to expected output
Ranked causes
- 1. Printer paper size not correctly configured (common)
- 2. Printer resolution not correctly configured (common)
- 3. Printer color support misconfigured when not supported (occasional)
- 4. Printer not compatible with system (rare)
Verify printer paper size configuration
- Test point
- Setup Printers menu > Paper size setting
- Expected
- Paper size matches the media loaded in the printer
2 more checks with pass and fail actions, 4 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery malfunction patient safety
BATT MALF message with INOP tone; battery LED flashes. Alarms cannot be paused or switched off if monitor not on AC power.
Ranked causes
- 1. Battery status cannot be determined or battery-board communication failure (common)
- 2. System interface board defect (occasional)
- 3. Main board defect (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
MIB/RS232 gas module not functioning
Gas module connected to RS232 port is not responding; no communication with gas measurement module
Ranked causes
- 1. MIB/RS232 port not configured for gas module (common)
- 2. Gas module cable not connected or defective (common)
- 3. System interface board is defective (rare)
Verify MIB/RS232 port configured for gas module
- Test point
- Monitor configuration settings
- Expected
- Port is configured for gas module communication
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
MP5 not communicating with host monitor via MSL
MP5 and host monitor cannot exchange data or signals over MSL connection
Ranked causes
- 1. Hardware for MSL communication defective (occasional)
- 2. Internal MSL ribbon cable defective (occasional)
- 3. MSL board, system interface board, or main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG output at host monitor (MP5 in companion mode) patient safety
Host monitor does not receive or display ECG signal from companion MP5 monitor
Ranked causes
- 1. Hardware for MSL communication defective (occasional)
- 2. Internal MSL ribbon cable defective (occasional)
- 3. MSL board, system interface board, or main board defective (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG sync pulse output
ECG synchronization pulse is not being output from the monitor; external device does not receive sync signal
Ranked causes
- 1. Software not configured correctly for ECG sync (common)
- 2. Wrong cable connected (occasional)
- 3. No ECG pulse signal available from monitor (rare)
Verify ECG sync software configuration
- Test point
- Monitor ECG sync configuration settings
- Expected
- Software configured correctly for ECG sync per configuration guide
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Nurse call relay did not activate when monitor alarmed patient safety
Patient alarm on monitor was triggered but nurse call system did not activate; no alert sent to nursing station
Ranked causes
- 1. Relay configuration incorrect (common)
- 2. Cable connection loose or incorrect at monitor or nurse call system (common)
- 3. System interface board defective (rare)
Verify nurse call relay configuration
- Test point
- Monitor relay configuration settings (latency, trigger)
- Expected
- Relay configured correctly per configuration guide
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Other Bed Overview not available
Other Bed Overview feature not functioning or not accessible
Ranked causes
- 1. Incompatible telemetry adapter or IIC version (occasional)
- 2. Configuration problem with care group assignment (occasional)
- 3. Multicast filter misconfiguration (rare)
3 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Predictive temperature label not available
Predictive temperature (pTemp) label is not selectable in the monitor configuration menu
Ranked causes
- 1. Measurement deactivated in settings (common)
- 2. Connector defective between module and monitor (occasional)
- 3. Predictive temperature module or measurement connector defective (occasional)
Verify parameter is enabled in monitor menu
- Test point
- Setup > Parameters menu
- Expected
- Predictive temperature parameter is switched on and activated
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Recorder output not as expected
The content of the recording output does not match what was expected; print quality, content, or format is incorrect
Ranked causes
- 1. Monitor not configured properly (common)
- 2. Paper not inserted correctly (common)
- 3. Printhead dirty (occasional)
Review monitor recording configuration
- Test point
- Monitor setup menu for recorder settings
- Expected
- Recording parameters match desired output format and content
2 more checks with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Speaker malfunction INOP message
INOP message 'SPEAKER MALFUNCTION' is displayed on the monitor, indicating audio alarm issue
Ranked causes
- 1. Speaker cable disconnected (common)
- 2. Speaker defective (occasional)
- 3. Sound amplifier on main board defective (occasional)
Verify speaker cable connection
- Test point
- Speaker connector on main board
- Expected
- Cable fully seated and secure
1 more check with pass and fail actions, 3 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Abnormal status log entries indicating system reboot or event
Status Log shows multiple event entries with cold start (C) or hot start (H) events; system has experienced unexpected reboots or component resets
Ranked causes
- 1. Event occurred causing system reboot (unknown)
- 2. Recurring event pattern indicating repeating fault (unknown)
Access Status Log via Main Setup -> Revision menu
- Test point
- Monitor Status Log display
- Expected
- Status Log window displays event entries with codes and timestamps
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm occurs but no LED lights up patient safety
Alarm condition detected but visual indicator LED does not illuminate
Ranked causes
- 1. Environmental lighting too bright (common)
- 2. Main board defective (occasional)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery not operational
INOP tone indicates battery communication or power issue
Ranked causes
- 1. Battery defective or low charge (common)
- 2. System interface board fault (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Calibration: Touch panel
Touchscreen requires calibration after power loss during procedure or miscalibration; must complete without interruption or touch panel becomes deactivated.
Ranked causes
- 1. Monitor powered off during touch calibration (occasional)
- 2. Touchscreen accidentally mis-calibrated with wrong spot selected (occasional)
Verify touch calibration procedure completes without power interruption
- Test point
- Main Screen key and touchscreen
- Expected
- Calibration completes and touchscreen responsive
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
IIT adapter no network connection
No network icon or network icon flashing; IIT adapter not associated to central station
Ranked causes
- 1. Incorrect RF Access Code (common)
- 2. Communication problem between monitor and IIT adapter (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Individual parameter INOP message patient safety
Parameter-specific INOP messages such as CO2 EQUIP MALF, ECG EQUIP MALF, NBP EQUIP MALF, or similar are displayed
Ranked causes
- 1. Measurement module defective or disconnected (common)
- 2. Main board defective (occasional)
Identify the failed parameter module from INOP message
- Test point
- Display message
- Expected
- Specific INOP code (CO2, ECG, NBP, Pressure, RESP, SpO2, Temp, etc.)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Marker pulse not displayed on monitor patient safety
Monitor display does not show marker pulse indicator when defibrillator is connected; visual sync indicator is absent
Ranked causes
- 1. Defibrillator cable disconnected or defective (common)
- 2. Main board defective (rare)
Disconnect defib cable and connect known-good defib with good cable
- Test point
- Marker pulse and ECG OUT signal on good defib and monitor display
- Expected
- Marker pulse visible on monitor display and good defib
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Monitor not assigned to sector or duplicate equipment label
Monitor connected to LAN with IP address but bed not monitored at central; monitor label not set or duplicate label exists
Ranked causes
- 1. Monitor not assigned to a sector (common)
- 2. Duplicate Equipment Label on network (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
No ECG output signal to defibrillator patient safety
Defibrillator is not receiving ECG waveform output from monitor; no ECG signal or marker pulse on defibrillator display
Ranked causes
- 1. Defibrillator cable disconnected or defective (common)
- 2. Main board defective (rare)
Disconnect defib cable and connect known-good defib with good cable
- Test point
- ECG OUT and marker pulse signals at defibrillator
- Expected
- Marker pulse and ECG OUT signal present at good defib
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
No test sound issued during boot
Speaker does not produce test tone during system startup
Ranked causes
- 1. Speaker defective (common)
- 2. Main board defective (rare)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Poor print quality from recorder
Recorder printouts show poor print quality, faded output, or reduced image clarity
Ranked causes
- 1. Printhead dirty (common)
- 2. Printhead failure or paper feed problem (occasional)
Visually inspect and clean printhead
- Test point
- Recorder printhead
- Expected
- Printhead is clean and dry with no ink or debris accumulation
2 more checks with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Power loss alarm buzzer does not sound patient safety
When battery is removed and monitor is disconnected from AC power, the power loss alarm buzzer does not produce an audible alert
Ranked causes
- 1. Power loss alarm buzzer defective (occasional)
- 2. Communication problem between main board and power loss alarm buzzer (occasional)
Disconnect AC power and remove battery
- Test point
- Power connections
- Expected
- Monitor in unpowered state with buzzer active
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer not found message
Prompt message 'Print job could not be queued' is displayed and no print device is found
Ranked causes
- 1. Printer disabled in Setup menu (common)
- 2. Paper size mismatch between printer and report (common)
Verify printer is enabled in Setup Printers menu
- Test point
- Setup > Printers configuration
- Expected
- Correct printer is enabled and active
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printer remote device unavailable status
Status message displayed that printer on remote location (Remote 1, 2, or 3) is unavailable; print job stalled
Ranked causes
- 1. Print error on IntelliVue Information Center (IIC) (common)
- 2. Network connection to IIC not functioning (common)
Test network connection to IIC
- Test point
- Network connection status at monitor and IIC
- Expected
- Network connection active and responding
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Printing initiated but report not output
Status message 'Printing on device Remote 1... (Remote 2, Remote 3)' is displayed but no report is actually printed
Ranked causes
- 1. IIC print queue full; printer not powered on (common)
- 2. Printer paper tray empty (common)
Verify printer power is on
- Test point
- Printer power indicator
- Expected
- Printer shows power-on status and ready state
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Red Power On/Error LED blinks repeatedly (cyclic reboots) caution
Power On and Error LEDs flash repeatedly indicating system cycles through boot repeatedly
Ranked causes
- 1. Software fault in operating system (common)
- 2. Hardware failure detected (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Reference value mismatch between monitor and CalKey
Displayed reference value on the monitor does not match the value on the CalKey card for predictive temperature module
Ranked causes
- 1. Defective CalKey (occasional)
- 2. Defective CalKey or measurement assembly contacts (occasional)
Verify CalKey value reading on monitor
- Test point
- Monitor display during CalKey insertion
- Expected
- Reference value matches CalKey printed value
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR communication aborted with interference or invalid channel
SRR INTERFERENCE or SRR INVALID CHAN INOP message issued; communication fails
Ranked causes
- 1. SRR device out of range (common)
- 2. Telemetry device incompatible with SRR (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
SRR communication dropouts
Gaps in parameter waveforms; SRR INTERFERENCE INOP may be issued during communication
Ranked causes
- 1. RF interference present (common)
- 2. Too many SRR devices on one channel (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Telemetry device disconnected INOP
INOP message 'TELE DISCONNECTED' displayed indicating telemetry device connection lost
Ranked causes
- 1. Telemetry cable disconnected or fallen off (common)
- 2. SRR link disrupted between telemetry device and monitor (common)
Inspect cable connection between telemetry device and monitor
- Test point
- Cable connectors at both ends
- Expected
- Cable fully seated and secure on both device and monitor
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
WLAN adapter no network connection
No network icon or network icon flashing; WLAN adapter not associated to central station
Ranked causes
- 1. WLAN settings mismatch (common)
- 2. Communication problem or weak RSSI signal (occasional)
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
AC Power LED or Power On/Error LED remains off after pressing power button
Neither AC nor Power On/Error LED indicator turns on when power button pressed
Ranked causes
- 1. Main board defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm lamp not functioning patient safety
CHECK ALARM LAMPS INOP message displayed; visual alarm indicator LED not working or not visible
Ranked causes
- 1. Main board defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Alarm LEDs do not illuminate
Alarm indicator LEDs remain off during boot or operation
Ranked causes
- 1. Main board defective (rare)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery LED shows red with blinking
Battery indicator LED flashing red when battery connected
Ranked causes
- 1. Battery charge depleted or battery defective (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery low warning
BATT LOW message with INOP tone; less than 20 minutes operating time remaining
Ranked causes
- 1. Battery charge depleted (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery symbol is not displayed
Battery indicator icon or status symbol does not appear on monitor display
Ranked causes
- 1. Monitor not configured for battery operation (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Battery temperature too high caution
CHECK BATT TEMP message with INOP tone
Ranked causes
- 1. Ventilation blocked or monitor exposed to excessive heat (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Charger malfunction
CHARGER MALFUNCT message with INOP tone; battery LED may flash
Ranked causes
- 1. Battery charger circuit defect (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Image sticking on LCD display
Static image remains temporarily visible on display after content changes; ghost image or residual image visible on screen
Ranked causes
- 1. Static image displayed for extended period on LCD (common)
Switch display off and back on
- Test point
- LCD display screen
- Expected
- Sticking image clears and display shows current content
2 corrective actions and part numbers are in ServiceForge. Connect ServiceForge to your AI
Individual measurement or application not functioning correctly
Specific patient parameter (e.g., specific waveform, derived value) or monitoring application (e.g., arrhythmia detection, trending) shows questionable readings or fails to operate
Ranked causes
- 1. Parameter or application not configured correctly or sensor issue (unknown)
Consult Instructions for Use for parameter-specific troubleshooting
- Test point
- Individual parameter or application operation
- Expected
- Parameter functions per design specification
1 more check with pass and fail actions, 2 corrective actions, and part numbers are in ServiceForge. Connect ServiceForge to your AI
Invalid leadset INOP
INOP message 'INVALID LEADSET' displayed for telemetry device
Ranked causes
- 1. Leadset issue with telemetry device (unknown)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
Measurement selection icon does not change to SRR
SRR measurement type not selectable or available on monitor
Ranked causes
- 1. SRR interface defective or incompatible (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
NG-LAN to central not functioning
Network connection from MP5 to central station or network is not operational
Ranked causes
- 1. MSL connection active disables NG-LAN at MP5 (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No connectivity to IIC with no prompt or error message
Monitor has no connection to IntelliVue Information Center but no error message displayed
Ranked causes
- 1. Switch configuration or firmware issue (common)
1 corrective action and part numbers are in ServiceForge. Connect ServiceForge to your AI
No IP address assigned
Monitor has LAN connection but no IP address assignment; wireless found access point or wired hears traffic but no DHCP/static IP
Ranked causes
- 1. DHCP failure or static IP not configured (unknown)
No LAN connection
Monitor does not have a LAN connection; wireless monitor cannot find access point or wired monitor cannot hear anything on its LAN connection
Ranked causes
- 1. Access point not found or no LAN traffic detected (unknown)
More symptoms
- Other Bed Alarms not appearing
- Power On/Error LED remains red continuously
- pTemp Check Probe INOP issued
- pTemp Equip Malf INOP issued
- pTemp No Sensor INOP issued
- pTemp Transducer Malf INOP issued
- Recorder reports door open when closed
- Recorder reports out of paper when not out
- Replace battery or battery low INOP for telemetry
- SRR device assignment fails with interference INOP
- SRR device assignment not possible
- Support tool unable to access monitor data or perform diagnostic functions
- Telemetry device not recognized by SRR adapter
- Telemetry device unsupported INOP
- Telemetry equipment malfunction INOP
- Touch position input invalid or inaccurate
- Touchscreen deactivated after power loss during calibration
- Touchscreen mis-calibrated during calibration procedure
- Wireless monitor losing central contact
19 more symptoms with ranked causes and checks are in ServiceForge. Connect ServiceForge to your AI
Error codes
| Code | Meaning | Trigger | Severity |
|---|---|---|---|
BATT EMPTY | Battery is depleted, estimated operating time <= 10 minutes | Remaining battery power drops to critical level | critical |
BATT INCOMPAT | Incompatible or incorrect battery type installed | Battery is not correct model for this monitor | error |
BATT LOW | Battery charge critically low | Battery operating time drops below 20 minutes | warning |
BATT MALF | Battery malfunction | Monitor cannot determine battery status or communication lost with battery | critical |
CHARGER MALFUNCT | Battery charger malfunction | Monitor detects problem with internal battery charger circuit | error |
CHECK ALARM LAMPS | Alarm lamp circuit malfunction | Main board cannot drive alarm indicator LEDs | critical |
CHECK BATT TEMP | Battery temperature excessive | Battery temperature detected above safe operating threshold | warning |
CHECK INTERNAL VOLTAGE | Problem with too low voltages (5V, 12V) in the monitor | Voltages below nominal threshold, causing alarm lamps, display or interfaces to malfunction | warning |
CHECK MONITOR TEMP | Temperature inside monitor is too high | Internal temperature exceeds safe operating threshold | warning |
INOP | Monitor inoperable condition | Various system failures (battery, communication, hardware) | critical |
MONITOR LABEL DUPLICATE | Duplicate equipment label on network | Another monitor has same equipment label | error |
NO CENTRAL MONITORING | No connection to central monitoring system | Monitor loses connection to IntelliVue Information Center | warning |
SETTINGS MALFUNCTION | Problem with settings storage or cloning process | Settings memory corrupted or configuration transfer failed | error |
SRR INTERFERENCE | Short Range Radio interference detected | RF interference or too many devices on same SRR channel | warning |
SRR INVALID CHAN | Short Range Radio invalid channel | SRR device assigned to invalid or unavailable channel | error |
UNSUPPORTED LAN | Network configuration not supported or misconfigured | Network infrastructure mismatch, IP conflict, or connectivity failure | error |
WIRELESS OUT OF RANGE | Wireless monitor out of access point range | Wireless monitor loses signal to all available access points | warning |
Likely causes and actions for each code are in ServiceForge. Connect ServiceForge to your AI